LAB 2: Introduction to Lab Techniques I

Total Page:16

File Type:pdf, Size:1020Kb

LAB 2: Introduction to Lab Techniques I Teaching Data Analysis and Presentation– ABLE Poster - D. Hougen-Eitzman - Carleton College Instructor's Note: Students will be required to plot the data from the Bradford Asssy into a standard curve, fit a line to the data, and use the equation for the line. These skills were introduced in the first lab computer session. LAB 2: Introduction to Lab Techniques I. INTRODUCTION wavelength (λ) absorbed by a substance. It In this laboratory, you will be introduced does this by measuring the intensity of light (or reintroduced) to some basic techniques after it has passed through the substance. of modern biology- microscopy, The essential components of the instrument spectrophotometry and micropipetting. In are shown below in Figure 2.2. today’s lab we will practice the general use We make use of the phenomenon that the of the compound light microscope. In absorption of light at a given wavelength is addition, we will conduct an experiment that related to the concentration of the absorbing will introduce/reinforce the use of chemical. For instance, if I0 is the intensity spectrophotometers and micropipettors, as of the incident light (the light entering the well as exercise your computer skills. sample) and I is the intensity of the transmitted light (the light leaving the II. PROTEIN ASSAY EXPERIMENT sample), then the absorbance is the relative In this lab exercise, we will practice two amount of light that is absorbed by the important and use lab techniques: using a sample: spectrophotometer and micropipetting. In A = log I0 /I the exercise, you will use these methods for a common type of assay in biochemistry - Thus, if the intensity of light coming out of determining the protein concentration of an the sample is the same as the light going in, unknown sample. then no light has been absorbed and A = log (1) = 0. On the other hand, if the The determination of protein concentration absorbance is high, this means that very is frequently required in biochemical work. little light passed through the solution. In For instance, if we are comparing the the visible spectrum, you can think of water activity of 2 different enzyme preparations, as having an absorbance close to 0, while and find one to display substantially more milk has a large absorbance. Note that the activity than other, we really can't conclude relationship between absorbance and anything until we have ascertained how proportion of light transmitted is much of the each enzyme is present logarithmic, so a 10 times reduction in (remember that enzymes are proteins). transmitted light results in an increase in A. Spectrophotometry absorbance (A) of only 1. Spectrophotometric techniques are The conversion of absorbance to techniques based on the differential concentration of absorbing substance is absorption of light by different chemicals. straightforward: These techniques serve a wide array of A = ε c Beer-Lambert Law functions in biology. They can be used to l determine the concentration of many where A is absorbance, ε is the extinction compounds, such as DNA and proteins, and coefficient (a property of the compound that they can be used to measure enzyme is doing the absorbing), c is the activity. concentration of the absorbing material, and The machine we'll be using is a l is the length of the light path, usually 1 cm. spectrophotometer. It measures This little equation is so important that it has absorbance, the amount of light of a given 2-1 Teaching Data Analysis and Presentation– ABLE Poster - D. Hougen-Eitzman - Carleton College been made into a law -- the Beer-Lambert Law or sometimes Beer's Law. 2-2 Teaching Data Analysis and Presentation– ABLE Poster - D. Hougen-Eitzman - Carleton College B. Use of Micropipettors Setting volumes on the pipetmen: Throughout this term, you will be required To use the pipettors, first set the volume to to pipette accurately and reproducibly. This pipet using the dial on the pipet. See the is not as easy as it sounds. In fact, next to table below for example settings on the failing to read and think about the lab pipettors. directions, pipetting mistakes are the greatest source of heartache in the laboratory. for the P1000: To facilitate this important task, we will be 1 0 0 using highly accurate and sophisticated (and 0 5 2 expensive) automatic pipettors. There are several brands available, but we will be 0 0 5 using the Pipetman. You will need to make use of 3 different micropipettors: equals equals equals 1000 µl or 500 µl or 250 µl or 1) the P1000, which measures 200 to 