<<

Surgical Technique

SL-PLUS™ MIA SL-PLUS™ SLR-PLUS™

Contents

Indications / contraindications...... 2 Warnings and precautions...... 4 Preoperative planning...... 5 Surgical technique...... 6 Postoperative treatment...... 16 Explantation of the stem...... 17 Implants...... 18 Demo Implant...... 21 Dimensions...... 22 Instruments...... 28 Sterilization...... 38

N.B. The surgical technique described in this brochure is the procedure suggested by the authors for uncomplicated surgery. The surgeon must, however, decide which procedure is the most suitable and effective for each individual patient. 1 Indications/contraindications

Indications SL-PLUS™ MIA SL-PLUS™ SLR-PLUS™

All femur types can be treated with the exception of those with extreme curvature, e.g. after angulation osteotomies. In these cases, a corrective osteotomy beforehand may be appropriate.

• Advanced degeneration of the hip joint as a result of degenerative, post-traumatic or rheumatoid arthritis • Fracture or avascular necrosis of the femoral head • Conditions resulting from previous surgery such as osteosynthesis, joint reconstruction, arthrodesis, hemiarthroplasty or hip total prosthesis

The SLR-PLUS revision stem is designed for prosthetic arthroplasty in revision surgery. Indications SLR-PLUS™

• Dysplasia • Revisions of cemented and non-cemented implants due to aseptic or non-aseptic loosenings, osteolysis, dislocations or impingements • Insufficient soft tissue balancing

Indications for Lateral Stem SL-PLUS™

• Varus femur forms • Trumpet shape of the proximal femur (champagne flute) • Insufficient soft tissue balancing • Status after previous operations (e.g. displacement osteotomy, pelvic osteotomy)

Lateral Stem SL-PLUS™ MIA SL-PLUS™ SLR-PLUS™

The Lateral stem design has a much greater offset (see Dimensions on p. 22, 24, 26) and, at 123°, a much smaller CCD angle than the standard stem (131°). In the case of coxa vara, the CCD angle is smaller than normal (126°). Implanting a stem prosthesis with a CCD angle of 131° or more can result in a medialization and/or lengthening of the leg. By implanting a lateralizing stem, the lever arms can be reconstructed in such a way that the strength of the pelvic-trochanter muscles is restored. At the same time, an optimal soft tissue tension is achieved. This minimizes the risk of an inadequate gait and the risk of luxation, if required by the preoperative planning.

2 Contraindications SL-PLUS™ MIA SL-PLUS™ SLR-PLUS™

• Acute or chronic infections, local or systemic • Local infections of the area operated • Severe muscle, nerve or vascular diseases that endanger the extremity in question • Lacking bone substance or inadequate bone quality that endangers a stable seating of the prosthesis • All concomitant disease that may endanger the function of the implant, such as: - Any allergies to the implant materials - Renal insufficiency - Cardiac insufficiency (e.g. as a result of increased metal ion levels in the blood) • Revision with extensive bone defects extending for distal to the lesser trochanter and thereby reducing the stable anchoring of the stem to less than half

Contraindications SLR-PLUS™

• Bone defects following tumor resections • Periprothetic fractures

Important It will be seen that SLR-PLUS Revision stem may be used in a wide variety of cases. The limitation are complex revisions in which even such adjunctive measures as cerclages or wires fail to stabilize the device in the proximal part of the femur. In such cases, a long revision stem; e.g. REDAPT™ should be used, which obtains fixation in the distal half of the host bone.

3 Warnings and precautions

The following may not be implanted under any circumstances • Implant components that have been damaged or scratched • Implants that have been handled inappropriately or processed in a way that is not part of the surgical technique • Implants that have already been used • Implants where the packaging and/or labelling is damaged or not intact

Please return the affected devices to the appropriate Smith&Nephew representative or sales office.

Reliable fit of femoral ball heads on stem tapers The taper connection can only be reliably and firmly seated if the surface of the ball head cone and the surface and structure of the hip stem taper are completely intact. The disposable plastic cap protecting the stem taper from damage shall not be removed until the trial ball is attached. To ensure that the ball head performs as required, it is essential to take great care when attaching it to the stem taper: Clean and dry the neck taper with a clean, sterile cloth. Place the prosthetic femoral head on the neck taper and firmly impact with the femoral head impactor and a several times. Never reuse a femoral ball head that has been impacted onto a stem cone and then removed.

Restrictions on head/insert combinations • BIOLOX®forte/delta ceramic inserts must only be combined with BIOLOX®forte/delta ceramic ball heads.

Restrictions on head/taper combinations • Material: Stainless steel (FeCrNiMoNbN) heads and stainless steel stems should only be used together. Neither should be used with other metal components. • Sizes: all ball head sizes of Smith&Nephew Orthopaedics AG can safely be used with all hip stems of Smith&Nephew Orthopaedics AG. For combinations with components of other manufactures (e.g. Smith & Nephew Inc.), the surgeon must verify if the combination is allowed as listed in the compatibility matrix (lit. no. 04758) available on www.smith-nephew.com/compatibilitymatrix. Ball heads should only be combined with hip stems of identical taper dimensions.

Femoral ball head revisions For the revision of a femoral ball head, apply the following: • The corresponding insert has to be revised as well. • Replacement only by a metal ball head or a special ceramic revision femoral ball head featuring a metallic taper adaptor (BIOLOX® OPTION). Do not assemble a standard ceramic head on a used taper; the ceramic head may fracture from irregularities on the femoral component taper. • In the case of a revision involving a fractured ceramic component, femoral ball head and insert have to be revised. Remove all loose identifiable fragments and thoroughly irrigate and lavage the operative site. A special ceramic revision femoral ball head coupled with a metallic taper adaptor should be used if the femoral taper is intact. Metal or Oxinium femoral ball heads with a polyethylene insert shall not be used for revising fractured ceramic components. In case the taper is damaged or no appropriate ceramic revision femoral ball head is available, the femoral stem must be revised to provide a suitable femoral taper to attach a new ceramic ball head.

4 Preoperative planning SL-PLUS™ MIA SL-PLUS™ SLR-PLUS™

Preoperative planning is essential to determine the correct orientation and size of implant. AP and axial x-rays are required for this purpose. The offset and neck length to be achieved with the prosthesis are determined with the aid of x-ray templates.

