Selection guide

Model Optical System Wavelength Range Bandwidth

EMC-11D-V Single Beam 325-1000 nm 4 nm

EMC-11-UV Single Beam 200-1000 nm 4 nm

EMC-16PC-V Single Beam 320-1100 nm 4 nm

EMC-16PC-UV Single Beam 190-1100 nm 4 nm

EMC-31PC-UV Single Beam 190-1100 nm 2 nm

EMC-32PCS-UV Single Beam 190-1100 nm 0.5/1/2/4/5 nm Scanning function

EMC-61PC-UV Double Beam 190-1100 nm 1.8 nm

EMC-61PCS-UV Double Beam 190-1100 nm 0.5/1/2/4/5 nm Scanning function

EMC-NANO-UV Single Beam 190-1100 nm 4 nm 0.2 ~ 2.0 µl

“PC” means with PC control software | Standard central beam height: 15 mm

M. Wave Basic software included in: EMC-11D-V, EMC-11-UV

M. Wave Professional software included in: EMC-16PC-V, EMC-16PC-UV

UV/VIS Analyst software included in: EMC-31PC-UV, EMC-31PCS-UV EMC-61PC-UV, EMC-61PCS-UV EMC-NANO-UV

EMCLAB Spectrophotometers are tested with DAkkS (German body of accreditation) certified UV/VIS Reference Materials NIST traceable:

 Photometric Accuracy  Wavelength Accuracy  Spectral Resolution  Stray Light TRACEABLE

All EMCLAB Spectrophotometers are supplied with a EMCLAB Works Calibration Certificate.

E-Mail: [email protected] | www.emc-lab.de 3

Visible Spectrophotometer EMC-11D-V

EMC-11D-V is the only model of manually setting wavelength, but precise design and high quality components ensures excellent performance. It is widely used in high schools and colleges for general analysis and experiments.

Features:

 Large LCD screen (128*64 Dots)  Wavelength can be read out from the screen directly  Auto Zero and Blank  Parallel port, printed directly  Large sample compartment, it can be accommodate 5-100 mm path length with optional holders  Pre-aligned design ensures the user can change lamp conveniently  The optional PC software M. Wave Basic based on Windows can be expand the applications to Standard Curve & Kinetics  High quality silicon photometric diode detector and 1200 lines/mm grating ensure high accuracy and precision

Model EMC-11D-V

Wavelength Range 325-1000 nm Spectral Bandwidth 4 nm Optical System Single Beam, grating 1200 lines/mm Wavelength Accuracy ±2 nm Wavelength Repeatability 1 nm Photometric Accuracy ≤±0.5 % T or ±0.003A@1A Photometric Range 0-200 % T, -0.3 - 3A, 0-1999 Conc Stray Light 0.3 % T Stability ±0.004A/h @500 nm Display 128*64 LCD Photometric Mode T, A, C, F Detector Silicon Photodiode Standard Cell Holder 4-position 10 mm cell changer Sample Compartment Standard 10 mm path length Light Source Tungsten lamp Output USB Port & Parallel Port (Printer) Power Requirement AC 85V~265V 50/60 Hz Dimensions (LxWxH) 440 x 340 x 200 mm Weight 8 kg

4 E-Mail: [email protected] | www.emc-lab.de

UV / VIS Spectrophotometer EMC-11-UV

EMC-11-UV spectrophotometer is the ideal instrument for education and QC . Using your standard sample solutions, you can get a standard curve on the large LCD screen. It is widely used in colleges and enterprises for general quantitative analysis and experiments.

