The ROXI Colorimeter & Fluorimeter. Laboratory Application I

Total Page:16

File Type:pdf, Size:1020Kb

The ROXI Colorimeter & Fluorimeter. Laboratory Application I The ROXI Colorimeter & Fluorimeter. Laboratory Application I. Colorimetric measurements via Beer’s Law. Required Supplies & Costs: RGB LED; $1.95 Light Sensors; $3.95 ea 3-way switch; $6.54 300 ohm resistor; $0.10 Arduino Uno; $24.95 1. Sensor Description. The sensor consists of a common cathode RGB light emitting diode (Sparkfun Electronics, https://www.sparkfun.com/products/105 ), and two logarithmic scale analog light sensors (Adafruit Industries, https://www.adafruit.com/products/1384 ). Peak wavelengths for emission by the LED are specified as 467.5, 520, 625 nm by the manufacturer. The light sensors are powered by 2.3-6 VDC, and produce an output voltage (Vout) that is linear with the logarithm of the light intensity incident upon the sensor. The light sensors are glued onto the side of a cuvette holder that was 3D printed. Two light sensors (shown in blue in figure below) are present. The first at 180 deg. to the incident beam is for transmittance / absorbance measurements. The second detector is placed at 90 deg. angle for fluorescence or light scattering measurements. A circuit diagram for the apparatus is shown below. A 3-way switch (NKK Switch, HS13Y) distributes power to the RGB LED with current controlled via a 300-ohm resistor at the common cathode. Each light detector is powered via 5VDC, and ground is common. Two analog input channels of the Arduino are devoted to voltage measurements from the sensors. light detectors cuvette holder Vcc GND Ω Ω Ω Ω Signals Vcc GND V GND S = 5V R= 300 R= 3-way switch to distribute 5V to anode of desired color Figure 1. Schematic of the colorimeter / fluorimeter setup. D) A) B) C) Figure 2. Photograph of (A) the Adafruit light sensor chip; (B) view from above of the 3D printed cuvette holder, (C) the cuvette holder with LED and light detectors affixed with JB Weld, (D) the colorimeter installed within the ROXI case. The label was printed and laminated prior to gluing to the plexiglass. Three thru holes are designed into the cuvette holder to allow for placement of the LED and two light detectors. Peak wavelengths for emission by the LED are specified as 467.5, 520, 625 nm by the manufacturer. 2. Calibration of the Colorimeter. After assembly, we desired to devise a method to calibrate, or check the accuracy of the light sensor by using samples of known transmittance. To accomplish this, various plastic filters manufactured by Lee Filters were used. The manufacturer of these filters kindly distributes transmittance spectra for their products (shown below), allowing a simple means to achieve a calibration standard. Several optical filters could be placed into the optical path sequentially and the change in detector signal observed. The accepted transmittance values and measured voltages were then plotted and an exponential fit applied to the data. The resulting best-fit function was subsequently used to determine % transmittance for samples. This allows a transmittance standard to establish the relationship between voltage and transmittance. Data collected for an example trial experiment is shown on the next two pages. Information Regarding the Optical Filters LED color Measured Known (courtesy of Lee Filters) Voltage (V) Transmission Green 1.6023 1.5 % Red 2.2295 83% Blue 0.98 0% Green 0.5488 0% Red 1.5092 8% Blue 0.4949 0% Green 1.9845 12 % Red 0.3577 0% Blue 1.744 5% Green 2.5431 84 % Red 2.1021 47% Blue 2.2687 n.a. Green 2.4794 60 % Red 2.1021 12% Blue 2.6019 79% Green 1.9649 8% Red 0 0% Blue 2.4206 42% Green 2.3373 50% Red 0 0% Blue 2.5774 72% Green 1.9943 8% Red 2.0139 35% Blue 2.4059 38% Green 2.2589 30% Red 1.911 22% Blue 