Radiopacity Measurement of Restorative Resins Using Film and Three Digital Systems for Comparison with ISO 4049: International Standard

Total Page:16

File Type:pdf, Size:1020Kb

Radiopacity Measurement of Restorative Resins Using Film and Three Digital Systems for Comparison with ISO 4049: International Standard Bull Tokyo Dent Coll (2015) 56(4): 207–214 Original Article Radiopacity Measurement of Restorative Resins Using Film and Three Digital Systems for Comparison with ISO 4049: International Standard Rishabh Kapila1), Yukiko Matsuda1), Kazuyuki Araki1), Tomohiro Okano1), Keiichi Nishikawa2) and Tsukasa Sano1) 1) Division of Radiology, Department of Oral Diagnostic Sciences, Showa University School of Dentistry, 2-1-1 Kita-senzoku, Ota-ku, Tokyo 145-8515, Japan 2) Department of Oral and Maxillofacial Radiology, Tokyo Dental College, 2-9-18 Misaki-cho, Chiyoda-ku, Tokyo 101-0061, Japan Received 7 April, 2014/Accepted for publication 29 May, 2015 Abstract This study compared Ultra Speed Occlusal Film (USOF) and 3 digital systems in determining the radiopacity of 5 different restorative resins in terms of equivalents of aluminum thickness. Whether those digital systems could be used to determine whether radiopacity was in line with International Organization for Standardization (ISO) recommendations was also investigated. Disks of each of 5 restorative resins and an aluminum step wedge were exposed at 65 kVp and 10 mA on USOF and imaged with each digital system. Optical density on the film was measured with a transmission densitometer and the gray values on the digital images using Image J software. Graphs showing gray value/optical density to step wedge thickness were constructed. The aluminum equivalent was then calculated for all the resins using a regression equation. All the resins were more radiopaque than 1 mm of aluminum, and therefore met the ISO 4049 recommendations for restorative resins. Some resins showed statistically higher aluminum equivalents with digital imaging. The use of traditional X-ray films is declining, and digital systems offer many advantages, including an easy, fast, and reliable means of determining the radiopacity of dental materials. Key words: Radiopacity — Restorative resins — Digital system Introduction It is important to evaluate restorations accurately and in detail, which means the Radiopacity is a prerequisite of dental dentist must be able to ascertain whether materials, including restorative composite secondary caries or marginal defects are resins. A standardized method for measuring present, what the contours of the restorations the radiopacity of dental materials was are, whether there is contact with adjacent established3). teeth, and whether there are any cement 207 208 Kapila R et al. Table 1 Light cure composite resin materials Material Manufacturer Sorare® P GC Dental (Tokyo, Japan) Gradia® DirectX GC Dental (Tokyo, Japan) Estelite® Sigma Quick Tokuyama Dental Corp. (Tokyo, Japan) Clearfil® AP-X Kuraray Medical Inc. (Tokyo, Japan) Beautiful® II Shofu Inc. (Kyoto, Japan) overhangs or interfacial gaps, for example. films and densitometers or spectrophotome- Therefore, the materials used in such resto- ters1). A digital system for dentistry was first rations need to be sufficiently radiopaque introduced in 1989, since which time digital to allow them to be distinguished from radiography has found its way into dental background enamel and dentin. Radiopaque practice. Several types of sensor are available materials offer a number of advantages over for imaging: a charge-coupled device (CCD), those that are radiolucent: for example, they a complementary metal oxide semiconductor make detection of recurrent dental caries and (CMOS), and photo-stimulable phosphor visualisation of the radiographic interface plates. The most important advantages of between the materials and the tooth substrates digital radiographic systems are that they are easier5). more sensitive than silver halide film and Several factors may affect the