RF Calibrator (Ver.2), Disease Control, Atlanta, GA

Total Page:16

File Type:pdf, Size:1020Kb

RF Calibrator (Ver.2), Disease Control, Atlanta, GA KAMIYA BIOMEDICAL COMPANY further information, refer to "Decontamination of Laboratory Materials Required But Not Supplied Sink Drains to Remove Azide Salts," in the Manual Guide- K-ASSAY Safety Management No. CDC-22 issued by the Center for Calibrators: K-ASSAY RF Calibrator (Ver.2), Disease Control, Atlanta, GA. Cat. No. KAI-231C. RF (Ver.2) Purified water REAGENT PREPARATION For the Quantitative Determination of Human Rheumatoid Factor (RF) in Serum and Plasma Two Reagent Clinical Chemistry Analyzer: Reagents are ready to use and do not require reconstitution. Capable of accurate absorbance readings Cat. No. KAI-230 Capable of accurately dispensing the required volumes STORAGE AND HANDLING Capable of maintaining 37°C INTENDED USE calibrators with known concentrations of proteins and using the instrument’s data reduction capability or All reagents should be stored refrigerated (2-8°C). Return all Assay Procedure The K-ASSAY RF (Ver.2) assay is for the quantitative manually plotting the change in absorbance versus reagents to 2-8°C promptly after use. Unopened reagents determination of human IgG rheumatoid factor antibodies in concentration. Concentration of the controls and patient can be used until the expiration date indicated on the Note: Allow all reagents and specimens to warm to room patient serum or plasma (citric acid, EDTA, or lithium samples are interpolated from the calibration curve. package and bottle labels. temperature. Mix all reagents gently before using. heparin) based on immunoturbidimetric assay. The presence The K-ASSAY RF (Ver.2) assay should be run using the of IgG RF antibodies, in conjunction with clinical findings and REAGENT STABILITY An example of automated application: K-ASSAY RF Calibrator (Ver.2). These 6 calibrators are other laboratory tests, is an aid in the diagnosis of used to prepare a calibration curve for quantifying the levels rheumatoid arthritis (RA). FOR IN VITRO DIAGNOSTIC Discard reagents if they become contaminated. Evidence of Sample 3.0 µL of rheumatoid factor present in the patient's serum or plasma USE. cloudiness or particulate material in solution is cause to ↓ sample. discard. • ← R1 (Buffer Reagent) 100 µL INTRODUCTION AND SUMMARY ↓ 37 °C, 5 min. KIT COMPOSITION Opened reagents can be used for 1 month if stored at • ← R2 (Latex Suspension) 100 µL 2-8°C. The K-ASSAY RF (Ver.2) assay is intended for the ↓ 37 °C, 5 min. Reagents (Liquid Stable) quantitative determination of human IgG rheumatoid factor 2-point endpoint, 700 nm INSTRUMENT (RF) antibodies by immunoturbidimetric assay. The R1: Buffer Reagent 1 x 20 mL rheumatoid factor (an autoimmune antibody in the patient's Automated Method (Example) Phosphate Buffer Measurement of absorbance is to be made with an serum or plasma) interacts with the aggregated human IgG instrument able to accurately read absorbances. Refer to the in the reagent forming immune complexes. The immune Chemistry Parameters for Automatic Analyzer R2: Latex Suspension 1 x 20 mL instrument manual from the manufacturer regarding the complexes cause an increase in light scattering, which Suspension of latex particles sensitized following: INSTRUMENT Abbott Architect c8000 correlates with the concentration of rheumatoid factor. with human γ globulin (0.7 mg/mL) ASSAY NAME ( RFv2 ) Autoantibodies of the IgG, IgM, or IgA isotype, which are Phosphate Buffer a) Use or function ASSAY TYPE ( PHOTOMETRIC ) reactive with the crystallizable fraction (Fc) of IgG, are called b) Installation procedures and requirements ASSAY AVAILABILITY ( ENABLED ) rheumatoid factors. Rheumatoid factor is found in 50-79% of WARNINGS AND PRECAUTIONS c) Principles of operation REACTION MODE ( END UP ) adults with classical rheumatoid arthritis. Quantification of d) Performance characteristics, operating instructions WAVELENGTH ( 700 ) / ( NONE ) rheumatoid factor has been shown to be useful in the clinical FOR IN VITRO DIAGNOSTIC USE. Rx only. e) Calibration