Gufor Primarideliy Health Care Clinicsn in Ieraq

DISCLAIMER This guideline has made possible through support provided by the U.S. Agency for International Development (USAID) under Primary Health Care Project in Iraq (PHCPI) implemented by University Research Co,. LLC. This guideline has been developed in Iraq in close collaboration with the Ministry of Health (MoH) in 2013 Table of contents Table of contents ...... 1 Acronyms ...... 6 1. Introduction ...... 9 2. Setting up a primary health care laboratory………………………………………..10 3. Safety in the Laboratory ...... 12 3.1 Physical Hazards ...... 12 3.2 Chemical Hazards ...... 13 3.3 Biological Hazards ...... 14 3.4 Safety Measures ...... 15 3.5 Cleaning, disinfection, sterilization and disposal of biohazard waste...... 19 3.6 Disposalof BiohazardWaste ...... 22 4. ...... 26 4.1Collection & Handling of blood ...... 26 4.2Part One: Automated Blood Count Techniques ...... 29 4.2.1 Hemoglobin Concentration ...... 30 4.2.2 Count ...... 30 4.2.3 Counting Systems ...... 30 4.2.4 Reliability of Electronic Counters ...... 32 4.2.6 Packed Cell Volume and Mean Cell Volume ...... 35 4.2.7 Variationin Red Cell Volumes: ...... 36 4.2.8 Total White Blood Cell Count ...... 37 4.2.9 AutomatedDifferential Count ...... 37 4.2.10 New White Cell Parameters ...... 39 4.2.11 Platelet Count ...... 39 4.2.12 Mean Platelet Volume ...... 40 4.2.13 Reticulocytes Count ...... 41 4.2.14 Immature Reticulocyte Fraction ...... 41 4.2.15 Calibrationof Automated Blood Cell Counters ...... 42 4.2.16 Flagging of Automated Blood Counts ...... 43 4.3 Part Two: Manual Techniques ...... 44 4.3.1 Haemoglobin Estimation ...... 44 4.3.2 Packed Cell Volumeor Haematocrit ...... 49 4.3.3 Total White Blood Cell Count ...... 522 4.3.4 PlateletCount ...... 566 4.3.5 Erythrocyte Sedimentation Rate ...... 60

1 | Page PHCPI Laboratory Standards Guideline

4.3.6 ReticulocyteCount ...... 644 4.3.7 Preparation of blood films ...... 677 4.3.8. Antiglobulin or Coombs' Test ...... 700 4.3.9. ABO & Rh Group ...... 722 5. Serology ………………………………………………………………………………...74

5.1Rapid Plasma Reagin (RPR) TEST ...... 744 5.2 TPHA(Treponema pallidum Haemagglutination Assay ) TEST ...... 777 5.3Widal Test (( test))...... 800 5.4Rose Bengal Test for Brucellosis ...... 844 5.5C-Reactive Protein (CRP) Test ...... 877 5.6ASO latex slide agglutination test ...... 89 5.7 Rheumatoid Factor (RF) ...... 900 5.7 Toxoplasmosis Test ...... 933 5.8Infectious Mononucleosis test ...... 977 5.9 Rota Virus Latex Test ...... 99 5.10Pregnancy Test ...... 1000 5.10.1 Rapid test for detection of human Chorionic Gonadotrophin in Urine . 1000 5.10.2 Direct hCG Latex Pregnancy Kit ...... 1011 6. Bacteriology ...... 1033 6.1General urine examination ...... 1033 6.1.1 Macroscopical Examination of Urine ...... 1055 Bilirubin Test ...... 10909 Bence - Jones protein Test...... …...109

Specific gravity test ...... 10910 6.1.2 Microscopical Examination of Urine ...... 1144 6.2 General Stool Examination ...... 1266 6.2.1Macroscopical examination ...... 1277 6.2.2 Microscopic Examination ...... 1300 6.3Urine Culture ...... 1344 6.4Throat Swab...... 1465 6.5 Nasal Swab ...... 1555 6.6. SMART Tests ...... 1600 6.6.1 Helicobacter pylori Serum/Whole Blood ...... 1600 6.6.2Salmonella TyphiAntigen………………………………………………………..1633

2 | Page PHCPI Laboratory Standards Guideline

6.6.3 Vibrio CholeraeRapid Visual Antigen Detection Test ...... 1666 6.7Preparation of CultureMedia ...... 1722 6.7.1 Blood agar: Tryptic Soya Agar(TSA) with 5% sheep blood ...... 1733 6.7.2 Chocolate agar with TSA base and growth supplement ...... 1744 6.7.3 MacConkey agar (MAC) ...... 1755 6.7.4 Mueller-Hinton agar ...... 1766 6.7.5 Cary-Blair medium, Amies medium, and Stuarts medium ...... 1777 6.8 Staining of bacteriological smears ...... 1777 6.8.1 Gram’s Stain ...... 1777 6.8.2 Ziehl-Neelsen’s stain (for acid fast bacilli) ...... 1788 7 Parasitology ...... 1800 7.1 Malaria parasite diagnosis ...... 1800 8. Biochemistry ...... 1888 8.1 Blood glucose ...... 1888 8.2 Function Test ...... 1911 8.2.1 Alanine Aminotransferase (ALT) ...... 1911 8.2.2 Aspartate Aminotransferase (AST) ...... 1966 8.2.3 Alkaline Phosphatase ...... 2022 8.2.4 Total Serum Bilirubin ...... 2044 8.3 Renal Function test ...... 2077 8.3.1 Blood Urea ...... 2077 8.3.2 Serum Creatinine ...... 2100 8.3.3 Uric acid ...... 2122 8.4 Lipid Profile...... 2144 8.4.1 Total Cholesterol ...... 2144 8.4.2 Triglycerides ...... 2177 8.4.3 HDL Cholesterol ...... 21919 8.4.4 LDL Cholesterol ...... 2222 Annex 1 ...... 2266 References: ...... 2377

3 | Page PHCPI Laboratory Standards Guideline

List of Figures

Figure 1: Plan for a one room laboratory……………………………………………………10

Figure 2: Indicates another possible arrangement of a peripheral laboratory ...... 11 Figure 3: Common symbols of hazards ...... 17 Figure 4: Rinsing the ………………………………………………………………….22

Figure 5:Anterior surface of the left arm showing veins most suitable for venepuncture ...... 27 Figure 6: Venepuncture technique using needle and syringe...... 28 Figure 7: Venepuncture technique using an evacuated container ...... 28 Figure 8:The areas of the foot of a baby or infant which are suitable for obtaining capillary blood...... 28 Figure 9: Effect of threshold discrimination...... 31 Figure 10: Setting Discrimination Thresholds ...... 34 Figure 11: Counting chamber ...... 544 Figure12: Counting Chamber ...... 58 Figure 13: ...... 722 Figure 14: Results of Pregnancy Test (PT) ...... 101 Figure 15: Adhesive bag…………………………………………..………………………. 103

Figure 16: Urine color chart……………………………………………………………….. 105

Fig17: common urine cells ...... 1188 Fig18: different casts seen in urine ...... 11919 Fig19: common urine crystals ...... 1222 Fig20: Infectious agents ...... 125 Figure21: Ova of Enteric Parasites ...... 1344 Figure 22: Entamoeba histolytica & Giardia lamblia (HPF) ...... 1344 Fig 23: Identification of Streptococci and Enterococci Flowchart ...... 1422 Fig 24: Presumptive Identification of Staphylococcus Species Flowchart ...... 1433 Fig 25: Identification of Enterobacter flow chart……………………………………………144

Fig 26: Identification of Glucose non-fermenting gram negative rods………………….…..145

4 | Page PHCPI Laboratory Standards Guideline

Figure 27: Gram stain smear of Vincent Angina……………………………………………149

Figure 28: Corynebacterium diphtheriae cells stained by Albert technique……………….153

Figure 29: A flow diagram for the preliminary identification of staphylococci ...... 1588 Figure 30: Interpretation of Smart Test Results ...... 1622 Figure 31: Interpretation of Salmonella Smart Test ...... 165 Figure 32: Vibrio Cholera Rapid Visual Antigen Detection test…………………………..169

Figure 33:Visual Interpretation of Rapid Dipstick Test Results ...... 1722 Figure 34: P. Falciparum ...... 1855 Figure 35: P. Ovale ...... 1866 Figure 36: P. Vivax...... 1877

List of Tables

Table1: Dilutions of haemoglobin cyanide (HiCN)…………………………………………..46

Table 2: Factors affecting the microhematocrit………………………………………………51

Table 3: TPHA results reading ...... 79

Table 4:RecordWidal (agglutination) test…………………………………………………. 81 Table 5: Determination of anti-Salmonella species antibodies titre (Widal test) ...... 822 Table 6: Example of Widal slide method ...... 822 Table 7: indication of significant serum titers ...... 833 Table 8: Determination ofanti-Brucella antibodies titre (Rose Bengal Test) ...... 866 Table 9: Examples of Rose Bengal results ...... 86 Table 10: Reagents of Infectious Mononucleosis Test ...... 977 Table 11: Degree of Albumin Precipitation ...... 1099 Table 12: Differentiation of β-haemolytic streptococci ...... 1522 Table 13: Biochemical identification of clinically significant Corynebacteria…………….154

Table 14: Culture media, conditions and organisms for all nasal specimens ...... 1566 Table 15: Differentiation of medically important species of Staphylococcus ...... 1577 Table 16: Expected Test Result Outcomes for the Dipstick and interpretaion of test results 170

5 | Page PHCPI Laboratory Standards Guideline

Acronyms

AAFB Acid and Alcohol Fast Bacilli AIDS Acquired Immune Deficiency Syndrome ALP Alkaline Phosphatase ALT Alanine Aminotransferase AST Aspartate Aminotransferase ATCC American Type Culture Collection BHSP Basic Health Services Package CRP C‐ Reactive Protein CSF Cerebro‐Spinal Fluid CV Coefficient of Variation DNA Deoxyribonucleic Acid EDTA Ethylenediaminetetraacetic Acid ELISA Enzyme‐Linked Immunosorbent Assay ESR Erythrocyte Sedimentation Rate GIT Gastro Intestinal Tract Hb Hemoglobin hCG human Chorionic Gonadotropin Hct Hi Meth hemoglobin HiCN Hemiglobincyanide HIV Human Immune defficiency Virus HPF High Field Power LDL Low Density Lipoprotein IDDM Insulin Dependent Diabetes Miletus Ig Immunoglobulin IHA Indirect Haemogglutination IU International Unit KIA Klinger Iron Agar LDH Low Density Lipids LPS Lipopolysaccharides MAC MacConky

6 | Page PHCPI Laboratory Standards Guideline

MCH Mean Corpuscular Hemoglobin MCV MIU Motility‐Indole‐Urase MOH Ministry of Health MPV Mean Platelet Volume MPXI Mean Peroxidase Activity Index MRSA Methicillin Resistant Staph aureus MSA Mannitol Salt Agar NIDDM Non‐Insulin Dependent Diabetes Mellitus NRBS Nucleated Red Blood Cells O.D Optical Density PCV Packed Cell Volume PDW Platelet Distribution Width PHC Primary Health Care PPE Personal Protective Equipment

R Reactive RBCs Red Blood Cells RDW Red cell Distribution Width RF Rheumatoid Factor RNA Ribonucleic Acid RPM Round Per Minute RPR Rapid Plasma Reagin SD Standards Deviation SHb Sulphemoglobin SMART Sensitive Membrane Antigen Rapid Test SOPs Standard Operating Procedures SPA Suprapubic Aspirate SR Strong Reactive TCBS Thiosulfate Citrate Bile Salts TNCC Total Nucleated Cell Count TPHA Treponema Pallidum Particle Agglutination TSA Tryptic Soya Agar URC University Research Company USAID United States Agency of InternationalDevelopment

7 | Page PHCPI Laboratory Standards Guideline

UTI Urinary Tract Infection VSA Viral Structural Antigens VDRL Venereal Disease Research Laboratory WBC White Blood Cells XLD Xylose Lysine Desoxycholate

8 | Page PHCPI Laboratory Standards Guideline

1. Introduction The Iraqi Ministry of Health (MoH) is undergoing health reform and moving forward with reorganizing and stabilizing its health services. Ensuring access to routine, high-quality, and equitable healthcare has emerged as a critical need, and the Government of Iraq is responding with vigor and commitment to improving the quality of primary health care (PHC) services. To assist with these efforts, USAID awarded University Research Co. LLC (URC) the four-year Primary Health Care Project in Iraq (USAID/PHCPI) to support the MoH in achieving its goal of improving the quality of care for all Iraqis. The project aims to strengthen health sector capacity to provide essential PHC services and expand availability of such services throughout the country, with increasing coverage for vulnerable and at-risk populations, such as children under five years, women, and internally displaced persons. The "Laboratory Standard Operating Procedures (SOPs)," often called the "Methods," discusses how the experiment occurred. Documenting the procedures of the laboratory experiment is important not only so that others can repeat the results but also so that you can replicate the work later, if the need arises. Historically, laboratory procedures have been written as first-person narratives as opposed to second-person sets of instructions. The Basic Health Services Package (BHSP) for Iraq was considered in developing this guideline as all the SOPs for the investigations are listed in the BHSP.

9 | Page PHCPI Laboratory Standards Guideline

2. Setting up a primary health care laboratory A one room laboratory: Figure 1 sets out the possible arrangement of a peripheral medical laboratory attached to a health center. It shows a laboratory suitable for carrying out some or all of the techniques described in the manual. The plan is limited to one room, since often this is all the space that is available for the laboratory. The room should measure at least 5m X 6m

Figure 1: Plan for a one room laboratory

Alternative plan for a one-room laboratory ( check figure2) 1: outpatient’s table; 2: hand-operated centrifuge; 3: ; 4: hematology area; 5: ; 6: water bath ; 7: electric centrifuge; 8: syphilis serology and biochemistry area; 9: reagent refrigerator; 10: reagent shelf; 11: glassware shelf; 12: balance; 13: staining box; 14: area for examination of sputum specimens; 15: ; 16: sinks; 17: waste sink; 18: bed for patients; 19: record-keeping area; 20: area for examination of stool specimens; 21: area for examination of urine specimens; 22: area for reception of specimens; 23: gas .

10 | Page PHCPI Laboratory Standards Guideline

Figure 2: Indicates another possible arrangement of a peripheral laboratory.

11 | Page PHCPI Laboratory Standards Guideline

3. Safety in the Laboratory Quite common hazards and accidents can occur in the laboratory. The need and importance of laboratory safety should be thereal concern of the laboratory. Laboratory accidents andhazards are controlled by the use of:  Simple precautions  Foresight (prudence)  Safety devices

Above all a ‘real concern’ or a ‘built in concern’ or ‘safetymindedness’ for oneself and the other fellow working at thenext bench is very essential. Laboratory accidents range from minor injury, illness or loss ofbody parts to death. There are many factors that contribute tolaboratory accidents. These include:  Poorly designed laboratory  Overcrowding of materials  Poor training  Lack of concentration  Noisy and untidy working environment  Carelessness and neglect  Overwork and fatigue  Hot and humid climatic conditions  Hurrying to finish work on time  Emergency condition (especially during night hours)

Source of Laboratory Hazards  Physical  Chemical  Biological

3.1 Physical Hazards A. Poorly Designed Laboratory Buildings Due attention should be given in the design of . The management, the laboratory personnel and thearchitecture should be involved in the standard design of thelab. B. Burns Burns may be caused by:  Flammable chemicals and stains, or by reagents catchingalight.

12 | Page PHCPI Laboratory Standards Guideline

 Fires from spirit lamps, Bunsen burners, lighted tapers (e.g. when heating Ziehl-Neelsen stain), or from faultyelectrical equipment or overloaded circuits. Spirit burnersshould not be used in direct sunlight because in brightlight the flame can be difficult to see.  Corrosive chemicals being spilt on the skin or ingestedwhen mouth –pipetting. C. Electric shock Electric shock can be caused by:  Faulty electrical circuits.  Incorrect installation of equipment.  Touching exposed live wires. D. Cuts Cuts may be caused by:  Breakages.  Using glassware that is cracked or has damaged edges.  Walking on glass chippings.

3.2 Chemical Hazards A. Toxic harmful chemicals Inhaling fumes from toxic chemicals Ingesting toxic chemicals by mouth pipetting Skin contact B. Explosive chemicals Injury from explosions can be caused by: Incompatible chemical exploding Leaking gas C. Flammable chemicals causing fire Burns D. Kinds of chemicals a. Corrosive (strong acids & alkalis)  Concentrated sulphuric acid  Nitric acid  Sodium hydroxide  Potassium Hydroxide b. Toxic irritating chemicals-cause death or serious ill health if swallowed, inhaled, and by skin contact.  Potassium cyanide – Chloroform  Barium chloride - Sodium azide

13 | Page PHCPI Laboratory Standards Guideline

c. Flammable chemicals  Ether – Acetone  Romanowsky stains - Methanol d. Explosive chemicals  Picric acid e. Carcinogens  Chemicals that cause cancer through ingestion,inhalation, skin contact  Proven carcinogen chemicals include benzidin, otoulidine, and Nitrophenol.  The risk is proportional to the length of exposure, frequency of exposure, and concentration of thechemical.

3.3 Biological Hazards Laboratory acquired infections Infection can be caused by:  Pathogens being inhaled in aerosols (airborne droplets) when snap-closing specimen containers, dispensing orpipetting infectious fluids, or centrifuging infectiousmaterial in open buckets. Aerosols may also be formedand inhaled following breakages or after spilling infectiousfluids. Breakages in centrifuges can be particularlyhazardous if the centrifuge is opened before the aerosolshave settled.  Pathogen being ingested from contaminated fingers, or infood that has been contaminated, e.g. by being stored in a laboratory refrigerator. Care should be taken to avoidthe fingers or other parts of the body touching infectedmaterial. Mouth-pipetting specimens and cultures is oneof the commonest ways of ingesting pathogens.  Pathogens entering the skin through needle punctures, cuts, scratches, insect bites, sores or other open skinlesions. Laboratory workers must always handle infectedneedles with great care.  Pathogens can also be acquired from unclean or nondisinfectedroom floors and walls, water taps andlaboratory benches.  Pathogens are acquired directly through careless contactswith patients or carrier staff through breathing, handcontact, etc.

14 | Page PHCPI Laboratory Standards Guideline

3.4 Safety Measures

Safely Designed and Organized Laboratory It is clear that a poorly designed laboratory and overcrowdingcan increase the risk of laboratory accidents. It is, therefore, important to know how the laboratory should be designed withregard to safety considerations. The following are some of the features.

A. Adequate floor, bench and storage space B. A floor that is well constructed with a surface that is nonslip, impermeable to liquids, and resistant to thosechemicals used in the laboratory. C. Walls those are smooth, free from cracks, impermeable toliquids and easily washable. D. A door at each end of the lab so that lab staff will not betrapped should a firebreak out. E. Adequate ventilation with windows that can be opened. F. Sectioning of the laboratory into separate rooms withplaces for patients, visitors, and reception of specimens. G. Bench surface that are without cracks, impervious, washable, and resistant to disinfectants and chemicals. H. Suitable storage facilities that include a well-ventilated, fire proof, locked store, for the storage of flammablechemicals. I. A good supply of gas, water, electric power & wall electricpoints J. Provision of protective safety cabinets, fire extinguishersat accessible points, and adequate waste disposal area, etc.

Safe use of Laboratory Equipment A. Positioning  Suitable and ideal place for operation  Avoid overcrowding of a bench with equipment  Position equipment correctly that requires special facilitieslike ventilation, shield from sunlight, and great care.

B. Installation

 Should be carried out by the supplier or by the health unitelectrician or trained lab equipment technician.  Important points to consider for safe installation  Ensure that the voltage of the new equipment is thesame as that of the electricity supply.

15 | Page PHCPI Laboratory Standards Guideline

 Check that the power required by the instrument doesnot exceed the power supply circuit of the lab.  Make sure that the equipment is wired correctly, andthe wiring system have grounded conductor.

Safe use of Electrical Equipment The supplier should demonstrate the use of an apparatus. Ifthis is not possible, the operation and service manual shouldbe carefully studied before the equipment is operated. Points to consider with regard to the safe use of electricalequipment: 1. Hands should be dry completely, and also the floor onwhich the operator is standing. 2. The electric supply must be disconnected whenperforming any maintenance and at the end of the day’swork. 3. If a fuse should blow, do not automatically put in a newone until the circuit is checked.

Safe use and Storage of Chemicals andReagents Even in the smallest lab, dangerous chemicals are useddirectly or incorporated into stains and reagents. Theseinclude highly flammable chemicals such as ether or methanol, highly corrosive chemicals such as phenol orsulphuric acid, or toxic and harmful chemicals such asformaldehyde solution. The correct handling and storage of hazardous chemicals isessential to prevent injury and damage. It is particularlyimportant to keep chemicals out of direct sunlight and avoidoverheating in chemical stores and the laboratory. Overheating can decompose many chemicals, causeexplosions, or the formation of toxic fumes.

Labeling of dangerous chemicals and reagents To reduce accidents caused by chemicals, many countrieshave introduced legislation, requiring manufacturers to labeldangerous chemicals with hazard symbols and to providesimple safety instructions. The six accepted danger symbols currently in use are toxic, corrosive, explosive, oxidizing, highly flammable, and harmful or irritant. The safe use and storage of these hazardouschemicals is presented in detail in the lecture note introduction to med lab.

16 | Page PHCPI Laboratory Standards Guideline

Highly Oxidizing Corrosive Toxic Harmful or Flammable Irritant

Figure 3: Common symbols of hazards

Preventing Laboratory Infection All specimens received in the lab should be regarded aspotentially pathogenic. For example, a blood specimen sentfor measuring hemoglobin may contain highly infectiousorganisms. Laboratory acquired infections can be prevented by:  Practicing personal hygiene  Wearing of laboratory coat and protective gloves  Safe handling of specimens and infectious materials.  Avoiding mouth-pipetting  Disposing safely of specimens and contaminatedmaterial.  Being immunized against highly infectious pathogens

Practice of Personal Hygiene

This includes:  Washing of hands and arms with soap and water beforeand after work  Wearing protective clothing and gloves wheneverpossible.

 Not wearing protective clothing outside of the lab area.

 Changing gloves when cut or infected.Covering any cuts, insect bites open sores, or wounds with a water proof adhesive dressing.  Wearing closed shoes and not walking barefoot.

17 | Page PHCPI Laboratory Standards Guideline

 Not eating, drinking, chewing gum, smoking or applyingcosmetics in any part of the lab.

Safe Handling and Disposal of Specimens Special precautions should be taken when collectingspecimens, especially blood specimens, and when testingspecimens and handling infected material. Safety measures involved are:  Careful handling and disposal of materials used forcollecting specimens.  Making contaminated materials non-infectious by usingappropriate decontamination methods.  Wearing gloves and a plastic apron when collecting bloodsuspected of having a highly infectious disease like AIDS.  Avoid contamination of fingers, other body parts, andworking surface.  Specimens suspected of having hepatitis, and AIDS must be labeled ‘HIGHRISK’.

Strict Prohibition of Mouth Pipetting Pathogens may be ingested during mouth pipetting, either bydirect aspiration or from the mouth ends of whichhave been contaminated from fingers or benches. Accidentscaused by mouth pipetting include infection, poisoning, chemical burns, and other injuries from chemicals. There aremany inexpensive and simple ways to measure and dispensesafely without mouth pipetting.

Miscellaneous  Immunization- protective inoculations against certainpathogenic organisms are necessary.  The use of signs- display suitable safety signs bothprohibitive (don’t) and commands (do).  First aid- basic practical training in first aid helps to reducesuffering and consequences of serious accidents.

Elements of a Laboratory SafetyProgram A successful laboratory safety program requires theparticipation of persons at every level of the laboratorystaff.  Safety does not occur only with the appointment of safetyofficer or safety committee.  Safety does not occur by having employees solelyperform their assigned duties in manner which they feel ismost efficient.  Laboratory safety requires the full participation of everymember of the staff.

Management Responsibility  Establish a policy relative to the design andimplementation of his safety program.  Delegate authority for implementing the program

18 | Page PHCPI Laboratory Standards Guideline

 Provide a safe and healthful work place  Provide fund for the implementation of the program  Assess the program- establish a mechanism to ensuresafety  Establish safety committee- regular report, recommendations regarding need for modification of theprogram.

Safety Officer Responsibility  Technical advisor to the program  Assist in the development of safe work method  Advise management on safety issues  Assist safety committee  Provide a variety of communication, e.g. hazard notice,safety data  Review a variety of plans which include facility designs, special equipment purchase in relation to safety.

Supervisor Responsibility  Train the staff in lab practices required for safe conduct ofwork.  Evaluate regularity of the laboratory facilities, equipment, personnel and work place.  Correct unsafe condition as fire hazards, physical hazards, and defective equipment.

Employees’ Responsibility  Use of safe equipment  Report of malfunctioning of equipment  Report injuries or exposure  Report hazard or unsafe condition to supervisors  Receiving proper safety training.  Ask questions if uncertain about safety regulations

3.5 Cleaning, disinfection, sterilization and disposal of biohazard waste.

Cleaning glassware  Instructions for cleaning:Wear protective clothing and gloves when cleaning lab equipmentglass containers (Erlenmeyer flasks, beakers, test-tubes)  pipettes

Glass containers New glassware Glassware that has never been used may be slightly alkaline.

19 | Page PHCPI Laboratory Standards Guideline

In order to neutralize it:

 Prepare a bowl containing 3 liters of water and 60 ml of concentrated hydrochloricacid (i.e. a 2% solution of acid).  Leave the new glassware completely immersed in this solution for 24 hours.  Rinse twice with ordinary water and once with demineralized water.  Dry.

Dirty glassware  Preliminary rinsing: Rinse twice in cold or lukewarm water (never rinse bloodstained tubes in hot water). If the glassware has been used for fluids containing protein, it should be rinsedimmediately and then washed (never allow it to dry before rinsing).  Soaking in detergent solution: Prepare a bowl of water mixed with washing powder or liquid detergent. Put therinsed glassware in the bowl and brush the inside of the containers with a test-tube brush Leave to soak for 2–3 hours.

Rinsing Remove the articles one by one. Rinse each one thoroughly under the tap, thensoak them all in a bowl of ordinary water for 30 minutes. Rinse each article in a stream of clean water. (Do not forget that traces of detergentleft on glassware can lead to false laboratory results.)

Draining Place containers (beakers, flasks, measuring cylinders) on the pegs of a drainingrack. Place test- tubes upside-down in a wire basket.

Drying Place the glassware in wire baskets and dry in a hot-air oven at 60 °C. Alternatively, place the baskets in a sunny spot in the laboratory and cover them with a fine cloth.

Plugging The clean dry glassware should be put away in a cupboard to protect it from dust.It is recommended that glass containers be plugged with non-absorbent cotton wool or their mouths covered with small caps made from newspaper or,preferably, thin sheets of paraffin wax or clinging plastic, if available.

Pipettes Immediate rinsing Once a pipette has been used, rinse it immediately in a stream of cold water toremove blood, urine, serum, reagents, etc.

20 | Page PHCPI Laboratory Standards Guideline

Soaking in water After rinsing, place the pipettes in a large, plastic measuring cylinder (or bowl) fullof water. If the pipettes have been used to measure infected material, leave them ina cylinder full of disinfectant solution (e.g. a quaternary ammonium compound.

Soaking in detergent and rinsing Follow the instructions given above for soaking and rinsing of laboratoryglassware.

Blocked pipettes Put blocked pipettes in a cylinder filled with dichromate cleaning solution. Slide them carefully into the solution and leave for 24 hours.

The next day, pour the dichromate solution into another cylinder (it can be usedfour times). Hold the cylinder containing the pipettes under the tap and rinse thoroughly. Remove the pipettes one at a time. Check that the obstruction has been washedaway. Rinse again. Leave to soak in ordinary water for 30 minutes, then change the water and soakfor a further 30 minutes. Warning: Dichromate cleaning solution is highly corrosive and should be used withextreme care. If it is accidentally splashed on the skin or clothing or into the eye(s), wash at once with large quantities of water.

Drying Dry heat-resistant glass pipettes in a hot-air oven at 60 °C and ordinary pipettes inan at 37 °C. Alternatively, leave pipettes to air-dry.

Using the vacuum pump This is a small instrument made of metal, plastic or glass that is attached to thewater tap.

Turn the water on hard to drive a strong jet through the pump. This causes airto be sucked into the side arm of the pump and the rubber tubing attached toit.

Fit this rubber tubing over the tip of the pipette.

3. Dip the other end of the pipette into the rinsing liquid (water or detergentsolution), which is sucked through the pipette and discharged by the pump into the sink (check the figure below).

21 | Page PHCPI Laboratory Standards Guideline

Figure 4: rinsing the pipet

3.6 Disposalof BiohazardWaste A. General rules Prior to any treatment, all biohazard wastes, including those to be incinerated, should be enclosed in a puncture-resistant, biohazard bag that is color-codedor labeled with the biological hazard symbol.

Untreated biohazard waste is not to be disposed of in the municipal waste (system) stream. All biohazard waste must be treated by chemical disinfectionor autoclaving before in any way being disposed of in the municipal wastestream.

After disinfection, but before disposal in the municipal waste stream, alltreated biohazard wastes (apart from liquids) should be enclosed in anunmarked outer bag that is not red or labeled with the biohazard symbol. Anybiohazard waste that has been treated as described below and packaged suchthat it is clearly evident that the waste has been effectively treated, is notsubject to be treated as biohazard waste and may be collected, transported forincineration or disposed of as municipal waste.

The person handling the emptying of waste bins, waste or sharpscontainers must be careful not to touch anything without protective clothing and must use gloves to avoid contracting infections from the waste.

22 | Page PHCPI Laboratory Standards Guideline

The waste bins, sharps containers, etc., must be clearly indicated/marked: Biohazard wastes.

B. Laboratory waste Discarded swabs, cotton, sample containers and other biohazard but non-sharpobjects used for sample collection should be placed in a covered trashcontainer lined with an autoclavable biohazard yellow bag. Visibly soiled or splashedtourniquets and Vacutainer holders should be discarded and new equipmentused.

Disposable plastics such as tips, culture plates, micro plates, petri dishes, testtubes, etc., used for testing of samples should be disposed of in a coveredwaste bin lined with an autoclavable biohazardyellow bag, autoclaved andtransported for incineration.

Pipettes should be soaked in special pipette baskets in 10 % household bleachovernight, before being thrown in a waste bin lined with an autoclavablebiohazard yellow bag, autoclaved and incinerated.

Reusable glassware such as cylinders, flasks and beakers should be disinfectedwith 10 % household bleach over night before washed and autoclaved.

C. Sharps All sharp implements used in the laboratory need to be handled carefully.They can easily break the skin and increase the risk of infection withinfectious agents. Broken glassware must also be handled as sharps. The laboratory shouldhave a special box for broken glassware If needles are used, they should never be recapped before disposal.Recapping needles may lead to somebody accidentally pricking him orherself. Dispose of all sharps directly into a secureperforated sharps disposal container lid. The lid only allows the sharps into the container but notout. The sharps container should be puncture-resistant, leak proof on thesides and bottom, and color-coded or labeled with a biohazard symbol. Note: When selecting sharps containers, look for special safety features such aslids that lock tight for safe disposal and a container that can be sterilized by steam,gas, or chemicals. If sharps containers are not specifically constructed to beautoclaved, the resulting mass of melted plastic is extremely hazardous due to theneedles that often protrude. When full, seal top, , then arrange for incineration. Decontamination is achieved by autoclaving the sharps container.Incineration destroys the sharps completely hence eradicating the chancesof accidental injury to people in the community.

23 | Page PHCPI Laboratory Standards Guideline

D. Residual body fluids Ensure that all tubes/containers containing residual body fluids are properlysealed to avoid spillage. These are then stored temporarily in waste binslined with disposal bags. Transfer the disposal bags, together with contents, into autoclaving bags. Then place a piece of autoclavable tape on the disposal bag to seal it and autoclave as required. After the autoclaving cycle is completed, incinerate the wholepackage to completely destroy all contents.

E. Chemical wastes The following procedure is adopted for most chemicals that are used in lowconcentrations. Collect all liquid chemical waste in properly labeled bottles with a littleconcentrated disinfectant. Keep monitoring the rising level of waste in the trap bottles. Never fill thetrap bottles to the very top to avoid spillage. Empty the contents of the trap bottle down the drain, preferably a specialsink in the laboratory, and wash down with more disinfectant, liquid soapand a large volume of water. Minimum disinfection time of any liquid Biohazard Waste is 30 minutes. Note: Some disinfectants react with organic compounds, acids, or bases. Consider thiswhen selecting your disinfectant before commencing your work. Safety Precautions Anybody handling Biohazard Wastes should always put on laboratoryprotective clothing and gloves and keep to the laboratory safety practices toavoid viral and other infectious disease transmissions. Any spills or splashes of infectious material should be immediately cleaned upwith absorbent material using an approved disinfectant such as dilute 10%bleach or chlorhexidine solution.

Decontaminating bench tops and other surfaces Bench tops should be wiped with a disinfectant (a phenolic disinfectant, 1%sodium hypochlorite [bleach], or 70% isopropyl alcohol) routinely after workingwith infectious agents or clinical specimens or after spills, splashes, or othercontamination by infectious materials. Solutions of disinfectants should bemaintained at the work station. Decontamination of spills The following procedure is recommended for decontaminating spills. • Isolate the area to prevent anyone from entering.

• Wear gloves and protective clothing (e.g., a gown or lab coat, shoes, and a mask [if the spill may contain a respiratory agent or if the agent is unknown]).

24 | Page PHCPI Laboratory Standards Guideline

• Absorb or cover the spill with disposable towels. • Saturate the towels with an appropriately diluted intermediate or high-level disinfectant (e.g., a phenolic formulation or household bleach). • Place disinfectant-soaked towels over the area and leave them in place for at least 15 minutes before removing and discarding them. • Wipe area using clean disinfectant-soaked towels and allow area to air dry. • Place all disposable materials (including towels) used to decontaminate the spill into a biohazard container. • Handle the material in the same manner as other infectious waste.

25 | Page PHCPI Laboratory Standards Guideline

4. Hematology

4.1Collection & Handling of blood General precautions:

 Always wear disposable plastic gloves, especially if there is any hand cut.  Always be careful when handling syringes, needles & lancets.  Always use disposable syringes, needles & lancets.  Always take care of specimen identification & request forms.

Technical errors:

 Faults in sample Collection (See below).  Faults in sample Transport & Storage (identification error, how long delay, what temp.).

Technical errors from sample collection:

 Insufficient or excess anticoagulant.  Inappropriate anticoagulant.  Inadequate mixing with anticoagulant.  Patient &/or specimen identification error.  Delay in transit to lab. Venous blood collection:

The phlebotomist should enter a training program. Venous blood, mixed with an anticoagulant, is preferred for most hematological examinations.

Best site is the antecubital vein Fig.4, 5, 6 & in obese subject may use the dorsum of the hand.

Needle gauge: for adults 21 (0.8mm), for children 23 (0.6mm).

Make sure of proximal & distal fixation of the vein.

Skin preferably cleaned with 70% alcohol.

Steps to facilitate obtaining a venous blood sample are:

 Warming the patient’s arm.

26 | Page PHCPI Laboratory Standards Guideline

 Using a tourniquet (should be loosened once entered the vein to avoid Tapping the skin over the site of the vein.

 Asking the patient to open & clench his fist several times.

Advantages of using the tourniquet:

- Blood filling of the vein. - Proximal fixation of the vein. Disadvantages of using the tourniquet: - Hemoconcentration. - Red cell hemolysis.

Figure 5: Anterior surface of the left arm showing veins most suitable for venepuncture.

27 | Page PHCPI Laboratory Standards Guideline

Figure6: Venepuncture technique using needle and syringe.

Figure 7: Venepuncture technique using an evacuated container

The distal end of the needle has been screwed into the holder and the proximal needle has then been unsheathed and inserted into a suitable vein.

Figure 8: The areas of the foot of a baby or infant which are suitable for obtaining capillary blood.

28 | Page PHCPI Laboratory Standards Guideline

4.2 Part One: Automated Blood Count Techniques A variety of automated instruments for performing blood counts are in widespread use. Semi automated instruments require some steps (e.g., dilution of a blood sample) to be carried out by the operator. Fully automatedinstruments require only that an appropriate blood sample is presented to the instrument. Semiautomatic instruments often measure a small number of components (e.g., WBC and Hb). Fully automated multichannel instruments usually measure from 8 to 20 components, including some variables that have no equivalent in manual techniques. Automated instruments usually have a high level of precision, which, for cell counting and cell-sizing techniques, is greatly superior to that achievable with manual techniques. If instruments are carefully calibrated and their correct operation is ensured by quality control procedures, they produce test results that are generally accurate. When blood has abnormal characteristics, the results for one or more parameters may be aberrant; instruments are designed so that such inconsistent results are “flagged” for subsequent review. The abnormal characteristics that lead to inaccurate counts vary between instruments, so it is important for instrument operators to be familiar with the types of factitious results to which their instruments are prone.

Blood cell counters may have automated procedures for sample recognition (e.g., by bar- coding), for ensuring that adequate sample mixing occurs, for taking up the test sample automatically, and for detection of clots or inadequately sized samples. Ideally, blood sampling is carried out by piercing the cap of a closed tube so that samples that carry an infection hazard can be handled with maximum safety.

Laboratories performing large numbers of blood counts each day require fully automated blood counters capable of the rapid production of accurate and precise blood counts, including platelet counts and differential counts, either three-part or five- to seven-part. The sample throughout required varies with the workload and the timing of arrival of blood specimens in the laboratory, but for most large laboratories a throughput of 100 or more samples per hour is required. Sample size and the availability of a “predilute” mode are particularly relevant if the laboratory receives many paediatric specimens. Choice of an instrument for an individual laboratory, as well as for point-of-care sites outside the laboratory, should take account of capital expenditure and running costs, including maintenance and reagents; size of instrument; requirements of services such as water, compressed air, drainage, and an electricity supply with stable voltage;environmental disturbance by generation of heat, vibration, and noise; any influence on performance by the

29 | Page PHCPI Laboratory Standards Guideline ambient temperature and humidity; storage requirements for the often bulky reagents; ease of operation; and the likely level of support that can be expected from the manufacturer.

Most automated instruments, however, count for a specified period of time rather than on an exact volume of blood; they therefore require calibration by means of the direct counts derived from instruments counting cells in a defined volume of diluted blood. For some variables, instruments are calibrated by the manufacturer, but others require calibration in the laboratory. Performance characteristics of an instrument vary over time, so periodic recalibration is needed; both when quality control procedures indicate the necessity and when certain components are replaced.

4.2.1 Hemoglobin Concentration Most automated counters measure haemoglobin by a modification of the manual HiCN method with cyanide reagent or with a nonhazardous chemical such as sodium lauryl sulphate, which avoids possible environmental hazards from disposal of large volumes of cyanide-containing waste. Modifications include alterations in the concentration of reagents and in the temperature and pH of the reaction. A nonionic detergent is included to ensure rapid cell lysis and to reduce turbidity caused by cell membranes and plasma lipids. Measurements of absorbance are made at a set time interval after mixing of blood and the active reagents but before the reaction are completed.

4.2.2 Red Blood Cell Count Red cells and other blood cells can be counted in systems based on either aperture impedance or light-scattering technology. Because large numbers of cells can be counted rapidly, there is a high level of precision. Consequently, electronic counts have rendered the RBC and the red cell indices derived from it (the MCV and the MCH) of much greater clinical relevance than was possible when only a slow and imprecise manual RBC was available.

4.2.3 Counting Systems

Impedance Counting Impedance counting, first described by Wallace Coulter in 1956 depends on the fact that red cells are poor conductors of electricity, whereas certain diluents are good conductors; this difference forms the basis of the counting systems used in Beckman–Coulter, Sysmex, Abbott, Roche, and a number of other instruments.

30 | Page PHCPI Laboratory Standards Guideline

For a cell count, blood is highly diluted in a buffered electrolyte solution. The flow rate of this diluted sample is controlled by a mercury siphon (as in the original Coulter system) or by displacement of a tightly fitting piston. This result is a measured volume of the sample passing through an aperture tube of specific dimensions (e.g., 100 mm in diameter and 70 mm in length). By means of a constant source of electricity, a direct current is maintained between two electrodes, one in the sample beaker or the chamber surrounding the aperture tube and another inside the aperture tube. As a blood cell is carried through the aperture, it displaces some of the conducting fluid and increases the electrical resistance. This produces a corresponding change in potential between the electrodes, which lasts as long as the red cell takes to pass though the aperture; the height of the pulses produced indicates the volume of the cells passing through. The pulses can be displayed on an oscillograph screen. The pulses are led to a threshold circuit provided with an amplitude discriminator for selecting the minimal pulse height, which will be counted. The height of the pulses is used to determine the volume of the red cells.

Figure 9: Effect of threshold discrimination (horizontal axis) in separating cell signals from background noise.

Light Scattering Red cells and other blood cells may be counted by means of electro-optical detectors. A diluted cell suspension flows through an aperture so that the cells pass, in single file, in front

31 | Page PHCPI Laboratory Standards Guideline of a light source; light is scattered by the cells passing through the light beam. The scattered light is detected by a photomultiplier or photodiode, which converts it into electrical impulses that are accumulated and counted. The amount of light scattered is proportional to the surface area and therefore the volume of the cell so that the height of the electrical pulses can be used to estimate the cell volume. The high-intensity coherent laser beams used in current instruments have superior optical qualities to the noncoherent tungsten light of earlier instruments. Sheathed flow allows cells to flow in an axial stream with a diameter not much greater than that of a red cell; light can be precisely focused on this stream of cells. Electro- optical detectors are used for red cell sizing and counting in Bayer-Technicon systems and for white cell differential counting in a number of other instruments.

