4. Measurement of Electrical Current, Voltage and Resistance

Total Page:16

File Type:pdf, Size:1020Kb

4. Measurement of Electrical Current, Voltage and Resistance Measurement of Electrical Current, Voltage and Resistance / 1 4. MEASUREMENT OF ELECTRICAL CURRENT, VOLTAGE AND RESISTANCE 4.1. PRINCIPLES OF MEASUREMENTS Electrical voltage and current are two important quantities in an electrical network. The voltage is the effort variable without which no current is available. It is measured across an electrical circuit element or branch of a circuit. The device that measures the voltage is the voltmeter. The current is the flow variable that represents net motion of the charged particles (electrons in solids, ions in a liquid) in a given direction. The product of the two yields the instantaneous electrical power. The ratio of the voltage to the current is the impedance. The current is measured by an ammeter (also called an ampermeter). Ammeters are connected in series with the load to measure the current in the load. Eventually, the ammeters require breaking the current loop to place it into the RT A circuit. The voltmeter connection is rather easy since it + IL V is connected without disturbing the circuit layout. VT RL - VL Therefore, most electrical measurements require determination of the voltage rather than the current due Figure 4.1. Connections for an the ease of measurement. Connections of ammeters and ammeter and a voltmeter. voltmeters are illustrated in Figure 4.1. The current generates a magnetic field around the current carrying conductor. It is also possible to check out the size of the current by sensing the magnetic field strength. This is carried out by clamp-on type ammeters. The electrical resistance of a circuit component is measured using an ohmmeter that applies a voltage across and determines the current passing through the component. Voltmeters and ammeters display the results as deflections of dials on calibrated screens or numerical values on alphanumeric displays as illustrated in Figure 4.2. Both types are connected to the circuit via sensing Figure 4.2. Analog and digital voltmeters leads and indicate the voltage. However, their Measurement of Electrical Current, Voltage and Resistance / 2 internal operations and user interfaces are different. The first type forms the analog meters that will be discussed in this chapter. The second category will be discussed in the next chapter under the title of digital voltmeters. 4.2. MC BASED MEASURING INSTRUMENTS 4.2.1. MC in Analog Electrical Measuring Instruments The standard MC instrument indicates positive DC currents (IMC) as deflection on the scale. The galvanometer displays both positive and negative currents. The moving IFSD RMC coil is usually made up of a very thin wire. The maximum current that + - gives full‐scale deflection IFSD is in the order of 0.1 to 10 mA and coil VMC resistance RMC 10 to 1000 Ω. The maximum deflection angle is about Figure 4.3. Model of MC 100°. The current through the moving coil IMC is limited by the IFSD. A voltage drop VMC = IMCRMC occurs across the coil. The moving coil can represented by the full‐scale deflection current IFSD and coil resistance RMC as shown in Figure 4.3. 4.2.2. Ammeters Basic DC Ammeter (Ampermeter) The current capacity of the meter can be expended by adding a resistor in RMC parallel with the meter coil as shown in Figure 4.4. The input current is I FSD + - shared between the coil resistance RMC and the parallel resistance that is VMC called the shunt RSH. As the maximum input current IT flows in, the coil takes RSH IFSD and remaining (IT ‐ IFSD) is taken by the shunt resistor. Voltage developed IT (IT - IFSD) across the meter is RM MC = RIV MCFSD = ( T − )RII SHFSD ..........................................(4.1) Figure 4.4. DC Ammeter. The meter resistance RM seen between the input terminals is VMC RM == MC // RR SH ................................................................................................. (4.2) IT Example 1 Calculate the multiplying power of a shunt of 200 Ω resistance used with a galvanometer of 1000 Ω resistance. Determine the value of shunt resistance to give a multiplying factor of 50. 9 Ifsdx1000 = (IT – Ifsd)x200 yielding IT = 6xIfsd. 