UV-Vis Spectroscopy

Total Page:16

File Type:pdf, Size:1020Kb

UV-Vis Spectroscopy Ultraviolet and visible spectroscopy Dr. Ahmad Najjar Philadelphia University Faculty of Pharmacy Department of Pharmaceutical Sciences 1st semester, 2020/2021 Spectrophotometry Spectroscopy is a general term referring to the interactions (mainly absorption and emission) of various types of electromagnetic radiation with matter. Spectrophotometry is a method to measure how much a chemical substance absorbs or emits light by measuring the intensity of light (electromagnetic radiation). It refers to the use of light (electromagnetic radiation) to measure chemical concentrations. Electromagnetic spectrum refers to the full range of all frequencies of electromagnetic radiation, which is refers to the waves of the electromagnetic field, propagating through space carrying electromagnetic energy. Electromagnetic radiation Electromagnetic radiation (EMR) has been described in terms of a stream of photons that travel in a wave-like pattern. Each photon contains a certain amount of energy, and all electromagnetic radiations consists of these photons. All electromagnetic radiations travels in a straight line at the Crest Crest speed of light (3 x 108 m/s). The only difference between the various types of electromagnetic radiations is the amount of energy found in the photons. Electromagnetic radiations are described by several terms Trough such as wavelength, frequency, wave number and energy. λ wavelength (units : m,cm, μm, nm) ν 1/λ ν frequency (units of cycles/sec, sec1, Hertz c ν λ ν /ν c ν wavenumber (number of waves per cm; unit : cm1) Energy (E) h ν h h c ν λ 8 1 velocityof light in vacuum c ν λ 3.00 10 m.s h is planck's constant 6.62x10-34 J.s Electromagnetic radiation Crest Crest Electromagnetic wave is also characterized by several fundamental properties, including its velocity, amplitude, phase angle, polarization, and direction of propagation. Trough Wave amplitude measures the magnitude of oscillation of a particular wave. Larger amplitude means higher energy and lower amplitude means lower energy. Amplitude is important because it tells you the intensity or brightness of a wave in comparison with other waves. Power, P, and Intensity, I, of light give the flux of energy from a source of EMR. • P is the flux of energy per unit time • I is the flux of energy per unit time per area Electromagnetic radiation spectrum Electromagnetic radiation in the domain ranging between 180 and 780 nm, has been studied extensively. This portion of the electromagnetic spectrum, designated as the ‘UV/Visible’. Generally provide little structural information but is very useful for quantitative measurements. Legend: γ = Gamma rays HX = Hard X-rays SX = Soft X-Rays EUV = Extreme-ultraviolet NUV = Near-ultraviolet Visible light (colored bands) NIR = Near-infrared MIR = Mid-infrared FIR = Far-infrared EHF = Extremely high frequency (microwaves) SHF = Super-high frequency (microwaves) UHF = Ultrahigh frequency (radio waves) VHF = Very high frequency (radio) HF = High frequency (radio) MF = Medium frequency (radio) LF = Low frequency (radio) VLF = Very low frequency (radio) VF = Voice frequency ULF = Ultra-low frequency (radio) SLF = Super-low frequency (radio) ELF = Extremely low frequency(radio) Problem 1: Calculate the wavenumber of a beam of IR radiation with a wavelength of 3μm. Problem 2: The frequency of a radiation is 3x1012 s-1. Calculate the wavelength of the radiation. Problem 3: Calculate the energy of 530-nm photon of visible radiation J 10 x 3.75 = / h = h = E Answer: 흀 풄 19 19 - m 10 = / = Answer: 풄 흀 4 4 - = 3,333 cm 3,333 = 1/ = wavenumber Answer: 흀 1 - Spectrophotometric methods A group of techniques that relies on the quantitative interaction of EMR and matter, these are mainly: . Absorption (excitation process) . Emission (deactivation (or de-excitation) process): Incandescence (luminescence after thermal heating) Chemiluminescence (luminescence after chemical reaction) Photoluminescence (luminescence after light absorption): Fluorescence: Resonance fluorescence Non-resonance fluorescence Phosphorescence Note: deactivation of absorbed energy could also be done through non-radiative (radiationless) process, such as relaxation, internal conversion (quenching) and intersystem crossing. Spectrophotometric methods: Absorption When a photon is absorbed by an analyte, it is "destroyed," and its energy is