4.2.4 Reliability of Electronic Counters Electronic counts are precise, but care needs to be taken so that they are also accurate. The recorded count on the same sample may vary from instrument to instrument and even between different models of the same instrument. Inaccuracy may be introduced by:

1-coincidence (i.e., by two cells passing through an orifice simultaneously and being counted as one cell). 2-pulse being generated during the electronic dead time of the circuit. 3-Recirculation of cells that have already been counted. 4-red cellagglutination (which causes a clump of cells to be counted as one Cell). 5-by the counting of bubbles, lipid droplets, microorganisms, or extraneous Particles as cells. 6-Faulty maintenance may lead to variation in the volume aspirated or the flow rate. 7-Single-channel instruments may have their thresholds set incorrectly, and multichannel instruments may be incorrectly calibrated. Coincidence correction; in some instruments, this is done automatically by electronic editing. Errors of coincidence can be detected by carrying out a series of measurements at various dilutions of the same specimen, plotting the data on graph paper, and then extrapolating the graph to the baseline for the true value. Alternatively, the need for coincidence correction can be avoided by having the dimensions and flow characteristics of the aperture through which the cells pass such that cells can only pass in single file; this may be achieved by sheath flow or hydrodynamic focusing in which diluted blood is injected into a sheath of fluid as it flows into the sensing zone. This induces the cells to pass through the centre of the sensing zone in single file and free of distortion. Coincidence can be more effectively reduced with sheathed flow and precisely focused light in an electro-optical detector than in an impedance counter so that less dilution of the blood sample is needed.

Electrical impulses generated by recirculation of cells can be eliminated by electronic editing.

32 | Page PHCPI Laboratory Standards Guideline recirculation of cells in the region of the aperture can be prevented by “sweep flow” in which a directed stream of diluent sweeps cells and debris away from the aperture, thus preventing cells from being recounted and debris from being counted as cells.

Inaccurate counts consequent on red cell agglutination are usually the result of cold agglutinins. They are recognized as erroneous because of an associated marked factitious elevation of the MCV. A correct count can be achieved by prewarming the blood sample and, if necessary, also prewarming the diluent.

A correct RBC and, particularly, a correct measurement of the MCV are dependent on the use of an appropriate diluent. For impedance counters, pH, temperature, and rate of ionization have to be standardized and remain constant because changes alter the electrical field and may lead to artefactual alterations in the size, shape, and stability of the blood cells in the diluent. Diluents must be free of particles and give a background count of less than 50 particles in the measured volume. The correct diluent for each individual instrument must be used; other diluents, even those made by the same manufacturer, may not be interchangeable. Any laboratories using diluents other than those recommended by the manufacturer of the instrument must satisfy themselves that no error is being introduced.

For red cell counting in simple single-channel counters a suitable diluent requires a pH of 7.0–7.5 and osmolality of 340 ± 10 mmol. Physiological saline (9 g/l NaCl) or phosphate- buffered saline, which have the advantages of simplicity and ready availability, can be used as a red cell diluent, provided that the counts are performed immediately after dilution to avoid errors owing to sphering. Commercial solutions of saline (for intravenous use) are usually particle-free. Other solutions may require filtration through a 0.22- or 0.45-mm micropore filter to remove dust.

Setting Discrimination Thresholds An accurate RBC requires that thresholds be set so that all red cells, but a minimum of other cells, are included in the count. Some counters have a lower threshold but no upper thresholdso that white cells are included in the “RBC.” Because the WBC is usually very low in relation to the RBC, this is not usually of practical importance; however, an appreciable error can be introduced if the WBC is greatly elevated, particularly if the patient is also anaemic. The setting of the lower threshold is of considerable importance because it is necessary to ensure that microcytic red cells are included in the count without also counting large platelets.

Current multichannel instruments, both impedance counters and counters using light- scattering technology, have thresholds that are either precalibrated by the manufacturer or are

33 | Page PHCPI Laboratory Standards Guideline automatically adjusted, depending on the characteristics of individual blood samples. Single- channel impedance instruments capable of performing a direct RBC require setting of thresholds so as to separate pulses generated by red cells from background noise and from pulses generated by platelets. This is done by adjusting the aperture current and the pulse amplification. A simple method is to dilute a fresh blood sample and carry out successive counts on the suspension, while the lower threshold control is moved incrementally from its maximum to its minimum position. At the maximum position, the count should be zero or close to zero, and the counts will increase as the amplitude is reduced. The counts at each setting are plotted on arithmetic graph paper.

Figure 10: Setting Discrimination Thresholds Method to establish working conditions of cell counters. The correct setting of the threshold (at arrow) is intended to exclude noise pulses without loss of the signal pulses produced by the blood cells.

The correct threshold setting is at the left of the horizontal part of the graph before the line begins to slope. It is important to check that the setting selected is valid for microcytic cells. The threshold can be defined more precisely for an individual sample by means of a pulse height analyser linked to the counting system. The lower threshold is correctly set if beyond

34 | Page PHCPI Laboratory Standards Guideline this point there are less than 0.5% of the counts at the peak (mode) of the pulse size distribution curve.

4.2.6 Packed Cell Volumeand Mean Cell Volume Modern automated blood cell counters estimate PCV by technology that has little connection with packing red cells by centrifugation. It is sometimes convenient to use different terms to distinguish the manual and automated tests, and for this reason the International Council for Standardization in Haematology has suggested that the term “haematocrit” (Hct) rather than PCV should be used for the automated measurement. However, it should be noted that, in the past, the terms “packed cell volume” and “haematocrit” have been used interchangeably for the manual procedure.

Sources of error include uneven cell distribution in the sample; too many or too few cells in the sample; subjective judgments as to whether a given cell falls within the defined counting area; contamination of the hemocytometer; variation in how a sample is loaded; and differences in user technique.

With automated instruments, the derivation of the RBC, PCV, and MCV are closely interrelated. The passage of a cell through the aperture of an impedance counter or through the beam of light of a light-scattering instrument leads to the generation of an electrical pulse the height of which is proportional to cell volume. The number of pulses generated allows the RBC to be determined, as discussed earlier. Pulse height analysis allows either the MCV or the PCV to be determined. If the average pulse height is computed, this is indicative of the MCV, and the PCV can be derived by multiplying the estimated MCV by the RBC. Similarly, if the pulse heights are summated, this figure is indicative of the PCV, and the MCV can, in turn, be derived by dividing the PCV by the RBC.

Automated instruments require calibration before the PCV or MCV can be determined. Calibration of the PCV can be based on manual PCV determinations. Alternatively, the MCV can be calibrated by means of the pulse heights generated by latex beads, stabilize cells, or some other calibrant containing particles of known size; however, unfixed human red cells that are biconcave and flexible will not necessarily show the same characteristics in a cell counter as latex particles or some other artificial calibrant. Aperture-impedance systems measure an apparent volume that is greater than the true volume, being influenced by a “shape factor; this factor is less than 1.1 for young, flexible red cells; is between 1.1 and 1.2 for fixed biconcave cells; and is about 1.5 for spheres, whether they be fixed cells or latex spheres.

The MCV, and therefore the PCV, as determined by an automated counter, will vary with certain cell characteristics other than volume. As indicated earlier, such characteristics include

35 | Page PHCPI Laboratory Standards Guideline shape, which in turn is partly determined by flexibility. With impedance counters, the normal disc-shaped red cell becomes elongated into a cigar shape as it passes through the aperture; this is caused by deformation in response to shear force, which occurs in cells of normal flexibility. Cells with a reduced haemoglobin concentration undergo more elongation than normal cells; this leads to a reduced “shape factor,” a reduced pulse height in relation to the true size of the cell, and underestimation of the MCV. Conversely, cells with abnormally rigid membranes and cells such as spherocytes with a high haemoglobin concentration will undergo less deformation than normal and the MCV will be overestimated. Earlier light-scattering instruments also underestimated the volume of red cells with a reduced haemoglobin concentration because light scattering was affected by the haemoglobin concentration. These artefacts are seen even with normal red cells of varying haemoglobin concentration but are more apparent with red cells from patients with defects in haemoglobin synthesis such as those from patients with iron deficiency. Light-scattering instruments have been developed to avoid artefacts of this type. Cells are isovolumetrically sphered so that their light-scattering characteristics are uniform and should follow the laws of physics.

The automated MCV and PCV are prone to certain errors that do not occur or are less of a problem with manual methods. These include those resulting from microclots or partial clotting of the specimen, extreme , and the presence of cryoglobulins or cold agglutinins; the last is a relatively common cause of factitious elevation of the MCV because clumps of cells are sized as if they were single cells. Because the RBC is underestimated, the PCV is less affected, although it is also inaccurate. It is rare for warm agglutinins to cause a similar problem. Sickling may cause a factitious increase in MCV and PCV, whereas alterations in plasma osmolarity occurring, for example, in severe hyperglycemias also cause factitious elevation of the MCV and PCV.

4.2.7 Variationin Red Cell Volumes: Red Cell DIistribution Width

Automated instruments produce volume distribution histograms that allow the presence of more than one population of cells to be appreciated. Instruments may also assess the percentage of cells falling above and below given MCV thresholds and “flag” the presence of an increased number of microcytes or macrocytes. Such measurements may indicate the presence of a small but significant increase in the percentage of either microcytes or macrocytes before there has been any change in the MCV.

The RDW is derived from pulse height analysis and can be expressed either as the standard deviation (in fl) or as the coefficient of variation (CV) (%) of the measurements of the red cell

36 | Page PHCPI Laboratory Standards Guideline volume. Current Beckman-Coulter and Bayer-Technicon instruments express the RDW as the SD, and Sysmex instruments express it as either the SD or the CV. The normal reference range is in the order of 12.8 ± 1.2% as CV and 42.5 ± 3.5 fl as SD. However, widely different ranges have been reported; therefore it is important for laboratories to determine their own reference ranges. The RDW expressed as the CV has been found of some value in distinguishing between iron deficiency (RDW usually increased) and thalassaemia trait (RDW usually normal) and between megaloblastic anaemia (RDW often increased) and other causes of (RDW more often normal).

4.2.8 Total White Blood Cell Count The total WBC is determined in whole blood in which red cells have been lysed. The lytic agent is required to destroy the red cells and reduce the red cell stroma to a residue that causes no detectable response in the counting system without affecting leucocytes in such a manner that the ability of the system to count them is altered. Various manufacturers recommend specific reagents, and for multichannel instruments that also perform an automated differential count use of the recommended reagent is essential.

Fully automated multichannel instruments perform WBCs by impedance or light-scattering technology or both. Residual particles in a diluted blood sample are counted after red cell lysis or, in the case of some light-scattering instruments, after the red cells have been rendered transparent. Thresholds are set to exclude normal platelets from the count, although giant platelets are included. Some or all of any nucleated red cells present are usually included, so that when nucleated red cells are present the count approximates more to the TNCC than to the WBC.

Factitiously low automated WBCs occasionally occur as a consequence of leukocyte agglutination, prolonged sample storage, or abnormally fragile cells (e.g., in leukaemia). Factitiously high counts are more common and usually result from failure of lysis of red cells. With certain instruments this may occur with the cells of neonates or be consequent on uraemia or on the presence of abnormal haemoglobin such as haemoglobin S or haemoglobin C; high counts may also be the result of microclots, platelet clumping, or the presence of a cryoglobulin.

4.2.9 AutomatedDifferential Count Most automated differential counters that are now available use flow cytometry incorporated into a full blood counter rather than being stand-alone differential counters. Increasingly, automated blood cell counters have a differential counting capacity, providing either a three- part or a five- to seven-part differential count. Counts are performed on diluted whole blood

37 | Page PHCPI Laboratory Standards Guideline in which red cells are either lysed or are rendered transparent. A three-part differential count assigns cells to categories usually designated: (a) “Granulocytes” or “large cells”; (b) “” or “small cells”; and (c) “Monocytes,” “mononuclear cells,” or “middle cells.” In theory, the granulocyte category includes eosinophils and basophils, but in practice it is common for an appreciable proportion of cells of these types to be excluded from the granulocyte category and to be counted instead in the monocyte category. Five- to seven-part differential counts classify cells as neutrophils, eosinophils, basophils, lymphocytes, and monocytes and in an extended differential count may also include large immature cells (composed of blasts and immature granulocytes) and atypical lymphocytes (including small blasts). Automated instruments performing three-part or five-to seven-part differential counts are able to “flag” or reject counts from the majority of samples with nucleated red cells, myelocytes, promyelocytes, blasts, and atypical lymphocytes. To a lesser extent, instruments incorporating a three-part differential count, although not capable of enumerating eosinophils or basophils, are able to flag a significant proportion of samples that have an increased number of one of these cell types. Automated differential counters using flow cytometry classify far more cells than is possible with a manual differential count. Automated counts are consequently much more precise than manual counts; however, with certain cell categories—specifically monocytes and basophils—the degree of precision is sometimes less than would be expected for the number of cells counted, indicating that such cells are not always classified in a consistent manner. The accuracy of automated counters is less impressive than their precision. With all types of counters, unusual cell characteristics or ageing of a blood specimen can lead to misclassification of cells. Although the majority of samples containing abnormal cells are “flagged,” this is not invariably so; the presence of nucleated red cells, immature granulocytes, atypical lymphocytes, and blasts (even occasionally quite large numbers of blasts) may not give rise to a “flag.” However, human observers performing a 100-cell manual differential count also miss significant abnormalities. In general, automated counts have compared favourably with routine manual counts, especially if the instruments are assigned only two functions—performing differential counts on normal samples and “flagging” abnormal samples. If morphological abnormalities are flagged, microscopic examination of a stained should always be undertaken. In the presence of a significant number of NRBC, the total count is neither a true “WBC” nor a true “TNCC” and the absolute WBC counts calculated from the total will necessarily be somewhat erroneous.

38 | Page PHCPI Laboratory Standards Guideline

Instruments currently in use that count NRBC and correct the WBC for NRBC interference include the Abbott CellDyn 4000, the Sysmex XE2100, and the Beckman-Coulter LH750. Differential counters based on pattern recognition in stained blood films were initially preferred by many haematologists, but they were relatively slow, and because they could count only a small number of cells in a reasonable time, the precision of the automated count was no better than that of a manual count. However, with improved computing technology and with the use of artificial neural networks, such instruments (e.g., DiffMaster, CellVision AB, SE-223 70, and Lund, Sweden) are now capable of providing a useful differential count on blood samples containing abnormal cells. Up to 30 films an hour can be processed and reviewed, and abnormal cells can be reclassified if required.

4.2.10 New White Cell Parameters Many instruments are able to “flag” the presence of atypical or “variant” lymphocytes by features such as alteration in size and in impedance or light-scattering characteristics. Automated white cell counters can also analyse cell characteristics by novel technologies and identify cell types by features that differ greatly from those used when a blood film is examined visually. It is possible, for example, to identify eosinophils by the ability of their granules to polarize lightor to detect a left shift or the presence of blasts by the reduced light scattering of the nuclei of more immature granulocytes. There is also the potential to produce information that is not directly analogous with that available from a manual differential count. Instruments that incorporate a cytochemical reaction give information on enzyme activity expressed as the mean peroxidase activity index (MPXI); standard range is -10 to +10. An increased MPXI has been observed in infections, in some myelodysplasias and leukaemias, in the acquired immune deficiency syndrome (AIDS), and in megaloblastic anaemia, whereas a reduced MPXI occurs in inherited and acquired neutrophil peroxidase deficiency.It has been suggested that the parameters used for defining leucocyte types might also allow detection of the presence of malaria pigment as a screening test in areas where malaria is prevalent.

4.2.11 Platelet Count Platelets can be counted in whole blood using the same techniques of electrical or electro- optical detection as are used for counting red cells. An upper threshold is needed to separate platelets from red cells, and a lower threshold is needed to separate platelets from debris and electronic noise. Recirculation of red cells near the aperture should be prevented, as pulses produced may simulate those generated by platelets. Three techniques for setting thresholds have been used: (a) Platelets can be counted between two fixed thresholds (e.g., between 2 and 20 fl;

39 | Page PHCPI Laboratory Standards Guideline

(b) Pulses between fixed thresholds can be counted with subsequent fitting of a curve and extrapolation so that platelets falling outside the fixed thresholds are included in the computed count. (c) Thresholds can vary automatically, depending on the characteristics of individual blood samples, to make allowance for microcytic or fragmented red cells or for giant platelets. A new method for platelet counting by flow cytometry has been developed Platelets in a blood sample are labelled fluorescently with a specific monoclonal antibody or combination of antibodies, and by measuring the RBC: platelet ratio the platelet count can be calculated. Suitable antibodies to platelet antigens are CD41, CD42, and CD61. This method using CD41 and CD61 has been adopted by the International Council for Standardization in Haematology as the reference method some instruments now provide an automated immunological platelet count for diagnostic use. Although these instruments can count platelets down to levels of 10 × 109/l or less, it should be noted that precision at these levels is often poor with CVs of 22– 66% being observed and with mean counts differing appreciably between instruments. Factitiously low automated platelet counts may be the result of giant platelets being identified as red cells, EDTA-induced platelet clumping, and satellitism. Misleadingly high platelet counts may be due to markedly microcytic or fragmented red cells, to cell fragments in leukaemia, or to bacteria or fungi.

4.2.12 Mean Platelet Volume The same techniques that are used to size red cells can be applied to platelets. The calculated mean platelet volume (MPV) is very dependent on the technique of measurement and on length and conditions of storage prior to testing the blood. When MPV is measured by impedance technology, it has been found to vary inversely with the platelet count in normal subjects. If this curve is extrapolated, it has been found that data fit the extrapolated curve when thrombocytopenia is caused by peripheral platelet destruction; however, the MPV is lower than predicted when thrombocytopenia is caused by megaloblastic anaemia or bone marrow failure. The MPV is generally greater than predicted in myeloproliferative disorders, but differentiating essential thrombocythaemia from reactive thrombocytosis on this basis has not been very successful. Other platelet parameters that can be computed by automated counters include the platelet distribution width (PDW), which is a measure of platelet , and the “plateletcrit,” which is the product of the MPV and platelet count and, by analogy with the haematocrit, may be seen as indicative of the volume of circulating platelets in a unit volume of blood. The PDW has been found to be of some use in distinguishing essential thrombocythaemia (PDW increased) from reactive thrombocytosis (PDW normal). The plateletcrit does not appear to provide any information of clinical value.

40 | Page PHCPI Laboratory Standards Guideline

4.2.13 Reticulocytes Count Automated reticulocyte counts have been developed by using the fact that various dyes and fluorochromes combine with the RNA of reticulocytes.Following binding of the dye, fluorescent cells can be enumerated using a flow cytometer. Most fully automated blood counters now incorporate a reticulocyte counting capacity so that use of a stand-alone reticulocyte counter is no longer necessary and use of a general purpose flow cytometer is no longer appropriate. An international standard for this method has been published by ICSH. The dyes used in the different systems include auramine O (Sysmex), thiazole orange (ABX), CD4K 530 (Abbott), as well as nonfluorescent dyes such as oxazine 750 (Bayer-Technicon) and the traditional New methylene blue (Beckman-Coulter, Abbott). After staining, it is necessary to separate the reticulocytes from unstained red cells, and, because the dyes also combine with DNA of nucleated cells, these cells must also be excluded. The threshold for this exclusion is determined by the intensity of fluorescence and particle sizing. Although the separation of reticulocytes from mature red cells is not always clearcut, automated reticulocyte counts correlate well with manual reticulocyte counts, although absolute counts may differ because automated counts are dependent on the conditions of incubation and the method of calibrating the instrument. Precision is much superior to that of the manual count because many more cells are counted and the subjective element inherent in recognizing late reticulocytes is eliminated.Potential sources of inaccuracy are the inclusion of some leucocytes and platelets and, less often, Howell–Jolly bodies or malarial parasites in the “reticulocyte” count. Automated reticulocyte counts are fairly stable in blood that has been stored for 1–2 days at room temperature or up to 3–5 days at 4°C.

4.2.14 Immature Reticulocyte Fraction Fully automated instruments provide a measure of the various degrees of reticulocyte maturation because the most immature reticulocytes, produced when erythropoietin levels are high, have more RNA and fluoresce more strongly than the mature reticulocytes normally present in the peripheral blood. Parameters indicating reticulocyte immaturity have potential clinical relevance. For example, an increase in mean fluorescence intensity indicative of the presence of immature reticulocytes has been noted as an early sign of engraftment following bone marrow transplantation. The characteristics of reticulocyte output in different types of anemia’s can be especially appreciated from an output bivariate graph relating fluorescent intensity to reticulocyte count.

41 | Page PHCPI Laboratory Standards Guideline

As described earlier, low total count with a relatively high immature reticulocyte fraction (IRF) is indicative of a repopulating marrow, Whereas a reticulocytopenia with low IRF is typical of severe aplastic anaemia or renal failure A high total count with high IRF occurs in acute haemolysis and blood loss, whereas a low to normal total count with a high IRF occurs in dyserythropiesis and in early response to haematinics. The appearance of reticulocytes with high fluorescence also heralds response when severe aplastic anaemia is being treated with immunosuppressive therapy

4.2.15 Calibrationof Automated Blood Cell Counters The following methods are recommended for calibrating an automated blood cell counter 1. By using fresh normal blood specimens to which values have been assigned for Hb,

PCV, RBC, WBC, and platelet count by standardized reference methods 2. By use of a stable calibrant (either preserved blood or a substitute) to which values appropriate for the instrument in question have been assigned by comparison with fresh normal blood

3. By use of a commercial calibrant with assigned values suitable for the instrument in

question

For reasons of convenience and economy, control materials are commonly used as calibrants; but this practice is not recommended. Such materials are not sufficiently stable to serve as calibrants and their stated values are often approximations that are not assigned by reference methods. They are designed to give test results within a stated range over a stated period rather than a specific result. The procedure for assigning values to fresh blood samples and indirectly to a stable calibrant is as follows: 1. 4 ml blood specimens are obtained from three haematologically normal volunteers

and are anticoagulated with K2 EDTA. 2. The Hb value is assigned by using the haemiglobincyanide method and the mean of

two measurements.

3. The PCV is assigned by the microhaematocrit method, taking the mean of

measurements in four microhaematocrit tubes. 4. The RBC is assigned by performing counts on a single-channel aperture-impedance counter capable of performing a direct cell count; the mean of two dilutions, each counted twice, is used.

5. The MCV is assigned by calculation from the RBC and PCV.

42 | Page PHCPI Laboratory Standards Guideline

6. The WBC is assigned by performing counts on a single-channel aperture-impedance instrument capable of performing direct cell counts; the mean of two dilutions, each counted twice, is used.

4.2.16 Flagging of Automated Blood Counts “Flagging” refers to a signal that the specimen being analyzed may have a significant abnormality because one or more of the blood count variables are outside specified limits (usually 2SD) or there is a qualitative abnormality that requires a quality control check and/or additional investigation. This usually includes a blood film review. Although it is theoretically desirable for every blood count to include examination of a stained film, this is being challenged by increasing workloads requiring time- and cost-effective rationalization, as well as by the use of automated analyzers that report differential leucocyte counts on every specimen. Consequently, fewer blood films are being examined microscopically. Thus, a decision of when a blood film should be made, stained, and examined should take account of flagging and the need to ensure analytic reliability. This includes a check of any significant changes from a recent previous count (delta-check), as well as any specific clinical circumstances. The following is a guide to this selection. Blood count request: Is it a first time count or repeat count?

First time count: Is it a routine screening test or special category? If Routine: Analyzer report for blood count alone Film required if any flags are signaled

If Special category: Film required: 1. Diagnosed blood disease patients

2. Patients receiving radiotherapy and/or chemotherapy

3. Renal disease

4. Neonates

5. Intensive care unit 6. If special tests have also been requested for: infectious mononucleosis, haemolytic

anaemia, enzymopathy, abnormal haemoglobins

7. If the clinical details on the request form indicate lymphadenopathy, splenomegaly,

jaundice or suggest the possibility of leukaemia or lymphoma

8. Specific requests by clinician

Repeat count: Film required:

43 | Page PHCPI Laboratory Standards Guideline

1. Delta check positive when compared with previous record 2. Any flag occurs in present count 3. On each occasion for patients with known blood diseases, for neonates, and when

specifically requested by clinicians

4.3 Part Two: Manual Techniques

4.3.1 Haemoglobin Estimation

Introduction The hemoglobin concentration (Hb) may be estimated by measurement of its colour, by its power of combining with oxygen or carbon monoxide, or by its iron content. The procedures to be described are all colour or light-intensity matching techniques, which also measure, to a varying extent, any methaemoglobin (Hi) or sulphaemoglobin (SHb) that may be present.

Clinical significance This test is done to detect if the patient is suffering from anaemia or rather high Hb Concentration as in polycythaemia.

Principle: To measure hemoglobin concentration (Hb), a known volume of carefully mixed whole blood is added to a diluent which lyses red cells to produce a hemoglobin solution; lysis occurs because of the hypotonicity of the diluent. The Hb is then determined from the light absorbance (optical density) of the solution of hemoglobin or its derivative at a selected wavelength.

Specimen Venous or free-flowing capillary blood that has been anticoagulated with: ethylenediaminetetra-acetic acid (EDTA).

Glassware and equipment EDTA tubes, dispenser or micropipette, test tubes, spectrophotometer or colorimeter.

44 | Page PHCPI Laboratory Standards Guideline

Haemoglobinocyanide (Cyanmethhemoglobulin) Procedure The haemiglobincyanide (Cyanmethemoglobin) procedure is the internationally recommended procedure for determining the haemoglobinconcentration of blood. The basis of the procedure is dilution of blood in a solution containing potassium cyanide and potassium ferricyanide. Haemoglobin, Hi, and HbCO, but not SHb, are converted to HiCN. The absorbance of the solution is then measured in a spectrometer at a wavelength of 540 nm or a photoelectric colorimeter with a yellow–green filter.

Reagents: Diluent The original (Drabkin's) reagent had a pH of 8.6. The following modified solution, Drapkin- type reagent, as recommended by the International Committee for Standardization in Haematology, has a pH of 7.0–7.4. It is less likely to cause turbidity from precipitation of plasma proteins and requires a shorter conversion time (3–5 min) than the original Drabkin's solution, but it has the disadvantage that the detergent causes some frothing: Potassium ferricyanide (0.607 mmol/l) 200 mg

Potassium cyanide (0.768 mmol/l) 50 mg

Potassium dihydrogen phosphate (1.029 mmol/l) 140 mg

Nonionic detergent 1 ml

Distilled or deionized water To 1 litre

The pH should be 7.0–7.4 and must be checked with a pH meter at least once a month. The diluent should be clear and pale yellow in colour. When measured against water as a blank in a spectrometer at a wavelength of 540 nm, absorbance must be zero. If stored at room temperature in a brown borosilicate glass bottle, the solution keeps for several months. If the ambient temperature is higher than 30°C, the solution should be stored in the refrigerator but brought to room temperature before use. It must not be allowed to freeze. The reagent must be discarded if it becomes turbid, if the pH is found to be outside the 7.0–7.4 range, or if it has an absorbance other than zero at 540 nm against water blank.

Procedure

Make a 1 in 201 dilution of blood by adding 20 μl of blood to 4 ml of diluent, or according to the manufactures instructions. Place a rubberstopper on the tube containing the solution and invert it several times. Let the test sample stand at room temperature for at least 5 min (to ensure

45 | Page PHCPI Laboratory Standards Guideline the complete conversion of haemoglobin to haemiglobinocyanide), and then pour it into a and read the absorbance in a spectrometer at 540 nm or in a photoelectric colorimeter with a suitable filter, against a water blank. The absorbance of the test sample must be measured within 6 hours of its initial dilution. The absorbance of a commercially available HiCN standard (brought to room temperature if previously stored in a refrigerator) should also be compared to a water blank in the same spectrometer or photoelectric colorimeter as the patient sample. The standard should be kept in the dark, and, to ensure that contamination is avoided, any unused solution should be discarded at the end of the day on which the ampoule is opened.

Calculation of Haemoglobin Concentration

Preparation of Standard Graph and Standard Table

When many blood samples are to be tested, it is convenient to read the results from a standard graph or table relating absorbance readings to haemoglobin in g/l for the individual instrument. This graph should be prepared each time a new photometer is put into use or when a bulb or other components are replaced. It can be prepared as follows:

Prepare five dilutions of the HiCN reference standard (or equivalent preparation) (brought to room temperature) with the cyanide–ferricyanide reagent according to the Table belowbecause the graph will be used to determine the haemoglobin measurements, it is essential that the dilutions are performed accurately.

Table 1.Dilutions of haemiglobincyanide (HiCN) reference solution for preparation of standard graph Haemoglobin [*] HiCN volume Tube (%) (ml) Reagent volume (ml)

1 100 (full strength) 4.0 (neat) None

2 75 3.0 1.0

3 50 2.0 2.0

4 25 1.0 3.0

5 0 None 4.0 (neat)

* As percent of haemoglobin in reference solution.

46 | Page PHCPI Laboratory Standards Guideline

The haemoglobin concentration of the reference preparation in each tube should be plotted against the absorbance measurement. For example, if the label on the reference preparation states that it contains 800 mg/l, (i.e. 0.8 g/l) and the procedure for haemoglobin measurement uses a dilution of 1:201, the respective haemoglobin concentrations of tubes 1–5 would be 160 g/l, 120 g/l, 80 g/l, 40 g/l, and zero.

Using linear graph paper, plot the absorbance values on the vertical axis and the haemoglobin values on the horizontal axis. The points should fit a straight line that passes through the origin. Providing that the standard has been correctly diluted, this provides a check that the calibration of the photometer is linear. From the graph, it is possible to construct a table of readings and corresponding haemoglobin values. This is more convenient than reading values from a graph when large numbers of measurements are made. It is important that the performance of the instrument does not vary and that its calibration remains constant in relation to haemoglobin measurements. To ensure this, the reference preparation should be measured at frequent intervals, preferably with each batch of blood samples.

Supplementary information

Inaccuracy in estimation of Hb due to: The slow conversion of carboxyhemoglobin to methemoglobin leads to overestimation of the Hb if the test is read at 3 minutes, since carboxyhemoglobin absorbs more light at 540 nm than does cyanmethemoglobin. This difficulty can be overcome by prolonging the reaction time to 30 min before reading. The difference between the 5 and 30 min readings can be used as a semiquantitative procedure for estimating the percentage of HbCO in the blood.

Spectrophotometers and photometers are both sensitive to the effect of turbidity, which may be caused by a high WBC, high concentrations of lipids or plasma proteins, or non-lysed red cells. Increased turbidity causes a factitiously elevated estimate of Hb.

When the WBC is high, turbidity effects are circumvented by centrifugation or filtration of the solution prior to reading the absorbance.

When turbidity is due to a high level of plasma protein, it can be cleared by the addition of either potassium carbonate or a drop of 25% ammonia solution.

When turbidity is due to hyperlipidemia, a blank can be prepared from the diluent and the patient’s plasma or the lipid can be removed by diethyl ether extraction and centrifugation.

Calculation of Haemoglobin Concentration

47 | Page PHCPI Laboratory Standards Guideline

Reference range

Haemoglobin

Men 150 ± 20 g/l

Women 135 ± 15 g/l

Units: g / l, mg/dl

48 | Page PHCPI Laboratory Standards Guideline

4.3.2 Packed Cell Volume Haematocrit

Introduction The haematocrit ×1000 is about three times the haemoglobin expressed in g/l. In conjunction with estimations of haemoglobin and red blood cell count (RBC), it can be used in the calculation of red cell indices. However, its use in under-resourced laboratories may be limited by the need for a specialized centrifuge and a reliable supply of capillary tubes.

Clinical significance The packed cell volume (PCV) can be used primarily as a simple screening test for anaemia, as a reference procedure for calibrating automated blood count systems, and as a rough guide to the accuracy of haemoglobin measurements.

Principle The centrifuge used for the capillary tubes provides a relative centrifugal force (g) of about 12000 g, and 5 min centrifugation results in precipitation of RBC in the distal end.

Specimen Anticoagulated blood sample, because the PCV gradually increases with storage, the test should be performed within 6 hours of collecting the blood sample, but a delay of up to 24 hours is acceptable if the blood is kept at 4°C. Failure to mix the blood sample adequately will produce an inaccurate result. The degree of oxygenation of the blood also affects the result because the PCV of venous blood is ∼2% higher than that of fully aerated blood (which has lost CO2 and taken up O2). To ensure adequate oxygenation and sample mixing, the free air space above the sample should be >20% of the container volume.

Glassware and equipment

Anticoagulant

K2-EDTA tube is recommended, because K3-EDTA causes shrinking of the red cells, reducing the PCV by about 2%. Anticoagulant concentration in excess of 2.2 mg/ml may also cause a falsely low PCV as a result of cell shrinkage.

49 | Page PHCPI Laboratory Standards Guideline

Capillary Tubes

Variation of the bore of the tubes may cause serious errors if they are not within the narrow limits of defined specifications that should be met by manufacturers: length 75 ± 0.5 mm; internal diameter 1.07–1.25 mm, wall thickness 0.18–0.23 mm; and bore taper not exceeding 2% of the internal diameter over the entire length of the tube.

Centrifuge

Centrifuges should be checked at intervals (at least annually) by a tachometer for speed and by a stopwatch for timer accuracy. Efficiency of packing should also be tested by centrifuging samples of normal and polycythaemic blood for varying times from 5 to 10 min to determine the minimum time for complete packing of the red cells.

Procedure

Microhaematocrit Procedure The microhaematocrit procedure is carried out on blood contained in capillary tubes 75 mm in length and having an internal diameter of about 1 mm. The tubes may be plain for use with anticoagulated blood samples or coated inside with 1 IU of heparin for the direct collection of capillary blood. The centrifuge used for the capillary tubes provides a centrifugal force of c 12000 g, and 5 min centrifugation results in a constant PCV. When the PCV is greater than 0.5, it may be necessary to centrifuge for a further 5 min.

Allow blood from a well-mixed specimen, or from a free flow of blood by skin puncture, to enter the tube by capillarity, leaving at least 15 mm unfilled. Then seal the tube by a plastic seal (e.g. Cristaseal, Hawksley, and Lancing, Sussex). Sealing the tube by heating is not recommended because the seals tend to be tapered and there is the likelihood of lysis. After centrifugation for 5 min, measure the proportion of cells to the whole column (i.e., the PCV) using a reading device.

Accuracy of Microhaematocrit The microhaematocrit procedure has an adequate level of accuracy and precision for clinical utility. Attention must be paid to a number of factors that may produce an inaccurate result.

Reading The test should be read as soon as possible after centrifugation because the red cells begin to swell and the interface becomes progressively more indistinct. To avoid errors in reading with

50 | Page PHCPI Laboratory Standards Guideline the special reading device, a magnifying glass should be used. White cells and platelets (the buffy coat) must be excluded as far as possible from the reading of the packed red cells. If a special reading device is not available, the ratio of red cell column to whole column can be calculated from measurements obtained by placing the tube against arithmetic graph paper or against a ruler.

Supplementary information

Plasma Trapping The amount of plasma trapped between red cells, especially in the lower end of the red cell column, and red cell dehydration during centrifugation generally counterbalance each other, and the error caused by trapped plasma is usually not more than 0.01 PCV units. Thus, in routine practice, it is unnecessary to correct for trapped plasma, but if the PCV is required for calibrating a blood cell analyser or for calculating blood volume, the observed PCV should be reduced by a 2% correction factor after it has been centrifuged for 5 min or for 10 min with polycythaemic blood. It is, however, preferable to use the surrogate reference procedure. Plasma trapping is increased in macrocytic anaemias, , thalassaemia, hypochromic anaemias, and sickle cell anaemia; it may be as high as 20% in sickle cell anaemia if all the cells are sickled.

Factors affecting the test:

Table 2. Factors affecting the microhaematocrit

51 | Page PHCPI Laboratory Standards Guideline

Reference range

Units

% or L / L

4.3.3 Total White Blood Cell Count

Introduction

Visual counting of blood cells is an acceptable alternative to electronic counting for white cell, while, it is not recommended for routine red cell counts because the number of cells that can be counted within a reasonable time in the routine laboratory will be too few to ensure a sufficiently precise result.

Principle: The principles of manual cell counts, the use of the haemocytometer counting chamber for manually counting white cells and platelets in a known surface area.

52 | Page PHCPI Laboratory Standards Guideline

Clinical significance

Some of the common Causes of leucocytosis:  Bacterial infections  Inflammation and tissue necrosis  Metabolic disorders (e.g. uraemia, eclampsia, acidosis, gout)  Neoplasms of all types (e.g. carcinoma, lymphoma, melanoma)  Acute haemorrhage or haemolysis  Drugs (e.g. corticosteroid therapy (inhibits margination): lithium, tetracycline)  Chronic myeloid leukaemia, myeloproliferative disease, polycythaemia vera, myelofibrosis, essential thrombocythaemia  Treatment with myeloid growth factors (e.g. G-CSF, GM-CSF)  Infectious mononucleosis, rubella, pertussis,  Infectious hepatitis, cytomegalovirus, HIV, herpes simplex or zoster  Tuberculosis, toxoplasmosis, brucellosis, syphilis  Chronic lymphoid leukaemias  Acute lymphoblastic leukaemia  Non-Hodgkin's lymphoma (some)  Thyrotoxicosis.

Some of the common Causes of leucopenia:  Selective neutropenia  Drug-induced  Systemic lupus erythematosus  Felty's syndrome  Hypersensitivity and anaphylaxis  Viral (e.g. hepatitis, influenza, HIV)  Fulminant bacterial infection (e.g. typhoid, miliary Tuberculosis)

Specimen Venous or free-flowing capillary blood that has been anticoagulated with ethylenediaminetetra- acetic acid (EDTA). Thorough mixing of the blood specimen before sampling is essential for accurate test results. Ideally, tests should be performed within 6 hours of obtaining the blood

53 | Page PHCPI Laboratory Standards Guideline specimen because some test results are altered by longer periods of storage. However, results that are sufficiently reliable for clinical purposes can usually be obtained on blood stored for up to 24 hours at 4°C.

Glassware and equipment

Counting Chambers

The visibility of the rulings in the counting chamber is as important as the accuracy of calibration, so that chambers with a “metallised” surface and Neubauer or Improved Neubauer rulings are recommended. These have nine 1 mm × 1 mm ruled areas, which, when covered correctly with the special thick cover glass, each contain a volume of 0.1 μl of diluted blood. Coverslips designed for mounting of preparations must not be used with counting chambers. The sample is introduced between the chamber and the coverglass using a pipette or capillary tube, and the preparation is viewed using a ×40 objective and ×6 or ×10 eyepieces. With Neubauer and Improved Neubauer chambers, count the cells in 4 or 8 horizontal rectangles of 1 mm × 0.05 mm (80 or 160 small squares) or in 5 or 10 groups of 16 small squares, including the cells that touch the bottom and left-hand margins of the small squares( see figure 11).

Figure 11: Improved NeubauerCounting chamber   Micropipette 

54 | Page PHCPI Laboratory Standards Guideline

Reagents:

Diluent To make the counting of white cells easier, diluted whole blood is mixed with a fluid to lyse the red cells and stain the white cell nuclei deep violet-black. The diluent is 2% (20 ml/l) acetic acid coloured pale violet with gentian violet.

Procedure Make a 1 in 20 dilution of blood by adding 0.1 ml of well-mixed blood (lack of adequate mixing is a major source of error) to 1.9 ml of diluent in a 75 × 10 mm plastic (or glass) tube. After sealing the tube with a lid or tightly fitting bung, mix the diluted blood in a mechanical mixer or by hand for at least 2 min by tilting the tube to an angle of about 120 degrees combined with rotation, thus allowing the air bubble to mix the suspension. Fill a clean dry counting chamber, with its coverglass already in position, without delay. This is simply accomplished with the aid of a plastic Pasteur pipette or a length of stout capillary glass tubing that has been allowed to take up the suspension by capillarity. Take care that the counting chamber is filled in one action and that no fluid flows into the surrounding moat.

Leave the chamber undisturbed on a bench for at least 2 min for the cells to settle, but not much longer, because drying at the edges of the preparation initiates currents that cause movement of the cells after they have settled. The bench must be free of vibrations, and the chamber must not be exposed to draughts or to direct sunlight or other sources of heat. It is important that the coverglass should be of a special thick glass and perfectly flat, so that when laid on the counting chamber, diffraction rings are seen. The coverglass should be of such a size that when placed correctly on the counting chamber the central ruled areas lie in the centre of the rectangle to be filled with the cell suspension.

Factors affecting the test: If any of the following filling defects occur, the preparation must be discarded and the filling procedure must be repeated using another clean dry chamber:

Overflow into moat

Chamber area incompletely filled Air bubbles anywhere in chamber area

55 | Page PHCPI Laboratory Standards Guideline

Any debris in chamber area

To obtain a coefficient of variation of 5%, it is necessary to count about 400 cells in practice, it is reasonable to count 100 white cells. To minimise distribution errors, count the cells in the entire ruled area (i.e., 9 × 0.1 μl areas in an Improved Neubauer counting chamber).

Counting in only one or two fields results in a wide variance that is reduced as more cells are counted. However, high precision is achieved only when thousands of cells are counted, which is only possible with automated cell counters.

Calculation

WBC/L

Thus, if N cells are counted in 0.1 μl, then the WBC/l is as follows:

(e.g., if 115 cells are counted, the WBC is 115 × 200 × 106/l = 23 × 109/l)

Range of White Blood Cell Count in Health

White blood cell count 4.0–10.0 × 109/l

Units × 109/l

4.3.4 Platelet Count Introduction Manual counts are used routinely in under-resourced laboratories, and they are still needed even in well-equipped laboratories for blood samples with a significant proportion of giant platelets. However, for all other samples, automated full blood counters produce platelet counts with a precision that is much superior to that of manual platelet counts.

56 | Page PHCPI Laboratory Standards Guideline

Principle: Manual platelet counts are performed by visual examination of diluted, lysed whole blood using a Neubauer or Improved Neubauer counting chamber as for total white cell counts.

Clinical significance This test is used for detection of thrombocytosis or thrombocytopenia which occurs in many diseases.

