<<

PROCEDURE Measurement 64 Techniques (Invasive)

Susan Scott PURPOSE: Cardiac output (CO) measurements are used to assess and monitor cardiovascular status. CO can be used in the evaluation of patient responses to various therapies, including fl uid management interventions, vasoactive and inotropic medication administration, and mechanical assist devices. CO measurements can be obtained either continuously or intermittently via a pulmonary (PA) catheter. CO measurements provide data that may be useful in directing and/or improving the care for critically ill patients with hemodynamic instability.

PREREQUISITE NURSING • volume is the amount of volume ejected KNOWLEDGE from either per contraction. Left ventricular is the difference between left ventricular • Understanding of normal anatomy and physiology of the end-diastolic volume and left ventricular end-systolic cardiovascular system and pulmonary system is neces- volume. Left ventricular stroke volume is normally 60 to sary. 100 mL/contraction. Major factors that infl uence stroke • Understanding of basic dysrhythmia recognition and treat- volume are , , and contractility. ment of life-threatening dysrhythmias is needed. • Right preload refers to the pressure in the right • Pathophysiologic changes associated with structural heart ventricle (RV) at the end of and is measured by disease (e.g., ventricular dysfunction from myocardial the RAP or CVP. Elevations in left heart fi lling pressures infarction, diastolic or systolic changes, and valve dys- may be accompanied by parallel changes in RAP, espe- function) should be understood. cially in patients with left systolic ventricular dysfunction. • Understanding of the principles of aseptic technique is nec- Other factors that affect RAP are venous return, intravas- essary. cular volume, vascular capacity, and pulmonary pressure. • Understanding of the PA catheter (see Fig. 72-1 ), lumens Right heart preload is increased in right , right and ports, and the location of the PA catheter in the heart ventricular infarction, tricuspid regurgitation, pulmonary and PA (see Fig. 72-2 ) is needed. , and fl uid overload. Right heart preload is • Pressure transducer systems (see Procedure 75) should be decreased in hypovolemic states. understood. • Left heart preload refers to the pressure in the LV at the • Competence in the use and clinical application of hemo- end of diastole and is measured by the PAOP. When LV dynamic waveforms and values obtained with a PA cath- preload or end-diastolic volume increases, the muscle eter including assessing normal and abnormal waveforms fi bers are stretched. The increased tension or force of and values for (RAP), pulmonary contraction that accompanies an increase in diastolic artery pressure (PAP), and occlusion fi lling is called the Frank-Starling law. According to the pressure (PAOP) is needed. PAOP may also be termed Frank-Starling law the heart adjusts its pumping ability pulmonary artery wedge pressure (PAWP). to accommodate various levels of venous return. Note: • Knowledge of vasoactive and inotropic medications and In patients with advanced chronic LV dysfunction and their effects on cardiac function, ventricular function, coro- remodeled (spherical or globular shaped LV instead nary vessels, and vascular smooth muscles is needed. of the normal elliptical-shaped LV), the Frank-Starling • CO is defi ned as the amount of blood ejected by the left law does not apply. In these patients, muscle fi bers of the ventricle (LV) per minute and is the product of stroke heart are already maximally lengthened; as a result, the volume (SV) and (HR). It is measured in liters heart cannot respond signifi cantly to increased fi lling or per minute. stretch with increased force of contraction. • Afterload refers to the force the ventricular myocardial CO=× SV HR fi bers must overcome to shorten or contract. It is the force • Normal CO is 4 to 8 L/min. The four physiological factors that resists contraction. The amount of force the LV must that affect CO are preload, afterload, contractility, and overcome infl uences the amount of blood ejected into the heart rate. systemic circulation. Afterload is infl uenced by peripheral

553 554 Unit II Cardiovascular System

Figure 64-1 Systematic assessment of the determinants of cardiac output may assist the clinician in defi ning the etiological factors of cardiac output alteration more precisely. (From Whalen DA, Keller R: Cardiovascular patient assessment. In Kinney MR, et al, editors: AACN clinical reference for critical care nurse , ed 4, St. Louis, 1998, Mosby, 227-319)

