Linseis Optical Dilatometer and Heating Microscope Brochure

Total Page:16

File Type:pdf, Size:1020Kb

Linseis Optical Dilatometer and Heating Microscope Brochure THERMAL ANALYSIS Optical Dilatometer DIL L74 Heating Microscope Since 1957 LINSEIS Corporation has been deliv- ering outstanding service, know how and lead- ing innovative products in the field of thermal analysis and thermo physical properties. Customer satisfaction, innovation, flexibility and high quality are what LINSEIS represents. Thanks to these fundamentals our company enjoys an exceptional reputation among the leading scientific and industrial organizations. LINSEIS has been offering highly innovative benchmark products for many years. The LINSEIS business unit of thermal analysis is involved in the complete range of thermo Claus Linseis analytical equipment for R&D as well as qual- Managing Director ity control. We support applications in sectors such as polymers, chemical industry, inorganic building materials and environmental analytics. In addition, thermo physical properties of solids, liquids and melts can be analyzed. LINSEIS provides technological leadership. We develop and manufacture thermo analytic and thermo physical testing equipment to the high- est standards and precision. Due to our innova- tive drive and precision, we are a leading manu- facturer of thermal Analysis equipment. The development of thermo analytical testing machines requires significant research and a high degree of precision. LINSEIS Corp. invests in this research to the benefit of our customers. 2 German engineering Innovation The strive for the best due diligence and ac- We want to deliver the latest and best tech- countability is part of our DNA. Our history is af- nology for our customers. LINSEIS continues fected by German engineering and strict quality to innovate and enhance our existing thermal control. analyzers. Our goal is constantly develop new technologies to enable continued discovery in Science. 3 OPTICAL DILATOMETER The Optical Research Dilatometer L74 was de- sample is not burdened with any force. For veloped to meet the demanding applications of soft samples or samples that melt during the the glass, ceramics, metal and energy industry. measurement , this leads to a result, which is not A high resolution CCD camera enables a visual distorted by the contact pressure. real time analysis of the sample expansion, eit- her as single frame or as video sequences. The big advantage of this method is that the DSC - Differential Scanning Calorimeter Option DSC - Differential Scanning Calorimeter Option for Optical Dilatometer 4 Several correction and analysis features are in- • determination of viscosity curves of glasses corporated into the LINSEIS Evaluation Software. • analysis of sintering warpage (sintering of The unique horizontal design enables most de- ceramic tapes and multilayers; co-sintering of manding applications. The special solid-liquid metal-ceramic bi- and multi-layers) adapter allows expansion / volume measure- • analysis of the infiltration behavior of metallic ments of solids, liquids and solid – liquid phase melts in ceramic materials (synthesis by melt- transitions. There is also a special sample holder infiltration) for measuring rigid foils available, which avoids • analysis of melt filtration by porous ceramic measurement errors due to pushrod forces like materials in a classical dilatometer. • investigation of the coating behaviour • analysis of the brazing behavior and brazing Application tests, solder developement, glas-metal or • analysis of sintering processes (densification ceramic sealing behavior as a function of temperature and time sintering kinetics) Industries • analysis of the thermal expansion behavior • glass (thermal expansion coefficient) • metal • investigation of the wetting- and spreading- • enamel coatings behaviour of metallic- and glass-melts, slags • ceramics and ashes on different substrates (metal, cera- • energy mic, glass) • optical determination of the drop shape and contact-angle determination of surface ten- sion up to high temperatures • analysis of the softening - and melting behavi- or (melting kinetic) • investigation of corrosion behavior (metallic- and glass-melts, slags etc. on refractory mate- rials) Dilatometer • analysis of the ash melting behavior 5 OPTICAL DILATOMETER