Magnetic Resonance Imaging, a Success Story for Superconductivity

Total Page:16

File Type:pdf, Size:1020Kb

Magnetic Resonance Imaging, a Success Story for Superconductivity FEATURES Magnetic Resonance Imaging a success story for superconductivity l René Aarnink1 and Johan Overweg2 - DOI: 10.1051/epn/2012404 l 1 Minimally Invasive Healthcare Department Head, Philips Research Eindhoven l 2 Principal Scientist, Tomographic Imaging Systems Department, Philips Research Hamburg Superconductivity is at the heart of MRI scanners, helping to give doctors an unprecedented view of structures deep within the human body. Nevertheless, challenges still exist in realizing the full potential of this important imaging modality, some of which relate to physics and some to economics. ithout superconductivity, one of to- The field strength generated by a current-carrying con- day’s most important medical imag- ductor can be calculated using Ampere’s Law. If the MRI ing modalities would not be possible. magnet were a simple long solenoid with uniform wind- Magnetic Resonance Imaging (MRI), ing density, this would reduce to: Wwhich depends on being able to align the weak magnetic moment of the protons in a patient’s body with a mag- B = µ0 I N/L netic field, requires the use of magnetic field strengths in m Working screen for Magnetic excess of 1T. This can only be achieved practically through where B = field strength in Tesla; I = current per turn; N = Resonance Imaging the use of superconducting electromagnets. number of turns; L = coil length and µ0 is the permeability 26 EPN 43/4 Article available at http://www.europhysicsnews.org or http://dx.doi.org/10.1051/epn/2012404 MAGNETIC RESONANCE IMAGING FEATURES of free space (to a close approximation, equal to that of shorter scan times, increasing patient throughput and the patient/air within an MRI scanner). reducing patient discomfort. There is therefore a trend This simple formula works out at approximately 1 × 106 toward higher field strength scanners, with 3T MRI scan- ampere-turns (around 10,000 turns of wire driven with ners already being used in routine clinical care and 7T 100 amps) for a typical MRI magnet length of 1.2 metres machines being used in (pre-)clinical research. and a field of 1 tesla. For several reasons, practical MRI However, 7T MRI scanners are extremely expensive and magnets actually have around 5 times as many ampere- will probably remain so, in particular because the current- turns. Firstly because of their small length-to-diameter carrying capacity of the superconductors employed de- ratio; secondly because of the increasing turns density creases rapidly above 5T, leading to a disproportionate needed towards the ends of the coil to make the field increase in the amount of con- homogeneous; and lastly because of the need for active ductor needed. In addition, the stray-field compensation (the magnet consists of two electro-magnetic (Lorentz) forces concentric coil sets with opposing polarity, shaped in that have to be contained become such a way that the outer coil to a large extent cancels a real problem, increasing with Zero boil-off magnets are the external magnetic field of the inner coil). the square of the field. The main now state of the art Without the use of superconducting materials such as problem with Lorentz forces on “ Niobium Titanium, cooled with liquid helium to below superconducting conductors is ” their critical temperature, the energy loss in such coils that they can lead to motion of the conductors, which would make modern MRI scanners impractical. Thanks causes friction heat. It only takes a few microjoules of to these materials, however, MRI scanners are already heat energy to raise a small piece of the superconductor commonplace in hospitals, providing superior soft-tissue above its critical temperature (the heat capacity at 4K is imaging to assist doctors in the diagnosis of a wide range extremely small) and if this non-superconducting zone of diseases, including cancer, Alzheimer’s, stroke, trauma starts to propagate through the coil, the entire stored en- injury and Parkinson’s disease. ergy of the magnet (many megajoules) is dumped into the liquid helium cooling the magnet (this is known as Field strength and uniformity a quench). MRI magnets are designed to survive such The characteristics of the magnetic field generated by an event without permanent damage, but re-cooling and an MRI scanner have a significant impact on its per- bringing the magnet back to the required field strength formance. The field strength has to be very uniform in is time consuming and costly. the imaging volume (typically the cross-section of the human body, with a spherical volume of around 45 to Cryostat design 50 cm diameter in the center of the magnet) and it has In most of today’s MRI magnets, the superconducting to remain extremely stable during the scan. In