<<

214 IEEE JOURNAL OF SOLID-STATE CIRCUITS, VOL. 50, NO. 1, JANUARY 2015 A Fully-Implantable Cochlear Implant SoC With Piezoelectric Middle-Ear Sensor and Arbitrary Waveform Neural Stimulation Marcus Yip, Rui Jin, Student Member, IEEE, Hideko Heidi Nakajima, Konstantina M. Stankovic, and Anantha P. Chandrakasan, Fellow, IEEE

Abstract—A system-on-chip for an invisible, fully-implantable cochlear implant is presented. Implantable acoustic sensing is achieved by interfacing the SoC to a piezoelectric sensor that de- tects the sound-induced motion of the middle ear. Measurements from human cadaveric ears demonstrate that the sensor can detect sounds between 40 and 90 dB SPL over the speech bandwidth. Ahighly-reconfigurable digital sound processor enables system power scalability by reconfiguring the number of channels, and provides programmable features to enable a patient-specific fit. A mixed-signal arbitrary waveform neural stimulator enables energy-optimal stimulation pulses to be delivered to the auditory nerve. The energy-optimal waveform is validated with in-vivo measurements from four human subjects which show a 15% to 35% energy saving over the conventional rectangular waveform. Prototyped in a 0.18 µm high-voltage CMOS , the SoC in 8-channel mode consumes 572 µW of power including stimulation. The SoC integrates implantable acoustic sensing, sound processing, and neural stimulation on one chip to minimize the implant size, and proof-of-concept is demonstrated with measurements from a human cadaver ear. Fig. 1. Block diagram of a conventional cochlear implant. Index Terms—Arbitrary waveform, cochlear implant, en- ergy-efficient, hearing loss, implantable, low-voltage, microphone, middle ear, piezoelectric, reconfigurable, SoC, stimulation, unitasshowninFig.1.Theexternal unit comprises a micro- ultra-low-power. phone to pick up sound, a sound processor to digitize and com- press the sound into coded signals, and a transmitter to send data wirelessly to the internal unit via a coil. The external unit I. INTRODUCTION also houses the battery which supplies power wirelessly to the implanted unit via the same coil that is used for data. Without S OF 2010, over 30 million people in the United States a continuous transfer of power (e.g., if the external unit is re- A suffer from sensorineural hearing loss [1] which arises moved), the implanted unit is unpowered and does not func- from disease in the inner ear or auditory nerve. For mild cases of tion. The implanted unit comprises a receiver and stimulator hearing loss, a hearing aid may provide adequate compensation. unit embedded in the skull, and an electrode array implanted However, for profound cases (i.e., greater than 90 dB of loss), a in the cochlea. Pulses of electrical current are modulated by the cochlear implant (CI) is necessary to restore hearing. received codes and delivered to the electrode array, triggering CIs use to directly stimulate the auditory nerve action potentials in the auditory nerve which are interpreted by fibers, thus bypassing the damaged hair cells in the cochlea. the brain as sound. Today’s state-of-the-art CIs consist of an external and internal Although today’s CIs are very successful in restoring hearing for many of the profoundly deaf, the external component results Manuscript received April 23, 2014; revised July 04, 2014; accepted August in several limitations. The device is cumbersome to wear, and 20, 2014. Date of publication September 25, 2014; date of current version De- it cannot be worn in the shower or while participating in water cember 24, 2014. This paper was approved by Guest Editor Yogesh Ramadass. This work was supported by NSERC and the Bertarelli Foundation. sports. It also raises concerns with aesthetics and social stigma. M.Yip,R.Jin,andA.P.Chandrakasan are with the Microsystems Tech- These reasons motivate the development of a fully-implantable nology Laboratories, Massachusetts Institute of Technology, Cambridge, MA cochlear implant (FICI) that is internalized and invisible. 02139 USA (e-mail: [email protected]). H. H. Nakajima and K. M. Stankovic are with Harvard Medical School, A FICI has three main requirements that are distinct from a Boston, MA 02115 USA, and also with the Eaton Peabody Laboratory, conventional CI. First, a FICI that is completely untethered (i.e., Massachusetts Eye and Ear Infimary, Boston, MA 02114 USA. without a coil that continuously provides power) requires an im- Color versions of one or more of the figures in this paper are available online at http://ieeexplore.ieee.org. planted battery that is rechargeable because of the volume and Digital Object Identifier 10.1109/JSSC.2014.2355822 power consumption constraints of the FICI, as well as the need

