THE NATIONAL MEASUREMENT SYSTEM FOR TEMPERATURE

James F. Schooley

Institute for Basic Standards National Bureau of Standards Washsington, D. C. 20234

December 1975

Issued August 1 976

U.S. DEPARTMENT OF COMMERCE, Elliot L Richardson, Secretary Edward O. Vetter, Under Secretary Dr. Betsy Ancker-Johnson, Assistant Secretary for Science and Technology

NATIONAL BUREAU OF STANDARDS, Ernest Ambler. Acting Director

THE NATIONAL MEASUREMENT SYSTEM

FOR TEMPERATURE

J. F. School ey Heat Division Institute for Basic Standards FOREWORD

CONCEPT OF A NATIONAL MEASUREMENT SYSTEM

"Concurrently with the growth and industrialization of this nation, there has developed within it a vast, complex system of measurement which has made possible the very growth that brought the system into being. This National Measurement System (NMS) stands today as one of the key elements in a worldwide measurement system that links all major nations together in a consistent, compatible network for communication and trade. "Briefly stated, the essential function of the NMS is to provide a quantitative basis in measurement for (i) interchangeability and (ii) decisions for action in all aspects of our daily life -- public affairs, commerce, industry, science, and engineering. "Our National Measurement System is one of a number of mutually interacting systems within our technologically based society that form the environment in which the individual citizen must live and function. Familiar examples are the conmuni cation, transporta- tion, educational, medical, and legal systems, all of which may be included under the general heading of social systems. "In view of the demonstrated value of the systems approach for the understanding and improve- ment of hardware such as computers and weapons, some of these social systems are being subjected to the same type of analysis. The National Measurement System, which evolved ' in this country with little formal recognition as a system, is now being examined in this way at the National Bureau of Standards (NBS) which undertook the study of NMS partly because of a growing realization of the all -pervasive nature and great economic importance of the nation's measurement activities, and partly because of the challenge to NBS in putting its splendid new facilities to optimum use for the benefit of the nation. Such optimum use can be approached only when NMS, of which NBS is a central element, and the services it requires for effective operation are sufficiently well understood. "Because the government has wisely refrained from assigning leadership of NMS to NBS by law or executive order, the Bureau of Standards must maintain this leadership through dem- onstrated competence and general acceptance of its capability. This situation presents both a challenge and a responsibility. The NBS must make a continuous effort to under- stand the structure and operation of the NMS, to assess the value of its services for national objectives, and to develop means for evaluating its effectiveness. "Our study of the National Measurement System is still in its early stages. There remains much work to characterize inputs and outputs, interactions with other social systems, involvement with national objectives, the functions of NMS elements, and couplings between the elements."

This description of the National Measurement System was first published by Dr. R. D.

Huntoon in Science, 1 58 , 67-71, October 6, 1967. Dr. Huntoon at that time was director of the Institute for Basic Standards^ of the National Bureau of Standards. For the reasons stated in the above quotation. Dr. Huntoon began and fostered systematic studies of the National Measurement System of which the following work is a part. CONTENTS

Fane EXECUTIVE SUMMARY ~V~

1. INTRODUCTION 3 2. STRUCTURE OF THE NATIONAL SYSTEM 4 2.1 Conceptual System of Temperature 4 2.2 Basic Thermometry Infrastructure 5 2.2.1 Documentary Specification System 5 2.2.1.1 Standardization Institutions 5 2.2.1.2 Survey of Documentary Standards 7 2.2.2 Thermometry Instrumentation System 8 2.2.2.1 Temperature Measurement Tools and Techniques 8 2.2.2.2 The Thermometry Instrumentation Industry 11 2.2.3 Reference Data 14 2.2.4 Thermometry Reference Materials 15 2.2.5 Science and People 15 2.3 Realized Thermometry Measurement Capabilities 16 2.4 Thermometry Dissemination and Enforcement Network 16 2.4.1 Central Thermometry Standards Authorities 16 2.4.2 State and Local Offices of Weights and Measures 18 2.4.3 Temperature Standards and Testing Laboratories and Services 18 2.4.4 Regulatory Agencies 20 2.5 Direct Measurements Transactions Matrix for Thermometry 21 2.5.1 Analysis of Thermometry Suppliers and Users 21 2.5.2 Highlights of Thermometry Users 21 3. IMPACT, STATUS, AND TRENDS OF THE TEMPERATURE MEASUREMENT SYSTEM 26 3.1 Impact of Thermometry Measurements 26 3.1.1 Functional, Technological and Scientific Applications 26 3.1.2 Economic Impacts - Costs and Benefits 28 3.1.3 Social, Human, Person-on-the-Street Impacts 28 3.2 Status and Trends of the Temperature Measurement System 28 4. SURVEY OF NBS SERVICES TO THERMOMETRY 34 4.1 The Past 34 4.2 The Present-Scope of NBS Thermometry Activities 34 4.2.1 Description of NBS Thermometry Services 34 4.2.1.1 Heat Division Activities, 1974 34 4.2.1.2 NBS Thermometry Activities Outside Heat Division 36 4.2.1.3 The NBS Thermometry Input-Output Table 37 4.2.2 Users of NBS Services 38 4.2.3 Alternative Sources 38 4.2.4 Funding Sources for NBS Thermometry Services 38 4.2.5 Mechanisms for Supplying Services 39 4.3 Impact of NBS Thermometry Services 40 4.3.1 Economic Impact of Major User Classes 40 4.3.2 Technological Impact of NBS Services 40 4.3.3 Pay-Off from Changes in NBS Services 40 4.4 Evaluation of NBS Thermometry Program 41 4.5 The Future of Thermometry at NBS 42 5. SUMMARY AND CONCLUSIONS 45 6. ACKNOWLEDGEMENTS 45 7. REFERENCES 46 APPENDIX A. Methodology of the Study 50 APPENDIX B. Summary of Background Documents 50 APPENDIX C. The 1927 International Temperature Scale 51 APPENDIX D. NBS Participation in Voluntary Standards Bodies' Thermometry Activities 55 APPENDIX E. Resistance 55 APPENDIX F. Other Temperature Sensors in More or Less Common Use 56

iii LIST OF FIGURES

Page Figure 2.2A Sketch of the structure of the International Electrotechnical Commission, The International Standards Organization, and the International Organization for Legal Metrology 6 Figure 2.2B Useful Sensor Ranges 10 Figure 2.2C of Water 12 Figure 2.4A The U. S. Navy Temperature Calibration Chain 18

LIST OF TABLES

Table 2.2A Primary Fixed Points on the IPTS-68 7 Table 2. 28 Thermocouple Characteristics, from ASTM E230-72 9 Table 2.2C Sensors in Common Use at Various 9 Table 2.4A The Structure of the General Conference on Weights and Measures .... 17 Table 2. 48 Range and Accuracy of U. S. Army Calibrations 19 Table 2.4C National Conference of Standards Laboratories, List of Commercial Facilities 20 Table 2.5A The Thermometry Community and Its Roles as Suppliers and Users of Thermometric Goods and Services 22 Table 2.58 Thermometry Input-Output Matrix 25 Table 4.2A N8S Thermometry Input-Output Table 37 Table 4. 28 Cryogenic Physics and Temperature Sections, Funding by Sources, Fiscal Year 1974 39 Table 4.3A N8S-National Needs Chart 41

iv .

THE NATIONAL MEASUREMENT SYSTEM FOR TEMPERATURE

J. F. School ey

Heat Division Institute for Basic Standards

January 1976

ABSTRACT

The National Temperature Measurement System reaches quite literally into all phases of American life. Comfort control in home, school, office and factory, health care, manufac- turing, food preparation and storage, and all forms of powered transportation are just a few of the many facets of America's human activities that depend for their trouble-free operation on the existence of reliable thermometry. The Heat Division of the National Bureau of Standards collaborates with the national laboratories of other nations in establishing an International Practical Temperature Scale which represents thermodynamic accuracy insofar as current scientific practice permits. The NBS is the only U.S. agency bearing this responsibility; in addition, only the NBS is responsible for disseminating the International Scale to U.S. scientific and technical activities. The Study of the National Temperature Measurement System shows that the NBS continues to contribute energetically to the quality of the International Scale. It further indicates that NBS maintains a consistent and deliberate effort not only to provide access to the Scale at the several levels required by U.S. science and technology, but also to participate in solving special problems in thermometry. These are problems for which the nature of NBS as an objective, expert resource in thermometry uniquely qualifies it to furnish effective, practical solutions. In briefly describing the many aspects of the National Temperature Measurement System, this Study attempts to portray its great diversity of products, services, and people. The System is largely organized, as might be expected, for the purpose of economic gain; the NBS enters the picture only in those areas where highly accurate or intricate thermometry is essential

V

THE NATIONAL MEASUREMENT SYSTEM FOR TEMPERATURE

J. F. School ey

Heat Division Institute for Basic Standards

January 1976

EXECUTIVE SUMMARY

Temperature is a well-defined, basic 2) The type of measurement which is scale quantity. It is perhaps the most needed is not one for which conmonly-measured of all the physical units calibration rationale exists. (the possible exception being that of time). Fast-response, hostile environ- This very ubiquity, however, has a double ment, and geometrically- effect. We desire (in principle) that we constrained thermometry require- should have access to very accurate thermom- ments exist in many fields of etry, but, on the other hand, we tend to endeavor; many of these are noted take the measurements for granted, so that in the full report. our practice of thermometry is often of 3) The calibration procedure which poorer accuracy than we suspect. exists is often expensive. The The direct temperature measurement and drugstore is an example control industry is a large one by any of this problem— a $10 standards, including several dozen multi- calibration of a $1 thermometer million dollar firms. The amount of is simply impractical. economic leverage which thermometry exer- Adding to these problems those which cises is multiplied greatly, however, by exist where the temperature scale is not the fact that good thermometry is essential adequate to today's needs gives one a fair to the comnercial success of many industries, picture of the opportunities which lie be- such as semiconductor, steel, chemicals and fore NBS in thermometry today. plastics manufacturing, medical testing and Again in brief terms, NBS thermometry care, and precision instrument manufacture programs lie in several general areas: and use. Temperature is acquiring a new 1) Generate an adequate scale below urgency in other areas as well, with the 20 K and above 1300 K. tremendous demand for new sources of energy; 2) Devise calibration and temperature with the more stringent pollution and standardization procedures which efficiency requirements under consideration will solve people's problems, for automobiles and aircraft; with the heavy rather than simply transmitting the new demands for safety and output being highest attainable accuracy. placed on nuclear power reactors; and with 3) Participate in the development of the desire to map and exploit the ocean's thermometry methods in areas where depths. glaring weaknesses exist. In briefest terms, NBS is meeting the The NBS temperature measurement activi- heaviest of today's demands. The temper- ties impinge on the U.S. temperature measure- ature scale is known and can be transmitted ment system directly in a multitude of ways. on demand with an accuracy which is satis- The most obvious, of course, is through the factory for most purposes in the range 20- calibration of precise temperature sensors. 1300 K. The calibration services offered by A second interaction is through NBS parti- NBS for platinum resistance thermometers, cipation on voluntary standards committees; for liquid-in-glass thermometers, and for that is a quite important activity, because thermocouples answer most of the direct it serves both to communicate to others needs for accuracy in that range. expertise and data originating at NBS and to Yet many human activities suffer a lack alert the NBS staff to problem areas. A of adequate thermometry, even in the same third major interaction with the temperature 20-1300 K range. The particular reasons for community arises through NBS staff parti- this situation are numerous, but they come cipation in the development of measurement down to a few general problems: methods—in general, this interaction occurs 1) Adequate thermometry exists, but it directly with one or more technical is not applied to the problem. The thermometry groups in industrial or scien- whole field of practical medicine tific laboratories. Many such instances suffers from this difficulty. are detailed in the report.

1 The National Measurement System study pressing and susceptible problems. We has been a valuable tool in our effort to intend to continue to use this approach to focus our limited resources on the most assess and, as necessary, to modify the NBS temperature program.

2 1 . INTRODUCTION problems can be alleviated and the costs reduced by basic or applied research or As the stewardess turned and walked to- other services which might be undertaken by ward the back of the plane, it banked the NBS. sharply and pitched forward. The pretty This study is part of a large enter- woman in Seat 3A clutched her baby tightly prise [2] which has a broad goal— to and glanced anxiously at her husband. document the extent to which the NBS is "What is it?" she asked. meeting its statutory responsibility to [3] "I don't know, replied her husband as 1) provide a central basis for a he instinctively steadied her and their complete and consistent system of baby with one hand and placed the other on physical measurement in the U.S., their sleeping son to his right. "We are 2) coordinate that system with those surely going down in a hurry". of other nations, and The plane dropped rapidly. Looking out 3) furnish essential services leading of the side window, the frightened couple to accurate and uniform physical could see the rising lights of a city. The measurement in science, industry woman gasped as the plane pulled up just as and commerce. a crash seemed inevitable. The lights of a It is a natural step to separate the runway wheeled into view and disappeared NBS activities according to the various units behind the nose of the aircraft as the of measurement, and to present the discussion wheels bumped onto the pavement. A flash- for each unit along the lines sketched in ing red light appeared beside the plane, the previous paragraph. Thus, this portion testifying to the presence of an emergency of the overall study will discuss the extent vehicle beside them. to which the NBS is meeting its statutory Then, as the plane began to slow, the responsibilities in the area of temperature voice of the pilot sounded over the cabin measurement. intercom. "We're sorry for the hasty land- The study of thermometry has a long ing, folks, but we received an engine and honorable history. Undoubtedly, the overheat alarm up there, and the book says fundamental significance of temperature to 'Bring it down while you still have two thermodynamics and its importance to the wings'. We'll check it out now and, if it study of material properties has led to its was a false alarm, we'll be back in the air preeminence as a subject of painstaking in no time. Meanwhile, please sit back, research and careful national and inter- relax, and listen to some music". national deliberation. For whatever reason, The woman's husband, a physicist, em- however, temperature is far and away the ployed at the National Bureau of Standards, most widely studied of all the standard audibly released his breath and, noticing units. Thus we shall find that the con- that his hands were still pressing forcibly ceptual bases for temperature and for its on his wife and children, he consciously measurement are soundly placed, and that released his grasp. Leaning back in his there is very substantial agreement on a seat, he pondered the rush of events which well -developed temperature scale. Further- had followed the placid removal of his more, the National Bureau of Standards has dinner tray by the stewardess. Noting the a distinguished history of leadership in absence of any visible disturbance in the thermometric affairs, and we shall see that engine nacelles looming darkly from the the NBS continues today to participate in wings on either side of the passenger com- the expanding frontier of thermometry. partment, he wondered what the real cost The very ubiquity of thermometry, of this particular engine overheat alarm however, has led to two problems which this sensor malfunction might be. First, there study will treat: one problem is that so was the extra landing— jet fuel, extra crew many temperature measurements are made and time, airport fees. Then, whatever missed so many thermometry facilities exist, that appointments might take place as a result even cataloging them, let alone their of the inevitable delay here—wherever they analysis, is nearly impossible; the other is were. that the familiarity of thermometry has, in The story related above [IJ is true in many cases, dimmed people's appreciation of all of its elements. It points up several the desirability of characterizing the features of the tale we want to tell in the accuracy of a given temperature measurement coming pages— temperature sensors do fail, and of the possible importance of that in spite of careful manufacture, selection, accuracy to the problem at hand. and calibration; these failures are costly, both in dollar and in human terms; and it is in part the job of the thermometry program at the NBS to consider whether such

3 There are other questions which we 2.1 Conceptual System of Temperature shall address as we explore the manifold components which make up the national mea- Temperature, like the quantities mass surement system for contact thermometry. and length, has a firm physiological basis, Among these are whether the people who want so that the history of thermometry reaches to measure temperature have available to to the ancient Greeks [1]. By the beginning them the thermometric equipment and the of the 17th century, Galileo had demonstrated accuracy required for a successful solution an air thermometer, and before that century to the problem at hand, what role the NBS ended, there were calls for thermometric can play in solving difficulties which arise fixed points and for standardization! in those situations, and whether adequate , a Pole, and , a Swede, mechanisms exist by which present de- were among the early contributors to ficiencies and future needs in the measure- temperature scales in the region between ment system can be evaluated and overcome ice and steam temperatures. through the NBS's efforts. Efforts to discuss temperature and thermometry in terms which are scientifical- ly recognizable today resulted, in the 19th 2. STRUCTURE OF THE century, in the definition of temperature NATIONAL TEMPERATURE MEASUREMENT SYSTEM as "the thermal state of a body with re- ference to its ability to communicate heat In this section the reader will begin to other bodies" by Maxwell [2]. Lord to see the awesome proportions of the U.S. (then William Thomson) placed the evaluation thermometry community. This giant takes on of temperature (or "thermometry) on a sound some of the aspects of a bustling city, with thermodynamic basis by considering the its own map, currency, banks, factories, efficiency of reversible heat engines [3]. emergencies, and policemen. This method, which is based on Carnot's The reader can only find his way through concept of heat engine cycles, permits the the community with the aid of maps--the definition of a thermodynamic zero of basic scale of temperature and the various temperature. thermometry specifications and regulations Clearly one can define a temperature issued by the General Conference of Weights scale by defining an of and Measures, the American National temperature and by simultaneously assigning Standards Institutes and others of the a numerical value to the temperature at "city fathers". The currency of this realm which a particular natural phenomenon is composed of temperature measurements at occurs. Alternatively, he can simultaneous- all levels of accuracy. The "really big ly assign numerical values to the temper- money"--the highest accuracy, International atures characteristic of two separate Practical Temperature Scale measurements-- phenomena [4]. Either technique defines the is rarely seen outside of the banks (the size of the unit of temperature and major calibration laboratories), but no one evaluates the scale at two temperatures, thus doubts that the city needs the "big money" leaving the temperatures at which all other just as much as it needs the small change phenomena take place as quantities to be de- of less precise measurements which is termined by experiment. readily seen throughout the city's factories While the science of thermodynamic and homes. temperatures was slowly advancing, the We all live in this community, we buy technologist busily set up temperature thermometers at home or at work, we measure scales on which the available natural temperature more or less accurately, we phenomena of melting ice, boiling water, complain when we are "shortchanged" by un- body temperature and the like could be necessarily inaccurate or expensive evaluated. As we mentioned above, thermometry and we exult when a shiny new Fahrenheit and Celsius were involved in coin of a more useful or less expensive this work, although they were certainly not measurement comes into circulation. For alone. As one might suppose, the scales the next few pages, then, let us take a which were proposed were suggested for quick tour through Thermometry, USA--our diverse reasons, and their evolutions were own hometown! not necessarily predictable. These efforts resulted in the Fahrenheit scale, chosen so that 100 degrees separated human body temperature from that of a cold winter day, and the Centigrade scale, which was defined so that one hundred degrees separated water's freezing and boiling points at 0 °C and 100 °C, respectively. The Fahrenheit scale was in use in the 18th century, but the temperature in , and Jq = 273.15 K [9]. Centigrade scale was not produced until 1850. We should note here that the satisfac- The thermometric principle of defining tion which thermometricians feel in knowing the temperature scale by assigning values to that International Practical [10] Temperature the temperatures of two natural phenomena Scale (IPTS) temperatures are (potentially, at (in this case the Centigrade scale assign- least) completely equivalent to thermodynamic ments) was chosen by the 31 -member Seventh temperatures must be tempered by the know- General Conference on Weights and Measures ledge that advances in thermometric research in 1927 as the basis for the first Inter- will likely result in periodic adjustments national Temperature Scale. (ITS). The prac- of the scale values, with consequent ad- itcal aim of the General Conference in intro- justments required in "science and ducing the Scale of 1927 was to provice a con- industry" [11]. venient reproducible scale for general use. The latest revision of IPTS [12], in The achievement of thermodynamic accuracy 1968, continues the practice of assigning was a difficult experimental problem, and values to a number of triple points, boiling the General Conference hoped to enhance points, and freezing points. The usefulness work-a-day thermometric accuracy by de- of these defining fixed points lies in their fining the scale in terms of platinum ease of use as check points against which a resistivity, thermocouple emf, and optical given thermometer may be calibrated. In pyrometry rather than in terms of gas fact, few of them are easily used [13], and thermometry or other techniques capable of consequently their practical value is thermodynamic accuracy. This idea appears diminished. Recently some progress has to have been quite a good one [5]. been made in the "fixed point business", Because of the insight which it gives however; this will be discussed in Section 4. into the nature of thermometry, the ITS of 1927 is included as Appendix C. We might note that the Scale was based upon a 2.2 Basic Thermometry Infrastructure proposal jointly presented by NBS, NPL (Great Britain) and PTR (Germany), and that 2.2.1 Documentary Specification System the U.S. announcement was written by G. K. Burgess during his tenure as Director of 2.2.1.1 Standardization Institutions NBS [6]. The General Conference considered There are five hierarchies of standards that the ITS needed its own group bodies which are concerned with thermometry. of specialists, so an Advisory One of these is headed by the General Con- Committee on Thermometry was formed ference on Weights and Measures (C.G.P.M.), in 1933 to address these problems directly. and it is devoted to the elucidation of the The Advisory Committee prepared a new (and, Kelvin Thermodynamic and of the Inter- above 630 °C, an improved) version of the national Practical Temperature Scales. A ITS in 1948 which the International second hierarchy generates many con- Committee on Weights and Measures recommended sensus standards for use in commerce, to the Ninth General Conference. The new industry, and other "public" situations, scale was adopted; it retained the definition ultimately leading to the International of the melting and boiling points of water Standards Organization and the International as 0 °C and 100 °C, although the name of the Electrotechnical Commission. Thirdly, there unit was changed to Celsius to honor the are various professional and manufacturing pioneering Swede [7]. Furthermore, the associations. A fourth hierarchy is comprised thermodynamic scale was given the name of of U.S. regulatory agencies; and finally, Kelvin and defined by setting an absolute there are government and commercial standards zero of temperature and assigning the value laboratories. (We should note that, al- 273.16 K to the triple point of water [8]. though our own concern is temperature and The awful truth that the practical thermometry, each of these hierarchies scale unit (°C) differed from the thermo- considers many other units as well J The dynamic unit (the Kelvin) resounded first, fourth, and fifth of these hierar- in the world's halls of state for twenty chies are primarily concerned with physical years. In 1968, however, the 13th General measurements per se and are discussed in Conference at last adopted identical bases Section 2.4. The remaining two are de- for the two scales. Henceforth, the scribed here. Kelvin would be defined as 1/273.16 of the temperature of the triple point of water, a) Commercial Consensus Standards and the degree Celsius was given an equal value by defining Celsius temperatures as The measurement of temperature in com- t = T - Tq, where T is the thermodynamic merce and industry has been a long-time