addition, coil is enclosed in a helium tank containing approxi- the proton polarization, which determines the strength mately 1000 litres of liquid when full. Magnets are usu- of the NMR signal, increases linearly with field strength. ally cooled down and filled only once, in the factory High field strengths therefore enable increased signal- where they are made, and remain cold for their entire to-noise ratios and better image quality. They also allow life (typically 10 years). How MRI worKS When a person is in an MRI scanner, the Gradient fields in the three dimensions powerful magnetic field and powerful hydrogen nuclei (i.e., protons, found in allow the scanner to work only with pro- radio waves. abundance in the human body as water) tons from a ‘slice’ at a time, allowing the MRI is used to image every part of the align with the strong magnetic field. A ra- creation of a whole volume that can be body, but is particularly useful in neu- dio wave at just the right frequency for the looked at in three dimensions. rological conditions, disorders of the protons to absorb energy pushes some of Contrast agents may be injected intrave- muscles and joints, for evaluating tu- the protons out of alignment. The protons nously to show enhancement of blood mors and showing abnormalities in the then snap back to alignment, producing a vessels, tumors or inflammation. Unlike heart and blood vessels. n detectable rotating magnetic field as they CT scanning, MRI uses no ionizing radia- do so. Since protons in different areas of tion and is generally a very safe proce- source: the body (e.g., fat versus muscle) realign at dure. Patients with some metal implants www.websters-online-dictionary.org/ different speeds, the different structures and cardiac pacemakers are prevented definitions/Magnetic+resonance+ of the body can be revealed. from having an MRI due to effects of the imaging EPN 43/4 27 FEATURES MAGNETIC RESONANCE IMAGING Any heat leak into this cold mass causes helium to high-vacuum space and the magnet’s suspension sys- evaporate (at a rate of around 1.4 litres per hour per tem has a low thermal conductivity, the single biggest watt of thermal load). If the cold mass is enclosed in a contributor to this heat load is thermal radiation. Older MRI magnets relied on two refrigerator-cooled shields in the vacuum space to reduce the heat load to less than 50 mW (the radiated heat decreases with the fourth power INCREASED Patient comfort AND acceSS of the surface temperature of the shield seen by the cold mass), resulting in a loss of less than 1 litre of helium per Some patients feel very claustrophobic in the bore of a conven- day. More modern refrigerators have about 1W of cooling tional cylindrical magnet MRI scanner, and the sudden loud noises power at 4K, allowing boil-off from the helium tank to be emitted by the scanner when the gradient coils are activated often stopped completely. These zero boil-off magnets are now worsens this feeling and can cause the patient to move suddenly. the state of the art. The rapidly increasing cost of helium Such movements can degrade image quality, so sedatives and/or more than justifies the extra cost of such a refrigerator. ear plugs are often used to calm the patient and minimize the effect. However, it takes about 7 kW of electrical power to drive the refrigeration system. One of the challenges in magnet design is therefore to make these refrigeration systems more efficient and reliable so that the loss of helium due to planned and un-planned downtime of the refrigeration system is mini- mized. When the magnet is being shipped or temporar- ily stored before installation, the refrigerator is also kept running whenever possible. With the escalating cost of helium, which is a by-product of natural-gas extraction, another important challenge for the future is minimizing the helium content of the mag- net. A few hundred litres of liquid helium or less would actually be sufficient to keep the magnet superconduct- ing, so the main challenge in developing magnets without a large helium buffer volume is to make the refrigeration system so reliable that it never fails, or so efficient that it can re-cool the magnet without using much external m The Panorama HFO, Philips’ high-field, wide open system, offers outstanding clinical helium after a failure. versatility with leading solutions for all clinical applications. And it is preferred by patients because of its comfortable space and wide open bore. High-temperature superconductors One way of eliminating the need for helium in MRI In an attempt to increase patient comfort and provide clinicians with scanners completely would be to use high-temperature better access to the patient, for example to perform MRI-guided clin- superconducting materials such as Yttrium Barium Cop- ical interventions, open-magnet MRI scanners have been developed. per Oxide (YBCO) tape.