acquired by the analyte. This energy promoting the analyte electron from a lower-energy state (ground state) to a higher-energy, (or excited) state. Na atomic HCOH molecular energy levels energy levels Spectrophotometric methods: Absorption . The energy levels have well-defined values (i.e., they are quantized). Absorption only occurs when the photon's energy matches the difference in energy, E, between two energy levels. A plot of absorbance as a function of the photon's energy (expressed as wavelength, ) is called an absorbance spectrum. Absorption spectrum of Hg gas Absorption spectrum of chlorophyll a Why atomic spectrum has sharper peaks!!! Molecular orbital Electrons in atoms exist in atomic orbitals (consist of electronic levels only) while electrons in molecules exist in molecular orbitals (consist of electronic, vibrational and rotational levels). Each molecular orbital has energy level represent electronic state S. Between each electronic states there lies several vibrational levels V, themselves also sub-divided into a collection of rotational levels R. Molecular orbital Spectrophotometric methods: Absorption A molecule absorbs a photon by undergoing an energy transition exactly equal to the energy of the photon. The energy captured during the photon absorption can be expressed as Etot = Erot + Evib + Eelec In atoms (no sub-levels present), the energy captured during the photon absorption can be expressed as Etot = Eelec The reason why electronic absorption bands are usually very broad is that many different vibrational and rotational levels are available at slightly different energies. Therefore, a molecule could absorb photons with a fairly wide range of energies and still be promoted from the ground electronic state to one particular excited electronic state. Spectrophotometric methods: Absorption Absorbing a photon of visible light causes a valence electron in the analyte to move to a higher-energy level. When an analyte absorbs infrared radiation one of its chemical bonds experiences a change in vibrational energy level. Spectrophotometric methods: Emission Emission of a photon occurs when an analyte in a higher-energy state returns to a lower-energy state. The higher-energy state can be achieved in several ways including thermal energy, radiant energy from a photon, or by a chemical reaction. Emission following the absorption of a photon is also called photoluminescence, and that following a chemical reaction is called chemiluminescence. Spectrophotometric methods: Emission In many cases molecules in the excited states loss some energy before emission process occurs. This will give emitted radiations with lower energies than those generated the absorption in first place. The loss of energy could be happened through several radiative or non- radiative deactivation processes such as: vibrational relaxation, internal or external conversions and through intersystem crossing process. In general Eemission = Eabsorption - Erelaxation (and/or -Einternal conversion and/or -Eintersystem crossing) Note that, as the E decrease, the will increase. Spectrophotometric methods: Jablonski diagram Spectrophotometric methods: Jablonski diagram . Absorption Radiative transition from ground state to higher states with the same spin quantum number (S0 S0+n). This may include changing vibrational levels also. Relaxation (R) Radiationless transition between vibrational levels in the same electronic state ( Vn Vn-1). The definition also cover the transition to Vn-2, Vn-3….until V0. Internal Conversion (IC) Radiationless transition between states with same spin quantum numbers ( Sn Sn-1). The transition includes changing the vibrational levels. Intersystem Crossing (ISC) Radiationless transition between states with different spin quantum numbers ( S1 T1). Changing vibrational levels is expected also. Fluorescence Radiation transition between states with the same spin quantum number (S1 S0). The transition may include changing in the vibrational levels. Phosphorescence Radiation transition between states with different spin quantum number (T1 S0) including changing vibrational levels. The UV/Vis spectrum UV/Vis spectrometers collect the data (transmittance or absorbance) over the required range of wavelengths and generate the spectrum of the compound under analysis as a graph. The spectrum exhibit peaks over the investigated wavelengths range. The wavelength at which the top of the