Tech Tip 60602

Total Page:16

File Type:pdf, Size:1020Kb

Tech Tip 60602 www.iotech.com TECHTIP 60602 Basic Strain Measurements using IOtech DAQ Hardware INTRODUCTION DISCUSSION Strain gages are sensing devices used in a variety of physical test and STRAIN MEASUREMENT CONFIGURATIONS measurement applications. They change resistance at their output Wheatstone Bridge terminals when stretched or compressed. Because of this character- To make an accurate strain measurement, extremely small resis- istic, the gages are typically bonded to the surface of a solid material tance changes must be measured. The Wheatstone bridge circuit is and measure its minute dimensional changes when put in com- widely used to convert the gage’s microstrain into a voltage change pression or tension. Strain gages and strain gage principles are often that can be fed to the input of the analog-to-digital converter used in devices for measuring acceleration, pressure, tension, and (ADC). (See Figure 1.). When all four resistors in the bridge are force. Strain is a dimensionless unit, defined as a change in length absolutely equal, the bridge is perfectly balanced and V = 0. But per unit length. For example, if a 1-m long bar stretches to 1.000002 out when any one or more of the resistors change value by only a m, the strain is defined as 2 microstrains. Strain gages have a fractional amount, the bridge produces a significant, measurable characteristic gage factor, defined as the fractional change in voltage. When used with an instrument, a strain gage replaces one resistance divided by the strain. For example, 2 microstrain applied or more of the resistors in the bridge, and as the strain gage to a gage with gage factor of 2 produces a fractional resistance undergoes dimensional changes (because it is bonded to a test change of (2x2)10-6 = 4x10-6, or 4 . Common gage resistance μΩ specimen), it unbalances the bridge and produces an output values typically range from 120 to 350 , but some devices are as Ω voltage proportional to the strain. low as 30Ω or as high as 3 kΩ. Figure 1: Full-Bridge Circuit R1 R3 + R1 R Motion – + 4 Vexc VOUT R2 R3 – R2 R4 The full-bridge circuit provides the largest output with minimum errors. All four arms of the bridge are active; two are in tension while the two on the opposite side are in compression. www.iotech.com • IOtech • 25971 Cannon Rd • Cleveland, OH 44146 • (440) 439-4091 • Fax (440) 439-4093 • [email protected] Full-Bridge Circuits measurement as long as the input amplifier has high input imped- Although half-bridge and quarter-bridge circuits are often used, ance. For example, an amplifier with a 100-MΩ input impedance the full-bridge circuit is the optimal configuration for strain produces negligible current flow through the measurement leads, gages. It provides the highest sensitivity and the fewest error minimizing voltage drops due to lead resistance. components, and because the full bridge produces the highest output, noise is a less significant factor in the measurement. For Half-Bridge Circuits these reasons, the full bridge is recommended when possible. When physical conditions do not allow mounting a full-bridge gage, a half bridge might fit. Typically, two strain gages are A full bridge contains four strain gages mounted on a test mounted on a test member, and two discrete resistors complete member as shown in Figure 1. Two gages are mounted on the top the bridge. The output voltage is: surface to measure tension and the other two are mounted on the opposite surface to measure compression when the beam is Equation 2: Half-Bridge Output Voltage forced downward. As the member deflects, the two gages in V = V (X/2) tension increase in resistance while the other two decreases, o ex unbalancing the bridge and producing an output proportional Where: Vo = bridge output voltage, V to the displacement. Upward motion reverses the roles of the Vex = excitation voltage applied to the bridge, V strain gages. The bridge output voltage is given by: X = relative change in resistance, ΔR/R Equation 1: Full-Bridge Output Voltage For a large ΔR, half-bridge and quarter-bridge circuits can intro- V = (V )(X) o ex duce an additional nonlinearity error. (See Figure 2.) Also, the readings are not accurate when the temperature coefficients Where: Vo = bridge output voltage, V among the bridge completion resistors and strain gages are V = excitation voltage applied to the bridge, V ex different and the resistances do not change proportionally with X = relative change in resistance, ΔR/R temperature. Furthermore, bridge completion resistors are not usually located near the strain gages, so temperature differences The bridge nulls out potential error factors such as temperature contribute additional errors. In systems with long lead wires, the changes because all four strain gages