Considerations of the Use of Triple Quadrupoles Or Ion Traps in Quantitative Applications Triple Stage Quadrupole API MS / MS Full Scan Products
Total Page:16
File Type:pdf, Size:1020Kb
Considerations of the use of Triple Quadrupoles or Ion Traps in Quantitative Applications Triple Stage Quadrupole API MS / MS Full Scan Products Q1 PARENT MASS ANALYZER IS Q3 PRODUCT PARKED, PASSING MASS ANALYZER PARENT IONS Q0 LENS ONLY PARENT IONS IS SCANNED, ENTER Q2 + / - IONS AND TRANSPORTS OF A SINGLE M / Z PASSING FULL COLLISION CELL NEUTRALS + / - IONS TO Q1 TO Q2 COLLISION SCAN PRODUCT COLLIDE WITH FORMED IN PARENT MASS CELL ION SPECTRA TO Ar GAS AND API SOURCE ANALYZER THE DETECTOR DISASSOCIATE E n e rg y ION SOURCE Q0 Q1 Q2Ar Q3 DETECTOR Triple Stage Quadrupole API MS / MS SRM Products Q1 PARENT MASS ANALYZER IS Q3 PRODUCT PARKED, PASSING MASS ANALYZER PARENT IONS Q0 LENS ONLY PARENT IONS IS PARKED, PASSING ENTER Q2 + / - IONS AND OF A SINGLE M / Z ONLY PRODUCT TRANSPORTS COLLISION CELL NEUTRALS TO Q2 COLLISION IONS OF A SINGLE + / - IONS TO Q1 & COLLIDE WITH FORMED IN PARENT MASS CELL M / Z TO THE API SOURCE Ar GAS THEN DETECTOR ANALYZER DISASSOCIATE E n e rg y ION SOURCE Q0 Q1 Q2Ar Q3 DETECTOR Trapology Column Charge Effluent 0 + + 0 - Ion Source - Ion Gate Helium + + Bath Gas Ion Scanned RF Injection + + + Ion Trap + Mass Analyzer Waveforms + + Computer Ion Current Detector An ion trap holds a fixed maximum number of ions and so can be filled and emptied a certain number of times during the evolution of a chromatographic peak. The length of time that the trap stays open to collect ions is determined by a pre-scan which measures total ion current (prevents space charging…) Pre-scan 5 sec. Across a peak 5 sec. wide the trap might fill and empty 5 times. So, a group of ions are collected ca. every 1 sec - each group is then ejected to the detector, smaller ions first. 1 sec 1 sec 1 sec 1 sec 1 sec 5 sec. Quadrupole MS scanning A quadrupole mass analyzer scans by ramping the DC/RF voltage in time across the quadrupoles. So, in any given time, ions of one m/z only are present and pass through to the detector - all others are rejected. Quadrupole MS scanning If one MS/MS scan between m/z 100 and m/z 500 is completed in one second, then each m/z will be allowed to pass for 2.5 ms. For the same peak, the quadrupole performs 5 complete scans from 100 - 500 m/z each taking 1 sec. 100 - 500 m/z 100 - 500 m/z 100 - 500 m/z 100 - 500 m/z 100 - 500 m/z 5 sec. Quad and trap full scan summary For a trap, each m/z (and all For the quadrupole, m/z at the same time) is/are each m/z is scanned collected for ca. 750 ms (sequentially) to the (taking pre-scan and interscan detector for times into account) and then 2.5 ms scanned to the detector. + + + + Ion traps have the edge over quads in full scan MS/MS For full scan MS/MS, a trap will give better sensitivity - because there are more ions representing each m/z arriving at the detector for each scan. How much more sensitive is the ion trap for full scan MS/MS? Accounting for pre-scan/interscan timing, the trap produces ca. 300 times (750 ms/2.5 ms) for the “collection” of each m/z compared to a quadrupole with the same ion source emittance (i.e. 2 orders of magnitude.) But wait…there’s more! What if I wanted to pass (filter) only one m/z ion to the detector (i.e. SRM) - then I could spend more time on that ion... Duty cycle considerations On a quadrupole, A trap will still only interscan times are collect ions in relatively small and so ‘batches’ - and pre- the quadrupole scan/interscan times remains fixed on that afford a duty cycle of one ion - a duty cycle about 75%. of close to 100%. + + The chromatographic time scale… While the beam instrument is continuously detecting one particular m/z - a trap builds a curve from an ‘average’ over each collection time - and the points are least frequent at the most important region for quantitation (the ‘take-off’.) In general… When comparing a triple quad and an ion trap with identical ion source emittances, the triple quadrupole device provides ~ an order of magnitude better LOQ (from better RSDs) for quantitation experiments than a trap, primarily due to integration effects. SRM 5pg Alprazolam with LCQDECA RT: 4.26 SN: 20 100 90 80 %RSD 6.11 70 60 50 40 30 Relative Abundance Relative 20 10 0 SRM 750fg Alprazolam with TSQ RT: 4.41 SN: 28 100 90 80 70 %RSD 1.87 60 50 40 30 20 10 0 0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 Time (min) Conclusions ◗ Both traps and triple quads can be used very effectively in qualitative and quantitative applications ◗ Making a decision on which to use should involve careful consideration of the application as well as budget ◗ In general: – Traps offer better sensitivity than triple quadrupoles in full scan product ion MS/MS mode (e.g., many qualitative applications, peptide sequencing) – Triple quads offer better performance for SRM experiments (e.g., in target compound quantitation) – It is important to also consider the ion source emittance and other factors that effect performance. These may not be equivalent across various vendor product lines, which could affect the accuracy of the broad generalizations stated above..