<<

PEDIATRICDENTISTRY/Copyright © 1988 by The AmericanAcademy of Pediatric Dentistry Volume10, Number2

Respiratory during pediatric : and capnography Jay A. Anderson, DDS, MDWilliam F. Vann, Jr., DMD,MS, PhD

Abstract Routine Respiratory Monitoring: A major concernin pediatric dentistry is maximizingrisk The Problem managementthrough optimal monitoring of respiratory What should constitute routine respiratory monitor- function during sedation techniques. This article examines ing for all pediatric patients undergoing conscious or the problemsinherent in respiratory monitoring in sedated deep sedation long has been a topic of debate. In 1985 the pediatric dental patients, the traditional methodsof respira- American Academy of Pediatrics and the American tory monitoring, and new technologies which are useful in Academyof Pediatric Dentistry jointly adopted the optimizing respira tory monitoring. The authors discuss tran- Guidelines for the Elective Use of ConsciousSedation, Deep scutaneous oximetry and pulse oximetry as possible monitors Sedation, and General in Pediatric Patients of oxygenation, and capnographyas a possible monitor for (1985). These guidelines require the use of a precordial apnea, airway obstruction, and developing . stethoscope, visual assessment of the child’s color, continuous monitoring of heart and , Risk management involves striving for maximum and the frequent assessment of head position to ensure patient safety and should be a primary concern to all a patent airway. Visual assessment of chest and abdomi- dentists and physicians practicing conscious or deep nal movementto monitor for effective respiratory ex- sedation. A major component of risk management for cursions is implied in the guidelines. These standard sedation or anesthesia involves physiologic monitor- methods should be used routinely to assess the rate, ing. Monitoring may be defined as continuous observa- depth, and adaquacy of . tion of data from specific organ systems to evaluate the While these time-tested methods are essential, each status of physiologic function. The purpose of monitor- has serious practical limitations. Visual assessment of ing is to permit prompt recognition of any deviation patients’ respiratory excursions, color, and general from normal, so corrective therapy can be instituted appearance requires close and continuous attention. For before morbidity ensues. Ideal monitors should be pediatric dental sedation, the sedationist is usually also continuous or "real time", rapidly responsive, noninva- the operator. Airway assessment is very difficult be- sive, accurate, dependable, convenient, and affordable. cause attention must be given to performance of de- The principal systems monitored during pediatric seda- tailed and sophisticated dental procedures. Even with tion are the central nervous, cardiovascular, and respi- constant attention, shallow respirations of a sedated ratory systems. child can be difficult to assess because of clothing, This discussion reviews monitoring of the respira- drapes, and restraining devices. Breath sounds may be tory system. There is compelling evidence that this is the difficult to hear with the precordial stethoscope, espe- most important system to monitor for pediatric pa- cially with the noise of the dental equipment. One is tients. Furthermore, major scientific advances are being often left to assess respiratory function by the patient’s made in this area which should be understood by all color and vital signs. In children this is a physiologically who use sedation and/or anesthesia, especially those dangerous method for respiratory monitoring, because whotreat pediatric patients. these changes do not occur until late in the process of