5 concern to professional associations and to countries accept them as legally binding for organizations which have arisen solely to national and international trade. These promote standardization. The latter group bodies are the International Standards has achieved wide acceptance because it Organization [14], the International Electro- encourages the participation of technical Commission, and the International manufacturers of thermometric apparatus, Organization for Legal Metrology [15]. The users of such equipment and objective third American National Standards Institute [16] parties; we shall discuss this latter group provides the U.S. representation to the fi rst. ISO and lEC, and it, in turn, derives much Three international bodies are concerned of its detailed recommendations from the with standards used in commerce, and their American Society for Testing and Materials recommendations are significant because many [17]. Figure 2.2A shows a sketch of these organizations.

lEC ISO OIML

International Electro- International Organization for International Organization technical Commission Standardization for Legal Metrology

1906 "standardization 1946 (1926 ISA) Federation of 1955 (Aug. 1972, U.S. of electrical apparatus" 54 Nat. Stds. Inst, "quality, joins) "recommends uniform production, trade." int'l. requirements to U.S. National Committee promote int'l. cooper- Member Bodies - for U.S. - ANSI to lEC - ANSI ation"

NBS Rep. by C. H. Page - Int. Conf . of Legal Met. Standing and Tech. Com. Standing and Tech. Com. - Int. Bureau of Legal Met and Subcommittees and Subcommittees lint. Committee of Legal Met. American National Standards ^ Working Groups Institute

Selects "American National American Society for Testing Standards" and Materials Board of Stds. Review develops "consensus" standards Company Member Council Twenty-four Member Body Council Board of Directors Consumer Council Technical Ad. Bds. H Technical Standing Committees Technical Committees -{Professional Associations

FIGURE 2.2A. Sketch of the structure of the lEC, the ISO, and OIML,

The work of these standards groups is to dictions", so that each may more or less develop standards for all manner of devices, simultaneously write conflicting recommenda- test methods and test equipment. The lEC tions regarding essentially identical and ISO seek agreements which will facili- materials or instruments. W. A. McAdams tate international trade--in fact, some Director of ASTM, pointed out recently that nations (most recently Denmark) [18] auto- the writing of international standards is matically accept these international more difficult in cases where the individual Standards as legally binding within their member nations have international standards, own boundaries. The United States has re- since these generally conflict in their cently joined OIML, whose function is to details; McAdams noted a trend toward recommend uniform metrological requirements, standardization first at the international which relate to legal standards in its level and advised ASTM to "Reexamine its member nations. role and effort in light of these trends" The most difficult problem which these [19]. If, in fact, international standard- groups face may well be that they have ization were to precede national standardiza- rather completely overlapping "juris- tion, rather impressive coordination problems might well develop. Currently ASTM has over Association (NEMA), the Society for Auto- 20,000 members who serve on 115 committees, motive Engineers (SAE), Underwriters and these members would no doubt continue Laboratory, Inc. (UL), the Scientific to be interested in the details of any Apparatus Makers Association (SAMA), and' international standards which concern the the Instrument Society of America (ISA). products of their companies. Also, NBS is involved in commercial 2.2.1.2 Survey of Documentary Standards standards activities, at several levels. Individual staff members serve on committees The actual documents produced by the and subcommittees of both the U.S. and the five hierarchies of standards bodies show a international bodies; a partial listing of great variety both in form and in number. such service is included as Appendix D. The General Conference of Weights and Measures has issued but three basic docu- b) Professional and Trade Associations ments in fifty years. These documents, however, are the three editions of the The second main category of voluntary International Practical Temperature Scale standards bodies comprises the professional and they represent the final authority in associations. This category can be further the field of thermometry. Table 2.2A shows subdivided into scientific associations and the primary fixed points on the IPTS-68, commercial associations. Among the former and the full text is noted in reference 12. group may be included the International Union The standards issued by the American of Pure and Applied Physics (lUPAP), the Society for Testing and Materials are the American Institute of Physics (AIP), the product of consensus agreements among manu- International Institute of Refrigeration facturers, users, and objective third par- (IIR), the National Conference of Standards ties, as we noted earlier. These standards Laboratories (NCSL), and the National are issued annually in a multivolume Book of Conference on Weights and Measures (NCWM). Standards published by ASTM from its offices The latter group includes (but is certainly at 1916 Race Street, Philadelphia, Penn- not limited to) the American Society for sylvania 19103. Many of these standards are Mechanical Engineers (ASME), the Electronics selected by the American National Standards Industries Association (EIA), the Institu- Institute to be designated as American tion for Electrical and Electronic Engineers National Standards. Standards so selected (IEEE), the National Electrical Manufacturers are noted in the ASTM listing and the

Table 2.2A

Primary Fixed Points on IPTS-68

State T (K) t (°C) J 68 thermodynamic zero temperature 0 -273.15

equilibrium hydrogen triple point 13 81 -259.34 equilibrium hydrogen, p = 33,330.6 N/m2 17 042 -256.108 equilibrium hydrogen boiling point 20 28 -252.87 neon boiling point 27 102 -246.048 oxygen triple point 54 361 -218.789 oxygen boiling point 90 188 -182.962 water triple point 273 16 0.01 water boiling point 373 15 100.00 zinc freezing point 692 73 419.58 silver freezing point 1235 08 961.93 gold freezing point 1337 .58 1064.43

7 .

information which they contain then forms 64% Fe, the alloy has a minimum coefficient the basis for possible standards consider- of thermal expansion and is called Invar. ation by the U.S. National Committees to the The alloy containing 18-27% Ni , on the other International Electrotechnic Commission and hand, has a coefficient some twenty times to the International Organization for Legal larger, so that the bonded pair forms a Metrology. sensitive, stable mechanical thermometer Professional associations are not capable of 0.03 °C precision in the range uniform as regards their issuance of 0-100 °C [21]. standards. Many of them undertake very few Similar in principle to the constant- standardization activities, whereas volume gas thermometer is the liquid- or others (for example the National Committee 1 iquid-and-vapor-fi 1 led pressure thermometer. for (Medical) Clinical Laboratory Standards) In this device a capillary tube connects offer substantial standardization guidance the sensing bulb to a pressure gauge or to a to their members. pressure-sensitive control device. The The temperature measurement standards filled thermometer, although capable of only issued by government agencies are in the moderate accuracy, is commonly used because so-called Military Specifications handbooks the pressure gauging and control is conve- and in the particular calibration handbooks nient and reliable. issued by the individual calibration Resistance thermometry came into common laboratories, as noted in the previous use late in the 19th century with the de- Section velopment of high quality galvanometers. H. L. Callendar developed thermometric 2.2.2 Thermometry Instrumentation System elements of platinum wire and derived specifications for evaluating the quality 2.2.2.1 Temperature Measurement Tools and of thermometers [22]. By 1927, much of the Techniques ITS could be referred to platinum resistance thermometry, and presently it is possible Many schemes for evaluating temperature to obtain measurements of the highest pre- have been used since Cro-Magnon yanked his cision from 20 K to perhaps 1300 K by using fist out of the fire and grunted at his platinum resistance thermometers [23,24]. mate to fetch some cold water. Not so many, In addition to platinum metal resistiv- however, that we cannot mention some of ity, a variety of other resistive thermom- those in common use today. Galileo's air eters has become useful over the years. A globe, inverted over a dish of water, was the brief listing of these is included in forerunner of today's gas thermometer, which Appendix E, since some of them will be is probably the most useful of the "thermo- referred to in later sections. dynamic" thermometers in the temperature Thermoelectric thermometry, on which is range 20-600 K. These instruments are based based the widely used thermocouple, depends on the ideal gas equation PV = nRT; commonly on the Seebeck, Peltier, and Thomson either constant-volume or constant-pressure effects [25]. The Peltier effect is the ratios are taken at two temperatures, with liberation or absorption of heat accompany- the final temperature, J 2> evaluated by the ing current flow through the junction of product of the filling, or initial temper- dissimilar metals. The Thomson effect ature, Tj, and the pressure (or volume) describes the reversible production of ratio heat accompanying the flow of current along T2 = Ti(P2/Pi) or Ti = Ti(V2/Vi). a homogeneous conductor whose temperature The trick is to find an ideal gas and to varies from one point to another. Seebeck 's avoid "pitfalls" [20]. observation that current will flow in a Fahrenheit and Celsius developed circuit consisting of dissimilar wires thermometers, which rely on the different connected in a loop in which the junctions thermal expansion properties of mercury and have unequal temperatures can be interpreted glass. Although the relative instability in terms of the Peltier and Thomson effects. of the glass limits their accuracy to 0.01 °C The overall circuit emf can be described by to 0.03 °C, such thermometers enjoy two Peltier coefficients at the junctions more extensive use today than any other and two Thomson coefficients in the two type. This situation is due to their low dissimilar wires. cost, their simplicity in construction and Since its discovery 150 years ago, use, and, for moderate accuracy work, their thermoelectric thermometry has come into reliability. These characteri sties apply extremely wide use. Platinum versus 90% equally well to another commonly used platinum- 10% rhodium is the standard in- expansivity thermometer, the bimetallic terpolating instrument of IPTS from 630 °C- element. These elements ordinarily are 1064 °C, and other thermocouples have been bonded strips of nickel steels; at 36% Ni adopted for many temperature measurements 8 and control applications. Table 2.2B shows the ASTM letter designations for thermo- couples [26].

Table 2. 28

*70 Thermocouple Characteristics, from ASTM CO on Symbol Elements (+/-) Useful temperature range

B Pt-30% Rh/Pt-6% Rh 0 to 1820 °C E Ni-10% Cr/constantan -270 to 1000 °C J Fe/constantan -210 to 1200 °C K Ni-10% RH/Ni-5% (Ai,Si) -270 to 1372 °C R Pt-13% Rh/Pt - 50 to 1768 °C S Pt-10% RH/Pt - 50 to 1768 °C T Cu/constantan -270 to 400 PC

By virtue of the radiation given off by There are many other methods by which a hot object, one can measure its temperature temperature can be measured in particular with a radiation pyrometer. The radiation ranges and with particular accuracies instrument in most common use at present (especially with low accuracy). These is the disappearing filament optical methods are listed in Appendix F, since pyrometer. In its use, the current through they comprise much of scientific and a filament is adjusted until its luminance industrial thermometry. A composite of matches that of a glowing object (usually temperature sensors appropriate for various at a wavelength near 650 nm). At that point, temperature ranges is given in Table 2.2C. the image of the filament blends into that of the object; the filament current control displays the measured temperature.

Table 2.2C

Sensors in Common Use at Various Temperatures Temperature Range Sensor (Approximate) Limiting Accuracy

0-20 K Semiconductor resistance ±1 mK

20-1000 K Platinum resistance ± 1 mK (at 20 K) to ±12 mK (at 1000 k) 20-1300 K Thermocouples ±10 mK (at 20 K) to ±200 mK (at 1300 K) 250-400 K Mercury-glass ±30 mK 250-400 K Bimetallic ±30 mK 250-500 K Thermistor Uncertain 1000 K Optical Pyrometer ±0.5 K to 20 K

The IPTS-68 designates the Planck the fixed points. These are shown in figure radiation law, the platinum-10% rhodium 2.2B, along with other sensors which, while platinum thermocouple, and the platinum re- not standard instruments, have come into si stance thermometer as the standard inter- common use because of a combination of low polating methods for temperatures between price and reasonable reliability.

9

. .

As we noted earlier, the triple point of proper account of the influence of the water is of central importance to thermom- measurement equipment used when considering etry; the definition of its temperature as the reliability of a particular thermometry 273.16 K (plus the thermodynamic zero of system. temperature) sets the "size" of the unit Much more remains to be said on the sub- temperature interval, and it thus provides jects of temperature and thermometry than the scale on which the thermodynamic temper- the trifle which comprises Section 2 of atures of all other natural phenomena must this study. However, we have attempted to be measured. The most common means of present here a concise discussion of the realizing the temperature of the triple scale of temperature as it is derived point of water involves the use of a cell scientifically and as it is defined in a like that shown in figure 2.2C. practical way, and enough information on the The particular cell shown is designed for evolution and practice of thermometry so use with standard platinum resistance that one can appreciate the diver- thermometers. When the ice mantle is sity and vitality of the activities to be correctly formed around the central tube described in the balance of the study. and the cell is surrounded by a mixture of crushed ice and water, the cell will furnish 2.2.2.2 The Thermometry Instrumentation the triple point temperature very pre- Industry cisely, typically the standard deviation of the platinum thermometer readings at the The temperature sensors and associated triple point is less than 0.14 mi 1 1 idegrees measuring equipment which form the basis of The simplicity of the triple point of the thermometry instrumentation industry water cell, coupled with its excellent have been enumerated to a large extent in thermometric reproducibility, makes it ideal the preceding Section. In turn, this equip- as the principal defining point for the ment forms the basis for measurement and temperature scale. All but the most modest control of a very great amount of processing thermometry facilities can readily incor- equipment, monitoring equipment, and durable porate such a cell for ready calibration of consumer goods--examples of these are the temperature sensors. control equipment for the fractionation of Although no other fixed point device is petroleum, the temperature monitors on the quite so easily used as the water triple coolant loop of a nuclear power plant, and point cell, many others, especially metal the several thermostats in a household freezing point cells, are in common use. cooking range. In this Section, however, The uses of such cells in thermometry are we want only to describe the primary described in the excellent NBS Monograph 126 thermometry instrumentation industry itself. "Platinum Resistance Thermometry", by J. L. This group of people provides the means Riddle, 6. T. Furukawa, and H. H. Plumb, by which the National Temperature Measure- from which figure 2.2B was abstracted. ment System makes the bulk of its day-to-day In practice, the sensors described in temperature measurements. It includes at this section mostly require auxiliary equip- least 25 companies with assets greater than ment: the optical pyrometer requires, in $1 million. Very few indeed are the people general, a calibrated lamp or voltage who construct their own thermometers and divider; the thermocouple requires a re- temperature control apparatus from scratch ference junction and, depending on the (but these few are responsible for nearly sensitivity needed for the particular all of the advances in the field). application, either a millivolt - or a Much routine thermometry is accomplished microvolt-range meter; and the platinum and using thermocouples, (however, resistance other resistance thermometers require thermometry, because of its higher signal- Wheatstone or other bridge circuits or an to-noise ratio, is rapidly gaining favor). equivalent means (such as a current- Many manufacturers provide preassembled potential measurement system) for measuring couples in various configurations [27], al- their resistances. By and large, only the though most people use thermocouple wire liquid-in-glass and the bimetallic thermom- provided with the appropriate calibration eters are free of this need. table to construct simple thermocouple As a result of the need for auxiliary configurations equipment, the performance of practical The existence of ASTM and other stan- thermometry generally reflects the quality dards for commercial devices will be noted of a system, comprising the sensor plus its later. These standards are quite effective associated equipment. For this reason, we in promoting uniform quality in thermocouple must concern ourselves to some extent with thermometry since they allow the user to the products and status of the electronics predict the uncertainty in a given temper- industry, and we must be careful to take ature measurement, and, usually, in a

11 particular control application as well. The great majority of manufacturing pro- cesses involving thermocouples require only coarse (perhaps 1%) control of temperature. Typical of such applications are heat treating furnaces; steam, hot water and hot gas plants; and bearing temperature monitor- ing. The automotive, power generation, metal fabrication, and motor manufacturing industries all employ these applications and use thermocouples in large quantities. In general, manufacturers of thermo- couples and thermocouple wire assure their own conformance to industry standards by using either thermocouples or platinum resistance thermometers which are routinely calibrated at NBS or at a secondary facility. In addition, emf-temperature tables (recently corrected by NBS to conform to IPTS-68) are available for all the common couples. The construction of automatic reading and recording devices and of temperature control devices for use with thermocouples is a major thermometric industry. Several major companies [28] provide recorders and meters which express common thermocouple outputs directly in temperature units, and these firms also make available control circuitry which will activate heaters, coolers, temperature limit devices, and the like in order to automate processes which involve processing at a particular temper- ature. The quality control exercised in this industry is often quite thorough, with individual checks on temperature accuracy (in addition to similarly careful checks on mechanical and electrical integrity) using high quality baths and measuring equipment [29]. We have found this industry, taken as a group, to be very conscious of its responsibilities and to be grateful for the active support of NBS through calibration services and consultation. Bimetallic thermometers, too, are widely used in thermometry and in temperature FIGURE 2.2C. Water triple point cell control applications [30]. One common household thermostat employs two bimetallic A. Heavy black felt shield against ambient thermometers; one for visual indication radiation. and the other in the control circuit [31]. B. Polyethylene tube for guiding the SPRT Automotive choke and radiator thermostat into the thermometer well. assemblies also employ bimetallic temperature C. Water vapor. control units. Calibration of bimetallic D. Borosilicate glass cell. units is generally rather coarse, due to the E. Water from ice bath. nature of the major applications. F. Thermometer well (precision bore). The fastest growing of the temperature G. Ice mantle. sensor industries appears to be the thermis- a H. Air-free water. tor industry. The thermistor* basically finding wide application in I. Aluminum bushing with internal taper at semiconductor, is upper end to guide the SPRT into the the medical area due to its small size and close-fitting inner bore. high sensitivity [32]. The same qualities J. Polyurethane sponge. make it a desirable sensor in many other K. Finely divided ice and water. areas, among which are automotive engine tem- peratures, surface temperatures of manifold systems, micrometeorology, process control, processing and health care are generally and air conditioning sensing and control. calibrated to about ±0.1 °F, and quite The upper limit for good stability in often the calibration is traceable to NBS. thermistors is about 300 °C. This industry This is not generally true of the household is made up of many small firms, so that re- unit production, however [39]. latively little research has been done to Optical pyrometers are extensively used characterize the response, stability, and in many industries. Subject to inaccuracies interchangeabil ity of these sensors. owing to uncertain emissivity corrections Nevertheless, much automatic reading and as well as to differences among particular control equipment is available for use with operators, optical pyrometers still are the thermistors. Some $10-20 million worth of thermometers chosen in cases where contact thermistor units are produced in the U.S., thermometry is especially difficult or where and perhaps one-fifth as many are imported radiation measurements are convenient and from Holland or Germany [33]. Since these reliable, as in surface thermometry. units are priced at perhaps 10(t each but Many other thermometer sensors are manu- serve much more expensive measuring and factured for particular applications, and control equipment, the thermistor temper- control devices are constructed for their ature control industry is a large one indeed. employment. Our experience has been that, Some notion of the rapid progress in the taken as an industry, thermometer and thermistor field can be gained from a recent temperature control equipment manufacturers note [34] that a -5 °C to 45 °C field employ relatively sophisticated personnel thermometer system has been developed for and equipment in order to assure the quality use with a voltage-to-frequency converter of their products in competition with other circuit employing a portable counter. suppliers of these units. In addition, some Another area of rapid growth is in of the individual manufacturers support miniature resistance thermometers [35]. thermometry research and development efforts The accuracy, sensitivity, and stability on a continuing basis. This activity has of standard platinum resistance resulted in the recent development of such thermometers has made them preeminent for advances as cryogenic temperature control- exacting thermometry up to 630 °C, and lers utilizing semiconductor temperature commercial thermometers are now made in sensors [40], liquid crystal sensors said many configurations, including some of to be sensitive to changes of 0.1 °F [41], millimeter dimensions. We have at present and radiation sensors which measure temper- limited data [36] on which to judge the atures of millimeter-size areas for trouble- extent of any sacrifice in sensor quality shooting electronic circuitry [42]. resulting from miniaturization or other A further measure of the instrumentation changes; however, a recent advertisement industry is found in the Standard In- offers 0.1 °C resolution and ±0.5 °C dustrial Classification Manual. (See accuracy (interestingly, the accuracy Appendix B). The following codes refer to specification is correctly defined as a thermometry: "total system" accuracy, a very encouraging 3822 Measuring and Controlling Instruments sign) [37]. 3823 Industrial Instruments for Process Liquid-in-glass thermometers are widely Variables used for temperature measurement; once 3829 Measuring and Controlling Devices, filled and calibrated, only column separa- Other tion and breakage cause obvious changes in These codes are further broken down into the response of these thermometers. Bumping subcategories for which meaningful economic or temperature cycling do, however, cause statistics are gathered by the Bureau of minute changes in bulb volume which are the Census. Representative entries follow: detectable in calibrations. Because of the (current Industrial Reports, 1972) relative delicacy and expense of control 38230 Electrical and Electronic Thermometers, designs based on liquid-in-glass, however, 38 companies with 1972 shipments exceeding most uses are found in temperature monitor- $100,000; total shipments $40 million. ing applications such as batch chemical 38230 Mechanical Thermometers, 40 companies processing, health care, and household with 1972 shipments exceeding $100,000; thermometry. There are several large total shipments $39 million. manufacturers of liquid-in-glass thermometers. 38230 Primary Thermocouples and Leads Wire, They sell some $5-6 million worth of ASTM 6 companies with 1972 shipments exceeding models, as noted earlier, perhaps $7-8 $100,000; total shipments $11 million. million worth of precision and research 38230 All other Primary Thermometers, 4 thermometers, and roughly the same value companies with 1972 shipments exceeding in student-group thermometers [38]. The $100,000; total shipments $10 million. liquid-in-glass units intended for chemical 38220 Thermostats for heating and cooling. 11 companies with 1972 shipments exceeding user, and standards groups suggested (both $100,000; total shipments $287 million. as groups and in individual correspondence) 38220 Temperature Controls for Appliances, appropriate temperature ranges and tabular 14 companies with 1972 shipments exceeding intervals for which reference data for each $100,000; total shipments $140 million. of the thermocouple types would be most The 1973 U.S. Industrial Outlook notes useful. In response to these requests, that in 1972 the Automatic Temperature staff members of the NBS (both in the Controls category accounted the following Temperature Section in the Gaithersburg statistics: laboratories and in the Division $700 million shipments in Boulder) constructed the new tables of 105 establishments counted Monograph No. 125. This was accomplished, 38,000 people employed in some cases, by acquiring new basic 4.2% (over million) exported calibration data on wire furnished by , $30 The outlook projected the shipment level manufacturers. In other cases, the of this category at $1 billion annually by calibration files were searched for appro- 1980. priate data which summarized and converted Other statistical data could be garnered to the new 1968 Temperature Scale. In all on the Automatic Temperature Controls cases, computer fitting techniques were category or on other thermometry categories used to obtain smooth tabular values of as well. However, we believe that the thermocouple e.m.f. versus temperature. identification of the sources of such data, These particular thermocouple reference together with representative figures show- tables are sold to the public both in the ing the enormous volume of activity just in U.S. and abroad through the U.S. Government this one facet of thermometry, serves to Printing Office. In an effort to avoid point up its importance to the U.S. economy. "going out of print" in less than five For the main bulk of industrial ship- years, the Printing Office ran off 20,000 ments, such as those categorized above, the copies in the first printing. quality of the included thermometric devices Monograph 125 is now in use by thermo- is controlled by adherence to American couple wire manufacturers, whose Quality Society for Testing and Materials standards Control laboratories run continuing cali- or to similar voluntary or military bration checks to ascertain that the wire specifications. In turn, these resistance as manufactured does yield the tabulated or thermocouple devices are referred emf versus temperature within limits to more accurate standards, and eventually specified by the appropriate manufacturing to the NBS. Unfortunately, neither user standard. The instrumentation industry nor manufacturer is always aware of lapses also uses Monograph 125 to determine that in which the calibration chain has been meter outputs (often given directly in broken, with consequent deficiencies in temperature) correspond to the tabulated the performance of the equipment. values. (We might note here that this industry specifically requested that a 2.2.3 Reference Data supplement to Monograph 125 be issued with the tabular temperatures on the Fahrenheit Much of the most accurate thermometry Scale, for those of their customers not is preserved on standard thermometers, as yet able to use the metric Celsius scale. we shall see in the next Section. Of the NBS complied with this request within most heavily used thermometry, however, one year.) Major users of thermocouples use the majority is preserved in the form of Monograph 125 in the calibration of thermo- reference tables. A notable example of couples employed in test stands, process these data is a recent NBS publication. control instrumentation, and materials data Monograph No. 125 "Thermocouple Reference eval uation. Tables Based on the IPTS-68". A short Thus we see, in the history of one docu- discussion of the history of this publica- ment, the manner in which a technical need tion may illuminate the process by which is observed and answered by the NBS staff information becomes Reference Data. in generating Reference Data in thermometry. In discussions with manufacturers and Similar Reference Data are now in the users of thermocouples and with the process of being obtained at NBS for small, corresponding American Society for Testing commercial thermistor and resistance and Materials committee over a period of thermometers. Some data obtained by years, it became clear that the thermocouple various industrial representatives are in industry relies most heavily on seven types existence but their value is restricted of thermocouple thermometers. With the since the interchangeabil ity , stability and advent of the International Practical Tem- accuracy limitations of the corresponding perature Scale of 1968, these manufacturer. sensors are not generally available or not

