Column Chromatography Column Chromatography

Total Page:16

File Type:pdf, Size:1020Kb

Column Chromatography Column Chromatography Column Chromatography Column chromatography 1- Apparatus: A) A vertical glass tube height /diameter ratio should be at least 20:1 or 50:1 B) A Reservoir of the solvent (eluent) C) Collecting vesseles 2- Chemicals : A- Adsorbent : 1) Insol. in solvents used. 2) Non-reactive with the sample or the solvent 3) White or colorless 4) Appropriate particle size 5) Uniform particle size Types of adsorbents 1- Alumina ( Al2O3 ) 2- Silica gel 3- Kieselguhr : (diatomaceous earth): Used as a carrier to the stationary phase in partition Chromatography. 4- Kaolin 5- Cellulose Powder (Partition Chromatography) 6- Sephadex gel 7- Polyamide (flavonoids) Types of columns 1- Gravity Columns: The mobile phase move through the stationary phase by gravity force. 2- Flash Columns (Air or nitrogen pressure): The mobile phase is pushed by stream of air or nitrogen using special device (Adaptors). 3- Low and Medium Pressure Columns (pumped): The movement of mobile phase is accelerated by using pumps that generate low or medium pressure. The increase in the flow rate shorten the time of separation. 4- High pressure Columns (HPLC): In this columns we use very fine silica gel so great increase in separation power. However, the flow rate of the mobile phase is severely decreased. High pressure pumps are used to push the solvent through the column which in this case must be made of stainless steel. I- Packing of the Column Dry Method: The stationary phase is packed dry and then wetted with the mobile phase. Wet Method: The stationary phase is made a slurry with the mobile phase, and is poured into the column & allowed to settle. 11 Mobile phase: It is a mixture of organic solvents (unusually one solvent only) the choice of the column mobile phase is achieved after TLC study in different solvent systems. Good solvent system must produce Rf value less than 0.6 for all materials to be separated by the column. If the system moves them more and produces higher Rf no separation will occur. Systems that do not move spots at all on TLC are not good for column separation. Isocratic system: Means using the same mobile phase from the beginning to the end of the separation. Gradient: The polarity of the system increased gradually during separation by increasing the proportion of the more polar solvent. A typical gradient may be start with CHCl3, followed by CHCl3/MeOH mixtures with gradual increase in % of MeOH till all spots are eluted from the system. II- Sample Application 1- Wet application: Dissolve the sample in the initial mobile phase and apply by pipette to the top of the column. This is very good method but in most of cases the samples are not soluble in the initial mobile phase. 2- Dry loading: Dissolve sample in any volatile solvent. The sample solution is then adsorbed on small weight of adsorbent and the solvent is allowed to evaporate. The dry adsorbent loaded with the sample is then applied to the column. I- Packing of the Column II- Sample Application III- Mode of Development III- Modes of Chromatographic Separation 1- Frontal Analysis This type of chromatographic development is rarely used and probably is of academic interest only. The sample is fed continuously onto the column as a dilute solution in the mobile phase in contrast to displacement and elution development, where distinct samples are placed on the system and the separation subsequently processed. Frontal analysis can only separate part of the first compound in a relatively pure state, each subsequent component being mixed with those previously eluted. 2- Elution development Is best described as a series of absorption- extraction processes which are continuous from the time the sample is injected into the distribution system until the time the solutes exit from it. Sample components travel through the column at rates determined by their retention on the solid packing Components are eluted from the column at different rates according to their distribution (or partition) coefficients. A complete separation is possible Elution Development E B + E E A + E E E= Eluate, B is more sorbed than A Elution analysis can be done in a number of ways: 1- Single-Solvent Elution : 2- Stepwise gradient elution Benzene Benzene – MeOH (9:1 ) Benzene – MeOH ( 8:2 ) Benzene – MeOH ( 7 : 3) 3- Continuous Gradient Elution : Advantages of gradient Elution 1- It minimize tailing and results in more discrete peaks and hence better resolution 2- Components of a mixture having widely different properties & polarities can be separated in a single operation . 