X-ray Templates - scale 1.15:1 • SL-PLUS MIA stem, Lit. No. 2202 x x • SL-PLUS stem, Lit. No. 2201 • SLR-PLUS stem, Lit. No. 2203

Today, digital templating is the new standard in many institutions. Smith & Nephew is supporting a number of third party digital pre-operative planning . Please contact your local Smith & Nephew representative for further information.

To determine the entry point in the medullary canal, the surgeon is advised to draw the femoral axis on the AP radiograph and extend this proximally. This line indicates how far laterally the box needs to be placed to open up the medullary canal, thus facilitating identification of the lateral entry point at operation.

It is also helpful to define the position of the stem within the canal. The distance x from the shoulder of the stem to the greater and lesser trochanter is measured and can serve as an additional leg length check.

Note X-ray templating is not as reliable for determining implant sizing in revision cases as they are in planning for primary implantations and can therefore only provide an approximate estimate of the actual stem size required.

5 Surgical technique

SL-PLUS™ MIA SL-PLUS™ Position of the patient and approach For the purpose of this surgical technique, the anterolateral, minimally invasive approach is chosen with the patient in the supine position.

Whether an anterolateral, lateral, posterolateral or posterior approach is used is at the surgeon’s discretion. The skin incision and muscle detachment depend on the selected approach. All implants and instruments also suit a minimally invasive approach.

Capsular incision and dissection Perform a blunt dissection between the tensor fasciae latae and the gluteus medius/minimus from the lateral side to expose the femoral neck.

One sharp and one blunt Hohmann retractor are placed laterally and medially respectively. The curved rectus tendon is exposed, undermined, divided and released from its attachment to the capsule.

The third retractor is placed between the ventral joint capsule and the overlying muscles (curved rectus tendon and the ileopsoas muscle).

The femoral neck is then exposed via an H-shaped incision of the joint capsule: • longitudinal incision, as far medially as possible, from the acetabular rim to the intertrochanteric line • proximal transverse incision of the acetabular labrum from approx. the nine o’clock to the three o’clock position • distal transverse incision extending along the intertrochanteric line

6 After the wing-like opening of the joint capsule, the dissection of the capsule can be continued by extending the distal incision along the intertrochanteric line in the direction of the lesser trochanter and the proximal incision medially and/or laterally.

Two blunt Hohmann retractors are positioned inside the joint. Problematic osteophytes on the acetabular rim are removed.

The technique selected by the surgeon for the neck resection (single or double osteotomy) depends on the patient (coxa vara/ valga).

Capsular release In order to facilitate alignment of the stem, the surgeon switches the leg to the “Figure 4” position and performs an additional release of the posterior capsule.

Now a retractor is placed between the capsule and the gluteus muscle holding the leg in a 10-20° abducted position. This facilitates the placement of the retractor and gives access to the lateral and cranial part of the capsule. In that way this part of the capsule should be released or resected.

The leg is hyperextended, adducted approximately 30–40° and externally rotated by 90°. The leg being operated on is placed under the other leg in a “figure of 4” position. Alternatively, the other leg is lowered and the leg being operated on is placed on top.

The proximal femur is mobilized with two retractors, a lateral trochanter retractor and a second retractor on the medial side of the femoral neck.

Capsular releases must be performed in the direction of both the lesser trochanter and the trochanteric fossa to the tip of the trochanter. Optionally, it can also be performed on the caudal rim of the acetabulum.

7 The cup is then implanted according to the corresponding surgical technique.

Stem Preparation In very muscular or obese patients, patients with a valgus femoral neck, or in cases where the proximal femur sits deep to the skin surface, further release of the posterior capsule or release of the piriformis tendon may be necessary to allow Inferior gemellus m. adequate mobilization of the femur prior to preparation of the Internal obturator m. implantation site. Superior gemellus m. Piriformis m.

Access to the medullary canal SL-PLUS™ MIA The box chisel (75006419/41000029) is placed close to the posterior cortex at the resection level. The box chisel is introduced in alignment with the femoral axis and a small block of bone is removed. If, after the neck resection, the box chisel is not used to expose the canal, splitting of the trochanter may occur during rasp insertion.

Do not impact the box chisel too deeply.

SL-PLUS™ SLR-PLUS™ The standard box chisel (75002322/110904) or the MIS Offset box chisel (75007254/600620) are available.

The MIA curved rasp facilitates opening of the diaphyseal medullary cavity. Removal of the intramedullary fat by suction to avoid fat embolism.

Further opening of the diaphyseal medullary cavity and probing of the diaphysis with a corresponding awl are recommended.

8 SL-PLUS™ MIA Use of the MIA guide rasp The aim of the MIA guide rasps is to ensure easier alignment of Medial the rasping bed, thereby preventing a varus implant seating. The MIA guide rasp should be introduced with the desired degree of anteversion, matching the planned rotation of the stem.

The rasp depth can be checked by means of the line markings 1 2 3 4 5 6 on the stem. These markings correspond to the shoulder level

Markings on implant shoulder of the respective SL-PLUS MIA implant size. During the rasping operation, care should be taken to restrict the depth of insertion of the MIA guide rasp to around one or two sizes smaller than the planned implant size. As soon as cortical contact is achieved the guide rasp should not be driven any further. An optional MIA guide rasp, size 01-3, is available for small stem sizes of 01 or 0.

When mounting the rasp on the slap hammer or rasping machine, ensure that the side marked “MEDIAL” is correctly oriented. If the medial and lateral sides are inadvertently reversed, the rasp handle may strike the medial aspect of the greater trochanter, forcing the rasp out of its planned neutral alignment.

Broaching of the femur The rasp cuts longitudinal grooves in the femoral cortex. The goal is to achieve extensive surface area contact to provide cortical support for the implant. By gradually increasing the depth of rasping within the medullary cavity, the area of cortical contact increases, along with the resistance to the advance of the rasp. As soon as the rasp is fully engaged within cortical bone, the pitch of the hammer blows rises. The critical resource for checking the correct size is the preoperative planning with the x-ray.

SL-PLUS™ MIA SL-PLUS™ Preparation of the bone bed up to implants of size 4 starts with a size 01 rasp. For implants of size 5 and above, a size 1 rasp can be used initially.

SLR-PLUS™ Start with the smallest rasp size 1.

