SPECIAL SCIENTIFIC REPORT-FISHERIES Na 349

Total Page:16

File Type:pdf, Size:1020Kb

SPECIAL SCIENTIFIC REPORT-FISHERIES Na 349 349 CHEMICAL ANALYSES OF MARINE AND ESTUARINE WATERS USED BY THE GALVESTON BIOLOGICAL LABORATORY SPECIAL SCIENTIFIC REPORT-FISHERIES Na 349 UNITED STATES DEPARTMENT OF THE INTERIOR FISH AND WILDLIFE SERVICE United States Department of the Interior, Fred A. Seaton, Secretary Fish and Wildlife Service, Arnie J. Suomela, Commissioner Bureau of Commercial Fisheries, Donald L, McKernan, Director CHEMICAL ANALYSES OF MARINE AND ESTUARINE WATERS USED BY THE GALVESTON BIOLOGICAL LABORATORY by Kenneth T. Marvin, Zoula P. Zein-Eldin, Billie Z. May and Larence M. Lansford Chemists Galveston, Texas United States Fish and Wildlife Service Special Scientific Report— Fisheries No. 349 Washington, D. C. June 1960 CONTENTS Introduction 1 Sample treatment prior to analysis 1 Sample storage containers 2 Analytical methods 2 Standard samples 2 Phosphate 3 Inorganic only 3 Total and inorganic 3 Total only 4 Nitrate-nitrite 5 Nitrite 5 Salinity 6 Copper 6 Sulfide 7 Oxygen 7 Total carbon dioxide 8 Ammonia 10 Chlorophyll 10 "Carbohydrates" 11 "Protein" (tyrosine equivalent) 12 Washing procedure for all analytical glassware 12 Literature cited 13 111 CHEMICAL AMLYSES OF MAP.INE AND ESTOARIKE mTERS USED BY TBE GALVESTON BIOLOGICAL lABORATORY by Kenneth T. Marvin, Zoula P. Zein-Eldin, Billie Z. May and Larence M. Lansford ABSTRACT This paper describes the chemical techniques and procedures used hy the Biological lahoratory of the U. S. Bureau of Commercial Jlsherles, Galveston^ Texas, for analyzing samples Involved In the chemical and hlo- logical survey of the marine and estuarlne waters of the Gulf of Mexico and also In the many laboratory and field studies and experiments that have heen made pertaining to the red tide investigation. During the past 12 years the staff of All samples collected from the M/Vs the U. S. Bureau of Commercial Fisheries, Kingf ish and Alaska that were subject to Galveston, Texas, has published papers per- bacterial and/or chemical deterioration were taining to the hydrography and chemistry of quick-frozen immediately after collection. the Gulf of Mexico and also problems rela- Ctollier and Marvin (1953) showed that signi- tive to the red tide. Included were data ficant changes in the phosphate ratio of collected from the M/V Pompano from 1949 to some waters occurred within 30 minutes after 1952 (Graham et al. 1954, and Marvin 1955a), collection, but that the ratio could be the MA Alaska from 1951 to 1953 (Collier effectively stabilized by quick-freezing. 1958), and the M/V Kingfish from 1954 to May (unpublished msinuscript) showed that date (Finucane and Dragovich 1959). variations in the carbohydrate-like raatericil in sea water could be minimized by similar The purpose of this paper is to pre- treatment. After quick-freezing, samples sent the chemical methods which were used were stored at below freezing temperatures to obtain the data utilized in these publi- until delivered to the chemistry laboratory cations. We have endeavored to describe for analysis. Our usual freezing bath con- the techniques and procedures in a form sisted of a vat of 38 percent ethylene gly- usable by chemical aids or technicians with col (commercial antifreeze) held at about a minimum of supervision. -20° C. in a quick-freeze cabinet. Freezing can reduce the air pressure within a sample container to a point where impurities around SAMPLE TREATMENT PRIOR TO ANALYSIS the cap-glass junction will be drawn into the container. We minimized this effect by No attempt was made to stabilize the sealing the junction with plastic insulation to Scotch #33). The seal samples collected from the M/V Pomp ano . tape (similar With the exception of salinity and total should not be allowed to come in contact phosphate analyses, all determinations were with the freezing solution. performed on board ship immediately after collection. Salinity, copper, eind total phosphorus samples did not require preservation. The semipermeable membrane under freezing latter, however, were "baited" with a pinch conditions. Samples frozen in this type of of thorium carbonate (Harvey 1948) to pre- container have shown as much as a 2-percent vent the accumulation of phosphorus on the decrease in salinity concentration within sides of the container. An alternative a year. Water from the same samples stored method that we found satisfactory and have at room temperature have shown a salinity used on occasion consisted of adding the increase of over 1 percent during the same amount of H2SO4 called for in the analytical period. procedure immediately after collection. Sulfide, CO2 , and O2 samples were analyzed Sample evaporation can virtually be immediately after collection. We found the eliminated by sealing containers with poly- practice of fixing O2 samples and titrating seal screw caps. Salinity samples, for them later resulted in an error because the example, have remained constant for as long