Using High Speed/High Resolution Size Exclusion Chromatography Separation of Polymeric Materials with Dynamic and Multi-Angle Light Scattering Detection

Jennifer Gough1, Robert Birdsall1, Damian Morrison1, John Champagne2 , Kyle Rogers2 1Waters Corporation, Milford, MA, USA, 2Wyatt Technology Corporation, 6330 Hollister Ave.; Santa Barbara, CA ,USA

OVERVIEW INTRODUCTION  Explore more detection with the Polymeric materials are an integral part of everyday tasks, and Waters® ACQUITY® Advanced there is great complexity behind the research and manufacturing of Chromatography® (APC™) these high quality and competitive products. From the simple task System and the Wyatt μDAWN™ of carrying food in biodegradable shopping bags to battling diseases multi-angle and quasi-elastic light with intricate drug delivery systems, these everyday tasks are scattering (MALS/QELS). possible through an exacting polymer process. As technologies advance, there is a drive to understand and predict material  Bring the high resolution separation capabilities of APC to light scattering properties through more expansive characterization and flexible analysis. analytical instrumentation. Creative analytical solutions, ranging from high speed chromatographic separation to advanced mass  Experience the simplicity of the APC/ spectrometry, have become the answer for complex product µDAWN hyphenation and the benefits knowledge.1, 2, 3 that are brought to polymer characterization. Traditional analytical solutions have often been the answer for polymer characterization. Coupling liquid chromatography (LC) with detectors such as a (RI), ultra violet or photo diode array (UV or PDA), evaporative light scattering (ELS) and fluorescence (FLR) have met traditional needs. Examples of creative analytical solutions include adding a secondary LC system for two dimensional (2D) chromatography or hyphenating the LC system with detectors such as dynamic light scattering (DLS) and multi- angle light scattering (MALS).4

LC separations with resolution and speed have not always been synonymous. Yet, with the innovative design of the Waters Advanced Polymer Chromatography (APC) system, including the robust separation column and low dispersion flow path, the system yields high resolution, short analysis times, and quick mobile phase changes. This high speed/high resolution separation technique works best with low dispersion detectors, availability and diversity of which continue to increase as the advantages of APC become more widely used in research and industry.5, 6, 7

In this study, the expansion of the Waters APC approach to the size exclusion separation is presented. The sample is diverted from the ACQUITY APC Core 1 system to a Wyatt µDAWN with MALS/QELS and RI detector while maintaining the required system dispersion control needed for the APC experiment (Figure 1).8

Figure 1. ACQUITY APC Core 1 System with post column flow to the µDAWN.

TO DOWNLOAD A COPY OF THIS POSTER NOTE, VISIT WWW.WATERS.COM/POSTERS 1 EXPERIMENTAL Conditions for demonstration measurements The test conditions are based on using a high speed, high resolution ACQUITY APC SEC Core 1 System coupled to Wyatt µDAWN and Optilab UT-rEX detectors. The two sytems are connected with seven thousanths inner diameter tubing and a low pressure poly fitting from the APC PDA outlet to the µDAWN inlet. The first experimental samples evaluated are a mixture of two narrow polystyrene (PS) standards and of a 200K dalton PS standard dissolved in terahydrofuran (THF). The second experimental samples are a bovine serum albumin (BSA) standard and a protein standard mixture in aqueous buffer solution. The separations are analyzed using the Wyatt µDAWN and RI. Each system‘s respective sofware is used for their respective instrument control, data collection and data processing. SEC conditions Organic Solvent Method System: Waters APC Core 1 System Eluent: THF Flow Rate: 0.7 mL/min Sample Concentration: 1 mg/mL of each narrow standard (2 mg/mL total) Injection Volume: 10 μL Column Temp.: 40 °C Columns: Waters 4.6 x 150 mm APC XT 450 Å, 125 Å and 45 Å in series Conc. Detector: Waters ACQUITY RI detector @ 40 °C (used only when disconnected from µDAWN) LS and RI Detector: Wyatt µDAWN and Optilab UT-rEX (settings below) Coupling Tubing: 0.007 inches ID Stainless Steel tubing Aqueous Method System: Waters APC Core 1 System Eluent: Aqueous Buffer: 150mM NaCl, 20mM Phosphate, pH=6.5 Flow Rate: 0.6 mL/min Sample Concentration: 1 mg/mL of each narrow standard Injection Volume: 2 μL Column Temp.: 25 °C Columns: Waters ACQUITY BEH SEC Protein 4.6 x 150 mm 200A Conc. Detector: Waters ACQUITY RI detector @ 40 °C (used only when disconnected from µDAWN) LS and RI Detector: Wyatt µDAWN and Optilab UT-rEX (settings below) Coupling Tubing: 0.007 inches ID Stainless Steel tubing