have the same temperature bridge completion resistors should be attached close to the coefficient and are located in close proximity on the specimen. The gages. However, this may not always be practical due to test resistance of the lead wire does not affect the accuracy of the fixture limitations or other physical conditions. Figure 2: Half-Bridge Circuit Bridge Strain completion resistors gages R R3 1 + R1 Motion – + Vexc VOUT R2 – R4 R2 In a half-bridge circuit, only two arms are active. Two strain gages are on the specimen while the two fixed resistors that complete the bridge are not. 2 www.iotech.com • IOtech • 25971 Cannon Rd • Cleveland, OH 44146 • (440) 439-4091 • Fax (440) 439-4093 • [email protected] Quarter-Bridge Circuits Figure 4: Kelvin Strain-Gage Bridge Circuit A quarter-bridge circuit uses one strain gage and three bridge RL completion resistors. The output voltage is: i≅0 RL Equation 3: Quarter-Bridge Output Voltage ie R Vo = Vex (X/4) 1 R4 Where: V = bridge output voltage, V o RL i≅0 V = excitation voltage applied to the bridge, V ex + + X = relative change in resistance, ΔR/R + Vsense R2 R3 Vexc Vout – – – This arrangement has the smallest output, so noise is a potential R problem. Furthermore, all the error sources and limitations in the half- L bridge circuit apply to the quarter-bridge circuit. (See Figure 3). RL ie RL PROCEDURE Excitation Source The Kelvin bridge circuit uses one pair of wires to provide the Accurate measurements depend on a stable, regulated, and low- excitation voltage directly at the bridge, and another to sense the noise excitation source voltage. A regulated source is necessary excitation voltage. A third pair of wires measures the bridge output because the output voltage of a strain gage is also proportional voltage. This arrangement removes the voltage-drop error in the to the excitation voltage. Therefore, fluctuations in the excita- excitation wires from the measured strain signal. tion voltage produce inaccurate output voltages. most standard circuits, the heat that each gage dissipates is less An ideal data acquisition system provides an excitation source than 100 mW, so it’s not usually a problem. This is especially for each channel, independently adjustable from 1.5 to 10.5 V true when the strain gage is bonded to a material that conducts with a current limit of 100 mA. An excitation voltage, V, used heat quickly, such as metal. However, because most wood, with a strain gage of resistance R, requires a current of I = V/R. plastic, or glass materials do not conduct heat away as rapidly, The resistance of a Wheatstone bridge measured between any use the lowest excitation voltage possible without introducing two symmetrical terminals equals the value of one of the noise problems. Also, heat can become a problem when the resistance arms. For example, four 350-Ohm arms make a 350- strain gages are uncommonly small, or numerous gages occupy Ohm bridge. The load current equals the excitation voltage a limited space. divided by the bridge resistance; in this case, 10 V/350 Ω = 0.029 A = 29 mA. Consider a Kelvin connection for applying the excitation volt- age. Because the excitation leads carry a small current, they drop Heating a correspondingly small voltage; V = I/Rl, which reduces the Resistive heating in strain gages also should be considered voltage reaching the bridge terminals. As illustrated in Figure 4, because the gages respond to temperature as well as stress. In Kelvin connections eliminate this drop with a pair of leads Figure 3: Quarter-Bridge Circuit Bridge Strain completion resistors gages R3 R1 + R1 Motion – + Vexc VOUT – R4 R2 A quarter-bridge circuit uses only one active arm and is the least sensitive of the three types. It is also the most prone to noise and errors. 3 www.iotech.com • IOtech • 25971 Cannon Rd • Cleveland, OH 44146 • (440) 439-4091 • Fax (440) 439-4093 • [email protected] added at the excitation terminals to measure and regulate the Conductors carrying such low level signals are susceptible to bridge voltage. For example, when ie = 50 mA, Rl = 5Ω, and the noise interference and should be shielded. Low-pass filters, combined voltage drop in the two leads is 500 mV, no voltage differential voltage measurements, and signal averaging are also drops in the sense wires. effective techniques for suppressing noise interference. Further- more, instrumentation amplifiers usually condition the ex- An IOtech strain-gage module uses a Kelvin connection to measure tremely low strain-gage signals before feeding them to ADCs. and regulate the voltage at the bridge. It supplies the voltage to the For example, a 10-V full-scale input provides 156 μV of resolu- strain gage with one pair of leads and measures it with another pair tion for a 16-bit ADC. The instrumentation amplifier gain as shown in Figure 5.
