Pol J Med Phys Eng 2007;13(1):1-12. PL ISSN 1425-4689 doi: 10.2478/v10013-007-0001-x website: http://www.pjmpe.waw.pl

Grzegorz Domański, Bogumił Konarzewski, Zdzisław Pawłowski, Krzysztof Zaremba, Janusz Marzec, Artur Trybuła, Robert Kurjata

A Simple Method of Determining the Effective Coefficient

Institute of Radioelectronics, Warsaw University of Technology, Nowowiejska 15/19, 00-665 Warsaw, Poland e-mail: [email protected]

This paper presents a simple method of determining the effective attenuation coefficient from steady-state diffuse measurements. Key words: optical parameters, effective attenuation coefficient, reflectance measurements.

Introduction

Visible and near- (NIR) interacts with biological tissues by absorption and [1, 2, 3, 4, 5]. It is important to know the optical properties of tissues in order to properly choose the measurement geometry in optical tomography [2, 3]. Also, optical properties themselves can potentially provide information about the tissue’s metabolic status or diagnose a disease [2, 3]. Optical properties of phantom tissues and biological tissues have been described by authors using various methods and techniques [6, 7]. However, it is still a problem of how to determine these optical properties quickly and relatively cheaply [6, 7]. The aim of this work was to develop a simple and cheap method for effective in vivo estimation of the attenuation coefficient. 2 Grzegorz Domański et al.

Theory and Modelling

In this section a simple model of radiation transport is described, which was used to derive an effective absorption coefficient from spatially resolved measurements of diffuse reflectance. A diffusion model was used to generate analytical expressions for the reflectance. For an infinitive medium the solution of the diffusion equation is given by [3]:

I exp(-m r ) F()r =×0 eff (1) 4pD r

F 2 where: is the fluence rate [W/cm ], I0 is the source intensity [W], D is the 1 diffusion coefficient [6] given by D = , m is an absorption coefficient [cm­1] mm+ ¢ a 3(as ) m ­1 mm¢ =- and s is the scattering coefficient [cm ], ss()1 g being the reduced scattering m coefficient, g is the anisotropy coefficient, and m = a is the effective attenuation eff D coefficient [7]. From equation (1) it can clearly be seen that in the diffusion theory for an m infinitive medium light attenuation in tissue depends only on eff rather than on the m absorption coefficient or the scattering coefficient alone. The knowledge of eff enables us to quickly evaluate light attenuation between a source and a detector.

A quotient of two photon fluence rates for two different source-to-detector distances r and r0 is given by

F r =--0 exp()m (rr ) F eff 0 0 r

Its natural logarithm is

æ F ö æ r ö lnç ÷ = lnç 0 ÷ --m (rr ) (2) ç F ÷ eff 0 è 0 ø è r ø

For large source-to-detector distances, the above function behaves asymptotically m as a linear function with a slope ­ eff, which makes its determination fast and easy. The Effective Attenuation Coefficient… 3

æ F ö Let the logarithm of the measured quotient of light intensity be denoted by y = lnç ÷ ç F ÷ è 0 ø and the distances from the source to the detector at the measured points by ri, where i =

1, …, n. The values y, measured at points ri, are denoted by yi. Then æ F ö æ r ö fr(,)lnmm= ç ÷ = lnç 0 ÷ -- (rr ) (3) effç F ÷ eff 0 è 0 ø è r ø

m In order to determine the value of eff, the sum of squares of the deviations of the m theoretical signal values y=f( eff,r)from the measured signals yi is minimized. Let the sum be written as n =-m 2 Syfrå()i (,)eff i (4) i =1 m Minimization of S in view of eff leads to dS = 0 (5) m d eff

By computing the derivative we obtain

n ¶Sf¶ 2 =-=å []()yf(,r)m ¶m ¶m i eff i effi =1 eff n é ¶f ù =-å 2()yf(,r)m - (,r)m = i eff i ê ¶m eff i ú i =1 ë eff û n =-(,r)m - 2()å()yfi eff ii rr0 i =1