1000 1 ml 0.5 ml 0.25 ml µl (microliters); 2) the P200, which measures 20 to 200 µl; for the P200: 3) and the P20, which measures 1 to 20 µl. 2 0 0 0 8 5 Fill out the following chart: 0 5 0 1 µl = ____ml; measure it with a P____ equals equals equals 10 µl = ____ml; measure it with a P____ 200 µl or 85 µl or 50 µl or 100 µl = ____ml; measure it with a P____ 0.2 ml 0.085 ml 0.05 ml 1000 µl = ____ml; measure it with a P____ for the P20: 2 1 0 Figure 2.3. All of 0 0 5 the micropipettors operate in the 0 5 5 same fashion. To choose the equals equals equals volume, hold the 20 µl or 10.5 µl or 5.5 µl or micropipettor 0.02 ml 0.0105 ml 0.0055 ml body in one hand and turn the volume adjustment knob until the correct volume shows on the digital indicator. The volumes are read from the top down. 2-3 Teaching Data Analysis and Presentation– ABLE Poster - D. Hougen-Eitzman - Carleton College Using the pipetmen: Rules for pipetting After you have dialed in the appropriate It is always good to have a list of rules. volume, attach a disposable tip to the shaft. Here is a list of rules for pipetting: The P1000 uses the large tips (usually blue or clear) and the P200 and P20 use the ⇒ Never rotate the volume adjuster smaller size (usually yellow). Next, depress beyond the upper or lower range of the plunger to the first stop. This part of the the pipettor. stroke is the calibrated volume displayed on the dial. Immerse the tip into the sample ⇒ Never force the volume adjuster. If liquid to a depth of several mm. Allow the force is required, you are doing plunger to return slowly to the up position. something wrong or the pipettor is Never let it snap up. (Why?) Withdraw the broken. In either case, see you tip from the liquid and remove any adhering instructor. In general, in laboratories liquid by touching to the inside of the tube as well as life, if force is required, holding the sample liquid. back off and think about it. To dispense the sample, place the tip end ⇒ Never use the pipettor without a tip against the side wall of the receiving tube (No duh!). and depress the plunger slowly to the first stop. Wait a second and then depress the ⇒ Never lay down the pipettor with plunger to the second stop. With the liquid in the tip. plunger depressed, remove the tip from the tube, allow the plunger to return to the top ⇒ Never let the plunger snap back after position and discard the tip. withdrawing or ejecting sample. The most common mistake in pipetting is using the second plunger stop to fill the ⇒ Never immerse the barrel into pipet – don't do this! :You should use the solution. first stop for filling the pipet, and the second stop for emptying it. Follow these rules throughout your life and you will be a great success in any molecular or biochemistry lab. 2-4 Teaching Data Analysis and Presentation– ABLE Poster - D. Hougen-Eitzman - Carleton College C. Making a Standard Curve Instructor's Note: Students will be required The Bradford Assay to plot the data from the Bradford asssy into The Bradford assay is, more or less, the a standard curve, fit a line to the data, and gold standard of protein concentration use the equation for the line. These skills determination. It is quick and can detect were introduced in the first lab computer protein concentrations as low as 1µg/ml. session. The assay takes advantage of the fact that the color of the dye Coomassie Brilliant For this lab, we'll determine protein Blue G-250 changes when it is bound to concentration of an unknown sample using a proteins. This color change alters the standard curve. We will first measure the amount of light absorbed by the solution, absorbance of several samples of known which in turn can be measured with the protein concentration. We can then draw a spectrophotometer. The Coomassie blue graph of the relationship between binds primarily to basic and aromatic amino absorbance and protein concentration. Once acid residues, especially arginine. This we have this standard curve, we can specificity does introduce the problem of measure the absorbance of an unknown varying sensitivity, depending on the amino sample and read its corresponding protein acid composition of the proteins, but this is concentration from the graph. Without an compensated for by its ability to detect such accurate standard curve, well, there is just small amounts of protein. (Under what not much point in going on. circumstances would the specificity of the Coomassie blue binding a potential Preparation of the Standard Curve problem?) The Bradford reaction comes in We will construct a standard curve 2 forms, one which can detect 200 µg/ml to according to the table below.
