Guide for Troubleshooting Immunofluorescence

Total Page:16

File Type:pdf, Size:1020Kb

Guide for Troubleshooting Immunofluorescence life.science.discovery. life.science.discovery. ™ Guide for Troubleshooting Immunofluorescence ● Both Positive and Negative Controls not visible through the fluorescent ● Increased incidence of weak positive specimens. (All specimens show microscope. increased fluorescent intensity.) ● Nothing seen on slide. ● Nonspecific staining with all cells apple green. ● Nonspecific staining with entire antigen substrate appearing green. ● Variation of endpoint titer of known Positive Control either weaker or stronger than expected and/or specimen results inconsistent. ● Nonspecific haze or film over entire antigen substrate in all controls and specimens. ● Positive Control stains at a weaker reaction than expected. (All specimens show decreased fluorescent intensity.) ● Nonspecific haze or film with only certain specimens. ● Negative Control stains positive and/or Positive Control stains negative. ● Nonspecific dense staining particularly around perimeter of well. ● Negative Control stains positive and/or repeated specimens not giving ● Antigen substrate cells distorted or nucleus of cells with punched out reproducible results. appearance. ● Cells visible but negative staining with all tests including Positive Control. ● Uneven staining within well. Scrapes or tears in antigen surface. ● More positive results than expected (false positives) with IgM testing. ● ● Irregular fluorescent staining pattern not closely resembling that seen ● More negative results than expected (false negatives) with IgM testing. with the Positive Control. ● Weaker results than expected with IgM testing. ● Results differ from those reported by another laboratory. Welcome to MBL Bion’s Troubleshooting in Immunofluorescence Guide. Our team of immunofluorescence specialists have worked together to identify common problems, possible causes, and possible solutions for the most commonly occurring issues in the staining and interpretation of IFA slides. 455 State Street, Suite 100 Des Plaines, IL 60016-2204 • T: 847.544.5044 F: 847.544.5051 • mblintl.com Problem Both Positive and Negative Controls not visible through the fluorescent microscope. Possible Causes Possible Solutions Check microscope with previously stained slide. If there is still nothing visible, Fluorescence microscope may not be functioning properly check filters for correct wavelength and/or replace or realign the bulb. Problem Nothing seen on slide. Possible Causes Possible Solutions • Do not touch antigen surface with hands or pipettes at any time. • Antigen wiped off during removal from pouch or during testing. • Follow rinse directions carefully. Do not focus PBS stream directly onto the • Overly vigorous use of PBS squeeze wash bottle during rinses. antigen well. • Microbial contamination of serum specimen will digest off the antigen • Take care to avoid contamination during specimen collection, separation and substrate. handling. • PBS pH too acid or too alkaline may cause loss of antigen adhesion • When making PBS, always mix well and leave overnight to dissolve. Check to slide. pH before use. Adjust pH if necessary. • Glassware dirty or not thoroughly rinsed free of detergents. • Clean glassware properly. Problem Variation of endpoint titer of known Positive Control either weaker or stronger than expected and/or specimen results inconsistent. Possible Causes Possible Solutions • Inconsistency in timing intervals and incubation temperatures. • Use consistent, uniform technique run to run and person to person. • Serum dilutions made improperly. Serum added at top of test tube and • Recheck serum dilution scheme. Prepare fresh dilutions with proper technique not mixed well into PBS. Serum amount not correct for dilution. (mixing serum and PBS, well at each step). Repeat test. • Pipettor delivery not accurate. • Pipettor calibration must be periodically checked. • Conjugate dilution made improperly. (When not using a Kit with the • Recheck dilution calculation. Prepare fresh solution and repeat. conjugate included.) • Always check expiration dates of reagents and never mix reagents from kit to • Reagents outdated or mixed between kits. kit or substitute from another manufacturer. Problem Positive Control stains at a weaker reaction than expected. (All specimens show decreased fluorescent intensity.) Possible Causes Possible Solutions • Check fluorescence microscope. Most