Bimonthly Report of Field Quality

Total Page:16

File Type:pdf, Size:1020Kb

Bimonthly Report of Field Quality

Report on field quality in the main LHC dipole collared coils: July-August 2002

E. Todesco, LHC-MMS-MA

This report gives data relative to field quality measured in collared coils during the period July 1st–August 31th 2002, comparison to beam dynamics targets and status of the production holding points. Updated graphs can be found in the LHC field quality observatory http://lhc-div- mms.web.cern.ch/lhc-div-mms/MMSPAGES/MA/Obs.html. Please note that the web address has changed.

What’s new

 Available measurements: 45 collared coils, 18 cold masses, 10 cryodipoles.  All the manufacturers are now producing collared coils with the new cross-section. Two collared coils with old cross-section still have to be measured (Ansaldo 11 and Noell 10). The total number of collared coils with the old cross-section will be 35 (Alstom 1-12 and 15, Ansaldo 1 to 11, Noell 1 to 11).

 We present for the first time graphs for the skew (a2, a3, a4) and for the even multipoles (b2 and b4). Control limits on these multipoles are based on the LHC Project Report 501 and on communications of S. Fartoukh to the Magnet Evaluation Board.  Control limits on collared coil data are set through correlations with measurements at 1.9 K of 10 cryomagnets made by LHC-MTA. Three more magnets have been added with respect to the previous report of July 1st 2002.

1. Measured magnets and assembly data

 7 collared coils have been measured (collared coils 39th to 45th) o 3 Alstom (HCMB__A001-01000018, 19 and 21) o 1 Ansaldo (HCMB__A001-02000014) o 3 Noell (HCMB__A001-01000011, 12 and 13)

50 45 s l i 40 o c 35 d e r 30 a l l o

c 25

f o

20 r e

b 15 m

u 10 N 5 0 Aug-00 Dec-00 Apr-01 Aug-01 Dec-01 Apr-02 Aug-02

L H C - M M S

Fig. 1: Number of measured collared coils versus time

 Cross section: Noell 11 have X-section 1 (the old one); all the others have X-section 2. 1  Shims (see table I and Fig. 2 for a summary over all the collared coils produced so far): o Azimuthal sizes of Ansaldo 14 coils are 0.1-0.15 mm larger than the nominal (both inner and outer layer), and therefore thinner shims have been used (see Table I). This was already observed in collared coil 12 and 13 (see report of May-June 2002). Actions have been taken by the project engineers to change the shims of the curing mould. We still expect a few coils with non-nominal shims. o Outer shim of Noell 11 (old cross-section) is 0.05 mm larger. This has a very limited impact on field quality. o The remaining collared coils have nominal shims.

Table I: Shims thickness and coil cross-section type of measured collared coils. Nominal shims: 0.2 mm inner layer, 0.8 mm outer layer

Shim (mm) Magnet Number Magnet name Inner Outer X-section 39th HCMB__A001 1000018 0.20 0.80 2 40th HCMB__A001 1000019 0.20 0.80 2 41st HCMB__A001 2000014 0.10 0.70 2 42nd HCMB__A001 1000021 0.20 0.80 2 43rd HCMB__A001 3000011 0.20 0.85 1 44th HCMB__A001 3000012 0.20 0.80 2 45th HCMB__A001 3000013 0.20 0.80 2

Firm 1 Collared coil 1.2 Firm 2 Firm 3 1.0 ) m

m 0.8 (

outer nominal (0.8 mm) s s e

n 0.6 k c i h t

m 0.4 i h s 0.2 inner nominal (0.2 mm) 0.0 0 5 10 15 20 25 30 35 40 45 50

Magnet number LHC-MMS Fig. 2: Thickness of the polar shims used in the collared coils

2. Magnetic length and transfer function

 Magnetic length of collared coils 39th to 45th are within targets (see Fig. 3). No difference between old and new cross-section is observable, as expected. A small systematic difference of about 10 units

2 between Noell and Ansaldo values is observable in the last 25 coils, Alstom values being placed in between. Collared coil 14.48 Firm 1 Firm 2 upper limit for single magnet (3 sigma) 14.47 Firm 3 ) m (

