1

Refractive index measurements of Ge

John H. Burnett*a, Simon G. Kaplana, Erik Stoverb, and Adam Phenisc aNational Institute of Standards and Technology, Physical Measurement Laboratory, 100 Bureau Dr., Gaithersburg, MD USA 20899; bM3 Measurement Solutions, Inc., 938 S. Andreasen Dr., Suite I, Escondido, CA USA 92029; cAMP Optics, 13308 Midland Rd., #1304, Poway, CA USA 92074

ABSTRACT

A program has been started at NIST to make high-accuracy measurements of the (IR) index properties of technologically important IR materials, in order to provide the IR optics community with updated values for the highest- quality materials now available. For this purpose, we designed and built a minimum-deviation-angle refractometry system enabling diffraction-limited index measurements for wavelengths from 0.12 µm to 14 µm. We discuss the apparatus and procedures that we use for IR measurements. First results are presented for germanium for the wavelength range from 2 µm to 14 µm, with standard uncertainties ranging from 2 × 10-5 near 2 µm to 8 × 10-5 near 14 µm. This is an improvement by about an order of magnitude of the uncertainty level for index data of germanium generally used for optic design. A Sellmeier formula fitting our data for this range is provided. An analysis of the uncertainty is presented in detail. These measurements are compared to previous measurements of Ge.

Keywords: index, index of refraction, , , refractometry, germanium, Ge, infrared

1. INTRODUCTION

Since germanium (Ge) first began to be developed for transistor and infrared (IR) optics applications in the late 1940s, the material quality has continually improved to meet increasingly stringent demands. For optics applications, this has driven the need for more accurate optical properties measurements, especially the index of refraction in the high IR transmission wavelength region (2 µm to 14 µm for dry air), to meet the requirements of optical design. Ge measurements by various methods in the 1950s reported absolute index measurement accuracies ranging from high 10-3 to a few parts in 10-4, with results between labs differing by as much as 1 percent.1,2,3,4 These materials had various doping and defect levels. The measurements by Salzberg and Villa3,4 in 1957 and 1958 for the range 2 µm to 16 µm have been used extensively for IR optics design for two decades since their publication, and to some extent even today.

By the mid-1970s, driven by the importance of Ge for IR applications, material quality and index measurement methods had improved to the point where reported index measurements for the range of 2 µm to 14 µm generally had stated uncertainty estimates in the mid to low 10-4 range.5,6 In particular, Icenogle et al.5 reported measurements for the refractive indexes of Ge over the range 2.5 µm to 12 µm and their variations with over the ranges 95 K to 298 K, with conservative estimates of the their standard uncertainties of ±6 × 10-4. These differed from previous measurements, e.g., those of Salzberg and Villa4, by as much as 5 × 10-3, when account is taken of the different measurement wavelengths and used. Icenogle et al.5 suggested that the difference may be due to different sample impurities. Edwin et al.6 reported measurements of a Ge sample at two laboratories over a subrange 8 µm to 14 µm by different deviation-angle methods, with a discrepancy of 3 × 10-4 between labs. These results differed from those of Icenogle et al. by ≤ 10 × 10-4 over the range of overlap (with the exception of longest wavelength index). In 1980, Li7 reviewed and tabulated all the measured index data on Ge and Si since 1949. Based on an analysis of all the data, he recommended index values for Ge for wavelength and temperature in the ranges 1.0 µm to 18 µm and 100 K to 550 K, in the form of an expression with the wavelength dispersion given by a single 1/λ2 term. The stated standard uncertainty for the index was ±2 × 10-3 over the wavelength and temperature range, consistent with the uncertainties claimed for the 1970s measurements discussed above. ______*[email protected]; phone 301-975-2679 2

Since the late 1970s, the results of Icenogle et al.5, and to some extent the previous results of Salzberg and Villa3,4, have been widely used in IR optics design, including incorporation in commercial optics design software. These values have been recommended in well-established index data reference books, such as the Handbook of Optical Constants of Solids8 and the Handbook of Infrared Optical Materials.9 With no substantial additions to these index data over the following 30 years, they remain the values IR optical designers must rely on, accepting a measurement uncertainty for each sample of at best (1 to 2) × 10-3 and sample-to-sample variation up to an order of magnitude higher.

There have been a number of different methods used to measure the index of bulk materials, including techniques based on refraction angle, interference in slabs, surface reflection, and ellipsometry. For the transparent region of IR materials, the methods reporting the highest accuracy are variations of the minimum-deviation-angle refractometry method, based on measuring the angle of refraction through a prism of the material set at the angle of minimum deviation, discussed below. The accuracy of this approached is limited by the diffraction of the beam through the prism. For a prism about 3 cm on a side, it is straightforward to calculate that the diffraction-limited index uncertainty is approximately 5 × 10-7 in the , which has been verified.10 This diffraction-limited uncertainty should scale with wavelength approximately two orders of magnitude to the mid 10-5 range in the mid-IR. This is an order of magnitude smaller than the uncertainties reported for Ge discussed above. Thus, with an optimized refractometer design, an order of magnitude improvement in index accuracies should be achievable. Besides the inherent value of this higher accuracy for more precise IR optical design, this level of accuracy would be of value to sort out index variations in samples of commercial material, feeding back to material fabricators to improve material quality.