Specimen Platelet counts are best performed on ethylenediaminetetra-acetic acid (EDTA)-anticoagulated blood that has been obtained by clean venipuncture. They can also be carried out on blood obtained by skin prick, but the results are less satisfactory than those on venous blood. Skin- prick platelet counts are significantly lower than counts on venous blood and less constant; a variable number of platelets are probably lost at the site of the skin puncture.

Glassware and equipment

Counting Chambers

The visibility of the rulings in the counting chamber is as important as the accuracy of calibration, so that chambers with a “metallised” surface and Neubauer or Improved Neubauer rulings are recommended. These have nine 1 mm × 1 mm ruled areas, which, when covered correctly with the special thick cover glass, each contain a volume of 0.1 μl of diluted blood. Coverslips designed for mounting of microscopy preparations must not be used with counting chambers. The sample is introduced between the chamber and the coverglass using a pipette or capillary tube, and the preparation is viewed using a ×40 objective and ×6 or ×10 eyepieces. With Neubauer and Improved Neubauer chambers, count the cells in 4 or 8 horizontal rectangles of 1 mm × 0.05 mm (80 or 160 small squares) or in 5 or 10 groups of 16 small squares, including the cells that touch the bottom and left-hand margins of the small squares( see figure 12).

57 | Page PHCPI Laboratory Standards Guideline

Figure12: Improved Neubauer Counting Chamber

 Microscope  Micropipette  Test tube

Reagents: 1% ammonium oxalate, which lyses red cells, as a diluent produces a higher and more accurate countthan use of formal citrate, which leaves red cells intact.

Procedure The diluent consists of 1% aqueous ammonium oxalate in which the red cells are lysed. This procedure is recommended in preference to that using formal-citrate as diluent, which leaves the red cells intact and is more likely to give incorrect results, when the platelet count is low.

Before diluting the blood sample, examine it carefully for the presence of blood clots. If these are present, a fresh specimen should be requested because clots will cause the platelet count to be artificially low. Make a 1 in 20 dilution of well-mixed blood in the diluent by adding 0.1 ml of blood to 1.9 ml of ammonium oxalate diluent (10 g/l). Not more than 500 ml of diluent should be made at a time, using carefully clean glassware and fresh glass-distilled or deionised

58 | Page PHCPI Laboratory Standards Guideline water. If possible, the solution should be filtered through a micropore filter (0.22 μm) and kept at 4°C. For use, a small part of the stock is refiltered and dispensed in 1.9 ml volumes in 75 × 12 mm tubes.

Mix the suspension on a mechanical mixer for 10–15 min. Fill a Neubauer counting chamber with the suspension, using a stout glass capillary or Pasteur pipette. Place the counting chamber in a moist and leave untouched for at least 20 min to give time for the platelets to settle.

Examine the preparation with the ×40 objective and ×6 or ×10 eyepieces. The platelets appear under ordinary illumination as small (but not minute) highly refractile particles if viewed with the racked down; they are usually well-separated, and clumps are rare if the blood sample has been skillfully collected. To avoid introducing into the chamber dirt particles, which might be mistaken for platelets, all equipment must be scrupulously clean. Platelets are more easily seen with the phase-contrast microscope. A special, thin-bottomed (1 μm) counting chamber is best for optimal phase-contrast effect. The number of platelets in one or more areas of 1 mm2 should be counted. The total number of platelets counted should always exceed 200 to ensure a coefficient of variation of 8–10%.

Calculation

Thus, if N is the number of platelets counted in an area of 1 mm2 (0.1 μl in volume), the number of platelets per litre of blood is:

Range of platelet Count

Platelet count 280 ± 130 × 109/l

Factors affecting the test:

The errors associated with manual cell counts are technical and inherent.

Technical errors can be minimized by avoiding the following:

59 | Page PHCPI Laboratory Standards Guideline

 Poor technique in obtaining the blood specimen

 Insufficient mixing of the blood specimen

 Inaccurate pipetting and the use of badly calibrated pipettes or counting

chambers

 Inadequate mixing of the cell suspension

 Faulty filling of the counting chamber

 Careless counting of cells within the chamber

Units × 109/l

4.3.5 Erythrocyte Sedimentation Rate

Introduction

Inflammatory response to tissue injury (the acute-phase response) includes alteration in serum protein concentration, especially increases in fibrinogen, haptoglobin, caeruloplasmin, immunoglobulins (Ig), and C-reactive protein (CRP), and decrease in albumin. The changes occur in acute infection, during active phases of chronic inflammation, with malignancy, in acute tissue damage (e.g., following acute myocardial infarction), or with physical injury.

Principle: Essentially it is the measurement after 1 hour of the sedimentation of red cells in diluted blood in an open-ended of 30 cm length mounted vertically on a stand.

Clinical significance Measurement of the acute-phase response is a helpful indicator of the presence and extent of inflammation or tissue damage and response to treatment. The usual tests are estimation of CRP and measurement of the erythrocyte sedimentation rate (ESR);

Specimen Either collect venous blood in ethylenediaminetetra-acetic acid (EDTA) and dilute a sample accurately in the proportion of 1 volume of citrate to 4 volumes of blood, or collect the blood directly into the citrate solution.

60 | Page PHCPI Laboratory Standards Guideline

Glassware and equipment The recommended tube is a straight glass or rigid transparent plastic tube 30 cm in length and not less than 2.55 mm in diameter. The bore must be uniform to within 5% throughout. A scale graduated in mm extends over the lower 20 cm. The tube must be clean and dry.

Specially made racks with adjustable levelling screws are available for holding the sedimentation tubes firmly in an exactly vertical position. The rack must be constructed so that there will be no leakage of the blood from the tube.

Reagents:

Diluent prepare a solution of 109 mmol/l trisodium citrate (32 g/l Na3Ca6H5O7.2H2O). Filter through a micropore filter (0.22 mm) into a sterile bottle. It can be stored for several months at 4°C but must be discarded if it becomes turbid through the growth of moulds.

Procedure Conventional Westergren Procedure The test should then be carried out on the diluted sample within 4 hours of collecting the blood, although a delay of up to 6 hours is permissible provided that the blood is kept at 4°C. EDTA blood can be used within 24 hours if the specimen is kept at 4°C, provided that 1 volume of 109 mmol/l (32 g/l) trisodium citrate is added to 4 volumes of blood immediately before the test is performed.

Mix the blood sample thoroughly and then draw it up into the Westergren tube to the 200 mm mark by means of a teat or a mechanical device; mouth suction should never be used. Place the tube exactly vertical and leave undisturbed for exactly 60 min, free from vibrations and draughts and not exposed to direct sunlight.

Calculation Read to the nearest 1 mm the height of the clear plasma above the upper limit of the column of sedimenting cells. The result is expressed as ESR = X mm in 1 h. A poor delineation of the upper layer of red cells may sometimes occur, especially when there is a high reticulocyte count.

61 | Page PHCPI Laboratory Standards Guideline

Range in Health There is a progressive increase with age, but older than 70 years it is difficult to define a strictly healthy population for determining normal values.

Erythrocyte sedimentation rate ranges in health

Age (years) 95% Upper limit (mm in 1h)

Men

17–50 10

51–60 12

61–70 14

>70 about 30

Women

17–50 12

51–60 19

61–70 20

>70 about 35

Pregnancy

First half 48 (62 if anaemic)

Second half 70 (95 if anaemic)

In the newborn, the ESR is usually low. In childhood and adolescence, it is the same as for normal men with no differences between boys and girls. It is increased in pregnancy, especially so in the later stages, and dependent of anaemia.

Factors affecting the test:

Modified Procedures The test could be affected by: Length of Tube The overall length of the tube is not a critical dimension for the test provided that it fits firmly in an appropriate holding device. The tube must, however, be long enough to ensure that packing of the cells does not start before the test has been completed.

62 | Page PHCPI Laboratory Standards Guideline

Temp: should be between 18-25 C Hemolysis: affect the rate of sedimentation

Plastic Glass Tubes

A number of plastic materials (e.g., polypropylene and polycarbonate) are recommended as substitutes for glass in Westergren tubes. Nevertheless, not all plastics have similar properties, and it must be demonstrated that the ESR with the chosen tubes is reproducible and not affected by the plastic.

Disposable Glass Tubes

Disposable glass tubes should be supplied clean and dry and ready for use. It is necessary to show that neither the tube material nor the manufacturer's cleaning process affect the ESR.

Capillary Procedure

Short tubes of narrower bore than in the standard tube are available mainly for tests on infants. These are, however, no longer in general use, and it is necessary to establish normal ranges or a correction factor to convert results to an approximation of ESR by the Westergren procedure.

Time

Sedimentation is measured after aggregation has occurred and before the cells start to pack usually at 18-24 min. From the rate during this time period the sedimentation that would have occurred at 60 min is derived and converted to the conventional ESR equivalent by an algorithm.

Sloping Tube

Red cells sediment more quickly when streaming down the wall of a sloped tube. This phenomenon has been incorporated into automated systems in which the end-point is read after 20 min with the tube held at an angle of 18 degrees from the vertical. This has been shown to give results comparable to the conventional procedure.

63 | Page PHCPI Laboratory Standards Guideline

Anticoagulant

EDTA blood can be used without citrate dilution, at least if packed cell volume (PCV) is below 0.36 (haemoglobin < 110 g/l); less precise results are obtained when the PCV is higher. The readings from undiluted samples must then be adjusted as for the standardised procedure (see below).

Because of the biohazard risk of blood contamination inherent in using open-ended tubes, it is now recommended that, where possible, a closed system be used in routine practice. Manual procedures are available that avoid transfer of the blood into the sedimentation tube. Automated closed systems use either blood collected in special evacuated tubes containing citrate or EDTA blood. A sample is taken up through a pierceable cap and then automatically diluted in the system if this is required. Some systems use sloping tubes at an angle of 18 degrees to obtain results rapidly, and one model of the Ves-matic (Diesse) also incorporates centrifugation.

Whenever a different procedure or tube is planned, a preliminary test should be carried out to check precision and to compare results with those obtained by the standardised procedure described in the following section.

Units mm / hour

4.3.6 Reticulocyte Count

Introduction

Reticulocytes are juvenile red cells; they contain remnants of the ribosomal ribonucleic acid (RNA) that was present in larger amounts in the cytoplasm of the nucleated precursors from which they were derived. Ribosomes have the property of reacting with certain basic dyes such as azure B, brilliant cresyl blue, or New methylene blue (see below) to form a blue or purple precipitate of granules or filaments.

Principle

On exposure of unfixed cells to certain dyes that contains only basic component of the stain, such as brilliant cresyl blue or new methylene blue, the ribosomes are precipitated and stained by the dye, to appear as a reticular network. With these stains red cells stain a pale greenish- blue while the reticulum stains bluish-purple.

64 | Page PHCPI Laboratory Standards Guideline

Clinical significance

The number of reticulocytes in the peripheral blood is a fairly accurate reflection of erythropoietic activity, assuming that the reticulocytes are released normally from the bone marrow and that they remain in circulation for the normal time period.

Specimen

Venous blood that has been anticoagulated with ethylenediaminetetra-acetic acid (EDTA). Thorough mixing of the blood specimen before sampling is essential for accurate test results. Ideally, tests should be performed within 6 hours of obtaining the blood specimen because some test results are altered by longer periods of storage. However, results that are sufficiently reliable for clinical purposes can usually be obtained on blood stored for up to 24 hours at 4°C.

Glassware and equipment

EDTA tubes, dispenser or micropipette, test tubes, glass slides, Microscope.

Reagents

Staining Solution

Dissolve 1.0 g of New methylene blue or azure B in 100 ml of iso-osmotic phosphate buffer pH 6.5.

Procedure

Deliver 2 or 3 drops of the dye solution into a 75- × 10-mm plastic tube by means of a plastic Pasteur pipette. Add 2–4 volumes of the patient's EDTA-anticoagulated blood to the dye solution and mix. Keep the mixture at 37°C for 15–20 min. Resuspend the red cells by gentle mixing, and make films on glass slides in the usual way. When dry, examine the films without fixing or counterstaining.

The exact volume of blood to be added to the dye solution for optimal staining depends on the RBC. A larger proportion of anaemic blood, and a smaller proportion of polycythaemic blood, should be added than of normal blood. In a successful preparation, the reticulofilamentous

65 | Page PHCPI Laboratory Standards Guideline material should be stained deep blue and the nonreticulated cells should be stained diffuse shades of pale greenish blue. Films should not be counterstained. The reticulofilamentous material is not better defined after counterstaining, and precipitated stain overlying cells may cause confusion. Moreover, Heinz bodies will not be visible in fixed and counterstained preparations. If the stained preparation is examined under phase contrast, both the mature red cells and reticulocytes are well defined. By this technique, late reticulocytes characterized by the presence of remnants of filaments or threads are readily distinguished from cells containing . Satisfactory counts may be made on blood that has been allowed to stand (unstained) for as long as 24 hours, although the count will tend to decrease after 6–8 hours unless the blood is kept at 4°C.

Counting Reticulocytes

An area of film should be chosen for the count where the cells are undistorted and where the staining is good. A common fault is to make the film too thin; however, the cells should not overlap. To count the cells, use the ×100 oil-immersion objective and, if possible, eyepieces provided with an adjustable diaphragm. If eyepieces with an adjustable diaphragm are not available, a paper or cardboard diaphragm, in the centre of which has been cut a small square with sides about 4 mm in length, can be inserted into an eyepiece and used as a less convenient substitute.

The counting procedure should be appropriate to the number of reticulocytes present. Very large numbers of cells have to be surveyed if a reasonably accurate count is to be obtained when only small numbers of reticulocytes are present. When the count is less than 10%, a convenient procedure is to survey successive fields until at least 100 reticulocytes have been counted and to count the total red cells in at least 10 fields to determine the average number of red cells per field.

Calculation

Number of reticulocytes in n fields = x

Average number of red cells per field = y

Total number of red cells in n fields = n × y

Reticulocyte percentage = [x ÷ (n × y)] ×

100%

Absolute reticulocyte count = % × RBC

66 | Page PHCPI Laboratory Standards Guideline

Thus, when the reticulocyte percentage is 3.3 and the RBC is 5 × 1012/l, the absolute reticulocyte count per litre is as follows: [3.3/100] × 5 × 1012 = 165 × 109

It is essential that the reticulocyte preparation be well spread to ensure an even distribution of cells in successive fields.

When the reticulocyte count exceeds 10%, only a relatively small number of cells will have to be surveyed to obtain a standard error of 10%.

Factors affecting the test:

It is essential that the reticulocyte preparation be well spread and well stained. Other important factors that affect the accuracy of the count are the visual acuity and patience of the observer and the quality and resolving power of the microscope. The most accurate counts are carried out by a conscientious observer who has no knowledge of the supposed reticulocyte level, thus eliminating the effect of conscious or unconscious bias.

Range of Reticulocyte Count in Health

The range of reticulocyte in adults and children is 50–100 × 109/l (0.5–2.5%). In infants (full term, cord blood) it is 2–5%.

Carry out a rough platelet count to determine the approximate concentration, and add an appropriate amount of the suspension to preserved blood. Mix well for 20 min and, with continuous mixing, dispense into sterile containers. Cap and seal. At 4°C, the preparation should have a shelf life of 3-4 months. Before use, resuspend by thorough shaking by hand, followed by mechanical mixing for approximately 15 min.

4.3.7 Preparation of blood films

Principle

Although we can judge the proportions of red cells and white cells in a tube of sedimented blood, we get far more information if the blood is carefully mixed and a thin layer is spread on a glass slide to form a blood film. The blood cells are then preserved by exposure to the alcohol methanol, a process known as fixation. The fixed film of blood is stained with a mixture of

67 | Page PHCPI Laboratory Standards Guideline several dyes so that the individual cells can be recognized when they are examined with a microscope.

Specimen

Blood films can be prepared from fresh blood with no anticoagulant added or from ethylenediaminetetra-acetic acid (EDTA)-anticoagulated blood. Heparinized blood should not generally be used because its staining characteristics differ from those of EDTA-anticoagulated blood. Good films can be made in the following manner, using clean slides, if necessary wiped free from dust immediately before use. Slides should measure 75 × 25 mm and approximately 1 mm thick; ideally, they should be frosted at one end to facilitate labelling, but these are more expensive.

Glassware and equipment

EDTA tubes, dispenser or micropipette, spreader, slides, microscope.

Reagents

Leishman's Stain

Weigh out 0.2 g of the powdered dye, and transfer it to a conical flask of 200–250 ml capacity. Add 100 ml of methanol and warm the mixture to 50°C for 15 min, occasionally shaking it. Allow the flask to cool and filter. It is then ready for use, but it will improve on standing.For daily use, store the stain in an airtight (prevent moisture entering the stain) amber (semi- opaque) container. Keep in a cool place (not refrigerated) and never in direct sunlight.

The stock stain should be kept in a tightly stoppered light opaque (e.g. amber) container in a cool dark place. Renew every 3 months or earlier if indicated. To obtain optimum color reaction, some suggest that 3–5 days should be allowed before using freshly made stain.

Buffered Water

Make up 50 ml of 66 mmol/l Sörensen's phosphate buffer of the required pH to 1 litre with water at a pH of 6.8. An alternative buffer may be prepared from buffer tablets, which are

68 | Page PHCPI Laboratory Standards Guideline available commercially. Solutions of the required pH are obtained by dissolving the tablets in water and testing with a pH meter.

Manual Procedure

Spreading Blood Films

First, make a spreader from a glass slide that has a smooth end. Using a glass cutter, break off one corner of the slide, leaving a width of about 18 mm as the spreader. A spreader can be used repeatedly unless the edge becomes chipped, but it must be thoroughly washed and dried between films.

Place a small drop of blood in the centre line of a slide about 1 cm from one end. Then, without delay, place a spreader in front of the drop at an angle of about 30 degrees to the slide and move it back to make contact with the drop. The drop should spread out quickly along the line of contact. With a steady movement of the hand, spread the drop of blood along the slide. The spreader must not be lifted off until the last trace of blood has been spread out; with a correctly sized drop, the film should be about 3 cm in length. It is important that the film of blood finishes at least 1 cm before the end of the slide.

The thickness of the film can be regulated by varying the pressure and speed of spreading and by changing the angle at which the spreader is held. With anaemic blood, the correct thickness is achieved by using a wider angle, and, conversely, with polycythaemic blood, the angle should be narrower. The ideal thickness is such that on microscopy there is some overlap of red cells throughout much of the film's length. The leucocytes should be easily recognizable throughout most of the film. With poorly made films the leucocytes will be unevenly distributed, with monocytes and other large leucocytes being pushed to the end and the sides of the spread. An irregular streaky film will occur if the slide is greasy, and dust on the surface will cause patchy spots.

The films should be allowed to dry in the air. In humid conditions the films may be exposed to a current of warm air (e.g., from a hairdryer), but this should be in a microbiological safety hood.

69 | Page PHCPI Laboratory Standards Guideline

Labeling Blood Films

The film should be labelled immediately after spreading. Write either a laboratory reference number or the name of the patient and the date in pencil on the frosted end of the slide or on the film itself (writing on the thickest part, which is least suitable for microscopic examination). A label written in pencil will not be removed by staining. A paper label should be affixed to the slide later. If blood films are to be stored for future reference, apply the paper label in such a manner that it is easily read when the slides are filed.

Fixing Blood Films

Dry the films in the air. Leave for 3 min in the fixative this must be done without delay and the films should never be left unfixed for more than a few hours. If films are sent to the laboratory by post, it is preferable that, when possible, they are thoroughly dried and fixed before dispatch.

Staining Blood Films

Romanowsky stains are used universally for routine staining of blood films, and satisfactory results can be obtained.

Air dry the film and then flood the slide with the stain. After 2 min, add double the volume of water and stain the film for 5–7 min. Then wash it in a stream of buffered water until it has acquired a pinkish tinge (up to 2 min). After the back of the slide has been wiped clean, set it up right to dry.

When several batches of films are being stained in succession, the staining solution should be renewed at intervals (e.g., after each 50 slides). Loss of staining power is usually the result of precipitation of the eosin Y, and this will result in the nuclei staining blue instead of purple.

Carry out a rough platelet count to determine the approximate concentration, and add an appropriate amount of the suspension to preserved blood. Mix well for 20 min and, with continuous mixing, dispense into sterile containers. Cap and seal. At 4°C, the preparation should have a shelf life of 3-4 months. Before use, resuspend by thorough shaking by hand, followed by mechanical mixing for approximately 15 min.

4.3.8. Antiglobulin or Coombs' Test The antiglobulin test is probably the most important test in serology and blood banking. The antiglobulin test (Coombs' test) was introduced by Coombs, Mourant, and Race in 1945 as a method for detecting “incomplete” Rh antibodies (i.e., IgG antibodies capable of sensitizing red cells but incapable of causing

70 | Page PHCPI Laboratory Standards Guideline agglutination of the same cells suspended in saline) as opposed to “complete” IgM antibodies, which do agglutinate saline-suspended red cells. Coombs' test is divided into direct and indirect antiglobulin tests:  In the direct antiglobulin test (DAT), the patient's cells, after careful washing, are tested for sensitization that has occurred in vivo.  In the indirect antiglobulin test (IAT), normal red cells are incubated with a serum suspected of containing an antibody and subsequently tested, after washing, for in vitro– bound antibody. The DAT is used to demonstrate in vivo attachment of antibodies to red cells, as in autoimmune hemolytic anemia, alloimmune HDN, and alloimmune hemolysis following an incompatible transfusion. The IAT has wide application in blood transfusion serology, including antibody screening and identification and cross-matching. DAT: A spin tube technique is recommended for the routine antiglobulin test. The test should be carried out in glass tubes. Plastic tubes are not recommended because they may adsorb IgG, which could neutralize anti-IgG of the antiglobulin reagent. Use a blood sample in EDTA anticoagulant. If a cold-reacting autoantibody is present, the patient's red cells should be washed three times in a large volume of saline warmed to 37°C to wash off cold antibodies and obtain a smooth suspension of cells. There is no risk of washing off adsorbed complement components. However, the washing process should be accomplished as quickly as possible and the test should be set up immediately afterward because bound warm antibody occasionally elutes off the cells when they are washed and false-negative results may be obtained. “Saline” refers to 9 g/l NaCl buffered to pH 7.0. DAT Method: Make 5% suspension of red cells that have been washed three times in saline. If failed to wash the red cells properly then the antihuman globulin antisera may be neutralized by immunoglobulins or complement in the surrounding serum or plasma and cause a false negative result. Add 1 drop of the cell suspension to 2 drops of anti-human globulin reagent. Centrifuge for 1 minute at 150g. Examine for agglutination after gently re-suspending the button of cells. A concave mirror and good light help in macroscopic readings. If the result appears to be negative, confirm this microscopically.

71 | Page PHCPI Laboratory Standards Guideline

Figure 13: Coombs Test

4.3.9. ABO & Rh Group The test used to determine the blood ABO and Rh group

Materials

1- Slide 2- Anti-A Monoclonal sera 3- Anti-B Monoclonal sera 4- Anti-D Monoclonal sera 5- EDTA blood

Method

1- Add one drop of each colored anti-sera on the slid 2- To each anti-sera add one drop of EDTA blood 3- Mix the blood with anti-sera 4- Rotating the slide by hand many times read the result within 2-5 mintes 5- Record the result

72 | Page PHCPI Laboratory Standards Guideline

6- If Rh negative, test it by tube method: A. Add one volume of AntiD sera to test tube. B. Add one volumeof 2% red cells suspension. C. Mix suspension by traping the tube. D. Leave the suspension undisturbed for 15 min and read result.

Discrepancies

A- False Positive 1- formation 2- Auto-immune hemolytic anemia B- False Negative Occurred when the anti-sera is impotent or expired

73 | Page PHCPI Laboratory Standards Guideline

5. Serology

5.1 Rapid Plasma Reagin (RPR) TEST

Principle

Syphilis is a chronic venereal infection caused by the spirochaete micro-organism Treponema pallidum. As the organism cannot be cultured on artificial media, diagnosis of syphilis depends on the detection of specific antibodies by serological tests in combination with clinical findings. The RPR test is “non-treponemal” in that the antibodies detected are not specific for T. pallidum, although their presence in patient’s serum or plasma is strongly associated with infection by the organism. This test measures antibody (IgG and IgM) produced in response to lipoidal material released from damaged host cells as well as to lipoprotein-like material released from the spirochaetes. These antibodies tend to disappear after successful treatment of the infection.

The RPR test consists of modified VDRL antigen containing microparticulate carbon, which aggregates in the presence of reagin type antibodies in serum or plasma, indicating a positive result. The aggregation can be read macroscopically. Non-reactive samples typically appear as a smooth non-aggregated pattern, which may form buttons in the centre of the test area.

Material 1. Carbon antigen reagent 2. Positive Control 3. Negative Control 4. Disposable Test Slides 5. Disposable Pipette Stirrers (pipstirs) 6. Dispenser 7. Needle 8. Specimen collection container 9. Timer 10. Automatic card rotator (100 rpm) 11. (0.85%) physiological saline (semi quantitative test only). 12. Test tubes for sample dilution (semi quantitative test only).

74 | Page PHCPI Laboratory Standards Guideline

Procedure

a. Qualitative Test 1. Allow all reagents, controls and samples to reach room temperature before use. 2. Draw the sample into the pipstir taking care not to transfer any cellular elements. 3. Hold the pipstir over a test card circle and allow one drop (50µl) of specimen to fall onto the test card. It is important to hold the dispenser in a vertical position whilst dispensing the sample. 4. Spread the specimen evenly over the entire test circle using the mixing (broad) end of the pipstir. 5. Shake the vial of carbon antigen reagent to ensure even mixing. 6. Attach the needle to the dispenser. Withdraw sufficient carbon antigen for the number of tests being performed. 7. Keeping the dispenser in a vertical position allow one drop to fall onto the specimen. Do not mix. Return unused antigen from the dispenser to the glass vial. 8. Rotate the tile for 8 minutes preferably on an automatic rotator at 100 rpm. 9. Read and interpret results macroscopically in good light.

b. Semi Quantitative Test 1. Prepare doubling dilutions of the sample from the undiluted specimen to 1:32 using physiological saline. Mix well. 2. Using the pip stirs, place one drop (50µl) of each dilution onto a separate test card circle. 3. Using the mixing (broad) end of the pipstir spread each dilution evenly over the test circle, starting at the highest dilution (1:32), proceeding to the lowest (1:2). 4. Continue as from 5 in the Qualitative Test. 5. After 8 minutes rotation, read the test and note the last circle in the dilution series that has a positive result. 6. If the highest dilution tested (1:32) is reactive, proceed with a further dilution series by preparing doubling dilutions of the sample from 1:32 to 1:512 using physiological saline. Mix well and continue as from 2in the semi quantitative test.

75 | Page PHCPI Laboratory Standards Guideline

Interpretation

Quality Control

It is recommended that the kit positive and negative controls are run with each batch of test specimens. For the assay to be valid the positive control provided should give a strong positive pattern and the negative control provided should give a clearly negative result.

Results

Qualitative Test

Positive result

Reactive (positive) samples display characteristic agglutination ranging from slight (weak- reactive) to intense (reactive). A strong positive reaction is seen as large aggregates in the centre of the test circle. Weakly positive reactions are seen as small aggregates around the edge of the test circle.

Negative result

Negative results show no aggregates. The carbon antigen either remains in a smooth suspension or forms a distinct button.

Semi Quantitative Test

Results may be graded from strong to non-reactive and the titre expressed as the reciprocal of the last dilution showing a positive reaction.

1. Strong Reactive (SR): Large clumps of carbon particles with a clear background.

2. Reactive (R): Large clumps of carbon particles, more dispersed than strong reactive.

3. Weak Reactive (WR): Small clumps of carbon particles with light grey background.

4. Trace Reactive (TR): Slight clumping of carbon particles, typically seen as a button of aggregates in the centre of the test circle or dispersed around the edge of the test circle.

5. Non-Reactive (NR): A smooth grey pattern or a button of non-aggregated carbon particles in the centre of the test circle.

Reactive samples should be recorded as antibody positive and must be subjected to further tests to determine the presence or absence of specific anti-Treponemal antibody.

76 | Page PHCPI Laboratory Standards Guideline

Limitations

 As with all reagin tests the RPR test may give a small proportion of false positive results. Diseases such as infectious mononucleosis, leprosy, lupus erythematous, vaccinia and viral pneumonia can cause such reactions.  Reactive RPR test specimens should be tested with further serological tests (i.e. TPHA and FTA-abs) since, as with any serological procedure, the diagnosis should not be made on a single reactive result.  As with other serological tests, the RPR test cannot distinguish between syphilis and other pathogenic Treponemal infections e.g. Yaws.

5.2 TPHA (Treponema pallidum Haemagglutination Assay) TEST

Principle

Serological screening tests for syphilis usingcardiolipin and lecithin as antigens are simple to perform but biological falsepositive reactions occur frequently because the tests use non- treponemalantigens. TPHA reageants are used to detect human serum antibody to T.pallidumby means of an indirect haemagglutination (IHA) method. Preserved avianerythrocytes are coated with antigenic components of pathogenic T.pallidum. These Test Cells agglutinate in the presence of specific antibodiesto T.pallidum, and show characteristic patterns in microtitration plates. Any non-specific reactions occurring are detected using the Control Cells, which areavian erythrocytes not coated with T.pallidum antigens. Non-specific reactions mayalso be absorbed out using these Control Cells. Antibodies to non-pathogenic treponemes are absorbed by an extract of Reiter’streponemes, included in the cell suspension. Test results are obtained in 45-60minutes and the cell agglutination patterns are both easily read and long lasting. TPHA test has been shown to be a convenient and specific test for thediagnosis of treponemal infection, having a specificity and a sensitivity comparable to that of the FTA-ABS test (Fluorescent treponemal antibody absorptionTest). It requires minimumlaboratory equipment and is very simple to perform.

77 | Page PHCPI Laboratory Standards Guideline

Material

1. Test cells; preserved avian erythrocytes sensitised with T.pallidum antigen. 2. Control cells; preserved avian erythrocyte. 3. Diluent. 4. Positive control serum; (prediluted 1:20) this will give an equivalenttitre of 1/640:/2560 in the quantitative test. 5. Non-reactive control serum; (prediluted 1:20). 6. Accurate pipettes for delivering 10, 25, 75 and 190 microlitres. 7. U-Well microtitration plates.

Procedure Qualitative method Each sample requires 3 wells of a microtitration plate.

1. Add 190µl of diluent to Well 1. 2. Add 10µl serum to Well 1. 3. Using a micropipette, mix contents of Well 1 and transfer 25µl to Wells 2 and3. 4. Ensure that the Test and Control Cells are thoroughly resuspended. Add 75µl of control cells to Well 2. Add 75µl of Test Cells to Well 3. 5. Tap the plate gently to mix the contents thoroughly. 6. Incubate 45-60 minutes at room temperature. (Keep the plate away from heat, direct sunlight and any source of vibration). 7. Read results. Results are stable for 24hrs if the plate is covered and the above precautions are observed.

Quantitative test

Each sample requires 8 Wells of a microtitration plate. Labelled A through to H.

1. Add 25µl of diluent to Wells B to H inclusive. 2. Transfer 25µl of 1:20 serum dilution from screening test to Wells A and B. 3. Take 25µl of diluted serum from Well B and serially dilute from Wells B to H inclusive in 25µl aliquots, discarding 25µl of diluted serum from Well H. 4. Ensure that the Test Cells are thoroughly resuspended. Add 75µl of Test cells to wells A to H inclusive. This will give a dilution of serum of 1/ 80 in Well A through 1/ 10240 Well H.

78 | Page PHCPI Laboratory Standards Guideline

5. Shake the plate gently to mix the contents thoroughly. 6. Incubate for 45-60 minutes at room temperature. (Keep the plate away from heat, direct sunlight and any source of vibration). 7. Read results. Results are stable for 24hrs. If the plate is covered and the above precautions are observed.

Table 3: TPHA result reading RESULTS TEST CELLS CONTROL CELLS

Strong Positive Full cell pattern covering the No agglutination tight bottom of the well. Button

Weak Positive Cell pattern covers approx. 1/3 of No agglutination tight well bottom Button

Indeterminate Cell pattern shows a distinctly open No agglutination tight centre Button

Negative Cells settled to a compact bottom, No agglutination tight typically with a small clear centre. Button

Non-specific * Positive reaction Positive reaction

Non-specific absorption

1. Add 100µl of test serum to a small tube then add 400µl of Control Cells. Mix well and stand for 1 hour. 2. Centrifuge for 15 minutes at 1000 rpm and test the supernatant by the qualitative method.

Note: The sample is now at 1/5, this should be taken into account when preparingthe dilutions.If the result is repeatedly non-specific the sample should be tested by anothermethod e.g. FTA-ABS.

Interpretation of results

Strong positive reactions may show some folding at the edge of the cell mat. When the Test well is positive, the Control well should be observed.

79 | Page PHCPI Laboratory Standards Guideline

The Control cells should settle to a compact button. They should not be used as acomparison for Non-Reactive serum patterns since the Control Cells will give amore compact pattern than the Test Cells. Agglutination in the Control well indicates the presence of non-specific agglutininsin the sample, the test should be reported as INVALID. A serum that gives thisresult may be absorbed using the Control Cells as detailed under Non-specificabsorption.A doubtful reaction with Test Cells should be reported as INDETERMINATE. This result may indicate a low level of antibody in early primary syphilis or yaws.

This sample should be first retested in the qualitative test then a further sampleshould be tested at a later date to determine whether or not there is a rising titre. It isalso advisable to perform another confirmation test (FTA-ABS) to complete the profile of the test serum.

5.3Widal Test ((agglutination test))

Introduction The widal agglutination test for the diagnosis of the enteric fever, The patients' serum is tested directly for antibody by entire a slide of tube method, These test are either qualitative and semi quantitative. The rapid slide test is used primarily as a screening procedure especially useful when large numbers of sera must be examined. The tube test should be used to confirm positive result obtained by the slide test. The degree to which agglutination occurs depends upon the concentration of the antigen, the amount of antibody present, the composition of the salt solution and the temperature. Sera from normal patients may show positive agglutination with widal antigen due the previously immunization, past infection, or the present of antibodies to related antigen. Titer detected as a result of active infection or recent immunization. It is therefore, necessary to evaluate two or more serum samples, taken at 3 to 5 days intervals after the onset of disease. A progressive increase in titers is the prime evidence of recent infection or immunization.

Martials 1. antigens: Antigen are nonviable bacteria cells(salmonella)in 0.5% phenolized saline and stained 2. Serological pipettes, 0.2 ml with 0.1 ml gradation. 0.1 ml capacity 3. Class agglutination slide with bow 4. Wooden applicator sticks 5. Physiological saline (0.85% NaCl)

80 | Page PHCPI Laboratory Standards Guideline

6. Rotatory (if available)

Procedure: - Specimen collection: Collect 5-10 ml of blood in the sterile plain tube without anticoagulant. Allow the blood to clot for 20-30 min, serum maybe kept for up to 24 hours at 2.8 c. If sample to be kept for longer periods , it should be frozen, do not refreeze one it thawed serum

The antigen should be allowed to warm to room temperature prior to use in test. A- Slide method 1- deliver 0.08,0.04,0.02,0.01 and 0.005 ml of test serum on to separate ring, using 0.2 ml pipette 2- Shake the antigen vial in order to thoroughly mix the contents. 3- Add one drop of antigen to each circle, mix antigen and serum, using a new tooth pick or applicator for each circle. 4- Rock the slide for 1-3 min (lock the kit procedure) or using rotatory shaker if available. 5- The degree of agglutination observe in each ring is estimated as follow. 6- Dilution giving a 2+ agglutination is the end point or titer of that serum.

Note: storage kit: Widal kit, store at 2-8°C, antigen are ready for use as supplied

Measure titer: The titer is the reciprocal of the highest final dilution procedure a 2 +agglutination

Table 4: RecordWidal (agglutination) test

Degree of agglutination

100% back ground clear to slightly hazy 4+

75% back ground slightly cloudy 3+

50% back ground moderately cloudy 2+

25% back ground cloudy 1+

None negative

81 | Page PHCPI Laboratory Standards Guideline

Table 5: Determination of anti-Salmonella species antibodies titre (Widal test)

The titer is reciprocal of the highest dilution producing (2+) agglutination

Serum ml correlation dilution

0.08 1:20

0.04 1:40

0.02 1:80

0.01 1:160

0.005 1:320

Table 6: Example of Widal slide method

Slide test result

Serum ml 0.08 0.04 0.02 0.01 0.005 Serum readying titers or endpoint Dilution 1:20 1:40 1:80 1:160 1:320

Agglutination reaction

Serum No 1 ++++ ++++ ++++ ++ - 1:160

Serum No 2 ++++ +++ ++ - - 1:80

Serum No 3 +++ ++ ++ _ _ 1:80

82 | Page PHCPI Laboratory Standards Guideline

Note: 1-If auto-agglutination of antigen is suspected (Show by low power microscope) the vial antigen is unstable and should be discarded. 2-patints, occasionally fail to develop any serum agglutinin

The chart below give approximate indication of the significant of serum titers.

Table 7: indication of significant serum of anti-Salmonella antibody titre

Serum agglutinins

Disease Febric antigen appear maximum titer and significance widal antigen

Typhoid fever Salmonella O group D 7-10 Days 3-5 week 1:80 (in early stage)=suspicious 1:160 and rising= strong suspicious

Para typhoid Those characterized by fever Salmonella O group A prolonged fever and salmonella Salmonella O group B typhoid like symptoms infection present antibodies of titers similar to above

Note:

1-The minimal titers defined as positive of the O typhoid antigen and H antigen must be determined for individual geographic areas. 2-Cross-reaction occur with both non Styphi group dsalmonella and salmonella from other groups

83 | Page PHCPI Laboratory Standards Guideline

3-A single elevated titer for (O) equal to or greater than (1:320) or H equal to or greater then (1:640) is considered positive.

5.4 Rose Bengal Test for Brucellosis Introduction Human brucellosis –undulant or Malta fever: Brucellosis is systemic infection in which any organ or system of body can be involved. Human become infected by three primary routes 1-ingestion of infected unpasteurized animal milk products 2-inhaltion of infected aerosolized particles 3-Direct contact with infected animal parts through rupture of skin and mucous membrane including the conjunctiva. Brucellosis is one of the most commonly reported bacterial infection acquired in laboratory , within laboratory aerosolizing is primary mechanism of the transmission, so it is required biosafety level 3 precaution for isolation in Lab .

The main way of diagnosis of brucellosis is by serological test. Because of the difficulty of isolating the organism, the serum agglutination test (SAT) is widely used and detected antibodies for Br abroteus and Br melitensis. Rose Bengal test (agglutination test) for brucellosis it is remain an inexpensive sensitive and specified test for acute case , but it is not useful in following up of patients or in diagnosis relapses .

Principle The Rose Bengal an agglutination test reaction used for the diagnosis of brucellosis. The patients’serum is tested directly for antibody by either a slide or tube method. The test are qualitative and semi qualitative. A number of antigens are available commercially for use is the rapid slide test. The test is performed by mixing serum and antigen in ruled squired on a glass slide. This test is widely used and detected antibodies to Brucellosis aborteas, Br. Melitensis. In case of a cute infection a titer of 1:160 or greater in rapid slide method are consider diagnosis if this result fits the clinical and epidemiologic findings. Materials:

1. AntigenBrucellosis aborteas, nonviable. Bacterial cells in 0.05% phenolized saline and stained. 2. Serological pipettes 0.2 ml with 0.1 ml graduation. 0.1 ml capacity

84 | Page PHCPI Laboratory Standards Guideline

3. Glass agglutination slide with row 4. Wooden applicator sticks 5. Physiological saline(0.85% NaCl) 6. Rotator shaker if available

Procedure

Specimen collection

Collect 5-10 ml of blood in the sterile plain tube without anticoagulant. Allow blood to clot for 20-30 mint, serum may be kept for up to 24 hours at 2-8 C° e . If samples to be kept for longer period, it should be frozen don’t refreeze once themed serum

The antigen should be allow to warm to room temperature prior to use in test.

1. Delivers 0.08, 0.04, 0.02, 0.01 and 0.005 ml of test serum on to separate ring using 0.2 ml pipette. 2. Shake the antigen vial in order to thoroughly mix the contents. 3. Add one drop of antigen to each circle mixer antigen and serum, using a new tool pick or applicator for each circle. 4. Rock the slide for 1-3 min (lock the kit procedure) or using rotatory shaker if available. 5. The degree of agglutination observe in each ring is estimated as follow.

Record agglutination Degree of agglutination 100% back ground clear to slightly hazy 4+ 75% back ground slightly cloudy 3+ 50% back ground moderately cloudy 2+ 25% back ground cloudy 1+ None negative

Note: Storage kits Rose Bengal kit store at 2-8C° antigen are ready for use as supplied.

85 | Page PHCPI Laboratory Standards Guideline

Table 8: Determination of antibrucella antibody titer (Rose Bengal Test) The titer is reciprocal of the highest dilution producing (2+) agglutination

Serum Ml correlation dilution

0.08 1:20

0.04 1:40

0.02 1:80

0.01 1:160

0.005 1:320

The serum dilation giving a 2+ agglutination is the end point or titer of the serum. Measure titer is reciprocal of the highest final dilution producing (2+) agglutination Example of rose bagel test result are shown in the flowing table

Table 9: Examples of Rose Bengal results Slide test result

Serum titers or Serum (ml) 0.08 0.04 0.02 0.01 0.005 endpoint readying

dilution 1:20 1:40 1:80 1:160 1:320

Agglutination reaction

Serum No 1 ++++ ++++ ++++ ++ - 1:160

Serum No 2 ++++ +++ ++ - - 1:80

Serum No 3 +++ ++ ++ _ _ 1:80

86 | Page PHCPI Laboratory Standards Guideline

Limitation of the procedure

1- The serum agglutination test SAT measure both antibodies IgG and IgM in serum of patient 2- After cure the antibodies for brucellosis, usually persistent for more than one year 3- Use 2-mercaptoethanol (detect IgG only) to identify chronic or relapse Brucellosis cases. 4- Any titer high than 1:40 with 2-mercaptoethanol lgG indicative active infection (i.e. relapses or chronic infection). 5- A single titer of greater than 160 or more is considered significant. 6- Prozone phenomena may give false negative result which is possible because of the high antibody concentration so serum dilution is recommended to avoid this. 7- False positive reaction may result from cross-reaction with antibodies of cholera, yersinia or francisella. 8- False negative and false positive reaction can be avoided by routinely dilution the serum beyond at 1:320

5.5 C-Reactive Protein (CRP) Test Principle

C - reactive protein (CRP) in patient’s sera has been found in association with acute infections, necrotic conditions and a variety of inflammatory disorders. There is a strong correlation between serum levels of CRP and the onset of the inflammatory process. Monitoring the levels of CRP in patient’s sera indicates the effectiveness of treatment and the assessment of patient recovery.