common error is improperly • Recheck filter wavelengths. Replace or realign bulb. aligned bulb. • Store kit with all components as instructed in refrigerator. • Deterioration of all reagents particularly the antigen substrate due to Handle serum specimens sterilely and refrigerate until tested. For long term improper storage such as excessivel heat. • storage aliquot and freeze at -20 ˚C, or lower. • Deterioration of serum specimen due to improper storage (excessive heat or multiple freeze/thaw cycles). • Recheck dilution calculations. Prepare fresh solution and repeat test. • Conjugate dilution made too weak (when not using a kit with conjugate • Obtain fresh conjugate. Always store away from direct light. Cover moist included). chamber with paper towel to block light during restain. • Conjugate activity lost due to excessive exposure to strong light and/or • Check pouch for punctures. Puffiness indicates integrity not compromised. heat during storage or use in test. Remove slides from sealed pouch 5 to 30 minutes before use in test. • Antigen deterioration from punctured pouch or slides removed from • Check antigen under light microscope. Replace substrate. pouch too long before testing. • Settling of antigen substrate in center of well causes compression of • Use components furnished. Do not substitute reagents. cells which results in quenching. • Follow directions for PBS preparation carefully. Always mix well to dissolve • pH of any reagent (conjugate, PBS, mounting medium) too acid (below and pH before using. Distilled water gives more consistent results than pH 7.0). deionized water. • PBS not prepared correctly. • When not using a kit with counterstain included, establish directions for its • Evans Blue Counterstain solution is too strong quenching specific use as part of the last buffer rinse or as part of the conjugate diluent. positive fluorescence. • Allow time for reagents to equilibrate to room temperature. Cold reagents • Reagents not brought to room temperature prior to use in test. take longer to react. • Substrate not incubated long enough with specimen or conjugate. • Recheck correct time for all steps. Do not shorten incubation periods. • Excessive wash times may leach out antigenic material. • Recheck wash times. Do not leave substrate in buffer wash longer than instructed. • Mounting medium dried out between slide and coverslip. • Add additional mounting medium and/or recoverslip. • Excess mounting medium doesn’t allow tight contact between antigen substrate and coverslip for best light pathway. • Drain excess mounting medium by holding edge of slide against absorbent paper. Do not push down on coverslip. • Slides left overnight before reading may have deterioration in fluorescent intensity, especially if exposed to excessive light and/or heat. • Store slides in refrigerator away from direct light and read as soon as possible. Problem Negative Control stains positive and/or Positive Control stains negative. Possible Causes Possible Solutions Use of wrong control or reagents. Check reagent labels and repeat test. 455 State Street, Suite 100 Des Plaines, IL 60016-2204 • T: 847.544.5044 F: 847.544.5051 • mblintl.com 2 Problem Negative Control stains positive and/or repeated specimens not giving reproducible results. Possible Causes Possible Solutions • Always change pipettes or pipettor tips between use of controls and specimens. • Cross-contamination of Controls and/or specimens. Always return proper caps to reagent vials and specimen containers. • Running of controls and/or specimens between wells. • Use smaller drops to stay within well perimeters. • Splashing of excess control and/or specimens from well to well due to Follow rinse directions carefully. Alternative rinse method is to shake off excess over vigorous rinsing with wash bottle. • reagent and gently agitate slide in jar of fresh PBS. Problem Cells visible but negative staining with all tests including Positive Control Possible Causes Possible Solutions Reagents used in wrong sequence or one or more reagents not added. Review procedure and repeat test following staining steps precisely. Problem More positive results than expected (false positives) with IgM testing. Possible Causes Possible Solutions Pretreatment of serum, using ion exchange chromatography or IgG Presence of Rheumatoid Factor. immunoprecipitation, to remove IgG interference. Problem More negative results than expected (false negatives) with IgM testing. Possible Causes