F:\IJCAS\Sept 2010 Ijcas\3Aasth

Total Page:16

File Type:pdf, Size:1020Kb

F:\IJCAS\Sept 2010 Ijcas\3Aasth Aastha Ahuja et al. /International Journal of Chemical and Analytical Science 2010, 1(9),205-207 Review Article Available online through ISSN: 0976-1209 www.ijcas.info Advances In HPLC Column Packing *Aastha Ahuja, Rahul Chib, Mukul Sonker, Geet Sethi, Shweta Gupta , Anju Hooda Delhi Institute of Pharmaceutical Sciences and Research (DIPSAR), University of Delhi Pushp vihar, sector-3, New Delhi-110017 Received on: 20-05-2010; Revised on: 16-06-2010; Accepted on:15-07-2010 ABSTRACT High Performance Liquid Chromatography (HPLC) is a highly improved form of column chromatography in which instead of a solvent being allowed to drip through a column under gravity, it is forced through under high pressures of up to 400 atmospheres(500-5000 p.s.i). The sensitivity and range of the technique depends on the choice of column and on the efficiency of the overall system. Column technology has seen great developments over the years. The transition from large porous particles and pellicular materials to small porous particles occurred in the early 1970s, when micro particulate silica gel (10 -mm dp) came into light and appropriate packing methods were developed. Monolithic columns are a promising alternative to packed columns in the future but much of the research is yet to be done. The use of CAPILLARY COLUMNS (4 mm) has increased in recent years, in part because small-diameter columns use much less solvent and provide higher sensitivity. While speaking of the future of packaging development, silica gel with chemically bonded phases will be around for a long time. The field is a promising one with great scope for research and development. Keywords: HPLC, column packing, Perfusion chromatography,Monoliths. INTRODUCTION HPLC was developed in the late 1960s and early 1970s. Today, it is widely not adsorbed elutes faster. Thus, the adsorption decides the retention time of the applied for separations and purifications in a variety of areas including phar- molecule. After passing through the column the mobile phase enters into the maceuticals, biotechnology, environmental, polymer and food industries. [1] HPLC detector attached to the printer, where the molecule is detected and print is has over the past decade become the method of choice for the analysis of a wide taken for graphical separation of the molecule. Column technology has devel- variety of compounds. The main advantage of HPLC over GC is that the oped alongside the development of HPLC instrumentation the new “HPLC” analytes need not necessarily be volatile, so macromolecules are suitable for instruments could develop up to 6,000psi (400 bar) of pressure, and included HPLC analysis. [2] improved detectors and columns. With continued advances in performance, the name was changed to HPLC from liquid chromatography (LC). Today, trace Principle: concentrations of compounds, as low as “parts per trillion” (ppt), are easily In isocratic HPLC the analyte is forced through a column of the stationary obtained. [4] phase (usually a tube packed with small round particles with a certain surface chemistry) by pumping a liquid (mobile phase) at high pressure through the History Of HPLC column. The sample to be analyzed is introduced in a small volume to the stream Very few reliable chromatographic methods were available to the scientists, of mobile phase and is retarded by specific chemical or physical interactions before the 1970’s. During 1970’s, a number of techniques including open-col- with the stationary phase as it traverses the length of the column. Nature of the umn chromatography, paper chromatography, and thin-layer chromatography, analyte, stationary phase and mobile phase composition accounts for the retar- were used to carry out chemical separations. But these were however inadequate dation. The time at which a specific analyte elutes (comes out of the end of the for quantifying compounds and for resolution between similar compounds. Be- column) is called the retention time and is considered a reasonably unique fore the advent of HPLC, low-pressure, gravity-fed columns with large, irregu- identifying characteristic of a given analyte. Pressure increases the linear veloc- larly shaped porous particles such as silica gel or alumina were frequently used. ity (speed) of the component providing it less time to diffuse within the column, And, because of their high surface area, these porous particles had excellent which leads to improved resolution in the resulting chromatogram. Any miscible sample capacity. However, these large particle sizes gave poor column effi- combinations of water or various organic liquids (the most common are metha- ciency. Thus, Pressure liquid chromatography began to be used, during this time, nol and acetonitrile) serves as the solvent system. Water may contain buffers or to decrease flow through time, thereby reducing purification time of compounds salts to assist in the separation of the analyte components, or compounds such being isolated by column chromatography. as