Photomultiplier Tubes Construction and Operating Characteristics Connections to External Circuits PHOTOMULTIPLIER TUBES Construction and Operating Characteristics

INTRODUCTION Variants of the head-on type having a large-diameter hemi- Among the photosensitive devices in use today, the photo- spherical window have been developed for high multiplier tube (or PMT) is a versatile device that provides ex- experiments where good angular acceptability is important. tremely high sensitivity and ultra-fast response. A typical photo- multiplier tube consists of a photoemissive (photocath- Figure 2: External Appearance ode) followed by focusing , an multiplier and a) Side-On Type b) Head-On Type an electron collector () in a tube, as shown in Fig- PHOTO- ure 1. SENSITIVE AREA When light enters the , the photocathode emits photoelectrons into the vacuum. These photoelectrons are then PHOTO- directed by the focusing towards the electron SENSITIVE AREA multiplier where are multiplied by the process of sec- ondary emission. The multiplied electrons are collected by the

H A M A M anode as an output signal. A T S U R928 R 3 M H A M A M A T S U AD E MA IN DE IN JA PA N Because of secondary-emission multiplication, photomulti- JAPA N plier tubes provide extremely high sensitivity and exceptionally low among the photosensitive devices currently used to detect radiant energy in the , visible, and near TPMSC0028EA TPMOC0083EA regions. The also features fast time re- sponse, low noise and a choice of large photosensitive areas. Figure 3: Types of Photocathode This section describes the prime features of photomultiplier a) Reflection Mode tube construction and basic operating characteristics. REFLECTION MODE PHOTOCATHODE Figures 1: Cross-Section of Head-On Type PMT DIRECTION OF LIGHT FOCUSING ELECTRODE PHOTOELECTRON SECONDARY TPMSC0029EA LAST STEM PIN ELECTRON VACUUM b) Transmission Mode (10- 4 Pa) DIRECTION OF LIGHT e- SEMITRANSPARENT PHOTOCATHODE

FACEPLATE STEM DIRECTION ELECTORON ANODE OF LIGHT MULTIPLIER () PHOTOELECTRON PHOTOCATHODE TPMHC0084EB

TPMHC0006EA CONSTRUCTION The photomultiplier tube generally has a photocathode in ei- The superior sensitivity (high current amplification and high ther a side-on or a head-on configuration. The side-on type re- S/N ratio) of photomultiplier tubes is due to the use of a low-noise ceives incident light through the side of the glass bulb, while in electron multiplier which amplifies electrons by a cascade sec- the head-on type, it is received through the end of the glass bulb. ondary electron emission process. The electron multiplier con- In general, the side-on type photomultiplier tube is relatively low sists of from 8, up to 19 stages of electrodes called dynodes. priced and widely used for spectrophotometers and general pho- There are several principal types in use today. tometric systems. Most of the side-on types employ an opaque 1) Circular-cage type photocathode (reflection-mode photocathode) and a circular- The circular-cage is generally used for the side-on type of cage structure electron multiplier which has good sensitivity and photomultiplier tube. The prime features of the circular-cage high amplification at a relatively low supply . are compactness and fast time response. The head-on type (or the end-on type) has a semitransparent photocathode (transmission-mode photocathode) deposited upon the inner surface of the entrance window. The head-on type provides better spatial uniformity (see page 7) than the side-on type having a reflection-mode photocathode. Other fea- tures of head-on types include a choice of photosensitive areas from tens of square millimeters to hundreds of square centime- ters. TPMOC0077EB

- 1 - 2) Box-and-grid type 6) Microchannel plate (MCP) This type consists of a train of quarter cylindrical dynodes The MCP is a thin disk consisting of millions of micro and is widely used in head-on type photomultiplier tubes be- glass tubes (channels) fused in parallel with each other. Each cause of its relatively simple dynode design and improved channel acts as an independent electron multiplier. The MCP uniformity, although time response may be too slow in some offers much faster time response than the other discrete dy- applications. nodes. It also features good immunity from magnetic fields and two-dimensional detection ability when multiple are used.

TPMOC0078EA

3) Linear-focused type TPMOC0082EA The linear-focused type features extremely fast response time and is widely used in head-on type photomultiplier tubes 7) Metal channel type where time resolution and pulse linearity are important. The Metal channel dynode has a compact dynode costruction manufactured by our unique fine machining tech- nique. It achieves high speed response due to its narrower space between each stage of dynodes than the other type of conventional dynode construction. TPMOC0079EA It is also adequate for position sensitive measurement. 4) Venetian blind type The venetian blind type has a large dynode area and is ELECTRON primarily used for tubes with large photocathode areas. It of- fers better uniformity and a larger pulse output current. This structure is usually used when time response is not a prime consideration.

TPMOC0084EA

TPMOC0080EA In addition, hybrid dynodes combining two of the above dy- 5) Mesh type nodes are available. These hybrid dynodes are designed to The mesh type has a structure of fine mesh electrodes provide the merits of each dynode. stacked in close proximity. This type provides high immunity to magnetic fields, as well as good uniformity and high pulse linearity. In addition, it has position-sensitive capability when SPECTRAL RESPONSE used with cross- anodes or multiple anodes. The photocathode of a photomultiplier tube converts energy of incident light into photoelectrons. The conversion efficiency (photocathode sensitivity) varies with the wavelength of the inci- dent light. This relationship between photocathode sensitivity and wavelength is called the spectral response characteristic. Figure 4 shows the typical spectral response of a bialkali photo- multiplier tube. The spectral response characteristics are deter- mined on the long wavelength side by the photocathode material and on the short wavelength side by the window material. Typical spectral response characteristics for various types of photomulti- ELECTRON ELECTRON ELECTRON plier tubes are shown on pages 88 and 89. In this catalog, the longwavelength cut-off of spectral response characteristics is de- fined as the wavelength at which the cathode radiant sensitivity becomes 1% of the maximum sensitivity for bialkali and Ag-O-Cs , and 0.1% of the maximum sensitivity for multialkali photocathodes.

COARSE MESH TYPE FINE-MESH TYPE Spectral response characteristics are typical curves for repre- sentative tube types. Actual data may be different from type to TPMOC0081EA type. - 2 - Figure 4: Typical Spectral Response of Head-On, Bialkali 6) High temperature bialkali or low noise bialkali Photocathode (Na-K-Sb) This is particularly useful at higher operating tempera- 100 tures since it can withstand up to 175°C. A major application is in the oil industry. At room temperatures, this photocathode operates with very low , making it 10 ideal for use in counting applications. 7) Multialkali (Na-K-Sb-Cs) The multialkali photocathode has a high, wide spectral re- CATHODE RADIANT 1 sponse from the ultraviolet to near infrared region. It is widely SENSITIVITY used for broad-band spectrophotometers. The long wave- length response can be extended out to 930nm by special QUANTUM EFFICIENCY (%) photocathode processing. 0.1 8) Cs-Te, Cs-I CATHODE RADIANT SENSITIVITY (mA/W) These materials are sensitive to vacuum UV and UV rays but not to visible light and are therefore called solar blind. Cs- 0.01 Te is quite insensitive to wavelengths longer than 320nm, 200 400 600 800 and Cs-I to those longer than 200nm. WAVELENGTH (nm)

TPMOB0070EA WINDOW MATERIALS The window materials commonly used in photomultiplier PHOTOCATHODE MATERIALS tubes are as follows: The photocathode is a photoemissive surface usually consist- 1) ing of alkali metals with very low work functions. The photocath- This is frequently used glass material. It transmits radia- ode materials most commonly used in photomultiplier tubes are tion from the near infrared to approximately 300nm. It is not as follows: suitable for detection in the ultraviolet region. For some appli- 1) Ag-O-Cs cations, the combination of a bialkali photocathode and a The transmission-mode photocathode using this material low-noise borosilicate glass (so called K-free glass) is used. is designated S-1 and sensitive from the visible to infrared The K-free glass contains very low (K2O) which range (300 to 1200nm). Since Ag-O-Cs has comparatively can cause background counts by 40K. In particular, tubes de- high thermionic dark emission (refer to "ANODE DARK CUR- signed for scintillation counting often employ K-free glass not RENT" on page 8), tubes of this photocathode are mainly only for the faceplate but also for the side bulb to minimize used for detection in the near infrared region with the photo- noise pulses. cathode cooled. 2) UV-transmitting glass (UV glass) 2) GaAs(Cs) This glass transmits ultraviolet radiation well, as the name GaAs activated in cesium is also used as a photocathode. implies, and is widely used as a borosilicate glass. For spec- The spectral response of this photocathode usually covers a troscopy applications, UV glass is commonly used. The UV wider spectral response range than multialkali, from ultravio- cut-off is approximately 185nm. let to 930nm, which is comparatively flat over 300 to 850nm. 3) Synthetic silica 3) InGaAs(Cs) The synthetic silica transmits ultraviolet radiation down to This photocathode has greater extended sensitivity in the 160nm and offers lower absorption in the ultraviolet range infrared range than GaAs. Moreover, in the range between compared to fused silica. Since thermal expansion coefficient 900 and 1000nm, InGaAs has much higher S/N ratio than Ag- of the synthetic silica is different from which is used for O-Cs. the tube leads, it is not suitable for the stem material of the 4) Sb-Cs tube (see Figure 1 on page 1). Borosilicate glass is used for This is a widely used photocathode and has a spectral the stem, then a graded seal using glasses with gradually response in the ultraviolet to visible range. This is not suited different thermal expansion coefficients are connected to the for transmission-mode photocathodes and mainly used for re- synthetic silica window. Because of this structure, the graded flection-mode photocathodes. seal is vulnerable to mechanical shock so that sufficient care 5) Bialkali (Sb-Rb-Cs, Sb-K-Cs) should be taken in handling the tube. These have a spectral response range similar to the Sb- 4) MgF2 (magnesium fluoride) Cs photocathode, but have higher sensitivity and lower noise The crystals of alkali halide are superior in transmitting than Sb-Cs. The transmission mode bialkali photocathodes ultraviolet radiation, but have the disadvantage of deliques- also have a favorable blue sensitivity for flashes cence. Among these, MgF2 is known as a practical window from NaI (Tl) , thus are frequently used for radia- material because it offers low deliquescence and transmits tion measurement using scintillation counting. ultraviolet radiation down to 115nm. - 3 - Figure 5: Typical Transmittance of Various Window Materials test sheets accompanying the tubes usually indicate these pa- rameters except for tubes with Cs-I or Cs-Te photocathodes,

100 which are not sensitive to lamp light. (Radiant sensitiv- ity at a specific wavelength is listed for those tubes instead.) Both the cathode and anode luminous sensitivities are ex- pressed in units of A/lm (amperes per lumen). Note that the lu- BOROSILICATE UV- GLASS men is a unit used for luminous flux in the visible region and TRANSMITTING GLASS therefore these values may be meaningless for tubes which are 10 MgF2 sensitive beyond the visible region. (For those tubes, the blue sensitivity or red/white ratio is often used.) SYNTHETIC QUARTZ TRANSMITTANCE (%) Figure 6: Typical Human Eye Response and Spectral Energy Distribution of 2856K Tungsten Lamp