14.46 h t g n e

l average

14.45 c i t e n

g 14.44 a M 14.43 lower limit for single magnet (3 sigma) 14.42 0 5 10 15 20 25 30 35 40 45 50

Magnet number LHC-MMS

Fig. 3: Magnetic length of the measured collared coils

 Main field in the central part of Ansaldo 14 (collared coils 41 st in Fig. 4) is 4 sigma lower than the average. This is partly due to the non-nominal shims (0.1 mm less on both layers), as it happened in Ansaldo 12 and 13 (36th and 37th in the figure, see also report of May-June 2002). When this effect is corrected, collared coils 41st fall in the lower part of the 3 sigma range (see Fig. 5). A corrective action will be taken: ferromagnetic laminations will be added to increase the magnetic length.

3  Difference between average main field in old and new cross-section is small (less than 5 units), as expected from simulations (see Fig. 5).  Noell coils have a main field of about 15 units higher than Alstom or Ansaldo (see Fig. 5). This systematic difference is half of the allowed range at three sigma (30 units).  All produced collared coils fit within the 3 sigma limit when data are reduced to nominal shims (see Fig. 5).

Collared coil 0.598 Firm 1 Firm 2

) Firm 3 A k

/ 0.597 upper limit for single magnet (3 sigma) T (

n o

i average t c

n 0.596 u f

r e f s

n 0.595 lower limit for single magnet (3 sigma) a r T

0.594 0 5 10 15 20 25 30 35 40 45 50

Magnet number LHC-MMS

Fig. 4: Average main field in the straight part of the measured collared coils

Data reduced to Collared coil 0.598 Firm 1 nominal shims Firm 2

) Firm 3 A k

/ 0.597

T upper limit for single magnet (3 sigma) (

n o i t c

n 0.596 u f

r systematic X-section 1 e f

s systematic X-section 2 n 0.595 a r lower limit for single magnet (3 sigma) T

0.594 0 5 10 15 20 25 30 35 40 45 50

Magnet number LHC-MMS

Fig. 5: Average main field in the straight part of the measured collared coils. Data reduced to nominal shims

 Integrated transfer function of magnets 39th to 45th is within the 3 sigma budget of the allowed random per arc (see Fig. 6).  When data are reduced at nominal shims (see Fig. 7), one finds some systematic difference (around 20 units) between Noell and Ansaldo-Alstom. This is well within the total width of the band allowed by beam dynamics (at three sigma) in the hypothesis of a complete mixing of the manufacturers.

4 Collared coil 8.66 Firm 1

) Firm 2 A

k Firm 3 / 8.64 m upper limit for single magnet (3 sigma) T (

. t

c 8.62 average n u f

. f

s 8.60 n a r t

. lower limit for single magnet (3 sigma) r

g 8.58 e t n I 8.56 0 5 10 15 20 25 30 35 40 45 50

Magnet number LHC-MMS Fig. 6: Integrated transfer function in the measured collared coils.

Data reduced to Collared coil 8.66 Firm 1 nominal shims )

A Firm 2 k / Firm 3 m 8.64 T (

upper limit for single magnet (3 sigma) n o i t

c 8.62 n u f

r

e systematic X-section 1 f 8.60 s systematic X-section 2 n a r

t lower limit for single magnet (3 sigma)

.

r 8.58 g e t n I 8.56 0 5 10 15 20 25 30 35 40 45 50

Magnet number LHC-MMS

Fig. 7: Integrated transfer function in the measured collared coils. Data reduced at nominal shims.

3. Estimated coil waviness

 Coil waviness estimated from the variation of the multipole along the axis is still anomalous for Alstom 18 and 19 (see Fig. 8, collared coils 39th, 40th). Corrective actions have been taken and Alstom 21 (see Fig. 8, collared coil 42th) is within the 30 micron limit, but still considerably higher than previous Noell and Ansaldo, that were around 15 microns.  Some small deterioration of coil waviness has been observed in Noell 12 and 13 (see Fig. 8, collared coils 44th and 45th).

