V.8.00

STERIL Wall-mounted UVGI air sterilizer

SFA2030F // SFA2045F CONTENTS

SECTION PAGE

1- SYMBOLS 2

1.1- WARNING IDENTIFICATION 3

2- GENERAL WARNINGS 3

2.1- GENERAL CHECK 4

2.2- TRANSPORT DAMAGE IN ITALY FREIGHT PRE-PAID 4

2.3- TRANSPORT DAMAGE OUTSIDE ITALY FREIGHT PRE-PAID 5

2.4- TRANSPORT AND STORAGE CONDITIONS 5

3- INSTALLATION SAFETY RULES 6

3.1- WORKING ENVIRONMENT 7

4- SAFETY REQUIREMENTS 8

4.1- SAFETY REQUIREMENTS 9

5- PRODUCT DESCRIPTION 10

5.1- INTENDED USE 10

5.2- STANDARDS/APPROVALS 11

5.3- TECHNICAL SPECIFICATIONS 12

5.4- NAMEPLATES 13

5.5- PRODUCT PROFILE 14

5.6- STRUCTURE OF AIR TREATMENT 15

6- INSTALLATION 15

6.1- ELECTRICAL PREPARATION 15

6.2 - WALL MOUNTING 16

6.3 - RECOMMENDED POSITIONING IN A DENTAL TREATMENT ROOM 18

7- INSTRUCTIONS FOR USE 18

7.1 - REMOTE CONTROL OVERVIEW 19

7.2 - INSERTING REMOTE CONTROL BATTERIES 19 7.3 - RESET WEEK 20 ENGLISH

7.4 - SETTING THE TIME 20

7.5 - LCD DISPLAY 21

7.6 - MEMORY FUNCTION 21

7.7 - SETTINGS INSTRUCTIONS 22

8- CLEANING 29

8.1- RECOMMENDED DISINFECTANT 29

9- ROUTINE MAINTENANCE 30

10- TROUBLESHOOTING 31

11- EXTRAORDINARY MAINTENANCE 32

11.1- MAINTENANCE TABLE RESERVED FOR TECHNICAL SUPPORT 33

11.2- TECHNICAL TROUBLESHOOTING 34

11.3- REPLACING THE PHOTOCATALYTIC FILTER 35

11.4- CLEANING OR REPLACING THE UVC LAMP 36

11.5- REPLACING THE FUSES 37

12- RELATED ITEMS 38

13- DISPOSAL 38

14- WARRANTY 39

DECLARATION OF CONFORMITY 40

CERTIFICATE OF WARRANTY 41

1 1- SYMBOLS

ENGLISH CE MARKING. This product bears the CE mark according to the rules indicated in the standards EN 60335 - CEI EN 62233 - CEI EN 61000 - CEI EN 55014 and the directives 2014/30/EU - 2014/35/EU.

Protection type against direct and indirect contacts: Class I.

EARTHING SYMBOL. This symbol indicates the presence of a protective earthing contact. INSTRUCTIONS FOR OPERATION. This symbol on the product indicates that the user ma- nual must be consulted before using the equipment. HAZARD SYMBOL. It indicates a dangerous situation that might result in moderate or severe/fatal injuries unless avoided. WARNING SYMBOL. It indicates a hazardous situation that might result in moderate to minor injury or property damage. DO NOT TOUCH WITH WET HANDS SYMBOL. This symbol indicates that you should not touch the device with wet hands.

UVC HAZARD SYMBOL: artificial optical radiation (AOR) hazard

GENERAL INFORMATION SYMBOL. This symbol indicates information that allows you to use the device more effectively. DISPOSAL SYMBOL. This symbol found on the product indicates that the equipment cannot be treated as normal household waste in accordance with Directives 2002/95/EC, 2002/96/EC and 2003/108/EC. FRAGILE SYMBOL. This symbol indicates that the product inside the box is fragile. Avoid impacts. PROTECT FROM SYMBOL. This symbol indicates to keep the box away from water and moisture. UP SYMBOL. Transport and store the product only in the direction indicated by the arrows. DO NOT STACK SYMBOL.

TEMPERATURE LIMIT SYMBOL. From -10° to 45°C.

HUMIDITY LIMIT SYMBOL. From 10 to 90%. ATMOSPHERIC PRESSURE LIMIT SYMBOL. From 500 to 1060 hPa.

2 1.1- WARNING IDENTIFICATION

It indicates a dangerous situation that might result in moderate or severe/fatal injuries unless avoided.

It indicates a dangerous situation that might result in moderate or mild injuries or da- ENGLISH mage to property unless avoided.

This symbol indicates that you should not touch the device with wet hands.

This symbol indicates an artificial optical radiation (AOR) hazard. Possible damage to eyes and skin.

This icon represents useful information. 2 - GENERAL WARNINGS

• Read this manual carefully before carrying out any operation on the product, follow the warnings contained in it and keep it for future reference. • This manual is intended to provide the user with instructions for correct installation and use of the product. • The product must be used in compliance with the procedures contained in the user ma- nual and never for purposes other than those specified therein. • The user is legally responsible for the installation, operation and maintenance of the equipment itself. • The manual describes all product versions and optionals, therefore not all instructions are applicable to your product. • The information, technical specifications, illustrations contained in this publication • are not binding. • Tecnomed Italia s.r.l. has adopted a policy of constant product improvement and reser- ves the right to make changes to this manual or to the product itself without prior notice, as long as such changes do not affect safe use of the device. • Keep the manual within reach. • Furthermore, any kind of reproduction or appropriation of the text and/or images in the manual is absolutely forbidden, therefore some instructions, specifications and images contained in this manual may differ slightly from the product purchased. • All the material contained in this manual is owned by Tecnomed Italia and/or by • companies representing it. The images are not binding and are purely illustrative. • If the appliance is resold, it must be sent to the new owner with this manual. • The original text of this manual is in English.

DANGER! Modifying the appliance is strictly forbidden. Tecnomed Italia s.r.l. assumes NO responsibility for personal injury and/or direct or indirect property damage resulting from non-compliance with this requirement.