Trifluoroacetic acid which acts as an ion pairing agent In spite of being considered somewhat mature, new developments in HPLC still continue. There Flow rates still continued to remain inconsistent. HPLC was developed in the have been improvements in column construction, packing material design, mid- 1970’s and made better with the development of column packing materials bonded phase chemistry and formats. In addition to this, new phases have and the convenience of on-line detectors. In late 1970’s, new methods including extended the pH range (high and low), providing it more versatility. [3] reverse phase liquid chromatography became popular. It made separation be- tween very similar compounds easier. HPLC was frequently used for the separa- Instrumentation: tion of chemical compounds, by the 1980’s. The new techniques made separa- The heart of a HPLC system is the column. The column contains the particle tion, identification, purification and quantification much easier, better and re- that contains the Stationary phase. Pump forces the mobile phase to pass producible. Computers and automation further added to the convenience of through the column and injector injects the mixture to be separated into the HPLC. flowing mobile phase. Molecules that are adsorbed maximum by stationary phase are eluted slowest through the column, whereas the molecules which are The last decade saw vast improvements in the form of development of micro- columns, and other specialized columns. The dimensions of the typical HPLC column are: 250 mm in length with an internal diameter between 3-5 mm and *Corresponding author. that of micro-columns, or capillary columns, ranges from 3 µm to 200 µm. Fast Aastha Ahuja HPLC makes use of a column that is shorter than the characteristic column, Delhi Institute of Pharmaceutical Sciences and Research (DIPSAR), with a length of about 3 mm, and they are packed with smaller particles. [5, 6] University of Delhi,Pushp vihar, sector-3, New Delhi-110017 Tel.: + 91-9899430775 E-mail:[email protected] International Journal of Chemical and Analytical Science Vol.1.Issue 9.September 2010 205-207 Aastha Ahuja et al. /International Journal of Chemical and Analytical Science 2010, 1(9),205-207 Development In Porous Packings Comparison of current packing types in HPLC Porous packings have been in favor throughout the history of HPLC. The transition from large porous particles and pellicular materials to small porous c d particles occurred in the early 1970s, when micro particulate silica gel (10 -mm dp) came in the scene and appropriate packing methods were developed. Until Diffusive pore spherical materials were developed and perfected, till the 1970s irregularly shaped micro particulate packings were in vogue. The spherical packings had Through pore several advantages over their predecessors, such as; these could be packed more homogeneously, gave better efficiencies, and could be manufactured in higher purity. [7] Totally porous particle Perfusion particle The major surface area of the particle is contained within the diffusive pores Comparison of current packing types in HPLC which dominate the typical porous packing (Figure-1). In a porous particle, solutes get transferred from the moving mobile phase, outside of the particles, Thin porous e Porous f into the stagnant mobile phase within the pores, to interact with the stationary Shell phase. Particle size reduction improves both the interparticle mass transfer as layer well as the intraparticle mass transfer. After this interaction, the solute molecule must diffuse out of the particle and continue its journey down the column. Such a mass transfer takes place many thousands of times as the differential separa- tion process advances and the solute is eluted from the column. The mobile phase in which the solute was originally present, moves down the column ahead Poroshell particle of the solute, while the solute spends its time in the diffusive pores. This slow Non -porous silica (NPS) rate of mass transfer into and out of the porous particle is a major source of band or nonporous resin (NPR) broadening in HPLC. The advantage of using smaller particles is that they Fig.1. Pictorial comparison of liquid chromatorgraphy column packing shorten the path length of this diffusion process, improve mass transfer, and provide better efficiency. But however, congruent with the improvement in a. Large, totally, porous particles (100 +m m) (normally irregular shaped) efficiency was the decrease in column permeability; that is, an increase in b. Superficially porous particle (50 m m) ( also called as pellicular or porous layer bead column backpressure. The increase in pressure is proportional to the inverse of c. Microparticulate totally porous particle (5 m m) the square of the diameter of the particle. Thus, halving particle diameter will d. Perfusion particle (12 m m) increase the column head pressure by a factor of four. A current trend in HPLC e. Non-porous silica (NPS) or nonporous resin (NPR) for high-throughput separations is to use shorter columns with smaller particles f.
