NEW METHODS FOR PATHOGENS : A REVLEW DANIEL Y. C. FUNG The Pennsylvania State University

Intr ductioii

Interests in new methods for identification of bacterial pathogens have increased greatly in recent years due to the awareness of, the need for, and the potential of rapid methods and automation in Micro- biology. In 1971 a report by the U .So National Institute of General Medical Sciences presented the status of mechanization and automation in the clinical laboratory; autcnnation in microbiology was one of the topics reviewed (Kinney, 19-71). Richardson (1972) summttrized the developments of automation in the dairy Laboratories. These topics have been discussed in detail in the International Symposium on Rapid Methods and Automation in Microbiology first held L~IStockholm, Sweden, 1973, and again in Cambridge University, England, 1976. Another symposium was held in Kiel, Germny, 1974 with the title Automation of Microbio1ogr;ical Food Analysis. TheIn proceedings of the First International Symposium were published in two volumes : Automation in Microbiology and Innrmnolom (Heden and Illeni, 1975a) and New Approaches to the Identification of Microorganisms (Heden and Illeni, 1975b). The purpose of this article is to summarize the highlights of some automated microbiological tests and rapid methods relevant to the food industry. Hopefully some of these methods could be adapted and used routinely by the food industry.

New Approaches and Automation in Microbioloa Generally the approach to developing new nethods for pathogens involved either a search for new ideas and related instrument develop- ment or improvement of existing bacteriological methods. The impetus for the development of new ideas and instruments mbly comes from clinical needs for rapid detection of organisms and antibiotic sensitivity testing. There are a rimer of new approaches to studying pathogens which involve measuring growth of organisms in the presence of various antibot ics . One way of improvbg antibiatic sensitivity testing in liquid medium is to miniaturize the entire procedure and design ways to dilute or dispense varying concentrations of antibiotics into miniaturized wells; into these wells a standardized nWer of the pathogens under investi- gation is inoculated by a convenient multipoint inoculator. After growth period the minimal inhibition concentration (MIC) of various

* Presented at the 29th Annual Reciprocal Meat Conference of the American Meat Science Association, 1976. 285 antibiotics could be ascertained by reading the endpoint of no growth in each series of wells. A machine called the Dptiterl (Dynatech Lab. Inc ., Alexandria, VA) can automatically dilute antibiotics and add nutrient and test organisms into the wells for rapid MIC studies in the Microtiter plates (a plastic plate with 8 x 12 wells). This machine can perform a variety of immunological and serological tests also. A machine named the Autotiter (Ames Corapany, Elkhart, ll?) can also perform similar features. Recently the Dynatech Lab. Inc. designed a machine that can dispense varying concentrations of antibiotics into the Microtiter plate for MIC determinations. Along with an automatic multipoint inoculator system and 'a convenient reading aid this system offers a convenient package for the determination of MIC; the machine is called MIC-2000. Various other dilution machines and procedures exist for rapid dilution of antibiotic and the addition of other liquids to the dilution series (e.g., Medimixes by Linbro Chemical Co. Inc., New Haven, CT; Titertek by Flow Lab., Rockville, MD; Bioreactor by Biomeca, Geneva, Switzerland; etc ) . Another way of improvement of antibiotic sensitivity test is to design methods to quickly detect light scattering changes in nutrient solution with and without antibiotics under rigidly controlled condition. An elaborate system, the AutoBac 1 system was designed for this purpose (Praglin et al., 1975; McKie et al., Bacterial isolates are first dispensed into liquid and adjusted1975). to the correct turbidity before introduced to a specially designed l3-cell chamber (a cuvette module) with I2 of the cells containing the desired antibiotic disc in each; one cell serves as the growth control. The cuvette module is then incubated in the machine. After 3-4 hrs, light scattering ability of each cell is monitored and matched against the control cell. A computer printout from the machine indicates the antibiogram of the organism under study. When prolonged incubation time Is required, the module is reincubated and scanned at a Later time. Laboratory evaluation for the AutoBac 1 system versus the conventional Kirby-Bauer method showed awagreement for Gram-negative organisms and a 93% agreement with (Stubb and Wicker, 1976). However some major discrepancies were observed for some organisms on certah anti- biotics by these authors as well as by Greenberry, et al. (1976). Due to the efficiency of the system, an attempt was made to use antimicrobial susceptibility profiles generated by the AutoBac 1 system in cooperation with a suitable computer program to rapidly identify bacterial isolates. A 95-97$ accuracy of identification were reported (Sielaff et al., 1976) when 481 clinical isolates were studied using this system as compared with conventional identification procedures.

1 Naming of machine and manufacturer in this article is for reference only 286

An ambitious system--The AutoMicrobic System--which can detect, identify and enumerate bacteria and yeast in urine samples was recently (1976) announced by McDonnel Douglas Astronautic Co. (East St. Louis, MD). This system consists of a small cuvette with 20 tiny culture wells; 16 of which contain freeze-dried, highly selective media (patent pending). Introduction of dilute urine into the cuvette was done by a pneumatically controlled system. The cuvettes are then placed in a reader-. Final printed results are obtained in I2 hrs after automatic measurement of the change in optical absorption of each well from the initially obtained valve (Jones 6 e., 1976) Microbiological evaluation of the system shawed an overall high accuracy (96 or above) compared to conventional methods for most test organisms (Algridge 5 e.,1976; Somrenwirth, 1976). This system seems to have good promise as a high capacity (240 clinical specimens) automated microbiological identification system.

