AAS, GFAAS, ICP Or ICP-MS? Which Technique Should I Use? an Elementary Overview of Elemental Analysis
Total Page:16
File Type:pdf, Size:1020Kb
1069 primer reprint 11/29/01 2:41 PM Page 3 AAS, GFAAS, ICP or ICP-MS? Which technique should I use? An elementary overview of elemental analysis. 1069 primer reprint 11/29/01 2:41 PM Page 4 Start here. If your laboratory is performing elemental analysis on solutions, there are a number of mature, effective spectrometric techniques you could use, including flame and furnace AA (Atomic Absorption), ICP (Inductively Coupled Plasma), and ICP-MS (ICP Mass Spectrometry). Selecting the most appropriate tool for the job can sometimes appear to be a daunting task, especially since there is considerable overlap of capabilities. In fact, all of the techniques may be able to perform your particular analysis at acceptable levels of accuracy and precision. However, by evaluating the relative strengths and weaknesses of each of the techniques, particularly as they may apply to the practical and performance requirements of your analysis, it is possible to establish a rational basis for decision making. In other words, it needn’t be all that difficult. Even if you’re new to elemental analysis, this booklet will get you well on your way to selecting the best technique (or techniques) to meet your lab’s needs. Understanding how each technique works. While it isn’t necessary to grasp all of the complex technological details of each of the various elemental analysis instruments, for comparison purposes it is helpful to have a general understanding of basic principles. Let’s start at the beginning, with atomic absorption. 2 1069 primer reprint 11/29/01 2:42 PM Page 5 Percent Relative Standard Deviation (% RSD) Short-term stability The standard deviation expressed as a percentage of the total. The ability of the instrument to read back the value you specified as the value of the standard after one hour’s drift, ppb usually around 2%. An abbreviation for parts per billion. 1 part in 1,000,000,000 or one microgram per liter. Spectrometer Meter = to measure. Spectro = Spectrum. For optical spec- ppm trometers this is a series of colored bands dispersed from An abbreviation for parts per million. 1 part in 1,000,000 white light by means of a refracting material (prism) or by or one milligram per liter. means of a diffractive device (grating). Detectors can be placed at exact locations in the spectrum to measure the ppt presence of the desired element. An abbreviation for parts per trillion. 1 part in 1,000,000,000,000 or one nanogram per liter. A mass spectrometer spectrum is a series of signals registering Plasma on the detector with increasing mass-to-charge ratio, each A very high-temperature ionized gas composed of electrons analyte having a distinct peak at a given mass-to-charge and positively charged ions. ratio. Precision Spray chamber A measure of the spread of a series of repeat measurements The nebulizer sprays a fine mist of the sample into the spray about their mean value. This is normally expressed as the chamber. The spray chamber removes the large droplets from standard deviation or percent relative standard deviation. the mist that are highly variable in supply from one moment to It is important to note that a small spread (that is, a low the next. The spray chamber, therefore, must select only the standard deviation) does NOT give any information on the tiniest of the droplets and all the large droplets must be accuracy of the result. It is possible for a result to display pumped away. Only 3% of the droplets are allowed to reach excellent precision, yet be inaccurate. the plasma and 97% of the sample goes to the waste container. Reproducibility Torch The ability to read back (repeat) a standard or known solution A very hot flame. A source of bright light. The quartz or glass as an unknown sample and obtain the value that you specified. tubes that make up a plasma torch. The outside tube is the Reasons that you don’t get the same value back as you told the torch coolant tube which typically has a flow of 15 to 20 lpm instrument include: drift, calibration fit, or noise in the data. of Ar gas. This flow keeps the torch below its melting point. The middle tube is the auxiliary gas tube. The gas flow through Resolution (Optical Spectrometer) this tube is typically between 0 and 3 lpm of Ar gas. This gas The width of an analytical peak at half its height expressed pushes the plasma higher in the coil and is used to keep the in nanometers. A measure of the instrument’s ability to sepa- plasma away from the sample injector tube. The innermost rate two overlapping peaks. At the point of resolution, two tube is the