amount of free iodine in the supposedly as a year and a half when sealed in this fixed sample increased with time. manner, and we strongly recommend their use The rate of reactions, which includes the development of color, depends among ANALYTICAL METHODS other things on temperature. For this rea- son the analyst must be sure that all sam- The colorimetric procedures that fol- ples and standards are brought to the same low refer to the Fisher A. C. electrophoto- temperature before adding reagents. Stand- meter, the instrument used for the determi- ard samples were interspersed among the nation of the majority of samples collected regular samples rather than being run as a from the M/V Kingf ish and also the MA Alaska . separate group. We have found that this The colorimetric analyses of samples col- practice reduces operator bias. lected from the M/V Pompano , however, were carried out visually with Nessler tubes. SAMPLE STORAGE CONTAINERS Standard Samples The selection of proper sample con- With a few exceptions, the methods tainers is of considerable importance. We described below require the use of labora- have found, after considerable experimenta- tory-prepared standard samples. Initially tion, that pyrex is equal to or superior to we employed distilled water in the prepara- soft glass, particularly for the storage of tion of these. However, it became evident total and inorganic phosphate and nitrate- that the percent recovery depended, to some nitrite samples. During our survey work extent, on the type of water. For this (M/V Pomp ano ) , those samples that were not reason we endeavored, when possible, to era- analyzed on board ship (salinity and total ploy used portions of the samples for the phosphorus) were stored in citrate bottles. preparation of standards. If a sufficient During the initial collection programs in- volume cannot be obtained for this purpose, volving the M/V Alaska , frozen samples were it is recommended that the percent recovery stored in "Kimble" type screw-cap 25 by 200 be determined by "spiking" a few samples mm. (Culture tubes. Salinity and unfrozen with the standard solution. total phosphate samples were stored in 4- ounce prescription "Duroglas" bottles. It Standard samples were prepared from was during the latter phase of the Alaska compounds of known purity and formula program that pyrex was found to be superior weight that contained the element, nutrient, to the softer glass, particularly for the etc., under consideration. The compounds analyses mentioned. Subsequent collections selected were readily soluble but were not (M/V Kingf ish ) employed pyrex storage con- hygroscopic or deliquescent to the extent tainers for all samples except salinities that they absorbed large amounts of atmos- which were stored in 4-ounce prescription- pheric moisture. The compounds used were type bottles. either bone dry or of a known moisture content. This enabled all weights to be We did not use polyethylene containers reduced to a bone-dry basis. Final amounts because the material acts as an adsorbent, used in standard samples were obtained by reducing the reliability of an analysis. dilution of the relatively concentrated Further, polyethylene becomes an effective stock solution. : Two compounds were usually employed vigorously. Clean each cap-glass junction in the preparation of each series of stand- with a jet of distilled water, remove caps, ard samples. We discouraged the practice and dispense 25-milliliter portion into the of using stock solution more than once. The sample flasks with a large bore 25-milli- practice of starting each series of standard liter graduated pipette. samples from the "dry" compounds rather than from preserved stock solutions, and also of 5. Add 1 drop of stannous chloride using two different compounds to a series reagent to the first sample flask and then of standard samples, enabled the analyst to swirl the flask thoroughly. Thirty seconds detect systematic errors that could reduce later, add a drop to the second flask and the accuracy of his data. swirl. Continue at 30-second intervals until all flasks have been treated. To correct for glassware and reagent contamination, blanks consisting of "pure" 6. Determine the density of the blue water were treated in the same manner as color that will develop if there is any unknown samples. Any color that developed phosphate present with a Fisher a.c. elec- was due to impurities which were corrected trophotometer equipped with a 650 A filter by subtracting the blank sample readings and a 23-milliliter cell. Color-density from those of the regular samples. System- determinations are made 30 minutes after the atic errors can be introduced by using addition of the stannous chloride reagent. a low quality distilled water in preparing If analyst cannot check a sample every 30 blank samples. We have found that a satis- seconds, increase the time interval in item factory grade can be prepared by distilling 5 as necessary. tap water two times in a borosilicate glass still.