procedures including materials and / or 1,2,3,4,5,6,7,8 LAST READ REQUIRED ( 33 ) diagnosis and prognosis of rheumatoid arthritis. equipment to be used Not to be used internally in humans or animals. Normal ABSORBANCE RANGE ( ) ( ) f) Operational precautions, limitations, and hazards Rheumatoid factor has been measured using a variety of precautions exercised in handling laboratory reagents should SAMPLE BLANK TYPE ( SELF ) g) Service and maintenance information methods, including agglutination, latex fixation, be followed. FLEX READ TIME ( 32 ) - ( 33 ) nephelometric, and enzyme-linked immunoabsorbent COLOR CORRECTION 9,10 Do not mix or use reagents from one test kit with those from SPECIMEN COLLECTION AND PREPARATION ( ) - ( ) assay. The K-ASSAY RF (Ver.2) uses a latex-enhanced READ TIME a different lot number. immunoturbidimetric assay format. It is recommended that specimen collection be carried out in BLANK READ TIME ( 19 ) - ( 20 ) Do not use reagents past their expiration date stated on each accordance with CLSI document M29-A3. No method can REAGENT ( RF ) PRINCIPLE OF TEST reagent container label. offer complete assurance that human blood samples will not DILUENT ( NONE ) Do not pipette by mouth. Avoid ingestion and contact with transmit infection. Therefore, all blood samples should be DILUTION DISP. MODE ( TYPE 1 ) The K-ASSAY RF (Ver.2) assay quantifies the skin. considered potentially infectious. REAGENT VOLUME R1:( 100 ) R2:( 100 ) rheumatoid factor in the patient's serum or plasma based WATER VOLUME R1:( ) R2:( ) on latex-enhanced immunoturbidimetric assay. Reagents contain human source material. Each donor unit Serum or plasma (EDTA or lithium heparin) test samples DISPENSE MODE ( TYPE 1 ) ( TYPE 1 ) Calibrators, controls, and patient samples are pipetted into used in the preparation of this product has been tested by an must be collected in the manner routinely used for clinical DILUTION NAME ( STANDARD ) sample cups. Microvolumes of samples and reagent FDA approved method and found non-reactive for HBsAg laboratory tests. Freshly drawn serum is preferred and SAMPLE ( 3.0 ) diluent are automatically pipetted into individual cuvettes. and antibody to HIV. However, it is not possible to guarantee should be used within the day of collection. Samples may that any human source material is free of these or other also be stored refrigerated (4-8°C) for 8 days or at -20°C for DILUTION FACTOR ( ) Following an initial incubation and measurement of sample 11 infectious agents. up to 3 months. Avoid repeat freeze/thawing of specimens. DEFAULT DILUTION ( SELECTED ) blank, gamma globulin containing aggregated human IgG REACTION CHECK ( RATE RATIO ) is added to the cuvettes. The sample (autoantibodies) Therefore, all products containing human source material Use plastic tubes for storing the samples, do not use MAX. ABS. VARIATION ( 0.01 ) solution and gamma globulin reagent (antigen) are then should be handled in accordance with good laboratory glass. CALIBRATION METHOD ( SPLINE ) mixed in the reaction cuvettes. Insoluble antigen-antibody practices and appropriate control. See the National Institute REPLICATES ( 3 ) (immune) complexes form. The immune complexes cause of Health Manual, "Biosafety in Microbiology and Biomedical PROCEDURE RESULT CONC. UNITS ( IU / ML ) an increase in light scattering, which correlates with the Laboratories," 2nd ed., 1988. concentration of rheumatoid factor. Following an Materials Supplied RF Calibrators (Ver.2) includes 6 levels to be used as Reagents in this kit contain < 0.1% w/v sodium azide as a incubation period lasting approximately 5 minutes, the calibrators. Input the concentrations for each calibrator. preservative. Sodium azide may form explosive compounds absorbance change of the solution is measured. Reagent 1 (R-1) Buffer Reagent 1 x 20 mL in metal drain lines. When disposing of reagents through Parameters for other automated analyzers are available. Reagent 2 (R-2) Latex Suspension 1 x 20 mL A calibration curve is generated by assaying a series of plumbing fixtures, flush with copious