Polarimeter Buyer's Guide

Total Page:16

File Type:pdf, Size:1020Kb

Polarimeter Buyer's Guide Polarimeter Buyer’s Guide Understanding Polarimeters Selecting the Right Polarimeter Illumination Method 3 Polarimeters use different lamp sources for transmitting Measurement Range and Accuracy Polarimeters are precise optical instruments that measure purity polarized light. Some models employ halogen lamps which 1 Purchasing a new polarimeter is a major decision. Choose the and concentration through the optical activity exhibited in have a short life span, generate heat, and require servicing model that offers the best measurement range and accuracy inorganic and organic compounds. Compounds are considered while others have LED illumination. LED illumination does not to meet present and future application needs. Be aware to be optically active if linearly polarized light is rotated when require maintenance or generate heat and is environmentally that many polarimeters offer a very limited measurement passing through them. This optical rotation in general means friendly and energy efficient. All Reichert polarimeters feature range that may limit readings. Also, most importantly, there that the polarization of the direction of light is rotated at LED illumination. a specific angle proportionate to the concentration of the are many polarimeters on the market that have a non-linear accuracy specification. In other words, within the limited optically active substance being tested. Maintenance measurement range of these instruments, the accuracy Optical rotation is measured in degrees of angle. At point 4 Operating costs are important to consider when evaluating specification can degrade to different levels across that range. zero, a polarizer and analyzer are set in an angle of 90 degrees instruments. Many companies offer expensive service plans. toward each other, which means that no light reaches the Reichert builds robust instruments that provide reliable detector (0% transmission). When an optically active substance Be sure the polarimeter has an accurate linear response over the entire measurement range. service for decades without requiring repair contracts. Backed is then placed in the measuring tube, the transmission will rise by longstanding knowledge in the analytical instruments as a result of the plane polarization. To measure the angle, the industry, Reichert polarimeters are manufactured for reliability analyzer is rotated up to the point where the transmission of the Competitor Reichert and longevity. detector is again at minimum. Polarimetry is a mature technique and today’s automatic Sample Tube Lengths, Material Composition instrumentation provides features and benefits that purely 5 Polarimetric applications utilize different sample tube lengths. manual-optical systems do not. Busy labs processing multiple Longer tube lengths provide greater accuracy. But sometimes samples have the option of automated data capture, variable shorter tube lengths are needed if samples are limited in wavelength and temperature with circular polarimeters having Accuracy range quantity or very expensive. Tube lengths range from 10 to Accuracy range 0 0 0.005 accuracy and 0.001 resolution. These instruments are 0 0 +360 to-360 +90 to-90 200mm depending on the application. Material composition suited for quality control, process control, and research for of tubes is also important. Many sample types work well demanding applications in many industries including beverage, with a stainless steel tube. Samples that are corrosive are cosmetic, chemical, food, essential oils and pharmaceutical. better suited for glass. Reichert provides a wide range of tube lengths and materials to meet all needs. Which Polarimeter is Right for Your Lab? Competitor’s polarimeters Reichert polarimeters are full With this overview, Reichert makes it easy to determine accuracy can degrade over a range instruments, with an Software limited measurement range. accurate linear response over 6 which polarimeter suits your needs. Identify answers to the the entire 3600 degree range. When selecting a polarimeter, ensure the software features steps below to determine the specific requirements of your meet your needs. Reichert offers an intuitive, simple interface applications, in your industry, and with your budget. And then, for quickly and accurately attaining measurements. The simply pick the appropriate Reichert polarimeter from the chart 2 Measurement Response Time measurement screens are simplified and Reichert software on the back. In today’s labs, speed is critical - with fewer resources processing provides custom scales and methods, and is compliant with more samples, technicians wait time must be minimized. 