Sintering The investigation of sintering processes using that the sintering process can take place wit- pushrod dilatometers has one disadvantage: hout any influence. If you compare the results There is always a force in z-direction that can of the same substance that was sintered on an cause a certain sinter direction. The optical di- optical and on a pushrod dilatometer, there can latometer instead is able to measure a sinter be big differences because of this effect. process completely contact free and guarantees Sintering behaviour of ceramic tapes (ZrO2) 1.00 0.95 0 0.90 L / 0.85 0.80 0 200 400 600 800 1000 1200 1400 1600 Temperature [°C] 1.00 0.95 0 0.90 -0.001 0 )/dt 0 L / 0.85 -0.002 d(L/L 0.80 -0.003 0.75 -0.004 -0.005 240 260 280 300 320 340 360 380 Time [min] 6 HEATING MICROSCOPE Applications • Ash fusion microscopy mass there is a pre norm DIN CEN/TS 15730 that can • Observation and analysis of sintering processes be used as well. It defines four characteristic tempe- • Contact angle determination ratures under oxidizing atmosphere. • Microscopy at high temperatures and under different atmospheres Functional Principle The principle of determination of the ash melting be- Ash melting microscope havior is based on a sample that is melted under de- Ash melting of a coal sample fined temperature ramps. The melting temperatures 1 2 for stone coal are up to 1600°C, for bio mass and wood they are at around 1200°C to 1300°C and for culm Ash melting Ash microscope containing bio mass they are even lower. The evalua- 3 4 tion is done automatically according to DIN 51730 by monitoring of a shadow profile of the sample where 1 – Softening temperature = edges are getting round the change of geometry is documented. By means of 2 – spheric temperature = round shape of height that is as big as its baseline operating geometric factors, the resulting characteri- 3 – half sphere temperature = half sphere shape of stic temperatures can be determined and be used for height that is half the size of its baseline 4 – flow temperature = sample is nearly liquid, has comparison of different coal qualities. only one third of the size of half sphere point General Fields of Application Solid ash fuels consist of inorganic compounds like si- • Measurements according to DIN 51730 (1984, licon oxide and alkali oxides. Due to the high diversi- 1998) / ISO 540-1995 ty of ash compositions it is nearly impossible neither • Characteristic temperatures plus start of sinte- to define a universal melting point nor to determine ring / sintering point a fuel specific melting point. The melting behavior of • Ash melting behavior ashes is heavily dependent on its chemical composi- • Sintering behavior tion. The melting procedure of ashes takes place in a • Dilatometric curves (e.g. shape, area) broader temperature range. Even if the slagging pro- • Softening- and melting behavior, moistening behavior blems of power plants are different from each other, • Viscosity curve the melting behavior of ashes is useful to get a qua- Suitable for analysis of coal ashes, bio ashes, litative analysis of the slagging behavior of different slags, as well as ceramic, enamel, clay ceramics, coals what gives a result that can be applied to the dental ceramics, grinding discs, special cera- corresponding boiler. The ash melting behavior of mics, fire proof ceramics, glass, steal, soldering coals is determined according to DIN 51730, for bio pastes, stainless steel and fluxing agents. 7 SOFTWARE Features -Software In respect to thermal and mechanical sample • Quenching Dilatometer Software treatment numerous different matematical • CHT / CCT / DCCT / TTT Diagrams functions can be selected. • Thermo Kinetics Software • User-friendly • Multi-methods analysis (DSC TG, TMA, DIL, etc.) Control and Evaluation • Zoom function Software • Online help menu • Software package compatible with Windows 7 • Report generator and 8 • Data export to MS Excel • Software for creation of CHT, CCT, DCCT and • Export and import of data ASCII TTT diagrams • Program capable of text editing • all necessary measuring parameter are inclu- • Data security in case of power failure ded in the menu structure • Thermocouple break protection • Unlimited number of forming steps during • Repetition measurements with minimum measurement parameter input • free choice of all control