Recommended publications
  • Introduction to Cryostat Design
    Introduction to Cryostat Design J. G. Weisend II July 2014 Course Goals Provide an introduction to cryostat design Examine cryostats as integrated systems rather then a group of subsystems Provide background information and useful data Provide pointers to resources for more detail Show examples Introduction to Cryostat Design 7/7/14 2 J. G. Weisend II ICEC 2014 Syllabus Introduction Cryostat Requirements Materials Selection Thermal Isolation and Heat Leak Structures Safety Instrumentation Seals, Feed throughs and Bayonets Transfer Lines Examples References and Resources Introduction to Cryostat Design 7/7/14 3 J. G. Weisend II ICEC 2014 Introduction What is a cryostat? A device or system for maintaining objects at cryogenic temperatures. Cryostats whose principal function is to store cryogenic fluids are frequently called Dewars. Named after the inventor of the vacuum flask and the first person to liquefy hydrogen Introduction to Cryostat Design 7/7/14 4 J. G. Weisend II ICEC 2014 Introduction Cryostats are one of the technical building blocks of cryogenics There are many different types of cryostats with differing requirements The basic principles of cryostat design remain the same Before we can do anything else we have to define our requirements Introduction to Cryostat Design 7/7/14 5 J. G. Weisend II ICEC 2014 E158 LH2 Target Cryostat Introduction to Cryostat Design 7/7/14 6 J. G. Weisend II ICEC 2014 Cryostat Requirements Maximum allowable heat leak at various temperature levels This may be driven by the number of cryostats to be built as well as by the impact of large dynamic heat loads (SCRF or target cryostats) Alignment and vibration requirements Impact of thermal cycles Need to adjust alignment when cold or under vacuum? Alignment tolerances can be quite tight (TESLA : +/‐ 0.5 mm for cavities and +/‐ 0.3 mm for SC magnets) Introduction to Cryostat Design 7/7/14 7 J.
    [Show full text]
  • Versatile Cryogen-Free Cryostat for the Electromagnetic Characterization Of
    Saniour et al. EPJ Techniques and Instrumentation (2020) 7:3 EPJ Techniques and https://doi.org/10.1140/epjti/s40485-020-00055-2 Instrumentation RESEARCH ARTICLE Open Access Versatile cryogen-free cryostat for the electromagnetic characterization of superconducting radiofrequency coils Isabelle Saniour1*† , Michel Geahel1,2,3†, Javier Briatico2, Cornelis J. van der Beek3,4, Georges Willoquet1, Laurène Jourdain1, Bertrand Baudouy5, Gilles Authelet5, Jean-Christophe Ginefri1, Luc Darrasse1 and Marie Poirier-Quinot1 * Correspondence: isabelle.saniour@ universite-paris-saclay.fr Abstract Isabelle Saniour and Michel Geahel are Co-first authors. The use of high temperature superconducting (HTS) radio frequency (RF) coils in †Isabelle Saniour and Michel Geahel Magnetic Resonance Imaging (MRI) greatly improves the signal-to-noise ratio (SNR) contributed equally to this work. in many biomedical applications and particularly in micro-MRI. However, a detailed 1Université Paris-Saclay, CEA, CNRS, Inserm, BioMaps, Orsay, France understanding of the electrical behavior of HTS coils is important in order to Full list of author information is optimize their performance through MR experiments. This paper presents a simple available at the end of the article and versatile cryogen-free cryostat designed to characterize the RF properties of HTS coils prior to their use in MRI. The cryostat can be used at temperatures from 50 K to 300 K, with a control precision of approximately 3 mK at 70 K, and can measure the RF electrical power transmitted to an HTS coil over a range from 1 μW to 10 W. The quality factor and resonance frequency of the tested HTS coil are determined as a function of the temperature and the power it dissipates.
