<<

Evaluation and Characteristic of Modifed Ventilator Under Hyperbaric Conditions During Volume- controlled Ventilation

Cong Wang Beijing Tiantan Hospital Lianbi Xue (  [email protected] ) Beijing Tiantan Hospital https://orcid.org/0000-0002-4506-7968 Jialong Liu Beijing Aeonmed Co Ltd Liyun Chang Beijing Aeonmed Co Ltd Qiuhong Yu Beijing Tiantan Hospital Yaling Liu Beijing Tiantan Hospital Ziqi Ren Beijing Tiantan Hospital Ying Liu Beijing Tiantan Hospital

Research

Keywords: Hyperbaric , Ventilator, Volume-control ventilation, Gas density, Airway resistance

Posted Date: August 2nd, 2021

DOI: https://doi.org/10.21203/rs.3.rs-636953/v1

License:   This work is licensed under a Creative Commons Attribution 4.0 International License. Read Full License

Page 1/14 Abstract

Purpose:The stability of the modifed ventilator (Shangrila590, Beijing Aeonmed Company, Beijing, China) was evaluated under hyperbaric conditions during volume-controlled ventilation in this study by Michigan test lung (5601i, Grand Rapids, MI, US).

Methods:Experiments were performed inside the multiplace hyperbaric chamber at 1.0, 1.5 and 2.0 atmospheres absolute (ATA). The modifed ventilator placed inside the hyperbaric chamber was connected to the test lung. During volume-controlled ventilation (VCV), data for the test lung were collected by a personal computer outside the hyperbaric chamber. The preset volume (VTset) of the ventilator (400-1000 ml) and the resistance and compliance of the testing lung were adjusted before the experiments at every ambient . With every test setting, the tide volume (VT), inspiratory airway peak pressure (Ppeak) and minute volume (MV) displayed by the ventilator and the test lung were recorded by the computer. We compared the ventilator and test lung data under 1.0, 1.5 and 2.0 ATA to evaluate the stability of the modifed ventilator.

Results:The variation in VT in the test lung and the ventilator at different ambient changed within a narrow range, and the differences were statistically signifcant. In every test setting, changes in the MV of the ventilator were limited and acceptable, with signifcant differences at different ambient pressures. However, Ppeak increased obviously, as detected by the ventilator and test lung at higher during VCV.

Conclusions:The modifed Shangrila590 ventilator can work well in a hyperbaric chamber. It can provide relatively stable VT and MV during VCV with VTset from 400 ml to 1000 ml when the ambient pressure increases from 1.0 ATA to 2.0 ATA. The raised ambient pressure will lead to increased gas density, which may result in more airway resistance and higher Ppeak during VCV.

Key Messages

The modifed ventilator can work well with sable VT and MV during VCV in hyperbaric condition. Ppeak increased obviously because of the raised airway resistance due to high ambient pressure.

Introduction

Hyperbaric oxygen (HBO) involves treatment with a high fraction of inspired oxygen (FiO2) at higher than , whereby increased pressures depend on guidelines and indications. HBO therapy is widely used in critically ill patients in the intensive care unit (ICU), such as for acute carbon monoxide poisoning, sickness, arterial gas embolism, radiation-induced tissue injury, and acute traumatic ischemic injury [1,2]. HBO therapy remains among the safest used today under the common range of the 2-3ATA environment [3]. Arterial of oxygen (PaO2) can be improved through increased pressure of oxygen (PO2) in inspired gas physiologically,

Page 2/14 especially during HBO treatment. According to the gas law, PO2 in inspired can be elevated not only by increasing FiO2 but also by increasing the pressure of breathing gas. Under a normal baric environment, we can only limitedly raise PO2 by increasing FiO2 [4]. Alternatively, we can elevate PO2 by increasing ambient pressure and FiO2, which can raise the oxygenated efciency of pulmonary tissue in the hyperbaric chamber. When the ventilator is placed in the hyperbaric chamber with the same parameters as outside the chamber, pulmonary will be markedly improved if mechanical ventilation can provide stable pulmonary ventilation. However, the ordinary ventilator used in the ICU does not work well during HBO therapy because of the high ambient pressure. This limits mechanical ventilation carried out during HBO therapy. Overall, broadening the feld of mechanical ventilation may beneft patients on mechanical ventilation who need HBO therapy.