Radiometry and microcalorimetry are two newer approaches to rapidly determine growth of bacteria in clinical and food samples. The radio- metric method relies on the fact that in a suitable medium containing radioactive glucose (or other sugars) and bacteria, radioactive C02 will be liberated due to the utilization of sugar by most coamon pathogens, The liberated radioactive C02 are then collected and allowed to pass through an ionization chamber. The degree of ionization depends on the amount of radioactive C02 collected and this is trans- lated to "Growth index" on the machine, Bactec, designed and marketed by Johnston Lab. Inc . (Cockeyville, MD) is a completely automated system for radiometric determination of microbial growth. A Large body of literature exists for radiometry detection of bacterial growth in clinical and environmental specimens (Schort et al. , 1973; Rosner, 1975; Buda and Bromn, 1975; Foley and Broman, 1975; Previte et al., 1976; etc.). This technique is basically designed for fermentative bacteria. In order to make use of this procedure for "non-fermentors," a special radionretric medium was developed which allowed the rapid detect ion of Pseudomonas and Alcaligenes (Previte et al. , 1975 ) . Bacterial activities in liquid and food systems can also be measured by small cha es in heat production due to bacterial growth (Sachs and Menefee, 19727 When more refined instruments are used, microcalorimetry can even be used to identify bacteria (Staples et al., 1973; Russel et al, 1975). An attempt to use radiometry and microcalorimetry for the rapid detection of food borne microorganisms was made by Lampi --et al. ( 1974) e Beside studying pure cultures, they reported that there is a high degree of correlation between viable cell count and radiometric measurements as meat loaf underwent spoilage; they did not repart practical usefulness of microcalorimetry in food systems.

When bacteria grow in a liquid medium there is a change of electrical conductivity; this can be measured by a procedure called electrical impedance measurement due to microbial activity. This principle has been exploited for the studying of microbial growth in media with and without antibiotic for antibiotic sensitivity testing (Cady, 1975; Ur and Brown, 1975). Detection of bacteria in blood and spinal fluids in children by this machine (Bactometer 2 Bactomatic Inc., Palo Alto, CA) was reported by Khan, et al.*claimed that average detection time was 8.5 hrs compared with 24 hrs needed for the conventional method. Other applications of Bactometer 2 include monitoring of anaerobic bacteria (Hardy et al.,detection of fecal coli (Mischak g.,1976). Measure of the degree of light generated by the luciferin-luciferase system in the presence of bacterial ATP offers another procedure to detect microbial growth in liquid samples (Chappella and Levin, 1968). The method has been used to detect bacterial growth in blood cultures and urine (Schtock et al., 1976; Gutekunst et al., 1976). One important aspect of this assay is the need to destroy ATP of nonbacterial origin in the biological samples before performing the bacteria ATP measurement. Lab-Line ATP Photometer (Isb-Line Plaxa, Melrose Park, IL) and Wont 760 luminescence Biometer (Wont, Wilmington, DE) are two instruments designed for this test.

Identification of bacteria, especially anaerobic organisms, by gas chromatography is well known (Holdermen and Moore, 1972; Dowel1 and Hawkin, 1972; Mitruka, 1975). Application of gas-liquid chromatography to biological samples to ascertain microbial presence and activities have also been reported, e.g., the measurements of Lactic acid in spinal fluid of bacterial meningitis (Brook et al., 1976; Bricknell et al., 1976). Pyrolysis in combination with gas-liquid chromatography provides another means of rapid and automated bacterial identification (Meuzelaar et al., 1975; Kistemaker et al., 1975; Oxborrow et al., 1976). &king use of this combination procedure an instrument called Bacterial ID system TOO series is marketed by Chemical Data System, Inc., Oxford, PA.

An ingenious and quite unique way of identification of microorganisms is provided for by a technique called differential light scattering. In this system a laser beam is applied to a bacterial suspension; the scattered light (due to the various components of the bacterial cell) is collected by a rotating detector. The intensity of light collected at various angles provides a specific pattern. Since different bacteria scatter light differently, the patterns can be used to identify the organisms when compared to standard "prints" of known cultures. More- Over, when the bacterial cells are challenged with certain antibiotics the cell structures undergo changes, this can readily be detected by this system. Using this technique antibiotic sensitivity testing can also be performed. For greater detail of the principles and applications of this method, the article by Wyatt should be consulted (Wyatt, 1973). The commercial products for this test are Differential I, I1 and I11 marketed by Science Spectrum, Santa Barbara, CA.

There are no doubt many other ideas and techniques for rapid and novel ways to identify bacteria (Goldschmidt, 1970). Most of these tests are designed for pure culture studies and antibiotic sensitivity tests. More work are needed to determine the usefulness of these methods in food systems where mny other interfering compounds are present to msk the identify of pathogens in the natural habitats. 288

Diagnostic Kits

Another area of research and development of new methods for pathogens centers on convenient and rapid diagnostic kits. The one closest to the conventional test is the R/B system (Diagnostic Research Inc., Roslyn, NY) which consists of the basic 2 tube R/B system using 8 agar-base reactions and the 2 tube Expanders system which utilizes another 6 solid media tests. A special sterile needle is used to stab-streak the suspect colony into the tubes. Reading of reactions are made after 18 to 24 hrs at 35 to 37 C incubations. The results can be compared with the Enteric Analyzer for ease of inter- pretation. Reliability of this system compared with conventional system in the identification of enteric was cfted at 90-96$ (Isenbery and Painter, 1971; Smith & g.,197la and 1971b). Studies by O'Donnel --et al. (1970) and Brown (1973) indicated a high 99-106 accuracy when certain species were examined while Martin --et al. (1970) could only ascribe a poor 45% accuracy. (In citing accuracy, the author only used accurate identification of the organisms under study. A discussion on the accuracy of individual tests between systems is not made due to limitation of space; the readers are referred to the original articles for such informtions).

