SAFETY TEST REPORT

MEASUREMENT AND TEST REPORT For

Gaomi shuangyun labor protection supplies Co. ltd.

Dong Mou Dong Cun, Gaomi Economic Development Zone, 261000 City, Province,

Models: KN95

March 14, 2020

This Report Concerns: Equipment Type:

Original Report Jeliton Protective Face Mask

Test Engineer:

Report Number: CIC200314067-ZJ

Test Date: March 12-13, 2020

Reviewed By:

Prepared By: CIC Testing Technology Co., Ltd. No.102, Floor 1, Building A, No. 26 of Xinhe Road, Xinqiao Community, Xinqiao Street, Baoan District, Shenzhen, Guangdong, China. Tel: 4008-707-283 Fax: +86-0755-23208027 Web: http:// www.soncap-saso.com

Note: This test report is limited to the above client company and the product model only. It may not be duplicated without prior written consent of Shenzhen CIC Testing Technology Co., Ltd. Page 2 of 13 Report No. CIC200314067-ZJ

TEST REPORT EN 149 Respiratory Protective Devices - Filtering Half Masks to Protect Against Particles - Requirements, Testing, Marking

Report reference No...... : CIC200314067-ZJ

Tested by (+signature)...... : Dana Huang ………………see cover page………………… Approved by (+ signature)... : Hale Zhao ………………see cover page………………… Date of issue...... : March 13,2020

Testing laboratory Name……………………… : Shenzhen CIC Testing Technology Co., Ltd Address…………………… : No.102, Floor 1, Building A, No. 26 of Xinhe Road, Xinqiao Community, Xinqiao Street, Baoan District, Shenzhen, Guangdong, China. Test location...... : (Same as above)

Client Name………………………. : Gaomi shuangyun labor protection supplies Co. ltd. Dong Mou Dong Cun, Gaomi Economic Development Zone, 261000 Weifang Address……………………. : City, Shandong Province,China Test specification

Standard...... : EN 149: 2001+ A1: 2009

Test procedure...... : CE-PPE Procedure deviation...... : N.A. Non-standard test method.. : N.A. Test Report Form No………………..: EN_149_A TRF originator ...... : SGS Fimko Ltd Master TRF ...... : 2006-06 Copyright © 2006 IEC System for Conformity Testing and Certification of Electrical Equipment (IECEE), Geneva, Switzerland. All rights reserved. This publication may be reproduced in whole or in part for non-commercial purposes as long as the IECEE is acknowledged as copyright owner and source of the material. IECEE takes no responsibility for and will not assume liability for damages resulting from the reader's interpretation of the reproduced material due to its placement and context. Test item

Description...... : Jeliton Protective Face Mask Model No...... :KN95

Trade Mark...... : JELITON Manufacturer...... : Gaomi Renmin Labor Protection Appliance Factory Dong Mou Dong Cun, Gaomi Economic Development Zone, 261000 Address...... : Weifang City, Shandong Province,China

Classification...... : FFP2 NR D Page 3 of 13 Report No. CIC200314067-ZJ

Test case verdicts Test case does not apply to the test object...... :N(.A.) Test item does meet the requirement...... : P(ass) Test item does not meet the requirement...... :F(ail)

Testing Date of receipt of test item ……………………………..: March 12, 2020 Date(s) of performance of test ……………………...: March 12-13, 2020 Laboratory sample number...... : Sample appearance and function are in normal Yes condition, yes or no...... : Ambient temperature...... : 24.9℃ Ambient humidity...... : 64.2%

Summary of testing: The submitted sample were tested and found to compliance with requirements of the standards EN 149: 2001+ A1: 2009

General remarks

This test report shall not be reproduced except in full without the written approval of the testing laboratory.

The test results presented in this report relate only to the item tested.

"(see remark #)" refers to a remark appended to the report.

"(see appended table)" refers to a table appended to the report.

Throughout this report a comma is used as the decimal separator.

General product information: 1. Instructions and equipment marking related to safety is applied in the language that is acceptable in the country in which the equipment is to be sold. Page 4 of 13 Report No. CIC200314067-ZJ

Copy of marking plate:

Jeliton Protective Face Mask Model: KN95 Classification : FFP2 NR D EN 149: 2001+ A1: 2009

Gaomi Renmin Labor Protection Appliance Factory. MADE IN CHINA Page 5 of 13 Report No. CIC200314067-ZJ EN 149 Clause Requirement + Test Result - Remark Verdict

5 Classification P

Particle filtering half masks are classified Complied with standard, see P according to their filtering efficiency and their appended. maximum total inward leakage. There are three classes of devices: -FFP1 N

-FFP2 P

-FFP3 N

6 Designation P

Particle filtering half masks meeting the P requirements of this European Standard. Year of publication, classification

7 Requirements P

7.1 In all tests all test samples shall meet the Complied see bellow P requirements 7.2 Nomial values and tolerances P

Unless otherwise specified,the values stated in Actual using value is clear P this European Standard are expressed as normal values. 7.3 Visual inspection P

