ISSSD 2014 April 13 to 16th, 2014. Cusco, Peru

VALIDATION OF MEAN GLANDULAR DOSE VALUES PROVIDED BY A DIGITAL BREAST TOMOSYNTHESIS SYSTEM IN BRAZIL

Bruno Beraldo Oliveira1, Lucas Paixão1, Sabrina Donato da Silva1 Maria Helena Araújo Teixeira2, Maria do Socorro Nogueira1/3

1 Post-graduation in Sciences and Technology of Radiations Minerals and Materials - CDTN/CNEN Presidente Antônio Carlos, 6627 31270-901, Belo Horizonte, Brazil [email protected]

2 Dr Maria Helena Araújo Teixeira Clinic Guajajaras, 40 30180-100, Belo Horizonte, Brazil [email protected]

3 Development Center of Nuclear Technology - CDTN/CNEN Presidente Antônio Carlos, 6627 31270-901, Belo Horizonte, Brazil [email protected]

213

ISSSD 2014 April 13 to 16th, 2014. Cusco, Peru

Abstract

Digital breast tomosynthesis (DBT) is an emerging imaging modality that provides quasi-three-dimensional structural information of the breast and has strong promise to improve the differentiation of normal tissue and suspicious masses reducing the tissue overlaps. DBT images are reconstructed from a sequence of low-dose X-ray projections of the breast acquired at a small number of angles over a limited angular range. The Hologic Selenia Dimensions system is equipped with an amorphous Selenium (a-Se) detector layer of 250 μm thickness and a 70 μm pixel pitch. Studies are needed to determine the radiation dose of patients that are undergoing this emerging procedure to compare with the results obtained in DBT images. The mean glandular dose (DG) is the dosimetric quantity used in quality control of the mammographic systems. The aim of this work is to validate DG values for different breast thicknesses provided by a Hologic Selenia Dimensions system using a DBT mode in comparison with the same results obtained by a calibrated 90X5-6M-model

Radcal . DG values were derived from the incident air kerma (Ki) measurements and tabulated conversion coefficients that are dependent on the half- value layer (HVL) of the X-ray spectrum. Voltage and tube loading values were recorded in irradiations using W/Al anode/filter combination, automatic exposure control mode and polymethyl methacrylate (PMMA) slabs which simulate different breast thicknesses. For Ki measurements, the ionization chamber was positioned at 655 mm from the focus and the same radiographic technique values were selected with the manual mode. DG values for a complete procedure ranged from 0.9 ± 0.1 to 3.7 ± 0.4 mGy. The results for different breast thicknesses are in accordance with values obtained by DBT images and with acceptable levels established by the Commission of the European Communities (CEC) and the International Atomic Energy Agency (IAEA). This work contributes to determine the reliable radiation dose received by the patients and validate the values provided by this DBT system.

Keywords: Digital breast tomosynthesis; mean glandular dose; incident air kerma

214

ISSSD 2014 April 13 to 16th, 2014. Cusco, Peru

1.- INTRODUCTION

Two-dimensional (2D) mammography currently is the gold standard for detecting breast cancer in an early stage and is used in screening programs in many countries [Van Schie et al., 2013]. It is a low cost, fast, noninvasive technique that involves low doses of . However, 2D images results in tissue superposition [Sechopoulos 2013]. Digital breast tomosynthesis (DBT) is an emerging imaging modality that provides quasi- three-dimensional (3D) structural information of the breast and has strong promise to improve the differentiation of normal tissue and suspicious masses reducing the tissue overlaps [Lu et al., 2013]. DBT can reduce false positive findings and allow 3D localization of a lesion in the breast. The possibility of achieving these results at a dose level roughly equivalent to that used in conventional mammography has generated interest in the application of tomosynthesis for breast imaging [Wu et al., 2013]. The Hologic Selenia Dimensions system is equipped with an amorphous Selenium (a-Se) detector layer of 250 μm thickness and a 70 μm pixel pitch. For planar mammography, the system uses W/Rh and W/Ag anode/filter combinations while in DBT mode, the system changes the filtration to 700 μm of Al thickness. A Hologic proprietary filtered back projection (FBP) algorithm is used to reconstruct images from 15 low-dose X-ray projections of the breast acquired over 15º [Cockmartin et al., 2013]. DBT technique did not have widespread clinical use for many years but large-area digital detectors became available, and DBT gained a resurgence of research and clinical interest in the last decade [Li et al., 2013a]. The imaging performance of DBT is complicated mainly by the detector efficiency, the need for precise geometrical alignment, and the detection of X-ray scatter. None of DBT systems available incorporates the anti-scatter grid or a correction technique as a solution for reduction of the X-ray scatter signal that can result in loss of contrast and accuracy [Sechopoulos 2013; Sechopoulos et al., 2013]. The problem of a limited number of projections acquired is one of the challenges in attempting to reconstruct a 3D breast volume from 2D projection images [Wu et al., 2013]. DBT is characterized by anisotropic spatial resolution, with high spatial resolution in the planes parallel to the detector, and a considerably lower resolution in the perpendicular