Wyatt Detector Conditions LS Detector: Wyatt µDAWN LS Temp: Ambient RI Detector: Optilab UT-rEX (25°C)

Chromatographic System Control Waters Empower 3 FR3 Software and Wyatt ASTRA Software

Using High Speed/High Resolution Size Exclusion Chromatography Separation of Polymeric Materials 2 with Dynamic and Multi-Angle Light Scattering Detection RESULTS AND DISCUSSION The polymer samples for this study are chosen to define the separation of a polymer standard mixture and a single narrow polymer standard. The protein standards are chosen for their common use in biological and pharmaceutical fields. Comparison of the RI detector signals of the mixed polymer standards are used to assess RI resolution and analysis quality throughout the analyte flow path. The single protein is an excellent example of a biological macromolecule for confirmation of detector resolution, and the protein standard mixture adds confidence to instrumental analysis flexibility from organic solvent to aqueous chromatography within an eight hour day. Below in Figure 2 is a chromatogram of a PS narrow standard mixture of three molecular weights. The overlaid RI analysis reveals a consistent retention time for the APC and Optilab detectors. The RI data is generated from importing the APC and the Wyatt data into a spreadsheet through text files. Data confidence is increased through the comparable instrumental signal analysis in the RI chromatogram. Figure 3 shows the 200K Dalton (Da) PS correlation function and control curve from the µDAWN, which provides the hydrodynamic radius value. The MALS/QELS detection, in conjunction with RI detector, enable many calculations such as the absolute molecular weight, calculated mass, and mass fraction (Table 1). The constant RI detector response is indicative of peak/sample homogeneity (Figure 4). These polymer properties reveal material characteristics that can be helpful in determining behavior of the polymer in a product matrix or solution.9-11

12 -7.12E-04

10 -7.14E-04 APC RI -7.16E-04 8 dRI Optilab

-7.18E-04 6 -7.20E-04

RIU 4 RIU µ -7.22E-04 2 -7.24E-04

0 -7.26E-04 3 5 7 9 11

-2 -7.28E-04

-4 -7.30E-04 Time (min)

Table 1. Molecular weight customizable table of 200K Figure 2. APC RI signal in blue, and Optilab RI signal in Da PS. red of PS standard mix.

Figure 3. MALS/QELS correlation and control graphs: Figure 4. RI is constant across 200K Da PS LS peak: 200K Da PS. consistent signal across peak.

Using High Speed/High Resolution Size Exclusion Chromatography Separation of Polymeric Materials 3 with Dynamic and Multi-Angle Light Scattering Detection The advantages gained by polymer characterization with the µDAWN and Optilab can also be applied to the polymer- like characteristics of proteins. The BSA analysis is demonstrated below in the table and in the chromatogram, bringing confirmation to the resolution through the exhibited monomer, dimer and trimer (Table 2 & Figure 5). The hydrodynamic radius of the BSA in Figure 6 can be used to predict the behavior of the analyte in solution. The protein standard mixture, seen in Figure 7, demonstrates the separation and detection capability of the combined systems and their flexibility to operate with an organic solvent method (normal phase) or an aqueous method (reverse phase).10, 11 The red horizontal lines near each peak in the chromatograms below represent the molar mass distribution, and these can have linear, convex or concave shape. In this study, the BSA and protein mix exhibited some non-linear behavior in their molar mass distribution. This curvature is related to detector response of an analyte. A constant detector response is indicative of a homogenous sample that is baseline resolved from neighboring peaks in the chromatogram. Deviation from a constant detector response or non-linear response can be a result of increased peak heterogeneity or non-resolved peaks.12

Table 2. Wyatt ASTRA calculations for the BSA molecular weight of peaks 1, 2, and 3.