Not all medical devices can be subjected to hyperbaric chambers; indeed, most life support technologies, such as hemofltration, electrical defbrillation and extracorporeal membrane oxygenation, are at present incompatible with the hyperbaric environment [5,6].

Pneumatical ventilators can operate in hyperbaric environments safely, but they cannot provide stable VT, respiratory rate (RR) or MV with raised ambient pressure [7,8]. Stable operation of a ventilator should be the guarantor for patients on mechanical ventilation, not only in the ICU but also in the hyperbaric chamber. Modern ventilators tend to be electropneumatical and electronic. In recent years, a series of bench tests have been carried out on ventilators under hyperbaric conditions during different ventilation models, such as VCV and pressure-controlled ventilation (PCV) [9,10]. Nevertheless, most electropneumatical ventilators cannot function well in hyperbaric chambers. In the early stage, researchers focused on empirically predicting changes in ventilation parameters under specifc high pressures and then adjusted the parameters of the ventilator to manually compensate for the changes. With the understanding of respiratory mechanics in hyperbaric environments and the development of modern ventilator technology, modifed ventilators for hyperbaric conditions have been developed. Modifed ventilators can adjust the compensation by itself automatically, precisely and rapidly when ambient pressure changes, for example, Siaretron 1000 Iper (Bologna, Italy) [10,11]. However, the type of ventilator for hyperbaric conditions is limited globally. In China, we do not have our own ventilator for hyperbaric conditions with independent intellectual property rights. In brief, we frst modifed the Shangriala590 ventilator (Beijing, China) and then performed a series of bench tests in a hyperbaric chamber before putting it to use clinically. In this study, we detected the stability of VT and MV by a test lung during VCV in a hyperbaric chamber, and we clarifed the characteristics of hyperbaric ventilation according to the respiratory mechanics parameters.

Methods

The ventilator

The Shangrila590 ventilator is an electropneumatic ventilator from Beijing Aeonmed Company that is commonly used in the ICU in China. According to the safety regulations of medical hyperbaric chambers

Page 3/14 in China [12,13], the pneumatic part was placed in the chamber, and the electronic part was assembled out of the chamber. The two parts of the ventilator were connected with wire through the chamber without leakage, allowing doctors to operate the ventilator outside the hyperbaric chamber. Ventilator engineers modifed the algorithm and replaced some components of the ordinary ventilator to make the ventilator work reliably and safely in a hyperbaric environment.

The test lung

We used the Model 5601i Adult/Infant PneuView System (Michigan Instrument, Grand Rapids, MI, US) to detect ventilation parameters. The detection system contains Michigan test lung and PneuView data collection software. Model 5601i houses an electronic interface module, which converts the pressure signal from the test lung to digital data and transfers these data to a personal computer. PneuView data collection software serves as the link between the test lung and the computer.

The critical care multiplace hyperbaric chamber

It has been proven that a multiplace hyperbaric chamber is better suited for HBO treatment of critically ill patients than a monoplace hyperbaric chamber because it permits appropriate ICU equipment to be used inside the chamber by staff [14]. The critical care hyperbaric chamber [GY3800-A (GY3800 M2-D), Yantai Hongyuan Oxygen Industrial Inc., Yantai, China] is a multiplace hyperbaric chamber with an automated operation system equipped with electrocardiogram monitors, ventilators, transcutaneous oxygen (O2) and carbon dioxide (CO2) tension monitors, syringe drivers, and infusion pumps, among others, to ensure the continuous treatment of ICU patients. Our chambers have three compartments, two ICU chambers and a prechamber between them, and have the capacity for 24 seated people or 8 gurneys.

The experimental confguration

We calibrated the ventilator and the test lung at atmospheric pressure before the experiments. The test lung was located inside the hyperbaric chamber and connected to the pneumatic part of the ventilator. The digital data detected by the test lung were passed by electrical penetration and wires through the bulkhead to the personal computer outside (Figure 1). According to the parameters shown in the Calibration Specifcation for Ventilators [15], the ventilator was adjusted by the doctors outside, and the resistance and compliance of the test lung were regulated by the staff inside according to the test settings in Table 1.

Experimental procedure

The hyperbaric chamber pressure rose sequentially to 1.0 ATA, 1.5 ATA and 2.0 ATA. At every pressure stage, the ventilator was operated in different VTset (400-1000 ml), f 20 BPM, I/E 1:2, PEEP 2 cmH2O, FiO2 40 %), with the corresponding resistance and compliance of the test lung provided in Table 1. The steady state of the ventilator after regulation was 2 minutes. Data were collected by the software for 20 cycles at

Page 4/14 every setting. The in the hyperbaric chamber was maintained at a steady temperature of 25 °C to 26 °C.

