CH455 Analytical Chem: Lab Excerpts — Spring 2005 From “ Procedures”

Real World Samples is one of the more common ways the “real world” sees chemistry. Drug testing, vehicle emission testing, blood alcohol readings, DNA tests, the test for explosives done at airport security gates, all these are examples of analytical chemistry. Where possible our experiments are designed for you to test some of “your” world in our experiments. If appropriate, in the Chemicals & their Location section of the experiment, you will be told what type of sample to bring. Bringing your own sample is not a requirement, simply an option. We will always have samples for you to test (except for the urine sample, which someone will have to provide). You should not supply your own sample unless the procedure tells you that this is an option for the particular experiment. From “Introduction to the use of Analytical Equipment”

Pipettes are used to transfer measured volumes of liquids between containers. You will be using TD pipettes; to deliver means that the filled with solution will deliver the rated volume, with only the aid of gravity, not using a bulb to blow out the contents of the pipette. There are many types of pipettes, each designed for a specific function and mode of usage, three of which are listed below: Transfer or Volumetric pipettes These are used to transfer an accurately known volume of liquid from one container to another. They are shaped like a piece of tubing with a bulb and have one fiducial mark. If the bulb is centered, it is commonly called a volumetric pipette, while if the bulb is close to the delivery tip, it is called an Ostwald-Folin pipette. These are the most accurate pipettes available and range from 1 – 250 mL. Measuring or Graduated pipettes These pipettes are simply a piece of uniform diameter tubing drawn out to a delivery tip with graduations similar to those found on a . If the graduations extend all the way to the tip, they are called serological pipettes; if the graduations stop before the constriction for the tip, they are called Mohr pipettes. These are less accurate than transfer pipettes, but more accurate than graduated cylinders. The volume range of these pipettes is generally 0.1 – 25 mL. Automatic Pipettes These pipettes are equipped with a plunger that controls the delivery of fluid by the positive displacement of a plunger, and are equipped with disposable plastic tips. In some varieties, the delivery volume is variable, and controlled by a dial on the barrel of the pipette. The primary benefit of these pipettes is that they allow rapid, repetitive delivery of sample or reagent when multiple analyses are required. The volume of these pipettes range from 0.02 – 25 mL.

CH455 Analytical Chemistry Week Three Lab Handout Page (1) Spring 2005 Using a volumetric pipette For the best analytical results, pipettes should be used in a uniform manner. By always holding the pipette the same way, allowing the same amount of time for the pipette to drain, etc., errors in the volume will usually be minimized. The correct procedure for their use is as follows: 1. Clean your pipette and rinse with distilled water. Check the tip and return to the stockroom if damaged. 2. Rinse the inside of the pipette with the solution to be transferred to avoid dilution.

3. Use a pipette bulb to fill the pipette by suction to about 2 cm above the fiducial mark. If the bulb has filled with air, but your pipette is not yet full, remove the bulb from the pipette, replacing it with the tip of your forefinger. Squeeze the air out of the bulb and replace on the top of the pipette to continue filling. 4. Remove the pipette bulb, replacing it with the tip of your forefinger. 5. Using the thumb and other fingers, rotate the barrel of the pipette. Using your forefinger to control the flow of liquid out of the pipette, carefully lower the meniscus to the fiducial mark. 6. Touch off drops hanging from the pipette tip and wipe excess fluid from the outside of the pipette. 7. Move your receiving container close to your initial container. Transfer the pipette tip from one container to the other and allow the liquid to flow out of the pipette into the receiving container. 8. While draining, hold the pipette vertically. When flow stops, hold the tip against the inner wall of the container for a further 10 seconds to allow the final drops to exit. 9. A small volume of solution will be left in the pipette tip. This is normal. (a) if you are using a TD pipette (the only sort we have in our teaching labs), its volume is calibrated to take this into account; thus, this small volume should not be blown out; however, (b) if you were using a TC pipette, its volume is calibrated to include this small volume, and so the last bit of solution must be blown out from the tip.

Using a fixed volume automatic pipette

You will use a fixed volume automatic pipette in one of your labs. The 25 µL fixed volume automatic pipette has an accuracy of ±0.15 µL and a precision of 0.13 µL. The correct procedure for using this pipette is as follows: 1. Attach a new disposable tip to the pipette shaft. Press firmly to ensure a positive airtight seal. 2. Hold the pipette in one hand. Depress the plunger to the first stop. 3. Holding the pipette vertically, immerse the disposable tip into the sample liquid to 5–10 mm below the surface and release the plunger slowly to fill the tip. Never let it snap up!