5 Collared coil 0.090 Firm 1 Firm 2 Firm 3 ) m m (

0.060 s s e n i v a w

l i 30 micron target o

c 0.030

f o

0.000 0 5 10 15 20 25 30 35 40 45 50

Fig. 8: Estimated coil waviness in theMagnet straight part number of the measured collared coils. LHC-MMS

4. Summary of systematics

 Best estimates of skew and even normal systematics are given in Fig. 9, with an error at 95% confidence limit (two sigma). All the multipoles are within specifications. Details are given in Sections 5 and 6. Collared coil - 3 Specifications systematic vs targets Measured 2

1

0 0 s s e e r r u u t t

-1 1 2 1 2 r r e e e e e e r r r r p p u u u u a a t t t t

r r r r h h e t e e e -2 a2 a3 a4 t 6 p p o p p o a a b a a b

2 2 2 4 4 4 b b b b b -3 b Fig. 9: Best estimate for systematic skew multipoles and even normal multipoles (markers) versus beam dynamics limits (red line). An error of two sigma (95% confidence limit) is associated to the best estimates of systematics.  Best estimates for systematic odd multipoles are shown in Fig. 10. In the left part, raw data are plotted. This gives the actual situation for the manufactured collared coils: b3 and b5 are larger than the upper specifications of 3 and 1 units respectively.  In the left part of Fig. 10, data are reduced to nominal shims and separated according the two cross- sections (33 collared coils have cross-section 1, 12 have cross-section 2). The change of cross- section under-corrected b3 and b5, and overcorrected b7. Errors associated to the best estimate for the systematic are still large for b3 and b5, but it is likely that odd multipoles are outside the specification. Details are given in Section 7.

2 2

0 9 0 9

-2 b5 b7 -2 b5 b7

-4 -4 Specifications Specifications -6 -6 Measured X-s 1 Measured Measured X-s 2 -8 -8

Data reduced -10 b3 -10 b3 to nominal shims Fig. 10: Best estimate for systematic odd normal multipoles (markers) versus beam dynamics limits (red line). An error of two sigma (95% confidence limit) is associated to the best estimates of systematics. Raw data (left) and data reduced to nominal shims and separated according to different cross-sections (right). The cross-section change has considerably improved the foreseen machine performaces: a LHC made with dipoles with cross-section 1 would have had severe limitations, due to b3 and b5:  At high field a systematic b3 of 7 units would have caused major problems in the chromatic correction (maximum specification is 3 units).  At injection a systematic b5 of 2.1 units would have given major problems in dynamics aperture (maximum specification is 1.1 units). On the other hand, a machine made with dipoles with cross-section 2 could have minor performance limitations due to b5 and b7:  At injection a systematic b5 of 1.3 units could reduce dynamic aperture (maximum specification is 1.1 units).  At injection a systematic b7 of 0.35 units could reduce dynamic aperture (maximum specification is

0.1 units). Simulations summarized in LHC project report 501 show that a systematic b7 of 0.22 units may have a very limited impact on dynamic aperture at injection (reduction of 0.4-0.8 sigma), whilst a

systematic b7 of 0.4 units reduces the dynamic aperture of 2 sigma.  At high field a systematic b3 of 3.5 units is outside the specification of 0.5 units, but it is within the hard limit of 4.2 units given by the strength of chromatic correctors. Notwithstanding this relevant improvement, we are still far from optimal values. The best estimates for the correction needed on the cold mass is the following (for the collared coil, values should be multiplied by a factor 1.2):  b3 correction: -3.5 units  b5 correction: -0.5 units  b7 correction: -0.35 units

7 At the moments two options are considered: a further correction of the cross-section or an increase of the mid-plane insulation. The action on b5 and b7 is critical since the needed correction is 1 to 1.5 times the natural sigma of the multipoles, given by manufacturing tolerances. At the same time, actions are being taken to understand the drift in b3 (see section 7.1) and the unexpected jump of b7 (see section 7.3).

5. Systematic skew multipoles

 Systematic skew multipoles a2 a3 and a4 are within beam dynamics limits (see Figs. 11-13). We have a large margin for the a3, whilst beam dynamics limits are tighter for a2 and a4.

