<<

and calibration of Outline photon and beams • General considerations with cavity ion chambers • Calibration of ion chambers Chapter 13 – For photon beams F.A. Attix, Introduction to Radiological and – For electron beams Dosimetry • Reference dosimetry of photon beams Almond et al., AAPM’s TG-51 protocol for clinical reference • Reference dosimetry of electron beams dosimetry of high- photon and electron beams, Med. Phys. 26, pp.1847-1870, 1999 McEwen al., Addendum to the AAPM’s TG-51 protocol, Med. Phys. 41, pp. 041501-1-20, 2014

Introduction Ion chamber calibration

• The success of depends on • Ion chamber can serve as an absolute dosimeter if its the accuracy of a prescribed dose delivery gas mass is known • This necessitates high accuracy in the • of the commercially manufactured ion chambers dosimetry of high-energy photon and are not constructed with exactly known sensitive electron beams volume, therefore they require calibration • Two aspects are involved: • National laboratories maintain standard – proper calibration of the measuring instruments chambers and calibrated g-ray beams (ionization chamber and electrometer) • Regional calibration laboratories (ADCL - Accredited – characterization of clinical beams Dosimetry Calibration Laboratories in US) provide calibration services for general-use instruments for a fee

Ion chamber calibration Ion chamber calibration

• Starting from an ion chamber calibrated free-in-air • Three approaches to ion chamber calibration: for one (exposure or air ) and – Exposure Nx transferring this information to obtain another – Dose in cavity gas N – old TG-21 protocol quantity, absorbed dose to water, based on a gas measurement in a phantom introduces complexity – Absorbed dose in water ND– new TG-51 protocol and possible errors • Beam dosimetry can be done • To overcome these complexities, primary – In free space standards laboratories have developed standards for absorbed dose to water in photon beams from – In water phantom (need correction for 60Co and accelerator beams and these have an perturbation due to chamber insertion) uncertainty of 1% or

1 TG-51 protocol General formalism • Prescribes a methodology for clinical reference • Given NQ (in Gy/C or Gy/rdg), the absorbed- dosimetry D, dose to water calibration for an ion chamber • Applies to photon beams with nominal located in a beam of quality Q between 60Co and 50 , and electron beams • Under reference conditions: with nominal energies between 4 and 50 MeV Q Q • Uses ion chamber calibrated in terms of absorbed D w  MN D ,w (Gy) 60 dose to water in a Co beam Q where D w is the absorbed dose to water (in Gy) at • Sets up certain well-defined reference conditions the point of measurement of the ion chamber when • Starting point: an ion chamber with calibration it is absent and M is the fully corrected factor directly traceable to national standards of electrometer reading in coulombs (C) or meter absorbed dose (may be done through ADCL) units (rdg)

General formalism: kQ General formalism: kQ

• Usually absorbed-dose calibration factors will be • Recall Spenser-Attix cavity theory, stating: 60 obtained for reference conditions in a Co beam med  L  M  W  • Define the quality conversion factor, kQ, such that D  D    P , D  med air   correction s air   60  m e Q Co   air air   air N D ,w  k Q N D ,w (Gy/C or Gy/rdg) • The quality conversion factor k is chamber specific w P – accumulates Q  L   corrections   P various chamber-   correction s  • Using kQ, gives Q  related corrections 60 N   air  Q Co D ,w   Q (, dose gradient, D  Mk N (Gy) • Thus k Q  60  change in electron w Q D ,w Co  w  N D ,w  L  fluence inside the   P   correction s  chamber, etc.)    air  60   Co

General formalism: kQ General formalism: kQ • For electron beams the quality conversion factor kQ contains two components: • For photon beams, the protocol provides Q k Q  Pgr k R values of kQ for most cylindrical ion 50 Q chambers used in reference dosimetry • P gr is necessary only for cylindrical chambers – corrects for the ionization gradient at the measurement point (extended list in Addendum to TG-51) – depends on the radius of the chamber cavity and • Plane-parallel chambers are not included – must be measured by the user, the protocol provides a procedure for measuring PQ in the user’s electron beam because there is insufficient information gr • kR50 is a chamber-specific factor, a function of electron about wall correction factors in photon beam quality as specified by R50 (depth in water where beams other than 60Co beams dose falls off to 50% of maximum dose)