peak occurs is called max (lambda max). Some compounds show more than one max. Spectrum profile is affected by several conditions like : sample state, pH, solvent nature, presented metal ions, temperature and concentration. The UV/Vis spectrum . Other examples: . The recorded spectra of compounds in the condensed phase, whether pure or in solution, generally present absorption bands that are both few and broad, while those spectra obtained from samples in the gas state yield spectra of detailed ‘fine structure’. Electronic transitions of organic compounds Organic compounds represent the majority of the studies made in the UV/Vis. The observed transitions involve electrons engaged in or or non-bonding n electron
Recommended publications
  • A) Single-Beam
    Lecture 2 Spectrophotometer Spectrophotometry is the basis for many of the instruments used in clinical chemistry. The primary reasons for this are ease of measurement, satisfactory accuracy and precision, and the suitability of spectrophotometric techniques to use in automated instruments. Spectrophotometer measures light absorption by a liquid substance at various wavelengths. The Components of unknown material can be determined, or the concentration of a number of known substances can be measured. Types of Spectrophotometer Ultraviolet (UV) Spectrophotometers. Uses ultraviolet light of wave lengths from 200 nm to 350 nm. light) of wave lengths from 350 nm to 700 nm. Spectrophotometer Block Diagram a) Single-beam b) Double-beam Most common Spectrophotometer 1. Photodiode 2. Connection wire 3. Lamp 4. Filter/Detector 5. On/Off switch and zero transmission adjustment knob 6. Wavelength selector/Readout 7. Sample chamber 8. Transmittance/absorbance control 9. Absorbance/Transmittance scale 1. Light Sources Tungsten lamp: Vis. near IR (320 nm~2500 nm) Deuterium arc lamp: UV (200~400 nm) Uses a tungsten filament and anode placed on opposite sides of a nickel box structure designed to produce the best output spectrum. Unlike tungsten lamps, the filament is not the source of light in deuterium lamps. Instead an arc is created from the filament to the anode. The arc created excites the molecular deuterium contained within the bulb to a higher energy state. The deuterium then emits light as it transitions back to its initial state Its continuous spectrum is only from 180 nm to 370 nm. Light Intensity of Tungsten and Deuterium lamps A problem with tungsten lamps is that, during operation, the tungsten progressively vaporizes from the filaments and condenses on the glass envelope.
    [Show full text]
  • UV-VIS Nomenclature and Units
    IN804 Info Note 804: UV-VIS Nomenclature and Units ​ Ultraviolet-visible spectroscopy or ultraviolet-visible spectrophotometry (UV/VIS) involves the spectroscopy of photons in the UV-visible region. It uses light in the visible and adjacent ​ near ultraviolet (UV) and near infrared (NIR) ranges. In this region of the electromagnetic ​ spectrum, molecules undergo electronic transitions. This technique is complementary to ​ ​ ​ fluorescence spectroscopy, in that fluorescence deals with transitions from the excited state to the ground state, while absorption measures transitions from the ground state to the excited state. UV/VIS is based on absorbance. In spectroscopy, the absorbance A is defined as: ​ ​ ​ ​ (1) where I is the intensity of light at a specified wavelength λ that has passed through a sample ​ ​ (transmitted light intensity) and I is the intensity of the light before it enters the sample or ​ ​0 incident light intensity. Absorbance measurements are often carried out in analytical chemistry, ​ ​ ​ ​ since the absorbance of a sample is proportional to the thickness of the sample and the concentration of the absorbing species in the sample, in contrast to the transmittance I / I of a ​ ​ ​ ​0 s ample, which varies exponentially with thickness and concentration. The Beer-Lambert law is ​ used for concentration determination. The term absorption refers to the physical process of absorbing light, while absorbance ​ refers to the mathematical quantity. Also, absorbance does not always measure absorption: if a given sample is, for example, a dispersion, part of the incident light will in fact be scattered by the dispersed particles, and not really absorbed. Absorbance only contemplates the ratio of transmitted light over incident light, not the mechanism by which light intensity decreases.