amounts of water. For K-ASSAY RF (Ver.2) 1 Rev. 2020-10-29 CALIBRATION Multi-Site Precision Sample INTERFERENCE LABELING SYMBOLS 1 2 3 4 For samples with RF concentrations less than 600 IU/mL, a Testing was performed on an Abbott Architect c8000 Lot Number multi-point calibration curve using the K-ASSAY RF Mean (IU/mL) 14.13 21.24 101.69 532.83 analyzer according to the CLSI EP07-3rd edition guideline Calibrator (Ver.2), Cat. No. KAI-231C, should be used. In our N 75 75 75 75 with the following results. Reagent testing on an Abbott Architect c8000, the calibration curve Expiration or “Use By” Date was stable for 43 days. It is recommended that the user S.D. 0.64 0.54 0.75 4.44 Repeatability Ascorbic Acid No interference up to 50 mg/dL confirm calibration frequency as this may depend on the C.V. % 4.5 2.5 0.7 0.8 Bilirubin, Conjugated No interference up to 20 mg/dL Catalog Number instrument and type/number of other assays being performed. S.D. 1.02 1.03 1.61 7.86 Bilirubin, Unconjugated No interference up to 20 mg/dL In Vitro Reproducibility For Diagnostics Use C.V. % 7.2 4.9 1.6 1.5 Chyle No interference up to 1,500 FTU 2-8 °C Temperature Limitation. TRACEABILITY Hemoglobin No interference up to 500 mg/dL Store between 2 and 8 degrees C Accuracy / Correlation The values of the K-ASSAY RF Calibrator (Ver.2) are Total Cholesterol No interference up to 400 mg/dL Potential Human Biohazard traceable to the NIBSC Rheumatoid Arthritis Serum, Testing was performed on an Abbott Architect c8000 Total Triglycerides No interference up to 1,000 mg/dL 64/002. Based upon this standardization, results are analyzer according to the CLSI EP9-A3 guideline. A Manufacturer reported in International Units (IU/mL). comparison of the K-ASSAY RF (Ver.2) assay and EXPECTED VALUE company A’s RF assay was performed with the following Consult Package Insert for Instructions for Use QUALITY CONTROL results: The expected value as reported is < 11 IU/mL.
Recommended publications
  • Retrospective Approach to Evaluate Interferences in Immunoassay
    Retrospective approach to evaluate interferences in immunoassay Jayesh Warade Laboratory Services, Meenakshi Mission Hospital and Research Center, Madurai, Tamilnadu, India INFO ABSTRACT Corresponding author: Background Jayesh Warade Laboratory Services Despite the increase in sensitivity and specificity of Meenakshi Mission Hospital immunoassay technique over years, analytical inter- and Research Center Madurai, Tamilnadu ference remains to be major area of concern. India E-mail: [email protected] The interfering substances are endogenous substances that are natural, polyreactive antibodies as heterophil- Key words: ic or auto antibodies, or human anti-animal antibodies heterophilic - antibody, immunoassay, interferences, cross reactive together with other unsuspected binding proteins that are unique to the individual. Interfering substances can interfere with the reaction between analyte and reagent antibodies in immunoassay resulting in false positive or negative values. This ultimately results in misinterpretation of patients reports and finally to wrong course of treatment. Objective In our study, we used a retrospective approach to find out the extent of interferences and type of interfer- ences in some cases during our routine practice. Method The immunoassay reports which were clinically not correlating were retrospectively evaluated after dis- cussion with the clinician. Over a period of six month a total of 42 samples were evaluated for interference for Page 224 eJIFCC2017Vol28No3pp224-232 Jayesh Warade Retrospective approach to evaluate interferences in immunoassay different immunoassay parameters such as Beta for prior extraction, immunoassays may lack ad- HCG, Estradiol, CA 125, AFP, prolactin, Hepatitis equate specificity and accuracy (1). B Surface antigen (HbSAg) and troponin I. The Developing immunoassays for the quantifica- samples were treated with commercially avail- tion of an analyte in a buffer solution has its able antibody blocking agents and were reana- own challenges, nonetheless quantification of lyzed.