Recommended publications
  • White Paper Absorbance Or Fluorescence: Which Is the Best
    WHITE PAPER No. 40 Absorbance or Fluorescence: Which Is the Best Way to Quantify Nucleic Acids? Natascha Weiß1, Martin Armbrecht1 ¹Eppendorf AG, Hamburg, Germany Executive Summary When performing molecular experiments on nucleic acids, it is a basic requirement to determine the concentration as well as the quality of the sample. The standard method which serves this purpose is UV-Vis spectrophotometry – but is it always the right choice? In the present White Paper, the advantages and disadvantages of this technique will be described and it will be compared with nucleic acid quantification via fluorescence. In this context, it will be discussed which situations warrant the use of which one of the two methods. The decision will generally depend on the condition of the sample as well as on the requirements of downstream applications. Since the advantages of both methods complement each other well, it is most practical to have the ability to perform both methods on a single instrument in a flexible manner. Introduction Experiments involving nucleic acids are a mainstay in any evaluated by measuring the sample at additional wave molecular laboratory. DNA and RNA are isolated from micro- lengths (230 nm, 280 nm) and calculating the purity ratios, organisms and from cells of higher order organisms in order i.e. the ratios of the values obtained at 260/230 nm and at to be employed in a broad variety of processing steps and 260/280 nm, respectively. In this way, it will be evident analyses. It is crucial for any type of downstream applica- whether cellular debris or remainders of reagent used during tion that a defined amount of nucleic acid is used and that purification such as proteins, sugar molecules, certain salts the sample is free from contaminations that may impact the or phenols, are present in the solution, as these will generate experiments.
    [Show full text]
  • On Food, Spectrophotometry, and Measurement Data Processing (Keynote Lecture)
    12th IMEKO TC1 & TC7 Joint Symposium on Man Science & Measurement September, 3–5, 2008, Annecy, France ON FOOD, SPECTROPHOTOMETRY, AND MEASUREMENT DATA PROCESSING (KEYNOTE LECTURE) Roman Z. Morawski Warsaw University of Technology, Faculty of Electronics and Information Technology, Institute of Radioelectronics Warsaw, Poland [email protected] Abstract: Spectrophotometry is getting more and more Despite philosophical abnegation of food issues, the often the method of choice not only in laboratory analysis of practice of food preparation and refinement has flourished (bio)chemical substances, but also in the off-laboratory for centuries, and inspired research-and-invention-oriented identification and testing of physical properties of various minds. Today, we may speak about a fully developed products, in particular – of various organic mixtures discipline of science and technology. Enough to say that the including food products and ingredients. Specialized Institute of Food Technologists – the largest international, spectrophotometers, called spectrophotometric analyzers are non-profit professional organization involved in the designed for such applications. This keynote lecture is on advancement of food science and technology – is the state of the art and developmental trends in the domain encompassing 23 000 members worldwide. Its Committee of spectrophotometric analyzers of food with particular on Higher Education provided the following definition of emphasis on wine analyzers. The following issues are food science: "Food science is the discipline in which the covered: philosophical and methodological background of engineering, biological, and physical sciences are used to food analysis, physical and metrological principles of study the nature of foods, the causes of deterioration, the spectrophotometry, the role of measurement data processing principles underlying food processing, and the improvement in spectrophotometry, food analyzers on the market and of foods for the consuming public" [1].
    [Show full text]
  • ELISA Plate Reader
    applications guide to microplate systems applications guide to microplate systems GETTING THE MOST FROM YOUR MOLECULAR DEVICES MICROPLATE SYSTEMS SALES OFFICES United States Molecular Devices Corp. Tel. 800-635-5577 Fax 408-747-3601 United Kingdom Molecular Devices Ltd. Tel. +44-118-944-8000 Fax +44-118-944-8001 Germany Molecular Devices GMBH Tel. +49-89-9620-2340 Fax +49-89-9620-2345 Japan Nihon Molecular Devices Tel. +06-6399-8211 Fax +06-6399-8212 www.moleculardevices.com ©2002 Molecular Devices Corporation. Printed in U.S.A. #0120-1293A SpectraMax, SoftMax Pro, Vmax and Emax are registered trademarks and VersaMax, Lmax, CatchPoint and Stoplight Red are trademarks of Molecular Devices Corporation. All other trademarks are proprty of their respective companies. complete solutions for signal transduction assays AN EXAMPLE USING THE CATCHPOINT CYCLIC-AMP FLUORESCENT ASSAY KIT AND THE GEMINI XS MICROPLATE READER The Molecular Devices family of products typical applications for Molecular Devices microplate readers offers complete solutions for your signal transduction assays. Our integrated systems γ α β s include readers, washers, software and reagents. GDP αs AC absorbance fluorescence luminescence GTP PRINCIPLE OF CATCHPOINT CYCLIC-AMP ASSAY readers readers readers > Cell lysate is incubated with anti-cAMP assay type SpectraMax® SpectraMax® SpectraMax® VersaMax™ VMax® EMax® Gemini XS LMax™ ATP Plus384 190 340PC384 antibody and cAMP-HRP conjugate ELISA/IMMUNOASSAYS > nucleus Single addition step PROTEIN QUANTITATION cAMP > λEX 530 nm/λEM 590 nm, λCO 570 nm UV (280) Bradford, BCA, Lowry For more information on CatchPoint™ assay NanoOrange™, CBQCA kits, including the complete procedure for this NUCLEIC ACID QUANTITATION assay (MaxLine Application Note #46), visit UV (260) our web site at www.moleculardevices.com.