0018-9200 © 2014 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information. YIP et al.: A FULLY-IMPLANTABLE COCHLEAR IMPLANT SOC WITH PIEZOELECTRIC MIDDLE-EAR SENSOR 215 to avoid future surgeries for battery replacement. The number charge-per-day usage model. For a 5 gram ultra-capacitor with of charges per day must also be limited to once or twice a day an energy density of 5 W hr/kg, 12 hours of continuous usage to minimize user impact. As a result, a FICI requires ultra- would require the FICI to consume just 1 mW of power as- low-power sound processing and energy-efficient neural stim- suming 50% conversion efficiency in the power management ulation. Secondly, recent state-of-the-art ICs are typically de- unit. The majority of the 1 mW power budget will be consumed signed for external microphone-based CIs and do not require by the electrode impedance and electrode drivers [6] during the the neural stimulator to be on the same chip [2], [3]. In contrast, process of electrically stimulating the auditory nerve. Assuming thesizeofaFICIcouldbenefit from monolithic integration of typical stimulation power of approximately 750 W [2], [3], the signal processing and stimulation circuits. Thirdly, an im- this leaves approximately 250 W for the implantable sensor plantable acoustic sensor with adequate sensitivity and band- front-end circuits and sound processor. width is required to replace the external microphone. 2) Implantable Acoustic Sensor: A key enabler for a FICI This paper presents a system-on-chip (SoC) for a FICI system is an implantable acoustic sensor that is able to sense that addresses the above issues. First, low-power implantable external sound pressure waves from within the body. Recently, acoustic sensing is achieved by interfacing the SoC to a piezo- totally invisible middle ear implants (MEIs) have been devel- electric sensor that is mounted at the umbo of the malleus within oped to treat conductive hearing loss1 [7], [8]. MEIs typically the middle ear, and this is demonstrated with measurements use an implantable sensor to detect the mechanical motion of from human cadaveric temporal bones. Second, a highly-re- the ossicles, using the ear as a natural microphone. The sensor configurable sound processor enables system power scalability readout can be amplified and fed to an output transducer which by scaling the number of spectral channels. Third, simulations drives the stapes (the stirrup-shaped bone of the ossicles) with with a computational auditory nerve fiber model from [4] are increased vibration to compensate for hearing loss [7]. In this used to determine energy-optimal biphasic stimulation wave- work, we apply the sensors found in MEIs to a FICI system; forms [5], which are delivered to the nerves in the cochlea with instead of driving the output transducer of a MEI, we use the an arbitrary waveform neural stimulator. The energy savings sensor readout as an input to the sound processor of a FICI from alternate stimulation waveforms are validated in-vivo in which stimulates the cochlea directly. four human CI subjects with loudness perception tasks. The Prior work on implantable sensors has looked at MEMS resulting stimulation power savings over the conventional accelerometers, but they have limited sensitivity and require rectangular pulse transfer directly to overall system power milli-Watt power consumption [1]. Alternative approaches savings because stimulation power typically dominates [5]. include magnetic sensors [9] and subcutaneous microphones The SoC integrates an implantable acoustic sensor front-end, [10], but they suffer from incompatibility with magnetic reso- sound processor, and neural stimulator on one chip to minimize nance imaging (MRI) and unwanted body noise respectively. the implant size and demonstrate proof-of-concept for a FICI. This work leverages a piezoelectric sensor [8] because of its It should be noted that the wireless charger and power manage- small size, low-power operation, and superior sensitivity. ment unit is not included in the scope of this work. Considering that the loudness of normal conversational Section II presents the requirements and architecture of the speech is around 60 to 70 dB SPL2 [1], [11] and that the FICI. Section III describes the implantable acoustic sensor and dynamic range of speech can be up to 50 dB [12], the sensor the front-end circuit implementation, and Section IV discusses should be able to detect sounds from 40 dB SPL (quiet library) the reconfigurable sound processor design. Section V discusses to 90 dB SPL (busy roadside). Furthermore, the sensor should energy-efficient neural stimulation waveforms, and presents de- also support a bandwidth that spans the frequencies specificto tails of the stimulator architecture. Section VI presents proto- cochlear implant hearing from a few hundred Hz up to 5 kHz type measurement results, Section VII discusses potential use [13], [14]. cases and future work, and Section VIII concludes the paper. 3) Number of Spectral Channels: The choice of the number of spectral channels should strike a balance between the speech II. SYSTEM REQUIREMENTS AND ARCHITECTURE DESCRIPTION recognition performance of CI users and the hardware com- This section highlights the main high-level requirements and plexity and power consumption in a FICI. In [15], Shannon et specifications of the system, and describes the system architec- al. used acoustic simulations with normal hearing listeners in ture of the FICI. quiet and showed that as few as 3 to 4 channels of spectral in- formation can result in good speech recognition performance. A A. System Requirements similar study with CI users in quiet was presented in [16] where 1) Power Consumption: Ultra-low power consumption can the authors found that the average performance improved as the ease energy storage requirements which can translate to pos- number of stimulation electrodes was increased from 1 to 4, but sible size reduction of the implant, or increased time between no differences were observed between 7-, 10-, or 20-electrode recharge. Lithium-ion batteries possess high energy density but 1Conductive hearing loss occurs when there is damage to the middle ear suffer from limited (on the order of 1000’s) recharge cycles which blocks the conduction of sound waves toward the inner ear (cochlea). In which would eventually necessitate replacement via surgery. contrast, sensorineural hearing loss occurs when there is damage to the cochlea. MEIs treat conductive hearing loss only and require the cochlea to be intact and Ultra-capacitors, on the other hand, can be cycled on the order of functional, whereas CIs are used to treat sensorineural hearing loss. 10 times and may be more suitable for an implantable system 2Sound pressure level (SPL) in units of dB SPL is a logarithmic measure of provided that the limited energy density still permits a one- sound pressure with respect to a reference of Pa . 216 IEEE JOURNAL OF SOLID-STATE CIRCUITS, VOL. 50, NO. 1, JANUARY 2015

Fig. 2. Block diagram of the fully-implantable cochlear implant SoC. processors. For CI users in noise, the best CI listeners improved The ADC output is processed by a 0.6 V reconfigurable sound their performance only up to 7 electrodes, while CI users with processor that implements the well-known Continuous Inter- low levels of speech recognition did not benefitfrommorethan leaved Sampling (CIS) sound processing strategy [21] in which 4 electrodes [17]. Although user performance depends also on each electrode is stimulated synchronously in an interleaved the speech processing strategy, the effective number of spec- manner with an amplitude determined by the output of each tral channels possible with electrically-evoked hearing is usu- channel of the processor. The interleaved stimulation drastically ally limited by the current spreading between electrodes [18]. In reduces interaction between electrodes which allows for a high this work, the target number of channels is 8 which provides ad- rate of stimulation that is important for preserving temporal in- equate spectral resolution while minimizing the hardware com- formation in speech [15]. plexity for a proof-of-concept system. The processor in this work is designed to be extremely re- configurable in order to enable system power scalability as well B. Architecture Description as patient-specific fitting capability. First, the number of chan- nels can be configured to 8, 6, or 4 to enable a power-perfor- A block diagram of the proposed SoC is shown in Fig. 2 mance tradeoff. The filter bank has reconfigurable coefficients [19]. The system is separated into three main subsystems: 1) a to adjust the filter bandwidths for the three modes of operation, piezoelectric sensor front-end (PZFE), 2) a low-voltage recon- and multi-rate signal processing is leveraged to reduce power figurable sound processor, and 3) an energy-efficient arbitrary and area. The bandwidth of the processor covers 300 Hz to waveform neural stimulator and high-voltage electrode switch 5.5 kHz, and the filter cut-off frequencies of the logarithmi- matrix. cally-spaced channels are based on [13], [14] to emulate the The PZFE conditions the signal from the sensor which is a tonotopic structure of hearing. Furthermore, processor settings measure of the sound-induced motion of the umbo. The PZFE like global channel gain, type of rectification, and amount of operates from a 1.5 V analog supply and comprises three stages: compression are all programmable. The processor also has the achargeamplifier (CA) to interface to the piezoelectric sensor, capability to adjust the volume level for each channel individ- a programmable-gain amplifier (PGA), and a single-ended to ually to provide additional patient fitting capability. The pro- differential ADC driver. A mid-rail reference voltage bi- cessor outputs 6 bit data at an analysis rate of 1 kHz, which ases the sensor and sets the DC operating point of the PZFE. represents the logarithmically compressed energy in each fre- The ADC driver stage also provides DC level shifting from quency band. This value is used to modulate a train of electrical mV down to mV which is the input current pulses (1,000 pulses/sec per electrode) that is delivered common-mode of the ADC. The signal is then digitized by a to the corresponding electrode. fully-differential low-power 16 kS/s 9 bit successive approx- The interleaved operation of the CIS strategy conveniently imation register (SAR) ADC operating from a digital supply allows for a single current source to be interleaved between all voltage of 0.6 V [20]. electrodes. This is accomplished by using a high-voltage elec- YIP et al.: A FULLY-IMPLANTABLE COCHLEAR IMPLANT SOC WITH PIEZOELECTRIC MIDDLE-EAR SENSOR 217