1 100 100 120 160 200 240 300 400 500

WAVELENGTH (nm) TUNGSTEN 80 TPMOB0076EB LAMP AT 2856K

As stated above, spectral response range is determined by 60 the photocathode and window materials. It is important to select an appropriate combination which will suit your applications. 40

VISUAL SENSITIVITY RELATIVE VALUE (%) 20 RADIANT SENSITIVITY AND QUANTUM EFFICIENCY

As Figure 4 shows, spectral response is usually expressed in 0 terms of radiant sensitivity or quantum efficiency as a function of 200 400 600 800 1000 1200 1400 wavelength. Radiant sensitivity (S) is the photoelectric current WAVELENGTH (nm) TPMOB0054EB from the photocathode, divided by the incident radiant power at a given wavelength, expressed in A/W (amperes per watt). Quan- tum efficiency (QE) is the number of photoelectrons emitted from BLUE SENSITIVITY AND RED/WHITE RATIO the photocathode divided by the number of incident . It is For simple comparison of spectral response of photomulti- customary to present quantum efficiency in a percentage. Quan- plier tubes, cathode blue sensitivity and red/white ratio are often tum efficiency and radiant sensitivity have the following relation- used. ship at a given wavelength. The cathode blue sensitivity is the photoelectric current from the photocathode produced by a light flux of a tungsten lamp at S × 12400 QE = × 100% λ 2856K passing through a blue filter (Corning CS No. 5-58 pol- ished to half stock thickness). Since the light flux, once transmit- Where S is the radiant sensitivity in A/W at the given wavelength, ted through the blue filter cannot be expressed in lumens, blue λ and is the wavelength in nm (nanometers). sensitivity is conveniently expressed in A/lm-b (amperes per lu- men-blue). The blue sensitivity is an important parameter in scin- tillation counting using an NaI (Tl) scintillator since the NaI (Tl) LUMINOUS SENSITIVITY scintillator produces emissions in the blue region of the spec- Since the measurement of the spectral response characteris- trum, and may be the decisive factor in energy resolution. tic of a photomultiplier tube requires a sophisticated system and The red/white ratio is used for photomultiplier tubes with a much time, it is not practical to provide customers with spectral spectral response extending to the near infrared region. This pa- response characteristics for each tube ordered. Instead cathode rameter is defined as the quotient of the cathode sensitivity mea- or anode luminous sensitivity is commonly used. sured with a light flux of a tungsten lamp at 2856K passing The cathode luminous sensitivity is the photoelectric current from through a red filter ( IR-D80A for the S-1 photocathode or the photocathode per incident light flux (10-5 to 10-2 lumens) from R-68 for others) divided by the cathode luminous sensitivity with a tungsten filament lamp operated at a distribution temperature the filter removed. of 2856K. The anode luminous sensitivity is the anode output current (amplified by the process) per inci- dent light flux (10-10 to 10-5 lumens) on the photocathode. Al- though the same tungsten lamp is used, the light flux and the applied voltage are adjusted to an appropriate level. These pa- rameters are particularly useful when comparing tubes having the same or similar spectral response range. final - 4 - Figure 7: Transmittance of Various Filters Figure 8: Typical Current Amplification vs. Supply Voltage

104 109 100

TOSHIBA R-68 103 ANODE LUMINOUS 108 80 CORNING SENSITIVITY CS-5-58 (1/2 STOCK 2 7 60 THICKNESS) 10 10

40 101 106 TRANSMITTANCE (%) TOSHIBA 20 IR-D80A 100 105 CURRENT AMPLIFCATION

0 ANODE LUMINOUS SENSITIVITY (A / lm) 200 400 600 800 1000 1200 10-1 104 WAVELENGTH (nm) CURRENT AMPLIFICATION ()

TPMOB0055EB 10-2 103 200 300 500 700 1000 1500

SUPPLY VOLTAGE (V) TPMOB0058EA CURRENT AMPLIFICATION (GAIN) Photoelectrons emitted from a photocathode are accelerated by an so as to strike the first dynode and produce ANODE DARK CURRENT secondary electron emissions. These secondary electrons then A small amount of current flows in a photomultiplier tube even impinge upon the next dynode to produce additional secondary when the tube is operated in a completely dark state. This output electron emissions. Repeating this process over successive dy- current, called the anode dark current, and the resulting noise node stages, a high current amplification is achieved. A very are critical factors in determining the detectivity of a photomulti- small photoelectric current from the photocathode can be ob- plier tube. As Figure 9 shows, dark current is greatly dependent served as a large output current from the anode of the photomul- on the supply voltage. tiplier tube. Current amplification is simply the ratio of the anode output Figure 9: Typical Dark Current vs. Supply Voltage current to the photoelectric current from the photocathode. Ide- (AFTER 30 MINUTE STORAGE) ally, the current amplification of a photomultiplier tube having n 101 dynode stage and an average secondary emission ratio δ per stage is δn. While the secondary electron emission ratio δ is given by 100 δ =A • Eα

α where A is constant, E is an interstage voltage, and is a coeffi- 10-1 cient determined by the dynode material and geometric struc- ture. It usually has a value of 0.7 to 0.8. When a voltage V is applied between the cathode and the anode ANODE DARK CURRENT (nA) of a photomultiplier tube having n dynode stages, current amplifi- 10-2 cation, µ, becomes

V α n µ = δn = (A Eα)n = A { ()} 10-3 n+1 400 600 800 1000 1200 1400 n A αn αn = V = K V APPLIED VOLTAGE (V) (n+1)αn TPMOB0071EA

Since photomultiplier tubes generally have 9 to 12 dynode Major sources of dark current may be categorized as follows: stages, the anode output varies directly with the 6th to 10th 1) of electrons power of the change in applied voltage. The output signal of the Since the materials of the photocathode and dynodes photomultiplier tube is extremely susceptible to fluctuations in have very low work functions, they emit thermionic electrons the voltage, thus the power supply must be very even at room temperature. Most of dark currents originate stable and provide minimum ripple, drift and temperature coeffi- from the thermionic emissions, especially those from the cient. Various types of regulated high-voltage power supplies photocathode as they are multiplied by the dynodes. Cooling designed with this consideration are available from Hamamatsu . the photocathode is most effective in reducing thermionic emission and, this is particularly useful in applications where - 5 - low dark counts are essential such as in . 4) Leakage current (ohmic leakage) Figure 10 shows the relationship between dark current Leakage current resulting from the glass stem base and and temperature for various photocathodes. Photocathodes socket may be another source of dark current. This is pre- which have high sensitivity in the red to infrared region, espe- dominant when the photomultiplier tube is operated at a low cially S-1, show higher dark current at room temperature. voltage or low temperature. The flatter slopes in Figures 9 Hamamatsu provides thermoelectric coolers (C659 and and 10 are mainly due to leakage current. C4877) designed for various sizes of photomultiplier tubes. Contamination from dirt and moisture on the surface of the tube may increase the leakage current, and therefore should Figure 10: Temperature Characteristics of Dark Current be avoided.

10-5 5) Field emission When a photomultiplier tube is operated at a voltage near 10-6 R316 the maximum rated value, electrons may be emitted from (HEAD-ON TYPE, Ag-O-Cs) electrodes by the strong electric field and may cause noise 10-7 pulses. It is therefore recommended that the tube be oper-

10-8 R374 ated at a voltage 20 to 30% lower than the maximum rating. (HEAD-ON TYPE, MULTIALKALI) The anode dark current decreases with time after the tube is 10-9 placed in a dark state. In this catalog, anode dark currents are measured after 30-minute storage in a dark state. 10-10

ANODE DARK CURRENT (A) 10-11 R3550 ENI (EQUIVALENT NOISE INPUT) (HEAD-ON TYPE, 10-12 LOW-NOISE BIALKALI) ENI is an indication of the photon-limited signal-to-noise ratio. R6095 It refers to the amount of light usually in watts or lumens neces- (HEAD-ON TYPE, BIALKALI) 10-13 sary to produce a signal-to-noise ratio of unity in the output of a -60-40 -20 0 20 40 photomultiplier tube. ENI is expressed in units of lumens or watts. TEMPERATURE (°C) For example the value of ENI (in watts) is given by TPMOB0065EB

2) of residual gases ( feedback) 2q Idb µ ∆f ENI = (watts or lumens) Residual gases inside a photomultiplier tube can be ion- S ized by collision with electrons. When these strike the where photocathode or earlier stages of dynodes, secondary elec- q = electronic charge (1.60 ×10-19 coul.) trons may be emitted, thus resulting in relatively large output Idb = anode dark current in amperes after 30-minute noise pulses. These noise pulses are usually observed as storage in darkness afterpulses following the primary signal pulses and may be a µ = current amplification ∆ problem in detecting light pulses. Present photomultiplier f = bandwidth of the system in hertz (usually 1 hertz) S = anode radiant sensitivity in amperes per watt at tubes are designed to minimize afterpulses. the wavelength of interest or anode luminous 3) Glass scintillation sensitivity in amperes per lumen When electrons deviating from their normal trajectories strike the glass envelope, scintillations may occur and dark For the tubes listed in this catalog, the value of ENI may be calcu- pulses may result. To minimize this type of dark pulse, photo- lated by the above equation. Usually it has a value between 10-15 multiplier tubes may be operated with the anode at high - and 10-16 watts or lumens. age and the cathode at ground potential. But this is inconve- nient to handle the tube. To obtain the same effect without difficulty, Hamamatsu provides "HA coating" in which the EFFECTS glass bulb is coated with a conductive paint connected to the Most photomultiplier tubes are affected by the presence of cathode. (See "GROUND POLARITY AND HA COATING" magnetic fields. Magnetic fields may deflect electrons from their on page 10.) normal trajectories and cause a loss of gain. The extent of the loss of gain depends on the type of photomultiplier tube and its orientation in the magnetic field. Figure 11 shows typical effects of magnetic fields on some types of photomultiplier tubes. In gen- eral, tubes having a long path from the photocathode to the first dynode are very vulnerable to magnetic fields. Therefore head- on types, especially large diameter tubes, tend to be more ad- versely influenced by magnetic fields.

