Accepted Manuscripts are published online shortly after acceptance, before technical editing, formatting and proof reading. Using this free service, authors can make their results available to the community, in citable form, before we publish the edited article. We will replace this Accepted Manuscript with the edited and formatted Advance Article as soon as it is available.
You can find more information about Accepted Manuscripts in the Information for Authors.
Please note that technical editing may introduce minor changes to the text and/or graphics, which may alter content. The journal’s standard Terms & Conditions and the Ethical guidelines still apply. In no event shall the Royal Society of Chemistry be held responsible for any errors or omissions in this Accepted Manuscript or any consequences arising from the use of any information it contains.
www.rsc.org/methods Page 1 of 6 Analytical Methods
1 2 Analytical Methods RSC Publishing 3 4 5 TECHNICAL NOTE 6 7 8 9 An improved method for BTEX extraction from 10 charcoal 11 Cite this: DOI: 10.1039/x0xx00000x 12 13 Raffaele Cucciniello a, Antonio Proto a,*, Federico Rossi a, Nadia 14 b c 15 Marchettini , Oriana Motta Received 00th January 2015, 16 Accepted 00th January 2015 17 Abstract In this paper we propose a simple procedure for the extraction of BTEX (benzene, 18 DOI: 10.1039/x0xx00000x toluene, ethylbenzene and xylenes) from activated charcoal. For this purpose synthetic samples 19 www.rsc.org/ were prepared in laboratory and real samples were collected in a polluted environment using 20 21 passive sampling. The proposed extraction procedure is based on the use of an appositely Manuscript 22 designed soxhlet apparatus which dimensions are as small as the cartridge of the passive 23 samplers, in order to accommodate this latter and permit to use a small amount of solvent. The 24 25 selection of solvent was based on the strong request to substitute carbon disulfide, which is by far 26 the solvent of choice, with less hazardous substance in order to enhance operator protection. 27 Among the solvents tested the best results were found by using dichloromethane that, under 28 29 optimized conditions, led to a complete recovery (>97%) of each analyte (BTEX), both in the
30 synthetic and the real samples. The extracting procedure, with CH 2Cl 2 in the soxhlet last about 20
31 minutes which also results in a reduced operating time. Accepted 32 33 34 35 36 37 38 Introduction Occupational carbon disulfide exposure has been associated with a variety of health effects stimulating some 39 Aromatic hydrocarbons, such as benzene, toluene, ethylbenzene and xylenes (commonly called BTEX) are widely present in the epidemiological researchers to study the potential relationships
40 Methods environment. These compounds can be generally found in between the exposures and neurological, reproductive, 41 gaseous phase and have a key role in the chemistry of the cardiovascular or endocrine dysfunctions as well as eye and 42 troposphere [1,2]. Major sources of aromatics to the urban skin irritations. It has been reported that CS 2 induces diffuse 43 environment include vehicle exhaust (mobile sources), vascular encephalopathy, parkinsonism and peripheral 44 petroleum refining processes, residential heating, waste polyneuropathy [10,11,12]. The US Occupational Safety and 3 45 incineration, coke ,aluminium production (stationary sources) Health Administration (OSHA) indicates for CS 2 12.4 mg/m as 46 and gas fuel stations [3,4,5]. The presence of BTEX in the an 8 h TWA concentration [9]. However Godderis et al. [13] demonstrated that diminished neurobehavioral functions in 47 urban atmosphere has a direct impact on human health because of their carcinogenic and mutagenic potential [6], therefore they workers currently exposed to CS 2 could be detected even if 48 3 need to be monitored periodically. The techniques used for average exposure was as low as ≤ 10 mg/m in a relatively short 49 BTEX monitoring are usually based on their sorption in a tube time (<10 years exposure). Therefore they concluded that the 50 current TLV insufficiently protects against neurological effects