215

ISSSD 2014 April 13 to 16th, 2014. Cusco, Peru direction. However, this low spatial resolution is deemed to be enough to substantially reduce the issue of tissue superposition [Sechopoulos 2013]. The Food and Drug Administration (FDA) approved in 2011 the DBT systems using a 2D imaging alone or a combination of 2D and 3D imaging (Combo mode) [Gur et al., 2009; Philpotts et al., 2012]. The selection of Combo mode may result in more radiation exposures than with conventional mammography [Li et al., 2013b]. Radiation dose is an important concern in screening mammography.

Regulations require medical physicists to evaluate the Mean Glandular Dose (DG) as a part of quality control of services [Li et al., 2013a]. Studies are needed to determine the radiation dose of patients that are undergoing this emerging procedure to compare with the results obtained in DBT images.

The aim of this work was to validate DG values for different breast thicknesses provided by a Hologic Selenia Dimensions system using a DBT mode in comparison with the same results obtained by a calibrated 90X5-6M-model Radcal ionization chamber.

2.- MATERIALS AND METHODS

DG values were estimated from the incident air kerma (Ki) for exposure of patients using tabulated conversion coefficients that are dependent on the half-value layer (HVL) of the X-ray beam [IAEA 2007]. Measurements were carried using a Hologic Selenia Dimensions system located in a mammography clinic of Belo Horizonte city, Brazil. The medical physicist’s quality control tests of the DBT system used was in according to the quality assurance program for digital mammography established by the International Atomic Energy Agency (IAEA) [IAEA 2011]. The distance between the X-ray focal spot and the breast support was 700 mm. Irradiations were carried out using the automatic exposure control (AEC) and the DBT mode for polymethyl methacrylate (PMMA) slabs ranging from 20 to 70 mm of thickness. PMMA spacers of 1, 2, 5, 10 and 20 mm of thickness were used over the slabs to represent breasts of different attenuations (Figure 1). The W/Al anode/filter combination was used

216

ISSSD 2014 April 13 to 16th, 2014. Cusco, Peru and the values of tube voltage (kV) and tube loading (mAsAEC) were recorded for each exposure.

Figure 1.- Setup for exposure parameters determination.

2.1.- Ki and HVL measurements

The Radcal 90X5-6M ionization chamber was centred laterally at 60 mm from the chest wall and 45 mm from the breast support. The Ki was measured using the manual mode for the same parameters selected to expose different thickness of PMMA slabs and spacers. For different values of mAs, the kerma (MAEC) values were estimated by extrapolation from a measurement taken at the closest available manual mAs setting (Mmanual) according to Equation 1 [IAEA 2011].

 mAs  M  M  AEC  AEC manual   (1)  mAsmanual 

217

ISSSD 2014 April 13 to 16th, 2014. Cusco, Peru

For Ki calculations, the Equation 2 were used, where Ki is the incident air kerma measured without backscatter at the surface of the slab of PMMA, Nmammo is the value of the calibration factor for beam quality, and kTP is the correction factor for temperature and pressure [IAEA 2011].

Ki  M AEC Nmammo kTP (2)

In these calculations, for correction of Ki values to the location of the entrance surface, the

Equation 3 were used, where Ki,n is Ki value obtained at a determined thickness, d is the distance between focus and breast support and n is the total thickness of PMMA slabs and spacers [IAEA 2011].

2  d  45  Ki,n  Ki,45    d  n  (3)

The same setup was maintained and HVL values of the beam were calculated by adding high purity (99.999%) aluminum filtrations of 0.25, 0.35 and 0.5 mm of thickness positioned on the compression plate. The Equation 4 was used to obtain the acceptable HVL values depending of the tube voltage, where C is 0.25 for W/Al combination [IAEA 2011].

kV kV  0.03  HVL   C 100 100 (4)

2.2.- DG measurements

In DBT systems, the DG is the sum of the doses received from individual projections. DG calculations were carried out using the Equation 5, where g, c, s and t() are conversion factors [Dance et al., 2011; IAEA 2011].

218

ISSSD 2014 April 13 to 16th, 2014. Cusco, Peru

DG   Ki g c s t  (5)

The factor g gives the DG for a breast having a 50% fibroglandular and 50% fat composition and is tabulated against breast thickness and HVL. The factor c allows the calculation for breast of different glandularity and is tabulated against HVL and thickness for typical breast compositions. The factor s gives a correction that depends on the anode/filter combination [Dance 1990; Dance et al., 2000]. The factor t() is the tomo factor for projection angle  [Dance et al., 2011; IAEA 2011]. DG calculations were also carried out using the Equation 6, where the factor T gives a correction for a complete DBT examination [Dance et al., 2011].