The Enterotube system consists of I1 agar-base tests housed in a compartmented plastic tube (Roche Diagnostics, Nutley, NJ) . Inoculation is done by touching an isolated bacterial colony with the tip of a special needle of the kit and pulling the needle through the kit. As the needle is pulled through the chambers bacteria are deposited into various media. After incubation (38-24 hrs) appropriate reagents are added to some chanibers and reactions are recorded. To facilitate identification of unknown, a convenient booklet (ENCISE) with special codes and keys is available for keying out the enterics. A computer consultation service is available for helping technicians to identify cultures. Thh system is easy to use and seems to have wide acceptance. The disadvantage is that only I1 tests are available and the short shelf-life. The original system was designed for Enterobacteriacea; recently (1976) a new system is marketed for the identification of non- fermentors. Compared with conventional tests the Fhterotube in combination with ENCISE ID system did very well (caaccuracy) in providing rapid and efficient identification of (Grunbery et e.,1969; Elston et ale, 1971; Tomfobrde e.t; al., 1973; Painter and Isenbery, 1973; Wust and Kayser, 1974; Isenberg et al., 1975). Both the R/B system and the Enterotube suffer from short shelf- life because of dehydration of agar and the lack of flexibility of application

Dehydrated media or paper discs impregnated with media will prolong shelf-lif'e of the diagnostic kits. The Auxotab system (Wilson Diagnostic Inc., Glenwood, IL) consists of a series of carded bacterial identification systems each containing ten reagent filled capillary chambers. After the addition of cell suspension, results of reactions can be read and common pathogens identified within 7 hrs. These tests depend on enzymatic reactions of the cell suspension; therefore aseptic technique is not necessary since bacterial cell division does not occur in this system. Stability of this system is about 18 months. This system has been used to study species of Enterobacteriaceae (Cottereau et al., 1967; Barbe, 1969) and a wide range of aerobic and anaerobic organisms (Buissiere et al., Joubert et al., 1968). More recent evalwtions indicate that1967; the Auxotab system is accurate at 87s (Rhoden et al., 1973a) and after improvements, advanced to 96% (Rhoden -et a1-*, 1973b). Another dehydrated system is the API system (Analytab Product, Inc ., Carle Place, NY) which consists of the API 20 system and the API 50 system, the formr has 20 tests and the latter has 50 tests. The dehy- drated media. are housed in small plastic chambers in a convenient plastic card. Suspensions of the organisms are introduced to the individual chambers. After 18-24 hrs at 37 C and approprbte reagents introduced, the reactions are read. Data are matched with the API Coder and API Profile Register for identification. To facilitate accurate identification, the company has a computer service to help technicians to interpret data either by direct usage of the computer bank or utilization of telephone service for informtion. Overall accuracy of this system was cited at 93-96$ (Washington et al., 1971; Smith et e.,1972). A new anaerobic system was recently introduced (API 2OA): the accuracy was cited at 9@ (Hauser and Zabranslfy, 1976). The API 50 system is useful for studying biotypes at intrasubspecific level (Tonrfohred and D'hto, meAPI system is also adapted for the rapid identification of yeast (Burgard and Sommenwirth, 1976). This system is easy to use and has long shelf-life (2 years) but the system is not sterile, therefore SOE difficulty may develop when tests are incubated for 48 hrs . The next two systems to be described havC stability in storage (1 to 2 years) as well as flexibility. "he PathoTec system (General Diagnostic, Morris Plain, NJ) is sold separately as 12 different paper strips each impregnated with appropriate substrates. An isolate is first suspended in liquid and then pipetted into a series of I2 tubes each containlng one paper strip. Reactions are allowed to occur and changes of color on the strips are recorded. Since only enzymatic reactions are utilized, the resvlts can be read in 4 hrs and aseptic techniques are not necessary. Also since the strips are sold separately the user can chose the strips they need for testing. The accuracy of the system is approximately 95% (Blazeic et &., 1973; Rosner, 1973). The advantage of PathoTec system is speed7 hrs) and flexibility; however the need to set up the tests and subsequent cleaning of test tubes make this system less desira5le for technicians in terms of operational convenience. It seems easy enough to make a kit for this system but in so doing flexibility of the system will have to be sacrificed.

The newest entry into the market is the Minitek system of BBL (Baltimore, MD). It consists of a l2-we11 plastic plate, a dispenser (which allows ease and flexibility of dispensing substrate impregnated paper discs into the plastic plate), and a convenient plunger to introduce 0.1 ml of bacterial suspension into the small wells. After incubation time the reactions are recorded and results compared with a color Coder to aid identification of unknowns. At least 35 different substrates are available for use in the system; thus providing the user with a good choice of tests. The accuracy of this system is about 97% compared to the conventional method (Hansen et al., 1974; Kiehn et al., 1974; Hansen and Stewart, 1976). The Minitek system can also be used for anaerobic organisms; of 108 clinical isolates tested all were correctly identified compred with the more tedious conventional anaerobic cultivation techniques (McCarthy et al., 1976).

