Spectrophotometry Laboratory

Total Page:16

File Type:pdf, Size:1020Kb

Spectrophotometry Laboratory CELL BIOLOGY SPECTROPHOTOMETRY Spectrophotometry Laboratory INTRODUCTION hemoglobin protein, these molecules are colorless in solution, that is, they do not absorb visible light. Spectrophotometry refers to the measurement of However, both nucleic acids and proteins absorb radiant energy as a function of wavelength. A radiation in the ultraviolet region of the electro- spectrophotometer consists of a monochromator and a photocell connected electrically to a meter or recorder. The monochromator provides a discrete narrow wavelength band from a selected part of the electromagnetic spectrum (ultraviolet, visible, infrared). The beam of light is focused on the photocell and the light intensity is measured. Light intensity is compared before and after a solution is placed in the beam of light. Spectrophotometry is useful in two ways. (1) It can be used to characterize and identify a com- pound. When the instrument is used for this purpose light absorption is measured over a range of wavelengths. Generally a graph, called an absorption spectrum, is made plotting the light absorption versus wavelength. (2) Spectrophoto- magnetic spectrum. The figure above shows metry can also be used to determine the concen- characteristic spectra for salmon sperm DNA and tration of a compound in solution. When the egg albumin, a protein. Notice first that while the instrument is used for this purpose the amount of heights of the two absorption maxima are the same, light absorbed by a solution at one wavelength is they occur at different wavelengths, 280 nm for the determined. Both of these uses of spectrophoto- protein and 260 nm for the DNA. Second, the metry will be demonstrated in this exercise. concentration of the two solutions is very different: 0.1% for protein and 0.007% for DNA. These differences result from the number and nature of the Spectrophotometry identifies compounds absorbing groups in the two compounds. DNA absorption at 260 nm is caused by the purine and Oxyhemoglobin and deoxyhemoglobin differ in pyrimidine bases in the molecule. Protein composition only by the presence of oxygen bound absorption at 280 nm is caused by the amino acids to the heme of oxyhemoglobin. Binding of oxygen tryptophan and tyrosine in the molecule. Purine and to the large hemoglobin moiety makes a profound pyrimidine bases are stronger absorbers per residue difference in the absorption spectrum of these than the amino acids. More bases are also found per molecules. The comparison in this laboratory of unit weight of nucleic acid than are the tryptophan these two “red” heme compounds with one another and tyrosine per unit weight of protein. As a result, points out how spectrophotometry can be used to nucleic acids have a higher absorption coefficient identify molecules. (optical density per unit concentration) than For further illustration and example consider proteins. most proteins and nucleic acids. In contrast to the 1 CELL BIOLOGY SPECTROPHOTOMETRY Measurements & Laws of Spectrophotometry Spectrophotometry determines concentration The amount of light absorbed may be expressed As a demonstration of the use of spectrophoto- in two different ways. One way is as the percent of metry to determine concentration, the concentration the total incident intensity that passes through the of a protein will be established. Because, as sample mentioned above, most proteins do not absorb in the visible spectrum and this laboratory uses only %T = I/I x 100% o visible light spectrophotometers we must use a where Io is the incident intensity and I, the intensity means to “color” the protein so it will absorb in the of light leaving the sample. The other way of visible spectrum. A simple protein assay will be expressing the amount of light absorbed is the used; the Biuret test is specific for the presence of optical density (O.D.) or absorbance (A) peptide bonds. Known concentrations of proteins will be used to construct a standard curve, which is A = log (Io/I) = – log T. a plot of absorbance versus concentration. Using If a parallel beam of monochromatic light passes this curve you will be able to obtain the through a homogeneous absorbing medium, the concentration of protein of unknown concentration. intensity the radiation is decreased by a constant fraction for each unit thickness of the medium. This relationship between amount of radiation absorbed PROCEDURES and thickness of the absorbing substance is known as Lambert’s Law. 1. Qualitative Absorption In solutions, the absorbance is proportional to the number of molecules of absorbing substance per Direct the hand spectroscope toward outdoor unit volume of solution. Beer’s Law states that light. Adjust it such that a complete spectrum can there is a linear relationship between concentration be observed. Then place colored filters in front of in moles