14 known. Such data, when available, will thermometric properties can be utilized play a role similar to that of Monograph throughout the Measurement System. 125. Hardly data in the usual sense of the 2.2.5 Science and People word, the Reference Functions of the 1968 Practical Temperature Scale International In considering the National Temperature constitute a primary source nonetheless Measurement System infrastructure, one must are used in the interpolation of the which first realize that temperature is not of between fixed points of temperature scale direct interest to most people who measure These functions temperature information. it. For the most part, people who measure are found in the journal Metrologia 5 35 temperature want to obtain a reproducible, they are employed with the capsule (1969); readily interpretable result for a clinical standard platinum resistance thermometer laboratory blood test, or they want to en- to obtain values of temperature in the sure that a batch of steel, semiconductors, range 13.81 K to 273.15 K. or plastic meets its specifications, or they want to optimize the efficiency of a steam 2.2.4 Thermometry Reference Materials boiler, of a turbine, or of a chemical pro-

cess . The primary reference materials of It is true that many people are thermometry comprise the pure materials interested in the specific temperatures at used in fixed-point devices and in which various physical and chemical thermometers. For example, NBS Special phenomena take place, and occasionally Publication 260 lists two Standard Refer- they relate these temperatures to ence Materials, high-purity zinc and tin, theoretical derivations; in addition the which are suitable for use in the construc- temperature measurement industry is directly tion of freezing-point cells. These cells, interested in temperature -gev se. On properly prepared and used, may then serve balance, however, the less concerned the to define calibration temperatures on the average person has to be with the measure- International Practical Temperature Scale ment process and the thermometry system of 1968 [43]. The high-purity platinum he is using, the happier he is. For this wire which is used in constructing standard reason, the primary objective of thermo- resistance thermometers also constitutes a metricians and of thermometry equipment thermometry reference material. Similarly, manufacturers is to make temperature measure- the thermoelectric standard Pt 67, soon to ment as accurate, as automatic, and as be designated as a Standard Reference efficient as they can. Material by NBS, is a valuable reference The sciences of Physics and Chemistry material in thermometry. both underlie thermometry--rather , one might In one case, a fixed-point device may better note that thermometry underlies both be considered to comprise a thermometry of these fields and most of the engineering reference material. This is the Supercon- sciences as well! Many of ductive Thermometric Fixed Point Device, the fundamental advances in thermometry have arisen as which has been designated by NBS as Standard scientists reached for new tools with which Reference Material 767—when the device they could probe the mysteries of nature, leads are connected to a simple measuring and it is still true that circuit, the user may rely on the thermo- most professional thermometrists have accompanying metric reproducibility of the included interests in one of a variety of scientific areas. superconductive samples [44]. The scientific underpinning Somewhat less accurate thermometry of thermom- etry to be found in reference materials are under development is a great kaleidoscope of scientific and technical in the Physical Chemistry Division of NBS societies and their journals. The principal publications [45]. These are standards for differential thermal analysis. are Metrologia, the international journal of metrology, the Many scientific and industrial labora- Review of Scientific tories maintain "old faithful" materials Instruments, the Physical Review, and the Journal of Chemical which they employ as thermometry reference Physics. Corresponding materials. Often, however, these materials to these in technical areas are journals such are of uncertain purity or are known to be as Instrumentation Tech- nology and the NBS Monograph impure, so that they enjoy only a very local series. The usefulness. A value of the NBS Standard American Institute of Physics and its member Reference Materiel program lies in the fact societies, the Instrument Society of that their products, being generally of very America, and a wealth of scientific and high purity and careful characterization^ engineering organizations contribute can be readily duplicated. Thus their personnel to the study of thermometry.

15 2.4 Thermometry Dissemination and Enforce- 2.3 Realized Thermometry Measurement Capabilities ment Network

The network of standardization insti- The ease with which precise and accurate tutions has already been examined temperature measurement can be provided in Section 2.2.1. Here we shall note varies considerably with the temperature the dissemination and enforcement institutions. range and with the application of the user. The most-used range of temperature is 2.4.1 Central Thermometry Standards the range 0-200 °C; this is where most Authorities human activities take place. The reader will appreciate that these activities in- a) International organizations con- volve far and away the most numerous cerned with thermometry--The General Con- measurements of temperature of any part of ference on Weights and Measures, including the national system. However, the "quality" its component bodies (the International of thermometry required in these activities Committee for Weights and Measures, the is, for the most part, adequately met by International Bureau of Weights and Measures, the cheapest and least accurate sensors available— the humble mercury-in-glass the Advisory Committee on Thermometry, and its Working Groups). This body is respon- and bimetallic thermometers and the sible for the construction and promulgation thermocouple. Offering accuracies of one degree or so and precisions of perhaps of the International Practical Temperature one-tenth degree, these are the counterpart Scale. of the "dollar watch"—cheap but adequate. Arising from the treaty of the Meter, There is an aspect of each of the a 17-nation convention meeting in Paris in items in the preceding paragraph to which 1875, were a General Conference on Weights the free-and-easy remarks made there do not and Measures, an International Committee apply. This aspect is that of the on Weights and Measures, and an International professional in each of the areas named. Bureau of Weights and Measures [46]. The It is true that we can live and work General Conference meets at six-year inter- effectively in buildings whose temperatures vals, when it considers agenda items which vary substantially from the norm; however, are introduced by one or more Advisory the professional building designer can do Committees, including (since 1937) an Ad- his job correctly only with the benefit visory Committee on Thermometry (le Comity of thermometry which is much more precise Consultatif de Thermometrie, or C.C.T.). than our own requirements. The same is Table 2.4A shows a sketch of the true of the automotive designer, of General Conference structure: appliance manufacturers, and of the medical profession. The many different applications of temperature measurement means that a great deal of effort is involved in getting a comprehensive overview of the needs for temperature precision and accuracy. The major aspects of thermometry capabilities have been noted already. They appear in summary form in Tables 2.2B and 2.2C and in figure 2.2B.

16 . . .

Table 2.4A

General Conference on Weights and Measures

41 Nations subscribe, Conference meets at about six-year intervals - 14th meeting was October 1971

International Committee on Weights and Measures

18 Nations, elected by the General Conference; it is the executive body; meets at two-year intervals at BIPM.

Advisory Committee on Thermometry International Bureau of Weights and Measures

12 Major National Standards Laboratories Paris, Pare de St. Cloud. and some six individual advisors; 10th Its missions are: Session was May, 1974 at BIPM. USNBS 1. Fundamental scale research. Representative at 10th Session 2. International scale comparisons. R. P. Hudson, Division 221 3. Coordination of measurement H. H. Plumb, Temperature Section techniques Its Missions are: 4. Coordination of physical constants I. Recommend scale improvement based on measurements working group observations. 2. Study thermometric techniques.

Working Group 1 Preston-Thomas, NPL, NBS (H. H. Plumb), IMM; Revisions of IPTS-68 Working Group 2 IMGC, CNAM, NRC; secondary fixed points, simplified techniques Working Group 3 NPL, PTB, NBS (L. A. Guildner); Thermodynamic Temperatures above 100 K Working Group 4 KOL, Swenson, NSL, IMPR; Thermodynamic Temperatures below 100 K Working Group 5 NBS (R. P. Hudson), Japan, NPL; Practical Thermometry Below 30 K

As can readily be seen by a glance at b) U.S. Thermometry Authorities Table 2.4A, the thermometry work of the General Conference and its subsidiary National Bureau of Standards. The NBS organizations is the definition of the provides temperature calibrations based on thermodynamic scale and the promulgation of the International Practical Temperature an International Practical Scale as closely Scale of 1968, and it provides consultative duplicating thermodynamic temperatures as services in thermometry to science, to can be accomplished within the limit of industry, and to standards institutions at practical utility. Within the United States, all levels. the National Bureau of Standards is re- sponsible for coordinating this effort. c) Other National Authorities Much of the actual scale work is done at the NBS, so that such efforts constitute a Many nations maintain primary standards substantial part of the NBS thermometry. laboratories, some of which are also legal A second international standards metrological institutions. Among the more authority is the International Organization prominent national standards laboratories for Legal Metrology and its component bodies are those of England, Canada, Australia, (the International Bureau for Legal France, Germany, Japan, Russia, Italy, and Metrology, the International Committee for The Netherlands. Legal Metrology, and its Secretariats, Technical Committees, and Working Groups). The OIML issues International Recommendations on technical data and instruments which are the concerns of Legal Metrology in member countries. See Section 2.2.1.

17 2.4.2 State and Local Offices of Weights . National Aeronautics and Space and Measures Administration laboratories at Huntsville, Alabama and at Orlando, The state offices of weights and mea- Florida sures for the most part are inactive in • Standards and calibration laboratories thermometry. Exceptions to this rule are maintained by major thermometry in- the states of Connecticut, Massachusetts, strument manufacturers and major and Michigan. These are "Seal States" (any users of thermometry. thermometers sold in the state must have the approval of the State Office of Weights a) Government and Commercial and Measures). The offices in these states Temperature-Cal ibration Faci 1 ities maintain calibration laboratories and ex- amine a great many thermometers each year. A great many facilities exist for the calibration of thermometers and other 2.4.3 Temperature Standards and Testing standard instruments. Of special interest Laboratories and Services are facilities run by various government agencies for the calibration of their own The principal thermometry standards and their contractors' instruments, and laboratories (apart from those described commercial facilities which calibrate in- elsewhere) exist in Federal agancies, in struments for all comers on a fee basis. industrial instrumentation and manufactur- We have examined in some depth the ing concerns, and, in a few cases, in system of calibration laboratories operated commercial testing laboratories. by the U.S. Navy. This system is shown Some of the most prominent of these are: schematically in figure 2.4A. NBS calibrates • Department of Defense calibration instruments which are used in the primary centers, strategically located to or Type I standards laboratories to calibrate serve the Navy, Army, or Air Force periodically all the secondary (Type II) needs laboratory standards. In turn, these are • Energy Research and Development Ad- used to calibrate the 300-odd "calibration" ministration laboratories at Sandia facilities which serve the Navy's own Laboratories, Savannah River and Oak and its contractors' needs. Ridge National Laboratory

NAVY FIELD CALIBRATION LABS U.S. NAVY CONTRACTORS

~ 100 ~ 200

TYPE II ALAMEDA, POMONA, INDIANAPOLIS, NEWPORT, NORFOLK, LABS OAHU. PENSACOLA

METROLOGY ENGINEERING CENTER POMONA. CALIFORNIA

TYPE I EASTERN STANDARDS LABORATORY WESTERN STANDARDS LABORATORY WASHINGTON, D.C. SAN DIEGO

NBS

IPTS

FIGURE 2.4A. U.S. Navy Temperature Calibration Chain.

18 The Army, as well, has a highly devel- Techniques oped calibration system. The calibration "Platinum resistance thermometer (PRT) laboratory at Redstone is the Army Standards certifications are performed by taking Laboratory and it represents the highest measurements at the triple point of water level (primary reference) of calibration (.01 °C), the freezing point of tin standards capability within the Army and (231.97 °C), and the freezing point of zinc provides primary certification service for (419.58 °C). The resistance of the thermom- Army calibration activities in the Con- eter at each temperature is determined with tinental United States (CONUS) and for the a G-4 Mueller Temperature Bridge having Army Calibration Laboratories overseas. resolution to 1 micro-ohm. Army Calibration Laboratories provide CONUS "Platinum-platinum 10% rhodium thermo- calibration service within designated geo- couples are certified by taking measurements graphical areas: Anniston Army Depot, at the freezing point of anitmony (630.7 °C), Alabama; Letterkenny Army Depot, the freezing point of silver (961.9 °C), and Pennsylvania; Lexington-Blue Grass Army the freezing point of gold 1064.4 °C). Depot, Kentucky; Pueblo Army Depot, Measurements are referenced to the ice point Colorado; Sacramento Army Depot, California; and voltages are determined with a potentio- Tobyhanna Army Depot, Pennsylvania; and meter having resolution to .01 microvolt. Tooele Army Depot, Utah. The freezing points of tin, lead, zinc, The Redstone temperature facility is aluminum, and copper are also available for directly dependent upon NBS. The following the calibration of thermocouples. discussion is taken from reference 47, p. 99 "Cryogenic temperature sensors are "Standard certified by comparison with a platinum "NBS traceability is maintained by a resistance thermometer in a stirred bath. group of Myers type platinum resistance Bath media range from ethylene glycol thermometers which are periodically cali- (-40 °C) to salt (500 °C)". brated by NBS. Working standards are re- The range and accuracy of various gularly checked against triple point of Army thermometry capabilities are shown in water cells (.01 °C) to insure that the Table 2.4B [48]. measurement system is not drifting.

Table 2.4B

Temperature Range and Accuracies

Range Accuracy (±) Item/Parameter Conventional S.I. Unit Conventional S.I. Unit

Liquid in Glass Thermometer -38 to 500 °C 235 to 773 °K 0.01 °C 0.01 °K

Differential Hz in Glass Thermometer 0 to 100 °C 273 to 373 °K 0.005 °C 0.005

Platinum Resistance Thermometer -40 to 500 °C 90 to 773 °K 0.002 °C 0.002 °K

Thermocouple -40 to 1450 °C 233 to 1723 °K 0.25% for Pt-Pt 0.25% for Rd to 1300 °C Pt-Pt Rd or 4 microvolts to 1573 °K

Black Body Radiation Source Temperature (see thermocouple data)

Cryogenic Thermometer -269 to -181 °C 4 to 92 °K 0.03 °C 0.03 °K

19 1

Similar calibration operations are main- Table 2.4C tained by the U.S. Air Force, by the USAEC, and by NASA. Of course, the question of NCSL Listing of whether all of these government calibration Commercial Calibration Facilities maintain the levels of facilities actually Primary Secondary accuracy to which they aspire can only be settled by Measurement Assurance Programs (MAP) Aerojet Electro Systems of the sort that are now in progress with Company X X Sandia Laboratory, Redstone Arsenal, Leeds All is Chalmers Manu- and Northrup, and others. Even more facturing Company X X difficult to answer are the questions of American Geophysical and whether the stated levels of accuracy are Instrument Company, Inc. X X required by the needs of the using AVCO-Lycoming X organizations, and whether they actually AVCO Corporation X X obtain the accuracy they need. Bendix Corporation X The National Conference of Standards Boeing Aerospace X X Laboratories has prepared a useful Directory Boeing Wichita X X of Standards Laboratories [49] which lists Burroughs Corporation X X commercial facilities offering temperature Cannon Instrument Company X calibrations for a fee. Table 2.4C contains Colorado Engineering Expt. a listing of these companies. Station X X

The purpose of displaying this list is E.G. and G. , Inc. X to show the extent of calibration activities. Electro-Scientific NBS has contact with this group primarily Industry, Inc. X X through its calibration services and MAP John Fluke Manufacturing activities. The latter determine actual Company X scale transferral accuracy and represents General Dynamics Corporation X X a large undertaking. General Electric Company X Hercules, Inc. X Hewlett Packard Company X X 2.4.4 Regulatory Agencies Honeywel X X ITT Research Inst. X X The principal Federal regulatory agency Leeds and Northrup X X which is involved heavily in thermometry is Clifton (Litton Ind.) X the Food and Drug Administration. This Lockheed-California Company X X agency, like the three "Seal State" Weights LTI Research Foundation X and Measures offices, has the responsibility LTV Electrosystems, Inc. X X for regulating the sale of medical electronic McDonnell Douglas thermometers and for the thermometric aspects Astronautics X X of health instrumentation. Martin Marietta Corporation X X The nature of the activities of the re- National AstroLabs/Metro- gulatory agencies often demands rather quick lonies X decisions on the safety or reliability of a North American Rockwell X X given product. Because of this problem, these NAR, Rocket Dyne Division X agencies occasionally request information Sanders Associates, Inc. X which is difficult for NBS to supply on a Singer General Precision, short-term basis. A recent case in point is Inc. X that of medical electronic thermometers, which Sperry Electronic Tube come under the purview of FDA; to evaluate Division X the quality of particular instruments (let SSCO Standards Lab X alone the evaluation of the quality of the Teledyne McCormick Selph X manufacturing, which was the information Teledyne Systems Company X X actually desired) appeared to require a TRW Systems Group X X substantial amount of time, if NBS were to Tucker Electronics Company X conduct an orderly, statistical investigation. Varian Associates X Yet FDA felt constrained to quickly address Volumetrics X X the question of the reliability of these Westinghouse Electric thermometers, since the public safety was in- Company X X volved. Similar situations, of course, can be expected to recur in this and other areas, and the problem offers a challenge to NBS to aid these agencies effectively.

20 .