Statistical analysis

Multiple analyses of variance were used to evaluate the impact of the ventilator and the test lung, different ambient pressure conditions, and different VTset (400-1000 ml) of VCV. Post hoc comparisons were performed for multiple comparisons when Mauchly’s test of sphericity showed signifcance. A p value smaller than 0.05was considered signifcant. We used SPSS 19.0 to perform the statistical analysis and GraphPad Prism 5 to prepare graphs.

Results

During VCV with every VTset, VT displayed by the ventilator itself was higher than that detected by the test lung in the same ambient pressure, especially for the VTset 800 ml groups and VTset 1000 ml groups (Table

2). With increased ambient pressure, VT decreased not only in the test lung groups but also in the ventilator groups. Changes in VT in the test lung and ventilator at different ambient pressures seemed stable, and the differences were statistically signifcant (Figure

2). Compared with VTset, downtrends of VT displayed by the ventilator were 0.5-1 % for the VTset 400 ml group, 1-2.6 % for the VTset 500 ml group, 4-6.5 % for the VTset 600 ml group, 5-8 % for the VTset 800 ml group and 4.7-7.1 % for the VTset 1000 ml group when the ambient pressure increased from 1.0 ATA to 2.0 ATA. Meanwhile, we evaluated the MV displayed by the ventilator during VCV at different ambient pressures (Figure 3). At every VTset, changes in MV were limited to the range 1.0-2.0 ATA, and there were signifcant differences at different ambient pressures (Table 3).

With increased ambient pressure during VCV at a fxed level of VTset, we observed increased Ppeak displayed by the ventilator but stable Ppeak detected by the test lung (Figure 4). Statistical analyses showed that Ppeak displayed by the ventilator was higher than that detected by the test lung for the same

VTset group and ambient pressure (Table 4). In ventilator groups with fxed VTset, Ppeak increased obviously at raised ambient pressure with multiple comparisons, but the difference was not statistically signifcant in the test lung groups. At high ambient pressure (1.5-2.0 ATA), uptrends of Ppeak were 7-13 % for the VTset 400 ml group, 9-19 % for the VTset 500 ml group, 9-20 % for the VTset 600 ml group, 6-12 % for the VTset 800 ml group and 8-18 % for the VTset1000 ml group compared with Ppeak displayed by the ventilator at normal ambient pressure (1.0 ATA).

Discussion

Advanced investigation showed that ordinary ventilators used at normal atmospheric pressure cannot maintain stable VT during VCV when operated at high atmospheric pressure. Inspiratory fow provided by

Page 5/14 the ventilator will decrease with increasing ambient pressure [7,8,9,10]. The reason is that during HBO therapy, the ambient pressure is raised by compression air, which results in a high breath gas density and has no infuence on breath gas viscosity. High breath gas density results in more turbulent fow in peripheral airways according to an increased Reynold’s value (>1000, Figure 5). To obtain the same inspiratory fow, turbulent fow produces higher airway resistance than laminar fow. Hence, more driving pressure (△P) must be provided by the ventilator to overcome the higher airway resistance; otherwise, it will lead to decreased inspiratory fow. Unless this phenomenon is technically compensated for, hypoventilation may occur due to decreased VT and MV [11,17]. To maintain adequate VT and MV, we need to increase the inspiratory fow as the chamber pressure increases through manual regulation or automatic compensation [7,8,9,10,16].

Evaluating VT during VCV at high ambient pressure

During VCV, the modifed Shangrila590 ventilator can provide more △P to overcome the greater resistance and achieve constant VT and MV, even though VT and MV decreased within a narrow range. In the high VTset group (800-1000 ml), the decline in VT as the ambient pressure increased was greater than that in the low VTset group (400-600 ml) (Table 2, Figure 2). However, the degree of decline was smaller than that in previous research at the same ambient pressure scale, which resulted in a 20-56 % decline in

VT [7,9]. Nonetheless, compared with the hyperbaric ventilator Siaretron IPER 1000, a 6.5-20

% increase in VT during VCV under 1.0 ATA to 2.2 ATA ambient pressure occurred, which is CE-certifed for hyperbaric use in Europe [10]. A modifed Penlon Nufeld 200 has been used in a monoplace hyperbaric chamber, fxed outside the chamber, with a 30 % decrease in VT with ambient pressure from 1.0 ATA to 2.0 ATA [16].