CH455 Analytical Chemistry Week Three Lab Handout Page (2) Spring 2005 4. Withdraw the top from the sample liquid. If any liquid drops remain on the outside of the disposable tip, touch them to the inside of the liquid container to remove them from the tip surface, or wipe the tip carefully with a Kimwipe, taking care not to touch the tip orifice. 5. To dispense the sample, touch the tip end against the side wall of the receiving vessel and depress the plunger slowly to the first stop. Wait one second, and then press the plunger all the way to the second stop, expelling any residual liquid in the tip. 6. With the plunger fully depressed, withdraw the pipette from the vessel carefully. 7. Allow the plunger to return to the UP position. 8. Discard the tip. A fresh tip should be used for each sample to prevent sample carryover.

Viscosity and Temperature Effects Solution viscosity affects the rate at which the solution drains from the pipette. It will also affect the amount that adheres to the walls of the pipette. No matter what the solution is, some solution remains on the walls. Pipettes are calibrated to take this effect into account for solutions with properties similar to distilled water. Solutions that are more viscous require that the pipettes be recalibrated. The National Institute of Standards and Technology has published standard draining times for various size pipettes. If a solution falls outside the NIST standard times, it is a good idea to calibrate all pipettes. Temperature will affect the volume delivered since density varies with temperature. Such errors are very small (about a few tenths of a percent) and if all solutions in an analysis are at the same temperature, these errors will tend to cancel.

Volumetric Flasks Volumetric flasks are used to prepare exact volumes of primary standard solutions (e.g., EDTA) or exact dilutions of samples (e.g., unknowns). They are fragile and expensive. Do not heat solutions in a ; it will not return to its original calibrated volume on cooling. Volumetric flasks are calibrated TC or to contain the stated volume when filled so that the bottom of the liquid meniscus is level with the fiducial mark on the neck of the flask. Since all the liquid cannot be removed from them (some always sticks to the sides), they are not suitable for "measuring out" volumes of liquids. There is considerable error if they are used in a to deliver manner. When dissolving a solid in a volumetric flask, do not fill it to the mark until all the solid has dissolved. Solvation often results in a change of volume of the solution and this should be allowed to occur before the final volume adjustment is made. It is much easier to mix the contents of a flask before it has been filled to the mark. To mix the solution well, you should ensure the stopper is well seated before repeatedly inverting and shaking the flask. When filling a volumetric flask, there is no margin for error. The typical operation is to add the substance to be diluted, mix well, fill the flask to the bottom of the neck with diluent and then add diluent CH455 Analytical Chemistry Week Three Lab Handout Page (3) Spring 2005 slowly, using a Pasteur pipette, until the bottom of the meniscus reaches the mark. Read the meniscus at eye level to avoid parallax errors. The volumetric flask must be filled exactly to the mark, not just below or above it. If too much diluent is added, the process should be redone using more care. You should not simply remove the excess liquid above the fiducial mark. From “Lab Reports”

Computers The data analysis for most experiments can be done more quickly using a computer. If you do not know how to use a spreadsheet (e.g., Excel), or a graphing programme (e.g., Kaleidagraph), make the time to learn at the start of the semester. UT has a helpful Excel tutorial and is the Quickstart guide provided by the producers of Kaleidograph. Ask your TA for assistance. We recommend that you do all calculations in Excel, but to cut and paste the tables of numbers you wish to graph into Kaleidograph and use that program to make your graphs. A free trial copy of Kaleidograph for both PC and Mac may be downloaded from . Saving and exporting of data is disabled, and graphs printed with the trial version have the words “demo copy” in the background of the graph; however, some creative students in the past have found ways to deal with these limitations. We will not penalize you for printing graphs that indicate you are using a trial copy of this software. A number of experiments produce data in digital form, so you will have to use a computer to analyze this data. Whenever possible we will connect the instrument computers to the internet. We recommend that you upload your data directly into your UT Webspace . Your files are then stored so that you can access them from anywhere, with less danger of losing your data than if you relied solely on a floppy or zip disk; however, you should also bring your own disks to copy your data files in case the internet connection is not working. Do not erase your original data from the hard drive. If something goes wrong with your copy, you can always come back into the lab to make a new copy. We expect you to use your UT WebSpace to store all computer files generated by this course for the duration of the semester, for review at any time by the course instructor. There are several computer labs that you can use to work on your labs. The physical chemistry computer lab (WEL 2.108) contains a network of iMacs and a printer, available for your use during times listed on the door to the room. Kaleidagraph, Excel, and Mathematica are available on all computers in this room. Computer use is on a first-come, first-serve basis; preference goes to people doing work associated with physical and analytical labs. The computer room also contains a small library of useful references. Please do not remove these books from the room. The TAs who work in the physical chemistry computer room are physical chemistry TAs who are under no obligation to help you with your analytical work. The Welch microcomputer facility contains both Macs and PCs in rooms WEL 2.302/6, and is open for student use when no classes are scheduled in those rooms. During the day, student use is restricted to people enrolled in CH courses, i.e., you. To use these labs you must have a UT IF account