Features:

 Large LCD screen (128*64 Dots)  Can display total 50 groups of data, 3 groups per screen.  Can display standard curve and the curve equation  System can also save the test results. Total groups of data and 200 standard curves can be saved; it is convenient for check and reload.  Data can be restored after a sudden power cut  Auto setting wavelength  Tungsten lamp & Deuterium lamp can be turned on/off individually to extend lifetime  Pre-aligned design makes it convenient to change lamps  Large sample compartment, it can be accommodate 5-100 mm path length cuvettes with optional holders. A variety of optional accessories are available  The optional application software M. Wave Professional provides complete control of the spectrophotometer through the Built-in USB port. You can achieve the following functions: I. Quantitative II. Kinetics III. Wavelength Scan IV. Multi Wavelength V. DNA/Protein

Model EMC-11UV

Wavelength Range 200-1000 nm Spectral Bandwidth 4 nm Optical System Single Beam, grating 1200 lines/mm Wavelength Accuracy ±2 nm Wavelength Repeatability 1 nm Photometric Accuracy ≤±0.5 % T or ±0.003A@1A Photometric Range 0-200 % T, -0.3 - 3A, 0-9999 Conc. Stray Light 0.3 % T Stability ±0.002A/h @500 nm Detector Silicon Photodiode Standard Cell Holder 4-position 10 mm cell changer Sample Compartment Standard 10 mm path length cuvette Light Source Tungsten & Deuterium lamp Output USB Port & Parallel Port (Printer) Power Requirement AC 110/220V 50/60 Hz Dimensions (LxWxH) 480 x 360 x 220 mm Weight 14 kg

E-Mail: [email protected] | www.emc-lab.de 5

UV / VIS Spectrophotometer EMC-16 series

EMC-16 series are simple-to-use instruments with advanced performance; its stray light is only 0.05%T. The local stand-alone software provides functions of Photometry, Quantitative Test, Kinetics and System Utilities functions.

Features:

 Large LCD screen (128*64 Dots)  System can also save the test results. Total 200 groups of data 200 standard curves can be saved in the RAM. Convenient for check and reload.  Data can be restored after a sudden power cut  Auto setting wavelength  Tungsten lamp & Deuterium lamp can be turned on/off individually to extend lifetime  Pre-aligned design makes it convenient to change lamps  Large sample compartment, it can be accommodate 5-100 mm path length cuvettes with optional holders. A variety of optional accessories are available  The application software M. Wave Professional provides complete control of the spectrophotometer from a computer through the Built-in USB port. It can expand to the following functions: I. Quantitative, II. Kinetics, III. Wavelength Scan, IV. Multi Wavelength, V. DNA/Protein

Functions:

1. Photometry, 2. Quantitative, 3. Kinetics, 4. System Utilities

Model EMC-16PC-V EMC-16PC-UV

Wavelength Range 320-1100 nm 190-1100 nm Spectral Bandwidth 4 nm Optical System Single Beam, grating 1200 lines/mm Wavelength Accuracy ±0.5 nm Wavelength Repeatability 0.3 nm Photometric Accuracy ≤±0.5 % T or ±0.003A@1A Photometric Range 0-200 % T, -0.3 - 3A, 0-9999 Conc. Stray Light ≤0.05 % T@360 nm ≤0.05 % T@260, 360 nm Stability ±0.002A/h @500 nm Detector Silicon Photodiode Display Graphic LCD (128+64 dots) Keyboard 22 Membrane keypad Standard Cell Holder 4-position 10 mm cell changer Sample Compartment Standard 10 mm path length cuvette Light Source Tungsten lamp Tungsten & Deuterium lamp Output USB Port & Parallel Port (Printer) Power Requirement AC 110/220V 50/60 Hz Dimensions (LxWxH) 490 x 370 x 220 mm Weight 12 kg 14 kg

6 E-Mail: [email protected] | www.emc-lab.de

UV / VIS Spectrophotometer EMC-16 series

Series Local Control Software

Main Menu

Move the cursor on the function menu you want, then press ENTER key to go into the corresponding interface

1. Basic Mode

Absorbance, Transmittance or Concentration measurements. It can display and save 200 groups of data, 5 groups per screen

2. Quantitative 2.1 Standard Curve A most 9 standard sample can be used to establish a standard curve. The curve and its equation will be displayed on the screen simultaneously. You can measure your unknown concentration solutions by the curve. Total 200 curves can be saved in the RAM 2.2 Coefficient Method If the coefficient k & b in the formula C=k*a + b is known, you can input them directly by the button, and then test your unknown solutions.

