<<

dew point.qxp 6/6/2005 1:07 PM Page 2 DEW POINT MEAS IN METAL Accurate measurement of etal heat treatments such as annealing, , and carboni- triding require furnace atmospheres that have controlled water a gas or atmosphere dew vapor concentrations. Measurement of atmosphere dew point tem- point can have a M perature is widely used to provide meaningful data to help heat treaters monitor water vapor concentration and optimize the process. Con- significant effect on quality densation (optical chilled mirror) hygrometers, metal oxide sensors, and thin- and process in heat film polymer sensors are among the available technologies. Common heat treating terms are defined in the sidebar. Some of them de- treating. Several scribe processes in which atmosphere control is a distinct “science.” Metallur- measurement methods gists have the knowledge and ability to change a process “recipe” to alter the characteristics of the material being heat treated, and the moisture parameters and instrument types are of the process are very important to final properties. Repeatability also is re- available. quired: Every time the product is heated, the energy used must produce the desired result. Dew point monitoring of these processes is critical since expen- ditures for the gasses and energy needed to maintain their atmospheres are Ken Soleyn primary cost drivers. Reworking a product rejected due to an incorrect atmos- GE Infrastructure Sensing phere analysis is often not an option. Billerica, Massachusetts A heat treating furnace is not an ideal environment for dew point monitoring sensors and probes. The involved are beyond the operating limits of most sensors, and the furnace environment may outgas or entrain chemical vapors and contaminants such as oils, salts, and particulate. Special sampling systems typically are required to enable dew point sensors to be used for furnace atmospheres. Sensors also require a program of periodic calibration against a reference standard. This article compares the major dew point measurement techniques and provides additional information about atmosphere sampling, system main- tenance, and instrument calibration.

A Dew Point Primer Dew point is defined as the temperature to which a gas mix- ture at a given pressure must be cooled to achieve the saturation point of the water vapor contained in the mixture. At the saturation point the gas is holding the maximum amount of water in the gaseous state (water vapor). Any water in excess of the saturation point will condense to liquid. In addition, any sur- face that is below the dew point temperature will acquire condensation. Below 0°C (32°F), frost (ice crystals) will form on surfaces colder than the satura- tion temperature. The dew point temperature can be directly correlated to the partial pressure of water vapor contained in the gas. Equation 1 yields the vapor pressure over water, while Eq. 2 yields the vapor pressure over ice. Since there are differ- ences between the two saturation vapor pressures and because supercooled water can exist below 0°C (32°F), the two terms dew point and frost point are differentiated. However, the general practice in industry is to refer to ei- ther measurement as “dew point.”

–6 esw = (1.0007 + 3.46 × 10 P) × 6.1121exp[17.502T/(240.97 + T)] Eq. 1

–6 esi = (1.0003 + 4.18 × 10 P) × 6.1115exp[22.452T/(272.55 + T)] Eq. 2

es = Saturation water vapor pressure, millibars (esw, water; esi, ice) P = Barometric or process pressure, millibars T = Temperature, °C 54 HEAT TREATING PROGRESS • JULY 2005 dew point.qxp 6/6/2005 1:07 PM Page 3 UREMENT

103

Inserting the dew point or frost , millibar

point temperature, respectively, in s Equation 1 or 2 yields the existing 101 water vapor pressure, e. Inserting the ambient temperature in the equations yields the appropriate saturation 10–1 water vapor pressure (Fig. 1). The measurement of dew point and its conversion to water vapor pressure is commonly used to derive other hu- 10–3

midity parameters when the gas com- Saturation water vapor pressure, e position, temperature, and pressure are known. A discussion of basic hu- 10–5 midity parameters follows: –80 –40 0 40 80 Relative humidity (%RH): The Ambient temperature, °C ratio of the existing water vapor pres- sure, e, to the saturation water vapor pressure, es, both in millibars. The pa- Fig. 1 — Plot of ambient temperature vs. rameter is dimensionless and commonly expressed as a percent. saturation water vapor pressure.