radiopacity allow exposure to radiation to be reduced. In of dental materials, but composition seems digital imaging, the gray scale has an inverse to be most important. Additional factors relationship with optical density, with white include material thickness, type of X-ray film being allotted a value of 255 (for an 8-bit used, and alteration of powder/liquid ratio of image) and black 0. Unless performed with luting materials1). It has been demonstrated great care, traditional film development can that the radiopacity of dentin is approximately produce significant variations in the final equivalent to that of aluminum of the same radiograph. The results of digital imaging, on thickness, and that enamel has approximately the other hand, are more consistent4). twice the radiopacity of aluminum of the same The aim of the present study was to thickness9). According to the International determine the radiopacity of 5 different restor- Organization for Standardization (ISO), the ative resins in terms of equivalent aluminum radiopacity of a material should be equal to or thickness by using an ultra-speed film and greater than the same thickness of aluminum, 3 different digital systems. Whether those and should not be less than 0.5 mm of any digital systems could be used to determine value claimed by the manufacturer3). whether radiopacity was in line with ISO One of the most highly recommended recommendations was also investigated. methods of measuring radiopacity involves using an aluminum step wedge as a reference standard. The ISO and American National Materials and Methods Standards Institute/American Dental Associa- tion have published standardized procedures 1. Preparation of specimens for quantifying the radiopacity of several types The 5 restorative resins used in the study of dental material which use a ≥98% pure are summarized in Table 1. All were obtained aluminum step wedge as a reference8). directly from the maker and were manu- Radiographic image density is commonly factured in accordance with ISO standards. evaluated by means of conventional X-ray The test specimens were prepared by using Digital Evaluation of Radiopacity 209 Table 2 Digital imaging systems used in study System Manufacturer Charged-coupled device Kodak RVG 5000 (Rochester, NY, USA) Imaging plate Soredex Digora (Tuusula, Finland) Imaging plate Array Corporation Arcana (Tokyo, Japan) the method described in ISO 4049:20093). Disk-shaped specimens 10 mm in diameter and 1.0 mm in thickness were made from Fig. 1 Images obtained each of the 5 resins. The thickness of each A: Ultra Speed Occlusal Film; B: Arcana; C: Digora; and D: RVG. specimen was measured to an accuracy of ±0.01 mm using a micrometer. An aluminum step wedge was also fabricated according to ISO recommendations with an aluminum RVG owing to the small size of the imaging purity exceeding 98%. The aluminum step plate and sensor, respectively. To reduce wedge had steps of 1.0, 1.5, 2.0, 3.0, 4.0, 6.0, variation in the intensity of radiation in the 8.0 and 10.0 mm in thickness. central X-ray beam due to the anode heel effect, the film and specimen were placed 2. Film and digital systems perpendicular to the cathode-anode axis. Kodak Ultra Speed Occlusal Film (USOF: Exposure was set at 65 kVp and 10 mA, with a DF-50, Carestream Health, Rochester, NY, USA) 30-cm focus to film distance. The X-ray unit was used. Table 2 shows the 3 digital imaging was kept in the same position throughout systems employed: Arcana, RVG, and Digora. the experiment. Exposure time was 0.05 and 0.25 sec for RVG, Digora, and Arcana, and 3. Exposure conditions 0.25 sec for USOF. These exposure parameters In ISO 4049:2009, it is recommended that were selected so that the individual steps of the dental X-rays be taken with a single-phase unit aluminum step wedge would be clearly visible with a total filtration of 1.5 mm aluminum on the images produced. Three exposures and capable of operating at 65±5 kVp. were made for each of the settings for the Currently, single