Possible Solutions Pretreatment of serum, using ion exchange chromatography or IgG Competition from IgG specific antibodies. immunoprecipitation, to remove IgG interference. Problem Weaker results than expected with IgM testing. Possible Causes Possible Solutions • Incubation time too short. • Serum should be incubated 90 minutes and conjugate incubated 60 minutes. • Inadequate incubation temperature. • Incubate at 37°C • Separation system inadequate (especially with extremely high IgG antibody • Check separation system procedure from manufacturer. titers). Problem Increased incidence of weak positive specimens. (All specimens show increased fluorescent intensity.) Possible Causes Possible Solutions Over-reading
Recommended publications
  • Two-Photon Excitation, Fluorescence Microscopy, and Quantitative Measurement of Two-Photon Absorption Cross Sections
    Portland State University PDXScholar Dissertations and Theses Dissertations and Theses Fall 12-1-2017 Two-Photon Excitation, Fluorescence Microscopy, and Quantitative Measurement of Two-Photon Absorption Cross Sections Fredrick Michael DeArmond Portland State University Let us know how access to this document benefits ouy . Follow this and additional works at: https://pdxscholar.library.pdx.edu/open_access_etds Part of the Physics Commons Recommended Citation DeArmond, Fredrick Michael, "Two-Photon Excitation, Fluorescence Microscopy, and Quantitative Measurement of Two-Photon Absorption Cross Sections" (2017). Dissertations and Theses. Paper 4036. 10.15760/etd.5920 This Dissertation is brought to you for free and open access. It has been accepted for inclusion in Dissertations and Theses by an authorized administrator of PDXScholar. For more information, please contact [email protected]. Two-Photon Excitation, Fluorescence Microscopy, and Quantitative Measurement of Two-Photon Absorption Cross Sections by Fredrick Michael DeArmond A dissertation submitted in partial fulfillment of the requirements for the degree of Doctor of Philosophy in Applied Physics Dissertation Committee: Erik J. Sánchez, Chair Erik Bodegom Ralf Widenhorn Robert Strongin Portland State University 2017 ABSTRACT As optical microscopy techniques continue to improve, most notably the development of super-resolution optical microscopy which garnered the Nobel Prize in Chemistry in 2014, renewed emphasis has been placed on the development and use of fluorescence microscopy techniques. Of particular note is a renewed interest in multiphoton excitation due to a number of inherent properties of the technique including simplified optical filtering, increased sample penetration, and inherently confocal operation. With this renewed interest in multiphoton fluorescence microscopy, comes increased interest in and demand for robust non-linear fluorescent markers, and characterization of the associated tool set.
    [Show full text]
  • Introduction to Flow Cytometry Principles Data Analysis Protocols Troubleshooting
    Flow Cytometry ipl.qxd 11/12/06 11:14 Page i Introduction to Flow Cytometry Principles Data analysis Protocols Troubleshooting By Misha Rahman, Ph.D. Technical advisors Andy Lane, Ph.D. Angie Swindell, M.Sc. Sarah Bartram, B.Sc. Your first choice for antibodies! Flow Cytometry ipl.qxd 11/12/06 11:14 Page ii Flow Cytometry ipl.qxd 11/12/06 11:14 Page iii Introduction to Flow Cytometry Principles Data analysis Protocols Troubleshooting By Misha Rahman, Ph.D. Technical advisors Andy Lane, Ph.D. Angie Swindell, M.Sc. Sarah Bartram, B.Sc. Flow Cytometry ipl.qxd 11/12/06 11:14 Page 2 Preface How can I explain what flow cytometry is to someone that knows nothing about it? Well, imagine it to be a lot like visiting a supermarket. You choose the goods you want and take them to the cashier. Usually you have to pile them onto a conveyor. The clerk picks up one item at a time and interrogates it with a laser to read the barcode. Once identified and if sense prevails, similar goods are collected together e.g. fruit and vegetables go into one shopping bag and household goods into another. Now picture in your mind the whole process automated; replace shopping with biological cells; and substitute the barcode with cellular markers – welcome to the world of flow cytometry and cell sorting! We aim to give you a basic overview of all the important facets of flow cytometry without delving too deeply into the complex mathematics and physics behind it all. For that there are other books (some recommended at the back).