    [Show full text]
  • Case Reports & Imaging Journal False Negative Cryptococcus Latex
    Annangi S, et al. J Case Repo Imag 2017 1: 002 Case Reports & Imaging Journal Case Report False Negative Cryptococcus Latex Agglutination Test Despite Correcting for Conventional Reasons: A Case Report Srinadh Annangi1, Snigdha Nutalapati1 and Mesfin Fransua2* 1Department of Internal Medicine, Morehouse School of Medicine, Atlanta, GA, USA 2Department of Internal Medicine, Division of Infectious Diseases, Morehouse School of Medicine, Atlanta, GA, USA *Corresponding author: Mesfin Fransua, Department of Internal Abstract Medicine, Division of Infectious Disease, Morehouse School of Med- Cryptococcus meningitis needs high index of suspicion given its icine, 720, Westview Dr, Atlanta, GA, USA, Tel: +1 4047561366; subtle presentation. Though the gold standard for diagnosis is cul- E-mail: [email protected] ture of cerebrospinal fluid, given the delay in culture results, treat- ment is usually initiated with positive latex agglutination test or Indian Received Date: November 28, 2016 ink staining. A 50 years old male with HIV/AIDS (Human immuno- deficiency virus infection/ acquired immune deficiency syndrome) Accepted Date: March 28, 2017 non compliant with antiretroviral therapy presented complaining of throbbing non radiating occipital headache. Physical examination Published Date: April 12, 2017 is unremarkable and computerized tomography of head showed no evidence of mass lesions or raised intracranial pressure. Cryptococ- Citation: Annangi S, Nutalapati S, Fransua M (2017) False Negative cal meningitis is suspected especially given CD4 count of 19 cells/ Cryptococcus Latex Agglutination Test Despite Correcting for Conven- mcL. Cerebrospinal Fluid (CSF) Indian ink stains and Cryptococcus tional Reasons: A Case Report. J Case Repo Imag 1: 002. antigen latex agglutination test from pronase treated CSF sample was negative.
    [Show full text]
  • (P30) Test Because Spermatozoa Production
    ~. ANALYSIS OF SEMINAL FLUID IDENTIFICATION OF SEMEN-SPECIFIC PROTEIN p30 BY ABAcard@PSA (p30) Test Because spermatozoa production in males can be affected by aspermia, vasectomy or other conditions, forensic scientists have recognized the need for seminal fluid diagnostic tests which did not rely upon sperm cell presence. p30 is a 30,000 Dalton semen glycoprotein of prostatic origin that is also known as PSA or prostate specific antigen. The range of PSA is 200,000 to 5.5 million ng/ml of semen. The sensitivity of the ABAcard@PSA test is 4 ng/ml and therefore seminal fluid diluted up to 1 in a million should be detectable. The ABAcard@PSA test has been tested against numerous biological fluids of men and women. Since no cross reactivity has been reported to date, this supports the hypothesis that P30 is a male-specific protein. For this reason, the detection of the P30 antigen in a forensic stain is strong evidence that the stain is seminal in nature. The determination of the presence of semen may be made by using the ABAcard@ PSA (p30) test. In this test 200~1of sample is added to the sample well'S', and allowed to soak in. If PSA is present in the specimen, it will react with the mobile monoclonal ~ antihuman PSA antibody and a mobile antigen antibody complex is thus formed. This mobile antibody-antigen complex migrates through the absorbent device towards the test area 'T'. In the test area 'T', a polyclonal antihuman PSA antibody is immobilized. This immobilized antibody captures the above complex so that an antibody-antigen- antibody sandwich is formed.