Recommended publications
  • J . Org. Chem., Vol. 43, No. 14, 1978 2923 A
    Notes J. Org. Chem., Vol. 43, No. 14, 1978 2923 (11) Potassium ferricyanide has previously been used to convert vic-l,2-di- carboxylate groups to double bonds. See, for example, L. F. Fieser and M. J. Haddadin, J. Am. Chem. SOC., 86, 2392 (1964). The oxidative dide- carboxylation of 1,2dicarboxyiic acids is, of course, a well-known process. See inter alia (a) C. A. Grob, M. Ohta, and A. Weiss, Helv. Chirn. Acta, 41, 191 1 [ 1958); and (b) E. N. Cain, R. Vukov, and S. Masamune, J. Chern. SOC. D, 98 (1969). 1 I I L <40 26- 40- 63- 40 63 200 Rapid Chromatographic Technique for Preparative Figure 2. Silica gel particle size6 (pm):(0) rh: (0) r/(w/2). Separations with Moderate Resolution W. Clark Still,* Michael Kahn, and Abhijit Mitra Departm(7nt o/ Chemistry, Columbia Uniuersity, 1Veu York, Neu; York 10027 ReceiLied January 26, 1978 We wish to describe a simple absorption chromatography technique for the routine purification of organic compounds. 0 1.0 2 0 3.0 4.0 Large scale preparative separations are traditionally carried Figure 3. Eluant flow rate (in./min). out by tedious long column chromatography. Although the results are sometimes satisfactory, the technique is always time consuming and frequently gives poor recovery due to band tailing. These problems are especially acute when sam- ples of greater than 1 or 2 g must be separated. In recent years 41 several preparative systems have evolved which reduce sep- aration times to 1-3 h and allow the resolution of components having Al?f 1 0.05 on analytical TLC.
    [Show full text]
  • Synthesis and Applications of Monolithic HPLC Columns
    University of Tennessee, Knoxville TRACE: Tennessee Research and Creative Exchange Doctoral Dissertations Graduate School 8-2005 Synthesis and Applications of Monolithic HPLC Columns Chengdu Liang University of Tennessee - Knoxville Follow this and additional works at: https://trace.tennessee.edu/utk_graddiss Part of the Chemistry Commons Recommended Citation Liang, Chengdu, "Synthesis and Applications of Monolithic HPLC Columns. " PhD diss., University of Tennessee, 2005. https://trace.tennessee.edu/utk_graddiss/2233 This Dissertation is brought to you for free and open access by the Graduate School at TRACE: Tennessee Research and Creative Exchange. It has been accepted for inclusion in Doctoral Dissertations by an authorized administrator of TRACE: Tennessee Research and Creative Exchange. For more information, please contact [email protected]. To the Graduate Council: I am submitting herewith a dissertation written by Chengdu Liang entitled "Synthesis and Applications of Monolithic HPLC Columns." I have examined the final electronic copy of this dissertation for form and content and recommend that it be accepted in partial fulfillment of the requirements for the degree of Doctor of Philosophy, with a major in Chemistry. Georges A Guiochon, Major Professor We have read this dissertation and recommend its acceptance: Sheng Dai, Craig E Barnes, Michael J Sepaniak, Bin Hu Accepted for the Council: Carolyn R. Hodges Vice Provost and Dean of the Graduate School (Original signatures are on file with official studentecor r ds.) To the Graduate Council: I am submitting herewith a dissertation written by Chengdu Liang entitled, “Synthesis and applications of monolithic HPLC columns.” I have examined the final electronic copy of this dissertation for form and content and recommend that it be accepted in partial fulfillment of the requirements for the degree of Doctor of Philosophy, with a major in Chemistry.