All these commercial kits have been studied extensively both by the manufacturers and impartial investigators as to the ability to identify unknowns. Depending upon the investigators, many showed reliability of go$ or above after improvements are made. (One commercial product not mentioned in this review was withdrawn from the market due to consistence poor mrket testing and evaluations). A comprehens ive study of five test kits--APP, Enterotube, Auxotab, Pathotec, and R/B--versus the conventional tests was made by Nord et al. They tested 329 strains of Enterobacteriaceae with all six methods and concluded that as far as accurate identification of unknown is concerned the ranking was as follows: Enterotube (gfi), AS($$), R B (91$), PathoTec (7s), and Auxotab (42$). In another study, Maussa h75) using E, PathoTec, and Enterotube compared with conventional tests reported the follawing results : Enterotube ( 99.2$), API- (512.25)~ and PathoTec (80.5%). It should be emphasized that these studies were done a few years ago before many important improvements and modifications were made by PathoTec and Auxotab. A recent comparison between ASand Minitek showed that Minitek gave 97.65 accuracy and API- 95.3% (Aquino and Dowell, 1975). A comparison of these 2 methods against conventional methods for identification of 175 anaerobes showed that the percentage correlation of both positive-negative reactivities with the API- anaerobic system ranged from 70-99.4$ and with the Minitek from 97.1- lo@ ( Hansen and Stewart, 1976). When and Minitek systems were compared to conventional techniques for isolates, an accuracy of 85% for both methods was reported (Shayegani et a1., 1975). An interesting "double blind" identification study was made by three hospital laboratories each contributed 130 isolates and each tested 100 randoanly chosen "blinds" from the pool of cultures, using the conventional method, R/B system and API 20 system. The results showed that all three methods provided identical genus and species identification with more than 96 of the isolatcs (Ayres et al., 1976).

It is of interest to know also the relative costs of these systems compared with conventional procedures in the identification of isolates. The only data available are for the identification of isohtes in the family Enterobacteriaceae. For the conventional 7-tube (10 test) setup and lT-tube (20 test) setup, the cost per isolate identification is $3.60 and $7.98, respectively. For a similar identification using API 10s and API 20E the cost is $2.33 and $3.02, respectively (Robertson, et al., 1976]Estimation of costs for each identification by N. A. Cox7personai communication) is as follows: API 20 ($2.17), ($4.88), Enterotube ($1.85), Minitek 12 test m),Minitek (&.74), and R/B 4 tube system ($2.03). The values are obtained for purchasing of lots of 100 for each system. It seems that these commerciai. diagnostic kits are less expensive and more convenient than the conventional procedures while offering accurate identification of bacteria isolates.

Miniatur ized Microbiological Techniques

hny of the recently developed diagnostic kits can be called minia- turized microbiological techniques . Summaries of miniaturized micro- biological methods have been made by Hartman (l968), and Fung and Hartman ( 1972, 1975 1 The system developed by Fung (1969) for the identification of bacterial isolates involves miniaturization of tests as well as multiple inoculation of mny organisms to a variety of test substrates. The general procedure of this system involves obtaining pure cultures, preparation of a master plate, aseptic multiple inoculation of cultures into liquid or onto solid media, observation of biochemical changes and collection of data and interpretation.

Microtiter plates (96 wells, each holds 0.35 ml) or vessels with many wells can be used for cultivation of organisms. Sterilization of the plates can be achieved by socking the plates in 500 ppm hypochlorite solution (1hr) and rinsing the plates with sterile distilled water. Ultraviolet and CO-60 irradiation (Jape-William, 1975) are also convenient ways to sterilize the plates. Liquid, semi-solid, and solid media can be introduced into the wells of the plate by a hand pipette or by auto- mtic pipetting machine. A large petri dish (15 x 150 mm) can also be used for solid media.

In order to facilitate mass transfer of test cultures into substrates in the Microtiter plate or onto solid media in petri dishes a multiple inoculator can be constructed by fixing 96 stainless steel pins (27 mm long) into woodblocks or other materials; the design of the 96 pins is patterned after the 8 x 3.2 configuration of the Microtiter plate. Pin heads are suitable for inoculation into liquid media and solid agar surfaces while pin points are suitable for stabbing into agar or semi- solid agar. Sterilization of the device is by alcohol-flaming. The sterile device can be used to change inocula from a Master plate containing test organisms (mximum of 96 unknown) to either Microtiter plates containing suitable substrates or petri dishes containing solid agar.

Routine use of this procedure usually involves collection of a Large number of unknown cultures. The cultures are gruwn overnight in broth and a few drops of the culture are introduced into the wells of a Microtiter plate to form a mster plate. From the master plate the organisms are Cransferred to suitable solid or liquid media by use of the multiple inoculation device. Incubation time can be shortened to 6-8 hrs for many tests. Interpretation of results is by use of pre- designed charts and tables of individual investigators. The author has used this procedure for carbohydrate tests (F'ung and Miller, 1970), IMViC tests (F'ung and Miller, 1972), litmus milk test (Rug and Hartmn, 1975) effect of dyes on bactgrial growth (Fung and Miller, 1973), and others. Organisms tested included a variety of Gram-positive and Gram-negative organisms of Public Health significance. The use of this system for rapid and convenient mass identification of Streptococci (Hahn, 1974), Enterobacteriaceae (Newsom, 1973), fish pathogens (Chen , 1976), Streptococci, lactobacilli, and Gram-negative rods from animal sources (Jayne-Williams, 1975 and 1976), anaerobic bacteriology (Wilkins et al. , 1975a and 1975b) etc . have been reported. The type of tests described by these authors covered the whole gamut of indicating that most microbiological tests can be miniaturized.