parameters • Evaluation of current measurement • Specification of temperature-time force-gradi- • Curve comparison up to 32 curves ent and strain rate and deformation degree • Storage and export of evaluations • Control rate input by user or selection of indus- • Programmable gas control try parameter, i.e. quenching rates based on • Statistical evaluation package T 8/5 times • Smoothing of total or partial measurement • Automatic force control on the basis of user • Tangent intersection determination (automa- defined functions tic or manual) • individual commentaries • Free scaling • heat up and cool down speeds The information of a thermo analytical measure- • end of the heating curve as well as duration of ment can be increased when using the broad holding temperature range of specialized Software. • programmable heating / cooling and isother- mal segments Software options • function menus are easy to handle • Specific Heat determination (Cp) • Graphical evaluation software with many func- • Rate
Recommended publications
  • Protocol for Use of Differential Scanning Calorimeter (NSF/Epscor Proteomics Facility @ Brown University)
    MicroCal VP-DSC Geoff Stetson/ Page Laboratory/ 2007 Protocol for Use of Differential Scanning Calorimeter (NSF/EPSCoR Proteomics Facility @ Brown University) 1. Equipment (photos) 2. Getting Started Degasser - Turn on the degasser. - Make sure the metal valve on the top of the lid is closed. - Set the temperature to 20-25°C. - Place a small stir bar in a couple of plastic vials. - Fill vials 2/3 full with Milli-Q water. - Place vials into slots on top of degasser. - Turn the stirrer speed knob to - 10 or 11 o’clock. - - Place the lid on firmly, and while pressing down turn on the degasser o (Note: If you flip the switch to on, the degasser will run until switched off. If the switch is flipped to timer, the degasser will run for eight minutes). - Degas the sample for 8-15 minutes. o Turn off the vacuum. - Open the metal valve on top of the lid in order to release the vacuum created underneath. - Remove the plastic vials. 1 MicroCal VP-DSC Geoff Stetson/ Page Laboratory/ 2007 DSC - Unscrew top. - Remove contents of of the sample cell and the reference cell using the glass filling syringe. o Insert the funnel into the top of the cell. o Slowly put the syringe into the cell until it gently touches the bar across the top of the funnel, then remove the liquid. Repeat. o (Note: Be careful when putting anything into the cells. The machine is extremely sensitive.) - Fill glass syringe with degassed Milli-Q water. - Rinse out both cells with the Milli-Q water (3x).
    [Show full text]
  • Soda Can Calorimeter
    Soda Can Calorimeter Energy Content of Food SCIENTIFIC SCIENCEFAX! Introduction Have you ever noticed the nutrition label located on the packaging of the food you buy? One of the first things listed on the label are the calories per serving. How is the calorie content of food determined? This activity will introduce the concept of calorimetry and investigate the caloric content of snack foods. Concepts •Calorimetry • Conservation of energy • First law of thermodynamics Background The law of conservation of energy states that energy cannot be created or destroyed, only converted from one form to another. This fundamental law was used by scientists to derive new laws in the field of thermodynamics—the study of heat energy, temperature, and heat transfer. The First Law of Thermodynamics states that the heat energy lost by one body is gained by another body. Heat is the energy that is transferred between objects when there is a difference in temperature. Objects contain heat as a result of the small, rapid motion (vibrations, rotational motion, electron spin, etc.) that all atoms experience. The temperature of an object is an indirect measurement of its heat. Particles in a hot object exhibit more rapid motion than particles in a colder object. When a hot and cold object are placed in contact with one another, the faster moving particles in the hot object will begin to bump into the slower moving particles in the colder object making them move faster (vice versa, the faster particles will then move slower). Eventually, the two objects will reach the same equilibrium temperature—the initially cold object will now be warmer, and the initially hot object will now be cooler.