    [Show full text]
  • Enzyme Analysis by Spectrophotometry - Plate Reader Method
    Enzyme Analysis by Spectrophotometry - Plate Reader method Enzyme Extraction Equipment and reagents: Machine/Product Reference (Company, Type, …) Centrifuge – cooled Eppendorf 5415R with F45-24-11 rotor Mixer Mill / Cryo Mill Retsch MM 400 with 2x PTFE Adapter rack for 10 reaction vials 1.5 and 2.0 ml. Microtube Vortex Ika Vortex 1 Plate reader BioTek PowerWave HT with Gen 5 software Microcentrifuge tubes Safe-lock, 1.5 and 2.0 ml Pipettes (+Multistep / Multichannel) 1-5 ml; 0.1-1 ml; 5-100 µl 8-tube strip PCR tubes Crushed Ice Liquid Nitrogen (+ thermos jar) PVP (Polyvinylpyrrolidone, PVP40) Sigma PVP40T TRIS (Tris(hydroxymetyl)-aminoethane) -1 99 %, 121.14 g.mol Sigma T1378 Na 2-EDTA (disodium salt dehydrate) -1 99 %; 372.2 g.mol Sigma ED2SS DTT (1,4-Dithiothreitol) -1 ≥ 99 %; 154.24 g.mol Sigma 43815 Hydrochloric Acid concentrated HCl -1 -1 -1 37 %; 1,19 g.ml ; 36.46 g.mol = 12.08 mol.l VWR 20252.290 Most buffers can be prepared in advance and kept at 4°C Requirement: the samples must be collected in 1.5 or 2.0 ml Eppendorf tubes at harvest. Prepare reagent working solutions: Extraction buffer (0.1 M TRIS ; 1 mM Na 2-EDTA; 1 mM DTT, pH 7.8) Dissolve 12.114 g TRIS (121.14 g/mol) + 0.3722 mg EDTA (372 g/mol) + 0.1542 g DTT (154.2 g/mol) in 900 ml distilled water, adjust to pH 7.8 with HCl (use 5M HCl) and dilute to 1 liter. Keep the extraction buffer on ice.
    [Show full text]
  • Microlab 500-Series
    MicroLab 500-series Getting Started 2 Contents CHAPTER 1: Getting Started Connecting the Hardware…………………………………………………………………………….…...6 Installing the USB driver…………………….…………………………………………………...………..6 Installing the Software…………………………………………………………………………..………...8 Starting a new Experiment………………………………………………………………………..……….8 CHAPTER 2: Setting up a MicroLab Experiment Adding Input Sources Sensor Variables……………………………………………………………………………….…………10 Adding a Sensor ………………………………………………………………………………………….10 Calibrating a Sensor………………………………………………………………………………………11 The Calibration Module………………………………………………………………………………......11 Adding a Calibration Point……………………………………………………………………………….11 Correlating the Data……………………………………………………………………………………....12 Adding More Variables………………………………………………………………………………...…14 Formula Variables………………………………………………………………………………………...15 Designing an Experiment The Programming Steps………………………………………………………………………..………..17 Displaying the Data…………………………………………………………………………..…….……..17 Controlling an Experiment Starting the Experiment…………………………………………………………………….………..…..18 Stopping the Experiment…………………………………………………………………….………......18 Repeating an Experiment………………………………………………………………………………..18 Switches A & B……………………………………………………………………………………………19 Analyzing the Data Graph Analysis…………………………………………………………………………..………..………19 Add a Curve Fit………………………………………………………………………….....……19 Data Analysis…………………….…………………………………………………………………..……22 CHAPTER 3 - Spectrophotometry (522/516) Calibrating the Spectrophotometer Reading a Blank………...…………………………………………………………………………..…….23 3 Reading
    [Show full text]
  • 2017 Automated Liquid Handling Resource Guide
    The Lab Manager AUTOMATED LIQUID HANDLING RESOURCE GUIDE What You Need to Know When Buying an Automated Liquid Handling System BY RYAN ACKERMAN Upgrading Stand-Alone Automated Liquid Handling Systems to Workstations BY ANGELO DEPALMA, PhD Small-Volume Liquid Handling Requires Advanced Technology BY MIKE MAY, PhD 20 10 Common Liquid Handling Errors & How to Avoid Them 17 BY LAB MANAGER The Latest in Pipette Tip Design BY ANGELO DEPALMA, PhD Automated Liquid Handling Product Finder BY LAB MANAGER Automatic Liquid Handling Resource Guide 2017 What You Need to Know When Buying an Automated Liquid Handling System Automated liquid handlers come in a seemingly endless variety of configurations, with many different specifications. By Ryan Ackerman MAINTENANCE TIP: AUTOMATED LIQUID HANDLING The signs that you should get your automated liquid handler serviced are fairly obvious. If the system is constantly experiencing glitches or producing inconsistent, unreliable results, it’s probably time to do some maintenance. Mechanical problems are another hint that you may want to call your service technician—these issues can include pipette tips being out of calibration, the deck not being “framed” correctly, belts being worn out, or the pipette stages being out of alignment. Why is it important to know the typical sample How does the sensitivity to contamination affect the type volume being used? of automated liquid handler required? Automated liquid handlers come in a seemingly endless variety of As the instrumentation used to analyze samples becomes more configurations, with many different specifications. An important one sensitive, the methods used to transport and prepare the samples to consider is what minimum, or maximum dispensing volume is must become resistant to contamination to ensure no carryover correct for your processes.