DANGER! Tecnomed Italia s.r.l. assumes no responsibility, express or implied, and cannot be held liable for injuries to persons and/or direct or indirect damage to property, resulting from the non-observance of what is reported in the entire manual and/or as a result of incor- rect procedures related to the installation and/or use of the appliance and its accessories and/or resulting from poor and/or lack of cleaning and/or maintenance.

3 2.1 - GENERAL CHECK

ENGLISH WARNING! Upon delivery of the product, check that there are no signs of impact andthat the packaging is intact. Otherwise, follow the indications in paragraph 2.2 or 2.3. 2.2 - TRANSPORT DAMAGE IN ITALY FREIGHT PRE-PAID Upon delivery, if the package/parcel shows visible signs of damage, the recipient must or may sign the delivery note conditionally, reserving the right to claim for any damages. Ita- lian law prescribes that “the unconditional acceptance of delivery of the goods transpor- ted by the consignee, together with the payment of what is due to the carrier, extinguishes the causes of action arising from the contract, except in case of fraud or gross negligence of the carrier” (art. 1698 Italian Civil Code). To conditionally accept a parcel it is necessary to use specific fields in the document (delivery note) that the carrier will ask you to sign upon delivery. Each carrier has its own procedure for signing conditionally, you will need to ask them how to proceed. If the packaging is visibly damaged upon delivery, proceeds follows: 1. Check the appearance and the condition of the packages. Check that it is not da- maged, intact, and that everything is there: if the document indicates there are several packages, check that they are all there. The recipient and the employee of the transport company sign the delivery note adding that a more detailed inspection is required, clearly stating the reason: - “goods conditionally accepted subject to detailed inspection” - “package with packaging tape added, goods conditionally accepted subject to detai- led inspection” - “cardboard package with obvious signs of warping/folding, goods conditionally accep- ted subject to detailed inspection” - “package externally damaged and/or partially opened, goods conditionally accepted subject to detailed inspection” 2. Leave the product and the packaging as they are. Document the damaged package/ parcel with photos and/or videos. 3. Do not use the product. 4. Report the damages to the transport company. 5. Report the damage to Tecnomed Italia s.r.l. ([email protected]). 6. In no case must the product be returned to Tecnomed Italia s.r.l. before having received a reply and authorisation to do so. 7. Send the signed delivery note to Tecnomed Italia srl. 8. Leave the product and the packaging as they are. 9. Do not use the product.

Note: If you are concerned that there may be damage hidden and not visible from the outside, please conditionally sign anyway.

If the product is damaged, but no damage to the packaging was detected at the time of delivery, follow these steps: 1. Inform the transport company no later than 7 days after delivery. 2. Report the damage to Tecnomed Italia s.r.l. ([email protected]) by sending visual proof of the unpacked product. 3. Leave the product as it is

WARNING! If the recipient fails to comply with any of the provisions stated above, the damage shall be considered to have arisen after delivery.

4 2.3 - TRANSPORT DAMAGE OUTSIDE ITALY FREIGHT PRE-PAID

WARNING! Tecnomed Italy srl shall not be liable for damages occurred during transit. Check the goods immediately upon receipt!

If the packaging is visibly damaged upon delivery, follow these steps: ENGLISH

1. The recipient must acknowledge the loss or damage on the delivery note. The recipient and the employee of the transport company sign the delivery note. Only on the basis of this proof can the recipient assert the claim for replacement for damage by the transport com- pany. 2. Leave the product and the packaging as they are. 3. Do not use the product.

If the product is damaged, but no damage to the packaging was detected at the time of delivery, follow these steps: 1. Notify the transport company no later than 30 days after delivery. 2. Leave the product and the packaging as they are. 3. Do not use the faulty product.

WARNING! If the recipient fails to comply with any of the aforementioned provisions, the damage shall be considered as if arising after delivery.

WARNING! If the goods travel carriage forward, the responsibility lies between the buyer and the carrier. Tecnomed Italia will have no responsibility.

2.4 - TRANSPORT AND STORAGE CONDITIONS

The product inside the packaging is fragile. Avoid shocks.

Keep the package away from water and humidity.

Transport and store the product only in the direction indicated by the arrows.

Do not stack.

Temperature limit: from -10° to 45° C.

Humidity limit: from 10 to 90%.

Atmospheric pressure limit: from 500 to 1060 hPa.

5 3- INSTALLATION SAFETY RULES ENGLISH DANGER! Do not operate the appliance if light shines out of it. IT MAY RESULT IN EYE AND SKIN DAMAGE.

DANGER! The product is intended for indoor use only, it is not suitable for wet or humid environments.

DANGER! Do not install the appliance in places where there are anaesthetic or flamma- ble gases, or in places with harmful chemical air.

DANGER! The appliance must not be used in a potentially explosive atmosphere.

DANGER! You must not cover, block or obstruct, even partially, the air inlet and outlet vents- doing so will compromise the effectiveness of the disinfection.

DANGER! Install the appliance in a place protected from impact or accidental splashes of water or liquids.

DANGER! Do not install the device near heat sources or in places with high temperatu- res. Install it in such a way that there is adequate air circulation around it.

DANGER! Before proceeding with installation, check the device condition; if it shows signs of damage caused during transport, do not start it and contact Tecnomed Italia s.r.l. as per paragraph 2.1.

DANGER! Faulty components must only be replaced with original spare parts.

DANGER! Modifying the appliance is strictly forbidden. Tecnomed Italia s.r.l. assumes NO responsibility for personal injury and/or direct or indirect property damage resulting from non-compliance with this requirement.

DANGER! Respect the operating conditions specified in the “Technical Specifications” paragraph and do not exceed the prescribed values.

WARNING! Before connecting the power cable, make sure that the electrical contacts are perfectly dry. If necessary, use compressed air to dry them.

DANGER! The electrical system of the premises in which the purchased product is used must comply with CEI 64-8/7 standards and subsequent amendments. The appliance must be installed by qualified personnel only.

DANGER! To avoid the risk of electric shock, this appliance must only be connected to an electrical system with protective earth.

WARNING! Check that the power supply voltage indicated on the label is the same as the mains voltage.

