Basics of Mass Spectroscopy the Roots of Mass Spectroscopy (MS)

Total Page:16

File Type:pdf, Size:1020Kb

Basics of Mass Spectroscopy the Roots of Mass Spectroscopy (MS) Spectroscopy Beauchamp 1 Basics of Mass Spectroscopy The roots of mass spectroscopy (MS) trace back to the early part of the 20th century. In 1911 J.J. Thomson used a primitive form of MS to prove the existence of isotopes with neon-20 and neon- 22. Current, easy-to-use, table-top instruments of today are a very recent luxury. In less than a day, you could be running samples on a mass spectrometer. However, it would take you longer to learn the many intricacies of MS, something we cannot pursue in a book such as this. We will mainly look at electron impact mass spectrometry (EI) and briefly mention chemical ionization (CI) as they pertain to determining an organic structure. The technique of MS only requires very small amounts of sample (g-ng) for high quality data. For that reason, it is the preferred method to evaluate product structures in combinatorial chemistry, forensic laboratories and with complicated biological samples. Generally, in these situations, you have some indication of the structure(s) possible. MS can be coupled to separation techniques such as gas chromatography (GC) and high pressure liquid chromatography (HPLC) to make a combination technique (GC-MS and LC-MS). GC can separate components in relatively volatile mixtures and HPLC can separate components in relatively less volatile mixtures. There are also options for direct inlet of solid samples and sampling methods for high molecular weight biomolecules and polymers. But, these are beyond the scope of this book. MS is different from the other spectroscopies (UV-Vis, IR, NMR) in that absorption or emission of electromagnetic radiation is not used. Rather, the sample (molecule) is ionized by some method (often a high energy electron beam = electron impact = EI). An electron is knocked out of a bonding molecular orbital (MO), forming a radical cation. Dications and anions can also be formed, but we will not consider these possibilities. high EI mass spec RH+ energy RH 2 e- e- + radical cation The cations formed are accelerated in a high voltage field, focused and separated by mass to charge ratio (m/z or m/e) using a magnetic and/or electric fields. A detector indicates the intensity of each mass signal and the mass data (x axis) are plotted against this intensity (y axis) to produce a spectrum similar to that shown below. It is also possible that this same data can be printed in a tabulated, numerical form (shown in the side box). The most useful information from the MS is the molecular weight (the M+ peak), which can indicate what the formula is. The formula provides the degree of unsaturation, which gives important clues to the possible structures (rings and pi bonds). Fragment peaks that are detected provide hints as to the nature of the carbon skeleton, heteroatoms and functional groups present. The most abundant peak (largest) in the mass spectrum is called the base peak. It is assigned a value of 100% and all other detectable masses are indicated as a percent of the base peak. The molecular weight peak is called the mass peak or molecular ion peak or parent peak and symbolized with an M. Since this peak is a radical cation, it often also has a + or + . (plus sign and a dot) superscript as well. We will use M+. There is often ambiguity in the other fragment peaks because of high energy rearrangements that are possible. It is usually very difficult to assign a structure to a completely unknown molecule based solely on mass spectroscopy. But a mass spectrum can help provide a very important piece of the puzzle, the molecular weight. y:\files\classes\Spectroscopy Book home\1 Spectroscopy Workbook, latest MS full chapter.doc Spectroscopy Beauchamp 2 base peak = largest peak in MS spectrum = 100% peak, other peaks are reported as a percent of this peak molecular ion = M = M+ = M+ = parent peak Only specific isotopic masses are found in the molecular formula. We do not see “average” masses that are listed in the periodic table. Also present will be M+1, M+2, etc. peaks due to other isotopes. On low resolution MS these peaks can help decide what the molecular formula is. In the MS example below, some of the peaks are very ‘logical’ (57, 85 and 91 are logical) and some are less so (39, 41, 42, 51 and 55). It is also true that peaks that are ‘logical’ are sometimes small or