3. Kinetics This mode may be used for time course scanning or reaction rate calculation. Abs. VS Time graphs is displayed on the screen in real time. It can most record 1000 data.

4. System Utilities Lamp management, time & data set, obtain dark current, calibrate wavelength, default system, and some other system functions.

E-Mail: [email protected] | www.emc-lab.de 7

UV / VIS Spectrophotometer EMC-3 series

The EMC-3 series are advanced single beam spectrophotometers. They provide excellent performance for measurements in the range of 190 to 1100 nm. The memory is 32K. They are suitable for pharmaceutical, biochemical and clinical lab applications as well as routine applications such as quantitative analysis, kinetics, wavelength scan, multi-wavelength and DNA/Protein analysis. The UV/VIS Analyst software based on Microsoft Windows® makes the instruments versatile. The instruments provide excellent performance for measurements.

Features:

 Fixed or variable slits (Bandwidth)  Sealed, solvent-resistant tactile keypad with alpha-numeric entry for file names and units  Pre-aligned deuterium lamp for easy lamp replacement. The status of the lamp may be monitored  Online software upgrading via internet helps to keep it updated  Data Download-to-PC software expands the data storage to unlimited  The PC model comes standard with Windows® based application software UV-VIS Analyst  Data can be saved by USB memory device directly

Model EMC-31PC-UV EMC-32PCS-UV

Wavelength Range 190-1100 nm Spectral Bandwidth 2 nm 0.5/1/2/4/5 nm Optical System Single Beam, grating 1200 lines/mm Wavelength Accuracy ±0.5 nm ±0.3 nm Wavelength Repeatability 0.3 nm 0.2 nm Scan Speed Hi, MED, LOW, MAX. 3000 nm/min Photometric Accuracy ≤±0.5 % T or ±0.005A@1A Photometric Range 0-200 % T, -0.3 - 3A, 0-9999Conc. Stray Light 0.05 % T@220, 340 nm Stability ±0.002A/h @500 nm Detector Silicon Photodiode Display 5 inches LCD (320*240 dots) Baseline Flatness ±0.002A (200-1000 nm) Standard Cell Holder 4-position 10 mm cell changer Sample Compartment Standard 10 mm path length cuvette Light Source Tungsten & Deuterium lamp (Pre-aligned) USB type A port for USB memory device (Right side) Output USB type B port for optional computer connectivity (Back)

Parallel port for printer Power Requirement AC 110/220V 50/60 Hz Dimensions (LxWxH) 490 x 370 x 220 mm 589 x 428 x 240 mm Weight 14 kg 20 kg

8 E-Mail: [email protected] | www.emc-lab.de

UV / VIS Spectrophotometer EMC-6 series

The EMC-6 series are advanced double beam spectrophotometers. The two detectors measure sample and reference respectively and simultaneous for optimizing measurement accuracy. They provide excellent performance for measurements in the range of 190 to 1100 nm. The memory is 32K. They are suitable for pharmaceutical, biochemical and clinical lab applications as well as routine applications such as quantitative analysis, kinetics, wavelength scan, multi- wavelength and DNA/Protein analysis. The UV/VIS Analyst software based on Microsoft Windows® makes the instruments versatile. The instruments provide excellent performance for measurements.

Features:

 Fixed or variable slits (Bandwidth)  Sealed, solvent-resistant tactile keypad with alpha-numeric entry for file names and units  Pre-aligned deuterium lamp for easy lamp replacement. The status of the lamp may be monitored  Online software upgrading via internet helps to keep it updated  Data Download-to-PC software expands the data storage to unlimited  The PC model comes standard with Windows® based application software UV-VIS Analyst  Data can be saved by USB memory device directly

