An Element for Your Success
Unabhängige,Independent consultantsberatende and und planning planende engineers Ingenieure for water, waste water, solid waste and environmental für Wasser-, Abwassertechnologies-, Abfall- und Umwelttechnik
Hannover Oldenburg ∙ Erfurt Hannover ∙ Bremen Erfurt
COD Determination
Hannover Oldenburg ∙ Erfurt Hannover ∙ Bremen Erfurt
Quick COD (LAR) Measurement Method: • Thermal combustion at 1200°C which guarantees the complete oxidation of the sample without reagents.
• Then measured with an NDIR detector
Reliability: • High reliability – 98% availability
Sensitivity: Measurement range: • 10-150 mg/l COD • 100-2,000 mg/l COD • 500-5,000 mg/l COD • 1,000-8,000 mg/l COD • 20,000-250,000 mg/l COD • More ranges on request
Hannover Oldenburg ∙ Erfurt Hannover ∙ Bremen Erfurt
Quick COD (LAR)
Calibration: • Not specified
Performance: Feautures and accessories: • Catalyst-free technique • Reagent-free analysis • No filtration necessary • No memory effects • Very low maintenance and operational costs • Auto start option • Analyser available minimum 98% • Multi stream measurement (optional) • Monitors up to six streams
Hannover Oldenburg ∙ Erfurt Hannover ∙ Bremen Erfurt
Quick COD (LAR)
Equipment Configuration: • Designed for wall mounting, outdoors or indoors. Rail- and stand-mounting options
are available. • weatherproof, and lockable. Dimensions • Width 700 x height 1,020 x depth 520 mm Weight • 115 kg (standard)
Sample Volume: 10-600 microliter small
Sample Preparation • Maintenance- free particle cutter • Optional homogeniser for the continuous homogenisation of samples
Hannover Oldenburg ∙ Erfurt Hannover ∙ Bremen Erfurt
Quick COD (LAR)
Response Time: • 1-2 minutes
Particle Tolerance: • Max. 0.8mm according to ISO 8245
Measurement interval • 1 – 3 minutes - depending on the application
Reagent Consumption • Only distilled water and a small amount of inexpensive acid required
Reagent consumption: • No reagent
Hannover Oldenburg ∙ Erfurt Hannover ∙ Bremen Erfurt
CAA- Continuous Auto Analyzer (PAI)
Measurement Method: • Real continuous automatic monitoring of a wide range of parameters in liquid sample using these analytical techniques: • Potentiometry • Conductivity • Photometry • Fluorimetry Reliability: • Not specified Sensitivity: • not specified.
Calibration • Manual/automatic (optional)
Hannover Oldenburg ∙ Erfurt Hannover ∙ Bremen Erfurt
CAA- Continuous Auto Analyzer (PAI)
Performance:
Feautures and accessories:
Accessories • Not specified Special features • Contrary to other techniques which analyze different samples, analyse the same sample continuously. • Cacluation of contaminant loads even if rapid changes in concentration may occur.
Sample Condition: • Pressure: 0 bar with sample pump. Higher pressures on demand • Flow rate: application dependant • Temperature: >0-50 º C (application dependant)
Hannover Oldenburg ∙ Erfurt Hannover ∙ Bremen Erfurt
CAA- Continuous Auto Analyzer (PAI)
Equipment configuration:
• Layout of CAA for ethylone oyide in glycol/water mixtures
Hannover Oldenburg ∙ Erfurt Hannover ∙ Bremen Erfurt
CAA- Continuous Auto Analyzer (PAI)
Dimensions:
• Housing: wall mounting, stainless steel and plastic
• Width ±400 mm x depth: 270 mm x heigt: 910 mm • Weight: approximately 30 kg
Information transfer: • Transmitting results to process control an analog output ( (0)4- 20 mA) is used.
Equipment safety: • Two separated housings to protect the electronics from any hazardous vapour and liquids.
Hannover Oldenburg ∙ Erfurt Hannover ∙ Bremen Erfurt
CAA- Continuous Auto Analyzer (PAI)
Operation : Inbuilt toch screen for manual control of CAA analyser . • Results are displayed in numerical or graphical
presentation
Operation Notification and Failure: • System fault: in case of system failure the alarm contact is triggered. • Threshold alarm (optional): relays for threshold alarm are available. • Response Time: • Not specified
Particle Tolerance: • Not specified
Measurement interval • continuous
Hannover Oldenburg ∙ Erfurt Hannover ∙ Bremen Erfurt
Liquiline System CA 80 (EH)
Measurement Method: • Established COD dichromate method directly comparable to laboratory results.
Reliability: ± 10 % according to potassium hydrogen phthalate test Sensitivity: • 10-5,000 mg/l COD • 40-20,000 mg/l COD
Calibration • Automatic calibration and cleaning User-configurable measuring, cleaning and calibration intervals • Zero standard: 500 ml per month given a calibration interval of 48 h. (no dilution module) • Standard 1: approx. 90 ml per month given a calibration interval of 48 h (no dilution module)
Hannover Oldenburg ∙ Erfurt Hannover ∙ Bremen Erfurt
Liquiline System CA 80 (EH) Calibration Interval: • 1 h to 90 days, depending on the application and ambient condition
Performance: Features and accessories: Accessories – Kit CA8x/CAT860: wall mounting – Kit CA8x/CAT860: housing bottom – Kit CA8x: door with window – … Sample Characteristics: • Volume: 8 ml per measurement • Temperature: 4 to 40 °C (39 to 104 °F) • Consistency: Low solids content, aqueous, homogenized
Measuring interval: • Continuous (approx. 55 min), adjustable 30 min to 24 h
Hannover Oldenburg ∙ Erfurt Hannover ∙ Bremen Erfurt
Liquiline System CA 80 (EH)
Equipment Safety: Electrical safety • According to EN/IEC 61010-1:2010, Class I equipment
• Low voltage: overvoltage category II • For installations up to 2000 m (6500 ft) above MSL
Information Transfer: • Not specified
Particle Tolerance: • Not specified
Detection Limit: • Not specified
Ambient temperature range • 5 to 40 °C (41 to 104 °F) • Humidity 10 to 95%, not condensing
Hannover Oldenburg ∙ Erfurt Hannover ∙ Bremen Erfurt
Comparison
LAR PAI EH
+++ + +++ Reliability
+ ++ +++ Sensitivity
+++ ++ ++ Performance