per liter and absorbance, and only holds if the slit and observe what wavelengths are the specific absorption per molecule does not vary attenuated. with concentration. These two laws may be combined into the Beer-Lambert Law 2. Operation of the Digital “Spectronic 20D+” A = αcl 1. Plug in the spectrophotometer. where α is the molar absorption coefficient, c is the 2. Turn it on with left front knob (%T). concentration of solute in moles per liter, and l is 3. Set wavelength with the knob on top. the path length in centimeters. Most spectrophoto- 4. Push the Filter selector lever (bottom, front) to meters have a linear % T scale; however, the correspond to the appropriate wavelength. convenience of this scale is not the most important 5. When the machine has warmed up (5 minutes) consideration. The last formula indicates that check to see that the “Transmittance” MODE is absorbance is linearly related to concentration. set. Therefore, absorbance is the most relevant scale for 6. Adjust the Transmittance to 0.0 using the LEFT most spectrophotometer applications. knob (%T) on the front of the machine. 7. Press the MODE button to read “Absorbance”. 8. Insert blank into sample holder, close the lid. 9. Adjust the Absorbance to 0.00 with RIGHT front knob (100%T/0A). 2 CELL BIOLOGY SPECTROPHOTOMETRY 10. Without changing any settings, remove the of the unknown concentration will be treated with blank, insert the sample and close the lid. Biuret reagent. The absorbance of all the samples 11. Read the absorbance. will be determined. The absorbance of the known 12. You may continue to place other samples in the concentrations will then be used to construct a machine and read their absorbance, or if you standard curve that will be used to determine the change the wavelength, repeat steps 8 through concentration of the unknown. 11. Note that all pipettings should be accurately done. Use one 2-ml pipette for the known concentration and its dilutions and a second 2-ml 3. Quantitative Absorption Spectra pipette for the unknown concentration. Prepare and examine hemoglobin and erythrosin 1. Place seven cuvettes in a rack. Label one “B” solutions. for blank and the remainder label 1-6. Pipette 2 1. Place about 10 ml of distilled water in a clean ml of distilled water into “B” and cuvettes 2-5. test tube. Using a Pasteur pipette, add about 2 Note do not add water to cuvette 1. drops of sheep blood. Cover the tube with 2. Pipette 2 ml of the 2.5 mg/ml stock albumin parafilm and mix by inverting several times. solution into cuvette 1. Pipette an additional 2 2. Divide this solution between two cuvettes. Set ml into cuvette 2. This produces a two-fold one aside; this is the “oxyhemoglobin”. dilution. 3. Add a small pinch of sodium hydrosulfite to the 3. Mix the protein and water in cuvette 2 by other tube. Cover this tube with parafilm again drawing the solution into the pipette and gently and invert several times. This is the blowing it back into the cuvette several times. “deoxyhemoglobin”. The color should be 4. Next withdraw 2 ml from cuvette 2 and add that noticeably bluer than in the other tube. to cuvette 3. Mix as before and transfer 2 ml 4. The oxyhemoglobin and deoxyhemoglobin from cuvette 3 to cuvette 4. samples are now ready to run. 5. Mix again and transfer 2 ml to cuvette 5. When 5. Fill two additional cuvettes about one third full cuvette 5 has been mixed, remove 2 ml and with the erythrosin solutions. One cuvette discard. Each cuvette should now contain should have the 1/20,000 concentration and the exactly 2 ml. Note also that each cuvette now second should have the 1/60,000 concentration. contains half the protein of the preceding 6. Use a fifth cuvette as a blank. cuvette in the series. 7. Using a Spec. 20, measure the absorbance in 10 6. Pipette 2 ml of the “unknown” protein nm increments of oxyhemoglobin, deoxyhemo- concentration into cuvette 6. globin, and the 1/20,000 and 1/60,000 erythrosin 7. To each cuvette now add 3 ml Biuret reagent solutions. Cover the range from 400 to 650 nm. and mix. Incubate at 37°C for 30 minutes. Think about how you want to do this with the 8. Using the blank first, read the absorbance of all fact that the “Spec” has to be “blanked” after the protein cuvettes at 550 nm in the Spec. 20. each change in wavelength. 4. Quantitative Determination of Protein Determine the concentration of a protein sample. A stock protein solution will be serially diluted by factors of two. These solutions along with a sample 3 CELL BIOLOGY SPECTROPHOTOMETRY DATA ANALYSIS Qualitative Absorption Compare your observations of the spectrum with the different filters. What region(s) of the spectrum would you expect a blue solution to absorb? A green solution? Can you make a generalized statement about “color” and absorbed wavelengths? Quantitative Absorption Plot all data for oxyhemoglobin, deoxyhemo- globin, and the two erythrosin concentrations on this graph.