phase X-ray units are being USOF and 3 digital systems. replaced by constant-potential machines, which The images produced on USOF were pro- yield a longer contrast scale and reduce patient cessed immediately in an automatic processor exposure to radiation8). In the present study, (Dent-X, AFP Imaging, Elmsford, NY, USA) X-rays were produced by an inverter-type operating at 28°C with a Dent-X developer constant-potential dental X-ray unit, the Max- and fixer (Dent-X, AFP Imaging). The imaging DC70 ( J. Morita, Kyoto, Japan), with spot sizes plates produced with Digora and Arcana were of 0.8 and 2.5 mm aluminum filtration for scanned immediately, whereas direct digital each exposure. images were displayed with RVG (Fig. 1). All 5 disk shaped specimens were placed on the USOF along with the aluminum step 4. Image analysis wedge. All 5 disk shaped specimens were Optical density on the USOF images was also subsequently placed on the imaging plate measured by using a transmission densitom- of the Arcana. Only one resin disk shaped eter (Konica Densitometer PDA-15; Konica specimen at a time was used for Digora and Minolta Holdings, Tokyo, Japan) with an 210 Kapila R et al. Fig. 3 Sample graph with polynomial equation used to obtain equivalent of aluminum thickness in mm sheet (ver. 2007), with the Y axis equalling Fig. 2 Sample image used for mean gray value the step wedge thickness in millimeters and measurement in Arcana using Image J software X axis the corresponding gray value/optical 50×50-pixel regions of interest were set in each density of the step wedge. A regression step and resin disks. equation was obtained from each graph using the polynomial method to the third degree up to 10 decimal places. The regression curve and reliability of each plot exceeded 0.99, aperture of 1 mm in diameter. The optical indicating that there was no deviation from densities were measured for all 1 mm thick disk linearity for the different thicknesses of the resin specimens and each step of aluminium aluminum step wedge. According to ISO step wedge on each film. Each film was recommendations, the slope of the graph measured three times.
Recommended publications
  • WHO Manual of Diagnostic Imaging Radiographic Anatomy and Interpretation of the Musculoskeletal System
    The WHO manual of diagnostic imaging Radiographic Anatomy and Interpretation of the Musculoskeletal System Editors Harald Ostensen M.D. Holger Pettersson M.D. Authors A. Mark Davies M.D. Holger Pettersson M.D. In collaboration with F. Arredondo M.D., M.R. El Meligi M.D., R. Guenther M.D., G.K. Ikundu M.D., L. Leong M.D., P. Palmer M.D., P. Scally M.D. Published by the World Health Organization in collaboration with the International Society of Radiology WHO Library Cataloguing-in-Publication Data Davies, A. Mark Radiography of the musculoskeletal system / authors : A. Mark Davies, Holger Pettersson; in collaboration with F. Arredondo . [et al.] WHO manuals of diagnostic imaging / editors : Harald Ostensen, Holger Pettersson; vol. 2 Published by the World Health Organization in collaboration with the International Society of Radiology 1.Musculoskeletal system – radiography 2.Musculoskeletal diseases – radiography 3.Musculoskeletal abnormalities – radiography 4.Manuals I.Pettersson, Holger II.Arredondo, F. III.Series editor: Ostensen, Harald ISBN 92 4 154555 0 (NLM Classification: WE 141) The World Health Organization welcomes requests for permission to reproduce or translate its publications, in part or in full. Applications and enquiries should be addressed to the Office of Publications, World Health Organization, CH-1211 Geneva 27, Switzerland, which will be glad to provide the latest information on any changes made to the text, plans for new editions, and reprints and translations already available. © World Health Organization 2002 Publications of the World Health Organization enjoy copyright protection in accordance with the provisions of Protocol 2 of the Universal Copyright Convention. All rights reserved.