sample injector tube which injects the very small peaks of equal height are said to be resolved if you can see a droplets of sample into the base of the plasma fireball. These 1/2 height or greater dip between the tops of the two peaks. droplets come from the spray chamber. The droplets rise in the torch where they are dried, and the various steps of molecular Resolution (Mass Spectrometer) decomposition and elemental excitation take place. The width of an analyte’s mass peak at 5% of its peak height – for a quadrupole ICP-MS this is usually <0.8 mass units. Wavelength The distance between two peaks of a wave. When Thermo Sample and skimmer cone Elemental refers to wavelengths we usually are speaking of In ICP-MS, ions are sampled from the bulk plasma initially wavelengths of light. through an orifice in the sample cone into the expansion chamber, then by a second orifice in the skimmer cone into Wavelength flexibility the ion lens region of the mass spectrometer. The ability of an instrument to “see,” or measure any wave- length without having to change any hardware. Sample introduction system The apparatus used to turn the sample into an ultra-fine aerosol that can be introduced into the atomizer without over- whelming it with either too much sample or particles which are too large, either of which will extinguish the plasma or flame by taking too much thermal energy from it. Sample throughput The rate at which samples can be analyzed and the results printed. Usually measured as samples per hour. 19 1069 primer reprint 11/29/01 2:42 PM Page 6 Dilution Glossary To reduce the concentration of a sample solution by the Accuracy addition of an appropriate acid or organic solvent. This may The degree of agreement between the measured result be done to reduce the concentration of an analyte so that it and the accepted or known value. If quantified, this may be falls within the dynamic range of the instrument or to reduce expressed as a percentage difference. It is important to note the total dissolved salt concentration of an aqueous solution that this is NOT the same as precision. It is possible for a result or viscosity of an organic solution so that it can be nebulized. to be accurate but not precise, or very precise but not accurate. Dynamic range Aerosol The concentration range over which the instrument continues An extremely fine mist generated by a nebulizer. This is the to read a higher signal for an increase in the concentration of form in which the sample solution is presented to the flame or the measured material and can be calibrated. plasma for volatilization and subsequent measurement. Hydride generation accessory Analyte An accessory designed to generate free hydrogen that is allowed The element whose concentration is being determined. The to bubble through the sample liquid. Certain elements react atomic spectroscopy techniques discussed in this primer deter- with the hydrogen gas to form hydrides of those elements. mine the concentration of an analyte element by comparing the Elemental hydrides are in the form of a gas so it is easy to sepa- instrumental response for the sample to that of a series of stan- rate them from the liquid. These gases are allowed to enter the dard solutions containing the analyte element of interest. sample cell where they are atomized. Because all of the element in the gas is exposed to the source, the signal is much larger Atomic absorption than if it were introduced as a liquid using a nebulizer (only A method that measures the concentration of atoms of an about 5-10% efficiency). Therefore, much lower concentrations element by passing light, emitted by a hollow cathode lamp of the element (such as As, Se, Sb and Pb) can be detected. of that element, through a cloud of atoms from a sample. Only those atoms that are the same as those in the lamp will absorb Ions the light from the lamp. A reduction in the amount of light An atom which has lost one or more electrons becoming posi- reaching the detector is seen as a measure of the concentration tively charged due to the fact that it now has more protons of that element in the original sample. than it has electrons. Normally atoms are neutral in charge, electrons = protons, but if an atom loses an electron it becomes Atomic emission an ion and can be directed by electromagnetic fields. A method that measures the concentration of atoms or ions of an element in a sample, using the light those atoms emit when Laser accessory heated to very high temperatures. Since the intensity of the An accessory designed to use a UV laser to ablate (erode away) light that is emitted by these atoms or ions is proportional to the surface of conductive or non-conductive materials.