Model EMC-61PC-UV EMC-61PCS-UV

Wavelength Range 190-1100 nm Spectral Bandwidth 1.8 nm 0.5/1/2/4/5 nm Optical System Double Beam, grating 1200 lines/mm Wavelength Accuracy ±0.3 nm Wavelength Repeatability 0.2 nm Scan Speed Hi, MED, LOW, MAX. 3000 nm/min Photometric Accuracy ≤±0.5 % T or ±0.005A@1A Photometric Range 0-200 % T, -0.3 - 3A Stray Light 0.05 % T@220, 340 nm Stability ±0.001A/h @500 nm Detector Silicon Photodiode Display 5 inches LCD (320*240 dots) Baseline Flatness ±0.001A Standard Cell Holder Standard 10 mm single cell holder (2 pcs.) Sample Compartment Standard 10 mm path length cuvette Light Source Tungsten & Deuterium lamp (Pre-aligned) USB type A port for USB memory device (Right side) Output USB type B port for optional computer connectivity (Back) Parallel port for printer Power Requirement AC 110/220V 50/60 Hz Dimensions (LxWxH) 589 x 428 x 240 mm Weight 22 kg

E-Mail: [email protected] | www.emc-lab.de 9

NANO Spectrophotometer EMC-NANO-UV

EMC-NANO-UV the new model for fast, easy, accurate and economic measuring of minimum volumes of 0.2 µl to 2 µl with high quality components ensures excellent performance. It is used in DNA, RNA, PCR, protein analyzing within the range from UV to VIS range.

Features:

 Compact instrument with on-board software as well as PC application software UV/VIS-Analyst  Multifunctional EMC-cell holder for Nano volume and standard cuvette optical path length 10 mm, optional with magnetic stirrer  Large 5’’ LCD screen (320*240 Dots)  0.2-2.0 µl sample volume for DNA, RNA, PCR, protein …  One step – one movement drop close turn and measure  Wavelength from 190-1100 nm  Pre-aligned design ensures the user can change lamps conveniently  Additional USB port for memory extension  High quality silicon photodiode detector ensures high accuracy and precision

The special multifunctional cell holder for Nano volume and standard cuvette optical path length 10 mm:

 Simple-to-use  Multifunctional EMC-cell holder – 2 in 1  Horizontal light path – high precision, no lost of energy

Model EMC-NANO-UV

Wavelength Range 190-1100 nm Spectral Bandwidth 4 nm Optical System Single Beam, grating 1200 lines/mm Wavelength Accuracy ±0.5 nm Wavelength Repeatability 0.3 nm Photometric Accuracy ≤±0.5 % T or ±0.004A@1A Photometric Range 0-200 % T, -0.3 - 3A, 0-9999 Conc. Stray Light 0.05 % T@220, 340 nm Stability ±0.0008A/h @500 nm Detector Silicon Photodiode Multifunctional Cell Holder 0.5 – 2 µl Volume & Standard cuvette 10 mm path length Scan Speed Hi, MED, LOW – MAX: 3000nm/min Light Source Tungsten & Deuterium lamp USB type A port for USB memory device (Right side) Output USB type B port for optional computer connectivity (Back) Parallel port for printer Power Requirement AC 110/220V 50/60 Hz Dimensions (LxWxH) 490 x 370 x 220 mm Weight 14 kg

Recommended

Article No. Volume (µl) Increment (µl) R/A% (CV%) Tip EMC-5720005 0.1-2.5 0.05 4.0 (2.0) 1

10 E-Mail: [email protected] | www.emc-lab.de

UV / VIS Spectrophotometer EMC-3 & EMC-6 series

Series Local Control Software

All methods are included as built-in standard; this eliminates the need of software. Online software update via internet. The local control software includes functions such as: Photometry, Quantitative, Kinetics, Wavelength Scan, Multi-wavelength Test and System Utilities

Standard Curve

Up to 10 standard solutions may be used to establish calibration equation curve. There is a choice of four methods for fitting curve through the calibration points: linear fit, Linear fit through zero, square fit and cubic fit.

Wavelength Scan

The wavelength scan intervals are 0.1, 0.2, 0.5, 1, 2, 5 nm and High, Medium and Low scan speeds are available. Scan speeds vary from 100 to 3000 nm/min. Wavelengths are scanned for high to low so that the instrument stand-by at high wavelength. This minimizes the degradation of UV sensitive samples. Precise control of filter and lamp changes means that their effects are not seen on the final scan. Post-run manipulation includes re-scaling axes, curve tracking and peak picking.