2 General formalism: kQ General formalism • In an electron beam, the dose is given by • The factor kR50 is written as the product of: Q Q 60 Co k R  k R k ecal D  MP k  k N (Gy) 50 50 w gr R50 ecal D ,w

• kecal is the photon-electron conversion factor (fixed • The reference depth for electron-beam dosimetry is for a given chamber model), it is the value needed to at d = 0.6R – 0.1 cm, is essentially at the 60 Q ref 50 Co N ecal convert N Dw , into Dw , , the absorbed-dose depth of dose maximum for beams with energies <10 MeV but is deeper for higher-energy beams calibration factor in an electron beam of quality Qecal • At this depth the protocol can use of stopping- • k´R50, is the electron beam quality conversion factor, beam quality dependent, and converts into N Q power ratios, accounting for the realistic (not mono- D ,w energetic) energy distributions of electron beams

General formalism General formalism

• Cylindrical chambers are preferred dosimeters • To use this formalism one starts by obtaining • The protocol allows and provides data to carry through the above approach using plane-parallel an absorbed-dose to water calibration factor chambers, although there is evidence that minor for an ion chamber in a 60Co beam construction details significantly affect the response • The next step is to determine the quality of these detectors in 60Co beams, making the conversion factor, k , for the chamber being measurements or calculations of kecal uncertain Q • Plane-parallel chambers should be cross calibrated used in high-energy electron beams against calibrated – This step requires characterization of the beam cylindrical chambers quality Q

Obtaining an absorbed-dose to Obtaining an absorbed-dose to water calibration factor water calibration factor • The absorbed-dose calibration factor is defined as

60 Co • The ion chamber should be checked for any 60 Co D w N Dw,  (Gy/C or Gy/rdg) problems before it is sent for calibration M 60 Co • The ion chamber and the electrometer with where D w is the absorbed dose to water (in Gy) in the calibration laboratory’s 60Co beam at the point of measurement which it is to be used should both be of the ion chamber in the absence of the chamber calibrated, possibly as a single unit • It applies under standard environmental conditions of 22 °C, 101.33 kPa, and relative humidity between 20% and 80%, • All ranges of the electrometer that are respectively (in the US and Canada) routinely used for clinical reference • It must be traceable to the user’s national primary standard for absorbed dose to water dosimetry should be calibrated

3 Chamber waterproofing Measurement phantoms

• A chamber is calibrated and used clinically in water • Clinical reference measurements must be performed • Equivalent waterproofing techniques must be used in a water phantom with dimensions of at least 30  for measurements in the user’s beam and in the 30  30 cm3 (non-water phantoms are prohibited) calibration laboratory • If the beam enters through the • If a chamber is not inherently waterproof (preferred plastic wall of the water phantom method) it requires extra waterproofing sleeves and the wall is >0.2 cm thick, all depths should be scaled to water- • A waterproofing sleeve should minimize air gaps near the chamber wall (0.2 mm) and should be equivalent depths by measuring made of PMMA 1 mm thick from the outside face of the wall with the phantom full of water and accounting for the wall density

Charge measurement Shutter timing error – Co-60 only • The fully corrected reading from an ion • Any shutter timing error must be accounted for if needed chamber, M, is given by • If a beam shutter is used with a timer that closes the shutter M  P P P P M (C or rdg) when a preset has elapsed, the t ion TP elec pol raw measured by the timer may not agree exactly with the t´ representing the where Mraw is the raw ion chamber reading in coulombs, C, or the instrument’s reading units (rdg) shutter-open period • This can be detected by making two – PTP is the –pressure correction; measurements X1 and X2 for different – corrects for incomplete ion collection efficiency timer settings, t1 and t2 and – Ppol corrects for polarity effects calculating beam shutter timing error: – P takes into account the electrometer’s calibration factor elec 2 t1  X 1t 2 if the electrometer and ion chamber are calibrated separately   X 2  X 1