9 For IT=50xIfsd, 1000xIfsd=(50-1)xIfsdxRsh yielding Rsh =1000/49 = 20.41 Ω Measurement of Electrical Current, Voltage and Resistance / 3 Multi­Range Ammeter Switch The parallel resistance (shunt) can be changed to suit poles different full‐scale current requirements as indicated in the previous example. The function can be accommodated by Rotary switch using a set of resistors and selecting them one by one. The arm switch however must be of make‐before‐break type Figure 4.5. Make‐before‐break type (Figure 4.5) that makes the contact with the new position switch. before it breaks the old connection. This eliminates the chance of forcing the full input current through the moving coil during changing the position of the switch. Example 2 Design a multi‐range DC ammeter using the basic movement with an internal resistance RMC= 50 Ω and full‐scale deflection current IMC= IFSD= 1 mA. The ranges required 0‐10 mA, 0‐50 mA, 0‐100 mA and 0‐500 mA as illustrated in Figure 4.6. 9 VMC = IMCxRMC = 50 mV 9 For range-1 (0-10 mA) RSH1= 50/9 =5.56 Ω IT IFSD RMC 0 500 mA RSH1 0 → 500 mA 0 100 mA 0 50 mA Rotary RSH2 0 10 mA 0 → 100 mA selector R switch SH3 0 → 50 mA R SH4 0 → 10 mA Multi-range ammeter scale Multi-range ammeter circuit Figure 4.6. A multi-range ammeter circuit and scale for example 2. 9 For range-2 (0-50 mA) RSH2= 50/49 =1.02 Ω 9 For range-3 (0-100 mA) RSH3= 50/99 =0.505 Ω 9 For range-4 (0-500 mA) RSH4= 50/499 =0.1 Ω Example 3 Design a multi‐range DC ammeter using the basic movement with an internal resistance RMC= 50 Ω and full‐scale deflection current IMC= IFSD= 10 mA. The ranges required 0‐0.1 A, 0‐1 A, 0‐10 A and 0‐ 100 A. Measurement of Electrical Current, Voltage and Resistance / 4 9 VMC = IMCxRMC = 500 mV 9 For range-1 (0-0.1 A) RSH1= 500/90 = 5.56 Ω 9 For range-2 (0-1 A) RSH2= 0.5/0.99 = 0.505 Ω 9 For range-3 (0-10 A) RSH3= 0.5/9.99 = 0.05 Ω 9 For range-4 (0-100 A) RSH4= 0.5/99.99 = 0.005 Ω 4.2.3. Voltmeters IFSD RS RMC Basic DC Voltmeter The moving coil can be used as a voltmeter by adding a series + - + - VS VMC resistance RS as illustrated in Figure 4.7. The input voltage is VM divided between the coil resistance RMC and RS. Current passing + - RM through both resistors is IMC which is limited by the full‐scale deflection current IFSD of the coil. The full‐scale input voltage Figure 4.7. Basic DC voltmeter. VM = IFSD(RS+RMC) ................................................. (4.3) The input impedance seen is RM = RS + RMC ..................................................................................................................... (4.4) However, with RS>>RMC, RM is approximately equal to RS and VM ≅ IFSDRS. Example 4 The coil of a moving coil voltmeter is 4 cm long and 3 cm wide and has 100 turns on it. The control spring exerts a torque of 2.4x10-4 N-m when the deflection is 100 divisions on the full scale. If the flux density of the magnetic filed in the air-gap is 0.1 Wb/m2, estimate the resistance that must be put in series with the coil to give one volt per division. The resistance of the voltmeter coil may be neglected. -4 -4 9 TEM = TSP ⇒ 2.4x10 = 100x0.1x12x10 xIFSD ⇒ IFSD =20 mA. Therefore, current per division is 0.2 mA. Assuming that RMC is negligibly small compared to RS : RS = 5 kΩ Example 5 A moving coil instrument gives full-scale deflection of 10 mA when the potential difference across its terminals is 100 mV. Calculate: a. The shunt resistance for a full scale corresponding to 100 mA; 9 RSH = 100 / 90 = 1.11 Ω b. The resistance for full scale reading with 1000 V; 9 RMC = 100 /10 = 10 Ω; RS + RMC = (1000 / 10) kΩ = 100 kΩ yielding RS = 100 kΩ (RMC is negligible) Measurement of Electrical Current, Voltage and Resistance / 5 c. The power dissipated by the coil and by the external resistance in each case. 2 9 Power dissipated by the coil, PC = IM xRMC = 1 mW; 2 9 PSH = VM /RSH = 9 mW 2 9 PS = VM /RS = 10 W. Multi­Range Voltmeter The series resistance can be changed to suit different full‐scale voltage requirements as shown in Figure 4.8. Resistors are organized either in parallel fashion (conventional connection) as in the case Voltage to be measured RS2 RS1 RMC RS4 IFSD 0 → 1000 V RMC 1 RS3 0 → 100 V RS3 2 V R M S2 0 → 50 V 3 Rotary 4 selector RS1 0 → 10 V RS4 0 → 1000 V switch Multi-range voltmeter circuit Multi-range voltmeter circuit Parallel connection Series connection Figure 4.8. Parallel and series resistance connections for a multi-range voltmeter. of ammeter and selecting them one by one or all connected in series like a voltage divider (modified connection). The switch however must be of break‐before‐ Switch make type (Figure 4.9) that breaks the contact with the old poles position before it makes it with the new position. This Rotary eliminates the chance of forcing a current larger than the switch arm full‐scale current through the moving coil during changing Figure 4.9.