Kinetics

This mode may be used for scanning time course or reacting rate calculations. Abs. VS time graphs is displayed on the screen in real time. Wait time and measurement time up to 12 hours may be entered with time intervals of 0.5, 1, 2.5, 10, 30 seconds and 1 min. Post-run manipulation includes re-scaling, curve tracking and selection of the part of the curve required for the rate calculation. Rate is calculated using a linear regression algorithm before multiplying by the entered factor.

Multi-Wavelength

Up to 10 wavelengths may be entered, allowing the measurement of multiple wavelengths on a series of samples.

DNA/Protein Test

Concentration and DNA purity are calculated by absorbance ratios 260 nm/280 nm or 260 nm/230 nm with optional subtracted absorbance at 320 nm DNA concentration = 62.9*A260-36.0*A280 or 49.1*A260-3.48*A230 Protein concentration = 1552*A260-757.3*A280 or 183*A260-75.8*A230 Other wavelengths and factors may be entered.

E-Mail: [email protected] | www.emc-lab.de 11

Spectrophotometer Software

M. Wave Professional Software for EMC-16 series

M. Wave Professional application software is based Microsoft Windows, the instrument can be controlled by PC software through built-in USB communication port, which makes the UV/VIS series with more functions and easy to control.

Quantitative

Use up to 20 standards to establish standard curve. Three methods for fitting a curve: 1. Linear fit 2. Linear through zero 3. Square fit

Kinetics

The Kinetics mode may be used for scanning time course or reacting rate calculations. Abs. VS time graphs is displayed.

Wavelength Scan

Automatically records peaks and valleys. The quantity of the curves stored is unlimited. Post-run manipulation and processing includes 1. Re-scaling axes, curve 2. Smoothing, combination, zooming, overlap… 3. 1st to 4th derivative

Multi-wavelength Test

You can set up to 20 wavelengths to measure a sample

DNA/Protein Test

Optional two formulas: DNA concentration = 62.9*A260-36.0*A280 or 49.1*A260-75.8*A230 You can also enter other wavelengths and factors to calculate.

12 E-Mail: [email protected] | www.emc-lab.de

Spectrophotometer Software

UV/VIS Analyst Software for EMC-3 series, EMC-6 series & EMC-NANO

The PC application software UV/VIS Analyst takes the features of the stand-alone version plus more powerful data processing, expanded data collecting, and storage capability. It comes standard with EMC-3 and EMC-6 series.

The PC application software offers: 1. Photometric Mode 2. Quantitative Test (Standard curve) 3. Wavelength Scan 4. Kinetics 5. DNA/Protein 6. Multi-Wavelength 7. System Utility

Quantitative Test

Use up to 20 standards to establish standard curve. Four methods for fitting a curve: 1. Linear fit 2. Linear through zero 3. Square fit 4. Cubic fit

Wavelength Scan

Automatically records peaks and valleys. The quantity of channels is unlimited. You can simultaneously store as many as desired. Post-run manipulation and processing includes: 1. Re-scaling axes, curve 2. Smoothing, combination, zooming, overlap… 3. 1st to 4th derivative

Kinetics (Abs. VS time)

The Kinetics mode may be used for scanning time course or reacting rate calculations. Abs. VS time graphs are displayed on the screen in real time. Wait time, measurement time and time intervals may be entered. Post-run manipulation includes re-scaling, curve tracking and selection of the part of the curve required for the rate calculation. Rate is calculated using a linear regression algorithm before multiplying by the entered factor.

E-Mail: [email protected] | www.emc-lab.de 13

Spectrophotometer Software

DNA/Protein Test

Concentration and DNA purity are quickly and easily calculated. Absorbance ratios 260 nm/280 nm with optional subtracted absorbance at 320 nm DNA concentration = 62.9*A260-36.0*A280 Protein concentration = 1552*A260-757.3*A280

Other wavelengths and factors may be entered.