The visual inspection shall also include the Clear marking is provided, P marking and the information supplied by the see sample body manufacturer. 7.4 Packaging P Masks shall be offered for sale packaged in such Distinct design and warning P a way that they are protected against mechanical are made on packaging, see damage and contamination before use. sample body 7.5 Material P Materials used shall be suitable to withstand Comfortable wearing, when P handling and wear over the period. Any material releasing no hazards is from the filter media released shall not constitute produced a hazard or nuisance for the wearer. 7.6 Cleaning and disinfecting N

The materials used shall withstand the Single-use equipment N cleaning and disinfecting 7.7 Practical performance P

The particle filtering half mask shall undergo Complied, see bellow test P practical performance tests under realistic conditions. 7.8 Finish of parts Soft equipment N Page 6 of 13 Report No. CIC200314067-ZJ EN 149 Clause Requirement + Test Result - Remark Verdict

Parts likeyl to come into contact with the wearer N shall have no sharp edges or burrs 7.9 Leakage P

7.9.1 Total inward leakage P

The laboratory tests shall wearer to protect with Enough safe condition is P high probability against the potential hazard to be provide expected. Exercise results for total inward leakage shall be FFP2, See below test table P not greater than 25% for FFP1 11% for FFP2 5% for FFP3 7.9.2 Penetration of filter material P

Meet the requirements of Table 1 FFP2 P Sodium chloride test: 20% Paraffin oil test: 20% 7.10 Compatibility with skin P

Materials that may come into contact with the Have no irritation or adverse P wearer’s skin shall not be known to be likely to effect to skin and health cause irritation or any other adverse effect to health. 7.11 Flammability Have no such hazard P

The material used shall not present a danger for P the wearer and shall not be of highly flammable nature. 7.12 Carbon dioxide content of the inhalation air N

The carbon dioxide content of the inhalation air <1.0% P (dead space) shall not exceed an average of 1,0 %(byvolume). 7.13 Head harness

Head harness shall be designed so that mask The design is considered P can be doned and removed easily. Head harness shall be adjustable or self- The design is considered P adjusting and sufficiently robust to hold the mask firmly in position . 7.14 Field of vision P

The field of vision is acceptable if determined so Clear field of vision when P in practical performance tests. wearing 7.15 Exhalation valve(s) P

A particle filtering half mask may have one or One valve provided N more exhalation valve(s) and shall function correctly in all orientations. Page 7 of 13 Report No. CIC200314067-ZJ EN 149 Clause Requirement + Test Result - Remark Verdict

If an exhalation valve is provided it shall be Clearly function P protected against or be resistant to dirt and mechanical damage and may be shrouded or may include any other device Exhalation valve(s) shall continue to operate Complied, see below P correctly after a continuous exhalation flow of 300 l/min over a period of 30 s. Exhalation valve housing is attached to the Enough strong P faceblank, and withstand axially a tensile force of 10 N applied for 10 s. 7.16 Breathing resistance P

The breathing resistances apply to valved and Complied, see below test P valveless and shall meet the requirements table 7.17 Clogging N

7.17.1 General Single-use device N

For single-use devices clogging test is an N optional test. Devices designed to be resistant to clogging, N shown by a slow increase The specified breathing resistances shall not be N exceeded before the required dust load of 833 mg·h/m3 . 7.17.2 Breathing resistance N

7.17.2.1 Valved particle filtering half masks N

7.17.2.2 Valveless particle filtering half masks N

After clogging the inhalation and exhalation N resistances shall not exceed -FFP1:3mbar -FFP2:4mbar -FFP3:5mbar at 95 l/min continuous flow. N

7.17.3 Penetration of filter materia N

All types claimed to meet the clogging N requirement shall also meet the penetration requirements given in 7.9.2 after the treatment. 7.18 Demountable parts No any such part N

All demountable parts (if fitted) shall be readily connected and secured, where possible by hand.

8Testing P

8.1 General P

No special measuring devices and methods are Common methods P specified, commonly used devices and methods Page 8 of 13 Report No. CIC200314067-ZJ EN 149 Clause Requirement + Test Result - Remark Verdict

shall be used.

8.2 Visual inspection P

The visual inspection is carried out appropriate Considered P by the test house prior to laboratory or practical performance tests. 8.3 Conditioning P

8.3.1 Simulated wearing treatment P

A breathing machine is adjusted to 25 cycles/min 25 cycles/min P and 2,0 l/stroke. 2,0 l/stroke. For testing, a saturator is incorporated in the a saturator incorporated by P exhalation line between the breathing machine breathing machine and the and the dummy head, dummy head The spilling out of the dummy’s mouth and Incline considered P contaminating the particle filtering half mask the head shall be incline 8.3.2 Temperature conditioning P

Exposet masks to the following thermal cycle: Complied P

a) for 24 h to a dry atmosphere of (70 ± 3) ℃; P

b) for 24 h to a temperature of (-30 ± 3) ℃; P Allow to return to room temperature for at least 4 5h P h between exposures and prior to subsequent testing.