2.5235 59% Blank – 100% T Green -530 nm 2.5921 100% Blank – 100% T Blue -467 nm 2.6558 100% Blank – 100% T Red -630 nm 2.2785 100% Blue Channel 120 100 y = 0.0172e 3.2352x 80 R² = 0.9975 60 40 % Transmission % 20 0 0 1 2 3 sensor voltage (V) Green Channel 120 100 y = 0.0022e 4.1755x 80 R² = 0.9832 60 40 % Transmission % 20 0 0 1 2 3 sensor voltage (V) Red Channel 120 100 y = 0.0339e 3.4571x 80 R² = 0.9683 60 40 Transmission 20 0 0 1 2 3 sensor voltage (V) 3. Experiment: Demonstration of Beer’s Law. Introduction. The purpose of this experiment is to demonstrate the Beer-Lambert law relationship using the colored dye bromthymol blue and the ROXI colorimeter. Transmittance is defined as the fraction of incident electromagnetic power that is transmitted through a sample: ͽ Transmittance: T = ͽt where I, I 0 are the transmitted and incident radiant powers, respectively. All spectrometers measure transmittance, but chemists frequently convert to absorbance since this variable is linearly related to the concentration of the absorbing substance: Absorbance: A= -log (T) A= 2 - log (%T) The Beer-Lambert law of optical absorption describes the linear relationship between absorbance (A) and concentration (c). Beer’s law: A = ϵ × b × c In this equation, ε is a wavelength dependent number that describes the ability of the substance to absorb light, and b is the pathlength (cm) light travels through the sample. There is flexibility with what units of concentration can be used, however, if molarity is chosen ε is known as the molar absorptivity (L / mol cm). Procedure. 1. Dissolve approx. 150 mg of bromthymol blue (M.W. = 624.38 g/mol) in a 250.00 mL volumetric flask (water is solvent). Compute the exact molarity of the solution. 2. Obtain five 100 mL volumetric flasks. Transfer 1 mL, 2 mL, 3 mL, 4 mL, and 5 mL of the stock solution to volumetric flasks 1-5, respectively. 3. Fill the volumetric flasks to the mark using pH=10 ammonium chloride / ammonia buffer and mix well. Compute the exact concentrations of these solutions by using the dilution factors. 4. Turn on the ROXI colorimeter and set the color dial to red. Confirm the 100% T has been set. Recalibrate using transmittance standards if necessary. Zero the blank (buffer solution). 5. Use the ROXI colorimeter to measure the absorbance of the solutions using the red channel. 6. Use the data and concentrations to prepare a Beer’s law plot of the data. 7. Obtain a solution of unknown concentration and test the unknown solution and measure the absorbance using the colorimeter. 8. If desired, the absorbance values measured with the ROXI colorimeter can be compared to measurements made with a laboratory spectrophotometer for quality control. Sample Data. Mass of Bromthymol Blue used (g) 0.161 Molar mass of Bromthymol blue (g/mol) 624.38 Moles of Bromthymol blue (mol) 0.0002578 Molarity of solution (mol/L) 0.001031 standard concentration Absorbance Absorbance measured via laboratory solutions (mol/L) measured via ROXI spectrophotometer (reference method) A 0.00001031 0.48 0.48 B 0.00002062 0.83 0.67 C 0.00003093 1.41 0.96 D 0.00004124 2 1.28 E 0.00005155 2.47 1.60 Roxi Colorimeter Spectrophotometer @ 630 nm 3 y = 49989x - 0.11 2 y = 27430x + 0.148 2.5 R² = 0.9934 R² = 0.9902 2 1.5 1.5 1 Abs./ A Abs./ 1 Abs./ A Abs./ 0.5 0.5 0 0 0 0.00002 0.00004 0.00006 0 0.00002 0.00004 0.00006 Conc. / M Conc. / M laboratory Unknown analysis ROXI spectrophotometer (reference method) Absorbance for unknown solution 1.1141 0.8110 Unknown concentration from 0.00002445 0.00002417 calibration curve (mol/L) 4. Conclusion & Discussion. From the results of the analysis and the graph, it can be seen that the concentration was linearly proportional to absorbance as expected. The unknown solution tested had a true concentration