Recommended publications
  • MGM's Jawaharlal Nehru Engineering College
    MGM’s Jawaharlal Nehru Engineering College Laboratory Manual Electrical Power Transmission & Distribution For Second Year (EEP) Students Manual made by Prof. P.A. Gulbhile AuthorJNEC,Aurangabad. FORWARD It is my great pleasure to present this laboratory manual for second year EEP engineering students for the subject of Electrical Power Transmission & Distribution. Keeping in view the vast coverage required for visualization of concepts of Electrical Power Transmission with simple language. As a student, many of you may be wondering with some of the questions in your mind regarding the subject and exactly what has been tried is to answer through this manual. Faculty members are also advised that covering these aspects in initial stage itself, will greatly relieved the minfuture as much of the load will be taken care by the enthusiasm energies of the students once they are conceptually clear. H.O.D. LABORATORY MANNUAL CONTENTS This manual is intended for the second year students of Electrical Electronics & Power engineering branch in the subject of Electrical Power Transmission & Distribution. This manual typically contains practical/Lab Sessions related layouts, sheets of different parts in electrical power transmission & distribution system as well as study of different models of transmission line, measurement of parameters of transmission line and various aspects related the subject to enhance understanding. Although, as per the syllabus, only descriptive treatment is prescribed ,we have made the efforts to cover various aspects of electrical Power Transmission & Distribution subject covering types of different transmission lines , their circuit diagrams and phasor diagrams, electrical design of overhead transmission line , electrical design of overhead transmission line, elaborative understandable concepts and conceptual visualization.
    [Show full text]
  • Physics 517/617 Experiment 1 Instrumentation and Resistor Circuits
    Physics 517/617 Experiment 1 Instrumentation and Resistor Circuits 1) Study the operation of the oscilloscope, multimeter, power supplies, and wave generator. For the oscilloscope you should try to understand the function of all the knobs or buttons on the front panel. Some buttons may have several functions and some of the functions will not be relevant for this class. Try to follow the examples in p. 38-42 of the instruction manual. 2) Verify Ohm’s law by measuring and then plotting voltage vs. current for a resistor. Fit your graph(s) to extract the measured resistance. Use a resistor of your choice. Repeat the measurement with a resistor of a much higher value (e.g. 10-100X) than your previous choice. Use a DC power supply for the circuit. 3) Measure the DC resistance of your multimeter (on voltage scale) using a resistor divider, which consists of two resistors in series with one of them being the multimeter resistor. How does your measurement of the multimeter’s resistance compare to the specs of the meter? Note: You do not need to measure the current in this experiment. 4) The RMS (Root Mean Square) value of a voltage (or current) is defined as t 1 2 VRMS = Ú V dt t 0 Show that VRMS = V0 / 2 for a sine wave voltage, V = V0 sinwt. Note: The multimeter only measures the RMS value of a voltage or current. † 5)† This exercise is intended to make you† familiar with some of the very useful functions of the oscilloscope. Send a 1 kHz sine wave with an amplitude of 1 V and DC offset of 1 V into the scope.