CRP latex particles are coated with antibodies to human CRP. When the latex suspension is mixed with serum containing elevated CRP levels on a slide, clear agglutination is seen within 2 minutes.

Material

1- Suspension of polystyrene latex (approximately 1%) particles coated with Anti-CRP antibodies. 2- Positive Control. Serum containing CRP antibodies. 3- Negative Control. Serum free of CRP antibodies. 4- Stirrers 5- Plastic slide

87 | Page PHCPI Laboratory Standards Guideline

6- Micro-pipettes capable of dispensing 50μl. 7- Isotonic saline (0.9% NaCI)

Reagent Preparation o o All reagents should be brought to room temperature (20 C to 25 C) and mixed gently to resuspend latex prior to use. Do not induce foaming. The test slide should be thoroughly cleaned before use as traces of detergent or prior specimen may affect the result.

Recommended cleaning procedure:

1. Used cards must be immediately immersed in a disinfectant solution. Follow disinfectant manufactures guidelines. 2. The reaction circles must be physically rubbed with non-abrasive material to ensure removal of possible adhering particles. 3. Thoroughly rinse in purified water. 4. Allow reaction card to dry. 5. Spray cards with a 70% alcohol solution. 6. Allow the alcohol to evaporate prior to re-use.

Procedure

Qualitative Method 1. Allow kit reagents and patient serum to come to room temperature. 2. Transfer one drop (50μl) of patient’s serum to the test circle on the slide. 3. Shake the latex reagent, then using the dropper provided, add one drop of suspension to the test circle. 4. Mix the drops using a disposable stirrer ensuring coverage of the test circle with the mixture. 5. Gently and evenly, rock and rotate the test slide for 2 minutes whilst examining the test slide for agglutination.

Semi Quantitative Method 1. Using isotonic saline prepare serial dilutions of the patients serum (1/2, 1/4, 1/8, 1/16, 1/32, 1/64 and so on) 2. Transfer one drop (50μl) of each serum dilution to the test circle on the slide.

88 | Page PHCPI Laboratory Standards Guideline

3. Shake the latex reagent, then using the dropper provided, add one drop of suspension to the test circle. 4. Mix the drops using a disposable stirrer ensuring coverage of the test circle with the mixture. 5. Gently and evenly, rock and rotate the test slide for 2 minutes whilst examining the test slide for agglutination.

Results and interpretation Examine the test slide under a strong light source after 2 minutes. Kit controls or known level value samples should be tested with each test run. The kit negative control should give a negative result after 2 minutes. The kit positive control should give a positive result at a titre of 1/4 +/- one double dilution after 2 minutes. If levels of controls or users known samples do not give expected results, test results must be considered invalid.

Qualitative Method A positive result is indicated by the obvious agglutination pattern of the latex, in a clear solution. A negative result is indicated by no change in the latex suspension on the test slide.

Latex CRP kit has a detection limit of 6mg/litre of CRP in the patient’s serum. Positive results will be obtained at a CRP serum concentration above 6mg/litre and negative results will be obtained at 6mg/Litre and below.

Semi-Quantitative Method The serum CRP concentration can then be calculated approximately by multiplying the dilution factor (i.e 2, 4, 8 or 16) by the detection limit, i.e. 6, to give the number of mg/Litre concentration e.g. if the agglutination titre appears at 1/8 the approximate serum CRP concentration is 8 x 6 = 48 mg/Litre.

Titres of 500 mg/Litre have been detected withCRP with no prozone (hook) effect.

5.6 ASO latex slide agglutination test Principle

Streptococcal infections are very common in all populations, and a high percentageof people will have antibodies against streptococci. The β-haemolyticgroup A streptococci produce two haemolysins: oxygen-labile streptolysin Oand oxygen-stable haemolysin S. Only reduced (non-oxidized) streptolysin Ois immunogenic and is used for the test. The antistreptolysin O test is basedon the fact that patients with Streptococcus pyogenes (group A streptococcal) infections develop antibodies that inhibit the haemolytic activity of streptolysinO. The antibodies are usually long-lasting and a single increased titreis not an indication of a current

89 | Page PHCPI Laboratory Standards Guideline infection. Only a fourfold or greater rise intitre on successive serum samples taken 10–14 days apart should be consideredindicative of recent infection. This test is mainly used in the diagnosis ofacute rheumatic fever, acute glomerulonephritis and other post-streptococcal diseases.

There are two types of commercial antistreptolysin O test kits:

The ASO latex slide agglutination test is used to screen sera to identify those with raised ASO titres (200 IU or higher).

The ASO tube test is a haemolysis inhibition test that is used to determine ASO antibody titre in serum samples that are positive in the ASO latex slide agglutination test. Atitre of less than 50 IU does not confirm the diagnosis of acute rheumatic fever.

Materials and reagents

1- Disposable cards, with 6 wells eachDisposable dropper 2- Positive control serum 3- Sensitized latex reagent (with streptolysin O) 4- Applicator sticks

Procedure 1- Dilute the serum 1: 20. (Or according to kit literature). 2- Place 1 drop of the serum solution in a well on the disposable card. 3- Use a new dropper to add 1 drop of sensitized latex reagent. 4- Use an applicator stick to mix the two drops and spread them over the entire well. 5- Examine for agglutination within 2 minutes.  A positive reaction appears as a fine agglutination within 2 minutes.  A negative reaction shows no agglutination.

5.7 Rheumatoid Factor (RF) Principle

Rheumatoid Factor (RF) is found in sera of patients with Rheumatoid Arthritis and is believed to be IgM antibodies directed against the patient’s own immunoglobulin G.

RF is a rapid latex agglutination test kit for the detection of (RF) in human serum. RF latex particles are coated with specially purified human gamma globulin, when the latex suspension

90 | Page PHCPI Laboratory Standards Guideline is mixed with serum containing elevated RF levels on a slide, clear agglutination is seen within 2 minutes. RF has a detection limit of 8 IU/ml of RF in the patient serum.

Material

1- Suspension of Polystyrene latex particles (approximatly 1.25%) coated with suitably modified Fc fraction of IgG in stabilizing buffer. 2- Positive Control. Serum containing Rheumatoid Factor antibodies. Working Strength. 3- Negative Control. Serum free of Rheumatoid Factor antibodies. 4- Stirrers 5- Plastic slide 6- Micropipettes capable of dispensing 50 μl. 7- Isotonic saline (0.9% NaCI)

Specimen collection amd preparation

Obtain a sample of venous blood from the patient and allow a clot to form and retract. Centrifuge clotted blood sample and collect clear serum. Fresh serum samples are required. Do not use haemolysed, contaminated or lipaemic serum for testing as this will adversly affect the results. o o Serum may be stored at 2 C to 8 C for up to 48 hours prior to testing. If longer storage is o required, store at –20 C for up to 6 weeks. Thawed samples must be mixed prior to testing. Do not repeatedly freeze-thaw the specimens as this will cause false results.

Reagent preparation o o All reagents should be brought to room temperature (20 C to 25 C) and mixed gently prior to use. Do not induce foaming. The test slide should be thoroughly cleaned before use as traces of detergent or prior specimen may affect the result. Recommended cleaning procedure:

1- Used cards must be immediately immersed in a disinfectant solution. Follow disinfectant manufactures guidelines. 2- The reaction circles must be physically rubbed with non-abrasive material to ensure removal of possible adhering particles. 3- Thoroughly rinse in purified water. 4- Allow reaction card to dry. 5- Spray cards with a 70% alcohol solution. 6- Allow the alcohol to evaporate prior to re-use.

91 | Page PHCPI Laboratory Standards Guideline

Assay Procedure

Qualitative Method 1. Allow kit reagents and patient serum to come to room temperature.

2. Transfer 50μl of patient’s serum to the test circle on the slide.

3. Shake the latex reagent, then using the dropper provided, add one drop of suspension to the test circle.

4. Mix the drops using a disposable stirrer ensuring coverage of the test circle with the mixture.

5. Gently and evenly, rock and rotate the test slide for 2 minutes whilst examining the test slide for agglutination.

Semi Quantitative Method

1. Using isotonic saline prepare serial dilutions of the patients serum (1/2, 1/4, 1/8, 1/16, 1/32, 1/64 and so on)

2. Transfer 50μl of each serum dilution to the test circle on the slide.

3. Shake the latex reagent, then using the dropper provided, add one drop of suspension to the test circle.

4. Mix the drops using a disposable stirrer ensuring coverage of the test circle with the mixture.

5. Gently and evenly, rock and rotate the test slide for 2 minutes whilst examining the test slide for agglutination.

Results and interpretation

Examine the test slide under a strong light source after 2 minutes. Kit controls or known level value samples should be tested with each test run. The kit negative control should give a negative result after 2 minutes. The kit positive control should give a positive result at a titre of 1/4 +/- one double dilution after 2 minutes. If levels of controls or users known samples do not give expected results, test results must be considered invalid.

Qualitative method

A positive result is indicated by the obvious agglutination pattern of the latex, in a clear solution. A negative result is indicated by no change in the latex suspension on the test slide.

92 | Page PHCPI Laboratory Standards Guideline

Positive results will be obtained at a RF serum concentration of 8 IU/ml or more and negative results will be obtained at a RF concentration below 8 IU/ml.

Semi – Quantative Method

The serum RF concentration can then be calculated approximately by multiplying the dilution factor (i.e 2, 4, 8 or 16) by the detection limit, i.e. 8, to give the number of IU/ml concentration e.g. if the agglutination titre appears at 1/8 the approximate serum RF concentration is 8 x 8 = 64 IU/ml.

Titres of 1074 IU/ml have been detected with Avitex RF with no prozone (Hook) effect.

5.7 Toxoplasmosis Test Toxoplasma IgG ELISA

Introduction

Toxoplasma gondii is an intracellular protozoan parasite with a worldwidedistribution. Although cats are the definitive host, human infection resultsfrom ingestion of contaminated soil, careless handling of cat litter, and ingestion of raw or undercooked meat or transmission from mother tofoetus through the placenta. When a seronegative woman becomes infected with T. gondii duringpregnancy, the organism is often transmitted to the foetus. Infectionduring the first trimester may lead to spontaneous abortion, stillbirth, or overt disease in the neonate.

The Toxoplasma IgG kit is a rapid ELISA designed for the qualitativeor quantitative detection of IgG antibodies to Toxoplasma gondii inhuman serum.

Principle of the test

Diluted serum or plasma specimens (1:100) are incubated for 20 minutes to allow specific antibodies to T. gondii to bind to the antigencoatedwells. After washing away unbound antibodies and otherserum constituents, T. gondii specific IgG is detected using rabbitanti- human IgG conjugated to horseradish peroxidase. After 20minutes incubation, unbound conjugate is removed by washing, andTMB enzyme substrate is added for 10 minutes. A blue colourdevelops if antibodies to T. gondii are present. Addition of stopsolution gives a yellow colour and the optical densities of controls, standard(s) and samples are measured using a reader.

93 | Page PHCPI Laboratory Standards Guideline

Materials

1. Microplate 96 wells in 12 X 8 break-apart strips, pre-coated withT.gondii purified antigens enriched for P30 antigen

2. Reagent 1: Sample Diluent 100mM Tris-buffered saline, pH 7.2 withantimicrobial agent, 10ml, (blue), concentrate (x10)

3. Reagent 2: Wash Buffer 100mM Tris-buffered saline with detergent, pH 7.2, 100 ml, concentrate (x10)

4. Reagent 3: Conjugate (peroxidase conjugated rabbit anti-humanIgG), 12 ml, (red), ready to use

5. Reagent 4: TMB Substrate, 12 ml, ready to use

6. Reagent 5: Stop solution, 12 ml, ready to use

7. Standards1: (for quantitative assays), 15 IU/ml, 50 IU/ml; 150IU/ml (blue), ready to use

8. Standard1: (for qualitative assays) 8 IU/ml, (yellow), ready to use

9. Positive control1: 100 IU/ml (red), 1ml, ready to use

10. Negative control: 1 ml (green), 1ml, ready to use

11. Distilled or de-ionised water

Equipment

1- 10mm X 60mm tubes for dilution, 2- Pipettes 10μl, 100μl, 1000μl; 3- Repeating dispenser 100μl, 4- Microplate reader with 450nm filter, 5- Microplate washing device.

Samples

Only freshly drawn and properly refrigerated sera or plasma shouldbe used in this assay. Avoid haemolysed, lipemic or bacterialcontaminated sera. Sera should be stored at 2-8ºC for no longer than5 days. If delay in testing is anticipated, store test sera at –20ºC.Avoid multiple freeze-thaw cycles.

94 | Page PHCPI Laboratory Standards Guideline

Method

Ensure that all materials are at room temperature before beginningthe procedure. We recommend that the standards and the controlsare always run in duplicate. Samples may be run singly or induplicate.

1. Assemble the number of strips required for the assay.

2. The sample diluent X10 concentrate contains 0.09%sodium azide as preservative. Prepare sufficient workingstrength diluent for the assay run. However, if the workingstrength diluent is to be stored for more than 1 week, addsodium azide (0.9g/L). Store unused sample diluentconcentrate and dilute sample diluent at 2 – 8oC. Dilute theSample Diluent (Reagent 1) 1:9 in distilled water to makesufficient buffer for the assay run e.g. add 10ml sample diluentconcentrate to 90 ml water.

3. Dilute patient samples 1:100 (e.g. 5μl serum plus 0.5 ml diluent). It is important to dispense all samples andcontrols into the wells without delay. Therefore ensure thatall samples are ready to dispense.

4. For qualitative determinations, dispense 100 μl of thenegative control, the 8 IU/ml standard, the positive controland the diluted patient sample into the wells. Forquantitative determinations, use sample diluent as 0 IU/mland additionally dispense the 15 IU/ml, 50 IU/ml and 150IU/ml standards.

5. Place the strips into the incubation bag provided andincubate for 20 minutes at room temperature. During allincubations, avoid direct sunlight and close proximity toany heat sources.

6. Dilute the Wash Buffer (Reagent 2) 1: 9 in distilled water tomake sufficient buffer for the assay run e.g. add 50ml washbuffer concentrate to 450ml water. The diluted wash buffer isstable for two months at 2 - 8°C.

7. After 20 minutes, decant or aspirate the well contents andwash the wells 3 times using an automatic plate washer orthe manual wash procedure (see below). Careful washingis the key to good results. Blot the wells on absorbentpaper before proceeding. Do not allow the wells to dry out.

95 | Page PHCPI Laboratory Standards Guideline

Manual Wash Procedure:

Empty the wells by inversion. Using a multi-channel pipetteor , fill the wells withash buffer. Empty byinversion and blot the wells on absorbent paper. Repeatthis wash process two more times.

8. Dispense 100μl of Conjugate (Reagent 3) into each well.

This reagent is color coded red. Keep all pipettes andother equipment used for Conjugate completely separatefrom the TMB Substrate reagent! Incubate the wells for 20minutes in the incubation bag at room temperature.

9. After 20 minutes, discard the well contents and carefullywash the wells four times with wash buffer. Ensure that thewells are completely washed. Blot the microplate onabsorbent paper to remove final drops of wash fluid. Donot allow the wells to dry out.

10. Using a repeating dispenser, rapidly dispense 100μl of TMBSubstrate (Reagent 4) into each well. Incubate the plate for10 minutes.

11. Add 100μl of Stop Solution (Reagent 5) to each well. Toallow equal reaction times, the stop solution should beadded to the wells in the same order as the TMB Substrate.

12. Read the optical density in a microplate reader within 10minutes.

Interpretation

Qualitative determinations

Negative samples: OD < 8 IU/ml OD

Positive samples: OD >/= 8 IU/ml OD

Quantitative determinations

Plot the optical densities of the standards against their respectiveconcentrations. Draw a line to join the points. Read theconcentrations of unknowns from this graph. Concentrations below 8IU/ml are considered negative; concentrations above 8 IU/ml areconsidered positive for anti- toxoplasma IgG.A negative result indicates no current or previous infection with T.gondii. Such individuals are presumed to be susceptible to primaryinfection. However see Limitations below.A positive result indicates a current or previous infection with T.gondii.

96 | Page PHCPI Laboratory Standards Guideline

5.8 Infectious Mononucleosis test

Principle Infectious Mononucleosis is a qualitative determination of the nonspecific heterophile antibodies that result from primary infection by Epstein – Barr Virus and causing the clinical presentation of Infectious mononucleosis. The IM- Latex is a slide agglutination test also can get semi-quantitative detection of Heterophile antibodies. Latex particles coated with antigenic extract of beef erythrocyte membranes and are agglutinated when mixed with samples containing IM heterophile antibodies.

Materials and reagents Preparedness:

 Reagents

Table 10: Reagents of Infectious Mononucleosis Test Latex Latex particles coated with antigenic extract of beef erythrocytes membranes, phosphate buffer, pH 7.2. Sodium azide , 0,95 g/L. Control +ve Human serum with anti – IM antibodies titer ≥ 1/4. Sodium azide , 0.95g/L Control -ve Animal serum. Sodium azide , 0.95g/L

Disposable slides Storage: when stored tightly closed at 2-8 oCdo not freeze. (According to kit literature)

 Samples: Fresh serum. Stable 7 days at 2-8oC or 3 months at – 20oC. Samples with presence of fibrin should be centrifuged. Do not use highly hemolized or lipemic samples.  Equipment : 1- Mechanical rotator with adjustable speed at 80-100 r.p.m. 2- Centrifuge. 3- Automatic micropipette

97 | Page PHCPI Laboratory Standards Guideline

Methods:

 Qualitative Method : 1- Allow reagents and samples to reach room temperature. 2- Place 50 ul of the sample and one drop of each Positive and Negative controls into separate circles on the slide test. 3- Swirl the IM – Latex reagent gently before using and add one drop (50ul) next to the samples to be tested. 4- Mix the drops with stirrer, spreading them over the entire surface of the circle. Use different stirrers for each sample. 5- Place the slide on a mechanical rotator at 80-100 r.p.m. for 2 minutes .False positive results could appear if the test is read later than two minutes.  Semi quantitative method: 1- Make serial two ford dilutions of the sample in 9 g/L saline solution. 2- Proceed for each dilution as in the qualitative method.

Results and Interpretations in reports:

Examine macroscopically the presence or absence of visible agglutination immediately after removing the slide from the rotator. The presence of agglutination indicates a titer ≥ 1/28 of anti IM Heterophil antibodies by the Davidson method. The titer in the semi- quantitative method is defined as the highest dilution showing a positive result. The following are the limitations of the test: 1- False positive results may be obtained in some geographical areas where the horse serum is used as a prophylactic measure (vaccination). Also this could be seen in patients with leukemia, Burkitts Lymphoma, Pancreatic carcinoma, viral hepatitis, CMV infections and others.

2- False negative results also encountered and in such Patients either repeat testing after interval of several days to get a high heterophil antibody titer or look for specific antibodies against viral structural antigens like anti VCA IgM.

Quality Control:

Positive and Negative controls are recommended to monitor the performance of the procedure, as well as the comparative pattern for a better result interpretation.

98 | Page PHCPI Laboratory Standards Guideline

5.9 Rota Virus Latex Test

Principle of the method and clinical significance: Rapid latex agglutination assay for the detection of in faecal samples this virus has been shown to be principle causative agent of gastro-enteritis. Patients may harbor up to 108 virus particles per gram of faeces.

Materials and reagents Preparedness:  Material insert in kit: Test reagent latex, control negative ,Positive bovine rota-virus control inactivated ( ready to use ),5 x concentrated extraction buffer pH 7.2,Agglutination slides ,pipette  Specimen collection and storage: Prepare a 10% suspension of the faecal sample by adding 0.1 g /0.1 ml of sample to 1 ml of extraction buffer in screw capped vial. Mix well and stand in room temperature for 2 minutes. Specimen may be stored 24 hours at the time of the test .If storage is more than 24 hour the specimen should be stored at -20 oCin freezer.  Safety warnings: Do not pipette by mouth, use disinfectant 70% alcohol, waste must be treated as biohazardous and incinerated.

Method:

1- Add 10 ml of extraction buffer to 40ml de-ionize water. 2- Add 2 gms of faecal sample to 2 ml of extraction buffer. 3- Mix well on . 4- Stand at room temperature for 10 minutes. 5- Centrifuge test sample for 10 minutes about 800 g. 6- Place one drop of the supernatant from the faecal sample onto the circle of test slide and use the 2 circle for control reagent. 7- Add one drop of control latex reagent next to the drop of faecal sample on circle 1. 8- Add one drop of test latex reagent next to drop of faecal sample on circle 2. 9- Spread the control latex reagent and test latex reagent in circle 1, 2 by disposable pipette. 10- Gently tilt the test slid backward and forward for 2 minutes.

Results and Interpretations in reports:

1- Positive result is indicated by the visible agglutination of the latex particles which occur within few seconds.

99 | Page PHCPI Laboratory Standards Guideline

2- Negative result is indicated by milky appearance without any visible aggregation of the latex particles.

5.10 Pregnancy Test

5.10.1 Rapid test for detection of human Chorionic Gonadotrophin in Urine Human chorionic gonadotrophin (hCG), a glycoprotein hormone secreted by viable placental tissue during pregnancy, is excreted in urine approximately 20 days after the last menstrual period. The levels of hCG rise rapidly reaching peak levels after 60-80 days. The appearance of hCG in urine soon after conception and its rapid rise in concentration makes it an ideal marker for the early detection and confirmation of pregnancy. Test kitutilizes the principle of Immunochromatography, a unique two-site immunoassay on a membrane. As the test sample flows through the membrane assembly within the test device in place of the dipstick, the coloured monoclonal anti-hCG-colloidal gold conjugate complexes with the hCG in the sample. This complex moves further on the membrane to the test region where it is immobilised by the monoclonal anti-hCG coated on the membrane, leading to the formation of a pink coloured line which confirms a positive test result. Absence of this coloured line in the test region indicates a negative test result. The unreacted conjugate, unbound complex, if any, and the colloidal gold conjugated rabbit IgG moves further along the membrane and are subsequently immobilised by the goat anti-rabbit IgG coated on the control region forming a pink line. This control line serves to validate the test results.

Specimen collection and preparation Urine samples collected at any time may be used, however, it is recommended that to maximize hCG concentration, the first voided morning specimen should be used. The urine should be collected in a clean dry container (plastic or glass) which must be free from detergent. Urine specimens should be as fresh as possible and it is preferable to test within 24 hours of collection. The sample may be stored for longer periods (72 hours) prior to use, if stored, store at 20°C to 80°C.

Filtration or centrifugation is generally not necessary for urine used in thistest, however, if a sample is very turbid, centrifugation or filtration may be necessary. (The use of supernatant from turbid samples allowed to sediment naturally prior to use may negate the need for sample preparation – this does not affect the hCG concentration.)

Reagent preparation

Devices and samples should be brought to room temperature (20 to 25)°C and mixed gently prior to use.

100 | Page PHCPI Laboratory Standards Guideline

In case the pouch has been stored at 4°C to 8°C, allow at least 30 minutes for the device to come to room temperature. Check the color of the desiccant. It should be blue. If it has turned colorless or faint blue, discard the device and use another device.

Assay Procedure

1. Open the pouch and remove the device. Once opened, the device must be used immediately. 2. Dispense two drops of urine specimen into the sample well „S‟ using the dropper provided. 3. Read the results at the end of fifteen minutes for serum or five minutes for urine.

Results and interpretation Negative: Only one colored line appears on the control region „C‟ only

Positive: A distinct colored line appears on the control region „C‟ and on the test region “T”.

Figure 14: Results of Pergnancy Test (PT)

The test should be considered invalid if no line appears. Repeat the test with a new device. Depending on the concentration of hCG in the specimen, positive results may appear as early as 30 seconds. Negative results must be confirmed only at the end of fifteen minutes for serum or five minutes for urine.

5.10.2 Direct hCG Latex Pregnancy Kit Detectable levels of Chorionic Gonadotropin (hCG) in urine startat 5 mIU/ml during the first week of gestation and rises to100,000 mIU/ml at 2 to 3 months. The hCG level doublesapproximately every 2.2 days during the first trimester1. Valuesdecline from 10% to 15% of peak concentrations during 2nd and3rd trimesters2. Direct hCG Latex kit contains one reagent of latexparticlescoated with monoclonal antibodies to hCG. The reagent is mixed with the urine samples.

101 | Page PHCPI Laboratory Standards Guideline

MATERIALS

1. Latex reagent. 2. Positive and negative controls. 3. Reaction slide and stirring sticks.

SPECIMEN COLLECTION AND PREPARATION Generally, the first morning urine contains the highestconcentration of the hCG hormone and therefore, it is morerecommended for testing. However, urine collected at otherperiods can also be used. The urine in this case shouldhave been kept in 2-8°C and used within 72 hours fromcollection time.

PROCEDURES 1. Bring reagents to room temperature. 2. Place 50μl of patient urine, one drop positive and onedrop negative controls into different circles of the slide. 3. Add one drop of latex reagent directly to each sample. 4. Mix using the supplied sticks and spread the mixtureover the entire circle. 5. Gently rock the slide. Agglutination may be observedafter two minutes. Direct light source may help to observe the results.

Reading the results Presence of agglutination within two minutes indicates positive reaction.

Lack of agglutination within two minutes indicates negative reaction.

102 | Page PHCPI Laboratory Standards Guideline

6. Bacteriology

6.1General urine examination Introduction:

The examination of urine provides a wide variety of useful medical information regarding the disease involving the kidney and lower urinary tract. Both functional (physiologic) and structural (anatomic) disorders of the kidney and lower urinary tract may be elucidated.as diagnostic and monitoring, we believe laboratory urine tests will remain an essential part of clinical medicine. The basic routine analysis consist of 4 parts: specimen evaluation, gross (physical) examination, chemical screening, and sediment examination.

Clinical significance

Examination of urine is a fundamental investigation in patients in whom kidney disorders or infections of the urinary tract are suspected. There are also many patients who exhibit no clinical symptoms, but in whom previously unrecognized urinary tract infections can be diagnosed by urine examination

Collection of urine specimens

Containers for the collection of urine should be wide-mouthed, clean and dry. And we can also be used the adhesive bag for infants (Fig.??) .If the urine specimen has to be transported for any length of time it should contain an appropriate preservative to prevent bacterial overgrowth or hatching of viable ova .The specimen can be preserved for microscopic examination of the deposit by adding 8–10 drops of formaldehyde, 10% solution per 300 ml of urine. Urine preserved in this way is not suitable for other tests. Urine to be examined under the microscope must be freshly passed into a clean dry vessel. A midstream urine specimen is the most useful. Urine stored in a refrigerator may contain an excess of precipitated salts and will not be suitable for microscopy.

Figure (15) Adhesive bag for infanturine collection

103 | Page PHCPI Laboratory Standards Guideline

Principle the elements of routine urinalysis include macroscopical examination the evolution of physical characteristics ( color, odor, and specific gravity ) pH and chemical examination ( protein , glucose. Keton bodies, bilirubin, urobilinogen. Bile pigment and bile salt) and microscopic inspection of centrifuge sediment for R.B.C pus, cast, crystal, etc.

Material (glassware& Equipment):

1- Centrifuge 2- Microscope 3- Urine cup 4- Conical centrifuge tube 5- Slides 6- Cover slide 7- Watch glasses 8- Dropper 9- Forceps 10- Indicator paper of limited pH range: for the 5.0–7.0 range and for the 6.0–8.0 range. 11- Test tube 12- Burner 13- Urinometer 14- Pasteur pipette

REAGENTS

1- 20 % 5-sulfosalicylic acid 2- Ehrlich reagent 3- Acetic acid 4- Lougal iodine 5- Dipstick strip (reagent strip): should be considered as a first choice because of its practical, easy and cost effective.

104 | Page PHCPI Laboratory Standards Guideline

6.1.1 Macroscopical Examination of Urine Appearance

Urine is normally clear straw-yellow in color. More concentrated urine may appear dark yellow, the presence of blood cells or excess salts may make the urine appear cloudy.Pigments from bile substances may make the urine appear deep yellow or brown.Urine can occasionally appear colorless. Report the appearance as clear or cloudy, colorless, pale yellow, deep yellow or brown.

Color Urine varies in appearance, depending principally upon a body's level of hydration, as well as other factors. Normal urine is a transparent solution ranging from colorless to amber but is usually a pale yellow. In the urine of a healthy individual the color comes primarily from the presence of urobilin. Urobilin in turn is a final waste product resulting from the breakdown of heme from hemoglobin during the destruction of aging blood cells.

Colorless urine indicates over-hydration, generally preferable to dehydration (though it can remove essential salts from the body). Colorless urine in drug tests can suggest an attempt to avoid detection of illicit drugs in the bloodstream through over-hydration (see Fig.16)

Figur (16) Urine color Chart

105 | Page PHCPI Laboratory Standards Guideline

 Colorless very dilute urine caused by polyuria , diabetes insipidus  Cloudy caused by phosphate, carbonate urates. , leukocytes. Red cell bacteria. fungi ,spermatozoa ., mucin  Dark yellow urine is often indicative of dehydration.  Yellowing/light orange may be caused by removal of excess B vitamins from the bloodstream.  Certain medications such as rifampin and phenazopyridine can cause orange urine.  Bloody urine is termed hematuria, a symptom of a wide variety of medical conditions  Dark orange to brown urine can be a symptom of jaundice, rhabdomyolysis, or Gilbert's syndrome.  Black or dark-colored urine is referred to as melanuria and may be caused by a melanoma.  Pinkish urine can result from the consumption of beets.  Greenish urine can result from the consumption of asparagus.  Reddish or brown urine may be caused by porphyria (not to be confused with the harmless, temporary pink or reddish tint caused by beeturia).  Milky caused by lipiduria  Blue urine stains can be caused by blue diaper syndrome

.

Odor

Normally with aromatic odor. Abnormally: Offensive odor due to pus increase. Acetone odor due to ketones increase (ketonuria).

Reaction (PH)

Principle

Colored indicator paper is dipped in the urine (or placedin a and a few drops of urine are added to it).The color changes according to the PH.The paper is then compared with a standard control chartgiving the corresponding pH value.

106 | Page PHCPI Laboratory Standards Guideline

Method

The urine specimen must be tested within 1 hour ofcollection

1- Place a strip of universal indicator paper in a watch glass, let a few drops of fresh urine fall from the dropper on tothe paper alternatively, and dip the test paper directly into the urine inthe receptacle. 2- Pick the strip of paper up with forceps.Compare the colorobtained with those shown on thestandard chart .Read off the pH unit given forthe color that matches the test paper most closely. 3- According to the result obtained, select a strip of indicatorpaper for the corresponding limited range. For example,if the pH is 6, use indicator paper for the range 5.0–7.0.If the pH is 7 or more, use indicator paper for the range6.0–8.0.The pH of urine is normally about 6.0 (range 5.0–7.0). AcidpH values (4.5–5.5) are observed in some forms of diabetes, muscular fatigue and acidosis. Alkaline pH values (7.8–8.0) are common in patients with infections of the urinarytract and in people on a vegetarian diet. Determination of the pH of urine is useful for the identification of crystalline

Detection of Glucose in urine

Introduction:

Glucose may appear in urine at different blood glucose level, varying in individuals’ blood level, Glomerular, blood flow, tubular reabsorption rate and urine flow influence its appearance. Glucosuria usually occur when blood level is more than 180 to 200 mg/dl

Principle

Glucose is the most commonly found sugar substance in urine, particularly in diabetic patients and patients suffering from chronic renal failure.

Method

Glucose in urine can be detected using a urine dipstick, the dipsticks are placed into the urine and immediately removed. They are then compared with a comparison chart after an appropriate time that is also specified on the chart

The color changes observed on the dipstick will give a semi-quantitative estimation of the amount of substance present. This can be reported as negative, Trace,+ (160-195),++ (>200), +++ (250-300). Dipsticks must be stored according to the manufacturer’s instructions

107 | Page PHCPI Laboratory Standards Guideline

Detection and estimation of protein

Principle:

Protein is not normally found in urine, this is because kidneys ensure that large molecules such as protein remain in the blood. However, the kidneys sift out the smaller impurities and expel them through the urine. In case some protein does get through, the body usually re- absorbs it and uses it as an energy source. However, if there is too much protein in the blood, then some amount of protein will be present in the urine. This can happen even if the kidneys are functioning properly Elevated protein levels are observed in the urine of patientswith:

 urinary Schistosomiasis  chronic renal disease  pyelonephritis  diabetes mellitus systemic disorders (lupus erythematosus) multiple myeloma  However, orthostatic proteinuria, a form of functional proteinuria usually seen in young men, which occurs on standing up and disappears on lying down, has nopathological significance

Method

Urine for Albumin test done by one of the following:-

 Acetic acid test 1. Prepare 1 ml of urine in a test tube then heat gently till ebullition. 2. If the solution becomes turbid that indicates the presence of albumin or amorphous urate. So you must put acetic acid20% for insurance if the turbidity precipitated so it is amorphous urate, if still turbid so it is albumin in urine.

 Sulfosalicylic acid test

1-To approximately 3ml of supernatant urine in test tube and one equal amount of 3%SSA invert in the mix 2- Let to stand exactly 10 minute 3-Invert again twice. 4-observe the degree of precipitation and grade the results according to the following:

108 | Page PHCPI Laboratory Standards Guideline

Table 11: Degree of Albumin Precipitation

Negative No turbidity Trace perceptible + Distinct turbidity but no discrete granulation ++ Turbidity with granulation but no flocculation +++ Turbidity with granulation and flocculation ++++ Clumps of precipitated protein or solid precipitate

Bence-Jones protein test

The Bence-Jones protein urine test is a test for the protein by the same name. The Bence-Jones protein is a protein that is a globulin that is excreted in the urine. The significance of finding the Bence-Jones protein in urine is that it is an indication of some neoplasm in the body. This is usually in the case of cancers like bone marrow cancer. The presence of the protein can also be an indication of the presence of anemia, bone disease, and renal failure. The test is also used to screen for a rare condition called Waldenstrom’s macroglobulinemia a kind of lymphoma involving the white blood cells.

Reasons The Bence-Jones protein test is a test that is used to diagnose kidney failure, bone disease, and bone marrow cancer. This is one of the preliminary tests that are conducted before more invasive techniques are performed like biopsies. This may also be followed by other tests for tumor markers. Anemia testing may be done be a hematocrit. Cancers of the bone marrow can be particularly hard to identify in the initial stages because it is within the bone and does not present itself as any kind of specific swelling or symptoms. The only way that it can be diagnosed is with the help of tumor markers and more invasive tests later on.

Procedure The test is done by taking a urine sample from the patient. This should ideally be done in the morning when the concentration of urine is at its optimal amount. The urine test for Bence- Jones protein is then conducted using a method called electrophoresis. This is basically a method of concentrating urine. The substitute for this is to simply heat the urine. After heating

109 | Page PHCPI Laboratory Standards Guideline the urine, the Bence-Jones protein will then precipitate and crystallize at a certain temperature before dissolving back into the urine at the boiling point

Test: - Centrifuge the urine and pour 5ml of the supernatant fluid into a test tube - Add acetic acid if the urine alkaline until the urine become acidic - Place in the tube and heat oven or Bunsen burner flam slowly .watching for any cloudiness appear. Bence Jjones protein being appear at 40-60C

and disappear over 70C allow the urine to cool to about 60°C, the Bence-Jones protein reappear.

Specific gravity Test clinical significance

Urinalysis is a parameter commonly used in the evaluation of kidney function and can aid in the diagnosis of various renal diseases (eg diabetes insipidus).

Principle

Specific gravity (which is directly proportional to urine osmolality which measures solute concentration) measures urine density, or the ability of the kidney to concentrate or dilute the urine over that of plasma. Dipsticks are available that also measure specific gravity in approximations. Most laboratories measure specific gravity with a refractometer Specific gravity between 1.002 and 1.035 on a random sample should be considered normal if kidney function is normal. Since the specific gravity of the glomerular filtrate in Bowman's space ranges from 1.007 to 1.010, any measurement below this range indicates hydration and any measurement above it indicates relative dehydration

Method

 The Urinometer vessel is filled three fourth full with urine ( minimum volume of urine 15 ml)  The urinometer is inserted with spinning motion  Make sure that is floating freely ( which reading the urinometer be sure that it is not touching the sides or the bottom of cylinder avoid surface bubble read a bottom of the meniscus  we can identifying the specific gravity by using dipstick

110 | Page PHCPI Laboratory Standards Guideline

Bilirubin Test

Introduction

Bilirubin, a product of hemoglobin catabolism, is characterized by its distinctive yellow pigment. The presence of bilirubin in urine is always abnormal. In most healthy individuals the amount of conjugated bilirubin excreted is not detected by the strips. In cases when bilirubin is elevated and is conjugated, it will be detected by the test strip.It is important to note that unconjugated bilirubin cannot be excreted by the kidneys because it is bound to albumin and is not soluble in water. In the liver, bilirubin combines with glucuronic acid through the action of a glucuronyl transferase to form water soluble bilirubin diglucuronide. Under normal circumstances, conjugated bilirubin passes from the and then to the intestinal tract. Intestinal bacteria reduce conjugated bilirubin to urobilinogen. Approximately half of the urobilinogen is excreted in the feces; most of the other half is recirculated through the liver. A small amount of urobilinogen bypasses the liver and is excreted in the urine

Normally bilirubin is not found in a routine urine test. This test is done to detect abnormally high urine concentration of direct (conjugated) bilirubin and help to identify the cause of jaundice.

Clinical significance

The presence or absence of bilirubin in urine must be correlated with serum bilirubin test, the specimen appearance is dark with yellow foam

Detection Bilirubin in urine 1.Prepare 1 ml of urine in a test tube then put 2 -3 dropsLugol Iodine on the wall of test tube gently. 2. A violet ring appeared on the surface of urine 3. Shake the tube, then solution becomes violet or brownish 4- if violet ring appeared on the surface of urine mean positive result

Urobilinogen test

Urobilinogen testis directly associatedwith the function of liver and any disorder or infection related to thisorgan can be traced in theurobilinogentestnormal specimen contain small amount of urobilinogen (0.1 to 1 Ehrlich units per dl of urine) this test done to aid diagnosis of extra hepatic obstruction such as blocking of the common bile duct and to aid differential diagnosis of hepatic and pre hematologic disorders (haemolysis of R.B.C).

111 | Page PHCPI Laboratory Standards Guideline

Principle

The test for urobilnogen is based on the Ehrlich Aldehyde Reaction. P- dimethylaminobenzaldehyde in an acid medium with a color enhancer reacts with urobilinogen to form a pink-red color. The urine chemical reagent strip reactivity increases with increasing temperature

Method

Urobilinogen Analysis

1- Add 2.5ml of afresh specimen of urine into a tube 2- Add o.25ml of Ehrlich reagent 3- Allow to stand for 5 minute 4- A deep color of the urine indicate urobilinogen is present in increased amounts. A fine pink or brown color indicates that urobilinogen present in normal amount.

Note: the optimum temperature for testing is 22-26 °C. a freshly voided sample is always best for ensuring optimal results.

We can also be use the dipstick to identify the urobilinogen in urine

False-Positive and False-Negative Urobilinogen Results

False-positive results: A false positive urobilinogen reaction may occur with the chemical reagent strip method when substances known to react with Ehrlich's reagent.

False positive:

Sulfonamides and p-aminosalicylic acid are present in the urine. Drugs that contain Azo dyes, such as Azo Gantrisin®, have a gold color that masks the reaction, causing a false positive reaction. Atypical color reactions may be obtained in the presence of high concentrations of p-aminobenzoic acid. The chemical reagent strip urobilinogen test cannot detect porphobilinogen in a urine specimen. Porphobilinogen is a molecule formed during the synthesis of the heme portion of hemoglobin. False-negative resultsDue to the instability of urobilinogen.

112 | Page PHCPI Laboratory Standards Guideline

False negative: A false negative result may occur using the chemical reagent strip method if the urine specimen has remained at room temperature for an extended period of time exposed to light. A false negative result may also occur if formalin is present.

Detection of ketones bodies Introduction Normal urine does not contain ketone bodies. Acetone and other ketones bodiesmay appear in urine:  In severe or untreated diabetes( diabetic ketoacidosis )  In certain other conditions (dehydration, vomiting, starvation and following strenuous exercise). Principle

When sodium nitroprussideis added tourine containing ketone bodies, a purple color is produced.

Materials and reagents

1- Test-tubes 2- Test-tube rack 3- Measuring cylinder, 10ml 4- Dropping pipette 5- Sodium nitroprusside crystals 6- Acetic acid 7- Ammonia.

Method

1- Just before carrying out the test, place a sufficient amount of sodium nitroprusside crystals into a test-tube to cover the bottom 2- Add 5ml of distilled water. Shake well until the crystals are almost dissolved. (Not all the crystals are expected to dissolve as the solution is saturated.) 3- Measure 10ml of urine into another test-tube. 4- Add four drops of acetic acid to the urine, followed by 10 drops of freshly prepared Sodium nitroprusside solution and mix well

113 | Page PHCPI Laboratory Standards Guideline

5- Holding the tip of the pipette against the side of the tube, let 20 drops (1 ml) ofammonia solution flow on to the surface of the liquid Wait for 5 minutesbefore readinga positive result may be obvious before this time

Result

If the result is positive purple ring appears on top of the urine. If the result is negative, no color change occurs Reporting the results of the test for detection of ketone bodies in urine a: Positive reaction; b: negative reaction. Substances in urine Color change Result None = Negative Pink ring = + Red ring= ++ Purple ring =+++ Ketone bodies in urine can also be detected using a urine dipstick

6.1.2 Microscopical Examination of Urine Detection of abnormal elements Urine contains cells and crystals in suspension that can be collected by centrifugation or by allowing the urine to stand and the suspended particles to form a sediment. The resulting urinary deposit can be examined under the microscope In certain diseases of the urinary tract, the urinary deposits are considerably altered.