94 RESPIRATORY MONITORING FOR PEDIATRIC SEDATION: Anderson and Vann respiratory compromise when it may be too late to In summary, hypoxemia is a serious problem in the prevent an emergency situation leading to morbidity. sedation of pediatric dental patients and none of the time-honored methods of monitoring these patients is Hypoxemia reliable as a warning of hypoxemia. We conclude that Hypoxemia has been shown to be the leading cause there is a critical need for moreprecise monitoringof the of morbidity and mortality during anesthesia (Bendixin during pediatric sedation. Specifi- and Laver 1965). Hypoxemiais defined as a low partial cally, monitoring should detect respiratory depression pressure of oxygen in the and may be caused by early, to prevent hypoxemiafrom occurring, rather than such conditions as a failure of oxygen supply, pulmo- detecting physiologic changes that occur as a result of nary disease, cardiovascular collapse, hypoventilation, hypoxemia. apnea (cessation of breathing), or airway obstruction. Hypoxemia develops readily during deep sedation Respiratory Monitoring: The Questions techniques used for dentistry in young adults (Dionne et al. 1981). It also has been shownto be the leading cause There are two principal questions to be answered by of death in two studies that focused specifically on respiratory monitoring: (1) Is the patient breathing? dental sedation and anesthesia in adults (Tomlin 1974; (i.e., Is apnea or airway obstruction present?); and (2) A1-Kishaliet alo 1978). ventilatory exchange adaquate? (i.e., Are oxygen and Evidence is overwhelming that hypoxemia is also being exchanged adaquately?). For the major complication in the sedation of pediatric continuous intraoperative monitoring during sedation, dental patients. Goodson and Moore (1983) reported these two questions should be addressed separately. several cases of life-threatening reactions after sedation for pediatric dental procedures. Although morbidity Monitoring Apnea or Airway Obstruction and mortality were induced by drug overdosage, close Several monitoring devices have been suggested to scrutiny of the cases reveals that the commonfactor that assess whether or not the patient is breathing (exchang- led to death or severe morbidity was respiratory failure ing air). These include capnography, rib cage/abdomi- leading to hypoxemia. nal wall plethysmography, and nasal thermisters. Each The development of hypoxemia in pediatric dental will detect the presence or absence of normal air ex- patients is almost certainly more subtle and insidious change more precisely than observation or auscultation than most practitioners realize. Mooreet al. (1984) re- alone. In studies on the detection of sleep apnea, Keener ported that 25% of children sedated with 60 mg/kg and Phillips (1985) found the use of capnography to chloral hydrate and 50% /oxygen could more reliable than the nasal thermister. Rib cage/ab- not maintain a patent airway. Houpt et al. (1985) re- dominal respiratory inductive plethysmography meas- ported partial airway obstruction and transitory reduc- ures a difference in electrical impedance between sur- tions in respiration in a study that compared several face electrodes placed on the chest or abdominal wall sedation regimens, including one with as little as 50 during respiratory movements, generating a respira- mg/kg chloral hydrate and 50% nitrous oxide/oxygen. tory waveform. Because chest and abdominal wall Recently, Mueller et al. (1985) evaluated oxygenation movements continue and often become exaggerated in sedated pediatric dental patients. Oxyhemoglobin during airway obstruction, this technique will not de- saturations below physiologically acceptable levels tect early obstruction reliably (Peabodyet al. 1979). were reported for a high percentage of children sedated The most promising technology for detection of with alphaprodine or with chloral hydrate plus 50% these respiratory problems appears to be capnography nitrous oxide/oxygen. This finding was significant which is the continuous analysis of the carbon dioxide because it demonstrated positively that children be- content of respired gases. Capnography recently has come hypoxemic even when supplemented with an risen to the forefront as a means of detecting apnea and enriched oxygen mixture. Furthermore, despite epi- airway obstruction. The simplest capnographs merely sodes of hypoxemia, there were no detectable changes detect the presence of carbon dioxide as CO2 is exhaled in vital signs or mucosalcolor (Mueller et al. 1985). This and therefore will indicate whether or not the patient is finding is not surprising because the signs and symp- breathing. This type of unit is the least expensive, but toms of hypoxemia often do not become evident clini- provides the least information. The more commonly cally until very low and dangerous levels of oxygen used capnographs are more sophisticated in that the tension develop (Manninen and Knill 1979). Expert carbon dioxide detected is also quantified. This proves observers are unable to recognize hypoxemia consis- to be a valuable source of additional data regarding tently until the arterial oxygen tension (PaO2) falls ventilatory adaquacy. below 40 mm Hg (Comroe and Bethelho 1947). The In capnography gas is suctioned continuously from normal value for PaO2 when breathing room air is 90- the airway for analysis. A waveform(Fig I -- next page) 100 mmHg. is produced that maybe analyzed to detect the presence