    [Show full text]
  • Realization of AC Bridges Using Labview. Details: Explanation
    Realization of AC Bridges Using LabView. Introduction: AC Bridges are used to measure the values of unknown resistance, inductance, and capacitance. Although, AC bridges are believed to be very convenient and provides accurate result of the measurement. The construction of the bridges is very simple. The bridge has four arms, one AC supply source and the balance detector. It works on the principle that the balance ratio of the impedances will give the balance condition to the circuit which is determined by the null detector. The bridges have four arms, two have non-inductive resistance and the other two have inductances with negligible resistance. Software Used: LabView LabVIEW is essentially a graphical programming language (technically it’s a development environment, and the language is “G”, but in common usage it’s a language). Instead of typing words like with C++,Python, or other text-based languages, you place and connect visual objects around your screen. Using LABVIEW the AC bridges can be Solved at an ease with proper visualization and insights. The graphical programming approach enables us to build or Solve AC bridges for different number of applications such as measuring the frequency of the audio signals, filtration of undesirable signals etc. Details: AC Bridges that have been analyzed using LabView are: (1) Maxwell’s Inductance Bridge. (2) DeSauty’s Capacitance Bridge. (3) Wheatstone’s Resistance Bridge. Explanation: Maxwell’s Inductance Bridge: A Maxwell bridge is a modification to a Wheatstone bridge used to measure an unknown inductance in terms of calibrated resistance and inductance or resistance and capacitance. Let us consider w=314, Let us consider E to be a constant (let it be 100V).
    [Show full text]
  • Department of Physics Second Allied Physics Iii
    KUNTHAVAI NAACHIYAR GOVERNMENT ARTS COLLEGE FOR WOMEN, THANJAVUR. DEPARTMENT OF PHYSICS SECOND ALLIED PHYSICS III 18K4CHAP3 UNIT - I 1. Dr. S. SNEGA, DEPARTMENT OF PHYSICS, THANJAVUR. UNIT - II 2. Dr. N. GEETHA, DEPARTMENT OF PHYSICS, THANJAVUR. UNIT-III 3. Ms. D.S. VASANTHI DEPARTMENT OF PHYSICS, THANJAVUR. UNIT-I: CURRENT ELECTRICITY AND NUCLEAR PHYSICS Kirchhoff’s law –Wheatstone’s bridge – Metre Bridge – Carey foster’s bridge – Measurement of specific resistance – Potentiometer – Calibration of low range voltmeter Nucleus – Nuclear Size – Charge – Mass and Spin – Shell Model – Nuclear fission and fusion – Liquid drop model – Binding energy – Mass defect KIRCHHOFF’S RULES Ohm’s law is useful only for simple circuits. For more complex circuits, Kirchhoff ’s rules can be used to find current and voltage. There are two generalized rules: i) Kirchhoff ’s current rule ii) Kirchhoff ’s voltage rule. Kirchhoff’s first rule (Current rule or Junction rule) It states that the algebraic sum of the currents at any junction of a circuit is zero. It is a statement of conservation of electric charge. All charges that enter a given junction in a circuit must leave that junction since charge cannot build up or disappear at a junction. Current entering the junction is taken as positive and current leaving the junction is taken as negative. I1 + I2 – I3 – I4 – I5 = 0 (or) I1 + I2 = I3 + I4 + I5 Kirchhoff’s Second rule (Voltage rule or Loop rule) It states that in a closed circuit the algebraic sum of the products of the current and resistance of each part of the circuit is equal to the total emf included in the circuit.