    [Show full text]
  • UV-Vis) Absorption Vs
    授課教師: Professor 吳逸謨 教授 Warning: Copyrighted by textbook publisher. Do not use outside class. Principles of Instrumental Analysis Section III – Molecular Spectroscopy Chapter 13 Introduction to Ultraviolet-Visible Absorption Spectrometry + Chapter 14 Applications of Ultraviolet-Visible Absorption Spectrometry Reference: p. 423, Visible and UV spectra, in “Organic Chemistry” textbook – Solomons, 3rd Ed] 1 Molecular Spectroscopy Ultraviolet–visible (UV-Vis) absorption vs. fluorescence Spectroscopy Ultraviolet–visible spectroscopy or ultraviolet-visible spectrophotometry (UV-Vis or UV/Vis) refers to absorption spectroscopy or reflectance spectroscopy in the ultraviolet-visible spectral region. This means it uses light in the visible and adjacent (near-UV and near- infrared [NIR]) ranges. The absorption or reflectance in the visible range directly affects the perceived color of the chemicals involved. In this region of the electromagnetic spectrum, molecules undergo electronic transitions. Fluorescence spectroscopy is based on molecular emission: The “UV-Vis” technique is complementary to fluorescence spectroscopy, in that fluorescence deals with transitions from the excited state to the ground state, while UV-Vis absorption measures transitions from the ground state to the excited state.[1] 2 (2007/3) UV-Vis Spec 儀器 -實驗課 3 FIGURE 6-3 Regions of the electromagnetic spectrum. (For UV-Vis, λ = 100~500 nm, 500~1000 nm) 4 Ch6 An Introduction to Spectrometric Methods P.135 Apendix: Chap. 7E - Radiation Transducers p. 191 Read the texts in Chap. 7 7E-2 Photon Transducers for optical spectroscopy – Barrier-Layer Photovoltaic Cells – Vacuum Phototubes – Photomultiplier tubes (PMT), picture on p. 195 – Silicon Photodiodes (Fig. 7-32) [semiconductor type] [A silicon photodiode transducer consists of a reverse- biased p-n junction.] (The latter two types of transducers are more commonly used in UV-Vis.) – We will discuss more details later.
    [Show full text]
  • Replacing the AMOR with the Minidoas in the Ammonia Monitoring Network in the Netherlands
    Atmos. Meas. Tech., 10, 4099–4120, 2017 https://doi.org/10.5194/amt-10-4099-2017 © Author(s) 2017. This work is distributed under the Creative Commons Attribution 3.0 License. Replacing the AMOR with the miniDOAS in the ammonia monitoring network in the Netherlands Augustinus J. C. Berkhout1, Daan P. J. Swart1, Hester Volten1, Lou F. L. Gast1, Marty Haaima1, Hans Verboom1,a, Guus Stefess1, Theo Hafkenscheid1, and Ronald Hoogerbrugge1 1National Institute for Public Health and the Environment (RIVM), P.O. Box 1, 3720 BA Bilthoven, the Netherlands anow at: Royal Netherlands Meteorological Institute (KNMI), P.O. Box 201, 3730 AE De Bilt, the Netherlands Correspondence to: A. J. C. Berkhout ([email protected]) Received: 16 October 2016 – Discussion started: 3 March 2017 Revised: 31 August 2017 – Accepted: 16 September 2017 – Published: 2 November 2017 Abstract. In this paper we present the continued develop- nitrogen oxides or sulfur dioxide. These aerosols contribute ment of the miniDOAS, an active differential optical ab- to the total burden of particulate matter and may have public sorption spectroscopy (DOAS) instrument used to measure health effects (Fischer et al., 2015). When deposited on na- ammonia concentrations in ambient air. The miniDOAS has ture areas, ammonia causes acidification and eutrophication, been adapted for use in the Dutch National Air Quality Mon- leading to a loss of biodiversity. Intensive animal husbandry itoring Network. The miniDOAS replaces the life-expired leads to high ammonia emissions in the Netherlands. Hence, continuous-flow denuder ammonia monitor (AMOR). From it is of great importance to have a correct understanding of September 2014 to December 2015, both instruments the sources and sinks of ammonia in the Netherlands and of measured in parallel before the change from AMOR to the processes that determine ammonia emissions to the air, miniDOAS was made.
    [Show full text]
  • Fundamentals of Modern UV-Visible Spectroscopy
    Fundamentals of modern UV-visible spectroscopy Primer Tony Owen Copyright Agilent Technologies 2000 All rights reserved. Reproduction, adaption, or translation without prior written permission is prohibited, except as allowed under the copyright laws. The information contained in this publication is subject to change without notice. Printed in Germany 06/00 Publication number 5980-1397E Preface Preface In 1988 we published a primer entitled “The Diode-Array Advantage in UV/Visible Spectroscopy”. At the time, although diode-array spectrophotometers had been on the market since 1979, their characteristics and their advantages compared with conventional scanning spectrophotometers were not well-understood. We sought to rectify the situation. The primer was very well-received, and many thousands of copies have been distributed. Much has changed in the years since the first primer, and we felt this was an appropriate time to produce a new primer. Computers are used increasingly to evaluate data; Good Laboratory Practice has grown in importance; and a new generation of diode-array spectrophotometers is characterized by much improved performance. With this primer, our objective is to review all aspects of UV-visible spectroscopy that play a role in obtaining the best results. Microprocessor and/or computer control has taken much of the drudgery out of data processing and has improved productivity. As instrument manufacturers, we would like to believe that analytical instruments are now easier to operate. Despite these advances, a good knowledge of the basics of UV-visible spectroscopy, of the instrumental limitations, and of the pitfalls of sample handling and sample chemistry remains essential for good results.