Fig. 3. (a) Block diagram and (b) photograph of the measurement setup and discrete prototype used to characterize the piezoelectric sensor mounted on the middle ear of a human cadaveric temporal bone. trode switch matrix to select the active electrode and control the is positioned by a micro-manipulator external to the bone. The direction of current flow. The SoC is designed to be used with sensor is clamped at one end like a cantilever, while the other monopolar electrode arrays, where a common return electrode end is placed at the umbo of the malleus. As the umbo vibrates is used to provide a return path for all electrodes. Furthermore, a back and forth, it exerts a force that bends the sensor which in 0.6 V digital controller provides the control signals for the cur- turn generates a charge across its terminals which is converted rent DAC and switch matrix which allows the waveform of the to an output voltage by a charge amplifier. stimulationpulsestobeprogrammedtoanyarbitraryshape. Fig. 4(a) and (b) show that the measured umbo velocity and The stimulator has high voltage compliance in order to sensor readout are very linear with the sound pressure level in accommodate up to 1 mA of stimulation current through the the ear canal. Fig. 4(c) shows the measured output spectrum of electrode-tissue interface which may have a few kilo-Ohms thechargeamplifier from 200 Hz to 10 kHz and it can be seen of impedance. The stimulation current is drawn from a high that the sensor is able to detect sounds over a 50 dB dynamic voltage supply ( 5 V to 10 V), and the switch matrix range from 40 to 90 dB SPL. is driven with high-voltage logic operating from 7V to 12 V. Level shifters are used to interface between the 0.6 V B. Piezoelectric Sensor Front-End and domains. Lastly, the current DAC circuits operate The details of the 3-stage PZFE of the SoC are shown at the from a supply voltage of 3.3 V. top of Fig. 2, where the piezoelectric sensor is modeled as a Thevenin voltage source and series capacitance .Since III. IMPLANTABLE PIEZOELECTRIC ACOUSTIC SENSOR the charge amplifier of the first stage dominates the noise per- This section first describes the characterization of the piezo- formance of the front-end, its signal transfer function and noise electric sensor on the middle ear of a human cadaveric tem- analysis is provided next. poral bone using a discrete prototype, followed by the design 1) Charge Amplifier Transfer Function: Fig. 5(a) shows the and analysis of the sensor front-end of the SoC. charge amplifier (stage 1) of the PZFE with noise sources, and Fig. 5(b) shows the equivalent block diagram, where , , A. Sensor Characterization and are the noise from , , and the op-amp respectively, In order to investigate the performance of the middle-ear and is the open-loop transfer function of the op-amp. In mounted piezoelectric sensor, human cadaveric temporal bones order to determine the transfer function from the piezoelectric were provided by the Eaton-Peabody Laboratory at the Mass- sensor voltage tothechargeamplifier output , the charge achusetts Eye & Ear Infirmary (Boston, MA), and PZT-5A amplifier input voltage can be referred back to through piezoceramic material (Piezo Systems Inc., Woburn, MA) was the following transfer function, , used. A block diagram of the prototype and measurement setup where . For the frequencies of interest where the is shown in Fig. 3(a). Swept-sine measurements were made loop gain is large, the closed-loop transfer function is given by using an audio amplifier and speaker connected to a probe tube where . that funnels sound into the ear canal of the temporal bone. Ear Therefore, the transfer function of stage 1 from to is given canal pressure ( ) and umbo velocity ( ) were mon- by itored with a probe microphone and laser Doppler vibrometer (Polytec) respectively. A discrete prototype was used to record (1) the output voltage of the sensor ( ). All three outputs were recorded by LabVIEW and the transfer characteristics from which gives the desired band-pass characteristic necessary to to to were calculated. pass only the frequencies relevant to speech for a CI (300 Hz Fig. 3(b) shows the photograph of the measurement setup. to 6 kHz). The high-pass and low-pass corners are set by The temporal bone is held by a specimen holder and the sensor and respectively, and the mid-band gain is simply 218 IEEE JOURNAL OF SOLID-STATE CIRCUITS, VOL. 50, NO. 1, JANUARY 2015

Fig. 4. (a) Umbo velocity and (b) charge amplifier output voltage versus ear canal sound pressure at 0.5, 1, 2, and 4.7 kHz. (c) Spectrum of the charge amplifier output for sound pressure levels from 40 to 90 dB SPL.

Fig. 5. (a) Equivalent circuit for the charge amplifier including noise sources, and (b) the corresponding block diagram. for . Note that the negative polarity of the charge am- Following similar analysis, the noise spectral density of re- plifier has been ignored here for simplicity. To accommodate a ferred to is . Finally, range of typical sensor sizes ( 0.2 to 3 nF), is a 3 bit the noise transfer function of the op-amp referred to is switched-capacitor tunable from 6 pF to 66 pF which provides programmable gain in 3 dB steps, and is set to 88.4 M such that (300 Hz) for all values of . Finally, is a 4 bit switched-resistor with logarithmically-spaced values from 1 k to 100 k to ensure (6 kHz) for typical values of . (4) 2) Charge Amplifier Noise Analysis: The noise transfer functions referred to can be determined by calculating the for ,where since noise transfer function to the output andthendividingby . Therefore, the noise of the op-amp is magnified by .For , the noise transfer function referred to is which is more dominant at lower frequencies. which evaluates to From the above analysis, it can be seen that the noise spec- tral density of , , and the op-amp thermal noise all have a characteristic, and that the resistor noise is reduced for larger values of . However, the op-amp noise is independent (2) of because it depends on which is generally chosen to be fixed. For typical values of ,thenoisefrom for . As a result, the noise spectral density of referred is negligible because of its large value, and the relative contribu- to is tions of noise from and the op-amp vary depending on . The total integrated noise from simulation for nF and 3nFare2.5 V and 1.7 V respectively, which are lower than the minimum expected signal of approximately 3 V at (3) 40 dB SPL as determined by the discrete prototype. YIP et al.: A FULLY-IMPLANTABLE COCHLEAR IMPLANT SOC WITH PIEZOELECTRIC MIDDLE-EAR SENSOR 219

Fig. 6. Block diagram of the 0.6 V digital reconfigurable multi-rate CIS sound processor.