%RH = (e/es) × 100 Eq. 3 Absolute humidity (“AH”): The mass of water vapor per unit volume. Common absolute humidity units are g/m3 or grains/ft3 (1 grain = 1/7000 lb, or ~65 mg).

AH, g/m3 = 216.7e/(T + 273.16) Eq. 4 Volumetric mixing ratio (“VMR”): The ratio of volume of water vapor to dry carrier gas. The parameter is dimensionless. Common units are ppmv (parts per million by volume).

6 VMR, ppmv = [e/(P – e)] × 10 Eq. 5 Mass mixing ratio (“MMR”): Also called specific humidity, this is the ratio of the mass of water vapor to the mass of the dry carrier gas. The parameter is dimensionless. Common units are ppmw (parts per million by weight).

6 MMR, ppmw = [18e/MW(P – e)] × 10 Eq. 6

MW = Molecular weight of the carrier gas, g/mole (MWair = 28.96 g/mole)

Dew point preferred: In the heat treating industry, dew point is the param- eter of choice because it provides meaningful data. For example, an error of only a few degrees Celsius in a dew point measurement can represent an error of 0.1 to 0.2% in the potential, which can result in a deviation in the carbon content of the final product of as much as 25%. The focus of this article now shifts to dew point measurement methods.

Condensation Hygrometers Condensation (optical chilled mirror) hygrometers represent a direct method of measuring dew point or frost point. The basic design consists of a conden- sation target (the mirror) that is cooled until dew or frost is detected. Feedback control maintains the condensation surface at an equilibrium point such that the mass of dew or frost is constant. The resultant temperature of the con- HEAT TREATING PROGRESS • JULY 2005 55 dew point.qxp 6/6/2005 1:07 PM Page 4

GaAs IR emitters virtually 100% of the IR signal is re- Optical balance Photodetectors ceived by the photodetector. As dew or ice condenses on the mirror the re- Operational amplifier flected light scatters, decreasing the (op-amp) amount of light received by the pho- todetector. A closed-loop controller regulates the light received by the Gas flow photodetector to a constant value by controlling the heat pumped away Thermoelectric from the mirror. When the light signal heat pump control Emitter is controlled to a constant value, the regulation mass of dew or frost on the mirror is also constant. The resultant temper- Dew point temperature indicator ature is equal to the dew or frost Precision 188.8 point. four- PRTD Power supply Rhodium or platinum mirror Chilled mirror hygrometers are Thermoelectric cooler (TEC) heat pump very precise and repeatable. The wetted components are inert and the Fig. 2 — Schematic of a two-stage, optical chilled mirror dew point sensor (condensation hy- sensors do not drift over time. Typ- grometer). They can measure dew points from –80 to 85°C (–110 to 185°F). ical accuracy is ±0.2°C (±0.4°F) dew/frost point. This explains why they are used as reference standards for calibrating other types of hu- midity instruments. Maintenance: The condensing sur- face (mirror) must remain clean. There are two basic types of contam- inants: soluble (salts) and insoluble physical contaminants. Soluble con- taminants will dissolve in the dew/frost layer and increase the vapor pressure of the condensate, re- sulting in higher dew point readings. Physical contaminants will either Fig. 3. — Chilled mirror hygrometers are available with a variety of sensors designed for spe- scatter or adsorb the light signal. cific measurement ranges. Cleaning the mirror and rebalancing Chilled mirror densation target is, by definition, the reflected light signal against a ref- equal to the dew or frost point tem- erence emitter/detector pair can mit- hygrometers are perature. These hygrometers have the igate the effects of contaminants. precise and repeatable. ability to measure dew point from –80 GE developed the PACER system to 85°C (–110 to 185°F). The specific for automatically cleaning and rebal- The wetted components range is governed by the amount of ancing chilled mirrors. The PACER are inert and the sensors cooling and the ability to heat the cycle first utilizes rapid cooling to de- do not drift over time. sensor for measurement to high dew velop a thick dew/frost layer, and points. then rapid heating to flash evaporate Typical accuracy is Details: The optical chilled mirror some contaminants. Any contami- ±0.2°C (±0.4°F) hygrometer (Figures 2 and 3) utilizes nants that remain will accumulate in a solid-state thermoelectric heat discrete “islands” such that approxi- dew/frost point. pump, which may be augmented mately 85% of the mirror is clean. The with liquid coolant or refrigerant to optical control system is then “bal- cool a metal mirror (typically made anced” against a reference IR emitter/ of rhodium, platinum, or stainless detector pair, which effectively ). A four-wire platinum resistance negates the effect of the residual con- temperature detector (PRTD) is ther- taminants. While the PACER works mally coupled to the mirror. An in- very well to mitigate contamination, frared (IR) emitter illuminates the eventually the mirror will need to be mirror at a 45° angle and a photode- manually cleaned. The instrument’s tector positioned at a complimentary optical circuitry alerts the user to the 45° angle receives the light reflected need for manual cleaning by moni- by the mirror. When the mirror is dry, toring the gain ratio of the photode- 56 HEAT TREATING PROGRESS • JULY 2005 dew point.qxp 6/6/2005 1:07 PM Page 5