Important Since the first rasp determines the position of all the following rasps, its orientation is critical for exact positioning of the stem.

9 SLR-PLUS™ If the intraoperatively determined bone conditions indicate that the shaft could fracture, the use of a wire or cerclage band is recommended to stabilize the bone before starting the preoperative process with the rasp. This helps to considerably reduce the risk of fracture during bone preparation and when impacting the implant and should therefore, ideally, always be used.

For correct alignment, the rasp is inserted along the anatomical axis of the stem. If the operator deviates from this axis there is a risk of subsequent varus positioning of the stem in the femur.

The mountable MIA trial rasps are used to create an implantation site of the correct size and alignment for the femoral implant, as described below.

Adapters with differing offsets are available to accommodate the selected surgical approach and/or patient size. Please refer to the instrument sets on page 30, 34, 36.

All modular adapters can be connected to the slap hammer or the MIA Rasping Machine (IMT Woodpecker) or the modular knock plate (75000642/21000378).

At the start of the rasping process, the depth of the rasp must be checked to ensure that it is not inserted too deeply. It is Bone loss in calcar arch extremely important to understand that the osteotomy is unrelated to the final position of the implant. Because of the initially reduced rasp resistance, there is a tendency for surgeons to insert the smaller rasps too deeply into the femur. This will result in excessive enlargement of the implantation site and lead to gaps on the medial aspect between the bone and the final implant.

10 The subsequent rasp is introduced into the cavity until resistance is felt. The rasp is then driven laterally and distally into the femur using the (slap) hammer or the rasping machine. This process is repeated with sequential rasp sizes until an adequate depth is achieved for the desired implant size. If the double offset adapters are used, care should be taken to ensure that these instruments are used on the corresponding side of the patient.

SL-PLUS™ MIA The rasp is introduced into the cavity along a slightly arc-shaped path.

Attention should be paid to the anteversion and varus/valgus alignment of the slap hammer or the rasping machine with respect to the femoral axis. Insertion of the rasps in a varus inclination involves the risk of perforation and/or fracture of the lateral cortex of the femur.

If rasping is carried out using the slap hammer or the MIA Rasping Machine, the weight of these instruments helps ensure the precise longitudinal alignment of the rasp within the femur. It is important that lateral pressure is continuously applied to the rasping machine to ensure that the rasp moves in line with the axis of the canal and does not seat in a varus position.

Unlike the SL-PLUS™ and SLR-PLUS™ stem, the SL-PLUS MIA™ rasp does not enter and exit the canal along the femoral axis, but rather along a curved arc.

11 The shoulder of the rasp corresponds to the shoulder height of the implant and should correspond to the preoperatively determined distance to the greater trochanter (marked x). If the greater trochanter is not present anymore, the lesser trochanter x can be taken instead.

Only in rare cases does the planned prosthesis size not correspond to the size rasped at operation. If there is a discrepancy of two or more sizes, the rasp may not have reached the necessary depth because of incorrect angulation or the presence of an obstacle in the canal. In these cases the prosthesis would be too small and stable cortical anchorage would not be ensured in the long term. If doubt exists, intraoperative radiography or fluoroscopy is necessary so that the obstruction can be assessed.

If the preplanned implant size goes too deep into the femoral canal, a fracture (fissure) should be excluded by intraoperative radiography or fluoroscopy.

SLR-PLUS™ In most cases, however, the existing proximal osteolysis and the laxity of scar tissue has to be taken into account. To avoid postoperative luxation, the prosthesis should be positioned so as to ensure sufficient soft tissue tension. This can also cause the shoulder of the stem to settle proximal to the original trochanter apex.

The situation may arise that the planned stem size does not correspond to the intraoperatively rasped size. If there is a difference of two or more sizes, the rasp may not have reached the necessary depth due to either being inserted at the wrong angle to the longitudinal axis of the femur or due to an obstacle, such as residues of bone cement or protrusions of bone. In these cases the stem would be too small and cortical anchoring would not be ensured. Residual foreign bodies or bony protrusions could fracture the bone diaphysis. If in doubt, a special x-ray examination should be carried out (image intensifier)

Once the optimum size and stability of the rasp and its height position are achieved, the offset adapter is removed from the mounted trial rasp.

12 Trial Reposition The neck module is attached to the trial rasp by hand.

SL-PLUS™ MIA SL-PLUS™ Standard neck modules are available for trial rasp sizes 01–0, 1–6 and 7–12. Matching “lateral” neck modules are available for trial rasp sizes 1–6 and 7–12.

SLR-PLUS™ Standard and Lateral trial rasp sizes 1-6 and 7-11 with matching neck modules are available.

It is important to ensure that the neck module fits flush with the trial rasp and that it clicks into position.

The use of a Dual Mobility cup or retention in the cup, however, can significantly reduce postoperative luxation, even if an SLR-PLUS™ lateral stem is used. Because of the retaining effect of such an inlay, resulting in increased luxation resistance, it may in some cases also be possible to use a shorter neck length. This can prevent unintentional leg lengthening.

Trial Reposition The selected trial ball head can either be fitted to the neck module in advance or in situ. Ideally fitting of a M/+4 or L/+8 trial ball head is recommended to leave room for up-/down- sizing on the final stem.

The joint is reduced and the leg length, soft tissue tension and range of motion are checked. During the initial operations, it is recommended that the surgeon obtain AP and lateral intraoperative radiographs (image intensifier) to verify the position and size of the rasp in both planes.

If necessary the trial ball head and/or the neck module (standard or lateral) should be replaced until a satisfactory result is achieved.

If luxation is still possible even though the lateralized modular neck is used and the rasp is securely seated, check if the next sized stem should be used. This then generally protrudes further proximally out of the femoral bone compared to the smaller size and can therefore lead to leg lengthening. The patient should be made aware of this risk before the surgery.

13 The neck module can either be removed manually from the trial rasp or with a Kocher retractor.

The rasp adapter is connected to the trial rasp. The connected rasping machine or the slap hammer is used to remove the trial rasp.

SL-PLUS™ MIA

Removal of the rasp, as with its introduction, must be performed along a curved arc to minimize disturbance of the bone bed and to avoid fractures in the vicinity of the trochanter.

Implantation of the stem SL-PLUS™ MIA The stem of the correct size is introduced manually as deep as possible into the canal. The SL-PLUS MIA stem is then seated with the MIA stem impactor (75007255/600621), using appropriately measured strokes to minimize the risk of femoral fractures.