(the force opposing blood fl ow within sympathetic neural stimulation, or decreased body tem- the vessels), systolic , systolic stress, and perature. Increased heart rate can be triggered by factors systolic impedance. Peripheral resistance is affected by that cause catecholamine release, such as hypoxia, hypo- the length and radius of the blood vessel, arterial blood tension, pain, or anxiety. The more rapid the heart rate, pressure, and venous constriction or dilation. The systolic the less time is available for adequate diastolic fi lling, force of the heart is increased in conditions that cause which can result in a decreased CO. Because multiple (increased afterload), including aortic factors regulate cardiac performance and affect CO, these stenosis, hypertension, or hyperviscosity of blood (e.g., factors must be assessed ( Fig. 64-1 ). polycythemia). The systolic force of the heart is decreased • adjusts the CO to an individual ’ s body size in conditions that cause or decrease the (square meter of body surface area), making it a more viscosity of blood (e.g., ). Right ventricular after- precise measurement than CO. load is calculated as pulmonary vascular resistance. Left • Refer to Table 64-1 for normal hemodynamic values and ventricular afterload is calculated as systemic vascular calculations. resistance. • At the bedside, CO measurements are obtained through a • Contractility is defi ned as the ability of the myocardium PA catheter via the intermittent bolus CO method or the to contract and eject blood into the pulmonary or systemic continuous CO method. vasculature. Contractility is increased by sympathetic • Thermodilution CO measures right ventricular outfl ow; neural stimulation, and the release of calcium and norepi- therefore, intracardiac right-to-left shunts, as well as tri- nephrine is decreased by parasympathetic neural stimula- cuspid and pulmonic valve insuffi ciency, can result in tion, acidosis, and hyperkalemia. Contractility and heart inaccurate measurements.3,38 rate can be infl uenced by neural, humoral, and pharmaco- • The thermodilution cardiac output (TDCO) method: an logical factors. injectate solution of a known volume (10 mL) and tem- • In addition to stroke volume, CO is affected by heart rate. perature (room or cold temperature) is injected into the Normally, of the parasympathetic and sympathetic right atrium (RA) through the proximal port of the PA nervous system regulate heart rate through specialized catheter. The injectate exits the catheter into the RA, cardiac electrical cells. Heart rate and rhythm are infl u- where it mixes with blood and fl ows through the RV to enced by neural humoral, and pharmacological factors. the PA. A thermistor located at the tip of the PA catheter Decreased heart rate can be the result of factors such as detects the change in blood temperature as the blood increased parasympathetic neural stimulation, decreased passes the tip of the catheter in the PA. The CO is 64 Cardiac Output Measurement Techniques (Invasive) 555

TABLE 64-1 Hemodynamic Parameters Parameters Calculations Normal Value Body surface area (BSA) Weight (kg) × height (cm) × 0.007184 Varies with size (range = 0.58–2.9 m2 ) CO HR × SV 4–8 L/min Stroke volume (SV) CO × 1000 / HR 60–100 mL/beat Stroke volume index (SVI) SV / BSA 30–65 mL/beat/m 2 Cardiac index (CI) CO / BSA 2.5–4.5 L/min/m 2 Heart rate (HR) 60–100 beats/min Preload (CVP) or RAP 2–6 mm Hg Left atrial pressure (LAP) 4–12 mm Hg Pulmonary artery diastolic pressure (PADP) 5–15 mm Hg PAOP 4–12 mm Hg RVEDP 0–8 mm Hg LVEDP 4–10 mm Hg Afterload Systemic vascular resistance (SVR) MAP − CVP/RAP × 80 / CO 900–1400 dynes/s/cm − 5 SVR index (SVRI) MAP − CVP/RAP × 80 / CI 2000–2400 dynes/s/cm − 5 /m2 Pulmonary vascular resistance (PVR) PAMP − PAOP × 80 / CO 100–250 dynes/s/cm − 5 PVR index (PVRI) PAMP − PAOP × 80 / CI 255–315 dynes/s/cm − 5 /m2 Systolic blood pressure 100–130 mm Hg Contractility (EF): Left LVEDV × 100 / SV 60–75% Right RVEDV × 100 / SV 45–50% Stroke work index: Left SVI (MAP − PAOP) × 0.0136 50–62 g-m/m2 /beat Right SVI (MAP − CVP) × 0.0136 5–10 g-m/m2 /beat Pressures: MAP + 1 − 70–105 mm Hg DBP 3 (SBP DBP) PAMP + 1 − 9–16 mm Hg PADP 3 (PASP PADP)

CO, Cardiac output; DBP, diastolic blood pressure; MAP, ; LVEDP, left ventricular end-diastolic pressure; RVEDP, right ventricular end-diastolic pressure; PAMP, pulmonary artery mean pressure; PAOP, pulmonary artery occlusion pressure; LVEDV, left ventricular end-diastolic volume; RVEDV, right ventricular end-diastolic volume; PASP, pulmonary artery systolic pressure; PADP, pulmonary artery diastolic pressure; SBP, systolic blood pressure. Adapted from Tuggle D: Optimizing : Strategies for fl uid and medication titration in . In Carlson K, editor: AACN advanced critical care nursing , St Louis, 2009, Saunders, 1106; and Ahrens T: Hemodynamic monitoring, Crit Care Nurs Clin N Am 11:19-31, 1999.

calculated as the difference in temperatures on a time the second thermistor improved accuracy compared versus temperature curve. with Fick CO measurements and also improved preci- • CO can be calculated from PA catheters with two types of sion or repeatability of CO measurements in both cold thermistors: and room temperature.4,26 In one study, cold injectate ❖ A single thermistor has one inline temperature sensor had excellent precision with the standard single- near the tip of the catheter that lies in the PA when in thermistor PA catheter. Researchers concluded that proper position. the dual-thermistor PA catheter provided the greatest ❖ A dual thermistor has two inline temperature sensors, benefi t in decreasing measurement variability when one in the right atrium/superior caval (immediately room temperature injections were used to measure CO.4 above the injectate port opening) and one near the tip of • The change in temperature over time is plotted as a curve the catheter (same position as single thermistor). and displayed on the bedside monitor screen. CO is math- Because a temperature sensor is located in the right ematically calculated from the area under the curve and is atrium, there is no need to enter a “correction factor” or displayed digitally and graphically on the monitor screen “computation constant” into the computer to account ( Fig. 64-2 ). The area under the curve is inversely propor- for the loss in thermal indicator (heat) from the hub of tional to the rate of blood fl ow. Thus a high CO is associ- the RA injectate port to the RA. Investigators found that ated with a small area under the curve, whereas a low CO 556 Unit II Cardiovascular System