6 WARNING! In order to guarantee the functionality and safety of the product and avoid damage caused by wear, it is necessary to carry out routine and extraordinary main- tenance at pre-established intervals, by contacting your local Tecnomed Italia s.r.l. dealer. For further information, see the paragraph “Maintenance” on page 28. ENGLISH

DANGER! Tecnomed Italia s.r.l. assumes no responsibility, express or implied, and can- not be held liable for injuries to persons and/or direct or indirect damage to property, resul- ting from the non-observance of what is reported in the entire manual and/or as a result of incorrect procedures related to the installation and/or use of the appliance and its acces- sories and/or resulting from poor and/or lack of cleaning and/or maintenance.

DANGER! Do not disassemble the device. In case of malfunction, please contact an authorised Tecnomed Italia s.r.l. centre.

WARNING! Tecnomed Italia s.r.l. stipulates that in order to ensure device operational safety, installation, maintenance and repairs of this product must only be carried out by technicians authorised by Tecnomed Italia s.r.l.

DANGER! Any technician or worker who is not authorised by Tecnomed Italia s.r.l. and who proceeds to modify/alter the product by replacing parts or components with others not used by the manufacturer, assumes the responsibility similar to that of the manufacturer. TECNOMED ITALIA S.R.L. IS ONLY RESPONSIBLE FOR THE SAFETY AND GUARANTEES THEPRODUCT IF WHAT IS WRITTEN IN THIS MANUAL IS OBSERVED. For any request, always indicate the date of purchase, the device model and serial num- ber. For information on technical support contact:[email protected]

3.1 - WORKING ENVIRONMENT

ENVIRONMENTAL REQUIREMENTS: • Environmental temperature: from +10°C to + 40°C. • Optimum environmental temperature: from 15°C to 30°C. • Relative air humidity: from 30 to 75%. • Ambient pressure from 600 to 1060 hpa

7 4 - SAFETY REQUIREMENTS ENGLISH DANGER! Do not operate the appliance if light shines out of it. IT MAY RESULT IN EYE AND SKIN DAMAGE

DANGER! The product is intended for indoor use only, it is not suitable for wet or humid environments.

DANGER! Do not install the appliance in places where there are anaesthetic or flamma- ble gases, or in places with harmful chemical air.

DANGER! The device must not be used in a potentially explosive atmosphere.

DANGER! You must not cover, block or obstruct, even partially, the air inlet and outlet vents- doing so will compromise the effectiveness of the disinfection.

DANGER! Use the appliance in a place protected from impact or accidental splashes of water or liquids.

DANGER! Do not use the device near heat sources or in places with high temperatures. Use it in such a way that there is adequate air circulation around it.

DANGER! Do not operate the appliance if it shows signs of damage. Contact Tecno- med Italia s.r.l. as per paragraph 2.1.

DANGER! Use the device only for the intended use that is foreseen for it. Failure to comply with this may cause serious injury to the patient, the operator and damage to the device..

DANGER! The electrical system of the premises in which the purchased product is used must comply with CEI 64-8/7 standards and subsequent amendments. The appliance must only be installed by technical personnel authorised by Tecnomed Italia s.r.l.

DANGER! Always check the appliance and the accessories are in proper working or- der. Always check that there is no water underneath the device. Do not use if there are operating faults. In case of anomalies, contact a service centre authorised by Tecnomed Italia s.r.l.

DANGER! Modifying the appliance is strictly forbidden. Tecnomed Italia s.r.l. is NOT responsible for personal injury and/or direct or indirect property damage resulting from non-compliance with this requirement.

WARNING! In order to guarantee the functionality and safety of the product and to avoid damage caused by wear, it is necessary to carry out routine and extraordinary main- tenance (by contacting a centre authorised by Tecnomed Italia s.r.l.), at pre-established intervals. For further information, see the paragraph “Maintenance” on page 28.

8 DANGER! For maintenance and cleaning, disconnect the appliance by unplugging it from the wall socket.

WARNING! Only use products for care and cleaning authorised by Tecnomed Italia s.r.l. Unsuitable products can damage the surface of the device, affect its operation, cause contamination, cause injury to patients and/or the operator. Tecnomed Italia s.r.l. ENGLISH assumes NO responsibility for personal injury and/or direct or indirect property damage re- sulting from non-compliance with this requirement. For further information, see the para- graph “Cleaning” page 27.

DANGER! In the event of a malfunction that is not described in this documentation, put the product out of service and contact an authorised Tecnomed Italia s.r.l. technician.

DANGER! Do not disassemble the device. In case of malfunction, please contact an authorised Tecnomed Italia s.r.l. centre.

DANGER! Tecnomed Italia s.r.l. assumes no responsibility, express or implied, and cannot be held liable for injuries to persons and/or direct or indirect damage to property, resulting from the non-observance of what is reported in the entire manual and/or as a result of incor- rect procedures related to the installation and/or use of the appliance and its accessories and/or resulting from poor and/or lack of cleaning and/or maintenance.

4.1 SAFETY REQUIREMENTS Tecnomed Italia s.r.l. assumes no responsibility, express or implicit, and cannot be held liable for injuries to persons and/or direct or indirect damage to property, resulting from the non- observance of what is reported in the entire manual and/or as a result of incorrect procedures related to the installation and/or use of the appliance and its accessories and/or resulting from poor and/or lack of cleaning and/or maintenance. Tecnomed Italia s.r.l. cannot be held responsible, expressly or implicitly, for any type of injury to persons and/or damage to property, carried out by the user of the product and its accessories, due to the following: • If the appliance is used for purposes other than those for which it is intended; • If the appliance is not used in accordance with all instructions and requirements described in this manual; • Lack of spare parts to use in the event of a breakdown due to faults or problems; • Use of accessories not authorised by Tecnomed Italia s.r.l.; • If the electrical system where the appliance is installed does not comply with the standard of appropriate application and requirements; • If the assembly and/or any repairs have been carried out by personnel not authorised by Tecnomed Italia s.r.l.; • If the environmental conditions in which the device is stored do not meet the requirements indicated in the chapter “Transport and storage conditions”.

