Functions of Anesthesia Reservoir Bag in a Breathing System

Total Page:16

File Type:pdf, Size:1020Kb

Functions of Anesthesia Reservoir Bag in a Breathing System Professional article Zdrav Vestn | May – June 2017 | Volume 86 Metabolic and hormonal disorders id PROFESSIONAL ARTICLE Functions of anesthesia reservoir bag in a breathing system Miljenko Križmarić Abstract Faculty of Medicine, This article addresses the different functions of anaesthesia reservoir bag in a circle anesthesia University of Maribor, breathing system. A main purpose of the contribution was to explain complex interaction between Maribor the reservoir bag and fresh gas flow during mechanical ventilation. The anesthesia reservoir bag is a collapsible gas container which is an essential component of most anesthesia breathing system. Correspondence: The anesthesia reservoir bag permits manual ventilation and acts as a visual or tactile indicator of Miljenko Križmarić, e: miljenko.krizmaric@ spontaneous breathing. The bag was excluded from traditional breathing system when the mechani- um.si cal ventilator was in use. Discrepancies between the set and actual tidal volume can occur. However, on some anaesthesia workstation systems, such as the Dräger Primus, the reservoir bag is an integral Key words: part of the breathing system during mechanical ventilation, where it serves as a reservoir for oxygen anesthesia breathing and anaesthetic gases. In mechanically ventilated patients, gases enter the reservoir bag from the systems; circle fresh gas flow during inspiratory phase, when the decoupling valve closes. The safe administration anesthesia breathing of general anaesthesia requires understanding of the technological advances in highly sophisticated systems; anesthesia reservoir bag; fresh gas anaesthetic equipment. decoupling Cite as: Cite as: Zdrav Vestn. 2017; 86:226–35. Zdrav Vestn. 2017; 86:226–35. Received: 14. 9. 2016 Accepted: 5. 3. 2017 1. Introduction The anaesthesia reservoir bag is an The reservoir bag is most often placed important component in most breath- between the expiratory valve and CO2 ing systems used in anaesthesia. Figure absorber. Such a configuration allows 1 presents a traditional or classical circle for the dust from absorbent beads to be anaesthesia breathing system that com- pushed into the inspiratory limb during prises 7 different components: manual or mechanical ventilation (3). • inflow of fresh gases, As seen in Figure 1, the inflow of • inspiratory and expiratory unidirec- fresh gases at inspiratory phase is di- tional valve; rected to the patient through the inspi- • an inspiratory and expiratory corru- ratory valve, the inspiratory limb (tube) gated tubes (limbs) and the Y-piece. Expired air containing • Y-piece; carbon dioxide (CO2) flows through • adjustable pressure-limiting valve the expiratory limb through the expi- • CO2 absorber with absorbent; ratory valve and fills the reservoir bag. • reservoir bag (1-2). When the pressure in the bag increases Functions of anesthesia reservoir bag in a breathing system on line edition METABOLIC AND HORMONAL DISORDERS Figure 1: Traditional dotok svežih plinov anaesthesia breathing (Fresh gas flow) system. absorber CO2 ven�lator (Ven�lator) inspiracijska zaklopka (CO2 absorber) (Inspiratory valve) inspiracijski krak nastavek Y (Inspiratory limb) (Y connect) preklopnik (Bag/ven�lator selector switch) ekspiracijski krak bolnik (Ekspiratory limb) (Pa�ent) dihalni balon (Reservoir bag) ekspiracijska zaklopka (Ekspiratory valve) razbremenilni ven�l (Adjustable pressure-limi�ng valve) above the value set on the pressure- the tube is attached with greater force limiting valve, the respiratory gases and thus less likely to slip off. are released from the breathing system ISO 5362 is the international ISO stan- through this pressure-limiting valve dard that defines the characteristics of into the