    [Show full text]
  • Validation of Spectrophotometric Microplate
    a Food Science and Technology ISSN 0101-2061 DDII: httpI://dx.doi.org/10.1590/1678-457X.36216 Validation of spectrophotometric microplate methods for polyphenol oxidase and peroxidase activities analysis in fruits and vegetables Érica Sayuri SIGUEMDTD1, Jorge Andrey Wilhelms GUT1,2* Abstract Enzymes polyphenol oxidase (PPD) and peroxidase (PDD) play important roles in the processing of fruits and vegetables, since they can produce undesirable changes in color, texture and flavor. Classical methods of activity assessment are based on cuvette spectrophotometric readings. This work aims to propose, to validate and to test microplate spectrophotometric methods. Samples of apple juice and lyophilized enzymes from mushroom and horseradish were analyzed by the cuvette and microplate methods and it was possible to validate the microplate assays with satisfactory results regarding linearity, repeatability, accuracy along with quantitation and detection limits. The proposed microplate methods proved to be reliable and reproducible as the classical methods besides having the advantages of allowing simultaneous analysis and requiring a reduced amount of samples and reactants, which can beneficial to the study of enzyme inactivation in the processing of fruits and vegetables. Keywords: microplate; polyphenol oxidase; peroxidase; enzymatic assay; apple juice. Practical Application: Rapid assessment of PDD/PPD activities in fruits and vegetables with reduced amounts of reactants. 1 Introduction Monitoring the activity of enzymes polyphenol oxidase 2002). The selection of the appropriate substrate is also important, (PPD, E.C.1.14.18.1) and peroxidase (PDD, E.C.1.11.1.7) is an seeing that the colored compounds formed from the oxidation important control point for harvesting, storing and processing products have their maximum absorption at different wavelengths of fruits and vegetables.
    [Show full text]
  • Q & a – Accuracy and Precision
    Food Analysis – FScN 4312W Laboratory: Assessment of Accuracy and Precision Key to Questions 1. Theoretically, how are standard deviation, coefficient of variation, mean, percent relative error, and 95% confidence interval affected by: (1) more replicates, and (2) a larger size of measurement? Was this evident in looking at the actual results obtained using the volumetric pipettes and the buret, with n = 3 vs n = 6, and with 1mL vs 10mL? Ans: (1) As the sample size increases (more replicates) - Calculated mean approaches the true mean - Standard deviation is inversely proportional to the square root of sample size, hence it decreases - CV decreases, as standard deviation approaches 0 for larger sample size - % Relative error decreases as calculated mean approaches true mean - 95% confidence interval narrows down (2) Larger size of measurement - Mean is close to true mean - Standard deviation might be larger due to larger size - CV decreases - % Relative error decreases, as calculated mean is close to true mean - 95% confidence interval narrows down Not all of the above assumptions were confirmed in this experiment according to the data observed. One reason could be that the sample size was very small (n = 3 and n = 6), to notice significant differences, another reason could be human error and variation from one person’s measuring technique to another. 2. Why are percent relative error and coefficient of variation used to compare the accuracy and precision, respectively, of the volumes from pipetting/dispensing 1 and 10mL with the volumetric pipettes and buret in parts 2 and 3, rather than simply the mean and standard deviation, respectively? Ans: Mean calculated from the readings gives the calculated mean, which may differ from the true mean.