Recommended publications
  • Use of Laboratory Equipement
    USE OF LABORATORY EQUIPEMENT C. Laboratory Thermometers Most thermometers are based upon the principle that liquids expand when heated. Most common thermometers use mercury or colored alcohol as the liquid. These thermometers are constructed as that a uniform-diameter capillary tube surmounts a liquid reservoir. To calibrate a thermometer, one defines two reference points, normally the freezing point of water (0°C, 32°F) and the boiling point of water (100°C, 212°F) at 1 tam of pressure (1 tam = 760 mm Hg). Once these points are marked on the capillary, its length is then subdivided into uniform divisions called degrees. There are 100° between these two points on the Celsius (°C, or centigrade) scale and 180° between those two points on the Fahrenheit (°F) scale. °F = 1.8 °C + 32 The Thermometer and Its Calibration This section describes the proper technique for checking the accuracy of your thermometer. These measurements will show how measured temperatures (read from thermometer) compare with true temperatures (the boiling and freezing points of water). The freezing point of water is 0°C; the boiling point depends upon atmospheric pressure but at sea level it is 100°C. Option 1: Place approximately 50 mL of ice in a 250-mL beaker and cover the ice with distilled water. Allow about 15 min for the mixture to come to equilibrium and then measure and record the temperature of the mixture. Theoretically, this temperature is 0°C. Option 2: Set up a 250-mL beaker on a wire gauze and iron ring. Fill the beaker about half full with distilled water.
    [Show full text]
  • Laboratory Glassware N Edition No
    Laboratory Glassware n Edition No. 2 n Index Introduction 3 Ground joint glassware 13 Volumetric glassware 53 General laboratory glassware 65 Alphabetical index 76 Índice alfabético 77 Index Reference index 78 [email protected] Scharlau has been in the scientific glassware business for over 15 years Until now Scharlab S.L. had limited its sales to the Spanish market. However, now, coinciding with the inauguration of the new workshop next to our warehouse in Sentmenat, we are ready to export our scientific glassware to other countries. Standard and made to order Products for which there is regular demand are produced in larger Scharlau glassware quantities and then stocked for almost immediate supply. Other products are either manufactured directly from glass tubing or are constructed from a number of semi-finished products. Quality Even today, scientific glassblowing remains a highly skilled hand craft and the quality of glassware depends on the skill of each blower. Careful selection of the raw glass ensures that our final products are free from imperfections such as air lines, scratches and stones. You will be able to judge for yourself the workmanship of our glassware products. Safety All our glassware is annealed and made stress free to avoid breakage. Fax: +34 93 715 67 25 Scharlab The Lab Sourcing Group 3 www.scharlab.com Glassware Scharlau glassware is made from borosilicate glass that meets the specifications of the following standards: BS ISO 3585, DIN 12217 Type 3.3 Borosilicate glass ASTM E-438 Type 1 Class A Borosilicate glass US Pharmacopoeia Type 1 Borosilicate glass European Pharmacopoeia Type 1 Glass The typical chemical composition of our borosilicate glass is as follows: O Si 2 81% B2O3 13% Na2O 4% Al2O3 2% Glass is an inorganic substance that on cooling becomes rigid without crystallising and therefore it has no melting point as such.