scavenging system. On expira- anaesthetic reservoir bags ISO 5362:2006 tion, the respiratory gases fill the reser- – Anaesthetic reservoir bags) (7). In ac- voir bag, and on inspiration these gases cordance with this standard, the maxi- return to the patient from the reservoir mum permissible gas leakage in reser- bag, thereby passing through the CO2 voir bags of original nominal volume of absorber, which absorbs carbon diox- 1 l is 10 ml/min. In reservoir bags with a ide (CO2). The valves direct the circular volume of more than 1 l, the maximum one-way flow of respiratory gases (4). permissible gas leakage is 25 ml/min. In contrast to the presented traditional The leakage is measured at a pressure of Circle-breathing system, different man- 3 ± 0.3 kPa. Reservoirs can be made of ufacturers may have individual compo- natural rubber and contain latex (often nents placed differently. reusable) or are made of latex-free mate- The disadvantage of the circular rials and are disposable (the note that the breathing system is in a large number of reservoir is latex-free should be indicat- different components where malfunc- ed on the packaging). The force that the tions can occur (5). These occur due to reservoir is required to withstand during incorrectly connected connection sites, the linear load testing of the reservoir leakage of gas, obstruction of the tube, neck is 40 ± 4 N (7). The tolerances for cracks in the filters, or because the tube the nominal volume range within ± 15 %. may accidentally slip off the coupling. Connection to the breathing system is When connecting the tube, it is neces- made by means of two connectors with sary to keep in mind that the tube should internal diameters of 15 mm (15F, female be pushed and rotated (the manufac- conical fitting) and 22 mm (22F, female turer in the user manual talks about the conical fitting). The material of the res- “push and twist” procedure). In this case ervoir must be antistatic (i.e. not prone on line edition Zdrav Vestn | May – June 2017 | Volume 86 PROFESSIONAL ARTICLE ration function is then taken over by a mechanical ventilator. 2. Reservoir bag and spontaneous breathing The most important function of a reservoir bag during patient’s sponta- neous breathing is the collection of re- spiratory gases. In the patient’s inspira- tion of 500 ml of gas for 1 second, a high average flow of 30 l/min (0.5 l × 60 s) is generated; the maximum values in the middle of inspiration are higher by Figure 2: Breathing to static electricity buildup) and should 30–40 % (12). The inflow of fresh gases reservoir bags of different volumes and not slip when held in hands. The material does not allow for such high flows, so couplings (Intersurgical should also not be permeable to inspira- the gasses must be collected in a respi- Ltd). tion anaesthetics and should not absorb ratory reservoir bag. During spontane- or adsorb them. The upper part of the ous breathing, the patient inhales fresh reservoir with the coupling is called the respiratory gases along with gases from “neck” and the opposite part is the “tail”. the reservoir bag, which are devoid of The tail, which is a tubular extension of CO2 as they pass through the CO2 ab- the reservoir, must be at least 2 cm long. sorber and the inspiratory valve, with When open (i.e. open tail), it functions the expiratory limb being closed via the as a relief mechanism in the linear Ma- expiratory one-way valve. During expi- pleson F breathing system. The reservoir ration, the reservoir bag is filled with ex- bag has an ellipsoid shape, which facili- haled air and the inflow of fresh gases as tates grip (7). the inspiratory valve is closed (Figure 1). Reservoir bags intended for paediat- When the reservoir bag is filled up to its ric use generally have a nominal volune nominal volume, the excess gases are re- of 0.5 or 1 litre (Figure 2, three lower res- leased through