of 2.47 x 10 -5 M, very close to the measured values of 2.445 x 10-5 M and 2.417 x 10 -5 M obtained by the ROXI colorimeter and the laboratory spectrometer. The concentration of an unknown sample could be determined with good accuracy (here 1%) using the ROXI colorimeter. However, the ROXI colorimeter data yielded a slope of the Beer’s law plot considerably different from the laboratory spectrophotometer. The cause of the discrepancy is likely that the laboratory spectrophotometer was operated at 630 nm, a slightly larger wavelength than the LED. The slightly larger wavelength is on the rapidly falling edge of the absorption spectrum for the dye used. This will lead to a smaller slope of the Beer’s law plot. Laboratory Application II. Fluorescence measurements. Introduction. For a demonstration of the ROXI fluorescence measurement we have prepared several different standard solutions of sodium fluorescein dye, placed them within a cuvette and illuminated the samples using the blue LED channel at λ=467 nm. The #735 Velvet Green (Lee Filters) plastic filter matches the emission spectrum of fluorescein well, and this filter was placed in front of the fluorescence detector to reduce excitation light. It was found that the observed signal voltage plotted vs. the logarithm of the concentration of fluorescein produced a linear response that could be used for simple analysis. Emission Filter Spectrum Sodium Fluorescein excitation and emission spectra Wavelength (nm) Procedure. 1. Preparation of standard solution. Place approx. 0.025g sodium fluorescein in a 0.25L volumetric flask. Dilute with D.I. water (approx. conc.= 100 µg / ml). 2. Preparation of standards for analysis. Transfer 0.01mL, 0.02 mL, 0.1 mL, 0.2 mL, 0.35 mL, 0.5 mL, 1mL from the stock solution to a 10 mL volumetric flask labeled A, B, C, D, E, F, G, respectively.
Recommended publications
  • WPA CO 7500 Colorimeter User Manual
    WPA CO 7500 Colorimeter User Manual Biochrom Ltd Certificate No. 890333 Declaration of Conformity This is to certify that the WPA CO 7500 Colorimeter Part number 80-3000-43 (mains only) 80-3000-44 (mains / battery) manufactured by Biochrom Ltd. conform to the requirements of the following Directives-: 73/23/EEC & 89/336/EEC Standards to which conformity is declared EN 61 010-1: 2001 Safety requirements for electrical equipment for measurement, control and laboratory use. EN 61326: 1998 Electrical equipment for measurement, control and laboratory use – EMC requirements Signed: Dated: 24th March 2003 David Parr Managing Director Biochrom Ltd Postal address Telephone Telefax Biochrom Ltd +44 1223 423723 +44 1223 420164 22 Cambridge Science Park Milton Road e mail: [email protected] website: http://www.biochrom.co.uk Cambridge CB4 0FJ England Registered in England No: 3526954 Registered Office: 22 Cambridge Science Park, Milton Road, Cambridge CB4 0FJ, England. CONTENTS Unpacking, Positioning and Installation 1 OPERATION 2 Introduction 2 Using the Instrument 3 Making an absorbance or %T measurement 4 Making a kinetics measurement 4 TROUBLE SHOOTING NOTES 5 ACCESSORIES, CONSUMABLES AND SPARES 6 OUTPUT OF RESULTS 6 Use with serial printer 6 Use with PC 6 Use with chart recorder 6 MAINTENANCE 7 General maintenance 7 Changing a filter 7 Replacing the light bulb 8 SPECIFICATION AND WARRANTY 9 Unpacking, Positioning and Installation • Ensure your proposed installation site conforms to the environmental conditions for safe operation: Indoor use only Temperature 5°C to 35°C Maximum relative humidity of 80 % up to 31°C decreasing linearly to 50 % at 40°C If this equipment is used in a manner not specified or in environmental conditions not appropriate for safe operation, the protection provided by the equipment may be impaired and instrument warranty withdrawn.