    [Show full text]
  • PML Brochure
    PHYSICAL MEASUREMENT LABORATORY Gauging nature on all scales NIST.GOV/PML PHYSICAL MEASUREMENT LABORATORY (PML) The Physical Measurement Laboratory (PML), a major frequency, electricity, temperature, humidity, pressure operating unit of the National Institute of Standards and vacuum, liquid and gas flow, and electromag- and Technology (NIST), sets the definitive U.S. standards netic, optical, acoustic, and ionizing radiation. PML for nearly every kind of measurement in modern life, collaborates directly with industry, universities, profes- sometimes across more than 20 orders of magni- sional and standards-setting organizations, and other tude. PML is a world leader in the science of physical agencies of government to ensure accuracy and to measurement, devising procedures and tools that make solve problems. It also supports research in many continual progress possible. Exact measurements are fields of urgent national importance, such as manu- absolutely essential to industry, medicine, the research facturing, energy, health, law enforcement and community, and government. All of them depend on homeland security, communications, military defense, PML to develop, maintain, and disseminate the official electronics, the environment, lighting and display, standards for a wide range of quantities, including radiation, remote sensing, space exploration, length, mass, force and shock, acceleration, time and and transportation. IMPACTS ❱ Provides 700 kinds of calibration services ❱ Numerous special testing services NIST.GOV/PML | 2 NIST.GOV/PML
    [Show full text]
  • Chemostat Culture for Yeast Experimental Evolution
    Downloaded from http://cshprotocols.cshlp.org/ at Cold Spring Harbor Laboratory Library on August 9, 2017 - Published by Cold Spring Harbor Laboratory Press Protocol Chemostat Culture for Yeast Experimental Evolution Celia Payen and Maitreya J. Dunham1 Department of Genome Sciences, University of Washington, Seattle, Washington 98195 Experimental evolution is one approach used to address a broad range of questions related to evolution and adaptation to strong selection pressures. Experimental evolution of diverse microbial and viral systems has routinely been used to study new traits and behaviors and also to dissect mechanisms of rapid evolution. This protocol describes the practical aspects of experimental evolution with yeast grown in chemostats, including the setup of the experiment and sampling methods as well as best laboratory and record-keeping practices. MATERIALS It is essential that you consult the appropriate Material Safety Data Sheets and your institution’s Environmental Health and Safety Office for proper handling of equipment and hazardous material used in this protocol. Reagents Defined minimal medium appropriate for the experiment For examples, see Protocol: Assembly of a Mini-Chemostat Array (Miller et al. 2015). Ethanol (95%) Glycerol (20% and 50%; sterile) Yeast strain of interest Equipment Agar plates (appropriate for chosen strain) Chemostat array Assemble the apparatus as described in Miller et al. (2013) and Protocol: Assembly of a Mini-Chemostat Array (Miller et al. 2015). Cryo deep-freeze labels Cryogenic vials Culture tubes Cytometer (BD Accuri C6) Glass beads, 4 mm (sterile; for plating yeast cells) Glass cylinder Kimwipes 1Correspondence: [email protected] © 2017 Cold Spring Harbor Laboratory Press Cite this protocol as Cold Spring Harb Protoc; doi:10.1101/pdb.prot089011 559 Downloaded from http://cshprotocols.cshlp.org/ at Cold Spring Harbor Laboratory Library on August 9, 2017 - Published by Cold Spring Harbor Laboratory Press C.
    [Show full text]
  • White Paper Absorbance Or Fluorescence: Which Is the Best
    WHITE PAPER No. 40 Absorbance or Fluorescence: Which Is the Best Way to Quantify Nucleic Acids? Natascha Weiß1, Martin Armbrecht1 ¹Eppendorf AG, Hamburg, Germany Executive Summary When performing molecular experiments on nucleic acids, it is a basic requirement to determine the concentration as well as the quality of the sample. The standard method which serves this purpose is UV-Vis spectrophotometry – but is it always the right choice? In the present White Paper, the advantages and disadvantages of this technique will be described and it will be compared with nucleic acid quantification via fluorescence. In this context, it will be discussed which situations warrant the use of which one of the two methods. The decision will generally depend on the condition of the sample as well as on the requirements of downstream applications. Since the advantages of both methods complement each other well, it is most practical to have the ability to perform both methods on a single instrument in a flexible manner. Introduction Experiments involving nucleic acids are a mainstay in any evaluated by measuring the sample at additional wave molecular laboratory. DNA and RNA are isolated from micro- lengths (230 nm, 280 nm) and calculating the purity ratios, organisms and from cells of higher order organisms in order i.e. the ratios of the values obtained at 260/230 nm and at to be employed in a broad variety of processing steps and 260/280 nm, respectively. In this way, it will be evident analyses. It is crucial for any type of downstream applica- whether cellular debris or remainders of reagent used during tion that a defined amount of nucleic acid is used and that purification such as proteins, sugar molecules, certain salts the sample is free from contaminations that may impact the or phenols, are present in the solution, as these will generate experiments.