- 6 - Figure 11: Typical Effects by Magnetic Fields Perpendicular Hamamatsu provides photomultiplier tubes using fine mesh to Tube Axis dynodes. These tube types exhibit much higher immunity to ex- ternal magnetic fields than the photomultiplier tubes using other 120 dynodes. In addition, when the light level to be measured is

110 28 mm dia. rather high, or type photomultiplier tubes can be SIDE - ON TYPE used in hishly magnetic fields. 100

90

80

(%) SPATIAL UNIFORMITY 70 Spatial uniformity is the variation of sensitivity with position of 60 incident light on a photocathode. 13 mm dia. 50 HEAD - ON TYPE Although the focusing electrodes of a photomultiplier tube LINEAR - FOCUSED ( TYPE DYNODE ) 40 are designed so that electrons emitted from the photocathode or RELATIVE OUTPUT

30 38 mm dia. dynodes are collected efficiently by the first or following dynodes, HEAD - ON TYPE CIRCULAR CAGE some electrons may deviate from their desired trajectories in the 20 ( TYPE DYNODE ) focusing and multiplication processes, resulting in a loss of col- 10 lection efficiency. This loss of collection efficiency varies with the 0 -0.5 -0.4 -0.3 -0.2 -0.10 0.10.2 0.3 0.4 0.5 position on the photocathode from which the photoelectrons are emitted and influences the spatial uniformity of a photomultiplier TPMOB0017EB MAGNETIC FLUX DENSITY (mT) tube. The spatial uniformity is also determined by the photocath- When a tube has to be operated in magnetic fields, it may be ode surface uniformity itself. necessary to shield the tube with a magnetic shield case. In general, head-on type photomultiplier tubes provide better Hamamatsu provides a variety of magnetic shield cases. To ex- spatial uniformity than side-on types because of the photocath- press the effect of a magnetic shield case, the magnetic shield- ode to first dynode geometry. Tubes especially designed for ing factor is used. This is the ratio of the strength of the magnetic applications have excellent spatial uniformity, because uniformity is the decisive factor in the overall perfor- field outside the shield case, Hout, to that inside the shield case, Hin. It is determined by the permeability µ, the thickness t (mm) mance of a gamma camera. and inner diameter D (mm) of the shield case, as follows: Figure 13: Examples of Spatial Uniformity Hout 3µt (a) Head-On Type (b) Side-On Type = Hin 4 D (R6231-01 for gamma camera applications) Reflection-mode photocathode It should be noted that the magnetic shielding effect de- 100 PHOTO- creases towards the edge of the shield case as shown in Figure CATHODE ANODE (TOP VIEW) 50 SENSITIVITY (%) 12. It is recommended that the tube be covered by a shield case 050 100 longer than the tube length by at least half the tube diameter. 0 ANODE SENSITIVITY (%)

Figure 12: Edge Effect of Magnetic Shield Case

PHOTO- 100 CATHODE EDGE EFFECT

t LONGER THAN r 50 GUIDE KEY

2r PMT 0 ANODE SENSITIVITY (%) L (R6231-01 for gamma camera applications) 1000

100 TPMHC0085EB TPMSC0030EC

10

1 SHIELDING FACTOR(Ho/Hi) r r TPMOB0011EA

- 7 - TEMPERATURE CHARACTERISTICS DRIFT AND LIFE CHARACTERISTIC By decreasing the temperature of a photomultiplier tube, While operating a photomultiplier tube continuously over a dark current originating from thermionic emission can be re- long period, anode output current of the photomultiplier tube may duced. Sensitivity of the photomultiplier tube also varies with the vary slightly with time, although operating conditions have not temperature. In the ultraviolet to visible region, the temperature changed. This change is reffered to as drift or in the case where coefficient of sensitivity usually has a negative value, while near the operating time is 103 to 104 hrs it is called life characteristics. the long wavelength cut-off it has a positive value. Figure 14 Figure 16 shows typical life characteristics. shows temperature coefficients vs. wavelength of typical photo- Drift is primarily caused by damage to the last dynode by multiplier tubes. Since the temperature coefficient change is heavy electron bombardment. Therefore the use of lower anode large near the long wavelength cutoff, temperature control may current is desirable. When stability is of prime importance, the be required in some applications. use of average anode current of 1µA or less is recommended.

Figure 14: Typical Temperature Coefficients of Anode Sen- Figure 16: Examples of Life sitivity

PMT:R1924 1.5 SUPPLY VOLTAGE:1000 V INITIAL ANODE CURRENT:10 µA C] ° 1 BIALKALI 100 MULTIALKALI Sb-Cs

0.5 Cs-Te GaAs (Cs)

50 0 TEMPERATURE COEFFICIENT

Ag-O-Cs RELATIVE ANODE CURRENT (%)

FOR CATHODE SENSITIVITY [% / -0.5 MULTIALKALI Sb-Cs 0 1 10 100 1000 10000 -1 200 300 400 500 600 700 800 900 1000 1100 1200 TIME (hour)

WAVELENGTH [nm] TPMHB0448EA

TPMOB0013EA TIME RESPONSE HYSTERESIS In the measurement of pulsed light, the anode output signal A photomultiplier tube may exhibit an unstable output for sev- should reproduce a waveform faithful to the incident pulse wave- form. This reproducibility is greatly affected by the electron tran- eral seconds to several tens of seconds after voltage and light sit time, anode pulse rise time, and electron transit time spread are applied, i.e., output may slightly overshoot or undershoot be- (TTS). fore reaching a stable level (Figure 15). This instability is called As illustrated in Figure 17, the electron transit time is the time hysteresis and may be a problem in and interval between the arrival of a delta function light pulse (pulse other applications. width less than 50ns) at the photocathode and the instant when Hysteresis is mainly caused by electrons being deviated from the anode output pulse reaches its peak amplitude. The anode their planned trajectories and electrostatically charging the dy- pulse rise time is defined as the time required to rise from 10% to node support and glass bulb. When the applied voltage 90% of the peak amplitude when the whole photocathode is illu- is changed as the light input changes, marked hysteresis can minated by a delta function light pulse (pulse width less than 50 occur. As a countermeasure, many Hamamatsu side-on photo- ps). The electron transit time has a fluctuation between indi- multiplier tubes employ "anti-hysteresis design" which virtually vidual light pulses. This fluctuation is called transit time spread (TTS) and defined as the FWHM of the frequency distribution of eliminate hysteresis. electron transit times (Figure 18) at single photoelectron event. The TTS is an important factor in time-resolved measurement. Figure 15: Hysteresis Measurement The time response characteristics depend on the dynode structure and applied voltage. In general, tubes of the linear-fo- cused or circular-cage structure exhibit better time response than tubes of the box-and-grid or venetian blind structure. MCP- PMTs, which employ an MCP in place of conventional dynodes, offer better time response than tubes using other dynodes. For I max. Ii example, the TTS can be significantly improved compared to I min. normal photomultiplier tubes because a nearly parallel electric ANODE CURRENT field is applied between the photocathode, MCP and the anode. 0567 Figure 19 shows typical time response characteristics vs. ap- TIME (MINUTE) plied voltage for types R2059 (51mm dia. head-on, 12-stage, lin- TPMOC0071EA ear-focused type) . - 8 - Figure 17: Anode Pulse Rise Time and Electron Transit Time VOLTAGE-DIVIDER CONSIDERATION Interstage voltages for the dynodes of a photomultiplier tube DELTA FUNCTION LIGHT are usually supplied by a voltage-divider circuits consisting of series-connected . Schematic diagrams of typical volt- RISE TIME FALL TIME age-divider circuits are illustrated in Figure 20. Circuit (a) is a basic arrangement (DC output) and (b) is for pulse operations. 10% ANODE Figure 21 shows the relationship between the incident light level OUTPUT 90% SIGNAL TRANSIT TIME and the average anode output current of a photomultiplier tube using the voltage-divider circuit (a). Deviation from the ideal lin- earity occurs at a certain incident level (region B). This is caused TPMOB0060EA by an increase in dynode voltage due to the redistribution of the voltage loss between the last few stages, resulting in an appar- Figure 18: Electron Transit Time Spread (TTS) ent increase in sensitivity. As the input light level is increased, the anode output current begins to saturate near the value of the current flowing through the (region C). Therefore, TYPE NO. : R2059 *FWHM=550ps it is recommended that the voltage-divider current be maintained 104 *FWTM=1228ps at least at 20 times the average anode output current required

103 from the photomultiplier tube.

102 Figure 20: Schematic Diagrams of Voltage-Divider Circuits (a) Basic arrangement for DC operation 1 10 PHOTOCATHODE ANODE RELATIVE COUNT

100

1R 1R 1R 1R 1R 1R 1R 1R 1R 1R 1R -5 -4 -3 -2 -1 0 1 2 3 4 5 -HV TIME [ns] (b) For pulse operation PHOTOCATHODE TPMHB0126EB ANODE

Figure 19: Time Response Characteristics vs. Supply 1R 1R 1R 1R 1R 1R 1R 1R 1R 1R 1R Voltage -HV C1 C2 C3

TYPE NO. : R2059 TACCC0030EB 10 2

TRANSIT TIME

Figure 21: Output Characteristics of a PMT Using Voltage- Divider Circuit (a) 10 1 10

C RISE TIME TIME (ns) 1.0

B 10 0 ACTUAL CURVE

T. T. S 0.1 IDEAL CURVE TO DIVIDER CURRENT 0.01 A 500 1000 1500 20002500 3000

SUPPLY VOLTAGE (V) RATIO OF AVERAGE OUTPUT CURRENT

TPMOB0059EB 0.001 0.001 0.01 0.1 1.0 10

TACCB0005EA

LIGHT FLUX (A.U.)

- 9 - Generally high output current is required in pulsed light appli- the housing for a photomultiplier tube and when using an electro- cations. In order to maintain dynode potentials at a constant static or magnetic shield case, extreme care is required. value during pulse durations and obtain high peak currents, large In addition, when using foam rubber or similar material to are used as shown in Figure 20 (b). The val- mount the tube in its housing, it is essential that material having ues depend on the output charge. If linearity of better than 1% is sufficiently good insulation properties be used. This problem can needed, the capacitor value should be at least 100 times the out- be solved by applying a black conductive layer around the bulb put charge per pulse, as follows: and connecting to the cathode potential (called HA Coating), as shown in Figure 23. I t C > 100 (farads) As mentioned above, the HA coating can be effectively used V to eliminate the effects of external potential on the side of the where I is the peak output current in amperes, it is the pulse width bulb. However, if a grounded object is located on the photocath- in seconds, and V is the voltage across the capacitor in . ode faceplate, there are no effective countermeasures. Glass In high energy physics applications where a high pulse output is scintillation, if it occurrs in the faceplate, has a larger influence on required, as the incident light is increased while the interstage volt- the noise. It also causes deterioration of the photocathode sensi- age is kept fixed, output saturation will occur at a certain level. This tivity and, once deteriorated, the sensitivity will never recover to is caused by an increase in the electron density between the the original level. To solve these problems, it is recommended electrodes, causing effects which disturb the elec- that the photomultiplier tube be operated in the cathode ground tron current. As a corrective action to overcome space charge scheme, as shown in Figure 24, with the anode at a positive high effects, the voltage applied to the last few stages, where the elec- voltage. For example, in scintillation counting, since the tron density becomes high, should be set at a higher value than grounded scintillator is directly coupled to the photomultiplier the standard voltage distribution so that the voltage gradient be- tube, it is recommended that the cathode be grounded, with a tween those electrodes is enhanced. For this purpose, a so- high positive voltage applied to the anode. Using this scheme, a called tapered bleeder circuit (Figure 22) is often employed. Use coupling capacitor Cc is used to separate the high positive volt- of this tapered bleeder circuit improves pulse linearity 5 to 10 age applied to the anode from the signal, making it impossible to times better than that obtained with normal bleeder circuits obtain a DC signal output. (equally divided circuits). Hamamatsu provides a variety of socket assemblies incorpo- Figure 23: HA Coating rating voltage-divider circuits. They are compact, rugged, light- weight and ensure the maximum performance for a photomulti- GLASS BULB plier tube by simple wiring. CONDUCTIVE PAINT (SAME POTENTIAL AS CATHODE) Figure 22: Tapered Bleeder Circuit INSULATING PHOTOCATHODE ANODE PROTECTIVE COVER SIGNAL OUTPUT RL CONNECTED TO CATHODE PIN 1R 1R 1R 1R 2R 3R 2.5R

