Vítor Rodrigues do Nascimento

Licenciatura em Ciências da Engenharia Biomédica

Evaluation of for temperature measurement at the wards in a university .

Dissertação para obtenção do Grau de Mestre em Engenharia Biomédica

Orientador: Heikki Teriö, Ph.D. R&D Manager, Karolinska Institutet Co-orientador: Mário Secca, Ph.D, FCT-UNL

Júri:

Presidente: Prof. Doutora Maria Adelaide de Almeida Pedro de Jesus Arguente(s): Prof. Doutor Hugo Gamboa Vogal(ais): Prof. Doutor Mário Forjaz Secca

Vítor Rodrigues do Nascimento

Licenciatura em Ciências da Engenharia Biomédica

Evaluation of thermometers for ear temperature measurement at the wards in a university hospital.

Dissertação para obtenção do Grau de Mestre em Engenharia Biomédica

Orientador: Heikki Teriö, Ph.D. R&D Manager, Karolinska Institutet Co-orientador: Mário Secca, Ph.D, FCT-UNL

Universidade Nova de Lisboa - Faculdade de Ciencias e Tecnologia Lisboa - 2012

I

Evaluation of thermometers for ear temperature measurement at the wards in a university hospital.

Copyright © 2012 Vítor Rodrigues do Nascimento, FCT/UNL, UNL.

A Faculdade de Ciências e Tecnologia e a Universidade Nova de Lisboa têm o direito, perpétuo e sem limites geográficos, de arquivar e publicar esta dissertação através de exemplares impressos reproduzidos em papel ou de forma digital, ou por qualquer outro meio conhecido ou que venha a ser inventado, e de a divulgar através de repositórios científicos e de admitir a sua cópia e distribuição com objectivos educacionais ou de investigação, não comerciais, desde que seja dado crédito ao autor e editor.

Faculdade de Ciências e Tecnologia and Universidade Nova de Lisboa have the perpetual and unlimited geographic right to archive and publish this dissertation through printed or digital versions or any other known or still to be invented way, and to disclose it through scientific repositories and to admit its copy and distribution for educational or research purposes, non- commercial, as long as due credit is given to its author and editor.

II

Abstract

Since thermometers were banned due to environmental concerns, started to use electronic thermometers for measuring body temperature. Body temperature can be measured from different body parts, although the least invasive and quickest is preferred and therefore eardrum measurements are frequently taken. However, lately the staff feels that the taken measurements are not accurate. A new purchasing agreement for the purchase of these devices renders a good opportunity to study further the use of these devices at the wards of the university hospital, study their maintenance process, identify what performance is essential for the clinical usage, the parameters that are essential to measure and also identify ear thermometers in the market that can be used for comparative study.

Temperature measurements were taken with the help of an infrared ear , Covidien Genius2, in its calibration blackbody device at the R&D department of the Huddinge Hospital in order to verify accuracy claims. This data were compared against other studies and measurements of other infrared ear thermometers devices, the Braun ThermoScan Pro 4000 and also a digital contact thermometer, Welch Allyn Suretemp Plus, applied to different body sites. Informal meetings also took place in order to get more information about the devices and to know where they were used and repaired.

It was found that Genius2 measured temperature accurately when compared with a blackbody radiator. Regarding the Braun, it showed an accurate estimate of core temperature in comparison to invas