    [Show full text]
  • On Food, Spectrophotometry, and Measurement Data Processing (Keynote Lecture)
    12th IMEKO TC1 & TC7 Joint Symposium on Man Science & Measurement September, 3–5, 2008, Annecy, France ON FOOD, SPECTROPHOTOMETRY, AND MEASUREMENT DATA PROCESSING (KEYNOTE LECTURE) Roman Z. Morawski Warsaw University of Technology, Faculty of Electronics and Information Technology, Institute of Radioelectronics Warsaw, Poland [email protected] Abstract: Spectrophotometry is getting more and more Despite philosophical abnegation of food issues, the often the method of choice not only in laboratory analysis of practice of food preparation and refinement has flourished (bio)chemical substances, but also in the off-laboratory for centuries, and inspired research-and-invention-oriented identification and testing of physical properties of various minds. Today, we may speak about a fully developed products, in particular – of various organic mixtures discipline of science and technology. Enough to say that the including food products and ingredients. Specialized Institute of Food Technologists – the largest international, spectrophotometers, called spectrophotometric analyzers are non-profit professional organization involved in the designed for such applications. This keynote lecture is on advancement of food science and technology – is the state of the art and developmental trends in the domain encompassing 23 000 members worldwide. Its Committee of spectrophotometric analyzers of food with particular on Higher Education provided the following definition of emphasis on wine analyzers. The following issues are food science: "Food science is the discipline in which the covered: philosophical and methodological background of engineering, biological, and physical sciences are used to food analysis, physical and metrological principles of study the nature of foods, the causes of deterioration, the spectrophotometry, the role of measurement data processing principles underlying food processing, and the improvement in spectrophotometry, food analyzers on the market and of foods for the consuming public" [1].
    [Show full text]
  • BROOKFIELD DIAL READING VISCOMETER with Electronic Drive
    BROOKFIELD DIAL READING VISCOMETER with Electronic Drive Operating Instructions Manual No. M00-151-I0614 SPECIALISTS IN THE MEASUREMENT AND CONTROL OF VISCOSITY with offices in : Boston • Chicago • London • Stuttgart • Guangzhou BROOKFIELD ENGINEERING LABORATORIES, INC. 11 Commerce Boulevard, Middleboro, MA 02346 USA TEL 508-946-6200 or 800-628-8139 (USA excluding MA) FAX 508-946-6262 INTERNET http://www.brookfieldengineering.com TABLE OF CONTENTS I. INTRODUCTION .....................................................................................5 I.1 Components .......................................................................................................5 I.2 Utilities ................................................................................................................6 I.3 Specifications .....................................................................................................6 I.4 Set-Up ................................................................................................................7 I.5 IQ, OQ, PQ .........................................................................................................7 I.6 Safety Symbols and Precautions .......................................................................8 I.7 Cleaning .............................................................................................................8 II. GETTING STARTED ..............................................................................9 II.1 Operation ...........................................................................................................9
    [Show full text]
  • ELISA Plate Reader
    applications guide to microplate systems applications guide to microplate systems GETTING THE MOST FROM YOUR MOLECULAR DEVICES MICROPLATE SYSTEMS SALES OFFICES United States Molecular Devices Corp. Tel. 800-635-5577 Fax 408-747-3601 United Kingdom Molecular Devices Ltd. Tel. +44-118-944-8000 Fax +44-118-944-8001 Germany Molecular Devices GMBH Tel. +49-89-9620-2340 Fax +49-89-9620-2345 Japan Nihon Molecular Devices Tel. +06-6399-8211 Fax +06-6399-8212 www.moleculardevices.com ©2002 Molecular Devices Corporation. Printed in U.S.A. #0120-1293A SpectraMax, SoftMax Pro, Vmax and Emax are registered trademarks and VersaMax, Lmax, CatchPoint and Stoplight Red are trademarks of Molecular Devices Corporation. All other trademarks are proprty of their respective companies. complete solutions for signal transduction assays AN EXAMPLE USING THE CATCHPOINT CYCLIC-AMP FLUORESCENT ASSAY KIT AND THE GEMINI XS MICROPLATE READER The Molecular Devices family of products typical applications for Molecular Devices microplate readers offers complete solutions for your signal transduction assays. Our integrated systems γ α β s include readers, washers, software and reagents. GDP αs AC absorbance fluorescence luminescence GTP PRINCIPLE OF CATCHPOINT CYCLIC-AMP ASSAY readers readers readers > Cell lysate is incubated with anti-cAMP assay type SpectraMax® SpectraMax® SpectraMax® VersaMax™ VMax® EMax® Gemini XS LMax™ ATP Plus384 190 340PC384 antibody and cAMP-HRP conjugate ELISA/IMMUNOASSAYS > nucleus Single addition step PROTEIN QUANTITATION cAMP > λEX 530 nm/λEM 590 nm, λCO 570 nm UV (280) Bradford, BCA, Lowry For more information on CatchPoint™ assay NanoOrange™, CBQCA kits, including the complete procedure for this NUCLEIC ACID QUANTITATION assay (MaxLine Application Note #46), visit UV (260) our web site at www.moleculardevices.com.