Polarity corrections Electrometer correction factor • Polarity effects vary with beam quality and other conditions such as cable position • If the electrometer is calibrated separately from the • It is necessary to correct for these effects each time clinical ion chamber, the electrometer correction factor, Pelec, reference dosimetry is performed is just the electrometer calibration factor, correcting + - • Taking reading with both polarities applied, M raw and M raw the electrometer reading to true coulombs +-  MMraw raw  • It is common practice in the US to calibrate ion P  pol chambers and electrometers separately 2M raw

+ - • It is common practice in Canada to calibrate them as • Mraw (one of M raw or M raw) is the reading corresponding to the charge collected for the reference dosimetry measurements in a unit, in which case Pelec =1.00 the clinic (should be the same as for the chamber calibration) • Also Pelec =1.00 for cross-calibrated plane-parallel • Polarity correction should be less than 0.3% (Addendum to chambers since it cancels out of the final equations TG-51 allows for 0.4%)

4 Corrections for ion-chamber Standard environmental conditions collection inefficiency • Since calibration factors are given for standard • The recombination correction factor Pion is used to environmental conditions of T0 = 22°C and P0 = 101.33 correct ion chamber readings for lack of complete kPa (1 atmosphere), one corrects charge or meter collection efficiency readings to standard environmental conditions by • Pion is a function of the dose per pulse in an 273.2 T 101.33 accelerator and thus will change if either the pulse PTP  rate for a fixed dose rate, or the dose rate is changed 273.2 22.0 P • The correction must be measured in each set of • It is assumed that the relative humidity is always in the experimental conditions for which clinical reference range of 20% to 80%, with the reading error ±0.15% dosimetry is being performed • Chambers require time (usually 5 to 10 min) to reach • The value of Pion should be less than 1.05 with their surroundings

Measuring Pion Measuring Pion • The standard two-voltage techniques should be used: • Although initial recombination may dominate, for the charge produced by the ion chamber is measured in continuous (i.e., 60Co) beams, the two-voltage formula the beam of interest when two different (by at least a gives an estimate of the general recombination

factor of 2) bias voltages are applied 2 1. VV /  PV HL • Let VH be the normal operating voltage for the detector ion  H  2 MMVVHL// (always the higher of the two voltages in these raw raw  HL measurements), and MH be the raw chamber reading raw • For pulsed or pulsed-swept beams with Pion < 1.05 • After measuring MH the bias voltage is reduced to raw 1. VV / PV HL VL, and once the chamber readings have reached ion  H  HL L MMVVraw// raw  HL equilibrium (takes several minutes), M raw is measured

Beam quality specification Beam quality specification

• For both photon and electron beams from accelerators, • Beam quality is characterized by a parameter the beam quality must be specified in order to related to the central-axis depth-dose curves for the determine the correct value of the quality conversion beam factor, kQ or the electron quality conversion factor, – For photons it is %dd(10) – the percentage depth dose at 10 cm depth in water due to photons only k´R50 – For it is R – the depth in water in cm at which • For a 60Co beam the factor k = 1.000 by definition 50 Q the absorbed dose falls to 50% of the maximum dose • Beam quality must be measured each time clinical • It is essential to use SSD=100 cm when reference dosimetry is performed for accelerator establishing the beam quality for photon and beams electron beams because %dd(10) and R50 are functions of SSD

5 Depth of measurement Depth of measurement photon beam electron beam • The point of measurement for a cylindrical chamber is on the central axis of the chamber and this is always placed at the reference depth when measuring dose at an individual point • The effective point of measurement is upstream of the point of measurement due to the predominantly forward direction of the secondary electrons • Effect of shifting depth-ionization data measured with cylindrical chambers

• This results in shift of the depth-dose curve upstream upstream a) by 0.6 rcav for photon beams and b) by 0.5 rcav for electron beams (to shallower depth) (rcav=1.0 cm). The raw data are shown by curve I (long dashes), shifted data by curve II (solid line). Electron beam curve must be further corrected