SL-PLUS™ SLR-PLUS™ Seated with the Impactor (75002321/110903)

If using a Ti/HA coated stem, the Ti/HA coating leads to a total oversize of 0.7 mm and, particularly if hard bone is present, can prevent the implant from being inserted as deeply as the rasp. In this case, it is advisable to recheck the leg length, soft tissue tension and range of motion with the trial ball head. This oversize is not usually relevant for clinical purposes and can be offset with a shorter ball head. Otherwise, the implant bed must be enlarged accordingly with the last rasp.

The protective cap remains mounted on the cone during impaction.

Please note It is not sufficient to press the stem in solely by hand.

It is not permitted to impact the stem deeper than the prepared bony bed or to correct the position once the stem is in the cone bed as this would inevitably split the femoral stem. It is recommended to recheck the leg length, soft tissue tension and range of motion with the trial ball head.

14 Ball head introduction Before positioning the ball head, carefully wash the stem cone with water and then dry.

The ball head is then fitted by screwing it lightly into place.

The ball head is firmly impacted using the head impactor (75002160/110242). The joint is then reduced, manipulated and retested to ensure proper functioning.

Note Please carefully read the chapter “Reliable fit of femoral ball heads on stem tapers” on page 4. Ceramic heads must never be impacted using a metal instrument.

Each femoral stem has a standard 12/14 cone for coupling with OXINIUM™, ceramic or CoCr ball heads supplied by Smith & Nephew Orthopaedics AG or Smith & Nephew Inc.

Wound closure Insert Redon drains and close the wound. Position the leg in slight abduction.

15 Postoperative treatment

SL-PLUS™ MIA SL-PLUS™ Postoperative rehabilitation should be provided in accordance with the standard procedures of the respective hospital. For the SL-PLUS MIA™ as well as the SL-PLUS™ stem, immediate weight bearing is possible. Definitive osseointegration is not achieved until 3 months postoperatively.

SLR-PLUS™ It should, however, be noted that revision surgery requires individual postoperative treatment, depending on the degree of preoperative bone destruction and the intraopera­tively achieved stability of the implant. In some cases the operated hip can be mobilized immediately and be partially load bearing, although mobilization without full load bearing is indicated for some time. During physiotherapy there is also the risk of postoperative luxation of the head out of the cup to be considered, as the stability can be considerably affected by soft tissue defects due to previous surgery.

16 Explantation of the stem

SL-PLUS™ MIA The SL-PLUS MIA stem can be extracted using the M6 extraction screw (75002165/110249).

SL-PLUS™ SLR-PLUS™ Stems can be extracted using the M8 extraction screw (75002325/110911).

In case of difficulty, an extraction block secured with the M8 extraction screw is also available.

Both extraction screws, M6 and M8, can be used in connection with either the slap hammer or the rasping machine.

Please note Ensure that the extraction screw is inserted axially.

The removal of a well osseointegrated stem should be done by the following four steps: 1. Break the bone bridges in the proximal (Ti-/HA-coated) part of the stem by using a thin flat chisel. 2. Make a short cut in the cortical bone with a thin at the distal end of the stem. Both ends of this cut should be drilled with a Ø4.0mm to avoid fissure following this cut. This cut should be placed in the ventral - lateral part of the femur corticalis. 3. Spread the bone edges of this cut carefully from the implant using a chisel with gentle wedge action. Doing so, you will break the bone bridges between the distal end of the implant and the surrounding bone. 4. Try to remove the implant stem with the appropriate extrac- tion tools as described above.

17 SL-PLUS™ MIA Implants

SL-PLUS MIA INTEGRATION-PLUS™ Cementless Stem Material: Ti-6Al-7Nb ISO 5832-3, Cone 12/14 With Ti-plasma/hydroxyapatite coating proximally:

Standard Stem CCD 131° Lateral Stem CCD 123°

SAP No. Art. No. Size SAP No. Art. No. Size 75000172 11000422 01 - - - 75000173 11000423 0 - - - 75000174 11000424 1 75000186 11000436 1 75000175 11000425 2 75000187 11000437 2 75000176 11000426 3 75000188 11000438 3 75000177 11000427 4 75000189 11000439 4 75000178 11000428 5 75000190 11000440 5 75000179 11000429 6 75000191 11000441 6 75000180 11000430 7 75000192 11000442 7 75000181 11000431 8 75000193 11000443 8 75000182 11000432 9 75000194 11000444 9 75000183 11000433 10 75000195 11000445 10 75000184* 11000434 11 75000196* 11000446 11 75000185* 11000435 12 75000197* 11000447 12 * Special size (optional) * Special size (optional)

18 Implants SL-PLUS™ Implants

SL-PLUS Cementless Stem Material: Ti-6Al-7Nb ISO 5832-3, Cone 12/14

Standard Stem CCD 131° Lateral Stem CCD 123°

SAP No. Art. No. Size SAP No. Art. No. Size 75002717 11420S 01 - - - 75002719 11421S 0 - - - 75002695 11401S 1 75002748 11531S 1 75002697 11402S 2 75002750 11532S 2 75002699 11403S 3 75002752 11533S 3 75002701 11404S 4 75002756 11534S 4 75002703 11405S 5 75002758 11535S 5 75002705 11406S 6 75002760 11536S 6 75002707 11407S 7 75002762 11537S 7 75002709 11408S 8 75002764 11538S 8 75002711 11409S 9 75002766 11539S 9 75002713 11410S 10 75002768 11540S 10 75002714* 11411S 11 75002769* 11541S 11 75002715* 11412S 12 75002770* 11542S 12 * Special size (optional) * Special size (optional)

SL-PLUS INTEGRATION PLUS Cementless Stem Material: Ti-6Al-7Nb ISO 5832-3, Cone 12/14 With Ti-plasma/hydroxyapatite coating proximally

Standard Stem CCD 131° Lateral Stem CCD 123°

SAP No. Art. No. Size SAP No. Art. No. Size 75002720 11459 01 - - - 75002722 11460 0 - - - 75002723 11461 1 75002773 11569 1 75002724 11462 2 75002777 11570 2 75002725 11463 3 75002778 11571 3 75002726 11464 4 75002779 11572 4 75002727 11465 5 75002780 11573 5 75002728 11466 6 75002781 11574 6 75002729 11467 7 75002782 11575 7 75002730 11468 8 75002783 11576 8 75002731 11469 9 75002784 11577 9 75002732 11470 10 75002785 11578 10 75002733* 11471 11 75002786* 11579 11 75002734* 11472 12 75002787* 11580 12 * Special size (optional) * Special size (optional)