21CFR Part 11 compliance. Reichert offers the fastest measurements for a given accuracy. Reichert Polarimeters Polar1 Polar2 Polar3 Technical Specifications Polar1 - #10836-576 Polar2 - #10836-572 Polar3 - 10836-568 Measurement Scales Optical Rotation, Specific Rotation,0 Z International Sugar Scale, % Concentration User Definable Scales 7 1,000 1,000 Measuring Range ± 3600 / ± 2590Z ±900 / ±2590Z ± 3600 / ± 2590Z Resolution 0.010 / 0.020Z 0.0010 / 0.010Z 0.0010 / 0.010Z Accuracy ± 0.010 / ± 0.030Z ± 0.0050 / ± 0.0150Z Reproducibility ± 0.010 / ± 0.030Z ± 0.0050 / ± 0.0150Z Sensitivity Up to OD 2 Up to OD 5 Wavelengths 589nm standard 589nm standard, others on request Response Time 6 to 8 sec. over the entire measuring range ≤ 4 sec. over entire measuring range Tube lengths 10 to 200 mm length Up to 100mm length 10 to 200 mm length Tube Materials Glass, acid-proof stainless steel, stainless steel with temperature sensor Temperature Interface Temperature measurement NTC Sensor Accuracy ± 0.030C ±0.10C ± 0.030C Temperature range 00C to 990C 100C to 400C 00C to 990C For more information Resolution 0.010C contact your VWR representative or visit Other Information VWR.com® Light Source LED Display LCD display, monochromatic 4.3" color touchscreen 7" color touchscreen Universal Display (optional) Universal Display (optional) Interfaces RS232 (2x), Parallel (1x) RS232, USB, Ethernet RS232 (1x), USB A (4x), USB B (1x), Ethernet Dimensions 25.5 x 12.4 x 6.2 inches 17 x 12 x 6.3 inches 28.75 x 14.5 x 6.3 inches (650 x 315 x 160mm) (430 x 300 x 160mm) (730 x 370 x 160 mm) Weight 26lbs. (11.8kg) 22lbs. (9.97kg) 40 lbs. (18.3kg).
Recommended publications
  • Chirality and Enantiomers
    16.11.2010 Bioanalytics Part 5 12.11.2010 Chirality and Enantiomers • Definition: Optical activity • Properties of enantiomers • Methods: – Polarimetry – Circular dichroism • Nomenclature • Separation of enantiomers • Determination of enantiomeric excess (ee) Determination of absolute configuration • Enantiomeric ratio (E‐value) 1 16.11.2010 Definitions Enantiomers: the two mirror images of a molecule Chirality: non‐superimposable mirror‐images •Depends on the symmetry of a molecule •Point‐symmetry: asymmetric C, Si, S, P‐atoms •Helical structures (protein α‐helix) Quarz crystals snail‐shell amino acids Properties of Enantiomers – Chemical identical – Identical UV, IR, NMR‐Spectra – Differences: • Absorption and refraction of circular polarized light is different – Polarimetry, CD‐spectroscopy • Interaction with other chiral molecules/surfaces is different – Separation of enantiomers on chiral columns (GC, HPLC) 2 16.11.2010 Chiral compounds show optical activity A polarimeter is a device which measures the angle of rotation by passing polarized light through an „optical active“ (chiral) substance. Interaction of light and matter If light enters matter, its intensity (amplitude), polarization, velocity, wavelength, etc. may alter. The two basic phenomena of the interaction of light and matter are absorption (or extinction) and a decrease in velocity. 3 16.11.2010 Interaction of light and matter Absorption means that the intensity (amplitude) of light decreases in matter because matter absorbs a part of the light. (Intensity is the square of amplitude.) Interaction of light and matter The decrease in velocity (i.e. the slowdown) of light in matter is caused by the fact that all materials (even materials that do not absorb light at all) have a refraction index, which means that the velocity of light is smaller in them than in vacuum.