After comparing it to zero, we obtain

n æ æ r ö ö 2åç y - lnç 0 ÷ + m=(r - r )÷ (r - r ) 0 (6) ç i ç r ÷ eff ii00÷ i =1 è è i ø ø

From the above we finally have

n æ æ r öö n åç y - lnç 0 ÷÷(r -r )+m å (r -r )2 = 0 (7) ç i ç r ÷÷ i 00eff i i =1 è è i øø i=1 4 Grzegorz Domański et al.

m The value of eff, which minimizes the cost function S, can be computed from the following formula

n æ æ r öö ç - ç 0 ÷÷(r - r ) åç yi lnç ÷÷ i 0 = è r ø m =- i 1 è i ø eff n (8) (r - r ) 2 å i 0 i=1

m With the aim of finding the uncertainty of the estimated parameter eff, let us assume that the yi measurement is a random variable with the variance var(yi), and the m random variables yi are independent for different values of i. The variance var( eff) is then given by

2 n æ ¶m ö var(m ) = åç eff ÷ var(y ) (9) eff ç ¶ ÷ i i =1 è yi ø

Its derivative is given by

¶m rr- eff =- i 0 (10) n ¶ 2 yi - ) å(rrk 0 k=1

m The variance var( eff)is

n - ) 2 å(var()rri 0 yi m = i =1 var(eff ) 2 (11) æ n ö - ) 2 ç å(rri 0 ÷ è i =1 ø

s m The standard deviation ( eff)is

n - ) 2 å(var()rri 0 yi sm()var()== m i =1 (12) eff eff n - ) 2 å(rri 0 i =1 The Effective Attenuation Coefficient… 5

The variance var(yi) is expressed by the measured light intensity

2 é æ F öù é ¶ æ æ F ööù var(y )= var lnç i ÷ = ê ç lnç i ÷÷ú var(F ) = i ê ç FF÷ú ¶ ç ç F ÷÷ i ë è 00øû ë i è è øøû 2 æ F 1 ö var(F ) = ç 0 ÷ var(F ) = i ç FF÷ i F 2 è i 0 ø i

If we assume that the main source of the uncertainty of the measured photon fluence rate F is random optical contact of the detector with the tissue, then the standard deviation of the measured value of F is proportional to F, with the coefficient of proportionality of a sa()FF= (13)

The variance var(yi) is expressed by the variance of the measured photon fluence rate var(F ) var(y ) ==i a 2 (14) i 2 F i

m The variance of the estimated effective attenuation coefficient var( eff)is

n å(rr- ) 22a i 0 a 2 var(m ) = i =1 = (15) eff 2 n n 2 æ 2 ö - ) - ) å(ri r0 ç å(rri 0 ÷ i =1 è i =1 ø

m The standard deviation of eff is then given by a sm( = eff ) (16) n - ) 2 å(rri 0 i =1

For a semi-infinitive medium the solution of the diffusion equation is given by

é --+m 22 -m ++22 + ù I exp() (zz ) r exp( ()zz0 2 z r) F(,rz )= 0 ×-ê eff 0 eff b ú (17) 4pD ê -+22 ++22 + ú ë ()zz0 r ()zz0 2 zb r û 6 Grzegorz Domański et al.

1 1+ R where z = , z is the distance to the extrapolated boundary, zD=2 eff , 0 m¢ b b - s 1 Reff n =---21 + + + = tissue Rnneff1440.... rel 0 710 rel 0668 00636 n rel , nrel is the relative refraction nair index of the tissue compared to that of air. m For this model the estimated value of parameter x = eff is found by minimizing the cost function E(x)

n =-2 Ex()å() f (,) rii x y (18) i =1 where æ F(rx,)ö fr(,)ln x= ç i ÷ (19) i ç F( ÷ è rx0 ,)ø dE The condition for the minimum is =0. dx Then, the value of the derivative is given by

n ¶ dE =-f å 2()fr(,)ii x y (20) dx i =1 ¶x

¶f F(rx,) 1 æ ¶F(rx,) ¶F(r ,)x ö =-0 ç i F(rx,) 0 F(rx,)÷ ¶ F( F(2 ¶ 0 ¶ i x rxi ,) rx0 ,)è x x ø