    [Show full text]
  • I INVESTIGATION of METHODS for DETECTING NEEDLE INSERTION INTO BLOOD VESSELS by Ehsan Qaium BS in Mechanical Engineering, Virgi
    INVESTIGATION OF METHODS FOR DETECTING NEEDLE INSERTION INTO BLOOD VESSELS by Ehsan Qaium BS in Mechanical Engineering, Virginia Tech, 2012 Submitted to the Graduate Faculty of the Swanson School of Engineering in partial fulfillment of the requirements for the degree of Master of Science in Mechanical Engineering University of Pittsburgh 2018 i UNIVERSITY OF PITTSBURGH SWANSON SCHOOL OF ENGINEERING This thesis was presented by Ehsan Bin Qaium It was defended on April 19,2018 and approved by William Clark, PhD, Professor Jeffrey Vipperman, PhD, Professor Cameron Dezfulian, MD, Associate Professor Thesis Advisor: William Clark, PhD, Professor ii Copyright © by Ehsan Qaium 2018 iii INVESTIGATION OF METHODS FOR DETECTING NEEDLE INSERTION INTO BLOOD VESSELS Ehsan Qaium, M.S. University of Pittsburgh, 2018 Peripheral intravenous IV (pIV) placement is the mainstay for providing therapies in modern medicine. Although common, approximately 107 million difficult pIV placements each year require multiple attempts to establish IV access. Delays in establishing IV access lead to increased patient pain, delayed administration of life saving medicine, and increased cost to the institution. Current solutions involve using visual vein finders, ultrasounds and a central line if peripheral IV insertion attempts fail. The objective of this study was to investigate methods by which entry into a blood vessel could be detected, and to design and test a novel medical device that increases the likelihood of successful pIV placement on the first attempt. Two types of measurement methods (static and transient) were investigated in this study. Static measurement involved measurements performed with a multimeter and a Wheatstone bridge. The multimeter measurement was unsuccessful due to the effect of polarization.
    [Show full text]
  • Basic Bridge Circuits
    AN117 Dataforth Corporation Page 1 of 6 DID YOU KNOW ? Samuel Hunter Christie (1784-1865) was born in London the son of James Christie, who founded Christie's Fine Art Auctioneers. Samuel studied mathematics at Trinity College and, upon graduation, taught mathematics at the Royal Military Academy for almost 50 years. Christie made many contributions to magnetic science, such as the dependence of magnetic forces on temperature and solar ray effects on terrestrial magnetism. In 1833, he published a paper on the magneto-electric conductivity of various metals, illustrating how wire conductance varies inversely with length and directly as the square of wire diameter. Embedded in this paper was the description of a circuit used to measure and compare wire conductance. Charles Wheatstone (an English physicist and inventor) recognized the value of Christie's circuit; he was the first to put this circuit, which bears his name, to extensive use and to develop many significant applications for it. To this day (162 years later), the Wheatstone bridge remains the most sensitive and accurate method for precisely measuring resistance values. Samuel Christie never got recognition for his bridge circuit. At a royalty of five cents for every bridge circuit used, imagine what Christie’s original circuit invention would be worth today. Basic Bridge Circuits Preamble Analytical investigations throughout this document focus Bridge circuits have been in use for well over 150 years. on the R-ohm type bridge, which means all bridge To date, the bridge is still the most economical circuit resistors are “R” ohms when not exposed to the field technique for accurately measuring resistance.
    [Show full text]
  • Shunt Calibration of Strain Gage Instrumentation
    MICRO-MEASUREMENTS Strain Gages and Instruments Tech Note TN-514 Shunt Calibration of Strain Gage Instrumentation I. Introduction the gage. Furthermore, the electrical contacts for inserting the resistor can introduce a significant uncertainty in The need for calibration arises frequently in the use the resistance change. On the other hand, decreasing of strain gage instrumentation. Periodic calibration is the resistance of a bridge arm by shunting with a larger required, of course, to assure the accuracy and/or linearity resistor offers a simple, potentially accurate means of of the instrument itself. More often, calibration is necessary simulating the action of a strain gage. This method, known to scale the instrument sensitivity (by adjusting gage factor as shunt calibration, places no particularly severe tolerance or gain) in order that the registered output correspond requirements on the shunting resistor, and is relatively conveniently and accurately to some predetermined input. insensitive to modest variations in contact resistance. It is An example of the latter situation occurs when a strain also more versatile in application and generally simpler to gage installation is remote from the instrument, with implement. measurable signal attenuation due to leadwire resistance. In this case, calibration is used to adjust the sensitivity Because of its numerous advantages, shunt calibration is of the instrument so that it properly registers the strain the normal procedure for verifying or setting the output signal produced by the gage. Calibration is also used to set of a strain gage instrument relative to a predetermined the output of any auxiliary indicating or recording device mechanical input at the sensor.