IV. RECONFIGURABLE SOUND PROCESSING and half-wave rectification are possible. Half-wave rectifi- The sound processor in this work implements the CIS sound cation has the potential to reduce stimulation power since processing strategy [21] because it is the most ubiquitous the average envelope is smaller, but at the cost of poten- strategy among CI manufacturers. The two main objectives for tially lower sound quality [14]. the design of the sound processor are 1) ultra-low-power op- • Channel gain and compression factor: Following enve- eration and 2) highly-reconfigurable features to enable system lope detection, global gain and dynamic range compression power scalability and patient-specific fitting capability. The first are applied. The global gain can be set in octaves from goal is accomplished by leveraging ultra-low-voltage digital to with 3 bits, and the signal is compressed according to processing at 0.6 V to maximize energy-efficiency. The second ,where is the compres- goal is addressed with a flexible digital architecture featuring a sion factor. Logarithmic compression is needed because multi-rate reconfigurable filter bank and highly-programmable of the well-known loudness growth function of electrical processor parameters. hearing which shows a linear relationship between sound The block diagram of the reconfigurable CIS sound processor intensity in dB SPL and electrical stimulation intensity in is shown in Fig. 6. The processor spectrally decomposes the Amperes. Evidence shows that different amounts of com- signal with a logarithmically-spaced multi-rate filter bank to pression can be beneficial [12], and therefore three settings emulate natural hearing. The envelope of each filter output is are available: (default), 128, and 16. For low extracted, downsampled, and logarithmically-compressed to fit compression ( ), the stimulation power is roughly the patient’s electric hearing dynamic range, and each channel linear with gain, while for high compression ( ), has patient-specific volume settings. By clock-gating the unused the stimulation power scales linearly with the logarithm of channels, both the processor power and stimulator power (which the gain. dominates the SoC power) scale linearly with the number of • Volume settings: Each channel has individual threshold channels. (THR) and most-comfortable-level (MCL) settings that can be used to fit the dynamic range of the stimulus current A. Programmable Features for each electrode based on the user. Each channel has 3 bit Aside from being able to select between 8-, 6-, or 4-channel programmability in both THR and MCL, and stimulation modes, other programmable processor parameters can affect the power increases with higher THR and MCL settings. user performance and power consumption of the CI and they are B. Reconfigurable Filter Bank Architecture highlighted next. • Rectification: The type of rectificationusedintheen- A logarithmically-spaced filter bank requires higher fre- velope detector can affect speech recognition scores and quency channels to be wider in bandwidth, while low frequency sound quality [14], and therefore both full-wave (default) channels need to be narrow and more selective (i.e., higher 220 IEEE JOURNAL OF SOLID-STATE CIRCUITS, VOL. 50, NO. 1, JANUARY 2015

Fig. 7. Structure of the reconfigurable FIR filter (Type 3) used for channels A, C, E, G that can be reconfigured into 3 modes: 14-, 16-, and 20-tap used in 4-, 6-, or 8-channel modes.

filter order). In this work, high order filters are avoided by effective filter order is much higher because of multi-rate signal using multi-rate signal processing to achieve the narrow low processing. The effective frequency responses of the filter bank frequency filters in a power- and area-efficient manner. As (at kHz) reconfigured in 4-, 6-, and 8-channel modes shown in Fig. 6, the ADC data is decimated in 3 stages using are shown in Fig. 8. efficient 19-tap half-band FIR filters, resulting in data rates of 2, 4, 8, and 16 kHz which are used for channels A/B, C/D, E/F, V. E NERGY-EFFICIENT ARBITRARY WAVE FOR M and G/H respectively. NEURAL STIMULATOR The filter bank is implemented with FIR filters for their linear In this section, we consider the design of the neural stim- phase which can have a positive effect on sound quality and ulator which delivers electrical current to the nerve fibers of speech intelligibility. When the sound processor is configured the cochlea. The typical power consumption of the stimulator in 8-channel mode, all channels are active. In 6- and 4-channel can often be a few milli-Watts [3], [22] which can represent modes, the subsets of channels (D, H) and channels (B, D, F, greater than 90% of the total SoC power given that the PZFE and H) are clock-gated respectively. The cut-off frequencies of the sound processor have micro-Watt power consumption. There- individual filters vary with the channel mode and they are based fore, any power savings in the stimulator translate directly to on [13], [14]. In order to achieve reconfigurability in the filter responses, overall system power savings. the filter bank leverages three types of FIR filters with different A. Computational Nerve Fiber Simulations levels of reconfigurability. Fig. 7 shows the most reconfigurable filter (Type 3) which can be programmed to three different filter Most neural stimulators today deliver charge-balanced lengths: 14, 16, or 20 taps used in the 4-, 6-, or 8-channel mode biphasic rectangular current pulses as shown by the green curve (using control signals , ,and )toad- in Fig. 9(a), where the first (cathodic) phase excites the nerve just the frequency selectivity. Since the filters are symmetric, fiber and the second (anodic) phase provides charge balancing. the filter is folded to reduce the number of multiplications by The rectangular waveform has been widely adopted for its half. The coefficients are quantized to 8 bit precision which is simplicity and ease of generation with a simple current source. the minimum possible without significantly affecting the de- However, studies have shown that alternate waveforms have sired frequency response, and word lengths are optimized for the potential to excite nerves with reduced energy [5], [23]. the given coefficients. Based on [5], a heuristic search was applied to a computational The Type 3 FIR filter is used for channels A, C, E, and G model of an auditory nerve [4] to seek out an energy-optimal which are active for all three modes. The Type 2 FIR filter used waveform with CI-specific parameters.3 The waveform was for channels B and F has two levels of reconfigurability and can constrained to be charge-balanced with 10 time steps/phase, but be programmed to have either 16 or 20 taps used in the 6- or unlike [5], no constraint was placed on the shape of either phase 8-channel modes only. Finally, the Type 1 FIR filter used for of the pulse to allow both phases to be co-optimized. Fig. 9(a) channels D and H is a fixed 20-tap filter because it is used in the (blue curve) shows the energy-optimal waveform after 10,000 8-channel mode only. search iterations at a phase width of 25 s, and it is 28% more Finally, taking channel A for example, it can be shown that energy-efficient than the conventional rectangular waveform its effective filter response at kHz is given by at threshold. The energy-optimal waveform in this work is somewhat different from the truncated Gaussian shapes in [5] (5) which may be attributed to a different nerve fiber model as well where and represent the half-band and channel as the lack of constraint on the shape of the anodic phase. AFIRfilters at their respective downsampled data rates. Al- 3Typical CIs use biphasic waveforms with a phase width between 25 and though is only a 20-tap filter (in 8-channel mode), the 50 s. YIP et al.: A FULLY-IMPLANTABLE COCHLEAR IMPLANT SOC WITH PIEZOELECTRIC MIDDLE-EAR SENSOR 221

Fig. 8. Effective frequency response of the multi-rate filter bank at 16 kHz reconfigured in (a) 4-channel, (b) 6-channel, and (c) 8-channel modes.