completely missing (119). Many of the other peaks will be explainable with certain assumptions about fragmentations discussed later in this chapter. Tabulated Data 91 57 85 119 Mass percent O 1-phenyl-2-hexanone 27 6 C12H16O , MW = 176 28 2 29 24 57 85 = base peak 39 7 100 41 26 Many smaller 42 1 peaks are not 43 1 shown, but listed 50 1 in data table to 51 3 75 the left. 55 3 57 99 58 5 percent 60 1 relative 50 63 3 intensity 65 11 91 77 2 85 100 (base) 29 41 86 6 25 M+ 89 2 65 peak 90 2 27 176 91 36 39 58 86 92 92 6 0 176 7 = M+ 177 1 0 25 50 75 100 125 150 175 200 mass m = charge e Spectroscopy Beauchamp 3 Typical MS Instrument Features. The moving charged cations (R-H+) can be made to curve in their direction of flight in a magnetic or electric field. The amount of curvature is determined by the mass (m) of the ions as shown in the following equations (assuming the charge, e, is constant = +1). The magnetic field (B) and/or accelerator plate voltage (V) can be altered to cause each possible mass to impact the detector. The charged masses must survive about 10-6 to 10-5 seconds to make this journey to the detector. Often there is some rational feature to explain each peak’s special stability that allows it to last long enough to reach the detector, where it becomes part of the data we examine. We will look at some of these features later in this discussion. We will not discuss other possibilities, such as metastable ions or +2 and negatively charged ions. Our main goal in this book is interpretation. 2 2 m = B r m = mass e 2V e = charge (usually +1) B = size of magnetic field r = radius of curvature V = voltage on accelerator plate r mV 1 = e B Besides just seeing a positively charged mass at the detector, we must resolve it from nearby mass values. MS instruments can be either low resolution (LRMS) or high resolution (HRMS). Low resolution MS instruments can generally resolve single amu values as high as about 2000 amu’s (e.g. they can distinguish 300 amu from 301 amu). An atomic mass unit is defined as 1/12 the mass of a neutral carbon-12 atom (12C = 12.0000, by definition). High resolution MS instruments can resolve masses as close as the fourth decimal place (XXX.XXXX). With such accuracy, an exact molecular formula can be determined by a computer. A molecular formula can also be obtained from LRMS, y:\files\classes\Spectroscopy Book home\1 Spectroscopy Workbook, latest MS full chapter.doc Spectroscopy Beauchamp 4 through a slightly more involved procedure. HRMS instruments tend to be more expensive and less common. Exact Masses We need to be precise in our calculation of possible masses for each collection of atoms because the atoms in any cation hitting the detector are specific isotopes. The atomic weights listed in the periodic table are average weights based on the abundance and mass of all of the naturally occurring isotopes of each element. For example, the atomic weight of bromine in the periodic table is 79.9, even though there is no bromine isotope with a mass of 80. The 79.9 atomic weight is a result of an approximate 50/50 mixture of two stable isotopes of mass 78.9 and 80.9. Because of this complication, we will require data on the exact masses and the relative abundance of the common isotopes that we expect to encounter. Those most useful to us in organic chemistry and biochemistry are listed below. Average Element Atomic Weight Nuclides Exact Mass Relative Abundance* hydrogen 1.00797 1H 1.00783 100.0 2 H (D) 2.01410 0.015 carbon 12.01115 12C 12.00000 100.0 13 C 13.00336 1.11 nitrogen 14.0067 14N 14.00307 100.0 15 N 15.00011 0.37 oxygen 15.9994 16O 15.9949 100.0 17 O 16.9991 0.04 18 O 17.9992 0.20 fluorine 18.9984 19F 18.9984 100.0 silicon 28.086 28Si 27.9769 100.0 29 Si 28.9765 5.06 30 Si 29.9738 3.36 phosphorous 30.974 31P 30.9738 100.0 sulfur 32.064 32S 31.9721 100.0 33 S 32.9715 0.79 34 S 33.9679 4.43 chlorine 35,453 35Cl 34.9689 100.0 37 Cl 36.9659 31.98 bromine 79.909 79Br 78.9183 100.0 81 Br 80.9163 97.3 iodine 126.904 127I 126.9045 100.0 *The most abundant nuclide is assigned 100% and the others assigned a fractional percent of that value. Coincidently, in the examples listed in the table above with more than one isotope, the lowest mass isotope is the 100% isotope. Spectroscopy Beauchamp 5 Obtaining a molecular formula from a HRMS is relatively straight forward Each possible molecular mass is unique when calculated to 3-4 decimal places and computers can do the calculations for us.