B

Figure 64-2 A, Examining cardiac output curves to establish reliability of values. B, Normal cardiac output curve with rapid upstroke and smooth progres- A sive decrease in temperature sensing. (B: From Ahrens T: Hemodynamic monitor- ing, Crit Care Nurs Clin North Am 11[1]:28, 1999.)

is associated with a large area under the curve (Fig. 64-3, proximal lumen port). When a PA catheter is properly A ). placed, the thermal fi lament section of the catheter is • The thermistor near the distal tip of the catheter detects located in the RV. This fi lament emits a pulsed low heat the temperature change and sends a signal to the CO energy signal in a 30- to 60-second pseudorandom computer and bedside monitor. The computer calculates binary (on/off) sequence, which allows blood to be the CO with the modifi ed Stewart-Hamilton equation, and heated and the heat signal adequately processed over the CO number is displayed on the monitor screen. The time as blood passes through the ventricle. A bedside average result of three to fi ve measurements is used to computer constructs thermodilution curves detected determine CO. from the pseudorandom heat impulses and measures • Accuracy of TDCO is dependent on adequate mixing of CO automatically. The computer screen displays blood and injectate, forward blood fl ow, steady baseline digital readings updated every 30 to 60 seconds that temperature in the PA, and appropriate procedural tech- refl ect the average CO of the preceding 3 to 6 minutes. nique. 3,19 In addition, loss of thermal indicator (heat), The CCO eliminates the need for fl uid boluses, reduces respiratory artifact, and hemodynamic instability can contamination risk, and provides a continuous CO cause variability from one injection to another.19,29,32 trend. 1,2,32 • Traditionally cardiac output measurements have been rec- ❖ Because the CCO computer constantly displays and ommended to be performed at end-expiration, though frequently updates the CO, treatment decisions can be there is research suggesting that this is not necessary. 20,24 expedited. Derived hemodynamic calculations (e.g., Additional research is needed. cardiac index and systemic vascular resistance) can be • Commercially available closed-system delivery sets (CO- obtained with greater frequency, thereby providing up- Set, Edwards Lifesciences, LLC, Irvine, CA) can be used to-time information in assessment of response to thera- with both cold and room temperature injectate ( Figs. 64-4 pies that affect hemodynamics.1 and 64-5 ). • CCO has been compared with TDCO, transesophageal • The Continuous Cardiac Output (CCO) method proceeds Doppler scan technique, and aortic transpulmonary tech- as follows: nique to determine its precision. Study results all show ❖ Continuous measurement of CO can be performed small bias, limits of agreement, and 95% confi dence without the need for injected fl uid. The CO can be limits, refl ecting that CCO provides accurate measure- obtained with a heat-exchange CO catheter. This cath- ment of CO and is a reliable method.1,2,6,22,32,39,50 eter has a membrane that allows for heat to exchange • Adequate mixing of blood and indicator (heat) is neces- with blood in the right atrium. sary for accurate CCO measurements. Conditions that ❖ The PA catheter with CCO capability contains a 10-cm prevent appropriate mixing or directional fl ow of the indi- thermal fi lament located close to the injection port cator or blood include intracardiac shunts or tricuspid (15 to 25 cm from the tip of the catheter, near the regurgitation. 64 Cardiac Output Measurement Techniques (Invasive) 557

Figure 64-3 A, Variations in the normal cardiac output curve seen in certain clinical conditions. B, Abnormal cardiac output curves that produce an erroneous cardiac output value. (From Urden LD, Stacy KM, Lough ME: Critical care nursing: Diagnosis and management, ed,7, St Louis, 2014, Mosby.)

• The CCO method is based on the same physiological prin- measurements are available, but large infusions of fl uid ciple as the TDCO method (indicator-dilution technique). are discouraged.8,15 • The TDCO method uses a bolus of injectate as the indica- • Because bolus injections are not needed with the CCO tor for measurement of CO. The CCO method uses heat method, the prevalence of user error is theoretically reduced.14 signals produced by the thermal fi lament as the indicator. • The CCO catheter can be used to obtain both CCO and The CCO computer provides a time-averaged rather than TDCO measurements. instantaneous CO reading. CCO values are infl uenced by • The CCO does not refl ect acute changes in CO values the same principles as TDCO. because the updated value on the monitor display is an • The heated thermal fi lament has a temperature limit to a average of 3 to 6 minutes of data. A delay of approxi- maximum of 44 °C (111.2 °F). When calibrated by the mately 10 or more minutes to detect a change of 1 L/min manufacturer, CCO computers produce reliable calcula- in CO may occur. When monitoring a patient with an tions within a temperature range of 30 to 40 °C (86 to unstable condition that is being aggressively treated with 104 °F) or 31 to 43 °C (87.8 to 109.4 °F). An error message medication or other therapies, one should be aware of the appears if the temperature in the PA is out of range. Follow delay in data displayed. manufacturer ’ s guidelines. • Infusions through proximal lumens should be limited EQUIPMENT to maintenance of patency of the lumen. Concomitant infusions through the proximal lumen can theoretically • Nonsterile gloves affect CCO measurements by altering the PA temperature. • Cardiac monitor Studies have shown that such infusions can cause varia- • Hemodynamic monitoring system (see Procedure 75 ) tions in TDCO measurements.19,47 • PA catheter (in place) • To date, no published data describing the effect of concur- • CO computer or module rent central line infusions on the accuracy of CCO • Connecting cables 558 Unit II Cardiovascular System