9 5- PRODUCT DESCRIPTION ENGLISH 5.1 - INTENDED USE The Steril Fan line includes biological air treatment equipment, ensuring the elimination of all pathogens such as , viruses, fungi and moulds present in the air. The internal UVC lamps and a series of special filters allow the device to guarantee constant room sanitising. Furthermore, it should be noted that, every time air passes through the germicidal chamber, it is purified and up to 99.99% of germs are killed. This appliance is intended for air disinfection only. Any other use is to be considered improper and therefore dangerous. The equipment is intended for use in hospital emergency rooms, operating rooms, infusion rooms, delivery rooms, intensive care units, general wards, health practices, dental practices, pharmacies, grocery stores, companies and waiting rooms.

DANGER! This appliance is not intended for use in areas exposed to the risk of explosion.

DANGER! The product is intended for indoor use only, it is not suitable for wet or humid environments. The user must comply with the following: 1 - Use only EC certified devices in perfect state. 2 - Protect himself, the patients and any third parties from any hazards. 3 - Avoid any contamination of the product. National regulations in force when using the product, in particular: 1 - Current provisions on occupational safety. 2 - Current measures on accident prevention.

DANGER! Use the device only for the intended use that is foreseen for it. Failure to comply with this may cause serious injury to the patient, the operator and damage to the device.

DANGER! Tecnomed Italia s.r.l. assumes no responsibility, express or implied, and can- not be held liable for injuries to persons and/or direct or indirect damage to property, resul- ting from the non-observance of what is reported in the entire manual and/or as a result of incorrect procedures related to the installation and/or use of the appliance and its acces- sories and/or resulting from poor and/or lack of cleaning and/or maintenance.

10 5.2 - STANDARDS/APPROVALS

Medical device The STERIL FAN device bears the CE mark. The STERIL FAN device is a device with electrical protection class I.

Equipment safety: The inductor and light wave controller have signal feedback and anti - ENGLISH interference control.

The STERIL FAN device also complies with the following standards:

Numero Titolo HOUSEHOLD AND SIMILAR ELECTRICAL APPLIANCES - SA- EN60335-2-65:2003 + FETY - PART 2: SPECIAL REQUIREMENTS FOR AIR CLEANING A1:2008 + A11:2012 APPLIANCES

EN60335-1:2012 + HOUSEHOLD APPLIANCES AND SIMILAR APPLIANCES - SAFETY AC:2014 + A11:2014 + PART 1: GENERAL REQUIREMENTS A13:2017

STANDARDS ON MEASUREMENT METHODS FOR ELECTROMA- EN62233:2008 GNETIC FIELDS OF HOUSEHOLD APPLIANCES AND SIMILAR EQUIPMENT WITH REGARD TO HUMAN EXPOSURE.

EN61000-3-2:2014 STANDARDS ON ELECTROMAGNETIC COMPATIBILITY (EMC) EN61000-3-3:2013 - REQUIREMENTS FOR HOUSEHOLD APPLIANCES, ELECTRICAL EN55014-1:2017 TOOLS AND SIMILAR EQUIPMENT EN55014-2:2015

2014/35/EU LOW VOLTAGE DIRECTIVE

2014/30/EU ELECTROMAGNETIC COMPATIBILITY DIRECTIVE

11 5.3 - CARATTERISTICHE TECNICHE Codice SFA2030F SFA2045F ENGLISH Flow rate m3/h 800 m3/h 1200 m3/h Room coverage m2 30 m 45 m Room coverage m3 90 m 135 m Max. Watt power 95 W 120 W Primary dust filter YES (code SFA20T.1) YES (code SFA20T.1) Carbon filter NO NO TiO2 photocatalytic filter SINGLE (cod. SFA20T.2) SINGLE (cod. SFA20T.2) UVC lamp 1X20W (code SFA20T.3) 1X20W (code SFA20T.3) intensity 20102 um/cm2 20102 um/cm2 UVC wavelength 253.7 nm 253.7 nm HEPA filter NO NO Recirculation speed 3 speeds 3 speeds Max noise db S 55 db S 55 db MANCA FOTO TARGHETTA Operation manual-programmable manual-programmable Remote control ALL THREE LEDS DL1 DL2 ALL THREE LEDS DL1 DL2 DL3 DL3 Critical operation alarm YES - 3 alarms YES - 3 alarms Lamp life 12000 h (approx. 4 years) 12000 h (approx. 4 years) External emission µvc at 1 cm S 5 µw/cm2 S 5 µw/cm2 Microbiological elimination% 99.90% 99.90% emission 0,003 mg/m3 0,003 mg/m3 Weight 11 kg 11 kg Dimensions 1000x330x200 mm 1000x330x200 mm (wxdxh) (wxdxh)

Sterilisation effect (disinfection time with air volume at maximum power 30/90 min) - Percent of staphylococcus aureus eliminated S99.90% - Percent of natural bacteria and viruses in the air eliminated S90.0%

The method of physical sterilisation and purification is adopted. During the operation of the device no gases and harmful substances are generated and the device can be used with people present.

Fusibili

Identificazione Valore Dimensione Protezione Posizione F2,5 FUSE1-2 5x20 230VAC Under the right side cover TIME-DELAY

12 Packing dimensions: • Code SFA2030F: PACK 1 TECNOMED ITALIA - Box Dimensions 400X1000X270 mm (width x depth x height). Weight 13 kg (approx).