    [Show full text]
  • A Guide to Colorimetry
    A Guide to Colorimetry Sherwood Scientific Ltd 1 The Paddocks Cherry Hinton Road Cambridge CB1 8DH England www.sherwood-scientific.com Tel: +44 (0)1223 243444 Fax: +44 (0)1223 243300 Registered in England and Wales Registration Number 2329039 Reg. Office as above Group I II III IV V VI VII VIII Period 1A 8A 1 2 1 H 2A 3A 4A 5A 6A 7A He 1.008 4.003 3 4 5 6 7 8 9 10 2 Li Be B C N O F Ne 6.939 9.0122 10.811 12.011 14.007 15.999 18.998 20.183 11 12 13 14 15 16 17 18 3 Na Mg 3B 4B 5B 6B 7B [---------------8B-------------] 1B 2B Al Si P S Cl Ar 22.99 24.312 26.982 28.086 30.974 32.064 35.453 39.948 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 4 K Ca Sc Ti V Cr Mn Fe Co Ni Cu Zn Ga Ge As Se Br Kr 39.102 40.08 44.956 47.9 50.942 51.996 54.938 55.847 58.933 58.71 63.546 65.37 69.72 72.59 74.922 78.96 79.904 83.8 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 5 Rb Sr Y Zr Nb Mo Tc Ru Rh Pd Ag Cd In Sn Sb Te I Xe 85.47 87.52 88.905 91.22 92.906 95.94 [97] 101.07 102.91 106.4 107.87 112.4 114.82 118.69 121.75 127.6 126.9 131.3 55 56 57* 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86 6 Cs Ba La Hf Ta W Re Os Ir Pt Au Hg Ti Pb Bi Po At Rn 132.91 137.34 138.91 178.49 180.95 183.85 186.2 190.2 192.2 195.09 196.97 200.59 204.37 207.19 208.98 209 210 222 87 88 89** 104 105 106 107 108 109 110 111 112 114 116 7 Fr Ra Ac Rt Db Sg Bh Hs Mt 215 226.03 227.03 [261] [262] [266] [264] [269] [268] [271] [272] [277] [289] [289] 58 59 60 61 62 63 64 65 66 67 68 69 70 71 * Lanthanides Ce Pr Nd Pm Sm Eu Gd Tb Dy Ho Er Tm Yb Lu 140.12 140.91 144.24 145 150.35 151.96 157.25 158.92 152.5 164.93 167.26 168.93 173.04 174.97 90 91 92 93 94 95 96 97 98 99 100 101 102 103 ** Actinides Th Pa U Np Pu Am Cm Bk Cf Es Fm Md No Lr 232.04 231 238 237.05 239.05 241.06 244.06 249.08 251 252.08 257.1 258.1 259.1 262.11 The ability to analyse and quantify colour in aqueous solutions and liquids using a colorimeter is something today’s analyst takes for granted.
    [Show full text]
  • AMSA Meat Color Measurement Guidelines
    AMSA Meat Color Measurement Guidelines Revised December 2012 American Meat Science Association http://www.m eatscience.org AMSA Meat Color Measurement Guidelines Revised December 2012 American Meat Science Association 201 West Springfi eld Avenue, Suite 1202 Champaign, Illinois USA 61820 800-517-2672 [email protected] http://www.m eatscience.org iii CONTENTS Technical Writing Committee .................................................................................................................... v Preface ..............................................................................................................................................................vi Section I: Introduction ................................................................................................................................. 1 Section II: Myoglobin Chemistry ............................................................................................................... 3 A. Fundamental Myoglobin Chemistry ................................................................................................................ 3 B. Dynamics of Myoglobin Redox Form Interconversions ........................................................................... 3 C. Visual, Practical Meat Color Versus Actual Pigment Chemistry ........................................................... 5 D. Factors Affecting Meat Color ............................................................................................................................... 6 E. Muscle
    [Show full text]
  • C201, C301 and C401 Multiparameter Colorimeters Multiparameter Measurement for Convenient Field Testing
    Colorimeters – Optical C201, C301 and C401 Multiparameter Colorimeters Multiparameter measurement for convenient field testing Convenient, foil pack reagents – Follows US EPA method 330.5 for wastewater, and Standard Method 4500-Cl G for drinking water Simple to use – One-time blanking for multiple measurements Waterproof and dustproof IP67 housing – Even the sample chamber is waterproof Colorimeters include Compact size – Take your Oakton colorimeter case and reagents anywhere! No detachable parts and no light shield required – Vial cap prevents stray light from affecting readings Specifications Large, easy-to-read display – Large, custom Model C201 colorimeter C301 colorimeter C401 colorimeter LCD displays readings with units of measure and Chlorine (free and total) user-friendly message codes that guide meter Range 0 to 1.99 ppm; 2.0 to 6.0 ppm 0 to 1.99 ppm; 2.0 to 6.0 ppm 0 to 1.99 ppm; 2.0 to 6.0 ppm operation Resolution 0.01 ppm; 0.1 ppm 0.01 ppm; 0.1 ppm 0.01 ppm; 0.1 ppm Advanced power-supply management – Accuracy ±0.02 ppm, ±0.2 ppm ±0.02 ppm, ±0.2 ppm ±0.02 ppm, ±0.2 ppm Measures over 1200 samples with a single set of pH batteries, delivering quick, stable results in less Range — 5.9 to 8.2 pH 5.9 to 8.2 pH than six seconds. Also features auto-off function. Resolution — 0.1 pH 0.1 pH Accuracy — ±0.1 pH ±0.1 pH Sturdy carrying case with Cyanuric acid accessories – Compact carrying case Range — — 5 to 90 ppm includes all items necessary for colorimeter Resolution — — 1 ppm measurements and protects the meter when Accuracy — —