¶f 1 æ ¶F(rx,) ¶F(rx,) ö =-ç i F(rx,) 0 F(rx,)÷ ¶ F( F( ¶ 0 ¶ i xrxrxi ,)0 ,)è x x ø

¶f ¶F(rx,) 11¶F(rx,) =-i 0 (21) ¶ ¶ F( ¶ F( x xrxi ,) xrx0 ,)

The value of the second derivative is given by

2 dE2 n éæ ¶f ö ¶ 2 f ù = å 2êç ÷ +-()fr(,) x y ú (22) 2 ¶ ii¶ 2 dx i =1 ëêè x ø x ûú The Effective Attenuation Coefficient… 7

m The approximate value of x = eff obtained from (8) can be improved by iteration

æ dE ö ç ÷ è dx ø xx¢ =- (23) æ 2 ö ç dE÷ ç 2 ÷ è dx ø

Let replace function F from (17) by solving the diffusion equation for a semi-infinitive medium, written as

- - I éexp()xR exp()xR2 ù F(,rx )=×1 ê 1 - ú (24) p 4 D ë R1 R2 û

=-22 + =++22 + where Rzzr10(), Rzzzr20()2 b .

If R10 = R1(r0), R1i = R(ri), R20 = R2(r0), R2i = R2(ri), function f can be written as

é exp(-xR ) exp(-xR )ù ê 1i - 2i ú R R = ê 1i 2i ú fr(,)lni z (25) êexp(-xR ) exp(-xR )ú ê 10 - 20 ú ë R10 R20 û

If exp(-xR ) exp(-xR ) j = 1i - 2i i R1i R2i exp(-xR ) exp(-xR ) j = 10 - 20 0 R10 R20 then

æ j ö f = lnç i ÷ ç j ÷ è 0 ø 8 Grzegorz Domański et al.

¶j ¶j i j - 0 j ¶f j ¶ 0 ¶ i ¶j 11¶j = 0 xx=-i 0 (26) ¶ j j 2 ¶j ¶j x i 0 xxi 0

¶j i =-exp( -xR ) + exp( - xR ) (27) ¶x 1i 2i ¶j 0 =-exp( -xR ) + exp( - xR ) (28) ¶x 10 20

¶ 2 f é¶j2 11¶j æ ¶j öù é¶ 2j 11¶j æ ¶j öù =+-ê i i ç i ÷ú - ê 0 +-00ç ÷ú ¶ 2 ¶ 2 j ¶ ç j 2 ¶ ÷ ¶ 2 j ¶ ç j 2 ¶ ÷ xxë i xxè i øû ë x 0 xxè 0 øû

2 2 ¶ 2 f 11é¶j2 æ ¶j ö ù 1é¶j2 1 æ ¶j ö ù =-ê i ç i ÷ ú - ê 0 --ç 0 ÷ ú (29) ¶ 2 j ¶ 2 j ¶j¶ 2 j ¶ xxi ëê i è x ø ûú 0 ëê x 0 è x ø ûú

¶j2 i =---RxRRxRexp( ) exp( ) ¶x 2 11ii22ii

¶j2 i =---RxRRxRexp( ) exp( ) ¶x 2 10 10 20 20

The variance of x can be computed from

2 dE n ¶ var(x ) = å dx s 2 ()y (30) ¶ 2 i i =1 yi dE dx 2

dE ¶ 1 é ¶ æ dE ö dE2 ¶ æ dE2 ö dE ù dx = ç ÷ - ç ÷ 2 2 ê 2 2 ú ¶ 2 ¶ è ø ¶ yiidE æ dEö ë y dx dx yi è dx ø dx û 2 ç ÷ dx è dx 2 ø The Effective Attenuation Coefficient… 9

¶ dE df =-2 ¶ yi dx dx

¶ dE2 df2 =-2 ¶ 2 2 yi dx dx

2 4a¶2 n f dE2 dE df2 var(x ) = - (31) 4 å 2 2 2 ¶ æ ö i =1 x dx dx dx ç dE÷ è dx 2 ø

The above equations serve as a background theory for estimating the measuring error.