Collared coil 6 Firm 1 Firm 2 Firm 3

) 4 s t i n u

( 2 upper limit for systematic

t r

a systematic p

0 t h g

i lower limit for systematic a

r -2 t s

2

a -4

-6 0 5 10 15 20 25 30 35 40 45 50

Magnet number LHC-MMS & MTA

Fig. 11: Average a2 in the straight part of the collared coils (dots), best estimate for systematic in each aperture (solid lines), and beam dynamics limits for the systematic (red line) based on correlations with 10 cryodipoles. Firm 1 Collared coil 3 Firm 2 Firm 3 upper limit for systematic 2 ) s t i n u

( 1

t r a p

0 t

h systematic g i a

r -1 t s

3 lower limit for systematic a -2

-3 0 5 10 15 20 25 30 35 40 45 50

Magnet number Collared coilLHC-MMS & MTA 1.0 Firm 1 Fig. 12: Average a3 in the straight part of theFirm collared 2 coils (dots), best estimate for systematic in each aperture (solid lines), and beam dynamics limits for the systematic (red line) based on correlations with 10 cryodipoles. Firm 3 ) s t i 0.5 n u (

t upper limit for systematic r a p

0.0 systematic t h g i lower limit for systematic a r t s -0.5 4 a

-1.0 8 0 5 10 15 20 25 30 35 40 45 50

Magnet number LHC-MMS & MTA Fig. 13: Average a4 in the straight part of the collared coils (dots), best estimate for systematic in each aperture (solid lines), and beam dynamics limits for the systematic (red line) based on correlations with 10 cryodipoles.

6. Systematic even multipoles

For each multipole subject to beam dynamics specifications, we present two separated plots for the systematic per aperture, and a plot of the systematic per beam, i.e. the average of both apertures (that should be zero due to two-in-one symmetry).

6.1 Normal quadrupole  The systematic per aperture is within specifications in both apertures (see Figs. 14 and 15).  Noell 12 and 13 (collared coils 43rd and 44th) have rather large values of normal quadrupoles. The problem is under investigation. Noell 14 collared coil features better values of the normal quadrupole (see Figs. 14 and 15, collared coil 45th).

Collared coil 3 Firm 1 Firm 2 ) s t i 2 Firm 3 n u (

1

1 .

p upper limit for systematic a

t

r 0 a systematic p

t

h -1 g i a r t s

-2

2 lower limit for systematic b -3 0 5 10 15 20 25 30 35 40 45 50

Magnet number LHC-MMS & MTA

9 Fig. 14: Average b2 in the straight part of the aperture 1 collared coils (dots), best estimate for systematic per aperture (black line), and beam dynamics limits for the systematic (red line) based on correlations with 10 cryodipoles.

Firm 1 Collared coil 3 Firm 2

) Firm 3 upper limit for systematic s t i 2 n u (

2

1 .

p systematic a

t

r 0 a p

t lower limit for systematic h

g -1 i a r t s -2 2 b -3 0 5 10 15 20 25 30 35 40 45 50

Magnet number LHC-MMS & MTA

Fig. 15: Average b2 in the straight part of the aperture 2 collared coils (dots), best estimate for systematic per aperture (black line) and beam dynamics limits for the systematic (red line) based on correlations with 10 cryodipoles.

 The systematic per beam normal quadrupole is within specifications (see Fig. 16).

Firm 1 Collared coil 3 Firm 2 Firm 3

) 2 s t i n u

( 1

t upper limit for systematic over two apertures r a p

0 t systematic over two apertures h g i

a lower limit for systematic over two apertures

r -1 t s

2 b -2

-3 0 5 10 15 20 25 30 35 40 45 50

Magnet number LHC-MMS & MTA

Fig. 16: Average b2 in the straight part of collared coils (dots), best estimate for systematic per beam (black line) and beam dynamics limits for the systematic (red line) based on correlations with 10 cryodipoles. Collared coil 1.0 Firm 1 6.2 Normal octupole Firm 2 )

 The systematics per aperture is within specifications in both apertures (see Figs. 17 and 18). t

i Firm 3

 The systematicn per beam is also within specifications (see Fig. 19). u

( 0.5

1

. upper limit for systematic p a

t systematic r 0.0 a p

t h g i a

r -0.5 t

s lower limit for systematic

4 b -1.0 10 0 5 10 15 20 25 30 35 40 45 50

Magnet number LHC-MMS & MTA Fig. 17: Average b4 in the straight part of the aperture 1 collared coils (dots), best estimate for systematic per aperture (black line), and beam dynamics limits for the systematic (red line) based on correlations with 10 cryodipoles.