Recommended publications
  • Made to Measure. Practical Guide to Electrical Measurements in Low Voltage Switchboards V
    Contact us A 250 500 200 150 V (b) 100 (a) 50 0 t Made to measure. Practical guide to electrical measurements in low voltage switchboards A 250 500 ABB SACE The data and illustrations are not binding. We reserve 200 the right to modify the contents of this document on the 150 Una divisione di ABB S.p.A. basis of technical development of the products, 100 Apparecchi Modulari without prior notice. 50 0 Viale dell’Industria, 18 Copyright 2010 ABB. All rights reserved. - 1.500 - CAL. 20010 Vittuone (MI) Tel.: 02 9034 1 Fax: 02 9034 7609 bol.it.abb.com www.abb.com V 80 V 60 2CSC445012D0201 - 12/2010 (f) 40 50 Hz 20 0 t Made to measure. Practical guide to electrical measurements in low voltage switchboards table of Made to measure. Practical guide to electrical measurements contents in low voltage switchboards 1 Electrical measurements 5.3.2 Current transformers ......................................................... 37 5.3.3 Voltage transformers ......................................................... 38 1.1 Why is it important to measure? .......................................... 3 5.3.4 Shunts for direct current .................................................... 38 1.2 Applicational contexts .......................................................... 4 1.3 Problems connected with energy networks ......................... 4 6 The measurements 1.4 Reducing consumption ........................................................ 7 1.5 Table of charges .................................................................. 8 6.1 TRMS Measurements
    [Show full text]
  • Keithley Instrumentation for Electrochemical Test Methods and Applications ––
    Keithley Instrumentation for Electrochemical Test Methods and Applications –– APPLICATION NOTE Keithley Instrumentation for Electrochemical Test Methods and Applications APPLICATION NOTE With more than 60 years of measurement expertise, Keithley Cyclic Voltammetry Instruments is a world leader in advanced electronic test Cyclic voltammetry (CV), a type of potential sweep method, instrumentation. Our customers are scientists and engineers is the most commonly used electrochemical measurement in a wide range of research and industrial applications, technique, which typically uses a 3-electrode cell. Figure including many electrochemistry tests. Keithley manufactures 1 illustrates a typical electrochemical measurement circuit products that can source and measure current and voltage made up of an electrochemical cell, an adjustable voltage accurately. Electrochemistry disciplines that employ source (V ), an ammeter (A ), and a voltmeter (V ). The Keithley instrumentation include battery and energy storage, S M M three electrodes of the electrochemical cell are the working corrosion science, electrochemical deposition, organic electrode (WE), the reference electrode (RE), and the counter electronics, photo-electrochemistry, material research, electrode (CE). The voltage source (V ) for the potential sensors, and semiconductor materials and devices. Table 1 S scan is applied between the WE and CE. The potential (E) lists some of the test methods and applications that employ between the RE and WE is measured with the voltmeter, and Keithley products. the overall voltage (VS) is adjusted to maintain the desired Table 1. Electrochemistry test methods and applications potential at the WE with respect to the RE. The resulting Methods and Measurement current (i) flowing to or from the WE is measured with the Capabilities Applications ammeter (AM).