Recommended publications
  • A Simple Spectrophotometric Method for the Determination Of
    J Clin Pathol: first published as 10.1136/jcp.14.2.202 on 1 March 1961. Downloaded from 202 Technical methods usual laboratory filters may give falsely increased values.) Of the clear and colourless filtrate 10 ml. is pipetted into A a 25 ml. measuring flask and 10 ml. reagent (3) added. simple spectrophotometric After mixing dilute with distilled water to the mark and method for the determination of renew mixing. After 10 minutes the mixture is ready for reading. The colour is stable, according to Watt and urea in blood and urine Chrisp, for 11 days. READING T. K. WITH, TOVE DREYER PETERSEN, AND BIRGIT PETERSEN From the Central Laboratory, Svend- The reactions are read in a spectrophotometer at 420 nm. borg County Hospital, Denmark with a blank prepared in the same way as the reaction mixture but with 3 ml. of distilled water instead of 3 ml. of blood. The urea standard (reagent 4) is treated in the Several methods for urea in determining blood and urine same way as the blood. Cuvettes of 1 cm. thickness are are in use in clinical laboratories, but none is ideal in suitable. mass routine The analysis. spectrophotometric method If p, s, and b are the extinctions of the the based on a modified Ehrlich sample, reagent proposed by Watt standard, and the blank respectively, the urea concen- and Chrisp (1954) for pure solutions is therefore of tration is interest because it can be modified for use in mass clinical analyses of blood and urine. Brown (1959) has l00(p-b)/(s-b) mg./100 ml.
    [Show full text]
  • I. Direct Titration of Sulfate, II. High Precision Spectrophotometric Analysis Max Quentin Freeland Iowa State College
    Iowa State University Capstones, Theses and Retrospective Theses and Dissertations Dissertations 1955 I. Direct titration of sulfate, II. High precision spectrophotometric analysis Max Quentin Freeland Iowa State College Follow this and additional works at: https://lib.dr.iastate.edu/rtd Part of the Analytical Chemistry Commons Recommended Citation Freeland, Max Quentin, "I. Direct titration of sulfate, II. High precision spectrophotometric analysis" (1955). Retrospective Theses and Dissertations. 14746. https://lib.dr.iastate.edu/rtd/14746 This Dissertation is brought to you for free and open access by the Iowa State University Capstones, Theses and Dissertations at Iowa State University Digital Repository. It has been accepted for inclusion in Retrospective Theses and Dissertations by an authorized administrator of Iowa State University Digital Repository. For more information, please contact [email protected]. INFORMATION TO USERS This manuscript has been reproduced from the microfilm master. UMi films the text directly from the original or copy submitted. Thus, some thesis and dissertation copies are in typewriter face, while others may be from any type of computer printer. The quality of this reproduction is dependent upon the quality of the copy submitted. Broken or indistinct print, colored or poor quality illustrations and photographs, print bleedthrough, substandard margins, and improper alignment can adversely affect reproduction. In the unlikely event that the author did not send UMI a complete manuscript and there are missing pages, these will be noted. Also, if unauthorized copyright material had to be removed, a note will indicate the deletion. Oversize materials (e.g., maps, drawings, charts) are reproduced by sectioning the original, beginning at the upper left-hand comer and continuing from left to right in equal sections with small overlaps.