the adjustable pressure- ervoir bags), and for adults a volume of 2 limiting valve. In this case, the pressure or 3 l (two upper reservoir bags in Figure limiting outflow valve in the Dräger 2). The lower two reservoir bags in Fig- Primus anaesthesia workstation is in ure 2 have couplings with a standardised the position marked “Spont” (fully open internal diameter of 15 mm (15F) while valve). other reservoir bags have a standardised During spontaneous breathing, the internal diameter of 22 mm (22F). reservoir bag is a visual monitor; it is In the traditional circle breathing sys- slightly deflated on inspiration and re- tem, a reservoir bag has a function only filled on expiration. The reservoir bag is in spontaneous breathing or in manual also a tactile monitor, as its soft structure positive pressure ventilation. During enables us to feel its filling or emptying mechanical ventilation, it is excluded by touch (23). from the breathing system by bag/ven- The monitoring of spontaneous tilator selector switch (Figure 1) and in breathing depends on several factors: the this mode it has no function. The respi- shape of the reservoir bag, its size, the Functions of anesthesia reservoir bag in a breathing system on line edition METABOLIC AND HORMONAL DISORDERS to that of spontaneous breathing. In as- sisted ventilation, the onset of inspirato- ry phase can be determined by tactile or visual monitoring of the reservoir bag. 4. Safety pressure The reservoir bag has a safety function which protects the patient’s lung from barotrauma. It is the most flexible part of the breathing system. During gas filling, it is the reservoir bag that first shows evi- dence of filling. When the pressure limit- ing outflow valve is closed, the reservoir Figure 3: A protective level of inflation, the flow of fresh gases, bag is filled beyond the nominal volume, cage on the inside of the 22F coupling. and the patient’s respiratory capacity. but nevertheless it must maintain a safe- The accurate assessment of the pa- ty pressure within the limits below 60 cm tient’s respiratory volume cannot be H2O (7). Laplace’s law shows the correla- done simply by observing the reservoir tion between the tension on the surface bag (8).
Recommended publications
  • Maintenance of Airway Pressure During Filter Exchange Due to Auto-Triggering
    Maintenance of Airway Pressure During Filter Exchange Due to Auto-Triggering Joakim Engstro¨m MSc CCRN, Henrik Reinius MD, Camilla Fro¨jd PhD CCRN, Hans Jonsson MSc CCRN, Go¨ran Hedenstierna MD PhD, and Anders Larsson MD PhD BACKGROUND: Daily routine ventilator-filter exchange interrupts the integrity of the ventilator circuit. We hypothesized that this might reduce positive airway pressure in mechanically ventilated ICU patients, inducing alveolar collapse and causing impaired oxygenation and compliance of the respiratory system. METHODS: We studied 40 consecutive ICU subjects (P /F ratio < 300 mm aO2 IO2 Hg), mechanically ventilated with pressure-regulated volume control or pressure support and > PEEP 5cmH2O. Before the filter exchange, (baseline) tidal volume, breathing frequency, end-inspiratory plateau pressure, and PEEP were recorded. Compliance of the respiratory system was calculated; F , blood pressure, and pulse rate were registered; and P ,P , pH, and base IO2 aO2 aCO2 excess were measured. Measurements were repeated 15 and 60 min after the filter exchange. In addition, a bench test was performed with a precision test lung with similar compliance and ؎ resistance as in the clinical study. RESULTS: The exchange of the filter took 3.5 ؎ 1.2 s (mean SD). There was no significant change in P (89 ؎ 16 mm Hg at baseline vs 86 ؎ 16 mm Hg at 15 aO2 or in compliance of the respiratory system (41 ؎ 11 (24. ؍ min and 88 ؎ 18 mm Hg at 60 min, P ؍ ؎ ؎ mL/cm H2O at baseline vs 40 12 mL/cm H2O at 15 min and 40 12 mL/cm H2O at 60 min, P .32).