    [Show full text]
  • New Target Cryostat for Experiments with Negative Muons
    ! XJ9900001 OBbEflMHEHHbM MHCTMTYT flflEPHbIX MCCJlEflOBAHMM flyoHa D13-98-146 V.F.Boreiko, V.M.Bystritsky, V.I.Datskov, A.N.Fedorov, V.N.Pavlov, V.A.Stolupin, A.Del Rosso1, R.Jacot-Guillarmod1, F.Mulhauser1, L.A.Rivkis2 NEW TARGET CRYOSTAT FOR EXPERIMENTS WITH NEGATIVE MUONS Submitted to «Nuclear Instruments and Methods* 'institute of Physics, University of Fribourg, Switzerland 2A.A.Bochvar All-Russia Research Institute of Inorganic Materials, Moscow, Russia 30-07 1998 Introduction The study of processes yielding charge nonsymmetric muonic molecules like pxZ (z = p,d,t and Z = He, Li, Be) and nuclear fusion started about fifteen years ago. Theoretical and experimental interest have increased these past five years [1-13]. Due to some experimental difficulties, the muon catalyzed fusion (pCF ) is essentially measured in deuterium-helium mixtures. The nuclear fusion from the charge nonsymmetric molecule pd 3He may occur via two different channels, pd3He -4 a + p(14.7 MeV) + p (1) -4 psLi +7(16.4 MeV), (2) whereas the pd4He molecule fuses via a single channel pd*He -4 p6U + 7(1.48 MeV). (3) The proposed apparatus should be able to easily detect those fusion products. In addition, the 6.85 keV soft X rays resulting form the decay of the pd4He and pd3He molecules should be measured, as well as the muon decay electrons. The detection of so many different types of particles, with totally different paths in matter, is a non-trivial constraint in the design of the apparatus. The different processes which occur when a muon enters a deuterium-helium mix­ ture are strongly dependent on the mixture density.
    [Show full text]
  • Design and Fabrication of the 1.9 K Magnet Test Fa- Cility at BNL, and Test of the First 4 M-Long MQXF Coil
    Mon-Af-Po1.01-05 1 Design and Fabrication of the 1.9 K Magnet Test Fa- cility at BNL, and Test of the First 4 m-Long MQXF Coil J. Muratore, M. Anerella, P. Joshi, P. Kovach, A. Marone, P. Wanderer Abstract— The future high luminosity upgrade of the Large Had- vertical superconducting magnet test facility of the SMD at ron Collider (LHC) at CERN will include twenty 4.2 m-long BNL has been upgraded to perform testing in superfluid He at Nb3Sn high gradient quadrupole magnets which will be compo- 1.9 K and 1 bar, the operational condition at the LHC. This nents of the triplets for two LHC insertion regions. In order to test these and four pre-production models, the vertical supercon- has involved extensive modifications of the 40-year old, 4.5 K ducting magnet test facility of the Superconducting Magnet Divi- cryogenics plant and vertical test facility at the SMD. These sion (SMD) at Brookhaven National Laboratory (BNL) has been upgrades include new piping, compressors, and other critical upgraded to perform testing in superfluid He at 1.9 K and 1 bar, components; a new vertical test cryostat with a top plate and the operational condition at the LHC. This has involved extensive hanging fixture which can accept larger diameter and longer modification of the 4.5 K cryogenics plant, including piping, magnets, up to an actual length of 5 m; a 1.9 K heat exchang- compressors, and other upgraded components; a new vertical test cryostat which can accept larger diameter magnets; a modern- er; a newly designed warm bore tube; and a lambda plate de- ized power supply system upgraded with IGBT switches and fast signed with a novel sealing scheme to provide for both shutoff capability, and that can supply 24 kA to test high field strength and minimum heat loss.
    [Show full text]
  • Cryostat Design
    Published by CERN in the Proceedings of the CAS-CERN Accelerator School: Superconductivity for Accelerators, Erice, Italy, 24 April – 4 May 2013, edited by R. Bailey, CERN–2014–005 (CERN, Geneva, 2014) Cryostat Design V. Parma1 CERN, Geneva, Switzerland Abstract This paper aims to give non-expert engineers and scientists working in the domain of accelerators a general introduction to the main disciplines and technologies involved in the design and construction of accelerator cryostats. Far from being an exhaustive coverage of these topics, an attempt is made to provide simple design and calculation rules for a preliminary design of cryostats. Recurrent reference is made to the Large Hadron Collider magnet cryostats, as most of the material presented is taken from their design and construction at CERN. Keywords: superconductivity, cryogenics, vacuum, heat transfer, MLI, construction technologies. 1 Introduction The term cryostat (from the Greek word κρύο meaning cold, and stat meaning static or stable) is generally employed to describe any container housing devices or fluids kept at very low temperatures; the notion of ‘very low temperature’ generally refers to temperatures that are well below those encountered naturally on Earth, typically below 120 K. The very first cryostats were used in the pioneering years of cryogenics as containers for liquefied gases. The invention of the first performing cryostats is generally attributed to Sir James Dewar (Fig. 1), and hence cryostats containing cryogenic fluids are nowadays also called dewars. In 1897 Dewar used silver-plated double-walled glass containers to collect the first liquefied hydrogen. Even though the heat transfer phenomena were not well mastered during his époque, Dewar understood the benefits of thermal insulation by vacuum pumping the double-walled envelope, as well as shielding thermal radiation by silver-plating the glass walls.