Multi-wavelength

Up to 20 wavelengths can be selected and multiple samples can be measured. (auto cell changer is required to run multiple samples automatically)

14 E-Mail: [email protected] | www.emc-lab.de

Spectrophotometer Cuvettes

Cuvettes and quartz glass

Standard Premium Optical path Volume Outside dim Cuvette Art. No. Art. No. length (mm) (µl) (H x W x D mm)

Macro cuvettes:

EMC-S0001 EMC-100OS Cuvette square glass 10 3500 45 x 12,5 x 12,5 EMC-S0002 EMC-105OS Cuvette square glass 50 17500 45 x 12,5 x 52,5 Cuvette square EMC-S0003 EMC-100QS quartz glass 10 3500 45 x 12,5 x 12,5 Cuvette square EMC-S0004 EMC-105QS quartz glass 50 17500 45 x 12,5 x 52,5

Semi-Micro cuvettes:

- EMC-108OS Cuvette glass 10 1000 45 x 12,5 x 12,5 Cuvette glass EMC-S0005 - (black side walls and base) 10 1000 45 x 12,5 x 12,5 - EMC-108QS Cuvette quartz glass 10 1000 45 x 12,5 x 12,5 Cuvette quartz glass EMC-S0011 EMC-108QSB (black side walls and base) 10 1000 45 x 12,5 x 12,5

Micro cuvettes:

Cuvette glass EMC-S0007 EMC-104OSB (black side walls and base) 10 700 45 x 12,5 x 12,5 - EMC-104QS Cuvette quartz glass 10 700 45 x 12,5 x 12,5 Cuvette quartz glass EMC-S0013 EMC-104QSB (black side walls and base) 10 700 45 x 12,5 x 12,5 - EMC-108QSM Cuvette quartz glass 10 500 45 x 12,5 x 12,5 Cuvette quartz glass - EMC-08QSMB (black side walls and base) 10 500 45 x 12,5 x 12,5 Cuvette glass EMC-S0006 - (black side walls and base) 10 350 45 x 12,5 x 12,5 Cuvette quartz glass EMC-S0012 - (black side walls and base) 10 350 45 x 12,5 x 12,5 Cuvette quartz glass - EMC-108QSM15 averture 5 x 2 mm 10 100 45 x 12,5 x 12,5

Flow-through cuvettes:

EMC-S0009 - Cuvette glass 10 480 45 x 12,5 x 12,5 - EMC-175OS Cuvette glass 10 750 45 x 12,5 x 12,5 EMC-S0010 - Cuvette quartz glass 10 480 45 x 12,5 x 12,5 - EMC-175QS Cuvette quartz glass 10 750 45 x 12,5 x 12,5

Standard optical glass: 340 – 2500 nm Premium special optical glass: 320 – 2500 nm Standard quartz glass: 190 – 2500 nm Premium quartz glass: 200 – 2500 nm

E-Mail: [email protected] | www.emc-lab.de 15

Spectrophotometer Disposable Cuvettes

Disposable Cuvettes UV/VIS

New plastic coupled with modern methods of production allow photometric measurement well into the UV range. Ultra-precision molds with several cavities – one per cuvette – are used for producing EMC-disposable cuvettes under controlled room conditions. Thus, in any one injection molding run several cuvettes are produced simultaneously. In order to prevent deviations in extinction coefficient values from occurring in any one run due to increased stray light, the cuvettes are automatically sorted according to their numbered cavities of origin. Each package contains only cuvettes that have been produced in any one cavity. The corresponding cavity number is noted on each package. Thus, even for extensive analyses, the same cuvettes can be used.