    [Show full text]
  • Information Technology Laboratory Technical Accomplishments
    CONTENTS Director’s Foreword 1 ITL at a Glance 4 ITL Research Blueprint 6 Accomplishments of our Research Program 7 Foundation Research Areas 8 Selected Cross-Cutting Themes 26 Industry and International Interactions 36 Publications 44 NISTIR 7169 Conferences 47 February 2005 Staff Recognition 50 U.S. DEPARTMENT OF COMMERCE Carlos M. Gutierrez, Secretary Technology Administration Phillip J. Bond Under Secretary of Commerce for Technology National Institute of Standards and Technology Hratch G. Semerjian, Jr., Acting Director About ITL For more information about ITL, contact: Information Technology Laboratory National Institute of Standards and Technology 100 Bureau Drive, Stop 8900 Gaithersburg, MD 20899-8900 Telephone: (301) 975-2900 Facsimile: (301) 840-1357 E-mail: [email protected] Website: http://www.itl.nist.gov INFORMATION TECHNOLOGY LABORATORY D IRECTOR’S F OREWORD n today’s complex technology-driven world, the Information Technology Laboratory (ITL) at the National Institute of Standards and Technology has the broad mission of supporting U.S. industry, government, and Iacademia with measurements and standards that enable new computational methods for scientific inquiry, assure IT innovations for maintaining global leadership, and re-engineer complex societal systems and processes through insertion of advanced information technology. Through its efforts, ITL seeks to enhance productivity and public safety, facilitate trade, and improve the Dr. Shashi Phoha, quality of life. ITL achieves these goals in areas of Director, Information national priority by drawing on its core capabilities in Technology Laboratory cyber security, software quality assurance, advanced networking, information access, mathematical and computational sciences, and statistical engineering. utilizing existing and emerging IT to meet national Information technology is the acknowledged engine for priorities that reflect the country’s broad based social, national and regional economic growth.
    [Show full text]
  • Enzyme Analysis by Spectrophotometry - Plate Reader Method
    Enzyme Analysis by Spectrophotometry - Plate Reader method Enzyme Extraction Equipment and reagents: Machine/Product Reference (Company, Type, …) Centrifuge – cooled Eppendorf 5415R with F45-24-11 rotor Mixer Mill / Cryo Mill Retsch MM 400 with 2x PTFE Adapter rack for 10 reaction vials 1.5 and 2.0 ml. Microtube Vortex Ika Vortex 1 Plate reader BioTek PowerWave HT with Gen 5 software Microcentrifuge tubes Safe-lock, 1.5 and 2.0 ml Pipettes (+Multistep / Multichannel) 1-5 ml; 0.1-1 ml; 5-100 µl 8-tube strip PCR tubes Crushed Ice Liquid Nitrogen (+ thermos jar) PVP (Polyvinylpyrrolidone, PVP40) Sigma PVP40T TRIS (Tris(hydroxymetyl)-aminoethane) -1 99 %, 121.14 g.mol Sigma T1378 Na 2-EDTA (disodium salt dehydrate) -1 99 %; 372.2 g.mol Sigma ED2SS DTT (1,4-Dithiothreitol) -1 ≥ 99 %; 154.24 g.mol Sigma 43815 Hydrochloric Acid concentrated HCl -1 -1 -1 37 %; 1,19 g.ml ; 36.46 g.mol = 12.08 mol.l VWR 20252.290 Most buffers can be prepared in advance and kept at 4°C Requirement: the samples must be collected in 1.5 or 2.0 ml Eppendorf tubes at harvest. Prepare reagent working solutions: Extraction buffer (0.1 M TRIS ; 1 mM Na 2-EDTA; 1 mM DTT, pH 7.8) Dissolve 12.114 g TRIS (121.14 g/mol) + 0.3722 mg EDTA (372 g/mol) + 0.1542 g DTT (154.2 g/mol) in 900 ml distilled water, adjust to pH 7.8 with HCl (use 5M HCl) and dilute to 1 liter. Keep the extraction buffer on ice.
