Allen K. Vickers Applications Engineer Manager December 6, 2000
PLOT Column Selection
(or How to Retain Your Gases)
Telephone Number: 816-650-0621 Chair Person: Lisa Lloyd WCOT vs. PLOT
Stationary Chromatographic Stationary Type Phase Process Phases
WCOT Liquid or Gas / Liquid Polysiloxanes gum partition PEG
PLOT Solid Gas solid Porous Polymers,
adsorption Al2O3, Zeolites, etc.
Dial 1-816-650-0621 for e-Seminar Audio Why Use PLOT Columns?
• Enhanced retention • Enhanced selectivity • Permit higher initial oven temperatures than WCOT
Dial 1-816-650-0621 for e-Seminar Audio PLOT Columns
"Solid" Porous Layer Fused Silica Tubing Ideal for the analysis of gases due to their increased retention (k) and unique selectivity (α) compared to WCOT
Dial 1-816-650-0621 for e-Seminar Audio WCOT Ethylene Analysis
Dial 1-816-650-0621 for e-Seminar Audio PLOT Ethylene Analysis
Dial 1-816-650-0621 for e-Seminar Audio Selectivity Interactions in PLOT Phases
Shape / Size Surface
Zeolites Al2O3 Porous Polymers
Bonded Carbon Bonded Silica Molecular Sieves
Dial 1-816-650-0621 for e-Seminar Audio J&W / Agilent PLOT Columns
• GS-Alumina
• HP-PLOT Al2O3 “M”
• HP-PLOT Al2O3 “S”
• HP-PLOT Al2O3 “KCl” • HP-PLOT MoleSieve • GS-CarbonPLOT • HP-PLOT Q • HP-PLOT U • GS-GasPro
Dial 1-816-650-0621 for e-Seminar Audio Al2O3 PLOT Columns
•GS-Alumina
• HP-PLOT Al2O3 “M”
• HP-PLOT Al2O3 “S”
• HP-PLOT Al2O3 “KCl”
•Different "polarities" due to different salt deactivation treatments
Dial 1-816-650-0621 for e-Seminar Audio Surface Interactions in PLOT Columns
+ + + + + + + + + + + + + neutral δ- δ+
Gas Flow δ- δ+
+ +
+ + + + + + + + + + +
• Vapor pressure always plays a leading role in solute interactions
Dial 1-816-650-0621 for e-Seminar Audio Dial 1-816-650-0621 for e-Seminar Audio Propylene Impurities
Dial 1-816-650-0621 for e-Seminar Audio Crude 1,3-Butadiene
Dial 1-816-650-0621 for e-Seminar Audio Extended HC Mixture
Dial 1-816-650-0621 for e-Seminar Audio Al2O3 PLOT Column Dimensions Available
• “M”, “S” and “KCl” • all available in 0.25, 0.32 and 0.53 mm ID • 15, 30 and 50 meter lengths (depending on ID)
• GS-Alumina • available in 0.53 mm ID • 15, 30 and 50 meter lengths
Dial 1-816-650-0621 for e-Seminar Audio 0.25 mm ID PLOT Al2O3 “S”
Dial 1-816-650-0621 for e-Seminar Audio Al2O3 Column Considerations
• Will adsorb water and CO2, affecting retention and resolution. • Recondition at 200°C to restore performance. • Not recommended for analysis of “polar” compounds such as alcohols, sulfurs, ketones, etc. • Upper temperature limit of 200ºC
Dial 1-816-650-0621 for e-Seminar Audio Break Number 1
For Questions and Answers Press *1 on Your Phone to Ask a Question
Dial 1-816-650-0621 for e-Seminar Audio Molecular Sieve PLOT Columns
Size Selectivity
)ORZ
Zeolite "Backbone"
*UDSKLFÃRIÃW\SHÃ.),ÃPROHFXODUÃVLHYH REWDLQHGÃIURPÃWKHÃ,QWHUQDWLRQDOÃ=HROLWH $VVRLFDWLRQÃZHEÃVLWHÃ ZZZL]RRQOLQHRUJ Dial 1-816-650-0621 for e-Seminar Audio Molecular Sieve PLOT Columns
• 5Å zeolite for the analysis of fixed and permanent gases:
H2, O2, N2, CO, CH4 He, Ne, Ar, Kr, Xe
SF6
Dial 1-816-650-0621 for e-Seminar Audio Permanent Gases
Dial 1-816-650-0621 for e-Seminar Audio Noble Gases
Dial 1-816-650-0621 for e-Seminar Audio Sulfur Hexafluoride