CH455 Analytical Chemistry Week Three Lab Handout Page (4) Spring 2005 validated for PRS (Print server); the same account that you need to have access to TELESYS or to use the SMF facility. Excel is available on all computers in this lab. Several licensed copies of the Mac version of Kaleidagraph are available on disk from the proctor. Kaleidagraph is also available in the Physics computer lab.

The Internet is used extensively in this course. We will expect you check email sent to your official UT email address on a daily basis, and will use it as the primary means of communication during the semester. The CH455 course web page is extensive and copies of any lab handouts will be available to download. Your lab report and exam grades will be accessible via eGradebook, accessible via UT DIrect. It is your responsibility to check your grades regularly to ensure there have been no errors made. From “Useful Analysis Techniques” Graphical Analysis

Formatting All graphs must be computer generated. All graphs must be of a suitable size, scale, and axes choice, so that the relevant features of the data are easily seen. Show units, have a title, label both axes, explain abbreviations, and have a legend identifying any symbols used if the graph shows more than one dataset. Do not have any shading in the graph background or gridlines (the Excel defaults). The title of your graph should be more descriptive than “Y” vs. “X”. Some experiments use a computer to collect your experimental data. A plot of this data usually results in a spectrum, voltammogram, chromatogram, cooling curve, or something similar whose shape is important. In this case use the computer plot preferences to hide all data points, and just show the line. In Kaleidograph, use the Line Plot option; in Excel use the XY (Scatter) option and pick the appropriate chart sub-type (Do not use the Excel Line option, it will not allow you to specify the x axis). Some experiments involve collecting a set of data that you will compare to a mathematical equation used to describe the phenomenon you were investigating. Data analysis of these experiments will involve plotting your data, finding the line of best fit, and deriving quantities from the slope and intercept of that line. If your plot is a series of points that you hope show some relationship, such as a linear calibration curve, then use the computer plot preferences to show the data points, and the line of best-fit. Do not draw any lines that “join the dots” — this is not good science. In Kaleidagraph, use the Scatter Plot option; in Excel use the XY (Scatter) option and pick the appropriate chart sub-type. Use the Series tab to Add a plot and select the appropriate cells for the X and Y axes.

CH455 Analytical Chemistry Week Three Lab Handout Page (5) Spring 2005 Viscosity vs. Concentration n sp/ conc Linear (n sp/ conc) To the right are two examples of 0.600

y = -0.0554x + 0.5125 plots of the same dataset. The first was R 2 = 0.0539 created using the default settings of Excel. 0.500 To make a more appropriate graph for a 0.400 scientific report, the following changes were made to create the second plot: the 0.300 grey background, grid lines, and legend 0.200 removed; both axis ranges reduced so that data fills the entire graph and there isn’t 0.100 wasted “white space”; the fonts sizes were

0.000 increased to be more legible; a more 0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 Conc informative title and axis labels were added; units were added to the axis labels and their significant figures were fixed; the “join-the-dots” line was removed; the equation of the line of best fit was edited to include the names of the variables.

CH455 Analytical Chemistry Week Three Lab Handout Page (6) Spring 2005 Potentiometric of a HCl–H3PO4 mixture Potentiometry will be used to detect the endpoints of the titration of a mixture containing unknown amounts of hydrochloric (strong) and phosphoric (weak polyprotic) acids. Your standardized NaOH will be used as titrant. A good understanding of the experimental logic will be needed to simultaneously analyze two components of a mixture. You will use a more advanced mathematical treatment of data – using first and second derivative graphs – to aid in the detection of the endpoints. You will be graded on your accuracy. This experiment is performed in pairs.