    [Show full text]
  • Second Harmonic Imaging Microscopy
    170 Microsc Microanal 9(Suppl 2), 2003 DOI: 10.1017/S143192760344066X Copyright 2003 Microscopy Society of America Second Harmonic Imaging Microscopy Leslie M. Loew,* Andrew C. Millard,* Paul J. Campagnola,* William A. Mohler,* and Aaron Lewis‡ * Center for Biomedical Imaging Technology, University of Connecticut Health Center, Farmington, CT 06030-1507 USA ‡ Division of Applied Physics, Hebrew University of Jerusalem, Jerusalem 91904, Israel Second Harmonic Generation (SHG) has been developed in our laboratories as a high- resolution non-linear optical imaging microscopy (“SHIM”) for cellular membranes and intact tissues. SHG is a non-linear process that produces a frequency doubling of the intense laser field impinging on a material with a high second order susceptibility. It shares many of the advantageous features for microscopy of another more established non-linear optical technique: two-photon excited fluorescence (TPEF). Both are capable of optical sectioning to produce 3D images of thick specimens and both result in less photodamage to living tissue than confocal microscopy. SHG is complementary to TPEF in that it uses a different contrast mechanism and is most easily detected in the transmitted light optical path. It also does not arise via photon emission from molecular excited states, as do both 1- and 2-photon excited fluorescence. SHG of intrinsic highly ordered biological structures such as collagen has been known for some time but only recently has the full potential of high resolution 3D SHIM been demonstrated on live cells and tissues. For example, Figure 1 shows SHIM from microtubules in a living organism, C. elegans. The images were obtained from a transgenic nematode that expresses a ß-tubulin-green fluorescent protein fusion and Figure 1 also shows the TPEF image from this molecule for comparison.
    [Show full text]
  • A Multicenter Analysis of Subjectivity of Indirect Immunofluorescence Test in Antinuclear Antibody Screening
    Arch Rheumatol 2019;34(3):326-333 doi: 10.5606/ArchRheumatol.2019.7310 ORIGINAL ARTICLE A Multicenter Analysis of Subjectivity of Indirect Immunofluorescence Test in Antinuclear Antibody Screening Vildan TURAN FARAŞAT1, Talat ECEMİŞ1, Yavuz DOĞAN2, Aslı Gamze ŞENER3, Gülfem TEREK ECE4, Pınar ERBAY DÜNDAR5, Tamer ŞANLIDAĞ6 1Department of Medical Microbiology, Manisa Celal Bayar University, Faculty of Medicine, Manisa, Turkey 2Department of Medical Microbiology, Dokuz Eylül University, Faculty of Medicine, Izmir, Turkey 3Department of Medical Microbiology, Katip Çelebi University, Faculty of Medicine, Izmir, Turkey 4Department of Medical Microbiology, Izmir Medicalpark Hospital, Izmir, Turkey 5Department of Public Health, Manisa Celal Bayar University, Faculty of Medicine, Manisa, Turkey 6Department of Medical Microbiology, Manisa Celal Bayar University, Faculty of Medicine, Manisa, Turkey ABSTRACT Objectives: This study aims to evaluate the interpretation of the antinuclear antibody (ANA)-indirect immunofluorescence (IIF) test results based on the interpreter-related subjectivity and to examine the inter-center agreement rates with the performance of each laboratory. Patients and methods: The ANA-IIF testing was carried out in a total of 600 sera and evaluated by four laboratories. The inter-center agreement rates were detected. The same results given by the four centers were accepted as gold standard and the predictive values of each center were calculated. Results: The inter-center agreement was reported for ANA-IIF test results from 392 of 600 (65.3%) sera, while 154 of 392 results were positive. Four study centers reported 213 (35.5%), 222 (37.0%), 266 (44.3%), and 361 (60.2%) positive test results, respectively. In terms of the patterns, the highest and lowest positive predictive values were 72.3% and 42.7%, respectively, while the highest and lowest negative predictive values were 99.6% and 61.5%, respectively.