Pediatric Dentistry: June, 1988 - Volume10, Number2 95 tube was inserted through a small needle puncture in NORMAL CAPNOGRAM the nasal and threaded up the nasal prong so that the sampling tube lay just inside the nostril when the cannula was in place (Figs 2a, b—below). This study revealed that the capnograph provided a consistent B waveform, representing the breath-to-breath ventila- tory pattern. When apnea or airway obstruction oc- curred, the PetCO2 dropped immediately to zero, allow- ing for immediate detection of these entities. Addition- ally, arterial blood samples were obtained for compari- son of PaCO2 with simultaneously recorded PetCO2 values. The absolute value of the PetCO2, measured via nasal prongs, was not reliable for reflecting the exact PaCO2 at any given moment in all patients. The PetCO2 values did provide a good "trend" indicator for the de- tection of developing hypoventilation as they rose. The A: begins accuracy of this monitor in predicting PaCO2 in nonin- B-C: Plateau = outflow of alveolar gas tubated, spontaneously breathing patients requires further study. C: End-Tidal CO2 In summary, capnography provides two valuable FIG 1. Capnography produces a waveform by the continuous parameters for continuous respiratory monitoring: (1) analysis of respired gas for CO2. The presence of the waveform The continuously displayed CO2 concentration pro- implies exhalation of gases from the . The end-tidal CO2 vides a breath-to-breath representation of the presence (point C) corresponds to alveolar gas which may correlate or absence of air flow (i.e., functional respirations), and closely with the PaCO2. thus serves as an effective monitor for apnea and airway obstruction; and (2) the PetCO2 value serves as a trend or absence of ventilation (i.e., apnea or airway obstruc- indicator of the depth or adaquacy of respiration. It is tion) and the depth of respiration (hypo- or hyperventi- important to note that these changes may be detected by lation) by analysis of the end-tidal CO2 (PetCO2) which capnography long before detectable changes in oxy- is the concentration of CO2 measured at the terminal genation, especially during oxygen supplementation. portion of the exhalation curve. It is displayed con- A nasal nitrous oxide hood can be placed directly tiuously by most capnographs. It represents gas coming over the sampling tube-nasal prong assembly for use from the alveoli and has been shown to correlate closely during the administration of nitrous oxide conscious with the PaCO2 ( of CO2 in arterial sedation. Because CO2 sampling occurs during exhala- blood) in intubated patients (Burton 1966). Gases may tion, the flow of nitrous oxide and oxygen will not be sampled at the endotracheal tube connector in intu- prevent CO2 sampling. Several companies are introduc- bated patients or at the nares in nonintubated patients. ing devices designed for sampling in nonintubated Anderson et al. (1987) evaluated capnography as a patients. Finally, capnography has been used with suc- respiratory monitor during outpatient general anesthe- cess in pediatric patients for the assessment of airflow in sia for oral . In these nonintubated patients, sleep studies to detect sleep apnea (Keener and Phillips respired gases were sampled via a modified nasal oxy- 1985). Further study is needed to determine the useful- gen cannula placed at the nares. The plastic sampling ness of capnography for respiratory monitoring during

FIGS 2a and b. Respired gases are sampled for capnography by inserting the sampling tube through a nasal oxygen cannula and placed at the patient's nares.