Firm 1 Collared coil 1.0 Firm 2

) Firm 3 s t i upper limit for systematic n u

( 0.5

2

.

p systematic a

t

r 0.0 a p

t

h lower limit for systematic g i a

r -0.5 t s

4 b -1.0 0 5 10 15 20 25 30 35 40 45 50

Magnet number LHC-MMS & MTA

Fig. 18: Average b4 in the straight part of the aperture 2 collared coils (dots), best estimate for systematic per aperture (black line) and beam dynamics limits for the systematic (red line) based on correlations with 10 cryodipoles.

Firm 1 Collared coil 1.0 Firm 2 Firm 3 ) s t i 0.5 n u

( upper limit systematic over two apertures

t r

a systematic over two apertures p

0.0 t h g i a

r lower limit systematic over two apertures t s -0.5 4 b

-1.0 0 5 10 15 20 25 30 35 40 45 50

Magnet number LHC-MMS & MTA

11 Fig. 19: Average b4 in the straight part of collared coils (dots), best estimate for systematic per beam (black line) and beam dynamics limits for the systematic (red line) based on correlations with 10 cryodipoles.

7. Systematic odd multipoles

7.1 Normal sextupole

 Data from Noell collared coil with the new X-section confirm previous results from Alstom and Ansaldo (see Figs. 20 and 21).  The cross section correction shifted down the normal sextupole from around 2.2 units (excluding the data from collared coil 1 to 15 that experienced an upward trend) to –1.8 units, in agreement with simulations (-3.9 units).  Due to the positive trend, our estimate for systematic in X-section 2 is 0.7 units out of the limit (see fig. 21). The associated error is 0.6 units (95% confidence level, see Fig. 10).  The low systematic difference between firms observed in X-section 1 seems to be preserved in X- section 2; Alstom collared coils feature a lower b3 with respect to Noell and Ansaldo.  Cryodipoles with the new X-section should feature 3.5 units of b3 at high field; this is outside the specification but within the hard limit of 4.2 units given by chromaticity correctors.

Collared coil 15 Firm 1 Firm 2 Firm 3

) 10 s t i n u

( 5

t

r average a p

0 t upper limit for systematic h g i a

r -5 t s

3

b -10 lower limit for systematic

-15 0 5 10 15 Data20 reduced25 to30 35 Collared40 45coil 50 10 Firm 1 nominal shims Magnet number LHC-MMS & MTA Fig. 20: Average b3 in the straight part of the Firmcollared 2 coils (dots), best estimate for systematic (solid lines), and beam dynamics limits for the systematic (red line) based on correlations with 10 cryodipoles.Firm 3 )

s 5 t i

n systematic X-section 1 u (

t

r 0 a p

t upper limit for systematic

h systematic X-section 2 g

i -5

a aim of X-section r

t correction s

3 lower limit for systematic

b -10

-15 12 0 5 10 15 20 25 30 35 40 45 50

Magnet number LHC-MMS & MTA Fig. 21: Average b3 in the straight part of the collared coils (dots), best estimate for systematic (solid lines), and beam dynamics limits for the systematic (red line) based on correlations with 10 cryodipoles. Data reduced at nominal shims and separated according to X-section type. 7.2 Normal decapole

nd  Alstom 21 (collared coils 42 in Figs. 22 and 23) features a very high b5 (around 0.7 units more than previous collared coils 38th –40th).  Measured b5 in new X-section Noell collared coils is larger than expected: the average is close to Alstom values, whilst in the previous X-section it was much lower, being close to Ansaldo (see Fig. 23, collared coils 44th and 45th).  Best estimate for systematic b5 in new X-section is 0.25 units larger than the upper allowed limit. Part of this difference is due to a positive trend of 0.2 units experienced from collared coil 9 th (where the correction was defined) to 20th. The remaining 0.35 units are due to a measured effect of the correction (-1.0 units) lower than expected (-1.35 units). We still have a large error associated to the estimate of the new X-section systematic: 0.24 units(at 95% confidence level), i.e. nearly the width of the allowed range (see Fig. 10).