    [Show full text]
  • HELIUM-3 for Researchers CIA CRYO Industries of America, Inc
    26 years of making cryogenics HELIUM-3 for Researchers www.cryoindustries.com CIA CRYO Industries of America, Inc. He-3 Cryostats - Sample in vacuum or top loading into vapor/liquid • Lowest base temperature due to advanced synthetic charcoal technology • Versatile 3-way charcoal sorb cooling system for lowest vibration and sub cooling • Flex circuits for quick & easy wire installation/removal and automatic thermal anchoring • Can be integrated into our high efficiency optical & non-optical cryostats, superconducting magnet systems, storage (transport) dewars or into your existing system. Helium-3 Introduction The basic principles of He3 inserts are indicated by the following three steps, 1,2,3. Warm the charcoal to release the adsorbed He3 gas which is then condensed by the 1 K POT. The 1liquid helium-3 collects in the He-3 pot, cooling the sample. Lowering the charcoal temperature to 4 K cryopumps the liquid He3 lowering its temperature . The isotope He3 is used instead of He4 because it does not exhibit film creep and a lower pressure can be 2reached (lower pressure means lower temperature). 3Collect the He3 gas back into the sorb and reuse - over and over and over again. He-3 gas He-3 gas in sorb back in sorb No gas left in charcoal sorb 1 K POT Condensing Pumped He-4 surfaces He-3 pot (empty) Liquid He-3 He-3 pot Sample mount (full) (empty) - 2 - Advanced Sorb Technology New charcoal technology provides for the higher pumping speeds - needed to cool the sample to less than 260 mK. 1200 Cooling Power Versus Temperature 1000 He-3 (SV std) W) μ 800 Screenshot taken during actual test of cryostat no.
    [Show full text]
  • Introduction to Cryostat Design J
    IntroductionIntroduction ToTo CryostatCryostat DesignDesign J. G. Weisend II National Science Foundation SLAC National Accelerator Laboratory CourseCourse GoalsGoals ProvideProvide anan introductionintroduction toto cryostatcryostat designdesign ExamineExamine cryostatscryostats asas integratedintegrated systemssystems ratherrather thenthen aa groupgroup ofof subsystemssubsystems ProvideProvide backgroundbackground informationinformation andand usefuluseful datadata ProvideProvide pointerspointers toto resourcesresources forfor moremore detaildetail ShowShow examplesexamples Introduction to Cryostat Design J. G. Weisend II 11/10/2008 ICEC 2008 2 SyllabusSyllabus IntroductionIntroduction CryostatCryostat RequirementsRequirements MaterialsMaterials SelectionSelection ThermalThermal IsolationIsolation andand HeatHeat LeakLeak StructuresStructures SafetySafety InstrumentationInstrumentation Seals,Seals, FeedFeed throughsthroughs andand BayonetsBayonets TransferTransfer LinesLines ExamplesExamples ReferencesReferences andand ResourcesResources Introduction to Cryostat Design J. G. Weisend II 11/10/2008 ICEC 2008 3 IntroductionIntroduction WhatWhat isis aa cryostat?cryostat? AA devicedevice oror systemsystem forfor maintainingmaintaining objectsobjects atat cryogeniccryogenic temperatures.temperatures. CryostatsCryostats whosewhose principalprincipal functionfunction isis toto storestore cryogeniccryogenic fluidsfluids areare frequentlyfrequently calledcalled DewarsDewars.. NamedNamed afterafter thethe inventorinventor ofof
    [Show full text]
  • INTRODUCTION to LABORATORY CRYOGENICS M.N. Jirmanus, Ph.D
    INTRODUCTION TO LABORATORY CRYOGENICS By M.N. Jirmanus, Ph.D. Janis Research Company, Inc. 2 Jewel Drive Wilmington, MA 01887 Tel: (978) 657·8750 Fax: (978) 658·0349 .' Preface This work is meant for scientists, students, and laboratory personnel who have had little or no experience in cryogenics. It concentrates primarily on cryogenic systems that are commercially available for operation between 1.5 K and 300 K (room temperature). For those users who want to design and build their own equipment, or who are interested in a more detailed analysis of cryogenic systems, a number of excellent references are listed at the end of this booklet. The first section discusses the vacuum requirements for laboratory dewars and variable temperature cryostats. This is followed by a section on liquid helium and liquid nitrogen dewars, and another section on variable temperature cryostats. The next section concentrates on superconducting magnets that are combined with variable temperature cryostats