96 RESPIRATORY MONITORING FOR PEDIATRIC SEDATION: Anderson and Vann pediatric sedation for dentistry. PtcO2 values varied considerably. Trend detection was inconsistent and response time was extremely variable Monitoring Adequacy of Ventilatory Exchange and slow, occasionally taking as long as 15 min to The second basic question posed in respiratory stabilize following changes in inspired oxygen concen- monitoring involves the assessment of the adequacy of tration. They concluded that the transcutaneous oxygen the ventilatory exchange that is occurring. Adequate monitor was an inadequate monitor of arterial oxygena- alveolar ventilation is evaluated most precisely via arte- tion during anesthesia. rial blood gas analysis. Blood gas analysis, though very Anderson et al. (1987) evaluated the transcutaneous accurate, is not practical for continuous respiratory oxygen monitor during ultralight general anesthesia for monitoring because it is invasive, inconvenient, and dentistry. Although a high correlation was demon- expensive. It is also retrospective, i.e., the values ob- strated between PaO2 and PtcO2 overall, a considerable tained from the laboratory represent the state of arterial lag time was demonstrated between changes in PaO2 blood several minutes earlier. and accurate changes in PtcO2 during any period when Advances in noninvasive, continuous monitoring of PaO2 was changing rapidly. This same phenomenon the respiratory system have occurred in recent years. was demonstrated by Kafer et al. (1981). Hypoxemia is the leading cause of mortality and mor- Accurate monitoring of respiratory status is critical bidity during dental sedation and anesthesia, and its during periods of apnea, airway obstruction, and severe signs and symptoms do not occur until late. Thus, a respiratory depression. For this reason the transcutane- sensitive, continuous method of monitoring oxygena- ous oxygen monitor does not appear to be optimal for tion is desirable. This parameter represents the "bottom use during dental sedation. Still, the ability to monitor line" in respiratory monitoring. Several monitors have the PaO2 remains an attractive concept and technologi- been introduced for this purpose, including the tran- cal changes could make transcutaneous oxygen moni- scutaneous oxygen monitor, ear oximeter, and pulse toring a better option in the future. oximeter. Pulse Oximetry Transcutaneous Oxygen Monitor The most recent and promising development for The transcutaneous oxygen monitor can monitor respiratory monitoring is the pulse oximeter. It func- PaO2 continuously and noninvasively over the entire tions by placing a pulsating vascular bed between a 2- physiologic range of oxygenation. The instrument util- wavelength red and infrared light source and a photo- izes a heated polarographic electrode that is sealed to diode detector (Fig 3). the patient's skin. The heat causes vasodilitation of cu- Light absorption varies taneous capillaries, producing a state of arterialization with arterial pulsation, of blood flow. This causes the oxygen tension of the the wavelength of light blood at the skin to increase to levels similar to arterial used, and the oxyhemo- PaOr Oxygen from the blood diffuses through the skin globin saturation. Using and reacts with the electrode, which measures the POr spectrophotometric The transcutaneous oxygen monitor has several analysis, the oximeter clinical disadvantages. It is expensive and requires ex- determines the ratio of tensive calibration and a lengthy warm-up period. Also, oxygenated (red) hemo- FIG 3. Pulse oximeter probe with arterialization of the skin may cause burns. Beech and globin to deoxygenated light source and photodiode detec- Lytle (1982) evaluated the monitor during dental anes- (blue) hemoglobin and tor. thesia and concluded that it was useful and accurate, displays oxyhemoglobin though accuracy was determined by only 11 random saturation (SaO2). Artifacts that otherwise would be blood samples. caused by tissue and venous blood are eliminated. Cassidy et al. (1986) used transcutaneous oxygen Pulse oximetry has been shown to be accurate during monitoring in children to assess the respiratory effects steady state conditions in young healthy volunteers of nitrous oxide sedation. They reported that it seemed (Yelderman and New 1983), intensive care patients to offer a reliable modality of early detection of ventila- (Kim 1984), and during general anesthesia (Brodsky et tion-related complications. They noted rapid responses al. 1985). The usefulness of pulse oximetry has been in transcutaneous oxygen tension (PtcO2) following reported for dental extractions under general anesthe- changes in delivered oxygen concentration, but did not sia (Beeby and Thurlow 1986) and for pediatric dental verify accuracy. They noted that stabilization of the sedation (Mueller et al. 1985; Moody et al. 1986). These electrode required 8-15 min following placement. studies suggest that the pulse oximeter is a useful Knill et al. (1982) assessed the transcutaneous oxy- monitor and that it is more sensitive for detecting hy- gen monitor during anesthesia and reported that stable poxemia than visual assessment and vital sign changes.

Pediatric Dentistry: June, 1988 ~ Volume 10, Number 2 97 Accuracy was not assessed in these studies. in combination with hemoglobin and is reflected as Anderson et al. (1988) evaluated the accuracy hemoglobin saturation (SaO2). Each hemoglobin mole- pulse oximetry during ultralight general anesthesia for cule is capable of carrying 4 oxygen molecules. The young, healthy adults undergoing third molar extrac- combination of oxygen and hemoglobin results in a tions. Patients were not intubated and presented a change in the shape of the hemoglobin molecule and unique monitoring challenge because periods of hyper- thus a change in its affinity for oxygen. Becauseof these and hypoventilation occurred commonlyas boluses of changes, the PaO2 and the SaO2 are not linearly related, intravenous anesthetic agents were given. Apnea, air- but rather are related by the oxyhemoglobin dissocia- way obstruction, and laryngospasm may easily occur tion curve (Fig. 4). during ultralight general anesthesia, and patients often move, vocalize, and shiver. Oxyhemoglobinsaturations 100- obtained using 4 brands of pulse oximeters were com- pared with arterial saturations measured in vitro from 122 arterial blood samples to determine accuracy. De- Z 80- spite the rigorous conditions imposed by this technique, 3 of 4 pulse oximeters were accurate in predicting arte- 60- rial hemoglobin saturation over the range evaluated (70-100%). The fourth pulse oximeter tended to significantly 40- underestimate SaO2. Kagle et alo (1987) reported the same degree of inaccuracy in an evaluation of this oximeter. As a result, the software of the oximeter was 20- modified by the manufacturer, resulting in a significant improvement in accuracy.