    [Show full text]
  • Introducing Automation to Your Lab
    Introducing Automation to Your Lab A step-by-step reference guide for the 21st Century biologist Written by Opentrons © OPENTRONS 2019 | INTRODUCING AUTOMATION TO YOUR LAB 1 INTRODUCTION Table of Contents CONTENT FIGURES Introduction Dispelling Myths About Lab Automation 03 Figure 1 Busting The Top 5 Automation Myths 03 Chapter 1 The Overall Benefits Of Lab Automation 05 Figure 2 Top 5 Benefits Of Lab Automation 05 Chapter 2 Choosing The Best Workflow To Automate 06 Figure 3 Your Workflow Is A Good Candidate For 06 Automation If... Chapter 3 Precision And Lab Automation 08 Figure 4 Translating Manual Processes To Automation 07 Chapter 4 How Much Liquid To Move? 09 – Volume Ranges Figure 5 Levels of Throughput 09 – Low, Medium, or High Throughput? Figure 6 Robot Pricing Table 12 Chapter 5 Space Considerations 10 Figure 7 The Opentrons Approach: 14 – Sterility Figure 8 Next Steps When Purchasing An Automated 16 – Size Lab Robot Chapter 6 Cash Considerations 11 Chapter 7 Developing Your Protocol 14 Chapter 8 Choosing The Right Vendor 15 Conclusion Next Steps On The Path To Automation 16 Appendix Resources 17 © OPENTRONS 2019 | INTRODUCING AUTOMATION TO YOUR LAB 2 INTRODUCTION Dispelling the Myths of Lab Automation “Repetitio est mater studiorum,” says the Latin FIGURE 1 proverb: “Repetition is the mother of all learning.” Busting the top 5 automation myths Scientists live this ancient saying every day, winning hard-earned discoveries by repeating the same process over and over. In some cases, this simply MYTH REALITY builds a large enough sample size to create statistical significance in the results.
    [Show full text]
  • 2021 Product Guide
    2021 PRODUCT GUIDE | LIQUID HANDLING | PURIFICATION | EXTRACTION | SERVICES TABLE OF CONTENTS 2 | ABOUT GILSON 56 | FRACTION COLLECTORS 4 | COVID-19 Solutions 56 | Fraction Collector FC 203B 6 | Service Experts Ready to Help 57 | Fraction Collector FC 204 7 | Services & Support 8 | OEM Capabilities 58 | AUTOMATED LIQUID HANDLERS 58 | Liquid Handler Overview/selection Guide 10 | LIQUID HANDLING 59 | GX-271 Liquid Handler 11 | Pipette Selection Guide 12 | Pipette Families 60 | PUMPS 14 | TRACKMAN® Connected 60 | Pumps Overview/Selection Guide 16 | PIPETMAN® M Connected 61 | VERITY® 3011 18 | PIPETMAN® M 62 | Sample Loading System/Selection Guide 20 | PIPETMAN® L 63 | VERITY® 4120 22 | PIPETMAN® G 64 | DETECTORS 24 | PIPETMAN® Classic 26 | PIPETMAN® Fixed Models 66 | PURIFICATION 28 | Pipette Accessories 67 | VERITY® CPC Lab 30 | PIPETMAN® DIAMOND Tips 68 | VERITY® CPC Process 34 | PIPETMAN® EXPERT Tips 70 | LC Purification Systems 36 | MICROMAN® E 71 | Gilson Glider Software 38 | DISTRIMAN® 72 | VERITY® Oligonucleotide Purification System 39 | REPET-TIPS 74 | Accessories Overview/Selection Guide 40 | MACROMAN® 75 | Racks 41 | Serological Pipettes 43 | PLATEMASTER® 76 | GEL PERMEATION 44 | PIPETMAX® CHROMATOGRAPHY (GPC) 76 | GPC Overview/Selection Guide 46 | BENCHTOP INSTRUMENTS 77 | VERITY® GPC Cleanup System 46 | Safe Aspiration Station & Kit 47 | DISPENSMAN® 78 | EXTRACTION 48 | TRACKMAN® 78 | Automated Extraction Overview/ 49 | Digital Dry Bath Series Selection Guide 49 | Roto-Mini Plus 80 | ASPEC® 274 System 50 | Mini Vortex Mixer 81 | ASPEC® PPM 50 | Vortex Mixer 82 | ASPEC® SPE Cartridges 51 | Digital Mini Incubator 84 | Gilson SupaTop™ Syringe Filters 86 | EXTRACTMAN® 52 | CENTRIFUGES 52 | CENTRY™ 103 Minicentrifuge 88 | SOFTWARE 53 | CENTRY™ 117 Microcentrifuge 88 | Software Selection Guide 53 | CENTRY™ 101 Plate Centrifuge 54 | PERISTALTIC PUMP 54 | MINIPULS® 3 Pump & MINIPULS Tubing SHOP ONLINE WWW.GILSON.COM 1 ABOUT US Gilson is a family-owned global manufacturer of sample management and purification solutions for the life sciences industry.