Vigilance II monitor Sterile injectate solution (user supplied)

Edwards Lifesciences Lid Swan-Ganz thermodilution catheter Thermistor Distal Model Balloon Mounting lumen tab EDSL 70CC2 container Non-vented supplied cable IV spike Thermistor Snap clamp Inline placed connector To IV/pressure Flow through cooling tube monitoring housing Injectate delivery tubing Cooling Tubing coils coil 10 cc syringe

Check valve Thermal tubing Proximal Balloon injectate inflation hub User supplied Temperature Model 93-521 cooling valve 3-way stopcock & probe container mounting bracket continuous flush device (supplied separately) Figure 64-4 Closed injectate delivery system. Cold temperature injectate. (From Edwards Life- sciences, LLC, Irvine, CA.)

Additional equipment, to have available as needed, includes ception of cold solution and may decrease anxiety associ- the following: ated with the procedure . • Bolus thermodilution ❖ Injectate temperature probe ❖ Injectate solution PATIENT ASSESSMENT AND ❖ 10-mL prefi lled syringes PREPARATION ❖ Injectate solution bag with intravenous (IV) tubing and three-way Luer-Lok stopcock Patient Assessment ❖ Closed CO injectate system • Assess the patient ’ s history of medication therapy, includ- ❖ Ice (for cold injectate only) ing medication allergies, recent bolus therapies, and current ❖ Nonvented caps for stopcocks medications. Rationale: Medications can infl uence CO • Setup for CCO measurements. • Syringe holder or automatic injector device • Assess the patient ’ s medical history for the presence of • Printer coronary artery disease, , and left or • Dispensing port right ventricular dysfunction. Rationale: Medical history provides baseline information regarding cardiovascular PATIENT AND FAMILY EDUCATION performance. • Assess current intracardiac waveforms (e.g., PAP, RAP, • Explain the procedure for CO and the reason for its mea- PAOP). Rationale: This assessment ensures the PA cath- surement. Include expectations related to sensations during eter is positioned properly. the procedure. (The patient should not experience pain or • Assess the patient ’ s vital signs, fl uid balance, heart and lung discomfort.) Rationale: Explanation decreases patient sounds, skin color, temperature, level of consciousness, and family anxiety. Preparatory information of sensations peripheral , cardiac rate and rhythm, and hemody- decreases patient fear of the impending procedure. namic values. In patients with advanced systolic heart • Explain the monitoring equipment involved, the frequency failure, assess for pulsus alternans (alternating strong and of measurements, and the goals of therapy. Rationale: weak pulses). Rationale: Clinical information provides Explanation encourages the patient and family to ask ques- data regarding blood fl ow and tissue . Abnormali- tions and voice specifi c concerns about the procedure. ties can infl uence the variability of CO measurements. • Explain any potential variations in temperature the patient • Ensure that no medication is being infused via the proxi- may or may not experience if a cold injectate is used. mal port. Rationale: Medications infused via this port will Rationale: This explanation acknowledges the varying be bolused into the patient if the bolus CO method is physical responses to the injectate and the possible per- utilized. This is not an issue with the CCO method. 64 Cardiac Output Measurement Techniques (Invasive) 559

Vigilance II monitor Sterile injectate solution (user supplied)

Edwards Lifesciences Swan-Ganz thermodilution Thermistor catheter Non-vented Distal IV spike Balloon lumen

70CC2 cable Snap clamp

Thermistor connector Flow through To IV/pressure housing monitoring

10 cc syringe Injectate delivery tubing Check valve

Balloon User supplied inflation Temperature 3-way stopcock & valve probe continuous flush device Figure 64-5 Closed injectate delivery system. Room temperature injectate. (From Edwards Life- sciences, LLC, Irvine, CA.) • Ensure that the patient and family understand preprocedural Patient Preparation information. Answer questions as they arise, and reinforce • Verify that the patient is the correct patient using two information as needed. Rationale: Understanding of previ- identifi ers. Rationale: Before performing a procedure, the ously taught information is evaluated and reinforced. nurse should ensure the correct identifi cation of the patient • Assist the patient to the supine position. Rationale: CO for the intended intervention. measurements are most accurate in the supine position.

Procedure for Measurement of Cardiac Output With the Closed or Open Thermodilution Method Steps Rationale Special Considerations 1 . HH 2 . PE 3. Select the injectate delivery Both systems are reliable. 32,37,40 A closed system may eliminate cost system: open or closed method. and time expenditures of individual (Level B * ) syringe preparation. The closed system has infection control benefi ts because of reduction of multiple entries into the system.34 4. Select cold or room temperature Room temperature injectate may be The acceptable temperature range for injectate. used for most patients. Research on cold and room temperature injectate ( Level B * ) room temperature versus cold varies by system (manufacturer). injectate supports the accuracy of Generally, room temperature is either method.7,9,12,23,41,42,45,48 18–25 °C, and cold is 0–12 °C. Cold injectate may improve the accuracy of CO measurement for patients with low or high CO. 45

* Level B: Well-designed, controlled studies with results that consistently support a specifi c action, intervention, or treatment.