    [Show full text]
  • Experiment 7
    Experiment 11 Beer’s Law OUTCOMES After completing this experiment, the student should be able to: determine the wavelength (color) of maximum absorbance for a solution. examine the relationship between the absorbance and concentration of a solution. use absorbance data to determine the concentration of an unknown solution. DISCUSSION Colored compounds obtain their particular colors due to the wavelengths of light that they absorb and reflect (or transmit). Thus, a red solution transmits red light and absorbs other wavelengths. If one graphs how much light is absorbed by a compound at each wavelength, an absorption spectrum is obtained for the compound showing areas of high and low absorbance. For our red solution, the red wavelengths will have low absorbances (high transmittance) and the blue or green wavelengths will likely have higher absorbances (lower transmittance). If we focus on one particular wavelength for a compound, we can use the absorbance of light to determine the concentration of a solution. In general, a higher concentration of a colored solution has more light-absorbing compounds and so it will absorb more light (or transmit less light) than a solution of lower concentration. Thus, when a graph of absorbance vs. concentration is plotted for solutions of known concentration, a direct relationship should result as shown in Figure 1. This direct relationship is known as Beer’s Law and is shown by the equation A = εcl where A is the measured absorbance, ε is the molar absorptivity (a constant for the particular solute you are analyzing that is related to how strongly the solute absorbs light of a particular wavelength), c is the molar concentration of the solute, and l is the path length or the distance that the light travels through the cuvette (usually 1 cm).
    [Show full text]
  • POCKET COLORIMETER™ Analysis System Instruction Manual
    46760-88 POCKET COLORIMETER™ Analysis System Chlorine (Cl2) Instruction Manual © Hach Company, 1991–2001. All rights reserved. Printed in the U.S.A. te/dk 6/01 8ed 2 TABLE OF CONTENTS CERTIFICATION ................................................................................................. 5 SAFETY PRECAUTIONS .................................................................................. 9 Use of Hazard Information .................................................................................. 9 Precautionary Labels ......................................................................................... 10 SPECIFICATIONS ............................................................................................. 11 OPERATION ........................................................................................................ 13 GENERAL DESCRIPTION ............................................................................. 15 Safety Precautions ............................................................................................. 18 Battery Installation ............................................................................................ 18 Operation ........................................................................................................... 20 CALIBRATION ................................................................................................... 23 User-Entered Calibration ................................................................................... 23 Instrument Calibration
    [Show full text]
  • Standard Operating Procedures
    Sassafras Samplers Monitoring Program Standard Operating Procedures 1.0 Introduction The Standard Operating Procedures (SOP) outlined in this manual are designed to help guide and sustain a high quality water monitoring program. The procedures are utilized by the, RIVERKEEPER®, ShoreRivers employees and interns, and volunteers to assure that each individual or sampling crew follow the same protocols. Quality assurance in a monitoring program is essential for providing representative and accurate data for a water body, which is vital for the continuation of a monitoring program. The SOP’s are listed by section, with added instructions for safety and tips on keeping up proper maintenance of testing equipment. All volunteers are to follow these guidelines and procedures to ensure reliable, high quality data. The RIVERKEEPER®, is to use the SOP as a tool for training new volunteers during an initial training class and during coordinated periodic training sessions. 2.0 Safety Safety to staff and volunteers is of the upmost importance to ShoreRivers. There are hazards in both the field and the lab, so please use caution. For instance, during inclement weather conditions, the sampling site may become a hazard due to wet and/or slippery conditions, high winds, or torrential water that can be dangerous to the individuals sampling. It is also recommended that more than one person be on site in case of an emergency. If at any time a volunteer feels that unsafe conditions exist, they are advised to terminate that activity immediately. The tests that are performed on water samples require the use of reagents that can become hazardous if handled incorrectly.