for laboratory experiments, requiring relatively large magnetic fields. section five describes closed cycle refrigerator cryostats that require no liquid cryogens, and the last section concentrates on experimental techniques, thermometry, and automatic temperature control. This last section also includes a few tables that are helpful in estimating heat loads on the cold stage and attached sample. More detailed information and experimental data are available from the referencea.listed.at.the end of this section. The figures (drawings) spread throughout sections 2 through 5, show a variety of dewar and cryostat designs, based primarily on our experience at Janis. These designs have evolved over a 30 year period, and represent a typical cross section of laboratory units that can be used for a wide variety of experiments.
    [Show full text]
  • Normal Cryogenic Fluid Behavior (Emphasis on Helium, but Not Superfluid, Which Will Be Covered Later)
    Normal Cryogenic Fluid Behavior (emphasis on helium, but not superfluid, which will be covered later) Tom Peterson, SLAC June 2019 Outline • Cooling modes for superconducting devices • Forced flow cooling • Two-phase flow and pool boiling • Fluid dynamics • Saturated bath thermodynamics June, 2019 Cryogenic Fluids 2 USPAS Tom Peterson Large-scale cooling of superconducting devices • Physicists and engineers designing a large-scale liquid helium system typically must design the cooled components (magnets or RF cavities, their containers, and the interfaces to them) • Cooling mode, heat transfer, pressure drops, cool- down, warm-up and non-steady or upset system operations all must be considered as part of the component design • The cooled devices must be viewed as part of the cryogenic system June, 2019 Cryogenic Fluids 3 USPAS Tom Peterson Cooling modes in large-scale cryogenic systems • Pool boiling helium I used in superconducting RF for HERA (DESY), LEP (CERN), KEKB (KEK, Japan), CESR (Cornell) • Forced flow of subcooled or supercritical helium I for cooling superconducting magnets (Tevatron, HERA, SSC, RHIC, ITER) • Stagnant, pressurized helium II (the Tore Supra tokamak in France demonstrated the technology, LHC magnets) • Saturated helium II (CEBAF, FLASH and Eu-XFEL at DESY, SNS at Oak Ridge, FRIB at MSU, LCLS-II at SLAC, and more) • This list also illustrates the extent to which superconductivity and cryogenics have become standard technology for accelerators June, 2019 Cryogenic Fluids 4 USPAS Tom Peterson Helium phase diagram
    [Show full text]
  • High Pressure Techniques at Low Temperatures
    Pramina-J. Phys., Vol. 28 No. 5, May 1987, pp. 435-469. © Printed in India. High pressure techniques at low temperatures S RAMASESHAN, G PARTHASARATHY and E S R GOPAL* Raman Research Institute, Bangalore 560080, India Instrumentation and Services Unit, Indian Institute of Science, Bangalore 560012, India *Also Department of Physics Abstract. The article reviews the experimental techniques used in high pressure-low temperature investigations to study a variety of physieo-chemical phenomena. The general principles of producing high pressures at low temperatures, the methods of measuring P and T, the materials used for construction and the diamond anvil cell (DAC) are briefly given first. Specific pieces of apparatus to measure the mechanical properties, phase equilibria, thermal properties, electrical properties, magnetic phenomena, optical and Raman/IR spectroscopic behaviour as well as M6ssbauer spectra are then discussed. While instrumentation is the main emphasis of the article, a few illustrative examples of interesting observations are also indicated. Over 250 current papers are cited. Keyword~ High pressure techniques; low temperatures; diamond anvil cell; mechanical properties; phase equilibria; thermal properties; electrical properties; magnetic phenomena; Raman spectra; M6ssbauer spectra; instrumentation. PACS No. 07"35 1. Introduction The study of the behaviour of materials at high pressures and low temperatures has been useful not only in understanding the properties of these materials but also in the observation of new features of the physicochemical properties. It is now believed that almost all materials will become metallic at sufficiently high pressures; therefore a combination of high pressures with low temperatures is a logical step in pursuing this phenomenon.