Required Reading D.C. Harris, Quantitative Chemical Analysis (6th ed., W. H. Freeman, NY, 2003) pp. 179–87, 204, 232–8, 139–40. Pipettes, Analytical Lab Manual, pp. xxx–xxxii. Volumetric Flasks, Analytical Lab Manual, pp. xxxii – xxxiii. Graphical Analysis, Analytical Lab Manual, p. xxxvii.

PreLab Quiz Topics In addition to being able to explain the purpose of your experiment, the general procedure steps, the use of all chemicals in this experiment and any specific hazards , your prelab quiz may include explanations of any of the following terms: strong and weak acid, polyprotic acid, pH, conjugate acids and bases, fraction of dissociation. You should be able to describe what you expect to see if you titrate a polyprotic acid with a strong base and know all relevant chemical reactions for this experiment. Describe the chemical and/or physical processes that will occur when you reach the endpoint of your titration. Explain how we can use first and second derivative analysis to determine the endpoints of your . Explain how we can titrate a mixture and determine quantities of both components in a single titration. There is a lot of data manipulation in the data analysis for this experiment — too much to do by hand, or with a calculator. We recommend you use Excel or a similar spreadsheet program. (It is difficult to use Kaleidograph to do this analysis.) If you do not know how to use a spreadsheet, ask your TA to show you in office hours, before the day your report is due. There are some helpful tutorials on Excel on the UT web at . Look up and record the three ionization steps of phosphoric acid in your notebook, along with their respective equilibrium constants. Because these equilibrium constants differ from each other by more than a factor of 10 000, there will be no more than two phosphate species present to any significant degree at any given pH. These two phosphate species will form a conjugate acid-base pair. As pH changes, protons are removed from (or added to) the phosphoric acid in a stepwise manner. The complete removal of each proton will be signaled by a distinct endpoint. We will observe two of the endpoints. These two endpoints are used to determine the phosphoric acid concentration by using the volume of standardized base

CH455 Analytical Chemistry Week Three Lab Handout Page (7) Spring 2005 needed to get from: 1) the start to the first endpoint, 2) from the first to the second endpoint, and 3) from the start to the second endpoint. If a second acid (in our case, the strong acid HCl) is present in addition to the phosphoric acid, it will also react with base according to its equilibrium constant. The HCl proton and the first proton of H3PO4 will react with base simultaneously (HCl actually reacts first, but no distinct endpoint is observed). The result is that the volume of standard base used to reach the first endpoint represents the sum of the HCl and H3PO4 concentrations, while the volume of base used to get from the first endpoint to the second is a measure of the H3PO4 concentration alone. From this information, the concentrations of both HCl and H3PO4 may be calculated. This paragraph is important, if you don't understand it, read it again! The endpoints for this experiment cannot be easily detected using a visual (colour) indicator; however, the endpoints are clear when you measure the potential of a suitable electrode as a function of titrant volume. You will measure the pH of the solution as a function of the volume of titrant added, using a combination pH electrode consisting of a glass indicator electrode and an Ag|AgCl electrode. A pH meter measures activity (i.e., pAH+), which can be converted into the actual concentration of hydrogen ions; however, during a titration all we care about is the location of inflection points. Whether we measure pH or pAH+, the inflection points will occur at the same titrant volumes. Chemicals and their Location

Stockroom Unknown Hydrochloric – Phosphoric acid mixture

Balance Room Your Drawer EDTA dihydrate, disodium salt Sodium hydroxide, standardized soln

Equipment and its Location

Stockroom Above/on lab bench Burette Electrode holder Magnetic stir bar Magnetic stir plate Weighing bottle pH meter & electrodes

Safety Issues and Chemical Hazard Information Physical Hazards Health Hazards Hydrochloric acid water-reactive, corrosive toxic Phosphoric acid corrosive none Sodium hydroxide water-reactive, corrosive toxic, irritant Both concentrated acids are highly corrosive. Be careful. Wear gloves while working with these chemicals.

CH455 Analytical Chemistry Week Three Lab Handout Page (8) Spring 2005 Procedure

Preparation for your next experiment 1. CH455 Water Hardness Prep (Required): You need to make your standard EDTA solution that will be used for your experiment next week. It takes a while to dissolve, so start this preparation closer to the beginning rather than the end of today’s lab session. If there is not a spare balance available, proceed with today’s experiment, but remember to come back and make this solution some time today. Follow Steps 1 & 2 in the Water Hardness Procedure.

