Review 821 European Society of Hypertension recommendations for conventional, ambulatory and home blood pressure measurement Eoin O’Brien, Roland Asmar, Lawrie Beilin, Yutaka Imai, Jean-Michel Mallion, Giuseppe Mancia, Thomas Mengden, Martin Myers, Paul Padfield, Paolo Palatini, Gianfranco Parati, Thomas Pickering, Josep Redon, Jan Staessen, George Stergiou and Paolo Verdecchia, on behalf of the European Society of Hypertension Working Group on Blood Pressure Monitoring Journal of Hypertension 2003, 21:821–848 Correspondence and requests for reprints to Professor Eoin O’Brien, Blood Pressure Unit, Beaumont Hospital, Dublin 9, Ireland. Keywords: blood pressure measurement, conventional blood pressure E-mail:
[email protected] measurement, ambulatory blood pressure measurement, home blood pressure measurement, sphygmomanometers, European Society of Received 31 January 2003 Accepted 3 February 2003 Hypertension recommendations Please refer to the Appendix for afilations. Introduction tension and cardiovascular disease. The content is Over the past 20 years or so, the accuracy of the divided into four parts: Part I is devoted to issues conventional Riva-Rocci/Korotkoff technique of blood common to all techniques of measurement, Part II to pressure measurement has been questioned and efforts conventional blood pressure measurement (CBPM), have been made to improve the technique with auto- Part III to ambulatory blood pressure measurement mated devices. In the same period, recognition of the (ABPM) and Part IV to self blood pressure measure- phenomenon