When we focused on the accuracy of VT displayed by the ventilator, it was not exactly equal to

VT detected by the test lung. For VTset between 400-600 ml, the VT displayed by the ventilator may overestimate the actual VT Otherwise, for VTset between 800-1000 ml, the VT displayed by the ventilator may underestimate the actual VT. The difference in VT displayed by the ventilator and by the test lung was narrow during VTset between 400-600 ml, but it was wide during VTset between 800-1000 ml. However, the accuracy of the test lung in previous studies was controversial at high ambient pressure [7,9]. We used a water tank-simulated lung in pre-experiments, which can roughly refect the true value of ventilator

VT. The data between the ventilator and the water tank-simulated lung seemed good, but the numerical precision of the water tank-simulated lung was low. For statistical analysis, we decided to choose the Michigan test lung (5601i) for the test equipment, according to previous research [7].

During VCV, the goal of maintaining constant VT is to take in enough O2 and ensure expiration of CO2.

Factors affecting the expiration of CO2 include not only Vt but also and high PaO2. During

HBO therapy, PaO2 is much higher than that under normobaric conditions. Under hyperbaric conditions, respiratory resistance leads to decreased breath gas fow and enlarged dead space. These may reduce of CO2 [18,19]. In addition to the stable operation of ventilators, it is essential to monitor the

Page 6/14 expiratory volume, arterial partial pressure of carbon dioxide (PaCO2), transcutaneous carbon dioxide tension (PTCCO2), or end-tide carbon dioxide partial pressure (PETCO2) [11,20].

Changes in Peak during VCV at high ambient pressure

Our data showed that the Ppeak displayed by the ventilator increased obviously during VCV with fxed

VTset in the process of ambient pressure rise (Table 4, Figure 4). As Ppeak can refect inspiratory resistance, the ventilator can provide more △P to overcome increased airway resistance to maintain stable VT. Our data shown in Figure 4 supports this mechanism. However, changes in Ppeak detected by the test lung were gentle because of the different detected positions of the ventilator and the test lung. The breathing gas fow was buffered when detected by the test lung.

Side effect of HBO in pulmonary system and preventive measures

In general, there is no risk of pulmonary (PBT) in patients with normal lungs during HBO therapy. Based on Boyle’s Law, there is potential for PBT due to lung overinfation during decompression when disease is present, such as asthma or chronic obstructive pulmonary disease (COPD) with active bronchospasm, mucous plugging, and bullous lung disease. Additionally, pneumothorax (PTX) is a potentially life-threatening phenomenon, especially given the increased risk of tension PTX during decompression. All candidates for HBO therapy must be screened for pulmonary disease to avoid increasing the risk of PBT and PTX [21].

Continuous exposure of the lungs to elevated PaO2, either at normobaric or hyperbaric pressure, leads to toxic effects of O2. Pulmonary O2 toxicity can be avoided if O2 is provided at the proper dose [5,22]. When

FiO2 is continuously high in normobaric environments, the lungs are at risk of O2 toxicity: (a) high FiO2 levels promote the formation of absorption atelectasis in the absence of nitrogen; (b) high FiO2 levels also induce ROS-mediated damage; and (c) another side effect of hyperoxemia is the rise in PaCO2 [22].

In the hyperbaric chamber, we can obtain higher PO2 by increasing ambient pressure with lower FiO2 to avoid absorption atelectasis. In rats, HBO exposure caused signifcant oxidative stress in the frst 24 h. However, these effects were resolved at the end of the tenth day of HBO treatment [23]. There are two pathways for the development of CO2 intoxication. PCO2 is increased in inspired breathing gas or expiration of produced CO2 is insufcient [17]. Increased PCO2 in inspired gas may occur when gas exchange occurs in the hyperbaric chamber. To prevent raised PCO2 levels, hyperbaric chambers must be fushed continuously with breathing gas. Increased breathing resistance in the hyperbaric chamber may decrease the expiration of produced CO2. Maintaining stable pulmonary ventilation and monitoring

PaCO2 by blood gas, PETCO2, and PTCCO2 must be established during HBO therapy [11,17,21].