Set-up 2. Because of the limited number of pH meters, this experiment will be carried out in groups of two or three. Decide whose standardized NaOH solution you will use, yours or your partner's (choose based on volume available and reliability of the standardization). Record the concentration of this solution. (Your TA will let you know the correct value if your standardization was inaccurate.) Clean, rinse, and fill your burette with the NaOH. Record the initial reading of the burette to the nearest 0.02 mL. 3. Collect a pH meter from the Stockroom. Note that the electrode is stored in a solution that protects the electrode and keeps it functioning well. Do not leave the electrode exposed to air for any long periods of time, either have it in the storage solution, or in one of your experimental solutions. Whenever you change the solution the electrode is in, you should rinse the electrode well using your squeeze bottle of distilled water so that you do not cross-contaminate either the storage container or your solutions. 4. You will use beakers rather than Erlenmeyer flasks for this titration, so you can fit both the burette tip and the pH electrode into the titration container. Make sure you use a beaker that is large enough. Set up the apparatus using an empty beaker as a guide. The beaker sits on the stirplate with a stirbar inside it. Once you have solution in your beaker, you will use the electrode holder to position the electrode in the beaker so that the stir bar will not bump it and so that the electrode doesn't interfere with the titration process.

Sample Preparation 5. Obtain your unknown acid mixture from the Stockroom in a clean, dry, labeled 100 mL beaker and ask them which one you have received (it will have an identifying letter or number). Use your 25 mL volumetric pipette to transfer 25.0 mL aliquots of your unknown solution into three clean, dry, labeled beakers. Dilute each sample with 40 mL of distilled water. No colour indicator is used for this titration. 6. Position the first beaker on the stirplate. Add the stirbar, and set the stirrer so it gently mixes the solution. You do not want a whirlpool o' death. Position the electrode and burette over the beaker.

CH455 Analytical Chemistry Week Three Lab Handout Page (9) Spring 2005 Potentiometric Titration 7. This titration differs from previous experiments — in this case, you continue to add titrant past the endpoint. The endpoint position is determined later from your data. The pH of your solution will vary rapidly close to each endpoint, and be fairly constant in between endpoints. Titrate your first unknown, adding 1.00 mL portions of your titrant, and recording the burette reading and the pH after each addition. This titration should be fairly quick and approximate. You will not collect a lot of data, but you get an idea of the position of the two endpoints, by identifying the two regions where the pH is changing rapidly. 8. Now that you know the approximate positions of the endpoints, titrate your other two unknown samples. To get a good value for your endpoints you must obtain a large number of data points near the endpoints of the titration. Start titrating using relatively large aliquots of base (1–2 mL), recording pH and burette reading after each addition. As the pH starts to change, add enough titrant in a single step to change the pH by 0.1 units. Stop and wait for the stir bar to mix the solution well, before recording each data point. Continue titrating in this way, until you have passed both endpoints and a negligible pH change is observed for each additional mL of titrant added.

Shutdown procedures IMPORTANT NOTE: Save your NaOH solution in case you find that something is wrong with your results from today's experiment. 9. Make sure you have recorded information for the sample you studied, in this case, the letter or number identifying your unknown. 10. Put all waste in the appropriate waste containers. Clean and dry all glassware. Clear up your work area. Return items to the Stockroom.

Waste Disposal Waste Chemicals Waste Container Location Unused HCl, dilute or concentrated Acid Waste cabinet Unused NaOH, dilute or concentrated Base Waste cabinet Excess unknown solution Acid Waste cabinet All solutions after titration Potentiometric Neutral Waste cabinet

Data Analysis

Endpoint determination 1. Tabulate the data for your second and third titration runs in a spreadsheet. (See the sample spreadsheet shown below.) The first two columns will be the burette reading and the pH at each data point. The third column is the total volume of base added up to that point, found by subtracting the initial burette

CH455 Analytical Chemistry Week Three Lab Handout Page (10) Spring 2005 reading from the reading at each data point. Correct for cases where you had to refill your burette. The first titration was approximate only and will not be used in your calculations. Sample Excel spreadsheet A B C D E F G 1 Burette pH Total Vol. DpH Average D2pH Avg. avg. 2 reading NaOH added DVol volume DVol2 volume 3 20.75 1.89 0 (B4-B3) (C4+C3) 4 = = 25.35 2.01 =A4-$A$3 (C4-C3) 2 (B5-B4) (C5+C4) (D5-D4) (E5+E4) 5 = = = = 29.12 2.21 =A5-$A$3 (C5-C4) 2 (E5-E4) 2