    [Show full text]
  • Borrelia Igg ELISA Kit ( Cat # KA1465 V.01 ) For
    www.abnova.com Intended Use The kit is intended for the detection of IgG antibodies to the pathogenic borrelia strains ( B.afzelii, B. garinii and B. burgdorferi sensu stricto ) in human serum and cerebrospinal fluid. The detection of antibodies is one of the laboratory tests that help to diagnose Lyme disease (LD). The diagnosis of LD is based on the combination of clinical examination and laboratory testing. Anti-borrelia IgG antibodies are detectable 6 weeks after infection. The level of IgG antibodies usually increases with disease progression to the 2 nd stage. During the 3 rd stage of the disease the antibody levels in serum are high and can remain elevated for several years. However, clinical symptoms of LD are similar to other diseases, therefore the serological methods are of use in differential diagnosis of neuroinfections, arthropathies, carditis and skin diseases. Test Principle Borrelia IgG ELISA assay is a solid-phase immunoanalytical test. The polystyrene strips are coated with the mixture of recombinant antigens. Anti-borrelia antibodies in serum samples bind to the immobilized antigens. The serum antibodies that do not bind are washed away and those that formed complexes with the antigens are later on recognised by animal anti-human IgG antibodies labelled with horseradish peroxidase. The presence of labelled antibodies is revealed by an enzymatic reaction with a chromogenic substrate. Negative sera do not react and the mild change in colour, if present, may be attributed to the reaction background. Kit Components ELISA break-away strips (turquise) coated with specific recombinant antigens 96 wells 1.3 mL Standard A, 3 AU/mL 1) , r.t.u.
    [Show full text]
  • Colorimeter a  K  C (Beer’S Law) (Order Code COL-BTA) Where K Is a Proportionality Constant
    with a constant cell width, one can assume and b to be constant. This leads to the equation Colorimeter A k C (Beer’s law) (Order Code COL-BTA) where k is a proportionality constant. This equation shows absorbance to be related directly to concentration and represents a mathematical statement of Beer’s law. The Vernier Colorimeter is designed to determine the Beer’s law is discussed in more detail below. concentration of a solution by analyzing its color In this guide and in some of our computer programs, transmittance is expressed as intensity. The color of a solution may be inherent or percent transmittance or %T. Since T=%T/100, the formula can be rewritten as derived by adding another reagent to it. The A log(100 / %T ) or A 2 log%T Colorimeter measures the amount of light transmitted through a sample at a user-selectable wavelength. Beer’s Law Using the front panel arrow keys, you may choose from four wavelengths: 430 nm, In general, absorbance is important 470 nm, 565 nm, and 635 nm. Features such as automatic sensor identification and because of its direct relationship with one-step calibration make this sensor easy to use. concentration according to Beer’s law. Many experiments in chemistry How the Colorimeter Works and biology are based on this Light from a LED light source passes through a Source Detector concept. To obtain a Beer’s law cuvette containing a solution sample, as shown in curve, several standards (solutions of Figure 1. Some of the incoming light is absorbed LED Transmitted known concentration) are prepared light by the solution.