    [Show full text]
  • Removal of Foreign Bodies in the Upper Gastrointestinal Tract in Adults: European Society of Gastrointestinal Endoscopy (ESGE) Clinical Guideline
    Guideline Removal of foreign bodies in the upper gastrointestinal tract in adults: European Society of Gastrointestinal Endoscopy (ESGE) Clinical Guideline Authors Michael Birk1, Peter Bauerfeind2, Pierre H. Deprez3, Michael Häfner4, Dirk Hartmann5, Cesare Hassan6, Tomas Hucl7, Gilles Lesur8, Lars Aabakken9, Alexander Meining1 Institutions Institutions are listed at end of article. Bibliography This Guideline is an official statement of the European Society of Gastrointestinal Endoscopy (ESGE). It DOI http://dx.doi.org/ addresses the removal of foreign bodies in the upper gastrointestinal tract in adults. 10.1055/s-0042-100456 Published online: 2016 Recommendations Endoscopy © Georg Thieme Verlag KG Nonendoscopic measures Endoscopic measures Stuttgart · New York 1 ESGE recommends diagnostic evaluation based 7 ESGE recommends emergent (preferably within ISSN 0013-726X on the patient’s history and symptoms. ESGE re- 2 hours, but at the latest within 6 hours) thera- commends a physical examination focused on peutic esophagogastroduodenoscopy for foreign Corresponding author the patient’s general condition and to assess signs bodies inducing complete esophageal obstruc- Alexander Meining, MD Department of Internal of any complications (strong recommendation, tion, and for sharp-pointed objects or batteries in Medicine I, University of Ulm low quality evidence). the esophagus. We recommend urgent (within 24 Albert-Einstein-Allee 23 2 ESGE does not recommend radiological evalua- hours) therapeutic esophagogastroduodenoscopy 89081 Ulm tion for
    [Show full text]
  • Medical Imaging Louis Kreel
    Postgrad Med J: first published as 10.1136/pgmj.67.786.334 on 1 April 1991. Downloaded from Postgrad Med J (1991) 67, 334 - 346 i) The Fellowship of Postgraduate Medicine, 1991 Reviews in Medicine Medical imaging Louis Kreel Department ofDiagnostic Radiology, Prince of Wales Hospital, Shatin, N. T., Hong Kong Introduction Physicists have tapped the electromagnetic spec- latitude with lower radiation and minimal repeat trum to great effect. Each energy wave-form that examinations. Dark rooms for film processing will penetrates tissues has a corresponding imaging disappear. Fluoroscopy too will be fully automated system, starting with Roentgen's radiograph of his and digitized, and all radiographs will show the full wife's hand in 1895. Since then gamma rays, range of radiodensities from lung, through soft X-rays, protons, ultrasound and radiofrequency tissues to bone with one exposure. This will entail coupled to a magnetic field have been grafted onto new equipment, film, cassettes and imaging systems computers to produce anatomical images varying at considerable expense. in type and detail. More recently these modalities Furthermore, each passing year sees the intro- have produced sophisticated physiological data of duction of new scanners with shorter scanning considerable interest. times and higher spatial resolution, producing Thus the words 'digital' and greater detail and 'computer' have more information. Millisecond copyright. invaded radiology departments, not only with computed tomography (CT), cardiac cine, scanners, of which they are integral components, magnetic resonance imaging (MRI) down to but also in conventional fluoroscopy and radio- breath-holding time, and the technology of CT graphy. The change is as revolutionary as that of applied to nuclear medicine (NM), yielding single steam engine to internal combustion engine or photon emission CT and proton emission tomog- automobile to aeroplane.
    [Show full text]
  • Energy Subtraction Methods As an Alternative to Conventional X-Ray Angiography" (2016)
    Western University Scholarship@Western Electronic Thesis and Dissertation Repository 10-26-2016 12:00 AM Energy Subtraction Methods as an Alternative to Conventional X- Ray Angiography Christiane S. Burton The University of Western Ontario Supervisor Ian A. Cunningham The University of Western Ontario Graduate Program in Medical Biophysics A thesis submitted in partial fulfillment of the equirr ements for the degree in Doctor of Philosophy © Christiane S. Burton 2016 Follow this and additional works at: https://ir.lib.uwo.ca/etd Part of the Medical Biophysics Commons, and the Physical Sciences and Mathematics Commons Recommended Citation Burton, Christiane S., "Energy Subtraction Methods as an Alternative to Conventional X-Ray Angiography" (2016). Electronic Thesis and Dissertation Repository. 4213. https://ir.lib.uwo.ca/etd/4213 This Dissertation/Thesis is brought to you for free and open access by Scholarship@Western. It has been accepted for inclusion in Electronic Thesis and Dissertation Repository by an authorized administrator of Scholarship@Western. For more information, please contact [email protected]. Abstract Cardiovascular diseases (CVD) are currently the leading cause of death worldwide and x-ray angiography is used to assess a vast majority of CVD cases. Digital subtraction an- giography (DSA) is a technique that is widely used to enhance the visibility of small vessels obscured by background structures by subtracting a mask and contrast image. However, DSA is generally unsuccessful for imaging the heart due to the motion that occurs between mask and contrasted images which cause motion artifacts. An alternative approach, known as dual-energy or energy subtraction angiography (ESA) is one that exploits the iodine k- edge by acquiring images with a low and high kV in rapid succession.