    [Show full text]
  • VENTILATION THIS IS Evita® V600 Evita® V800
    THIS IS VENTILATION Evita® V800 Evita® V600 D-18935-2010 D-2669-2019 2 | Our mission: Improving outcomes in Intensive Care of patients who are artificially ventilated for 41% at least 14 days, will survive the next year.1 As your specialist in acute care, we focus on reducing mortality rates in the ICU. Supporting patient outcome and increasing staff satisfaction in the ICU via connected technologies and services that help achieve therapeutic goals faster and safer is what drives us. AVOIDING ICU-ACQUIRED WEAKNESS AVOIDING COGNITIVE IMPAIRMENT • Start the weaning process as early as • Provide a comfortable and supportive environment possible to help reduce ventilation time. that helps your patient feel calm and at ease. • Encourage spontaneous breathing, which • Turn the ICU into a healing environment and helps train of respiratory muscles and start enable the patient to feel more comfortable in early mobilisation activities. a family-friendly integrated surrounding. • Improve sedation management and optimise patient interaction. PREVENTION OF VALI / ARDS • Protect the lung with personalised ventilation strategies. • Support spontaneous breathing at any time to provide a seamless transition from controlled ventilation to patient contribution. • Improve clinical results with decision supporting insights from multiple sources. 1) Damuth et al. Lancet Respir Med 2015 D-2795-2019 | 3 1) Damuth et al. Lancet Respir Med 2015 D-3460-2018 4 | All ventilation strategies combined in one device Evita® V600 HIGH-END VENTILATION Evita devices have supported you for over 25 years with high quality standards, the ability to configure and upgrade your machine, and advanced training and service concepts. Experience the next level of ventilator operation.
    [Show full text]
  • Anesthesia Delivery System Consideration for Use As a Ventilator
    March 23, 2020 Mindray A-Series Anesthesia Delivery System Consideration for use as a Ventilator Dear Valued Customer: In response to the COVID-19 pandemic, the US Food and Drug Administration (FDA) has provided guidance1 that allows for the use of anesthesia delivery systems for continuous ventilation. This is due to the increased demand for mechanical ventilation that is estimated to exceed the available ventilators in Intensive Care Units (ICUs) in the US. While Mindray does not promote anesthesia delivery systems for continuous ventilation, this information is provided in direct response to the current public health crisis. It is advised that the user becomes completely familiar with the information provided and considers all risks and benefits prior to using the anesthesia system for continuous ventilation. When considering the use of an anesthesia delivery system for continuous ventilation, the following functional differences should first be noted: 1) The indications for use are different (please reference the appropriate product Operator’s Manual at https://www.mindraynorthamerica.com/technical-documents/). Specifically, an anesthesia system is used, primarily, in Operating Rooms (ORs) where the case may last for a relatively short period or for several hours, under worst-case conditions. A continuous stand-alone ventilator is used, primarily, in the ICU and may operate continuously for several days. 2) Anesthesia systems are used, primarily, for mandatory ventilation on sedated and muscle- relaxed patients. Continuous stand-alone ventilators are used for spontaneous breathing support. 3) Manual mode on the anesthesia system is used for manually ventilating a patient or letting a patient breathe spontaneously, and is only available on the anesthesia systems (when the APL valve is set to “SP”, the PEEP will be zero (0) and the patient can respire spontaneously).
    [Show full text]
  • Severe Respiratory Acidosis Resulting from a Compromised Coaxial Circuit
    ISSN: 2377-4630 Dang et al. Int J Anesthetic Anesthesiol 2018, 5:072 DOI: 10.23937/2377-4630/1410072 Volume 5 | Issue 2 International Journal of Open Access Anesthetics and Anesthesiology CASE REPORT When Equipment Misbehaves: Severe Respiratory Acidosis Result- ing from a Compromised Coaxial Circuit 1* 1 2 1 Check for Anh Q Dang , Marc Rozner , Nicole Luongo and Shital Vachhani updates 1Department of Anesthesiology & Perioperative Medicine, The University of Texas MD Anderson Cancer Center, USA 2Department of Anesthesiology-CRNA Support, The University of Texas MD Anderson Cancer Center, USA *Corresponding author: Anh Q Dang (MD), Department of Anesthesiology & Perioperative Medicine, The University of Texas MD Anderson Cancer Center, 1400 Holcombe Boulevard, Unit 409, Houston, TX 77030, USA, Tel: (713)-792-6911, Fax: (713)-745-2956 ratory and expiratory unidirectional valves, inspiratory Abstract and expiratory corrugated tubing, a Y connector, an The construction of a coaxial circle breathing circuit enclos- overflow valve (also called a pop-off or adjustable pres- es the inspiratory limb within the expiratory limb to preserve heat, humidity, anesthetic gas, and oxygen. However, the sure-limiting valve), and a reservoir bag. Circle breath- construction also makes kinks or breaks in the enclosed ing circuits facilitate the use of low-flow systems, and inspiratory limb difficult to detect. With kinking or discon- their low resistance makes them appropriate for both nection of the central inspiratory limb, the patient may re- pediatric and adult patients. breathe expiratory gases, leading to a significant increase in dead space and concomitant respiratory acidosis. We A coaxial system (King Circuits, Noblesville, IN) is a describe a clinical case in which a disconnection in the in- circle breathing circuit that encloses the inspiratory limb ner inspiratory limb of a King (King Circuits, Noblesville, IN) coaxial breathing circuit resulted in severe respiratory aci- within the expiratory limb.
