Basic Vortex Mixer 88882011 & 88882012

In the United States: In Canada: For customer service, call 1-800-766-7000 For customer service, call 1-800-234-7437 To fax an order, use 1-800-926-1166 To fax an order, use 1-800-463-2996 To order online: thermofisher.com To order online: thermofisher.ca Operating Manual Find out more at thermofisher.com Revision A . 04 01 2021

©2021 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. Contents

Section 1 Important Information 1 Disposing of Product 2

Section 2 Introduction 3 Feature 3

Section 3 Inspection and Installation 4 Packing List 4 Connections 4 Structure Diagram 4

Section 4 Overview 5 Specifications 5 Environmental Conditions 5 Safety Instructions 6

Section 5 Operation 7 Control Panel 7 Installation of Accessories 8 Settings 10

Section 6 Safety Tips and Maintenance 11

Section 7 Troubleshooting 13

Section 8 Optional Accessories 14

Section 9 Warranty 15

Section 10 Compliance 16 Section 1 Important Information

Ignoring the following warnings could operating and request the service. the main power immediately and request cause serious injuries or even fatal the service if water may be in the product. accidents. Keep out of the direct sunlight. It may influence product life and proper Do not let the product take any strong Check the voltage, phase and capacity of operation. shock or vibration. It could cause power supply on the ID plate before abnormal operation or trouble. It may installation. Connect properly. Do not use the machine at places where deteriorate the ability of the product moisture is high and flooding can operation and not obtain correct results. Power supply must be properly happen. grounded. Abnormal grounded Do not sprinkle insecticide or flammable connection causes serious damage. Do not assemble, repair, modify on your spray on the product. Use smooth Grounded connection must not be on the own, this product may not work well and cloths. Cleaning with solvent can cause water pipe and gas pipe. electrical shock may possible. fire and deformity. Use provided power cord. Power cord: Changes to the product could result in Power off while product cleaning. It may Wall outlet with grounded terminal product inefficiency. Also this will void cause electric shock or fire. power cord 250V 10A. warranty. Which also indicates a hazardous situation, if not avoided may Do not drop or allow the machine to fall. Do not install the product in a place that result in minor or moderate injury. It will cause wrong operation and gas could leak. Do not use in a place that malfunction. has industrial oil smoke or metallic dust. Do not put heavy things on the power cord. It causes fire or electric shock. Do not Do not put the machine on the cord. It Only accessories that meet the use the machine near to places where may take off the wire coating and cause manufacturer's specifications shall be explosion could happen due to organic electric shock or fire. used. evaporating gases. Do not touch it with wet hands and place the main plug correctly. It could cause Explosive materials: acid, esther, nitro Disposing of Product compound. the electric shock or injuries. Installing power outlet near instrument Dispose of the unit by separating the Inflammable materials: salt peroxides, plastic, motor and electronics and inorganic peroxide, salt acids. may be convenient. Allow adequate spacing around the product to ensure proper dispose according to local laws and Check equipment for permissible ventilation, accessibility to the power regulations. environmental conditions when using switch and for allowing the instrument to inside of Temperature and Humidity run without impediment. Chamber or . It can be the cause of fire or trouble by stirrer Do not install the Vortex Mixer near electricity, electronic, and damage of machinery generating high frequency noise. motor. Avoid installation close to high frequency- welding machine, sewing machine, or Mixer's permissible environmental mass SCR controller. condition. Temperature 5°C to 40°C, Maximum relative humidity 85%. Do not inject any liquid and inflammable things inside of product. Unplug if there is a strange sound, smell and/or smoke from the product. Stop Do not pour water or put liquid on the top of the product when cleaning. Disconnect 1 Section 1 | Important Information Basic Vortex Mixer 2 Section 2 Introduction Section 3 Inspection and Installation