Procedure continues on following page 560 Unit II Cardiovascular System

Procedure for Measurement of Cardiac Output With the Closed or Open Thermodilution Method—Continued Steps Rationale Special Considerations 5. Select the injectate bolus amount An injectate of 10 mL may be used Volumes of 5 mL may necessitate (generally 10 mL). (Level B * ) for most patients.12,31,36 additional injections because of greater variability of individual measurements.31 6. Connect the CO cable to the PA Prepares the system. catheter. 7. Select the computation constant The computation constant is a Carefully select the correct consistent with the type and size correction factor determined by the computation constant for the type of the PA catheter, injectate catheter manufacturer that corrects and catheter size, injectate volume, volume, and injectate for the gain of indicator (heat) that and cold or room temperature temperature. Confi rm the injectate occurs as the injectate moves injectate. delivery system. through the catheter from the hub Confi rm the setting on the CO of the injectate port to the injection computer/monitor. port opening in the RA. Recheck the computation constant The catheter manufacturer provides a before each series of CO table to determine the correct measurements. computation constant. Follow the manufacturer guidelines. The computation constant must be accurate for valid and reliable CO measurements. 8. Connect the CO computer to the Supplies the energy source. power source if it is a stand-alone device or turn on the CO computer or module. 9. Note the temperature of the The injectate temperature should be at Follow manufacturer ’ s injectate (on the computer or least 10 °C less than the patient’ s recommendations. monitor screen). core temperature.27 10. Position the patient supine, with Studies of patients in the supine CCO values may be accurate with the the head of the bed elevated no position with the head of bed fl at or patient ’ s head of the bed elevated to more than 20 degrees. elevated up to 20 degrees have not 45 degrees. 11,17 (Level B * ) shown signifi cant differences in The patient ’ s medical condition and TDCO measurements.18,21,25,50 level of instability may determine Consistency in patient position may positioning. increase stability in consecutive CO Position should be documented and readings. communicated. Consistent positioning when obtaining CO measurements over time decreases measurement variability. Cautiously use CO values obtained from lateral positions. The lateral recumbent position increases variability in CO measurements.10,51 11. Verify the position of the PA Proper positioning of the PA catheter Improper positioning of the PA catheter by assessing both the RA ensures that the distal thermistor is catheter tip may result in false and PA waveforms for proper located in the PA. values.14,15,22,28 waveform contours. The distal thermistor sensor calculates the time-temperature data. Excessive coiling of the PA catheter in the RA or RV can result in poor positioning of the distal thermistor in relation to the injectate port.22

*Level B: Well-designed, controlled studies with results that consistently support a specifi c action, intervention, or treatment. 64 Cardiac Output Measurement Techniques (Invasive) 561

Procedure for Measurement of Cardiac Output With the Closed or Open Thermodilution Method—Continued Steps Rationale Special Considerations 12. Observe the patient ’ s cardiac rate A rapid heart rate or dysrhythmias and rhythm. may decrease CO and lead to variability in CO measurements. 13. If possible, consider restricting TDCO measurements obtained during infusions delivered through the administration of other infusions introducer or other central lines. can cause variability in CO (Level C * ) measurements (by as much as 40% higher). 19,50 14. Remove PE and discard used Reduces the transmission of supplies. microorganisms; Standard Precautions. 15. HH

* Level C: Qualitative studies, descriptive or correlational studies, integrative reviews, systematic reviews, or randomized controlled trials with inconsistent results.

Procedure for Closed-Method of Syringe Preparation and Cardiac Output Determination Steps Rationale Special Considerations Follow Steps 1–13 of the Procedure for Measurement of CO 1 . HH 2 . PE 3. Obtain the injectate solution 4. Aseptically connect the IV tubing Prepares the system. to the injectate solution. 5. Hang the IV injectate solution on Eliminates air from the tubing. an IV pole; prime the tubing. 6. Remove the sterile cap from the Prepares for the injectate connection. proximal lumen of the PA catheter. 7. Connect the injectate tubing to Connects the injectate solution to the the proximal lumen of the PA PA catheter. catheter via a three-way Luer-Lok stopcock (see Figs. 64-4 and 64-5 ). 8. Connect the injectate syringe to The syringe is used for solution Connect the system so that the CO the three-way stopcock (see Figs. injection. syringe is in a straight line with the 64-4 and 64-5 ). PA catheter to decrease resistance with injection of solution. To ensure accurate readings, when using a multiport PA catheter the proximal port should be used rather than the venous infusion port. 38 9. Connect the inline temperature Measures the injectate temperature. Verify temperature. probe (see Figs. 64-4 and 64-5 ). 10. If using cold injectate, set up the If using cold injectate, cool the Refer to manufacturer ’ s cold injectate system (e.g., injectate solution to 0–12 °C recommendations. CO-Set closed injectate system; (32–53 °F). Cold injectate may be proarrhythmic see Fig. 64-4 ). in some patients.35,46 11. Turn the stopcock so that it is Prepares for injection. open to the injectate solution (closed to the patient) and withdraw 10 mL of the injectate solution into the syringe. Procedure continues on following page 562 Unit II Cardiovascular System