Depth of measurement Depth of measurement • For cylindrical and spherical chambers the shift is taken as 0.6r for photon beams and 0.5r for • Using these measurements as depth-ionization cav cav curves ignores any variations in P and P with electron beams, where rcav is the radius of the ion pol ionization chamber cavity depth and for electron beams it also ignores any • The shifted curves are taken as the depth-ionization variations in the electron fluence correction factor curves for cylindrical chambers • Since well-guarded plane-parallel chambers • For plane-parallel chambers, the center of the front minimize these variations with depth, they are (upstream) face of the chamber air cavity is the point preferred for measuring electron beam depth- of measurement, no shift is needed ionization curves

Depth of measurement Depth of measurement • For photon beams the variation in stopping-power • In contrast to the depth-dose curve measurement, ratio is negligible past dmax (<0.1%) and thus the depth-ionization curve is treated as a depth-dose curve for measurements of absolute dose at the reference – These same techniques should be used to determine any depth in both electron and photon beams, a clinical photon beam depth-dose curve cylindrical chamber’s point of measurement is • In order to determine depth-dose curves for electron placed at the referenced depth (10 cm for photons

beams, the depth-ionization curve must be further and dref for electrons) corrected for the significant change in the stopping- • The gradient effects are included implicitly in the power ratio with depth beam quality conversion factor k for photons and – This conversion is not needed in the protocol except to Q explicitly by the term PQ for electrons transfer the dose from dref to dmax if necessary gr

6 Beam-quality specification for Beam-quality specification for photon beams photon beams • The percentage depth dose at 10 cm depth in a water • For beams with energies less than 10 MV, with %dd(10)<75% the value of %dd(10) measured in the open phantom due to photons only, %dd(10)X , is defined for a field size of 1010 cm2 at the phantom surface at an SSD of beam is the beam quality, %dd(10)X 100 cm • For beam energies of 10 MV and above (and all FFF beams according to the Addendum), the value of %dd(10)X for the • At higher energies (about 10 MV and above), the electrons open beam is obtained from %dd(10) parameter: from the accelerator may significantly affect this dose Pb %dd (10) 0.8905 0.00150% dd (10) % dd (10) • Placing a 1 mm thick lead foil below the accelerator head X Pb Pb (about 50 cm from the phantom surface) reduces the  foil at 50 cm, %dd (10)Pb  73% electrons contamination to a negligible level, thus %dd(10)Pb is obtained (it actually replaces unknown contamination with • If %dd(10)Pb is less than the thresholds given in the that due to Pb-foil) equations, then %dd(10)x=%dd(10)Pb

Beam-quality specification for Beam-quality specification for photon beams electron beams • Alternative formula to correct for electron contamination for • For the purposes of reference beam dosimetry, beam higher energy beams (up to %dd(10)=89%): quality in electron beams is specified by R50, the depth

% dd 10 x  1.267 % dd 10   20 .0 in water (in cm) at which the absorbed dose falls to 50% of • Here %dd(10) is measured for an open beam the maximum dose for a • Leads to an error in k <0.25% Q beam, which has a field size • The Addendum advocates use of this formula in place of Pb: “Although TG-51 clearly states that the foil must be removed for the dose measurement step, on the phantom surface there is anecdotal evidence of confusion as to when the lead foil must be used” 1010 cm2 (2020 cm2 for • TG-51 is not very sensitive to accurate accounting for electron R > 8.5 cm, i.e., E > 20 contamination (low error in k even if Pb foil is not used) 50 Q MeV) at an SSD of 100 cm