Figure 5. LS and RI chromatogram of BSA: horizontal lines denote molar mass distribution. Figure 6. MALS/QELS correlation and control graphs: BSA.

Figure 7. LS and RI chromatogram of protein standard mix: horizontal lines denote molar mass distribution.

Using High Speed/High Resolution Size Exclusion Chromatography Separation of Polymeric Materials 4 with Dynamic and Multi-Angle Light Scattering Detection CONCLUSIONS References 1. Xian, Jun Loh; for Personal Care Products  Coupling the APC, high resolution / high speed SEC and Cosmetics, Royal Society of Chemistry, Core 1 system to the Wyatt μDAWN and RI Cambridge, U.K., 2016. detectors is easily managed with the selection of 2. Srivastava, Apurva, et al. (2016), “Polymers in Drug the correct tubing diameter. Delivery”. Journal of Biosciences and Medicines 4: 69  The impact of the Wyatt detectors on -84. chromatographic band spread is minimized with 3. Montaudo, M. S., Puglisi, C., Samperi, F. and their small inter-detector dispersion design capable Montaudo, G. (1998), “Application of size exclusion of interfacing with most standard ultra high chromatography matrix-assisted desorption/ performance LC systems. ionization time-of-flight to the determination of molecular masses in polydisperse polymers”. Rapid  In the example studied, the benefit of adding the Commun. Mass Spectrom., 12: 519–528. MALS/QELS and RI to the SEC experiment offers the 4. Provder, Theodore; Urban, Marek W.; Barth, Howard ability to calculate absolute molecular weight and G.; Hyphenated Techniques in Polymer hydrodynamic radius. These calculations can predict Characterization, American Chemical Society, polymer behaviors in solutions/product matrices and Washington, DC 1994. bring confidence to sample quality through extended quantification.9-11 5. “ACQUITY Advanced Polymer Chromatography System”, Waters Corporation, 9 November 2016,  The high speed chromatography offered by the APC http://www.waters.com/webassets/cms/library/ is matched well to the experimental conditions docs/720004570en.pdf available with the μDAWN enabling greater 6. Kenrick, Sophia; Berges, Aym; “Absolute Molar Mass characterization for the complex properties of in UHPLC via the µDawn and UT-rEX”, Wyatt polymers and proteins. Technology Corporation, 18 November 2016, http:// www.wyatt.com/files/literature/app-notes/sec-mals- proteins/absolute-molar-mass-uhplc.pdf 7. “ACQUITY APC Columns: Increased Speed and Resolution for Aqueous and Organic Polymer Separations”, Waters Corporation, 9 November 2016, http://www.waters.com/webassets/cms/library/ docs/720004580en.pdf 8. “Online MALS-QELS (Quasi-Elastic Light Scattering)”, Wyatt Technology Corporation, 18 November 2016, http://www.wyatt.com/files/literature/app-notes/ wyattqels-hydrodyn.pdf 9. “International Symposium on GPC/SEC and Related Topics”, Amsterdam, N.L., 26 September 2016 10. Van Krevelen, D.W.; Te Nijehuis, K.; Properties of Polymers. 4th Ed., Elsevier Science, Amsterdam, N.L., 2009. 11. Pan, S., Ahirwal, D., Nguyen, D.A., Sridhar, T., Sunthar, P., Prakash, J.R. (2014) “Viscosity Radius of Polymers in Dilute Solutions: Universal Behavior from DNA Rheology and Brownian Dynamics Simulations”, Macromolecules, Issue 21, Volume 47, 7548-7560. 12. Slotbloom, D.J., Duurkens, R.H., Olieman, K., Erkens, G.B. (2008), “Static Light Scattering to Characterize Membrane Proteins in Detergent Waters, ACQUITY, Advanced Polymer Chromatography, and Solution”, Methods, Issue 46, 73-82. The Science of What’s Possible are registered trademarks of Waters Corporation. APC and IonSABRE are trademarks of Waters Corporation. All other trademarks are the property of their respective owners.