19 SLR-PLUS™ Implants

SLR-PLUS Cementless Stem Material: Ti-6Al-7Nb ISO 5832-3, Cone 12/14

Standard Stem CCD 131° Lateral Stem CCD 123°

SAP No. Art. No. Size SAP No. Art. No. Size 75002795 11801S 1 75001926 11000151 1 75002797 11802S 2 75001927 11000152 2 75002799 11803S 3 75001928 11000153 3 75002801 11804S 4 75001929 11000154 4 75002803 11805S 5 75001930 11000155 5 75002805 11806S 6 75001931 11000156 6 75002807 11807S 7 75001932 11000157 7 75002809 11808S 8 75001933 11000158 8 75002811 11809S 9 75001934 11000159 9 75002813 11810S 10 75001935 11000160 10 75002815 11811S 11 75001936 11000161 11

SLR-PLUS INTEGRATION-PLUS Cementless Stem Material: Ti-6Al-7Nb ISO 5832-3, Cone 12/14 With Ti-plasma/hydroxyapatite coating proximally:

Standard Stem CCD 131° Lateral Stem CCD 123°

SAP No. Art. No. Size SAP No. Art. No. Size 75100370 11861 1 75000198 11000448 1 75100371 11862 2 75000199 11000449 2 75002816 11863 3 75000200 11000450 3 75002817 11864 4 75000201 11000451 4 75002818 11865 5 75000202 11000452 5 75002819 11866 6 75000203 11000453 6 75002820 11867 7 75000204 11000454 7 75002821 11868 8 75000205 11000455 8 75002822 11869 9 75000206 11000456 9 75002824 11870 10 75000207 11000457 10 75002825 11871 11 75000208 11000458 11

20 Demo Implants

SL-PLUS™ MIA SAP No. Art. No. Size Description 75023398 91000428 5 SL-PLUS MIA INTEGRATION-PLUS Standard 75023399 91000440 5 SL-PLUS MIA INTEGRATION-PLUS Lateral

SL-PLUS™ SAP No. Art. No. Size Description 75007771 D11405S 5 SL-PLUS Standard 75007774 D11535S 5 SL-PLUS Lateral 75007773 D11467 7 SL-PLUS INTEGRATION-PLUS Standard 75007775 D11573 5 SL-PLUS INTEGRATION-PLUS Lateral

SLR-PLUS™ SAP No. Art. No. Size Description 75007777 D11805S 5 SLR-PLUS Standard 75007629 91000155 5 SLR-PLUS Lateral 75007779 D11865 5 SLR-PLUS INTEGRATION-PLUS Standard

21 SL-PLUS™ MIA Dimensions All dimensions in mm

Specifications Neck Height Standard Size Stem Stem M/L CCD CCD Size XS/–3 S/+0 M/+4 L/+8 XL/+12 XXL/+16 length I length II width angle angle Standard Lateral 01 128 109 26 131 01 17 19 21 24 27 32 0 132 113 27 131 0 18 19 22 25 27 33 1 137 117 28 131 123 1 18 20 23 25 28 33 2 141 121 29 131 123 2 19 21 23 26 29 34 3 145 124 30 131 123 3 20 21 24 27 29 35 4 150 128 32 131 123 4 20 22 25 27 30 35 5 154 132 33 131 123 5 21 23 26 28 31 36 6 159 136 34 131 123 6 22 24 26 29 32 37 7 163 140 35 131 123 7 23 25 27 30 32 38 8 168 144 37 131 123 8 24 25 28 31 33 39 9 173 148 39 131 123 9 24 26 29 32 34 39 10 178 152 40 131 123 10 25 27 30 33 35 40 11 183 157 42 131 123 11 26 28 31 34 36 41 12 188 162 44 131 123 12 27 29 32 35 37 42

Offset Standard Lateral Size XS/–3 S/+0 M/+4 L/+8 XL/+12 XXL/+16 Size XS/–3 S/+0 M/+4 L/+8 XL/+12 XXL/+16 01 28 30 33 36 39 42 01 0 29 31 34 37 40 43 0 1 30 32 35 38 41 44 1 35 38 41 44 48 51 2 31 33 36 39 42 45 2 37 39 42 46 49 52 3 32 34 37 40 43 46 3 38 40 44 47 50 54 4 33 35 38 41 44 47 4 39 42 45 48 52 55 5 34 37 40 43 46 49 5 41 43 47 50 53 57 6 36 38 41 44 47 50 6 42 45 48 51 55 58 7 37 39 42 45 48 51 7 44 46 50 53 56 60 8 38 41 44 47 50 53 8 45 48 51 55 58 61 9 40 42 45 48 51 54 9 47 50 53 56 60 63 10 41 43 46 49 53 56 10 49 51 55 58 61 65 11 43 45 48 51 54 57 11 51 53 56 60 63 66 12 44 46 50 53 56 59 12 52 55 58 61 65 68

Neck length Standard Lateral Size XS/–3 S/+0 M/+4 L/+8 XL/+12 XXL/+16 Size XS/–3 S/+0 M/+4 L/+8 XL/+12 XXL/+16 01 17 20 24 28 32 36 01 0 18 21 25 29 33 37 0 1 19 22 26 30 34 38 1 26 29 33 37 41 45 2 20 23 27 31 35 39 2 27 30 34 38 42 46 3 21 23 27 31 35 39 3 28 31 35 39 43 47 4 22 24 28 32 36 40 4 29 32 36 40 44 48 5 22 25 29 33 37 41 5 30 33 37 41 45 49 6 23 26 30 34 38 42 6 32 34 38 42 46 50 7 24 27 31 35 39 43 7 33 36 40 44 48 52 8 25 28 32 36 40 44 8 34 37 41 45 49 53 9 26 29 33 37 41 45 9 35 38 42 46 50 54 10 27 30 34 38 42 46 10 37 40 44 47 52 56 11 29 31 35 39 43 47 11 38 41 45 49 53 57 12 30 33 37 41 45 49 12 39 42 46 50 54 58