    [Show full text]
  • Mechanical Ventilation
    Fundamentals of MMeecchhaanniiccaall VVeennttiillaattiioonn A short course on the theory and application of mechanical ventilators Robert L. Chatburn, BS, RRT-NPS, FAARC Director Respiratory Care Department University Hospitals of Cleveland Associate Professor Department of Pediatrics Case Western Reserve University Cleveland, Ohio Mandu Press Ltd. Cleveland Heights, Ohio Published by: Mandu Press Ltd. PO Box 18284 Cleveland Heights, OH 44118-0284 All rights reserved. This book, or any parts thereof, may not be used or reproduced by any means, electronic or mechanical, including photocopying, recording or by any information storage and retrieval system, without written permission from the publisher, except for the inclusion of brief quotations in a review. First Edition Copyright 2003 by Robert L. Chatburn Library of Congress Control Number: 2003103281 ISBN, printed edition: 0-9729438-2-X ISBN, PDF edition: 0-9729438-3-8 First printing: 2003 Care has been taken to confirm the accuracy of the information presented and to describe generally accepted practices. However, the author and publisher are not responsible for errors or omissions or for any consequences from application of the information in this book and make no warranty, express or implied, with respect to the contents of the publication. Table of Contents 1. INTRODUCTION TO VENTILATION..............................1 Self Assessment Questions.......................................................... 4 Definitions................................................................................
    [Show full text]
  • Anaesthetic Machine Anatomy
    Anaesthetic Machine Anatomy Year Group: BVSc3 + Document Number: CSL_A00 Equipment list: Anaesthetic Machine Anatomy Equipment for this station: • Anaesthetic machine • Name labels • Function labels Considerations for this station: • Do not attempt to attach cylinders or connect the oxygen pipeline, this machine is for reference only and is NOT a working machine. • The first time you try to complete this task it may be worth refreshing your memory of the anaesthetic machine by reading the section of this booklet marked ‘Answers’. Anyone working in the Clinical Skills Lab must read the ‘CSL_I01 Induction’ and agree to abide by the ‘CSL_I00 House Rules’ & ‘CSL_I02 Lab Area Rules’ Please inform a member of staff if equipment is damaged or about to run out. Clinical Skills: Anaesthetic Machine Anatomy 1 2 3 Using the name labels On the bottom of the name On some of the function provided, name each part of label, place a function label labels there are additional the anaesthetic machine (match the circular tabs). questions. (match/stick the white square Place the correct answers in velcro tab to the yellow the space provided (match square tab). the semi-circular tabs). 4 5 You will need to lift the lid Once you have placed all of to find all of the the labels, use the components! information on the following pages of this booklet to check your answers. Here are some online resources and tutorials that you may find useful: 1. http://mhra.gov.uk/learningcentre/AnaestheticMachines/player.html 2. https://www.youtube.com/watch?v=1LY0eAzrIrE ANSWERS: Anaesthetic Machine Anatomy ANSWERS The following pages contain the answers i.e.