Work of breathing in hyperbaric environments is also a concern. Combined with the breathing equipment itself, the work of breathing will be increased compared to breathing the same gas in a normobaric environment [6]. For patients on mechanical ventilation, the endotracheal tube diameter is critical with regard to its effect on airway pressure and work of breathing [18]. During HBO therapy, we must consider

Page 7/14 that a high breath gas density induces high airway resistance, which cannot be avoided in hyperbaric chambers. In addition to sputum aspiration and exchange for large endotracheal intubation, we can decrease airway resistance by prolonging the inspiratory time appropriately and using a oxygen mixture to decrease the gas density. Additionally, we can reduce the high airway resistance and breathing work by downregulating ambient pressure or upregulating the support pressure of the ventilator.

Conclusions

In summary, during HBO therapy with raised ambient pressure from 1.0 ATA to 2.0 ATA, a modifed

Shangrila590 ventilator made in China can provide stable VT and MV during VCV with VTset from 400 ml to 1000 ml. During VCV, Ppeak increased obviously because of the raised ambient pressure. We will evaluate other ventilator models, such as PCV, in the future. Advanced detection of this ventilator operating in an environment of more than 2.0 ATA will be carried out in our further study.

Abbreviations

ATA: atmospheres absolute; VCV: volume-controlled ventilation; VTset: preset tide volume; VT: tide volume; Ppeak: inspiratory airway peak pressure; MV: minute volume; HBO: hyperbaric oxygen; FiO2: inspired oxygen; ICU: intensive care unit; PaO2: arterial partial pressure of oxygen; PO2: pressure of oxygen; PCV: pressure-controlled ventilation; O2: oxygen; CO2: carbon dioxide; △P: driving pressure;

PaCO2: arterial partial pressure of carbon dioxide; PTCCO2: transcutaneous carbon dioxide tension;

PETCO2: end-tide carbon dioxide partial pressure; PBT: pulmonary barotrauma; COPD: chronic obstructive pulmonary disease; PTX: pneumothorax.

Declarations

Ethics approval and consent to participate

Not applicable. This study didn’t involve human participants, human material, or human data. This study has been granted an exemption from our national ethics approval (Beijing Tiantan Hospital Medical Ethic Committee).

Consent for publication

Not applicable.

Availability of data and materials

All data used and/or analysed during the study are available from the corresponding author on reasonable request.

Competing interests

Page 8/14 The authors declare that have no competing interests.

Funding

None.

Authors’ contributions

LX designed the study. CW, QY and YL completed the literature review. LX and CW redacted the protocol. LC performed the measurements and the records. JL and LC modifed the ventilator. CW undertook the data statical analysis, and wrote the manuscript. ZR and YL operated and maintained the hyperbaric chambers. All authors read and approved the fnal manuscript.

Acknowledgements

The authors thank all the subjects for their participation in this research.

Authors’ information

1Department of Hyperbaric Oxygen, Beijing Tiantan Hospital, Capital Medical University, A zone, No.199, Nansihuan West Road, Fengtai District, Beijing, China. 2Beijing Aeonmed CO., LTD., Building No.9, Unit 26, Outer Ring West Road, Fengtai District, Beijing, China. *Corresponding Author, E- mail: [email protected], Tel.: +86-010-59976898.

References

1. Lindell K, Weaver MD, FCCP FACP, FCCM. Hyperbaric oxygen in the critically ill. Crit Care Med. 2011;39(7):1784–91. 2. Patrick M, Tibbles MD, Johns. Edelsberg MD, M.P.H. Hyperbaric- . The New England Journal of . 1996;334(25):1642–7. 3. Marvin Heyboer III, Deepali Sharma W, Santiago N, McCulloch. Hyperbaric Oxygen Therapy: Side Effects Defnedand Quantifed. Adv Wound Care (New Rochelle). 2017 Jun 1;6(6):210–224. doi: 10.1089/wound.2016.0718. 4. Jones MW, Brett K, Han N, Wyatt HA. Hyperbaric Physics. In: StatPearls [Internet]. Treasure Island (FL): StatPearls Publishing; 2021 Jan. 2020 Oct 27. 5. Millar IL1. Hyperbaric intensive care technology and equipment. Diving . 2015;45(1):50–6. 6. Kot J. Medical devices and procedures in the hyperbaric chamber. Diving Hyperbaric Medicine. 2014;44(4):223–7. 7. Stanga DF, Beck G, Chimiak BA,JM. Evaluation of respiratory support devices for use in the hyperbaric chamber. Navy Experimental Diving Unit, NEDU TR 03–18, 321 Bullfnch Rd. Panama City,