Fig. 9. (a) Energy-optimal stimulation waveform at 25 s from the heuristic search using the computational nerve fiber model. Rectangular and exponen- tial waveforms are included for comparison. (b) Perceived loudness versus energy delivered per phase from four human subjects.

B. Validation With In-Vivo Measurements because it mimics the optimal waveform closely in the cathodic phase, but the anodic phase was constrained to be symmetric In order to validate the modeling results and determine to the cathodic phase due to test limitations. The waveforms the impact of alternate waveforms on auditory perception in were alternated in pseudo-random order, and the amplitude was humans, the rectangular waveform was compared against the swept from threshold to just beyond the maximum comfortable exponential waveform shown in Fig. 9(a) by conducting a level (CL) in 50 A steps on a middle electrode. The subjects psychophysical loudness perception test on four subjects with were asked to rate the loudness on a scale of 0 to 25, with 8 Advanced Bionics CIs.4 The exponential waveform was used and 22 being the minimum and maximum CL respectively. 4Tests were conducted under the Massachusetts Eye & Ear Infirmary IRB Fig. 9(b) shows the average perceived loudness of both wave- protocol #94-01-003. forms versus the energy delivered (per Ohm of the electrode 222 IEEE JOURNAL OF SOLID-STATE CIRCUITS, VOL. 50, NO. 1, JANUARY 2015

Fig. 10. (a) Schematic of the high-voltage electrode switch matrix during the cathodic phase of electrode 2. (b) Schematic of the fast-settling 6 bit current steering DAC.

Fig. 11. Timing diagram for the digital control of the electrode switch matrix. resistance) for each of the 4 subjects. Overall, to achieve the 1) High-Voltage Electrode Switch Matrix: The CIS sound same loudness within the comfortable range between 8 and processing strategy permits a single current DAC to be inter- 22, the alternate exponential waveform requires approximately leaved among all electrodes using the high-voltage switch ma- 15% to 35% less energy than the rectangular waveform. trix shown in Fig. 10(a). The terminals of the intracochlear elec- trodes are designated by (for 1to8),and is the C. Arbitrary Waveform Stimulator Architecture common return electrode of a monopolar electrode array. A Basedonthenervefiber simulations and in-vivo measure- high-frequency electrode model between and ment validation of alternate waveforms, this section describes models the impedance of the electrode-tissue-electrode inter- the architecture of the arbitrary waveform stimulator in the SoC. face. Electrode is active when is asserted, and the switches YIP et al.: A FULLY-IMPLANTABLE COCHLEAR IMPLANT SOC WITH PIEZOELECTRIC MIDDLE-EAR SENSOR 223

Fig. 12. Ultra-low-voltage digital control of the stimulator.

, , ,and are used to control the direction of current and large voltage compliance. It provides 6 bits of reso- flow during the cathodic and anodic phases for each electrode. lution at a full-scale of 1 mA which is typically sufficient During each phase, the value of the current is determined by a 6 for CIs. ensures that the DAC current is simply bit current DAC ( [5:0])whichisdrivenbyadigitalwave- . Detailed analysis of the VCR in [24] form controller to realize any arbitrary waveform. Since the shows that is linear with which drives the gate of switch matrix works like a H-bridge, current always flows from themaintriodedevice . to are auxiliary devices to ground. In between the cathodic and anodic phases, an controlled by [1:3] and are used to linearize . All devices optional switch ( ) can be used to insert an inter-phase gap. are 3.3 V transistors except for which is a high-voltage Upon completion of each pulse, shorts the electrode to re- device that connects to the high-voltage switch matrix. move any residual charge. Although not pictured, a DC blocking In this work, current-steering is used to achieve the settling capacitor (220 nF) is placed in series with the electrodes to en- time required to generate arbitrary waveforms at a 25 sphase sure that no DC current flows into the tissue for safety reasons. duration with 10 steps/phase. The input code D[5:0] from the Fig. 11 shows the timing diagram of the switch matrix con- digital waveform controller steers binary-weighted currents to trol over a complete stimulation cycle. The start of each cycle which are then mirrored to the resistor string on the output begins on the rising edge of (1 kHz) which generates a branch. The generated control voltages [3:0] are linear with pulse and also asserts whichisusedtoen- D[5:0] as desired. Finally, the current DAC is power-gated from able the 3.3 V supply from which the current DAC circuits op- 3.3 V after all electrodes have been stimulated so that its power erate. Stimulation is enabled when is asserted, and the scales linearly with the number of channels. electrode selection signals are generated by a state machine 3) Low-Voltage Digital Waveform Controller: In order to that is clocked by . Note that rises half a cycle before minimize the power overhead of arbitrary waveform genera- in order to provide adequate time for the current DAC tion, the digital waveform controller operates at 0.6 V, and level circuits to power up. Stimulation is complete on the positive shifters are used to interface to the high-voltage logic which edge of which de-asserts and . drives the switch matrix. Fig. 12(a) shows the electrode selec- 2) Current Steering DAC: The current DAC shown in tion state machine that generates to and other signals Fig. 10(b) is based on the voltage-controlled resistor (VCR) that govern the stimulation cycle. Control signals topology [24] which is chosen for its high output impedance and are used to reconfigure the state machine between 224 IEEE JOURNAL OF SOLID-STATE CIRCUITS, VOL. 50, NO. 1, JANUARY 2015

TABLE I MEASURED PERFORMANCE SUMMARY FOR EACH SUBSYSTEM OF THE FICI SOC

waveform weight (w00 to w15) is multiplied with dstimX to generate D[5:0] which drives the current DAC. The shape of the cathodic and anodic phases are determined by w00 to w07 and w08 to w15 respectively, for a maximum of 8 steps/phase.

VI. MEASUREMENT RESULTS A prototype SoC was fabricated in a 0.18 m high-voltage CMOS process, and the die micrograph is shown in Fig. 13. The chip including pads measures 3.6 mm 3.6 mm, while the activeareais3.36mm . This section presents the measured re- sults, and a summary of the performance is provided in Table I.