The advantage of this system are low cost of operation, flexibility of tests, mass production of data for many unknowns, speed of reactions and savings of space and time. The disadvantages of this system include the need to have manual dexterity, experience, and training. Also for laboratories testing limited numbers of organisms at one time, setting up a complete system is inconvenient. An automatic machine can reduce the need for dexterity and the design of "kits" geared for the specific laboratories will reduce the number of complete plates to be prepared, i.e., design 8 sets of 12 tests in one Microtiter plate for studying 8 unknowns at a time.

C onc lus ion A variety of methods are discussed in this review on the search of new methods for the identification of pathogens. So= of these methods are applicable to the Food Industry Laboratories and others have some potential for adaptation to these laboratories. It is up to the laboratory directors to decide which procedure(s) or mthod(s) to use in place of or in conjuction with the conventional techniques. To venture into a new method, a director must be well informed and trained concerning the new mekhod and should know about the advantages and dis- advantages of the method as well as the knowledge of the potential and limitations of the system in relation to his particular needs and concerns.

References

Aldridge, C., S. Gibson, J. Ianham, M. Meyer, R. Vannest, Po Jones and R. Charles. 1976, An automted microbiological detection/DD system. 11. Microbiological consideration Abs . Ann. Mtg her SOC. Microbiol. 1976:C74. Aquino, T. I. and C . Dowell. 1975. A comparison of Minitek (BBL) and API (Analytab) for identification of Enterobacteriaceae. Abs. Ann. Mtg. Amer. SOC. Microbiol. 1975:Cll. 293

Ayres, L. W., T. L. Gavan, I. Rutherford and D. L. Taylor. 1976. A comparative study of three methods of identification of Enterobacter - iaceae. Abs. Ann. Wg. Amer. SOC.Microbiol. 1976:~17. Barbe, J. 1969. Organisation methodique de l'etude des caracterers enzymatiques des bacteries de la tribu des Klebsielleae. Thesis Faculte Mixte de Medecine et de Pharmacie de Marseille.

Bhzevic, D. J., P. C. Schreckenberger and J. M. btsen. 1973. Evaluation of the PathoTec !'Rapid I-D System. 'I Appl . Microb io1. 26 :ago.

Bricknell, K. S., P. T. Sugihara and J. Brook. 1976, Optimizing methylation conditions for gas -liquid chromatography assay of lactic acid in biological samples. Abs . Ann. Mtg. Amer . SOC. Microbiol. 1976 :c~c@.

Brook, I., K. S. Bricknell, G. Overturf and S. M. Finegold. 1976. Abnormalities in spinal fluid (CSF) detected by gas liquid chroma- tography (GU) in meningitis patients. Abs. Ann. Mtg. Amer. SOC. Microbiol. 1976:Cl07.

Brown, W. J. 1973. Evaluation of the Four tube R/B system for identi- f ication of Serratia mrcescens and Serratia liquefaciens . her. J. Med . Technol. 39:272. Buda, D. and R. Broman. 1975. Radiometric a say of gentamicin in blood by measurement of bacterial evolution of "CO, and 14C-urea. Abs. Ann. Mtg. her. SOC. Microbiol. 1975:c28 Buissiere, J., A. Fourcard and L. Colobert. 1967. Usage de substrats synthetiques pour 1' etude de l'equipment enzymatique des microorganisms . C. R. Acad. Sci. 264:415. Burgard, L. E. and A. C . Sonnenwirth. 1976. Evaluation of the API 20 clinical yeast system (API 2oC) for yeast identification. Abs. Ann. Mtg . her. SOC. Microbiol. 1976:C20. Cady, P. 1975. Rapid automated bacterial identification by impedance measurement. In New Approaches in the Identification of Microorganisms, Heden, C R. and T. Illeni, eds. 1975. John Wiley & Sons, New York. Chapter 6.

Chappelle, E. W. and G. V. Levin. Use of the firefly bioluminescent reaction for rapid detection and 1968.counting of bacteria. Biochemical Med. 2:41.

Chen, R. K., J. H. Nelson, L. D. Potocki, and S. Lee. 1976. A mechanized system for identification of fish pathogenic bacteria. Abs. Ann. Mtg. Amer. SOC. Microbiol. 1976:C44.

Cottereau, P., J. Buissiere and G. DeSaint-Aubert . 1967. La Colibacillose du veau. Bull. de la SOC. des Sci. Vet. de Lyon. 6:501. 294

Dowell, V. R. and T. M. Hawkins. 1972. Laboratory Methods in Anaerobic Bacteriology. U.S .D.H.E.W ., Center for Disease Control, Atlanta, GA. Elston, H. R., J. A. kudo, J. P. Stanek and M. Schaab. 1971. Multi- biochemical test system for distinguishing enteric and other gram- negative bacilli . Appl . Microbiol . 22 :408. Foley, T. and R. Broman. 1975. Radiometric identification of pathogenic Neisseria by 14C02 production from labelled sugar. Abs . Ann. Mtg. Amer . Soc . Microbiol. 1975 :C27. Fung, D .Y .C . Rapid methods for determining staphylococcal toxins and Salmonellae1969. associated with poultry products. R.D. Thesis. Iowa State University Library, Ames, IA. Fung, D .Y .C . and P. A. Hartmn. 1972. Rapid characterization of bacteria, with emphasis on Staphylococcus aureus. Can. J. Microbiol. 18 :1623 Fung, D.Y .C . and P. A. Hartman. 1975. Miniaturized microbiological techniques for rapid characterization of bacteria. In New Approaches to the Identsication of Microorganisms, Heden, C. R. and T. Illeni, eds . John Wiley and Sons, New York. Chapter 21. Fung, D. Y .C. and R. D. Miller. 1970. Rapid procedure for the detection of acid and gas production by bacterial cultures. Appl. Microbiol. 20: 527 Fung, D .Y .C . and R. D. Miller. 1972. Miniaturized techniques for IMViC tests. J. Milk and Food Technol. 35:328.