    [Show full text]
  • Laboratory Equipment Used in Filtration
    KNOW YOUR LAB EQUIPMENTS Test tube A test tube, also known as a sample tube, is a common piece of laboratory glassware consisting of a finger-like length of glass or clear plastic tubing, open at the top and closed at the bottom. Beakers Beakers are used as containers. They are available in a variety of sizes. Although they often possess volume markings, these are only rough estimates of the liquid volume. The markings are not necessarily accurate. Erlenmeyer flask Erlenmeyer flasks are often used as reaction vessels, particularly in titrations. As with beakers, the volume markings should not be considered accurate. Volumetric flask Volumetric flasks are used to measure and store solutions with a high degree of accuracy. These flasks generally possess a marking near the top that indicates the level at which the volume of the liquid is equal to the volume written on the outside of the flask. These devices are often used when solutions containing dissolved solids of known concentration are needed. Graduated cylinder Graduated cylinders are used to transfer liquids with a moderate degree of accuracy. Pipette Pipettes are used for transferring liquids with a fixed volume and quantity of liquid must be known to a high degree of accuracy. Graduated pipette These Pipettes are calibrated in the factory to release the desired quantity of liquid. Disposable pipette Disposable transfer. These Pipettes are made of plastic and are useful for transferring liquids dropwise. Burette Burettes are devices used typically in analytical, quantitative chemistry applications for measuring liquid solution. Differing from a pipette since the sample quantity delivered is changeable, graduated Burettes are used heavily in titration experiments.
    [Show full text]
  • BLAUBRAND® Volumetric Instruments and Density Bottles
    BLAUBRAND® Volumetric Instruments and Density Bottles Testing Instructions (SOP) March 2015 1. Introduction The standard DIN EN ISO 4787 describes both the design and the testing of the volumetric instruments of glass. The following Testing Instructions describe how to apply the ISO standard in practice. We recommend a testing every 1-3 years. The interval depends on the using of aggressive media and the cleaning procedure. These Instructions may also be used as a basis for the supervision of testing devices to DIN EN ISO 9001, DIN EN ISO 10012 and DIN EN ISO/IEC 17025. The test of the Density Bottles was effected on the basis of DIN EN ISO 4787. Meniscus adjustment with BLAUBRAND® Volumetric Instruments Meniscus adjustment Meniscus adjustment with ring mark with Schellbach stripe Read at the lowest point of the meniscus. Read at the point where the two arrows touch. Meniscus adjustment reflexion of liquid surface meniscus ring mark dark paper (p.e. black, blue) 2 2. Preparation for testing 2. Clear identification of the volumetric instrument to be tested Batch number, individual serial number, trademark, nominal volume and tolerances are directly printed ⇒ The test starts with a clear identification of the on every BLAUBRAND volumetric instruments. volumetric instrument in the test record. 2.1 Copy Test Record (see page 13) 2.2 Serial number/Identification number ⇒ Enter into Test Record All BLAUBRAND® volumetric instruments always carry a batch number, e.g., 13.04, or an individual serial number in the case of individual certificates, e.g., 13.040371 (year of production 2013, Batch No.