As de facto regulatory agencies, the 4 - essential College of American Pathologists and the D. In the upper right corner, the American Hospital Association participate adequacy of the activity is indicated. in the accreditation of hospitals through 0 - entirely adequate the certification of the quality of their 1 - could be improved thermometry capability. By and large, this 2 - marginal certification procedure comes down to 3 - serious deficiencies evidence of treaceability to the NBS. 4 - really unsatisfactory E. In the lower right, the primary 2.5 Direct Measurements Transactions Matrix type of thermometry activity supplied is for Thermometry coded according to the letter equivalents in Table 2.5A. 2.5.1 Analysis of Thermometry Suppliers and Users 2.5.2 Highlights of Thermometry Users

In order to attempt a codification, or The major users of thermometry, as evaluation of the thermometry activities of might be expected, are the manufacturing the U.S. in summary form, it is necessary concerns of categories (4) and (14), al- to categorize the various components--as though all of the categories listed in Table indeed we have done in the preceding Sections. 2.5A use two or more of the letter-coded For the purposes of this Section, we thermometry goods and services. shall tabulate fifteen sectors of the The number of classes of such thermom- thermometry conmunity in an order which etry users is so large, as we noted in the is common to all portions of the NBS Measure- introduction to this study, that we can ment System study. In this Table 2.5A are scarcely hope to list them, let alone listed representative entries, their discuss their individual needs. Let us principal roles as suppliers of thermometric rather submit that in the general case, sensors, instruments, data, or information, their requirements for accuracy and and all the corresponding roles as users. precision are not high and are adequately We have also indicated, where appropriate. met by available thermometry. Standard Industrial Classification codes Ultimately, of course, even the drug which apply to these entries. store thermometer manufacturer relies on NBS In addition to Table 2.5A, describing for the integrity of his product. One such the various supplier and user roles in the supplier referred to NBS as "the heart of National Temperature Measurement System, (his business). We could not function with- we have attempted to construct a matrix out NBS" [50]. The reason that this is so describing the transactions which occur is that both voluntary standards and most between supplier and user groups. In this legal requirements refer specifically to Direct Measurements Transactions Matrix, NBS as the final authority in matters of Table 2.5B, groups described in Table 2.5A temperature. This situation leads to a are listed in columns 1-25 in their roles continuing need for NBS to maintain calibra- as users of thermometry, and in rows 1-25 tion services for the various sensors and as suppliers. The code to the entries is as ranges in common use. fol lows: Besides its role as an "invisible" A. In the center of each matrix partner in much of our daily lives, temper- element, the numbers indicate the magnitude ature figures more prominently in many other of the activity between supplier and user areas. It is the life-blood of the temper- None - trivial ature controls industry, of course, and it

1 - minor plays a strong role in the broader classifi- 2 - moderate cations of general controls and of precision 3 - important instrumentation through the frequent pre- 4 - major sence of temperature stability or temperature B. In the lower left corner, any change regulation requirements. The petroleum in- in the activity level is shown. dustry is the biggest single user of liquid- N - declining in-glass thermometers, whose sales volume is 0 - stable $5-6 million in ASTM-specified models alone. 2 - growing The chemicals industry, steel and automotive 4 - growing explosively equipment manufacturing (in fact, the general C. In the upper left corner, the level class of manufacturing) all feature multitudes of importance of the activity is indicated of carefully-maintained standards and quality 1 - purely convenience control laboratories which uniformly include 2 - strongly desirable temperature standards. Power companies; 3 - no real alternative the aerospace industry; the Department of

21 Table 2.5A

The Thermometry Community and Its Roles as Suppliers and Users of Thermometric Goods and Services

Key to Supplier/User Entries A - Thermometric Reference Data and Compilations of Such Data B - Temperature Calibrations C - Temperature Sensors and Associated Instruments D - Thermometric Instruction and Consultation E - "Thermometry Needs" Information F - New Thermometry Methods G - Thermometry Texts and Literature H - Temperature Control Devices J - Temperature Standards and Regulations K - Money for Thermometric Research and Development L - Durable Goods Incorporating Temperature Control

Sector Suppl ies Uses

(1 The Knowledge Community Scientific Organizations AD A E F C Academic Institutions G F Professional Societies H Technical Publishers L SIC Codes: 2721, 2731, 2741, 7391, 8071, 8221, 8231, 8911, 8922, 8999

(2) International Metrological Organizations General Conference of Weights and Measures; A B A Standards Laboratories of England, France, Australia, Holland, Italy, Japan, USSR, Canada

(3) Documentary Specification Organizations International Electrotechnical Commission, International Organization for Legal Metrology, American National Standards Institute, American Society for Testing and Materials, Department of Defense Military Specification Activities

(4) The Instrumentation Industry Manufacturers of temperature sensors measurement equipment, temperature control equipment SIC Codes: 3822, 3823, 3829

(5) National Bureau of Standards Plant Division, Instrument Shops Division, Electricity Division, Mechanics Division, Heat Division, Optical Physics Division Cryogenics Division, Electromagnetics Division, Time and Frequency Division, Analytical Chemistry Division, Polymers Division, Metallurgy Division, Inorganic Materials Division, Reactor Radiation Division, Physical Chemistry Division, Engineering and Product Standards Division, Electronic Technology Division, Center for Building Technology, Center for Fire Research SIC Codes: 8922

22 Table 2.5A (Continued) Sector Supplies Uses Authorities (6) Other U.S. National Standards This is a null entry for thermometry. and A B (7) State and Local Offices of Weights Measures C D J

(8) Standards and Testing Laboratories Army, Navy, Air Force, National B D A B Aeronautics and Space Administration, C D Energy Research and Development G H Administration, Major Instrument and J User Corporations Calibration Laboratories SIC Codes: 9611, 9661, 9711

(9) Regulatory Agencies U.S. Food and Drug Administration, J K A B College of American Pathologists, C D American Hospital Association G SIC Codes: 9431 B 10) U.S. Department of Defense E F A (except Certification and K CD Calibration Laboratories) F G H J K L SIC codes: 9711

11) Other Government Agencies (except Calibration Laboratories) Energy Research and Development E F A B Administration, National Aeronautics K CD and Space Administration, Health, F G Education and Welfare Department, H J Department of Conmerce, United States K L Department of Agriculture, Interior Department, Housing and Urban Develop- ment

SIC Codes: 9531 , 9532, 9661

(12) State and Local Government (except C D Offices of Weights and Measures H L Essentially a null entry

(13) Industrial Trade Associations Scientific Apparatus Makers' Association, E D J Instrument Society of America, Society for Automotive Engineers, National Electrical Manufacture Association, etc. SIC Code: 8911

(14) Intermediate Industrial Users or Suppl iers Manufacturers of durable goods of all D E A B types (e.g., laboratory furnaces, F H CD household ranges, washers, dryers, K L E F air conditioners, airplanes, automobiles G H and parts); manufacturers of chemicals, J K plastics, fertilizers, etc.; mining and L petroleum companies; transportation companies; industrial and household repair companies; (and many others)

23 Table 2.5A (Continued) Sector Supplies Uses

(14) (Continued)

SIC Codes: Truly too numerous to list!

(15) General Public E • Each of us as individuals rather than 6 H as employees of a particular organization L

Defense, State Weights and Measures Bureaus great numbers of people involved. and other governmental agencies all need to Even among such an erudite group as re- measure temperature to a greater or lesser search scientists (and we do not exclude accuracy. For many of these portions of NBS!), individuals occasionally unnecessar- the NMS, temperature is an important ily degrade the quality of their research parameter which helps govern whether an by failure to apply careful thermometric (or

instrument, a process, or a product will other standard) techniques . Corporate ex- meet specified criteria, and the thermometric amples can be chosen almost at will from activity amounts to using a well-known the ranks of manufacturers of all kinds of sensor in a standard way to ensure adequate commodities; two such examples might illus- results. In other cases, primarily research trate the nature of the problem: in one and development activities, a stable sensor case an experimental laboratory in the does not exist, and the problem is avoided automotive industry only now (and with non- by widening tolerances or the group in- professionally trained personnel) is volved becomes a (probably unwilling) party attempting to base engine design upon to thermometric research. temperature and pressure profiles; in The group of customers for temperature another, a major power producer expressed instrumentation is so broad as to be no oonoem that uncertainty in temperature apparently of little value as a subject for measurement might well be decreasing the discussion--if the reader considers the head- thermodynamic operating efficiency by ing for a moment, he will realize that just several percent, thus wasting dwindling fuel about everybody belongs to it. The automo- supplies. tive manufacturing industry, power manu- Balancing these examples, of course, are facturers, and the petroleum, plastics, others in which thermometry development is chemicals, glass, aerospace, electrical underway to enhance the quality of a pro- appliance, and electronics industries are duct--one corporation, for example, is using temperature as a process parameter. studying the miniature platinum resistance In answer to the obvious question, we hasten thermometer in order to improve its to observe that an important characteristic differential thermal analyzer [51], and a is shared within this group. This large computer firm is active in thermometry characteristic is an attitude toward tempera- in order to make its semiconductor manu- ture measurement as a necessity to be subdued facturing more efficient, requiring ±1 °C as painlessly as possible. Of course, the for 16 hours at 1200 °C [52]. Nonetheless, quality of thermometry which is accomplished it would appear that effective work in this by members of this group varies widely, and area by NBS will be very productive in terms our feeling, based upon conversations with of increased quality of thermometry. many individuals, is that this quality The opportunities for NBS to make con- depends very strongly on the equipment and tributions in this area are varied and information which is at hand. In the case economically impressive. To cite a few of corporate activities, especially, cost examples, a major film maker reported a need considerations also play a large part in for temperature reproducibility at the ten determining the nature of such "incidental" millidegree level and a desire for stability thermometry. at the one millidegree level from 0-65 °C, The significance of this perhaps obvious both presently beyond their capability [53]; observation is that any activity which a kiln requires, for optimum output, con- raises the level of awareness of the utility tinuous monitoring of an air atmosphere and availability of adequate thermometry, or within 5 °F at 1800 °F, again beyond its pre- which raises the quality of thermometry sent capability [54]; most flow measurements available at a given cost, will have far- require temperature calibration, for reaching effects simply because of the accuracy as well as for subsidiary pollution 24 ) ,

DIRECT UJ 80) MEASUREMENTS 38) cc TESTING Labs) Publ.) NATIONAL Z2 TRANSACTIONS U COMMUNITY tn -J et of FEDERAL LOCAL S Educ. Go MATRIX FOR & UJ Gp PUBLIC < & SERVICES E <_) e: o U. OWM's) _1 O oO Stds. R SERVICES Major Maior AND Soc. INSTRUMENTATION U.S.

^MANUFACTURING ORGANIZATIONS oO uo v> INTERNATIONAL ORGANIZATIONS ASSOCIATIONS LABORATORIES S METROLOGICAL AUTHORITIES UJ 1— DOCUMENTARY department GOVERNMENT INDUSTRIAL INDUSTRIAL REGULATORY GOVERNMENT STANDARDS UJ (_> 3: STANDARDS "STANDARDS KNOWLEDGE -•(Science, CIVILIAN AGENCIES AGENCIES ^INDUSTRY ^AGENCIES GENERAL h- C3 HEALTH (excl. Sdefense (excl. TRADE -. STATE (SIC (SIC ?^ 3^ \ 5 7 2 3 1 1 COMMUNITY 2 2 1 2 1 1 1 KNOWLEDGE O 1 3 3 3 2 (Science, Educ. , 4 2 3 , G Prof. Snr. & Publ . A A D A D D ? D 2 INTERNATIONAL 3 3 1 3 1 2 1 METROLOGICAL 2 3 1 3 1 3 ORGANIZATIONS A A A F A 2 J 3 DOCUMENTARY 3 3 2 1 1 1 1 1 ? STANDARDIZATION 1 3 4 1 1 1 1 1 1

ORGANT/ATIONS ? J ? A ?. F ? F

4 INSTRUMENTATION 2 1 2 1 4 4 1 3 1 2 1 2 1 4 1 4 1 2 1 4 1 4 1 4 3 4 3 INDUSTRY 3 2 3 4 4 ? 3 2 4 4 2 3 3 4 4 (SIC Ma.ior Gp 38) 2 C 2 C 2 D 2 H 2 C 2 C 2 H ? H ? H 2 C ? n ? C. 4 C ?

5 3 3 4 2 2 1 2 3 2 3 2 3 2 3 2 3 2 3 2 1 2 1 2 1 1 1 NR'^ 2 2^ 3 4 3 3 3 3 2 3 3 2 2 3 1 A 2 A 2 D 2 A 2 A 2 B ? A 2 A n D 2 D 2 D ? n 6 OTHER U.S. NATIONAL STANDARDS AUTHORITIES

7 STATE & LOCAL 1 1 1 1 1 1 2 1 2 1 2

1 1 1 OFFICES OF 1 1 7 1

WEIGHTS & MEASURES 2 E D 2 D ? 0 ? ,1 ? ,1 8 STANDARDS & TESTING 2 2 2 3 2 3 2 3 2 LABORATORIES 2 3 2 4 3 2 1 1 AND SERVICES D E E 2 B B B

9 REGULATORY 2 4 1 4 3 1 3 1 4 3 ? AGENCIES 1 1 1 2 2

(excl. OWM's) 2 E ? J 2 J 2 ,1 2 ,]

10 DEPARTMENT OF 1 1 2 1

DEFENSE 1 2 1 3 4 1 (excl. Stds. Labs) K 2 E '

11 CIVILIAN FEDERAL 2 1 1 2 1

GOVERNMENT 1 1 2 2 1 3 2 3 1 AGENCIES E D D 12 STATE & LOCAL

GOVERNMENT 1 1 AGENCIES

13 INDUSTRIAL 1 2 1 2 2 1 1 2 1 7 7 TRADE 1 2 1 2 1 1 1 ASSOCIATIONS A E E E D E A

14 INDUSTRIAL 3 1 3 2 3 3 2 1 2 1 2 1 2 1 2 1 2 1 2 1 3 1 2 3 MANUFACTURERS 3 3 3 1 1 1 2 2 1 2 4 2 E 2 E 2 E E E E e E E E A

15 HEALTH SERVICES 2 1 2 1 2 3 2 1 2 1 3 1 4 3

(SIC Major Gp 80) 1 1 3 3 1 1 4 4 E E 2 E ? F E n 2

""^ 2 1 1 GENERAL PUBLIC 2 3 3 2 1 2 2 4 ? 2 n Key to Matrix Entries

C - Importance of Transactions (In)adequacy of Services

1 - Purely convenience - No improvements needed 2 - Strongly desirable - Could be improved 3 - No real alternatives - Marginal 4 - Essential - Serious deficiencies - Out of control B - Rate of Change C D

N - Declining A 0 - Stable 2 - Growing B E 4 - Growing Explosively A - Magnitude of Transactions

E - Thermometry Services Supplied 0 - Trivial

1 - Minor See Table 2.5A 2 - Moderate ? - Unknown 3 - Important X - Not studied 25 4 - Major Blank = 0 measurements, and this is simply not avail- touches virtually every person in the U.S. able; National Marine Fisheries Service, of The quality of thermometry routinely avail- NOAA, has found that a fraction of a degree able governs his health, his comfort, and change in water temperature can "radically often his manner of earning a living. The alter" the feeding habits of tuna. This major manufacturing industries depend to a finding has implications both for the value great extent on high-precision temperature of oceanographic temperature mapping and measurement and control. Finally, scientif- for pollution studies. One resulting ic research in many areas produces useful activity taking place is called the Mid- data only to the extent that it employs Ocean Dynamics Experiment [55]. sound, accurate thermometry. Some examples We have already discussed some of the of these applications follow. nuclear power reactor activities. Here the need is for stable, accurate fuel center- a) Thermometry in Medicine line temperatures in a hot, intensely radio- active environment, and similar stability Clinical laboratories are becoming in hot liquid sodium (LMFBR) or hotter highly automated in their testing and a gaseous helium (HTGR) [56]. In the geo- large fraction of the results obtained are thermal power search, the emphasis is on sufficiently sensitive to temperature to stability and accuracy of rock temperatures, demand that small, stable, highly sensitive, both to identify potential sources of geo- accurate, and rapidly-responding temperature thermal power and to monitor the depletion, sensors be used. Those sensors should also with time, of a given source [57]. An be capable of producing electrical signals accurate, quick method for determinina the that can be digitized in order to be used enthalpy of liquified natural gas is an with data recording and analysis systems. essential feature of equitable custody In instruments for enzyme rate analysis transfer, and this is not available. in which samples of only a few microliters The ability of a competent instrumenta- are used, the requirements for accuracy, tion scientist to influence the growth of response time and size of the temperature medical practice has been described in terms sensor are paramount, and in instruments for of "forcing a definition of the real re- acidity and blood gas analysis, temperature quirements of a measurement need to avoid measurements in the immediate vicinity of simply making gadgets for the M.D." [58]. the measuring electrodes require small, Another opportunity is in the coming use of accurate sensors. In routine clinical tests fluidic sensors to measure turbine inlet and that are presently being performed, tempera- outlet temperatures [59]; such a device is ture errors produce such a variation in the in increasing use in the aircraft industry, results of enzyme tests that comparisons but, though it is both durable and stable, among laboratories, and even within a given its temperature characteristics are re- laboratory, are useless. This is a sad latively unknown [60]. state of affairs, especially when one Many applications of infrared sensing realizes that inadequate thermometry in have been developed, and each of these clinical laboratory tests may result in contains the potential for temperature incorrect diagnosis and improper treatment mapping on both a fine and a macro scale. of the patient. Such activities as thermal pollution At the present time, 20% of all monitoring, plant and soil field studies, temperature measurements of patients in paper processing, rubber and textiles hospitals and in physician's offices are curing, printing and painting, food pro- taken using rapid, disposable and inexpen- cessing, and microcircuit analysis are all sive temperature sensors which are rapidly currently in operation. However, the displacing conventional mercury-in-glass temperature data which would allow thermometers and, yet, these sensors have correlation with other variables is very not been adequately evaluated! qualitative, because of uncertainties in Just as in the case of the electronic calibration [61]. fever thermometers, the temperature sensors in use today throughout the medical field 3. IMPACT, STATUS, AND TRENDS OF THE appear to have many of the required quali- NATIONAL TEMPERATURE MEASUREMENT SYSTEM ties, but they have never been adequately characterized. In addition to having 3.1 Impact of Thermometry Measurements sensors with the desired properties, it is necessary to be able to check their 3.1.1 Functional, Technological and calibrations periodically in the field or Scientific Applications to enable new calibrations to be made accurately. The best way of doing this We have already noted that thermometry would be through the use of a set of 26 .

standard temperature-reference-points in the and if a person is near a fire, then one range of biological interest and a practical side (say his front) is exceedingly hot simple mechanism to relate the patient whereas the other side (his back) may be temperature measuring instruments and the excessively cold. As a result of this ex- various devices utilized in the clinical change of radiant energy, the person is most laboratory to those reference points. uncomfortable but since the globe thermometer None exists at the present time. measures the mean radiant energy over 4 ir Miniature temperature sensors which steradians, it would indicate a temperature are insensitive to electromagnetic fields which would be pleasingly comfortable to the are needed in order to be able to determine person. This, then, points up the problem the temperature distributions produced in with such a thermometer. tissues being irradiated by the electro- We have noted earlier, of course, that magnetic waves in diathermy (a medical the home, office, and factory are heated and treatment in which heat is produced in the cooled according to temperature-sensitive tissues by a high frequency electric controls which operate furnaces, fans, re- current). Temperature mapping the tissues frigeration units, and sometimes ventilating is required in order to set standards of windows or doors. safe levels of electromagnetic radiation from the diathermy machines. This technique c) Thermometry in Manufacturing is required since direct measurements of electromagnetic field intensities in tissue One may consult the Standard Industrial are difficult to perform, and, in any case, Classification Codes noted in Table 2.5A for temperature is the critical variable inside an indication of the groups of manufacturers the tissue. Heretofore, such temperature which employ thermometry. measurements have not been performed in a The uses to which temperature measure- satisfactory way. ment amd control are put include steel making (not only to adjust the furnace b) Thermometry in Building Design and temperature itself, but also to monitor and Comfort control the temperature of gas streams, cooling ingots and coke ovens as well), Human thermal comfort has always been manufacturer of both ferrous and non ferrous a consideration in the design of buildings, metal products (chiefly in heat-treating whether homes or businesses. In the "good furnaces), and food preparation (much of old days" when energy was plentiful and the U.S. food preparation industry utilizes inexpensive and buildings did not need to assembly-line manufacturing principles, be so carefully designed, the thermostat techniques, and instrumentation) in the room or building could be set such which involves storage temperature that one would be comfortable at any given measurement and control as well as location in the room. When his location was process control changed to a different part of the building A great variety exists not only in the or even a different part of the same room, wide-ranging products of U.S. manufacturing, however, it is very likely that a change but also in the thermometry techniques which in the thermostat setting would be necessary it employs. Liquid in glass or metal in order for the person to be made comfor- thermometer systems, bimetallic, resistance, table again. Since energy required to keep and thermoelectric thermometers, and one comfortable is now very expensive and radiation pyrometers are used individually becoming more and more so with likelihood or in concert to measure, record, and control of an energy shortage, well insulated temperature in manufacturing. buildings designed with human thermal comfort maintained by a minimum use of d) Thermometry in Science energy are a prime consideration. A number of different types of The variety of thermometers employed in thermometers have been used by design scientific research and development is engineers and probably the one most nearly as great as that used in manufactur- widely used at the present time is the ing. High-precision resistance or quartz globe thermometer. This is a very crude oscillator thermometers are used in calori- thermometer system which attempts to measure metric studies and in other physical and the mean radiant energy over a solid angle chemical thermal data acquisition. Less of 4 TT steradians. Such a thermometer is precise but more versatile or smaller not a good indicator of human thermal thermometers are employed in basic studies comfort. This can easily be seen by a where temperature is not of primary impor- simple example. The human body is, to a tance. The range of temperature which is first approximation, a 2-dimensional system measured and controlled in scientific experimentation ranges from above the melting level of quality of thermometry. (We have point of the most refractory material (as, not reached the millenium in this particular for example, in research on thermonuclear area, however, as we have seen!) They relate fusion) to within one degree of absolute also to his safety, as temperature control zero. Innovation is the order of the of dangerous systems such as chemical pro- day in much of this v^^ork, and most advances cessors and boiler plants has improved to in thermometry arise from the efforts of the point that serious plant accidents are scientists to measure and control temperature. rare. They relate as well to his comfort and standard of living, insofar as he owns a 3.1.2 Economic Impacts - Costs and Benefits thermostated oven, toaster, automobile, home, air conditioner, laboratory, hot water Some idea of the costs involved in heater, office, greenhouse, or any manufac- the use of thermometry equipment by the tured item for which the materials of con- National Temperature Measurement System struction owe their properties to the careful has been given by figures quoted earlier in temperature control which is now so widely this Study. For example the annual sales available. of thermometers exceed $5 million just in ASTM-specified liquid-in-glass; electrical 3.2 Status and Trends of the Temperature and electronic thermometer sales in 1972 Measurement System exceeded $40 million; over $10 million worth of thermistors are sold annually; annual In coming to grips with the National sales of thermostats for heating and cooling Temperature Measurement System, its current exceed $280 million. status and trends, we called upon the per- Such figures indicate the economic sonal knowledge accumulated by our staff magnitude of thermometry (which is clearly over the years. We found that we had great enormous in itself and possesses a great stores of data on sensors, their stability deal of leverage to boot, since thermometry and calibration properties, and that we had comprises only a small portion of the value a limited knowledge of the needs of various of manufactured goods), but they do not people. address the individual questions of the The real core of the National Measure- benefit which may be derived by a given ment System is, of course, people. Only company or industry upon introducing more people and the purposes of people given effective thermometry. This latter type of meaning to thermometric devices and the information is difficult even for an numbers which come from them. individual company to obtain in a reliable In attempting to analyze the "National way. Still less visible is the cost of not Measurement System" for temperature, we introducing improvements in technology such realized that only by becoming as thermometry. conversant with the activities of people in This study has benefited from the ad- many different occupations could we begin vice of trained economists in numerous ways, to understand their interrelationships and from simple familiarization with economic to identify their capabilities and needs. statistics to the principles of technical- The best way to familiarize ourselves economic forecasting. A detailed discussion with the activities of the National of the cost-benefit relationship in the Temperature Measurement System is National Temperature Measurement System is by face-to-face discussion and laboratory or out of the question as a practical matter plant visits. In such extended contact and, we think, beyond the proper scope of we obtained some or all of the following this study. Even this brief overview of types of information: the System, however, makes abundantly 1) The nature of the group's thermom- clear the general situation that thermometry etry interest; investments in the National Temperature 2) the relevance of this interest to Measurement System are large and growing and other group's activities and to that the use of automatic thermometry equip- national problems; ment is increasing steadily. 3) the practical precision and accuracy desired, and that realized; 3.1.3 Social, Human, Person-on-the-Street 4) relevant economic data--costs or Impacts sales of calibrations, sensors, thermometry or associated equipment; The ma.ior benefits of thermometry in 5) impact of their work upon national, many social terms relate, for example, to the industrial, or social problems; precision and efficacy of medical treatment; 6) value of NBS thermometric services people are alive today because U.S. Industry to their own activities, and par- has provided U.S. medicine with a certain ticular needs which NBS might be