3-Displacement Development Displacement chromatography employs three basically different mobile phases, such as the carrier, displacer, and regenerant (s). The carrier serves to facilitate the load; and to dissolve the sample. The function of the displacer is to displace the sample components from the stationary phase; 3-Displacement Development During this step the individual components also displace each other. By the end of the displacement step, the displacer occupies the stationary phase. Finally, the regenerant removes the displacer from the stationary phase and it prepares the column for the separation of the next sample. In Summary: •The developer is contained in the mobile phase. •The mobile phase must be more sorbed than any sample component •A single pure band of the first component is obtained •There is always an overlap zone for each succeeding component. Displacement analysis D B { A + B A Displacement analysis, D= Displacer which more sorbed than B, which more sorbed than A The following terms are used : Eluent: The chromatographic solvent used in elution analysis. Effluent: All liquid flowing out of the column Eluate: The volume of the effluent that contains eluted substance. Retention volume (Retardation volume): This is the volume of pure solvent flowing-out of the column, before the substance in question leaves it Monitoring the column: Usually fractions of certain volume are collected, evaporated to small volume and spotted on TLC. Similar fractions are collected together for more purification or crystallization. In bioassay guided fraction the fractions are monitored by the bioassay then by TLC. 31 Evaporation of similar collected fractions 32 Factors affecting separation Factors due to Stationary Phase: 1- Particle size of the stationary phase: Reducing the particle size increases the surface area and improve separation. However, reduction of the particle size will decrease the flow rate of the mobile phase. In HPLC we use very fine particles to get very good separation. The flow rate problem is solved by the use high pressure pumps to push the mobile phase through the stationary phase. Columns are made of stainless steel to withstand the high pressure. 2- Adsorbent type: The choice of the suitable adsorbent is very important. 3- Uniformity of packing of the column: If the stationary phase is not packed uniformly then the bands will be irregular and less uniform resulting in poor separation. 4- Concentration of the mixture: the proper ratio between sample to be separated and the amount of stationary phase is very important too much samples resulted in bad separation. Factors due to Mobile Phase: 1- Selection of the proper mobile phase: Very polar mobile phase will wash out all components without any separation. On the other hand very non polar mobile phase will result in broad band and poor separation. 2- Rate of flow: Slower flow rate usually resulted in a better separation and narrower bands. 3- Consistency of flow: The continuous flow of the mobile phase during the whole experiment gives better separation than interrupting the flow then continue it later. Factors due to Columns: Column dimensions: Increasing the length of the column improve separation. However, that usually leads to slower flow rate. Column temperature: Increasing the temperature usually reduces the adsorption power of the stationary phase and increase elution speed. This may leads to decrease in the efficiency of separation. Normal Phasa Chromatography Reversed Phase Chromatography Polar stationary Phase : Silica Non polar stationary Phase (ODS) Non polar mobile phase Octadecylsilane (C18) Polar mobile phase Non polar compounds elute faster polar compounds elute faster than than polar non polar When increase polarity of mobile When increase polarity of mobile phase elution time will increase phase elution time will decrease It can not be reused It can be reused Example of mobile phase: hexane – Example of mobile phase: methanol – chloroform – methylene chloride actonitrl- water .