    [Show full text]
  • Polarimetric Detection of Enantioselective Adsorption by Chiral Au Nanoparticles – Effects of Temperature, Wavelength and Size
    Nanomaterials and Nanotechnology ARTICLE Polarimetric Detection of Enantioselective Adsorption by Chiral Au Nanoparticles – Effects of Temperature, Wavelength and Size Invited Article Nisha Shukla2, Nathaniel Ondeck1, Nathan Khosla1, Steven Klara1, Alexander Petti1 and Andrew Gellman1* 1 Department of Chemical Engineering, Carnegie Mellon University, Pittsburgh, PA, USA 2 Institute of Complex Engineered Systems, Carnegie Mellon University, Pittsburgh, PA, USA * Corresponding author(s) E-mail: [email protected] Received 01 November 2014; Accepted 07 January 2015 DOI: 10.5772/60109 © 2015 The Author(s). Licensee InTech. This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/3.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. Abstract 1. Introduction R- and S-propylene oxide (PO) have been shown to interact The importance of chirality arises from the need of the enantiospecifically with the chiral surfaces of Au nanopar‐ pharmaceutical industry, and other producers of bioactive ticles (NPs) modified with D- or L-cysteine (cys). This compounds, for chemical processes that are enantioselec‐ enantiospecific interaction has been detected using optical tive and thereby produce enantiomerically pure chiral polarimetry measurements made on solutions of the D- or products. [1-4] This is one of the most challenging forms of L-cys modified Au (cys/Au) NPs during addition of chemical synthesis and is critical to the performance of racemic PO. The selective adsorption of one enantiomer of bioactive chiral products. Ingestion of the two enantiomers the PO onto the cys/Au NP surfaces results in a net rotation of a chiral compound by a living organism can result in of light during addition of the racemic PO to the solution.
    [Show full text]
  • Chapter 1: Optical Activity and Polarimetry
    Chapter 1: Optical Activity and Polarimetry Chiral, non-racemic compounds have the ability to rotate polarized light. This phenomenon is known as circular birefringence, and is where the phrase “optically active” comes from. The specific rotation of a material is an intrinsic, constant value at a given temperature and wavelength of light; specific rotation can be found by using polarimetry. 1.1 Optical Activity Normal light contains a mixture of all vibrational directions that are perpendicular to the direction of travel. We can use polarized filters to block most of the components of normal light and only allow through one direction of polarization (Figure 1-1). Figure 1-1: Polarized light vibrates only in one plane, while normal light vibrates in all planes. When we use two filters in series, the first one is called the polarizer (because it converts the light to polarized light) and the second is called the analyzer (because it allows you to analyze the light you just polarized). If the polarizer and analyzer are parallel to each other, the polarized light gets through fine. But if they’re at right angles, no light can pass through the analyzer, as shown in Figure 1-2. Figure 1-2: Whether polarized light can pass through the analyzer depends on its orientation to the polarizer. Since optical activity gives a compound the ability to rotate plane-polarized light, we can put a sample of the compound in between the polarizer and analyzer. This will rotate the light, allowing some of it to get through the analyzer even when the filters are at right angles (Figure 1-3).
    [Show full text]
  • Inverted Sugar Syrup - Wikipedia
    Inverted sugar syrup - Wikipedia https://en.wikipedia.org/wiki/Inverted_sugar_syrup Inverted sugar syrup Inverted sugar syrup (also called invert syrup , or simply invert Invert sugar sugar ) is an edible mixture of two simple sugars—glucose and fructose— that is made by heating sucrose (table sugar) with water. [1] It is thought to be sweeter than table sugar, [2] and foods that contain it retain moisture and crystallize less easily. Bakers, who call it invert syrup , may use it more than other sweeteners.[3] Though inverted sugar syrup can be made by heating table sugar in water alone, the reaction can be sped up by adding lemon juice, cream of tartar or other catalysts often without changing the flavor noticeably. Identifiers The mixture of the two simple sugars is formed by a process of hydrolysis of sucrose. This mixture has the opposite direction of optical rotation as the CAS Number 8013-17-0 (http://ww original sugar, which is why it is called an invert sugar. w.commonchemistry. org/ChemicalDetail.a spx?ref=8013-17-0) Contents ChEMBL ChEMBL1201647 (ht Chemistry tps://www.ebi.ac.uk/c Optical rotation Definition and measurement hembldb/index.php/c Definition of the inversion point ompound/inspect/Ch Chirality and specific rotation EMBL1201647) Effects of water ChemSpider none Mixtures in general Fully hydrolyzed sucrose ECHA 100.029.446 (https:// Partly hydrolyzed sucrose InfoCard echa.europa.eu/subs Monitoring reaction progress tance-information/-/s Production ubstanceinfo/100.02 9.446) Shelf life In other foods and products PubChem 21924868 (https://pu CID See also bchem.ncbi.nlm.nih.