C1 C2 C3 TPMOC0015EA

-HV TACCC0035EB Figure 24: Cathode Ground Scheme

GROUND POLARITY AND HA COATING PHOTOCATHODE ANODE Cc SIGNAL The general technique used for voltage-divider circuits is to OUTPUT ground the anode with a high negative voltage applied to the

1 R2 R3 R4 R5 R6 R7 cathode, as shown in Figure 20. This scheme facilitates the con- R TACCC0036EB nection of such circuits as ammeters or current-to-voltage con- + HV version operational to the photomultiplier tube. How- ever, when a grounded anode configuration is used, bringing a grounded metallic holder or magnetic shield case near the bulb of the tube can cause electrons to strike the inner bulb wall, re- sulting in the generation of noise. Also, for head-on type photo- multiplier tubes, if the faceplate or bulb near the photocathode is grounded, the slight conductivity of the glass material causes a current to flow between the photocathode (which has a high negative potential) and ground. This may cause significant dete- rioration of the photocathode. For this reason, when designing - 10 - SINGLE PHOTON COUNTING A typical pulse height distribution (PHD) for the output of pho- Photon counting is one effective way to use a photomultiplier tomultiplier tubes is shown in Figure 28. In this PHD, the lower tube for measuring very low light levels. It is widely used in astro- level discrimination (LLD) is set at the valley trough and the up- nomical photometry and chemiluminescence or biolumines- per level discrimination (ULD) at the foot where the output pulses cence measurement. In the usual application, a number of pho- are very few. Most pulses smaller than the LLD are noise and tons enter the photomultiplier tube and create an output pulse pulses larger than the ULD result from cosmic rays, etc. There- train like (a) in Figure 25. The actual output obtained by the mea- fore, by counting pulses between the LLD and ULD, accurate surement circuit is a DC current with a fluctuation as shown at light measurements becomes possible. In the PHD, Hm is the (b). mean height of the pulses. It is recommended that the LLD be set at 1/3 of Hm and the ULD at triple Hm. In most cases, however, Figure 25: Overlapping Output Pulses the ULD setting can be omitted. (a) Considering the above, a clear definition of the peak and val- ley in the PHD is a very significant characteristic for photomulti- plier tubes for use in photon counting.

TIME Figure 28: Typical Pulse Height Distribution (b)

SIGNAL PULSE + NOISE PULSE TIME NOISE PULSE TPMOC0073EA COUNTS When the light intensity becomes so low that the incident photons are separated as shown in Figure 26. This condition is called a single photon (or photoelectron) event. The number of LLD Hm ULD output pulses is in direct proportion to the amount of incident light PULSE HEIGHT and this pulse counting method has advantages in S/N ratio and TPMOC0076EA stability over the DC method averaging all the pulses. This pulse counting technique is known as the photon counting method. SCINTILLATION COUNTING Figure 26: Discrete Output Pulses (Single Photon Event) Scintillation counting is one of the most sensitive and effec- tive methods for detecting radiation. It uses a photomultiplier tube coupled to a transparent crystal called scintillator which pro- duces light by incidence of radiation.

TIME Figure 29: Diagram of Scintillation TPMOC0074EA

REFLECTIVE PHOTOCATHODE COATING Since the photomultiplier tube output contains a variety of PHOTOELECTRONS noise pulses in addition to the signal pulses representing photo- ANODE electrons as shown in Figure 27, simply counting the pulses with- DYNODES out some form of noise elimination will not result in an accurate RADIATION measurement. The most effective approach to noise elimination SOURCE is to investigate the height of the output pulses. SCINTILLATOR PMT OPTICAL COUPLING Figure 27: Output Pulse and Discrimination Level (USING SILICONE OIL etc.)

TPMHC0052EB

DARK CURRENT COSMIC RAY PULSE PULSE In radiation measurements, there are two parameters that ULD should be measured. One is the energy of individual PULSE HEIGHT SIGNAL PULSE and the other is the amount of particles. Radiation measure- ments should determine these two parameters.

LLD When radiation enters the scintillator, it produce light flashes in response to each . The amount of flash is proportional

TIME to the energy of the incident racliation. The photomultiplier tube

TPMOC0075EB detects individual light flashes and provides the output pulses - 11 - which contain information on both the energy and amount of c) 60Co+NaI (Tl) pulses, as shown in Figure 30. By analyzing these output pulses × using a multichannel analyzer (MCA), a pulse height distribution 10000 ( 2" dia. 2" t ) (PHD) or energy spectrum is obtained, and the amount of inci- dent particles at various energy levels can be measured accu- 5000 rately. Figure 31 shows typical PHDs or energy spectra when COUNTS gamma rays (55Fe, 137Cs, 60Co) are detected by the combination of an NaI(Tl) scintillator and a photomultiplier tube. For the PHD, 0500 1000 it is very important to have distinct peaks at each energy level. CHANNEL NUMBER This is evaluated as pulse height resolution (energy resolution) and is the most significant characteristic in radiation particle TPMOB0087EA measurements. Figure 32 shows the definition of energy resolu- tion taken with a 137Cs source. Figure 32: Definition of Energy Resolution

Figure 30: Incident Particles and PMT Output

TIME b

a H SCINTILLATOR H

NUMBER OF PULSES 2 THE HEIGHT OF OUTPUT PULSE IS PROPORTIONAL PMT TO THE ENERGY OF PULSE HEIGHT INCIDENT PARTICLE.

TPMOB0088EA CURRENT TIME Figure 33: Spectral Response of PMT and Spectral - sion of Scintillators

102 BGO

CsI (Tl) Figure 31: Typical Pulse Height Distributions (Energy Spectra) BaF2

55 NaI (Tl) a) Fe+NaI (Tl) 101

1000 ( 2" dia. × 2" t )

BIALKALI

100 500 QUANTUM EFFICIENCY (%) COUNTS OF VARIOUS SCINTILLATOR (%) RELATIVE EMISSION DISTRIBUTION

10-1 0500 1000 200 300 400 500 600 700 800 CHANNEL NUMBER WAVELENGTH (nm) b) 137Cs+NaI (Tl) TPMOB0073EA

10000 ( 2" dia. × 2" t ) Pulse height resolution is mainly determined by the quantum efficiency of the photomultiplier tube in response to the scintilla- 5000 tor emission. It is necessary to choose a tube whose spectral COUNTS response matches with the scintillator emission. In the case of thallium-activated iodide, or NaI(Tl), which is the most

0500 1000 popular scintillator, head-on type photomultiplier tube with a CHANNEL NUMBER bialkali photocathode is widely used. - 12 - Connections to External Circuits

LOAD RESISTANCE alone. From this we see that the upper limit of the load resis- Since the output of a photomultiplier tube is a current signal tance is actually the input resistance of the and that and the type of external circuit to which photomultiplier tubes are making the load resistance much greater than this value does usually connected has voltage inputs, a load resistance is used not have significant effect. While the above description assumed to perform a current-voltage transformation. This section de- the load and input impedances to be purely resistive, in practice, scribes considerations to be made when selecting this load resis- stray , input and stray in- tance. Since for low output current levels, the photomultiplier fluence phase relationships. Therefore, as frequency is in- tube may be assumed to act as virtually an ideal constant-current creased, these circuit elements must be considered as com- source, the load resistance can be made arbitrarily large, thus pound impedances rather than pure resistances. converting a low-level current output to a high-level voltage out- From the above, three guides can be derived for use in se- put. In practice, however, using very large values of load resis- lection of the load resistance: tance creates the problems of deterioration of frequency re- 1) In cases in which frequency response is important, the sponse and output linearity described below. load resistance should be made as small as possible. 2) In cases in which output linearity is important, the load Figure 34: PMT Output Circuit resistance should be chosen such that the output voltage

PHOTOCATHODE ANODE is below several volts. SIGNAL OUTPUT 3) The load resistance should be less than the approximate Ip RL CS of the external amplifier.

-HV HIGH-SPEED OUTPUT CIRCUIT TACCC0037EB For the detection of high-speed and pulsed light signals, a If, in the circuit of Figure 34, we let the load resistance be RL coaxial cable is used to make the connection between the photo- and the total of the capacitance of the photomultiplier tube anode multiplier tube and the electronic circuit, as shown in Figure 36. to all other electrodes, including such stray capacitance as wiring Since commonly used cables have characteristic impedances of capacitances be Cs, the cutoff frequency fc is expressed by the 50 Ω or 75 Ω, this cable must be terminated in a pure resistance following relationship. equivalent to the characteristic impedance to provide and ensure -free transmission for the signal 1 fc = waveform. If a matched transmission line is used, the imped- 2πCs RL ance of the cable as seen by the photomultiplier tube output will From this relationship, it can be seen that, even if the photo- be the characteristic impedance of the cable, regardless of the multiplier tube and amplifier have very fast response, response cable length, and no distortion will occur in signal waveforms. will be limited to the cutoff frequency fc of the output circuit. If the If proper matching at the signal receiving end is not achieved, load resistance is made large, at high current levels the voltage the impedance seen at the photomultiplier tube output will be a drop across RL becomes large, affecting a potential difference function of both frequency and cable length, resulting in signifi- between the last dynode stage and the anode. As a result, a loss cant waveform distortion. Such mismatched conditions can be of output linearity (output current linearity with respect to incident caused by the connectors used as well, so that the connector to light level) may occur. be used should be chosen with regard given to the frequency range to be used, to provide a match to the coaxial cable. Figure 35: Amplifier Internal Resistance When a mismatch at the signal receiving end occurs, all of the pulse energy from the photomultiplier tube is not dissipated (1) (2) at the receiving end, but is partially reflected back to the photo- PMT P PMT P CC multiplier tube via the cable. While this reflected energy will be DYn SIGNAL DYn SIGNAL Rin Rin OUTPUT OUTPUT fully dissipated at the photomultiplier tube when an impedance RL CS RL CS match has been achieved at the tube, if this is not the case, be- cause the photomultiplier tube itself acts as an open circuit, the TACCC0017EA energy will be reflected and, thus returned to the signal-receiving end. Since part of the pulse makes a round trip in the coaxial In Figure 35, let us consider the effect of the internal resis- cable and is again input to the receiving end, this reflected signal tance of the amplifier. If the load resistance is RL and the input is delayed with respect to the main pulse and results in wave- impedance of the amplifier is Rin, the combined parallel output form distortion (so called ringing phenomenon). To prevent this resistance of the photomultiplier tube, Ro, is given by the follow- phenomenon, in addition to providing impedance matching at ing equation. the receiving end, it is necessary to provide a resistance matched to the cable impedance at the photomultiplier tube end RL Rin Ro = RL + Rin as well. If this is done, it is possible to virtually eliminate the ring- ing caused by an impedance mismatch, although the output This value of Ro, which is less than the value of RL, is then the pulse height of the photomultiplier tube is reduced to one-half of effective load resistance of the photomultiplier tube. If, for ex- the normal level by use of this impedance matching . ample, RL=Rin, the effective load resistance is 1/2 that of RL - 13 - Figure 36: Typical Connections Used to Prevent Ringing This relationship is derived for the following reason. If the input impedance of the is extremely large,