• Code SFA2045F: PACK 1 TECNOMED ITALIA - Box Dimensions 400X1000X270 mm (width x depth x height). Weight 13 kg (approx). ENGLISH

5.4 - NAMEPLATES The nameplate is located on the left side of the device:

Data on the plate

• Importer’s name • Manufacturer’s name • Model • Reference code • Serial number • Power • Voltage • Consumption • Air circulation flow • Protection class • Regulatory symbols • Brand

13 5.5 - PRODUCT PROFILE

ENGLISH The equipment is composed of a primary filter, a UVC light wave inductor with TiO2 photocatalytic filter, a tangential fan and a radio-controlled electronic control unit. The structure is as follows:

A B A C D

E F

A. Primary filters B. TiO2 photocatalytic filter C. UVC germicidal lamp D. PCB main drive E. Swing motor F. Tangential fan

H

K

I

G J

G. PCB Display H. Infrared sensor I. Cover J. Directional flaps K. Power cable 1.4 mt

14 5.5 - PRODUCT PROFILE 5.6 - STRUCTURE OF AIR TREATMENT

The equipment is composed of a primary filter, a UVC light wave inductor with TiO2 photocatalytic filter, a tangential fan and a radio-controlled electronic control unit. The structure is as follows: ENGLISH A B A C D

1 2 3 Primary filter UVC rays light Photocatalytic E Removes dust Purifies the F and impurities Sterilises the air

A. Primary filters 6- INSTALLATION B. TiO2 photocatalytic filter C. UVC germicidal lamp D. PCB main drive DANGER! Carefully read the paragraph “Installation safety regulations” page 4 before E. Swing motor proceeding. F. Tangential fan DANGER! Before proceeding with the assembly/mounting operations it is necessary to make sure that the wall is able to support the device. The expansion plugs supplied must H ONLY be used with the following wall types: concrete, natural stone, concrete or brick. THEY ARE NOT SUITABLE FOR OTHER TYPES OF MATERIAL.

K DANGER! Tecnomed Italia s.r.l. assumes no responsibility, express or implied, and can- not be held liable for injuries to persons and/or direct or indirect damage to property, resul- ting from the non-observance of what is reported in the entire manual and/or as a result of incorrect procedures related to the installation and/or use of the appliance and its acces- sories and/or resulting from poor and/or lack of cleaning and/or maintenance.

6.1 - ELECTRICAL PREPARATION To operate the appliance, only a 230/240 Volt - 50 Hz power supply is required.

1. Socket power supply connection. I A schuko socket can be installed near the appliance.

G J

G. PCB Display H. Infrared sensor I. Cover J. Directional flaps K. Power cable 1.4 mt

15 2. Fixed mains power supply connection Create a mains supply conduit on the wall with a 3x1.5 mm3 cable at the point

ENGLISH indicated.

Mains power supply 3 x 1,5 mm

6.2 - WALL MOUNTING

1. Position the wall support bracket level to determine the installation position of the machine and mark the 3 holes for drilling.

560 mm

2. Drill the wall with a 10 mm wall bit.

3. Fix the bracket on the wall with the 3 fisher screws supplied

In case of mounting on plasterboard, use special fisher screws and fix the bracket with 5 expansion plugs.

4. Secure the panel to the wall with washers and M8 nuts.

16 5. Hook Steril Fan onto the wall bracket.

6. In case of socket power supply connection, make sure that the socket is earthed and connect the power supply by inserting the schuko plug. In the case of fixed mains power supply connection, cut off the plug and connect the cable using an insulated terminal block. ENGLISH

7. As soon as it is powered, the time (not synchronised) and the ambient temperature will appear on the display and a short acoustic signal will be heard.

WARNING! In order to guarantee appliance operation, a minimum distance of 50 cm from the false ceiling is recommended. The optimal positioning height of the appliance from the floor varies from 2.3 to 2.5 metres.

min 0,5 mt

min 2,3 mt - max 2,5 mt

6.3 - RECOMMENDED POSITIONING IN A DENTAL TREATMENT ROOM

If possible, install STERIL FAN on the wall at the foot of the chair.

17 ENGLISH

STERIL FAN

7- INSTRUCTIONS FOR USE

DANGER! Carefully read the paragraph “Installation safety regulations” page 6 before proceeding to use.

WARNING! If used in manual mode, switch off the machine at the end of the day using the “P” key on the remote control.

WARNING! When the machine is running, doors and windows must be closed.

WARNING! In the event of a malfunction that is not described in this documentation, put the product out of service and contact an authorised Tecnomed Italia s.r.l. techni- cian.

DANGER! Before proceeding with use, carefully read all the previous paragraphs, paying particular attention to the paragraph “Safety requirements” page 8.

DANGER! Tecnomed Italia s.r.l. assumes no responsibility, express or implied, and can- not be held liable for injuries to persons and/or direct or indirect damage to property, re- sulting from the non-observance of what is reported in the entire manual and/or as a re- sult of incorrect procedures related to the installation and/or use of the appliance and its accessories and/or resulting from poor and/or lack of cleaning and/or maintenance.

18 7.1 - REMOTE CONTROL OVERVIEW

A B ENGLISH

F C

A. Working mode (see points G, N, O) E D B. Function enabled C. Manual timer D. Clock E. Operating week F. Air speed R G G. Parameter forward /increase key H H. Parameter back/decrease key I. Manual mode setting Q J. Direction of recirculation K. UVC lamp on L. Timed settings P I M. Unit number key N. Positive or negative ioniser* O. Recirculation speed setting O J P. Timed mode setting Q. Cyclical setting key - 1.5h ON / 1.5h OFF N K R. ON/OFF *Not present on the SFA2030T/SFA2045T models M L

7.2 - INSERTING REMOTE CONTROL BATTERIES

- 2 x 1,5 volt AAA batteries (not included). - Press the symbol and slide the battery compartment cover to the opposite side to open it. - Make sure the battery polarity is correct ( ).

19 7.3 - RESET WEEK ENGLISH - Open the battery compartment. - Press the “CLOCK” button, the number of the week flashes. - Press the ” ” and “ ” buttons on the remote control to se- lect the number..

E G H

7.4 - SETTING THE TIME

D G H

- Open the battery compartment. - Press the “CLOCK” button twice, the “00” in “00:00” will flash to change the time.

- Press the ” ” and “ ” buttons on the remote control to set the hour. - Press the “CLOCK” button to confirm the hour; the “00“ of the minutes will flash to change the minutes. - Press the ” ” and “ ” buttons on the remote control to set the minutes - The entered settings will be stored by pressing the “CLOCK” button again or automati- cally after 10 seconds. The time will also be automatically synchronised on the wall unit display.

Note: the “RESET” button restores the factory settings.