    [Show full text]
  • CHAPTER-7 Underground Cables
    CHAPTER-7 Underground Cables Construction of Underground Cables: Fig.shows the general Construction of Underground Cables. The various parts are : Cores or Conductors: A cable may have one or more than one core (conductor) depending upon the type of service for which it is intended. For instance, the 3-conductor cable shown in Fig. 11.1 is used for 3- phase service. The conductors are made of tinned copper or aluminium and are usually stranded in order to provide flexibility to the cable. Insulation: Each core or conductor is provided with a suitable thickness of insulation, the thickness of layer depending upon the voltage to be withstood by the cable. The commonly used materials for insulation are impregnated paper, varnished cambric or rubber mineral compound. Metallic sheath: In order to protect the cable from moisture, Conductor gases or other damaging liquids (acids or alkalies) in the soil and atmosphere, a metallic sheath of lead or aluminium is provided over the insulation as shown in Fig. 11.1 Bedding: Over the metallic sheath is applied a layer of bedding which consists of a fibrous material like jute or hessian tape. The purpose of bedding is to protect the metallic sheath against corrosion and from mechanical injury due to armouring. Armouring: Over the bedding, armouring is provided which consists of one or two layers of galvanised steel wire or steel tape. Its purpose is to protect the cable from mechanical injury while laying it and during the course of handling. Armouring may not be done in the case of some cables. Serving: In order to protect armouring from atmospheric conditions, a layer of fibrous material (like jute) similar to bedding is provided over the armouring.
    [Show full text]
  • Applying the Wheatstone Bridge Circuit by Karl Hoffmann
    Applying the Wheatstone Bridge Circuit by Karl Hoffmann W1569-1.0 en Applying the Wheatstone Bridge Circuit by Karl Hoffmann Contents: 1 Introduction ...................................................................................................................................1 2 Elementary circuits with strain gages..........................................................................................5 2.1 Measurements on a tension bar..............................................................................................6 2.2 Measurements on a bending beam.........................................................................................7 2.3 Measurements on a twisted shaft ...........................................................................................8 3 Analysis and compensation of superimposed stresses, e.g. tension bar with bending moment .........................................................................................................................................10 4 Compensation of interference effects, especially temperature effects....................................14 5 Compensation of the influence of lead resistances. ..................................................................17 6 Table on different circuit configurations ..................................................................................20 7 Some remarks on the limits of the compensation of interference effects ...............................22 8 The linearity error of the Wheatstone bridge circuit...............................................................23
    [Show full text]
  • A Conversion of Wheatstone Bridge to Current-Loop Signal Conditioning for Strain Gages
    https://ntrs.nasa.gov/search.jsp?R=19970024858 2020-06-16T02:20:48+00:00Z View metadata, citation and similar papers at core.ac.uk brought to you by CORE provided by NASA Technical Reports Server NASA Technical Memorandum 104309 A Conversion of Wheatstone Bridge to Current-Loop Signal Conditioning For Strain Gages Karl F. Anderson NASA Dryden Flight Research Center Edwards, California 1995 National Aeronautics and Space Administration Dryden Flight Research Center Edwards, California 93523-0273 A CONVERSION OF WHEATSTONE BRIDGE TO CURRENT-LOOP SIGNAL CONDITIONING FOR STRAIN GAGES Karl F. Anderson* NASA Dryden Flight Research Center P. O. Box 273 Edwards, California 93523-0273 ABSTRACT Alcal current change through the gage during calibration, A Current loop circuitry replaced Wheatstone bridge cir- gage current, A cuitry to signal-condition strain gage transducers in more than 350 data channels for two different test programs at Iref reference resistance current, A NASA Dryden Flight Research Center. The uncorrected J circuit board jack test data from current loop circuitry had a lower noise level than data from comparable Wheatstone bridge circuitry, JP circuit board jumper were linear with respect to gage-resistance change, and OUT were uninfluenced by varying lead-wire resistance. The output current loop channels were easier for the technicians to set OVP over-voltage protection up, verify, and operate than equivalent Wheatstone bridge R resistors on the circuit card channels. Design choices and circuit details are presented in this
