<<

KERRY GREENE-DONNELLY Kerry Greene-Donnelly has over 18 years of experience in Medical . Kerry has experience and is certified through the ARRT in , , Computed and Quality Management. Kerry has an AAS in Radiologic Technology, a BPS in Health Service Management and a MBA. Currently Kerry is an Assistant Professor at Upstate Medial University in Syracuse NY. Kerry has taught for 12 years in the Sciences program, as well as, , and Assistant programs. Kerry's research interests are in CT dosimetry and image quality  Spatial Resolution  Ability to image small objects that have high subject contrast (1).

 CT has moderate spatial resolution 20 lp/cm (1)  Contrast Resolution

 Ability to distinguish between and image similar tissues (1)  CT has excellent low contrast detectability 0.25-0.5 % difference in tissue (2)  Temporal Resolution

 Ability to freeze motion of the object being imaged (2)  CT has moderate temporal resolution, ~300ms for full segment reconstruction down to 75ms with 4 segment reconstruction. (2)

 In plane  x/y axis

 Through plane  Z axis

 Focal Spot Size  Detector Size  Scanner Geometry  Sampling Frequency  Reconstruction Algorithm  Field of View (FOV)

 Inversely related to detail.  Increased detail requires decreased FSS  FSS is not a selection made by the technologist it is mA dependant and automatically selected.  Flying Focal spots (Siemens Somatom) allow for slight change in focus, increase in data points and sampling,. Increasing detail in the z axis or through

plane. (3)  Inversely related to detail.  Increased detail requires decreased size or space between detector elements.  Detector size or spacing is not a selection made by the technologist it is dependant on the make and model of CT scanner.

 Directly related to detail.  Increased detail requires increased scan geometry.  Describes the arc or amount of tube movement.

Several types. (2)  One 360 most common, even dose distribution  Two 180 average 2 data sets, increased quality  One 180 ½ scan, fast, low dose, used in peds  One 400 over scan, increased data and quality

 Directly related to spatial resolution  Increased sampling increases detail in an image  Number of projections is also known as the “sampling frequency”  This is accomplished by increasing the sampling rate or the speed at which the 128 vs. 1024 projections detectors are read.

Bsc Biomedical Sciences  Directly related to spatial resolution  As the spatial frequency of a filter increases the detail increases.  High frequency kernels may be know as “sharp, edge, or FC 50+.

Bsc Biomedical Sciences  Pixel- inversely related to detail, smaller is better  Matrix- directly related to detail, bigger is better Technologist do not control pixel or matrix directly  FOV- inversely related to detail, smaller is better Technologist controls FOV used for scanning /viewing  Pixel = FOV/ matrix

500 mm FOV 100 mm FOV Computed Tomography 3rd,Seeram 2009, Elsevier Publishing.  Slice Thickness  Pitch

 Inversely related to spatial resolution  Increased slice thickness, scan or viewing, decreases detail in an image.  Thick slices have an increased SSP and more partial volume averaging.

 Inversely related to spatial resolution  Increases in pitch will result in a decrease in image detail and increase in partial volume averaging,  It is important in MSCT to use a pitch that will provide non overlapping spirals, i.e. 4 slice with a pitch of 1.5, we will have interlacing spirals not over lacing.  Object Size  Object attenuation difference  Noise  kVp  mA(s)  Pitch  Slice thickness

 Object Size is directly related to contrast resolution, larger objects are easily visualized.  Object attenuation is directly related to contrast resolution, high attenuation objects are easily seen.  Small, low attenuating objects are the most difficult to image. Orange arrow

health.siemens.com  Variation in pixel values of a homogenous material over a specific area. Often described as a mottle or grainy appearance.

 Noise is caused by low count in an image. Anything that decreases the exposure to the patient, and therefore the detectors will increase noise.

 Noise is measured with a 20 cm water phantom, taking an ROI and viewing the standard deviation within the area.

 SD alone cannot fully account for the noise in an image, one can manipulate factors so that the SD is equal between images; however, the perception of noise or quality is very different- this concept is called noise power spectrum. (2)  mA  S  kVp  Filtration  Pixel size  Slice thickness  Detector efficiency  Interpolation scheme

Bsc Biomedical Sciences  Indirectly related to noise, and directly related to contrast resolution.  Increases in mAs, decrease noise and increase the ability to see slight differences in attenuation.  To half the noise in an image the mAs must increase 4 fold. i.e. 50 mAs must increase to 200mAs to decrease noise to ½ the original.  mAs is under the control of the technologist.  Indirectly related to noise, and directly related to contrast resolution.  Increases in mAs, decrease noise and increase the ability to see slight differences in attenuation.  One must remember that kVp has significant impact on PE absorption and attenuation in the subject.  kVp is under the direct control of the technologist, but due to its complex relationship with contrast resolution it should not be a first step to decreasing noise in an examination.  Indirect relationship with noise.  As the slice or pixel size decreases noise will increase and contrast resolution will decrease.  Scan slice thickness is often very thin, 1mm. While viewing is thicker, 4mm. Averaging 4 1mm slices into one thick 4mm will increase signal by 4X, but noise will increase √4 =2X.  Thicker slices or larger pixels will have better efficiency with respect to signal to noise.