Over the last several years a program has been established at NIST to pursue these goals for materials of importance to the IR optics community. In this paper we discuss the minimum-deviation-angle refractometry system developed at NIST and the procedures designed to achieve diffraction-limited index measurements in the IR out to 14 µm. Achieving this performance requires substantially reducing numerous sources of uncertainty compared to previous measurements, in some cases by an order of magnitude or more, as will be discussed. We will present our index measurements of a sample of high-quality Ge for wavelengths from 2 µm to 14 µm at the temperature of 22.000 °C. We fit these data to a Sellmeier formula and compare the values to previous measurements.

2. MEASUREMENT

The well-established minimum-deviation-angle refractometry method is based on determining the angle δ (λ) that a collimated beam of known wavelength λ refracts through a prism of known apex angle α.11,12 If the prism is oriented so that the beam travels symmetrically through the prism with respect to the apex angle, then the refraction angle (deviation angle) is at a minimum and the refractive index of the prism material nmat(λ) relative to the index of the gas surrounding the prism ngas(λ) is given by the following formula:

(λ) = (λ) (λ) = sin + (λ) 2 sin( 2). (1)

𝑛𝑛𝑟𝑟𝑟𝑟𝑟𝑟 𝑛𝑛𝑚𝑚𝑚𝑚𝑚𝑚 ⁄𝑛𝑛𝑔𝑔𝑔𝑔𝑔𝑔 ��𝛼𝛼 𝛿𝛿 �⁄ �⁄ 𝛼𝛼⁄ A substantial benefit of measuring the deviation angle δ (λ) at the angle of minimum deviation is that the measurement is at an extremum condition, making it fairly insensitive to the alignment of the prism, a substantial source of error. Due to the conceptual simplicity of the index determination, not depending on any models other than Snell’s law, it is relatively secure from subtle sources of systematic errors. The accuracy of the determination of nmat(λ) simply depends on the accuracy of the of the determination of the parameters, λ, ngas(λ), α, and δ (λ). The uncertainties in α and δ (λ) depend in part on the geometric and surface properties of the prism, along with the prism index homogeneity and isotropy. We give some discussion on those aspects of our implementation of the minimum deviation angle technique relevant to the uncertainties of these various components.

2.1 NIST Minimum-Deviation-Angle Refractometer

A custom minimum deviation angle refractometer system used for high-accuracy index measurements in the visible and ultraviolet, was modified with sources, detectors, and-all reflective optics appropriate for operation in the IR. A schematic diagram of the system is shown in Figure 1. The narrow-band IR for the δ (λ) measurements was generated by filtering radiation from a broadband 1200 °C blackbody source by a grating monochromator and from the a 3.39 µm HeNe 3

laser. The grating orders were separated by the prism sample. The entire path of the radiation from the source to the detector was enclosed in sealed housings purged with dry N2 gas, to enable measurements in the low transmission regions of air with substantial H2O and CO2 content.

θ/2θ Stacked Prism Goniometers Temperature Entrance Slit Control Unit Purge Box Sample Exit Slit Prism

M5 M6 M3,M4 Monochromator Focusing Optics Collimating Optics 1 m Focal Length

M8 Purge Housing Entrance Slit Exit Slit Gas Temperature Purge Box Controlled M9,M10 M2 Blackbody M7 HgCdTe Source Detector 1200 °C Chopper L-N2 Cooled M1

Figure 1. Schematic diagram of the top view of the NIST minimum-deviation-angle refractometer system with all reflective optics. Mirror M1 collects radiation from the blackbody source and images it, via M2, into a monochromator. The output of the monochromator is imaged into the higher refractometer entrance slit via a pair of parabolic mirrors M3 and M4. The radiation through the entrance slit is turned and angled down by plane mirror M5 and collimated by spherical mirror M6. The collimated beam is refracted by the prism and imaged to the exit slit by spherical mirror M7 and plane mirror M8. This radiation is re-imaged into the HgCdTe detector by a pair of parabolic mirrors M9 and M10.

The radiation from the blackbody source was chopped and imaged into a McPherson 2051 monochromator with a focal length of 1 m. A grating with 100 grooves per millimeter, blazed near 9 µm was used. Appropriate grating orders were chosen for each wavelength desired to ensure the grating was used near the blaze angle for high efficiency. The wavelength output of the monochromator was calibrated with the first six orders of HeNe 633 nm laser and the first four orders of a 3.39 µm HeNe laser. The small wavelength non-linearity of the monochromator was removed by the generated calibration curve. The wavelength was accurate to within ±0.2 nm for the full 2 µm to 14 µm range. The monochromator slit widths were chosen for each wavelength to be as large as possible to maximize the signal relative to the background, while ensuring that the bandwidth contributed less than 10% of the overall uncertainty budget for the index.