    [Show full text]
  • Monolithic Stationary Phases for Capillary Electrochromatography Based on Synthetic Polymers: Designs and Applications Frantisek Svec, Eric C
    Reviews Monolithic Stationary Phases for Capillary Electrochromatography Based on Synthetic Polymers: Designs and Applications Frantisek Svec, Eric C. Peters1), David Sy´kora, Cong Yu, Jean M. J. Fre´chet* Department of Chemistry, University of California, Berkeley, CA 94720-1460, USA Ms received: September 29, 1999; accepted: November 3, 1999 Key Words: Capillary electrochromatography; monolithic columns; synthetic polymers; stationary phase Summary Monolithic materials have quickly become a well-established station- the use of a solid stationary phase and a separation mechan- ary phase format in the field of capillary electrochromatography ism characteristic of HPLC based on specific interactions of (CEC). Both the simplicity of their in situ preparation method and solutes with a stationary phase. Therefore CEC is most com- the large variety of readily available chemistries make the monolithic monly implemented by means typical of both HPLC (packed separation media an attractive alternative to capillary columns packed with particulate materials. This review summarizes the con- columns) and HPCE (use of electrophoretic instrumentation). tributions of numerous groups working in this rapidly growing area, To date, both columns and instrumentation developed specifi- with a focus on monolithic capillary columns prepared from syn- cally for CEC remain scarce. thetic polymers. Various approaches employed for the preparation of the monoliths are detailed, and where available, the material proper- Although numerous groups around the world prepare CEC ties of the resulting monolithic capillary columns are shown. Their columns using a variety of approaches, the vast majority of chromatographic performance is demonstrated by numerous separa- these efforts mimic in one way or another standard HPLC tions of different analyte mixtures in variety of modes.
    [Show full text]
  • Identification Criteria for Qualitative Assays Incorporating Column Chromatography and Mass Spectrometry
    WADA Technical Document – TD2010IDCR Document Number: TD2010IDCR Version Number: 1.0 Written by: WADA Laboratory Committee Approved by: WADA Executive Committee Approval Date: 08 May, 2010 Effective Date: 01 September, 2010 IDENTIFICATION CRITERIA FOR QUALITATIVE ASSAYS INCORPORATING COLUMN CHROMATOGRAPHY AND MASS SPECTROMETRY The ability of a method to identify a compound is a function of the entire procedure: sample preparation; chromatographic separation; mass analysis; and data assessment. Any description of the method for purposes of documentation should include all parts of the method. The appropriate analytical characteristics shall be documented for a particular assay. The Laboratory shall establish criteria for identification of a compound. 1.0 Sample Preparation The purpose of the sample preparation and chromatographic separation is to present a relatively pure chemical component from the sample to the mass spectrometer. The sample purification step can significantly change both the performance of the chromatographic system and the mass spectrometer. For example, a change in extraction solvent can selectively remove interferences and matrix components that might otherwise co-elute with the compound of interest. In addition, selective preparation procedures such as immunoaffinity extraction or fractions collected from high performance liquid chromatography separation can provide a solution that is nearly devoid of any other compounds. 2.0 Chromatographic Separation 2.1 Gas Chromatography • For capillary gas chromatography, the retention time (RT) of the analyte shall not differ by more than two (2) percent or ±0.1 minutes (whichever is smaller) from that of the same substance in a spiked urine sample, Reference Collection sample, or Reference Material analyzed contemporaneously; • Alternatively, the laboratory may choose to use relative retention time (RRT) as an acceptance criterion, where the retention time of the peak of interest is measured relative to a chromatographic reference compound (CRC).
    [Show full text]
  • I. Chromatography
    Chromatography & Electrophoresis I. Chromatography • Chroma: “color” + graph: “measure” Separation Techniques – Developed in 19th century to separate pigments from plants and dyes Purifying & analyzing macromolecules – Refined technique now used to separate, (e.g., proteins & DNA) from a mixture analyze, & purify wide range of compounds I. Chromatography • Still called “chromatography” even though no longer limited to separating colored molecules II. Electrophoresis Mixture of macromolecule solutes from Chromatography sample dissolved in mobile phase Stationary phase: gel, beads, or other solid medium with specific absorptive characteristics Mobile phase: fluid solvent that flows over/through the stationary phase Separation of macromolecule solutes Solutes interact with stationary phase from mixture in original sample — ↑ interaction Æ Ø migration rate based upon relative rates of migration through the stationary phase Ø interaction Æ ↑ migration rate •Solutes that interact strongly with the stationary phase do not migrate very far Types of interactions: Interactions may be modified by manipulating •Solutes that interact with the stationary phase • Size migrate slower than the mobile phase • Charge the mobile phase: ∆ pH • Polarity/hydrophobicity ∆ Ionic strength • Non-specific binding ∆ Polarity •Solutes that do not interact with the stationary phase • Specific binding migrate at the same rate as the mobile phase ∆ Competitive binding Bruce Heyer 1 Chromatography & Electrophoresis Column chromatography Column chromatography • Solid phase as beads or particles packed in a cylinder • Solid phase as beads or particles packed in a cylinder Load sample containing mixture of macromolecules onto top of solid phase bed Pattern of bands reveals which • Elution macromolecules were in the sample mixture – Allows collecting separated fractions containing purified macromolecules Elute mobile phase fluid from bottom of column while adding fresh mobile phase to top Other types of chromatography II.