    [Show full text]
  • Radiodensity Evaluation of Dental Impression Materials in Comparison to Tooth Structures
    www.scielo.br/jaos Radiodensity evaluation of dental impression materials in comparison to tooth structures Rodrigo Borges FONSECA1, Carolina Assaf BRANCO2, Francisco HAITER-NETO3, Luciano de Souza GONÇALVES4, Carlos José SOARES56, Mário Alexandre Coelho SINHORETI7, Lourenço CORRER-SOBRINHO7 1– DDS, MSc, PhD, Professor, Federal University of Goiás, Dental School, Restorative Dentistry Area, Goiânia, GO, Brazil. 2– DDS, MSc, PhD student, São Paulo University, Dental School, Ribeirão Preto, SP, Brazil. 3– DDS, MSc, PhD, Professor, State University of Campinas, Piracicaba Dental School, Department of Oral Radiology, Piracicaba, SP, Brazil. 4– DDS, MSc PhD student at State University of Campinas, Piracicaba Dental School, Department of Dental Materials, Piracicaba, SP, Brazil. 5– DDS, MS, PhD, Professor, Federal University of Uberlândia, Dental School, Department of Operative Dentistry and Dental Materials, Uberlândia, MG, Brazil. 6- DDS, MS, PhD, Professor, Federal University of Paraíba, Dental School, Department of Restorative Dentistry, João Pessoa, PB, Brazil. 7- DDS, MS, PhD, Professor, State University of Campinas, Piracicaba Dental School, Department of Dental Materials, Piracicaba, SP, Brazil. Corresponding address: Rodrigo B. Fonseca - Faculdade de Odontologia - Universidade Federal de Goiás - Praça Universitária esquina com 1ª. Avenida, s/n, Setor Universitário - Goiânia - GO - Brazil - 74605-220 - e-mail: [email protected] - Phone: +55-62-32096325 - Fax: +55-62-32096054. ["#$%"&%"' ABSTRACT n the most recent decades, several developments have been made on impression materials’ Icomposition, but there are very few radiodensity studies in the literature. It is expected that an acceptable degree of radiodensity would enable the detection of small fragments left inside gingival sulcus or root canals. Objective: The aim of this study was to determine the radiodensity of different impression materials, and to compare them to human and bovine enamel and dentin.
    [Show full text]
  • Machine-Learning-Based Multiple Abnormality Prediction with Large-Scale Chest Computed Tomography Volumes
    Machine-Learning-Based Multiple Abnormality Prediction with Large-Scale Chest Computed Tomography Volumes Rachel Lea Draelosa,b, David Dovc, Maciej A. Mazurowskic,d,e, Joseph Y. Loc,d,f, Ricardo Henaoc,e, Geoffrey D. Rubind, Lawrence Carina,c,g aComputer Science Department, Duke University, LSRC Building D101, 308 Research Drive, Duke Box 90129, Durham, North Carolina 27708-0129, United States of America bSchool of Medicine, Duke University, DUMC 3710, Durham, North Carolina 27710, United States of America cElectrical and Computer Engineering Department, Edmund T. Pratt Jr. School of Engineering, Duke University, Box 90291, Durham, North Carolina 27708, United States of America dRadiology Department, Duke University, Box 3808 DUMC, Durham, North Carolina 27710, United States of America eBiostatistics and Bioinformatics Department, Duke University, DUMC 2424 Erwin Road, Suite 1102 Hock Plaza, Box 2721 Durham, North Carolina 27710, United States of America fBiomedical Engineering Department, Edmund T. Pratt Jr. School of Engineering, Duke University, Room 1427, Fitzpatrick Center (FCIEMAS), 101 Science Drive, Campus Box 90281, Durham, North Carolina 27708-0281, United States of America gStatistical Science Department, Duke University, Box 90251, Durham, North Carolina 27708-0251, United States of America Address for correspondence: Rachel Lea Draelos Email: [email protected] Department of Computer Science 308 Research Drive Durham, North Carolina 27708-0129 United States Abstract Machine learning models for radiology benefit from large-scale data sets with high quality labels for abnormalities. We curated and analyzed a chest computed tomography (CT) data set of 36,316 volumes from 19,993 unique patients. This is the largest multiply-annotated volumetric medical imaging data set reported.