    [Show full text]
  • T E M P E R a T U
    THE HIGH TEMPERATURE HEAT CONTENTS OP 0 MOLYBDENUM AND TITANIUM AND THE LOW TEMPERATURE HEAT CAPACITIES OP TITANIUM DISSERTATION Presented In Partial Fulfillment of the Requirements for the Degree Doctor of Philosophy in the Graduate School of The Ohio State University By CHARLES WILLIAM KOTHEN, B.A. // The Ohio State University 1952 Approved By: Adviser i TABLE OF CONTENTS E&gft INTRODUCTION ............................ 1 THEORETICAL ............................. 3 HISTORICAL .............................. 6 PART I The High Temperature Heat Contents of Molybdenum and Titahium ................... 13 Introduction ...................... 13 Apparatus .......................... 14 Measurements and Calculations ........ 32 Errors ... ......................... 45 Experimental Results ............... 49 PART II Low Temperature Heat Capacity of Titanium.. 61 Introduction ...................... 61 Apparatus ........................... 62 Measurements and Calculations ....... 71 Errors ................... 74 Experimental Results ............... 75 ACKNOWLEDGMENTS .......................... 80 APPENDIX I Physical Constants, Drop Calorimeter Data .... 61 APPENDIX II Low Temperature Calorimeter Data ..... 82 APPENDIX III Standard Lamp Calibration ............ 83 3182G G TABLE OF CONTENTS, (cont.) Page APPENDIX IV Bibliography ................. 85 AUTOBIOGRAPHY .......................... 89 iii LIST OF ILLUSTRATIONS 1 Vacuum Furnace 15 2 Dropping Mechanism 19 3 Improved Calorimeter 20 4. Modified Calorimeter, II 21 5 Drop Calorimeter Electrical Circuits
    [Show full text]
  • Adiabatic Dewar Calorimeter
    I.CHEM.E. SYMPOSIUM SERIES NO. 97 ADIABATIC DEWAR CALORIMETER T.K.Wright'and R.L.Rogers* A simple calorimeter has been developed that enables chemical reaction runaway conditions to be directly determined, under the low heat loss conditions found in full scale chemical plants. Since the calorimeter provides temperature time data in the near absence of environmental heat losses the data can be simply analysed to yield heats of reaction and chemical power output. The latter are used either in conjunction with plant natural cooling data to assess reactor stability or at higher temperatures to size reactor vents. If the reaction mechanisms are known or adequate assumptions can be made then the temperature-time data can also be processed to yield reaction kinetics constants for simulation purposes. Keywords: Hazards, Exotherms, Adiabatic Calorimeter, kinetics INTRODUCTION Evaluation of chemical reaction hazards requires the detection of exotherms/gas generation likely to lead to reactor overpressure. Some form of small scale scanning calorimetry is generally used for the initial detection of the exotherm and gas generation and a temperature of onset will be determined which is dependent on the sensitivity of the equipment, but on a 10-20gm scale exothermlcity will generally be detected at self heating rates of 2-10°C/hr - approximately 3-10 watts/lit. Depending on apparent exotherm size and proximity to process temperature or any likely excursions then secondary testing may be required to display more accurately (a) The minimum temperature above which the reactor will be unstable on the scale used. (b) The consequences of the exotherm - heat of reaction/adiabatic rise/ pressure developed/venting requirement.
    [Show full text]
  • Simple Calorimeter for Heats of Fusion. Data on The
    U.S. Department of Commerce, Bureau of Standards RESEARCH PAPER RP607 Part of Bureau of Standards Journal of Research, Vol. 11, October 1933 A SIMPLE CALORIMETER FOR HEATS OF FUSION. DATA ON THE FUSION OF PSEUDOCUMENE, MESITYLENE (« AND 0), HEMIMELLITENE, o- AND m-XYLENE, AND ON TWO TRANSITIONS OF HEMIMELLITENE By Frederick D. Rossini abstract A vacuum flask with a thermoelement serves as a simple calorimeter for measuring heats of fusion quickly and economically, with an accuracy of a few percent. The following heats of fusion (with estimated uncertainties), in k-cal. per mole, were obtained: pseudocumene, — 44.1° C, 2.75±0.06; hemimelli- tene,-25.5° C, 2.00±0.05; mesitylene (a), — 44.8° C, 2.28±0.06; mesitylene (0), -51.7° C., 1.91±0.05; o-xylene,-25.3° C, 3.33±0.07; m-xylene,-47.9° C, 2.76 ±0.05. Hemimellitene was found to have two transitions below the freez- ing point, with the following heats of transition, in A>cal. per mole: hemimelli- tene (7-»0),-58±2° C.,0.28±0.04; hemimellitene (P^a),- 46 ± 1° C.,0.36±0.04. CONTENTS Page I. Introduction 553 II. Apparatus and method 553 III. Materials 554 IV. Standardization experiments 555 V. Experimental data 557 VI. Conclusion 559 I. INTRODUCTION The simple calorimeter described here was assembled in order to provide a means for measuring as quickly and economically as prac- ticable, and with an accuracy of a few percent, the heats of fusion of certain hydrocarbons for which there are no data.