    [Show full text]
  • Column Chromatography Methods of Analysis In
    The Journal of Undergraduate Neuroscience Education (JUNE), Spring 2005, 3(2):A36-A41 ARTICLE Column Chromatography Analysis of Brain Tissue: An Advanced Laboratory Exercise for Neuroscience Majors William H. Church Chemistry Department and Neuroscience Program, Trinity College, Hartford, Connecticut 06106 Neurochemical analysis of discrete brain structures in to useful laboratory techniques such as standard solution experimental animals provides important information on preparation and calibration curve construction, centrifugation, synthesis, release, and metabolism changes following quantitative pipetting, and data evaluation and graphical behavioral or pharmacological experimental manipulations. presentation. Typically, only students that participate in Quantitation of neurotransmitters and their metabolites independent neuroscience research are familiar with these following unilateral drug injections can be carried out using types of quantitative skills. The usefulness of this type of standard chromatographic equipment typically found in most experimental design for understanding behavioral or undergraduate analytical laboratories. This article describes pharmacological effects on neurotransmitter systems is an experiment done in a six session (four hours each) emphasized through a final report requiring a comprehensive component of a neuroscience research methods course. The literature search. experiment provides advanced neuroscience students experience in brain structure dissection, sample preparation, and quantitation of catecholamines
    [Show full text]
  • Recent Trends in the Application of Chromatographic Techniques in the Analysis of Luteolin and Its Derivatives
    biomolecules Review Recent Trends in the Application of Chromatographic Techniques in the Analysis of Luteolin and Its Derivatives Aleksandra Maria Juszczak 1 , Marijana Zovko-Konˇci´c 2 and Michał Tomczyk 1,* 1 Department of Pharmacognosy, Faculty of Pharmacy, Medical University of Białystok, ul. Mickiewicza 2a, 15-230 Białystok, Poland; [email protected] 2 Department of Pharmacognosy, University of Zagreb, Faculty of Pharmacy and Biochemistry, A. Kovaˇci´ca1, 10000 Zagreb, Croatia; [email protected] * Correspondence: [email protected]; Tel.: +48-85-748-5694 Received: 1 October 2019; Accepted: 8 November 2019; Published: 12 November 2019 Abstract: Luteolin is a flavonoid often found in various medicinal plants that exhibits multiple biological effects such as antioxidant, anti-inflammatory and immunomodulatory activity. Commercially available medicinal plants and their preparations containing luteolin are often used in the treatment of hypertension, inflammatory diseases, and even cancer. However, to establish the quality of such preparations, appropriate analytical methods should be used. Therefore, the present paper provides the first comprehensive review of the current analytical methods that were developed and validated for the quantitative determination of luteolin and its C- and O-derivatives including orientin, isoorientin, luteolin 7-O-glucoside and others. It provides a systematic overview of chromatographic analytical techniques including thin layer chromatography (TLC), high performance thin layer chromatography (HPTLC), liquid chromatography (LC), high performance liquid chromatography (HPLC), gas chromatography (GC) and counter-current chromatography (CCC), as well as the conditions used in the determination of luteolin and its derivatives in plant material. Keywords: luteolin; hyphenated techniques; chromatography 1. Introduction Luteolin (Figure1) is a yellow dye commonly found in fresh plants.