    [Show full text]
  • Speed Limit: How the Search for an Absolute Frame of Reference in the Universe Led to Einstein’S Equation E =Mc2 — a History of Measurements of the Speed of Light
    Journal & Proceedings of the Royal Society of New South Wales, vol. 152, part 2, 2019, pp. 216–241. ISSN 0035-9173/19/020216-26 Speed limit: how the search for an absolute frame of reference in the Universe led to Einstein’s equation 2 E =mc — a history of measurements of the speed of light John C. H. Spence ForMemRS Department of Physics, Arizona State University, Tempe AZ, USA E-mail: [email protected] Abstract This article describes one of the greatest intellectual adventures in the history of mankind — the history of measurements of the speed of light and their interpretation (Spence 2019). This led to Einstein’s theory of relativity in 1905 and its most important consequence, the idea that matter is a form of energy. His equation E=mc2 describes the energy release in the nuclear reactions which power our sun, the stars, nuclear weapons and nuclear power stations. The article is about the extraordinarily improbable connection between the search for an absolute frame of reference in the Universe (the Aether, against which to measure the speed of light), and Einstein’s most famous equation. Introduction fixed speed with respect to the Aether frame n 1900, the field of physics was in turmoil. of reference. If we consider waves running IDespite the triumphs of Newton’s laws of along a river in which there is a current, it mechanics, despite Maxwell’s great equations was understood that the waves “pick up” the leading to the discovery of radio and Boltz- speed of the current. But Michelson in 1887 mann’s work on the foundations of statistical could find no effect of the passing Aether mechanics, Lord Kelvin’s talk1 at the Royal wind on his very accurate measurements Institution in London on Friday, April 27th of the speed of light, no matter in which 1900, was titled “Nineteenth-century clouds direction he measured it, with headwind or over the dynamical theory of heat and light.” tailwind.
    [Show full text]
  • Electronic Voltmeters and Ammeters - Alessandro Ferrero, Halit Eren
    ELECTRICAL ENGINEERING – Vol. II - Electronic Voltmeters and Ammeters - Alessandro Ferrero, Halit Eren ELECTRONIC VOLTMETERS AND AMMETERS Alessandro Ferrero Dipartimento di Elettrotecnica, Politecnico di Milano, Italy Halit Eren Curtin University of Technology, Perth, Western Australia Keywords: currents, voltages, measurements, standards, analog voltmeters, digital voltmeters, microvoltmeters, oscilloscopes Contents 1. Introduction. 2. Analog Meters 2.1. DC Analog Voltmeters and Ammeters 2.2. AC Analog Voltmeters and Ammeters 2.3. True rms Analog Voltmeters 3. Digital Meters 3.1. Dual-Slope DVMs 3.2. Successive-Approximation ADCs 3.3. AC Digital Voltmeters and Ammeters 3.4. Frequency Response of AC Meters 4. Radio-Frequency Microvoltmeters 5. Vacuum-Tube Voltmeters and Oscilloscopes 5.1. Analog Oscilloscopes 5.2. Digital Storage Oscilloscopes (DSOs) 5.3. Portable Oscilloscopes 5.4. High-Voltage Oscilloscopes Appendix Glossary Bibliography Biographical Sketches Summary Voltage UNESCOand current measurements are – esse EOLSSntial parts of engineering and science. Instruments that measure voltages and currents are called voltmeters and ammeters, respectively. ThereSAMPLE are two distinct types of voltmeterCHAPTERS and ammeter, which differ from each other by the operating principle that they are based on: electromechanical instruments and electronic instruments, which also include oscilloscopes. Electromechanical voltmeters and ammeters, including thermal-type instruments, represent early technology, but still are used in many applications. Basic elements of voltages and currents from the basic physical principles have been introduced in the electromechanical voltage and current measurements section. Also, voltage and currents standards have been dealt with in detail in other articles. ©Encyclopedia of Life Support Systems (EOLSS) ELECTRICAL ENGINEERING – Vol. II - Electronic Voltmeters and Ammeters - Alessandro Ferrero, Halit Eren In this article, modern electronic voltmeters and ammeters are discussed.