    [Show full text]
  • Automated Evaluation of Radiodensities in a Digitized Mammogram Database Using Local Contrast Estimation
    Rowan University Rowan Digital Works Theses and Dissertations 12-31-2004 Automated evaluation of radiodensities in a digitized mammogram database using local contrast estimation Min Taek Kim Rowan University Follow this and additional works at: https://rdw.rowan.edu/etd Recommended Citation Kim, Min Taek, "Automated evaluation of radiodensities in a digitized mammogram database using local contrast estimation" (2004). Theses and Dissertations. 1173. https://rdw.rowan.edu/etd/1173 This Thesis is brought to you for free and open access by Rowan Digital Works. It has been accepted for inclusion in Theses and Dissertations by an authorized administrator of Rowan Digital Works. For more information, please contact [email protected]. Automated evaluation of radiodensities in a digitized mammogram database using local contrast estimation by Min Taek Kim A Thesis Submitted to the Graduate Faculty in Partial Fulfillment of the Requirements for.the Degree of MASTER OF SCIENCE Department: Electrical and Computer Engineering Major: Engineering (Electrical Engineering) Approved: Members of the Committee In Charge of Major Work For the Major Department For t e College Rowan University Glassboro, New Jersey 2004 ABSTRACT Mammographic radiodensity is one of the strongest risk factors for developing breast cancer and there exists an urgent need to develop automated methods for predicting this marker. Previous attempts for automatically identifying and quantifying radiodense tissue in digitized mammograms have fallen short of the ideal. Many algorithms require significant heuristic parameters to be evaluated and set for predicting radiodensity. Many others have not demonstrated the efficacy of their techniques with a sufficient large and diverse patient database. This thesis has attempted to address both of these drawbacks in previous work.
    [Show full text]
  • High-Pitch CT Pulmonary Angiography in Third Generation Dual-Source CT: Image Quality in an Unselected Patient Population
    RESEARCH ARTICLE High-Pitch CT Pulmonary Angiography in Third Generation Dual-Source CT: Image Quality in an Unselected Patient Population Bastian O. Sabel*, Kristijan Buric, Nora Karara, Kolja M. Thierfelder, Julien Dinkel, Wieland H. Sommer, Felix G. Meinel Institute for Clinical Radiology, Ludwig-Maximilians-University Hospital, Munich, Germany * [email protected] Abstract Objectives OPEN ACCESS To investigate the feasibility of high-pitch CT pulmonary angiography (CTPA) in 3rd genera- Citation: Sabel BO, Buric K, Karara N, Thierfelder tion dual-source CT (DSCT) in unselected patients. KM, Dinkel J, Sommer WH, et al. (2016) High-Pitch CT Pulmonary Angiography in Third Generation Dual-Source CT: Image Quality in an Unselected Patient Population. PLoS ONE 11(2): e0146949. Methods doi:10.1371/journal.pone.0146949 Forty-seven patients with suspected pulmonary embolism underwent high-pitch CTPA on a rd Editor: Chin-Tu Chen, The University of Chicago, 3 generation dual-source CT scanner. CT dose index (CTDIvol) and dose length product UNITED STATES (DLP) were obtained. Objective image quality was analyzed by calculating signal-to-noise- Received: September 4, 2015 ratio (SNR) and contrast-to-noise ratio (CNR). Subjective image quality on the central, Accepted: December 22, 2015 lobar, segmental and subsegmental level was rated by two experienced radiologists. Published: February 12, 2016 Copyright: © 2016 Sabel et al. This is an open Results access article distributed under the terms of the Median CTDI was 8.1 mGy and median DLP was 274 mGy*cm. Median SNR was 32.9 in Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any the central and 31.9 in the segmental pulmonary arteries.