In summary, the pulse oximeter is very accurate and I I I I I I I I I I rapidly responsive. It is extremely convenient because it 20 40 60 80 100 requires no calibration, warm-uptime, or tissue prepa- ration. Multiple probes are available for different sites PaO= mmHg such as the ear, finger, toe, and nose. Pulse oximetry does have potential shortcomings. OXY-HEMOGLOBINDISSOCIATION CURVE Any reduction in vascular pulsations in the area being monitored by the oximeter probe will decrease the instrument’s ability to function. Such conditions could FIc 4. The oxyhemoglobindissociation curve reveals the result from , hypotension, vasoconstictors, nonlinear relationship between PaO2 (partial pressure of or direct vascular compression. Also, the oximeter may oxygenin blood) and oxyhemoglobinsaturation (SaO2). misinterpret movementof the probe as a pulse, produc- ing motion artifact. Motion artifact maybe a problem in The "S" shape of the oxyhemoglobin dissociation uncooperative pediatric patients who are moving or curve is important for physiologic uptake and delivery struggling, or with involuntary motion such as shiver- of oxygen in the body. In the lungs, hemoglobin is ing. Beeby and Thurlow (1986) reported that pulse rapidly and almost totally saturated over a wide range oximetry was not useful in 4 of their 30 patients under- of PaO2 (flat portion of the curve), while at the tissues going dental treatment due to motion artifact. Cur- large amount of oxygen is unloaded as desaturation rently, attempts are being made to eliminate this prob- occurs over a relatively small drop in PaO2 on the steep lem by incorporating ECGanalysis in the oximeter to portion of the curve. An understanding of the relation- determine the time interval during which the oximeter ship between PaO2 and the SaO2 is essential for those should detect arterial pulsations. This technology may using pulse oximetry clinically. Pulse oximetry is lim- eliminate muchof the motion artifact problem in pedi- ited in that changes in oxygenation are not detected atric dental patients in the future. until the PaO2 falls to the point where oxyhemoglobin The primary disadvantage of pulse oximetry is that desaturation occurs, that is, the 70-80 mmHg range, it measures oxyhemoglobin saturation rather than where the steep portion of the oxyhemoglobindissocia- PaO2. Oxygen is carried in the blood in 2 forms, that tion curve is rapidly approached. Whenone is breathing dissolved in the plasma and that bound to hemoglobin. room air, the normal PaO2is in the 90-100 mmHg range, The portion dissolved in the plasma is reflected by the which corresponds to an SaO2 of 96-100%. Because a PaO2 value and represents only a small fraction of the PaO2 of 60 mmHg corresponds to an SaO~ of approxi- total oxygen content. The majority of oxygen is carried mately 90%, the PaO2 must fall from approximately 100