Recommended publications
  • An Introduction to Isotopic Calculations John M
    An Introduction to Isotopic Calculations John M. Hayes ([email protected]) Woods Hole Oceanographic Institution, Woods Hole, MA 02543, USA, 30 September 2004 Abstract. These notes provide an introduction to: termed isotope effects. As a result of such effects, the • Methods for the expression of isotopic abundances, natural abundances of the stable isotopes of practically • Isotopic mass balances, and all elements involved in low-temperature geochemical • Isotope effects and their consequences in open and (< 200°C) and biological processes are not precisely con- closed systems. stant. Taking carbon as an example, the range of interest is roughly 0.00998 ≤ 13F ≤ 0.01121. Within that range, Notation. Absolute abundances of isotopes are com- differences as small as 0.00001 can provide information monly reported in terms of atom percent. For example, about the source of the carbon and about processes in 13 13 12 13 atom percent C = [ C/( C + C)]100 (1) which the carbon has participated. A closely related term is the fractional abundance The delta notation. Because the interesting isotopic 13 13 fractional abundance of C ≡ F differences between natural samples usually occur at and 13F = 13C/(12C + 13C) (2) beyond the third significant figure of the isotope ratio, it has become conventional to express isotopic abundances These variables deserve attention because they provide using a differential notation. To provide a concrete the only basis for perfectly accurate mass balances. example, it is far easier to say – and to remember – that Isotope ratios are also measures of the absolute abun- the isotope ratios of samples A and B differ by one part dance of isotopes; they are usually arranged so that the per thousand than to say that sample A has 0.3663 %15N more abundant isotope appears in the denominator and sample B has 0.3659 %15N.
    [Show full text]
  • Mass Spectrometry: Quadrupole Mass Filter
    Advanced Lab, Jan. 2008 Mass Spectrometry: Quadrupole Mass Filter The mass spectrometer is essentially an instrument which can be used to measure the mass, or more correctly the mass/charge ratio, of ionized atoms or other electrically charged particles. Mass spectrometers are now used in physics, geology, chemistry, biology and medicine to determine compositions, to measure isotopic ratios, for detecting leaks in vacuum systems, and in homeland security. Mass Spectrometer Designs The first mass spectrographs were invented almost 100 years ago, by A.J. Dempster, F.W. Aston and others, and have therefore been in continuous development over a very long period. However the principle of using electric and magnetic fields to accelerate and establish the trajectories of ions inside the spectrometer according to their mass/charge ratio is common to all the different designs. The following description of Dempster’s original mass spectrograph is a simple illustration of these physical principles: The magnetic sector spectrograph PUMP F DD S S3 1 r S2 Fig. 1: Dempster’s Mass Spectrograph (1918). Atoms/molecules are first ionized by electrons emitted from the hot filament (F) and then accelerated towards the entrance slit (S1). The ions then follow a semicircular trajectory established by the Lorentz force in a uniform magnetic field. The radius of the trajectory, r, is defined by three slits (S1, S2, and S3). Ions with this selected trajectory are then detected by the detector D. How the magnetic sector mass spectrograph works: Equating the Lorentz force with the centripetal force gives: qvB = mv2/r (1) where q is the charge on the ion (usually +e), B the magnetic field, m is the mass of the ion and r the radius of the ion trajectory.