SELECTED HEAT TREATING TERMS atmosphere controlled furnaces. In many heat treating operations, alyst. For low surface concentrations of carbon, the dew point the atmosphere must be controlled to prevent workpieces from may be adjusted to 0°C (30°F) or higher. oxidizing and/or decarburizing. Steel becomes more active as the temperature increases, and severe oxidation of carbon ATMOSPHERES AND GASES steel begins at about 425°C (795°F). Above 1200°C (2190°F), the argon: Provides an excellent inert atmosphere. It is used for gas- oxidation rate increases exponentially. At high temperatures shielded arc welding and for heat treatment of “exotic” alloys. the carbon in steel also can react with the atmosphere to lower Generally, argon must be delivered at a dew point of less than the carbon content. –60°C (–75°F) and an oxygen content of less than 20 ppm. carbon potential. A measure of the ability of an environment con- blast gas. Used in the steel industry to oxidize (burn) the coke that taining carbon to alter or maintain, under prescribed conditions, melts the ore to produce molten metal. Blast gas has to be moist the carbon level in steel. Control of carbon potential is impor- in order to carry heat from exhaust gasses into the furnace and tant in carburizing furnaces. Excessive carbon will permeate to maintain burn efficiency. Steam is typically added to control the grain structure of the , causing embrittlement and even- this mix, which means the gas must be analyzed for moisture tual component failure. Preheat or burn-off muffles require good content. The absolute humidity range is typically 7 to 50 g/m3 atmosphere control to flush out these contaminants. (3 to 22 grains/ft3), or dew points of 0 to 35°C (30 to 95°F). products of combustion. These result when fuel mixed with air is commercial atmospheres: Nitrogen is used in many heat burned. If fuels such as methane and propane are burned in op- treating applications, sometimes replacing endothermic atmos- timized proportions with air, the by-products might be ideal pheres. Nitrogen serves as a pure, dry, inert gas that can pro- for certain heat treated products. If an excess of air exists (a lean vide efficient purging and blanketing. Typical specifications re- atmosphere), loose scale may form. When an excess of fuel is quire the nitrogen to be delivered at dew point temperatures used (a rich atmosphere), a tight, adherent oxide forms. Note between –60 and –80°C (–75 and –110°F). Nitrogen is also used that water vapor is a by-product of combustion. as a carrier gas for carbon control atmospheres in many com- mercial heat treating applications. Nitrogen is mixed with hy- PROCESSES drogen in a 90-10 blend, where the serves as a re- annealing. A generic term for heat treatment that consists of heating ducing gas. to and holding metal at a suitable temperature then cooling at dissociated ammonia. Dissociated ammonia (N2 + H2) is produced a suitable rate to remove stresses, to induce softness, to alter from anhydrous ammonia (NH3) by raising the temperature and toughness, to produce electrical, magnetic, or other to 900–980°C (1650–1795°F) in a catalyst filled retort. The gas physical properties, to refine the crystal structure, to remove is then cooled for metering and transport. Dissociated am- gases, or to produce a specific microstructure. The temperature monia atmospheres are about 75% H2 and 25% N2,with less of the annealing operation and the rate of cooling depend upon than 300 ppm residual ammonia at a dew point below –60°C the material being heat treated and the purpose of the treatment. (–75°F). The atmosphere provides a dry, carbon-free source of brazing. Proper control of the furnace atmosphere is