Recommended publications
  • J . Org. Chem., Vol. 43, No. 14, 1978 2923 A
    Notes J. Org. Chem., Vol. 43, No. 14, 1978 2923 (11) Potassium ferricyanide has previously been used to convert vic-l,2-di- carboxylate groups to double bonds. See, for example, L. F. Fieser and M. J. Haddadin, J. Am. Chem. SOC., 86, 2392 (1964). The oxidative dide- carboxylation of 1,2dicarboxyiic acids is, of course, a well-known process. See inter alia (a) C. A. Grob, M. Ohta, and A. Weiss, Helv. Chirn. Acta, 41, 191 1 [ 1958); and (b) E. N. Cain, R. Vukov, and S. Masamune, J. Chern. SOC. D, 98 (1969). 1 I I L <40 26- 40- 63- 40 63 200 Rapid Chromatographic Technique for Preparative Figure 2. Silica gel particle size6 (pm):(0) rh: (0) r/(w/2). Separations with Moderate Resolution W. Clark Still,* Michael Kahn, and Abhijit Mitra Departm(7nt o/ Chemistry, Columbia Uniuersity, 1Veu York, Neu; York 10027 ReceiLied January 26, 1978 We wish to describe a simple absorption chromatography technique for the routine purification of organic compounds. 0 1.0 2 0 3.0 4.0 Large scale preparative separations are traditionally carried Figure 3. Eluant flow rate (in./min). out by tedious long column chromatography. Although the results are sometimes satisfactory, the technique is always time consuming and frequently gives poor recovery due to band tailing. These problems are especially acute when sam- ples of greater than 1 or 2 g must be separated. In recent years 41 several preparative systems have evolved which reduce sep- aration times to 1-3 h and allow the resolution of components having Al?f 1 0.05 on analytical TLC.
    [Show full text]
  • Synthesis and Applications of Monolithic HPLC Columns
    University of Tennessee, Knoxville TRACE: Tennessee Research and Creative Exchange Doctoral Dissertations Graduate School 8-2005 Synthesis and Applications of Monolithic HPLC Columns Chengdu Liang University of Tennessee - Knoxville Follow this and additional works at: https://trace.tennessee.edu/utk_graddiss Part of the Chemistry Commons Recommended Citation Liang, Chengdu, "Synthesis and Applications of Monolithic HPLC Columns. " PhD diss., University of Tennessee, 2005. https://trace.tennessee.edu/utk_graddiss/2233 This Dissertation is brought to you for free and open access by the Graduate School at TRACE: Tennessee Research and Creative Exchange. It has been accepted for inclusion in Doctoral Dissertations by an authorized administrator of TRACE: Tennessee Research and Creative Exchange. For more information, please contact [email protected]. To the Graduate Council: I am submitting herewith a dissertation written by Chengdu Liang entitled "Synthesis and Applications of Monolithic HPLC Columns." I have examined the final electronic copy of this dissertation for form and content and recommend that it be accepted in partial fulfillment of the requirements for the degree of Doctor of Philosophy, with a major in Chemistry. Georges A Guiochon, Major Professor We have read this dissertation and recommend its acceptance: Sheng Dai, Craig E Barnes, Michael J Sepaniak, Bin Hu Accepted for the Council: Carolyn R. Hodges Vice Provost and Dean of the Graduate School (Original signatures are on file with official studentecor r ds.) To the Graduate Council: I am submitting herewith a dissertation written by Chengdu Liang entitled, “Synthesis and applications of monolithic HPLC columns.” I have examined the final electronic copy of this dissertation for form and content and recommend that it be accepted in partial fulfillment of the requirements for the degree of Doctor of Philosophy, with a major in Chemistry.
    [Show full text]
  • Monolithic Stationary Phases for Capillary Electrochromatography Based on Synthetic Polymers: Designs and Applications Frantisek Svec, Eric C
    Reviews Monolithic Stationary Phases for Capillary Electrochromatography Based on Synthetic Polymers: Designs and Applications Frantisek Svec, Eric C. Peters1), David Sy´kora, Cong Yu, Jean M. J. Fre´chet* Department of Chemistry, University of California, Berkeley, CA 94720-1460, USA Ms received: September 29, 1999; accepted: November 3, 1999 Key Words: Capillary electrochromatography; monolithic columns; synthetic polymers; stationary phase Summary Monolithic materials have quickly become a well-established station- the use of a solid stationary phase and a separation mechan- ary phase format in the field of capillary electrochromatography ism characteristic of HPLC based on specific interactions of (CEC). Both the simplicity of their in situ preparation method and solutes with a stationary phase. Therefore CEC is most com- the large variety of readily available chemistries make the monolithic monly implemented by means typical of both HPLC (packed separation media an attractive alternative to capillary columns packed with particulate materials. This review summarizes the con- columns) and HPCE (use of electrophoretic instrumentation). tributions of numerous groups working in this rapidly growing area, To date, both columns and instrumentation developed specifi- with a focus on monolithic capillary columns prepared from syn- cally for CEC remain scarce. thetic polymers. Various approaches employed for the preparation of the monoliths are detailed, and where available, the material proper- Although numerous groups around the world prepare CEC ties of the resulting monolithic capillary columns are shown. Their columns using a variety of approaches, the vast majority of chromatographic performance is demonstrated by numerous separa- these efforts mimic in one way or another standard HPLC tions of different analyte mixtures in variety of modes.