    [Show full text]
  • Topics in Immunopathology 2021
    TOPICS Immunopathology This document forms Appendix 7 to Trainee HandbookTOPICS for INImmunopathology IMMUNOPATHOLOGY and should be read in conjunction with Section 2 of the Handbook The document was prepared in April 2009 by A/Prof David Fulcher, with contributions from Dr Richard Wong, Dr Melanie Wong and A/Prof William Sewell and was revised in March 2011. RCPA - Topics in Immunopathology 1 Functions as a medical specialist in the laboratory 1.1. Knowledge and skills in immunobiology & pathogenesis of immunological disorders 1.1.1. Organisation of the immune system 1.1.1.1. Lymphoid tissues and organs: anatomy and function 1.1.1.2. Cells of relevance to the immune response, ontogeny, characteristics and functions 1.1.1.3. Molecules of relevance to the immune response: MHC, cytokines, chemokines, receptors, signalling pathways, vasoactive mediators, prostaglandins, leukotrienes 1.1.1.4. Genes of relevance to the immune response, including genes for above molecules and MHC, gene rearrangement in generation of antigen receptors 1.1.1.5. Structure and function relationships between the organs, tissues and cells that participate in immune responses 1.1.1.6. Cellular and molecular correlates of immune function 1.1.1.7. Comprehensive understanding of the ontogeny of the immune system and the relationship between the development of immune function to development of allergic diseases, autoimmunity, vasculitides and immunodeficiencies. 1.1.2. Immune mechanisms 1.1.2.1. Innate immunity: 1.1.2.1.1. NK cells 1.1.2.1.1.1. Cytokines relevant in NK cell development and function 1.1.2.1.1.2.
    [Show full text]
  • Technical Guide for ELISA - Protocols - Troubleshooting Table of Contents
    Technical Guide for ELISA - Protocols - Troubleshooting Table of Contents Table of Contents Page 1. Introduction - What’s an ELISA? ................................................................................................................ 4 2. Assay Formats - What are my choices? ...................................................................................................... 5 3. How to Choose an Assay Format - I know what I want to measure. What do I do? ..................................8 4. Factors to Consider When Developing an Assay - There are a lot but a few are critical .......................... 10 5. Performing an Assay - What do I need to know? ..................................................................................... 26 Table of Contents 6. Data Handling - What should I do with all this data? .............................................................................. 28 7. Troubleshooting - I’m having trouble. Now what? ................................................................................. 30 8. Resources - Where can I get more information? ...................................................................................... 34 9. Glossary ................................................................................................................................................. 35 10. Related Products .................................................................................................................................. 38 11. Notes ...................................................................................................................................................
    [Show full text]
  • 2020-JIM-ELISA-Array.Pdf
    Journal of Immunological Methods xxx (xxxx) xxxx Contents lists available at ScienceDirect Journal of Immunological Methods journal homepage: www.elsevier.com/locate/jim Adaption of a conventional ELISA to a 96-well ELISA-Array for measuring the antibody responses to influenza virus proteins and vaccines ⁎ Eric Waltaria, , Esteban Carabajala, Mrinmoy Sanyalb, Natalia Friedlandb, Krista M. McCutcheona a Chan Zuckerberg Biohub, San Francisco, CA, USA b Stanford ChEM-H and Department of Biochemistry, Stanford University School of Medicine, Stanford, CA, USA ARTICLE INFO ABSTRACT Keywords: We describe an adaptation of conventional ELISA methods to an ELISA-Array format using non-contact Piezo ELISA-Array printing of up to 30 spots of purified recombinant viral fusion proteins and vaccine on 96 well high-protein Infectious disease binding plates. Antigens were printed in 1 nanoliter volumes of protein stabilizing buffer using as little as 0.25 Protein array nanograms of protein, 2000-fold less than conventional ELISA. The performance of the ELISA-Array was de- Titer monstrated by serially diluting n = 9 human post-flu vaccination plasma samples starting at a 1/1000 dilution and measuring binding to the array of Influenza antigens. Plasma polyclonal antibody levels were detected using a cocktail of biotinylated anti-human kappa and lambda light chain antibodies, followed by a Streptavidin- horseradish peroxidase conjugate and the dose-dependent signal was developed with a precipitable TMB sub- strate. Intra- and inter-assay precision of absorbance units among the eight donor samples showed mean CVs of 4.8% and 10.8%, respectively. The plasma could be differentiated by donor and antigen with titer sensitivities 3 6 4 6 ranging from 1 × 10 to 4 × 10 , IC50 values from 1 × 10 to 9 × 10 , and monoclonal antibody sensitivities in the ng/mL range.