    [Show full text]
  • Testing the Chirality of Glucose Using a Homemade Polarimeter
    CALIFORNIA STATE SCIENCE FAIR 2016 PROJECT SUMMARY Name(s) Project Number Greta M. van den Bergh 36566 Project Title Testing the Chirality of Glucose Using a Homemade Polarimeter Abstract Objectives/Goals My objective is to come up with a way to test the Chirality of Glucose. Knowing the Chirality of a molecule is important when developing substances that will go into the body since the body is a chiral environment. Can a homemade polarimeter be used as a first step in measuring the chirality of Glucose, a known Chiral molecule? Also, will the angle of rotation of plane polarized light increase with a greater concentration of Glucose or a longer path length through the solution? Methods/Materials Polarizer, Analyzer, Light Corn Syrup (Glucose) in four concentrations (0%,25%, 50%, 100%), Water, Four PVC tubes glued to a pane of glass to put the solutions in, iPad as a light source, camera, Analyzer stand, Lazy Susan, 16'x16' butcher block paper, Ruler, Protractor. In this experiment I use an iPad for a constant light source, two polarizers, one to polarize the light before entering the sugar solutions, and the other to analyze the outgoing light. The polarizer is rotated relative to the analyzer in order to get the minimum light through the system. This is where the polarizer and analyzer are at 90deg to each other. This technique was used for each solution concentration to see the angle differences where the minimum light occurred. This will give us the degree of polarization rotation. The 0% solution is our reference point, because water is non-chiral.
    [Show full text]
  • Optical Polarimetry
    Optical Polarimetry In a typical polarimetry experiment, monochromatic light is passed through the sample. A sodium lamp is usually used as the light source and the wavelength of its D line is 589.3 nm. The light provided by the source is not polarized so its electromagnetic waves oscillate in all planes perpendicular to the transverse axis. After passing through a polarizing lens (prism), the only light remaining is oscillating in one plane (plane polarized), whose angle is determined by the angle of the lens itself. This light is then passed through a solution of an optically active compound, which results in a rotation of the plane of oscillation. A second polarizing lens (prism) is used in conjunction with a detector to find the angle of rotation. The magnitude of the rotation is not only determined by the intrinsic properties of the molecule, but also on the concentration, path length, and wavelength of light. These parameters should be familiar from use of a UV/Vis spectrometer. In order to standardize the optical rotations reported in the literature, a parameter has been defined as the specific rotation [α] : λ The parameters are α the rotation measured by the polarimeter in degrees meas λ the wavelength used to measure the rotation in nanometers l the path length in decimeters (usually this is 2 dm, note that 1 dm = 0.1 m or 10 cm) c the concentration measured in g/mL Normally [α] values are quoted in the literature at 20 ºC. There is a slight temperature λ dependence on optical rotation, which can be corrected using: [α] (20 ºC) = [α] (T) [1 + 0.0001(T - 20)] λ λ The percent optical purity (x) of a material is the ratio of the measured specific rotation, [α] , to the standard pure specific rotation, [α] , where x = [α] /[α] *100.
    [Show full text]
  • Chiral and Achiral Enantiomeric Separation of (±)-Alprenolol
    Open Chem., 2019; 17: 429–437 Research Article M.M. López Guerrero*, A. Navas Díaz, F. García Sánchez, H. Corrall Chiral and Achiral Enantiomeric Separation of (±)-Alprenolol https://doi.org/10.1515/chem-2019-0049 received September 25, 2018; accepted March 30, 2019. arrythmia [2,3,4]. Numerous substances found in nature, many of them pharmaceutical medicines such as Abstract: The chiral separation of enantiomers is crucial Alprenolol, are chiral [5]. Typically, only one enantiomer for pharmacovigilance within drug discovery. Although a is therapeutically active, and the other is either inactive large number of prescribed medications are marketed as or toxic. This was highlighted in the media by the 1961 pure enantiomers, this is not always the case and many Thalidomide drug scandal, which demonstrated a are in fact racemic mixtures. Drug scandals, such as that further need for research into the separation of these two of Thalidomide in 1961, provide a clear example of the enantiomeric forms [6,7]. It has been verified that for beta social and economic repercussions that can be caused blockers, including Alprenolol, the pharmacokinetics by negligence of these chiral compounds. Two high differ greatly between the two enantiomers [8]. For a class performance liquid chromatography (HPLC) methods of drug that is marketed as a “racemic dose”, only the are presented to determine, separate and quantitate a (-)-Alprenolol accounts for the majority of the beta blocking commonly prescribed chiral beta blocker, (-)-Alprenolol. activity [9]. A large number of pharmaceutical products The first method utilises a chiral column to physically are sold under the ruse of being pure enantiomers but are separate the two enantiomers of Alprenolol in 25 minutes, still in fact racemic mixtures [10].