Method Preparation of the deposit

1- Mix the urine specimen gently and pour 10 ml into a centrifuge tube. 2- Centrifuge the specimen at medium speed (1500RPM) for 5 minutes. 3- Pour off the supernatant by quickly inverting the tube without shaking. (The supernatant may be used for biochemical testing.) 4- Tap of the tube to re-suspend the deposit. 5- Transfer one drop of the deposit on to a slide using a Pasteur pipette and cover with a cover slip 6- Label the slide with the patient’s name or identification number.

114 | Page PHCPI Laboratory Standards Guideline

Microscopic examination

Using the X10 objective and with the condenser lowered, scan the cover slip all over to look for ova of Schistosoma haematobium when indicated. Using the X40 objective and with the condenser lowered or aperture reduced,scan the coverslip area again and report any findings as a quantitative value for each high-power field The following may be found in urine 1- Erythrocytes 2- Leukocytes 3- Epithelial cells 4- Casts 5- Fungi 6- Bacteria 7- Crystals 8- Parasite eggs and larvae( Schistosoma haematobium ) 9- Trichomonas vaginalis 10- Spermatozoa.

 Erythrocytes

Erythrocytes in urine may be:

(a) Intact: small yellowish discs, darker at the edges (8um); (b) Crenated: spiky edges, reduced diameter (5–6um); (c) Swollen: thin circles, increased diameter (9–10um). The shape of the cells often changes during storage of urineand does not have any diagnostic importance. In normal individual s occasional red blood cell (0-2/HPF) may be seen in microscopic examination of the sediment Note: Erythrocytes may be found in the urine of women if the specimen has been taken during the menstrual period.

 Leukocytes Found in urine may be:

(a) Intact: clear granular discs, 10–15mm (the nuclei may be visible); (b) Degenerated: distorted shape, shrunken, less granular,

115 | Page PHCPI Laboratory Standards Guideline

(c)Pus: clumps of numerous degenerated cells. The presence of many leukocytes, especiallyin clumps, indicates a urinary tract infection

How to express the quantity leukocytes found in urine deposits Place one drop of urine deposit on a slide and cover with a cover slip. Using the x40 objective, examine the deposit and count the number of leukocytes per microscope field (HPF).

Report the results as described:

Number of leukocytes per microscope field Result Cell/per high field Reporting Results 0-5 Normal

5-10 Few 10- 20 + 20-30- ++ 20-50 +++ `more than 50 Abundant

Epithelial Cell

Different types of epithelial cells can be observed in urine. Some of these are readily identifiable; however it is difficult to distinguish small transitional epithelial cells from WBC and renal tubular epithelial cells from transitional epithelial cells. Thus, all non-squamous cells in urine are considered to be of transitional origin. When we are unsure about the origin of the cells in urine, we can stain a urine sediment with Wright's stain (or Diff-quik) and perform a cytological examination on the urine sediment. Epithelial cells are subjectively semi-quantified in urine (usually under low power using the 10x objective) as: none seen, few, moderate, many.

116 | Page PHCPI Laboratory Standards Guideline

Transitional epithelial cells The urinary tract from the pelvis down the ureters to the bladder and the proximal urethra is lined by transitional epithelial cells. These cells vary in size and shape depending on the location from which they originate, e.g. those from the renal pelvis are more caudate whereas those from the bladder are more round to polygonal and vary in size. These cells naturally slough into the urine in quite low numbers, so none to a few transitional epithelial cells are seen in the urine from healthy animals. Note that this depends on the method of urine collection, since these cells will be sloughed (traumatically) when the bladder is catheterized. Transitional epithelial cells must be distinguished from WBC, because they both have the same granular appearance. In general, transitional epithelial cells (arrow in above image) are larger and have more irregular borders than WBC (which are uniformly more round, arrowhead in above image).

Squamous epithelial cells These can be keratinized or non-keratinized.Non-keratinized squamous epithelial cells originate from the distal urethra, prepuce and/or vagina. They are larger than transitional cells and have small central nuclei. They can be round or have one or more flat border. Keratinized squamous epithelial cells are from the skin or vulva and are large cells with angular borders. They may or may not have nuclei (see upper image to the right). Nuclei are more visible in cells when the urine is stained with Sedi-stain (see central panel on right). If in doubt about the origin of the cells, a Wright's stain (routine hematologic stain) can be performed on a urine sediment and demonstrates the central nuclei and angular borders of squamous epithelial cells (lower panel on right).Squamous cells are frequently seen as contaminants in voided urine samples and can also contaminate samples collected by catheterization . Renal tubular epithelial cells These are rarely seen in the urine and, as mentioned above, are very difficult to distinguish from transitional epithelial cells. If large numbers of smaller epithelial cells of uniform appearance (size and shape) are observed in the urine, a renal origin for these cells is suspected. Transitional epithelial cells tend to be more variable in size and shape (to some extent). Sloughing of large numbers of renal tubular epithelial cells would indicate renal tubular injury

117 | Page PHCPI Laboratory Standards Guideline

RBCs WBCs

Squamous epithelial cells Transitional epithelial cells

Neoplasia Fig17: common urine cells (as seen with the X40 objective)

Casts

Hyaline casts: These can be quite difficult to see in wet preparations of urine sediments with light microscopy, even with the condenser of the microscope racked down. They are much easier to visualize using phase contrast; however phase is usually not available on most

118 | Page PHCPI Laboratory Standards Guideline microscopes. They become more visible with regular light microscopy if fat sticks to the protein matrix (Tamm-Horsfall mucoprotein) that makes up the hyaline cast (hyaline with fat) or particulate material from degenerating cells is present within the cast matrix (hyaline to finely granular cast). Granular casts: are rather short casts filled with large granules, pale yellow in color, with rounded ends. The granules come from degenerated epithelial cells from the tubules of the kidney.  Fine granular casts have smaller granules that do not fill the cast .Do not confuse with hyaline casts, partly covered by amorphous phosphate crystals  Blood casts are filled with more or less degenerated erythrocytes, brownish in color. They are found in acute kidney disease.  Pus casts are completely filled with leukocytes  Hyaline casts, which may contain a few leukocytes suffering from kidney infection.  Epithelial casts are filled with pale yellow epithelial cells (To make the cells more distinct, add a drop of 10% acetic acid to the deposit.)  Epithelial casts have no diagnostic significance.  Fatty casts are very shiny yellowish casts; the edges are indented and distinct andthe ends are rounded They are soluble in ether but not in acetic acid Fatty casts are found in patients with severe kidney disease.  Pseudo castsDo not mistake for casts: 1- clumps of phosphate crystals, short and clear-cut 2- aggregations of translucent mucus, the ends tapering into threads

Fig18: different typeof casts seen in urine (as seen with the X40 objective)

119 | Page PHCPI Laboratory Standards Guideline

Crystals: Crystals have regular geometric shapes unlike amorphous debris, which is madeup of clumps of small granules with no definite shape except in very rarediseases, crystals in urine have no diagnostic significance. 1- Calcium oxalate (acid urine)  Size: 10–20µ or about 50µ  Shape: envelope-shaped or peanut-shaped  Color: colorless, very shiny. 2- Uric acid (acid urine)  Size: 30–150µ.  Shape: varies (square, diamond-shaped, cubical or rose-shaped).  Color: yellow or brownish-red. 3- Triple phosphates (neutral or alkaline urine)  Size: 30–150µ.  Shape: rectangular or like a fern leaf or star  Color: colorless, shiny 4- Urate crystals  Urates (alkaline urine) (Size: about 20µ.  Shape: like a cactus or a bundle of needles  Colour: yellow, shiny.  Urates are often found together with phosphates.  Calcium phosphate (neutral or alkaline urine) Size: 30–40µ.  Shape: like a star. Colour: colourless. 5- Calcium carbonate (neutral or alkaline urine)  Size: very small.  Shape: similar to millet or corn grains, grouped in pairs.  Colour: colourless.  If acetic acid, 10% solution is added, the crystals dissolve, giving off  Bubbles of gas. 6- Calcium sulfate (acid urine)  Size: 50–100µ.  Shape: long prisms or flat blades, separate or in bundles.  Calcium sulfate crystals can be distinguished from calcium phosphate crystals bymeasuring the pH of the urine.

120 | Page PHCPI Laboratory Standards Guideline

7- Amorphous phosphates (alkaline urine)  Amorphous phosphates appear as small, whitish granules, often scattered. 8- Amorphous urates (acid urine) Amorphous urates appear as very small, yellowish granules, which are grouped in compact clusters.They are not soluble in acetic acid, 10% solution, but dissolve if the urine is gently heated (Urine kept in the refrigerator often shows a heavy precipitate of urates.) Other crystalline deposits The following are rarely found in the urine. When present, however, they are found 9- Cystine (acid urine)  Size: 30–60µ.  Shape: hexagonal plates.  Color: colorless, very shiny. Cystine crystals are found only in fresh urine as they are soluble in ammonia. &they are soluble in acetic acid, 10% solution (one drop per drop of deposit).They are found in patients with cystin uria, a very rare hereditary disease. 10- Acetyl sulfonamides (neutral or acid urine)  Shape: varied, but often similar to sheaves of needles.  Acetyl sulfonamide crystals are found in the urine following treatment withsulfonamide drugs. The presence of these crystals should be reported as they cancause kidney damage.

Ammonium biurate Amorphous urate Bilirubin

121 | Page PHCPI Laboratory Standards Guideline

Calcium carbonate Calcium oxalate Magnesium ammonium dihydrate phosphate

Uric acid Calcium oxalate Calcium phosphate monohydrate

Cystine Tyrosine Flat plates resembling cholesterol Fig19: The most common crystals which may be found in urine (as seen with the X40 objective)

Fungi  Size: 5–12µ.  Shape: round or oval bodies of various sizes found together. Do not confuse witherythrocytes. Budding may be seen. Fungi are not soluble in acetic acid.  Fungi are occasionally present in urine containing glucose. Check that the urinespecimen is fresh. Bacteria In healthy persons the urine contains practically no organisms. Bacteria may befound in patients who have an infection of some part of the urinary tract. Urethritis, cystitis or nephritis), or where bacteria from an infection elsewhere in the body are excreted in the urine

122 | Page PHCPI Laboratory Standards Guideline

The urine is centrifuged at high speed and the resulting deposit is examined under the microscope (as described in section this is the most important part of the analysis). However, the deposit may also be used to make smears that are stained.

Parasite eggs and larvae The following may be found:  eggs of Schistosoma haematobium: found together with erythrocytes  Trichomonas vaginalis

Schistosoma haematobium; The two methods used for detection of ova of Schistosoma haematobium are sedimentation and filtration. The sedimentation method is less sensitive but is cheaper and simpler to perform. The filtration technique is used when quantitative information is required for epidemiological surveillance purposes.

Materials and reagents _ Microscope _ Microscope slides _ Coverslips _ Centrifuge _ Conical centrifuge tubes, 15 ml _ Filter holder, 13 or 25um diameter _ Membrane filter, 12–20um pore size (or equivalent) filter-paper _ Conical flask for urine collection _ Pasteur pipettes _ Plastic syringe, 10 ml _ Lugol iodine, 0.5% solution _ Formaldehyde, 37% solution.

Detection Method

Collection of urine specimens The number of ova in the urine varies throughout the day; it is highest in urine obtained between 10:00 and 14:00. The specimen should therefore be collected between these times and should consist of a single terminal urine specimen of at least 10 ml. alternatively, a 24- hour collection of terminal urine can be made

123 | Page PHCPI Laboratory Standards Guideline

The whole specimen must be examined, as it may contain only a few ova. Ask the patient to collect the urine in a clean flask or bottle. Examine the specimen at once. If the urine cannot be examined for an hour or longer, add 1 ml of undiluted formalin (37% formaldehyde solution) to each 100 ml of urine. This will preserve any eggs that might be present Note: If formalin is not available, 2 ml of ordinary household bleach can be added to each 100 ml of urine. A: Sedimentation method 1. Shake the urine specimen well and pour into the conical flask. 2. Allow the urine to sediment for 1 hour. Remove the supernatant and transfer the sediment into a centrifuge tube. Centrifuge at 2500rpmfor 2 minutes 3. Examine the deposit under the microscope for the presence of ova.

Do not increase the centrifugation time and do not exceed 2500rpmas this may disrupt the ova and release miracidia. Important: — process the specimen as soon as possible; — shake the container before pouring the urine specimen into the conical flask; — label slides and tubes carefully. B: Filtration method 1. Place a filter in the filter holder. 2. Agitate the urine sample gently and draw 10ml into the syringe Attach the syringe to the filter holder. 3. Expel the urine from the syringe through the filter over a bucket or sink 5. Disconnect the syringe from the filter holder. Using forceps, carefully remove the membrane filter or filter-paper and place it on a microscope slide. The nylon membrane and filter-paper should be placed face-up, while the polycarbonate membrane should be placed face-down. 6. Add one drop of Lugol iodine solution to improve the visibility of the eggs. 7. Examine the entire filter under the microscope at x 10 or x 40. Record the results as the number of eggs per 10 ml of urine. Microscopic examination; The eggs of Schistosoma haematobium are large, about 120–150mm long, and have a terminal spine at one end, an embryo (the miracidium) can be seen inside the egg. Sometimes it is necessary to determine whether the eggs are viable. This can be done if the specimen is fresh and no preservatives have been added. Look carefully at the eggs to see if

124 | Page PHCPI Laboratory Standards Guideline the embryos are moving. This is the best indication of viability. If no movement is seen, look for the “flame cells” There are four flame cells, one at each corner of the embryo. Use a ¥100 objective with slightly reduced illumination to look for the rapid movement of cilia (short hairs) in the flame cells. Reporting the results When the syringe filtration technique is used, the results may be reported according to egg count categories:

_ Light infection: 1–49 eggs per 10 ml of urine.

_ Heavy infection: > 50 eggs per 10 ml of urine. A third category, such as >500 eggs per 10 ml of urine, or >1000 eggs per 10 ml of urine, may be appropriate in areas where the intensity of infection frequently reaches this level (i.e. in more than 10% of cases).

Bacteria: Bacilli Bacteria: Cocci

Schistosoma haematobium( egg) Trichomonas vaginalis (trophozoites)

125 | Page PHCPI Laboratory Standards Guideline

Fig20: The most common Infectious agentswhich may be found in urine (as seen with the X40 objective)

Fat droplets: Uniform and red-tinged versus fat droplets are more variable in shape, slightly greenish- tinged (or refractile), have a darker edge, are more globular shape (this can be visualized when you focus up and down) and usually float to the top of the coverslip

6.2 General Stool Examination

Introduction

Many diseases are caused by infection with parasites. They are an important causeof diarrhea (which is a major health problem in developing countries. If acute diarrhea is caused by parasitic infection, this can be determined bymicroscopical examination of a stool specimen.

Clinical Significance

The Inspection of feces is important because it may lead to a diagnosis of parasitic infestation, obstructive jaundice, diarrhea, malabsorption, rectosigmoidal, obstruction, dysentery, ulcerative colitis, or bleeding.

Materials and reagents:

 Microscope  Microscope slides  Coverslips  Wooden applicators or wire loops (0.45mm, nickel–chromium alloy wire)  Grease pencils  Sodium chloride, 0.85% solution  Lugol iodine, 0.5% solution

Examination of stool specimens for parasites

Collection of specimens for GSE

126 | Page PHCPI Laboratory Standards Guideline

Inspect approximately 100 g of faeces in a clean, dry container without preservatives. A screw-top wide mouth container is most suitable. Make sure that anyadult worms or segments passed are included. For collection of stool specimens for bacteriological examination (e.g. for culture of cholera and entero-pathogenic bacteria). The container with specimen should be labeled, the urine and dirt should be excluded, urine will destroy any amoebic trophozoites and dirt will interfere with diagnosis  never leave the stool specimen exposed to the air in container without lead  never keep the stool specimen artificially warm and don’t leave in the sun

Using Cary–Blair transport medium for bacteriology

Cary–Blair transport medium will preserve many kinds of entericbacteria (cholera vibrios, other vibrios, salmonella, shigella, etc.) for up to 4 weeks. The uninoculated medium may be stored in a sealed bottle at room temperaturefor 8–12 weeks. 1. Dip a sterile cotton wool swab in the stool specimen 2. For infants or other patients who cannot produce a stool specimen, take a rectalswab. Moisten the swab with sodium chloride solution and introduce the swabinto the . Turn the swab several times with a circular movement 3. Place the swab in a bottle containing Cary–Blair medium (three-quarters full) and send it to the bacteriology laboratory. If you cannot send the swab immediately, and store it at room temperature. Important:  Never store the swab in the incubator.  Never store the swab in the refrigerator.

6.2.1 Macroscopical examination

Faecal samples are best described by their colour, consistency and presence orabsence of macroscopic blood, exudate and adult parasite. 1-Colour The color can be described as:  Black (bleeding from upper gastrointestinal tract is more likely to cause stool black)  Brown

127 | Page PHCPI Laboratory Standards Guideline

 pale yellow (fat)  white (obstructive jaundice)  green  bloody (especially blood originating from lower gut may cause the stool to be red )  clay (suggested diminution or absence of bile or presence of barium sulfate ) 2-Consistency The consistency can be described as:  formed (normal shape)  soft formed  Unformed or liquid (watery). The presence of blood or mucus, usually seen as streaks of red or white, should be noted. Blood may be present in certain medical conditions (e.g. ulcerative colitis, schistosomiasis). 3- Pus

Patient with chronic ulcerative colitis and chronic bacillary dysentery frequently pass large quantity of pus with the stool  Large amount of pus never combine with amoebic colitis.  No pus seen in watery stool as viral gastroenteritis.

4- Mucus  When the mucus is seen even in slightest quantity, it should be considered as abnormal.  Bloody mucus indicates inflammation of rectal canalAssociated with pus and blood is found in stool of patient with ulcerative colitis, bacillary dysentery, intestinal tuberculosis  Mucus with pus and blood indicates ulcerative colitis, bacillary dysentery, Ulcerative carcinoma of colon. Intestinal tuberculosis.

5 –Blood The presences of Blood in stool should not be ignored  Bleeding of upper GITmay give black- tarry appearance to stool.  Bleeding from lower GIT may givered colorstool.

Fecal occult blood testing:

Introduction

128 | Page PHCPI Laboratory Standards Guideline

Bleeding is a common clinical problem. Blood loss ranges from occult bleeding of which the patient is unaware to massive bleeding that anyone would notice (melena or bloody stool. Positive occult blood test is usually due to chronic GI blood loss, both upper and lower GI lesion such as peptic ulcer, carcinoma, and colon carcinoma. It is a screening test for digestive carcinoma especially for elderly people.

Blood in the stool may be the only symptom of colorectal cancer, but not all blood in the stool is caused by cancer. Other conditions that can cause blood in the stool include:

 Hemorrhoids. These are enlarged, swollen veins in the anus. Hemorrhoids can be formed inside the anus (internal hemorrhoids) or outside of the (external hemorrhoids).  Anal fissures: these are thin tears in the tissue that lines the anus (anal sphincter) up into the .  Colon polyps: these are growths of the tissue attached to the colon and often look like a stem or stalk with a round top.  Peptic ulcers: these sores are formed when the digestive juices made in the stomach eat away at the lining of the digestive tract.  Ulcerative colitis: this type of inflammatory bowel disease (IBD) causes inflammation and sores (ulcers) in the inner lining of the colon and rectum.  Gastroesophageal reflux disease (GERD): This is the abnormal backflow (reflux) of food, stomach acid, and other digestive juices into the .  Crohn”s disease: this type of inflammatory bowel disease causes inflammation and ulcers that may affect the deep layers of the lining of the digestive tract.

Hemoccult slide test for occult blood:

Procedure:

A very small stool specimen is collected on the tip of wooden applicator. A thin smear of specimen is applied inside the circle. The cover is closed and application

129 | Page PHCPI Laboratory Standards Guideline

is to be discarded. The specimen dry completely. Next the perforated window in the back of the slide is opened.

Two or three drops of developing solution are applied to the slide opposite the specimen. The results can be read after 30 seconds.

Results: Reaction produced blue colour= positive No detectable blue colour= negative Note: - Subject should avoid ingestion red meat for three days before and during the test. - Use of vitamin C, iron tablets should be avoided. - Slide should not be dehydrated.

6-Fecal pH test

A fecal pH test is one where a specimen of feces is tested for acidity in order to diagnose a medical condition, is normally alkaline. An acidic stool can indicate a digestive problem such as or a contagion such as E.coli or Rotavirus. Test procedure: The test is fast and can be performed. A patient must not be receiving antibiotics. At least half a mg of feces is collected and a strip of nitrazine paper is dipped in the sample and compared against a color scale. A pH of less than 5.5 indicates an acidic sample.

7-Fecal fat: It is raised in malabsorption.

6.2.2 Microscopic Examination Direct microscopic examination of faeces in saline or iodine suspension is usefulfor the following reasons:

 to detect motile trophozoites  to detect ova and cysts present in moderate numbers  to detect erythrocytes, cellular debris or excess fat

130 | Page PHCPI Laboratory Standards Guideline

Select unformed or liquid faeces when using direct microscopy for detection oftrophozoites. Formed stools rarely contain motile trophozoites. Also perform adirect examination of any external blood or mucus.

Method 1. Prepare a 1:1 mixture of 0.5% Lugol iodine solution. 2. Take a dry microscope slide and label it with the name or number of thepatient. 3. Put: one drop of sodium chloride solution warmed to 37°C in the middle ofthe left half of the slide; and one drop of the iodine–acetic acid solution in the middle of the right halfof the slide 4. Using an applicator or wire loop, take a small portion (about 2–3mmdiameter) of the stool. a) If the stools are formed, take the portion from the centre of the sampleand from the surface to look for parasite eggs. b) If the stools contain mucus or are liquid, take the portion from the mucuson the surface or from the surface of the liquid to look for amoebae.

5. Mix the sample with the drop of 0.9% sodium chloride solution on the slide. 6. Using the applicator or wire loop, take a second portion of stool from the specimenas described above and mix it with the drop of Lugol iodine, discard the applicator (or flame the wire loop) after use. 7. Place a coverslip over each drop (apply the coverslips as shown in toavoid the formation of air bubbles). 8. Examine the preparations under the microscope. For the saline preparationuse the x10thenx40 objectives and ax10 eyepiece. As the eggs and cysts arecolorless, reduce the amount of light using the condenser aperture or lowerthe condenser to increase the contrast.Examine the first preparation with the x10 objective, starting at the top left handcorner as indicated in Focus on the edge of one coverslip using the x10 objective and examine the whole area under each coverslip for the presence of ova and larvae of parasite. Then switch to the x40 objective and again examine the whole area of the coverslip over the saline for motile trophozoites and the area of the coverslip over the iodine for cysts. 9. Lugol iodine–acetic acid solution causes the trophozoite forms to become nonmotile.The nucleus is clearly stained but it may be difficult to distinguish betweentrophozoite and cystic forms. 10. Using a fine Pasteur pipette, allow a drop of 5% methylene blue solution to run under the coverslip over the saline preparation This will stain thenuclei of any cells present

131 | Page PHCPI Laboratory Standards Guideline

and distinguish the lobed nuclei of polymorphs fromthe large single nuclei of mucosal cells. 11. If a drop of eosin solution is added, the whole field becomes stained except for the protozoa (particularly amoebae), which remain colourless and are thus easily recognized.

Intestinal protozoa Protozoa are microorganisms consisting of a single cell. Intestinal protozoa may be found in stools in their motile form (trophozoites) or as cysts. Some intestinal protozoa are pathogenic, others are harmless. All these protozoa arefound throughout the world.

Identification of motile forms (trophozoites) The trophozoites of protozoa are motile  either because of slow movements of the cell (amoebae)  or because they have rapidly moving flagella (long whip-like threads) or cilia(numerous short hairs). Trophozoites are chiefly found in:  watery stools  stools containing mucus  Soft formed stools.

The following features are useful for the identification ofmotile forms of intestinal protozoa  size  cytoplasm  pseudopodia  nuclei  Ectoplasm  Endoplasm  Vacuoles  Inclusion bodies containing erythrocytes, bacteria, yeast cells, debris, etc.  Nuclear membrane (chromatin)  Nuclear karyosome  Flagella  Undulating membrane.

Identification of cysts Cysts are the resistant forms of certain intestinal amoebae, flagellates and ciliates. They are small, round and non-motile and may have one or several nuclei.

132 | Page PHCPI Laboratory Standards Guideline

Measurement of cysts is useful for the correct identification of species. Importance of cysts: the clinical importance of cysts varies from country to country. The cyst is theinfective form of the organism. Healthy persons may be asymptomatic carriers ofcysts and are therefore, a public health hazard. The most important problem in the laboratory is the precise identification of cysts

Taenia saginata Hymenolepis nana Diphyllobothrium latum

Ascaris lumbricoides Ancylostoma duodenale Trichuris trichuria

Strongyloides stercoralis Clonorchis sinensis Paragonimus westermani

133 | Page PHCPI Laboratory Standards Guideline

Schistosome mansoni ova Schistosome japonicum

Figure21: Ova of Enteric parasites

Entamoeba histolytica cyst Entamoeba histolytica trophozoites

Giardia lamblia cyst Giardia lamblia trophozoites

Figure 22: Entamoeba histolytica&Giardia lamblia (HPF)

6.3Urine Culture

Introduction Urine is the specimen most frequently submitted for culture. It also presents major problems in terms of proper specimen collection, transport, culture techniques, and interpretation of results. As with any other specimen submitted to the laboratory, the more information provided by the submitting physician the more able is the laboratory to provide the best possible culture data.

134 | Page PHCPI Laboratory Standards Guideline

The most common sites of urinary tract infection (UTI) are the urinary bladder (cystitis) and the urethra. From these sites the infection may ascend into the ureters (ureteritis) and subsequently involve the kidney (pyelonephritis). Females are more prone to infection of the urinary tract and also present the greater problem in the proper collection of specimens.

In both males and females, UTI may be asymptomatic, acute, or chronic. Asymptomatic infection can be diagnosed by culture. Acute UTI is more frequently seen in females of all ages; these patients are usually treated on an outpatient basis and are rarely admitted to hospital. Chronic UTI in both males and females of all ages is usually associated with an underlying disease (e.g., pyelonephritis, prostatic disease, or congenital anomaly of the genitourinary tract) and these patients are most often hospitalized. Asymptomatic, acute, and chronic UTI are three distinct entities and the laboratory results often require different interpretation.

Asymptomatic pyelonephritis in females may go undetected for some time, and is often only diagnosed by carefully performed quantitative urine culture. Chronic prostatitis is common and difficult to cure, and is often responsible for recurring UTI. In most UTI, irrespective of type, enteric bacteria are the etiological agents, being isolated far more frequently than any other organism. In about 10% of patients with UTI, two organisms may be present and both may contribute to the disease process. The presence of three or more different organisms in a urine culture is strong presumptive evidence of improper collection or handling of the urine specimen. However, multiple organisms are often seen in UTI in patients with indwelling bladder catheters

Expected pathogens - Enterococci - Escherichia coli - Mycobacterium tuberculosis - Other Enterobacteriaceae - Other staphylococci - Pseudomonas and other non-fermenters - Staphylococcus saprophyticus

Media and diagnostic reagents - Isolation and quantitative media

Blood agar CLED agar or chromogenic media MacConkey agar, with crystal violet, )

135 | Page PHCPI Laboratory Standards Guideline

- Identification media and diagnostic reagents Kligler iron agar (KIA) Kovacs reagent for indole test motility–indole–urease (MIU) medium oxidase test reagent lysine–decarboxylase broth (Möller) ONPG test reagent Simmons citrate agar For staphylococci and enterococci catalase test reagent (Hydrogen peroxide) coagulase plasma bile–aesculin agar (for enterococci) novobiocin (5mg) disc for differentiating negative-coagulase staphylococci

Specimen collection It cannot be overemphasized that considerable importance is attached to the method of collection of urine specimens, transport to the laboratory, and the initial efforts by the laboratory to screen and culture the urine. It is the responsibility of the laboratory to provide the physician with sterile, wide-mouthed, glass or plastic jars, beakers, or other suitable receptacles. Urine specimens may have been collected by a surgical procedure, e.g., suprapubic aspiration, cystoscopy, or catheterization. If not, the laboratory must insist on a clean-catch midstream urine specimen, particularly in females and children. Since urine itself is a good culture medium, all specimens should be processed by the laboratory within 2 hours of collection, or be kept refrigerated at 4 °C until delivery to the laboratory and processed no longer than 18 hours after collection.

Whenever possible, urine specimens for culture should be collected in the morning. It is advisable to ask the patient the night before to refrain from urinating until the specimen is collected.

Women A woman who is ambulatory should: 1. Wash her hands thoroughly with soap and water and dry them with a clean towel. 2. Undress in a suitable room, spread the labia, and cleanse the vulva and labia thoroughly, using sterile cotton gauze pads and warm soapy water wiping from front to rear.

136 | Page PHCPI Laboratory Standards Guideline

3. Rinse thoroughly with warm water and dry with a sterile gauze pad. During the entire process the patient should keep the labia separated and not touch the cleansed area with the fingers. 4. Pass urine, discarding the first part of the stream. Collect the remaining urine in the sterile container, closing the lid as soon as the urine has been collected. 5. Hand the clean midstream urine, in the closed container, to the nursing personnel for prompt delivery to the laboratory. For bedridden patients, the same procedure is followed, except that a nurse must assist the patient or, if necessary, do the entire cleansing procedure before requesting the patient to pass urine. In both situations every effort must be made to collect a clean-catch urine specimen in a sterile container and to ensure that it is delivered promptly to the laboratory together with information on the patient, clinical diagnosis, and the requested procedures.

Men A man who is ambulatory should: 1. Wash his hands. 2. Pull back the foreskin (if not circumcised) and pass a small amount of urine. 3. Still holding back the foreskin, pass most of the remaining urine into a sterile container. This is a midstream urine specimen. 4. Place the cover on the container and hand to nursing staff for prompt delivery to the laboratory. For bedridden patients: 1. If necessary, nursing personnel should pull back the foreskin, wash and dry the glans with soapy water and gauze pads. 2. With foreskin pulled back, the patient should pass a small amount of urine into a urinal. 3. The patient should then pass most of the remaining urine into the sterile container. The cover should be placed on the container and the specimen transported to the laboratory. Infants and children Collection of a clean-catch urine specimen from infants and children who are ill in bed or uncooperative can be a problem. Give the child water or other liquid to drink. Clean the external genitalia. The child can be seated on the lap of the mother, nurse, or ward attendant, who should then encourage the child to urinate and collect as much urine as possible in a sterile container. The container should then be covered and delivered to the laboratory for prompt processing.

137 | Page PHCPI Laboratory Standards Guideline

Culture and interpretation All urine specimens brought to the microbiology laboratory should be examined at once, or placed in a refrigerator at 4 °C until they can be examined. The examination procedure includes the following steps: 1. Examination of a Gram-stained smear.

2. A screening test for significant bacteriuria.

3. A definitive culture for urine specimens found to be positive in the screening test (step 2), and for all specimens obtained by cystoscopy, Suprapubic aspirate (SPA), or catheterization.

4. Susceptibility tests on clinically significant bacterial isolates.

Preparation and examination of a Gram-stained smear is a necessary part of the laboratory process. Using a sterile Pasteur pipette (one for each sample), place one drop of well mixed, uncentrifuged urine on a slide. Allow the drop to dry without spreading, heat-fix and stain. Examine under an oil-immersion lens for the presence or absence of bacteria, polymorphonuclear leukocytes, and squamous epithelial cells.

One or more bacterial cells per oil-immersion field usually implies that there are 105 or more bacteria per milliliter in the specimen. The presence of one or more leukocytes per oil- immersion field is a further indication of UTI. Non-infected urine samples will usually show few or no bacteria or leukocytes in the entire preparation. In specimens from females, the presence of many squamous epithelial cells, with or without a mixture of bacteria, is strong presumptive evidence that the specimen is contaminated with vaginal flora and a repeat specimen is necessary, regardless of the number of bacteria per oil-immersion field. If results are required urgently, the report of the Gram-stain findings should be sent to the physician with a note that the culture report is to follow.

Screening method The absence of leukocytes and bacteria in a Gram-stained smear of a clean-catch urine sample prepared as described above is good evidence that the urine is not infected. A urine specimen that is “negative” on careful examination of the Gram-stained smear does not need to be cultured. An alternative simple and effective screening test is the test strip for leukocyte esterase/nitrate reduction. The strip is dipped into the urine specimen as instructed in the package literature. Any pink colour is a positive reaction indicating the presence of leukocyte esterase and/or bacteria in excess of 105 per ml. Urine samples that are positive in the screening test should be cultured as soon as possible to prevent possible overgrowth by nonsignificant bacteria. If the strip does not develop a pink colour it is interpreted as a

138 | Page PHCPI Laboratory Standards Guideline negative screening test, is so reported, and no culture is indicated. The test strip may not be sensitive enough to detect bacterial counts of less than 105 per ml of urine.

Quantitative culture and presumptive identification

Calibrated loop technique There are several culture methods for the quantification of bacteria in urine. The easiest and most commonly used are the calibrated loop technique. Of these method using CLED or chromogenic media, are considered to be the most versatile and efficient for large numbers of specimens. Chromogenic media contain various substrates which permit presumptive identification of several common species through a change in either colony pigmentation or colour of agar. They perform satisfactorily compared to CLED and have the advantage that mixed cultures are easier to detect. However, chromogenic media from different manufacturers can vary in specificity, and are relatively expensiveCalibrated loop/surface streak method Mix the urine gently to avoid foaming. Dip the end of a sterile calibrated loop (eg 1 μL, 2 μL or 10 μL) in the urine to just below the surface and remove vertically, taking care not to carry over any on the shank. Use this to inoculate CLED or chromogenic and spread according to the number of specimens. A maximum offour samples per 9 cm plate is recommended for this method with a 1 μL or 2 μL loop, or two samples if using a 10 μL loop. If a 1 μL loop is used, one colony equals 1000 cfu/mL (ie 1 x 106 cfu/L). SPA, other surgically obtained urine, and urine samples with expected significant bacteriuria as low as 105 cfu/L (increased inoculum sizes are required), Inoculate 100 μL (0.1 mL) of specimen aseptically to a full CLED or chromogenic agar plate. Spread inoculum over entire surface of plate with a sterile loop or a spreader. Do not use a sterile swab which will absorb much of the inoculum. To isolate individual colonies, spread inoculum with a sterile loop.

No. of cfu/L = No. of cfu on plate x 104.

This semi quantitative method is only sensitive for screening down to106 cfu/L if a 5 μL or

10 μL loop is used (eg 5 or 10 colonies), or107 cfu/L if a 1 μL or 2 μL loop is used (eg 10 or 20 colonies). (See table below). Guidance on assessing colony counts:

139 | Page PHCPI Laboratory Standards Guideline

Interpretation of culture The basis studies for interpreting urine culture results showing that bacterial counts of ≥108 cfu/L (≥105 cfu/mL) are indicative of an infection and counts below this usually indicate contamination.

In specific patient groups, counts between 108 cfu/L (105 cfu/mL) and 105 cfu/L (102 cfu/mL) may be significant. A pure isolate with counts between 107 and 108 cfu/L (104-105 cfu/mL) should be evaluated based on clinical information or confirmed by repeat culture. Overall the confirmation of a UTI requires the demonstration of significant bacteriuria by quantitative culture (defined according to patient group or specimen type). Routine culture methods may not be sensitive enough to detect low bacteria levels (eg ≥107 cfu/L / ≥104 cfu/mL) and increased sensitivity will be achieved by increasing the inoculum size.

The following is a list of imperial units and their equivalents.

≥105 cfu/mL equivalent to ≥108 cfu/L

<105 cfu/mL equivalent to <108 cfu/L

104 cfu/mL equivalent to107 cfu/L

<104 cfu/mL equivalent to <107 cfu/L

103 cfu/mL equivalent to 106 cfu/L

<103 cfu/mL equivalent to <106 cfu/L

102 cfu/mL equivalent to 105 cfu/L

140 | Page PHCPI Laboratory Standards Guideline

Increased inoculum sizes are also required for persistently symptomatic patients without bacteriuria if the patient has recurrent ‘‘sterile pyuria’’, or for specimens where lower counts are to be expected, such as SPA or other surgically obtained urine.

Identification Identification should be performed as rapidly as possible. Isolation of uropathogens should be performed by a surface streak procedure on both blood and MacConkey agar using calibrated loops for semi-quantitative method and incubated aerobically at 37 °C for 24 hours, and those cultures which becomes negative at the end of 24 hrs incubations will be further incubated for 48 hours. Bacterial identification should be made using biochemical tests, namely indole, citrate, oxidase, H2S production, lysine decarboxylase, lactose fermentation, urea hydrolysis, gas production, catalase, coagulase, manitol fermentation and novobiocin susceptibility test (See Flowcharts below).

141 | Page PHCPI Laboratory Standards Guideline

Fig23: Identification of Streptococci and Enterococci Flowchart

142 | Page PHCPI Laboratory Standards Guideline

Fig 24: Presumptive Identification of Staphylococcus Species Flowchart

143 | Page PHCPI Laboratory Standards Guideline

Fig 25: Identification of Enterobacteriaceae Flowchart

144 | Page PHCPI Laboratory Standards Guideline

Fig 26: Identification of Glucose Non-Fermenting Gram negative Rods Flowchart

Susceptibility tests Susceptibility tests should only be performed on well isolated colonies of similar appearance that are considered significant according to the guidelines presented above. Susceptibility

145 | Page PHCPI Laboratory Standards Guideline tests are generally more important on cultures obtained from patients who are hospitalized or have a history of recurring UTI. Cultures from patients seen in the clinic with a primary UTI may not require a susceptibility test.

6.4Throat Swab The upper respiratory tract canbe the site of several types of infection:

 pharyngitis, sometimes involving tonsillitis, and giving rise to a “sorethroat”  nasopharyngitis  otitis media  Sinusitis  Epiglottitis.

Of all those infections, pharyngitis is by far the most frequent; in addition, theuntreated infection may have serious sequelae. Only pharyngitis will be consideredhere.

Bacteriological diagnosis of pharyngitis is complicated by the fact that theoropharynx contains a heavy, mixed, normal flora of aerobic and anaerobicbacteria. The normal flora generally outnumbers the pathogens and the roleof the bacteriologist is to distinguish between the commensals and thepathogens. Where possible only the latter should be reported to the physician.

Normal flora of the pharynx The normal flora of the pharynx includes a large number of species thatshould be neither fully identified nor reported when observed in throatcultures: • Viridans (α-haemolytic) streptococci and pneumococci • Nonpathogenic Neisseria spp. • Moraxella (formerly Branhamella) catarrhalis (this can also be a respiratorypathogen) • Staphylococci (S. aureus, S. epidermidis) • Diphtheroids (with the exception of C. diphtheriae) • Haemophilus spp. • Yeasts (Candida spp.) in limited quantity • Various strictly anaerobic Gram-positive cocci and Gram-negative rods, spirochaetes and filamentous forms. The throats of elderly, immunodeficient, or malnourished patients, particularlywhen they have received antibiotics, may be colonized by Enterobacteriaceae (Escherichia coli, Klebsiella spp., etc.) and by the nonfermentativeGram-negative groups (Acinetobacter spp. and Pseudomonas spp.). Suchpatients may also have in their pharynx a proliferation of S. aureus or of

146 | Page PHCPI Laboratory Standards Guideline

Candida spp., or other yeast-like fungi. Although these microorganisms do notcause pharyngitis, except in association with granulocytopenia, it is advisableto report such isolates to the clinician, as they occasionally indicate the existenceof (or may sometimes give rise to) a lower respiratory tract infection (e.g. pneumonia) or bacteraemia. However, an antibiogram should not be performedroutinely on these colonizing microorganisms.

Expected pathogens  Candida albicans (oropharynx)  Corynebacterium diphtheriae (throat and nose)  Haemophilus influenzae (ear and sinus)  Moraxella catarrhalis (ear and sinus)  Neisseria meningitides  Pseudomonas  Staphyloccoccus aureus (ear and sinus)  Streptococcus pneumoniae (ear and sinus)  Streptococcus pyogenes (group A, throat)

Media and diagnostic reagents Isolation media Isolation media

Blood agar (prepared from a glucose-free base) Chocolate agar Löffler coagulated serum or Dorset egg medium Tellurite blood agar Modified Thayer–Martin medium (for gonococci andmeningococci)

Diagnostic reagents

Diagnostic reagents

Bacitracin disc Catalase and coagulase tests reagents Optochin disc Carbohydrate degradation media for Neisseria spp.

147 | Page PHCPI Laboratory Standards Guideline

Oxidase test reagent V and XV factors (discs or strips) Tributyrin

Rapid diagnostic tests

Grouping kit for haemolytic streptococci

Bacterial agents of pharyngitis

Streptococcus pyogenes (Lancefield group A) is by far the most frequent causeof bacterial pharyngitis and tonsillitis. This infection is particularly prevalentin young children (5–12 years).

Non-group-A, b-haemolytic streptococci (e.g. groups B, C and G) are uncommoncauses of bacterial pharyngitis and if detected should be reported.

Corynebacterium diphtheriae is the cause of diphtheria, a disease that is endemicin many countries. It can reach epidemic proportions in countries where thevaccination programme has been interrupted. Characteristically (with a fewexceptions), C. diphtheriae causes a typical form of infection, characterized bya greyish-white membrane at the site of infection (pharynx, tonsils, nose, or larynx). Diphtheria is a serious disease and the diagnosis is made on the basisof clinical findings. The physician would then generally make a specificrequest to culture for diphtheria bacilli.