2. Make a single graph showing the two titrations, plotting the pH vs. Total volume added for each titration. Hide all data points and show just the line joining the points (see the Graphical Analysis section of this manual). The two endpoints for your titration occur where the pH is changing rapidly. Estimate the rough positions of the two endpoints as the midpoint of each vertical section of the graph (~9.8 mL NaOH in the Figure part a). Label the endpoints on your graph. Determine a smaller volume range that expands the area of interest of the titration; i.e., includes both endpoints. Plot all graphs for the rest of your analysis over this shorter volume range, using the same x-axis (volume) scale for each graph so that they can be directly compared to each other. 3. Each endpoint falls at the point where the slope of the titration curve is the greatest. Mathematically, this can be described as the point where the derivative (dpH⁄dV) is maximized. Make a new column in your spreadsheet and calculate the first derivative of your data by dividing the change in pH between two consecutive readings, by the change in added volume between the two consecutive readings (giving the change of pH per mL of added base); e.g.,

Volume pH added (mL) dpH 3.24 – 3.18 0.06 9.01 3.18 = = = 0.545 mL–1 9.12 3.24 dV 9.12 – 9.01 0.11

4. Make a new column in your spreadsheet and calculate an average volume to go with each derivative point, by finding the average of each pair of consecutive readings; e.g., for the data above (9.01 + 9.12) Vav = 2 = 9.065

5. Plot graphs of dpH⁄dV vs. Vav for each titration run, hiding all data points and showing just the line joining the points, to give you a first derivative curve for each titration. The first derivative of a curve shows the slope of the curve. The slope of your titration curve should peak at each of the endpoints (~9.78 mL NaOH in the Figure part b). Determine the two endpoints for each titration from your first derivative curves. Label the endpoints on your graphs.

CH455 Analytical Chemistry Week Three Lab Handout Page (11) Spring 2005 2 6. An alternative mathematical way to find the endpoints is to investigate the second derivative (d pH/dV2), the rate of change for the slope of the curve. Make a new column in your spreadsheet and calculate 2 d pH/dV2 by dividing the change in dpH⁄dV between two consecutive readings, by the change in the average added volume between the two consecutive readings; e.g.,

Volume pH Average dpH (ml–1) added Volume dV (mL) (mL) d2pH 0.864 – 0.545 .319 9.01 3.18 9.07 0.545 = = = 1.99 mL–2 9.12 3.24 9.23 0.864 dV2 9.23 – 9.07 0.16 9.34 3.43

7. Make a new spreadsheet column and calculate an "average average" volume to go with each second derivative point, by finding the average of each pair of consecutive average volumes; e.g.: (9.07 + 9.23) Vavav = 2 = 9.15 2 8. Plot graphs of d pH/dV2 vs. Vavav for each titration run, hiding all data points and showing just the line joining the points, to give you a second derivative curve for each titration. The second derivative of a curve shows the rate of change of the slope of a curve, which should be equal to zero whenever the first derivative is at a maximum (or a minimum). At each endpoint, the second derivative should cross the x- axis, i.e., equal zero (~9.77 mL NaOH in the Figure part c). The second derivative graph should give the most accurate determination of each endpoint. Determine the two endpoints for each titration from your second derivative curves. Label the endpoints on your graphs. (A good titration will have only a single crossing of the x-axis at each endpoint. If you have multiple crossings, use your best judgment to determine what value to choose.) 9. Use your graphs to determine your end point volumes as precisely as possible and explain how you chose them.

Concentration determination

10. Go back and reread the introduction of this lab, to make sure you understand what is going on. The first endpoint is due to NaOH neutralizing both acids, but the second endpoint is due only to the neutralization of the phosphoric acid. (Note that both acids react 1:1 with NaOH.) For each titration run, use the volume of NaOH needed to titrate from the first endpoint to the second endpoint, to calculate the number of moles of NaOH that reacted with phosophoric acid at the second endpoint, and hence the number of moles of phosphoric acid in your unknown sample. 11. For each titration run, use the volume of NaOH needed to titrate to the first endpoint to calculate the number of moles of NaOH that reacted with both unknown acids at the first endpoint, and hence the total number of moles of acid in your unknown sample. 12. The amount of phosphoric acid is the same at both endpoints; i.e., the number of moles of phosphoric acid in your unknown found from the second endpoint will be equal to the number of moles of phosphoric acid in your unknown at the first endpoint. Thus, determine the number of moles of

CH455 Analytical Chemistry Week Three Lab Handout Page (12) Spring 2005 hydrochloric acid by subtracting the number of moles of phosphoric acid found from the second endpoint from the total number of moles of acid found from the first endpoint. 13. Knowing the volume of your unknown aliquot, and the number of moles of each of the two acids in your unknown aliquot, calculate the molarity of the two acids, HCl and H3PO4, in your unknown sample.