    [Show full text]
  • Chapter 2 Experimental Methodology
    CHAPTER 2 Chapter 2 Experimental Methodology 2.1 Introduction 2.2 Synthesis of nanomaterials. 2.2.1 DC arc plasma reactor setup for material processing 2.2.2 Microwave processing of materials 2.2.2.1 Microwave Induced Plasma processing System setup 2.2.2.2 Multimode Microwave setup 2.3 Analytical techniques for characterization nanophase powders 2.3.1 X- ray diffraction study 2.3.2 X-Ray photoelectron spectroscopy 2.3.3 Surface area measurement (BET method) 2.3.4 Electron Microscopy (SEM and TEM) 2.3.4.1 Scanning electron microscopy (SEM) 2.3.4.2 Transmission Electron Microscopy (TEM) 2.3.5 DTA&TGA 2.3.6 Vibrating sample magnetometer (VSM) 2.3.7 Ultraviolet visible spectroscopy (UV/vis) 2.3.8 Raman spectroscopy 2.3.9 Photoluminescence spectroscopy 2.4 Conclusion 2.5 References. 36 CHAPTER 2 Chapter 2 Experimental Methodology 2.1 Introduction This chapter provides the detailed description of the experimental setups for (i) DC arc plasma (reactor) & (ii) microwave processing systems used for the synthesis of nanophase materials and (iii) other analytical-measuring methods such as of X-ray diffraction, X-ray photoelectron spectroscopy, surface area using BET method, scanning and transmission electron microscopy (SEM / TEM), UV-Visible absorption spectroscopy, thermal analysis (DTA / TGA), vibrating sample magnetometer (VSM), Raman and Photoluminescence spectroscopy. 2.2 Synthesis of nanomaterials 2.2.1 DC arc plasma reactor setup for material processing The experimental DC arc plasma setup used for the synthesis of various nano­ materials is as shown in figure 2.1 (a). The plasma reactor consisted of a multi port stainless steel chamber (12 inch in diameter) with a multiple gas inlet facility providing various mixtures of gases, which can be incorporated as plasma forming gas.
    [Show full text]
  • Replacing the AMOR by the Minidoas in the Ammonia Monitoring Network in the Netherlands A.J.C
    Atmos. Meas. Tech. Discuss., doi:10.5194/amt-2016-348, 2017 Manuscript under review for journal Atmos. Meas. Tech. Discussion started: 3 March 2017 c Author(s) 2017. CC-BY 3.0 License. Replacing the AMOR by the miniDOAS in the ammonia monitoring network in the Netherlands A.J.C. (Stijn) Berkhout1, Daan P.J. Swart1, Hester Volten1, Lou F.L. Gast1, Marty Haaima1, Hans Verboom1*, Guus Stefess1, Theo Hafkenscheid1, Ronald Hoogerbrugge1 5 1National Institute for Public Health and the Environment (RIVM), P.O. Box 1, 3720 BA Bilthoven, the Netherlands *Now at Royal Netherlands Meteorological Institute (KNMI), P.O. Box 201, 3730 AE De Bilt, the Netherlands Correspondence to: A.J.C. Berkhout ([email protected]) Abstract. In this paper we present the continued development of the miniDOAS, an active differential optical absorption spectroscopy (DOAS) instrument to measure ammonia concentrations in ambient air. The miniDOAS has been adapted for 10 use in the Dutch National Air Quality Monitoring Network. The miniDOAS replaces the life-expired continuous-flow denuder ammonia monitor (AMOR). From September 2014 to December 2015, both instruments measured in parallel before the change from AMOR to miniDOAS was made. The instruments were deployed on six monitoring stations throughout the Netherlands. We report on the results of this intercomparison. Both instruments show a good uptime of ca. 90%, adequate for an automatic monitoring network. Although both instruments 15 produce minute values of ammonia concentrations, a direct comparison on short timescales such as minutes or hours does not give meaningful results, because the AMOR response to changing ammonia concentrations is slow.