    [Show full text]
  • A Simple Atmospheric Electrical Instrument for Educational Use
    A simple atmospheric electrical instrument for educational use A.J. Bennett1 and R.G. Harrison Department of Meteorology, The University of Reading P.O. Box 243, Earley Gate, Reading RG6 6BB, UK Abstract Electricity in the atmosphere provides an ideal topic for educational outreach in environmental science. To support this objective, a simple instrument to measure real atmospheric electrical parameters has been developed and its performance evaluated. This project compliments educational activities undertaken by the Coupling of Atmospheric Layers (CAL) European research collaboration. The new instrument is inexpensive to construct and simple to operate, readily allowing it to be used in schools as well as at the undergraduate University level. It is suited to students at a variety of different educational levels, as the results can be analysed with different levels of sophistication. Students can make measurements of the fair weather electric field and current density, thereby gaining an understanding of the electrical nature of the atmosphere. This work was stimulated by the centenary of the 1906 paper in which C.T.R. Wilson described a new apparatus to measure the electric field and conduction current density. Measurements using instruments based on the same principles continued regularly in the UK until 1979. The instrument proposed is based on the same physical principles as C.T.R. Wilson's 1906 instrument. Keywords: electrostatics; potential gradient; air-earth current density; meteorology; Submitted to Advances in Geosciences 1 E-mail: [email protected] 1 1. Introduction The phenomena of atmospheric electricity provide an ideal topic for stimulating lectures, talks and laboratory demonstrations.
    [Show full text]
  • Electrical Measurements
    Name ________________________ Group #_______ Date _________ Partners ______________________ Electrical Measurements Experimental Objective The objective of this experiment is to become familiar with some of the electrical instruments. You will gain experience by wiring a simple electrical circuit and drawing its circuit diagram. You will learn the correct use of a digital multimeter as an ohmmeter, voltmeter and ammeter. Theory Current, I, is the flow of charge, measured in amps. Voltage, V, is the difference in electrical potential between two points, measured in volts. Resistance is the ratio of voltage across to the current flowing through it, measured in ohms, Ω. V R = I Wires are conductors with very low resistance designed to carry the current from one object to another. Resistors are objects with a moderately high resistance made of carbon films. Resistors are color coded to indicate the magnitude. A circuit diagram is a diagram that represents the electrical circuit using internationally accepted symbols. The diagram represents the electrical connections of the circuit, but not necessarily the bench layout of each item. A series connection consists of two or more components that are connected end to end with one another and the same current flows through each component. A parallel connection consists of components connected so that one end of all the components are connected together and the other ends are connected together such that the same voltage is applied across each component. Multimeters are used to measure current, voltage and resistance. The meters used in our lab will give you a digital reading and since they use a battery should always be turned off when not in use.
    [Show full text]
  • An Oscillating - Magnet Watt Balance
    An Oscillating - Magnet Watt Balance H. Ahmedov TÜBİTAK, UME, National Metrology Institute of Turkey E-mail:[email protected] Abstract We establish the principles for a new generation of simplified and accurate watt balances in which an oscillating magnet generates Faraday’s voltage in a stationary coil. A force measuring system and a mechanism providing vertical movements of the magnet are completely independent in an oscillating magnet watt balance. This remarkable feature allows to establish the link between the Planck constant and a macroscopic mass by a one single experiment. Weak dependence on variations of environmental and experimental conditions, weak sensitivity to ground vibrations and temperature changes, simple force measuring procedure, small sizes and other useful features offered by the novel approach considerable reduce the complexity of the experimental setup. We formulate the oscillating-magnet watt balance principle and establish the measurement procedure for the Planck constant. We discuss the nature of oscillating-magnet watt balance uncertainties and give a brief description of the Ulusal Metroloji Enstitüsü ( UME ) watt balance apparatus. 1. Introduction In 1975 Kibble described the principles of the first moving-coil watt balance [1,2], a two-part experiment which links the electrical and mechanical Sİ units by indirect conversion of electrical power into mechanical power. The watt balance principle and macroscopic electrical quantum effects: the Josephson effect and the quantum Hall effect [3, 4] established the link between a macroscopic mass and the Planck constant, the fundamental constant of the microworld [ 5 ]. This link provides a potential route to the redefinition of the kilogram [6, 7], the last base unit which is still defined by a manmade object, the international prototype of the kilogram.