50Ω OR 75Ω COAXIAL CABLE and the output current of the photomultiplier tube is allowed to flow into the input terminal of the amplifier, most of the current PMT RL (50 OR 75Ω will flow through Rf and subsequently to the operational amplifier Ω Ω MATCHING RESISTOR) 50 OR 75 CONNECTOR output circuit. Therefore, the output voltage Vo is given by the HOUSING ANTI-REFLECTION RESISTOR expression -Rf × Ip. When using such an operational amplifier, it TACCC0039EA is of course, not possible to increase the output voltage without Next, let us consider waveform observation of high-speed limit, the actual maximum output being approximately equal to pulses using an (Figure 37). This type of operation the operational amplifier power supply voltage. At the other end requires a low load resistance. Since, however, there is a limit to of the scale, for extremely small currents, limitations are placed the oscilloscope sensitivity, an amplifier may be required. by the operational amplifier offset current (Ios), the quality of Rf, For cables to which a matching resistor has been connected, and other factors such as the insulation materials used. there is an advantage that the cable length does not affect the Figure 38: Current-Voltage Transformation Using characteristics of the cable. However, since the matching resis- an Operational Amplifier tance is very low compared to the usual load resistance, the out- Rf put voltage becomes too small. While this situation can be rem- lp lp p Vo= -lp Rf edied with an amplifier of high gain, the inherent noise of such an - amplifier can itself be detrimental to measurement performance. PMT + In such cases, the photomultiplier tube can be brought as close OP-AMP V as possible to the amplifier and a load resistance as large as pos- TACCC0041EA sible should be used (consistent with preservation of frequency If the operational amplifier has an offset current (Ios), the response), to achieve the desired input voltage. above-described output voltage becomes Vo=-Rf(Ip+Ios), the off- set current component being superimposed on the output. Fur- Figure 37: With Ringing Suppression Measures thermore, the magnitude of temperature drift may create a prob-

PMT lem. In general, a metallic film resistor which has a low tempera- P DYn ture coefficient is used for the resistance Rf, and for high resis- RL tance values, a vacuum-sealed type is used. Carbon resistors, OSCILLOSCOPE with their highly temperature-dependent resistance characteris- tics, are not suitable for this application. When measuring such WIRING SHOULD BE extremely low level currents as 100 pA and below, in addition to AS SHORT AS POSSIBLE. TACCC0026EA the considerations described above, the materials used in the It is relatively simple to implement a high-speed amplifier us- circuit implementation require care as well. For example, materi- als such as bakelite are not suitable, and more suitable materials ing a wide-band video amplifier or operational amplifier. How- are Teflon, polystyrol or steatite. In addition, low-noise cables ever, in exchange of design convenience, use of these ICs tends should be used, since general-purpose coaxial cables exhibit to create problems related to performance (such as noise). It is noise due to mechanical changes. In the measurement of these therefore necessary to know their performance limit and take cor- low level currents, use of an FET input operational amplifier is rective action. recommended. As the pulse repetition frequency increases, baseline shift creates one reason for concern. This occurs because the DC sig- Figure 39: Frequency Compensation of an Operational nal component has been eliminated from the signal circuit by Amplifier coupling with a capacitor which does not pass DC components. If Cf this occurs, the reference zero level observed at the last dynode Cs stage is not the actual zero level. Instead, the apparent zero level SHIELD CIRCUIT is the time-average of the positive and negative fluctuations of the signal waveform. This will vary as a function of the pulse den- Rf sity, and is known as baseline shift. Since the height of the pulses above this baseline level is influenced by the repetition fre- - SIGNAL OUTPUT quency, this phenomenon is of concern when observing wave- + forms or discriminating pulse levels. OP-AMP. TACCC0042EA

OPERATIONAL AMPLIFIERS In Figure 39, if a capacitance Cf (including any stray capaci- tance) exists in parallel to the resistance Rf, the circuit exhibits a In cases in which a high-sensitivity ammeter is not available, time constant of (Rf × Cf), so that response speed is limited to the use of an operational amplifier will enable measurements to this time constant. This is a particular problem if Rf is large. Stray be made using an inexpensive voltmeter. This technique relies capacitance can be reduced by passing Rf through a hole in a on converting the output current of the photomultiplier tube to a shield plate. When using coaxial signal input cables, since the voltage signal. The basic circuit is as shown in Figure 38, for cable capacitance Cc and Rf are in the feedback loop, oscilla- which the output voltage, Vo, is given by the following relation- tions may occur and noise may be amplified. While the method ship. of avoiding this is to connect Cf in parallel to Rf, to reduce gain at high frequencies, as described above, this creates a time con- Vo = -Rf -Ip stant of Rf × Cf which limits the response speed. - 14 - HAMAMATSU PHOTONICS K.K., Electron Tube Center 314-5, Shimokanzo, Toyooka-village, Iwata-gun, Shizuoka-ken, 438-0193, Japan : (81)539-62-5248, Fax: (81)539-62-2205 http://www.hamamatsu.com/

Main Products Sales Offices Opto- ASIA: North Europe: HAMAMATSU PHOTONICS K.K. HAMAMATSU PHOTONICS NORDEN AB Photo IC 325-6, Sunayama-cho, Färögatan 7, Hamamatsu City, 430-8587, Japan S-164 40 Kista, Sweden Position Sensitive Detectors Telephone: (81)53-452-2141, Fax: (81)53-456-7889 Telephone: (46)8-703-29-50, Fax: (46)8-750-58-95 Image Sensors Infrared Detectors U.S.A.: Italy: Solid State Emitters HAMAMATSU CORPORATION HAMAMATSU PHOTONICS ITALIA S.R.L. CdS Photoconductive Cells Main Office Via della Moia, 1/E Pyroelectric Detectors 360 Foothill Road, P.O. BOX 6910, 20020 Arese, (Milano), Italy Photocouplers Bridgewater, N.J. 08807-0910, U.S.A. Telephone: (39)2-935 81 733, Fax: (39)2-935 81 741 Telephone: (1)908-231-0960, Fax: (1)908-231-1218 Photointerrupters, Photoreflectors E-mail: [email protected] Hong Kong: S&T ENTERPRISES LTD. Electron Tubes Western U.S.A. Office Room 404, Block B, Photomultiplier Tubes Suite 110, 2875 Moorpark Avenue Seaview Estate, Watson Road, Light Sources San Jose, CA 95128, U.S.A. North Point, Hong Kong Image Pick-up Tubes Telephone: (1)408-261-2022, Fax: (1)408-261-2522 Telephone: (852)25784921, Fax: (852)28073126 Image Intensifiers United Kingdom: Taiwan: X-Ray Image Intensifiers HAMAMATSU PHOTONICS UK LIMITED S&T HITECH LTD. Microchannel Plates Lough Point, 2 Gladbeck Way, Windmill Hill, 3F-6 No.188, Section 5, Naking East Road Fiber Optic Plates Enfield, Middlesex EN2 7JA, United Kingdom Taipei, Taiwan, R.O.C. Telephone: (44)181-367-3560, Fax: (44)181-367-6384 Telephone: (886)2-2753-0188, Fax: (886)2-2746-5282 Imaging and Processing Systems Video Cameras for Measurement France, Portugal, Belgium, Switzerland, Spain: KORYO ELECTRONICS CO., LTD. HAMAMATSU PHOTONICS FRANCE S.A.R.L. 9F No.3-61, Chung Hsiao Image Processing Systems 8, Rue du Saule Trapu, Parc du Moulin de Massy, E.Road, Shitzu, Taipei Hsien, Taiwan, R.O.C. Streak Cameras 91882 Massy Cedex, France Telephone: (886)2-698-1143, Fax: (886)2-698-1147 Optical Telephone: 33(1) 69 53 71 00, Fax: 33(1) 69 53 71 10 Optical Measurement Systems Republic of Korea: Imaging and Analysis Systems Swiss Office SANGKI TRADING CO., LTD. Richtersmattweg 6a Suite 431, Sunmyunghoi Bldg., CH-3054 Schu¨ pfen, Switzerland 24-2, Youid-Dong, Youngdeungpo-ku, Telephone: (41)31/879 70 70, Fax: (41)31/879 18 74 Seoul, Republic of Korea Telephone: (82)2-780-8515, Fax: (82)2-784-6062 Belgiam Office 7, Rue du Bosquet Singapore: B-1348 Louvain-La-Neuve, Belgium S&T ENTERPRISES (SINGAPORE) PTE. LTD. Telephone: (32)10 45 63 34, Fax: (32)10 45 63 67 Block 2, kaki Bukit Avenue 1, #04-01 to #04-04 Kaki Bukit Industrial Estate, Singapore 417938 Spanish Office Telephone: (65)7458910,Fax: (65)7418201 Cetro de Empresas de Nuevas Tecnologias Parque Tecnologico del Valles 08290 CERDANYOLA (Barcelona) Spain Telephone: (34)3 582 44 30, Fax: (34)3 582 44 31

Germany, Denmark, Netherland: HAMAMATSU PHOTONICS DEUTSCHLAND GmbH Arzbergerstr. 10, Information in this catalog is believed D-82211 Herrsching am Ammersee, Germany to be reliable. However, no Telephone: (49)8152-375-0, Fax: (49)8152-2658 responsibility is assumed for possible inaccuracies or omission. Danish Office Specifications are subject to change Naverland 2 without notice. No patent rights are DK-2600 Glostrup,Denmark granted to any of the circuits Telephone: (45)4346-6333, Fax: (45)4346-6350 described herein. ©1998 Hamamatsu Photonics K.K. Netherland Office Postbus 536, 3760 Am Soest, Netherland Telephone: (31)35-6029191, Fax: (31)35-6029304 JAN.1998

Quality, , and service are part of every product.

234 CHAPTER 9 APPLICATIONS

9. 1 Spectrophotometry

9. 1. 1 Overview

Spectrophotometry is a study of the transmission and reflection properties of material samples as a func- tion of wavelength, but the term commonly means chemical analysis of various substances utilizing photom- etry. Photometric instruments used in this field are broadly divided into two methods: one utilizes light ab- sorption, reflection or polarization at specific wavelengths and the other uses external energy to excite a sample and measures the subsequent light emission. Photomultiplier tubes have been most widely used in this field for years. Major principles used in spectrophotometry are classified as illustrated in Figure 9-1 below.

I0 I 0 ' I0 λ I 0 ' λ1 I0

λ1 . I0' λ2 . I0' λ . ) (a) Absorption or Transmission (b) Reflection ( 1 I0 ' and Fluorescence (λ2 . I0'' )

I0 I λ1 . I0 λ1 λ1

GAS GAS

(c) Atomic Absorption (d) Flame

λ1 . I0 No λ2 . I0' I0 λ1 λ . 3 I0'' O 3

(e) Emission (Direct Reader) (f) Chemiluminescence

TPMOC0017EB

Figure 9-1: Major principles of spectrophotometry Specific photometric instruments currently used are: 1)Visible to UV spectrophotometers (absorption, reflection) 2)Infrared spectrophotometers (absorption, reflection) 3)Far UV spectrophotometers (absorption, reflection) 4)Emission spectrophotometers 5)Fluorescence spectrophotometers 6)Atomic absorption spectrophotometers 7)Azimuthal, circular dichroism meters 8)Raman spectrophotometers 9)Densitometers, colorimeters and color analyzers etc. 9. 1 Spectrophotometry 235

9. 1 .2 Specific applications

The following paragraphs explain major, specific applications of spectrophotometers, divided into two methods utilizing absorption or emission.