    [Show full text]
  • Identification Criteria for Qualitative Assays Incorporating Column Chromatography and Mass Spectrometry
    WADA Technical Document – TD2010IDCR Document Number: TD2010IDCR Version Number: 1.0 Written by: WADA Laboratory Committee Approved by: WADA Executive Committee Approval Date: 08 May, 2010 Effective Date: 01 September, 2010 IDENTIFICATION CRITERIA FOR QUALITATIVE ASSAYS INCORPORATING COLUMN CHROMATOGRAPHY AND MASS SPECTROMETRY The ability of a method to identify a compound is a function of the entire procedure: sample preparation; chromatographic separation; mass analysis; and data assessment. Any description of the method for purposes of documentation should include all parts of the method. The appropriate analytical characteristics shall be documented for a particular assay. The Laboratory shall establish criteria for identification of a compound. 1.0 Sample Preparation The purpose of the sample preparation and chromatographic separation is to present a relatively pure chemical component from the sample to the mass spectrometer. The sample purification step can significantly change both the performance of the chromatographic system and the mass spectrometer. For example, a change in extraction solvent can selectively remove interferences and matrix components that might otherwise co-elute with the compound of interest. In addition, selective preparation procedures such as immunoaffinity extraction or fractions collected from high performance liquid chromatography separation can provide a solution that is nearly devoid of any other compounds. 2.0 Chromatographic Separation 2.1 Gas Chromatography • For capillary gas chromatography, the retention time (RT) of the analyte shall not differ by more than two (2) percent or ±0.1 minutes (whichever is smaller) from that of the same substance in a spiked urine sample, Reference Collection sample, or Reference Material analyzed contemporaneously; • Alternatively, the laboratory may choose to use relative retention time (RRT) as an acceptance criterion, where the retention time of the peak of interest is measured relative to a chromatographic reference compound (CRC).
    [Show full text]
  • Column Chromatography Methods of Analysis In
    The Journal of Undergraduate Neuroscience Education (JUNE), Spring 2005, 3(2):A36-A41 ARTICLE Column Chromatography Analysis of Brain Tissue: An Advanced Laboratory Exercise for Neuroscience Majors William H. Church Chemistry Department and Neuroscience Program, Trinity College, Hartford, Connecticut 06106 Neurochemical analysis of discrete brain structures in to useful laboratory techniques such as standard solution experimental animals provides important information on preparation and calibration curve construction, centrifugation, synthesis, release, and metabolism changes following quantitative pipetting, and data evaluation and graphical behavioral or pharmacological experimental manipulations. presentation. Typically, only students that participate in Quantitation of neurotransmitters and their metabolites independent neuroscience research are familiar with these following unilateral drug injections can be carried out using types of quantitative skills. The usefulness of this type of standard chromatographic equipment typically found in most experimental design for understanding behavioral or undergraduate analytical laboratories. This article describes pharmacological effects on neurotransmitter systems is an experiment done in a six session (four hours each) emphasized through a final report requiring a comprehensive component of a neuroscience research methods course. The literature search. experiment provides advanced neuroscience students experience in brain structure dissection, sample preparation, and quantitation of catecholamines
    [Show full text]