20 7.5 - DISPLAY LCD

1 12 11 2

10 3 ENGLISH

9 4

8 5

7 6

1- In stand-by, the time, temperature and any malfunction codes are displayed. Symbols 2, 5, 6, 7, 8, 9, 10 will light up after switching on. 2- “UV” indicates the operation of the UVC lamp. 3- “CLEAN” indicates that the primary filters need to be cleaned. 4- “ALARM” indicates a malfunction (see paragraph 10). 5- Indicates the operation of the tangential fan. 6- Indicates the speed of the recirculation air IN 3 STEP. 7- “REPLACE” indicates the need to replace the UVC lamp (see paragraph 10). 8- “RUN” indicates the running status. 9- “MANUAL” indicates the operation set in manual. 10- “TIMER” indicates timed operation. 11- “AUTO” indicates cyclic operation. 12- “IONIZER” indicates the operation of the air ioniser (function not present in the SFA2030T and SFA2045T models).

7.6 - MEMORY FUNCTION The controller maintains the operating status after the device is turned off.

21 7.7 - SETTINGS INSTRUCTIONS ENGLISH

Turn on the unit by pressing the button on the remote control - R - (ON/OFF)

- Manual operating mode

Pressing the manual button - I - turns on the logo “ ” on the main display. This indicates that the machine is set to manual operation. I

- Direction of air recirculation

By pressing the swing button - J - it is possible to change the air recirculation direction in a loop. The flaps will move continuously. If you want to stop the flaps in a certain position, press the button again.

J

22 - Turning on sterilising function ENGLISH

WARNING! If the UVC function is disa- bled, the equipment will not perform the sterili- sing action but only air dust filtration.

By pressing the button - K - the internal UVC lamp is activated and therefore sterilisation. The setting is visible on the displays (red circle).

K

- Shutdown countdown setting

Once the device has been activated, a shutdown countdown can be activated by pressing the - L - button. Pressing it once sets the automatic shutdown after one hour. Pressing it 2 times, 2 hours and so on up to a maximum of 5 hours. Each press increases the countdown by 1 hour. The time cycle will be: 2:00 → 3:00 → 4:00 → 5:00 → 1:00.

The example in the figure shows a setting of 2 L hours visible on the display for a few seconds (red circle))

After automatic shutdown, the clock symbol will appear on the remote control display inste- ad of the manual operation symbol.

23 - Multi-control function: It is possible to operate up to 16 Steril Fans with a single remote control. Press button - M -

ENGLISH repeatedly to select the number of the unit to control. By selecting, for example, unit 2, upon power-up, it will follow the settings previously set regardless of what the remote control display indicates. The example in figure 1 indicates the setting of unit 1, in fig. 2 the setting of unit 16.

Figure 1 Figure 2

M

- Ioniser function: Function not present in SFA2030T - SFA2045T models.

N

24 - Adjustment of the air recirculation flow rate:

By pressing the - O - key it is possible to select th- ree speed steps of the tangential fan and conse- quently the air recirculation flow rate. ENGLISH The setting is visible on the displays (red circle).

O

25 - Programming function: ENGLISH

Users can set the timing according to their needs.

Preset setting:

After selecting the - P - function by pressing once the - L - button the display of the remote control shows the preset switch-on time of pro- gram 1 (06:00), by pressing it again, the switch-off time (08:00) - the program number flashes (figure 1). P It is possible to change the switch-on and switch- off times of program 1 using the arrows ” ” and L “ ”

1-SWITCH-ON TIME Figure 1 0-SWITCH-OFF TIME

flashing program

26 The 5 factory preset times are as follows: ENGLISH

This means that if you select the programming button the unit will run all 5 programs in sequence, so it will work intermittently for 2 hours on and 2 hours off and after 24 hours it will switch off completely. It is possible to vary all the switch-on and switch-off times of the 5 programs as you wish.

It is also possible to program a customised 6th daily cycle. In fact, by repeatedly pressing the - L - button you will enter program 6 where it is pos- sible to set the on and off times.

The timing setting shows “--: --”; use the arrows ▲ and ▼ to enter lthe switch-on time and press - L - again for the switch-off time. Each time these buttons are pressed, the time is changed in 10 minute increments; keep the buttons pressed for 3 seconds to change the time more quickly.

1-SWITCH-ON TIME 0-SWITCH-OFF TIME

program 6 program 6 flashing flashing

27 ENGLISH

The example in the figure shows that program 6 has been set with the fol- lowing values: • Switch-on at 07:00 • Switch-off at 18:00

WARNING! Programs overlap, so unwanted programs need to be deleted. To delete a program, it is necessary to set --:-- the switch-on and switch-off times to 24:00 by pressing and holding the up arrow button ▲ .

Example: in figure 3 by deleting programs 1-2-3- 4-5, only program 6 will appear on the display. This means that by selecting the pro- gramming function the unit will only run program 6.

Figure 3

- Cyclic function

By pressing the - P - button automatic function, the unit will work cyclically 1.5 hours on and 1.5 hours off. The setting is visible on the displays Q (red box).

28 8 - CLEANING

DANGER! For maintenance and cleaning, disconnect the appliance by unplugging it from the wall socket. ENGLISH WARNING! The user is responsible for the care and cleaning of the product in order to minimise the risk of contamination. Use authorised cleaning products. Tecnomed Italia s.r.l. assumes no responsibility, expressed or implied, and cannot be held responsible for direct or indirect personal injury and/or damage to property, caused by incorrect cleaning and/or use of inappropriate material.

WARNING! The device must not come into contact with liquids such as water and solvents, aggressive or flammable substances and their vapours. Cleaning agents must not enter the unit as they may cause a short circuit or potential malfunction.

WARNING! Never spray directly on the surface; the disinfectant could penetrate into the openings, ruin the paint, damage the electrical parts. Use a cloth moistened with disin- fectant (it should not be excessively wet or dripping).

8.1 - RECOMMENDED DISINFECTANT

GREEN & CLEAN MK Ready-to-use alcohol-free disinfectant wipes for the disinfection of alcohol-sensitive surfaces of medical products (alcohol-sensitive surfa- ces of the dental unit, handles, leather/imitation leather covers). The long-lasting effect prevents the settling of new germs present in the air. Contains no aldehydes, phenols and phosphates. Tested according to VAH/DGHM criteria. CE 0044. New advanced combination of active ingredients: quaternary ammonium compounds. Spectrum of action: HBV-HIV-HCV: 1 minute Bactericide-fungicide: 1 minute. Codice 60030130: dispenser with 70 wipes + 750 ml SK bottle (MK START).