Collared coil 2.5 Firm 1 Firm 2 2.0 Firm 3 ) s t i 1.5 n

u average (

t 1.0 r a p

0.5 t

h upper limit for systematic g i 0.0 a r t s -0.5 5 b -1.0 lower limit for systematic -1.5 0 5 10 15 20 25 30 35 40 45 50

Magnet number LHC-MMS & MTA Fig. 22: Average b5 in the straight part of the collared coils (dots), best estimate for systematic (blue line), and beam dynamics limits for the systematic (red line) based on correlations with 10 cryodipoles.

Data reduced to Collared coil 2.5 Firm 1 nominal shims Firm 2 2.0 Firm 3 ) s t i 1.5 systematic X-section 1 n u (

t 1.0 r a

p systematic X-section 2 0.5 t h g i 0.0 a upper limit for systematic r

t aim of X-section s -0.5 correction 5 b -1.0 lower limit for systematic -1.5 13 0 5 10 15 20 25 30 35 40 45 50

Magnet number LHC-MMS & MTA Fig. 23: Average b5 in the straight part of the collared coils (dots), best estimate for systematic (blue line), and beam dynamics limits for the systematic (red line) based on correlations with 10 cryodipoles. Data are reduced to nominal shims and separated according to different cross-sections. 7.3 Normal 14-th pole th  Alstom 19 (collared coils 40 in Figs. 24 and 25) features a very high b7 (around 0.2 units more than previous Alstom collared coils 29th –39th).  Noel data confirm previous trends: new X-section collared coils have a systematic b7 of around 1.1 units, i.e. 0.3 units more than the upper limit. The associated error is small (0.04 units at 95% confidence level, see Fig. 10).  The b7 has been increased by the cross-section correction of about 0.48 units against a foreseen value of 0.18 units, i.e. 0.3 units more than expected. This feature is under investigation. In particular, the very low value of b7 observed in Ansaldo for the first X-section has been not found in the second one. Collared coil 1.5 Firm 1 Firm 2 Firm 3 ) s t i 1.0 n upper limit for systematic u (

t

r average a p

0.5 t h g

i lower limit for systematic a r t s 0.0 7 b

-0.5 0 5 10 15 20 25 30 35 40 45 50

Magnet number LHC-MMS & MTA

Fig. 24: Average b7 in the straight part of the collared coils (dots), best estimate for systematic (solid lines), and beam dynamics limits for the systematic (red line) based on correlations with 10 cryodipoles.

Data reduced to Collared coil 1.5 Firm 1 nominal shims Firm 2 systematic X-section 2 Firm 3 ) s

t aim of X-section i 1.0

n correction upper limit for systematic u (

t r

a systematic X-section 1 p

0.5 t h g

i lower limit for systematic a r t s 0.0 7 b

-0.5 0 5 10 15 20 25 30 35 40 45 50

Magnet number LHC-MMS & MTA Fig. 25: Average b7 in the straight part of the collared coils (dots), best estimate for systematic (solid lines), and beam dynamics limits for the systematic (red line) based on correlations with 10 cryodipoles. Data are reduced to nominal shims and separated according to different cross-sections.

14 8. Random multipoles  Random per manufacturer and global random (i.e., the standard deviation of the distribution of all magnets) are shown in Figs. 26 and 27.  Raw data (see Fig. 26) show an out of tolerance for b3 and b5. This is mainly due to the change of cross-section that shifted down these multipoles of 3 units and 1 unit respectively. The other parameters are within specifications, also in the hypothesis of a complete mixing.  When data are reduced to nominal shims and split according to the cross-section type, one observes a random b3 out of tolerance in the old X-section: this is due to the upward trend (see Section 7.1, Fig. 21). This is the only out of tolerance in the old X-section.