    [Show full text]
  • EXPERIMENT 15 TF Notes
    EXPERIMENT 15 TF Notes 1. Have students log onto LoggerPro3 and start heating water as soon as they arrive 2. This reaction is easily contaminated so students must never insert the thermometer directly into the cuvette containing the reaction. Instead they should rather measure the temperature of the water and equate that to the temperature of the solution in the cuvette. 3. We will be using hot gloves in this experiment. 4. Make sure you are familiar with the calculations necessary to fill out the chart. All the necessary equations are given at the top fo the page. Note that the temperature must be converted from °C to K. EXPERIMENT 15 Thermodynamics of Complex-Ion Equilibria Introduction Thermodynamic data for a reaction system provides researchers with information that is important from both theoretical and practical points of view. There are several thermodynamic properties that chemists pay close attention to when designing or carrying out experiments such as thermodynamic stability, the change in free energy of a reaction, and temperature dependence. For example, if a chemist wants to create a new type of solar cell that combines a semiconductor material with a novel conductive oxide and wants to make sure that the two materials will not react with each other, thermodynamics provide the answer. By finding the free energy change associated with the reaction, s/he can determine how stable the layers are in contact with each other and to what temperature. In this experiment, you will learn how to determine those parameters from a controlled experiment by using spectrometry to find concentration data at various temperatures.
    [Show full text]
  • Comparison of Histopathology, Immunofluorescence, and Serology
    Global Dermatology Cae Report ISSN: 2056-7863 Comparison of histopathology, immunofluorescence, and serology for the diagnosis of autoimmune bullous disorders: an update Seline Ali E1, Seline Lauren N1, Sokumbi Olayemi1* and Motaparthi Kiran2,3 1Department of Dermatology, Medical College of Wisconsin, Milwaukee, WI, USA 2Dermatopathology, Miraca Life Sciences, USA 3Department of Dermatology, University of Texas Southwestern Medical Center, Dallas, TX, USA Introduction In an ELISA, the target antigen of interest (such as the NC16a domain of BP180) is immobilized by physical adsorption or by The diagnosis of autoimmune bullous disorders (AIBDs) relies on antibody capture. When antibody capture is utilized, this is referred to several different diagnostic methods. These include histopathology, as “sandwich ELISA” because the target antigen is bound between the direct immunofluorescence (DIF), indirect immunofluorescence immobilizing antibody and the primary antibody. Primary antibodies (IIF), enzyme-linked immunosorbent assay (ELISA) and are present in the patient’s serum. Enzyme-linked secondary antibodies immunoblotting. When faced with a presumptive AIBD, the are then added which bind the Fc region of primary antibodies. most widely employed method for diagnosis by dermatologists is Substrate is added and converted by the enzyme into a signal. A a combination of histopathology and DIF. While DIF is still the resulting color change, fluorescence, or electrochemical signal is diagnostic method of choice for linear IgA bullous disease and IgA quantitatively measured and reported [5]. pemphigus, ELISA is a more accurate, cost-effective and less invasive method of diagnosis for several AIBDs including pemphigus vulgaris Western blot is synonymous with immunoblot. For this method, and foliaceus, based on currently available evidence [1-3].
    [Show full text]
  • Laboratory Equipment Reference Sheet
    Laboratory Equipment Stirring Rod: Reference Sheet: Iron Ring: Description: Glass rod. Uses: To stir combinations; To use in pouring liquids. Evaporating Dish: Description: Iron ring with a screw fastener; Several Sizes Uses: To fasten to the ring stand as a support for an apparatus Description: Porcelain dish. Buret Clamp/Test Tube Clamp: Uses: As a container for small amounts of liquids being evaporated. Glass Plate: Description: Metal clamp with a screw fastener, swivel and lock nut, adjusting screw, and a curved clamp. Uses: To hold an apparatus; May be fastened to a ring stand. Mortar and Pestle: Description: Thick glass. Uses: Many uses; Should not be heated Description: Heavy porcelain dish with a grinder. Watch Glass: Uses: To grind chemicals to a powder. Spatula: Description: Curved glass. Uses: May be used as a beaker cover; May be used in evaporating very small amounts of Description: Made of metal or porcelain. liquid. Uses: To transfer solid chemicals in weighing. Funnel: Triangular File: Description: Metal file with three cutting edges. Uses: To scratch glass or file. Rubber Connector: Description: Glass or plastic. Uses: To hold filter paper; May be used in pouring Description: Short length of tubing. Medicine Dropper: Uses: To connect parts of an apparatus. Pinch Clamp: Description: Glass tip with a rubber bulb. Uses: To transfer small amounts of liquid. Forceps: Description: Metal clamp with finger grips. Uses: To clamp a rubber connector. Test Tube Rack: Description: Metal Uses: To pick up or hold small objects. Beaker: Description: Rack; May be wood, metal, or plastic. Uses: To hold test tubes in an upright position.