29 9 - ROUTINE MAINTENANCE ENGLISH

DANGER! For maintenance and cleaning, disconnect the appliance by unplugging it from the wall socket.

WARNING! The user is responsible for product maintenance. Tecnomed Italia s.r.l. as- sumes no responsibility, express or implied, and cannot be held responsible for injuries to persons and/or direct or indirect damage to property, caused by lack of technical main- tenance and safety checks. The product can only be used if all technical safety checks have been passed.

The primary filters need to be cleaned every six months. Remove the filters as shown in the figure, wash them under running water and dry them with compressed air.

30 10- TROUBLESHOOTING

In case of fault, immediately turn off the power and pull out the plug. Consult the following troubleshooting guide before contacting a Tecnomed Italia authorised technician. ENGLISH

PROBLEMA CAUSE SOLUTION 1- Connect the power plug to the mains 1- Plug disconnected 2- Check the switches on the The unit’s display does 2- No power. electrical panel. not turn on 3- The fuse (s) is/are blown. 3- Contact an authorised Tecnomed Italia technician.

Clean the primary filters. Reset the week value as The usage time has reached “CLEAN” indicator per paragraph 7.3. If the the pre-set filter cleaning flashing warning continues, contact threshold. a Tecnomed Italia authorised technician. 1- Increase the speed by pressing the “0” button on the remote control. 1- Fan speed at minimum 2- Clean the primary filters, Low air circulation 2- Dirty filters if the warning continues, contact a Tecnomed Italia authorised technician

“REPLACE” indicator Contact an authorised Damaged UVC lamp flashing Tecnomed Italia technician.

Switch the appliance off and on again to remove TiO2 catalytic filter and UVC CH3 error the alarm for a period of 72 lamp need replacing. hours. Contact an authorised Tecnomed Italia technician.

Contact an authorised ER02 - ER03 errors Operating errors Tecnomed Italia technician.

31 11 - EXTRAORDINARY MAINTENANCE ENGLISH

DANGER! For maintenance and cleaning, disconnect the appliance by unplugging it from the wall socket.

DANGER! It is necessary to carry out technical safety checks and extraordinary main- tenance every year, which make it possible to identify electrical faults (for example da- maged cable insulation). These checks must be carried out exclusivelyby a Tecnomed Italia s.r.l authorised technician

WARNING! The user is responsible for product maintenance. Tecnomed Italia s.r.l. de- nies any liability, expressed or implied, and cannot be held liable for injuries to persons and/ or direct or indirect damages to property, deriving from failure to perform the safety techni- cal controls and maintenance operations. The product can be used only if all the technical safety controls had positive result.

DANGER! The technical checks must also be carried out after any modifications made to the machine, such as installing other components after the first start-up, in order to avoid problems that could affect electrical safety.

32 11.1 - MAINTENANCE TABLE RESERVED FOR TECHNICAL SUPPORT ENGLISH SIGNATURE TECHNICIAN’S CATALYTIC FILTER REPLACEMENT COD. SFA20T.2 CATALYTIC FILTER REPLACEMENT COD. SFA20T.2 CATALYTIC FILTER REPLACEMENT COD. SFA20T.2 CATALYTIC FILTER REPLACEMENT COD. SFA20T.2 CATALYTIC FILTER REPLACEMENT COD. SFA20T.2 UVC LAMP CLEANING UVC LAMP CLEANING UVC LAMP REPLACEMENT COD. SFA20T.3 UVC LAMP CLEANING UVC LAMP CLEANING OPERATIONS TO BE CARRIED OUT PRIMARY FILTERS CLEANING PRIMARY FILTERS CLEANING PRIMARY FILTERS REPLACEMENT (2 X COD. SFA20T.1) PRIMARY FILTERS CLEANING PRIMARY FILTERS CLEANING TECHNICIAN DATE 1st YEAR 2nd YEAR 3rd YEAR 4th YEAR 5th YEAR

33 11.2 - TECHNICAL TROUBLESHOOTING (RESERVED FOR TECHNICAL SUPPORT) ENGLISH

PROBLEMA CAUSE SOLUTION TIME RESET Press the manual ALARM” Operating time without Carry out button ERROR CODE maintenance has maintenance as - L - for 5 seconds ER02 exceeded 9999 hours from the 3rd year on the remote control

Press the manual “ALARM” Operating time without Carry out button ERROR CODE maintenance has maintenance as - L - for 5 seconds CH03 exceeded 9999 hours from the 3rd year on the remote control

Replace the UVC lamp and/ or inductor. Clean Press the manual The UVC lamp is damaged “ALARM” the catalytic filter button suddenly/false contact on ERROR CODE and primary filters - L - for 5 seconds IN3 connector of the pcb/ IN03 with compressed on the remote damaged inductor air/ check the IN3 control connector/ replace the lamp inductor

MAIN DISPLAY No power Replace fuses F1-F2 OFF

Press the manual button “I” for 5 seconds "REPLACE" UVC lamp needs replacing Replace lamp tube on the remote control

See routine Cleaning time limit "CLEAN" maintenance threshold reached paragraph

34 11.3 - REPLACING THE PHOTOCATALYTIC FILTER

- Open the upper hatch and remove the primary filters ENGLISH

- Unscrew the 2 Phillips screws from the right filter bracket

- Remove the filter and replace it with a new one (Cod. SFA20T.2)

WARNING! Whenever the filter is replaced, follow the extraordinary maintenance ta- ble regarding the work to be carried out on the UVC lamp (paragraph 10.1).

35 11.4 - CLEANING OR REPLACING THE UVC LAMP ENGLISH

- Open the top hatch and remove the primary filters

- Unscrew the 2 Phillips screws from the right filter bracket

- Pull out the filter to access the UVC lamp

- Remove the lamp from the electric socket «A» by cutting the 2 cable ties and turning it 90°.

A B A - Clean with a cloth soaked in alcohol or replace the UVC lamp (Code SFA20T.2) - Clean the reflecting mirror «B» of the photocatalytic cell with a cloth soaked in alcohol.