Welcome to use Thermo Scientific Vortex 2.1 Feature Before unpacking the unit, first check for After unpacking, check that all unit parts Mixers. Vortex mixers are used in biomedical It can be easily switched between touch damages in the package of the unit. and accessories are as listed below. engineering, physical and chemical analysis, and continuous modes and the Contact us or the agent from which you Then unpack the unit. Check carefully to purchased the unit if any components environmental monitoring, food, hygiene, corresponding light is on when working see if there were damages incurred were omitted. petrochemical, metallurgy and other fields, normally. during transit. for , sample tissue, chemical ◆It has speed adjustment display reagents, bacteria liquid and other substances function. mixed. It is suitable for test tubes, centrifugal 3.1 Packing List ◆It uses a knob to achieve stepless tubes, color tubes, enzyme plates, deep speed regulation. Description Catalog Number Figure hole plates, PCR plates and other containers. ◆It has a high sensitivity touch switch. ◆A variety of accessories can be Basic This product is intended to be used in a Vortex Mixer 88882011 88882012 setting and is not to be used selected to maximize the needs of in a clinical setting. customers. ◆It uses brushless DC motor, with accurate speed control, stability, safe Rubber Platform 1 1 and reliable features. ◆DC motor design is suitable for long- Power Adapter 1 N/A time operation, up to 1 week or longer. 120V ◆It has a power recovery function. Multi-Power ◆The housing material is ABS, which N/A 1 Adapter 220V has excellent heat resistance, corrosion resistance, impact resistance and chemical resistance.

3.2 Connections 3.3 Structure Diagram

Rubber Platform

Control Panel Knob Foot(8)

Power Housing Power Switch Adapter

3 Section 2 | Introduction Basic Vortex Mixer 4 Section 4 Overview

4.1 Specifications 4.3 Safety Instructions Rotation Speed Speed Range...... 200 to 3000rpm Please read the entire instruction manual Speed Accuracy...... ±5% before operating the Basic Digital Vortex. ¢ Orbit Diameter...... 4mm Operating Mode...... Touch /Continuous WARNING DO NOT use the Basic Drive Motor...... Brushless DC motor Vortex Mixer in a hazardous (Suitable for long-time operation) atmosphere or with hazardous materials for which the unit was Load Maximum Load……………………………………...... 0.5kg @ 800rpm not designed. Also, the user should be aware that the protection provided Size Overall Dimensions...... 205×151×131mm by the equipment may be impaired Platform Diameter...... ¢76mm if accessories used are no provided Packing Dimensions...... 290×270×240mm or recommended by the manufacturer, or are used in a manner not specified Weight Net Weight...... 6.5Kg(14.33lb) by the manufacturer. Gross Weight...... 6.8Kg(14.99lb) CAUTION ! To avoid electrical shock, completely cut off power to the unit Power Supply Input Requirement...... 100-240VAC~, 50/60Hz, 6W by disconnecting the power cord from Output Requirement...... 24V ,1A the unit or unplug from the wall outlet. Disconnect unit from the power Others Certification...... RoHS, WEEE, cCSAus, CE Mark supply prior to maintenance and IP Class...... IP32 service. Any spills should be removed promptly. Bio hazard spills should be cleaned using approved liquid 4.2 Environmental Conditions promptly. Solvent spills are a fire hazard. Stop the unit immediately, Application Environmental Conditions: indoor use and DO NOT operate until clean up Temperature...... 5 to 40℃ is complete and vapors have dissipated. Altitude...... ≤2,000 m DO NOT immerse the unit for cleaning. Humidity...... 20% to 85% DO NOT operate the unit if it shows signs of electrical or mechanical Storage Environmental Conditions damage. Temperature...... 0 to 60℃ Altitude...... ≤2,000 m Humidity...... 20% to 90%, non-condensing