    [Show full text]
  • The Measurement of Heat Release Rates by Oxygen Consumption Calorimetry in Fires Under Suppression
    The Measurement of Heat Release Rates by Oxygen Consumption Calorimetry in Fires Under Suppression BOGDAN Z. DLUGOGORSKI, JACK R. MAWHINNEY and VO HUU DUC National Research Counc~l,Institute for Research in Construction Ottawa, Ontario KIA OR6, Canada ABSTRACT A series of open-space fire experiments was conducted at the National Fire Laboratory (NFL) to validate the capabilities of the NFL room-size oxygen-consumption calorimeter, and to assess the importance of accounting for actual water vapour content in the exhaust gases in calculating heat release rates (HRR). Water spray was used to partially suppress some of the fires, and to add significantly to the humidity of the exhaust gases. The equations normally used in the fire research community for oxygen calorimeter assume unsuppressed fires, and that water vapour in the exhaust gases is due solely to the humidity of the incoming air and to combustion reactions. This paper derives the basic equations for computing heat release rates based on the principle of nitrogen balance. The general equations take into account all sources of water vapour, including incoming air, combustion reactions, and evaporation due to suppression. The equations are then simplified to i) neglect all humidity, and, ii) consider only the humidity of the incoming air. The predictions of the HRR from the three sets of equations are compared with the HRR calculated for unsuppressed fires and with the HRR obtained by measuring fuel consumption rates. As long as the water vapour content in the exhaust gases is less than 7 %, both simplified equations can be used to measure the HRR of partially suppressed fires, without significant error.
    [Show full text]
  • Calorimetry Lab
    P31220 lab Calorimetry Lab Purpose: Students will measure latent heat and specific heat. PLEASE READ the entire handout before starting. You won’t know what to do unless you understand how it works! Introduction: Calorimetry is the art of measuring energy. For example, determining how many calories are in a cheeseburger is done with a device called a “bomb calorimeter.” A sample of the food is burned in a closed container that is surrounded by water. The energy content of the food is determined from the temperature increase of the water jacket that surrounds the combustion chamber. In this lab, you will do two classic calorimetry experiments: measuring the latent heat of fusion of water, and measuring the specific heat capacities of two different metals. Both experiments will use the same apparatus. Apparatus: Fig. 1 shows the construction of the basic calorimeter. The calorimeter is designed to minimize heat flow between the inner cup and the outside world. Conduction of heat is eliminated by supporting the inner cup only by the thin, insulating phenolic (a type of plastic) ring, and by providing an insulating air space around the cup. Convection is eliminated by blocking air circulation with the solid ring and the lid. Radiation is eliminated by making the inner cup and outer jacket out of aluminum, which is mirror- bright to infrared radiation. Fig. 1: The calorimeter. To use the calorimeter, the inner cup is half filled with a known mass of water, and the temperature is measured. The sample is added, the temperature is measured again, and the desired quantity (latent heat or specific heat) is calculated.