    [Show full text]
  • Signature Redacted Department of Metallurgy Gau, R E Da Signature of Professor in Charge of Research
    THERMODYNAMIC PROPERTIES OF THE GROUP VIa SULFIDES: CrS, Mo2 S3 AND WS2 by JOHN PATRICK HAGER B. S., Montana School of Mines (1958) M. S. , School of Mines and Metallurgy University of Missouri (1960) Submitted in Partial Fulfillment of the Requirements for the Degree of DOCTOR OF SCIENCE at the MASSACHUSETTS INSTITUTE OF TECHNOLOGY August, 1969 Signature of Author Signature redacted Department of Metallurgy gau, r e da Signature of Professor in Charge of Research Signature of Chairman of Departmental Committee Graduate Research nchives Signature redacted INST. 7 DEC 2 1969 4IBRARIES -~ U ii THERMODYNAMIC PROPERTIES OF THE GROUP VIa SULFIDES: CrS, Mo 2S 3 and WS 2 by John Patrick Hager Submitted to the Department of Metallurgy on August 18, 1969, in partial fulfillment of the requirements for the degree of Doctor of Science. ABSTRACT The thermodynamic properties of the metal-saturated phase of the Group VIa sulfides have been determined through a study of the effect of gas composition on the sulfidizing action of (H2S g) + H 2 (g) ) mixtures passed over heated metal samples. The following equations for the standard free energy of formation of CrS (c) ' Mo2S3(c) and WS2(c) were obtained: AFCc) ( 250) = -48,190 ( 760) + 13.27( 0.52)T; cal/l/2 g-mole S2(g) (1375-1570 0 K) AF 0 ( 220) = -41,730( 890) + 17.39( 0.59)1R; cal/l/2 g-mole S2(g) 102S3 (c)2(g (1365-1610 0 K) AF ( 220) = -40,110( 920) + 18.64( 0.63)T; cal/l/2 g-mole S WS 2 (c) 2(g) (1370-15650K) An additional study of the thermodynamic properties of Cu2S ) pro- vided a means of evaluating the experimental technique.
    [Show full text]
  • Accuracy in Volume Measurements
    Activity 1: Accuracy in volume measurements Requirements ● 50 mL burette ● Water ● Burette clamp and stand ● 100 mL graduated cylinder ● Funnel ● 100 mL beaker Health and safety, and outline instructions 1. Set up a burette on a stand assembly (see Figure 1). Figure 1 2. Fill the empty burette with water. Make sure the meniscus of the water in the burette rests on the 0.00 mL line at the top of the burette (see Figure 2). 48 48 48 48 35 35 35 35 49 49 49 49 36 36 36 36 50 50 50 50 0 0 0 0 37 37 37 37 1 1 1 1 38 38 38 38 2 2 2 2 39 39 39 39 3 3 3 3 40 40 40 40 0 0 4 0 4 0 4 4 41 41 41 41 1 1 1 1 2 2 5 2 5 2 5 5 42 42 42 42 3 3 3 3 4 4 6 4 6 4 6 6 43 43 43 43 5 5 5 5 6 6 7 6 7 6 7 7 44 44 44 44 7 7 7 7 8 8 8 8 8 8 8 8 45 45 45 45 9 9 9 9 10 10 9 10 9 10 9 9 46 46 46 46 Figure 2 Closed ClosedClosedClosedClosedClosedOpenClosedOpenClosedClosedOpenClosedOpen Closed Closed 3. Place a 100 mL graduated cylinder beneath the tip of the burette. Open the stopcock and slowly release about 40 mL of water from the burette into the graduated cylinder. Record the exact volume of water dispensed from the burette in the appropriate column of the results table.