Beam-quality specification for Beam-quality specification for electron beams electron beams • To determine R50 one must first measure a central-axis • The beam quality specifier for the electron depth-ionization curve in a water phantom at an SSD of 100 cm beam, R50, is determined from measured I50 • For cylindrical chambers, correct using for gradient effects by shifting RII1.029  0.06 (cm) (for 2   10 cm) the curve upstream by 0.5rcav 50 50 50 • Next, point B at the level or

of 50% of the maximum RII501.059 50  0.37 (cm) (for 50  10 cm) ionization; the depth of point B

gives I50

7 Photon beam dosimetry Reference conditions

• Either an SSD or an SAD setup can be used, the field • In photon beams size is 1010 cm2 Q 60 Co D w  Mk Q N D ,w (Gy) • When using an SSD setup, the field size is defined at the surface of the phantom gives the absorbed dose to water under reference conditions • When an SAD setup is • Reference dosimetry for photon beams is performed in being used, the field size is an open beam (i.e., without trays, wedges, or blocks) defined at the detector with the point of measurement of the cylindrical ion position which is placed at chamber placed at the reference depth which is a 10 cm depth at the isocenter water-equivalent depth of 10 cm in a water phantom of the machine

Absorbed dose to water in Absorbed dose to water in

clinical photon beams: kQ clinical photon beams: kQ • Calculated values of • For a set of Reference chambers (definition will follow) k in accelerator and beams with flattening filter the Addendum recommends Q calculation of k using: beams as a function Q

of %dd(10)X for 3 5 2 k  A  B 10  % dd 10   C 10  % dd 10   cylindrical ion Q x x

chambers commonly for 63  % dd 10 x  86 used for reference dosimetry • For beam qualities below %dd(10)X = 63, users should linearly interpolate between the value tabulated for that 60 • For Co beams 60 beam quality and kQ = 1.000 at Co (%dd(10)X = 58) kQ=1.000

Absorbed dose to water in Absorbed dose at other depths clinical photon beams: k Q in clinical photon beams

Table from Addendum • Clinical reference dosimetry determines the absorbed dose to water at 10 cm depth • If this is not the reference depth used for clinical dosimetry calculations, one determines the corresponding dose at the appropriate depth • For SSD setups the clinical percentage depth-dose curves are used • For SAD setups the clinical tissue-phantom ratio (TPR) curves are used • The table contains values for cylindrical chambers currently manufactured • Plane-parallel chambers are not included due to insufficient information on wall corrections in photon beams other than 60Co

8 Electron beam dosimetry Electron beam dosimetry

• In electron beam • For electron beams with R50  4.3 cm Q Q 60 Co (incident energies of 10 MeV or less), well- D w  MP gr k R k ecal N D ,w (Gy) 50 guarded plane-parallel chambers are gives the absorbed dose to water under reference preferred and they may be used at higher conditions for the same number of monitor units energies as used to measure the charge M, at the point of • Plane-parallel chambers must be used for measurement of the ion chamber, in an electron beam of quality Q, specified by R beams with R50  2.6 cm (incident energies 50 of 6 MeV or less)

Absorbed dose to water in Reference conditions clinical electron beams • Clinical reference dosimetry for electron beams is performed in an open beam at the reference depth which is at a water-equivalent depth of • To calculate the absorbed dose one needs the values of the factors PQ , k´ , and k dRref0.6 50 0.1 (cm) gr R50 ecal • The point of measurement of the ion chamber is placed at • The values of kecal , a photon-electron conversion dref – we know mass-collision stopping power ratios (water factor, for a number of ion chambers are given in to air) at that depth tables II and III of the protocol 2 • For beams with R50  8.5 cm, the field size is 1010 cm at the phantom surface and for higher-energy beams it is • The selection of the beam quality Qecal is arbitrary 2020 cm2 and has been taken as R50 = 7.5 cm for the • SSD may be from 90 to 110 cm (range where stopping- purposes of the protocol power ratios are not affected)

Absorbed dose to water in Absorbed dose to water in clinical electron beams: k´R50 clinical electron beams: kecal

• k´R50 – electron beam quality conversion factor

• Calculated values for k´R50 as a function of R50 for Table II. cylindrical ion chambers Plane-parallel used for clinical reference chambers dosimetry in electron beams