    [Show full text]
  • Optical Rotation
    photoelastic modulators Polarimetry: Optical Rotation APPLICATION NOTE By Dr. Theodore (Ted) Oakberg on Dec. 13, 2004 Rev. by Dr. Baoliang (Bob) Wang on Jan. 18, 2010 One simple setup using a PEM for measuring optical rotation is shown in Figure 1. PEM 1f Reference Computer When linear polarized light passes through an “optically V active” or “chiral” material, a rotation of the plane of Lock-in polarization of the light may occur. This phenomenon is called AC DC Optical Rotation. The measurement of optical rotation of a Light Source Detector sample is called “Polarimetry” by chemists.1 Optically active 0° Sample 0° 45° Signal materials include chiral organic molecules such as dextrose Polarizer PEM Polarizer Conditioning Unit and crystals with asymmetric structures such as quartz. Figure 1. Optical Bench Set-up for Measurement of Optical Rotation Optical rotation can be interpreted by physicists as the result of circular birefringence. Circular birefringence is the If the light source is a laser, the optical setup is particularly difference in refractive indices for right and left circularly simple since no collimating and focusing lenses are required. polarized light. Linearly polarized light can be represented by The fi rst polarizer after the light source should be mounted in the linear combination of right and left circularly polarized light. a precision rotator to allow precise alignment with the PEM When a linearly polarized light beam enters an optically active retardation axis. For this confi guration (with no sample in place) sample, the circular birefringence of the sample produces there is no AC signal at the detector but there is a DC signal.
    [Show full text]
  • Optical Rotation
    Polarimetry Polarimetry is an instrumental analytical method using rotation of polarized light by some substances as a measure of their concentration in a solution. The instrument used is called a polarimeter. When it is adapted for measuring quality of sugar the name saccharimeter is used. In both instruments it is the rotation of polarized light by a substance in a solution which is measured. Usually, it is only one instrument which has two interchangeable scales, one labelled in angular degrees 0, the other in units 0Z, named International Sugar Scale (I.S.S.). Polarimeter Figure 1 shows a principle of a polarimeter set up and its main components together with their function. Unpolarized light from the light source is first polarized. This polarized light passes through a sample cell. If an optical active substance is in a sample tube, the plane of the polarized light waves is rotated. The rotation is noticed by looking through the analyser as a change in intensity of illumination. To reach the same illumination as was without an optical active sample the analyser must be turned around for an angle. Readings are taken in degrees (angle) α or sugar degrees 0Z. Figure 1: a) principle of a polarimeter set up b) components of a polarimeter 1 Figure 2 shows polarizer (P) and analyzer (A) in a perpendicular position with one another. Both have the structure on a molecular level that polarize an unpolarized light. If they are in a position shown no light passes through analyzer. Polaroid films for example have such a molecular structure that filters (block) all planes of light vibrations except one.
    [Show full text]
  • Design of the Life Signature Detection Polarimeter Lsdpol
    Design of the Life Signature Detection Polarimeter LSDpol Christoph U. Keller*a, Frans Snika, C. H. Lucas Pattya,b, Dora Klindžića,c, Mariya Krastevaa, David S. Doelmana, Thomas Wijnena,c, Vidhya Pallichadathc, Daphne M. Stamc, Brice-Olivier Demoryb, Jonas G. Kühnb, H. Jens Hoeijmakersb, Antoine Pommerolb, Olivier Pochd aLeiden Observatory, Leiden University, PO Box 9513, 2300 RA Leiden, The Netherlands; bPhysikalisches Institut, Universität Bern, Sidlerstrasse 5, 3012 Bern, Switzerland; cFaculty of Aerospace Engineering, Delft University of Technology, Kluyverweg 1, 2629 HS Delft, The Netherlands; dUniversité Grenoble-Alpes, CNRS, IPAG, 38000 Grenoble, France ABSTRACT Many biologically produced chiral molecules such as amino acids and sugars show a preference for left or right handedness (homochirality). Light reflected by biological materials such as algae and leaves therefore exhibits a small amount of circular polarization that strongly depends on wavelength. Our Life Signature Detection polarimeter (LSDpol) is optimized to measure these signatures of life. LSDpol is a compact spectropolarimeter concept with no moving parts that instantaneously measures linear and circular polarization averaged over the field of view with a sensitivity of better than 10-4. We expect to launch the instrument into orbit after validating its performance on the ground and from aircraft. LSDpol is based on a spatially varying quarter-wave retarder that is implemented with a patterned liquid-crystal. It is the first optical element to maximize the polarimetric sensitivity. Since this pattern as well as the entrance slit of the spectrograph have to be imaged onto the detector, the slit serves as the aperture, and an internal field stop limits the field of view.