  • Recent Trends in the Application of Chromatographic Techniques in the Analysis of Luteolin and Its Derivatives
    biomolecules Review Recent Trends in the Application of Chromatographic Techniques in the Analysis of Luteolin and Its Derivatives Aleksandra Maria Juszczak 1 , Marijana Zovko-Konˇci´c 2 and Michał Tomczyk 1,* 1 Department of Pharmacognosy, Faculty of Pharmacy, Medical University of Białystok, ul. Mickiewicza 2a, 15-230 Białystok, Poland; [email protected] 2 Department of Pharmacognosy, University of Zagreb, Faculty of Pharmacy and Biochemistry, A. Kovaˇci´ca1, 10000 Zagreb, Croatia; [email protected] * Correspondence: [email protected]; Tel.: +48-85-748-5694 Received: 1 October 2019; Accepted: 8 November 2019; Published: 12 November 2019 Abstract: Luteolin is a flavonoid often found in various medicinal plants that exhibits multiple biological effects such as antioxidant, anti-inflammatory and immunomodulatory activity. Commercially available medicinal plants and their preparations containing luteolin are often used in the treatment of hypertension, inflammatory diseases, and even cancer. However, to establish the quality of such preparations, appropriate analytical methods should be used. Therefore, the present paper provides the first comprehensive review of the current analytical methods that were developed and validated for the quantitative determination of luteolin and its C- and O-derivatives including orientin, isoorientin, luteolin 7-O-glucoside and others. It provides a systematic overview of chromatographic analytical techniques including thin layer chromatography (TLC), high performance thin layer chromatography (HPTLC), liquid chromatography (LC), high performance liquid chromatography (HPLC), gas chromatography (GC) and counter-current chromatography (CCC), as well as the conditions used in the determination of luteolin and its derivatives in plant material. Keywords: luteolin; hyphenated techniques; chromatography 1. Introduction Luteolin (Figure1) is a yellow dye commonly found in fresh plants.
    [Show full text]
  • General Methods All Reactions Were Performed Under Nitrogen Atmosphere (Using Standard Schlenk Technique Or Glove Box)
    Supplementary Material (ESI) for Chemical Communications This journal is (c) The Royal Society of Chemistry 2010 Supporting information General methods All reactions were performed under nitrogen atmosphere (using standard Schlenk technique or glove box). All reagents were used as received unless otherwise stated. Tetrahydrofuran was distilled from benzophenone/sodium under nitrogen and stored over sodium in a glove box. Dissolving the substrate to be defluorinated in THF is not necessary. Same results are obtained when used neat. The dilute reaction mixtures described for the defluorination reactions originate from the use of stock solution used to facilitate the addition of the accurate amount of substrate. Column chromatography was performed using Silica Gel 60, 0.04–0.06 mm. 1H, 13C and 19F NMR spectra were recorded at a 400 MHz JEOL Eclipse 400 instrument, operating at 400 MHz for 1H nuclei. An 800 MHz Varian instrument was used as complement for menthyl-2,3,3,3-tetrafluoropropanoate. Samples were dissolved in CDCl3 and chemical shifts are given in ppm (parts per million) relative to residual CHCl3 (7.26 ppm), α,α,α- trifluorotoluene (-63.6 ppm) was used as internal standard for 19F NMR. Gas chromatography/mass spectrometry analyses were performed on a DB-5 equivalent capillary column (length 30m, i.d. 25 μm) using helium as carrier gas. Injector temperature 300 °C. Temperature program: 40 to 330 °C (12 °C/min) with 4 minutes hold time. The MS detector consisted of an ion trap with 70 eV ionization. Preparation of SmI2 in THF Diiodoethane (23 mmol, 6.58 g) was added to dry THF (194 ml).