Page 9/14 FL 32407 – 7015, November 2003. 8. Lie Sui An. Ting LS, Chuan LC, Joang KS, Soh Chai Rick (2018) Performance of the Oxylog® 1000 portable ventilator in a hyperbaric environment. Diving Hyperbaric Medicine 48(2):102–6. 9. Stahl W, Radermacher P, Calzia E. Functioning of ICU ventilators under hyperbaric conditions - comparison of volume- and pressure-controlled modes. Intensive Care Med. 2000;26:442–8. 10. Lefebvre J-C, Lyazidi A, Parceiro M, Giuseppe F, Sferrazza Papa E, Akoumianaki D, Pugin D, Tassaux L, Brochard, Jean-Christophe M, Richard. Bench testing of a new hyperbaric chamber ventilator at different atmospheric pressures. Intensive Care Med. 2012;38:1400–4. 11. Jack Kot. Medical equipment for multiplace hyperbaric chambers Part II: Ventilator. European Journal of Underwater Hyperbaric Medicine. 2006;7(1):9–12. 12. Medical hyperbaric chamber pressurized with air (GB/T 12130 – 2005), General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China and Standardization Administration of the People's Republic of China. 2005, 14th Sep. 13. Supervision Regulation on Safety Technology for Hyperbaric Oxygen Chambers (TSG 24-2015), General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China, 2015, 20th Nov. 14. Lind F. A pro/con review comparing the use of mono- and multiplace hyperbaric chambers for critical care. Diving Hyperbaric Medicine. 2015;45(1):56–60. (). 15. Calibration Specifcation for Ventilators (JJF 1234–2018). General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China, 2018, 27th Feb. 16. LEWIS RP, SZAFRANSKI J, BRADFORD RH, SMITH AND HS G. G. CRABBE. The use of the Penlon Nufeld 200 in a chamber. An evaluation of its use and a clinical report in two patients requiring ventilation for carbon monoxide poisoning. Anaesthesia. 1991;46:767–70. 17. Welslau W. Physiologic Effects of Increased Barometric Pressure. In Handbook on hyperbaric medicine. Daniel Mathieu (ED.). Springer-Verlag, Dordrecht, 2006:pp. 41–3. 18. PhDa RA, Yohanan Daskalovic PhDa, Ofr Ertracht MSca, Yehuda Arieli PhDa, Yohai Adir MDa, Amir Abramovich MDa, Pinchas Halpern MD. (2011) Flow resistance, work of breathing of humidifers, and endotracheal tubes in the hyperbaric chamber. American Journal of 29, 725– 730. 19. Heather E, Held, Pendergast DR. Relative effects of submersion and increased pressure on respiratory mechanics, work, and energy cost of breathing. J Appl Physiol. 2013;114:578–91. 20. Asger Bjerregård and Erik Jansen. Monitoring carbon dioxide in mechanically ventilated patients during hyperbaric treatment. Diving Hyperbaric Medicine. 2012;42(3):134–6. 21. Asger Bjerregård E. Jansen. Monitoring carbon dioxide in mechanically ventilated patients during hyperbaric treatment. Diving Hyperb Med. 2012 Sep;42(3):134–6. 22. Allardet-Servent J, Sicard G, Metz V, Chiche L. Benefts and risks of oxygen therapy during acute medical illness: Just a matter of dose! Rev Med Interne. 2019 Oct;40(10):670–6.

Page 10/14 doi:10.1016/j.revmed.2019.04.003. 23. Sefka Körpınar H, Uzun. The Effects of Hyperbaric Oxygen at Different Pressures on Oxidative Stress and Antioxidant Status in Rats. Medicina (Kaunas) 2019 May 24;55(5):205. doi: 10.3390/medicina55050205.

Tables

Due to technical limitations, table 1 to 4 is only available as a download in the Supplemental Files section.

Figures

Figure 1

The experimental confguration

Page 11/14 Figure 2

Changes in tide volume during volume-controlled ventilation at different ambient pressure

Figure 3

Changes in minute volume during volume-controlled ventilation at different ambient pressure

Page 12/14 Figure 4

Changes in inspiratory airway peak pressure during volume-controlled ventilation at different ambient pressure

Page 13/14 Figure 5

Type of breathing gas fow

Supplementary Files

This is a list of supplementary fles associated with this preprint. Click to download.

Table.pdf

Page 14/14