98 RESPIRATORY MONITORING FOR PEDIATRIC SEDATION: Anderson and Vann mmHg to 60 mmHg for the saturation to fall from 96 to TABLE1. Relationship Between PaO2 and SaO2* 90%(Table 1). PaO2(mmHg) SaO2(%) Often patients undergoing dental sedation or gen- 500 100 eral anesthesia are receiving supplemental oxygen 400 100 (with or without nitrous oxide), thus the PaO2 may 300 100 range from 150 to above 600 mmHg. The PaO must fall 200 100 2 150 100 drastically before any change will be detected in the 120 100 SaO2. Barker et al. (1986) graphically demonstrated that 100 99 large decreases in oxygenation may occur without any 95 98 90 97 change being detected by pulse oximetry. The PaO2 fell 80 96 to less than 70 mmHg before significant desaturation 70 94 occurred in pulse oximetry. The pulse oximeter will not 65 93 60 91 warn of downwardtrends in PaO2 over the wide range 55 89 of oxygentensions above this level (Fig 5). In brief, when 50 85 oxyhemoglobin desaturation begins to occur, serious 45 8O 40 75 35 65 Post-Intubation * At pH 7.40, temperature 37°C, PCO240. 500. .500 2~ PaO respiratory depression may be present. Furthermore, 2--- PtcO more rapid desaturation may be imminent as the steep portion of the curve is approached. Therefore, during "F_ 400 4O0 o pediatric sedation, even a small change in saturation (e.g., 99 to 96%) must be noted quickly and evaluated ~ before further desaturation occurs. 300 x>" 300 In summary, those using pulse oximetry must real- ize that though the SaO2 scale ranges from zero to 100, PO2(rnmHg; even a small decrease in saturation requires close atten- O I \\ Intraoperative 200 I 200 tion and prompt treatment. This is critical in children because their high basal oxygen consumption and o \~ hypoxemia ¢- smaller size produces less oxygen reserve. This results in more rapid and profound desaturations than in 100 100 young, healthy adults. Because the danger zone for severe hypoxemia is below 90%, it maybe argued that the pulse oximeter will ! detect hypoxemia prior to any evidence of signs and symptoms. Thus, the pulse oximeter is a valuable monitor, especially considering its other advantages. 100 ~t~~==~=1~l~ ° 100 Despite the initially high cost, the pulse oximeter is 80 80 rapidly becoming a routine respiratory monitor during ~ anesthesia. The cost is currently falling as more compa- E 60 60 nies introduce pulse oximeters. Four pulse oximeters ~ SaO2 presently available on the market are illustrated in C~ %SAT40 40 ~ NSaO2 Figure 6 (next page). ~. 2 20 Summary ~ t .... I 1 I 0 Accurate, continuous and noninvasive respiratory 0 1 2 3 4 monitoring represents a critical need when pediatric TIME(hrs.) sedation is performed. In adult dental lbatients, the combination of pulse oximetry and capnography ap- FIG5. These data are from a 28-year-old female undergoinga pears to be the most ideal monitoring system available general anesthetic. Note the wide range of transcutaneous for the respiratory system during sedation or general oxygentension (PtcO2) and PaO2values. Althoughthe 2PtcO anesthesia. These monitors, combined with a precordial and PaO2 decreased markedly, from 460 to 86 mmHg, the stethoscope and observation, serve to answer the two SaO~showed no significant change. Only when the PtcO 2 primary questions of "Is the patient breathing?" and "Is decreased further to 47 mmHg did the SaO2 fall to 84%, a significant change. gas exchange adaquate?"

Pediatric Dentistry: June, 1988 - Volume10, Number2 99 FIG 6. Four available pulse oximeters.

Risk managementin the sedation of pediatric dental stered to dental outpatients in the upright position. Anaesthesia patients is of primary importance if dentists are to retain 32:184-88,1978. the priviledge of using sedation techniques. Strong Anderson JA, Clark PJ, Kafer ER: Use of capnography and transcuta- consideration should be given to the routine use of the neous oxygen monitoring during outpatient general anesthesia pulse oximeter during pediatric sedation. Further for oral surgery. J Oral Maxillofac Surg 45:3-10, 1987. evaluation is necessary to determine whether capnogra- phy also will prove to be a respiratory monitor that is Anderson JA, Lambert DM, Kafer ER, Dolan P: Pulse oximetry: evaluation of accuracy during outpatient general anesthesia for useful in the quest to assure safety for sedated child oral surgery -- comparison of four monitors. Anesth Prog 35:53- patients. 64, 1988.

Barker SJ, Tremper KK, Gamel DM: A clinical comparison of tran- Dr. Andersonis an assistant professor, oral and maxillofacial surgery scutaneous PO2 and pulse oximetry in the operating room. and , and Dr. Vann is an associate professor and Anesth Analg 65:805-8, 1986. chairman, pediatric dentistry, University of North Carolina. Reprint requests should be sent to: Dr. Jay A. Anderson, Dept. of Oral and Beeby C, Thurlow AC: Pulse oximetry during general anesthesia for Maxillofacial Surgery, CB 7450 Brauer Hall, University of North dental extractions. Br Dent J 160:123-25, 1986. Carolina, Chapel Hill, NC27599. Beech AD, Lytle JJ: Continuous transcutaneous oxygen tension American Academy of Pediatrics and American Academy of Pediat- monitoring during ultralight general anesthesia for oral surgery. ric Dentistry: Guidelines for the elective use of conscious seda- J Oral Maxillofac Surg 40:559-65, 1982. tion, deep sedation, and general anesthesia in pediatric patients. Pediatr Dent 7:334-37, 1985. Bendixin HH, Laver MB: Hypoxia in anesthesia: a review. Clin Pharmacol Ther 6:510-39, 1965. A1-Kishali T, Padfield A, Perks ER, Thornton JA: Cardiorespiratory effects of nitrous oxide-oxygen: halothane anaesthesia admini- Brodsky JB, Shulman MS, Swan M, Marks JBD: Pulse oximetry