    [Show full text]
  • 040 Heat of Hydration
    Heat of hydration HEAT OF HYDRATION. SEMI ADIABATIC METHOD Standard EN 196-9 COMPUTERIZED LANGAVANT CALORIMETER. Ref. 111-101238 For the determination of cement heat hydration. Features This equipment comprises a computerized system for the automatic data acquisition during the test. Test data can be managed by means of a specific test software WinLec32 (Test-Lang version), developed specifically by the R&D IBERTEST Dept. for the Langavant method. The system automates data collection and performs the calculations required by computer, thus avoiding possible errors in the manual data collection, providing test results provided reliable and reproducible. The test results are stored on computer files and can be retrieved at any time for reporting, comparison, statistical analysis, etc. The minimun configuration comprises the following elements › Set of 2 isolated calorimeter bottles (one for reference), supplied with official calibration certificate. › Set of 2 temperature probes type Pt-100, supplied with official calibration certificate. › Set of 50 disposable mortar test cans. › Electronic module, with 4 measuring channels, for connecting up to 4 calorimeter bottles to the PC (3 for testing + 1 for temperature reference). The module is linked to the PC via USB 2.0 › Next-generation PC (dual-core microprocessor or higher), keyboard, mouse, TFT widescreen, Windows® operating system, manuals and licenses. › WinLec32 IBERTEST Software License (Version Test-Lang), preinstalled on the provided PC, running on Windows® NOTE: All the calorimeter bottles have been calibrated and certified by the “Laboratoire Regional Ponts et Chaussées”. Each bottle is individually marked and comprises a metallic plate indicating its heat loss coefficient and heat mass.
    [Show full text]
  • Technology Data Sheet Nuclear Material (NM)
    IPNDV Working Group 3: Technical Challenges and Solutions Nuclear Material (1)—Technology Data Sheet September 14, 2016 Nuclear Material (NM) Technology Name: Calorimetry Physical Principle/Methodology of Technology: Calorimetry measures the thermal power output of heat-producing NM. The heat results from the radioactive decay of isotopes by alpha particle emission (for most Pu isotopes and 241Am) and by beta decay (for 241Pu and Tritium). 239Pu for instance decays to 235U with the emission of an alpha particle and releasing 5.15 MeV in energy. The energy loss through the emission of spontaneous fission neutrons on the other hand is many orders of magnitude smaller than the total disintegration of energy and loss due to gamma ray emission, representing only a small percent of the total disintegration energy of Pu isotopes. Calorimetry is most commonly used for Pu measurements due to the high heat output of most Pu isotopes. In addition, the build-up of 241Am as Pu ages significantly increases the power output. It is also possible to measure highly enriched uranium (HEU) when there is a high enough 234U content. For reference, isotopes that emit relatively large amounts of heat include 238Pu (560 W/kg), 239Pu (1.9 W/kg), 240Pu (6.8 W/kg), 241Pu (4.2 W/kg), 241Am (114 W/kg), and 234U (0.2 W/kg). Thermal powers ranging from 0.1 mW to 1000 W can be measured with calorimetry. Used independently, calorimetry confirms that an item emits heat and determines how much heat is being emitted. The heat measurement is very accurate and precise.
    [Show full text]
  • TAM AIR ISOTHERMAL CALORIMETRY High Sensitivity
    TAM AIR ISOTHERMAL CALORIMETRY High Sensitivity Multi-Sample TAM AIR The best choice Long-Term 3 or 8 for isothermal Temperature Channel calorimetry of Stability Option chemical reactions and metabolic processes The TAM Air isothermal calorimetry system provides the sample flexibility and measurement sensitivity to accurately characterize heat flow in a wide variety of materials. The 8-channel Lowest calorimeters are widely recognized as the instrument of choice Baseline for cement hydration analyses using ASTM C1702 and C1679. The 3-channel calorimeters have the required sensitivity to detect Drift low levels of metabolic activity required for applications such as soil remediation and food contamination analysis. Unmatched baseline stability and signal-to-noise performance makes the TAM Air the best choice for the analysis of medium to high heat- producing materials. TAM AIR A VERSATILE ISOTHERMAL CALORIMETER with HIGH PERFORMANCE & ROBUSTNESS The TAM Air is a flexible, sensitive analytical platform that is an ideal tool for large-scale isothermal calorimetry experiments. Capable of measuring several samples simultaneously under isothermal conditions, the TAM Air with its air-based thermostat and easily interchangeable 8-channel standard volume or 3-channel large volume calorimeters, is especially well-suited for characterizing processes that evolve or consume heat over the course of days and weeks. The TAM Air offers high sensitivity and long-term temperature stability for detecting and characterizing heat production or consumption in a wide variety of sample types and sizes. Applications include cement and concrete hydration, food spoilage, microbial activity, battery performance and more. Monitoring the thermal activity or heat flow of chemical, physical and biological processes provides information which cannot be generated with other techniques.