    [Show full text]
  • Detection of Circular Polarization in Light Scattered from Photosynthetic Microbes
    Detection of circular polarization in light scattered from photosynthetic microbes William B. Sparksa,1, James Houghb, Thomas A. Germerc, Feng Chend, Shiladitya DasSarmad, Priya DasSarmad, Frank T. Robbd, Nadine Mansete, Ludmilla Kolokolovaf, Neill Reida, F. Duccio Macchettoa, and William Martinb aSpace Telescope Science Institute, 3700 San Martin Drive, Baltimore, MD 21218; bCenter for Astrophysics Research, University of Hertfordshire, Hatfield AL10 9AB, United Kingdom; cNational Institute of Standards and Technology, 100 Bureau Drive, Gaithersburg, MD 20899; dCenter of Marine Biotechnology, University of Maryland Biotechnology Institute, 701 East Pratt Street, Suite 236, Baltimore, MD 21202; eCanada-France-Hawaii Telescope, 65-1238 Mamalahoa Highway, Kamuela, HI 96743; and fDepartment of Astronomy, University of Maryland, College Park, MD 20742 Edited by Rita R. Colwell, University of Maryland, College Park, MD, and approved March 24, 2009 (received for review October 10, 2008) The identification of a universal biosignature that could be sensed spectroscopic features have been shown to be consistent with remotely is critical to the prospects for success in the search for life those of radiation tolerant microbes (13). Beyond the Solar elsewhere in the universe. A candidate universal biosignature is System, methods will be needed to assess whether extrasolar homochirality, which is likely to be a generic property of all planets harbor life, and remote sensing is a necessity. Attention biochemical life. Because of the optical activity of chiral molecules, is being given to atmospheric composition disequilibria and to it has been hypothesized that this unique characteristic may biological pigmentation spectral features as biomarkers (14–16). provide a suitable remote sensing probe using circular polarization Typical disadvantages of these methods include model depen- spectroscopy.
    [Show full text]
  • Series 341/343 Polarimeters
    Series 341/343 Polarimeters the industry-standard for optical activity measurement PerkinElmer polarimeters – precision in optical analysis Series 341/343 Polarimeter – the industry standard. Polarimetry is an important technique for the electronics, results in an instrument which combines investigation and analysis of optically active the highest possible accuracy and precision with substances. Common applications include: simple operation. For routine measurements simply • Characterization of newly synthesized compounds select the wavelength, insert the sample and read the result – either on the integral display or in • Quality control of pharmaceutical drugs PerkinElmer’s unique POL WinLab™ software. • Precise quantitative analysis of sugars • Selective detection of optically active components Operating principle by HPLC As Figure 1 illustrates, monochromatic light passes • Monitoring and control of enzymatic reactions through the polarizer, the sample cell and the analyzer Based on decades of experience in the development and to a photomultiplier. In the optical null position, the manufacture of precision polarimeters, the PerkinElmer plane of polarization of the analyzer is perpendicular Series 341/343 polarimeters have been developed for to the direction of polarization of the incoming radia- these and many other applications. tion. When an optically active sample is placed into the beam path, a servo feedback system automatically The proven optical null principle, with automatic rotates the analyzer until the optical null position has rotation of the analyzer, is the basis of PerkinElmer’s been obtained again. The angle of rotation is measured reputation for excellence in polarimetry. This princi- by an optical encoder with a resolution of 0.0001˚ and ple, combined with state-of-the-art microprocessor is displayed with the correct sign of the rotation.
    [Show full text]