Dial 1-816-650-0621 for e-Seminar Audio HP-PLOT MoleSieve
• 0.32 and 0.53 mm ID • 15 and 30 meter lengths • Both thin and thick films available • Choose thin film for standard permanent gas analysis • Choose thick film for resolution of argon from oxygen at 35ºC • Upper temperature limit of 350ºC
Dial 1-816-650-0621 for e-Seminar Audio Molecular Sieve Column Considerations
• Very high retention stationary phase • Can be easily fouled by high MW hydrocarbons and polar compounds • Use pre-column or backflush if necessary
• Will adsorb water and CO2 • Columns can be reconditioned by holding at elevated temperature (350ºC) for several hours
Dial 1-816-650-0621 for e-Seminar Audio GS-CarbonPLOT
• Monolithic carbon layer • Unique selectivity • Phase formed in situ • No particle generation • Extended temperature limit of 360ºC • Not affected by water
Dial 1-816-650-0621 for e-Seminar Audio GS-CarbonPLOT Applications
• Light hydrocarbons (C1-C5) • Synthesis Gas
•H2, CO, CO2, CH4, N2/O2, H2O
• Trace CO, CO2 and acetylene in ethylene streams (simplification of ASTM D2505)
• Great resolution of air, CO, CO2 and C1-C2 hydrocarbon isomers • Light solvents
Dial 1-816-650-0621 for e-Seminar Audio C1 to C3 Hydrocarbons
GS-CarbonPLOT 1. Methane 30 m x 0.32 mm I.D., 1.5 µm 2. Acetylene 7 Carrier: He @ 30 cm/s 3. Ethylene Oven: 125°C for 2 min 125°C to 225°C at 20°/min 4. Ethane Injector: Split 1:20, 250°C 5. Propyne Detector: MSD, SIM 6. Propylene 7. Propane 2
4 3
5 6
1
0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50
Dial 1-816-650-0621 for e-Seminar Audio Natural Gas Mixture
GS-CarbonPLOT 1. Methane 30 m x 0.32 mm I.D., 1.5 µm 2. Acetylene Carrier: He @ 30 cm/s 3. Propane Oven: 125°C for 2 min 5 4. iso-Butane 125°C to 350°C at 20°/min 5. n-Butane Injector: Split 1:20, 250°C 3 Detector: MSD, SIM 6. iso-Pentane 7. n-Pentane
4
1 2
7
6
0 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00
Dial 1-816-650-0621 for e-Seminar Audio Permanent Gases Column: GS-CarbonPLOT 30 m x 0.32 mm I.D. x 3.0 µm J&W P/N: 113-3133 Carrier: Helium at 27 cm/sec (calculated off N2 at 35°C) Oven: 35°C isothermal Injector: 185°C, 1:30 split, 250 µL injection volume Detector: TCD, 150°C 1 1. Nitrogen and Oxygen (5% each) 2. Carbon monoxide (5%) 2 3. Methane (4%) 4. Carbon dioxide (5%)
3
4
5 2. 0 2. 5 3. 0 3. 5 4. 0 & Dial 1-816-650-0621 for e-Seminar Audio Solvents 1. Methylene Chloride GS-CarbonPLOT 2. Acetone 30 m x 0.32 mm I.D., 1.5 µm 3. Carbon Disulfide Carrier: He @ 30 cm/s 3 Oven: 125°C for 2 min 4. Ethyl Ether 125°C to 350°C at 20°/min 5. Ethyl Acetate Injector: Split 1:20, 250°C 6. Benzene Detector: MSD, SIM
6
5
1
2
4
6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00
Dial 1-816-650-0621 for e-Seminar Audio GS-CarbonPLOT
• 0.32 and 0.53 mm ID • 15, 30 and 60 meters • 0.32 mm ID available in thin and thick film • Upper temperature limit of 350º/360ºC
Dial 1-816-650-0621 for e-Seminar Audio Break Number 2
For Questions and Answers Press *1 on Your Phone to Ask a Question
Dial 1-816-650-0621 for e-Seminar Audio Porous Polymer Phases
• Retention is a combination of size and surface interactions between the analyte and the porous solid phase. • The polymers are generally homopolymers of divinylbenzene and divinyl-styrene.