WARNING! If there are vibrations it is always recommended to secure the lamp with 2 cable ties.

36 11.5 - REPLACING THE FUSES

- Unscrew the 2 Phillips screws of the cover

ENGLISH

- Unscrew the fuse holders and replace the fuses

37 12 - RELATED ITEMS

ENGLISH To find all items related to this product, visit:

www.dentalastec.it

13 - DISPOSAL

This equipment complies with the European Directive 2002/96/CE on waste electrical and electronic equipment (WEEE). By ensuring this product is disposed of correctly, you will help prevent potential negative consequences to the environment and human health, which could otherwise arise from improper handling of the product at the end

of its service life. The symbol indicates that the product should not be disposed of like normal waste, but it should be sent to the nearest authorised waste collection centre for electrical and electronic equipment. Disposal must be carried out in accordance with the applicable environmental regulations for waste disposal. The equipment must be properly prepared (disinfection/sterilisation) before dismantling/disposal. For more information on the treatment, recovery and recycling of this product, please contact the Ecology and Environment Department or your local waste collection service or your dental equipment distributor. Illegal dumping of the product by the user entails the administrative sanctions stated by current legislation.

38 14 - WARRANTY

With this document, the manufacturer certifies that the product has been built correctly, in compliance with the applicable national and European regulations in force. The product is covered by a 12-month warranty from the date of purchase specified on the stamp and on the valid fiscal document issued by the seller. According to art. 1495 c.c., the person who intends to make a warranty claim, must give notice of the failure within 8 days from its ENGLISH discovery. The warranty claim may be refused if the invoice has been changed, deleted or made illegible after the purchase. Disconnect the pneumatic, hydraulic and electrical power supply at the end of the day (if present). Tecnomed Italia srl shall not cover dam- ages caused by failure to comply with the indications above. Not covered by warranty: • labour expenses, staff travel expenses, transport expenses: • operations for the installation and connection of the machine to the various power supply sources; • all parts subject to normal wear and tear (such as light bulbs, carbon brushes, cordon o-rings, suction tubing, etc...); • all those parts that are faulty due to negligence or careless use and/or maintenance; • in-transit damages; • damages caused by limescale or dirt accumulation on hydraulic and pneumatic supply circuits (if present); • malfunctions due to failure to respect the routine cleaning intervals of the filters (if any); • malfunctions caused by third-party equipment (, suction systems, water sof- teners), or other circumstances unrelated to defects in the manufacturing of the device. The Warranty becomes void if: - the equipment presents damages due to fall, exposure to flames, liquid spills, lightning, natural disasters, other causes not attributable to manufacturing defects; - the installation and/or extraordinary maintenance operations and/or the technical con- trols have not been carried out by an authorised Tecnomed Italia s.r.l. technician; - the power supply connection is incorrectly made or the appropriate safety devices have not been installed; - the serial number/series or CE marking are removed, erased or tampered. The warranty covers the replacement, free of charge, of the parts of the device resulting to be defective at the origin due to manufacturing defects. The Customer cannot request the replacement of the entire product. In case of permanent damage or repeated failure (deriving from the same cause), the manufacturer reserves the right to decide whether the equipment will be replaced or not. The coverage for the new product provided by the warranty will remain valid until the end of the original contract. For warranty service the purchaser should contact the reseller, the authorised service centers or Tecnomed Italy s.r.l. This warranty does not give the Customer the right to claim for direct or indirect damages (of any nature) to persons or property resulting from machine malfunction. Tecnomed Italia srl denies any liability, expressed or implied, and cannot be held liable for injuries to persons and/or animals and/or damages to property deriving from failure to comply with the instructions in this manual (warnings, installation safety standards, safety provisions, instructions for installation and use, information for correct cleaning and sterili- sation, maintenance information, general information, technical specifications etc...). The manufacturer reserves the right to make changes or improvements without prior notice if these changes do not affect the safe use of the device. In the event of any dispute concerning the application of the warranty or the condition of the equipment delivered, the buyer may not suspend or delay the payment of the price in full or in installments. No compensation may be claimed by the buyer for equipment downtimes. If the part to be changed is not returned, it will be charged to the buyer. Also, please see page “PAYMENT TERMS AND CONDITIONS” available on our site www.dentalastec.it Any and all disputes that may arise will be settled by the competent Court of Pesaro.

39 DECLARATION OF CONFORMITY

ENGLISH CERTIFIED COPY OF THE ORIGINAL Subject: EC DECLARATION OF CONFORMITY Distributor name: TECNOMED ITALIA SRL Electrical class: Ⅰ

Type: SFA2030F // SFA2045F Model: VBY-B-800 // VBY-B-1200 Product description: WALL-MOUNTED UVGI AIR STERILIZER

I, the undersigned Luca Riccardi, as administrator of TECNOMED ITALIA s.r.l., based on the certificates issued by the manufacturer and now in my possession, hereby declare that the product described is conformity of the directives:

EN60335-2-65:2003 + A1:2008 + A11:2012 EN60335-1:2012 + AC:2014 + A11:2014 + A13:2017 EN62233:2008 EN61000-3-2:2014 EN61000-3-3:2013 EN55014-1:2017 EN55014-2:2015

and requirements:

2014/35/EU 2014/30/EU

This compliance is expressed through the marking:

Monte Porzio, 04/05/2020

Tecnomed Italia srl

Legal Representative Luca Riccardi

40 ENGLISH

Garanzia-Guarantee-Garantie-Garantía ______24____ Mesi-Months-Mois-Meses _ Ragione sociale studio- Business name- Raison sociale- Razón social _ Indirizzo- Address- Adresse- Direccion

Etichetta prodotto - Product label - Etiquette produit - Etiqueta producto

Timbro del rivenditore-Dealer’s stamp Cachet d’achat-Sello del revendedor

Data d’acquisto-Purchase date Date d’achat-Fecha de compra

Tecnomed Italia s.r.l. Via Salvador Allende n.2, 61040 Castelvecchio di Monte Porzio (PU) Italy Phone +39 0721 95 51 25 Fax +39 0721 95 52 29 - www.dentalastec.it

41 ENGLISH