    [Show full text]
  • The Liquid Helium Cryogenic System for the Superconducting Cavity in Srrc
    Proceedings of the 2001 Particle Accelerator Conference, Chicago THE LIQUID HELIUM CRYOGENIC SYSTEM FOR THE SUPERCONDUCTING CAVITY IN SRRC F. Z. Hsiao, M. C. Lin, Ch. Wang, D. S. Lee, and J. R. Chen, SRRC, Hsinchu, Taiwan B. Hilbert, and A. Praud, Air Liquide DTA, Sassenage, France Abstract r A 500MHz superconducting cavity will replace the o ss n e r recovery valve p mai current copper cavity and begin to operate in the m compressor box co beginning of the year 2003. A liquid helium cryogenic oil system provides the cavity at 4.5K a cooling capacity of removal refrigerator 255W without LN2 pre-cooling and a liquefaction rate of 110 liter/hour with LN2 pre-cooling. A safety factor of 1.5 is used to estimate the heat load from the gas tank dewar cryostat superconducting cavity and the heat loss from the transfer lines. With the LN2 pre-cooling, this cooling system warm He gas cold He gas liquid He provides a cooling capacity of up to 450W to cool down Figure 1: Configuration of the Air Liquide designed the additional superconducting Landau cavity. The helium cryogenic system. capacity of the system can be tuned using a frequency driver installed at the compressor station. The pressure provides the required 4.5K cooling capacity to the fluctuations of the dewar and of the suction line are kept superconducting cavity. This system will be a turnkey to the same stability requirement that of the cavity system, which is designed and manufactured by the cryostat to minimize the influence in cavity operation.
    [Show full text]
  • Cryostatcryostat
    149E CRYOSTATCRYOSTAT Head office and factory 1222-1 Yabata, Chigasaki, Kanagawa 253-0085 Japan Technical service division Uji Factory 60, Ujihinojiri, Uji, kyoto, 611-0021, Japan Osaka branch office 3-3-31 Uemura Nissei bldg., Miyahara, Yodogawa-ku, Osaka 532-0003 Japan Overseas group companies Korea 107, Hyeongokasndan-Ro, Cheongbuk-Myeon, Pyeongtaek-Si, Gyeonggi-Do, Korea, 451-831 For inquiries China http://www.ulvac-cryo.com/english/ ●The content of this catalog is subject to change without notice for improvement of the products. 1410000PL Mini Stat provides powerful support for the research and development in the cryogenic eld. CRT-006 SERIES For cryogenic research and development, cryostats or cryocoolers are needed to cool samples (specimen). For Optical Transmission Measurement CRT-006-1000 ULVAC CRYOGENICS Mini Stat is our unique cryostat that employs a compact cryocooler (Cryo-Mini) instead of cryogens that need to be replenished such as liquid nitrogen or liquid helium. For measuring samples with a spectroscope With electricity and water, Mini Stat can cool samples to cryogenic temperature by simple operation. In This model is used to measure samples in conjunction with a spectroscope, at temperatures from cryogenic to ambient. Coldhead addition, with temperature controllers, samples can be kept precisely at a desired temperature between Specifications Cryogenic Heater cryogenic and room temperature, enabling the measurement in a wide temperature range. thermocouple lead wire Coldhead HE05/UW404 D 510/S W115 D10 5/S W112 Ultimate Temperature 4K 12K 12K Heater to We offer wide variety of cryocoolers to provide the best choice for your desired temperature as well as we control Temperature 4K-300K±0.2K 12K-300K±0.2K 12K-300K±0.2K temperature Controllable Range design and manufacture cryostats that satisfy various kinds of requirements based on our rich experience Optical Window φ20×4 Sample and expertise.
    [Show full text]