(1) Utilizing Absorption A. UV, visible and infrared spectrophotometers When light passes through a substance, the light energy causes changes in the electronic state of the substance (electron transition) or induces characteristic vibration of the molecules, resulting in a loss of partial energy. This is referred to as absorption, and quantitative analysis can be performed by measuring the extent of absorption. The principle and simplified block diagram1) of an absorption spectrophotometer are shown in Figure 9-2.

CELL MONOCHRO- DETECTOR LIGHT MATOR I I INTENSITY 0 I0 I C C

LAMP SAMPLE DISPLAY & d RECORDER C: Concentration TPMOC0018EB

Figure 9-2: Principle and block diagram of a absorption spectrophotometer

There are various optical systems in use today for spectrophotometers. Figure 9-3 illustrates the optical system2) of a spectrophotometer using sequential plasma emission as the light source for cov- ering from the ultraviolet to visible and infrared range.

MIRROR

GRATING

TWO EXIT FILTER PMT SLITS WHEEL PLASMA TORCH

MIRROR ENTRANCE SLIT Hg LAMP

MIRROR

TPMOC0019EA

Figure 9-3: Optical system of a UV to visible spectrophotometer 236 CHAPTER 9 APPLICATIONS

B. Atomic absorption spectrophotometers The atomic absorption spectrophotometer employs special light sources (hollow cathode lamps) constructed for the respective elements to be analyzed. A sample is dissolved in solvent and burned for atomization, and light from the specific hollow cathode lamp is passed through the flame. The amount of light that is absorbed is proportional to the concentration of the sample material. Therefore, by comparing the extent of absorption between the sample to be analyzed and a standard sample measured in advance, it is possible to know the concentration of the specific element contained in the sample. A typical optical system3) used for atomic absorption spectrophotometers is shown in Figure 9-4.

SLIT CONCAVE MIRROR PMT GRATING

PLANE CONCAVE MIRROR MIRROR SLIT BEAM SPLITTER BURNER

PLANE MIRROR LENS LENS HOLLOW CATHODE LAMPS D2 LAMP

TPMOC0021EA

Figure 9-4: Optical system used for atomic absorption spectrophotometers

(2) Utilizing Emission A. Photoelectric emission spectrophotometers (direct readers) When external energy is applied to a sample, light emission occurs from the sample. Dispersing this emission using a monochromator, into characteristic spectral lines of elements and measuring their presence and intensity simultaneously, enables rapid qualitative and quantitative analysis of the elements contained in the sample. Figure 9-5 illustrates the block diagram4) of a photoelectric emis- sion spectrophotometer in which multiple photomultiplier tubes are used.

PMT

INTEGRATOR EXIT SLITS

GRATING

DATA ENTRANCE SLIT

CRT PRINTER

ICP SPARC ARC

EXCITATION SOURCE

TPMOC0022EA

Figure 9-5: Block diagram illustrating a photoelectric emission spectrophotometer 9. 1 Spectrophotometry 237

B. Fluorospectrophotometers The fluorospectrophotometer is mainly used for chemical analysis in biochemistry, especially in molecular biology. When a substance is illuminated and excited by visible or ultraviolet light, it may emit light with a wavelength longer than that of the excitation light. This light emission is known as fluorescence and its emission process5) is shown in Figure 9-6. Measuring the fluorescent intensity and spectra allows quantitative and qualitative analysis of the elements contained in the substance.

EXCITED 3 2 STATE 1 V=0 EXCITED TRIPET STATE

NORMAL 3 2 STATE 1 V=0 FLUORES- PHOSPHORES- ABSORPTION CENCE CENCE

TPMOC0023EA

Figure 9-6: Fluorescent molecular energy levels Figure 9-7 shows the structure6) of a fluorospectrophotometer using photomultiplier tubes as the detectors. This instrument roughly consists of a light source, excitation monochromator, fluorescence monochromator and fluorescence detector. A lamp is commonly used as the light source be- cause it provides a continuous spectrum output over a wide spectral range. The excitation and fluores- cence monochromators use the same diffraction grating or prism, as are used in general-purpose monochromators.

FLUORESCENCE EXCITATION MONOCHROMATOR MONOCHROMATOR

REF. PMT GRATING GRATING DIFFU- BEAM SER SPRITTER CHOPPER SAMPLE FILTER PMT

SAMPLE XENON CELL LAMP

TPMOC0024EA

Figure 9-7: Fluorospectrophotometer structure 238 CHAPTER 9 APPLICATIONS

9. 1. 3 Characteristics required of photomultiplier tubes

The following photomultiplier tube characteristics are required in this application. a) High stability b)Low dark current c) High signal-to-noise ratio d)Wide spectral response (ultraviolet to infrared) e) Low hysteresis f) Excellent polarization properties Side-on and head-on photomultiplier tubes having a multialkali photocathode and silica window are most frequently used in these applications. 9. 4 In-vitro Assay 247

9. 4 In-vitro Assay

The analysis and inspection of blood or urine samples collected out of a living body is referred to as in-vitro assay. It is used for physical checkups, diagnosis, and evaluation of drug potency. The in-vitro assay can be classified as shown in Table 9-3. Of these, the concentrations of most tumor markers, hormones, drugs and viruses which are classified under immunological assay are exceedingly low. This requires extremely high- sensitivity inspection equipment for frequently requiring use of photomultiplier tubes.

Sample Inspection Biochemistry Enzyme, protein, sugar, lipid

Immunology Tumor marker, serum protein, hormone, reagent, virus Hematology (Leukocyte, red corpuscle, hemoglobin, platelet) computation, classification, coagulation Microbiology Bacteria identification, susceptibility

Table 9-3: Classification of in-vitro inspection Immunoassay, a measurement technique making use of specificity of the antigen-antibody reaction, is widely used. The principles of immunoassay9) are illustrated in Figure 9-19 below, and the procedures of each method are explained in the subsequent paragraphs. 248 CHAPTER 9 APPLICATIONS

(a) Samdwitch Method

(1) Solid-phase antibody+sample (2) Removing liquid layer after (3) Adding labeled antibody antigen-antibody reaction LABEL OBJECT ANTIGEN - HORMONE LIQUID LAYER ISOTOPE TUMOR ENZYME MARKER ANTIGEN- REAGENT ANTIBODY VIRUS, ETC. REACTION

ANTIBODY FIXED IN VESSEL

(4) Removing upper layer after (5) Measuring number of labels antigen-antibody reaction

LIQUID MEASURING NUMBER OF LABELS LAYER Proportional to quantity of object SAND- antigen WITCHING

(b) Competitive Method (1) Solid-phase antibody (2) Introducing sample and (3) Removing liquid layer after labeled antigen antigen-antibody reaction

LABELED ANTIGEN OBJECT ANTIGEN LIQUID LAYER

COMPETITIVE BOND ANTIBODY FIXED IN VESSEL Sample antigen and labeled antigen combine respectively with antibody according to their (4) Measuring number of labels quantities

MEASURING NUMBER OF LABELS Inversely proportional to quantity of object antigen

TPMHC0023EA

Figure 9-19: Principles of immunoassay Figure 9-19 (a) is a technique known as the sandwich method. Step (1): Samples are introduced into a vessel in which antibodies responding to object antigens (hormones, tumor markers, etc.) are fixed (solid- phase antibody). Step (2): Antigen-antibody reaction occurs, and each object antigen combines with a solid- phase antibody. This reaction has an extremely high singularity and hardly ever occurs with a different anti- gen. After antigen-antibody reaction, the liquid layer is removed leaving the combined antigen and antibody. Step (3): Labeled antibodies are added, which combine with object antigens. Step (4): Antigen-antibody reaction occurs again so that the object antigen is sandwiched between the antibodies. Then the liquid layer is removed. Step (5): The quantity of labels is optically measured using a photomultiplier tube. 9. 4 In-vitro Assay 249

Figure 9-19 (b) is another technique called the competitive method. Step (1): Antibodies responding to object antigens are fixed on the bottom of a vessel. Step (2): Samples are added along with the labeled object antigens. Step (3): Competitive reaction in which object antigens and labeled antigens combine with labeled antibodies in proportional to their concentration, reaching a state of equilibrium. After the antigen-antibody reaction, the unnecessary upper layer is removed. Step (4): The quantity of labels is measured using a photo- multiplier tube. In the sandwich method, the higher the concentration of object antigens, the larger the signal. Conversely, in the competitive method, the higher the concentration of the object antigens, the lower the signal. Immunoassay can be further categorized according to the material used for labeling as follows (1) Using radioactive isotopes for labeling (Radioimmunoassay) (2) Using enzymes for labeling (Enzymeimmunoassay)

9. 4. 1 RIA (Radioimmunoassay) method

(1) Overview Radioactive isotope (RI) is used for the labeling as was explained above, and radiation (gamma or beta rays) emitting from the RI labels is detected by the combination of a scintillator and a photomultiplier tube, so that the object antigen can be quantitatively measured. Radioactive isotopes most frequently used for labeling are 3H, 14C, 57Co, 75Se, 125I and 131I. (See Table 9-4.)10) Of these, 125I offers useful characteristics for labeling and is very widely used. Because radioactive isotopes other than 3H and 14C emit gamma rays, sodium iodide crystals are used as a scintillator, providing a high conversion efficiency.

Radioisotope Half-life Energy Detection Method 3H 12.26 years β Liquid scintillation 14C 5730 years β Liquid scintillation 57Co 270 days γ Scintillation crystal 75Se 120.4 days γ Scintillation crystal 125I 60 days γ Scintillation crystal 137I 8 days β, γ Scintillation crystal

Table 9-4: Radioactive isotopes used for labeling in radioimmunoassay Recently, for in-vitro assays, the quantity of samples and the number of items to be measured are rapidly increasing. To meet this trend, the equipment for radioimmunoassay has been automated. A typical piece of automated equipment in use today is the well scintillation counter11) that makes use of sodium iodide scintil- lators having a well-like hole to enhance the conversion efficiency of the radiation into light. Measurements are made by automatically inserting test tubes, which contain a mixture of antigens and antibodies including labels, into each hole in the scintillator. (See Figure 9-20.) Each detector section including a scintillator is covered by lead shield to block extraneous radiation. Isotopes 3H and 14C can be used for labeling; however, because they emit extremely low-energy beta rays, a liquid explained in Section 9.5, is used to make measurements with these two isotopes. 250 CHAPTER 9 APPLICATIONS

(ENLARGED VIEW)

LEAD SHIELD SCINTILLATOR TEST TUBE

PMT SAMPLE CONT- PHOTO- ROLLER CATHODE PMT DATA COM- PRO- PUTER CESSOR

HIGH VOLTAGE HV PULSE ELECTRICAL AMP COUNT- POWER SUPPLY POWER HEIGHT ER SIGNAL (PULSE) SUPPLY DISCRIM.