Collared coil - random (r.m.s) vs targets 3.5

3.0 Firm 1 Firm 2 2.5 Firm 3 units/10 units Specifications 2.0 Global 1.5

1.0

0.5

0.0 L B BdL b a b a b a b a b 1 2 2 3 3 4 4 5 5 7

LHC-MMS

Fig. 26: Random component in the measured collared coils

Collared coil - random (r.m.s) vs targets 3.5 Firm 1 - Xs 1 Firm 2 - Xs 1 3.0 Data reduced Firm 3 - Xs 1 2.5 to nominal shims units/10 units Specifications 2.0 Firm 1 - Xs 2 1.5 Firm 2 - Xs 2 1.0 Firm 3 - Xs 2 0.5

0.0 L B BdL b2 a2 b3 a3 b4 a4 b5 a5 b7 1.0

LHC-MMS Fig. 27: Random component in the measured collared coils. Data reduced to nominal shims and split according to different cross-sections.

 The statistics for the new cross-section is still very poor and therefore some features could change when more data will be available; with the present set of data we make the following preliminary considerations: o Collared coils with the new cross-section feature a very stable integrated main field BdL and odd multipoles (see Fig. 27).

15 o Firm 1 has a large random b2, mainly due to the very high value measured in Alstom 13 (collared coil 29th in Fig. 16). o Firm 1 and Firm 3 have a large random a2 (see Figs. 27 and 11).  All the random components are expected to decrease by around 20% in the cold mass, due to the main field increase. Therefore, out of tolerances in a2 and b2 are likely to disappear in the cold mass.

9. Holding point results

Table II: results of the holding point for the measured collared coils

Collared coil Data at Answer to Answer To Magnet name measure CERN MMS-MD manufact. Result Comments 39rd HCMB__A001 1000018 03/07/02 04/07/02 05/07/02 05/07/02 Ok Coil waviness almost normal Strong coil waviness in aperture 1 (50 microns) – b7 higher than in previous 40th HCMB__A001 1000019 17/07/02 18/07/02 18/07/02 24/07/02 Ok-w ones Same as 2000013 and 12: more laminations will be added in the cold 41th HCMB__A001 2000014 24/07/02 24/07/02 24/07/02 25/07/02 Corr. Act. mass to recover good TF 42th HCMB__A001 1000021 24/07/02 24/07/02 24/07/02 25/07/02 OK Rather large value of b2 in both 43th HCMB__A001 3000011 19/08/02 20/08/02 21/08/02 22/08/02 Ok-w apertures Re-collared with 0.1 mm more shim in the outer layer – Effect on multipoles different from what expected in c1, b5, 36th HCMB__A001 2000013 08/08/02 08/08/02 08/08/02 12/08/02 HOLD b7 Rather large value of b2 in both 44th HCMB__A001 3000012 20/08/02 22/08/02 24/08/02 26/08/02 Ok-w apertures 45th HCMB__A001 3000013 23/08/02 23/08/02 24/08/02 26/08/02 Ok

 A corrective action will be taken on Ansaldo 14. It has been agreed to add magnetic laminations to recover an integrated main field within tolerances.  Ansaldo 2000013 that has been collared and measured in June 2002 has been de-collared for electric problems and a thicker shim on the external layer has been inserted (0.1 mm more). The effect on field quality is different from what expected from simulations (see Table III), especially for main field and b5. The collared coil has been not approved and investigations are in progress. The following steps will be taken: o All collared coils that will be de-collared for any reason, in any firm, will be measured before and after the de-collaring to have more statistics. o Components and assembly procedures of Ansaldo 2000013 will be analysed

Table III: Effect of a shim change of 0.1 mm on the outer layer: model, measurements on HCMB__A001-02000013, and discrepancy with respect to model

C1 b3 b5 b7 Model 4.0 1.6 -0.08 -0.02 02000013 Aperture 1 7.4 1.0 0.46 -0.11 02000013 Aperture 2 8.7 1.2 0.42 -0.13 02000013 Average 8.0 1.1 0.44 -0.12 02000013 Av.-model 4.0 -0.5 0.52 -0.10

10. Acknowledgements

Magnetic measurements have been taken through personnel and instrumentation of the LHC-MMS-IF section; in particular, G. Brun, G. Busetta, R. Camus, P. Galbraith, A. Musso, and G. Peiro. M. Zehner and R. Moresi measured collared coils at Noell. We wish to acknowledge J. Billan, B. Bellesia, L. Bottura, A. Devred, S. Pauletta, V. Remondino, L. Rossi, S. Sanfilippo, W. Scandale, E. Wildner for comments, valuable help and discussions.