28 .

able to supply. a single lab; and the medical community Travel restrictions, coupled with the really doesn't know how to set up standards. large expenditure which travel involves, They suggest that the Federal government have the effect of limiting the amount of guide the medical profession and the in- such face-to-face discussion. The strument manufacturers in this task [1]. effectiveness of telephone or mail contact A similar theme is sounded by R. S. Melville in eliciting the type of information we of the National Institute of General seek is not nearly so great, unless a Medical Sciences, who notes the lack of previous contact with the individual has adequate clinical laboratory standards and created an easy relationship. the formation of a Committee for the study As a consequence of several such of automation in medical laboratories; he familiarization visits, we were able to stresses the utility of microcalorimetry discern patterns which allow us to separate work at NBS (in the Thermochemistry Section) groups of people according to the manner in and of the Standard Reference Materials which their activities impinge on program [2]. Mai Schechter, writing in thermometry. Lab World, urges Health, Education and Underlying the whole structure of the Welfare and National Institute for National Temperature Measurement System is Occupational Health and Safety to work with the International Practical Temperature American National Standards Institute, the Scale. NBS thermometricians improve the National Committee for Clinical Laboratory scale with the help of academic researchers Standards, and NBS on comprehensive medical who, though an independent lot, bring solid standards [3]. Dr. Nathan Gochman, Chief accomplishments to strengthen the scale. of Clinical Chemistry for the Veterans It is only a little oversimplification Administration Hospital in San Diego, to describe the three NBS calibration focused on temperature as a vital parameter services as the main pillars of support for in a talk on automated clinical testing, in the rest of the structure: in fact we have which he pointed out that the use of enzyme found a universally high regard for these analyses and other temperature-dependent services and a desire to see them more tests enhance noticeably the requirement widely available. for temperature standardization and control The NBS provides calibration services [4]. As will be noted in Section 4, NBS is to the various sensor manufacturers, beginning a comprehensive program of although, as we have indicated, the measuring the relatively unknown thermistor industry needs special help. In characteristics of the ubiquitous thermistor, turn, the sensor manufacturers support the and of providing the technical leadership temperature control equipment manufacturers for medical standardization. and the large government and commercial The regulatory agencies, too, have temperature-cal ibration faci 1 ities difficulties. One example of this problem Finally, the commodity industries, both is a recent case before the New York Public large and small, rest strongly upon the Utilities Commission involving the re- combined bases of temperature control devices location of the official thermometer by the and sensors. Orange and Rockland Utilities, Inc.; the Another aspect of the National utility had moved the thermometer from a Temperature Measurement System concerns relatively protected position to an health and safety in thermometry. We found, unsheltered one, and sought rate relief of in this area, that the problem of medical $1 million based on the colder readings [5]. temperature standards is severe. Doctors An enormous current and developing J. H. V. Brown and D. J. Lowell of Health, problem centers around automotive pollution Education and Welfare discussed the question and mileage; we noted earlier that auto- of standardization and health care in a motive design is often unscientific, in cogent article which points out that, with part owing to poor thermometry in design. acequate standardization, the medical This leads to mediocre mileage-to-weight profession could well concentrate on main- ratios. Emission control also suffers taining the health of the 220 million from poor thermometry. Privately, the citizens rather than simply curing the 1.3 industry admits that health and safety million sick. There are no widespread considerations rank low in its priority criteria for manufacture of hospital equip- listing [6]. The regulatory agencies, ment to prevent electrical shock, they say; however, must decide on workable goals in inadequate control of pacemaking equipment the face of conflicting arguments; the auto is rampant; the safety of ultrasound and industry has been described by the microwave equipment is completely unevalu- Secretary of Commerce as "spending nearly ated; clinical tests are widely variant $2 billion in 1973 on safety and emission from hospital to hospital and even within standards" [7], and the Coordinating 29 .

Research Council-Air Pollution Research During the past year or so, we have Advisory Committee has been both criticized visited well over fifty facilities of as an industry slave [8] and praised as an various types. We attempted to visit independent force [9]. P. G. Hansel, Vice- power manufacturers, aerospace manufacturers, President of Research for an NCR subsidiary, regulatory agencies, medical research and recently called for major programs to en- health care facilities, government calibra- hance engine efficiency [10]. Accurate tion laboratories, oceanographic facilities, measurement of temperature and pressure in temperature control and temperature sensor each cylinder and an electronic gasoline manufacturers. In addition, we have control system are being considered as a attended numerous scientific, technical, mechanism for reducing emissions and im- and commercial standards meetings. proving mileage of automobiles. As a consequence of these visits, it The problems of the medical community became obvious to us that we could signifi- also occur, of course, with respect to cantly affect the quality of medical regulatory agencies. We have already alluded patient care in the U.S. by undertaking a to the uncertainty of ultrasonic and micro- combined technical and liaison program in wave diathermy safety. The reliability of medical thermometry. Reprogramming of electronic thermometers [12] is also regula- personnel to this end has enabled us to tory concern of FDA, as well as of several begin already a 3-4 person effort which is states. The principal challenge to NBS is more fully described in Section 4. to provide useful guidance to these agencies We have found, as might be expected, within their relatively short time require- that many parts of the National Temperature Measurement System have generated of ments . ways In the various power and energy areas, solving their own problems in temperature. we find the active people uncertain as to One case which is worth citing is that of how to proceed with measurement and control critically evaluated data, in which an problems. In the very active area of attempt is made to analyze existing data to nuclear power reactors, the question of arrive at "best values" for a given public safety is the subject of continuing quantity [17,18]. The NAS/NRC has sponsored controversy. Fortunately (for the NBS a Report on Physics in Instrumentation, in thermometry program, that is) the primary which the integrity and needs for thermom- controversy centers around radiation etry instrumentation are evaluated. Many safety [13]. However, the lack of really professional associates have sponsored adequate temperature measurement capability research groups, such as the American Gas has contributed to the difficulty of Association and the Electric Power Research introducing substantial quantities of nu- Institute, and many non-NBS groups sponsor clear power into the U.S. energy network meetings and research projects on [14]. The U.S. currently has about 17,500 thermometry and control. A few examples megawatts of nuclear power--about 10% of that of these are the ORNL Instrumentation and currently projected to 1980. This is about Controls Division Annual Information meeting; the same amount of power as is produced by a course on Measurement Systems Engineering TVA from non-nuclear sources [15,16]. given annually by Arizona State University, Finally, in the area of commodity and "Physics Opportunities in Energy Pro- manufacturing, the problems of health and blems", offered by the American Physical safety often are not readily visible, and Society in February 1974. they often are not willingly exposed to Although sufficient to give us a picture view. of the basic status and trends of the Another area of immediate concern is National Temperature Measurement System, that of industrial standards. We believe the field investigations conducted to date the keys to this situation are information are by no means sufficient either to and cooperation, with NBS data and technical delineate completely the natures of the advice connecting the individual manufacturer various groups or to point the way un- and standards groups and among the various equivocally to correct courses of future standards groups themselves. action. To these ends, we have outlined In forming these ideas, we drew upon further specific inquiries which we shall many years of combined experience of note later. service to the National Temperature ^ Measurement System (although we have not always thought of it in those terms) and upon familiarization visits to particular faci 1 ities

30 4. SURVEY OF NBS SERVICES TO THERMOMETRY The latter volume likewise is a mine of NBS expertise, this time in the general NBS has contributed to the National field of calorimetry, with many of the Temperature Measurement System since the thirty-odd papers being of fundamental Bureau was established in 1901. In this importance to thermometry. section, we will present some discussion of It is worthwhile to review the NBS early contributions and we will detail the contributions to the Fifth Temperature nature of present activities. It should be Symposium, [6] since a good idea of the noted at the outset that the present ther- level of NBS thermometry can thus be gained. mometry activities cross Institute as well R. P. Hudson, Chief of the Heat Division; as Division lines, and we made a definite H. J. Kostkowski, Chief of the Optical attempt to present a balanced view of NBS Radiation Section; H. H. Plumb, Chief of the thermometry without chauvinism. The reader Temperature SEction; and W. R. Tilley, can easily determine how well we succeeded Chief of the Office of Technical Publications by soliciting comments from outside the NBS served on the General Committee. H. H. Plumb Heat Division. was Chairman of the Program Committee, as- sisted by H. J. Kostkowski, R. D. Cutkosky 4.1 The Past (Electricity Division), and R. L. Powell (Cryogenics Division) in addition to the From the first International Temperature twenty or so members from outside NBS. Scale of 1927 (submitted by NBS, NPL, and In the technical program, NBS contribu- PTR, and described for the U.S. public by ted some 27 papers ranging in nature from G. K. Burgess, at that time director of very fundamental thermodynamic thermometry NBS) [1], the National Bureau of Standards to temperatures in building environments. has maintained a leading role in U.S. All of the foregoing Section represents thermometry: With the National Research only a portion of the thermometry activities Council, NBS cooperated in preparing the 1939 of NBS in the recent past. Consultations, Temperature Symposium and provided half of reports to sponsors, other publications and the papers on temperature scales and on contributions to other scientific meetings precision thermometry [2]; NBS established must be added to complete the picture. temperature calibration facilities for the various standard instruments (Over 500 firms 4.2 The Present--Scope of NBS Thermometry were listed among recent calibration Activities services customers). With the AIP and the U.S. Army, NBS sponsored the 1954 Temperature 4.2.1 Description of NBS Thermometry Symposium and again contributed key papers Services on fixed point apparatus and temperature scales [3]; similar comments apply to the In order to give the reader a flavor of 1961 and the 1971 Symposia; in addition, the range of thermometry activities at NBS NBS has disseminated tables, monographs we shall note recent accomplishments of the on measurement techniques, and original appropriate programs. In the next subsection research in thermometry through journal we shall review the thermometry programs articles, colloquia, seminars, and other within the Heat Division [7], and scientific meetings. subsequently other thermometry highlights Two very useful compilations of NBS within NBS. thermometry contributions are Vol. 2^ of NBS Special Publication 300, "Precision 4.2.1.1 Heat Division Thermometry Activities, Measurement and Calibration-Temperature" 1974 [4] and its companion volume on Heat [5]. The former contains an assortment of papers The new year brought a change in leader- on temperature scales including IPTS-48; ship to the Temperature Section. R. P. the 1939 Hoge-Brickwedde gas-thermometer Hudson, Chief of the Heat Division, acceded scale recorded on platinum resistance to H. H. Plumb's long-standing desire to thermometers between 14 °K and 83 °K; its return to a more active scientific role in correction in 1955; the 1965 NBS 2-20 K the Section's activities, thus bringing to velocity of sound scale of Plumb and an end an era that began in October, 1966. Cataland; and the 0.002-2 K scale based on Under Dr. Plumb's leadership, the the paramagnetism of cerous magnesium Temperature Section gained both stature and nitrate by Hudson and Kaeser. In addition, versatility. In 1966, over 50% of the there are resistance, thermoelectric, Section's effort was devoted to calibration liquid-in-glass, and spectroscopic activities; this figure is now below 20%. thermometry and optical pyrometry; in all, Projects have been initiated in calorimetric some twenty-five papers. measurement of thermodynamic temperatures.

31 .

in medical thermometry, and in a continuing the usual number. Some twenty-seven study of the national temperature measure- participants crowded the calibration ment system. The Section participated laboratories during the week of 10-14 March. closely in the formulation of the new In addition to providing both that seminar Temperature Scale [8], and it is pre- and a three-hour thermometry session which sently represented on two of the four formed a part of a local university's short working groups of the CCT which were in course in standards laboratory operation, existence in January, 1974, as well as on the calibration staff tested over one the temperature committees and subcommittees thousand precision thermometers and answered of both the American Society for Testing and a like number of written and verbal in- Materials and the American National quiries on thermometry to round out a very Standards Institute. satisfactory year. Several of the highlights of the year's The Section's work in thermocouple activity in the Temperature Section appear thermometry included the publication of the in the Minutes of the May, 1974 meeting long-awaited Thermocouple Reference Tables, of the CCT at the International Bureau Monograph 125, and highly successful study of Weights and Measures (BIPM) in S6vres, of a new high-temperature nickel-silicon- France [9]. The National Bureau of Standards chromium thermocouple system developed in was represented at that meeting by Australia. R. P. Hudson, Chief of the Heat Division, The low-temperature acoustic thermometer and by H. H. Plumb. Technical contribu- program yielded results of thermodynamic tions were submitted by staff members of measurements of the transition temperatures the Cryogenic Physics Section, of the of two elements of the NBS superconductive Equation of State Section, and of the Tem- fixed-point device. perature Section. A summary of the The medical thermometry program, begun Temperature Section contributions follows: only a year ago but now the largest in the Section, has made rapid progress toward its 1) A brief discussion of the cryostat goal of improving the nation's medical care and methods used in the calibration by improving the quality of thermometry in of capsule platinum resistance medicine. New precision clinical laboratory thermometers from 13.81 K to thermometers were developed, equipment and 90.188 K; procedures for characterization of ther- mistors and small platinum resistance ther- 2) recent gas thermometry results mometers were completed, collaborative work which provide information on the on an environmental scanning radiometer and differences between the IPTS-68 and on thermometry in medical diathermy was thermodynamic temperatures from accomplished, and exchange of equipment and the triple point of water to ideas took place between the medical ther- 415 K (142 °C). Working group 3, mometry group and staff members of of which L. A. Guildner is a member, hospitals, clinics, and medical instrument reported it; manufacturers Closely allied to the field of medical 3) a brief description of apparatus, thermometry, as well, is a new project currently under construction, for begun in the past year by W. S. Hurst. the comparison of standard platinum Partially funded by the National Institutes resistance thermometers (SPRT's) of Health, this project will be an experi- from 90 K to 900 K; mental study of thin-film capacitance ther-

mometry . 4) the results of a study of the R. J. Soulen is utilizing a sensitive freezing temperatures of six superconductive tunnel junction to determine experimental aluminum freezing- noise temperatures according to the Nyquist point cells; and theorem. This technique was suggested by R. A. Kamper of the Cryogenics Division. 5) editorial suggestions regarding At present Soulen 's results agree within ±2 changes in the text of the IPTS-68. mK from .02 - .04 K with results of yet Consideration of possible revisions another thermodynamic temperature determin- of the text was a particular duty ation undertaken by H. Marshak. These

of Working Group 1 of the CCT, of latter measurements involve the measurement which H. H. Plumb is a member. of the anisotropy of y-i^adiation emitted by 60-Co nuclei in a single crystal of cobalt; As usual, the Section offered a week- the various nuclear data determine the long seminar on precision thermometry. The anisotropy due to decay of nuclei in the attendance, however, was more than double magnetic energy levels, and the Boltzmann

32 .

distribution function governs the level similar program is underway with HUD, AEC, populations. NASA, and EPA cooperation [14]. Since 20- In addition to the cryogenic physics 25 percent of the nation's energy use goes discussed above. Heat Division programs for space and water heating or for lighting, include the generation of Superconductive significant power savings can result from Thermometric Fixed Points from 0.5 K to 7 K, these programs. developed as NBS Standard Reference The Center also has an experimental Material 767 by J. F. Schooley and R. J. townhouse in which the thermal balance is Soulen, Jr. [10], and study of paramagnetic measured. B. A. Peavy, D. M. Burch, J. D. sensors below 20 K by B. W. Mangum and Allen, and D. R. Showalter have instrumented D. B. Utton. The overall Cryogenic Physics the house heavily, including some 140 air- Thermometry Program is discussed in some temperature sensors and 58 wall and window detail in NBS Technical Note 830 "NBS surface temperature probes [15]. The CBT Cryogenic Thermometry and the Proposed environmental test chamber provides a temper- Cryogenic Extension of the IPTS". ature environment ranging from -50 °F to In this section, an indication has +150 °F. In one recent study, the fuel been given of the range of NBS activities in economy of nighttime reduction of the temperature standards. Appendix D lists thermostat set point was evaluated and most (dare one say alll) of the NBS staff shown to be quite substantial. members who officially participated in these In other, related CBT programs, T. Kusuda activities and F. J. Powell are studying thermal To this catalog we add a recent anecdote storage and heating cycles in buildings [16]; which may be illustrative of NBS' role in and J. Stern and S. Edelman are studying industrial standards. Having joined the plastic infrared detectors [17] and their International Organization for Legal use in scanning thermal detectors [18]. Metrology (OIML) in August, 1972 [11], the U.S. was given the opportunity to vote on b) High-Speed Thermophysics a current draft recommendation on electrical resistance thermometers composed of Pt, Cu, A. Cezairliyan, of IMR, has been study- and Ni (PR 41). In order to develop a ing for some time the simultaneous measure- position on the recommendation, J. Schooley ment of a variety of thermophysical distributed copies of it during the ASTM properties of refractory metals. In Resistance Thermometry (E-20) Committee addition to such information as electrical meeting in June, 1973. Other copies were resistivity and heat capacity, he obtains distributed to interested companies who were temperature values for the sample data by alerted by ASTM. In all, correspondence optical pyrometry. Recently, this technique took place with some two dozen firms. provided information on the thermal proper- Their comments on the draft were consolidated ties of niobium alloys in the range 1500- into a U.S. response with the help of 2700 K [19]. In addition. Dr. Cezairliyan Mr. E. D. Zysk, of Engelhard Industries, has studied the relation between radiance a representative to lEC, and Mr. J. E. French temperature and actual temperature at the of the Scientific Apparatus Manufacturers [20]. Association. Mr. W. E. Andrus, U.S. representative to OIML, communicated this c) Microcalorimetry (negative) response officially [12]. Microcalorimetry is a science which 4.2.1.2 NBS Thermometry Activities Outside is of particular value to research bio- Heat Division chemists and to clinical laboratories because of its ability to analyze a) Thermal Efficiency of Buildings extremely small samples. This very require- ment, of course, brings its own emphasis Temperature measurements play a large on the use of small, sensitive thermometer role in several parts of the Center for probes. NBS, in its Thermochemistry Section Building Technology programs. One of these. of IMR, has a very active group including Project Breakthrough [13], was commissioned G. T. Armstrong, E. J. Prosen, R. N. Goldberg, by HUD to examine the feasibility of R. N. Boyd, B. R. Staples, and M. V. Kilday. applying the principle of overall energy An NBS clinical microcalorimeter which was consumption to a building complex; all the described in a recent report [21] uses a "waste heat" from the power generators is to quartz thermometer (sensitive to ±10 yK)for be used for building air and water heating. overall temperature stability evaluation, P. R. Achenbach and J. B. Coble are attempt- and a thermopile is used in the primary ing to assess the economic features, includ- energy detection. ing the heat balance, of the complex. A 33 4.2.1.3 The NBS Thermometry Input-Output ment services. Provide information Table on future needs. FDA, HUD, EPA, OSHA Other Government Agencies - Provide direct The present activities of NBS in ther- requests for calibrations, consulta- mometry can be summarized in a variety of tions on thermometry problems. ways. Perhaps the most useful technique Provide information on future needs. would show a dollar value which each USN, USA, USAF, NOAA, AEC, NASA, HUD, portion of the program provides to the DOT, NIH various components of the national measure- Medical Groups - Provide direct requests for ment system, but this technique is absolutely calibrations, and consultation. Pro- beyond the competence of mere scientists, vide information on future needs. involving as it does an economic evaluation NCCSL, CAP, Medical Instrument Manu- of basic temperature scale studies which facturers, Individual hospitals, predate their useful consequences by many physicians, and medical researchers years. Two methods, however, are at once Thermometry Equipment Manufacturers - Provide feasible and useful. One of these is the direct requests for calibration and source-output diagram, and the other is a consultation services. Provide in- national needs-thermometry activities formation on state-of-the-art design. matrix. -100 firms, ISA, SAMA In the source-output table 4.2A, we University Researchers and Professional attempt to show the origin of the current Journals - Provide information on thermometry activities (which we have advanced thermometry principles and discussed briefly in this section) and the techniques. Provide collaboration on major outputs which result from these research and development problems. activities. The distribution of calibra- -1000 universities, -50 journals tions, research efforts, and consultations over the various user areas conspire to NBS Outputs, with Percentages make the estimates of the percentage out- puts very uncertain; however, these numbers 10% may be useful even in a qualitative way. NBS - Calibrations. Consultation, Colloquia. Reports. 10% Industrial Firms - Calibrations (-$75,000 in FY-73). Consultations. Reports. Table 4.2A 10% Voluntary Standards Groups - Meetings NBS Thermometry Request Sources attendance (-100 estimated in FY-74). Technical evaluation of standards pro- NBS - Provide direct requests for calibra- posals (-10 estimated in FY-74). tions and consultations. 5% Electricity Division, Physical Regulatory Agencies - Consultations - OSHA, Chemistry Division, Mechanics FDA, HUD, DOT, EPA. Reports. Division, Optical Physics Division, 25% Cryogenics Division, Analytical Other Government Agencies - Calibrations Chemistry Division, Polymers Division, (-$40,000 in FY-74). Reports on Electromagnetics Division, Metallurgy sponsored research (estimate Division, Inorganic Materials Division, $500,000). Electronic Technology Division, I 15% Structures, Materials, and Life Safety Medical Groups - Calibrations (-$15,000 Division in FY-74). Reports on sponsored Industrial Requests - For sensor development research. Consultations. and characterization, calibration, 10% thermometry consultation. -500 in- Thermometry Equipment Manufacturers - dividual firms, 10-12 trade associa- Consultations. Calibrations tions (-$20,000 in FY-74). Voluntary Standards Groups - Provide direct 15% requests for technical guidance in Scientific Research Journals - Scientific writing standards. Provides infor- papers (-50 in FY-74). Colloquia. mation on current and planned develop- ment in scale work. CCT, ASTM, ANSI, lEC, ISO, OIML Regulatory Agencies - Provide direct re- quests for consultation and measure-

