Appendix EQ, Test Equipment Recommendations and Calibration
Appendix EQ Test Equipment Recommendations & Calibration a) Clothes Washer
Figure EQ-1, The Clothes Washer
Type: Fisher and Paykal top loading clothes washer Model: GWL 11 US Serial: UKN627345 Input Voltage: 110 ~ 120 V, 60 Hz, 7 Amps
Created by Phie Theng Lee 37 Michigan Technological University Appendix EQ, Test Equipment Recommendations and Calibration b) Clothes Dryer Type: Kenmore front loading clothes dryer Model: prototype Input Voltage: 220 ~ 240 V, 60 Hz
The following list contains the specifications and calibration factors of the test equipment. The calibration factors were verified using known sources and were calculated according to the following:
Calfac = Refval / Outvol in Engineering Units per Volt
Calfac Transducer sensitivity
Refval reference value of the calibration source
Outvol output voltage of the transducer c) Apparatus to use with Clothes Washer PCB Laser Tachometer Measurement Range up to 30,000 rpm with 1 pulse per revolution Operating Range 90 degree incidence (60 inches apart) 60 degree incidence (15 inches apart) Excitation Voltage 18 to 30 VDC Temperature Range 14 ~ 122 degree F Mounting Thread 5/8-18 UNF 2A Weight 0.85 oz
PCB Tri-axial accelerometer Model # 356A15, Serial # 33675, 33363 and 33664 Sensitivity 100 mV/g Measurement Range +/- 50 g peak Output Bias Voltage 8 ~ 12 VDC Frequency Range 2 ~ 5000 Hz +/- 5% or 1.4 ~ 6500 Hz +/- 10% Temperature Range -65 ~ 250 degree F Resonant Frequency >= 25 kHz Settling Time < 5 sec within 10% of bias Spectral Noise 5 micro g / Hz @ 100 Hz Mounting Thread 10-32 female Weight 0.37 oz Calibration Factors Serial # 33675 (at bottom of wash tub), X = 85.071 g/V, Y = 86.785 g/V, Z = 83.799 g/V Serial # 33363 (at top of wash tub), X = 168.45 g/V, Y = 170.96 g/V, Z = 165.59 g/V Serial # 33664 (at the cabinet), X = 171.41 g/V, Y = 170.47 g/V, Z = 173.72 g/V
Created by Phie Theng Lee 38 Michigan Technological University Appendix EQ, Test Equipment Recommendations and Calibration
PCB Load Cell Model # 208A03, Sensitivity 10 mV/lb Measurement Range 500 lb compression or tension Output Bias Voltage 8 ~ 14 VDC Frequency Range 0.0003 ~ 36000 Hz - 5% Temperature Range -65 ~ 250 degree F Spectral Noise 38.7 micro lb / Hz @ 100 Hz Mounting thread 10-32 female Weight 0.80 oz Calibration Factor Serial # 5316 126.256 lbf/V Serial # 1909 115.422 lbf/V
Model # 208A02, Serial # 9212 Sensitivity 50 mV/lb Measurement Range 100 lb compression or tension Output Bias Voltage 8 ~ 14 VDC Frequency Range 0.001 ~ 36000 Hz - 5% Temperature Range -65 ~ 250 degree F Spectral Noise 9.6 micro lb / Hz @ 100 Hz Mounting thread 10-32 female Weight 0.80 oz Calibration Factor 27.894 lbf/V
Model # 208M164, Serial # 17199 Sensitivity Measurement Range Output Bias Voltage Frequency Range Temperature Range Spectral Noise 5 micro g / Hz @ 100 Hz Mounting thread 10-32 female Weight 0.37 oz Calibration Factor 32.295 lbf/V
PCB Pressure Sensor Model # 1505M13, Serial # 163 (at Hot Water Inlet) Sensitivity 0.008 V/psi Measurement Range 0 ~ 500 psi Test Sensor Output 1 ~ 5 V Zero Output 0.988 V Zero Output Error 0.3% FS Output @ Full Scale 5.007 V Full Scale Output 4.019 V Full Scale Output Error 0.5% FS
Created by Phie Theng Lee 39 Michigan Technological University Appendix EQ, Test Equipment Recommendations and Calibration
Calibration Factor 100 psi/V
Model # 1505M14, Serial # 164 (at Cold Water Inlet) Sensitivity 0.008 V/psi Measurement Range 0 ~ 500 psi Test Sensor Output 1 ~ 5 V Zero Output 0.994 V Zero Output Error 0.2% FS Output @ Full Scale 5.011 V Full Scale Output 4.017 V Full Scale Output Error 0.4% FS Calibration Factor 100 psi/V
Omega Fast Response T-type Thermocouples with Self-Adhesive Backing Model # SA1-T-72 Maximum Temperature 350 degree F Minimum Temperature -75 degree F Wire 30 AWG PFA Teflon coated
Omega T-type signal conditioner Model # CCT-24-0/200C, Linearized, 0.3% accuracy Outputs 0 ~ 10 V Overvoltage 75 Vdc Error 0.05 degree C @ 25 degree C ambient typical Measurement Range 0 ~ 200 degree C d) Apparatus to use with Clothes Dryer Omega Fast Response T-type Thermocouples with Self-Adhesive Backing Model # SA1-T-72 Maximum Temperature 350 degree F Minimum Temperature -75 degree F Wire 30 AWG PFA Teflon coated
Omega T-type signal conditioner Model # CCT-24-0/200C, Linearized, 0.3% accuracy Outputs 0 ~ 10 V Overvoltage 75 Vdc Error 0.05 degree C @ 25 degree C ambient typical Measurement Range 0 ~ 200 degree C
Created by Phie Theng Lee 40 Michigan Technological University Appendix EQ, Test Equipment Recommendations and Calibration e) Data Acquisition Device
Figure EQ-2, HP model 35655A (front end) capable of connecting or acquiring data up to 32 channels.
Created by Phie Theng Lee 41 Michigan Technological University