    [Show full text]
  • Good Practice Guide for Isotope Ratio Mass Spectrometry, FIRMS (2011)
    Good Practice Guide for Isotope Ratio Mass Spectrometry Good Practice Guide for Isotope Ratio Mass Spectrometry First Edition 2011 Editors Dr Jim Carter, UK Vicki Barwick, UK Contributors Dr Jim Carter, UK Dr Claire Lock, UK Acknowledgements Prof Wolfram Meier-Augenstein, UK This Guide has been produced by Dr Helen Kemp, UK members of the Steering Group of the Forensic Isotope Ratio Mass Dr Sabine Schneiders, Germany Spectrometry (FIRMS) Network. Dr Libby Stern, USA Acknowledgement of an individual does not indicate their agreement with Dr Gerard van der Peijl, Netherlands this Guide in its entirety. Production of this Guide was funded in part by the UK National Measurement System. This publication should be cited as: First edition 2011 J. F. Carter and V. J. Barwick (Eds), Good practice guide for isotope ratio mass spectrometry, FIRMS (2011). ISBN 978-0-948926-31-0 ISBN 978-0-948926-31-0 Copyright © 2011 Copyright of this document is vested in the members of the FIRMS Network. IRMS Guide 1st Ed. 2011 Preface A few decades ago, mass spectrometry (by which I mean organic MS) was considered a “black art”. Its complex and highly expensive instruments were maintained and operated by a few dedicated technicians and its output understood by only a few academics. Despite, or because, of this the data produced were amongst the “gold standard” of analytical science. In recent years a revolution occurred and MS became an affordable, easy to use and routine technique in many laboratories. Although many (rightly) applaud this popularisation, as a consequence the “black art” has been replaced by a “black box”: SAMPLES GO IN → → RESULTS COME OUT The user often has little comprehension of what goes on “under the hood” and, when “things go wrong”, the inexperienced operator can be unaware of why (or even that) the results that come out do not reflect the sample that goes in.
    [Show full text]
  • Comparison of the Characteristic Mass Fragmentations of Phenethylamines and Tryptamines by Electron Ionization Gas Chromatograph
    applied sciences Article Comparison of the Characteristic Mass Fragmentations of Phenethylamines and Tryptamines by Electron Ionization Gas Chromatography Mass Spectrometry, Electrospray and Matrix-Assisted Laser Desorption Ionization Mass Spectrometry Bo-Hong Chen, Ju-Tsung Liu, Hung-Ming Chen, Wen-Xiong Chen and Cheng-Huang Lin * Department of Chemistry, National Taiwan Normal University, 88 Sec. 4 Tingchow Road, Taipei 11677, Taiwan; [email protected] (B.-H.C.); [email protected] (J.-T.L.); [email protected] (H.-M.C.); [email protected] (W.-X.C.) * Correspondence: [email protected]; Tel.: +886-2-7734-6170; Fax: +886-2-2932-4249 Received: 18 April 2018; Accepted: 19 June 2018; Published: 22 June 2018 Abstract: Characteristic mass fragmentation of 20 phenethylamine/tryptamine standards were investigated and compared by means of matrix assisted laser desorption/time-of-flight mass spectrometry (MALDI/TOFM), gas chromatography–electron ionization–mass spectrometry (GC-EI/MS) and liquid chromatography–electrospray ionization/mass spectrometry (LC-ESI/MS) + methods. As a result, three characteristic peaks ([M] and fragments from the Cβ-Cα bond breakage) were found to be unique and contained information useful in identifying 2C series compounds based on the GC-EI/MS method. We found that the protonated molecular ion ([M+H]+) and two types of fragments produced from the α-cleavage and β-cleavage processes were useful mass spectral information in the rapid screening and confirmation of phenethylamine and tryptamine derivatives when ESI/MS and MALDI/TOFMS methods were applied. This assay was successfully used to determine samples that contain illicit drugs. Keywords: phenethylamine; tryptamine; MALDI/TOFMS; GC-EI/MS; LC-ESI/MS 1.
    [Show full text]
  • Coupling Gas Chromatography to Mass Spectrometry
    Coupling Gas Chromatography to Mass Spectrometry Introduction The suite of gas chromatographic detectors includes (roughly in order from most common to the least): the flame ionization detector (FID), thermal conductivity detector (TCD or hot wire detector), electron capture detector (ECD), photoionization detector (PID), flame photometric detector (FPD), thermionic detector, and a few more unusual or VERY expensive choices like the atomic emission detector (AED) and the ozone- or fluorine-induce chemiluminescence detectors. All of these except the AED produce an electrical signal that varies with the amount of analyte exiting the chromatographic column. The AED does that AND yields the emission spectrum of selected elements in the analytes as well. Another GC detector that is also very expensive but very powerful is a scaled down version of the mass spectrometer. When coupled to a GC the detection system itself is often referred to as the mass selective detector or more simply the mass detector. This powerful analytical technique belongs to the class of hyphenated analytical instrumentation (since each part had a different beginning and can exist independently) and is called gas chromatograhy/mass spectrometry (GC/MS). Placed at the end of a capillary column in a manner similar to the other GC detectors, the mass detector is more complicated than, for instance, the FID because of the mass spectrometer's complex requirements for the process of creation, separation, and detection of gas phase ions. A capillary column is required in the chromatograph because the entire MS process must be carried out at very low pressures (~10-5 torr) and in order to meet this requirement a vacuum is maintained via constant pumping using a vacuum pump.