22 Offset Lateral

XXL/+16 Offset Standard XL/+12 M/L width L/+8

M/+4

Neck height S/+0

XS/-3 Neck length Standard

Neck length Stem length I CCD angle Lateral

Stem length II

23 SL-PLUS™ Dimensions All dimensions in mm

Specification Neck Height Standard Size Stem Stem M/L CCD CCD Size XS/–3 S/+0 M/+4 L/+8 XL/+12 XXL/+16 length I length II width angle angle Standard Lateral 01 128 110 29 131 01 16 18 21 24 26 29 0 132 113 30 131 0 17 19 22 24 27 29 1 137 117 32 131 123 1 18 20 23 25 27 30 2 141 121 33 131 123 2 18 20 23 26 28 30 3 145 124 34 131 123 3 19 21 24 26 29 31 4 150 128 35 131 123 4 20 22 25 27 29 32 5 154 132 37 131 123 5 21 22 25 28 30 33 6 159 136 38 131 123 6 21 23 26 29 31 33 7 163 140 40 131 123 7 22 24 27 29 32 34 8 168 144 42 131 123 8 23 25 28 30 33 35 9 173 148 44 131 123 9 24 26 29 31 34 36 10 178 153 45 131 123 10 25 27 30 32 35 37 11 183 157 48 131 123 11 26 28 31 33 36 38 12 188 162 50 131 123 12 27 29 32 34 37 39

Offset Standard Lateral Size XS/–3 S/+0 M/+4 L/+8 XL/+12 XXL/+16 Size XS/–3 S/+0 M/+4 L/+8 XL/+12 XXL/+16 01 28 30 33 36 39 42 01 0 29 31 34 37 40 43 0 1 30 32 35 38 41 44 1 35 38 41 44 48 51 2 31 33 36 39 42 45 2 37 39 42 46 49 52 3 32 34 37 40 43 46 3 38 40 44 47 50 54 4 33 35 38 41 44 47 4 39 42 45 48 52 55 5 34 37 40 43 46 49 5 41 43 47 50 53 57 6 36 38 41 44 47 50 6 42 45 48 51 55 58 7 37 39 42 45 48 51 7 44 46 50 53 56 60 8 38 41 44 47 50 53 8 45 48 51 55 58 61 9 40 42 45 48 51 54 9 47 50 53 56 60 63 10 41 43 46 49 53 56 10 49 51 55 58 61 65 11 43 45 48 51 54 57 11 51 53 56 60 63 66 12 44 46 50 53 56 59 12 52 55 58 61 65 68

Neck Length Standard Lateral Size XS/–3 S/+0 M/+4 L/+8 XL/+12 XXL/+16 Size XS/–3 S/+0 M/+4 L/+8 XL/+12 XXL/+16 01 17 20 24 28 32 36 01 0 18 21 25 29 33 37 0 1 19 22 26 30 34 38 1 23 26 30 34 38 42 2 20 23 27 31 35 39 2 24 27 31 35 39 43 3 21 23 27 31 35 39 3 25 28 32 36 40 44 4 22 24 28 32 36 40 4 26 29 33 37 41 45 5 22 25 29 33 37 41 5 27 30 34 38 42 46 6 23 26 30 34 38 42 6 28 31 35 39 43 47 7 24 27 31 35 39 43 7 29 32 36 40 44 48 8 25 28 32 36 40 44 8 30 33 37 41 45 59 9 26 29 33 37 41 45 9 31 34 38 42 46 50 10 27 30 34 38 42 46 10 33 36 40 43 48 52 11 29 31 35 39 43 47 11 34 37 41 45 49 53 12 30 33 37 41 45 49 12 35 38 42 46 50 54

24 25 SLR-PLUS™ Dimensions All dimensions in mm

Specification Neck Height Standard

Size Stem Stem M/L CCD CCD Size XS/–3 S/+0 M/+4 L/+8 XL/+12 XXL/+16 length I length II width angle angle Standard Lateral 1 170 151 30 131 123 1 17 19 22 25 27 30 2 175 156 32 131 123 2 18 20 23 25 28 31 3 181 160 33 131 123 3 19 21 23 26 29 31 4 186 165 34 131 123 4 20 22 24 27 29 32 5 191 170 35 131 123 5 20 22 25 28 30 33 6 197 175 37 131 123 6 21 23 26 28 31 34 7 203 180 38 131 123 7 22 24 27 29 32 35 8 208 185 40 131 123 8 23 25 27 30 33 35 9 215 191 42 131 123 9 24 26 28 31 34 36 10 221 196 44 131 123 10 25 27 29 32 35 37 11 227 202 45 131 123 11 26 28 30 33 36 38

Offset Standard Lateral

Size XS/–3 S/+0 M/+4 L/+8 XL/+12 XXL/+16 Size XS/–3 S/+0 M/+4 L/+8 XL/+12 XXL/+16 1 30 32 35 38 41 44 1 35 38 41 44 48 51 2 31 33 36 39 42 45 2 37 39 42 46 49 52 3 32 34 37 40 43 46 3 38 40 44 47 50 54 4 33 35 38 41 44 47 4 39 42 45 48 52 55 5 34 37 40 43 46 49 5 41 43 47 50 53 57 6 36 38 41 44 47 50 6 42 45 48 51 55 58 7 37 39 42 45 48 51 7 44 46 50 53 56 60 8 38 41 44 47 50 53 8 45 48 51 55 58 61 9 40 42 45 48 51 54 9 47 50 53 56 60 63 10 41 43 46 49 53 56 10 49 51 55 58 61 65 11 43 45 48 51 54 57 11 51 53 56 60 63 66

Neck Length Standard Lateral

Size XS/–3 S/+0 M/+4 L/+8 XL/+12 XXL/+16 Size XS/–3 S/+0 M/+4 L/+8 XL/+12 XXL/+16 1 19 22 26 30 34 38 1 23 26 30 34 38 42 2 20 23 27 31 35 39 2 24 27 31 35 39 43 3 21 23 27 31 35 39 3 25 28 32 36 40 44 4 22 24 28 32 36 40 4 26 29 33 37 41 45 5 22 25 29 33 37 41 5 27 30 34 38 42 46 6 23 26 30 34 38 42 6 28 31 35 39 43 47 7 24 27 31 35 39 43 7 29 32 36 40 44 48 8 25 28 32 36 40 44 8 30 33 37 41 45 49 9 26 29 33 37 41 45 9 31 34 38 42 46 50 10 27 30 34 38 42 46 10 33 36 40 43 48 52 11 29 31 35 39 43 47 11 34 37 41 45 49 53