Procedure for Closed-Method of Syringe Preparation and Cardiac Output Determination—Continued Steps Rationale Special Considerations 12. Turn the stopcock so that it is Minimal handling (< 30 seconds) of Syringe holders or automatic injector closed to the injectate solution the syringe is recommended to devices are available and can be and open to the patient. Support avoid thermal indicator variation used. the stopcock with the palm of the that may introduce error into the nondominant hand to aid in CO calculation.5,8,28 injectate administration. 13. Activate the CO computer, and The CO computer or module needs to Follow manufacturer ’ s guidelines. wait for the “ready” message. be ready before injection of solution. 14. Before administering the bolus An abnormal baseline may increase Patients with advanced systolic heart injectate, observe for a steady variability in CO measurements and failure (low ejection fraction) are baseline temperature (e.g., the introduce error. 1 more susceptible to a wavering line before the CO curve begins initial baseline from unstable PA should be fl at without blood temperature. undulations) on the monitor screen (see Figs. 64-2B and 64-3A ). 15. Observe the patient ’ s respiratory End expiration is defi ned as the phase Follow institution standard. pattern. Prepare to begin of the respiratory cycle preceding administering the injectate at end the start of inspiration. expiration to decrease variance in Signifi cant variations in transthoracic CO measurements from the pressure during respiration can respiratory cycle affect CO by altering venous (Level C * ) return.20,40,44,45 16. Administer the bolus injectate Prolonged injection time may result A prolonged injection time interferes rapidly and smoothly in 4 in false low CO. with time and temperature seconds or less. Rates of 2–4 seconds for injection of calculations. 5–10 mL of injectate yield accurate One respiratory cycle is generally < 4 results. 5,13,43 seconds. One ventilation cycle on a ventilator is generally 4 seconds. 17. Assess the CO curve and value The CO curve must be a normal Normal CO is 4–8 L/min. on the monitor screen (see Figs. curve. Abnormal CO curves may also 64-2 and 64-3 ). A normal curve starts at baseline provide information about the (baseline must be a straight, fl at, patient ’ s clinical condition, such as nonwavering line) with a smooth tricuspid valve regurgitation. upstroke and a gradual downstroke. If the CO curve is not normal, the CO measurement obtained from the injection should be discarded. Abnormal contours of the curve may indicate improper catheter position. An abnormal CO curve may represent technical error. 18. Repeat Steps 3 through 17 (up to Discard all CO measurements that do Allow 60 seconds between each CO three times total for cold injectate not have normal CO curves or have measurement to ensure consistency and up to fi ve times total for wandering baselines. and accuracy. room temperature injectate).

* Level C: Qualitative studies, descriptive or correlational studies, integrative reviews, systematic reviews, or randomized controlled trials with inconsistent results. 64 Cardiac Output Measurement Techniques (Invasive) 563

Procedure for Closed-Method of Syringe Preparation and Cardiac Output Determination—Continued Steps Rationale Special Considerations 19. Determine the CO measurement Determines accurate CO value.49 by calculating the average of three measurements within 10% of a middle (median) value. (Level E * ) 20. Return the proximal stopcock at Resumes RA monitoring. the RA lumen to the original position. 21. Continue infusions delivered Continues therapy. through the introducer or other central lines. 22. Observe the PA and RA Continues hemodynamic monitoring. waveforms on the monitor. 23. Remove PE and discard used Reduces the transmission of supplies in appropriate microorganisms; Standard receptacles. Precautions. 24. HH 25. Determine the hemodynamic Assesses cardiac performance and Compare the values with prior values calculations. hemodynamic status. and determine whether the plan of care requires alterations.

* Level E: Multiple case reports, theory-based evidence from expert opinions, or peer-reviewed professional orga nizational standards without clinical studies to support recommendations.

Procedure for Open Method of Syringe Preparation and Cardiac Output Determination Steps Rationale Special Considerations Follow Steps 1–13 of the Procedure for Measurement of Cardiac Output 1 . HH 2 . PE 3. Prepare syringes or obtain Prepares the injectate for CO Prefi lled syringes may decrease manufactured prefi lled syringes measurements. variability related to injectate for CO determination. volume. A. Clean the injectate port of the Manufactured prefi lled syringes may D5W IV bag with an alcohol be stored per manufacturer’ s wipe. recommendations. B. Apply a dispensing port to the Solutions drawn up should be used bag ’ s injectate port immediately for CO measurements. C. Aseptically withdraw the Solution drawn up in the clinical area injectate solution from the IV (as opposed to under a laminar fl ow bag into three to fi ve 10-mL area) that has no preservative, syringes and cap securely. should not be stored for later use. A dispensing port negates the use of needles and reduces the incidence of accidental needlesticks. 4. Cold injectate: Cool the syringes Iced slush is used to cool syringes. Place syringes in a bag in the in ice. container, not directly into the slush. Handling of a cold syringe causes warming and hampers validity of CO measurements.5,8,28 Cold injectate may be proarrhythmic.35,46 Procedure continues on following page 564 Unit II Cardiovascular System