(active or passive devices) which is filled with a suitable Analytical 51 sorbing agent, usually activated charcoal, followed by solvent and should be lowered. The request to find less hazardous substituted was first caught 52 desorption with CS 2 and gas chromatographic determination 53 [7,8]. In Italy, at the best of our knowledge, more than 95% of by Bertoni and Ciuchini [14], who investigated on the 54 laboratories of environmental analysis follows this procedure feasibility of substituting CS 2, in the desorption of COV notwithstanding alternative techniques are well known but not sampled on activated charcoal adsorbents, with 55 dichloromethane. CH Cl is a good solvent for GC analysis 56 commonly used, among the others, thermal desorption and 2 2 accelerated solvent extraction [9], because of their high costs. because it has a low boiling point and it doesn’t interfere with 57 other analytes, it has an higher TLV (174 mg/m 3) and is less 58 expensive with respect to CS 2. The results of BTEX extraction 59 60 This journal is © The Royal Society of Chemistry 2015 J. Name ., 2015, 00 , 1-3 | 1 Analytical Methods Page 2 of 6 TECHNICAL NOTE Journal Name
1 with CH 2Cl 2 were not satisfying because the recovery ranged benzene is well separated from that of the solvent and easily 2 from 90% for benzene to 55% for o–xylene. They also tested integrable. 3 different solvents to increase the recovery, without achieving 4 appreciable results. Results 5 This study follows and completes the investigations of Bertoni Apparatus description and extraction conditions 6 and Ciuchini with particular attention to BTEX extraction from Preliminary tests were performed on laboratory prepared 7 passive devices filled with activated charcoal. We screened a samples in order to compare the extraction capacity of selected 8 large number of solvents and CH 2Cl 2 turned out to be the most solvents with carbon disulfide. Synthetic samples were efficient in our experimental conditions. prepared by injecting 1.0 µL of each compound (benzene, 9 Moreover, CH Cl is already used in ultrasonic extraction 10 2 2 toluene, ethyl benzene, o xylene, m xylene and p xylene) to 10 procedures for BTEX, as recommended by EPA standards g of activated charcoal into a closed chamber. The system was 11 [3,4,5]; this procedure is, in fact, a valid alternative to the kept closed for 24 hours and charcoal was finally removed and 12 employment of CS , though many laboratories still prefer 2 maintained under N 2 atmosphere before analysis. Analyses 13 Soxhlet extraction for its simplicity and low cost of the basic were conducted on 300.0 mg of sample, i.e. the same amount as 14 equipment [15]. in a RING passive sampler, which contains 2.65 µg benzene 15 We performed the extraction of BTEX with CH 2Cl 2 by using a and xylenes and 2.63 µg of toluene and ethyl benzene. Classic small Soxhlet appositely designed to contain the cartridges of 16 extraction with CS 2 was performed by desorbing the analytes the passive samplers. Recently Soxhlet has been defined as a directly into a glass tube for about 30 minutes with 2 mL of 17 “panacea” in the field of extraction [15], where the 18 solvent (CS 2, low benzene, Sigma Aldrich) following NIOSH development of new technologies has improved its performance method 1501 [8], than the extract was filtered on cellulose 19 opening the door to the utilization in a number of different based filter to remove carbon residues and successively injected 20 approaches. Optimizing the extraction procedure we obtained for the gas chromatographic analysis. The same procedure was 21 quantitative recoveries for each analyte after only eight followed for the other solvents; samples were extracted by Manuscript 22 extraction cycles, that last about 20 minutes. Moreover we using 2 mL of solvent for 30 minutes in a closed glass tube and 23 applied the same procedure to analyze real environmental successively analyzed by GC FID. The blank was prepared by 24 samples and found data comparable to CS 2 extraction in all using activated charcoal without any treatment. cases. 25 Figure 1 shows the performances of the various solvents tested, where the extraction capacity of each solvent was normalized 26 Experimental with respect to the extraction capacity of CS 2 (100%). 27 Reagents Dichloromethane resulted to give the best results with an 28 The organic solvents used to desorb BTEX from activated efficiency of about 80% with respect to CS 2. Since CH 2Cl 2 is 29 charcoal were pentane, isopropanol, acetonitrile, also suitable for gas chromatographic analysis (low boiling 30 dichloromethane, carbon disulfide, diethylether, chloroform and point and poor response to FID), we used it to test our proposed methylethylketone. Stock standards in all solvents were extraction procedure. 31 Accepted 32 prepared from benzene, toluene, ethylbenzene and o , m , p 33 xylenes. The solutions were kept at 4 °C. All reagents were purchased from Sigma Aldrich at reagent grades and used 34 without any purification. Carbon disulfide was bought form 35 Sigma Aldrich, declared as product with low benzene 36 contamination. 37 38 Samplers and field sampling 39 RING passive samplers filled with activated charcoal (300 mg) were obtained from Aquaria Research srl, Italy.