    [Show full text]
  • Hollow Cathode Lamps
    Hollow Cathode Lamps Cathodeon The Ultimate Source MADAtec Srl www.madatec.com Tel. +39-0236542401 Cathodeon HOLLOW CATHODE LAMPS CONTENTS 1 Hollow cathode lamps 5 See through hollow cathode lamps 6 Single element lamps 15 Multi element lamps 22 Getting the best from your hollow cathode lamps 24 Hollow cathode lamps supply 25 Periodic table 1 Hollow Cathode Lamps Cathodeon CATHODEON HOLLOW CATHODE DISCHARGE LAMPS Fig 2 Cathodeon Ltd with over 60 years’ experience in vacuum/gas Internal construction and piece parts HOLLOW electronic devices is the worlds leading specialist in spectral source technology. Now Europe’s undisputed leader in the manufacture of hollow cathode lamps having been at the CATHODE heart of their design and development since the early years of atomic absorption analysis in the 1960’s. LAMPS Cathodeon hollow cathode lamps are fitted as original equipment by many of the worlds foremost atomic absorption instrument manufacturers, and as replacements by discerning users the world over. The Cathodeon hollow cathode lamp programme includes 70 single element and the widest range of proven multi element combinations in standard 11/2” (37mm), and 2” (50mm) diameters designed to fit directly into Perkin Elmer AA APPLICATIONS equipment. Where appropriate these lamps are available These components combined create a mechanically consistent with data coded bases permitting the instrument to automatically Hollow cathode lamps are and robust lamp with the beam of light focused through the identify the lamp element. In addition a range of lamps centre of the window and free from spurious discharge problems. specifically designed for use with Smith Hieftje background gas discharge devices in correction systems is available.
    [Show full text]
  • Deuterium Lamp Layout NEWEST
    Deuterium Lamps Cathodeon The Ultimate Source Cathodeon DEUTERIUM LAMPS CONTENTS 1 Introduction 2 Internal structure 4 Envelopes 5 Window materials 7 Lamps in operation 8 Selection criteria 14 Fibre-coupled deuterium lamp 15 Industrial deuterium lamps 16 VUV lamps 17 Analytical VUV lamps 18 Power supply for deuterium lamps 19 Pulsed deuterium lamps 20 Power supply for pulsed deuterium lamps 21 High power deuterium lamp 22 Power supply and housing for high power deuterium lamp 23 Modulated deuterium lamp system 24 Useful notes 1 Deuterium Lamps Cathodeon CATHODEON DEUTERIUM LAMPS INTERNAL CONSTRUCTION Cathodeon Ltd, with almost 60 years experience in vacuum/gas INTRODUCTION filled electronic devices, is the world leading specialist in Fig 2a Standard deuterium lamp spectral source technology. Involved in the design and Anode Compartment manufacture of deuterium lamps since the earliest days of Arc Aperture instrumental spectroscopy, the company was ideally placed to Anode meet the challenges imposed on deuterium lamp performance Box by the low noise and drift requirements of HPLC during its Dimple Structure rapid growth in the early 1970’s. This long experience and Filament Shield expertise in deuterium arc lamp manufacture has resulted Cathode in an unrivalled range of products to meet many scientific Front Aperture needs. Continuing improvement to production techniques, Cathode quality control standards and test specifications ensure Dimple / Centre Plate Compartment that Cathodeon deuterium lamps remain the first choice Compartment of instrument manufacturers and discerning scientists APPLICATIONS throughout the world. For users wishing to combine light from another source with Deuterium arc lamps are the beam from the deuterium lamp, a “see through” facility is Fig 1 Spectral distribution offered.