also impor- reducing gas. Uses include bright and brazing, tant to the life of alloy components. For example, poor control bright heat treating of carbon and selected nickel and of the dew point in aluminum brazing can allow excess mois- copper alloys, and bright annealing of electrical components. ture and condensation of highly corrosive fluxes, leading to Dissociated ammonia is also used as a carrier gas in certain ni- failure due to corrosion. triding processes. bright annealing. A process usually carried out in a controlled fur- dry hydrogen atmospheres. Commercially available hydrogen is 98 nace atmosphere so that surface oxidation is reduced to a min- to 99.9% pure. Cylinder hydrogen may contain trace amounts imum and the surface remains relatively bright. In order to limit of water vapor and oxygen. Dry hydrogen is used in furnaces oxidation, the water vapor concentration must be limited. Bright for annealing stainless and low-carbon steels, electrical steels, annealing environments are typically purged with inert gases and several nonferrous metals. It is also used for “hard such as nitrogen, argon, or dry air. Typically, the dew point tem- metals” such as tungsten carbide and tantalum carbide, for perature must be less than –50°C (–60°F). brazing nickel, , and copper, for annealing metal . A process in which ammonia (NH3) added to a gas powders and sintering powder (P/M) parts, and in carburizing environment dissociates to produce hydrogen (H2) the direct reduction of metal ores. and nitrogen (N2). The addition of nitrogen has three important endothermic atmospheres: Endothermic atmospheres are produced effects: inhibits the diffusion of carbon, which favors produc- by generators that use air and a hydrocarbon gas as fuel. The two tion of a shallow case; enhances , which favors gases are mixed, slightly compressed, and passed through a production of a hard, wear-resistant case that is easily polished; chamber filled with nickel catalyst. The chamber is heated exter- and forms nitrides, which further enhance wear resistance. nally, hence the term endothermic. mixtures are carburizing. A process in which ferrous metal is brought into con- used as carrier gases in carburizing and carbonitriding applica- tact with an environment of sufficient carbon potential to cause tions (they offer a wide range of possible carbon potentials). Other absorption of carbon at the surface, and by diffusion to create applications include bright of steel, carbon restora- a carbon concentration gradient between the surface and the tion of steel and bars, and sintering powder that requires interior of the metal. Carburizing is usually done at 850 to 950°C a reducing atmosphere. (1560 to 1740°F) in an atmosphere consisting of any of several exothermic atmospheres. Exothermic gas is produced by combus- carrier gases, principally nitrogen, , and hy- tion of a hydrocarbon fuel such as methane or propane to main- drogen, to which hydrocarbon gases (or vaporized hydrocarbon tain a reaction temperature of 980°C (1795°F) for sufficient time liquid) have been added. Methane or natural gas (CH4) is the to reach equilibrium. Heat is obtained from the reaction, hence most commonly used source of carbon. For carburizing in the the term exothermic. The resultant gas is cooled and water vapor range of 0.8 to 1% C, the dew point temperature of the carrier is removed either by a refrigerated or desiccant dryer. gas is optimized at –7 to –1°C (19 to 30°F). Dew points below Exothermic atmospheres are used for clean and bright annealing –12°C (10°F) may lead to accelerated sooting of generator cat- Continued on page 61 HEAT TREATING PROGRESS • JULY 2005 57 dew point.qxp 6/6/2005 1:08 PM Page 6