40 Methods 41 Field sampling was conducted in the city of Salerno (Italy). The measurement site was a commercial area characterized by an 42 elevated traffic density. Each sampler was exposed for 72 h, at 43 a height of approximately 2 m above the street level. Average Figure 1 . BTEX recovery (%) (average of the average of three 44 temperatures were measured by a thermometer data logger with replicates) with different solvents. 45 ± 1.0°C accuracy. For each sampling two samplers were 46 exposed and, after the sampling, they were transported and 300.0 mg of contaminated charcoal were introduced into the 47 stored at less than 4°C until analysis and were processed within steel cartridges of passive samplers; the BTEX were 48 a week after their arrival in the laboratory. successively extracted by using a specifically designed mini 49 soxhlet apparatus (Microglass Heim s.r.l., Italy. Figure 2), in Chromatographic conditions order to enhance the extractive capacities of the solvent. The 50 Analysis was done by a GC FID (Agilent 7890A) equipped steel cartridges were employed to eliminate possible losses of Analytical 51 with HP 5 column (50 m×0.2 mm id, 0.25µm). GC oven was carbon from the siphon to the solvent reservoir. The soxhlet 52 programmed to 60 °C, 2 min hold, ramped to 140°C at a rate of extraction chamber was dimensioned to fit the passive sampler 53 6°C/min following a 2 min hold, then ramped to 250 °C at a cartridge (2 cm i.d. × 4 cm height) and to guarantee a complete 54 rate of 20 °C/min following a 10 min hold. Helium was used as contact between solvent and substrate. 10 mL of solvent were 55 carrier gas with a flow rate of 1 mL min −1 and splitless injection used for each extraction, after which samples were collected in 56 mode was used. The injection volume was 1 µL. It has to be a volumetric flask and, in case of CH 2Cl 2 loss, fresh solvent 57 underlined that in these experimental conditions, the peak of was added up to 10 mL before analysis.
58 59 60 2 | J. Name ., 2015, 00 , 1-3 This journal is © The Royal Society of Chemistry 2015 Page 3 of 6 Analytical Methods Journal Name TECHNICAL NOTE
1 Passive sampling and field measurement 2 After laboratory tests with synthetic samples, the soxhlet 3 extraction procedure was successively applied for the BTEX 4 extraction from passive samplers exposed in a high traffic 5 street. Radial diffusive sampler was first invented in 1990 by 6 Eickeler and Löffenholz [17] as a dosimeter for detecting gases 7 or aerosol. This tool is very useful for monitoring many classes 8 of substances such as VOCs, nitrous and sulfuric oxides, ammonia, volatile acids and, more recently, carbon dioxide 9 [18,19]. 10 Diffusive cartridges having a radial symmetry for a high 11 sampling rate, have been employed for monitoring common 12 pollutants concentration in atmosphere [19,20] and particularly 13 BTEX [21,22], the most common among the volatile organic 14 compounds generally present in a polluted environment. Radial 15 symmetry provides higher sampling rates than other types of 16 passive samplers.
17 This passive sampler consists of an adsorbent cartridge packed 18 Figure 2 . Soxhlet apparatus for BTEX (extraction chamber 2cm i.d. × 4 cm height, passive sampler cartridge 0,9 cm i.d × with 300 mg of activated charcoal inserted in a microporous 19 3,3 cm height, bottom flask volume 20 mL). polyethylene membrane. The compounds diffuse through the 20 membrane toward the cartridge driven by the gradient of 21 concentration between the ambient air and the inner cartridge. 3 Manuscript 22 Compound concentrations C (mg/m ) can be calculated by 23 applying an equation derived from Fick’s first law of diffusion: 24 mg m m mg 25 C mL 26 m 10 P t min min 27 28 Where m d is the adsorbed mass of the analyte sampled during 29 the time t while m b is the mass of the analyte on a non exposed 30 cartridge (blank), P is the diffusive uptake rate of the substances (73 mL/min for benzene, 66 mL/min for toluene, 60
31 Accepted mL/min for ethylbenzene, 58 mL/min for o xylene and 64 32 mL/min for m&p