Computed Tomography 3rd, Kalendar 2011, Publics Publishing.  Rotational speed  Specialized reconstruction algorithms  Gating studies

 Decreasing the affects of subject motion on image quality is best resolved with increased temporal resolution or faster scanning.  Decreased rotation times, from 1s to ½ sec is common. The fasted rotation times available are ~300ms.  Cardiac imaging requires excellent temporal resolution to visualize small cardiac vessels.  Specialized equipment and techniques address the time issue in cardiac imaging.  Cardiac studies are generally performed with a pitch of 0.2. Resulting in 5 images of the same .  Using segments from several rotations to build an image decrease the overall time. For example, 300ms full rotation, using 4 segments for reconstruction decrease image time to 75ms.  Gating tags each image to a point in the heart motion. The computer can then select only images from a specific point in the cardiac motion for reconstruction.  The image below shows a set rotation time, with a changing heart rate, as the heart rate increase so does the blur. Even at very low heart rates, 40 bpm, motion is evident.  Image quality and dose are dependent factors  Generally to increase image quality, dose increases.  Consideration must always be taken when balancing patient dose with a required level of quality.  Consultation with a Radiologist and Medical Physicist , will best assist the Technologist in producing acceptable image quality at the lowest dose possible (ALARA). CTDI is measured using a acrylic phantom and an , a pencil chamber. CTDI has several types

CTDI 100 Accounts for a slice and scatter over 100mm CTDI w Accounts for increased exposure to the skin surface and decreased exposure to the center CTDI vol Accounts for pitch in spiral imaging, increasing pitch decreases the exposure to the patient  Most scanners will provide an estimated CTDI vol, prior to the start of a scan.  An estimated CTDI vol will be given at the termination of the exam.  It is important to remember that these are estimates and NOT patient dose.  Patient dose maybe estimated using the AAPM report 204 and scanner generated CTDI  An additional value is DLP, this is the CTDI vol * the scan length.  CTDI is useful in comparing scanner to scanner or protocol to protocol. It is most often used in accreditation and QC of a scanner.  CTDI reports scanner output, the question is what amount of tissue is that going into.  For the same CTDI vol, a small patient will have a higher dose. –more radiation, less tissue.  CTDI vol tends to overestimate dose to a large patient and underestimate for small patient.  CTDI vol is a helpful metric, it is not the whole picture.

 Adult  Head 75 mGy  Abdomen 25 mGy  Pediatric  Abdomen 20 mGy (4)  Filtration  Decreases dose by filtering low energy rays 2.5mm Al minimum, most units have 6-9 mm AL, proper filtration also decreases beam hardening artifacts  Technologist only control the beam shaping filter by selecting the correct head or body protocol.

 Increasing kVp without a corresponding decrease in mAs, increases patient dose.  140kVp increases dose by 47% compared to 120 kVp .  However, if mAs is decrease appropriately dose is reduced by 20%.  Lower kVp should be selected for due to high subject contrast. Increased noise may be tolerable in high subject contrast areas.

 Increased mAs, mA or a decreased rotation time, increases patient dose.  Adjusting the mAs patient size is important for dose and quality. If the patient is 3-4 cm smaller than the norm, decrease mAs by half.  This illustrates how great a decrease in mAs is needed for children and small adults.

 Thin slices, increase dose as more need to be taken to cover a set area and increased mAs is required to maintain noise levels.  Thick slices decrease dose, less slices needed with less mAs.  Radiation is most efficiently used with larger number of slices taken per rotation.  16 , 1.25 mm slices are more dose efficient than 4, 1.25mm slices. (5)

 Pitch refers the ratio of table feed to beam width. When they are equal there is no gap or overlap in radiation.  Increment is used in axial imaging, again it is a ration of the slice thickness to table feed. It describes if slices are contiguous, overlapping or leaving gaps.  Overlap increases patient dose, where gaps decrease dose.

 Similar in nature to AEC in radiography, mA or dose modulation software will vary the exposure over various body parts to maintain image quality.  As with radiography, patients must be positioned properly in the gantry, have no metal implements and shields must not be in the FOV.  For an abdomen a ~5 cm off center in the X axis would result in ~30% increase in dose. A y axis miss center of ~4 cm would result in a ~75% increase in dose  The mAs is set both at a minimum and maximum level. The software then varies the mAs for the anatomy and view being acquired.

 May be called many things by different manufacturers. The mAs can be modulated in plane X/Y, angular modulation or through plan, z axis.

 Image noise level is selected and the mAs is modulated to maintain quality while saving on dose.

 Angular Modulation (ap vs. lat) Smartscan- GE DOM-dose-Phillips CareDose- Siemens

 Angular-Longitudinal Modulation (ap vs. lat and changing anatomy over the z-axis) Smart mA- GE Z-DOM –Phillips CareDose 4D- Siemens SureExposure- Toshiba

 1. Radiologic Sciences for Technologists 10th, Bushong 2013, Elsevier Publishing.  2. Computed Tomography 3rd,Seeram 2009, Elsevier Publishing.  3. Computed Tomography 3rd, Kalendar 2011, Publics Publishing.  4. www.acr.org  5. AAPM Computed Tomography Radiation Dose Education Slides 9/6/12