A. Piezoelectric Sensor Front-End

Fig. 13. Die micrograph of the prototype SoC. Fig. 14(a) and (b) show the measured gain response of the charge amplifier for 3.2 nF and 0.56 nF which span the expected values for reasonable sizes of the sensor. Fig. 14(c) channel modes. The state machine is triggered on ,andthe shows the combined response of the charge amplifier and PGA signals are shifted out serially with which is a delayed with 0.56 nF and 12 pF for various PGA settings. version of . This ensures that transitions just after the neg- Simulation results are shown with dotted lines and show good ative edges of and to avoid glitching in the switches. The agreement with measured results. Furthermore, the integrated control signals for the switch matrix are generated by gating noise over the sound processor bandwidth is 2.81 V and and with non-overlapping clocks , ,and as 1.93 V for 0.56 nF and 3.2 nF, respectively, which showninFig.12(b). is less than the minimum expected signal at 40 dB SPL. Finally, Fig. 12(c) shows the digital arbitrary waveform interface that controls the shape of the pulses delivered to the B. Reconfigurable Sound Processor electrodes. The channel select block selects the appropriate To demonstrate the reconfigurability in the number of chan- sound processor output (dstimX for X = A to H) based on nels of the processor, a logarithmic chirp signal was applied at . While each is asserted, a step counter (cnt) running at the input of the ADC. Fig. 15(a) shows the measured spectro- (a high-frequency waveform clock) keeps track of the gram at the output of the ADC, and Fig. 15(b), (c), and (d) show time step within the pulse. The value of cnt determines which the measured spectrogram of the processor configuredin4-, YIP et al.: A FULLY-IMPLANTABLE COCHLEAR IMPLANT SOC WITH PIEZOELECTRIC MIDDLE-EAR SENSOR 225

Fig. 14. Measured gain response of the charge amplifier (stage 1) of the PZFE with (a) nF and (b) nF. Panel (c) shows the combined response of the charge amplifier and PGA (stage 1 and 2). Simulation results are shown with dotted lines.

6-, and 8-channel modes. Since the processor features a loga- rithmically-spaced filter bank, its spectrogram looks linear as expected. A Matlab simulation of the 8-channel processor is shown in Fig. 15(e), showing good agreement with the mea- sured results.

C. Energy-Efficient Arbitrary Waveform Stimulator The measured INL and DNL are 0.21/+1.25 LSB and 0.14/+0.16 LSB, respectively which is adequate for neural stimulation applications. Fig. 16 shows the measured current and voltage from a model electrode ( k , nF) for (a) a rectangular waveform and (b) the energy-optimal wave- form for phase widths of 25 s and 50 s with 8 steps/phase. Even though the energy-optimal waveform requires a higher peak current than the rectangular waveform, it consumes less energy and generates a smaller electrode voltage. Fig. 16(c) shows a measurement of the interleaved current pulse trains at 1,000 pulses/sec per electrode through all 8 elec- trodes. In this measurement, the pulses are programmed to be rectangular with a phase width of 31.25 s such that the current DAC is active for 50% of the 1 ms period, and power-gated for the remainder of the period.

D. Power Consumption 1) Stimulator Power: Fig. 17 summarizes the measured stimulator power in 8-, 6-, and 4-channel modes while pro- cessing a clip of speech with V. The measurement also includes the power consumption of the low-voltage dig- ital waveform controller, level shifters, current DAC, and high-voltage switch matrix control circuits. The power in- creases with the duty cycle of the current DAC which scales Fig. 15. Measured spectrograms at the output of the (a) ADC, (b) 4-channel processor, (c) 6-channel processor, and (d) 8-channel processor when a loga- with the phase width and number of channels. Finally, the rithmic chirp signal is applied at the input. (e) Ideal Matlab simulation to com- energy-optimal waveform provides power savings of approx- pare against the measured results shown in (d). imately 22% and 29% at phase widths of 25 s and 50 s, respectively. 2) Overall SoC Power: Table II summarizes the total SoC meeting the 1 mW requirement. In 8-channel mode, the PZFE, power consumption with typical speech input using the energy- SAR ADC, and sound processor consume only 2% of the the optimal waveform at 31.25 s phase and nominal processor total power, while 98% of the power is consumed by the stim- settings. Reconfigurability in the number of channels allows ulator circuits. Therefore, the stimulation power savings from for a power-performance tradeoff, and the SoC consumes 572, the energy-optimal waveform transfer directly to the overall 425, and 281 W in 8-, 6-, and 4-channel modes, respectively, system. 226 IEEE JOURNAL OF SOLID-STATE CIRCUITS, VOL. 50, NO. 1, JANUARY 2015

Fig. 16. Measured current and voltage of a model electrode ( k , nF) with (a) a rectangular waveform, and (b) the energy-optimal waveform. (c) Measured current pulse trains at 1,000 pulses/sec through all electrodes in 8-channel mode.

TABLE II SUMMARY TABLE OF THE MEASURED SOCPOWER CONSUMPTION

generator and audio amplifier were used to generate a clip of speech (“her husband brought some flowers”) which was played into the ear canal of the temporal bone with a speaker at 70 dB SPL. The signal from the umbo-mounted sensor was detected by the PZFE and processed by sound processor. Fig. 18(a) shows the spectrogram and time-domain waveform of the input speech signal in the ear canal, and Fig. 18(b) shows the measured spec- trogram and reconstructed sound (the reconstruction process is based on sound synthesis in vocoder applications) from the pro- cessor in 8-channel mode. The output from the umbo-mounted Fig. 17. Measured total stimulator power across 8-, 6-, and 4-channel modes for phase widths of (a) 25 s and (b) 50 s. sensor and SoC preserves the temporal envelope information of the speech signal with the exception of some high frequency content, demonstrating hearing with a human cadaveric ear. E. System Demonstration With a Human Cadaveric Temporal Bone VII. DISCUSSION AND FUTURE WORK The FICI SoC was tested with a piezoelectric sensor mounted The eventual goal of a completely invisible CI will require at the umbo of a human cadaveric temporal bone. A function the development of other subsystems. Most notably, the user YIP et al.: A FULLY-IMPLANTABLE COCHLEAR IMPLANT SOC WITH PIEZOELECTRIC MIDDLE-EAR SENSOR 227