    [Show full text]
  • Electron Ionization
    Chapter 6 Chapter 6 Electron Ionization I. Introduction ......................................................................................................317 II. Ionization Process............................................................................................317 III. Strategy for Data Interpretation......................................................................321 1. Assumptions 2. The Ionization Process IV. Types of Fragmentation Pathways.................................................................328 1. Sigma-Bond Cleavage 2. Homolytic or Radical-Site-Driven Cleavage 3. Heterolytic or Charge-Site-Driven Cleavage 4. Rearrangements A. Hydrogen-Shift Rearrangements B. Hydride-Shift Rearrangements V. Representative Fragmentations (Spectra) of Classes of Compounds.......... 344 1. Hydrocarbons A. Saturated Hydrocarbons 1) Straight-Chain Hydrocarbons 2) Branched Hydrocarbons 3) Cyclic Hydrocarbons B. Unsaturated C. Aromatic 2. Alkyl Halides 3. Oxygen-Containing Compounds A. Aliphatic Alcohols B. Aliphatic Ethers C. Aromatic Alcohols D. Cyclic Ethers E. Ketones and Aldehydes F. Aliphatic Acids and Esters G. Aromatic Acids and Esters 4. Nitrogen-Containing Compounds A. Aliphatic Amines B. Aromatic Compounds Containing Atoms of Nitrogen C. Heterocyclic Nitrogen-Containing Compounds D. Nitro Compounds E. Concluding Remarks on the Mass Spectra of Nitrogen-Containing Compounds 5. Multiple Heteroatoms or Heteroatoms and a Double Bond 6. Trimethylsilyl Derivative 7. Determining the Location of Double Bonds VI. Library
    [Show full text]
  • Modern Mass Spectrometry
    Modern Mass Spectrometry MacMillan Group Meeting 2005 Sandra Lee Key References: E. Uggerud, S. Petrie, D. K. Bohme, F. Turecek, D. Schröder, H. Schwarz, D. Plattner, T. Wyttenbach, M. T. Bowers, P. B. Armentrout, S. A. Truger, T. Junker, G. Suizdak, Mark Brönstrup. Topics in Current Chemistry: Modern Mass Spectroscopy, pp. 1-302, 225. Springer-Verlag, Berlin, 2003. Current Topics in Organic Chemistry 2003, 15, 1503-1624 1 The Basics of Mass Spectroscopy ! Purpose Mass spectrometers use the difference in mass-to-charge ratio (m/z) of ionized atoms or molecules to separate them. Therefore, mass spectroscopy allows quantitation of atoms or molecules and provides structural information by the identification of distinctive fragmentation patterns. The general operation of a mass spectrometer is: "1. " create gas-phase ions "2. " separate the ions in space or time based on their mass-to-charge ratio "3. " measure the quantity of ions of each mass-to-charge ratio Ionization sources ! Instrumentation Chemical Ionisation (CI) Atmospheric Pressure CI!(APCI) Electron Impact!(EI) Electrospray Ionization!(ESI) SORTING DETECTION IONIZATION OF IONS OF IONS Fast Atom Bombardment (FAB) Field Desorption/Field Ionisation (FD/FI) Matrix Assisted Laser Desorption gaseous mass ion Ionisation!(MALDI) ion source analyzer transducer Thermospray Ionisation (TI) Analyzers quadrupoles vacuum signal Time-of-Flight (TOF) pump processor magnetic sectors 10-5– 10-8 torr Fourier transform and quadrupole ion traps inlet Detectors mass electron multiplier spectrum Faraday cup Ionization Sources: Classical Methods ! Electron Impact Ionization A beam of electrons passes through a gas-phase sample and collides with neutral analyte molcules (M) to produce a positively charged ion or a fragment ion.