5 Section 4 | Overview Basic Vortex Mixer 6 Section 5 Operation

This chapter covers the control panel and its operation. 5.2 Installation of Accessories Vessel Platform 5.1 Control Panel Only accessories that meet the manufacturer's specifications shall be Base Tray used. Installation of Rubber Platform 1. Place the Rubber Platform on the top ① of the base tray and align the inner Step1 Step2 ④ bump of the Rubber Platform is to ② the groove on the base tray. Pits match ③ ⑤ 2. Press evenly along the perimeter of the Rubber Platform so that the Step1 Detail View ⑥ Rubber Platform is stuck on the base tray. Installation of Tray Rubber Platform Microplate Tray are suitable for “continuous” mode and samples are mixed with Base Tray enzyme ELISA PLATE. ①. Run/Stop button: Run or stop the 1. Parallel place the Microplate Tray on instrument. Power Switch the top of the base tray. ②. Mode button: To switch running 2. Align the inner bump of the Microplate Step1 Step2 Tray is to the groove on the base tray. mode between "Touch" and ⑦ "Continuous". gently press both sides of the Microplate ⑧ ③. Continuous Mode indicator: When in Pits match Tray to snap it onto the base tray, and Continuous Mode, light up. please make sure it tighten. ④. Speed display window: The window ⑦. Gear I:When the switch is in “gear 1", Step1 Detail View displays set speed (when the the instrument power is on. instrument is in standby) or current ⑧. Gear O:When the switch is in “gear O”, Installation of Flat Bottom Vessel speed (when the instrument is the instrument power is off. Platform running). 1. Place the Flat Bottom Vessel Platform Flat angle Microplate Tray ⑤. Touch Mode indicator: When in on the top of the base tray and align match Touch mode, light up. the inner bump of the Flat Bottom Base Tray ⑥. Speed setting Knob: Turn the knob Vessel Platform is to the groove on clockwise or counterclokwise are the base tray. used to increase/decrease the set 2. Press evenly along the perimeter of Step1 speed of the instrument. the Flat Bottom Vessel Platform so that the Flat Bottom Vessel Platform 7 Section 5 | Operation is stuck on the base tray. Basic Vortex Mixer 8 Installation of Tube Holder Speed setting Push EVA Sponge Pad Caution: Turn the speed knob clockwise or When replacing the accessories, remove counterclockwise, the Speed display Push power from the instrument by unplugging. Tube Holder window to set the target speed, for Tube Holders are suitable for "Continuous" example: mode to mix samples by test tubes. 1 6 0 0 Step2 1. Parallel place the Tube Holder on the Step3 In this case, the set speed is 1600rpm. top of the base tray. Press the Speed setting knob, the digital 2. Align the inner bump of the Tube Holder 5.3 Settings tube starts flashing, turn the knob to is to the groove on the base tray. gently Please confirm the environmental condition adjust the display window value to press both sides of the Tube Holder to and the power voltage first, and then, connect 1600,then press the knob again within Groove snap it onto the base tray, and please the mixer power cord to the power outlet, 5s,the digital tube stop flashing, the alignment make sure it tighten. Ensure the machine is on a flat surface display window shows 1600,the speed 3. Install the corresponding EVA sponge Step2 Detail View prior to operating. setting is finished. pad on the Tube Holder, after inserting Note: the into the EVA sponge pad, Switch On Power Recovery (1). When mixing by ELISA PLATE, the the instrument can than be turned on. Press the power switch to position “I” Continuous Mode: If the power is cut off capacity of the ELISA PLATE shall and the display window is shown as suddenly while the instrument is in not exceed 1/2 of the maximum Tube Holder below: operation, the instrument will resume the capacity. Otherwise, there may be 0 2 0 0 parameters set before the power is cut liquid splash. off, but the instrument will not continue Flat angle Speed display window (2). When Microplate Trap(88882122) is match The corresponding indicator for the to run. used with Deepwellplate, the Base Tray running mode is on. Touch Mode: The power is cut off when capacity of the Deepwellplate shall the instrument is working. If the power is not exceed 2/3 of the maximum Mode Settings on again, the instrument will resume the

capacity, and the maximum speed Step1 In standby, press “ ” button to switch parameters set before the power is off, not exceed 650rpm. Push “Touch” mode or “Continuous” mode. In put the vessel onto the Rubber Platform (3). The Recommended speed values in “Touch" mode, the instrument starts or Flat Bottom Vessel Platform and Push the table are for reference only running automatically when the vessel press to run. (Section Optional Accessories). presses on the Rubber Platform or Flat Change the speed value according Bottom vessel Platform, and automatically Finish Operation Step2 to the actual operation of the stops when the vessel leaves the platform. After use, please press power switch to instrument. Groove Run In “Continuous” mode, press “ Stop ” button, turn off the power. Unplug the instrument alignment the instrument with tray, platform or tube and store it according to the storage holder starts running continuously, and guide.

Run press “ Stop ” button again, the instrument Step2 Detail View 9 Section 5 | Operation stops running. Basic Vortex Mixer 10 Section 6 Safety Tips and Maintenance