34 4.2.2 Users of NBS Services broad, general problems in thermometry measurement science. In thermometry these The major users of National Bureau problems are met by very small NBS efforts of Standards thermometry services — two-to four-person programs undertaken in have been identified in Section 2.5. collaboration with university, govern- They comprise the "Knowledge mental, or industrial groups. The efforts Community" (Scientific and Educational are commonly mounted in response to needs groups), Metro! ogical and Standards bodies, perceived as affecting an entire industry the Thermometry Instrumentation industry, or measurement area. The fact that the NBS itself, government and commercial NBS efforts in these general areas are calibration laboratories, various government miniscule in comparison with those of agencies, manufacturing groups, and health other government or industrial laboratories, services. Typical examples of their uses coupled with the fact that workers in the of NBS services are the employment by a particular areas involved are generally thermometry manufacturer of a calibrated motivated to solve their own problems, Standard Platinum Resistance Thermometer indicate that the NBS efforts here are to determine the accuracy of a production the least essential of its services. A temperature sensor-readout system; and the modicum of foresight and modest resources conversion by a new company of NBS instruc- would permit the NBS effort in applied tion on the manner of establishing a cali- thermometry to be duplicated elsewhere. bration laboratory, into a set of orders for The missing elements in this solution are space, environmental control, and precision those which NBS can contribute in special thermometry equipment. circumstances: organizational resources to bring particular problems to light as 4.2.3 Alternative Sources is done in periodic workshops and con- ferences generated by NBS; objective It is difficult to say how the various expertise to apply to important national thermometry responsibilities of the NBS would questions such as the quality of particular be met by the National Temperature Measure- energy-conserving measures or effective ment System in the unlikely circumstance means of improving U.S. medical care by that NBS were to cease to exist. application of more relevant standards of The fundamental difficulty with such a thermometry; and the combination of state- step would lie in the responsibility of NBS of-the-art facilities, expert staff, and to participate with other industrialized flexibility in programming to "bear down" nations of the world in maintaining a uniform on a particular problem before it becomes International Temperature Scale which is a bottleneck to progress by a whole based on thermodynamic measurements. This is industry. a very expensive undertaking, since it de- mands the long-term involvement of highly- 4.2.4 Funding Sources for NBS Thermometry qualified scientists. We know of no Services

organization--governmental , university, or industrial --which could commit sufficient Overall, NBS thermometry is funded resources to meet this responsibility. Thus about 50% from external sources and 50% by direct congressional appropriation. any such attempt would necessarily involve a consortium of organizations with the The outside funding support comes chiefly attendant management problems. from calibration income and the Executive A similar difficulty appears in meeting Departments of Health, Education, and the responsibility of Scale dissemination. Welfare, of Housing and Urban Development, A likely solution would entail the verifica- and of Defense. tion of Scale integrity by each of the major The distribution of this funding is industrial and governmental calibration not uniform throughout the NBS programs, laboratories. This solution carries with it however. Most of the outside funding the inherent problem that a given laboratory accrues to applied thermometry programs outside the Heat Division (excepting would have a financial stake in minimizing the expense involved in certifying the calibration income), while the major direct accuracy of its thermometric measurements. appropriation support goes to Heat Division scale research and dissemination activities. The lack of objectivity in such a system would almost certainly lead eventually to The following summary of Fiscal Year the Cryogenic Physics and the generation of a "National Bureau of 1974 funding of (the principal Regulatory Standards" somewhere within the Temperature Sections two federal government. thermometry activities of the Heat Division) this The easiest of NBS' current responsi- helps to illustrate situation. bilities to meet is that of contributing to 35 Table 4.2B

Cryogenic Physics and Temperature Sections Funding by Source, Fiscal Year 1974

Source Percent

NBS

Direct Appropriation 80

Standard Reference Materials 1

Other NBS Divisions 1

Total NBS 82

Other Government Agencies

National Oceanographic Administration 1 Food and Drug Administration 3 Department of Justice 4

Total OGA 8

Cal ibrations 10

Grand Total 100%

4.2.5 Mechanisms for Supplying Services written and distributed during discussions with the users of these services. The dissemination mechanisms by which NBS supplies its thermometry services take Basic and Applied Thermometry results many forms, reflecting the diversity of NBS and services are communicated through the activities. Most of these have been writing of reports-to-sponsor, NBS technical discussed elsewhere in this study and will publications, or research papers. In be mentioned here only briefly. addition, of course, lectures, seminars, Scale Research results are communicated and technical meetings serve as outlets for to the general scientific community through this type of information. In many cases, articles published in the open literature. particular devices make up a portion of In addition, direct communication is made these services. to the International Thermometry Committee Standard Reference Materials and Data (the "CCT") of the General Conference on results appear as items for sale through Weights and Measures both in written reports the Office of Standard Reference Materials and by verbal presentation at periodic (listed in the NBS Special Publication 260), meetings. or as reference tables or other data Cal ibrations are rendered upon published in the NBS Monograph series. thermometry equipment owned by governmental Voluntary Standards services generally and industrial laboratories. Test results appear indirectly. New or revised indus- are given in written form and occasionally trial standards often are based in part or analyzed as a class to yield information wholly upon data or measurement techniques on the overall precision and accuracy of originating at the NBS. Examples of this NBS temperature calibrations. These activity arise primarily because the NBS summaries are made available to the public thermometry staff actively participate in the in general through the NBS Technical Note work of the American Society for Testing and series. Materials, the American National Standards

Measurement Assurance Programs , Institute, and their international counter- Consultation and Instruction involve the parts. Writing and editing of publications NBS staff directly with individual govern- issued by or at the request of such groups mental, industrial, and medical thermom- often falls to NBS thermometrists because etrists. Test summaries, recommendations, of their great familiarity with the topic. handbooks and instructional materials are 4.3 Impact of NBS Thermometry Services temperature scale at all temperatures carries to the areas of competence in international 4.3.1 Economic Impact of Major User Classes trade and of the general improvement of the quality of U.S. goods and services, very The major thermometry users have been specific technological improvements result identified in Section 2. In terms of from NBS activities in thermometry. immediate economic importance, certainly Among these improvements are techniques the largest contribution comes from the of calibration and measurement, which enable manufacturers of durable goods. These higher quality to be delivered for less people build and sell refrigerators, air expense; new knowledge of the characteristics conditioners, automobiles, aircraft, stoves, of medical sensors, which permits their use furnaces, furniture, and many other products at new levels of accuracy; new types of high- which may include thermometers or thermo- temperature thermocouples, leading to advances metric controls or may simply require in the instrumentation and design of nuclear thermometric monitoring or control in reactors, high-temperature turbines and their manufacture. These people are NBS engines, and high-temperature industrial thermometry calibration customers. It is processes; new accuracy in cryogenic difficult to assign a value added by instrumentation; new techniques for NBS services to this group. differential thermal analysis and other Mining, agriculature, chemicals basically microcalorimetric applications; new manufacturing, and metals and plastics knowledge of fast temperature measurement producers regularly send their highest techniques; and more accurate and reliable quality thermometers to NBS for thermal analysis of walls, buildings, and thermometry calibrations. Again the dwelling complexes. economic impact may be stated simply as A National Needs-Thermometry Activities "large". Chart was devised in a series of meetings The automatic temperature controls of the Temperature Section staff as a con- industry is separately classified by the venient means of monitoring particular Census Bureau. The various components areas where NBS thermometry efforts have an of this industry annually ship goods worth impact upon various national needs. several hundreds of millions of dollars Table 4.3A contains a listing of various annually. By this time, our listing is areas of importance to the nation in the reaching groups whose representatives not first column. The second column contains only buy calibrations from NBS, but also a listing of many projects and services of consult frequently on measurement problems. the thermometrists at the NBS. Following As the user-group output becomes more each entry in column two is a list of specialized, its direct and immediate numbers which denote the national needs which economic impact is smaller, but its long- are, in part, met by the particular NBS range importance appears to be larger. service or project. The manufacturers of thermocouple thermom- This method of presentation has pro- eters, resistance thermometers, liquid-in- vided us with the ability to see glass thermometers, bimetallic and thermistor how our own programs impinge on national thermometers ship annually goods valued at problems and indicate that we are more $10-50 million in each subgroup. However, readily able to plan our programs with a these products govern the quality and perspective arising from consideration of performance of those enormous user classes those needs. listed earlier. Still longer range in economic impact 4.3.3 Pay-Off from Changes in NBS Services is the government-funded research and technology activity, containing the U.S. Over the years, NBS has begun new executive Departments, the national services and dropped others, with consequent laboratories, and federally funded labora- adjustments by the U.S. Temperature Measure- tories in the industrial sector. NBS helps ment System. One such change was the the people who staff these organizations in cessation of calibrating fever thermometers. ways we shall note in the next section; and This change helped bring down the price of although their immediate economic impact is such thermometers, and it simultaneously small (or negative!) their importance for forced the manufacturers to establish and the future of the U.S. is large. regulate for themselves their calibration. A second example of such a change is 4.3.2 Technological Impact of NBS Services the institution of medical thermometry as a program in the Temperature Section two In addition to the obvious influence years ago. This change has already begun to which basic research on the international show benefits in improved thermometry as well

37 . . . : ,

Table 4.3A

NBS- National Needs Chart

National Needs Thermometry Activities

1 Energy 1 Thermocouple Development 1 5, S, 11 2. Measurement Assurance 2. Miniature Resistance

Thermometer 1 5, 9, 10, 11

3. Meteorology 3. Noise Thermometry 1 9, 10

/I r 4. Oceanography 4. Radiation Thermometry 1 . 5, 9 , 10, 11

5. Health 5. Fixed points 1 5, 9 , 10 6. Safety 6. Thermistor Characteri- zation 5, 9, 10, 11 7. Environment 7. Transistor Thermometry 5 o n 1 r» 11 8. Consumer Protection 8. Cal ibrations 2, 5, 8, 9, 10, 11 (i.e., product integrity)

9. Industrial Standards 9. Tiny Sensors 5 10. An Accurate Temperature Scale 10. Mi crocal or i met ry 5, 9, 12 1 1 Temperature Control 11 Electronic Thermometers 5, 8 12. Temperature Measurement Systems 12. Liquid Crystals 5, 9, 10 13. Thermal Pollution 13. Quartz Thermometers 3, 4, 5 14. Safety in Thermometry 6 15. Theory of Thermometry 10, 11, 12 16. Regulatory Agencies 5, 6, 7, 8 17. Practical Thermometry Tips 2, 8, 9 18. Intercompari son of Standard Platinum Resistance Thermometers 2, 10 19. Gas Thermometry 9, 10 20. Transient and Dynamic Thermometry 1, 2, 3, 4, 9, 11, 12 21. New Equipment Development 2, 9, 11, 12 22. Cryogenic Sensors 1, 2, 10, 11, 12 23. Nuclear Resonance Thermometry 10, 12

as an increased awareness of the importance 4.4 Evaluation of NBS Thermometry Program of accurate thermometry to effective medical practice. The NBS thermometry program is contin- As thermometry projects at the NBS are ually evaluated by its own staff and by NBS completed, or reach a point where further management in periodical reviews. In work would bring diminished returns, the addition to these self-appraisals, an ex- institution of new projects becomes plicit evaluation of the program is supplied possible (in the present climate which pre- by a panel of experts on an annual basis. vails throughout the U.S. economy, NBS is The Evaluation Panels have been established making every effort to maintain a lively through the National Academy of Sciences in effective standards program without overall conjunction with the National Academy of growth in its staff). Engineering and the National Research

A large part of the advance planning Counci 1 prior to such opportunities concerns explicit The Heat Division Evaluation Panel is questions of scientific and technical pay- made up of some ten members, which in offs accompanying various options. These recent years have represented the following are necessarily balanced against questions organizations of available facilities and major strengths . Argonne National Laboratory in the thermometry staff. • University of Illinois Center for .

Advanced Study In the opinion of its line management,

• Pennsylvania State University, physics the major strength of the NBS thermometry department program is that it has competent staff.

• Eastman Kodak Company, di rectorate of Because of this fact, those areas which are research treated at all receive intelligent attention. • Hunter Associates Laboratory, Inc. A second strength lies in the variety of the

• Los Alamos Scientific Laboratory thermometry activities: The Heat Division

• General Electric Corporate Research staff performs calibrations, and basic and and Development applied research in standard sensors; the

• Lawrence Livermore Laboratory Physical Chemistry Division staff studies

• General Telephone and Electronics thermometry as applied to differential Sylvania Lighting Center thermal analysis and microcalorimetry; and

• Cornell University department the Center for Building Research studies of chemistry thermometry applied to energy flow in con-

. California Institute of Technology struction materials. department of physics The major weakness of the NBS Thermometry

• Bell Laboratories, Murray Hill programs is that so many areas of thermom-

• University of Rochester, department etry are not under study at all . A second of physics and astronomy weakness is that there is too little inter- • University of Wisconsin, Space Science action between the metrological thermometry and Engineering in the Heat Division and the applied An estimate of the opinion of the Heat thermometrists working elsewhere in NBS. Division Evaluation Panel on the quality of The Evaluation Panel of the Heat Division the thermometry program may be obtained from generally supports these findings, offering the following, quoted from the most recent high praise for the program as performed but report: regret at the thinness of its coverage of the "The Panel believes that the thermometry field. program in the Heat Division, which includes the work in the Temperature Section and in 4.5 The Future of Thermometry at NBS the Cryogenic Physics Section as well as the Optical Pyrometry project in the The considerations which have been Optical Radiation Section, is excellent. treated in this study demonstrate a need The objectives of the program are sound and for a number of activities for NBS they are appropriate. The staff working thermometry, some along well -traveled on thermometry are very competent and they paths and others in new directions, are enthusiastic about their work. The accomplishments of the program in terms of a) The National Temperature Measurement publications and the thermometer developments System Study made by the staff are significant in respect to both quantity and quality." In the first place, the assessment of The Panel recommended that the program the needs of the National Temperature Measure- should be more heavily funded and staffed, ment System is clearly a valuable tool in but expressed an understanding of NBS directing NBS programs, and we believe that limitations in that regard. Detailed it should be a continuing process. Field discussions of some individual projects visits, which currently occupy a great deal give considerable guidance to NBS in its of effort (and not a little money) are an efforts to maintain a lively, effective excellent source of information, much of program. which cannot be obtained in other ways. We The criteria by means of which NBS see this activity diminishing in intensity evaluates possible additions to its programs within the next two years, hov;ever, as we can be listed as follows: become more knowledgeable regarding the • Relevance to the basic program National Temperature Measurement System.

• Possible Scientific or Technical We expect then to continue the field visits Advancement at a reduced level • Adequacy of Resources (staff, equip- As personal contacts become more firm ment, and space) and widespread, we expect to make greater • Available NBS Leadership. use of telephone and correspondence as a means In addition to the above criteria, work of maintaining liaison with current activities contemplated for other agencies must be of in facilities which we have already visited. such a nature that NBS, its staff, or its This procedure will then form a substantial facilities are uniquely suited for the and continuing channel of communication. project. A particularly important part of the study activity should be closer liaison 39 between the Heat Division staff and the other defining points on the cryogenic scale should

NBS thermometry personnel . Where the pro- be pressed. The generation of a convenient, grams require collaboration, it seems to take accurate, and reliable thermometer for place in a satisfactory manner; however, transmitting the cryogenic scale to the lack of awareness and communication is cryogenics industry is a vital part of this apparent in many areas, and this problem is program, as well. The nuclear magnetic compounded by difficulties such as competi- resonance thermometer offers not only an tion for external funding. alternative cryogenic sensor, but a means of following current research trends toward b) Calibrations ever-lower temperatures. Another area of temperature scale Secondly, our studies show a continuing research involves the convenient, accurate need for a calibration capability in measurement of temperatures in the 600 - platinum, liquid, and thermocouple thermom- 2000 °C range. At present, neither eters at the highest attainable levels of thermocouple nor resistance thermometry accuracy to maintain a uniform scale of is completely satisfactory in this range, temperature for both science and technology. and as industrial needs expand, NBS clearly The construction of comparison can provide substantial help by continuing facilities for calibration between fixed to work in this area. points and the design and installation of Finally, NBS should respond to the automatic bridging and data logging equip- various thermometry needs which have become ment can be expected to permit increased evident in the National Temperature versatility in the calibration services. Measurement System study. We can assume One potential feature of such services is an that only a small fraction of the total extended measurement assurance program which problems have been discovered up to the will permit NBS to evaluate selected cali- present time, and that others will demand bration facilities, at their request, in the attention as the study continues. field. As part of the calibration activities, d) Medical Thermometry the capability for fixed-point determin- ations is necessary. As part of this In the area of Medical Thermometry, work activity, the possibility of utilizing has begun on the characterization of transitions in well -characterized solids thermistors and of commercial platinum re- will be examined. Such solid-state fixed sistance thermometers. In addition, B. W. points would be substantially easier to Mangum and S. D. Wood are actively building, realize and to promulgate than most of in collaboration with the medical community, those in existence at present. with the Food and Drug Administration, and with the Bureau of Radiological Health, a c) The International Temperature Scale further technical base for medical thermometry. The expected end product is Thirdly, work on the temperature standardization of medical thermometry with scale should be continued. In particular, a consequent dramatic improvement in the the gas thermometer investigation of the quality and uniformity of patient care in scale integrity should be pressed to an the U.S. ultimate overlap with radiation thermometry. The extension of the scale through the e) Needs for Improved Thermometry in cryogenic range should be brought to Energy fruition--if necessary, through NBS initiation of an international discussion Major changes are occurring in the world- to this end. The NBS acoustic thermometry wide production and use of energy. The has stimulated a similar effort at NPL in current crisis in oil -based energy has England, where acoustic thermometry results propelled the U.S. into careful studies of are being compared with others from gas alternative sources. Most of these studies and isotherm thermometry. The NBS noise indicate that new extensions of existing and gamma-ray anisotropy thermometry may both thermometer capabilities will be necessary overlap the acoustic thermometer range, if the U.S. is to generate appreciable new offering a direct comparison of three sources of energy. fundamentally different thermometers. Take, as an example, increased The superconductive fixed point program utilization of nuclear fission reactor may provide, in addition to the five points power. Present technology, while virtually presently existing between 7.2 K and 0.5 K, free of serious accidents, still is in such further points at 18 K, 13 K, 0.1 K, and a state that remote siting of power reactors 0.2 K; an evaluation of these as potential offers the best means of widespread accep- 40 .