TPMHC0024EA

Figure 9-20: Schematic block diagram illustrating a well scintillation counter

(2) Major characteristics required of photomultiplier tubes Photomultiplier tubes used in RIA must have the following characteristics. a) High energy resolution or pulse height resolution (PHR) b) High level of stability c) Low noise To obtain high energy resolution, the photomultiplier tube should have high quantum efficiency at the peak emission wavelength (410 nanometers) of the sodium iodide (NaI(Tl)) scintillator. (See Figure 9-16.) In addition, because this application field deals with quite a few samples that emit extremely small amounts of radiation, it is also very important that the photomultiplier tube exhibits sufficiently low noise.

9. 4. 2 EIA (Enzymeimmunoassay) method

(1) Overview An enzyme is used as a label utilizing the antigen-antibody reaction. As Figure 9-2112) shows, RIA offers exceptionally high sensitivity among various immunoassay techniques. However, because it uses radioactive isotopes, various restrictions are imposed on its use. On the other hand, though its sensitivity is inferior to RIA, EIA is more popular because of its easy use. EIA sensitivity is gradually increasing due to improvements in reagents and detection method. 9. 4 In-vitro Assay 251

Complement-fixation rection Hemagglutination reaction

Immunodiffusion Immuno- electrophoresis Nephelometry

Fluoroimmunoassay Enzyme-immunoassay Emission-immunoassay

Radioimmunoassay

10 1 100 10 1 100 10 1 100 10 1 100 10 1 mmol µmol nmol pmol fmol

TPMHC0025EA

Figure 9-21: Comparison of various immunoassay techniques and measurable concentration range (Use this comparison data just for a general guide.)

In the EIA procedure, an enzyme used as a label in the antigen-antibody mixture in the last step in Figure 9-19 is activated to create a product. Then, color or fluorescence emitted from the product is de- tected by a photomultiplier tube. (See Figure 9-22.)13) The extent of color or fluorescent intensity is propor- tional to the quantity of enzyme (enzyme concentration).

ENZYME REACTION SUBSTRATE PRODUCT

COLORLESS EMMISION/FLUORESCENCE ENZYME

TPMHC0026EA

Figure 9-22: Label enzyme reaction

(2) Major characteristics required of photomultiplier tubes a) High sensitivity at the wavelength of color or fluorescence emitted from the product of the enzyme reaction b) Low dark current c) High signal-to-noise ratio d) Compact size 252 CHAPTER 9 APPLICATIONS

9. 4. 3 Other immunoassay methods

(1) Overview Besides EIA, several non radioactive immunoassay techniques not using radioisotopes are under re- search and development. One of these is fluoroimmunoassay in which a fluorescent substance is used for labeling. The final remaining mixture of antigens and antibodies is irradiated by an excitation light and the resulting fluores- cence is measured with regard to the intensity, wavelength shift and polarization. This technique offers slightly higher sensitivity than that of EIA. Figure 9-23 shows the schematic of a fluorescence-polarization photometer,14) which is used for fluoroimmunoassay.

I2

or PMT LIGHT I1 (OUTPUT) SOURCE

1 2 λ1 λ emit I , I CONDENSER FILTER or PRISM (For λ emit) FILTER or PRISM (For λ excite) PRIMARY SECONDARY POLARIZER POLARIZER (FIXED) (ROTATING) SAMPLE CELL

TPMHC0027EB

Figure 9-23: Schematic presentation of a fluorescence-polarization photometer To achieve high sensitivity equal to RIA by using non-radioactive immunoassay, intensive research and development of emission-immunoassay has been carried out. This immunoassay uses a chemiluminous substance or bioluminous substance for labeling and allows the final remaining mixture of antigens and antibodies to emit light, which is detected by a photomultiplier tube. There are three types of emission- immunoassay methods, as follows: 1) Use of a chemiluminous substance such as luminol and acridinium for labeling 2) Use of chemiluminescence or bioluminescence for activation of the label enzyme used in EIA 3) Use of a catalyst for the bioluminescence reaction Methods 2) and 3) can be thought of as kinds of EIA techniques. As shown in Figure 9-21 previously, emission-immunoassay is a high sensitivity equivalent to RIA. 9. 4 In-vitro Assay 253

(2) Major characteristics required of photomultiplier tubes Fluoroimmunoassay a) High sensitivity at fluorescent wavelengths b) High level of stability c) Low dark current d) High signal-to-noise ratio e) Compact size

Emission-immunoassay a) High sensitivity at emission wavelengths b) Excellent single-photoelectron pulse height distribution c) Low dark current pulse d) High gain e) Compact size 256 CHAPTER 9 APPLICATIONS

9. 6 Biotechnology

In life science applications, photomultiplier tubes are mainly used for detection of fluorescence and scat- tered light. Major equipment used for life science includes cell sorters, fluorometers and DNA sequencers.

9. 6. 1 Overview

(1) Cell sorters When light is irradiated onto a rapidly flowing solution which contains cells or chromosomes, a scat- tered light or fluorescence is released from the cells or chromosomes. By analyzing this scattered light or fluorescence, it is possible to elucidate cell properties and structures and separate the cells based on these properties. This field is known as . A cell sorter like the one illustrated in Figure 9-26 is most frequently used. The cell sorter is an instrument that selects and collects only specific cells labeled by a fluorescent substance from a mixture of cells in a solution.

CELL

AIR PRESSURE LENS FILTER FILTER PMT

DETECTOR () SIGNAL PROCESSING LENS

ELECTRIC LASER BEAM DATA FIELD PROCESSING ELECTRIC− FIELD + DEFLECTING PLATE

RIGHT LEFT CELL COLLECTOR CELL COLLECTOR

FLASK

TPMOC0025EA

Figure 9-26: Major components for a cell sorter In a cell sorter, a fluorescent probe is first attached to the cells. The cells pass through a thin tube at a fixed velocity. When each cell passes through a small area onto which an intense laser beam is focused, the fluorescence is emitted from the cell and is detected by a photomultiplier tube. The photomultiplier tube outputs an electrical signal in proportion to the number of fluorescent molecules attached to each cell. At the same time, the laser beam light is scattered forward by the cell, and detecting this scattered light gives information on the cell volume. After processing these two signals, the cell sorter creates an electrical pulse that deflects a drop of liquid, containing the desired cell into one of the collection tubes. 9. 6 Biotechnology 257

(2) Fluorometers While the ultimate purpose of the cell sorter explained above is to separate cells, the fluorometer16) is used to analyze cells and chemical substances by measuring the fluorescence or scattered light from a cell or chromosome with regard to such factors as the fluorescence spectrum, fluorescence quantum efficiency, fluorescence anisotropy (polarization) and fluorescence lifetime. (See Figure 9-27.)

LIGHT SOURCE MONOCHROMATOR

PMT-C DIGITAL WAVELENGTH CONTROLLER

SHUTTER CELL FILTER

POLARIZER

PMT-A PMT-B

SAMPLE CHAMBER

FROM PMT-B or PMT-C AMP

COMPUTING CIRCUIT

PRINTER

TPMOC0026EA

Figure 9-27: Automatic fluorescence-depolarization photometer The basic configuration of the fluorometer is nearly identical with that of the fluorospectrophotometer and thus a description is omitted here. There are a variety of models of fluorometers which are roughly categorized into: filtering fluorescence photometers, spectrofluorescence photometers, compensated- spectrofluorescence photometers, fluoroanisotropy analyzers, and phase fluorescence lifetime measure- ment systems. Of these, the fluoroanisotropy analyzer is an instrument specially dedicated to measurement of fluorescence-depolarization. When performing research on biological samples such as proteins, nucleic acid and lipid membranes, rotational relaxation of a fluorescent molecule takes place only slowly and the fluorescence is polarized in most cases. It is still necessary to compensate for the effect of fluorescence depolarization when measuring quantum efficiency and spectrum. For this purpose, the automatic fluorescence-depolarization photometer uses a pair of photomultiplier tubes which detect the two polarized components at the same time. 258 CHAPTER 9 APPLICATIONS

(3) DNA sequencers This is an instrument used to decode the base arrangement of DNA extracted from a cell. The principle of a DNA sequencer is shown in Figure 9-28. An extracted DNA segment is injected onto gel electrophore- sis plate along with a fluorescent label which combines with the DNA. When an electric potential is applied across the gel, the DNA begins to migrate and separate based on size and charge. When the DNA segment reaches the position of the scanning line, it is excited by a laser, causing only the portion with the labeling pigment to give off fluorescence. This fluorescent light is passed through monochromatic filters and detected by photomultiplier tubes. Computer-processing of the position at which the fluorescence has occurred gives information on where the specific bases are located. The DNA sequencer is used for the genetic study of living organisms, research into the cause and treatment of genetic diseases and decoding of human genes.

PIGMENT-LABELED DNA MONOCHROMATIC FILTERS

REFLECTING SURFACE FLUORESCENT INTENSITY TIME SCAN LINE

PMT

LASER

PMT

MONOCHROMATIC SCAN FILTERS MIRROR FLUORESCENT INTENSITY TIME

TPMOC0027EA

Figure 9-28 Principle of a DNA sequencer

9. 6. 2 Major characteristics required of photomultiplier tubes

Because the photomultiplier tube detects very-low fluorescence emitted from a cell or DNA, the following characteristics are required as in the case of spectrophotometry. a) High stability b)Low dark current c) High signal-to-noise ratio d)Wide spectral response e) Low hysteresis f) Excellent polarization properties 9. 9 Spectrometry/Solid Surface Analysis 265

9. 9 /Solid Surface Analysis

Mass spectrometry is a technique used to identify and analyze the mass, makeup and minute quantity of a sample through the measurement of the difference in mass and movement of ions by exerting electric or magnetic energy on the sample which is ionized. Solid state surface analysis is used to examine the surface state of a sample through the measurement of photoelectrons, secondary electrons, reflected electrons, transmitting electrons, Auger electrons or X-rays which are generated as a result of interactions of incident electrons with atoms composing the sample, which take place when an electron beam or X-ray irradiates the sample.

9. 9. 1 Mass

(1) Overview23) 24) Mass spectrometers are broadly classified into two groups: one using magnetic force (magnet) and one not using magnetic force. Currently used mass spectrometers fall under one of the following four types. • Time of flight (TOF) type • Quadrupole (Q-Pole) or type • Magnetic field type • Ion cyclotron (FTICR) type Among these, the quadrupole (Q-Pole) type mass is most widely used and its block dia- gram is shown in Figure 9-39.

ION SOURCE Q-POLE ELECTRODE DETECTOR POWER SUPPLY POWER SUPPLY POWER SUPPLY

RECORDER

ION DETECTOR SOURCE Q-POLE ELECTRODE VACUUM ~10-4Pa

VACUUM PUMP

TEMC0019EA

Figure 9-39: Block diagram of a quadrupole (Q-Pole) type mass spectrometer When a sample is guided into the ionizer, it is ionized through the electronic ionization (EI method), (CI method), fast atomic bombardment (FAB method), electro-spray ionization (ESI method) or atmospheric pressure chemical ionization (APCI method). The ionized sample is sent to the analyzer section (quadrupolar electrodes) in which the sample is separated depending on the mass per charge count (m/z) by the DC voltage and high-frequency voltage applied to the quadrupolar electrodes. After passing through the analyzer section, the ions then reach the detector section where they are detected by an electron multiplier tube. 266 CHAPTER 9 APPLICATIONS

Mass spectrometers are often combined with a gas chromatograph or liquid chromatograph to build a gas chromatograph mass spectrometer (GC-MS) or liquid chromatograph mass spectrometer (LC-MS). Mass spectrometers are used to identify, measure and analyze the composition of various samples such as petrochemicals, fragrance materials, medicines, biogenic components and substances causing environ- mental pollution. Figure 9-40 shows the schematic drawing of a gas chromatograph mass spectrometer.

