NELAC Quality Systems Checklist s1

NELAP Quality Systems Checklist Chemistry On-site

Date:______

Assessor:______

Organization Name:______

Analytical Grouping: SVOCs______

Methods: 525.2______

Person(s) Interviewed:______

Records Reviewed:______

Category Comments

Blank

Terphenyl-D14 – recovery standard – add to each extract Confirmation

CCV Limit: ± 30%, ISTD ±50% from ICAL – beginning & end of run, IS>70%

Additional : One CCV per tune window. Continuing Tune window 12 hours,. DFTPP Calibration DDT Breakdown ≤20%. Resolution - Anthracene and phenanthrene should be separated by baseline. Benz[a]anthracene and chrysene should be separated by a valley whose height is less than 25% of the average peak height of these two compounds.

Data Analysis

Data Review

IDOC: ±30%, RSD <30% Demonstration of CDOC: ±30%, RSD <30% Capability

Duplicates

Equipment

Facilities

10 January 2006 1 of 5 03d04a0af544dbde780dfd9ac5405d5f.doc Category Comments

Glassware

Minimum # of Cal Points: 3, RSD Limit: 30% Correlation Coefficient: ≥ 0.990 Initial Calibration RSD of RF <30% ICV Limit: ± 30%

Limit: ± 30% LCS

LODV: 1-4x LOD LOQV: 1-2x LOQ LOD/LOQ

Marginal Exceedences

Limit: ± 30% Matrix Spike

Method Validation

Non-Standard Methods

Personnel

Proficiency Testing

Ion ratios: <20% QC

Reagents & Media

Records

Records-Electronic

10 January 2006 2 of 5 03d04a0af544dbde780dfd9ac5405d5f.doc Category Comments

Records-Raw Data

1L sample volume. Add 40-50mg Na2S2O3 . pH adjusted to < 2 with 6N HCl.

Sample Receiving Cyanazine: dechlorination/acidification immediately prior to analysis - analyze within 14 days; Atraton & Prometon: dechlorinate only.

Extraction Hold Time: 14 days Analytical Hold Time: 30 days Samples

Selectivity

SOP

Support Equipment

Surrogate Limit: ± 30% - 1,3-Dimethyl-2-Nitrobenzene, Perylene-d12, Triphenylphosphate Internal Standard Limit: ≥70% of mean ICAL – Acenaphthene-d10, Chrysene-d12, Phenanthrene-d10 Surrogates For SPE, add some of IS before extraction

Traceability

Work Cell

10 January 2006 3 of 5 03d04a0af544dbde780dfd9ac5405d5f.doc Category Comments

Additional Comments:

Tuning Criteria From To Relative to m/z 51 10 80 198 68 0 2 69 70 0 2 69 127 10 80 198 197 0 2 198 198 base peak 199 5 9 198 275 10 60 198 365 1 100 198 441 present less than 443 442 (alt. base peak) 50 100 198 443 15 24 442

Extraction – 1 L sample - from disk or cartridge – ethyl acetate, then MeCl2, evap down

10 January 2006 4 of 5 03d04a0af544dbde780dfd9ac5405d5f.doc 10 January 2006 5 of 5 03d04a0af544dbde780dfd9ac5405d5f.doc