    [Show full text]
  • Colorimeter Series User's Manual
    TTi ® Colorimeter Series User’s Manual Taylor Technologies, Inc. 800-TEST KIT (837-8548) www.taylortechnologies.com ® TTi Colorimeter Series User’s Manual TTi ® Colorimeter Series User’s Manual First Edition last revision to this edition 6/12/2017 Taylor Technologies, Inc. Sparks, Maryland Taylor Technologies, Inc., Sparks, Maryland 21152 © 2012 Taylor Technologies, Inc. All rights reserved. First edition published online 2012 For ready reference by the reader, the date of the last revision to the manual is posted on the title page. The latest revision date for each chapter or subsection is posted in parentheses after its title in the table of contents. Check these periodically to make sure your file copy is up to date. TABLE OF CONTENTS PART 1 GENERAL INFORMATION Safety and General Precautions (4/21/15) .............................................................................................1-3 Colorimeter Description ........................................................................................................................1-3 Features (10/19/12) .......................................................................................................................1-3 Optional Chemical Forms (4/19/13) ................................................................................................1-4 Components and Optional Accessories (6/12/17) ...................................................................................1-5 Colorimeter Reagent Packs (6/12/17) ....................................................................................................1-5
    [Show full text]
  • Instruction Manual C401/301/201/101/102/103/104/105
    Instruction Manual C401/301/201/101/102/103/104/105 Portable Colorimeter 68X357704 Rev 1 05/05 Technology Made Easy ... Preface This manual functions in two ways: first as a step by step guide to help you operate the waterproof C401/301/201/101/102/103/104/105 Colorimeter; second, it serves as a handy reference guide. It contains basic instructions that you must follow during the operation, care and maintenance of the instrument. The safety protection provided by this equipment may be impaired if it is used in a manner not described in this manual. It is recommended that all operators should read this manual prior to working with this instrument. Eutech Instruments/ Oakton Instruments cannot accept any responsibility for damage or malfunction to the meter caused by improper use of the instrument. The information presented in this manual is subject to change without notice as improvements are made, and does not represent a commitment on the part of Eutech Instruments Pte Ltd/ Oakton Instruments. Note: Eutech Instruments Pte Ltd/ Oakton Instruments reserves the right to make improvements in design, construction, and appearance of our products without notice. Copyright © 2004 All rights reserved. Eutech Instruments Pte Ltd Oakton Instruments Rev. 1 05/05 TABLE OF CONTENTS 1 INTRODUCTION..........................................................................................................................1 1.1 UNPACKING C401/301/201/101/102/103/104/105 COLORIMETER AND ACCESSORIES .............................2 1.2 DISPLAY .................................................................................................................................................3
    [Show full text]
  • Jenway Catalogue 2012
    Double Beam Spectrophotometers Equipment forAnalysis Life Science Spectrophotometers UV/Visible Spectrophotometers pH Meters Ion Meters Flame Photometers Fluorimeters Dissolved Oxygen Meters Conductivity Meters Colourimeters Bibby Scientific Limited Some of the most famous names in science... As one of the largest broad based manufacturers of benchtop laboratory equipment worldwide, Bibby Scientific Ltd provides internationally recognised brands with reputations for product quality and high performance. These four famous brands are now brought together in a single package to offer an excellent level of quality, service and support. A Bibby Scientific Company Electrothermal are the newest addition to the Bibby Scientific portfolio and are market leaders in heating mantle design and manufacture. The extensive Electrothermal range includes controlled, stirring, Bunsen and spill-proof mantles in various shapes and capacities. Alongside the heating mantle range, Electrothermal offer an extensive selection of stirrers and melting point apparatus. Jenway® manufactures a wide range of analytical scientific instruments including UV/Vis spectrophotometers, flame photometers, colorimeters, portable and laboratory meters for the measurement of dissolved oxygen, pH, conductivity and specific ions. The extensive Stuart® range includes blood tube rotators, colony counters, hotplates, hybridisation ovens, rockers, shakers, stirrers and water purification systems. Techne® is a world leader in the manufacture of temperature control equipment, including
    [Show full text]