    [Show full text]
  • Imaging Foreign Bodies: Ingested, Aspirated, and Inserted
    IMAGING/REVIEW ARTICLE Imaging Foreign Bodies: Ingested, Aspirated, and Inserted Hsiang-Jer Tseng, MD*; Tarek N. Hanna, MD; Waqas Shuaib, MD; Majid Aized, MD; Faisal Khosa, MD, MBA; Ken F. Linnau, MD, MS *Corresponding Author. E-mail: [email protected], Twitter: @DrJackTseng. Foreign bodies can gain entrance to the body through several mechanisms, ie, ingestion, aspiration, and purposeful insertion. For each of these common entry mechanisms, this article examines the epidemiology, clinical presentation, anatomic considerations, and key imaging characteristics associated with clinically relevant foreign bodies seen in the emergency department (ED) setting. We detail optimal use of multiple imaging techniques, including radiography, ultrasonography, fluoroscopy, and computed tomography to evaluate foreign bodies and their associated complications. Important imaging and clinical features of foreign bodies that can alter clinical management or may necessitate emergency intervention are discussed. [Ann Emerg Med. 2015;66:570-582.] Please see page 571 for the Editor’s Capsule Summary of this article. A podcast for this article is available at www.annemergmed.com. 0196-0644/$-see front matter Copyright © 2015 by the American College of Emergency Physicians. http://dx.doi.org/10.1016/j.annemergmed.2015.07.499 INTRODUCTION recognize that radiopacity and radiographic visibility are Foreign bodies are a common cause of emergency 2 different concepts. Radiopacity is an intrinsic feature of department (ED) visits and can gain entry into the human an
    [Show full text]
  • Nonvariceal Upper Gastrointestinal Bleeding EVIDENCE TABLE
    ACR Appropriateness Criteria® Nonvariceal Upper Gastrointestinal Bleeding EVIDENCE TABLE Patients/ Study Objective Study Reference Study Type Study Results Events (Purpose of Study) Quality 1. Feinman M, Haut ER. Upper Review/Other- N/A To review UGIB. UGIB is still associated with significant 4 gastrointestinal bleeding. Surg Clin North Tx morbidity and mortality. The cornerstone of Am. 2014;94(1):43-53. management is initial stabilization, followed by localization and treatment of the bleeding. Medical management and minimally invasive treatments are used primarily and are often successful. Surgery is reserved for patients who fail conservative management. 2. Geffroy Y, Rodallec MH, Boulay-Coletta Review/Other- N/A To provide a structured approach in the The use of multidetector CTA in the 4 I, Julles MC, Ridereau-Zins C, Zins M. Dx diagnosis of GIB with multidetector CTA. diagnostic workup of patients with acute GIB Multidetector CT angiography in acute continues to grow because of improvements in gastrointestinal bleeding: why, when, and scanning technology and favorable results how. Radiographics. 2011;31(3):E35-46. published in several recent articles. Positive contrast-enhanced MDCT can define with a high degree of accuracy the location and at times the cause of active GIB. Contrast- enhanced MDCT with ready availability offers clear advantages in anatomic detail and can provide a rapid assessment that reveals underlying bowel disease and can map anatomy for angiography or surgery. * See Last Page for Key Revised 2016 Singh-Bhinder/Kim Page 1 ACR Appropriateness Criteria® Nonvariceal Upper Gastrointestinal Bleeding EVIDENCE TABLE Patients/ Study Objective Study Reference Study Type Study Results Events (Purpose of Study) Quality 3.