    [Show full text]
  • General Methods All Reactions Were Performed Under Nitrogen Atmosphere (Using Standard Schlenk Technique Or Glove Box)
    Supplementary Material (ESI) for Chemical Communications This journal is (c) The Royal Society of Chemistry 2010 Supporting information General methods All reactions were performed under nitrogen atmosphere (using standard Schlenk technique or glove box). All reagents were used as received unless otherwise stated. Tetrahydrofuran was distilled from benzophenone/sodium under nitrogen and stored over sodium in a glove box. Dissolving the substrate to be defluorinated in THF is not necessary. Same results are obtained when used neat. The dilute reaction mixtures described for the defluorination reactions originate from the use of stock solution used to facilitate the addition of the accurate amount of substrate. Column chromatography was performed using Silica Gel 60, 0.04–0.06 mm. 1H, 13C and 19F NMR spectra were recorded at a 400 MHz JEOL Eclipse 400 instrument, operating at 400 MHz for 1H nuclei. An 800 MHz Varian instrument was used as complement for menthyl-2,3,3,3-tetrafluoropropanoate. Samples were dissolved in CDCl3 and chemical shifts are given in ppm (parts per million) relative to residual CHCl3 (7.26 ppm), α,α,α- trifluorotoluene (-63.6 ppm) was used as internal standard for 19F NMR. Gas chromatography/mass spectrometry analyses were performed on a DB-5 equivalent capillary column (length 30m, i.d. 25 μm) using helium as carrier gas. Injector temperature 300 °C. Temperature program: 40 to 330 °C (12 °C/min) with 4 minutes hold time. The MS detector consisted of an ion trap with 70 eV ionization. Preparation of SmI2 in THF Diiodoethane (23 mmol, 6.58 g) was added to dry THF (194 ml).
    [Show full text]
  • The LC Handbook Guide to LC Columns and Method Development Contents
    Agilent CrossLab combines the innovative laboratory services, software, and consumables competencies of Agilent Technologies and provides a direct connection to a global team of scientific and technical experts who deliver vital, actionable insights at every level of the lab environment. Our insights maximize performance, reduce complexity, and drive improved economic, operational, and scientific outcomes. Only Agilent CrossLab offers the unique combination of innovative products and comprehensive solutions to generate immediate results and lasting impact. We partner with our customers to create new opportunities across the lab, around the world, and every step of the way. More information at www.agilent.de/crosslab Learn more The LC Handbook – Guide to LC Columns and Method Development www.agilent.com/chem/lccolumns Buy online www.agilent.com/chem/store Find an Agilent customer center or authorized distributor www.agilent.com/chem/contactus U.S. and Canada 1-800-227-9770 [email protected] Europe [email protected] Asia Pacific [email protected] The LC India [email protected] Handbook Information, descriptions and specifications in this publication are subject to change without notice. Agilent Technologies shall not Guide to LC Columns and be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance or use of this material. Method Development © Agilent Technologies, Inc. 2016 Published in USA, February 1, 2016 Publication Number 5990-7595EN
    [Show full text]
  • Gas Chromatography (GC), High Performance Liquid Chromatography (HPLC), Ion Exchange Chromatography (IC), Super Critical Fluid Chromatography (SFC)
    Chromatography For S. Y. B. Sc. Sem-IV By: Dr Vipul B. Kataria The technique was invented by Russian botanist Mikhail tswett in 1906. The word Chromatography is derived from latin language and in this word chroma means “colour” and graphein means “to write”. It can be defined as a method of separating a mixture of components into individual component through equilibrium distribution between two phases. Essentially, it is based upon the differences in rate at which the components of mixture move through a porous medium (stationary phase) under the influence of gas or liquid (mobile phase). Some basic terminology Analyte: substance that to be analysed. Chromatogram: Visual output of chromatography. Stationary phase: Non-moving phase that provides support for mixture to move. Mobile phase: Moving phase to specific direction. Eluent: Solvent that carries analyte. Retention time: it is specific time required for a particular analyte to pass through the system. Classification of chromatography Chromatography can be classified into different way based upon different aspects. 1 | Page 1. Based upon geometry On the basis of geometry, chromatography can be classified into following. a. Planar: Paper chromatography and thin layer chromatography (TLC) b. Column: Gas chromatography (GC), High performance liquid chromatography (HPLC), Ion exchange chromatography (IC), Super critical fluid chromatography (SFC). 2. Based upon mobile phase On the basis of mobile phase, chromatography can be classified into following. a. Gas chromatography: Mobile phase is gas. b. Liquid chromatography: Mobile phase is liquid. c. SFC: Supercritical fluid chromatography. Mobile phase is CO2 in liquid form. 3. Based upon stationary phase On the basis of stationary phase, Chromatography can be classified into following.