    [Show full text]
  • Find the Molar Mass of Sodium Carbonate, Na 2CO3. Na 2 X
    Moles and Molar Mass The mole is the "counting unit" used by chemists to indicate the number of atoms, ions, molecules, or formula units present in a particular chemical sample. The mole is similar to other counting units that you've used before....pair (2), dozen (12), and gross (144). One mole of a compound contains Avogadro's number (6.022 x 1023) of molecules (molecular compound) or formula units (ionic compound). The molar mass of a compound tells you the mass of 1 mole of that substance. In other words, it tells you the number of grams per mole of a compound. The units for molar mass are, therefore, grams/mole. To find the molar mass of a compound: 1. Use the chemical formula to determine the number of each type of atom present in the compound. 2. Multiply the atomic weight (from the periodic table) of each element by the number of atoms of that element present in the compound. 3. Add it all together and put units of grams/mole after the number. Example: Find the molar mass of sodium carbonate, Na2CO3. Na 2 x 23.0 = 46.0 C 1 x 12.0 = 12.0 O 3 x 16.0 = 48.0 molar = 106.0 g/mole mass For many (but not all) problems, you can simply round the atomic weights and the molar mass to the nearest 0.1 g/mole. HOWEVER, make sure that you use at least as many significant figures in your molar mass as the measurement with the fewest significant figures.
    [Show full text]
  • Fundamentals of Biological Mass Spectrometry and Proteomics
    Fundamentals of Biological Mass Spectrometry and Proteomics Steve Carr Broad Institute of MIT and Harvard Modern Mass Spectrometer (MS) Systems Orbitrap Q-Exactive Triple Quadrupole Discovery/Global Experiments Targeted MS MS systems used for proteomics have 4 tasks: • Create ions from analyte molecules • Separate the ions based on charge and mass • Detect ions and determine their mass-to-charge • Select and fragment ions of interest to provide structural information (MS/MS) Electrospray MS: ease of coupling to liquid-based separation methods has made it the key technology in proteomics Possible Sample Inlets Syringe Pump Sample Injection Loop Liquid Autosampler, HPLC Capillary Electrophoresis Expansion of the Ion Formation and Sampling Regions Nitrogen Drying Gas Electrospray Atmosphere Vacuum Needle 3- 5 kV Liquid Nebulizing Gas Droplets Ions Containing Solvated Ions Isotopes Most elements have more than one stable isotope. For example, most carbon atoms have a mass of 12 Da, but in nature, 1.1% of C atoms have an extra neutron, making their mass 13 Da. Why do we care? Mass spectrometers “see” the isotope peaks provided the resolution is high enough. If an MS instrument has resolution high enough to resolve these isotopes, better mass accuracy is achieved. Stable isotopes of most abundant elements of peptides Element Mass Abundance H 1.0078 99.985% 2.0141 0.015 C 12.0000 98.89 13.0034 1.11 N 14.0031 99.64 15.0001 0.36 O 15.9949 99.76 16.9991 0.04 17.9992 0.20 Monoisotopic mass and isotopes We use instruments that resolve the isotopes enabling us to accurately measure the monoisotopic mass MonoisotopicMonoisotopic mass; all 12C, mass no 13C atoms corresponds to 13 lowestOne massC atom peak Two 13C atoms Angiotensin I (MW = 1295.6) (M+H)+ = C62 H90 N17 O14 TheWhen monoisotopic the isotopes mass of aare molecule clearly is the resolved sum of the the accurate monoisotopic masses for the massmost abundant isotope of each element present.