    [Show full text]
  • Fv1000 Fluoview
    Confocal Laser Scanning Biological Microscope FV1000 FLUOVIEW FLUOVIEW—Always Evolving FLUOVIEW–—From Olympus is Open FLUOVIEW—More Advanced than Ever The Olympus FLUOVIEW FV1000 confocal laser scanning microscope delivers efficient and reliable performance together with the high resolution required for multi-dimensional observation of cell and tissue morphology, and precise molecular localization. The FV1000 incorporates the industry’s first dedicated laser light stimulation scanner to achieve simultaneous targeted laser stimulation and imaging for real-time visualization of rapid cell responses. The FV1000 also measures diffusion coefficients of intracellular molecules, quantifying molecular kinetics. Quite simply, the FLUOVIEW FV1000 represents a new plateau, bringing “imaging to analysis.” Olympus continues to drive forward the development of FLUOVIEW microscopes, using input from researchers to meet their evolving demands and bringing “imaging to analysis.” Quality Performance with Innovative Design FV10i 1 Imaging to Analysis ing up New Worlds From Imaging to Analysis FV1000 Advanced Deeper Imaging with High Resolution FV1000MPE 2 Advanced FLUOVIEW Systems Enhance the Power of Your Research Superb Optical Systems Set the Standard for Accuracy and Sensitivity. Two types of detectors deliver enhanced accuracy and sensitivity, and are paired with a new objective with low chromatic aberration, to deliver even better precision for colocalization analysis. These optical advances boost the overall system capabilities and raise performance to a new level. Imaging, Stimulation and Measurement— Advanced Analytical Methods for Quantification. Now equipped to measure the diffusion coefficients of intracellular molecules, for quantification of the dynamic interactions of molecules inside live cell. FLUOVIEW opens up new worlds of measurement. Evolving Systems Meet the Demands of Your Application.
    [Show full text]
  • Manual: Zetasizer Nano Accessories Guide
    ! www.malvern.com Zetasizer nano series Malvern Instruments Limited Printed in England MRK1378-01 Enigma Business Park Q & Grovewood Road, Malvern Worcs, WR14 1XZ, U.K. SelAccessories Guide Tel: +44 (0) 1684 892456 Fax: +44 (0) 1684 892789 Malvern Zetasizer Nano accessories guide MAN0487 Issue 1.1 April 2013 Copyright © 2007 - 2013 Malvern Instruments Ltd. Malvern Instruments pursues a policy of continual improvement due to technical development. We therefore reserve the right to deviate from information, descriptions, and specifications in this publication without notice. Malvern Instruments shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance or use of this material. No reproduction or transmission of any part of this publication is allowed without the express written permission of Malvern Instruments Ltd. Head office: Malvern Instruments Ltd. Enigma Business Park, Grovewood Road, Malvern, Worcestershire WR14 1XZ United Kingdom. Tel + [44] (0)1684-892456 Fax + [44] (0)1684-892789 Zetasizer, Malvern and the 'hills' logo are registered trademarks in the UK and/or other countries, and are owned by Malvern Instruments Ltd. NIBS and M3-PALS are trademarks of Malvern Instruments. M3 is granted Euro Pat No: 1 154 266 DE FR. Windows is a registered trademark of Microsoft Corporation. Tygon is a registered trademark of Saint-Gobain Corporation. Hellmanex is a registered trademark of Hellma GmbH & Co. KG. Table of contents Introduction and accessory range Introduction . 1-1 Accessory range. 1-1 General cells and cuvettes Introduction . 2-1 Cuvette holder . 2-2 Cell and cuvettes . 2-3 Size and molecular weight cuvettes .