Fig. 18. (a) Spectrogram and time-domain waveform of the input speech signal (”her husband brought some flowers”) to the audio amplifier driving the speaker placed in the ear canal of the temporal bone. (b) Measured spectrogram and reconstructed sound from the SoC with the piezoelectric sensor mounted on a cadaver temporal bone. will need an external device (used a few times per day) ca- approaches, and enable voltage scaling down to 0.6 V to pable of two key functions. First, a wireless charging system maximize energy-efficiency. The number of channels can be is needed to recharge the implanted power management unit reconfigured between 8, 6, or 4 to enable a power-perfor- (PMU). The charging system should be able to bring the im- mance tradeoff, and all processor settings are programmable planted unit from empty to fully charged in just a few min- to ensure a patient-specific fit. Third, an auditory nerve utes to minimize the inconvenience to the user. In order to fiber model is used to determine an energy-optimal biphasic increase the time between recharge, the implanted PMU must stimulation waveform which was validated in-vivo with four be as efficient as possible which will require clever design human CI users and shown to provide 15% to 35% energy of DC-DC converters and regulators to generate the required savings. A mixed-signal arbitrary waveform stimulator is supply voltages for the SoC. If an ultra-capacitor is used as used to deliver energy-efficient current pulses to the auditory the energy storage element, the PMU would have the added nerve. In 8-channel mode, the SoC consumes just 572 Wof requirement of being able to accommodate a terminal voltage power, 98% of which is attributed to the stimulator. There- that will decrease with use. fore, the energy savings of the energy-optimal stimulation The second key function of the external device would be to waveform transfer directly to the overall system. The SoC provide wireless data transfer to the implanted unit to allow integrates implantable acoustic sensing, sound processing, the user to select from a range of programs as determined by and neural stimulation on one chip to minimize the implant their audiologist. Each program could have a different combi- size, and proof-of-concept is demonstrated by using the SoC nation of front-end and sound processor settings. Furthermore, and umbo-mounted sensor to detect a clip of speech played a wireless data link would also allow the user to manually se- into the ear canal of a human cadaver ear. lect high-performance or low-power modes corresponding to 8-channel or 6/4-channel modes respectively. Alternatively, a ACKNOWLEDGMENT smart power management system could monitor the charge of a battery or terminal voltage of an ultra-capacitor and adjust the The authors would like to acknowledge D. Eddington and number of channels in the processor to extend the time before a V. Noel for help with the stimulation waveform measurements, D. Perreault for helpful discussion, and the TSMC University recharge is required. Shuttle Program for chip fabrication.

VIII. CONCLUSION REFERENCES This paper presents a SoC with an invisible middle-ear [1]D.J.Young,M.A.Zurcher,M.Semaan,C.A.Megerian, sensor for a fully-implantable CI. First, a piezoelectric sensor and W. H. Ko, “MEMS capacitive accelerometer-based middle detects sound in the ear canal by converting the mechanical ear microphone,” IEEE Trans. Biomed. Eng., vol. 59, no. 12, motion of the middle ear into an electrical signal that is pp. 3283–3292, Dec. 2012. [2] W. Germanovix and C. Toumazou, “Design of a micropower current- captured by the PZFE. Measurements with human cadaveric mode log-domain analog cochlear implant,” IEEE Trans. Circuits Syst. ears show that the sensor is capable of detecting sounds from II, vol. 47, no. 10, pp. 1023–1046, Oct. 2000. 40 to 90 dB SPL over the bandwidth of interest. Second, [3] R. Sarpeshkar, M. W. Baker, C. D. Salthouse, J. J. Sit, L. Turicchia, and S. M. Zhak, “An analog bionic ear processor with zero-crossing detec- a highly-reconfigurable sound processor leverages digital tion,” in 2005 IEEE Int. Solid-State Circuits Conf. Dig. Tech. Papers, processing to provide greater programmability over analog San Francisco, CA, USA, Feb. 2005, pp. 78–79. 228 IEEE JOURNAL OF SOLID-STATE CIRCUITS, VOL. 50, NO. 1, JANUARY 2015