    [Show full text]
  • LC Columns and Accessories
    Thermo Scientific Chromatography Columns and Consumables Thermo Scientific Chromatography Columns and Consumables 2014-2015 2014-2015 Your world of chromatography predictability LC Columns and Accessories As a leader in LC column technology including silica, polymer and porous graphitic carbon manufacturing, bonded phase production and column packing for over 35 years, you can rely on the quality of Thermo Scientific HPLC products. Here we 2014-2015 showcase our latest and most comprehensive range of innovative columns, accessories and equipment for fast LC, analytical HPLC and biomolecule separations. Chromatography Columns and Consumables Thermo Scienti c Thermo Scienti 4 LC Columns and Accessories Section Contents Introduction . 4-001 LC Accessories. 4-158 Column Selection. 4-002 LC Equipment . 4-186 Column Protection . 4-016 LC Reagents . 4-193 Columns for Fast and Analytical LC . 4-022 LC and LC-MS Instrument Parts . 4-209 Columns for Biomolecules . 4-110 Technical Information. 4-231 Featured Products Accucore Columns Syncronis Columns Hypersil GOLD Columns Ultimate core performance - Consistent, reproducible separations, Outstanding peak shape for speed and selectivity combined column after column, time after time your separations PAGE 4-042 PAGE 4-074 PAGE 4-024 MAbPac Columns Acclaim PepMap Columns Viper Connectors Unrivalled resolution and ef ciency The standard for peptide Simple, dead volume free plumbing in the analysis of protein variants separations in proteomics of HPLC and UHPLC systems PAGE 4-126 PAGE 4-149 PAGE 4-158 www.thermoscienti c.com/chromatography 4-001 LC Column Selection Information in the following section will help you make an informed decision on the appropriate HPLC column for your application, based on stationary phase use, analyte properties, LC-MS requirements or USP specifications.
    [Show full text]
  • Handbook of Size Exclusion Chromatography
    Handbook of Size Exclusion Chromatography CHROMATOGRAPHIC SCIENCE SERIES A Series of Monographs Editor: JACK CAZES Cherry Hill, New Jersey 1. Dynamics of Chromatography, J. Calvin Giddings 2. Gas Chromatographic Analysis of Drugs and Pesticides, Benjamin J. Gudzinowicz 3. Principles of Adsorption Chromatography: The Separation of Nonionic Organic Compounds, Lloyd R. Snyder 4. Multicomponent Chromatography: Theory of Interference, Friedrich Helfferich and Gerhard Klein 5. Quantitative Analysis by Gas Chromatography, Josef Novák 6. High-Speed Liquid Chromatography, Peter M. Rajcsanyi and Elisabeth Rajcsanyi 7. Fundamentals of Integrated GC-MS (in three parts), Benjamin J. Gudzinowicz, Michael J. Gudzinowicz, and Horace F. Martin 8. Liquid Chromatography of Polymers and Related Materials, Jack Cazes 9. GLC and HPLC Determination of Therapeutic Agents (in three parts), Part 1 edited by Kiyoshi Tsuji and Walter Morozowich, Parts 2 and 3 edited by Kiyoshi Tsuji 10. Biological/Biomedical Applications of Liquid Chromatography, edited by Gerald L. Hawk 11. Chromatography in Petroleum Analysis, edited by Klaus H. Altgelt and T. H. Gouw 12. Biological/Biomedical Applications of Liquid Chromatography II, edited by Gerald L. Hawk 13. Liquid Chromatography of Polymers and Related Materials II, edited by Jack Cazes and Xavier Delamare 14. Introduction to Analytical Gas Chromatography: History, Principles, and Practice, John A. Perry 15. Applications of Glass Capillary Gas Chromatography, edited by Walter G. Jennings 16. Steroid Analysis by HPLC: Recent Applications, edited by Marie P. Kautsky 17. Thin-Layer Chromatography: Techniques and Applications, Bernard Fried and Joseph Sherma 18. Biological/Biomedical Applications of Liquid Chromatography III, edited by Gerald L. Hawk 19. Liquid Chromatography of Polymers and Related Materials III, edited by Jack Cazes 20.
    [Show full text]