26 Lateral XXL/+16 Standard XL/+12 L/+8 M/L Width M/+4 S/+0

Neck height XS/-3

Neck length Standard

CCD angle Neck length Lateral

Stem length I

Stem length II

27 SL-PLUS™ MIA Instrument Set

SL-PLUS™ MIA Basic Set Set No. SAP/Art. 75200838/0942010

SAP No. Art. No. Description Size 75002198 110450 Case Basic Instruments 75007661 990019 Easy Tray Lid Plastic 1 75002319 110901 Slap Hammer 2 75002320 110902 Extraction Block 3 75002325 110911 Extraction Screw M8 4 75002165 110249 Extraction Screw M6 5 75002160 110242 Head Impactor 6 75007255 600621 MIA Stem Impactor 7 75004495 21000138 MIA Curved Rasp 8 75006420 41000030 SL-PLUS MIA Guide Rasp 1-6 9 75100839* 75100839 Trial Ball Head 22 S/+0 75100840 75100840 Trial Ball Head 22 M/+4 75100841 75100841 Trial Ball Head 22 L/+8 75100842* 75100842 Trial Ball Head 22 XL/+12 75100843* 75100843 Trial Ball Head 28 XS/-3 75100844 75100844 Trial Ball Head 28 S/+0 75100845 75100845 Trial Ball Head 28 M/+4 75100846 75100846 Trial Ball Head 28 L/+8 75100847 75100847 Trial Ball Head 28 XL/+12 75100848 75100848 Trial Ball Head 28 XXL/+16 75100849* 75100849 Trial Ball Head 32 XS/-3 75100850 75100850 Trial Ball Head 32 S/+0 75100851 75100851 Trial Ball Head 32 M/+4 75100852 75100852 Trial Ball Head 32 L/+8 75100853 75100853 Trial Ball Head 32 XL/+12 75100854 75100854 Trial Ball Head 32 XXL/+16 75100855* 75100855 Trial Ball Head 36 XS/-3 75100856 75100856 Trial Ball Head 36 S/+0 75100857 75100857 Trial Ball Head 36 M/+4 75100858 75100858 Trial Ball Head 36 L/+8 75100859 75100859 Trial Ball Head 36 XL/+12 J 75002203* 110500 Adapter for trial rasp (SL-PLUS standard) * Optional Instruments

Optional:

SAP No. Art. No. Description Size 75210292 75210292 SET 40 mm Trial Ball Head XS/-4 to L/+8 75210293 75210293 SET 44 mm Trial Ball Head XS/ 1-6 to L/+8 75018153 21000594 SL-PLUS MIA Guide Rasp 01-3 75000645 21000397 MIA Curved Rasp 45° 75004615 21000265 MIA Start Rasp 75004683 21000359 MIA Rasping Machine (IMT Woodpecker)

28 8

7

1

4 3

J OPTIONAL

9

2 5 6

29 SL-PLUS™ MIA Trial Rasps Set Set No. SAP 75210409

Set No. SAP No. Description Size 75007312 600930 MIA Instrument Case for SL 75007661 990019 Easy Tray Lid Plastic 1 75004481 21000123 MIA Trial Rasp 01 75004482 21000124 MIA Trial Rasp 0 75004483 21000125 MIA Trial Rasp 1 75004484 21000126 MIA Trial Rasp 2 75004485 21000127 MIA Trial Rasp 3 75004486 21000128 MIA Trial Rasp 4 75004487 21000129 MIA Trial Rasp 5 75004488 21000130 MIA Trial Rasp 6 75004489 21000131 MIA Trial Rasp 7 75004490 21000132 MIA Trial Rasp 8 75004491 21000133 MIA Trial Rasp 9 75004492 21000134 MIA Trial Rasp 10 75004493* 21000135 MIA Trial Rasp 11 75004494* 21000136 MIA Trial Rasp 12 2 75004603 21000253 MIA modular neck/trial rasp stand. 01–0 Std. 75004604 21000254 MIA modular neck/trial rasp stand. 1-6 Std. 75004605 21000255 MIA modular neck/trial rasp stand. 7-12 Std. 3 75004606 21000256 MIA modular neck/trial rasp lat. 1–6 Lat. 75004607 21000257 MIA modular neck/trial rasp lat. 7-12 Lat. 4 75006419 41000029 MIA Box Chisel 5 75009352* SYS251374 Trochanter Retractor *Optional Instruments

Optional Rasp Adapters Set No. SAP No. Art No. Description Size 75200166 75007307 600920 Offset-Adapter 25mm 25mm 75200168 75007308 600921 Offset-Adapter 40mm 40mm 6 75200169 75007309 600922 Offset-Adapter 10mm 10mm 7 75007310 600923 Offset-Adapter 17mm left 17/13mm 75200171 8 75007311 600924 Offset-Adapter 17mm right 17/13mm 75004612 21000262 Offset-Adapter 60mm left 60/25mm 75210202 75004613 21000263 Offset-Adapter 60mm right 60/25mm 75210372 75004614 21000264 Offset Adapter 45° w. Plate 45° 75210406 75004468 21000107 Offset Adapter w. Plate 25mm 75000642 21000378 Modular Knock Plate