(Vincent angina) is a rare condition characterizedby a necrotic ulceration of the pharynx with or without formationof a pseudomembrane. It is associated, at the site of infection, with a heavymixed flora of strict anaerobes dominated by Gram negative fusiform rodsand spirochaetes, generally referred to as Fusobacterium spp. and Treponemavincentii, and possibly others. Although both species belong to the normalmouth flora, their presence in large numbers in a Gram-stained smear of ulceratedlesions should be reported as a “fusospirochaetal complex”. (See figure 27) This microscopicdiagnosis need not be confirmed by anaerobic culture, which is difficultand time-consuming.

148 | Page PHCPI Laboratory Standards Guideline

Figure 27: Gram stain smear of Vincent Angina

However, the presence of this complex does not excludethe need to search for other pathogens, particularly S. pyogenes.

Although small numbers of C. albicans or other Candida species may be partof the normal oral flora, oral candidiasis results when the number of organismsincreases considerably in certain pathological conditions, e.g. in malnourishedpremature babies, in immunodeficient adults (e.g. patients withHIV/AIDS), or in patients who have received broad-spectrum antimicrobialsor cancer therapy. The affected area—tongue, tonsils, throat or buccal mucosa—may be extremely red, or covered with white patches or a confluent grey-white membrane (thrush). The diagnosis of candidiasis is best made byfinding numerous yeast cells, some of them forming long mycelium-like filaments,in a Gram-stained smear of the exudate.Swabs from the upper respiratory tract may be submitted to the laboratory,not for the diagnosis of a clinical infection, but to detect a potential pathogenin a healthy subject, a pharyngeal or a nasal “carrier”. This should only bedone as part of well-defined epidemiological surveys.

The followingpathogens can give rise to a carrier state in the upper respiratory tract:

• Staphylococcus aureus. Sampling of patients and staff for nasal carriers issometimes performed as part of an investigation of hospital outbreaks ofMeticillin-resistant S. aureus (MRSA).

• Neisseria meningitidis. Carriage of meningococci may be very prevalent (20% or more) even at non-epidemic times. Identification of pharyngealcarriers of meningococci is rarely needed, and need not be performed priorto the administration of prophylactic antibiotics to family or other closecontacts of patients with meningococcal disease.

149 | Page PHCPI Laboratory Standards Guideline

• Streptococcus pyogenes. Carriage of this organism in low numbers may beprevalent, especially among schoolchildren (20–30%).

• Corynebacterium diphtheriae. The carrier rate of the diphtheria bacillus ishigh in non- vaccinated populations. In such communities, it may be justifiedto identify and treat carriers among the close contacts of a patient withproven diphtheria. Carriers are rare when an immunization programme iscorrectly implemented.

Collection and dispatch of specimens

Ideally, specimens should be collected by a physician or other trained personnel.

The patient should sit facing a light source. While the tongue is keptdown with a tongue depressor, a sterile cotton-wool swab is rubbed vigorouslyover each tonsil, over the back wall of the pharynx, and over any otherinflamed area. Care should be taken not to touch the tongue or buccal surfaces.

It is preferable to take two swabs from the same areas. One can be usedto prepare a smear, while the other is placed into a glass or plastic containerand sent to the laboratory. Alternatively, both swabs may be placed in the containerand dispatched to the laboratory. If the specimen cannot be processedwithin 4 hours, the swab should be placed in a transport medium (e.g. Amiesor Stuart).

Direct microscopy

The fusospirochaetal complex of necrotizing ulcerative pharyngitis (Vincentangina) and Candida are best recognized on a Gram-stained smear, whichshould be prepared if the physician makes a special request. The Gramstainedsmear is not useful for the detection of streptococci or Neisseria spp.

Moreover, the direct smear has poor sensitivity and specificity for the detectionof the diphtheria bacillus, unless the specimen has been collected withcare and is examined by an experienced microbiologist. In the absence of aphysician’s request or of clinical information, a Gram-stained smear shouldnot be made for throat swabs.

Culture and identification Culture for Streptococcus pyogenes Immediately upon receipt in the laboratory, the swab should be rubbed overone-quarter of a blood agar plate, and the rest of the plate streaked with asterile wire loop. The blood agar should be prepared from a basal agarmedium without glucose (or with a low glucose content), e.g. tryptic soy agar (TSA). Acidification of glucose by S. pyogenes inhibits the production

150 | Page PHCPI Laboratory Standards Guideline ofhaemolysin. Blood from any species, even human blood (fresh donor blood), can be used at a concentration of 5%. The plates should be filled to a depthof 4–5 mm. Sheep blood is preferred because it does indicate haemolysis ofsome commensal Haemophilus spp. and it gives no haemolysis with the zymogenesvariant of Enterococcus faecalis.

The recognition of β-haemolytic colonies can be improved, and their presumptiveidentification hastened, by placing a co-trimoxazole disc (as usedfor the susceptibility test) and a special low-concentration bacitracin disc overthe initial streaked area. Because S. pyogenes is resistant and many other bacteriaare susceptible to co- trimoxazole, this disc improves the visibility ofβ-haemolysis. Incubation in a candle-jar will detect most b-haemolytic streptococci.

A simple way to increase haemolysis is to stab the agar surface perpendicularlyby inserting the loop deep into the medium to encourage growth of subsurface colonies. After 18 hours and again after 48 hours of incubationat 35–37 ∞C, the blood plates should be examined for the presence of small (0.5–2 mm) colonies surrounded by a relatively wide zone of clear haemolysis.

After Gram-staining to verify that they are Gram-positive cocci, thecolonies should be submitted to specific identification tests for S. pyogenes (see figure 10). Forclinical purposes, presumptive identification of S. pyogenes is based on its susceptibility to a low concentration of bacitracin. For this purpose, a special differentialdisc is used containing 0.02–0.05 IU of bacitracin. The ordinary discsused in the susceptibility test, with a content of 10 units, are not suitable foridentification. A β-haemolytic streptococcus showing any zone of inhibitionaround the disc should be reported as S. pyogenes. If the haemolytic coloniesare sufficiently numerous, the presence or absence of an inhibition zone maybe read directly from the primary blood agar plate. If the colonies are lessnumerous, one or two should be picked from the primary plate, streaked onone-fifth of another plate to obtain confluent growth, and each inoculated areacovered with a bacitracin disc. After overnight incubation, the subculturesshould be read for inhibition zones.

In some laboratories this presumptive identification is confirmed by serologicaldemonstration of the specific cell wall polysaccharides. This can be doneeither by the classical precipitin method, or more rapidly by using a commercial kit for the rapid slide coagglutination or latex agglutination tests. Ifdesirable, bacitracin-resistant b-haemolytic streptococci can be further identifiedusing some simple physiological tests. Minute colonies ofb-haemolytic streptococci may be encountered, which, when grown and serologicallygrouped, react with group A antiserum. These streptococci are not considered to be S. pyogenes and are not associated with the serious infectionscaused by group A streptococci.

151 | Page PHCPI Laboratory Standards Guideline

In reporting the presence of S. pyogenes in a throat culture, a semiquantitativeanswer should be given (rare, +, + +, or + + +). Patients with streptococcalpharyngitis generally show massive growth of S. pyogenes, with colonies overthe entire surface of the plate. Plates of carriers generally show fewer than 20colonies per plate. Even rare colonies of b-haemolytic streptococci should beconfirmed and reported.

Table 12: Differentiation of β-haemolytic streptococci

Culture for Corynebacterium diphtheriae

Although the diphtheria bacillus grows well on ordinary blood agar, growthis improved by inoculating one or two special media:

• Löffler coagulated serum or Dorset egg medium. Although not selective, bothof these media give abundant growth of the diphtheria bacillus afterovernight incubation. Moreover, the cellular morphology of the bacilli ismore “typical”: irregularly stained, short to long, slightly curved rods, showing metachromatic granules, and arranged in a V form or in parallel palisades. Metachromatic granules are more apparent after staining withmethylene blue or Albert stain than with the Gram stain (see figure 28)

152 | Page PHCPI Laboratory Standards Guideline

Figure 28.Corynebacterium diphtheriae cells stained by Albert's technique.The barred appearance is due to the presence of polyphosphate inclusions called volutin. Note also the characteristic "Chinese-letter" arrangement of cells.

• A selective tellurite blood agar. This medium facilitates isolation when thebacilli are few in number, as is the case for healthy carriers. On thismedium, colonies of the diphtheria bacillus are greyish to black and arefully developed only after 48 hours. Suspicious colonies, consisting of bacilli with a coryneform morphology on the Gram-stained smear, should be subcultured to a blood agar plate to check for purity and for “typical”morphology. It should also be remembered that colonies of the mitisbiotype of C. diphtheria, which is the most prevalent, show a marked zoneof b-haemolysis on blood agar.

A presumptive report on the presence of C. diphtheriae can often be given atthis stage. However, this should be confirmed or ruled out by some simple biochemical tests and by demonstration of the toxigenicity (see table14 below).

As the latterrequires inoculation of guinea-pigs or an in vitro toxigenic test (Elek) and hasto be performed in a central laboratory, only rapid biochemical identificationwill be covered here. C. diphtheriae is catalase- and nitrate-positive. Urea is nothydrolysed. Acid without gas is produced from glucose and maltose, generally not from saccharose. The fermentation of glucose can be tested on Kliglermedium. Urease activity can be demonstrated on MIU and nitrate reductionin nitrate broth in the same way as for Enterobacteriaceae. For the fermentation of maltose and saccharose, Andrade peptone water can be used as a base with a 1% final concentration of each carbohydrate. Results can usually beread after 24 hours, although it may be necessary to reincubate for one night.

153 | Page PHCPI Laboratory Standards Guideline

It must be emphasized that the microbiology laboratory’s role is to confirmthe clinical diagnosis of diphtheria. Therapy should not be withheld pendingreceipt of laboratory reports.

Table 13. Biochemical identification of clinically significant corynebacteria

species CYS PYZ Nitrate Urea Glucose Maltose Sucrose Glycogen C. diphtheria v.gravis + _ + _ + + _ + v.mitis + _ + _ + + _ _ v.intermedus + _ + _ + + _ _ v.belfanti + _ _ _ + + _ _ C.ulcerans + _ _ + + + _ + C. pseudotuberculosis + _ _ + + + _ _ C. amycolatum _ + v V + v V _ C. imitans _ ± _ _ + + ± _ C. _ v + + _ _ _ _ pseudodiphtheriticum C. striatum _ + + _ + _ v _

Susceptibility testing

Routine susceptibility tests on throat or pharyngeal isolates are most often notrequired, and may even be misleading. The major pathogens involved inbacterial pharyngitis are S. pyogenes and C. diphtheriae benzylpenicillin anderythromycin are considered as the antimicrobials of choice to treat both typesof infection. In cases of diphtheria, treatment with antitoxin is also indicated.

154 | Page PHCPI Laboratory Standards Guideline

6.5 Nasal Swab Nasal colonization with Staphylococcus aureus increases the risk of staphylococcal infections at other sites of the body such as and dialysis access sites. It is also associated with recurrent skin infections and nosocomial infections in nurseries and hospital wards. S. aureus is a major cause of morbidity and mortality in haemodialysis patients4 as most patients carry the organism in their anterior nares.

Eradication of nasal carriage of S. aureus may be beneficial in certain clinical conditions such as recurrent furunculosis. Systemic, in addition to topical, treatment is appropriate for nasally colonised patients who have infection elsewhere. Topical antibacterial agents such as mupirocin and chlorhexidine/neomycin are preferred to systemic formulations when a patient is identified as a carrier.

Nose swabs may be used to investigate carriage of Methicillin Resistant Staphylococcus Aureus (MRSA).

Specimen Collection, Transport and Storage Safety considerations  Use aseptic technique.  Collect specimens in appropriate marked leak proof containers and transport specimens in sealed plastic bags.  Collect swabs into Amies transport medium with charcoal and transport in sealed plastic bags.  Compliance with postal and transport regulations is essential.  Collect specimens before antimicrobial therapy where possible.  Specimens should be transported and processed as soon as possible.  If processing is delayed, refrigeration is preferable to storage at ambient temperature. Delays of over 48hr are undesirable.

Correct specimen type and method of collection  Plain sterile cotton wool swab. Sample the anterior nares by gently rotating the swab over the mucosal surface.  Unless otherwise stated, swabs for bacterial and fungal culture should then be placed in Amies transport medium with charcoal.  Numbers and frequency of specimen collection are dependent on clinical condition of patient.

155 | Page PHCPI Laboratory Standards Guideline

Specimen Processing Inoculate each agar plate with a swabfor the isolation of individual colonies, spread inoculum with a sterile loop.

Table 14: Culture media, conditions and organisms for all specimens

Staphylococci Identification

Staphylococci are the bacteria most frequently associated with the productionof pus. Staphylococci grow well aerobically on blood agar and form opaquewhite to cream colonies, 1–2mm in diameter, after overnight incubation. They are unique in growing on media with a high salt content, such as Manitol salt agar (MSA). Theycan be differentiated from streptococci by their morphology and by the productionof catalase. Catalase production by staphylococci is shown by placingit in a drop of 3% hydrogen peroxide onto the colonies deposited on a cleanglass slide. The appearance of bubbles of oxygen is an indication of catalaseproduction.

For clinical purposes, staphylococci can be divided into those that producecoagulase and those that do not. The coagulase-producing staphylococcibelong to the species S. aureus, which is the species of greatest medical interest.

Of the several coagulase-negative species, only two will be consideredhere—S. epidermidis and S. saprophyticus.

Although S. aureus is part of the commensal microbial flora of the nose (40%of healthy adults are positive), skin, and intestinal tract, this species causesimpetigo, boils, abscesses, wound infection, infection of ulcers and burns,osteomyelitis, mastitis (breast abscess), pleural empyema, pyomyositis, toxicshock syndrome, and other types of pyogenic infection.S. epidermidis is also a common commensal of the skin, nose, and other mucousmembranes and possesses a very low pathogenicity. However, its presence inpus should not always be dismissed as skin contamination. Despite its lowinfectivity, S. epidermidiscan cause skin

156 | Page PHCPI Laboratory Standards Guideline infections at the site of an in-dwellingcatheter, cannula, or other device. Infections with S. epidermidis are particularlytroublesome in cardiac and orthopaedic surgery involving the insertionof prosthetic devices (artificial heart valves or artificial hips).

S. saprophyticus is recognized to be a common cause of urinary tract infectionsin young women, being second only to E. coli in some populations.

The distinctive features of the three main species of Staphylococcus are givenin Table (16).and flow chart in figure 26.

Table 15: Differentiation of medically important species of Staphylococcus

157 | Page PHCPI Laboratory Standards Guideline

Figure 29: Aflow diagram for the preliminary identification of staphylococci

In view of the importance of the coagulase test in the identification of S. aureus, this test is described here in detail. Coagulase is an enzyme that causes plasma to clot. Staphylococcal coagulase exists in two forms: bound coagulase orclumping factor, which is demonstrated in the slide test, and free coagulase, which is demonstrated in the tube test.

158 | Page PHCPI Laboratory Standards Guideline

• Slide test. On a clean slide, emulsify one or a few similar colonies ofstaphylococci in a drop of saline. The suspension must be fairly thick. Dipa straight wire into plasma and use this to stir the bacterial suspension.

Observe for clumping within 10 seconds. False-negative slide tests occurapproximately 10% of the strains of S. aureus. If the slide test is negativefor an isolate that seems to be pathogenic on other grounds (pigment, clinicalsource), it should be re-examined in the tube test.

• Tube test. Dispense a few drops (0.5 ml) of plasma into a sterile 12 \75 mm tube, and add two drops of the pure culture in broth. A suspensionwith an equivalent density may also be prepared directly fromgrowth on blood agar. Incubate the tube at 35 ∞C for 4–18 hours and thenexamine for clotting.

The plasma used in the coagulase test may be fresh human or rabbit plasmaobtained with ethylenediamine tetraacetic acid (EDTA). It should be stored inthe refrigerator in small amounts (1 ml), and its performance checked withcultures of S. aureus and S. epidermidis, run in parallel.

Antimicrobial susceptibility testing Prudent use of antimicrobials according to local and national protocols is recommended.

Reporting Procedure:

Culture Report presence or absence of specific pathogens, also report results of supplementary investigations: Negatives "Staphylococcus aureus NOT isolated". Positives "Staphylococcus aureus isolated".

159 | Page PHCPI Laboratory Standards Guideline

6.6. SMART Tests

6.6.1 Helicobacter pylori Serum/Whole Blood The exact role that H. pylori plays in gastrointestinal disease still needs to be precisely defined and is the subject of ongoing research. However, the prevalence rates for H. pylori infection as demonstrated by histological and bacteriological methods can approach 90% in patients who present clinical symptoms of the gastrointestinal diseases.

Clinical significancece: Helicobacter pylori has been associated with a variety of gastrointestinal diseases including gastritis, duodenal and gastric ulcer, non-ulcer dyspepsia, gastric adenocarcinoma and lymphoma.

Principle of the test This assay is a double antigen chromatographic lateral flow immunoassay. The test strip in the deviceincludes: 1) a burgundy-colored conjugate pad containing colloidal gold coupled with H. pylori antigens, and 2) nitrocellulose membrane containing a test line (T line) and a control line (C line). The T line iscoated with H. pylori antigens, and the C line is coated with goat anti-H. pylori antibody. The antigensused in this device are from H. pylori cell lysate.

When IgG antibodies specific to H. pylori are present in the specimen, the T line will become aburgundy-colored band. If antibodies to H. pylori are not present or are present below the detectablelevel, no T line will develop. The C line should always appear as a burgundy- colored band regardlessof the presence of antibodies to H. pylori. The C line serves as an internal qualitative control of the testsystem to indicate that an adequate volume of specimen has been applied and the flow occurred.

Materials and reagent • Test devices, each sealed in a pouch with a dropper pipette. • 1 bottle of wash buffer-7 ml PBS diluents with 0.02% sodium azide as a preservative. • 1 package inserts (Instruction for Use). •Lancet or other blood collection device (Vacutainer blood collection tubes, sterile needles and syringes). • Timer.

Specimen collection and storage

1- Serum • Follow standard laboratory procedures to collect serum specimens.

160 | Page PHCPI Laboratory Standards Guideline

• Serum specimens can be stored at (9-30) °C for 8 hours, at (2-8) oC for one week, and at (≤ - 20) °C or lower for long term storage. Repeatedly frozen and thawed specimens arenot recommended for this assay. • Any sediment in serum specimens should be removed by centrifugation. Avoid using any turbidspecimens, which may be contaminated by microorganisms.

2- Whole Blood • Finger stick sampling is recommended for this assay. • Middle or ring finger is the preferred puncture site. • Clean patient’s finger with an alcohol swab. Wait until it is dry. • Puncture the fingertip with the lancet. Wipe away first sign of blood. • Gently rub the hand from palm to finger to help form a drop of blood over the punctured site. • Use the provided pipette to pick up the blood, and apply one drop of the blood to the sample well ofthe device. Then, follow the procedure.

Procedure 1. Refrigerated specimens and other test materials, including devices, must be equilibrated to roomtemperature before testing. 2. Remove the device from its wrap pouch prior to performing the assay, label the device withidentification. 3. Add one drop of fresh blood or serum to the sample well marked "S". Allow about 30 seconds forthe specimen to be absorbed totally. Discard the first three drops of wash buffer from the washbuffer squeeze bottle. Then add three drops of wash buffer into the sample well. 4. Strong positive results may be observed in 2-3 minutes or depending on manufacturers. Weak positive results may take a longertime, up to 7 minutes. For the whole blood test, a slight hemolysis might be observed, but it doesnot interfere with the results.

Interpretation of results IMPORTANT: Interpret the results depending on manufacturers. The T line should always be interpreted independently of the C line.

161 | Page PHCPI Laboratory Standards Guideline

Figure 30: Interpretation of Smart Test Results

Positive: If both the C line and T line appear, the result indicates that the IgG antibodies specific to H. pylori aredetected and the result is positive. A faint line in test region indicates a borderline specimen, which should be re-tested using an alternative method for confirmation. Negative: If only the C line appears in the control region, the test indicates that no antibodies to H. pylori aredetected and the result is negative. Invalid: When no control line appears within 5 minutes, repeat the test with a new test device.

Quality Control Procedure • Built-in Control Features This test contains a built-in quality control feature, the C line. The appearance of the burgundy C line indicates that that an adequate volume of specimen and wash buffer has been applied and theflow occurred.

• External Quality Control External controls are recommended, positive and negative, to monitor the performance of the assay.

162 | Page PHCPI Laboratory Standards Guideline

6.6.2 Salmonella Typhi Antigen

The diagnosis of typhoid consists of isolation of the bacilli and the demonstration of antibodies. The isolation of the bacilli is very time consuming and antibody detection is not very specific. Other tests have been developed that take only 10-20 minutes and require only a small quantity of stool or one drop of serum to perform. It is the easiest and most specific method for detecting S. typhi infection.

Principle of the test

OneStep S. Typhi Antigen test is a qualitative immunochromatographic assay. The test employs a conation of monoclonal antibody/colloidal gold dyeconjugate and a polyclonal antibody immobilized on the solid phase. This will selectively identify S. typhi antigen associated S. typhi (typhoid) infection with a high degree of sensitivity andspecificity.

As the specimen flows through the absorbent pad in the sample well and through the antibody/colloidalgold complex any S. typhi antigen present in the sample binds to the conjugate forming anantigen/antibody complex. The sample and dye complex continue to migrate along the membrane tothe immobilized polyclonal antibody. In the presence of S.typhi, the polyclonal antibody captures thecomplex. This forms a visible pink/purple band in the (B) or test area of the card. If no antigen ispresent, there is no line formation in the (B) area. The remaining complex continues to migrate toanother immobilized antibody on the membrane in the (C) or Control area of the card, and is capturedwhich then forms a band indicating proper performance of the test.

Materials Each kit contains: 1. S. typhi Antigen test. Each cassette contains a test strip with S. typhi specific antibody on the test region of themembrane and colored S. typhi antibody-gold conjugate pad. 2. Fecal sample buffer – 2 bottles, 8 mL each. 3. Instruction for use. 4. Specimen collection container. 5. Timer.

Specimen collection and storage OneStep S. typhi Antigen test can be run on stool or serum samples.The test works best on fresh samples. If testing cannot be done immediately, they should be stored at (2-8) °C after collection for up to 3 days. If testing cannot be done within 3 days, serum can be storedfrozen at (–20) °C

163 | Page PHCPI Laboratory Standards Guideline or colder.Shipment of samples should comply with local regulations for transport of etiologic agents.

Stool and serum specimens should be collected in containers that do not contain media, preservatives,animal serum or detergents as any of these additives may interfere with the Cortez S. typhi AntigenTest.

Reagent preparation Bring all reagents, including test device, to room temperature (20-30) oCbefore use.

Stool specimen preparation Add about 1/4 gram stool to approximately 500μl of the extraction reagent provided (about 12 drops from thedropper vial provided). Mix well and allow to sit for 5 minutes or so to allow the large particles to settle. (Amount of stool, extraction reagent and appropriate wetting time depending on manufacturer).

Note: Watery or diarrhea specimens are inappropriate for testing.

Procedure 1. Bring all materials and specimens to room temperature. 2. Remove the test card from the sealed foil pouch. 3. for stool samples: use the provided pipette to transfer sample from the upper layer of the stoolextract and add 3 drops to the sample well (marked as “A”). 4. for serum samples: use the provided pipette to transfer the serum sample and add 3 drops to thesample well (marked as “A”). 5. Read the result at 20 minutes. A strong positive sample may show test band earlier. However, toconfirm a result is negative, it must wait 20 minutes to read the results.( See figure 28)

Note: The amount of S. typhi antigens present in serum is typically less than that in stool. This may decrease the sensitivity of the test when using serum depending how soon after the onset of the infection the test is performed. Early infection typically exhibits greater levels of the antigen in the serum than in later infection. To confirm serum results:

The use of a stool sample is recommended if serum is used first and a negative result is obtained and typhoid is still suspected, a second test run on a stool sample should be performed.

164 | Page PHCPI Laboratory Standards Guideline

Interpretation of results Positive result: A distinct pink colored band appears on test line regions, in addition to a pink line onthe control line region.

Negative result: No line appears in the test line region. A distinct pink line shows on the control lineRegion.

Invalid: The control line next to the test line does not become visible within 20 minutes after theaddition of the sample.

Figure 31: Interpretation of Salmonella Smart Test

Quality Control

1. The control band is an internal reagent and procedural control. It will appear if the test has beenperformed correctly and the reagents are reactive. 2. Good Laboratory Practice recommends the daily use of control materials to validate the reliability of the device.

Limitations 1. The test is for qualitative detection of S.typhi antigen in stool or serum sample and does notindicate the quantity of the antigens.

165 | Page PHCPI Laboratory Standards Guideline

2. The test is for in vitro diagnostic use only. 3. For samples that test positive (reactive) by Cortez S. typhi Antigen Test, more specificconfirmatory testing should be done. A definitive clinical diagnosis should only be made by thephysician after all clinical and laboratory finding have been evaluated. The use of a rapid test aloneis not sufficient to diagnose S. typhi infection even if antigen is present. Also, a negative result doesnot preclude the possibility of infection with S. typhi.

Specificity Negative stool samples from patients in areas where typhoid is relatively rare and would yield a typicalnegative population, showed no false positives when the test was read within 20 minutes as specified. Samples that were positive for s. paratyphi however were also negative as the antibodies used in the S. typhi rapid test are specific for S. typhi only.

6.6.3 Vibrio Cholerae Rapid Visual Antigen Detection Test

Clinical Significant Vibrio cholerae strains belonging to the O1 and O139 serogroups are capable of causing epidemic and pandemic cholera.Clinically, cholera may range from asymptomatic colonization to sever diarrhea with massive fluid loss, leading to dehydration, electrolyte disturbances, and death.

The O1 serogroup is subdivided into two serotypes, Ogawa and Inaba. Serogroup O139, which appeared in India in 1992, hasspread rapidly throughout Asian countries and is considered tobe the potential eighth pandemic strain of cholera. Prompt diagnosis of cholera is ofkey importance to initiate effectivetherapy and to institute proper epidemiological measures.

Several rapid diagnostic tests for cholera have been described.Some detect the cholera toxin. The others detectthe lipopolysaccharide (LPS) antigen of V. cholerae O1 or O139.

Principle This test is an immunochromatographic rapid visual antigen detection test for V. cholerae from stool specimens. The nitrocellulose membrane is coated with monoclonal antibodies to V. cholerae O1 and O139 lipopolysaccharide (LPS) as two distinct bands. When the test sample migrates through the nitrocellulose membrane, colloidal gold particles coupled with anti-V. CholeraeO1 and O139 LPS monoclonal antibodies bind to the respective antigens from the test

166 | Page PHCPI Laboratory Standards Guideline sample, if present. This antigen-antibody complex moves through the nitrocellulose membrane and binds to the corresponding immobilized antibodies against V. cholerae O1 and/or O139, forming a magenta red color band, which confirms positive results. A control test band should always appear, irrespective of a positive or negative test sample, to validate the test procedure.

Materials and equipment In countries at risk for epidemics of cholera, the laboratory’s first role is to be preparedfor a possible epidemic. It must have ready or have access to the supplies and equipment necessary to identify V. cholerae O1 and/or O139 even in remote locations. Contact the local laboratory, medical center, or field site where testing will be performed in advance to ensure all necessary supplies, equipment, and personal protective equipment (PPE) will be readily available.

Supplies and equipment necessary to perform the rapid testIncluded with the rapid test kit:

 Individually packaged dipsticks in aluminum pouch with desiccant Plastic, individual, single-use 5-mL test tubes for sample.  Disposable transfer pipettes (plastic droppers) for stool.  Foam test tube stand.  Sterile normal saline or distilled water Disposable, clean sample containers for stool specimen collection.  Plastic rack of appropriate size for test tubes (suggested).  Watch or timer.  Biosafety supplies/personal protective equipment (PPE):Gloves Goggles or  Lab coat.  Biohazard receptacle for infectious waste.  Disinfectant.

Storage The rapid test kit may be stored at ambient temperature (24–26)oC but it is recommended that the kit be stored at 4oC when not in use for 6 months or more (depending on manufacturer).

Specimen’s collection Fecal specimens should be collected in the early stages of any enteric illness, when pathogens are usually present in the stool in highest numbers, and before antibiotic therapy has started. Collect stool specimens from persons who are suspected of being part of an outbreak and who meet the following criteria: 1. Currently have “rice water stool.” 2. Had onset of illness less than 4 days before sampling. 3. Have not received antimicrobial treatment for the diarrheal illness.

167 | Page PHCPI Laboratory Standards Guideline

Collect stools from patients in clean (no disinfectant or detergent residue) containers with tight-fitting, leak-proof lids. Do not collect specimens from bedpans, as they may contain residual disinfectant or other contaminants. Stool should be refrigerated if possible and processed within a maximum of 2 hours after collection. Suitable specimens: Liquid fecal specimens, ideal “rice water stool”—clear, watery stools, likened towater from boiled rice, an appearance classically seen in cholera cases. Viscous, mucoid, or semisolid stools—acceptable but must be diluted before usewith normal sterile saline or sterile distilled water.

Unacceptable specimens: Solid stools—not recommended, as particulate matter from these specimens may clog the nitrocellulose membrane restricting migration of the sample. Rectal swabs—the dipstick assay has not been evaluated for use with rectal swabs.

168 | Page PHCPI Laboratory Standards Guideline

Procedure

1. Put on personal protective equipment (PPE): lab coat, gloves, goggles and face shield. Wear PPE at all times while handling the specimen 2. Collect stool sample in plastic cup. Label cup with patient identifier. 3. Remove 1 test tube from kit and place test tube in Styrofoam or plastic rack. Label the test tube. 4. Remove 1 plastic dropper from kit.

5. Fill dropper halfway (150–200 µL) with stool sample, and transfer to test tube. Do not put more liquid than this amount in the test tube. 6. Open packet with rapid test, Label the dipstick with patient identifier. 7. Place the dipstick in the test tube with the arrows facing DOWN. The end of the strip should be submerged in the stool. The arrows should remain ABOVE the level of the stool. 8. Wait 15–20 minutes for the test to complete. When complete, you will see a band near the top of the dipstick (internal control band). 9. Review and interpret results within 15–20 minutes: A. Positive, V. cholerae O1 B. Positive, V. cholerae O139 C. Positive, both V. cholerae O1 and O139 D. Negative for V. cholerae O1 and O139 E. Test did not work correctly (invalid test) Do not interpret results after 30 minutes. A B C D E

10. Place all waste in a double-lined plastic bag labeled “biohazard.”

Figure32: Vibrio Cholerae Rapid Visual Antigen Detection Test Instruction Sheet, for a graphic representation of the test procedure

169 | Page PHCPI Laboratory Standards Guideline

1. Allow time for the kit to come to room temperature before opening and using. 2. Label a clean, unused test tube with the patient identification code. Transfer 150– 200 L of the patient’s liquid stool specimen to this test tube. Note: Viscous, mucoid, or semisolid specimens should be diluted with sterile saline or distilled water before transfer to the test tube. 3. Tear open aluminum pouch at notched area. Remove the dipstick. 4. Label the dipstick with the patient identification code; take care to not touch the area marked with an arrow (dipping area). 5. Carefully place the dipstick in the test tube with stool with the arrows facing DOWN. The end of the strip should be submerged in the stool. Be sure the arrows remain ABOVE the level of the stool. 6. Leave the dipstick undisturbed for 10–15 minutes or depending on manufacturers. 7. within 15-20 minutes, a magenta red internal “control” band will appear indicating a valid test; this line indicates that the reaction is complete. The presence of at least one or two additional magenta red lines signifies the sample is positive for V. cholerae O1 and/or V. cholerae O139. 8. Read and record the result of the test within 15–20 minutes. Do not interpret results after 30 minutes. (Reading & interpretation the result depending on manufacturers).

Interpretation and reporting of results Interpret test results of each dipstick according to the manufacturer’s instructions. Guidelines for interpreting test results can be found in Table (17) The report should include any comments or information the person needs to properly evaluate the test results.

170 | Page PHCPI Laboratory Standards Guideline

Table 16: Expected Test Result Outcomes for the Dipstick and Interpretation of Vibrio Cholerae Rapid Visual Antigen Detection Test Dipstick Test Result Interpretation

POSITIVE (Figures 3-1 and 3-2, column Indicative that the sample is positive only for V. A): Appearance of two bands, one for the cholerae O1. control test band and one test band specific for V. cholerae O1 (Vc O1).

POSITIVE (Figures 3-1 and 3-2, column Indicative that the sample is positive only for V. B): Appearance of two bands, one for the cholerae O139. control test band and one test band specific for V. cholerae O139 (Vc O139).

POSITIVE (Figures 3-1 and 3-2, column Indicative that the sample is positive for both V. C): Appearance of three bands, one for the cholerae O1 and O139. control test band, one test band specific for V. cholerae O1 (Vc O1) and one test band specific for V. cholerae O139 (Vc O139)

NEGATIVE (Figures 3-1 and 3-2, column Indicative that the sample is negative for V. cholerae D): Appearance of only the control test O1 and O139. band.

INVALID (Figures 3-1 and 3-2, column Indicates a procedural error, deterioration of the E): Appearance of no control test band specimen/dipstick or the presence of particulate matter in the specimen preventing flow of fluid through the nitrocellulose membrane.

Quality control:

Common source of error Specimens that are not classified as “watery” may clog the nitrocellulose membrane strip, limit migration of the liquid, and prevent the dipstick reaction resulting in an invalid test.

171 | Page PHCPI Laboratory Standards Guideline

These specimens should be either diluted with sterile saline or distilled water or rejected as unacceptable specimens for testing.

A B C D E

Figure33: Visual Interpretation of Rapid Dipstick Test Results

Precautions Be sure to run the test exactly as instructed, adhering to all procedures. Use sterile technique throughout the test procedure. Wear gloves, goggles and face shield, and a lab coat when handling and testing all clinical specimens.

6.7 Preparation of CultureMedia With just a few diagnostic materials, a laboratory can make an importantcontribution to individual patient care through accurate etiological diagnosis. In most developing countries bacteriological laboratory practice is hamperedby a shortage of culture media and basic reagents, which are very costly toimport. However, the number of culture media and reagents that have to bepurchased can be reduced to the essential ones, through rational selection, ashas been the case with essential drug lists. Additionally, some simple media and reagents can be produced or prepared locally.

172 | Page PHCPI Laboratory Standards Guideline

Each batch of medium prepared from individual ingredients or each differentmanufacturer’s lot number of commercial dehydrated medium and eachbatch of medium prepared from individual ingredients should be qualitycontrolled before use. Immediately after preparation, each medium should betested, as appropriate, with a reference strain for proper growth characteristics as described for each medium. A record of all media preparation or purchase dates and quality control test results should be kept, and any unusual characteristic (e.g., the color of the medium orslow growth of test bacteria) should be noted. Many media call for the use of defibrinated blood. Agar media should be dispensed into 15 x 100-mm or 15 x 150-mm Petri dishesto a uniform depth of 3–4 mm; approximately 20-ml of liquid agar medium willachieve this depth in a 15 x 100-mm plate. If agar is cooled to 50°C prior topouring, condensation is minimized. After pouring, the plates should be kept atroom temperature for several hours to prevent excess condensation from formingon the covers of the dishes. Another means by which condensation will be reducedis if plates are stacked so that they cool more slowly. Alternatively, if when preparing selective media (e.g., MacConkey [MAC], xylose lysine desoxycholate[XLD], thiosulfate citrate bile salts [TCBS] agar, etc.), conditions are such thatthere is little chance that the cooling media will be contaminated, after the agar ispoured into the plates, the lids can be placed on the dish so that a small opening isleft to let the heat out, resulting in the formation of less condensation on the upperlid; the lid should remain slightly open like this for approximately 30 minutes, while the agar solidifies. If, however, it is likely that the agar will be contaminated ifthe lid is left partly open, the agar should be allowed to solidify with the lid closed. Note: Covering the agar while it is still hot will allow for the formation of asubstantial amount of condensation on the upper lid. If the plates containcondensation, the plates should be covered at room temperature for 24 hoursto allow the condensation to evaporate. After condensation has evaporated, theplates should be placed in an inverted position and stored in a plastic bag in aninverted position at 4°C.

6.7.1 Blood agar: Tryptic Soya Agar (TSA) with 5% sheep blood Sheep blood agar is used as a general blood agar medium, and consists of TSA plus5% sheep’s blood. The sheep blood agar plate should appear a bright red color. Ifthe plates appear dark red, the blood has been added when the agar was too hot; if this happens, the medium should be discarded and a new batch prepared.

A. Prepare TSA according to the instructions given on the label of the dehydrated powder. For convenience, 500 ml of molten agar can be prepared in a l-liter flask. Add 20 g of agar into 500 ml of water. Heat to dissolve.

173 | Page PHCPI Laboratory Standards Guideline

B. Autoclave at 121°C for 20 minutes. Cool to 60°C. C. Add 5% sterile, defibrinated sheep blood (i.e., add 25 ml sheep blood to 500 ml of agar). If a different volume of basal medium is prepared, the amount of blood added must be adjusted accordingly to 5% (e.g., 50 ml of blood per liter of medium). D. Dispense 20 ml into 15 x 100-mm Petri dishes. Allow the medium to solidify and dry out, place in a plastic bag, and store at 4°C.

Quality control:

Test each new, freshly prepared or purchased batch of blood agarplates for growth and hemolytic reaction with a strain of S. pneumoniae. Thecolonies are small and should appear grey to grey-green surrounded by a distinct greenish halo in the agar.

6.7.2 Chocolate agar with TSA base and growth supplement Chocolate agar with growth supplements is a medium that supports the specialgrowth requirements needed for the isolation of fastidious organisms (whenincubated in a 5% CO2 atmosphere). Chocolate agar contains a reducedconcentration of agar, which increases the moisture content of the medium. Supplemented chocolate agar should support the growth of H. influenzae. Chocolate agar slants for transport and short-term storage can be prepared in thesame manner as that described for agar plates, except that the medium is dispensedin 16 x 125-mm screw- cap tubes and slanted before solidifying. A. Use TSA as the basal medium. Prepare double strength (i.e., 20 g in 250 ml distilled water). Autoclave and then cool to 50°C. Use the thermometer to verify the cooling temperature. B. Prepare a solution of 2% hemoglobin (i.e., 5 g in 250 ml distilled water). Mix the hemoglobin in 5–6 ml of the distilled water to form a smooth paste. Continue mixing as the rest of the water is added. Autoclave, and cool to 50°C. C. Add the hemoglobin solution to the double-strength TSA and continue to hold at 50°C. D. • Alternative to steps a-c: If a hemoglobin solution is unavailable, an alternative is to add 5% sterile defibrinated sheep, rabbit, guinea pig, or horse blood (i.e., 5 ml blood per 100-ml agar) to full-strength TSA (i.e., 20 g in 500 ml distilled water). DO NOT use human blood. After the base medium has been autoclaved and cooled to 50°C, add the blood and place ina hot water bath at no more than 80°C for 15 minutes or until a chocolate color is achieved. Then cool to 50°C.

174 | Page PHCPI Laboratory Standards Guideline

E. After the hemoglobin solution or the defibrinated blood has been added to the base medium and the medium has cooled to 50°C, add growth supplement (e.g., IsoVitaleX or Vitox) to a final concentration of 1%.Mix the ingredients by gently swirling the flask; avoid forming bubbles. Dispense 15–20 ml in each 15 x 100-mm Petri dish.

Quality control:

All freshly prepared or purchased chocolate agar media should betested to determine the medium’s capacity to support growth of the bacteria to beisolated, particularly H. influenzae. If the medium does not support the growth ofone or all of the bacteria, the medium should be discarded, and a new batch ofmedium should be prepared or purchased.

• Chocolate agar should look brown to brownish-red in color. N. meningitides and H. influenzae should appear as a greyish, almost translucent film on theslant’s surface with no discoloring of the medium after 24 hours of incubation;S. pneumoniae should appear as small grey to grey-green colonies with avery distinct greenish discoloring of the medium.

If H. influenzae does not grow, the growth supplement (IsoVitaleX or its equivalent) may have been inadvertently omitted.C

6.7.3 MacConkey agar (MAC) MacConkey agar (MAC) is a differential plating medium recommended for use inthe isolation and differentiation of lactose-nonfermenting, gram-negative entericbacteria from lactose-ermenting organisms. Colonies of Shigella on MAC appearas convex, colorless colonies about 2–3 mm in diameter. S. dysenteriae 1 coloniesmay be smaller. S. Typhi colonies are flat, colorless and usually 2–3 mm indiameter. Several commercial brands of MAC are available.Most manufacturers’ prepareseveral formulations of MAC, which may vary in selectivity and thereby affect theisolation of Shigella. For example, some formulations of MAC do not contain crystal violet, a selective agent; these types are not as selective and should not beused for isolation of Shigella. Oxoid MacConkey Agar No. 3, Difco BactoMacConkey Agar, and BBL MacConkey Agar are all suitable.

A. Prepare MAC according to manufacturer’s instructions. [Note: MAC can also be prepared from individual ingredients, but this produces more lot-to-lot variation than preparation of a commercially available dehydrated formulation.] B. Sterilize the medium by autoclaving at 121°C for 15 minutes.

175 | Page PHCPI Laboratory Standards Guideline

C. Cool to 50°C and pour into Petri plates (to a uniform depth of 3–4 mm). D. Leave lids ajar for about 20 minutes so that the surface of the agar will dry. Close lids and store at 4°C for up to 1 month. If plates are to be stored for more than a few days, put them in a sealed plastic bag to prevent drying.