    [Show full text]
  • A Low-Voltage Current-Mode Wheatstone Bridge Using CMOS Transistors
    World Academy of Science, Engineering and Technology International Journal of Electronics and Communication Engineering Vol:5, No:10, 2011 A Low-Voltage Current-Mode Wheatstone Bridge using CMOS Transistors Ebrahim Farshidi However, the need for low-voltage/low-power CMOS analog Abstract—This paper presents a new circuit arrangement for a circuits is growing due to the downscaling of CMOS current-mode Wheatstone bridge that is suitable for low-voltage processes and also, in many situations, particularly in the integrated circuits implementation. Compared to the other proposed mixed analog-digital systems, it is desirable to implement the circuits, this circuit features severe reduction of the elements number, circuits in the MOS technology [12]. low supply voltage (1V) and low power consumption (<350uW). In addition, the circuit has favorable nonlinearity error (<0.35%), In this paper, a new CMWB design to overcome the above operate with multiple sensors and works by single supply voltage. problems is presented [13]. The circuit is designed in CMOS The circuit employs MOSFET transistors, so it can be used for technology, which exhibits flexible characteristics with standard CMOS fabrication. Simulation results by HSPICE show respect to other CM or VM devices, Its main features are high performance of the circuit and confirm the validity of the reducing circuit elements, area efficient implementation, proposed design technique. operation in single supply voltage, current output signal and possibility of low-voltage and low-power due to the fact that Keywords—Wheatstone bridge, current-mode, low-voltage, MOS transistors are employed. MOS. The paper is organized as follows. In section 2, basic I.
    [Show full text]
  • Precision Resistance Measurements
    ECGR 2155 Instrumentation and Networks Laboratory UNIVERSITY OF NORTH CAROLINA AT CHARLOTTE Department of Electrical and Computer Engineering EXPERIMENT 6 – PRECISION RESISTANCE MEASUREMENTS OBJECTIVES This experiment introduces four different methods for making precision resistance measurements: Benchtop multimeter 2-wire, Benchtop multimeter 4-wire (Kelvin), Wheatstone bridge, and LCR meter MATERIALS/EQUIPMENT NEEDED Agilent 34461A Digital Multimeter WINSCO GS-470 Wheatstone Bridge SM-1102A Galvanometer B&K LCR meter Resistors: 1Ω, 10Ω, 1kΩ INTRODUCTION It is sometimes necessary to make resistance measurements – particularly of very small resistances – with precision greater than that which can be obtained with a hand held multimeter. This is frequently encountered with sensors. For example, the resistance of a strain gauge must be measured very accurately in order to determine deflection of solid materials under mechanical load. Some methods for making precision resistance measurements are discussed below; 2-wire Resistance Measurements: Multimeters function as “Ohmmeters” when operated in resistance measurement mode. Inexpensive multimeters measure resistance by applying known voltage to the unknown resistance, measuring the resulting current, and then applying Ohm’s law. More accurate meters apply a known current and measure the resulting voltage. In either case, meter test leads can introduce substantial error into low resistance measurements. Because the wire leads have small, but finite resistance, voltage division occurs between the resistance being measured, and the resistance of the leads. Figure 6-1 shows a simplified schematic diagram of a meter connected to an unknown resistance (Runknown) by two wire leads of resistance Rlead. Applying Ohm’s law to the meter circuit: V =(R +=2 RR) I unknown lead measured When Rlead<<Runknown, the measurement error is small.
    [Show full text]