    [Show full text]
  • Mass Spectrometry (Technically Not Spectroscopy)
    Mass Spectrometry (technically not Spectroscopy) So far, In mass spec, on y Populati Intensit Excitation Energy “mass” (or mass/charge ratio) Spectroscopy is about Mass spectrometry interaction of energy with matter. measures population of ions X-axis is real. with particular mass. General Characteristics of Mass Spectrometry 2. Ionization Different variants of 1-4 -e- available commercially. 4. Ion detection 1. Introduction of sample to gas phase (sometimes w/ separation) 3. Selection of one ion mass (Selection nearly always based on different flight of ion though vacuum.) General Components of a Mass Spectrometer Lots of choices, which can be mixed and matched. direct injection The Mass Spectrum fragment “daughter” ions M+ “parent” mass Sample Introduction: Direc t Inser tion Prob e If sample is a liquid, sample can also be injected directly into ionization region. If sample isn’t pure, get multiple parents (that can’t be distinguished from fragments). Capillary Column Introduction Continous source of molecules to spectrometer. detector column (including GC, LC, chiral, size exclusion) • Signal intensity depends on both amount of molecule and ionization efficiency • To use quantitatively, must calibrate peaks with respect eltilution time ttlitotal ion curren t to quantity eluted (TIC) over time Capillary Column Introduction Easy to interface with gas or liquid chromatography. TIC trace elution time time averaged time averaged mass spectrum mass spectrum Methods of Ionization: Electron Ionization (EI) 1 - + - 1 M + e (kV energy) M + 2e Fragmentation in Electron Ionization daughter ion (observed in spectrum) neutral fragment (not observed) excited parent at electron at electron energy of energy of 15 eV 70 e V Lower electron energy yields less fragmentation, but also less signal.
    [Show full text]
  • An Introduction to Mass Spectrometry
    An Introduction to Mass Spectrometry by Scott E. Van Bramer Widener University Department of Chemistry One University Place Chester, PA 19013 [email protected] http://science.widener.edu/~svanbram revised: September 2, 1998 © Copyright 1997 TABLE OF CONTENTS INTRODUCTION ........................................................... 4 SAMPLE INTRODUCTION ....................................................5 Direct Vapor Inlet .......................................................5 Gas Chromatography.....................................................5 Liquid Chromatography...................................................6 Direct Insertion Probe ....................................................6 Direct Ionization of Sample ................................................6 IONIZATION TECHNIQUES...................................................6 Electron Ionization .......................................................7 Chemical Ionization ..................................................... 9 Fast Atom Bombardment and Secondary Ion Mass Spectrometry .................10 Atmospheric Pressure Ionization and Electrospray Ionization ....................11 Matrix Assisted Laser Desorption/Ionization ................................ 13 Other Ionization Methods ................................................13 Self-Test #1 ...........................................................14 MASS ANALYZERS .........................................................14 Quadrupole ............................................................15
    [Show full text]
  • An Organic Chemist's Guide to Electrospray Mass Spectrometric
    molecules Review An Organic Chemist’s Guide to Electrospray Mass Spectrometric Structure Elucidation Arnold Steckel 1 and Gitta Schlosser 2,* 1 Hevesy György PhD School of Chemistry, ELTE Eötvös Loránd University, Pázmány Péter sétány 1/A, 1117 Budapest, Hungary; [email protected] 2 Department of Analytical Chemistry, ELTE Eötvös Loránd University, Pázmány Péter sétány 1/A, 1117 Budapest, Hungary * Correspondence: [email protected] Received: 16 January 2019; Accepted: 8 February 2019; Published: 10 February 2019 Abstract: Tandem mass spectrometry is an important tool for structure elucidation of natural and synthetic organic products. Fragmentation of odd electron ions (OE+) generated by electron ionization (EI) was extensively studied in the last few decades, however there are only a few systematic reviews available concerning the fragmentation of even-electron ions (EE+/EE−) produced by the currently most common ionization techniques, electrospray ionization (ESI) and atmospheric pressure chemical ionization (APCI). This review summarizes the most important features of tandem mass spectra generated by collision-induced dissociation fragmentation and presents didactic examples for the unexperienced users. Keywords: tandem mass spectrometry; MS/MS fragmentation; collision-induced dissociation; CID; ESI; structure elucidation 1. Introduction Electron ionization (EI), a hard ionization technique, is the method of choice for analyses of small (<1000 Da), nonpolar, volatile compounds. As its name implies, the technique involves ionization by electrons with ~70 eV energy. This energy is high enough to yield very reproducible mass spectra with a large number of fragments. However, these spectra frequently lack the radical type molecular ions (M+) due to the high internal energy transferred to the precursors [1].
    [Show full text]