    [Show full text]
  • Laboratory Equipment Used in Filtration
    KNOW YOUR LAB EQUIPMENTS Test tube A test tube, also known as a sample tube, is a common piece of laboratory glassware consisting of a finger-like length of glass or clear plastic tubing, open at the top and closed at the bottom. Beakers Beakers are used as containers. They are available in a variety of sizes. Although they often possess volume markings, these are only rough estimates of the liquid volume. The markings are not necessarily accurate. Erlenmeyer flask Erlenmeyer flasks are often used as reaction vessels, particularly in titrations. As with beakers, the volume markings should not be considered accurate. Volumetric flask Volumetric flasks are used to measure and store solutions with a high degree of accuracy. These flasks generally possess a marking near the top that indicates the level at which the volume of the liquid is equal to the volume written on the outside of the flask. These devices are often used when solutions containing dissolved solids of known concentration are needed. Graduated cylinder Graduated cylinders are used to transfer liquids with a moderate degree of accuracy. Pipette Pipettes are used for transferring liquids with a fixed volume and quantity of liquid must be known to a high degree of accuracy. Graduated pipette These Pipettes are calibrated in the factory to release the desired quantity of liquid. Disposable pipette Disposable transfer. These Pipettes are made of plastic and are useful for transferring liquids dropwise. Burette Burettes are devices used typically in analytical, quantitative chemistry applications for measuring liquid solution. Differing from a pipette since the sample quantity delivered is changeable, graduated Burettes are used heavily in titration experiments.
    [Show full text]
  • Thermo Scientific Pipetting Guide
    Thermo Scientific Pipetting Guide Tips for Good Laboratory Pipetting Part of Thermo Fisher Scientific Thermo Scientific Thermo Scientific Contents Introduction 3 Pipetting Pipetting Guide Pipetting terms 3 Types of pipettes 4 General pipetting guidelines 6 Pipetting techniques 6 Tip information 7 Recommendations for pipetting different compounds 8 Ensuring optimum performance 8 Usage of Finpipette Novus 9 Factors affecting the accuracy of air displacement pipettes 10 Preventing cross-contamination 11 Maintenance of your Finnpipettes 1 Autoclaving 1 UV resistance 1 Calibrating your pipettes (incl. conversion table) 13 On-line pipette calibration 15 General guidelines for decontaminating pipettes 16 Chemical compatibility of plastics 18 Customer support 19 Trouble shooting 19 Thermo Scientific Over 35 Years of Innovation A leader in pipetting For over 35 years we have led the way in liquid handling products and microplate instrumentation. We have always ensured that innovation, ergonomics, accuracy, precision and safety are key aspects of our products’ designs. In 1971 we introduced Thermo Scientific Finnpipettes, the world’s first continuously variable micropi- pettes. In 1976 we introduced the world’s first multichannel pipette. Since then we have continuously improved our pipettes, always leading the way with ergonomic design. Over the last 35 years, innovations such as the improved finger rest, soft-touch tip ejection and super blow-out have made Finnpipettes increasingly user-friendly. Our intensive research program and commitment to our customers forms the foundation for future innovations. The new demands in pipette applications are the key drivers of our product development. To date, over 3 mil- lion Finnpipettes have been sold in 150 countries.
    [Show full text]