Discussion Questions 1. We didn't calibrate the pH meters before doing this experiment. Why doesn't this matter? 2. HCl has one proton and H3PO4 has three. We see only two endpoints. Which endpoints do we see and why don’t we see four endpoints? 3. Explain clearly why we used a potentiometric titration for this analysis. Could we have used an indicator instead of a pH meter for this analysis? 6.5 (a) 6 Endpoint 5.5

5

4.5

4

3.5

3 9 9.5 10 10.5 11 V(mL) 8 40 (b) (c) 7 Endpoint 20 6

5 0 4 3 -20 2 Endpoint -40 1

0 -60 9 9.5 10 10.5 9 9.2 9.4 9.6 9.8 10 10.2 10.4 V(mL) V(mL) Sample potentiometric titration data analysis: (a) raw data, (b) first derivative curve, and (c) second derivative curve.

CH455 Analytical Chemistry Week Three Lab Handout Page (13) Spring 2005 From “Determination of Water Hardness using Complexometric titration”

Procedure Important Note: Any metal ions present in your glassware will react with EDTA and lead to incorrect results. All glassware coming in contact with your EDTA solution has been cleaned for you by the Stockroom by rinsing with 50% nitric acid followed by a distilled water rinse.

At least one week before Lab

Standard EDTA Solution

1. The disodium salt of EDTA dihydrate Na2H2Y.2H2O (the free acid is practically insoluble) is a primary standard. It is predried and kept in desiccators in the balance room. To minimize contamination, use the spatula provided in the desiccator. Using a clean dry weighing bottle, weigh accurately, by difference, ~0.25 g of disodium EDTA, into an acid-washed 250 mL volumetric flask. Use a clean dry funnel to ensure all of the accurately weighed solid is quantitatively transferred into the flask. Use a small amount of distilled water to carefully rinse the funnel to transfer any solid adhering to the funnel into the flask. Add ~200 mL of distilled water and swirl periodically until the EDTA has dissolved. 2. After the EDTA has dissolved, dilute to the mark and shake thoroughly. Rinse your clean polyethylene 500 mL bottle with three small portions of your EDTA solution, swirling to ensure the entire inner surface of the bottle has been rinsed, (to remove metals from the bottle’s inner surface), and transfer the rest of your EDTA solution to the bottle for storage. (Polyethylene is preferable to glass for storage because EDTA solutions leach metal ions from glass containers, resulting in a change in the concentration of free EDTA.)

CH455 Analytical Chemistry Week Three Lab Handout Page (14) Spring 2005 Spring 2005 CH455/6 Lab Report Title Page Student Use

Name: Class TA:

Lab Partners (if any): Makeup TA (if any):

Class TA Use

Prelab Quiz Score (out of 3): Lab finished: (TA signature required before student leaves lab) Due 9 / 9:30 am; 2 pm; Experiment Date: Report Due Date: Time: 5 pm; 7 pm Report Received

Date: Time: Initials: Grading

Report Grade (/15): Late penalty (/15): Report inc. penalty (/15):

Regrade (/13.5): Final Grade (/15) (includes penalty & regrade):

Potentiometric titration of a HCl–H3PO4 mixture

Your report should follow the Lab Report format listed earlier in this manual. Show samples of all calculations, identify any unknowns, report the average with appropriate significant figures and answer the discussion questions. Don't forget to include your conclusion section. Include all graphs mentioned in this data analysis in your lab report.

Average

HCl concentration in unknown

H3PO4 concentration in unknown

Letter identifying your unknown Potentiometric titration of a HCl–H3PO4 mixture Report Grading Breakdown

Your Out of a Score / possible Pre-lab Quiz / 3 Calculations/Data Analysis / 3 Results/Accuracy / 3 Discussion & Conclusions / 3 Presentation/Impression / 3

Subtractions