Procedure for Open Method of Syringe Preparation and Cardiac Output Determination—Continued Steps Rationale Special Considerations 5. Remove the nonvented cap from Prepares the stopcock. the proximal lumen stopcock of the PA catheter. 6. Aseptically connect one of the Reduces the risk of introducing sterile CO injectate syringes onto microorganisms into the system. the proximal lumen stopcock of the PA catheter. 7. Turn the stopcock so that it is Prepares the system for injectate Minimize handling (< 30 seconds) closed to the fl ush solution and administration. open between the injectate syringe and the patient. Support the stopcock with the palm of the nondominant hand. 8. Connect the inline temperature Measures the injectate temperature. probe (see Figs. 64-4 and 64-5 ). 9. Activate the CO computer, and The CO computer or module needs to Follow manufacturer ’ s guidelines. wait for the “ready” message. be ready before injection of solution. 10. Before administering the bolus An abnormal baseline may increase Patients with advanced systolic heart injectate, observe for a steady variability in CO measurements and failure (low ejection fraction) are baseline temperature (e.g., the introduce error. 1 more prone to a wavering initial line before the CO curve begins baseline from unstable PA blood should be fl at without temperature. undulations) on the monitor screen (see Figs. 64-2B and 64-3A ). 11. Observe the patient ’ s respiratory End expiration is defi ned as the phase Follow institution standard. pattern. Prepare to begin of the respiratory cycle preceding administering the injectate at end the start of inspiration. expiration to decrease variance in Signifi cant variations in transthoracic CO measurements from the pressure during respiration can respiratory cycle affect CO by altering venous (Level C * ) return.20,40,44,45 12. Administer the bolus injectate Prolonged injection time may result A prolonged injection time interferes rapidly and smoothly in 4 in false low CO. with time and temperature seconds or less. Rates of 2–4 seconds for injection of calculations. 5–10 mL of injectate yield accurate One respiratory cycle is generally < 4 results. 5,13,43 seconds. One ventilation cycle on a ventilator is generally 4 seconds. 13. Assess the CO curve and value The CO curve must be a normal Normal CO is 4–8 L/min. on the monitor screen (see Figs. curve. Abnormal CO curves may also 64-2 and 64-3 ). A normal curve starts at baseline provide information about the (baseline must be a straight, fl at, patient ’ s clinical condition, such as nonwavering line) with a smooth tricuspid valve regurgitation. upstroke and a gradual downstroke. If the CO curve is not normal, the reading should be discarded. Abnormal contours of the curve may indicate improper catheter position. An abnormal CO curve may represent technical error.

* Level C: Qualitative studies, descriptive or correlational studies, integrative reviews, systematic reviews, or randomized controlled trials with inconsistent results. 64 Cardiac Output Measurement Techniques (Invasive) 565

Procedure for Open Method of Syringe Preparation and Cardiac Output Determination—Continued Steps Rationale Special Considerations 14. Repeat Steps 6 through 13 (up to Obtains CO measurements. Discard Allow 60 seconds between each CO three times total for cold injectate all CO measurements that do not measurement to ensure consistency and up to fi ve times total for have normal CO curves or have and accuracy. room temperature injectate). wandering baselines. Asepsis is essential as the stopcock is turned and syringes are exchanged between CO measurements. 15. Determine the CO measurement Determines accurate CO value.49 by calculating the average of three measurements within 10% of a middle (median) value. (Level E * ) 16. After the last injectate is Closes the system; maintains the completed: sterility of the system. A. Turn the right atrial lumen stopcock of the PA catheter so that the system is open between the patient and the transducer. B. Aseptically remove the last injectate syringe. C. Place a new, sterile, nonvented cap on the stopcock port. 17. Observe the PA and RA Continues hemodynamic monitoring. waveforms on the monitor. 18. Remove PE and discard used Removes and safely discards used supplies in appropriate supplies. receptacles. 19. HH 20. Determine hemodynamic Assesses cardiac performance and Compare values with prior values and calculations. hemodynamic status. determine whether the plan of care requires alterations.

* Level E: Multiple case reports, theory-based evidence from expert opinions, or peer-reviewed professional orga nizational standards without clinical studies to support recommendations.

Procedure for Measurement of Cardiac Output With Continuous Cardiac Output Method Steps Rationale Special Considerations 1 . HH 2 . PE 3. Turn on the CO computer or Provides energy. module. 4. Connect the CO cable to the PA Prepares equipment. catheter. 5. Observe PA waveforms (e.g., RA, Determines whether the PA catheter is The thermal fi lament should fl oat PA). in the correct position. freely in the RV to prevent the loss of indicator (heat) into the cardiac tissue. If the loss of indicator occurs, the CO value is overestimated, giving erroneous readings. Follow manufacturer ’ s guidelines. The device may not measure CO if the PA catheter is malpositioned. Procedure continues on following page 566 Unit II Cardiovascular System

