Pain Management Clinic Procedure Suite Surgical Pendant

CSU Light Load Arm System; Arm Length 500+500 with Media Column 1000

- Arm length measured from axle bearing to axle bearing

- To be able to retrofit the arm system in the future, at the end of the arms are easily removable endcaps

- For reasons of operating safety and –friendlyness is the armsystem equipped with extremely smooth ball bearings and a combination brake. The combination brake prevents the unintended runaway of a support arm during a breakdown of the pneumatic brake and ensure the safe continued operation of the service supply unit

- The braking system includes a color control concept for intuitive operation without the use of pictograms

- Arm system with two 500mm by 500 mm horizontal swivel arms, intermediate bearing and ceiling bearing

- Action radius 1000 mm

- Load capacity on media column min. 120 kg

- Height of media column 500mm in length - - Double braking system, consisting of friction brake and pneumatic brake In case of pressure drop / pressure failure, the arm system remains in it’s previously adjusted position

Media Column

- Height of media column 1000 mm - Fitting of installation components possible on 3 sides - Pivot bearing integrated into media column - Double braking system in the pivot bearing, consisting of fictionbrake and pneumatic brake - Swivel range 330° - Adaptation to ceiling height via distance tube attached to arm system - Mounting system for attachment of workstation components, e.g. shelves, drawers, side arms, equipment holders from front side - Working height of workstation components infinitely variable - Set control elements, consisting of 2 handles for comfortable operation of the supply unit - Ergonomically designed handles to support the user posi-tioning the system - Color coding of control elements adapted to other supply unit components (e.g. joints, keypads) allowing for intuitive operation - Mounting at any front plate of media column is possible - Control element, consisting of 2 handles, mounted at front rail system of media column, for comfortable operation of the supply unit - Positioning in front of patient’s infusion equipment via integrated extension unit - Ergonomically designed handles to support the user positioning the system - Control of arm system’s pneumatic brakes via integrated push buttons

- One-piece front plates without gaps

- Surfaces resistant against disinfectants

- Color of media column: anodized aluminium

- Color of media column RAL 9002

- Color of front plates: anodized aluminium

- Smooth, disinfectant-resistant surfaces

- Integrated front plates without gaps.

- Different color shades of the component plates in RAL

- Number of component spaces

- Two parallel front rails for easy adaptation and installation from consoles, drawers, beams or equipment cher without use of new sealing elements

Installation Components

Gas Fittings

Media column equipped with:

__1__ x O2 (oxygen)

__1__ x Air 5 (compressed air 5 bar)

__2__ x VAC (vacuum)

__1__ x AGSS (Anesthesia Gas Scavenging System)

High voltage electrical installation

Media column equipped with:

__10__ x electrical socket

__2__ x data points

Number of electrical circuits: ___10____

Low voltage electrical installation

Media column equipped with: - follow OT17

__2__ x telephone socket TAE 6F __5__ x EDP socket RJ 45

__1__ x nurse call (material supplied by customer)

3.2 Workstation Components

3.5.7 Shelf for mounting at front rail system of Media Column

Shelf for attachment to front rail system of previously described media column

- Surface area 440 x 340 mm (W x D)

- Load capacity min. 40 kg

- Thickness max. 22 mm

- Material: Aluminium

- Smooth, scratchproof, fade-resistant and antistatic surface

- No sharp edges and grooves

- Integrated rubber elements at all 4 edges for injury prevention

- Additional standard rails (25x10 mm) at left and right hand side

- Load capacity of each standard rail min. 10 kg

Drawer for mounting at shelf

Drawer for attachment to shelf, as previously described in item no. 3.5.7

- Dimensions 440 x 340 x 170 mm (W x D x H) - Load capacity min. 5 kg - - Dimensions 640 x 340 x 170 mm (W x D x H) - Load capacity min. 7 kg

Special drawer runner system with - exceptionally smooth sliding function

- self-closing function

Facilitated cleaning by - easy removal capability of complete drawer

- removable front fascia

- smooth surfaces without protruding handles CVP-Pole

- Support tube with height adjustable extension pole

- Length of support tube min. 1000 mm

- Length of extension pole min. 800 mm

- 4 hooks, cross orientation, for accommodation of infusion bottles

( ) ( )

Pre-installation components

Ceiling fixture set anchor / bolt mounting for ceiling supply unit

- adaptable on site to height of false ceiling, min. length 1200 mm

- containing all necessary fixation components, e.g. sub-ceiling flange, distance tubes, reinforcement set, inter-connecting flange

Ceiling fixture set anchor / bolt mounting for ceiling supply unit

- for direct fixation to concrete ceiling

- containing all necessary fixation components, e.g. sub-ceiling flange, reinforcement set, inter-connecting flange