    [Show full text]
  • Ocean Drilling Program Technical Note 5
    This publication has been superseded by ODP Technical Note 15: http://www-odp.tamu.edu/publications/tnotes/tn15/f_chem1.htm WATER CHEMISTRY PROCEDURES ABOARD JOIDES RESOLUTION - SCME GOMMENTTS Joris Gieskes Gail Peretsman Scripps Institution of Oceanography Ocean Drilling Program La Jolla, California 92093 Texas A&M University College Station, Texas 77843-3469 OCEAN DRILLING PROGRAM TEXAS A&M UNIVERSITY TECHNICAL NOTE NUMBER 5 MAY 1986 Philip D. Rabinowitz Director Audrey W. Meyer Manager of Science Operations Louis E. Garrison Deputy Director Material in this publication may be copied without restraint for library, abstract service, educational or personal research purposes; however, republication of any portion requires the written consent of the Director, Ocean Drilling Program, Texas A&M University, College Station, Texas 77843-3469, as well as appropriate acknowledgement of this source. Technical Note No. 5 First Printing 1986 Distribution Copies of this publication may be obtained frαn the Director, Ocean Drilling Program, Texas A&M University, College Station, Texas 77843-3469. In seme case, orders for copies may require payment for postage and handling. DISCLAIMER This publication was prepared by the Ocean Drilling Program, Texas A&M University, as an account of work perfoπned under the international Ocean Drilling Program which is managed by Joint Oceanographic Institutions, Inc., under contract with the National Science Foundation. Funding for the program is provided by the following agencies: Department of Energy, Mines and
    [Show full text]
  • Mélanie Burette
    Mélanie Burette French, born 16th June 1993 [email protected] [email protected] Position: Post-doctoral researcher (since January 2021) CRBM CNRS UMR5237 Montpellier, France Supervisor : Dr. Cécile Gauthier-Rouvière Academic achievements 2020: Doctor of Philosophy in Cell Biology Université de Montpellier, France IRIM CNRS UMR9004 Montpellier, France Supervisor : Dr. Matteo Bonazzi 2017: Master of Science in Microbiology and Immunology Université de Montpellier, France. 2015: Bachelor of Science in Ecology and Organismal Biology Université de Montpellier, France. 2011: A-level in Science Lycée Jean Renoir, France Awards and honors 2019: IRIM Institute Meeting Grant 2019: European Society of Endocrinoloy Meeting Grant for the 44th symposium on Hormones and Cell Regulation. 2018: Winner “Prix du meilleur poster” of the 16th meeting of PhD students in Chemical and Biological Sciences for Health CBS2 Doctoral School, Montpellier, France. List of Publications 1. Burette M. *, Bienvenu A.*, Inamdar K., Bordignon B., Swain J., Muriaux D., Martinez E., Bonazzi M. (Submitted) Identification of a new Coxiella lipid- binding effector protein by a vacuolar lipid profiling. Page 1 of 4 2. Martinez E., Huc-Brandt S., Brelle S., Allombert J., Cantet F., Gannoun-Zaki L., BURETTE M., Martin M., Letourneur F., Bonazzi M. and Molle V. (2020) The Coxiella burnetii secreted protein kinase CstK influences vacuole development and interacts with the GTPase-activating protein TBC1D5. Journal of Biological Chemistry, 295(21): 7391-7403, doi: 10.1074/jbc.RA119.01011. 3. Burette M. *, Allombert J. *, Lambou K., Maarifi G., Nisole S., Case ED., Blanchet FP., Hassen-Khodja C., Cabantous S., Samuel J., Martinez E., Bonazzi M.
    [Show full text]
  • Titrations Titrations Are Done Often to Find out the Concentration of One Substance by Reacting It with Another Substance of Known Concentration
    Titrations Titrations are done often to find out the concentration of one substance by reacting it with another substance of known concentration. They are often done with neutralisation reactions, but can be done with redox reactions. One substance (generally the one we don’t know the concentration) is put in the conical flask. It is measured using a volumetric burette pipette. The other substance is placed in the burette However, the standard phrase: titrate solution A with solution B means that A should be in the conical conical flask and B should be in the burette. flask A conical flask is used in preference to a beaker because it is easier to swirl the mixture in a conical flask without spilling the contents. pipette Method for Titration Make sure bottom of Using the pipette meniscus is on line on neck of pipette •rinse pipette with substance to go in it (often alkali). •pipette 25 cm3 of solution A into conical flask. The volumetric pipette will have a mark on its neck to show the level to fill to. The bottom of the meniscus should sit on this line. •touch surface of solution with pipette ( to ensure correct amount is added). A small amount of solution will be left in the pipette at this stage. The calibration of the pipette will take into account this effect. It should not be forced out. N Goalby chemrevise.org 1 Using the burette The burette should be rinsed out with substance that will be put in it. If it is not rinsed out the acid or alkali added may be diluted by residual water in the burette or may react with substances left from a previous titration.