tance. In turn, remote siting places heavy In addition we are probing the thermometric demands on the capability for low-loss aspects of the thermocline oceanographic transfer of electric power over long dis- heat pump power generation and that of tances and into major cities; in our opinion, geothermal, controlled thermonuclear this problem calls for the introduction in reactor, and solar power plants. It should large quantity of superconductive power be noted, however, that the scientific transmission lines. These lines, composed problems appear to be by far the simplest of a central conductor coaxial with a feature of increasing the efficiency of flowing coolant and an enclosing return existing power facilitips. Economic and conductor, are of such tremendous current- political considerations here ("who pays for carrying capacity that a superconducting it?") are not trivial gaseous-helium-cooled cable the size of a man's arm can carry as much power as some f) Industrial Standards twenty high voltage towers can support. They require cryogenic installation and, In the area of industrial standards, we probably, underground siting. One cable, will undertake basically a liaison role however, is expected to supply so much which has two tasks. First, catalog the power that, for example, Brookhaven National standardization activities of the various Laboratory expects that its coming experimen- groups which were noted in Section 2, and tal switchover to 100% superconducting power second, communicate directly with the cable will not even provide test data on individual industries of Section 2 and with the ultimate capacity of a single cable. the "standards industry" so that they can The reader should note well with respect cooperate in writing standards if they so to superconductive power transmission, that desire. The next few years are certain to the entire cable except for its outer shield be critical ones for U.S. industry because will be existing at a temperature completely of the competition of foreign manufacturers, beyond the range of the present International and part of the problem will arise through Scale! international standardization. NBS Superconductive power transmission participation in the standardization process figures heavily too, in our opinion, in the by providing technical advice and effective intensive use of solar and geothermal power. liaison can benefit both the American These sources do not occur near the great consumer and U.S. industry. cities whose power needs they may fulfill, A continuing and challenging problem area so that low-loss, high capacity, long lines exists in trying to promote the production of will be a necessity for their use. higher quality, safer, and more internationally Field calibration of accurate thermom- competitive products through industrial eters used for lifetime prediction for thermometry. This is an activity which con- geothermal sources will become quite cerns primarily the group noted in Section 2. important, in our opinion. Such on-site Both of these tasks are difficult. Al- calibration will go far to break the though our experience to date may be atypical, "calibration chain" effects of long time we found that many commodity manufacturers delays in calibration and concomitant loss are interested in thermometry primarily as of instrumental accuracy. it relates to the efficiency of manufacture; In the Energy Research and Development they show little interest in thermometry Administration, an Assistant Administrator applied to the ability of the product to for Conservation has been formed. Success, perform its function or to its durability. for this group, must be measured in terms The principal reason given for this attitude of thermodynamics. A whole new order of is an economic one, and the challenge is to thermometric accuracy, convenience, and provide the means by which the economic persistence must be considered if its question can be overcome. Tools which can efforts are to be effective and appreciated. be used in this effort include a reference Household energy use as well may call service for thermometry expressly intended for thermometric improvement. The growing and publicized to offer technical informa- trend to condominium and apartment dwellings tion (on sensors and measurement techniques) dictate consideration of central power and locator information (on geographic sources for heat and light, but how to availability of calibration and engineering monitor energy consumption? laboratories) to help eliminate that portion In the area of thermometry in energy, of thermometry problems which arises through we are investigating the employment of a lack of awareness of existing technology thermometry as one tool in enhancing the and a well -planned Measurement Assurance or efficiency of present power production Measurement System Calibration program to facilities and in effective safety and sample thermometry competence in "willing", control of fast breeder reactor plants. selected facilities. 41 5. SUMMARY AND CONCLUSIONS in addition to the author, have visited various facilities on its behalf. The To summarize the state of the National assistance of all of these people is Temperature Measurement System is to gratefully acknowledged. summarize the state of the U.S. economy, its science and its technology. Thermom- etry is literally an al 1 -prevading activity. In this Study we have attempted to portray the structure of the National Temperature Measurement System, beginning with the most basic concepts, noting the sensors and other thermometry tools, and sketching the composition of the major developers and users of thermometry. We have discussed the technological and economic aspects of thermometry as they relate to the life of the average American. Finally we have presented a picture of the thermometry activities of the NBS, attempting to show the many ways in which the Bureau provides technical resources to help the nation to progress in thermometry. We conclude that temperature measure- ment and control in the U.S. is sufficiently precise and accurate to accomplish its stated objective. Areas which are deficient (medicine, new energy sources) are areas which involve new technology or in which no clear mechanism exists to focus on problems and to eliminate them. Areas in which we see particular needs developing over the next few years are: • High-temperature thermometry measurement and control • Thermometry in cryogenics • Fast-responding sensors

• Time dependent measurements • Measurements in hostile environments

• Calibration of measurement systems • Calibration of reduced-accuracy thermometers • New working relations between na- tional and international

I standardization groups Building on the foundation of this study, NBS thermometry must maintain a lively interest in the activities of groups through- out the National Temperature Measurement System and continually evaluate and assess the priorities of potential NBS programs in the light of the needs of the entire System.

6. ACKNOWLEDGEMENTS

As noted at the outset, this study was written by a single individual. However, most of the members of NBS Sections 221.03 and 221.11 have contri- buted directly to the study, and many { people, including H. H. Plumb, G. T. Furukawa, R. J. Soulen, G. W. Burns, W. S. Hurst, H. Marshak, and B. W. Mangum

• 42

i 7. REFERENCES

Section 1. Introduction [8] The chief reason for the separate de- finitions, according to Hall (Ref. 7), [1] Private communication, colleague at NBS. is that whereas the 100 °C interval between the ITS defining points was [2] The National Measurement System study known to ±.001 °C (thus the ITS degree is a collection of over twenty in- Celsius was reproducible to ±10 ppm) dividual "micro studies" of activities the thermodynamic accuracy of the in the Institute for Basic Standards, triple point was ±.01 °K so that the U. S. NBS. degree Kelvin could be known only to ±30 ppm. [3] U.S. Department of Commerce Organi- zation Order 30-2B, [9] "The Committee", Metrologia 5, 35 (1969). Section 2. Structure of the System [10] This adjective was added in 1960. [1] A short history of thermometry can be found in an article by R. B. Lindsay in [11] In fact one such adjustment has al- Vol. 3, Part 1, p. 3 of "Temperature, ready been suggested by L. A. Guildner Its Measurement and Control in Science of NBS, who points out that his gas and Industry" edited by C. M. Herzfeld, thermometer only reads 99.97 °C at Reinhold, N. Y., 1962. the boiling point of water. This and other NBS contributions to current [2] H. T. Wenzel (then Chief, Pyrometry thermometry are discussed in Section Section, NBS) in the lead article in 4.

Vol. 1 of "Temperature, Its Measurement and Control", Reinhold, N. Y., 1941. [12] Metrologia 5, 35 (1969).

[3] W. Thomson, Phil. Mag. 33, 313 (1848). [13] At NBS, for example, only five of these exist. [4] The word "clearly" here may have been somewhat optimistic, although it seems [14] International Standards Organization clear to us. Reference 2 and an article "Memento 1972". Copies of this and by C. M. Herzfeld on p. 41 of the volume other documents relating to ISO might

referred to in reference 1 both discuss be obtained from the ISO at 1, rue de this point. Varembe', 1211 Geneve 20, Switzerland. The same address applies for the lEC. [5] Even today, neither thermodynamic [15] The NBS Technical News Bulletin, thermometry nor any fixed points other February 1973, contains on p. 47 a than the triple point and the boiling short discussion of the O.I.M.L. and point of water are in common use; the beginning of U.S. participation nevertheless, the standard instruments therein. used in practical thermometry re- produce the most heavily-used portion [16] The pre-1970 history of ANSI can be of the scale within about 100 ppm of traced through "The Magazine of thermodynamic temperatures. Standards" articles in January 1967, January 1968, September 1968, [6] G. K. Burgess, J. Res. NBS, 1, 635 November, 1968, April 1969, and (1928). October 1969. Presently, ANSI activities are described in the "ANSI [7] The scale is described in an article by Reporter", published biweekly by ANSI, J. A. Hall of NPL on p. 115 of Vol. 2 1430 Broadway, New York 10018. of "Temperature, Its Measurement and Control in Science and Industry, [17] ASTM, 1916 Race Street, Philadelphia, Reinhold, N. Y., 1955. 19103, publishes the "ASTM Standard- ization News", a monthly magazine which discusses its affairs in concise but informal fashion.

43 . .

[18] ASTM Standardization News, February [32] It is not unusual for thermistors to 1973, p. 4. vary in resistance by a factor 5 or 10 over a 100 °C interval. [19] Ibid., p. 32. [33] Private communication, thermistor [20] J. A. Beattie, Vol. 2, p. 63, of manufacturer. "Temperature", Reinhold, (1955). [34] Industrial Research, June 1973, p. 64. [21] W. D. Huston, Vol. 3, Part 2, p. 949 of "Temperature", Reinhold, (1962). [35] Many of the manufacturers listed in reference 27 fabricate these sensors. [22] E. F. Moeller, on p. 162 of the volume cited in reference 1, discusses [36] K. R. Carr, Vol. 4, Part 2 of "Temper- Callendar's work in the general con- ature" 1972, p. 971. Dr. Carr has text of precision resistance privately told the author of some thermometry. additional work on this topic.

[37] Commercial ad in Research/Develop- ment, November 1973, p. 37. [23] We shall discuss further, in Section 4, the paper by J. P. Evans and [38] Private communication, thermometer 5. D. Wood, Metrologia, 7, 108 (1971). manufacturer.

[24] G. T. Furukawa, Vol. 3, Part 2, p. 320 [39] The variation in the readings found on of "Temperature", Reinhold, 1962. "drug store thermometers", often as much as several degrees Fahrenheit, is [25] A comprehensive summary of the prin- probably a consequence of economics. ciples and practice of thermoelectric We shall see.

thermometry appear in Section 1 of Vol. 3, Part 2 of "Temperature" [40] Artronix Instrumentation, in the Reinhold, 1962. December 1973, Physics Today.

[26] ASTM Handbook of Standards, Part 30, [41] Private communication. 1972.

[27] The Thomas Register (Thomas Publish- [42] Private communication. ing Company, 461 Eighth Avenue, N. Y. 10001) is a good source of information [43] Catalog of NBS Standard Reference on commercial products and manu- Materials, NBS Special Publication 260, facturers. June 1975, p. 51

[28] One instrument manufacturer estimates [44] Preparation and Use of Superconductive privately that 40-45% of all sales are Fixed Point Devices, SRM 767, NBS concerned with temperature-control Special Publication 260-44, December apparatus. 1972.

[29] See the discussion in Sections 2.2.2.1 [45] Dr. K. Churney. Private communication. and 2.5. [46] See J. Terrien, Metrologia 1, 15 (1965). [30] See Section 2.2.2.1 and figure 2.2F for references to these various [47] "Men and Measurements - United States sensors Army Metrology and Calibration Center", prepared by the Program Management [31] This is a control thermostat manu- and Administrative Office of the U. S. factured specifically for household Army Metrology and Calibration Center, use. Redstone Arsenal, Alabama 35809, April 1970.

44 [48] Ibid., p. 98. [5] Wall Street Journal, 17 July, 1973.

[49] "A Directory of Standards Laboratories" [6] Private discussion with the chief of can be obtained from NCSL Directory an experimental automotive research Committee, c/o NBS 200.01, Washington, laboratory, 1973. D. C. 20234. [7] F. B. Dent, Commerce Today, August 6, [50] Comment by the president of a 1973, p. 25. thermometer company, during a visit by an NBS staff member. [8] Deborah Shapley, Science, 182, p. 732 (1973). [51] Private communication, via the 28th Annual Calorimetry Conference; 13-15 [9] P. C. White, Science, 182, 23 November June 1973 at Worcester, Massachusetts. 1973.

[52] Electronics, October 1973, p. 88. [10] P. G. Hansel, Industrial Research/ Development, p. 8, July 1973. [53] Private communication. [11] NBS Technical News Bulletin, January [54] Private communication. 1973.

[12] Brooklyn Thermometer Co., Inc., $60;

[55] Commerce Today, April 30, 1973, pp. AMI-Therm. , $15 and $20. 30-32. [13] Nuclear News, p. 39, September 1973. [56] Discussions with AEC and ORNL personnel. [14] Conversations with the Directorate of

[57] Private cormiunication , M. Smith, LASL. Reactor Development, AEC; with the Instrumentation and Controls Division, [58] W. Vannah (of Foxboro Corp.) in NAS- AEC; and with a representative of NRC Physics Instrumentation Report, Bechtel Corporation, nuclear engineers. 1971. [15] Nuclear News, September 1973, p. 53. [59] J. G. McMillan, R. H. Pamperin, SAE Congress, Detroit, January 1972. [16] 1973 Press Handbook, TVA.

[60] Private communication, industrial [17] An example of this activity in thermal scientist. conductivity is Y. S. Touloukian,

R. W. Powell , C. Y. Ho, and P. G. [61] T. Berlincourt, in NAS/NRC Physics Klemens, Thermophysical Properties Instrumentation Report, 1971; private Research Center Data Series, 1, IFI/

communication, H. J. Kostkowski, Chief, Plenum Data Corp. , 1970. Optical Radiation Section. [18] Even here, NBS enters the picture. Section 3 Through the National Standards Reference Data Service, NBS fosters [1] J. H. V. Brown and D. J. Lowell, IEEE a large proportion of this analysis. Transactions in Biomedical Engineering, Vol. BME-19, 331 (1972). Section 4

[2] R. S. Melville, Chemical Instrument- [1] G. K. Burgess, J. Res. NBS 1 , 635 (1928). ation 4, 63 (1972). [2] "Temperature", Vol. 1, 1941, Reinhold, [3] Mai Schechter, Lab World, May 1973, New York. p. 26. [3] "Temperature", Vol. 2, 1955, Reinhold, [4] Nathan Gochman, IMR Special New York. Colloquium, NBS, 23 March, 1973.

45 ,

[4] "Precision Measurement and Cali- [20] A. Cezairliyan, J. Res. NBS 77A, 333 bration-Temperature", NBS Special (1973). Publication 300, Vol. 2, 1968. [21] E. J. Prosen and R. N. Goldberg, NBS [5] "Precision Measurement and Calibration- Interim Report 73-180, April 1973. Heat", NBS Special Publication 300, Vol. 6, 1970.

[6] "Temperature", Vol. 4, 1972, Instru- ment Society of America, Pittsburgh.

[7] "Temperature Section Activity Summary, 1974" J. F. Schooley, ed. NBS Tech- nical Note 869, June 1975.

[8] The International Practical Temper- ature Scale of 1968", Metrologia 5, 35 (1969).

[9] The minutes are published (in French) by the International Bureau of Weights and Measures, Pavilion de Breteuil, F 92310, Sfevres, France.

[10] J. F. Schooley, R. J. Soulen, Jr., and G. A. Evans, Jr., NBS Special Publication 260-44, December 1972, and R. J. Soulen, Jr., J. F, Schooley, and G. A. Evans, Jr., Rev. Sci. Instr. 44, 1537 (1973).

[11] NBS Tech. News Bulletin, February 1973, p. 47.

[12] Sad to say, OIML approved the draft recommendation over the U. S. object- ions! Therefore, the end of this particular story will come only after the U. S. has attempted to amend the draft.

[13] NBS Tech. News Bulletin, May 1973.

[14] Commerce Today, 6 August 1973, p. 10.

[15] D. Burch, NBS Standard, 5 December 1973, p. 5.

[16] NBS Dimensions, August 1973, p. 181.

[17] J. Cohen, S. Edelman, and C. Vezzetti Nature Phys. Sc. 233, 12 (1971).

[18] NBS Tech. News Bulletin, December 1972, p. 277.

[19] A. Cezairliyan, J. Res. NBS 77A, 45 (1973).

46 : .

APPENDIX A. METHODOLOGY OF THE STUDY APPENDIX B. SUMMARY OF BACKGROUND DOCUMENTS The study of the National Temperature Measurement System was accomplished by 1. "Temperature, Its Measurement and Con- utilizing several sources of information: trol in Science and Industry", Vol. 4,

. The literature of thermometry, H. H. Plumb, Editor-in-Chief, Instrument especially useful because periodic symposia Society of America, Pittsburgh, 1972. on "Temperature, Its Measurement and Control This three-part volume is an essential in Science and Industry", for years held reference for understanding the position with NBS sponsorship, provides a detailed of thermometry at the present time. A summary of active areas of thermometry. significant history of the International The journal Metrologia, trade association Practical Temperature Scale by H. periodicals, and the scientific and en- Preston-Thomas, of the National Research gineering literature all were methodically Council Laboratories in Ottawa, opens consulted in order to develop a compre- the volume, and the volume covers the hensive overview of the state of thermometry topi cs

in the U. S. . Temperature Scales and Fixed Points

. Personal knowledge on the part of NBS . Radiation Thermometry staff members provided individual insights Resistance Thermometry

into the strengths and needs of U. S. . Magnetic and Quantum Thermometry

thermometry. These were augmented by a . Temperature Control and Calibration

series of visits to less well--known areas . Bridges and Potentiometers

of thermometry applications in order to . Thermoelectric Thermometry

build a collection of personal contacts which . Biology and Medicine will continue to provide information through . Geophysical and Astrophysi cal telephone and written communication. Thermometry Participation in scientific and 2. "The International Practical Temperature technical meetings, as well as the delivery Scale of 1958" Metrologia 5 35 (1969). of talks, instructional seminars, and tours, This document is the defining text for provided up-to-date information on areas the current International Temperature where thermometry is advancing, where it is Scale. It contains the assigned values stymied for lack of fundamental research and definitions of the scale fixed points results, and areas in which future efforts and the descriptions of the standard are most likely to be rewarding. interpolating sensors. Economic data were obtained from per- 3. "Standard Industrial Classification sonal contacts in government and industry. Manual", Office of Management and Budget. The rough impressions of many staff The SIC code defines industries in members were integrated to build a picture accordance with the composition and of the System, in an early draft of the study, structure of the economy and covers the and later efforts focused on areas where un- entire field of economic activities. certainty remained. Successive visits, 4. "U. S. Industrial Outlook Series" telephone discussions and correspondence Domestic and International Business generally served to bring out the requisite Administration, U. S. Department of information, so that we did not resort to Commerce. This series of reports is formal polling techniques. oriented toward the future economic patterns of the U. S. Detailed analysis of some 200 industries, covering about 85% of the total value of shipments in

the U. S. , are presented with projections to 1980. 5. "Annual Survey of Manufacturers Series" Social and Economic Statistics Adminis- tration, Bureau of the Census, U. S. Department of Commerce. This series con- tains reports of product shipments for perhaps 1000 classes of manufactured pro- ducts. It is a very useful tool for the economic examination of particular

i ndustries 6. "The U. S. Government Organization Manual". General Services Administration. A very useful guidebook to the Federal bureaucracy

47 .