    [Show full text]
  • The LC Handbook Guide to LC Columns and Method Development Contents
    Agilent CrossLab combines the innovative laboratory services, software, and consumables competencies of Agilent Technologies and provides a direct connection to a global team of scientific and technical experts who deliver vital, actionable insights at every level of the lab environment. Our insights maximize performance, reduce complexity, and drive improved economic, operational, and scientific outcomes. Only Agilent CrossLab offers the unique combination of innovative products and comprehensive solutions to generate immediate results and lasting impact. We partner with our customers to create new opportunities across the lab, around the world, and every step of the way. More information at www.agilent.de/crosslab Learn more The LC Handbook – Guide to LC Columns and Method Development www.agilent.com/chem/lccolumns Buy online www.agilent.com/chem/store Find an Agilent customer center or authorized distributor www.agilent.com/chem/contactus U.S. and Canada 1-800-227-9770 [email protected] Europe [email protected] Asia Pacific [email protected] The LC India [email protected] Handbook Information, descriptions and specifications in this publication are subject to change without notice. Agilent Technologies shall not Guide to LC Columns and be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance or use of this material. Method Development © Agilent Technologies, Inc. 2016 Published in USA, February 1, 2016 Publication Number 5990-7595EN
    [Show full text]
  • F:\IJCAS\Sept 2010 Ijcas\3Aasth
    Aastha Ahuja et al. /International Journal of Chemical and Analytical Science 2010, 1(9),205-207 Review Article Available online through ISSN: 0976-1209 www.ijcas.info Advances In HPLC Column Packing *Aastha Ahuja, Rahul Chib, Mukul Sonker, Geet Sethi, Shweta Gupta , Anju Hooda Delhi Institute of Pharmaceutical Sciences and Research (DIPSAR), University of Delhi Pushp vihar, sector-3, New Delhi-110017 Received on: 20-05-2010; Revised on: 16-06-2010; Accepted on:15-07-2010 ABSTRACT High Performance Liquid Chromatography (HPLC) is a highly improved form of column chromatography in which instead of a solvent being allowed to drip through a column under gravity, it is forced through under high pressures of up to 400 atmospheres(500-5000 p.s.i). The sensitivity and range of the technique depends on the choice of column and on the efficiency of the overall system. Column technology has seen great developments over the years. The transition from large porous particles and pellicular materials to small porous particles occurred in the early 1970s, when micro particulate silica gel (10 -mm dp) came into light and appropriate packing methods were developed. Monolithic columns are a promising alternative to packed columns in the future but much of the research is yet to be done. The use of CAPILLARY COLUMNS (4 mm) has increased in recent years, in part because small-diameter columns use much less solvent and provide higher sensitivity. While speaking of the future of packaging development, silica gel with chemically bonded phases will be around for a long time. The field is a promising one with great scope for research and development.
    [Show full text]
  • Handbook of Size Exclusion Chromatography
    Handbook of Size Exclusion Chromatography CHROMATOGRAPHIC SCIENCE SERIES A Series of Monographs Editor: JACK CAZES Cherry Hill, New Jersey 1. Dynamics of Chromatography, J. Calvin Giddings 2. Gas Chromatographic Analysis of Drugs and Pesticides, Benjamin J. Gudzinowicz 3. Principles of Adsorption Chromatography: The Separation of Nonionic Organic Compounds, Lloyd R. Snyder 4. Multicomponent Chromatography: Theory of Interference, Friedrich Helfferich and Gerhard Klein 5. Quantitative Analysis by Gas Chromatography, Josef Novák 6. High-Speed Liquid Chromatography, Peter M. Rajcsanyi and Elisabeth Rajcsanyi 7. Fundamentals of Integrated GC-MS (in three parts), Benjamin J. Gudzinowicz, Michael J. Gudzinowicz, and Horace F. Martin 8. Liquid Chromatography of Polymers and Related Materials, Jack Cazes 9. GLC and HPLC Determination of Therapeutic Agents (in three parts), Part 1 edited by Kiyoshi Tsuji and Walter Morozowich, Parts 2 and 3 edited by Kiyoshi Tsuji 10. Biological/Biomedical Applications of Liquid Chromatography, edited by Gerald L. Hawk 11. Chromatography in Petroleum Analysis, edited by Klaus H. Altgelt and T. H. Gouw 12. Biological/Biomedical Applications of Liquid Chromatography II, edited by Gerald L. Hawk 13. Liquid Chromatography of Polymers and Related Materials II, edited by Jack Cazes and Xavier Delamare 14. Introduction to Analytical Gas Chromatography: History, Principles, and Practice, John A. Perry 15. Applications of Glass Capillary Gas Chromatography, edited by Walter G. Jennings 16. Steroid Analysis by HPLC: Recent Applications, edited by Marie P. Kautsky 17. Thin-Layer Chromatography: Techniques and Applications, Bernard Fried and Joseph Sherma 18. Biological/Biomedical Applications of Liquid Chromatography III, edited by Gerald L. Hawk 19. Liquid Chromatography of Polymers and Related Materials III, edited by Jack Cazes 20.