16 Appendix A

In Figs. 2 to 25, collared coils are identified by a progressive number. The link between this number and the official name is given in the following table.

Table III: relation between magnet numbers used in Figs. 2-25 and official names

st 1st 1000001 21st 1000010 41 2000014 nd 2nd 1000002 22nd 1000011 42 1000021 rd 3rd 2000001 23rd 1000012 43 3000011 th 4th 3000001 24th 3000007 44 3000012 th 5th 1000003 25th 3000008 45 3000013 6th 3000002 26th 2000008 7th 2000003 27th 2000007 8th 1000004 28th 3000009 9th 1000005 29th 1000013 10th 3000003 30th 2000006 11th 2000002 31st 1000014 12th 1000006 32nd 1000015 13th 3000004 33rd 2000010 14th 2000005 34th 2000009 15th 1000007 35th 1000016 16th 1000008 36th 2000013 17th 3000005 37th 2000012 18th 3000006 38th 1000017 19th 1000009 39th 1000018 th 20th 2000004 40 1000019

Appendix B. Control chart for magnetic length and main field

Control limits for the magnetic length (see Fig. 3) are put at 3 times the specified sigma from the measured average. No target is assumed for the average magnetic length. The same approach is followed for the control limits of the main field in the straight part (see Figs. 4-7): they are put at 3 times the specified sigma from the measured average, and no target is assumed for the average main field. We recall that beam dynamics specifications are given in terms of the sigma of the integrated main field. Therefore, the only chart relevant for beam dynamics is in Fig. 5 and 7. We assume an equal share of the integrated main field spread (8 units) between magnetic length and main field (5 units each), in the hypothesis of a Gaussian sum of the spreads.

Appendix C. Control chart for the systematic

Best estimates for systematic shown in Figs. 9-25 are defined as the average of the averages of each manufacturer. This definition takes into account for the quotas of dipoles assigned to manufacturers (one third each).

Control limits for the systematic are given using the following formula for working out correlations between cc i h collared coil data bn and multipoles at injection bn or at high field bn h cc oh i cc oi bn =(bn /k) + bn bn =(bn /k) + bn cc cm where k=B1 / B1 = 1.18 is the multipole rescaling induced by the 18% increase of the main field due to the yoke as derived from magnetic measurements.

17 Appendix D. Remarks on the cross-section correction

We recall the aim of the cross-section correction:

B1=0 b3=3.3 b5=1.15 b7=0.15

These shifts are expected in the cold mass, whilst in the collared coil they are scaled by the factor k = 1.18 (see Appendix C). Therefore in the collared coil we aimed at

B1=0 b3=3.9 b5=1.36 b7=0.18.

The cross-section correction has been based on the best estimates of the systematics in the collared coil, on the correlations to measurements at 1.9 K, and on the beam dynamics acceptance ranges. Correlations also used data of prototypes (if homogeneous with pre-series) to increase statistics. Figs. 21 and 23 show that the cross-section correction carried out at collared coil 9 aimed at centring the allowed ranges for b3 and b5.

For the b7 the situation is different (see Fig. 25). This is due to some change in correlations, and to a change of the b7 acceptance ranges that has been carried out after the definition of the new X-section. We recall that due to the intense tracking campaign carried out in 2001, the tolerance for b7 at injection has been reduced from [-0.4,0.4] as presented in Villars, LHC Days, March 2001, to [-0.3,0.3] in June 2001 to the final value of [-0.3,0.1] in the LHC Project Report 501 published in August 2001. The cross-section correction computed in June 2001 has been based on the range [-0.3,0.3].

18

Recommended publications