tector to the reference detector. Manual cleaning consists of heating the mirror (the analyzer has a heat switch) and wiping the metal mirror with a cotton swab dipped in a cleaning solution or distilled water. The procedure takes less than three minutes. Stabilization: Condensation hy- grometers must physically cool the condensing surface to the dew point temperature. The amount of cooling at a given ambient temperature is re- ferred to as the “depression.” The amount of depression is influenced by the sensor “body temperature,” gas composition, and gas pressure. Fig. 4 — Schematic of how a metal oxide (Al2O3/Au) trace moisture sensor works, left. An Because condensation hygrometers actual sensor is shown at right. are under closed-loop control, they typically take time to stabilize at a given reading. There is some oscil- lation before the reading “settles out.” For measurements at low frost points (less than –25°C, or –13°F), chilled mirrors must sufficiently undercool to acquire a frost coating on the mirror’s surface. This means that chilled mirrors typically must have the capability of cooling 5 to 10°C (9 to 18°F) below the lowest value meas- ured. For dew points less than –60°C (–75°F), full stabilization may take a few hours. However, process upsets are rapidly detected because the in- strument will respond very quickly to moisture. Service life: The initial cost of chilled mirror hy- grometers is generally greater than that of other dew point instruments. Off- setting the higher price tag are reliability, repeatability, and long-time drift-free service. These factors make condensation hygrom- eters very competitive with other dew point instruments. The component in a chilled mirror hygrometer most likely to fail is the thermoelectric heat pump. Thermo- electric cooling performance can be measured in terms of watt density (W/cm2). Lowering the watt density extends the life of the heat pump; however, the trade-off is giving up the ability to measure to lower mois- ture levels. Watt density can be low- Fig. 5 — A variety of metal oxide trace moisture instruments are used in heat treating appli- ered using refrigeration or circulation cations. They include integrated transmitters, top; analyzers with remote probes, inset; and portable of a chilled fluid through the sensor’s battery-operated hygrometers, bottom. base to supplement thermoelectric 58 HEAT TREATING PROGRESS • JULY 2005 dew point.qxp 6/6/2005 1:08 PM Page 7