    [Show full text]
  • APSF/ASA Anesthesia Machines As ICU Ventilators
    Updated: May 7, 2020 APSF/ASA Guidance on Purposing Anesthesia Machines as ICU Ventilators Anesthesia machines are equipped with ventilators that in many cases are capable of providing life-sustaining mechanical ventilation to patients with respiratory failure. They are used for this purpose every day in the operating room. FDA approved labeling does not provide for using anesthesia ventilators for long term ventilatory support. Nevertheless, anesthesia ventilators are an obvious first-line backup during the COVID-19 pandemic when there are not sufficient ICU ventilators to meet the patient care needs. Local resources and constraints will impact how this solution can best be implemented. Anesthesia machines not currently being used may be located in your own hospital operating rooms, NORA locations, at nearby ambulatory surgery centers, nearby office-based surgery practices, and through your anesthesia equipment distributors. Guidance is available from the manufacturers, but it may not convey all of the clinical considerations. Anesthesia professionals will be needed to put these machines into service and to manage them while in use. Safe and effective use requires an understanding of the capabilities of the machines available, the differences between anesthesia machines and ICU ventilators, and how to set anesthesia machine controls to mimic ICU-type ventilation strategies. This document is intended to provide guidance on using anesthesia machines safely and effectively as ICU ventilators. Detailed information is provided below and a quick reference guide is available for downloading. The quick reference guide is intended to be a bedside tool and includes a suggested schedule for monitoring the effectiveness and safety of the anesthesia ventilator.
    [Show full text]
  • An Introduction to Mechanical Ventilation
    Update in Anaesthesia An introduction to mechanical ventilation Fran O’Higgins, Adrian Clarke Correspondence Email: [email protected] INTRODUCTION • Hypercarbia - PaCO2 > 8kPa, One of the main interventions offered in an Intensive • Decreasing conscious level (Glasgow Coma Care Unit (ICU) is advanced respiratory support. Score < 8), Respiratory An understanding of the indications and types of mechanical ventilation is therefore essential for anyone • Significant chest trauma, working in this environment. • Tidal volume < 5ml.kg-1 or vital capacity < 15ml.kg-1. Summary INDICATIONS FOR MECHANICAL VENTILATION Mechanical ventilation The main indication for mechanical ventilation is Causes of respiratory failure is the major invasive respiratory failure. However, other clinical indications intervention offered in the include: Inadequate gas exchange ICU. In low income countries, where the facilities exist, • prolonged postoperative recovery, • Pneumonia, pulmonary oedema, acute respiratory distress syndrome (ARDS). it is appropriate that this • altered conscious level, therapy is only offered to Inadequate breathing patients suffering reversible • inability to protect the airway, or • Chest wall problems e.g. fractured ribs, flail critical illness. The basics of • exhaustion, when the patient is likely to develop mechanical ventilation are chest, described, with particular respiratory failure. • Pleural problems e.g. pneumothorax, focus on overcoming The aim of mechanical/artificial ventilation is to haemothorax, common pitfalls. improve gas exchange, to reduce the work of breathing and to avoid complications while maintaining • Respiratory muscle failure e.g. myasthenia gravis, optimal conditions for the patient to recover from poliomyelitis, tetanus, their underlying illness. Whatever the indication • Central nervous system depression e.g. drugs, for respiratory support, the underlying condition of brain stem compression, head injury.