DG   Ki g c s T (6)

3.- RESULTS

Exposure parameters for different thickness of PMMA using the AEC mode and W/Al combination are shown in Table 1. The thickness and the fibroglandular proportion of equivalent breast were obtained by adding the respective thickness of PMMA spacers on the PMMA slabs.

Table 1.- Exposure parameters for each thickness of PMMA slabs. Thickness of Thickness of Fibroglandular Tube voltage Tube loading PMMA [mm] equivalent breast* proportion of [kV] [mAs] [mm] equivalent breast* [%] 20 21 97 26 41 30 32 67 29 42 40 45 41 30 57 50 60 20 33 77 60 75 9 36 86 70 90 4 42 69 * Data provided by IAEA [2011].

219

ISSSD 2014 April 13 to 16th, 2014. Cusco, Peru

3.1.- Ki and HVL measurements

Ki and HVL values for different thickness of PMMA, besides tolerances and HVL values provided by the DBT system are shown in Table 2. The relative expanded uncertainty (k =

2) of the Ki and HVL measurements was calculated as 6.7%.

Table 2.- Ki and HVL values for each thickness of PMMA slabs.

Thickness of Ki [mGy] HVL [mm Al] PMMA [mm] Provided Calculated Tolerance 20 1.87 ± 0.13 0.45 0.46 ± 0.03 0.29 - 0.51 30 2.88 ± 0.19 0.51 0.53 ± 0.04 0.32 - 0.54 40 4.26 ± 0.29 0.53 0.55 ± 0.04 0.33 - 0.55 50 7.25 ± 0.49 0.59 0.60 ± 0.04 0.36 - 0.58 60 10.20 ± 0.68 0.64 0.64 ± 0.04 0.39 - 0.61 70 12.07 ± 0.81 0.74 0.70 ± 0.05 0.45 - 0.67

3.2.- DG measurements

Calculated DG values with the respective uncertainties and different thickness of PMMA for individual projections and for a complete DBT examination are shown in Figures 2 and

3 respectively. The relative expanded uncertainty (k = 2) of the DG measurements was calculated as 10%. Figure 3 also indicates values provided by the DBT system and by Oliveira et al., [2011] using a computed radiography (CR) system beyond the acceptable and desired levels established by the Commission of the European Communities (CEC) and the International Atomic Energy Agency (IAEA) [CEC 2006; IAEA 2011].

220

ISSSD 2014 April 13 to 16th, 2014. Cusco, Peru

Figure 2.- DG values for each thickness of PMMA slabs and projection angles.

Figure 3.- DG values for each thickness of PMMA slabs.

221

ISSSD 2014 April 13 to 16th, 2014. Cusco, Peru

4.- DISCUSSION

Considering the uncertainties, HVL results are in accordance with the tolerance established by IAEA [2011] and with values provided by the DBT system. The validation of HVL values provided by the Brazilian DBT system images is important to obtain reliable DG values. In Figure 2, there is no statistical difference in DG values for the same PMMA thickness and different projections of the X-ray tube. The relative error between the minimum and maximum DG values was approximately 76%. In Figure 3, DG values ranging from 0.90 ± 0.09 to 3.75 ± 0.37 mGy. Considering the uncertainties, DG results are in accordance with the values provided by the DBT system images and the acceptable levels established by CEC [2006] and IAEA [2011]. Moreover, for larger thickness of PMMA, the calculated DG values are under the desired levels established. In both figures, the minimum DG values was registered at 20 mm of PMMA thickness and the results increases with the thickness of PMMA, due to higher X-ray attenuation in larger thickness of this material. In accordance with Oliveira et al., [2011] there are around 600 mammography equipments with CR systems in Brazil and according to data obtained by the manufacturer’s representatives approximately 15 Hologic Selenia Dimensions systems. Therefore, it is important to perform a comparison of the radiation dose of patients that are undergoing a CR procedure with this new radiographic technique.

Oliveira et al., [2011] have published information on assessment of DG and image quality in mammography using a CR system located in a Brazilian laboratory. Considering the uncertainties, their DG values are higher than the results obtained in this work only for the last two thickness of PMMA.

5.- CONCLUSIONS

In this work, DG values were calculated to compare with values provided by the DBT system and the reference levels established internationally. Considering the uncertainties, the results are according to DG values provided by the equipment. Moreover, the results for

222

ISSSD 2014 April 13 to 16th, 2014. Cusco, Peru each thickness of PMMA are in accordance with acceptable levels established by IAEA [2011] and CEC [2006]. The is an important part of quality control and crucial for the optimization of DBT procedures. Taking into consideration the fact that this DBT system showed previous results of medical physicist’s quality control tests in according to the quality assurance program for digital mammography established by IAEA [2011], this system has proved to be calibrated and optimized. Therefore, the aim of this work was fulfilled by consolidating the dosimetry in DBT systems in addition to validate DG values provided by this equipment.