cooling. For example, GE’s 1311 se- purposes. Dew point calibration data Unlike temperature or ries chilled mirror sensors are are typically stored in an embedded equipped with thermoelectric heat memory chip along with temperature pressure sensors where pumps that are thermally bonded to compensation data. Other instru- the sensing element can a gold-plated, water-cooled heat ex- ments provide a frequency table that changer. Circulating city water (or is programmed into an analyzer. Raw be installed in a protective chilled water) through the unit lowers frequency and temperature data are sheath or diaphragm that the watt density, extending the meas- reduced by microprocessor-based cir- is in thermal or pressure urement range, and lengthening life cuitry to provide a direct readout or expectancy. It is not unusual for a linear output and display of dew equilibrium with the test water-cooled chilled mirror to pro- point temperature. (Remember that environment, dew point vide 10 to 15 years of continuous other humidity parameters such as service. absolute humidity and volumetric sensors must actually mixing ratio can be derived from the come in contact with that Metal Oxide Sensors dew point.) Metal oxide sensors are typically Metal oxide sensors are usually cal- environment. Filters are used for trace dew point measure- ibrated against a primary reference a “trade-off” between ment, usually over a dew/frost point humidity instrument such as a chilled protecting the sensor from range of –90 to 10°C (–130 to 50°F). mirror hygrometer. Typical accuracy is ±2 to 3°C (±4 to Filter required: Since metal oxides contaminants and 5°F). The measurement technique is such as aluminum oxide are very hy- optimizing response time. based on the ability of absorbed water groscopic, they respond rapidly to the molecules to change the capacitance presence of water vapor but may take or impedance of a metal oxide. Alu- considerable time to dry out. This is minum oxide (or alumina, Al2O3) is why some metal oxide sensors are the one most widely used. Other sen- stored in a chamber surrounded by sors include silica and zirconia (SiO2 desiccant when not in use. A special and ZrO2). valve introduces process air to the A thin layer of oxide is first de- sensor chamber. The net result: shorter posited or grown on a metal base. The dry-out times. Generally, these sensors oxide is typically sputter coated with respond to moisture (wet-up) in a few a layer of porous gold, which serves seconds. This makes them useful for as an upper electrode and also pro- detecting process upsets. tects the metal oxide. The oxide is GE-developed metal oxide sensors highly porous (has a large surface (Fig. 5) include those that combine an area). When dry, the spaces within the ultrathin aluminum oxide layer with oxide are filled with air, which has a a porous gold protective coating. The dielectric constant of 1. In this state the sensor is pre-aged, resulting in drift oxide’s electrical capacitance is low of less than 2°C (4°F) per year. How- and its impedance is high. When water ever, drift is highly dependent on the vapor (dielectric constant = 80) is ab- environment being measured. Unlike sorbed, microcondensation occurs temperature or pressure sensors within the pores and the oxide’s ca- where the sensing element can be in- pacitance increases in proportion to stalled in a protective the surrounding water vapor pressure. sheath or diaphragm The pores function as microscopic par- that is in thermal or allel plate capacitors. This is shown pressure equilibrium schematically in Fig. 4, left. (When ca- with the test environ- pacitors are placed parallel, the total ment, dew point sen- capacitance is the sum of the capaci- sors must actually come tances of the individual capacitors.) in contact with that en- A metal oxide sensor’s signal con- vironment. Filters are a ditioning circuitry typically measures “trade-off” between capacitance or impedance as fre- protecting the quency (or a discreet count). Probes sensor from con- are calibrated by recording frequency taminants and vs. dew point at multiple values optimizing re- across the operating range. Many sponse time. GE’s Fig. 6 — A variety of instruments based on metal oxide probes also include a planar gold/alu- thin-film polymer capacitive sensors are used temperature sensor for compensation minum oxide sensors are in metal heat treating. HEAT TREATING PROGRESS • JULY 2005 59 dew point.qxp 6/6/2005 1:08 PM Page 8