    [Show full text]
  • Shared Ventilation in the Era of COVID-19: a Theoretical Consideration of the Dangers and Potential Solutions
    Research Article | Original Research Shared Ventilation in the Era of COVID-19: A Theoretical Consideration of the Dangers and Potential Solutions https://doi.org/10.4187/respcare.07919 Cite as: RESPCARE 2020; 10.4187/respcare.07919 Received: 7 April 2020 Accepted: 22 April 2020 This Fast Track article has been peer-reviewed and accepted, but has not been through the composition and copyediting processes. The final version may differ slightly in style or formatting and will contain links to any supplemental data. Alerts: Sign up at rc.rcjournal.com/alerts to receive customized email alerts when the fully formatted version of this article is published. Page 1 of 50 Respiratory Care Shared Ventilation in the Era of COVID-19: A Theoretical Consideration of the Dangers and Potential Solutions Jacob Herrmann, PhD1,3 Andrea Fonseca da Cruz, PhD2 Monica L Hawley, PhD3 Richard D Branson, MSc, RRT, FAARC4 David W Kaczka, MD, PhD3,5 1Department of Biomedical Engineering Boston University Boston, MA, USA 2Department of Anesthesia University of Iowa Iowa City, IA, USA 3OscillaVent, Inc. Iowa City, IA, USA 4Division of Trauma and Critical Care Department of Surgery University of Cincinnati Cincinnati, OH, USA 5Departments of Anesthesia, Biomedical Engineering, and Radiology University of Iowa Iowa City, IA, USA Author Contributions: Literature search, Analysis of data, Review of manuscript: JH, AFC, RDB, MLH, DWK. Study design, Data collection, Manuscript preparation: JH, AFC, MLH, DWK. Industry Relationships: DK: ZOLL Medical Corporation and Monitor Mask, Inc. RDB: Mallinckrodt, Ventec Life Systems, and ZOLL Medical Corporation David W. Kaczka, MD, Ph.D. The University of Iowa Hospital and Clinics Department of Anesthesia 200 Hawkins Drive Iowa City, IA 52242 Email: [email protected] 1 Respiratory Care Page 2 of 50 QUICK LOOK Current Knowledge: A single mechanical ventilator can support multiple patients connected in parallel as a last resort for increasing surge capacity.
    [Show full text]
  • Non-Invasive-Ventilation.Pdf
    Early Human Development 89 (2013) S25–S31 Contents lists available at ScienceDirect Early Human Development journal homepage: www.elsevier.com/locate/earlhumdev Non-invasive positive pressure ventilation in children☆ M. Pavone ⁎, E. Verrillo, V. Caldarelli, N. Ullmann, R. Cutrera Respiratory Unit, Pediatric Department, Bambino Gesù Children's Hospital, Piazza S. Onofrio 4, 00165 Rome, Italy article info abstract Keywords: Non-invasive positive pressure ventilation is increasinglyusedinchildrenbothinacuteandinchronicsetting.Clin- Non-invasive ventilation ical data supporting safety, efficacy and limitations in children are growing. Technical problems related to the ven- Positive airway pressure tilators performance and interfaces selection have not been fully resolved, especially for younger children. Non- Children invasive ventilation can be applied at home. Its use at home requires appropriate diagnostic procedures, accurate titration of the ventilators, cooperative and educated families and careful, well-organized follow-up programs. © 2013 Elsevier Ireland Ltd. All rights reserved. 1. Introduction patient-ventilator connection. For non-invasive ventilation, gases are conducted into the airways via an external interface, in the case of inva- An increasing number of children with chronic hypercapnic respira- sive ventilation through an endotracheal tube or tracheostomy [2,10]. tory failure are currently treated with non-invasive ventilation [1].Acute Non-invasive ventilation in children is indicated essentially for: accesses to the emergency department or in-hospital follow-up controls 1) Diseases due to increased respiratory load (intrinsic cardio- of patients undergoing home long-term non-invasive ventilation are pulmonary disorders, abnormalities of the upper airways, especially increasingly common occurrences [2–4]. skeletal deformities of the chest wall); 2) Disorders characterized by Non-invasive ventilation has some undoubted peculiarities.