    [Show full text]
  • CHM 130 - Accuracy and the Measurement of Volume
    CHM 130 - Accuracy and the Measurement of Volume PURPOSE: The purpose of this experiment is to practice using various types of volume measuring apparatus, focusing on their uses and accuracy. DISCUSSION: Volume measuring apparatus come in several different designs – graduated cylinders, volumetric flasks, pipets, burets, etc. Each design has a different application and a different accuracy. We are going to study these applications and the accuracy of the designs. In general, the less accurate the apparatus is, the easier and faster it is to use. So if great accuracy is not needed, why not be practical and use the fast and easy apparatus. In an experiment, the measurements made using a volume measuring apparatus should be at least as accurate as all the other measurements made in the experiment. For this reason, it is important to know the accuracy of different apparatus that are available. There are two kinds of errors in measurements. ACCURACY is the error associated with how close a measurement is to the true or actual value. If an instrument gives values that are very close to the true value we say that it is ACCURATE. Example: A graduated cylinder upon measuring the same sample three times gave 566 mL, 584 mL, and 541 mL. The average of these three values is 563.7 mL. If the true value was 563.688 mL, we would say that the average was accurate but the individual measurements were neither accurate nor precise. PRECISION is the error associated with how close several measurements of the same quantity are to each other.
    [Show full text]
  • Error Analysis Example Error Analysis Is Always a Difficult Area for Students
    Error Analysis Example Error analysis is always a difficult area for students. However, the careful consideration of experimental error is one of the important skills that we need to learn to be effective scientists. In the following discussion, the errors in a titration experiment are considered. The first section is a detailed look at how to determine the most important errors. The second section is an example of the corresponding text that would be written in a lab report for CH141-142. Determining the Important Errors The purpose of the error analysis section of the lab report is to determine the most important errors and the effect that those errors have on the final result. Random Errors: Random errors cause positive and negative deviations from the average value of a measurement. Random errors cancel by averaging, if the experiment is repeated many times. Upon averaging many trials, random errors have an effect only on the precision of a measurement. The effect of random errors is primarily on the precision. Every non-integer experimental measurement is a source of random error. The random error is estimated from the readability of the device. A table of typical measurements and the associated precision, under practical circumstances, is given below. For instrument readings, to avoid round-off error, report one extra significant figure and then underline the digit that is not significant. Volumetric flasks and pipettes precision (relative) significant figures 10 mL 0.03 mL 3 i.e. 10.0 mL 25 mL 0.03 mL 3 25.0 mL 50 mL 0.05 mL 3 50.0 mL 100 mL 0.08 mL 4 100.0 mL Auto-pipettors 10 L 0.05 L (0.5%) 3 i.e.
    [Show full text]
  • Titrette Digital Bottle-Top Burette Leaflet
    The next generation Digital Burette! Titrette® class A precision FIRST CLASS·BRAND With the Titrette® bottle-top burette, available in 10 ml, 25 ml, and 50 ml sizes, you can titrate quickly and reliably with highest precision, even in close quarters, with no power hookup needed – in the lab, in production, or in the field! I High accuracy Precise titration within the error limits of Class A I Smooth, low-force operation For sensitive, drop-wise titration I Compact design Light and stable I Easy handling No switching between 'fill' and 'titrate' I Simple maintenance Cleaning and part replacement right in the lab I PC interface (optional) Data transmission directly to the PC A closer look The control elements of the Titrette® bottle-top burette have an intuitive layout. Separate buttons for On/Off and Pause; CLEAR/Select button Clear button allows user to reset the display and select functions. Easy-grip hand wheel action and smooth precision gears for fast or drop-wise titration make handling extremely simple and reliable. PC interface (optional) (see ordering information) Digital display Replaceable batteries On/Off button Pause button Hand wheels Piston Inspection window Dispensing cylinder Titration and recirculating valve Valve block Freely rotating (Bottle thread GL 45) Titrating tube horizontally and vertically adjustable, with integrated discharge valve Recirculating tube Telescoping filling tube Screw cap Extras Four additional helpful electronic functions make your work easier. Hold down the CLEAR button to select the desired function: Adjustment with Calibration Easy Calibration schedule With Easy Calibration To save a date for the technology, you can next calibration, simply easily make calibra- store it under 'GLP'.
    [Show full text]