APPENDIX B. SUMMARY OF BACKGROUND PHYSIKALISCH-TECHNISCHE REICHSANSTALT DOCUMENTS 7. "The Thomas Guide to Manufacturers" INTRODUCTION Publishing Company, N.Y. 10001). (Thomas The experience of the Bureau of Standards, A comprehensive guide to products and as of the National Physical Laboratory and organizations of U. S. industry. of the Reichsanstalt , has for many years 8. "A Directory of Standards Laboratories", past indicated the necessity, for in- Prepared by the National Conference of dustrial purposes, of international agree- Standards Laboratories. This directory ment on a scale of temperatures ranging serves as a "classified index" of U. S. from that of liquid oxygen to that of standards laboratories. luminous incandescent bodies. As a result of discussion extending over a considerable period, agreement has been reached by the three laboratories, subject to possible APPENDIX C. RP 22, NBS Journal of Research, minor drafting amendments on the attached Vol. 1 , 635 (1928) specification for a practical scale, as THE INTERNATIONAL TEMPERATURE affording a satisfactory basis on which SCALE, George K. Burgess uniformity in certification of temperature In 1911, the directors of the national measurements for industrial purposes may laboratories of Germany, Great Britain, and be maintained. the United States agreed to undertake the It is to be understood that this proposal unification of the temperature scales in does not purport to replace the absolute use in their respective countries. This temperature scale which it is recommended endeavor was approved formally in 1913 by should be adopted, on principle, by the the Fifth General Conference on Weights and International Conference on Weights and Measures, and in 1921 the Sixth General Measures. It is intended merely to re- Conference voted to expand the field of present this scale in a practical manner activities of the International Committee with sufficient accuracy to serve the every and International Bureau by including day needs of the laboratories for the pur- physical constants, such as standard temper- pose of industrial certifications, and is atures, and the coordination of results to be regarded as susceptible of revision obtained in other institutions. This action and amendment as improved and more accurate was duly ratified by the several govern- methods of measurement are evolved. ments adhering to the conference. It is anticipated that this scale will Finally, in 1927, the Seventh General shortly be adopted by the three laboratories Conference, representing 31 nations, on the for the purposes indicated, and the attach- recommendation of the three laboratories ed draft is presented to the conference above mentioned, which had consulted with for consideration, with the recommendation the Leiden Cryogenic Laboratory as to low that it should be officially adopted with temperatures, adopted unanimously the follow- such amendments, if any, as may be agreed ing resolution, proposed by the International on, as the best practical realization at Committee of Weights and Measures: the present time of the ideal thermometrie Le Comite, reconnaissant 1' importance scale. pratique de la representation d'une 6che11e thermom^trique internationale, recommande Part I. DEFINITION OF THE INTERNATIONAL a la Conference d' accepter, a titre pro- TEMPERATURE SCALE visoire, les repferes de temperature, les 1. The Thermodynamic Centigrade Scale, formules d' interpolation et les methodes on which the temperature of melting ice, de mesure proposes d'un commun accord par and the temperature of condensing water les trois laboratories nationaux vapor, both under the pressure of one d'Allemagne, des Etats-Unis d'Amerique, et standard atmosphere, are numbered 0° and de Grande-Bretagne. 100°, respectively, is recognized as the Le Comite recommande aussi que le texte fundamental scale to which all temperature annexe soit maintenu a 1 'etude dans le measurements should ultimately be referable. programme des Conferences speciales de 2. The experimental difficulties in- Thermometrie qui seront tenues sous ses cident to the practical realization of the auspices thermodynamic scale have been expedient to The English text is as follows: adopt for international use a practical TEXT CONCERNING THE ADOPTION OF AN INTER- scale designated as the International Tem- NATIONAL TEMPERATURE SCALE SUBMITTED FOR perature Scale. This scale conforms with DISCUSSION BY THE BUREAU OF STANDARDS, the thermodynamic scale as closely as is NATIONAL PHYSICAL LABORATORY, AND possible with present knowledge, and is designed to be definite, conveniently and 48 .

accurately reproducible, and to provide atmospheric pressure means for uniquely determining any temper- (gold point) 1063 ature within the range of the scale, thus Standard atmospheric pressure is promoting uniformity in numerical state- defined as the pressure due to a column of ments of temperature. mercury 760 mm high, having a mass of 3. Temperatures on the international 13.5951 g/cm^, subject to a gravitational scale will ordinarily be designated as "°C", acceleration of 980.665 cm/sec^ and is equal but may be designated as "°C.(Int.)" if it to 1 ,013,250 dynes/cm2. is desired to emphasize the fact that this It is an essential feature of a practical scale is being used. scale of temperature that definite numerical 4. The International Temperature Scale values shall be assigned to such fixed is based upon a number of fixed and re- points as are chosen. It should be noted, producible equilibrium temperatures to however, that the last decimal place given which numerical values are assigned and for each of the values in the table is signi- upon the indications of interpolation ficant only as regards the degree of instruments calibrated according to a reproducibility of that fixed point on the specified procedure at the fixed temper- International Temperature Scale. It is not

atures . understood that the values are necessarily 5. The basic fixed points and the known on the Thermodynamic Centigrade Scale numerical values assigned to them for the to the corresponding degree of accuracy. pressure of one standard atmosphere are 7. The means available for interpolation given in the following table, together lead to a division of the scale into four with formulas which represent the temper- parts

ature (t ) , as a function of vapor pressure (a) From the ice point to 660 °C the (p) over'^the range 680 to 780 mm of mercury. temperature t is deduced from the resistance 6. Basic fixed points of the Inter- R. of a standard platinum resistance thermom- national Temperature Scale-- °C eter by means of the formula (a) Temperature of equilibrium = Rq (1 + At + Bt2). between liquid and gaseous

oxygen at the pressure of The constants R , A, and B of this formula one standard atmosphere are to be deterfnined by calibration at the (oxygen point)--- -182.97 ice, steam, and sulphur points, respectively. The purity and physical condition of the t = tyeo +0.0126 (p-760) -0.0000065 platinum of which the thermometer is made P (p-760)2 should be such that the ratio R,/R shall ° (b) Temperature of equilibrium not be less than 1.390 for between ice and air-saturated t = 100° and 2.645 for t = 444.6°. water at normal atmospheric (b) From -190° to the ice point, the pressure (ice point) — 0.000 temperature t is deduced from the resistance (c) Temperature of equilibrium R. of a standard platinum resistance thermom- between liquid water and its eter by means of the formula vapor at the pressure of one = + 1^^ + Bt2 + C (t-100)t3]. standard atmosphere (steam h D

point) 100.000 The constants R , A, and B are to be de- termined as specified above, and the t„ = t76o + 0.0367 (p-760) additional constant C is determined by cali- P -0.000023 (p-760)2 bration at the oxygen point. (d) Temperature of equilibrium The standard thermometer for use below 0° between liquid sulphur and C must, in addition, have a ratio R./R less ° its vapor at the pressure than 0.250 for t = -183°. ^ of one standard atmosphere (c) From 660 °C to the gold point, the (sulphur point) 444.60 temperature t is deduced from the electro- motive force e of a standard platinum vs = tyeo + 0.0909 (p-760) platinum-rhodi urn thermocouple, one junction ^ - 0.000048 (p-760)2 of which is kept at a constant temperature (e) Temperature of equilibrium of 0 °C while the other is at the temper- between solid silver and ature t defined by the formula liquid silver at normal e = a + bt + ct^. atmospheric pressure (silver point) 960.5 The constants a, b, and c are to be de- (f) Temperature of equilibrium termined by calibration at the freezing between solid gold and point of antimony, and at the silver and liquid gold at normal gold points. 49 (d) Above the gold point the temperature preferable, although gas may be used, but t is determined by means of the intensity the source of heat and all good conducting J2 of monochromatic visible radiation of material in contact with it must terminate wavelength x cm, emitted by a black body at at least 4 cm below the free surface of the the temperature t2, to the intensity or liquid sulphur. Above the source of heat radiation of the same wavelength emitted by the tube is surrounded with insulating a black body at the gold point, by means material. Any device used to close the end of the formula of the tube must allow a free opening for

J2 C2 1 1 equalization of pressure. ] = " The sulphur should contain not over 0.02 ^°9e — [t;336 (t + 273) -jY percent of impurities. Selenium is the im- The constant C2 is taken as 1.432 cm de- purity most likely to be present in grees. The equation is valid if (t + 273) quantities sufficient to affect the temper- is less than 0.3 cm degrees. ature of the boiling point. The radiation shield is cylindrical and Part II. RECOMMENDED EXPERIMENTAL open at the lower end, and is provided with PROCEDURE a conical portion at the top, to fit closely to the protecting tube of the thermometer. 1. OXYGEN The cylindrical part is 1.5 to 2.5 cm larger The temperature of equilibrium of liquid in diameter than the protecting tube of the and gaseous oxygen has been best realized thermometer and at least 1 cm smaller in experimentally by the static method, the diameter than the inside of the boiling tube. oxygen vapor-pressure thermometer being The cylinder should extend at least 1.5 cm compared with the thermometer to be beyond each end of the thermometer coil. standardized in a suitable low temperature There should be ample opening at the top of bath. the cylindrical and below the conical portion 2. ICE to permit free circulation of vapor. The The temperature of melting ice is real- inner surface of the shield should be a ized experimentally as the temperature at poor reflector. The shield may be made of which pure, finely divided ice is in sheet metal, graphite, etc. equilibrium with pure, air-saturated water In standardizing a thermometer the under standard atmospheric pressure. The sulphur is heated to boiling and the heating effect of increased pressure is to lower so regulated that the condensation line is the freezing point to the extnt 0.007 °C at least 1 cm above the top of the insul- per atmosphere. ating material. The thermometer with its 3. STEAM radiation shield is inserted in the vapor, The temperature of condensing water vapor and when the line of condensation again is realized experimentally by the use of a reaches its former level simultaneous ob- hypsometer so constructed as to avoid servations of resistance and barometric superheat of the vapor around the thermom- pressure are made. In all cases care should eter, or contamination with air or other be taken to prove that the temperature is impurities. If the desired conditions have independent of vertical displacements of the been attained, the observed temperature thermometer and shield. should be independent of the rate of heat 5. SILVER AND GOLD supply to the boiler, except as this may For standardizing a thermocouple, the affect the pressure within the hypsometer, metal to be used at its freezing point is and of the length of time the hypsometer contained in a crucible of pure graphite, has been in operation. refractory porcelain, or other material 4. SULPHUR which will not react with the metal so as to For the purpose of standardizing re- contaminate it to an appreciable extent. sistance thermometers, the temperature of Silver must be protected from access of condensing sulphur vapor is realized by oxygen while heated. adherence to the following specifications The crucible and metal are placed in an relating to boiling apparatus, purity of electric furnace capable of heating the sulphur, radiation shield, and procedure. contents to a uniform temperature. The boiling-tube is of glass, fused silica, The metal is melted and brought to a uni- or similar material, and has an internal form temperature a few degrees above its diameter of not less than 4 or more than melting point, then allowed to cool slowly 6 cm. The vapor column must be sufficient- with the thermocouple immersed in it as ly long that the bottom of the radiation described in the next paragraph. shield is not less than 6 cm below the top The thermocouple, mounted in a porcelain of the heat insulating material surround- tube with porcelain insulators separating ing the tube. Electric heating is the two wires, is immersed in the molten 50 metal through a hole in the center of the minations is given as 630.5°, but the tem- crucible cover. The depth of immersion perature of any particular lot of antimony should be such that during the period of which is to be used for standardizing the freezing the thermocouple can be lowered or thermocouple is to be determined with a raised at least 1 cm from its normal standard resistance thermometer. position without altering the indicated The procedure to be followed in using the emf by as much as 1 microvolt. During freezing point of antimony as a fixed tem- freezing, the emf should remain constant perature is substantially the same as that within 1 microvolt for a period of at specified for silver. Antimony has a mark- least five minutes. ed tendency to undercool before freezing. As an alternative to displacing the The undercooling will not be excessive if couple, as a means of testing the absence the metal is heated only a few degrees of the influence of external conditions above its melting point and if the liquid upon the observed temperature, both freez- metal is stirred. During freezing the ing and melting points may be observed and temperature should remain constant within if these do not differ by more than 2 micro- 0.1° for a period of at least five minutes. volts, the observed freezing point may be 8. SECONDARY POINTS considered satisfactory. In addition to the basic fixed points, 6. THE STANDARD PLATINUM RESISTANCE the temperatures of a number of other points THERMOMETER are available and may be used in the cali- The diameter of the wire should not be bration of secondary temperature measuring smaller than 0.05 or larger than 0.2 mm. instruments. These points and their tem- The platinum wire of the thermometer must peratures on the international scale are be so mounted as to be subject to the mini- listed below. The temperatures given are mum of mechanical constraint, so that those corresponding to a pressure of one dimensional changes accompanying changes of standard atmosphere. The formulas for the temperature may result in a minimum of variation of vapor pressure with temperature mechanical strain being imposed upon the are valid for the range from 680 to 780 mm. pi atinum. °C The design of the thermometer should be Temperature of equilibrium between such that the portion, the resistance of solid and gaseous carbon dioxide-- -78.5 which is measured, shall consist only of platinum, and shall be at the uniform temper- t^ = tyeo + 0.1443{t^ + 273.2) ature which is to be measured. This may be P logio(p/760) P accomplished by either of the accepted Temperature of freezing mercury -38.87 systems of current and potential, or com- Temperature of transition of pensating leads. sodium sulphate — 32.38 After completion, the thermometer should Temperature of condensing be annealed at a temperature of at least napthalene vapor 217.96 660°. 7. THE STANDARD THERMOCOUPLE t = tyeo + 0.208(t + 273.2) The platinum of the standard couple shall P logio (p/760) P be of such purity that the ratio R^./R is ° Temperature of freezing tin- 231 .85 initially not less than 1.390 for Temperature of condensing t = 100°. The alloy is to consist of 90 benzophenone vapor 305.9 percent platinum with 10 percent rhodium. The completed thermocouple must develop an t = tyeo + 0.194(t + 273.2) electromotive force, when one junction is P logio (p/760) P at 0° and the other at the freezing point Temperature freezing cadmium 320.9 of gold, not less than 10,200 nor more than — Temperature of freezing lead 327.3 10,400 international microvolts. The dia- — Temperature of freezing zinc 419.45 meter of the wires used for standard — Temperature of freezing antimony-- 630.5 thermocouples should lie between the values Temperature of freezing copper 0.35 and 0.65 mm. in a reducing atmosphere 1,083 The freezing point of antimony, specified Temperature of freezing palladium 1,555 for the standardization of the thermocouple, Temperature of melting tungsten 3,400 lies within the range of 0° to 660° where The Bureau of Standards, therefore, in the international scale is fixed by the common with the other national laboratories, indications of the standard resistance will use until further notice in its thermometer, and the numerical value of scientific work and for the calibration of this temperature is therefore to be de- instruments, the standard temperatures, termined with the resistance thermometer. interpolation formulas, and methods of In the appendix the result of such deter-

51 . .

measurement as laid down above by the 18. T. B. Douglas, 425.03: Member ASTM General Conference of Weights and Measures Committee E-20. on October 4, 1927. It is recommended that 19. J. F. School ey. Chief 221.11: Member scientific workers elsewhere conform to the ASTM Committee E-20. International Temperature Scale as above set forth. It is expected that international thermom- APPENDIX E. RESISTANCE THERMOMETERS etric conferences will be called, as 1. Doped germanium. (References: S. A. occasion requires, by the International Friedberg, p. 359, Vol. 2, "Temperature", Committee on Weights and Measures, so that Reinhold (1955). J. E. Kunzler, T. H. this temperature scale may be revised as Geballe, and G. W. Hull, Jr., p. 391, the need arises. Part 1, Vol. 3 "Temperature, Its Measure- Washington, September 12, 1928. ment and Control in Science and Industry", Reinhold, N. Y. (1962); W. R. G. Kemp, J. G. Collins, C. P. Pickup, R. APPENDIX D. Muijlwijk, p. 85, Part 1 , Vol . 4 "Temper- NBS Participation in Voluntary Standards ature", ISA, Pittsburgh, 1972). The Bodies' Thermometry Activities doped germanium resistance thermometer 1. R. P. Hudson, Chief 221: U. S. repre- is widely used to preserve the cryogenic sentative to 10th CCPM Advisory temperature scale in the same way that Committee on Thermometry session; platinum resistance is used at higher General Chairman of 5th Temperature temperatures. The reproducibility of Symposium, and Chairman of CCT Working selected germanium thermometers is better

Group 5. than 1 millikelvin from 4 K down to .02 2. H. H. Plumb, Program Chairman of 5th K, and the devices are useful to temper- Temperature Symposium and Editor-in- atures as high as 20 K. Basically semi- Chief of the Proceedings; Member conductors, these resistors show a

Working Group 1 of CGPM Advisory modified exponential temperature depen- Committee on Thermometry; Member NBS dence of resistivity, so that the useful Commodity Standards Committee; Member span of temperatu»-e is about a factor ASTM Committee E-20; Member, ISA. ten or twenty for a given unit. In 3. L. A. Guildner, 221.11: Member, Work- general they show pronounced magnetic ing Group 3 of CGPM Advisory Committee field dependences (L. J. Neuringer, on Thermometry. A. J. Perlman, L. G. Rubin, Y. Shapira, 4. G, W. Burns, 221.11: Member, ASTM RSI, 42, 9 (1971)). Committee E-20; Member ANSI Committee 2. Carbon radio resistors. (References: C-96; Member, ISA. J. R. Clement, E. H. Quinnell, RSI 23, 5. G. T. Armstrong, Chief 316.02: Member 213 (1952); W. C. Black, Jr., W. R. ASTM Committee E-20. Roach, J. C. Wheat! ey, RSI, 35, 587 6. W. S. Hurst, 221.11: Member ANSI (1964); A. C. Anderson, p. 773, Part 2, Committee C-96; Member, ISA. Vol. 4, "Temperature", ISA, Pittsburgh, 7. J. A. Wise, 221.11: Member ASTM 1972. H. Weinstock, J. Paupia, p. 785, Committee E-20. Ibid.) The carbon radio resistor is a 8. C. A. Douglas 232.11: Member, lES. semiconductor whose resistivity, like 9. J. P. Evans, 221.11: Member, ISA. that of doped germanium, obeys a modi- 10. R. K. Kirby, 312.06: Member ASTM fied exponential dependence. Its Committee E-1 reproducibility is inferior to that of 11. F. R. Kotter, 211.06: Member ASTM the germanium, but its trivial cost Committee on Temperature in Power allows its use in low-precision monitor- Systems ing and control applications below 100 K. 12. R. H. Kropschott, 275.00: Member Like germanium thermometers, carbon ANSI Committee C-16. thermometers are quite magnetroresisti ve. 13. L. R. Moffitt, 272.00: Member ANSI 3. Metals with small amounts of magnetic Committee C-105. impurities. (Reference: A. R. Harvey, 14. F. J. Powell ,462.00: Member ASTM S. Legvold, and D. T. Peterson, Phys. Committee on Heat Transfer. Rev. B4 400 3 (1971 )). A small temper- 15. R. L. Powell, 275.00: Member ASTM ature dependent resistivity which Committee E-20. occurs in Kondo alloys continues to

16. Richard Raybold, 425.03: Chairman, vary at temperatures of 1 K and below, Instrumentation Subcommittee, Oceano- raising the possibility of secondary graphic Measurements Society. thermometry by this means. 17. J. L. Riddle, 221.11: Member, ASTM 4. Metal films. (Reference: A. J. Laderman, Committee E-20. G. J. Hecth, and A, K. Oppenheim, Vol. 3, Part 2, p. 943 of "Temperature", Rein- APPENDIX F. OTHER TEMPERATURE SENSORS IN hold, 1962.) Metal fi Im thermometry MORE OR LESS COMMON USE can be used at temperatures from 1. Paramagnetic susceptibility (/)• cryogenic to the melting point of the Reference: T. C. Cetas and C. A. metal. The sensor is quite fast-res- Swenson, p. 57, Part 1, Vol. 4, "Temper- ponding, but generally not stable in ature", ISA, Pittsburgh, 1972. resistance. 2. Liquid vapor pressure. Reference: 5. Thermistors. (References: Sections 29 Section I-B, pp. 121-168, Ibid. and 30 by C. R. Droms and H. B. Sachse 3. Noise. Reference: R. A. Kamper, p. 349, Part 339 and p. 347 of in Vol. 3, 2, p. Ibid. "Temperature", Reinhold, 1962.) Basic- 4. Acoustic. References: (High temper- ally semiconductors, these sensitive atures) G. L. Innes, p. 689, Ibid.; sensors have come into extremely wide (Low temperatures) A. R. Colclough, use at temperatures up to 300 or 400 °C p. 365, Ibid. because of their small size and high 5. Resonant frequency of quartz. Reference: sensitivity. They will be further A. Benjaminson, and F. Rowland, p. 701, discussed in Section 3. Ibid. 6. Miniature platinum thermometers. 6. Electronic junction devices. Section (Reference: D. A. Lucas, Vol. 4, IV-F, pp. 1097-1136, Part 2, Ibid. Part 2, p. 963 of "Temperature", ISA, 7. Nuclear resonance. Section V-A, pp. These millimeter-size 1972.) 1197-1238, Ibid. thermometers can display repro- ducibilities of 10 ppm at the triple point of water, and can be used from 13.8 K to 500 K. 7. Nickel resistance thermometers. (Reference: D. A. Grant and W. F. Hickes, Vol. 3, Part 2, p. 305, "Temper- ature", Reinhold, 1962.) The use of nickel resistors to meet the ruggedness and interchangeabil ity demands of in- dustry has resulted in their growing use as thermometers. 8. Ceramic thermometers. (References: A. R. Anderson and T. M. Stickney,

Vol . 3, Part 2, p. 361 , and W. N. Lawless, Vol. 4, Part 2, p. 1143, "Temperature", Reinhold, 1962, and ISA, 1972, respectively). The first reference deals with ceramics as high- temperature resistance thermometers, and the second, with ceramics as cryogenic capacitors, thermometers which are nearly independent of magnetic fields.

53

)

NBS-114A iREv. 7-73)

U .S. DEPT. OF COMM. L PL!HLI( Al lON OR Kl^PORT NO. 2. Cuiv 't Accession 3. Recipient's Accession No. BIBLIOGRAPHIC DATA No. SHEET NBSIR75-932

4. TITLE AND SISHTH LI: 5, Put^li cation Date December 1975 The National Measurement System for Temperature 6. Performing Organization ( ode

221.11

7. AUTHOR(S) 8. Performing Organ. Report No. James F. Schooley NBSIR75-932 9. PKRFORMINC, ORGANIZATION NAME AND ADDRESS

NATIONAL BUREAU OF STANDARDS 2210110 DEPARTMENT OF COMMERCE 11. Contract/Grant No. WASHINGTON, D.C. 20234

12. Sponsoring Organization Name and Complete Addres.s (Street, City, State, ZIP) 13. Type of Report & Period Covered Same as Item 9. Final

14. Sponsoring Agency Cod(-

15. SUPPLEMENTARY NOTES

16. ABSTRACT (A 200-word or less factual summary ol most si^ilicant information. If document includes a significant

bibliography or literature survey, mention it here.)

The National Temperature Measurement System reaches quite literally into all phases of American life. Comfort control in home, school, office and factory, health care, manufac- turing, food preparation and storage, and all forms of powered transportation are just a few ofthe many facets of America's human activities that deoend for their trouble-free operation on the existence of reliable thermometry. The Heat Division of the National Bureau of Standards collaborates with the national laboratories of other nations in establishing an International Practical Temperature Scale which represents thermodynamic accuracy insofar as current scientific practice permits. The is NBS the only U.S. agency bearing this responsibility; in addition, only the NBS is responsible for disseminating the International Scale to U.S. scientific and technical activities. The Study of the National Temperature Measurement System shows that the NBS continues to contribute energetically to the quality of the International Scale. It further indicates that NBS maintains a consistent and deliberate effort not only to provide access to the Scale at the several levels required by U.S. science and technology, but also to participate in solving special problems in thermometry. These are problems for which the nature of NBS as an objective, expert resource in thermometry uniquely qualifies it to furnish effective, practical solutions. In briefly describing the many aspects of the National Temperature Measurement System, this Study attempts to portray its great diversity of products, services, and people. The System is largely organized, as might be expected, for the purpose of economic gain; the NBS enters the picture only in those areas where highly accurate or intricate thermometry is essential.

17. KEY WORDS (six to twelve entries; alphabetical order; capitalize only the first letter of the first key word unless a proper name; separated by semicolons Data; International Practical Temperature Scale; Measurements; National Measurement System; Standards; Thermometry