    [Show full text]
  • Novel Approaches to Enhancing Selectivity and Efficiency in Microscale Liquid Chromatography
    AN ABSTRACT OF THE DISSERTATION OF Patrick T. Vallano for the degree of Doctor of Philosophy in Chemistry presented on March 6, 2001. Title: Novel Approaches to Enhancing Selectivity and Efficiency in Microscale Liquid Chromatography. Abstract approved: Redacted for Privacy Vincent T. Remcho For a number of reasons, miniaturization of chromatographic columns has been a general trend over the past three decades. Methods designed to enhance selectivity and efficiency can offer improved separation power and speed, expanding on the advantages of miniaturized columns. This dissertation describes novel approaches in this direction, focusing on two areas: the development of affinity-type sorbents for capillary HPLC derived from molecular imprinted polymers (MIPs) and the study of perfusive electroosmosotic flow (EOF) and its effect on efficiency in capillary electrochromatography (CEC). MIPs are synthetic polymers capable of selectively binding a template molecule incorporated prior to polymerization. MIPs prepared using nortripyline,a tricyclic antidepressant drug, were employed to screen a simulated chemical library, consisting of a series of structural analogs and related compounds. A parameterwas introduced to quantify the selective retention of the analytes. Library compounds containing the maj or structural features of the template (ring structure and pendant 2° amine) exhibited the highest affinity for the MIP. The use of macroporous packings under conditions of electroosmotic perfusion can result in improved chromatographic efficiencies. In this work, the performance of CEC columns packed with particles having different nominal pore diameters was investigated. The results indicate that perfusive EOF can yield significant gains in efficiency and speed, especially when wide pore packings and dilute buffers are employed.
    [Show full text]
  • A Glimpse of Thin Layer Chromatography (TLC) and Column Chromatography
    A glimpse of Thin layer Chromatography (TLC) and Column Chromatography (Student reference) Thin layer Chromatography (TLC) Chromatography is a technique for separating two or more compounds or ions by the distribution between two phases in which one is moving and the other is stationary. These two phases can be consisted with solid-liquid, liquid-liquid or gas- liquid. Thin-layer chromatography (TLC) is a solid-liquid form of chromatography in which the stationary phase is normally a polar absorbent and the mobile phase can be a single solvent or combination of solvents. Different affinity of the analyte with the mobile and stationary phases achieves separation of complex mixtures of organic molecules by TLC. Purpose and advantages of TLC TLC is a quick, inexpensive, highly efficient and robust microanalytical (microscale) technique separation method which is associated with many advantages. determine the number of components in a mixture . high sample throughput in a short time . verify a substance’s identity . suitable for screening tests . monitor the progress of the reaction . rapid and cost-efficient optimisation of the separation due to easy change of mobile and stationary phase . determine appropriate conditions for column chromatography . analyze the fractions obtained from column chromatogrpahy . ready-to-use layer acts as storage medium for data Chromatographic adsorbents Eluting solvents for chromatography Stationary phase The common stationary phase is Silica gel with empirical formula SiO2. At the surface of the silica gel, oxygens are usually bound to protons. The presence of hydroxyl groups makes the surface of silica gel highly polar. Consequently, polar functionality in the organic analyte interacts strongly with the surface of the gel particle while nonpolar organic analytes interact only weakly.
    [Show full text]