The results also showed that DG increases with the thickness of PMMA because of the higher X-ray attenuation in larger thickness of equivalent breasts. There is no statistical difference in DG values for each PMMA thickness and individual projections with the X- ray tube ranging from -7.5º to +7.5º.

For thick and fat equivalent breasts, DG values are under the desired levels established. This work also shows that the adoption of a DBT system does not imply an increase in DG values when compared with results obtained by Oliveira et al. [2011] in a CR system.

Acknowledgments

The authors are thankful to CNEN and CAPES for the incentive with PhD’s fellowship and Dr. Maria Helena Araújo Teixeira for providing the mammography unit. This work was supported by FAPEMIG and Ministry of Science and Technology - MCT/Brazil, through the Brazilian Institute of Science and Technology (INCT) for Radiation Metrology in Medicine.

REFERENCES CEC - Commission of the European Communities. (2006). European Guidelines for Quality Assurance in Breast Cancer Screening and Diagnosis. Report EUR 14821 4th ed. European Commission. Office for Official Publications of the European Communities. Luxembourg.

223

ISSSD 2014 April 13 to 16th, 2014. Cusco, Peru

Cockmartin L; Bosmans H; Marshall NW. (2013). Comparative power law analysis of structured breast phantom and patient images in digital mammography and breast tomosynthesis. Medical Physics 40: 081920-1-081920-17.

Dance DR. (1990). Monte Carlo calculation of conversion factors for the estimation of mean glandular breast dose. Physics in Medicine and Biology 35: 1211-1219.

Dance DR; Skinner CL; Young KC; Beckett JR; Kotre CJ. (2000). Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocol. Physics in Medicine and Biology 45: 3225-3240.

Dance DR; Young KC; Van Engen RE. (2011). Estimation of mean glandular dose for breast tomosynthesis: Factors for use with the UK, European and IAEA breast dosimetry protocols. Physics in Medicine and Biology 56: 453-471.

Gur D; Abrams GS; Chough DM; Ganott MA; Hakim CM; Perrin RL; Rathfon GY; Sumkin JH; Zuley ML; Bandos AL. (2009). Digital Breast Tomosynthesis: Observer Performance Study. American Journal of Roentgenology 193: 586-591.

IAEA - International Atomic Energy Agency. (2007). Dosimetry in Diagnostic Radiology: An International Code of Practice. IAEA Technical Reports Series 457. Vienna.

IAEA - International Atomic Energy Agency. (2011). Quality Assurance Programme for Digital Mammography. IAEA Human Health Series 17. Vienna.

Li X; Zhang D; Liu B. (2013a). A parameterization method and application in breast tomosynthesis dosimetry. Medical Physics 40: 092105-1-092105-9.

Li X; Zhang D; Liu B. (2013b). Workload and transmission data for the installation of a digital breast tomosynthesis system. Medical Physics 40: 063901-1-063901-8.

Lu Y; Chan H-P; Wei J; Hadjiiski LM. (2013). A diffusion-based truncated projection artifact reduction method for iterative digital breast tomosynthesis reconstruction. Physics in Medicine and Biology 58: 569-587.

Oliveira MA; Dantas MVA; Santana PC; Squair PL; Gomes DS; Nogueira MS. (2011). Assessment of glandular dose and image quality in mammography using computerised radiography employing a polymethylmetacrilate breast simulator. Radiation Measurements 46: 2081-2085.

Philpotts L; Raghu M; Durand M; Hooley R; Vashi R; Horvath L; Geisel J; Butler R. (2012). Initial experience with digital breast tomosynthesis in screening mammography. Proccedings of the 2012 Annual Meeting of the American Ray Society. Vancouver, Canada.

Sechopoulos I. (2013). A review of breast tomosynthesis. Part I. The image acquisition process. Medical Physics 40: 014301-1-014301-12.

224

ISSSD 2014 April 13 to 16th, 2014. Cusco, Peru

Sechopoulos I; Bliznakova K; Fei B. (2013). Power spectrum analysis of the x-ray scatter signal in mammography and breast tomosynthesis projections. Medical Physics 40: 101905-1-101905-7.

Van Schie G; Wallis MG; Leifland K; Danielsson M; Karssemeijer N. (2013). Mass detection in reconstructed digital breast tomosynthesis volumes with a computer- aided detection system trained on 2D mammograms. Medical Physics 40: 041902-1- 041902-11.

Wu G; Mainprize JG; Yaffe MJ. (2013). Characterization of a constrained paired-view technique in iterative reconstruction for breast tomosynthesis. Medical Physics 40: 101901-1-101901-12.

225