100 RESPIRATORYMONITORING FOR PEDIATRIC SEDATION: Anderson and Vann during one ventilation. Anesthesiol63:212-14, 1985. KimSK, BaidwanBS, Petty TL: Clinical evaluation of a newfinger oximeter. Crit Care Med12:910-12, 1984. Burton GW:The value of carbon dioxide monitoring during anaes- thesia. Anaesthesia21:173-83, 1966. Knill RL, ClementJL, KieraszewiczHT, DodgsonBG: Assessment of two noninvasivemonitors of arterial oxygenationin anesthetized Cassidy DJ, Nazif MM,Zullo T, Ready MA:Transcutaneous oxygen man. Anesth Analg 61:582-86, 1982. monitoring of patients undergoing nitrous oxide-oxygenseda- tion. Pediatr Dent8:29-31, 1986. ManninenP, Knill RL: Cardiovascular signs of acute hypoxaemia and hypercarbia during enflurane and halothane anesthesia in ComroeJH, BethelhoS: Unreliability of cyanosis in the recognition man. Can AnaesthSoc J 26:282-87, 1979. of arterial anoxemia.Am J MedSci 214:1-6, 1947. MoodyEH, Mourino AP, Campbell RL: The therapeutic effective- DionneRA, Driscoll E J, GelfmanSS, SweetJB, Butler DP,Wirdzek PR: ness of nitrous oxide and chloral hydrate administered orally, Cardiovascular and respiratory response to intravenous diaze- rectally, and combinedwith hydroxyzinefor pediatric dentistry. pam, fentanyl, and in dental outpatients. J Oral J Dent Child53:425-28, 1986. Surg 39:343-49,1981. Moore PA, Mickey EA, Hargreaves JA, NeedlemanHL: Sedation in GoodsonJM, MoorePA: Life-threatening reactions after pedodontic pediatric dentistry: a practical assessmentprocedure. J AmDent sedation: an assessmentof narcotic, , and antiem- Assoc 109:564-69,1984. etic drug interaction. J AmDent Assoc107:239-45, 1983. Mueller WA,Drummond JN, Pribisco TA, Kaplan RF: Pulse ox- HouptMI, KoenigsbergSR, Weiss NJ, Desjardins PJ: Comparisonof imetry monitoringof sedated pediatric dental patients. Anesth chloral hydrate with and without promethazinein the sedation of Prog 32:237-40,1985. youngchildren. Pediatr Dent7:41-46, 1985. PeabodyJF, GregoryGA, Willis MM,Philip AG,Lucey JF: Failure of Kafer ER, Sugioka K: Respiratory and cardiovascular responses to conventional monitoringto detect apnea resulting in hypoxemia. hypoxemiaand the effects of anesthesia. Int AnesthClin 19:85- Birth Defects14:225, 1979. 122, 1981. TomlinPJ: Deathin outpatient dental anaesthetic practice. Anaes- Kagle DM,Alexander CM,Berko RS: Evaluation of the Ohmeda3700 thesia 29:551-70,1974. pulse oximeter; steady state and transient responsecharacteris- tics. Anesthesiol66:376-80, 1987. YeldermanM, NewW: Evaluation of pulse oximetry. Anesthesiol 59:349-52,1983. Keener T, Phillips B: Assessmentof airflow in sleep studies by oronasal CO2 detection. Chest 88:316, 1985.

Public: dentists are honest and ethical When asked to rate the honesty and ethical standards of various professions, dentists ranked second only to pharmacists. The July SRI Gallup poll showed that dentistry had grown in the public’s esteem since the poll was last taken in 1983. Then, dentists ranked fourth -- behind clergymen, pharmacists, and physicians. Doug Harris, the American Dental Association’s director of marketing services, said, "It’s the esteem in which dentistry is held by the public that is the profession’s best marketing tool. And, it’s important to the marketing of any practice that dentists fulfill the image that the public has of them." Of the 734 respondents to the 1987 poll, 64% rated the ethical standards of dentists either "very high" or "high." This was an increase of 13% over the 1983 poll, when only 51% of the respondents rated dentists in these categories.

Pediatric Dentistry: June, 1988 ~ Volume 10, Number 2 101