    [Show full text]
  • Experiment 0 an Introduction to the Equipment
    Experiment 0 An Introduction to the Equipment Objectives After completing Experiment 0, you should be able to: • Use basic electronic instruments • Determine the precision of a measurement • Select the scale that gives the most accurate reading • Give a qualitative description of electric potential (voltage), current, and resistance • Describe the uses of an electrometer, voltmeter, ammeter, ohmmeter, and multimeter • Use an electrometer and digital multimeter properly • Describe the precautions required to protect meters from damage. Introduction In Physics 116L, you will investigate the properties of electricity and magnetism with a variety of laboratory instruments. Unlike mechanics, for which the basic measurements of length, time, and mass are familiar, common quantities, electricity and magnetism involve unfamiliar quantities and require special instruments for their study. Some of the measurements are quite simple, for example the circuits of Experiments 5 and 6, but others are more subtle. Although you are certainly familiar with certain aspects of basic electricity -- shocks upon touching metal objects on dry days, the quantitative experiments are not trivial. You must understand a number of physical processes and phenomena to form a conceptual picture of what is happening in these experiments. We cannot explore electrostatics one step at a time as in the lectures; understanding even the simplest experiments requires the complete framework of electrostatics. These first three experiments, which introduce you to electrical instruments and the basic properties of electric charge, require considerable thought and care. If you are unfamiliar with these instruments, you may find them slightly intimidating at first. However, they are not really difficult to use. This first "experiment" is merely a set of exercises to enable you to experiment with the basic instruments and become comfortable with using them.
    [Show full text]
  • Basic Electrical Measurements
    3 Basic Electrical Measurements 3.1 Electromechanical measuring devices current is delivered to the coil by two springs – these springs are also used as the mechanisms generating There are several advantages of traditional returning torque for the pointer. electromechanical instruments: simplicity, reliability, low price. The most important advantage is that the majority of such instruments can work without any additional power supply. Since people’s eyes are sensitive to movement also this psycho-physiological aspect of analogue indicating instruments (with moving pointer) is appreciated. On the other hand, there are several drawbacks associated with electromechanical analogue indicating instruments. First of all, they do not provide output signal, thus there is a need for operator’s activity during the measurement (at least for the reading of an indicated value). Another drawback is that such instruments generally use moving mechanical parts, which are sensitive to shocks, aging or wearing out. Relatively low price of moving pointer instruments FIGURE 3.1 today is not as advantageous as earlier, because on the The example of moving coil indicating instrument (1- moving coil, 2 – market there are available also very cheap digital permanent magnet, 3 – axle, 4 – pointer, 5 – bearings, 6 – spring, 7 – correction of zero). measuring devices with virtual pointer. Regrettably, it can be stated that most of the The moving coil is placed into the gap between the electromechanical analogue instruments are rather of magnet poles and soft iron core, shaped in such a way poor quality. In most cases these instruments are not as to produce uniform magnetic field. The movement able to measure with uncertainty better than 0.5%.
    [Show full text]
  • The Hall Effect C1
    The Hall Effect C1 Head of Experiment: Zulfikar Najmudin The following experiment guide is NOT intended to be a step -by-step manual for the experiment but rather provides an overall introduction to the experiment and outlines the important tasks that need to be performed in order to complete the experiment. Additional sources of documentation may need to be researched and consulted during the experiment as well as for the completion of the report. This additional documentation must be cited in the references of the report. 1 RISK ASSESSMENT AND STANDARD OPERATING PROCEDURE 1. PERSON CARRYING OUT ASSESSMENT Name Geoff Green Position Chf Lab Tech Date 18/09/08 2. DESCRIPTION OF ACTIVITY C1 Hall Effect 3. LOCATION Campus SK Building Huxley Room 403 4. HAZARD SUMMARY Accessibility X Mechanical X Manual X Hazardous Handling Substances Electrical X Other Lone Working Yes No Permit-to- Yes No Permitted? Work Required? 5. PROCEDURE PRECAUTIONS Use of 240v Mains Powered Equipment Isolate Socket using Mains Switch before unplugging or plugging in equipment Accessibility All bags/coats to be kept out of aisles and walkways. Use of Electro-magnet See attached Scheme of Work 6. EMERGENCY ACTIONS All present must be aware of the available escape routes and follow instructions in the event of an evacuation 2 THE HALL EFFECT 1. Objectives The goal of this experiment is to make accurate measurements of the charge carrier density and carrier mobility in three different materials where the majority charge carriers are electrons (n-type) or holes (p-type) using the van der Pauw method.