• N, R, Q, T, U phases
Dial 1-816-650-0621 for e-Seminar Audio Agilent Porous Polymer PLOT Columns
• HP PLOT-Q • Bonded polystyrene-divinylbenzene • HP PLOT-U • Bonded divinylbenzene/ethylene glycol dimethacrylate • More polar than “Q”
Dial 1-816-650-0621 for e-Seminar Audio Dial 1-816-650-0621 for e-Seminar Audio Natural Gas on HP-PLOT Q
Dial 1-816-650-0621 for e-Seminar Audio HP-PLOT Q Volatile Solvents
Dial 1-816-650-0621 for e-Seminar Audio HP-PLOT Q Consecutive runs for up to 200 headspace injections of alcohol mixture. Excellent stability!!
Dial 1-816-650-0621 for e-Seminar Audio HP-PLOT Q After 150 consecutive direct water injections
Dial 1-816-650-0621 for e-Seminar Audio HP-PLOT U Volatile Solvents
Dial 1-816-650-0621 for e-Seminar Audio HP-PLOT U Impurities in Acetic Acid
Dial 1-816-650-0621 for e-Seminar Audio HP-PLOT U Volatile Primary Amines
Dial 1-816-650-0621 for e-Seminar Audio Porous Polymer Column Dimensions Available
• Both HP-PLOT Q and HP-PLOT U: • 0.32 and 0.53 mm ID • 15 and 30 meter lengths
Dial 1-816-650-0621 for e-Seminar Audio GS-GasPro
• Silica-based PLOT column • Bonded stationary phase • Unique selectivity • Highly inert to reactive compounds • Retention stability unaffected by water
Dial 1-816-650-0621 for e-Seminar Audio GS-GasPro
• Upper temperature limit of 260º/300ºC • No particle generation • Applications
•CO, CO2
•C1 to C12 hydrocarbons • trace sulfur analysis • hydrides (arsine, phosphine, germane, etc.) • halocarbons
• HCl, NH3
Dial 1-816-650-0621 for e-Seminar Audio Dial 1-816-650-0621 for e-Seminar Audio Dial 1-816-650-0621 for e-Seminar Audio Dial 1-816-650-0621 for e-Seminar Audio Dial 1-816-650-0621 for e-Seminar Audio Dial 1-816-650-0621 for e-Seminar Audio Dial 1-816-650-0621 for e-Seminar Audio Dial 1-816-650-0621 for e-Seminar Audio GS-GasPro Dimensions Available
• Currently only available in 0.32 mm ID • 5, 15, 30 and 60 meter lengths
Dial 1-816-650-0621 for e-Seminar Audio Considerations for PLOT Column Analysis
• Inlet issues • split versus direct injection • gas sampling valves • low dead volume • column ID and flow rate
• Detector issues • particle generation or “spiking”; particle traps • column ID and flow rate
Dial 1-816-650-0621 for e-Seminar Audio Considerations for PLOT Column Analysis
• Column issues • selectivity • capacity; overloaded peaks • inertness • temperature limits
• Column contamination • efficiency loss; “ghost peaks”; increase in bleed
• water, CO2, high molecular weight hydrocarbons? •Carrier gas purifiers
Dial 1-816-650-0621 for e-Seminar Audio Wrap-up E-Seminar Questions