[4] D. M. Whiten, “Electro-anatomical models of the cochlear implant,” Marcus Yip (M’13) received the B.A.Sc. degree in Ph.D. dissertation, Mass. Inst.Technol.,Cambridge,MA,USA, science from the University of Toronto, 2007. Toronto, ON, Canada, in 2007, and the M.S. and [5] A. Wongsarnpigoon and W. M. Grill, “Energy-efficient waveform Ph.D. degrees from the Massachusetts Institute of shapes for neural stimulation revealed with a genetic algorithm,” J. Technology, Cambridge, MA, USA, in 2009 and Neural Eng., vol. 7, no. 4, Aug. 2010. 2013, respectively. [6]S.K.KellyandJ.L.WyattJr.,“Apower-efficient neural tissue stim- He has held internships at Actel Corporation, ulator with energy recovery,” IEEE Trans. Biomed. Circuits Syst., vol. Snowbush Microelectronics, Texas Instruments, and OnChip Power Corporation. His research interests 5, no. 1, pp. 20–29, Feb. 2011. include low-power sensor front-ends, reconfigurable [7]M.Barbara,V.Manni,andS.Monini,“Totallyimplantablemiddle analog-to-digital converters, and mixed-signal ear device for rehabilitation of sensorineural hearing loss: Preliminary circuits and systems for wearable and implantable medical applications. experience with the esteem envoy,” Acta Oto-Laryngologica, vol. 129, Dr. Yip received the Natural Sciences and Engineering Council of Canada no. 4, pp. 429–432, Apr. 2009. (NSERC) postgraduate fellowship from 2007 to 2011. [8] K. Kroll, I. Grant, and E. Javel, “The Envoy totally implantable hearing system, St. Croix Medical,” Trends in Amplification, vol. 6, no. 2, pp. 73–80, 2002. [9] A. Maniglia et al., “The middle ear bioelectronic microphone for a totally implantable cochlear hearing device for profound and total hearing loss,” Amer.J.Otology, vol. 20, no. 5, pp. 602–611, 1999. [10] A. Jenkins et al., “U.S. phase I preliminary results of use of the Oto- logics MET fully-implantable ossicular stimulator,” Otolaryng. Head Rui Jin (M’13) received the S.B. degree in electrical Neck Surgery, vol. 137, no. 2, pp. 206–232, Aug. 2007. engineering and the M.Eng degree in electrical en- [11] C. J. James et al., Ear & Hearing, vol. 24, no. 2, pp. 157–174, Apr. gineering and from the Massachu- 2003. setts Institute of Technology, Cambridge, MA, USA, [12] F. G. Zeng et al., “Speech dynamic range and its effect on cochlear im- in 2013 and 2014, respectively. plant performance,” J. Acoust. Soc. Am.,vol.111,no.1, pp. 377–386, His research interests include wireless power elec- Jan. 2002. tronics, energy-efficient sensors, and emerging appli- [13] P. C. Loizou, M. Dorman, and Z. Tu, “On the number of channels cation prototyping. needed to understand speech,” J. Acoust. Soc. Am., vol. 106, no. 4, pp. 2097–2103, Oct. 1999. [14] K. Nie, A. Barco, and F. G. Zeng, “Spectral and temporal cues in cochlear implant speech perception,” Ear & Hearing, vol. 27, no. 2, pp. 208–217, Apr. 2006. [15] R. V. Shannon, F. G. Zeng, V. Kamath, J. Wygonski, and M. Ekelid, “Speech recognition with primarily temporal cues,” Sci., vol. 270, no. 5234, pp. 303–304, Oct. 1995. [16] K. E. Fishman, R. V. Shannon, and W. H. Slattery, “Speech recognition as a function of the number of electrodes used in the SPEAK cochlear implant speech processor,” J. Speech Lang. Hear. Res.,vol.40,no.5, Hideko Heidi Nakajima received the B.S., M.S., pp. 1201–1215, Oct. 1997. and Ph.D. degrees in from [17] Q. J. Fu, R. V. Shannon, and X. Wang, “Effects of noise and spectral the Boston University College of Engineering, resolution on vowel and consonant recognition: Acoustic and electric Boston, MA, USA, and the M.D. degree from the hearing,” J.Acoust.Soc.Am., vol. 104, no. 6, pp. 3586–3596, Dec. Boston University School of Medicine. She is currently an Assistant Professor in Otology 1998. and Laryngology, Harvard Medical School, and [18] A. Faulkner, S. Rosen, and L. Wilkinson, “Effects of the number conducts research at the Eaton-Peabody Laborato- of channels and speech-to-noise ratio on rate of connected ries of the Massachusetts Eye and Ear Infirmary, discourse tracking through a simulated cochlear implant speech Boston, MA. Dr. Nakajima’s interests include ad- processor,” Ear & Hearing, vol. 22, no. 5, pp. 431–438, dressing fundamental scientific questions about the Oct. 2001. auditory system, and developing new and improved methods to diagnose and [19] M. Yip, R. Jin, H. H. Nakajima, K. M. Stankovic, and A. P. treat human hearing disease. Her primary research involves human auditory Chandrakasan, “A fully-implantable cochlear implant SoC with mechanics, middle-ear prostheses and acoustical stimulation of the cochlea. piezoelectric middle ear sensor and energy-efficient stimulation in 0.18 m HVCMOS,” in 2014 IEEE Int. Solid-State Circuits Conf. Dig. Tech. Papers,SanFrancisco,CA,USA,Feb.2014, pp. 312–313. [20] M. Yip and A. P. Chandrakasan, “A resolution-reconfigurable 5-to-10-bit 0.4-to-1 V power scalable SAR ADC for sensor applica- tions,” IEEE J. Solid-State Circuits, vol. 48, no. 6, pp. 1453–1464, Jun. 2013. Konstantina M. Stankovic trained at Massachusetts [21] B. S. Wilson, C. C. Finley, D. T. Lawson, R. D. Wolford, D. K. Institute of Technology (B.Sc. degrees in physics and Eddington, and W. M. Rabinowitz, “Better speech recognition with , Ph.D. degree) and Harvard Medical School cochlear implants,” Letters to , vol. 352, pp. 236–238, Jul. (M.D. degree, residency in otolaryngology, postdoc- 1991. toral research fellowship, clinical fellowship in neu- [22] J. Georgiou and C. Toumazou, “A 126- W cochlear chip for a totally rotology – skull base surgery). implantable system,” IEEE J. Solid-State Circuits, vol. 40, no. 2, pp. She is currently Assistant Professor of Otology 430–443, Feb. 2005. and Laryngology at Harvard Medical School and [23] M. Sahin and Y. Tie, “Non-rectangular waveforms for neural stim- Associate Surgeon at Massachusetts Eye and Ear in ulation with practical electrodes,” J. Neural Eng., vol. 4, no. 3, pp. Boston. Her research program is cross-disciplinary 227–233, Sep. 2007. and focused on hearing loss. She combines tools of [24] M. Ghovanloo and K. Najafi, “A compact large voltage-compliance systems neuroscience with electronics, optics and molecular studies to develop high output-impedance programmable current source for implantable novel diagnostics, prognostics and therapeutics for deafness. microstimulators,” IEEE Trans. Biomed. Eng.,vol.52,no.1,pp. Dr. Stankovic serves on the Editorial Board of Otology and Neurotology and 97–105, Jan. 2005. as President of the American Auditory Society. YIP et al.: A FULLY-IMPLANTABLE COCHLEAR IMPLANT SOC WITH PIEZOELECTRIC MIDDLE-EAR SENSOR 229

Anantha P. Chandrakasan (F’04) received the His research interests include micro-power digital and mixed-signal inte- B.S., M.S., and Ph.D. degrees in electrical engi- grated circuit design, wireless microsensor system design, portable multimedia neering and computer sciences from the University devices, energy efficient radios and emerging . He is a co-author of California, Berkeley, CA, USA, in 1989, 1990, of Low Power Digital CMOS Design (Kluwer Academic Publishers, 1995), and 1994, respectively. Digital Integrated Circuits (Pearson Prentice-Hall, 2003, 2nd edition), and Since September 1994, he has been with the Mass- Sub-threshold Design for Ultra-Low Power Systems (Springer 2006). He is achusetts Institute of Technology, Cambridge, where also a co-editor of Low Power CMOS Design (IEEE Press, 1998), Design of he is currently the Joseph F. and Nancy P. Keithley High-Performance Microprocessor Circuits (IEEEPress,2000),andLeakage Professor of Electrical Engineering. He was the Di- in Nanometer CMOS Technologies (Springer, 2005). rector of the MIT Microsystems Technology Labo- He has served as a technical program co-chair for the 1997 International Sym- ratories from 2006 to 2011. Since July 2011, he has posium on Low Power Electronics and Design (ISLPED), VLSI Design ’98, and been the Head of the MIT EECS Department. the 1998 IEEE Workshop on Signal Processing Systems. He was the Signal Pro- Dr. Chandrakasan was a co-recipient of several awards including the 1993 cessing Sub-committee Chair for ISSCC 1999–2001, the Program Vice-Chair IEEE Communications Society’s Best Tutorial Paper Award, the IEEE for ISSCC 2002, the Program Chair for ISSCC 2003, the Technology Direc- Devices Society’s 1997 Paul Rappaport Award for the Best Paper in an EDS tions Sub-committee Chair for ISSCC 2004–2009, and the Conference Chair publication during 1997, the 1999 DAC Design Contest Award, the 2004 DAC/ for ISSCC 2010–2014. He is the Conference Chair for ISSCC 2015. He was an ISSCC Student Design Contest Award, the 2007 ISSCC Beatrice Winner Award Associate Editor for the IEEE JOURNAL OF SOLID-STATE CIRCUITS from 1998 for Editorial Excellence and the ISSCC Jack Kilby Award for Outstanding Stu- to 2001. He served on SSCS AdCom from 2000 to 2007 and he was the meet- dent Paper (2007, 2008, 2009). He received the 2009 Semiconductor Industry ings committee chair from 2004 to 2007. Association (SIA) University Researcher Award. He is the recipient of the 2013 IEEE Donald O. Pederson Award in Solid-State Circuits.