Please read the manual carefully and 8. For a single test tube mixing, please Cleaning Warning: The operator is not permitted follow the following safety guidelines put test tube in the middle hole of the In order to ensure the safe use of the to gain access using a tool, to a part before operating this instrument. rubber platform. For multi-test tube instrument, please follow the which in normal use may be hazardous 1. Keep this manual for reference. mixing, please always place the test manufacturer’s recommendations when live. 2. Professionals are required to operate tubes in the center and symmetrically. cleaning the instrument. the instrument. 9. Be careful of the risks that may arise. ● Unplug the power first when cleaning. 3. The instrument can only be used with ◆Flammable substances ● Wipe the instrument with a damp, soft the original power adapter. ◆Fragile substances cloth or non-corrosive cleaning agent 4. When operating the instrument, the 10. Unplug the power before installing ● Direct spray instruments are operator must choose to wear the accessories. prohibited. appropriate clothing to avoid toxic 11. Carefully handle the instrument to ● Make sure that the instrument is and harmful solvents produced as avoid any collision or impact. completely dry before operating it. follows: ● Please wear gloves when cleaning. ◆Harmful liquids splashed Warning: Avoid dripping detergent or ◆The harmful solids ejected water into the inside of the instrument ◆Body, hair, clothing is rolled or corroded during cleaning. 5. If the instrument becomes abnormal or higher vibration in use, please immediately Clean Spill reduce the speed accordingly, If the If accidental spillage of liquids caused by instrument becomes abnormal or mishandling or contained breakage occurs there is a higher vibration when in use, on the surface of the instrument, please immediately reduce the speed accordingly shut down the instrument and clean up or immediately stop the instrument the liquid immediately. If the liquid has but turning off the power. already spilled into the unit, cut off the 6. Ensure that the accessories and power supply first and immediately clean containers are installed correctly. up the liquid at the surface of the instrument. 7. Check that the instrument and the Place the instrument in a ventilated and accessories are in good condition dry environment for 24 hours before reuse. before each operation. If the instrument is not functioning after drying for 24 hours, please contact the manufacturer.

11 Section 6 | Safety Tips and maintence Basic Vortex Mixer 12 Section 7 Troubleshooting

Please refer to the following table to If the problem still exists, contact your e troubleshoot if any malfunction occurs. local sales representative. Figur Error Cause Solution

Power disconnected Connect the power Not working properly

Power switch off Switch on power AC 1.0A 1.8m AC 1.0A 1.8m

The instrument is not Place the instrument on a 82×16.5mm 80×9mm 100×54mm 100×54mm 100×68mm 100×68mm 100×68mm

stable horizontal, rugged platform - Dimensions ¢ ¢ 136×97×11.5mm ¢ ¢ ¢ ¢ ¢ 100~240V 100~240V Loud noise The sample in the test Fill an equivalent amount of Vibration large tube is imbalance sample in the test tube Abnormal sound Contact with other objects Remove the objects contacted of the instrument’s case 200~2000rpm Half full with lid Speed Range 200~3000rpm 200~3000rpm 200~1200rpm 200~2000rpm 200~1800rpm 200~1000rpm 200~1000rpm 200~900rpm - - - - √ √ √ √ √ √ √ √ √ Continuous ------√ √ ouch T Cat. No. 88882120 88882121 88882122 88882122 88882103 88882123 88882124 88882125 88882126 88882127 88870124 88870125 ube Holder ube Holder ube Holder ube Holder ube Holder y y with essel Platform ra ra ower Adapter 220V Section 8 Optional Accessories oplate T oplate T ower Adapter 120V Description Rubber Platform Flat Bottom V Micr Micr 10mm dia x19 T 12mm dia x 13 T 15mm dia x 12 T 20mm dia x 7 T 26mm dia x 4T Multi-P P Rubber Strips 13 Section 7 | Troubleshooting Section 9 Warranty Section 10 Compliance