    [Show full text]
  • Targeting HIV-1 Protease Autoprocessing for High
    www.nature.com/scientificreports OPEN Targeting HIV-1 Protease Autoprocessing for High- throughput Drug Discovery and Received: 3 August 2018 Accepted: 23 November 2018 Drug Resistance Assessment Published: xx xx xxxx Liangqun Huang1, Linfeng Li2, ChihFeng Tien1, Daniel V. LaBarbera2 & Chaoping Chen1 HIV-1 protease autoprocessing liberates the free mature protease from its Gag-Pol polyprotein precursor through a series of highly regulated autoproteolysis reactions. Herein, we report the development and validation (Z’ ≥ 0.50) of a cell-based functional assay for high-throughput screening (HTS) of autoprocessing inhibitors using fusion precursors in combination with AlphaLISA (amplifed luminescent proximity homogeneous assay ELISA). Through pilot screening of a collection of 130 known protease inhibitors, the AlphaLISA assay confrmed all 11 HIV protease inhibitors in the library capable of suppressing precursor autoprocessing at low micromolar concentrations. Meanwhile, other protease inhibitors had no impact on precursor autoprocessing. We next conducted HTS of ~23,000 compounds but found no positive hits. Such high selectivity is advantageous for large-scale HTS campaigns and as anticipated based on assay design because a positive hit needs simultaneously to be nontoxic, cell permeable, and inhibiting precursor autoprocessing. Furthermore, AlphaLISA quantifcation of fusion precursors carrying mutations known to cause resistance to HIV protease inhibitors faithfully recapitulated the reported resistance, suggesting that precursor autoprocessing is a critical step contributing to drug resistance. Taken together, this reported AlphaLISA platform will provide a useful tool for drug discovery targeting HIV-1 protease autoprocessing and for quantifcation of PI resistance. HIV-1 protease (PR) is one of the three viral encoded enzymes essential for productive viral replication.
    [Show full text]
  • Molar Pregnancy with False Negative Urine Hcg: the Hook Effect Pang Y P, Rajesh H, Tan L K
    Case Report Singapore Med J 2010; 51(3) : e58 Molar pregnancy with false negative urine hCG: the hook effect Pang Y P, Rajesh H, Tan L K ABSTRACT of β-hCG, and with no identifiable foetal poles on Urine pregnancy tests are widely used in ultrasonography examination. However, in some emergency departments as the first screening urinary β-hCG immunoassay systems commonly used test for patients of reproductive age presenting for pregnancy screening, the large amount of β-hCG with gynaecological problems in order to rule out secreted in a molar pregnancy may produce a false pregnancy. Conditions, such as complete molar negative result due to a phenomenon known as the high- pregnancy, which produces large amounts of dose “hook effect”. beta human chorionic gonadotropin (β-hCG), We report a case demonstrating this phenomenon. A may cause a false negative result due to an 47-year-old Malay woman was initially diagnosed with oversaturation of the assay system, known as uterine fibroids after her urine pregnancy test was found the “hook effect”. We report a case where the to be negative, but was subsequently found to have a exclusion of pregnancy by urine testing led to hydatidiform mole. the initial misdiagnosis of a molar pregnancy as a degenerative fibroid. Physicians need to CASE REPORT be reminded of the possibility of false negative A 47-year-old multiparous Malay woman presented to results with this commonly used test. Negative the Singapore General Hospital emergency department or inconclusive results in patients with a high with a two-month history of abdominal bloatedness, suspicion of pregnancy should be further mild epigastric discomfort and loss of appetite.