Fung, D.Y.C. and R. D. Miller. 1973, Effect of dyes on bacterial growth. Appl. Microbiol. 25:793. Goldschmidt, M. C . 1970. Instrumentation for microbiology-horizon unlimited. In Rapid Diamosis Methods in , Graber, C. D., ed. Williams and Wilkins Coo, Baltimore, MD. Chapter 24.

Greenberg, J. M., R. L. Gottschall and E. L. Randall. 1976. A comparison of AutoBac 1 results with early and conventional readings of Bauer-Kirby disc diffusion tests. Abs. Ann. Mtg. Amer. SOC. Microb io1. 1976 :C 111. Gutekunst, R. R., B. Jaffee and E. Chappelle. 1976. Detection of bacteria in urine by the firefly luciferase ATP assay. Abs. Ann. Mtg. her. Soc . Microbiol. 1976:C72. Hahn, G. 1975. Ein Schnelltest zur identifizierung von streptokokke. Archiv flu: lebensmittelhygiene 29:25. 295 Hansen, S. L., D. R. Hardesty and B. M. Myers. 1974. Evaluation of the BBL Minitek system for the ID of Enterobacteriaceae. Appl. Microbiol. 28:798. Hansen, S. L. and B. J. Stewart. A comparison of API, Minitek and CDC methods for the identification1976. of anaerobes. Abs. Ann. Mtg. Amer. SOC. Microbiol. 1976:C13. Hardy, D. L., M. E. Cox and L. H. Lindberg. 1976. Automated monitoring of anaerobic bacteria by impedance measurements. Abs . Ann. Mtg. Amer Soc . Microbiol. 1976 :C@. Hartmm, P . A. 1968. Miniaturized Microbiological Methods. Academic Press , New York . Hauser, K. J. and R. J. Zabransky. 1976. Evaluation of the API anaerobe system (MI 2OA) for the identification of anaerobes in a clinical laboratory. Abs. Ann. Mtg. Amer. SOC. Microbiol, 1976:C14. Heden, C . R. and T. Illeni, eds . 1975a. Automation in Microbiology and Immunology. John Wiley and Sons, New YorK. Heden, C R. and T. Illeni, eds . 197%. New Approaches to the Identi- fication of Microorganisms. John Wiley and Sons, New York. Holderman, L. V. and W .E.C. %ore. 1972. Anaerobe kboratory Manual. Virginia Polytechnic Institute, Blacksburg, VA.

Isenberg, H. D. and B. G. Painter. 1971. A comparison on conventional mthods, the R/B system and modified R/B system as guides to the major divisions of Enterobacteriaceae. Appl. Microbiol. 22 :1126. Isenberg, H. D., J. S. Scherber and J. D. Cosgrave. 1975. Clinical lab evaluation of the further improved enterotube and ENCISE 11. J. Clin. Microbiol. 2:139. Jayne-W illiam, D . J . 1975. Miniaturized methods for the characterization of bacterial isolates. J. Appl. Bacteriol. 38:305.

Jape-William, D. J. 1976. The application of miniaturized methods for the characterization of various organisms isolated from the anima,l gut. J. Appl. Bacteriol. 40: in press

Jones, R. W., C. Aldridge, R. A. Charles and 5. B. Frenkle, 1976. An automated microbiological detection/ID system. 1. System description. Abs. Ann. Mtg. her, SOC.Microbiol. 1976:C73. Joubert, L., J. Buissiere and C. Chirob. 1968. Chimiotype de Bacillus anthracis et enquette epidemiologique. Bull. SOC.Sci. Vet. Med. Cow. Lyon 70:149. Khan, W., G. Friedman, W. Rodriguez, G. Controni and S. Ross. 1976. Rapid detection of bacteria in blood and spinal fluids in children by electrical impedance method. Abs . Ann. Mtg . Amer . SOC. Microbiol. 1976 :C 70.

Kiehn, To E., KO Brennan and P. D. Ellner. 1974. Evaluation of the Mlnitek system for the identification of Enterobacteriaceae. Appl. Microbiol . 28 :668. Kinney, T. D. (Chairman of the committee) 1971. The mechanization, automation, and increased effectiveness of the clinical laboratory. National Institute of General Medical Sciences, National Institutes of Health, D.H.E.W. Pub. No. (NM) 72:145.

Kistemaker, P. G., H.L.C. Meuzelaar and M. A. Posthumus. 1975. Fast ID of microbiological samples by curie point pyrolysis mas spectro- metry. In New Approaches in the Identification of Microorganisms, Heden, C. R. and T. Illeni, eds. John Wiley and Sons, New York. Chapter 11.

Iampi, R. A., D. A. Mikelson, D. B. Rowley, J. J. Previte and R. E. Wells, 1974. Radiometry and microcalorimetry--technique for the rapid detect ion of f ocdborne microorganisms. Food Techno1 . 28 :52.

MartFn, W. J., R. J. Birk, P.K.W. Yu and J. A. Washington, 11. 1970. Identification of member of the family Enterobacteriaceae by the R/B system. Appl. Microbiol. 20:880.