THERMO FISHER SCIENTIFIC THIS WARRANTY IS EXCLUSIVE AND IN 1. European Union the instruction manual, may cause STANDARD PRODUCT WARRANTY LIEU OF ALL OTHER WARRANTIES, The European voltage models harmful inter f er ence to radio The Warranty Period starts two weeks WHETHER WRITTEN, ORAL OR IMPLIED. of this product meet all the communications. Operation of this from the date your equipment is shipped NO WARRANTIES OF applicable requirements of the European equipment in a residential area is likely to from our facility. This allows for shipping MERCHANTABILITY OR FITNESS FOR A Directives and therefore display the CE cause harmful interference in which time so the warranty will go into effect at PARTICULAR PURPOSE SHALL APPLY. Marking. These Directives are captured case the user will be required to correct approximately the same time your Thermo shall not be liable for any in the EU Declaration of Conformity the interference at his own expense. equipment is delivered. The warranty indirect or consequential damages which may be obtained from the 3.2 Canadian ISED IC Notice protection extends to any subsequent including, without limitation, damages manufacturer. This ISM digital apparatus complies with owner during the first year warranty relating to lost profits or loss of Canadian ICES-001. period. products. 2. Pr oduct Safety Cetappareil ISM estconformeá la norme During the first two (2) years, component Your local Thermo Sales Office is ready This product family has been tested to NMB-001 du Canada.owing optimum parts proven to be non-conforming in to help with comprehensive site applicable productsafety standards by a tray rotation. materials or workmanship will be preparation information before your Nationally Recognized Test Laboratory 3.3 Korean KC Registration (NRTL) and may bear the NRTL's mark of repaired or replaced at Thermo's equipment arrives. Printed instruction 사용자안내문 expense, labor included. Installation and manuals carefully detail equipment safety compliance to those applicable 이기기는엄무용환경에서사용할 calibration are not covered by this installation, operation and preventive standards. 목적으로적함성평가를받은기기 warranty agreement. The Technical maintenance. 로서가정용환경에서사용하는경우전파 간섬의우려가있습니다 Services Department must be contacted 3. Electromagnetic Compatibility If equipment service is required, please WARNING Statement: EMC Registration for warranty determination and direction 3.1 FCC Statement (USA) call your Technical Services Department is done on this equipment for business prior to performance of any repairs. Any changes or modifications at 1-866-984-3766, option number 2. use only. It may cause interference when Expendable items, glass, filters and not expressly approved by the We're ready to answer your questions on the product would be used in home. This gaskets are excluded from this warranty. party responsible for compliance could equipment warranty, operation, warning statement applies a product for void the user's authority to operate the Replacement or repair of components maintenance, service and special business use. parts or equipment under this warranty application. Outside the USA, please equipment. Note: This equipment has been tested shall not extend the warranty to either contact local Thermo Technical Services 4. Environmental Compliance the equipment or to the component part Department or local distributor for and found to comply with the limits for a 4.1 Evaluation of Chemicals - beyond the original warranty period. The warranty information. Class B digital device, pursuant to part Regulations and Directives Technical Services Department must 15 of the FCC Rules. These limits are 4.1.1 REACH - Europe give prior approval for return of any designed to provide reasonable Thermo Fisher Scientific is components or equipment. At Thermo's protection against harmful interference committed to meeting all option, all non-conforming parts must when the equipment is operated in a compliance obligations to evaluate, be returned to Thermo Fisher Scientific commercial environment. This communicate, and register any postage paid and replacement parts are equipment generates, uses, and can Substances of Very High Concern shipped FOB destination. radiate radio frequency energy and, if not installed and used in accordance with (SVHC), and finding alternates where 15 Section 9 | Warranty Basic Vortex Mixer 16 appropriate. exemptions of the regulation due to their companies in each EU Member State, 4.1.2 RoHS - Europe application. and this product should be disposed of Thermo Fisher Scientific is determined 4.1.3 RoHS - China or recycled through them. Further to reduce the impact we have on the This product complies with the information on our compliance with environment, and so can declare that requirements of the legislative act these Directives, the recyclers in your this product fully complies with the Administration on the Control of country, and information on Thermo European Parliament's RoHS2 Pollution Caused by Electronic Scientific products which may assist the (Restriction of Hazardous Substances) Information Products (ACPEIP). A label detection of substances subject to the Directive 2011/65/EU, with respect to of conformance, such as one of the RoHS Directive are available at the limitation for the following following, may be found on the product: www.thermofisher.com/ substances: 5 10 15 20 25 50 ·Lead (0.1 %) ·Mercury (0.1 %) 5. End of Life Care ·Cadmium (0.01 %) Some considerations and suggestions ·Hexavalent chromium (0.1 %) are listed below for proper disposal of ·Polybrominated biphenyls (PBB) (0.1 this product. While addressing these %) actions for safe recycling and disposal, ·Polybrominated diphenyl ethers please follow all guidelines,Safety Data (PBDE) (0.1 %) Sheets (SDS), or regulations applicable to your location. 2015/863 Annex II (RoHS2 Amendment, · In effect after July 22, 2021) This product has materials and components that may be recycled or ·Bis(2-ethylhexyl) phthalate (DEHP) reused according to local guidelines and (0.1 %) regulations. ·Butyl benzyl phthalate (BBP) (0.1 %) ·Clean up any chemical or biological ·Dibutyl phthalate (DBP) (0.1 %) safety hazards. ·Diisobutyl phthalate (DIBP) (0.1 %) WEEE Compliance. This Our compliance is witnessed by written product is required to comply declaration from our suppliers and/or with the European Union's component testing. This confirms that Waste Electrical & Electronic Equipment any potential trace contamination levels (WEEE) Directive 2012/19/EU. It is of the substances listed above are below marked with the following symbol. the maximum level set by the latest Thermo Fisher Scientific has contracted regulations or follow established with one or more recycling/disposal

17 Section 10 | Compliance Basic Vortex Mixer 18