1 GAS 2 I/F 3 MASS SPECTROMETER CHROMATOGRAPH

Q-POLE

LENS

SEPARATOR OFF AXIS DEFLECTOR

ELECTRON MULTIPLIER

TEMC0020EA

Figure 9-40: Schematic drawing of a gas chromatograph mass spectrometer.

(2) Major characteristics required of electron multiplier tubes Since the mass spectrometer measures and analyzes the sample in minute amounts, electron multiplier tubes should have the following characteristics. a) High gain b) Low noise c) Long operating life

9. 9. 2 Solid surface analyzers

(1) Overview25) Solid surface analyzers are broadly divided into two groups: one using electron beams to irradiate a sample and the other using X-rays. Major solid surface analyzers presently used are as follows. • Scanning electron (SEM) • Transmission (TEM) • Auger electron spectrometer (AES) • Electron spectrometer for chemical analysis (ESCA) When a sample is irradiated by electron beams, interactions of the incident electrons with atoms which compose the sample occur and generate various kinds of signals characterized by the particular atom. Figure 9-41 shows the kinds of signals obtained and the approximate depth at which each signal is gener- ated on the surface of the sample. 9. 9 Mass Spectrometry/Solid Surface Analysis 267

CATHODE SECONDARY INPUT LUMINESCENCE ELECTRON ELECTRON

X RAY

REFLECTED INPUT

ELECTRON

ELECTRON

AUGER

SECONDARY ELECTRON

AUGER ELECTRON (5~10 nm) CONTINUOUS ELECTRON

X-RAY REFLECTED

SAMPLE ELECTRON (50nm~100nm) INTERNAL

POWER

CHARACTERISTIC

FLUORES- X-RAY (5~10µm) ABSORBED CENCE

TRANSMITTED ELECTRON X-RAY

ELECTRON Quantum Generation Depth and Spatial Resolution (By Goldstein) TAPPC0078EA

Figure 9-41: Interactions of incident electrons with sample

Obtained signals are chosen to extract necessary information according to measurement purpose, which is then used for analyzing the surface of the sample. Among the four types of surface analyzers, the scanning electron microscope (SEM) is the most widely used and its structure and principle are illustrated in Figure 9-42.26)

(1)The sample is irradiated by an electron beam to generate secondary electrons which are then collected. MAGNETIC FIELD LENS (2)The sample is scanned by an electron beam, just like UPPER END TV scan.

LOWER (3)A gray level image is displayed END according to the quantity of DETECTOR secondary electrons. SECONDARY ELECTRONS ELECTRON (1) GUN SAMPLE INPUT MAGNETIC SCAN ELECTRON FIELD LENS ELECTRON (FOCUSING LENS) BEAM (2) SAMPLE

MAGNETIC SCAN FIELD LENS SCAN COIL (OBJECTIVE LENS)

SAMPLE S S S S L L L A B M M ABCD C D (3)

SECONDARY ELECTRON SECONDARY DETECTOR ELECTRON DETECTOR An electron beam emitted from the is narrowed into a needle-like form and focused onto the sample by the focusing lens and objective lens. TAPPC0079EA

Figure 9-42: Structure and principle of a scanning electron microscope 268 CHAPTER 9 APPLICATIONS

An electron beam emitted from the electron gun is accelerated at a voltage of 0.5 to 30kV. This acceler- ated electron beam is then condensed by electromagnetic lens action of the focusing lens and objective lens, and finally formed into a very narrow beam of 3 to 100nm in diameter, irradiating on the surface of a sample. Secondary electrons are then produced from the surface of the sample where the electron beam landed, and detected with a secondary electron detector. The electron beam can be scanned in the XY directions across the predetermined area on the surface of the sample by scanning the electromagnetic lens. A magnified image can be displayed on the CRT in synchronization with the signals of the secondary electron detector. Figure 9-43 shows the structure and operation of the secondary electron detector.

COLLECTOR INPUT SCINTILLATOR PNJ ELECTRON LIGHT PIPE A B

PMT PRE-AMP

S S

B B

SEI SAMPLE REF A S l EC ED : SECONDARY ELECTRON : REFLECTED ELECTRON

(a) Secondary electron detector (b) Secondary electron detector

TAPPC0080EA

Figure 9-43: Structure and operation of a secondary electron detector A typical secondary electron detector consists of a collector electrode, scintillator, light pipe, photomul- tiplier tube and . Voltage is applied to the collector electrode and scintillator at a level required to collect secondary electrons efficiently. Most of the secondary electrons produced from the sample enter the scintillator and are converted into light. This converted light then passes through the light pipe and is detected with the photomultiplier tube. Figure 9-44 shows the images of a broken surface of , observed with a scanning electron microscope.27)

Figure 9-44: Photographs of broken ceramic material taken with a scanning electron microscope

(2) Major characteristics required of photomultiplier tubes To detect low level light emitted from the scintillator, photomultiplier tubes must have the following characteristics. a) High stability b) Low dark current c) High quantum efficiency 290 CHAPTER 9 APPLICATIONS

References in Chapter 9

1) Japan Analytical Instruments Manufacturers' Association: Guide to Analytical Instruments, 3rd Edition, 37 (1986). 2) Thermo Jarrell Ash Corp. Ltd.: The Atom Scan25 Spectrometer. 3) H. Daidouji: The Spectroscopical Society of Japan - Measurement Method Series, 20, 129, Japanese Association of (1985). (Published in Japanese) 4) Japan Analytical Instruments Manufacturers' Association: Guide to Analytical Instruments, 3rd Edition, 42 (1986). (Published in Japanese) 5) Japan Analytical Instruments Manufacturers' Association: Guide to Analytical Instruments, 3rd Edition, 45 (1986). (Published in Japanese) 6) K. Anan, K. Konno, Z. Tamura, M. Matsuhashi, J. Matsumoto and M. Watanabe: Fundamental Biochemical Experimental Methods, 4, 32, Maruzen Corp. (1975). (Published in Japanese) 7) M. Yamamoto: Medical Electronics and Bioengineering, 24, 6, 54 (1986). (Published in Japanese) 8) M. Yamamoto: Positron CT Imaging - PET- Medical History, 127 (14), 1218-1225 (1983). 9) Y. Endo and K. Miyai: Protein, Nucleic Acid and Enzyme, Separate Volume 31, Enzyme Immunoassay, 13, Kyoritsu Publishing Corp. (1987). (Published in Japanese) 10) G. Kawashima: Introduction to Immunoassay, 29, Nanzandou (1987). 11) Japan Analytical Instruments Manufacturers' Association: Guide to Analytical Instruments, 3rd Edition, 228 (1986). (Published in Japanese) 12) G. Kawashima: Introduction to Immunoassay, 162, Nanzandou (1987). (Published in Japanese) 13) G. Kawashima: Introduction to Immunoassay, 83, Nanzandou (1987). (Published in Japanese) 14) T. Oda and H. Maeda: Protein, Nucleic Acid and Enzyme, Separate Volume 31, Enzyme Immunoassay, 243, Kyoritsu Publishing Corp. (1987). (Published in Japanese) 15) Association of Japanese Radiation Instrument Industry: General Guide to '84 Medical Radiation Instrument Technology. 198, Denshi Keisoku Publishing Company (1983). (Published in Japanese) 16) K. Kinoshita and T. Kamiyama: Fluorescence Measurement - Application to , 83, The Spectroscopi- cal Society of Japan (1983). 17) T. Hayashi: Photomultiplier Tubes For Use In High Energy Physics, 6 (1983). 18) Hamamatsu Photonics: Photomultiplier Tubes and Environmental Conditions (1986) 19) Hamamatsu Photonics: Photomultiplier Tubes and Environmental Conditions (1986) 20) Hamamatsu Photonics Data Sheet: Ruggedized High-Temperature Photomultiplier Tubes (1992). 21) Hamamatsu Photonics Data Sheet: Ruggedized High-Temperature Photomultiplier Tubes (1992). 22) Hamamatsu Photonics Data Sheet: Ruggedized High-Temperature Photomultiplier Tubes (1992). 23) M. Tsuchiya, M. Ohashi, T. Ueno: New Development of Mass Spectrometry, Modern Chemical, Extra Number 15, Tokyo Chemical Coterie. 24) T. Ueno, K. Hirayama, K. Harada: Biological Mass Spectrometry, Modern Chemical, Extra Number 31, Tokyo Chemical Coterie. 25) JEOL Ltd.: Introduction to the world of SEM (published in Japanese) 26) The Nikkan Kogyo Shimbun: Structure of /Wonder of Technology, No. 3, 42, 1996 (published in Japa- nese) 27) JEOL Ltd.: Introduction to the World of SEM (published in Japanese) 28) Japan Analytical Instruments Manufacturers' Association: Guide to Analytical Instruments, 3rd Edition, 171 (1986). (Published in Japanese) 29) Japan Analytical Instruments Manufacturers' Association: Guide to Analytical Instruments, 3rd Edition, 179 (1986). (Published in Japanese) 30) Japan Analytical Instruments Manufacturers' Association: Guide to Analytical Instruments, 3rd Edition, 133 (1986). (Published in Japanese) 31) Japan Analytical Instruments Manufacturers' Association: Guide to Analytical Instruments, 3rd Edition, 133 (1986). 32) Aroka Corp.: Gated Surface Monitor. 33) H. Shimizu, N. Takeuchi, Y. Sasano, N. Sugimoto, I. Matsui and N. Okuda: Oyo Butsuri (), 50, References 291

11, 1154 (1984). (Published in Japanese) 34) Hamamatsu Photonics Data Sheet: MCP-PMTs (Microchannel Plate - Photomultiplier Tubes) 35) I. Yamazaki, H. Kume, N. Tamai, H. Tsuchiya, K. Oba: Oyo Butsuri (Applied Physics), 50, 7, 702 (1985) 36) T. Yamauchi et al.: Jpn. J. Appl. Phys. 21, 2, 348 (1982). 37) T. Yamauchi et al.: Jpn. J. Appl. Phys. 21, 2, 350 (1982). 38) Y. Suzuki, A. Funabashi, A. Ogata and T. Matoba: Measurement and Control, 17, 1 (1977). 39) Japan Analytical Instruments Manufacturers' Association: Guide to Analytical Instruments, 3rd Edition, 143 (1986). 40) Japan Analytical Instruments Manufacturers' Association: Guide to Analytical Instruments, 3rd Edition, 143 (1986). 41) : Wafer Dust Inspection System WM-3 (1990). 42) Co., Ltd: TKF-106. 43) Y. Nakada: (Electronic Parts and Materials) '83, Separate Volume, 189, Kogyo Chosa Kai Publishing Co., Ltd. (1983). 44) Japan Analytical Instruments Manufacturers' Association: Guide to Analytical Instruments, 3rd Edition, 46 (1986). (Published in Japanese) 45) Japan Spectroscopic Co., Ltd.: NR-1800.