    [Show full text]
  • Breathing Circuit Compliance and Accuracy of Displayed Tidal Volume During Pressure-Controlled Ventilation of Infants: a Quality Improvement Project
    Accepted: 30 March 2017 DOI: 10.1111/pan.13164 RESEARCH REPORT Breathing circuit compliance and accuracy of displayed tidal volume during pressure-controlled ventilation of infants: A quality improvement project Todd A. Glenski | Carrie Diehl | Rachel G. Clopton | Robert H. Friesen Department of Anesthesiology, Children’s Hospital Colorado, University of Colorado Summary School of Medicine, Denver, CO, USA Introduction: Anesthesia machines have evolved to deliver desired tidal volumes Correspondence more accurately by measuring breathing circuit compliance during a preuse self-test Dr. Todd A. Glenski, Department of and then incorporating the compliance value when calculating expired tidal volume. Anesthesiology, Children’s Mercy Hospital, Kansas City, MO, USA. The initial compliance value is utilized in tidal volume calculation regardless of Email: [email protected] whether the actual compliance of the breathing circuit changes during a case, as Section Editor: Dr. Jerrold Lerman happens when corrugated circuit tubing is manually expanded after the preuse self- test but before patient use. We noticed that the anesthesia machine preuse self-test was usually performed on nonexpanded pediatric circuit tubing, and then the breathing circuit was subsequently expanded for clinical use. We aimed to demon- strate that performing the preuse self-test in that manner could lead to incorrectly displayed tidal volume on the anesthesia machine monitor. The goal of this quality improvement project was to change the usual practice and improve the accuracy
    [Show full text]
  • STANDARDIZED VOCABULARY for MECHANICAL VENTILATION V 9.12.19  Robert L
    STANDARDIZED VOCABULARY FOR MECHANICAL VENTILATION V 9.12.19 Robert L. Chatburn 2015 https://www.ncbi.nlm.nih.gov/pubmed/22004898 Key Definitions for Taxonomy Maxim BREATH 1. A breath is one cycle of positive flow (inspiration) and negative flow INSPIRATORY TIME (expiration) defined in terms of the flow-time curve EXPIRATORY TIME TOTAL CYCLE TIME I:E RATIO DUTY CYCLE TIDAL VOLUME ASSISTED BREATH 2. A breath is assisted if the ventilator provides some or all of the work of WORK OF BREATHING breathing TRANSRESPIRATORY PRESSURE LOADED BREATH EQUATION OF MOTION 3. A ventilator assists breathing using either pressure control or volume control VOLUME CONTROL based on the equation of motion for the respiratory system PRESSURE CONTROL TRIGGER 4. Breaths are classified according to the criteria that trigger (start) and cycle TRIGGER VARIABLE (stop) inspiration CYCLE CYCLE VARIABLE PATIENT TRIGGERING 5. Trigger and cycle events can be either patient-initiated or ventilator-initiated PATIENT CYCLING MACHINE TRIGGERING MACHINE CYCLING TRIGGER WINDOW SYNCHRONIZATION WINDOW PRESSURE TRIGGERING PRESSURE CYCLING VOLUME TRIGGERING VOLUME CYCLING FLOW TRIGGERING FLOW CYCLING MANDATORY BRATH 6. Breaths are classified as spontaneous or mandatory based on both the trigger SPONTANEOUS BREATH and cycle events CONTINUOUS MANDATORY VENTILATION 7. Ventilators deliver 3 basic breath sequences: CMV, IMV, and continuous INTERMITTENT MANDATORY VENTILATION spontaneous ventilation CSV CONTINUOUS SPONTANEOUS VENTILATION VC-CMV 8. Ventilator deliver 5 basic ventilatory patterns: VC-CMV, VC-IMV, PC-CMV, PC- VC-IMV IMV, PC-CSV PC-CMV PC-IMV PC-CSV TC-IMV TARGET 9. Within each ventilatory pattern, there are several types that can be TARGETING SCHEME distinguished by their targeting schemes (set-point, dual, bio-variable, servo, SET-POINT TARGETING SCHEME adaptive, optimal, and intelligent) DUAL TARGETING SCHEME BIOVARIABLE TARGETING SCHEME SERVO TARGETING SCHEME ADAPTIVE TARGETING SCHEME OPTIMAL TARGETING SCHEME INTELLIGENT TARGETING SCHEME MODE OF VENTILATION 10.
    [Show full text]