Materials and Methods

The surface light intensity distributions were measured in vivo for human forearms by means of a one-channel microprocessor system. The system consists of a light detector (silicon photodiode), a switched integrator and a microprocessor with a built-in 10-bit analogue-to-digital converter. The tissue under examination was coupled to a light detector by fiber optics. The device was controlled from a personal computer via a serial interface RS232. The light source was a LED-emitting light of the of about 660 nm. The block schematic of the microprocessor system for the measurements of optical parameters is shown in Figure 1.

Figure 1. A block schematic of the microprocessor system for measurements of optical parameters. 10 Grzegorz Domański et al.

Figure 2. Fixing the source on the tissue.

m Table 1. The results of eff coefficient in vivo determination for human forearms.

m ­1 Person Forearm eff [cm ]

A Left 2.03 ± 0.19

A Right 2.20 ± 0.19

B Left 2.52 ± 0.19

B Right 2.40 ± 0.19

C Left 2.25 ± 0.19

C Right 2.16 ± 0.19

D Left 2.59 ± 0.18

D Right 2.37 ± 0.19

E Right 2.05 ± 0.19 The Effective Attenuation Coefficient… 11

Figure 2 shows how the source was placed on the tissue examined. Both the source and the detector were placed on top of the forearm. The plastic cover served as a positioning support.

Results

m The optical parameter eff was measured in vivo on human forearms by means of our method. The persons studied were 25-56 year-old males. The surface light intensity m distributions were measured by means of our system, and then the eff coefficient estimations were computed using the theory described in (8). These values were used as m starting values for an iteration procedure given in (23). The final eff coefficient results for human forearms in vivo are shown in Table 1. The range of the estimated values of the effective attenuation coefficient was 2-3 cm­1. These results agree with those reported in literature [6, 7]. However, our values are probably slightly overestimated due to the inaccuracies in the applied model of light transport in tissue, given by equation (17). Using a more accurate model of light propagation, for example the Monte Carlo simulation, better values of the effective attenuation coefficient could be obtained.

Discussion and Conclusions

Our method is simple, fast and inexpensive. It allows finding the effective attenuation coefficient from light intensity measurements for at least two places on the tissue examined. A measurement theory was developed and tested with real data. The estimated values of the optical parameters can be used as starting values for an iteration procedure in the Monte Carlo simulation procedure.

References

[1] Arridge SR, Hebden JC. Optical imaging in Medicine: II. Modelling and reconstruction. Physics in Medicine and Biology 1997; 42: 841-853. [2] Frostig RD (editor). In Vivo Optical Imaging of Brain Function, CRC Press, Boca Raton, FL, 2002. 12 Grzegorz Domański et al.

[3] Tuchin VV (editor). Handbook of Optical Biomedical Diagnostics, SPIE Press, Bellingham, 2002. [4] Vo-Dinh T. Biomedical Photonics Handbook, CRC Press, Boca Raton, Fl, 2003. [5] Milstein AB, Oh S, Reynolds JS, Webb KJ, Bouman CA, Millane RP. Three-dimensional optical diffusion tomography with experimental data. Optics Letters 2002; 27(2) 95-97. [6] Kienle A, Glanzamann T. In vivo determination of the optical properties of muscle with time-resolved reflectance using a layered model. Phys Med Biol 1999; 44: 2689-2702. [7] Marquez G, Wang LV. White light oblique incidence reflectometer for measuring absorption and reduced scattering spectra of tissue-like turbid media. Optics Express 1997; 1: 454-460.