    [Show full text]
  • Microlab 500-Series
    MicroLab 500-series Getting Started 2 Contents CHAPTER 1: Getting Started Connecting the Hardware…………………………………………………………………………….…...6 Installing the USB driver…………………….…………………………………………………...………..6 Installing the Software…………………………………………………………………………..………...8 Starting a new Experiment………………………………………………………………………..……….8 CHAPTER 2: Setting up a MicroLab Experiment Adding Input Sources Sensor Variables……………………………………………………………………………….…………10 Adding a Sensor ………………………………………………………………………………………….10 Calibrating a Sensor………………………………………………………………………………………11 The Calibration Module………………………………………………………………………………......11 Adding a Calibration Point……………………………………………………………………………….11 Correlating the Data……………………………………………………………………………………....12 Adding More Variables………………………………………………………………………………...…14 Formula Variables………………………………………………………………………………………...15 Designing an Experiment The Programming Steps………………………………………………………………………..………..17 Displaying the Data…………………………………………………………………………..…….……..17 Controlling an Experiment Starting the Experiment…………………………………………………………………….………..…..18 Stopping the Experiment…………………………………………………………………….………......18 Repeating an Experiment………………………………………………………………………………..18 Switches A & B……………………………………………………………………………………………19 Analyzing the Data Graph Analysis…………………………………………………………………………..………..………19 Add a Curve Fit………………………………………………………………………….....……19 Data Analysis…………………….…………………………………………………………………..……22 CHAPTER 3 - Spectrophotometry (522/516) Calibrating the Spectrophotometer Reading a Blank………...…………………………………………………………………………..…….23 3 Reading
    [Show full text]
  • 2017 Automated Liquid Handling Resource Guide
    The Lab Manager AUTOMATED LIQUID HANDLING RESOURCE GUIDE What You Need to Know When Buying an Automated Liquid Handling System BY RYAN ACKERMAN Upgrading Stand-Alone Automated Liquid Handling Systems to Workstations BY ANGELO DEPALMA, PhD Small-Volume Liquid Handling Requires Advanced Technology BY MIKE MAY, PhD 20 10 Common Liquid Handling Errors & How to Avoid Them 17 BY LAB MANAGER The Latest in Pipette Tip Design BY ANGELO DEPALMA, PhD Automated Liquid Handling Product Finder BY LAB MANAGER Automatic Liquid Handling Resource Guide 2017 What You Need to Know When Buying an Automated Liquid Handling System Automated liquid handlers come in a seemingly endless variety of configurations, with many different specifications. By Ryan Ackerman MAINTENANCE TIP: AUTOMATED LIQUID HANDLING The signs that you should get your automated liquid handler serviced are fairly obvious. If the system is constantly experiencing glitches or producing inconsistent, unreliable results, it’s probably time to do some maintenance. Mechanical problems are another hint that you may want to call your service technician—these issues can include pipette tips being out of calibration, the deck not being “framed” correctly, belts being worn out, or the pipette stages being out of alignment. Why is it important to know the typical sample How does the sensitivity to contamination affect the type volume being used? of automated liquid handler required? Automated liquid handlers come in a seemingly endless variety of As the instrumentation used to analyze samples becomes more configurations, with many different specifications. An important one sensitive, the methods used to transport and prepare the samples to consider is what minimum, or maximum dispensing volume is must become resistant to contamination to ensure no carryover correct for your processes.
    [Show full text]