9 Marking --

9.1 Packaging P

The following information shall be clearly and Complied, clearly marked P durably marked on the smallest commercially available packaging or legible through it if the packaging is transparent. 9.1.1 The name, trademark or other means of See user manual P identification of the manufacturer or supplier. 9.1.2 Type-identifying marking. P

9.1.3 Classification: FFP1, FFP2, FFP3. FFP2 P

9.1.4 The number and year of publication of this See above P European Standard. 9.1.5 At least the year of end of shelf life. 2 years P

9.1.6 The sentence ‘see information supplied by the English used manufacturer’, at least in the official language(s) of the country of destination, or by using the pictogram as shown in Figure 12b. Page 9 of 13 Report No. CIC200314067-ZJ EN 149 Clause Requirement + Test Result - Remark Verdict

9.1.7 The manufacturer’s recommended conditions of See user manual P storage (at least the temperature and humidity) or equivalent pictogram, as shown in Figures 12c and 12d. 9.1.8 The packaging of those particle filtering half P masks passing the dolomite clogging test shall be additionally marked with the letter "D". 9.2 Particle filtering half mask P

Particle filtering half masks P

Complying with this European Standard shall be P clearly and durably marked with the following: 9.2.1 The name, trademark or other means of P identification of the manufacturer or supplier. 9.2.2 Type-identifying marking. P

9.2.3 The number and year of publication of this See above P European Standard. 9.2.4 The symbols FFP1, FFP2 or FFP3 according to FFP2 P class. 9.2.5 If appropriate the letter D (dolomite) in N accordance with clogging performance. This letter shall follow the class designation (see 9.2.4). 9.2.6 Sub-assemblies and components with N considerable bearing on safety shall be marked so that they can be identified.

10 Information to be supplied by the manufacturer P

10.1 Information supplied by the manufacturer shall be English P at least in the official language(s) of the country of destination. 10.3 The information supplied by the manufacturer See user manual P shall contain all information necessary for trained See user manual and qualified persons on - application/limitations; - the meaning of any colour coding; - checks prior to use; - donning, fitting; -use; - maintenance (e.g. cleaning, disinfecting), if applicable; -storage; - the meaning of any symbols/pictograms used of the equipment. 10.4 The information shall be clear and Clearly considered P comprehensible. If helpful, illustrations, part numbers, marking shall be added. Page 10 of 13 Report No. CIC200314067-ZJ EN 149 Clause Requirement + Test Result - Remark Verdict

10.5 Warning shall be given against problems likely to See user manual P be encountered, for example: - fit of particle filtering half mask (check prior to use); - it is unlikely that the requirements for leakage will be achieved if facial hair passes under the face seal; - air quality (contaminants, oxygen deficiency); - use of equipment in explosive atmosphere. 10.6 The information shall provide recommendations P as to when the particle filtering half mask shall be discarded. Page 11 of 13 Report No. CIC200314067-ZJ EN 149 Clause Requirement + Test Result - Remark Verdict

Attachments: test table

Table 8.5 Leakage test P

Models Sample 1 Sample 2 Sample 3 Sample 4 Sample 5 Item

Test subject walk (km/h) 6 6 6 6 6

Particle size 0.08~1.5 0.08~1.5 0.08~1.5 distribution(um) 0.08~1.5 0.08~1.5

NaCl flow rate 98~101 98~101 98~101 (L/min) 98~101 98~101

NaCl concentration before mask 7.9~8.3 7.9~8.3 7.9~8.3 7.9~8.3 7.9~8.3 (mg/m3)

NaCl concentration after 0.84 0.84 0.85 0.83 0.83 mask (mg/m3)

Note:Testarkvolumeis2m3 Test result total inward Leakage is 9.8%<20%

Table 8.9-1 Inhalation breathing resistance test at 30 L/min P

Models Sample 1 Sample 2 Sample 3 Sample 4 Sample 5 Item

Inhalation gas velocity 30 30 30 30 30 (L/min)

Maximum resistance (mbar) 0.54 0.54 0.53 0.55 0.54

Note: Maximum permitted resistance <0.6 mbar

Table 8.9-2 Inhalation breathing resistance test at 95 L/min P

Models Sample 1 Sample 2 Sample 3 Sample 4 Sample 5 Item Page 12 of 13 Report No. CIC200314067-ZJ EN 149 Clause Requirement + Test Result - Remark Verdict

Inhalation gas velocity 95 95 95 95 95 (L/min)

Maximum resistance (mbar) 1.86 1.86 1.85 1.86 1.85

Note: Maximum permitted resistance <2.1 mbar

Table 8.9-3 Exhalation breathing resistance test at 160 L/min P

Models Sample 1 Sample 2 Sample 3 Sample 4 Sample 5 Item

Exhalation (L/min) 160 160 160 160 160

Maximum resistance (mbar) 2.15 2.16 2.15 2.14 2.14

Note: Maximum permitted resistance < 3.0 mbar Page 13 of 13 Report No. CIC200314067-ZJ

Appendix Photo documentation

Photo 1

View:

[√ ]front

[]rear

[]rightside

[]leftside

[]top

[ ] bottom

[ ] internal

******** End of this report ********