    [Show full text]
  • 205383Orig1s000
    CENTER FOR DRUG EVALUATION AND RESEARCH APPLICATION NUMBER: 205383Orig1s000 MEDICAL REVIEW(S) Division of Medical Imaging Products NDA 205383; Oraltag (iohexol) (b) (4) Oral Solution Sponsor: Otsuka Pharmaceutical Development and Commercialization Reviewer: Harris E. Orzach, MD Date: March 12, 2015 Addendum to Primary Clinical Review The NDA application contains a Statement of Financial Certification and Disclosure, stating the following: “This NDA does not include the reports of any clinical trials sponsored by the applicant, nor did the applicant contract with one or more clinical investigators to conduct the published studies summarized in this NDA. Thus, the requirements of 21 CFR Part 54 do not apply. The applicant’s assessment is correct and fulfills the certification and disclosure requirement. No regulatory action is needed. Reference ID: 3717437 --------------------------------------------------------------------------------------------------------- This is a representation of an electronic record that was signed electronically and this page is the manifestation of the electronic signature. --------------------------------------------------------------------------------------------------------- /s/ ---------------------------------------------------- HARRIS E ORZACH 03/17/2015 LIBERO L MARZELLA 03/17/2015 Reference ID: 3717437 assemble (b) (4) dedicated to commercial production of iohexol (b) (4) oral solution. Batch analysis and stability studies submitted in the present application are not acceptable, (b) (4) . CMC: I agree with the CMC reviewer’s assessment that all the CMC issues have been addressed The applicant states that in the resubmission, they have addressed the issues documented in the Complete Response letter issued by the FDA. Specifically: 1. The manufacturing facility, Ultra Seal (USC) in New Paltz, New York, has done the following: Installation and qualification of the manufacturing equipment to be used for commercial production.
    [Show full text]
  • Using X-Ray Imaging Techniques to Determine Density of Foods Shivangi Mohan Kelkar Purdue University
    Purdue University Purdue e-Pubs Open Access Dissertations Theses and Dissertations Fall 2013 Using X-ray Imaging Techniques to Determine Density of Foods Shivangi Mohan Kelkar Purdue University Follow this and additional works at: https://docs.lib.purdue.edu/open_access_dissertations Part of the Bioresource and Agricultural Engineering Commons, and the Food Science Commons Recommended Citation Kelkar, Shivangi Mohan, "Using X-ray Imaging Techniques to Determine Density of Foods" (2013). Open Access Dissertations. 151. https://docs.lib.purdue.edu/open_access_dissertations/151 This document has been made available through Purdue e-Pubs, a service of the Purdue University Libraries. Please contact [email protected] for additional information. Graduate School ETD Form 9 (Revised 12/07) PURDUE UNIVERSITY GRADUATE SCHOOL Thesis/Dissertation Acceptance This is to certify that the thesis/dissertation prepared By Shivangi Mohan Kelkar Entitled Using X-ray Imaging Techniques to Determine Density of Foods Doctor of Philosophy For the degree of Is approved by the final examining committee: Martin Okos Chair Carol J Boushey Osvaldo Campanella Ganesan Narsimhan To the best of my knowledge and as understood by the student in the Research Integrity and Copyright Disclaimer (Graduate School Form 20) , this thesis/dissertation adheres to the provisions of Purdue University’s “Policy on Integrity in Research” and the use of copyrighted material. Approved by Major Professor(s): ___________________Martin Okos _________________ ____________________________________ Approved by: Bernard Engel/ R.L. Stroshine 10/07/2013 Head of the Graduate Program Date USING X-RAY IMAGING TECHNIQUES TO DETERMINE DENSITY OF FOODS A Dissertation Submitted to the Faculty of Purdue University by Shivangi Mohan Kelkar In Partial Fulfillment of the Requirements for the Degree of Doctor of Philosophy December 2013 i Purdue University West Lafayette, Indiana ii To my parents and my brother ii iii ACKNOWLEDGEMENTS I would like to express my deepest gratitude to my academic advisor Dr.
    [Show full text]