Table III. Cylindrical chambers

9 Absorbed dose to water in Absorbed dose to water in

clinical electron beams: k´R50 clinical electron beams: k´R50

• Calculated values of k´R50 • Calculated values of k´R50 at dref as a function of R50 at dref for high-energy for several common electron beams as a

plane-parallel chambers function of R50 for several common cylindrical chambers

Absorbed dose to water in Absorbed dose to water in clinical electron beams: k´ R50 clinical electron beams • For Farmer-like cylindrical chambers the • Calculated values of k´ following expression can be used for 2  R50  9 R50 cm with a maximum error of 0.2%: at dref for high-energy   R / 3.67  electron beams as a ke (cyl) 0.9905 0.0710 50 R50 function of R50 for several • For well-guarded plane-parallel chambers, the common plane-parallel following expression is an analytical chambers representation of the curve shown in the figures, i.e., for 2  R50  20 cm: 0.214 kR (pp) 1.2239 0.145   R50 50

Absorbed dose to water in Use of plane-parallel chambers clinical electron beams Q • The correction for gradient effects (i.e., P gr) is • For electron beam dosimetry the protocol allows not necessary for plane-parallel chambers and is for the use of plane-parallel chambers calibrated in 60 close to unity for cylindrical chambers when the a Co beam • However, since the 60Co calibration factors of at reference depth is at dmax, which is usually the case for electron beams below 10 MeV least some plane-parallel chambers appear to be very sensitive to small features of their Q • For cylindrical chambers P gr is determined as construction, it is recommended that plane-parallel M d 0.5 r  chambers be calibrated against cylindrical P Q  raw ref cav (for cylindrical chambers) gr chambers in a high-energy electron beam Mdraw ref 

10 Use of plane-parallel chambers Use of plane-parallel chambers

60 Co • From these measurements the product of kN ecal Dw , is • After determining the beam quality and the determined for the plane parallel chamber as reference depth in the high-energy electron beam cyl pp 60  D  to be used, measurements are made, in sequence, kNCo  w  ecalDw ,  pp   MkR  with the point of measurement of both the 50

cyl calibrated cylindrical chamber and the plane- 60 M PQ k k N Co  gr R50 ecal D , w  parallel chamber at dref  (G y/C ) pp  • While measuring with the cylindrical chamber,  MkR  50 Q P gr is measured as described above • Use of this product circumvents the need for obtaining the 60Co absorbed-dose calibration factor for the plane-parallel chamber

Absorbed dose at d in max Using other ion chambers clinical electron beams • The protocol provides kQ data for the majority of chambers • This protocol provides the reference dose at a depth of used in clinical reference dosimetry in North America dref which, for higher-energy beams, will not be at dmax • Other cylindrical chambers can be used by finding the where clinical normalization most often takes place closest matching chamber for which data are given

• To establish the dose at dmax one should use the • The critical features are, in order, the wall material, the clinical percentage depth-dose data for a given beam radius of the air cavity, the presence of an aluminum electrode, and the wall thickness and determine the dose at dmax from that at dref • As long as the wall material is matched and the chamber is • Methods for measuring electron-beam percentage ‘‘normal,’’ these matching data should be accurate to depth-dose curves are given in the AAPM TG-25 within 0.5%. protocol

Reference class ion chambers TG-51 worksheets

The protocol provides four worksheets: A. Photon Beams B. Electron Beams – Cylindrical Chambers C. for plane-parallel chambers D. Electron Beams using Plane-Parallel Chambers

Appendix A of the Addendum

11 Summary General formalism • Ion chamber calibration: absorbed dose to water calibration factors in TG-51 protocol electrons • Reference conditions • Reference dosimetry of photon beams photons Q 60 Co D w  Mk Q N D ,w (Gy) electrons • Reference dosimetry of electron beams

60 electrons D Q  MP Q k  k N Co (Gy) w gr R50 ecal D ,w

QA of medical accelerators • QA procedure for medical accelerators includes dosimetry as its integral part • Old protocol: TG-40 • New protocol: TG-142 – Adds recommendations for asymmetric jaws, MLC’s, and dynamic/virtual wedge – Separate tolerance for IMRT and SRS/SBRT machines – Includes procedures for on-board imaging QA (planar or cone-beam CT)

12