usually installed in a polished stain- fect; however, heavier hydrocarbons sample. The filter housing must also less steel housing that has a porous or oils will tend to coat the surface of be made of hydrophobic material sintered steel filter. An industrial non- the sensor over time, creating a bar- such as AISI type 316 stainless steel hygroscopic filter should be installed rier to water vapor. (UNS S31600). upstream of the sensor. Many metal oxide and capacitive In certain cases, a vacuum pump or sensors can be cleaned with solvents, aspirator must be used to extract the Thin-Film Polymer Sensors reconditioned, and placed back in sample and create sufficient flow Thin-film polymer capacitive hu- service. And as previously men- across the sensor. The nominal flow in midity sensors are typically capable tioned, it is always a good practice to 0.25 in. (6 mm) OD tubing is 1 scfh (0.5 of measuring dew points from –15 to use a nonhygroscopic filter upstream L/min). All sampling lines are typi- 150°C (5 to 300°F). GE’s MMR series from a metal oxide sensor. Typical cally stainless steel with compression transmitters and MDR series probes filter media include borosilicate glass, fittings. For ultralow levels, internally (see Fig. 6) incorporate a sensor that silicone oils (oil bath), sintered steel, electropolished tubing with VCR pres- consists of interdigitated metal elec- and polytetrafluoroethylene (PTFE). sure fittings are used. (VCR is a trade- trodes on a substrate. The sur- mark of Cajon Co., Macedonia, Ohio.) face is coated with a proprietary thin- Sampling System Required The sampling system might also in- film polymer and porous metal layer. Sampling systems must be used in clude sensors to measure other process As water molecules are adsorbed by heat treating applications. Typically, variables, such as oxygen, pressure, the polymer the capacitance changes the system must extract a sample of temperature, and flow rate. in proportion to the surrounding rel- the furnace atmosphere and cool it to In some cases the sampling system ative humidity. The sensor is coupled the operating range of the sensor. De- might be installed in a protective or to signal conditioning circuitry that pending on the temperature and dew heated enclosure (Fig. 7). If the am- produces a frequency output. A pre- point, the cooling mechanism may bient temperature at the location of cision RTD measures temperature simply be heat loss through a few feet the sensor is lower than the dew point and provides temperature compen- of stainless steel tubing (air to air) or of the sample gas, condensation will sation for the relative humidity signal. a “water-to-air” heat exchanger. A occur. In these applications, the In MMR-101 high-temperature in- filter must be used to remove partic- sample must be heated above the dew struments, each sensor is calibrated ulate and contaminants; however, the point temperature of the process gas. at multiple points across several tem- filter medium must not be one that The rule of thumb is to heat sample peratures to produce a matrix that is will remove or add water vapor to the lines, filters, and the sensor body to stored in memory. at least 10°C (20°F) above the highest Microprocessor circuitry, linear re- dew point encountered. gression analysis, and psychometric A number of GE probes and trans- formulas are used to convert raw fre- mitters can be supplied with either quency and temperature data to di- explosion-proof or intrinsically safe rect linear dew point and tempera- certifications for monitoring atmos- ture signals (4 to 20 mA). Typical pheres containing flammable or com- accuracy of a thin-film polymer ca- bustible gases, or for installation in pacitive sensor is ±2 to 3°C (±4 to 5°F) them. In certain applications, the an- dew point, depending on the oper- alyzers and probes are installed in ex- ating range. plosion-proof enclosures and sam- pling systems are equipped with Contaminants and Filters safety features such as flame arrestors. The readings of all humidity instru- Other instruments may have intrinsic ments are affected by exposure to safety systems, where the probe or chemical and physical contaminants. transmitter is operated in the haz- For example, polar compounds such ardous zone while the analyzer is in- as alcohols will change the capaci- stalled in the safe zone. The compo- tance when absorbed by the sensor. nents are interconnected via an If dissociated ammonia atmospheres intrinsic safety barrier. are being used, dew point readings will be higher due to ammonia carry- Calibration Ensures Accuracy over. Salts such as zinc oxide (en- Metal oxide and polymer capaci- trained from galvanizing operations) Fig. 7 — Sampling systems extract, cool, tance sensors provide secondary dew can adversely affect metal oxide sen- and clean furnace atmosphere samples. The point measurement. While secondary sors. Reactive or corrosive chemicals system must not add or remove water vapor dew point instruments do not pro- should be avoided. from the dew point test gas. For many fur- vide the accuracy and precision of a naces, a turnkey sampling system having sen- In general, nonpolar compounds sors for oxygen and other process parameters primary instrument such as a chilled such as straight-chain hydrocarbon in addition to those for dew point might be mirror hygrometer, they have a few vapors or liquids have very little ef- appropriate. advantages including lower initial 60 HEAT TREATING PROGRESS • JULY 2005 dew point.qxp 6/6/2005 1:08 PM Page 9