    [Show full text]
  • Previously Unrecognized Non-Reproducible Antibody-Antigen Interactions and Their Implications for Diagnosis of Viral Infections Including COVID-19
    bioRxiv preprint doi: https://doi.org/10.1101/2021.07.20.453011; this version posted July 20, 2021. The copyright holder for this preprint (which was not certified by peer review) is the author/funder. All rights reserved. No reuse allowed without permission. Previously unrecognized non-reproducible antibody-antigen interactions and their implications for diagnosis of viral infections including COVID-19 Jiaojiao Pan1,11†, Lan Yang1†, Yi Deng1,11†, Baoqing Sun2†, Li Zhang3†, Wenya Wu1,11, Jingzhi Li1, Hu Cheng1,11, Yiting Li1, Wenwen Xu1, Jiao Yang1, Yiyue Sun1,11, Hao Fei1, Qinghong Xue4, Youx in Zhou 5, Hui Wang6, Peiyan Zheng2, Hao Chen2, Fengcai Zhu3, Daxin Peng7, Dayong Gu8, Jun Han9, Jiwan Qiu10 and Hongwei Ma1* Affiliations: 1Division of Nanobiomedicine, Suzhou Institute of Nano-Tech and Nano-Bionics, Chinese Academy of Sciences, Suzhou 215123, China. 2Department of Allergy and Clinical Immunology, Guangzhou Institute of Respiratory Health, State Key Laboratory of Respiratory Disease, The First Affiliated Hospital of Guangzhou Medical University, Guangzhou, 510120, China. 3Department of Vaccine Clinical Evaluation, Jiangsu Provincial Center for Disease Prevention and Control, Nanjing 210009, China. 4China Institute of Veterinary Drug Control, Beijing 100081, China. 5Department of Neurosurgery & Brain and Nerve Research Laboratory, The First Affiliated Hospital of Soochow University, Suzhou, Jiangsu 215006, China. 6Department of Medical Laboratory, The Central Hospital of Wuhan, Tongji Medical College, Huazhong University of Science and Technology, Wuhan 430014, China. 7Colledge of Veterinary Medicine, Yangzhou University, Yangzhou, Jiangsu 225009, China. 8Department of Clinical Laboratory, Shenzhen Second People’s Hospital / the First Affiliated Hospital of Shenzhen University Health Science Center, Shenzhen, 518035, China.
    [Show full text]
  • And RBD-Antigen Based Immunoassays
    diagnostics Article Characterization of the Diagnostic Performance of a Novel COVID-19 PETIA in Comparison to Four Routine N-, S- and RBD-Antigen Based Immunoassays Alexander Spaeth 1,†, Thomas Masetto 2,3,† , Jessica Brehm 1, Leoni Wey 3,4, Christian Kochem 3, Martin Brehm 1, Christoph Peter 2 and Matthias Grimmler 3,4,* 1 MVZ Medizinische Labore Dessau Kassel GmbH, Bauhüttenstraße 6, 06847 Dessau-Roßlau, Germany; [email protected] (A.S.); [email protected] (J.B.); [email protected] (M.B.) 2 Institut für Molekulare Medizin I, Heinrich-Heine-Universität Düsseldorf, Universitätsstraße 1, 40225 Düsseldorf, Germany; [email protected] (T.M.); [email protected] (C.P.) 3 DiaSys Diagnostic Systems GmbH, Alte Straße 9, 65558 Holzheim, Germany; [email protected] (L.W.); [email protected] (C.K.) 4 Hochschule Fresenius, University of Applied Sciences, Limburger Straße 2, 65510 Idstein, Germany * Correspondence: [email protected]; Tel.: +49-(0)-6432-9146-486; Fax: +49-(0)-6432-9146-167 † Equal contribution. Abstract: In 2019, a novel coronavirus emerged in Wuhan in the province of Hubei, China. The severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) quickly spread across the globe, Citation: Spaeth, A.; Masetto, T.; causing the neoteric COVID-19 pandemic. SARS-CoV-2 is commonly transmitted by droplet infection Brehm, J.; Wey, L.; Kochem, C.; and aerosols when coughing or sneezing, as well as high-risk exposures to infected individuals by Brehm, M.; Peter, C.; Grimmler, M. face-to-face contact without protective gear. To date, a broad variety of techniques have emerged to Characterization of the Diagnostic assess and quantify the specific antibody response of a patient towards a SARS-CoV-2 infection.