McCarthy, L. R., M. M. Barefoot, C. P. Whittel and S. T. Carson. 1976. Evaluation of the Minitek differentiation system for the identification of anaerobic bacteria. Abs . Ann. Mtg. Amer . SOC. Microbiol. 1976:Cl5.

McKie, J. E., Jr., R. J. Borovoy, J. F. Dooley, G. R. Evanega, G. Kendoza, F. Meyer, M. Moody, D. E. Packer, J. Praglin and H. Smith. 1975. AutoBac 1 A 3-haur, automated antimicrobial susceptibility system. 11. Microbiological studies . In Autolnrtion in Microbiology and Immunology, Heden, C R. and T. Illeni, eds . John Wiley and Sons, New York. Chapter 16. Meuzelaar, H.L.C., P. G. Kistemaker and A. Tom. 1975. Differentiation of bacterial strains by f'ully automated curie point pyrolysis gas- liquid chromatography. In New Approaches to the Identification of Microorganisms, Heden C . R, and T Illeni, eds. John Wiley and Sons, New York. Chapter 10, Mischak, R. P., J. Shaw and P. Cady. 1.976. Specific detection of fecal coli by impedance measurement. Abs . Ann. Mtg . Amer . SOC. Microbiol. 1976:P3 . Mitruka, B. M. 1975. Rapid automated identification of microorganisms in clinical species by gas chromategraphy. In New Approaches to the Identification of Microorganisms, Heden, C . R. and T. Illeni, eds . John Wiley and Sons, New York. Chapter 8. 297

Moussa, R. S . 1975. fialuation of the API, the PathoTec, and the improved Enterotube systems for the identification of Enterobacteriaceae. In Mew Approaches to the Identification of Microorganisms, Heden, C. R. and T. Illeni, eds. Jahn Wiley and Sons, New York. Chapter 25 Newsom, S .W .B. 1973. Economic, easy, antibiotic sensitivity testing and biochemical typing for enterobacteria. Abs . of the Symposium on Rapid Methods and Automtion in Microbiology, Stockholm, Sweden. June 3-8, 1973. A49.

Nod, C. E., A. A. Lindberg and A. Dahlback. 1974. Evaluation of five test -kits --API, Auxotab, Enterotube, PathoTec and R/B--f or the identification of Enterobacteriaceae. Med. Microbiol, Immunol. 159: 211. O'Donnell, E. D., F. J. Kanf'man, E. D. Longo and P. D. Ellner . 1970. Evaluation of the R-B system for the identification of Enterobacter- iaceae. Am. J. Clin. Pathol. 53:145. Oxborrow, G. S., N. D. Fields and J. R. Puleo. 1976. Pyrolysis gas- liquid chromatograph studies of the genus Bacillus. Abs . Ann. Mtg. hr. Soc b Microb io1. 1976 :Q77 Painter, B. G. and H. D, Isenberg. 1973. Clinical laboratory experience with the improved Enterotube . Appl. Microbiol. 25:896. Praglin, J., A. C. Curtiss, D. K. Longhenry and J. E. McKie, Jr. 1975. AutoBac 1-system description. In Automation in Microbioloay and Immunology, Heden, C . R. and T . Illeni, eds John Wiley and Sons, New York. Chapter 15. Previte, J, J., C. T. Roskey and D, B, Rowley. 1976. Problem with radio-mtric coliform assays. Abs . Ann. Mkg. Amer . SOC. Microbiol. 1976 :Q73

Previte, J. J., D. B. Rowley and R. Wells. 1975. Improvement in a non-proprietary radiomtric medium to allow the detection of some Pseudomonas species and Alcaligenes faecalis Appl . Microbiol . 30: 339

Renner, E. D., L. A. Gatheridge and J. A. Washington, XI, 1973. Waluat ion of radiometric system for detecting bacteremia. Appl . Microbiol. 26:368. Rhoden, D . L., K. M. Tomfohrde, P. Bo Smith and A. Balms. 1973a. Auxotab--a device for identifying enteric bacteria. Appl. Microbiol. 25:284.

Rhoden, D. C., K. M. Tonfohrde, P. B. Smith and A. Balms. l9nb. Evaluation of the improved Awcotab system for identifying Ehtero- bacteriaceae. Appl. Microbiol. 26:215 Richardson, G. H. 1972. Automation in the dairy laboratory. J. Milk Food Technol. 35:279.

Robertson, E. A., G. C. Mach and J. D. Machery. 1976. Analysis of cost and accuracy of alternative strategies for Enterobacteriaceae identification. J. Clin. Microbiol. 3:421.

Rosner, R. 1973. Evaluation of the PathoTec rapid ID system and two additional experimental reagent-impregnated paper strips. Appl. Microbiol . 26 :890. Rosner, R. 1975. Comparison of isotonic and radiometric-hypertonic cultures for the recovery of organisms from cerebrospinal, pleural and synovial fluids. J. Clin. Pathol. 63:149.

Russel, W. J., J. F. Zettler, G. C. Bhnchard and E. A. Boling. 1975. Bacterial identification by mi-rocalorimetry. In New Approaches to the Identification of Microorganisms, Heden, C. R. and T. Illeni, eds . John Wiley and Sons, New York. Chapter 7. Sachs, L. E. and E. Menefee. 1972. Thermal detection of spoikige in canned foods. J. Food Sci. 37:928. Schort, J. R., W. C. Hess and G. V. Levin. 1973. Method for radiorespiro- metric detection of bacteria in pure culture and Fn blood. Appl. Microb io1. 26 :867.