    [Show full text]
  • Draft Atmospheric Pressure Effects on Cryogenic
    DRAFT ATMOSPHERIC PRESSURE EFFECTS ON CRYOGENIC STORAGE TANK BOIL-OFF J. P. Sassa and C. R. Fortier1' aNASA Kennedy Space Center, KT-E Kennedy Space Center, FL, 32899, USA "NASA Kennedy Space Center, NIE-F7 Kennedy Space Center, FL, 32899, USA ABSTRACT The Cryogenics Test Laboratory (CTL) at the Kennedy Space Center (KSC) routinely utilizes cryostat test hardware to evaluate comparative and absolute thermal conductivities of a wide array of insulation systems. The test method is based on measurement of the flow rate of gas evolved due to evaporative boil-off of a cryogenic liquid. The gas flow rate typically stabilizes after a period of a couple of hours to a couple of days, depending upon the test setup. The stable flow rate value is then used to calculate the thermal conductivity for the insulation system being tested. The latest set of identical cryostats, 1 000-L spherical tanks, exhibited different behavior. On a macro level, the flow rate did stabilize after a couple of days; however the stable flow rate was oscillatory with peak to peak amplitude of up to 25 percent of the nominal value. The period of the oscillation was consistently 12 hours. The source of the oscillation has been traced to variations in atmospheric pressure due to atmospheric tides similar to oceanic tides. This paper will present analysis of this phenomenon, including a calculation that explains why other cryostats are not affected by it. INTRODUCTION Cryostats are regularly used at the CTL to determine the thermal performance of various insulations. Tests are conducted in small 1O-L cylindrical cryostats or large 1000-L spherical tanks.
    [Show full text]
  • 7—Thermochemistry: Heat of Reaction
    7—THERMOCHEMISTRY: HEAT OF REACTION Name: ______________________________________________ Date: ________________________________________________ Section: _____________________________________________ Objectives • Measure the enthalpy of reaction for the decomposition of hydrogen peroxide • Measure the heat capacity of a Styrofoam cup calorimeter using the heat of neutralization of a strong acid with a strong base • Graph your temperature vs time data to find temperature change when solutions are mixed Pre-Laboratory Requirements • Read chapter 6 in Silberberg • Pre-lab questions (if required by your instructor) • Laboratory notebook—prepared before lab (if required by your instructor) Safety Notes • Eye protection must be worn at all times. • Hydrochloric acid and sodium hydroxide are caustic and should not come in contact with your skin or clothing. Wear gloves when handling these chemicals. A lab coat or lab apron is recommended. • The hydrogen peroxide used in this experiment is often used as an antiseptic. However, avoid unnecessary exposure to your skin and clothing. Do not splash hydrogen peroxide in your eyes. Discussion When a chemical reaction occurs, bonds are broken and new bonds are formed. The breaking of bonds requires energy while the formation of new bonds releases energy. For most spontaneous chemical reactions, the energy released by the new bond formation is greater than that required to break the old bonds. In this process, energy is released (usually as heat) into the environment. These reactions are called exothermic and a minus sign is put in front of the quantity of energy released. Some reactions require energy from the environment to proceed; these are called endothermic reactions. Because energy must flow from the environment into the reactants for the reaction to occur, a plus sign is put in front of the amount of energy absorbed.
    [Show full text]