    [Show full text]
  • Principles of Electrical Measurement
    PRINCIPLES OF ELECTRICAL MEASUREMENT Sensors Series Senior Series Editor: B E Jones Series Co-Editor: W B Spillman, Jr Novel Sensors and Sensing R G Jackson Hall Effect Devices, Second Edition R S Popovi´c Sensors and their Applications XII Edited by S J Prosser and E Lewis Sensors and their Applications XI Edited by K T V Grattan and S H Khan Thin Film Magnetoresistive Sensors S Tumanski Electronic Noses and Olfaction 2000 Edited by J W Gardner and K C Persaud Sensors and their Applications X Edited by N M White and A T Augousti Sensor Materials P T Moseley and J Crocker Biosensors: Microelectrochemical Devices M Lambrecht and W Sansen Current Advances in Sensors Edited by B E Jones Series in Sensors PRINCIPLES OF ELECTRICAL MEASUREMENT S Tumanski Warsaw University of Technology Warsaw, Poland New York London IP832_Discl.fm Page 1 Wednesday, November 23, 2005 1:02 PM Published in 2006 by CRC Press Taylor & Francis Group 6000 Broken Sound Parkway NW, Suite 300 Boca Raton, FL 33487-2742 © 2006 by Taylor & Francis Group, LLC CRC Press is an imprint of Taylor & Francis Group No claim to original U.S. Government works Printed in the United States of America on acid-free paper 10987654321 International Standard Book Number-10: 0-7503-1038-3 (Hardcover) International Standard Book Number-13: 978-0-7503-1038-3 (Hardcover) Library of Congress Card Number 2005054928 This book contains information obtained from authentic and highly regarded sources. Reprinted material is quoted with permission, and sources are indicated. A wide variety of references are listed.
    [Show full text]
  • Electrical Measurements Lab Manual
    ELECTRICAL MEASUREMENTS LABORATORY MANUAL B.TECH (IV YEAR – I SEM) (2020-21) Prepared by: Mr G SEKHAR BABU, Assistant Professor Mr I RAJA SHEKAR, Assistant Professor Mr R PRADEEP, Assistant Professor Mr K HARISH, Assistant Professor DEPARTMENT OF ELECTRICAL & ELECTRONICS ENGINEERING MALLA REDDY COLLEGE OF ENGINEERING & TECHNOLOGY (Autonomous Institution – UGC, Govt. of India) Recognized under 2(f) and 12 (B) of UGC ACT 1956 Affiliated to JNTUH, Hyderabad, Approved by AICTE - Accredited by NBA & NAAC – ‘A’ Grade - ISO 9001:2015 Certified Maisammaguda, Dhulapally (Post Via. Kompally), Secunderabad – 500100, Telangana State, India 1 | P a g e LIST OF EXPERIMENTS (R18A0288) ELECTRICAL MEASUREMENTS LABORATORY COURSE OBJECTIVES: The objectives of the course are to make the students learn about: To calibrate LPF Watt Meter, energy meter, P.F Meter using electro dynamo meter type instrument as the standard instrument. To determine unknown inductance, resistance, capacitance by performing experiments on D.C Bridges &A.C Bridges. To determine three phase active & reactive powers using single wattmeter method practically. Measurement of parameters of choke coil To determine the ratio and phase angle errors of current transformer and potential transformer. Measuring earth resistance, dielectric strength of transformer oil & Testing of underground cables. Part - A The following experiments are required to be conducted compulsory experiments: 1. Calibration and Testing of single phase energy Meter 2. Measurement of tolerance of batch of low resistances by Kelvin’s double bridge 3. Measurement of voltage, current and resistance using DC potentiometer 4. Schering Bridge and Anderson bridge. 5. Measurement of parameters of a choke coil using 3 voltmeter and 3 ammeter methods.
    [Show full text]