Quality control: For quality control of MAC, the following organisms should beadequate for confirmation of selective and inhibitory growth characteristics: • E. coli should produce pink to red colonies with good to excellent growth; and, • S. flexneri should produce colorless colonies with fair to good growth, butS. dysenteriae 1 colonies may be smaller.

6.7.4 Mueller-Hinton agar Mueller-Hinton agar is the recommended medium used for standardizedantimicrobial susceptibility testing of certain bacteria; the organisms in this document for which it is appropriate to use this formulation of Mueller-Hintonmedium (i.e., unsupplemented Mueller- Hinton) are S. Typhi, Shigella spp., andV. cholerae.

[Note: Several formulations of Mueller-Hinton agar are commercially available.

This laboratory manual suggests that Mueller-Hinton agar medium should notbe prepared from individual ingredients because this can diminish the quality.Commercial dehydrated Mueller-Hinton is carefully quality controlled before being released for sale.]

A. Follow manufacturer’s instructions to prepare medium. B. After autoclaving, cool medium to 50°C in a water bath. C. Measure 60–70 ml of medium per plate into 15 x 150-mm plates, or measure 25–30 ml per plate into 15 x 100-mm plates. Agar should be poured into flatbottom glass or plastic Petri dishes on a level pouring surface to a uniformdepth of 3–4 mm. using more or less agar will affect the susceptibility results. Agar deeper than 4 mm may cause false-resistance results, whereas agar less than 4 mm deep may be associated with a false-susceptibility report. D. Freshly prepared plates may be used the same day or stored in a refrigerator (at 2°– 8°C) for up to 2 weeks. If plates are not used within 7 days of preparation, they should be wrapped in plastic to minimize evaporation. Just before use, if excess moisture is on the surface, plates should be placed in an incubator (35°–37°C) until the moisture evaporates (usually 10–30 min).

Note: do not leave lids a jar because the medium is easily contaminated.

176 | Page PHCPI Laboratory Standards Guideline

Quality control:

Each new lot of Mueller-Hinton agar should be quality controlledbefore use by testing the E. coli ATCC 25922 standard strain for antimicrobial susceptibility testing. (This formulation of Mueller-Hinton agar can also be usedfor testing of gram-positive aerobes, in which case S. aureus ATCC 25923 can beused as a quality control strain.) The pH of each new lot of Mueller-Hintonshould be between 7.2 and 7.4; if the pH is outside this range, the pH of the medium should not be adjusted by the addition of acid or base, i.e., the batch ofMueller- Hinton plates should be discarded and a new batch of plates prepared. Ifthe pH for every batch is too high or low, the entire lot of dehydrated mediummay have to be returned to the manufacturer as unsatisfactory. Inhibition zonesizes / minimal inhibitory concentration (MIC) values for quality control areincluded in the antimicrobial susceptibility testing section of each pathogenspecificchapter.

6.7.5 Cary-Blair medium, Amies medium, and Stuarts medium Prepare each of these transport media according to the manufacturer’sinstructions. [Note: Several dehydrated formulations of Cary-Blair arecommercially available; some require the addition of calcium chloride and somedo not.] These media can also be prepared from individual ingredients; however, itis very difficult to make a well quality-controlled batch and so this manualrecommends purchasing them from a manufacturer.When the Cary-Blair medium is prepared, it should be dispensed into containersin sufficient volume so that swabs will be covered by at least 4 cm of medium. Forexample, 5- to 6-ml amounts may be dispensed into 13 x 100-mm screw cap tubes.With the caps loosened, sterilize the medium by steaming (not by autoclave) at100°C for 15 minutes. Tighten the caps after sterilization, and store the medium at15°–30°C.

These media are quite stable if stored in tightly sealed containers in a cool darkplace so that the medium does not dry out. Each may be used for up to 1 year as long as no loss of volume, visible contamination (e.g., foreign objects or bacterialgrowth), or color change is observed. Prepared Amie’s medium that has been storedfor longer than 9 months, however, should be freshly steamed and the charcoal re-suspendedbefore use.

6.8 Staining of bacteriological smears

6.8.1 Gram’s Stain Bacteria are classified according to their Gram staining reaction. Gram positive bacteria have thicker and denser peptidoglycan layers in their cell walls. Iodine penetrates the cell wall in

177 | Page PHCPI Laboratory Standards Guideline these bacteria and alters the blue dye to inhibit its diffusion through the cell wall during decolourisation. Gram positive bacteria must have an intact cell wall to produce a positive reaction. Gram negative cells which do not retain the methyl/crystal violet are stained by a counterstain26. Neutral red, safranin or carbol fuchsin may be used as the counterstain.

Gram’s Method for examination of smears 1. Prepare a smear and heat gently to fix. 2. Flood the slide with 0.5% methyl crystal violet and leave for 30 seconds 3. Tilt the slide, pour on sufficient (1%) Lugol’s iodine to wash away the stain, cover with fresh iodine and allow to act for 30 seconds 4. Tilt the slide and wash off the iodine with 95 - 100% ethanol or acetone until colour ceases to run out of the smear. 5. Rinse with tap water. 6. Pour on 0.1% counterstain (neutral red, safranin or carbol fuchsin) and leave to act for about 2 min. 7. Wash with water and blot dry.

Interpretation Positive Result Gram positive bacteria stain deep blue/purple.

Negative Result Gram negative bacteria stain pink/red.

Quality Control Bacteria A culture containing Gram positive and Gram negative bacteriamay be used for quality control.

6.8.2 Ziehl-Neelsen’s stain (for acid fast bacilli) This staining technique is used to demonstrate the presence of acid and alcohol fast bacilli (AFB) which have waxy envelopes that make them difficult to stain and decolourise.Auramine-phenol staining is more sensitive than Ziehl-Neelsen’s and is thus more suitable forassessment of smears from clinical specimens. Ziehl-Neelsen’s staining provides morphologicaldetails and is more useful for confirming the presence of AFB in positive cultures.

Method 1. Flood the slide with strong carbol fuschin. 2. Heat gently, and once slide is just ‘‘steaming’’ leave for 3-5 min.

178 | Page PHCPI Laboratory Standards Guideline

3. Rinse well with water. 4. Decolourise for 2-3 min with a (3% v/v) acid-alcohol solution (97% alcohol + 30%ethanol alcohol), rinse with water, and then replace with fresh acid-alcohol for 3- 4 min until the slide remains a faint pink colour. 5. Rinse well with water. 6. Counter stain with (1% w/v) methylene blue or malachite green for 30 s. 7. Rinse with water and allow todry. 8. Apply immersion oil and read with a transmitted light microscope.

Note: Follow manufacturer’s procedure if commercial kits are used.

Interpretation Positive Result Acid fast bacilli vary from 0.5-10 μm in length16 and stain red. Some may appear beaded.

Negative Result All other organisms and background material stain green if malachite green counterstain isused or blue if methylene blue counterstain is used.

179 | Page PHCPI Laboratory Standards Guideline

7 Parasitology

7.1 Malaria parasite diagnosis Introduction

Malaria is a serious, sometimes fatal disease resulting from infection with Plasmodium spp. transmitted by the bite of Anopheles mosquitoes. The clinical diagnosis, where malariais suspected based on the history, symptoms and clinical findings must always be confirmed by a laboratory diagnosis. Laboratory diagnosis of malaria involves identification of malaria parasite or its antigens/products in the blood of the patient. There are four species of the genus plasmodium responsible for the malarial parasite infections that commonly infect man, P.falciparum, P.vivax, P.malariae and P.ovale. The most important of these is P.falciparum because it can be rapidly fatal and is responsible for the majority of malaria related death.

Clinical significance

Malaria parasites are usually detected in blood films stained with Field or Geimsa stains. They may also be detected using an immunological procedure known as a dipstick test It is important for the prognosis and treatment of the disease that the species involved are identified in the laboratory. For routine malaria microscopy, a thin and a thick film are made on the same slide. The thick film is used for the detection of parasites, while the thin film is used in identifying the species of parasites.

Principle

Blood to be examined for malaria parasites is usually collected at a health centre. The most suitable time for collection is at the height of an episode of fever, when the parasites are most numerous in the blood. Blood specimens should always be collected before anti- malarial drugs are given. Cleaning the finger before collecting a capillary bloodsample using a lancet to puncture the ball of the finger.

180 | Page PHCPI Laboratory Standards Guideline

Materials and reagents

 Microscope  Clean glass microscope slides  Sterile blood lancets  Cotton wool  Grease pencil  Methanol  70% Ethanol.  Measuring cylinders, 10, 50 and 100ml  Beakers, 50 and 250ml  Staining troughs  Glass rods  Wash bottle  Slide forceps  Slide racks  Timer  Giemsa stain  Methanol in a drop bottle  Buffered water, pH 7.2 or distilled water

Preparation of thin and thick blood film: Thin film: Using another clean slide as a “spreader”, and with the slide with the blood drops resting on a flat, firm surface, touch the small drop with the spreader and allow the blood to run along its edge. Firmly push the spreader along the slide, away from the largest drops, keeping the spreader at an angle of 45° Make sure that the spreader is in even contact with the surface of theslide all the time the blood is being spread.

Thick film: Always handle slides by the edges, or by a corner, to make the thickfilm as follows: Using the corner of the spreader, quickly join the larger drops of blood and spread them to make an even thick film, allow the thick film to dry in a flat, level position protected from flies, dust and extreme heat. Label the dry film with a grease pencil by writing across the thicker portion of the thin film the patient’s name or number and date.

181 | Page PHCPI Laboratory Standards Guideline

Method

1. With the patient’s left hand palm upwards, select the third or fourth finger. (The big toe can be used with infants. The thumb should never be used for adults or children.) Use cotton wool lightly soaked in ethanol to clean the finger — using firm strokes to remove dirt and grease from the ball of the finger, dry the finger with a clean piece of cotton wool (or lint).

2. With a sterile lancet, puncture the ball of the finger, using a quickrolling action. By applying gentle pressure to the finger, express the first drop of blood and wipe it away with dry cotton wool. Make sure that no strands of cotton wool remain on the finger 3. Working quickly and handling clean slides only by the edges, collect the blood asfollows:  Apply gentle pressure to the finger and collect a single small drop of blood, about this size, on to the middle of the slide. This is for the thin film.  Apply further pressure to express more blood and collect two or three larger drops, about this size, on to the slide about 1 cm from the drop intended for the thin film Wipe the remaining blood away with cotton wool.

Staining blood films with Giemsa stain

Principle During staining of the blood film, the haemoglobin in the erythrocytes dissolves (dehaemoglobinization) and is removed by the water in the staining solution. All that remain are the parasites and the leukocytes, which can be seen under the microscope.

Methods:

First method: Rapid method for staining thick and thin blood films This method is suitable for rapid staining of thick films when urgent results are required. It uses much more stain than the regular method 1. Using forceps, place the slides back to back in a staining trough 2- Prepare a 3% Giemsa solution in buffered or distilled water, pH 7.2,in sufficient quantity to fill the number of staining troughs being used, mix the stain well.

182 | Page PHCPI Laboratory Standards Guideline

3- Pour the stain gently into the staining trough, until all the slides aretotally covered. Stain for 30–45 minutes out of sunlight. 4. Pour clean water gently into the trough to remove the deposit on thesurface of the staining solution 5. Gently pour off the remaining stain), and rinse again inclean water for a few seconds. Pour the water off.

Second method 1. Allow the thick film to dry thoroughly; if results are required urgently, dryingmay be hastened by fanning, or briefly exposing the slide to gentle heat such asthat from a microscope lamp. Care should be taken to avoid overheating, otherwisethe thick film will be heat-fixed 2. Fix the thin film by adding three drops of methanol, or by dipping it into acontainer of ethanol for a few seconds. To permit dehaemoglobinization, thethick film should not be fixed; therefore avoid exposure of the thick film to methanolor methanol vapour 3. Prepare a 10% Giemsa solution in buffered or distilled water, pH 7.2; if a smallquantity is being used, three drops of stain per ml of buffered water will give thecorrect concentration of Giemsa solution. One slide requires about 3ml ofmadeupstain. Mix the stain well with a . 4. Gently pour the stain on to the slides or use a pipette. Stain for 5–10 minutes. 5. Gently flush the stain off the slides by adding drops of clean water. Do not tip offthe stain and then wash, as this will leave a deposit of scum over the smears.

6. Place the slides in the slide rack to drain and dry, film side downwards; makingsure that the film does not touch the slide rack.

Microscopic examination Examine the slide under the microscope using the X100 objective. Malaria parasitesfound in the blood are at different stages of development somemalaria parasites have granules of pigments in their cytoplasm.

Thin blood films: In thin blood films, the infected erythrocytes may remain unchanged or have adifferent colors or shape, or may contain pink (“Schüffner’s”) or red (“James”) thin films can be used to identify the species of malariaparasite

183 | Page PHCPI Laboratory Standards Guideline

Note: In patients who have been suffering from malaria for a long time, monocytesmay be seen in the thin blood film; the cytoplasm often contains brown orgreenishblackbodies (siderophils). In patients who have recently received an injection of anantimalarial drug, the parasites stain poorly and appear distorted and indistinct.

Thick blood films: In thick blood films, the background should be clean and free from debris, as theinfected erythrocytes are lysed. The malaria parasites should have deep red chromatinand blue or pale purplish-blue cytoplasm. In thick films stained with Giemsa, thenuclei of leukocytes should be stained dark purple. Schüffner’s dots may be seenaround the malaria parasites.Thick blood films are used for estimating the parasite density, as described below (see figures 30, 31, 32).

Reporting results

If the result of the examination of the stained blood films is positive, specify:  the species of parasite found  the stage of development of the parasite  The parasite density. Blood films containing P. ovale and P. vivax may contain few parasites and thereforetake more time to examine under the microscope. However, it is necessary to differentiatethe two species, since they may reappear in the blood without reinfection. Patients infected with P. ovale or P. vivax require additional treatment to eradicatethe liver stages of these parasites. A patient can harbour more than one species of malaria parasite at the same time (e.g. P. falciparum and P. malariae or P. falciparum and P. vivax). If the result is negative, report as “no parasites found”.

184 | Page PHCPI Laboratory Standards Guideline

Figure 34: P. Falciparum 1: Normal red cell, 2-18: Trophozoites (among these) , 2-10:correspond to ring stage trophozoites , 19-26:Schizonts (26 is ruptured schizont), 27&28: Mature macrogametocytes , 29&30:Mature microgametocytes (male)

185 | Page PHCPI Laboratory Standards Guideline

Figure 35:P. Ovale 1: Normal red cell , 2-5: Young trophozoites (Rings) , 6-15:Trophozoites , 16- 23:Schizonts , 24:Macrogametocytes( female) , 25: Microgametocytes (male)

186 | Page PHCPI Laboratory Standards Guideline

Figure 36: P. Vivax 1: Normal red cell , 2-6: Young trophozoites (Ringstage parasites) , 7- 18:Trophozoites , 19-27:Schizonts , 28&29:Macrogametocytes( female) , 30: Microgametocytes (male)

187 | Page PHCPI Laboratory Standards Guideline

8. Biochemistry

8.1 Blood glucose Glucose is the chief source of energy in the body. The levels of this compound are balanced by digestion and absorption of carbohydrates in the intestine, its storage and release in the liver and its utilization in the muscle. Causes of Hyperglycemia:  Insulin – dependent type (IDDM)  Non – insulin dependent type (NIDDM)  non obese NIDDM  obese NIDDM  Gestational diabetes  Impaired glucose tolerance: Glucose intolerance associated with certain conditions and syndrome such as (chronic pancreatitis with endocrine and exocrine insufficiency, Cushing syndrome , acromegaly ) Causes of hypoglycemia:  Deficient glucose production  Ketotic hypoglycemia  Glycogen storage disorder  Deficiency of gluconeogenic enzyme  Galactosemia  Hereditary fructose intolerance  Leucine hypersensitivity  Endogenous hyperinsulinism  insulinoma

Glucose is determined after enzymatic oxidation in the presence of glucose oxidase. The hydrogen peroxide formed reacts, under catalysis of peroxidase, with phenol and 4- aminophenazone to form a red-violet quinoneimine dye. The intensity of the colour produced is directly proportional to the concentration of glucose in the sample. Glucose + O2+H2O ------GOD H2O2 + Gluconate

2H2O2+ Phenol+ 4-Amino-antipyrine H2O+ Quinonimine

188 | Page PHCPI Laboratory Standards Guideline

Specimen Serum, heparin plasma, or fluoride plasma may be used. Plasma or serum samples without preservatives should be separated from the cells or clot within a half hour of being drawn. Glucose in separated, un-hemolyzed serum is stable up to four hours at 25°C and up to 24 hours at 4°C.

Reagents and materials

1- Full automated auto analyzer or Spectrophotometer 2- Centrifuge 3- Gel tube (clot activator) or plain tube. 4- Spectrophotometer 5- Deionized water 6- Pipettes 7- Glucose standards 20 mg/dL 100 mg/dL, 200 mg/dL, 400 mg/dL 8- Heating block or water bath 37°C 9- Timer 10- Reagents :

Reagent 1 tris buffer pH 7 100 mmol/l Buffer solution Phenol 0.3 mmol Reagent 2 glucose oxidase 10 000 u/l Peroxidase 1 000 u/l 4-amino-antipyrine 2.6 mmol/l Reagent 3 standard glucose 100 mg/dl

Preparation: Dissolve the powder R2 with the Buffer R1

Procedure

 Wave length ………… 505 nm ( 490 – 550)  Temperature………….37°C

Blank Standard Sample Standard - 10 µl - Sample - - 10µl Working reagent 1 ml 1ml 1ml

189 | Page PHCPI Laboratory Standards Guideline

 Mix incubate 10 minutes at 37 °C or 30 minutes at room temperature ( 20 °C – 30 °C)  The color is stable 30 minutes  Linearity : up to 500 mg / dl

Calculation: Glucose concentration =O. D. sample / O.D. standard × concentration of standard. Quality Control Level one and level two serum controls are tested with each patient run. The level one control range is 70-85 mg/dL and the level two range is 271-306 mg/dL.

Refrence intervals The reference range for glucose is as follows:

Cord 45-96 mg/dL

Premature 20-60 mg/dL

Newborn 40-60 mg/dL

1 week 50-80 mg/dL

Child 60-100 mg/dL

Adult 74-100 mg/dL

>60 year 82-115 mg/dL

>90 year 75-121 mg/dL

Limitations 1- Serum and plasma must be separated from the red blood cells promptly to prevent glycolysis. Glucose will decrease approximately 7% per hour when left in contact with red cells. 2- Whole blood glucose is 12-15% less than serum glucose. 3- Venous blood glucose is approximately 5 mg/dL less than arterial or capillary blood glucose.

190 | Page PHCPI Laboratory Standards Guideline

8.2 Liver Function Test

8.2.1 Alanine Aminotransferase (ALT) Is found mainly in the liver, but also in smaller amounts in the kidneys , heart , muscles, and . ALT is measured to see if the liver is damaged or diseased. Low levels of ALT are normally found in the blood. But when the liver is damaged or diseased, it releases ALT into the blood stream, which makes ALT levels go up. Most increases in ALT levels are caused by liver damage. The ALT test is often done along with other tests that check for liver damage, including aspartate aminotransferase (AST), alkaline phosphatase, lactate dehydrogenase (LDH), and bilirubin. Both ALT and AST levels are reliable tests for liver damage. Causes of high ALT:  Alcohol abuse  Medications  Acute and Chronic hepatitis Principle: Kinetic determination of ALT activity, method recommended by IFCC 2-oxoglutarate +l-alanine-----ALT------ glutamate + pyruvate Pyruvate + NADH + H+------LDH------ l-lactate + NAD+

The rate of NADH consumption is determined photometrically and is direct proportional to the ALT activity in the sample

Specimen A. Recommended specimen: 100 μl of serum or heparinised plasma or EDTAplasma. Collect specimens by standard venipuncture technique. Heparin may beused as an anticoagulant for plasma specimens. Handle specimens in stopperedcontainers to avoid contamination and evaporation. Follow universal precautionswhen performing phlebotomy or handling patient specimens, calibrators, or otherserum-based products. Discard contaminated materials with infectious waste. B. Remove serum promptly from the clot. Haemolysed specimens should not be usedbecause ALT contamination from red cells will occur. DO NOT freeze thesample; this will cause a loss of ALT activity.

191 | Page PHCPI Laboratory Standards Guideline

C. If concentration is greater than the analyser range, dilute with an equal volume ofisotonic saline and re-analyses. Multiply the result by 2 to obtain the original ALTactivity.

Reagent and materials A. full automated auto analyzer or spectrophotometer B. Centrifuge C. Gel tube (clot activator) or plain tube. D. Reagents :

Reagent 1 tris buffer P H 7.5 100mmol/l Buffer reagent L-alalnine 500 mmol/l Reagent 2 NADH 0.18 mmol/l Substrate LDH 1200u/l Oxoglutarate 15 mmol/l

Preparation: Dissolve one vial of reagent 2 with 3 ml of R 1 buffer, this working reagent stable 13 days at 2 – 8 °C or 24 hours at 20 – 25 °C. Procedure:

 Wave length ------340 nm  Working reagent 1 ml mixed with sample 100 µl and wait 1 minute  Measure the extinction decrease per minute for 1 – 3 minutes Calculation: Δ OD/min * 1750 = u/l Quality Control 1. Run quality control materials in normal and abnormal ranges (low and high controls) at the beginning of each day. Always run a set of controls after calibrating the analyser. If the controls are out of range, do not report patient results until the problem is resolved and the patient samples are repeated with quality control samples within acceptable range. 2. Variability is expressed as standard deviation (SD) and coefficient of variation CV) and plotted on a Levy-Jennings graph.

192 | Page PHCPI Laboratory Standards Guideline

Reference value:

Women up to 31 u/l at 37 °C. Men up to 40 u/l at 37 °C.

Results 1- Panic values: are double or triple to the normal value. 2- Be careful. Reference value for the end point procedure is differed from the reference value for the kinetic procedure.

Limitations A. High total protein samples that are predominantly gamma globulins can increase ALT results. The sample should be diluted with an equal volume of isotonic saline and then reanalysed. B. In sera with very high activities, the initial absorbance may be very low because most of the NADH may have been consumed before the first readings. In this case, return the sample after dilution as described above.

Colorimetric Method (Reitman and Frankle) for determination of serum alanine aminotransferase (ALT or GPT)

Principle:

α-oxogluterate +l-alanine ------GPT----- l-glutamate +pyruvate

Alanine aminotransferase is measured by monitoring the concentration of pyruvate hydrazone formed with 2, 4 – dinitrophenyl-hydrazine (2, 4-DNP).

Reagent composition

Contents initial concentration of solution

R1 buffer

Phosphate buffer 100 mmol/l, ph 7.4

L-alanine 200 mmol/l

α-oxogluterate 2 mmol/l

R2 2, 4-dinitrophenylhydrazine 2 mmol/l

R3 sodium hydroxide 4 mol/l

193 | Page PHCPI Laboratory Standards Guideline

CAL. Pyruvate standard

Procedure:

Wavelength 546 nm (530 – 550 nm)

Cuvette 1 cm light path

Incubation temperature 37 ċ

1 – Measurement against reagent blank

Pipette into test tubes:

Reagent blank sample

Sample ------0.1 ml

Buffer (R1) 0.5ml 0.5 ml

Deionized water 1 ml -----

Mix, incubate for exactly 30 minutes at 37 ċ

2, 4-DNP (R2) 0.5 ml 0.5 ml

Mix, allow to stand for exactly 20 minutes at 20 to 25 ċ

Sodium hydroxide (R3) 5 ml 5 ml

Mix, read the absorbance of sample against the reagent blank after 5 minutes.

2- Measurement against sample blank (Ecteric sample only) :

Pipette into test tubes:

Sample blank sample

Sample ------0.1 ml

Buffer (R1) 0.5ml 0.5 ml

Mix, incubate for exactly 30 minutes at 37 ċ

2,4-DNP (R2) 0.5 ml 0.5 ml

Sample 0.1 ml ----

194 | Page PHCPI Laboratory Standards Guideline

Mix, allow to stand for exactly 20 minutes at 20 to 25 ċ

Sodium hydroxide (R3) 5 ml 5 ml

Mix, read the absorbance of sample against the sample blank after 5 minutes.

Standard curve:

Wavelength 490 – 560 nm

Cuvette 1 cm light path

Temperature 20 – 25 ċ

Pipette into test tubes:

Tube no. Pyruvate standard (ml) deionized water (ml) buffer solution (ml)

1 0.00 0.2 1.00

2 0.05 0.2 0.95

3 0.10 0.2 0.90

4 0.15 0.2 0.85

5 0.20 0.2 0.80

6 0.25 0.2 0.75

7 0.30 0.2 0.70

8 0.35 0.2 0.65

9 0.40 0.2 0.60

10 0.45 0.2 0.55

Mix and pipette into each tube 1 ml of reagent solution R2. Mix and incubate for 20 minutes at 20 to 25 ċ. Add 10 ml of sodium hydroxide to each group. Mix and read absorbance against blank (tube no. 1) after 5 minutes.

The absorbance of the increasing amounts of puruvate (0.05 – 0.45 ml pyruvate standard) correspond to the following transaminase activities in U / L.

195 | Page PHCPI Laboratory Standards Guideline

Tube no. GPT (U / L)

2 9

3 18

4 27

5 37

6 46

7 56

8 67

9 77

10 87

The standard curve is obtained by plotting the measured absorbance against the transaminase activities in U/L.

Normal value: serum up to 12 U/L

8.2.2 Aspartate Aminotransferase (AST) AST (Aspartate transaminase): AST is normally found in red blood cells, liver, heart, muscle tissue, pancreas, and kidneys. AST formerly was called serum glutamic oxaloacetic transaminase. Low levels of AST are normally found in the blood. When body tissue or an organ such as the heart or liver is diseased or damaged, additional AST is released into the bloodstream. The amount of AST in the blood is directly related to the extent of the tissue damage. After severe damage, AST levels rise in 6 to 10 hours and remain high for about 4 days. The AST test may be done at the same time as a test for alanine aminotransferase, or ALT. The ratio of AST to ALT sometimes can help determine whether the liver or another organ has been damaged. Both ALT and AST levels can test for liver damage. Causes of increase AST:  Alcohol abuse .  viral hepatitis,  acute MI

196 | Page PHCPI Laboratory Standards Guideline

Principle: Kinetic determination of ALT activity, method recommended by IFCC 2-oxoglutarate +l-aspartate-----AST------ glutamate + oxaloacetate Oxaloacetate + NADH + H+------MDH------ malate + NAD+ The rate of NADH consumption is determined photometrically and is direct proportional to the ALT activity in the sample. Specimen A. Recommended specimen: 100 μl of serum or heparinised plasma or EDTA plasma. Collect specimens by standard venipuncture technique. Heparin may be used as an anticoagulant for plasma specimens. Handle specimens in stoppered containers to avoid contamination and evaporation. Follow universal precautionswhen performing phlebotomy or handling patient specimens, calibrators, or otherserum-based products. Discard contaminated materials with infectious waste. B. Remove serum promptly from the clot. Haemolysed specimens should not be used because AST contamination from red blood cells will occur. Refrigerate specimens if analysis is not performed immediately. Freeze specimens if analysis is not performed within 48 hours. C. If concentration is greater than the analyser range, dilute with an equal volume of isotonic saline and reanalyse. Multiply the result by 2 to obtain the original AST activity.

Reagents and materials 1- full automated auto analyzer or spectrophotometer 2- Gel tube (clot activator) or plain tube. 3- Centrifuge 4- Reagents :

Reagent 1 tris buffer P H 7.8 80mmol/l Buffer reagent L-aspartate 200 mmol/l Reagent 2 NADH 0.18 mmol/l Substrate LDH 800u/l Oxoglutarate 12 mmol/l MDH 600 u/l

Preparation: Dissolve one vial of reagent 2 with 3 ml of R 1 buffer, this working reagent stable 21 days at 2 – 8 °C or 24 hours at 20 – 25 °C.

197 | Page PHCPI Laboratory Standards Guideline

Procedure:

 Wave length ------340 nm  Working reagent 1 ml mixed with sample 100 µl and wait 1 minute  Measure the extinction decrease per minute for 1 – 3 minutes Calculation: Δ OD/min * 1750 = u/l Reference value:

 Women up to 31 u/l at 37°C  Men up to 38 u/l at 37°C Quality Control A. Run quality control materials in normal and abnormal ranges (low and high controls) at the beginning of each shift. Always run a set of controls after calibrating the analyser. If the controls are out of range, do not report patient results until the problem is resolved and the patient samples are repeated with quality control samples within acceptable range. B. Variability is expressed as standard deviation (SD) and coefficient of variation (CV) and plotted on a Levy-Jennings graph.

Limitations A. High total protein samples that are predominantly gamma globulins can increase AST results. The sample should be diluted with an equal volume of isotonic saline and reanalysed. Highly active sera can have a very low initial absorbance because much of the NADH is already consumed prior to measurement. In such cases, dilute sample as specified. B. Analyser range: ≤ 150 U/L

Colorometric Method (Reitman and Frankle) for determination of serum aspartate aminotransferase (AST or GOT):

Principle:

α-oxogluterate +l-aspartate ------GOT----- l-glutamate +oxaloacetate

Aspartate aminotransferase is measured by monitoring the concentration of oxaloacetate hydrazone formed with 2, 4 – dinitrophenyl-hydrazine (2, 4-DNP).

198 | Page PHCPI Laboratory Standards Guideline

Reagent composition

Contents initial concentration of solution

R1 buffer

Phosphate buffer 100 mmol/l, ph 7.4

L-aspartate 100 mmol/l

α-oxogluterate 2 mmol/l

R2 2, 4-dinitrophenylhydrazine 2 mmol/l

R3 sodium hydroxide 4 mol/l

R4 Pyruvate standard

Procedure:

Wavelength 546 nm (530 – 550 nm)

Cuvette 1 cm light path

Incubation temperature 37 ċ

1 – Measurement against reagent blank

Pipette into test tubes:

Reagent blank sample

Sample ------0.1 ml

Buffer (R1) 0.5ml 0.5 ml

De ionized water 1 ml -----

Mix, incubate for exactly 30 minutes at 37 ċ

2, 4-DNP (R2) 0.5 ml 0.5 ml

Mix, allow to stand for exactly 20 minutes at 20 to 25 ċ

199 | Page PHCPI Laboratory Standards Guideline

Sodium hydroxide (R3) 5 ml 5 ml

Mix, read the absorbance of sample against the reagent blank after 5 minutes.

2- Measurement against sample blank (Ecteric sample only):

Pipette into test tubes:

Sample blank sample

Sample ------0.1 ml

Buffer (R1) 0.5ml 0.5 ml

Mix, incubate for exactly 30 minutes at 37 ċ

2, 4-DNP (R2) 0.5 ml 0.5 ml

Sample 0.1 ml ----

Mix, allow to stand for exactly 20 minutes at 20 to 25 ċ

Sodium hydroxide (R3) 5 ml 5 ml

Mix, read the absorbance of sample against the sample blank after 5 minutes.

Standard curve:

Wavelength 490 – 560 nm

Cuvette 1 cm light path

Temperature 20 – 25 ċ

Pipette into test tubes:

Tube no. Pyruvate standard (ml) deionized water (ml) buffer solution (ml)

1 0.00 0.2 1.00

2 0.05 0.2 0.95

3 0.10 0.2 0.90

4 0.15 0.2 0.85

200 | Page PHCPI Laboratory Standards Guideline

5 0.20 0.2 0.80

6 0.25 0.2 0.75

7 0.30 0.2 0.70

8 0.35 0.2 0.65

9 0.40 0.2 0.60

10 0.45 0.2 0.55

Mix and pipette into each tube 1 ml of reagent solution R2. Mix and incubate for 20 minutes at 20 to 25 ċ. Add 10 ml of sodium hydroxide to each group. Mix and read absorbance against blank (tube no. 1) after 5 minutes.

The absorbance of the increasing amounts of puruvate (0.05 – 0.45 ml pyruvate standard) correspond to the following transaminase activities in U / L.

Tube no. GPT U / L

2 6

3 11

4 16

5 20

6 25

7 31

8 37

9 44

10 52

The standard curve is obtained by plotting the measured absorbance against the transaminase activities in U/L.

Note: to construct a standard curve for AST dilute 1.5 ml of pyruvate standard with 4.5 ml ml buffer (R1) immediately before measurement.

Normal value: serum up to 12 U/L

201 | Page PHCPI Laboratory Standards Guideline

8.2.3 Alkaline Phosphatase Alkaline phosphatase is an enzyme used by the cells to speed up the time of a reaction. It is found in the bone, placenta, kidneys, small intestines and liver, but is primarily associated with liver disease and the obstruction of bile flow. Within the liver, alkaline phosphatase is located near or within the tubes that collect bile. Principle: Phenyl phosphate------ALP------ phenol + phosphate Free phenol liberate by hydrolysis of substrate reacts then with 4-amino-antipyrine in the presence of alkaline. Potassium ferricyanide to form a red – colored complex which absorbance measured at 510 nm is directly proportional to ALP activity. Sodium arsenate incorporated in the reagent abolishes further enzyme activity and prevent the dilution of the color inherent in earlier methods. Specimen Recommended specimen: 20 μl of serum or heparinised plasma. Collectspecimens by standard venipuncture technique. Handle specimens in stoppered containers to avoid contamination and evaporation. Follow universal precautionswhen performing phlebotomy or handling patient specimens, calibrators, or otherserum-based products. Discard contaminated materials with infectious waste.

Haemolysed specimens should not be used and samples more than 7 days old should not be used due to loss of the enzyme activity in 7 days.

If absorbance change per minute exceeds 0.250, dilute 0.1 ml of the sample with 0.5 ml of normal saline and repeat the assay using this dilution. Multiply the results by 6 to obtain the alkaline phosphatase activity.

Reagents and materials A. Full automated autoanalyzer or spectrophotometer. B. Gel tube (clot activator) or plain tube. C. Centrifuge D. Reagents :

202 | Page PHCPI Laboratory Standards Guideline

reagent 1 substrate buffer disodium phenylphosphate 5 mmol/l carbonate-biocarbonate buffer 50 mmol/l P H 10 stabilizer Reagent 2 phenol 20u/l Standard Reagent 3 4- aminoantipyrine 60 mmol/l Blocking reagent sodium arsenate 240 g/l Reagent 4 potassium ferricyanide 150 mmol/l Dye reagent

Calibration A. Empty and wash flow-through cuvette by pressing [WASH] on the LCD display. An automatic return to the select menu will occur. B. Push sip by pressing lever [P] to drain the tube system. Do not sip any solution. C. Measure exactly 1,000 μl of distilled water into a sample cup and sip distilled water by pushing sip-press lever [P]. The sipping volume of 1,000 μl is automatically measured in the flow-through cuvette. D. The length of transportation of the air and water will be displayed on the screen. E. Press [↵] to finish calibration. The program will return to method select menu.

Quality Control A. Run quality control materials in normal and abnormal ranges (low and high controls) at the beginning of each shift. Always run a set of controls after calibrating the analyser. If the controls are out of range, do not report patient results until the problem is resolved and the patient samples are repeated with quality control samples within acceptable range. B. Variability is expressed as standard deviation (SD) and coefficient of variation (CV) and plotted on a Levy-Jennings graph.

203 | Page PHCPI Laboratory Standards Guideline

Procedure

Serum blank Serum sample Standard Regent blank R1 2 ml 2ml 2ml 2ml Incubate for 5 minutes at 37 ċ Serum - 50 µl - - R2 - - 50 µl - Incubate for 15 minutes at 37 ċ R3 0.5 ml 0.5 ml 0.5 ml 0.5 ml Mix well R4 0.5 ml 0.5 ml 0.5 ml 0.5 ml Serum 0.5µl - - - D.W. - - - 50 µL

Mix well wait for 10 minutes in dark place, read at 510 nm against blank. Calculation: Conc.of ALP = O.D of serum / O.D. of standard * concentration of standard Linearity: up to 285 u /l Reference value: Children 17 – 142 u / l Adult 21 – 92 u /l Conv F.: kau/dl * 7.09 = u/l

8.2.4 Total Serum Bilirubin Bilirubin is formed from the haem fragment of haemoglobin released by damaged or aged RBC. Bonemarrow, liver and spleen are the sites of bilirubin production. Causes for an increase in total bilirubin may be divided into 3 categories:  Pre- hepatic: resulting from various haemolytic states;  Hepatic: resulting fromhepatitis, cirrhosis, and other causes of hepatic necrosis; \  Post-hepatic: resulting from an obstruction of the common bile or hepatic duct. Principle: Sulfanilic acid react with sodium nitrate to form diazotized sulfanilic acid in the presence of dimethyl sulfide. Total bilirubin react with dizotized sulfanilic acid formazobilirubin.

204 | Page PHCPI Laboratory Standards Guideline

Specimen A. Recommended specimen is 100 μl of serum or heparinised plasma. Collect specimens by standard venipuncture technique. Handle specimens in stoppered containers to avoid contamination and evaporation. Follow universal precautionswhen performing phlebotomy or handling patient specimens, calibrators, or otherserum-based products. Discard contaminated materials with infectious waste. B. Protect specimens from light and analyse as soon as possible after collection. Direct exposure to sunlight is reported to cause as much as 50% loss of bilirubin in one hour, especially when the specimen is kept in a capillary tube. Exposure to normal room light can result in a significant loss of serum bilirubin after 2 to 3 hours. C. If analysis is not performed immediately, samples can be refrigerated for up to 24 hours at (2–5) °C. D. Samples that come in contact with alcohol from sterile wipes may become haemolysed, which will increase the value of bilirubin. E. If concentration is greater than the analyser range, dilute with 5 volumes of normal saline and reanalyse. Multiply the result by 6 to obtain the original total bilirubin concentration. The results may show a positive bias of up to 30%.

Reagents and materials A. Spectrophotometer and bilirubimnometer for neonate. B. Centrifuge C. Gel tube (clot activator) or plain tube. D. Reagents : Reagent 1 sulfanilic acid 30 mmol/l Hydrochloric acid 150 mmol/l Dimethylsulfoxide 7 mol/l Reagent 3 sodium nitrate 20 mmol/l Reagent 4 Standard

Procedure: Working reagent: Mix Reagent 1 20 volumes Reagent 3 1 volume Stability in absence of light

205 | Page PHCPI Laboratory Standards Guideline

6 hours at 20-25 °C 2 days at + 4 °C Wave length ------555nm Temperature ------37°CRead against R1 sample blank Read against R1 + R3 (working reagent) sample and standard Calibration A. Empty and wash flow-through cuvette by pressing [WASH] on the LCD display. An automatic return to the select menu will occur. B. Push sip by pressing lever [P] to drain the tube system. Do not sip any solution. C. Measure exactly 1,000 μl of distilled water into a sample cup and sip distilled water by pushing sip-press lever [P]. The sipping volume of 1,000 μl is automatically measured in the flow-through cuvette. D. The length of transportation of the air and water will be displayed on the screen. E. Press [↵] to finish calibration. The program will return to method select menu.

Quality Control A. Run quality control materials in normal and abnormal ranges (low and high controls) at the beginning of each shift. Always run a set of controls after calibrating the analyser. If the controls are out of range, do not report patient results until the problem is resolved and the patient samples are repeated with quality control samples within acceptable range. B. Variability is expressed as standard deviation (SD) and coefficient of variation (CV) and plotted on a Levy-Jennings graph.

Procedure Calculations

Sample blank Sample Standard blank Standard Sample 50 µl 50 µl - - Calibrator(R4) - - 50 µl 50 µl Reagent(R1) 1 ml - 1 ml - Working - 1 ml - 1 ml reagent

206 | Page PHCPI Laboratory Standards Guideline

Mix and read the extinction after incubation for 5 minutes at 37°C Stability of color is 1 hour (ovoid direct sunlight)

Calculation: (O.D sample – O. D. sample blank) / (O. D. calibrator – O. D. calibrator blank) * n N = calibrator concentration Linearity: linear up to 20 mg / dl, 340 µmol/l

Reference value:

 Serum total bilirubin: < 1 mg/dl, < 17 µmol / l  Conv. F. : mg / dl * 17.1 = µmol/l

Limitations A. Specimens from haemodialysis patients should not be analysed for total bilirubin. B. Haemoglobin affects bilirubin results. C. Compounds that discolour serum, such as 4-aminosalicylic acid andphenazopyridine, may falsely increase bilirubin results. D. Results for predominantly unconjugated bilirubin (e.g., for neonates) may be up to 10% higher than the reference method.

8.3 Renal Function test

8.3.1 Blood Urea The liver produces urea in the urea cycle as a waste product of the digestion of protein , Elevated levels of blood urea are most commonly associated with renal disease, but may also result from dehydration, a high-protein diet, excess destruction ofbody proteins, and gastrointestinal diseases, especially with intestinal obstruction. Principle:

Urea + H2O ------urease------ CO2 + 2NH3

Salicylate and hypochlorite in the reagent react with the ammonium ions to form 2.2dicarboxy –indophenol. The quantity of green compound is proportional to the urea concentration.

207 | Page PHCPI Laboratory Standards Guideline

Specimen A. Recommended specimen is 10 μl of serum or EDTA plasma. Collect specimens by standard venipuncture technique. Heparin may be used as an anticoagulant for plasma specimens. Handle specimens in stoppered containers to avoid contamination and evaporation. Follow universal precautions when performingphlebotomy or handling patient specimens, calibrators, or other serum-basedproducts. Discard contaminated materials with infectious waste. B. Refrigerate specimens up to 3 days at 4°C if analysis is not performedimmediately. C. If concentration is greater than the analyser range, dilute with 9 volumes of isotonic saline or distilled water and reanalyse. Multiply the result by 10 to obtain the original urea nitrogen concentration.