    [Show full text]
  • 130-77 Instruction Manual Updated 10/20/2020 Ver
    OFITE HTHP Viscometer Part No. 130-77 Instruction Manual Updated 10/20/2020 Ver. 9 OFI Testing Equipment, Inc. 11302 Steeplecrest Dr. · Houston, Texas · 77065 · U.S.A. Tele: 832.320.7300 · Fax: 713.880.9886 · www.ofite.com ©Copyright OFITE 2015 Intro ..................................................................................................2 Table of Components ....................................................................................3 Contents Specifications .................................................................................5 Setup ................................................................................................6 Computer ....................................................................................6 Viscometer ..................................................................................7 Control Panel ..................................................................................8 Operation .........................................................................................9 Software Start ...............................................................................14 Calibration .....................................................................................18 Fluid Manager ...........................................................................22 Software ........................................................................................23 Options ......................................................................................23 Save Rate Settings
    [Show full text]
  • Two-Photon Excitation Fluorescence Microscopy
    P1: FhN/ftt P2: FhN July 10, 2000 11:18 Annual Reviews AR106-15 Annu. Rev. Biomed. Eng. 2000. 02:399–429 Copyright c 2000 by Annual Reviews. All rights reserved TWO-PHOTON EXCITATION FLUORESCENCE MICROSCOPY PeterT.C.So1,ChenY.Dong1, Barry R. Masters2, and Keith M. Berland3 1Department of Mechanical Engineering, Massachusetts Institute of Technology, Cambridge, Massachusetts 02139; e-mail: [email protected] 2Department of Ophthalmology, University of Bern, Bern, Switzerland 3Department of Physics, Emory University, Atlanta, Georgia 30322 Key Words multiphoton, fluorescence spectroscopy, single molecule, functional imaging, tissue imaging ■ Abstract Two-photon fluorescence microscopy is one of the most important re- cent inventions in biological imaging. This technology enables noninvasive study of biological specimens in three dimensions with submicrometer resolution. Two-photon excitation of fluorophores results from the simultaneous absorption of two photons. This excitation process has a number of unique advantages, such as reduced specimen photodamage and enhanced penetration depth. It also produces higher-contrast im- ages and is a novel method to trigger localized photochemical reactions. Two-photon microscopy continues to find an increasing number of applications in biology and medicine. CONTENTS INTRODUCTION ................................................ 400 HISTORICAL REVIEW OF TWO-PHOTON MICROSCOPY TECHNOLOGY ...401 BASIC PRINCIPLES OF TWO-PHOTON MICROSCOPY ..................402 Physical Basis for Two-Photon Excitation ............................
    [Show full text]
  • SPARQ-Ed Risk Assessment Sheet : Immunofluorescence
    SPARQ-ed Risk Assessment Sheet : Immunofluorescence Hazard Analyse / Evaluate Risk Overall Risk Category Description of Risk Source Current Controls Event Category Consequences Exposure Probability (see explanation on last page) Exposure PPE worn (gloves, closed Prob footwear, and safety Chemical exposure - VR R U O F C Exposure to Chemical Agents : Risk of Unusual : goggles provided). Use body contact, spills Unusual but eye and skin exposure to chemicals Immuno- AC Low Low Low Low Mod Subs of potentially harmful and splash and Other Contact Minor : General Possible: Less likely (such as fixation / permeabilisation fluorescence QP Low Low Low Low Low Mod Chemical chemicals/substances in inhalation (eg. 4% with Chemical or chemicals used would to occur with the buffer containing 4% staining not a UP Low Low Low Low Low Low fume cupboard/safety paraformaldehyde is Substance be irritants only. control measures but paraformaldehyde) throughout the common cabinet. MSDS available. irritant to eyes and not impossible. RP Low Low Low Low Low Low staining procedure. procedure. Well documented skin). C Low Low Low Low Low Low procedures. PI Low Low Low Low Low Low Personal precaution Exposure procedures to be Prob followed in case of VR R U O F C Sharps : Immunoflourescence is injury. PPE worn Sharps injury (cuts or Unusual : AC Low Low Low Low Mod Subs generally performed with sections on (wearing gloves, closed slicing of fingers) Immuno- Conceivable : Sharps slides or with cells on coverslips. shoes, lab coat and from broken Hitting Objects Minor : May require fluorescence injury is likely to QP Low Low Low Low Low Mod Immunofluorescence staining of cells Mechanical safety goggles are coverslips or pointed with Part of first aid for minor cuts.
    [Show full text]