costs. Since they do not require cooling to achieve low dew points, SELECTED the sensors can be operated at rela- tively high temperatures, and gener- HEAT TREATING TERMS ally are more rugged in industrial en- Continued from page 57 vironments. Aluminum oxide-based and clean hardening. Rich exothermic atmospheres are useful for annealing, instruments are considered industrial and for copper brazing of low-carbon steels, Cu-Ni alloys, gold alloys, and some workhorses, but do require a program . Applications for lean exothermic atmospheres include annealing of of periodic calibration to compensate aluminum and copper and their alloys, bluing of steel parts, silver brazing of for long-term drift. nonferrous alloys, and nonflammable blanketing during various industrial processes. One of the best ways to maintain natural atmospheres (air). Air consists of about 78% nitrogen, 21% accurate dew point sensors is to stock oxygen, 0.9% argon, and other trace gases. Air at room temperature varies in spare sensors and rotate them in and moisture content from about 0.3 to 3%, nominally. Although natural atmos- out of service. For example, if a heat pheres are strongly oxidizing, they may be acceptable when workpieces are treating facility has 20 measurement to be machined after heat treating. points, five spares might be stocked. steam atmospheres. Steam injection into furnaces is used for scale-free and stress relieving of ferrous metals in the temperature range of 345 to 650°C After a predetermined period of (655 to 1200°F). The steam causes a thin, hard, tenacious blue-black oxide to service, the five spares are installed form on the surface of the part. Prior to processing in a steam atmosphere, parts in place of five working sensors, must be clean and oxide-free. To prevent condensation and rusting, furnace in- which are subsequently recalibrated. ternal surfaces and the parts in the furnace must be at a temperature above 100°C Calibration process: Typical cali- (212°F). And air must be purged from the furnace to prevent formation of a brown coating instead of the desired blue-black oxide. bration services consist of providing vacuum atmospheres. Heating metal parts at pressures below atmospheric is used “as found” data. First, however, most for many components, composites, and metals. Vacuum heat calibration labs clean industrial treating prevents surface reactions such as oxidation and , re- probes prior to placing them in their moves surface contaminants such as oxide films and lubricant residue, degasses calibration systems. (These systems metals, removes dissolved contaminants from metals, and joins metals by brazing or diffusion bonding. are kept very clean, and probes re- turned from the field are a source of ware, software, and training to per- Selected References contamination.) Aluminum oxide form calibrations. The cost and type Practical Heat Treating, by Howard E. sensors are typically cleaned in sol- of calibrations should be considered Boyer: ASM International, Materials Park, vents such as hexane or toluene, then when selecting instrumentation. Im- Ohio, 1984, 243 p. dried in an oven. Calibration of the portant factors in amortizing the cost “The Theory and Operation of Optical Chilled Mirror Hygrometers for Hu- dew point instrument follows. The of an internal calibration system in- midity Calibration,” by Ken Soleyn: Cal process consists of comparing read- clude the volume of outsourced cali- ings against a reference standard Lab – The International Journal of Metrology, brations, personnel training, and June/July/August 2002. which is sampling the same air or gas. turnaround times. Calibrations are typically performed “The Effect of Air Laden Soluble Salts on In the field: In addition to calibra- Dew Point Measurement Using Conden- at multiple test points. tions in the lab, field checks on alu- sation Hygrometers,” by Ken Soleyn: Pro- There are two basic requirements minum oxide probes can be per- ceedings of the National Conference of Stan- for a calibration system: formed by comparison with readings dards Laboratories International, Salt Lake • A dew point generator capable of made using a portable instrument. In City, Utah, 2004. producing stable and controlled this case, the user typically has no Water Vapor Measurement: Methods and In- dew/frost points. Several types are control of the dew point and is simply strumentation, by Pieter R. Wiederhold: Marcel Dekker Inc., New York, N.Y., 1997. available. comparing the reading of an instru- ISBN 0-8247-9319-6. • A reference instrument. The one ment against a reference instrument most commonly used is a condensa- Glossary of Metallurgical Terms and Engi- having a known calibration. It is im- neering Tables: Prepared under the direc- tion (chilled mirror) hygrometer. portant to keep the portable instru- If the devices being tested are tion of the ASM Handbook Committee, ment calibrated because it often ASM International, Materials Park, Ohio, found to be “out of tolerance,” adjust- uses the same measurement tech- 1979, 93 p. ments might be made at either the in- nology as the device being Carburizing and Carbonitriding: Prepared strument or the analyzer by repro- tested. under the direction of the ASM Com- gramming a new “look table.” After mittee on Gas Carburizing, ASM Interna- adjustment, the probes are recali- tional, Materials Park, Ohio, 1977, 215 p. For more information: Ken Soleyn is brated and verified to be within ac- “Psychometrics,” Chapter 6 in 1989 product manager – test and calibration, ceptable tolerances. ASHRAE Handbook: Fundamentals: Amer- GE Infrastructure Sensing, 1100 Tech- ican Society of Heating, Refrigerating, and Heat treaters that have a significant nology Park Drive, Billerica, MA 01821; Air Conditioning Engineers Inc., 1989. number of probes might find it cost tel: 978/437-1000, ext 1924; fax: 978/437- ISBN 0-910110-56-5. effective to set up internal dew point 1032; e-mail: [email protected]. GE calibration systems. Suppliers such Infrastructure Sensing’s website: www. as GE offer the instrumentation, hard- gesensing.com. HEAT TREATING PROGRESS • JULY 2005 61