Schrock, C. G., M. J. Barza, J. W. Deming, G. L. Picciolo, E. W. Chappelle and C. Weinstein. 1976. Rapid detection of bacterial growth in blood cultures by means of ATP determination. Abs . Ann. Mtg . Amer . SOC. Microbiol. 1976 :C7l . Sielsff, B. H., E. A. Johnson and J. M. btsen. 1976. Computer- assisted bacterial identification utilizing antimicrobial susceptibility profiles generated by AutoBac 1. J. Clin. Microbiol. 3:105. Shayegani, M., M. E. Hubbard, T. Hiscott and D. McGlynn. 1975. Evaluation of the R/B and Minitek systems for identification of Enterobacteriaceae. J. Clln. Microbiol. 1:5d+ Smith, P. B., D. L. Rhoden, K. M. Tomfohrde, C. R. Dunn, A. Balows and G. J. Hermann. 197la. R/B enteric differentiation system for ID of Enterobacteriaceae. Appl. Microbiol. 21:1036.

Smith, P. B., K. M. Tomfohrde, D. L. Rhoden and A. Balms. 1971b. Evaluation of the modified R/B system for ID of Enterobacteriaceae. Appl . Microb io1. 22 :928 Smith, P. B., K. M. Tomfohrde, D. L. Rhoden and A. Balms. 1972. API system: a multitube micromethod for identification of Enterobacteriaceae. Appl . Microb iol. 24 :449 299 Sonnenwirth, A. C . 1976. Developmental studies of an automated micro- biology detection ard identification system. Abs * Ann. Mtg. SOC. Microbiol. 1976:C75

Staples, B. R., E. J. Prosen and R. N. Goldberg. 1973. Fine structure in thermal growth patterns of bacteria by microcalorimetry. MBSIR 73-181 report to National Institute of General Medical Sciences, National Institutes of Health, Bethesda, MD 20014.

Stubbs, K. G. and K. Wicher. 1976. laboratory evaluation of the automated antimicrobial susceptibility system AutoBac Abs . Ann. Mtg Amer . SOC. Microbiol . 1976 :C110 Tomfohrde, K. M. and R. F. D'Amato. 1976. Biochemical characteristics of typical and atypical isolates of Salmonella. Abs . Ann. Mtg. Amer. SOC. Microbiol. 1976:~39.

Tomfohrde, K. M., D. L. Rhoden, P. B. Smith and A. Balms. 1973. Evaluation of the redesigned Enterotube. Appl. Microbiol. 25:301.

Ur, A. and D.F.J. Brown. 1975. Monitoring of bacterial activity by impedance measurements. In New Approaches to the Identification of Microorganisms, Heden, C. R. and T. Illeni, eds. John Wiley and Sons, N& York: Chapter 5. Washington, J. A., 11, P .K.W . Yu and W . J. Martin. 1971. Evaluation of accuracy of multitest micromethod system for identification of Enterobacteriaceae. Appl . Microbiol. 22 :267. Wilklns, T. D. and C . B . Walker. 1975. Development of micromethod for identification of anaerobic bacteria. Appl Microb iol. 30:825.

WiUcins, T. D., C. B. Walker and W.E.C. Moore. 1975. Micromethod for identification of anaerobic bacteria: design and operation of apparatus Awl. Microbiol. 30:831.

Wust, J. and F. H. Kkyser. 1974. Ekaluaticm of the redesigned Enterotube and its interpretation system. Patho Microbiol. 40:316.

Wyatt, P. J. 1973. Differential light scattering techniques for micro- orbiology. In Methods In Microbiology, Vol. 8. Academic Press, New York . Bruce Langlois: Do we have any questions?

Bruce Tompkin, Swift and Company: Where do you place the relative importance of the kits that you have listed there as opposed to serolo- gical testing for stating whether or not a sample is Salmonella positive or negative?

I cannot tell you that definitely. All that I can D-: tell D-: you is, first of all, the kits have been tested and many of them have an accuracy of about 95 percent when compared to conventional methods. Each one of these probably will cost $2 per identification, where the conventimal method is about $8 per complete identification. So there is a relative difference in terms of mney. I can't tell you which is better for Salmnella. Maybe you how.

Bruce Tompkin: Well, which kit you use is immaterial in terms of what I'm asking, since they all are based upon biochemical reactions?

D ,Y .C Fung: Yes

Bruce Tompkin: Now, a serological reaction is more specific, so what would I use when I fall apart on serological tests. I don't know what it is, In other words, that would be a backup system. That would not be the primary identification system in terms of finding whether or not it is Salmonella or not.

D.Y.C. kula: Well, a lot of people have done a lot of comparative work and study on that. As far as I can tell, the API system is the most complete in terms of defending whether the organism is Salmonella or not. Now, there has been big competition in the past between API and Minitek in terms of the completeness of the biochemical test. They all say that you have to back up with physiological tests for Salmonella and Shigella.

Bruce Langlois: Thank you, Dan. otrr next speaker, the final speaker for this session, is a native of Minnesota. He got his B.A. and MS at the University of Minnesota and his Ph D. at Illinois. He served a tour of duty at North Carolina State at Raleigh for about four years, before he went back to Minnesota, where he is now Professor of Food Microbiology. Dr. Bus- will be covering "Bacterial Injury and Rec avery ." F. F. Busta: I'd like to express my appreciation to be invited to speak to your association today. It's an honor to talk to the Reciprocal Meat Conference, especially when I'm going to talk about one of my best and favorite loves: "Bacterial Injury and Recovery."