    [Show full text]
  • Glass Filters As a Standard Reference Material for Spectrophotometry Selection, Preparation, Certification, and Use of SRM 930 and SRM 1930
    — NATL INST. OF STAND & TECH R.I.C. NISI PUBLICATIONS A 111 DM 2A7D2S 5 NIST SPECIAL PUBLICATION 260-116 U.S. DEPARTMENT OF COMMERCE/Technology Administration National Institute of Standards and Technology Standard Reference Materials: Glass Filters as a Standard Reference Material for Spectrophotometry Selection, Preparation, Certification, and Use of SRM 930 and SRM 1930 R. Mavrodineanu, R. W. Burke, J. R. Baldwin, M. V. Smith, QC J. D. Messman, J. C. Travis, and J. C. Colbert 100 . U57 NO. 260-1 16 1994 The National Institute of Standards and Technology was established in 1988 by Congress to "assist industry in the development of technology . needed to improve product quality, to modernize manufacturing processes, to ensure product reliability . and to facilitate rapid commercialization . of products based on new scientific discoveries." NIST, originally founded as the National Bureau of Standards in 1901, works to strengthen U.S. industry's competitiveness; advance science and engineering; and improve public health, safety, and the environment. One of the agency's basic functions is to develop, maintain, and retain custody of the national standards of measurement, and provide the means and methods for comparing standards used in science, engineering, manufacturing, commerce, industry, and education with the standards adopted or recognized by the Federal Government. As an agency of the U.S. Commerce Department's Technology Administration, NIST conducts basic and applied research in the physical sciences and engineering and performs related services. The Institute does generic and precompetitive work on new and advanced technologies. NIST's research facilities are located at Gaithersburg, MD 20899, and at Boulder, CO 80303.
    [Show full text]
  • Chemical Analysis
    Chemical Analysis Modern Instrumentation Methods and Techniques Second Edition Francis Rouessac and Annick Rouessac University of Le Mans, France Translated by Francis and Annick Rouessac and Steve Brooks Chemical Analysis Second Edition Chemical Analysis Modern Instrumentation Methods and Techniques Second Edition Francis Rouessac and Annick Rouessac University of Le Mans, France Translated by Francis and Annick Rouessac and Steve Brooks English language translation copyright © 2007 by John Wiley & Sons Ltd, The Atrium, Southern Gate, Chichester, West Sussex PO19 8SQ, England Telephone (+44) 1243 779777 Email (for orders and customer service enquiries): [email protected] Visit our Home Page on www.wiley.com Translated into English by Francis and Annick Rouessac and Steve Brooks First Published in French © 1992 Masson 2nd Edition © 1994 Masson 3rd Edition © 1997 Masson 4th Edition © 1998 Dunod 5th Edition © 2000 Dunod 6th Edition © 2004 Dunod This work has been published with the help of the French Ministère de la Culture-Centre National du Livre All Rights Reserved. No part of this publication may be reproduced, stored in a retrieval system or transmitted in any form or by any means, electronic, mechanical, photocopying, recording, scanning or otherwise, except under the terms of the Copyright, Designs and Patents Act 1988 or under the terms of a licence issued by the Copyright Licensing Agency Ltd, 90 Tottenham Court Road, London W1T 4LP, UK, without the permission in writing of the Publisher. Requests to the Publisher should be addressed to the Permissions Department, John Wiley & Sons Ltd, The Atrium, Southern Gate, Chichester, West Sussex PO19 8SQ, England, or emailed to [email protected], or faxed to (+44) 1243 770571.
    [Show full text]
  • Radiometric Calibrations of Portable Sources in the Vacuum Ultraviolet
    Volume 93, Number 1, January-February 1988 Journal of Research of the National Bureau of Standards Radiometric Calibrationsof PortableSources in the Vacuum Ultraviolet Volume 93 Number 1 January-February 1988 Jules Z. Kiose, J. Mervin The radiometric calibration program their uncertainties. Finally, the calibra- Bridges, and William R. Ott carried out by the vacuum ultraviolet tion services are delineated in an ap- radiometry group in the Atomic and pendix. National Bureau of Standards Plasma Radiation Division of the Na- Gaithersburg, MD 20899 tional Bureau of Standards is presented in brief. Descriptions are given of the Key words: arc (argCn); arc (blackbody); primary standards, which are the hydro- arc (hydrogen); irradiance; lamp (deu- gen arc and the blackbody line arc, and terium); radiance; radiometry; Standards the secondary standards, which are the (radiometric); ultraviolet; vacuum ultra- argon mini- and maxi-arcs and the deu- violet. terium arc lamp. The calibrationmeth- ods involving both spectral radiance and irradiance are then discussed along with Accepted: October 2, 1987 1. Introduction desired wavelength. In this case the most direct procedure is to employ a standard source. The The vacuum ultraviolet (VUV) region of the source to be investigated as well as the standard spectrum has become important in several areas of source are set up in turn so that radiation from each research and development. These include space- source passes through the same monochromator based astronomy and astrophysics, thermonuclear and optical elements. The calibration is performed fusion research, ultraviolet laser development, and essentially by a direct substitution of the standard general atomic physics research. Applications of source for the one to be calibrated.
    [Show full text]