    [Show full text]
  • Cooperative Binding Mitigates the High-Dose Hook Effect
    Edinburgh Research Explorer Cooperative binding mitigates the high-dose hook effect Citation for published version: Roy, RD, Rosenmund, C & Stefan, MI 2017, 'Cooperative binding mitigates the high-dose hook effect', BMC Systems Biology, vol. 11, no. 1, pp. 74. https://doi.org/10.1186/s12918-017-0447-8 Digital Object Identifier (DOI): 10.1186/s12918-017-0447-8 Link: Link to publication record in Edinburgh Research Explorer Document Version: Publisher's PDF, also known as Version of record Published In: BMC Systems Biology General rights Copyright for the publications made accessible via the Edinburgh Research Explorer is retained by the author(s) and / or other copyright owners and it is a condition of accessing these publications that users recognise and abide by the legal requirements associated with these rights. Take down policy The University of Edinburgh has made every reasonable effort to ensure that Edinburgh Research Explorer content complies with UK legislation. If you believe that the public display of this file breaches copyright please contact [email protected] providing details, and we will remove access to the work immediately and investigate your claim. Download date: 02. Oct. 2021 Roy et al. BMC Systems Biology (2017) 11:74 DOI 10.1186/s12918-017-0447-8 RESEARCH ARTICLE Open Access Cooperative binding mitigates the high-dose hook effect Ranjita Dutta Roy1,2, Christian Rosenmund2 and Melanie I. Stefan3,4,5* Abstract Background: The high-dose hook effect (also called prozone effect) refers to the observation that if a multivalent protein acts as a linker between two parts of a protein complex, then increasing the amount of linker protein in the mixture does not always increase the amount of fully formed complex.
    [Show full text]
  • High-Throughput Incubation and Quantification of Agglutination
    G C A T T A C G G C A T genes Article High-Throughput Incubation and Quantification of Agglutination Assays in a Microfluidic System David Castro 1,* ID , David Conchouso 1 ID , Rimantas Kodzius 1,2,3 ID , Arpys Arevalo 1 and Ian G. Foulds 1,4 1 Computer, Electrical and Mathematical Sciences & Engineering Division (CEMSE), King Abdullah University of Science and Technology (KAUST), 4700 KAUST, Thuwal, Jeddah 23955-6900, Saudi Arabia; [email protected] (D.C.); [email protected] and [email protected] (R.K.); [email protected] (A.A.); [email protected] (I.G.F.) 2 Mathematics and Natural Sciences Department, The American University of Iraq, Sulaimani, Sulaymaniyah 46001, Iraq 3 Faculty of Medicine, Ludwig Maximilian University of Munich (LMU), 80539 Munich, Germany 4 Okanagan Campus, School of Engineering, Faculty of Applied Science, University of British Columbia, 3333 University Way, Kelowna, BC V1V 1V7, Canada * Correspondence: [email protected] Received: 30 April 2018; Accepted: 29 May 2018; Published: 4 June 2018 Abstract: In this paper, we present a two-phase microfluidic system capable of incubating and quantifying microbead-based agglutination assays. The microfluidic system is based on a simple fabrication solution, which requires only laboratory tubing filled with carrier oil, driven by negative pressure using a syringe pump. We provide a user-friendly interface, in which a pipette is used to insert single droplets of a 1.25-µL volume into a system that is continuously running and therefore works entirely on demand without the need for stopping, resetting or washing the system.
    [Show full text]