Reagents and materials A. full automated autoanalyzer or spectrophotometer B. Gel tube (clot activator) or plain tube. C. Centrifuge D. Reagent :

R1a urease >5000 u/l R1b phosphate buffer 120 mmol/l p H 7 Sodium salicylate 63.4 mmol/l Sodium nitroprusside 5 mmol/l EDTA 1.5 mmol/l R2 sodium hypochloride 18 mmol/l Sodium hydroxide 750 mmol/l CAL standard 8.05 mmol/l

Preparation:

Add 1 vial of of urease R1a to one bottle of phosphate buffer R1b

Procedure:

Wave length …….600 nm

Temperature ……25 ċ – 37 ċ

208 | Page PHCPI Laboratory Standards Guideline

Calibration A. Empty and wash flow-through cuvette by pressing [WASH] on the LCD display. An automatic return to the select menu will occur. B. Push sip by pressing lever [P] to drain the tube system. Do not sip any solution. C. Measure exactly 1,000 μl of distilled water into a sample cup and sip distilled water by pushing sip-press lever [P]. The sipping volume of 1,000 μl is automatically measured in the flow-through cuvette. D. The length of transportation of the air and water will be displayed on the screen. E. Press [↵] to finish calibration. The program will return to method select menu.

Quality Control A. Run quality control materials in normal and abnormal ranges (low and high controls) at the beginning of each shift. Always run a set of controls after calibrating the analyser. If the controls are out of range, do not report patient results until the problem is resolved and the patient samples are repeated with quality control samples within acceptable range. B. Variability is expressed as standard deviation (SD) and coefficient of variation (CV) and plotted on a Levy-Jennings graph.

Procedure

Blank Standard sample standard - 10µl - sample - - 10µl Working reagent 1 ml 1 ml 1 ml Mixture incubated for at least 3 minutes at 37 ċ or 5 min. at 20-25 ċ

Sodium 200 µl 200 µl 200 µl hypochloride mix. Incubate for at least 5 min. at 37 ċ or 10 min. at 20-25 ċ , measurement is against reagent blank .

Calculation:

O.D.sample / O.D. standard ×concentration of standard

Calculations The results are reported in mmol/L. No further calculation is necessary unless thespecimen has been diluted.

209 | Page PHCPI Laboratory Standards Guideline

Conversion Urea = 2.14 × BUN BUN = 0.466 × Urea

Results A. Expected values: (urea) 1.7–8.3 mmol/L B. Panic values: <1.07 mmol/L and >37.1 mmol/L C. Each laboratory should confirm these values.

Limitations A. Do not use plasma collected with sodium fluoride. B. Specimens that contain haemoglobin increase urea nitrogen. Haemoglobin of 50 mg/dL (slightly haemolysed) increases BUN levels below 28 mg/dL by 1 mg/dL. C. Ammonium ions have shown an increase in urea nitrogen equivalent to their nitrogen content.

8.3.2 Serum Creatinine Creatinine measurement is used in the diagnosis and treatment of renal diseasesand in monitoring patients on renal dialysis. Analysis of creatinine activity in 100 μl of serum or plasma is based on an enzymatic method that produces a coloured product. A. The 100 μl of patient sample mixed with picrate in alkaline medium produces a coloured complex with creatinine. The rate of formation of the complex is then measured. B. The creatinine present in the sample reacts with the picrate at 37°C, and the rate of complex formation is measured and is proportional to the creatinine concentration.

Specimen A. Recommended specimen is 100 μl of serum or heparinized plasma. Collect specimens by standard venipuncture technique. Heparin may be used as an anticoagulant for plasma specimens. Handle specimens in stoppered containers to avoid contamination and evaporation. Follow universal precautions whenperforming phlebotomy or handling patient specimens, calibrators, or otherserum-based products. Discard contaminated infectious waste. B. Refrigerate specimens at 4°C for 24 hours if analysis is not done immediately.

Reagents amd materials A. Photometer or full automated analyzer.

210 | Page PHCPI Laboratory Standards Guideline

B. Gel tube (clot activator) or plain tube. C. Centrifuge D. Picric acid 26 mmol/L E. NaOH 1.6 mmol/L F. Creatinine 2 mg/dL or 176.8 μmol/L

Calibration A. Empty and wash flow-through cuvette by pressing [WASH] on the LCD display.An automatic return to the select menu will occur. B. Push sip by pressing lever [P] to drain the tube system. Do not sip any solution. C. Measure exactly 1,000 μl of distilled water into a sample cup and sip distilled water by pushing sip-press lever [P]. The sipping volume of 1,000 μl is automatically measured in the flow-through cuvette. D. The length of transportation of the air and water will be displayed on the screen. E. Press [↵] to finish calibration. The program will return to method select menu.

Quality Control A. Run quality control materials in normal and abnormal ranges (low and high controls) at the beginning of each shift. Always run a set of controls after calibrating the analyser. If the controls are out of range, do not report patient results until the problem is resolved and the patient samples are repeated with quality control samples within acceptable range. B. Variability is expressed as standard deviation (SD) and coefficient of variation (CV) and plotted on a Levy-Jennings graph.

Procedure A. Mix reagents according to manufacturer’s recommendations. B. Label 2 tubes: in tube or standard tube pipette 1,000 μl of reagents solution and 100 μl of creatinine standard reagent; and in tube 2 or sample tube pipette 1,000 μl of reagents solution and 100 μl of sample. C. Mix them well and measure them immediately as indicated. D. Press ZERO and then push sip by pressing lever [P] to sip water and drain the system. E. When the screen displays MEASURE STANDARD, place cuvette with the standard solution into cuvette holder and aspirate the standard solution by pushing lever [P]. F. When the screen displays MEASURE SAMPLE, place cuvette with the sample solution into cuvette holder and aspirate the sample solution by pushing lever [P].

211 | Page PHCPI Laboratory Standards Guideline

G. Press [RESULT] and await possible delay while measuring the sample solution. The result will be displayed on top line. H. Print or write out the results. Press [WASH] to sip rinse solution or press [MEASURE] to measure the next sample by pushing sip lever [P].

Calculations The results are reported in μmol/L. No further calculation is necessary unless thespecimen has been diluted.

Results A. Expected values:  Males: 53–97 μmol/L  Females: 44–80 μmol/L B. Panic values: < 27 μmol/L and > 663 μmol/L C. Each laboratory should confirm these values.

Limitations A. EDTA and fluoride/oxalate anticoagulants will cause low creatinine results. B. Lidocaine: Patients on long-term lidocaine therapy may show an increase of up to 1.0 mg/dL. Ninety percent of patients receiving intravenous lidocaine will show less than a 0.3 mg/dL increase. C. Proline: Patients receiving hyperalimentation fluid that contains proline may show an increase of up to 2.0 mg/dL. D. TRIS buffer (control fluid) causes an approximate 50% decrease in results. E. Dipyrone (Metamizol) at 40 mg/dL shows a –0.6 mg/dL bias at a creatinine concentration of 1.0 mg/dL. F. N-acetylcysteine: Patients receiving N-acetylcysteine (Fluimucil, Mucomyst) intravenously have been reported to show a large negative bias.

8.3.3 Uric acid Uric acid is a chemical created when the body breaks down substances called purines. Purines are found in some foods and drinks, most uric acid dissolves in blood and travels to the kidneys, where it passes out in urine. High levels of uric acid in the body is called hyperuricemia.

Principle:

Uric acid +O2 +H2O------uricase ------ allantoine + CO2 + H2O2

212 | Page PHCPI Laboratory Standards Guideline

2H2O2 +4-aminophenazone+ 2-4dichloro- phenol-sulfonate ------peroxidase------ quinonemine

Reagents:

Reagent 1 phosphate buffer P H 7.4 50mmol/l

Buffer solution 2-4 DCPS 4 mmol/l

Reagent 2 uricase 7o u/l

Vial of enzyme peroxidase 660 u/l

4-aminophenazone 1 mmol/l

Reagent 3 uric acid 6 mg/dl

Standard

Preparation:

Dissolve the content of one bottle R 2 with the content of one bottle buffer R 1

Procedure:

Wave length------510 nm (490 – 550)

Blank Standard Sample

Working reagent 1 ml 1 ml 1 ml

Standard - 20µl -

Sample - - 20 µl

mix. Incubate 5 minutes at 37 °C or 10 minutes at 20-25°C the colour is stable for 30 minutes.

213 | Page PHCPI Laboratory Standards Guideline

Calculation:

Concentration= O.D.sample / O.D. standard * concentration of standard Referencevalue: women 2.5 – 6 mg/dl, 148 – 357 µmol / l Men 3.4 – 7 mg/dl, 200 – 416 µmol / l Conv. F.: mg/dl * 60 = µmol / l

8.4 Lipid Profile

8.4.1 Total Cholesterol Cholesterol is a waxy, fat-like substance that occurs naturally in all parts of the body. Your body needs some cholesterol to work properly. But if you have too much in your blood, it can stick to the walls of your arteries. This is called plaque. Plaque can narrow your arteries or even block them.

High levels of cholesterol in the blood can increase your risk of heart disease. Your cholesterol levels tend to rise as you get older. There are usually no signs or symptoms that you have high blood cholesterol, but it can be detected with a blood test. You are likely to have high cholesterol if members of your family have it, if you are overweight or if you eat a lot of fatty foods.

You can lower your cholesterol by exercising more and eating more fruits and vegetables. You also may need to take medicine to lower your cholesterol.

Causes of hypercholesterolemia: 1- Familial hypercholesterolaemia (FH) 2-Reduced metabolism due to thyroid problems (hypothyroidism) 3-kidney diseases 4-Diabetes particularly when poorly controlled 5-Alcohol abuse 6-being overweight – this is probably the commonest cause of high cholesterol levels.

Causes of hypocholesterolaemia: 1-Hyperthyroidism 2-Liver disease 3-Malnutrition 4-Cancer

214 | Page PHCPI Laboratory Standards Guideline

5-Chronic infections or inflammation

Principle:

Esters of cholesterol +H2O ------chloesterol esterase------> cholesterol + fatty acids Cholesterol +O2 ------cholesterol oxidase ------> cholest-4-en-one +H2O2 H2O2 +phenol+4-aminophenazone ------peroxidase ------> quinonimine

Specimen A. Recommended specimen 10 μl of serum, heparinised or EDTA plasma. Collect specimens by standard venipuncture technique. Heparin may be used as an anticoagulant for plasma specimens. Handle specimens in stoppered containers to avoid contamination and evaporation. Follow universal precautions whenperforming phlebotomy or handling patient specimens, calibrators, or otherserum-based products. Discard contaminated materials with infectious waste. B. If the result is over 19.3 mmol/L, dilute the sample before precipitation 1:2 with normal saline and multiply the results by 3 to obtain the original cholesterol concentration. C. Cholesterol remains stable in serum for 6 days at 4–25°C and up to 4 months at – 20°C. D. Lipemic specimens generate turbidity of the sample/reagent mixture, which leads to falsely elevated results.

Reagents and materials A. Spectrophotometer or full automated analyzer. B. Gel tube (clot activator) or plain tube. C. Centrifuge D. Reagents : Reagent 1 pipes p H 6.9 90 mmol/l Buffer solution Phenol 26 mmol/l Reagent 2 cholesterol oxidase 399 u/l Vial of two enzymes Peroxidase 1250 u/l Cholesterol oxidase 300 u/l 4-aminophenazone 0.4 mmol/l Reagent 3 standard 200 mg/dl

215 | Page PHCPI Laboratory Standards Guideline

Preparation:

 Dissolve the contents of one bottle R2 to the contents of one bottle buffer reagent R1

Sample: serum, heparinized plasma

Procedure:

 Wave length …………… 505nm ( 500 550)  Temperature …………... 37 ċ

Blank Standard sample standard - 10µl - sample - - 10µl Working reagent 1 ml 1 ml 1 ml  Mix. Incubate 5 minutes at 37 °C, the colour is stable for 30 min.

Calculation:

Cholesterol concentration = O.D.sample / O.D. standard * concentration of standard

Quality Control A. Run quality control materials in normal and abnormal ranges (low and high controls) at the beginning of each shift. Always run a set of controls after calibrating the analyser. If the controls are out of range, do not report patient results until the problem is resolved and the patient samples are repeated with quality control samples within acceptable range. B. Variability is expressed as standard deviation (SD) and coefficient of variation (CV) and plotted on a Levy-Jennings graph.

Calculations The results are reported in md/dl or in mmol/L. No further calculation is necessary unless thespecimen has been diluted.

Results A) Reference range : up to 200 mg / dl B) Each laboratory should confirm these value Limitations A. Bilirubin concentrations above 4 mg/dl interferes with tests B. If results are over 25.9mmol/l dilute the sample and repeat measument C. High concentration of ascorbic acid (> 2.5 mg/dL) will give lower values.

216 | Page PHCPI Laboratory Standards Guideline

D. Haemoglobin levels higher than 200 mg/dL do not interfere with the test.

8.4.2 Triglycerides Triglycerides protein is essential for growth, the production of new tissue, and therepair of injured tissue. An increase in triglycerides levels may be the result ofnephrosis, cholestasis, pancreatitis, cirrhosis, diabetes mellitus, and hepatitis. Adecrease is seen with malnutrition. The analysis is based on the enzymatic hydrolysis of triglycerides with lipases in aquinoneimine formed from hydrogen peroxide used as a coloured indicator. Theintensity of the colour is proportional to the amount of triglycerides in the sample.10 μl of patient sample is mixed and incubated with 1,000 μl of solution containing 4-aminoantipyrine and 4- chlorophenol in presence of hydrogen peroxide. Afterincubation for 10 minutes, a reaction between 4-aminoantipyrine and 4-chlorophenoland hydrogen peroxide takes place, forming quinoneimine, which is a colouredproduct. The amount of coloured compound formed is proportional to the amount oftriglycerides in the specimen.

Specimen A. Recommended specimen 10 μl of serum, heparinised or EDTA plasma. Collect specimens by standard venipuncture technique. Heparin may be used as an anticoagulant for plasma specimens. Handle specimens in stoppered containers to avoid contamination and evaporation. Follow universal precautions whenperforming phlebotomy or handling patient specimens, calibrators, or otherserum-based products. Discard contaminated materials with infectious waste. B. Avoid haemolysed specimens. If analysis is not performed immediately, freeze specimens at −25oC for up to 4 months or refrigerate at 2–8oC for up to 3 days. C. If concentration is greater than the analyser range, dilute 1 part of sample with 4 parts of isotonic saline or distilled water and reanalyse. Multiply the result by 5 to obtain the original triglycerides concentration.

Reagents and materials A. Photometer or full automated analyzer. B. Gel tube (clot activator) or plain tube. C. Centrifuge D. Reagent PIPES buffer (pH 7.5) 50 mmol/L  4-chlorophenol 5 mmol/L  4-aminoantipyrine 0.25 mmol/L  Magnesium ion 4.5 mmol/L  ATP 2 mmol/L

217 | Page PHCPI Laboratory Standards Guideline

 Lipases ≥1.3 U/ml  Peroxidase ≥0.5 U/ml  Glycerol kinase ≥0.4 U/ml  Glycerol-3-phpsphate oxidase ≥1.5 U/ml  Standard Triglyceride 2.28 mmol/L

Calibration A. Empty and wash flow-through cuvette by pressing [WASH] on the LCD display. An automatic return to the select menu will occur. B. Push sip by pressing lever [P] to drain the tube system. Do not sip any solution. C. Measure exactly 1,000 μl of distilled water into a sample cup and sip distilled water by pushing sip-press lever [P]. The sipping volume of 1,000 μl is automatically measured in the flow-through cuvette. D. The length of transportation of the air and water will be displayed on the screen. E. Press [↵] to finish calibration. The program will return to method select menu.

Quality Control A. Run quality control materials in normal and abnormal ranges (low and high controls) at the beginning of each shift. Always run a set of controls after calibrating the analyser. If the controls are out of range, do not report patient results until the problem is resolved and the patient samples are repeated with quality control samples within acceptable range. B. Variability is expressed as standard deviation (SD) and coefficient of variation (CV) and plotted on a Levy-Jennings graph.

Procedure A. Mix 2 tubes: one blank with 1,000 μl of standard reagent, and tube 2 with 10 μl of serum or plasma and 1,000 μl of reagent solution. B. Mix each well and incubate for 5 minutes at 37oC, and measure the absorbance of the standard reagent and sample immediately as indicated. C. Press ZERO and then push sip by pressing lever [P] to sip water and drain the system. D. When the screen displays MEASURE RB, place cuvette with the blank solution into cuvette holder and aspirate the blank solution by pushing lever [P]. E. When the screen displays MEASURE SAMPLE, place cuvette with the sample solution into cuvette holder and aspirate the sample solution by pushing lever [P].

218 | Page PHCPI Laboratory Standards Guideline

F. Press [RESULT] and await possible delay while measuring the sample solution. The result will be displayed on top line. G. Print or write out the results. Press [WASH] to sip rinse solution or press [MEASURE] to measure the next sample by pushing sip lever [P].

Calculations The results are reported in mmol/L. No further calculation is necessary unless thespecimen has been diluted.

Results A. Expected values: 0.11–2.15 mmol/L. B. Each laboratory should confirm these values.

Limitations A. Lipemic specimen usually generates turbidity of the sample reagent mixture, which leads to falsely elevated results. B. Ascorbate gives falsely low values.

8.4.3 HDL Cholesterol High -density lipoprotein (HDL): HDL particles are able to remove cholesterol from within artery atheroma and transport it back to the liver for excretion or re-utilization, which is the main reason why the cholesterol carried within HDL particles (HDL-C) is sometimes called “good cholesterol”. Those with higher levels of HDL-C seem to have fewer problems with cardiovascular disease, while those with low HDL-C cholesterol levels (less than 40 mg/dl or about 1mmol/L) have increased rates for heart disease).

Increased: High carbohydrate diet and hyper alpha-lipoproteinemia.

Decreased: hypertriglyceridemia, tangier disease, obstructive liver disease and severe malnutrition.

How to increase HDL level:

Diet and lifestyle: certain changes in life style may have a positive impact on raising HDL level:

 Aerobic exercise  Weight loss  Niacin (vitamin B3) supplementation  Smoking cessation  Removal of trans fatty acid from the diet

219 | Page PHCPI Laboratory Standards Guideline

 Addition of soluble fiber to diet  Consumption of omega-3 fatty acids such as fish oil or flax oil.

Principle

This technique uses a separation method based on the selective precipitation of VLDL, and LDL by phosphotungstic acid and MgCl, sedimentation of the precipitant by residual cholesterol remaining in the clear supernatant.

Specimen A. Recommended specimen 200 μl of serum, heparinised or EDTA plasma. Collect specimens by standard venipuncture technique. Heparin may be used as an anticoagulant for plasma specimens. Handle specimens in stoppered containers toavoid contamination and evaporation. Follow universal precautions whenperforming phlebotomy or handling patient specimens, calibrators, or other- based products. Discard contaminated materials with infectious waste. B. If the supernatant is not clear, dilute the sample before precipitation 1:1 with normal saline and multiply the results by 2 to obtain the original HDLconcentration. C. HDL cholesterol remains stable in serum for 6 days at 4–25°C and up to4 months at – 20°C. Serum must be separated from the blood clot as rapidly as possible.

Reagents

R1 phosphotungstic acid 0.63 mmol/l

Precipitating reagent MgCl 25 mmol/l

CAL: cholesterol standard 50 mg/dl

R2: cholesterol WR

Procedure:

1-precipitation:

Sample Standard Sample 0.2 ml 0.2 ml Precipitating agent 0.4 ml 0.4 ml

220 | Page PHCPI Laboratory Standards Guideline

Mix and wait for 10 minutes at room temperature. Centrifuge for 10 minutes at 4000 r.p.m Separate the clear supernatant within 2 hours.

2-Colorimetry:

Tube Blank Sample supernatant Standard supernatant Cholesterol WR 1.0 ml 1.0 ml 1.0 ml Supernatant - 50µl - Standard - - 50µl

Mix and wait for 10 minutes at room temperature. Read the absorbance of the supernatant and the standard at 500 nm against the reagent blank

Calculation:

O.D. supernatant / O.D. standard * concentration of standard

Expected values:

Level mg/dl Level mmol/l Interpretation <40 for men, <50 for <1.03 Low HDL cholesterol, heightened risk for women heart disease 40-59 1.03-1.55 Medium HDL level >60 > 1.55 High HDL level, optimal condition considered protective against heart disease.

Quality Control A. Run quality control materials in normal and abnormal ranges (low and high controls) at the beginning of each shift. Always run a set of controls after calibrating the analyser. If the controls are out of range, do not report patient results until the problem is resolved and the patient samples are repeated withquality control samples within acceptable range. B. Variability is expressed as standard deviation (SD) and coefficient of variation (CV) and plotted on a Levy-Jennings graph.

221 | Page PHCPI Laboratory Standards Guideline

Limitations A. High triglyceride level interferes with HDL measure; dilute the sample before precipitation 1:1 with 0.9% saline and multiply the results by 2. B. High concentration of ascorbic acid (> 2.5 mg/dL) will give lower values. C. Haemoglobin levels higher than 100 mg/dL and bilirubin levels higher than 10 mg/dL interfere with the test.

8.4.4 LDL Cholesterol Low density lipoprotein (LDL) is lipoprotein which is involved in the transport ofcholesterol to the cells. An increase in LDL levels may be the result of altered dietand vascular diseases. A decrease is seen in AIDS, haematological malignanciessuch as acute myeloid leukaemia (AML) or chronic myeloid leukaemia (CML), and disorders involving splenomegaly. The analysis is based on direct homogenous enzymatic assay for quantitativedetermination of LDL cholesterol combining two steps. The analysis involvesremoval of chylomicrons, very low density lipoprotein (VLDL), and high densitylipoprotein (HDL), followed by enzymatic determination of LDL in the sample. 10 μl of patient sample is mixed and incubated at 37oC for 5 minutes with 750 μl ofenzyme solution containing cholesterol esterase and cholesterol oxidase, which allowsthe removal of chylomicrons, VLDL, and HDL from the specimen by formation ofcholestenone and hydrogen peroxide. The LDL present in the sample is converted tocholestenone and hydrogen peroxide in the presence of cholesterol esterase, cholesterol oxidase, and specific surfactant. The chromogen in presence of hydrogenperoxide peroxidase is converted to a coloured dye (quinine) whose concentration ismeasured by colorimetric method.

Specimen A. Recommended specimen 10 μl of serum or plasma. Collect specimens by standard venipuncture technique. Heparin may be used as an anticoagulant for plasma specimens. Handle specimens in stoppered containers to avoid contamination and evaporation. Follow universal precautions when performing phlebotomy orhandling patient specimens, calibrators, or other serum-based products. Discardcontaminated materials with infectious waste. B. If the serum concentration of LDL exceeds the measuring range, dilute the sample before precipitation 1:1 with normal saline, repeat the test, and multiply the results by 2 to obtain the original LDL concentration. C. LDL cholesterol remains stable in serum for 6 days at 4–25°C and up to4 months at – 20°C.

222 | Page PHCPI Laboratory Standards Guideline

D. Serum must be separated from the blood clot as rapidly as possible.

Reagents and materials A. Photometer or full automated analyzer. B. Gel tube (clot activator) or plain tube C. Centrifuge D. Reagent (enzymes)  Good’s buffer (pH 7.0) 50 mmol/L  Cholesterol esterase 600 U/L  Cholesterol esterase 500 U/L  Catalase 600 kU/L  TOOS 2.0 mmol/L  Detergents 0.3% w/v  Preservatives<0.1% w/v  Substrate Peroxidase 4000 U/L  4-Aminoantipyrin 4 mmol/L  Good’s buffer (pH 7.0) 50 mmol/L  Sodium azide 0.05%  Detergents 1% w/v  Preservatives<0.1% w/v

Calibration A. Empty and wash flow-through cuvette by pressing [WASH] on the LCD display. An automatic return to the select menu will occur. B. Push sip by pressing lever [P] to drain the tube system. Do not sip any solution. C. Measure exactly 1,000 μl of distilled water into a sample cup and sip distilled water by pushing sip-press lever [P]. The sipping volume of 1,000 μl is automatically measured in the flow-through cuvette. The length of transportation of the air and water will be displayed on the screen. D. Press [↵] to finish calibration. The program will return to method select menu.

Quality Control A. Run quality control materials in normal and abnormal ranges (low and high controls) at the beginning of each shift. Always run a set of controls after calibrating the analyser. If the controls are out of range, do not report patient results until the problem is resolved and the patient samples are repeated with quality control samples within acceptable range.

223 | Page PHCPI Laboratory Standards Guideline

B. Variability is expressed as standard deviation (SD) and coefficient of variation (CV) and plotted on a Levy-Jennings graph.

Procedure A. Label 2 tubes and pipette 10 μl water and 750 μl of enzyme solution into tube 1 (blank), and into tube 2 pipette 10 μl of test sample and 750 μl of enzyme solution. Mix gently and incubate at 37oC for 5 minutes. B. Into each tube add 250 μl of substrate solution and mix them well and incubate for 5 minutes at 37°C, then measure absorbance of sample against reagent blank. C. When the screen displays MEASURE BLANK, press ZERO and then push sip by pressing lever [P] to sip water and drain the system. D. When the screen displays MEASURE RB, place cuvette with the blank solution into cuvette holder and aspirate the blank solution by pushing lever [P]. E. When the screen displays MEASURE SAMPLE, place cuvette with the sample solution into cuvette holder and aspirate the sample solution by pushing lever [P]. F. Press [RESULT] and await possible delay while measuring the sample solution. The result will be displayed on top line. G. Print or write out the results. Press [WASH] to sip rinse solution or press [MEASURE] to measure the next sample by pushing sip lever [P].

Calculations The results are reported in mmol/L. No further calculation is necessary unless thespecimen has been diluted.

Results A. Expected values: up to 130 mg/dl. B. Each laboratory should confirm these values.

NOTE: Direct measurement of LDL is required instead of the Friedwald, or“calculated”, method whenever triglyceride values exceed 400 mg/dL. The Friedwald or “calculated” method of LDL determination is determined by subtracting HDL and VLDL from the total cholesterol concentration. Total cholesterol = HDL + LDL + VLDL. VLDL is generally estimated by the triglyceride concentration divided by 5, but is inaccurate when triglyceride values exceed 400 mg/dL.

Limitations A. High triglyceride level interferes with LDL measure; dilute the sample before precipitation 1:1 with 0.9% saline and multiply the results by 2. B. High concentration of ascorbic acid (> 2.5 mg/dL) will give lower values.

224 | Page PHCPI Laboratory Standards Guideline

C. Haemoglobin levels higher than 100 mg/dL and bilirubin levels higher than 10 mg/dL interfere with the test.

225 | Page PHCPI Laboratory Standards Guideline

Annex 1

TABLE 1 - Outline of the properties of heat decontamination methods. For everyday laboratory purposes, autoclaving is the preferred method, unless the item cannot withstand the heat and/or moisture of autoclaving.

Principle/Conditions Advantages Disadvantages Uses

Dry Heat Thermal inactivation: Non- Less effective Materials that are destroys by oxidation corrosive than moist heat; damaged by, or Simple requires longer are impenetrable design and times and/or to, moist heat principle higher temperatures

Hot Air Oven  160-180?C for 2-4  penetrates  slow  anhydrous hours water- diffusion, materials, such as insoluble penetration oils, greases and materials powders  loading, (e.g., grease packing critical  laboratory and oil) to performance glassware,  less instruments  not suitable corrosive to for reusable  closed metals and plastics containers sharp instruments than steam

Red-heat  oxidation to ashes  rapid  initial contact  inoculating Flame (burning) with flame can loops, needles produce a viable aerosol

226 | Page PHCPI Laboratory Standards Guideline

 possibility of accidental fire

Incineration  oxidation to ashes  reduces  improper use  for (burning) volume of may lead to decontamination waste by up emission of of waste items  1-60 minutes: to 95% pathogens in prior to disposal temperatures may smoke in landfill exceed 1000?C  requires transport of infectious waste

 excess plastic (>20%) content reduces combustibility

Moist Heat Irreversible More rapid coagulation of and more (microbial) proteins effective than dry heat

Pasteurization  heating to below  can be  not reliably  milk and dairy boiling point used on heat sporicidal products (generally 77?C) for sensitive  some heat- up to 30 minutes liquids and sensitive medical medical equipment devices

 low cost

Tyndallization  heating to 80-100?C  resistant  time  heat sensitive (Fractional for 30 mins on spores consuming materials such as Sterilization) successive days, with germinate bacteriologic  not reliably and are media, solutions sporicidal killed on the of chemicals,

227 | Page PHCPI Laboratory Standards Guideline

incubation periods in second and biological between third days materials

Boiling  maximum  minimal  cumbersome:  small temperature obtainable equipment not practical for instruments and is approximately required everyday lab equipment 100?C 10-30 mins use

 not reliably sporicidal

autoclaving  steam under  minimal  loading and  penetration of pressure time packing critical sterile glassware, required to performance media and  121?C/15 psi for instruments 15-90 mins (gravity  most  shielding dirt displacement dependable must first be  autoclave) sterilant for removed decontamination lab use of reusable  132?C/27 psi for 4-  maintenance supplies and 20 minutes (pre- and quality equipment vacuum autoclave) control essential   damages decontamination heat-sensitive of infectious itmes waste

228 | Page PHCPI Laboratory Standards Guideline

TABLE 2 - Summary of concentrations used, contact times, advantages and disadvantages and uses of some of the halogen-releasing chemical germicides. The wide ranges of effective concentrations and contact times cited are due to a number of factors, including the interdependence of time and concentration, the variability in resistance of different microorganisms, the amount of organic material present and the desired effect (e.g., low- level vs high-level disinfection)

Effective Examples of Concentrations, Advantages Disadvantages Uses Contact Times

Chlorine  100-10,000  broad  toxic,  general Compounds: ppm (.01-1%) spectrum corrosive to disinfectant Sodium hypochlorite free chlorine skin and metals   waste liquids solution 1 (liquid  10-60 minutes inexpensive  unstable at bleach)  surface (>= 3,000 ppm optimum  widely decontamination for broad effective pH of available spectrum) 6  emergency  spill clean up  inactivated bactericidal by organic  instrument at low matter disinfection temperature  deteriorates under light and heat: shelf life of dilutions is less than 1 week

Calcium  as for liquid  as for  as for liquid  as for liquid hypochlorite2 bleach liquid bleach bleach above, bleach granules, powder, but more except shelf tablets stable life is longer

229 | Page PHCPI Laboratory Standards Guideline

NaDCC3 (Sodium  as for liquid  more  toxic,  as for liquid dichloroisocyanurate) bleach stable than corrosive bleach powder, granules, hypochlorites  inactivated tablets  stable at by organic pH 6.0 matter

Chloramine-T4  as for liquid  more  deteriorates  as for liquid (Sodium bleach stable, less under bleach tosylchloramide) affected by humidity, light powder or tablets organic and heat matter than hypochlorites

 longer activity than hypochlorites

Chlorine dioxide5  demand-  longer  aqueous  instrument release of activity than solutions disinfection chlorine dioxide other decompose  gas in situ chlorine under light sterilization of compounds germ-free  less animal corrosive, chambers less toxic than other chlorine compounds

 effective at pH 6-10

Iodine  30-1,000 ppm  broad  not  germicidal Preparations: (.003-.1%) free spectrum consistently soaps and Iodophors6 iodine sporicidal antiseptics

230 | Page PHCPI Laboratory Standards Guideline

 10-30 minutes  germicidal  efficacy  surface over a wide reduced by decontamination pH range organic matter  work surface  generally  some wipedown nonstaining, iodophor  instrument less toxic and solutions disinfection less irritating support growth than aqueous of or alcoholic Pseudomonas7 iodine solutions

231 | Page PHCPI Laboratory Standards Guideline

TABLE 3A Summary of recommended concentrations, contact times, advantages and disadvantages of non-halogen chemical germicides. The wide ranges of effective concentrations and contact times cited reflect the interdependence of time and concentration as well as factors such as resistance of the particular class or strain of target microorganism(s) and desired effect. Also, some germicides are available in combinations (e.g., glutaraldehyde/phenol or peracetic acid/alcohol mixtures) which are synergistic whereby the components in combination produce a greater antimicrobial effect than the sum of their individual effects.

Effective Examples of Concentrations Advantages Disadvantages Laboratory and Contact Uses Times

Alcohols  70-80%  low toxicity  rapid  skin ethanol evaporation disinfectant  rapid action limits contact (antiseptic)  60-95%  low residue time isopropanol  surface

 non-corrosive  flammable, decontamination  10-30 eye irratant minutes  benchtop,  may damage cabinet rubber, plastic, wipedown shellac

 ineffective against bacterial spores

Phenolic  400-50,000  tolerant of  pungent  instruments Compounds ppm (.05-1.5%) organic load, odour, and equipment "hard" dilution corrosive, disinfection water

232 | Page PHCPI Laboratory Standards Guideline

 10-30  leaves an some forms  disinfection minutes active residue toxic of floors and (may be other surfaces  not desirable on sporicidal;  antiseptic some surfaces) limited activity soaps and  against viruses lotions biodegradable  leaves a residual film (undesirable in culture systems)

 may support growth of bacteria1

Quaternary  500-15,000  combined  non  surface Ammonium ppm (.05-1.5%) detergent and sporicidal, decontamination Compounds germicidal limited activity  10-30  equipment activity against viruses, minutes wipedown mycobacteria  stable  antiseptic  most  working formulations formulations dilutions have available not readily low toxicity biodegradable  floors and walls  may support growth of bacteria2

Hydrogen Peroxide  3-30%  rapid action  limited  surface aqueous sporicidal decontamination  no residue solution activity  instruments  low toxicity and equipment

233 | Page PHCPI Laboratory Standards Guideline

 10-60   corrosive to minutes environmentally some metals safe  6% for 30  potentially minutes may explosive at kill spores high concentrations

 stock solutions irritating to skin and eyes

Peracetic Acid  .001-.3%  broad  pungent  instruments (PAA) aqueous spectrum odour and equipment solution  sporicidal at corrosive to  gas phase  gas phase: 2- low some metals sterilization of 4% temperatures chambers for  shelf life of germ-free  5-120  can tolerate dilutions is less animals minutes organic load than 1 week

 rapid action  stock solutions  nontoxic irritating to decomposition skin and eyes products  stock must  leaves no be protected residue from heat, light

 gas phase: respiratory irritant, fire hazard above 55?C

234 | Page PHCPI Laboratory Standards Guideline

Table 3B - Summary of recommended concentrations, contact times, advantages and disadvantages of non-halogen chemical germicides. The wide ranges of effective concentrations and contact times cited reflect the interdependence of time and concentration as well as factors such as resistance of the particular class or strain of target microorganism(s) and desired effect. Also, some germicides are available in combinations (e.g., glutaraldehyde/phenol or peracetic acid/alcohol mixtures) which are synergistic whereby the components in combination produce a greater antimicrobial effect than the sum of their individual effects.

Effective Examples of Concentrations Advantages Disadvantages Laboratory and Contact Uses Times

Aldehydes:

Glutaraldehyde  0.5-2.5%  broad  expensive  cold sterilant alkalinized spectrum and fixative  pH, aqueous  does not temperature  surface solution corrode dependent decontamination  2-30 mins; up metal  pungent  instruments, to 12 hours to  can odour equipment, kill all spores tolerate glassware  toxic: skin, organic load eye, respiratory tract irritant

 activated solutions have less than 2- week shelf life

235 | Page PHCPI Laboratory Standards Guideline

Formalin (37%  3-27%  broad  pungent  cold sterilant aqueous formalin (1- spectrum odour and fixative formaldehyde) 10%   skin, eye  surface formaldehyde) inexpensive and respiratory decontamination in 70-90% tract irritant alcohol  does not  instruments corrode  potential and equipment  10-30 metal carcinogen minutes (animal  can studies) tolerate organic load  may require 24 hrs or more to kill all spores

Formaldehyde (gas)  1-3 hours  as for  as for  on site formalin formulin decontamination of biological  effective  flammable safety cabinet penetration  porr HEPA filters penetration of  enclosed covered areas surfaces

Ethylene Oxide Gas  50-1200  broad  flammable,  heat or mg/L spectrum reactive moisture sensitive  1-12 hours  no heat or  toxic: supplies, moisture potential instruments and evolved carcinogen and equipment mutagen  penetrates  some packaging sterilized items materials may need more

236 | Page PHCPI Laboratory Standards Guideline

than 24 hours for outgassing

References: 1. JohnBernardHenry, Clin.diagnosis&managementbylabmethods,

19thedition, 1996.

2. Dacie and Lewis: PRACTICAL HAEMATOLOGY Tenth Edition.

3. Cedric Mims, Hazel M Dockrell, Richard V Goering etal medical Microbiology 3rd edition 2004 Mosby by Elsevier 4. Murray PR , Baron Ej , Pealler MA etal , editors Manual of clinical microbiology ed7,Washington, Dc 1999,ASM press 5. Rose NR, Friedman H., Manual of clinical immunology 2nd editor Washington, DC 1980 ASM press 6. Richard L Guerrant, David H walker and Peter F Weller Tropical infection disease principles, pathogens and practice Vol. 1 .1999 Churchill Livingstone. 7. Mandel GL. Bennett JE, Dolin R, editors, Principle and practice of infectious diseases ed5, Philadelphia 2000, Elsevier Churchill Livingstone.

8. Performance standard for antimicrobial susceptibility test M100-S22 VOL32, NO.3 ,2012 9. R.R Gillies, T.C DODDS bacteriological illustrated 1971 Churchill Livingstone.

10. Vandepitte J., Engbaek K., Rohner P., Piot P. And Heuck C. C. Basic laboratory procedures in clinical bacteriology / 2nd ed. Geneva, World Health Organization 2003 11. Mekonnen E. Health Laboratory Management andQuality Assurance Ethiopia Public Health Training Initiative, The CarterCenter, the

237 | Page PHCPI Laboratory Standards Guideline

Ethiopia Ministry of Health, and the Ethiopia Ministry of Education.2004

12. Manual for Laboratory Diagnosis of Common Epidemic Prone Diseases at District Public Health Laboratories, National Center for Disease Control (India) .September 2009.

13. Standard Operating Procedures for Laboratory Services Laboratory Department Coast Provincial General Hospital Ministry of Health, Kenya, March 2005.

14. Perilla M. J., Ajello G., Bopp C., Elliott J., Facklam R., Knapp J. S., Popovic T., Wells J., Scott F. Manual for the Laboratory Identification andAntimicrobial Susceptibility Testing of Bacterial Pathogens of Public Health Importance in the Developing World, Centers for Disease Control and Prevention: National Center for Infectious Diseases and World Health Organization: Department of Communicable Disease Surveillance and Response, 2003.

15. Manual of basic techniques for a health laboratory., 2nd ed.World Health Organization, Geneva ,2003 16. O’Connor, JH, Helicobacter pylori and gastric cancer: a review and hypothesis. Eur. J. Gastro. Hepa. 1992; 6: 103-109. 17. Mcguigan JE., Peptic ulcer and gastritis. In Harrison’s Principles of Internal Medicine, 12th Edition, 1988, Chapter 238, 1229-1248. 18. Cover TL. and Blaser MJ., Helicobacter pylori: A Bacterial Cause of Gastritis, Peptic Ulcer Disease, and Gastric Cancer, ASM News, 1995; 61: 21-26.

238 | Page PHCPI Laboratory Standards Guideline

19. Podolsky I, Lee E, Cohen R, Peterson WL. Prevalence of H. pylori in healthy subjects and patients with peptic diseases. Gastroenteroloty 1989: 96: Suppl: A394. Abstract. 20. Kist M., Immunology of Helicobacter pylori. In Helicobacter pylori in peptic ulceration and gastritis, edited by Marshall BJ., McCallum RW., and Guerrant RL., 1991, Chapter8, 92-110. 21. Ivanoff B. Typhoid fever, global situation and WHO recommendations. Southeast Asia J. Trop. Med. Public Health, 1995, 26:supp2 1-6 22. Parry CM, Hien TT Dougan G et al., Typhoid fever, N. Eng. J. Med. 2002, 347:1770-82. 23. Harris, JR., Cavallaro, EC., De Nobrega, AA., Dos S. Barrado, JCB., Bopp, C, Parsons, MB, Djalo, D, Fonseca, FG. dS, Ba, U, Semedo, A, Sobel, J. and Mintz, ED. (2009), Field Evaluation of Crystal VC® Rapid Dipstick Test for Cholera during a Cholera Outbreak in Guinea- Bissau. Trop Med Int Health 2009; 14(9):1117-112. 24. Nato F, Boutonnier A, Rajerison M, Grosjean P, Dartevelle S, Guenole A, et al. One-step immunochromatographic dipstick tests for rapid detection of Vibrio cholerae O1 and O139 in stool samples. Clin Diagn Lab Immunol 2003; 10:476–8. 25. Kalluri P, Naheed A, Rahman S, Ansaruzzaman M, Faruque AS, Bird M, et al., Evaluation of three rapid diagnostic tests for cholera: does the skill level of the technician matter? Trop Med Int Health. 2006; 11(1):49-55. 26. Mukherjee P, Ghosh S, Ramamurthy T, Bhattacharya MK, Nandy RK, Takeda Y, Nair GB, Mukhopadhyay AK. Evaluation of a rapid immunochromatographic dipstick kit for diagnosis of cholera

emphasizes its outbreak utility. Jpn

239 | Page PHCPI Laboratory Standards Guideline

Committee members involved in laboratory standards guideline for PHCs :‐  Dr. Faisal Ghazi Nasser / Ministry of Health.  Mohammed Ajil Samreen / Ministry of Health.  Dr. Zuhair Ibrahim Jassim / Ministry of Health.  Dr. Thana’a Hussein Mohammed Saleh / Ministry of Health.  Dr. Nawar Mazen Jawad Al‐Alwan / Ministry of Health.  Isra’a Abbas Fadhel / Ministry of Health.  Ali Hussein Heial / Ministry of Health.  Jalal Mohamed Qader / Ministry of Health.  Nadia Jawad Dawood / Ministry of Health.  Iman Yousef Rasheed / Ministry of Health.  Tha’er Kadhim Fayyadh / Ministry of Health.  Dr. Ghassan Ahmed Khalil / Ministry of Health.  Dr. Khalid Ahmed Mahmoud / Ministry of Health.  Dr. Bara’a Abdulwahid/ PHCPI.