 Very low variation of extinction values  Excellent optical transmission range  Cavity-sorted production  Optical path length 10 mm  Outside dim 45 x 12,5 x 12,5 mm  Cuvettes supplied in practical styrofoam racks  Stoppers supplied in a plastic bag

Volume Packaging Wavelength Art. No. Cuvette (ml) (pcs.) (nm)

Macro cuvettes:

EMC-S0028 PS (glass-clear Polystyrene) 4,0 10 x 100 340 – 900 EMC-S0030 UV 4,0 1 x 100 220 – 900 EMC-S0031 Solvent resistant against polar solvents 4,0 1 x 100 220 – 900 EMC-S0032 PMMA (Polymethylmethacrylate) 4,0 10 x 100 300 – 900

Semi-Micro cuvettes:

EMC-S0029 PS (glass-clear Polystyrene) 1,6 10 x 100 340 – 900 EMC-S0033 UV 1,6 1 x 100 220 – 900 EMC-S0034 Solvent resistant against polar solvents 1,6 1 x 100 220 – 900 EMC-S0035 PMMA (Polymethylmethacrylate) 1,6 10 x 100 300 – 900

Stoppers:

EMC-S0036 PP (Polypropylene) 1 x 1000

16 E-Mail: [email protected] | www.emc-lab.de

Spectrophotometer Reference Materials

UV/VIS certified Reference Materials

TRACEABLE EMC-SET-667UV EMC-SET-666

The UV/VIS certified reference materials enable the user to check quickly and easily his measurements results with the quality standard of the spectrophotometer. The measurements results are internationally comparable. The reference materials are DAkkS (German body of accreditation) certified traceable to primary references of NIST (National Institute of Standards and Technology) and in accordance with the most important Pharmacopoeias (EP, DAB; USP).

UV/VIS certified reference materials for testing:

 Wavelength accuracy  Photometric accuracy  Stray light  Spectral resolution

Art. No. Filters Wavelength (nm)

Set liquid filter for testing acc. to European Pharmacopeias:

EMC-SET-667UV Potassium chloride in pure water A*: 235; 257; 313; 350; 430 Pure water reference filter) Toluene in n-hexane W*: 241; 287; 361; 536; 640 n-hexane (reference filter) 60 mg Potassium dichromate in HClO4 (0.75 Abs.) S*: 200 (cut-off) 600 mg Potassium dichromate in HClO4 (1.0 Abs.) Perchlorid acid HClO4 (reference filter) R*: Scan 265-270 Holmium Oxide in perchloric acid

Set glass filter wavelength accuracy and photometric accuracy:

EMC-SET-666 Holmium Oxide glass filter A*: 440; 465; 546.1; 590; 635 Neutral density glass filter 0.25 Abs. Neutral density glass filter 0.5 Abs. W*: 279; 361; 453; 536; 638 Neutral density glass filter 1.0 Abs. Empty filter mount, aluminium frame EMC-SET-666-01 Didymium glass filter A*: 270; 280; 297; 320; 340; Neutral density glass filter 0.5 Abs. 440; 465; 546.1; 590; 635 Neutral density glass filter 1.0 Abs. W*: 329; 472; 512; 681; 875 EMC-SET-666-02 Holmium Oxide glass filter A*: 440; 465; 546.1; 590; 635 Neutral density glass filter 0.5 Abs. Neutral density glass filter 1.0 Abs. W*: 279; 361; 453; 536; 638 *A= Wavelength for absorbance, W= Wavelength for wavelength accuracy, S= Wavelength for stray light, R= Wavelength for spectral resolution

E-Mail: [email protected] | www.emc-lab.de 17

Spectrophotometer Reference Materials

UV/VIS Reference Materials with EMCLAB Works Calibration Certificate

EMC-SET-100 EMC-SET-101

The EMC-glass filters for testing the wavelength accuracy and photometric accuracy enable the user to check quickly and easily his measurement results with the quality standards of the spectrophotometer. All EMC-glass filters are supplied with EMCLAB Works Calibration Certificate.