Procedure for Measurement of Cardiac Output With Continuous Cardiac Output Method—Continued Steps Rationale Special Considerations 6. Position the patient supine with CCO measurements are most accurate Additional studies are needed to the head of the bed elevated up to in a supine position, but head-of- determine the effect of patient body 45 degrees. (Level C * ) bed angle can be varied for position on CCO measurements. comfort, between 0 and 45 Document body position at the time degrees.16 of hemodynamic data collection. 7. Check the heat signal indicator CCO systems assess the quality of CCO monitors provide messages for on the CO computer or module the measured thermal signal. troubleshooting signal-to-noise ratio per manufacturer ’ s Relationships are in response to interferences. recommendations. thermal noise or signal-to-noise Refer to manufacturer ratio. recommendations. Technologic advances suppress the effects of blood thermal noise.52 8. Note that the CCO values refl ect CCO measurements are averaged over CCO values are updated every 30–60 an average of the preceding 3–6 the preceding 3–6 minutes and are seconds. Continuous data collection minutes of data collection. not individual measurements. refl ects phasic changes in the respiratory cycle. CCO measurements are not timed to the respiratory cycle. 9. When documenting CCO values, Provides data regarding hemodynamic also document other status. hemodynamic fi ndings. 10. Compare the CCO value with the CCO is a global assessment parameter The CCO method eliminates many of patient ’ s current clinical status and must be appreciated as part of the potential user-related and and hemodynamic fi ndings. the patient ’ s total hemodynamic technique-related errors associated profi le at a given time. with the intermittent bolus CO method. Research shows clinically acceptable correlation between the TDCO technique and the CCO method in the steady state. 1,6,30,32,33 Future studies are needed to determine effi cacy in patients in various phases of acute hemodynamic instability and in specifi c patient populations. Also, the effects of changes in positioning need to be studied further, especially in patients with structural or functional heart damage. 11. Note: The CCO catheter system can be used to obtain TDCO by following Steps 1–13 of the Procedure for Measurement of Cardiac Output with the Closed or Open Thermodilution Method CO and then follow the steps for either the Closed or Open Method of Syringe Preparation and Cardiac Output Determination. 12. Remove PE and discard used Removes and safely discards used supplies in appropriate receptacles. supplies. 13. HH

* Level C: Qualitative studies, descriptive or correlational studies, integrative reviews, systematic reviews, or randomized controlled trials with inconsistent results. 64 Cardiac Output Measurement Techniques (Invasive) 567

Expected Outcomes Unexpected Outcomes • Accurate CO measurement are obtained • Inability to accurately measure CO • Hemodynamic profi le and derived parameters are • Erroneous readings because of technical, equipment, obtained with accuracy, whether through the or operator error continuous or intermittent method • Contamination of the system • Sterility and patency of the PA catheter is maintained • Occlusion of the proximal PA lumen

Patient Monitoring and Care Steps Rationale Reportable Conditions These conditions should be reported if they persist despite nursing interventions. 1. Maintain patency of the PA PA catheter patency is essential for • Inability to maintain PA catheter catheter. accurate monitoring. patency 2. Monitor RA and PA waveforms for Proper placement determines accurate • Abnormal RA or PA waveforms or confi rmation of proper catheter hemodynamic and CO values position. measurement. 3. Maintain the sterility of the PA Reduces the risk for catheter-related • Fever, site redness, drainage, or catheter. infections. symptoms consistent with infection 4. Calculate cardiac index, systemic Determines cardiac performance and • Abnormal cardiac index, systemic vascular resistance, and other current hemodynamic status. vascular resistance, or other parameters as prescribed or hemodynamic values indicated. 5. Monitor vital signs and respiratory Changes in vital signs or respiratory • Changes in vital signs status hourly and as indicated. status may indicate hemodynamic • Changes in respiratory status compromise. 6. Include the fl uid volume used in Additional volume given • Signs or symptoms of fl uid the TDCO in the patient ’ s total intermittently should be included in overload (e.g., respiratory distress, fl uid volume intake. the total intake for accurate fl uid crackles, increased PADP or PAOP, volume assessment. elevated , new or worsening S3 gallop, worsening edema) 7. Assess the patient ’ s response to Hemodynamic monitoring may • Signifi cant worsening or therapies. expedite treatment decisions. improvement in CO and hemodynamic parameters. • Pulmonary vascular resistance (PVR) and systemic vascular resistance (SVR) 8. If using a closed system delivery Reduces the incidence of infection. set (CO set), change the system components (tubing, syringe, stopcocks, and IV solution) every 96 hours with the hemodynamic monitoring system (see Procedure 75). Procedure continues on following page 568 Unit II Cardiovascular System

Documentation Documentation should include the following: • Patient and family education nitroprusside, nicardipine, or nesiritide]), and • CO, cardiac index, SVR, volume indicators (PAOP and intravascular volume (e.g., diuretics) RAP) • Signifi cant medical therapies or nursing interventions • CO curves that affect CO (e.g., intraaortic balloon pump or • Baseline PA blood temperature ventricular assist device therapies, volume expanders, • Continuous or intermittent bolus method position changes), vascular resistance, or • Volume and temperature of injectate intravascular volume (e.g., sedation, blood/blood • Concurrent headrest elevation, vital signs, and products, headrest elevation, fl uid restriction, sodium hemodynamic measurements restriction) • Titration or administration of medications that affect • Unexpected outcomes contractility (e.g., dobutamine, , , • Additional interventions, including psychosocial or ), vascular resistance (e.g., intravenous emotional/psychiatric interventions that might nitrates or arterial vasodilator therapy [e.g., infl uence hemodynamic trends

References and Additional Readings For a complete list of references and additional readings for this procedure, scan this QR code with any freely available smartphone code reader app, or visit http://booksite.elsevier.com/9780323376624 .