30 4

5

OPTIONAL

1

OPTIONAL

OPTIONAL 3 2

7

6

8

31 SL-PLUS™ SLR-PLUS™ Instrument Set

SL/SLR-PLUS™ Basic Set Set. No. SAP/Art. 75200142/0942000

SAP No. Art. No. Description 75002198 110450 Case Basic Instruments 1/4 75007661 990019 Easy Tray Lid Plastic 1 75002319 110901 Slap Hammer 2 75002320 110902 Extractor Block 3 75002321 110903 Impactor 4 75002322 110904 Box Chisel 5 75002325 110911 Extraction Screw M8 6 75002160 110242 Head Impactor 7 75100839* 75100839 Trial Femoral Head 22 S/+0 75100840 75100840 Trial Femoral Head 22 M/+4 75100841 75100841 Trial Femoral Head 22 L/+8 75100842* 75100842 Trial Femoral Head 22 XL/+12 75100843* 75100843 Trial Femoral Head 28 XS/-3 75100844 75100844 Trial Femoral Head 28 S/+0 75100845 75100845 Trial Femoral Head 28 M/+4 75100846 75100846 Trial Femoral Head 28 L/+8 75100847 75100847 Trial Femoral Head 28 XL/+12 75100848 75100848 Trial Femoral Head 28 XXL/+16 75100849* 75100849 Trial Femoral Head 32 XS/-3 75100850 75100850 Trial Femoral Head 32 S/+0 75100851 75100851 Trial Femoral Head 32 M/+4 75100852 75100852 Trial Femoral Head 32 L/+8 75100853 75100853 Trial Femoral Head 32 XL/+12 75100854 75100854 Trial Femoral Head 32 XXL/+16 75100855* 75100855 Trial Femoral Head 36 XS/-3 75100856 75100856 Trial Femoral Head 36 S/+0 75100857 75100857 Trial Femoral Head 36 M/+4 75100858 75100858 Trial Femoral Head 36 L/+8 75100859 75100859 Trial Femoral Head 36 XL/+12 8 75004495* 21000138 Curved Rasp * Optional Instruments

Optional: SAP No. Art No. Description 75210292 75210292 SET 40mm Trial Femoral Heads XS/-4 to L/+8 75210293 75210293 SET 44mm Trial Femoral Heads XS/-4 to L/+8 75007254 600620 MIS Offset Box Chisel (instead of 110904) 75004683 21000359 MIA Rasping Machine (IMT Woodpecker)

32 Instrument Set

3 2

4

6 5

7

1

8 OPTIONAL

33 SL-PLUS™ Instrument Set

SL-PLUS™ Trial Rasp Set Set. No. SAP 75210410

SAP No. Art No. Description 75007312 600930 Instrument Case 75007661 990019 Easy Tray Lid Plastic 1 75007293 600899 Trial Rasp 01 75007294 600900 Trial Rasp 0 75007295 600901 Trial Rasp 1 75007296 600902 Trial Rasp 2 75007297 600903 Trial Rasp 3 75007298 600904 Trial Rasp 4 75007299 600905 Trial Rasp 5 75007300 600906 Trial Rasp 6 75007301 600907 Trial Rasp 7 75007302 600908 Trial Rasp 8 75007303 600909 Trial Rasp 9 75007304 600910 Trial Rasp 10 75007305* 600911 Trial Rasp 11 75007306* 600912 Trial Rasp 12 2 75004603 21000253 Trial Neck Std 01–0 75004604 21000254 Trial Neck Std 1–6 75004605 21000255 Trial Neck Std 7–12 75004606 21000256 Trial Neck Lat. 1–6 75004607 21000257 Trial Neck Lat. 7–12 *Optional Instruments

Optional Rasp Adapters Set No. SAP No. Art No. Description Size 3 75200166 75007307 600920 Offset-Adapter 25mm 25mm 75200168 75007308 600921 Offset-Adapter 40mm 40mm 75200169 75007309 600922 Offset-Adapter 10mm 10mm 4 75007310 600923 Offset-Adapter 17mm left 17/13mm 75200171 5 75007311 600924 Offset-Adapter 17mm right 17/13mm 75004612 21000262 Offset-Adapter 60mm left 60/25mm 75210202 75004613 21000263 Offset-Adapter 60mm right 60/25mm 75210406 75004468 21000107 Offset Adapter w. Plate 25mm 75000642 21000378 Modular Knock Plate

34 5

3

4

2

1

35 SLR-PLUS™ Instrument Set

SLR-PLUS Trial Rasp Set Set. No. SAP 75210408

SAP No. Art. No. Description 75017292 21000569 Case Instruments 75007661 990019 Easy Tray Lid 75017269 21000558 Trial Rasp 1 1 75017270 21000559 Trial Rasp 2 75017271 21000560 Trial Rasp 3 75017272 21000561 Trial Rasp 4 75017273 21000562 Trial Rasp 5 75017274 21000563 Trial Rasp 6 75017275 21000564 Trial Rasp 7 75017276 21000565 Trial Rasp 8 75017277 21000566 Trial Rasp 9 75017278 21000567 Trial Rasp 10 75017279 21000568 Trial Rasp 11 2 75004604 21000254 Modular Neck Standard 1–6 75004605 21000255 Modular Neck Standard 7–12 3 75004606 21000256 Modular Neck Lateral 1–6 75004607 21000257 Modular Neck Lateral 7–12

Optional Rasp Adapters Set No. SAP No. Art. No. Description Size 75200166 75007307 600920 Offset-Adapter 25mm 25mm 75200168 75007308 600921 Offset-Adapter 40mm 40mm 4 75200169 75007309 600922 Offset-Adapter 10mm 10mm 5 75007310 600923 Offset-Adapter 17mm left 17/13mm 75200171 6 75007311 600924 Offset-Adapter 17mm right 17/13mm 75004612 21000262 Offset-Adapter 60mm left 60/25mm 75210202 75004613 21000263 Offset-Adapter 60mm right 60/25mm 75210372 75004614 21000264 Offset Adapter 45° w. Plate 45° 75210406 75004468 21000107 Offset Adapter w. Plate 25mm 75000642 21000378 Modular Knock Plate

36 OPTIONAL 5

OPTIONAL 4

OPTIONAL 6 2 3

1

37 Sterilization

Implants All the implants described in this Surgical Technique are sterile when they are delivered by the manufacturer. Resterilization is not allowed.

Instruments System components and instruments are not sterile when they are delivered. Before use they must be cleaned by the usual methods in accordance with internal hospital regulations and sterilized in an autoclave in accordance with the legal regulations and guidelines applicable in the relevant country. For detailed information please refer to leaflet No. 003389 [1363].

The correct settings are given in the instructions for use issued by the autoclave manufacturer. Instrument manufacturers and dealers accept no responsibility for sterilization of products by the customer.

38 Notes

39 40

Manufacturer www.smith-nephew.com Smith & Nephew Orthopaedics AG Oberneuhofstrasse 10d ™Trademark of Smith & Nephew. CH-6340 Baar ©2016 Smith & Nephew, Inc Switzerland 06956-en V2 0817

0123