UV/VIS reference materials for testing:

 Wavelength accuracy  Photometric accuracy

Art. No. Filters Wavelength (nm)

Set glass filter for testing the wavelength accuracy and photometric accuracy:

EMC-SET-100 Holmium Oxide glass filter F 10 A*: 440; 465; 546.1; 590; 635 Neutral density glass filter F 20, 30 % T, Abs. 0.5 Neutral density glass filter F 30, 20 % T, Abs. 0.7 W*: 279; 361; 453; 536; 638 Neutral density glass filter F 40, 10 % T, Abs. 1.0 Blank filter EMC-SET-101 Didymium glass filter F 70 A*: 270; 280; 297; 320; 340; 440; Neutral density glass filter F 20, 30 % T, Abs. 0.5 465; 546.1; 590; 635 Neutral density glass filter F 30, 20 % T, Abs. 0.7 Neutral density glass filter F 40, 10 % T, Abs. 1.0 W*: 329; 472; 512; 681; 875 Blank filter

Set glass filter for testing the photometric accuracy:

EMC-SET-102 Neutral density glass filter F 20, 30 % T, Abs. 0.5 Neutral density glass filter F 30, 20 % T, Abs. 0.7 440; 465; 546.1; 590; 635 Neutral density glass filter F 40, 10 % T, Abs. 1.0 Blank filter

Single glass filter for testing the wavelength accuracy:

EMC-SET-103 Holmium Oxide glass filter F 10 279; 361; 453; 536; 638

Single glass filter for testing the wavelength accuracy and photometric accuracy:

EMC-SET-104 Didymium glass filter F 70 A*: 270; 280; 297; 320; 340 W*: 329; 472; 512; 681; 875 *A= Wavelength for absorbance, W= Wavelength for wavelength accuracy

18 E-Mail: [email protected] | www.emc-lab.de

Spectrophotometer Accessories

Accessories

1 2 3 4 5 6

Art. No. Accessories Item

Cell Holder:

EMC-00020 4-cell holder square cuvette 10 x 10 mm, standard 1 EMC-00021 4-cell holder square cuvette 50 x 50 mm EMC-00022 Cell holder for micro cells EMC-00023 Single cell holder EMC-20024 Single cell holder, water-jacketed 2 EMC-00025 4-cell holder, water-jacketed

Peltier Systems:

EMC-00026 Peltier System, Constant-Temperature-System 3 EMC-00027 Sipper System 3 EMC-00028 Peltier /Sipper System, Constant-Temperature System 3

Lamps:

EMC-00011 Halogen lamp 6V10W (only for EMC-11D-V) 4 EMC-00012 Halogen lamp 12V20W 4 EMC-00013 Halogen lamp 12V20W (only for EMC-6 series) 4 EMC-00017 Deuterium lamp for UV 5

Cleaner:

EMC-S0020 Cleaning concentrate for cells/cuvettes and optical components, 1 L 6

E-Mail: [email protected] | www.emc-lab.de 19

Spectrophotometer Calibration Certificate

EMCLAB Spectrophotometers are supplied with Works Calibration Certificate for wavelength accuracy. The Reference Materials for this calibration are DAkkS (German body for accreditation) calibrated and NIST traceable from the DAkkS calibration D-K-18752-01-00.

20 E-Mail: [email protected] | www.emc-lab.de

Index

Product Description Page EMC-11D-V VIS Single beam 4 EMC-11-UV UV/VIS Single beam 5 EMC-16PC-V VIS Single beam 6 EMC-16PC-UV UV/VIS Single beam 6 EMC-31PC-UV UV/VIS Single beam 8 EMC-32PC-UV UV/VIS Single beam/Scanning 8 EMC-61PC-UV UV/VIS Double beam 9 EMC-61PCS-UV UV/VIS Double beam/Scanning 9 EMC-NANO-UV UV/VIS Single beam 10 EMC-M. Wave Professional PC Software for EMC-16 series 12 EMC-UV/VIS Analyst PC Software for EMC-3, EMC-6 series & EMC-NANO 13 EMC-Cuvettes Glass, Quartz, 10 mm, 50 mm, Standard/Premium 15 Plastic EMC-SETS UV/VIS Reference Materials 17 Spectrophotometer Accessories 19 Calibration Certificate 20

Other types on request | Technical changes reserved

E-Mail: [email protected] | www.emc-lab.de 21

Dealer:

EMCLAB Instruments GmbH Bismarckstrasse 120 47057 Duisburg | Germany Fon: +49 203 3064042 Fax: +49 203 3064044 E-mail: [email protected] Web: www.emc-lab.de