The narrow-band radiation from the monochromator was collected and focused into the entrance slit of the refractometer with a focal length of 0.5 m, by a pair of parabolic mirrors. The entrance and exit slit widths of the refractometer were chosen as above to maximize signal without degrading the index uncertainty above the diffraction limit. A folding mirror and spherical mirror collimated the radiation to the prism, and the refracted radiation was focused onto the exit slit by spherical mirror and folding mirror. A ray trace analysis of system showed that aberrations from using spherical mirrors for the slit height of 4 mm, was less than the diffraction spread. The optics of the refractometer had a f-number of 10.

The prism sample holder was mounted on a pair of stacked computer-controlled goniometers. The bottom goniometer, which holds the detector arm, had an encoder with absolute accuracy of ±0.2 arcseconds. This limits the accuracy of the deviation angle δ (λ). The top goniometer was controlled by the drive program to ensure that after initial minimum deviation alignment, the minimum deviation condition of the prism was maintained. The prism tilts were aligned by an autocollimator, and the minimum deviation alignment was determined by multiple scans at various prism angles at 3.39 µm.

Accurate temperature control for the prism at 22.000 °C was achieved by a feedback control loop in the following fashion. The prism was mounted between two hollow copper plates. Two water baths with water flow to the copper plates were maintained to 0.01 °C at temperatures less than one °C above and below the set point temperature of 22.000 °C. Water from the two baths was mixed a few inches before entering the copper plates. The mixing of the two fluids was 4

accomplished by a mixing value, with the mixing ratio determined by the error in the sample temperature from the set point, in a PID control loop. The sample temperature was determined by four platinum resistance thermometers mounted on the copper plates. These were calibrated by a standards-grade platinum resistance thermometer calibrated against the triple point of water and the melting point of gallium, according to the ITS-90 international temperature standards protocol.13 The standard uncertainty of the calibration of each of the thermometers used was 0.002 °C. During temperature control operation, all four thermometers were monitored and they were always within 0.002 °C of each other. The temperature control loop kept all four thermometers oscillating within the range 22.000 ±0.005 °C. The temperature of the nitrogen gas was controlled by a second, similar PID control loop. The gas temperature within the sample housing was monitored by two calibrated platinum resistance thermometers and was controlled within the range 22.000 ±0.020 °C.

The radiation exiting the refractometer was detected by a liquid-nitrogen-cooled HgCdTe detector using phase sensitive

(lock-in) detection. The deviation angle for each wavelength δ (λ) was determined by introducing each wavelength in turn into the entrance slit of the refractometer and scanning the detector arm, with the prism angle scanning at half the angular rate to ensure that the minimum-deviation condition was maintained. For each wavelength, the encoder-determined angle of the center of the peak (obtained by fitting) relative to that of the un-deviated beam, determined the deviation angle δ (λ). Measurements were made on both sides of prism repeatedly to cancel effects of absolute angle drifts. The uncertainty of the center of each peak was estimated conservatively to be 10% of the FWHM (full-width at half-peak-maximum). This uncertainty dominated the total estimated uncertainty for all wavelengths. The peak widths broadened as the wavelength increased, as expected from diffraction. Thus the total uncertainty increased as the wavelength increased, as discussed in Section 3.1.

2.2 Ge Sample

The Ge material measured was made into a prism with sides 39 mm long and 29 mm high, and an apex angle of approximately 15 degrees. Both prism transmitting surfaces had figure errors less than 5 nm RMS (root mean squared) (λ/120 RMS at 633 nm) over the 90% clear aperture. The apex angle was determined by use of stacked indexing tables and an autocollimator, making angle difference measurements on both sides of the prism with all combinations of index angles separated by 15 degree intervals. The standard deviation of all the measurements from the various indexing table angle combinations was 0.12 arcseconds. Including other sources of error, such as the surface figure errors, gave a standard uncertainty of the apex angle of 0.2 arcseconds.

The sample resistivity, determined by a 4-probe measurement, was 17 Ohm-cm, within the range about 10 Ohm-cm to 40 Ohm-cm generally targeted for optimal Ge transmission properties. The index homogeneity was measured on a 10 mm thick, 27 mm diameter Ge disk, taken from a location in the Ge ingot close to where the prism material was taken. The measurements were made at 3.39 µm, using a M3 Measurements Solutions MWAVE 339 IR phase-shifting interferometer, with a calibrated accuracy of 0.002 λ RMS wavefront error at 3.39 µm.14 Due to the very high temperature dependence of the index of 3.96 × 10-4 /°C,5 there was concern about temperature gradients in the sample giving rise to index inhomogeneities during the characterization. For the measurement, the sample was mounted between two large, copper blocks with thermally-connecting side plates and apertures. The result was 9.1 × 10-6 RMS, 72.8 × 10-6 PV (peak to valley). Due to possible residual temperature gradient effects, this represents an upper bound for the grown-in inhomogeneity.

We measured the transmittance through the center of the face of the prism with a 6 mm diameter aperture, using a Fourier- transform spectrometer and an integrating sphere to collect the transmitted light.15 We found that there was no change in transmittance structure using smaller apertures. The results are shown in Figure 2. Spectra show a transmittance of about 44% through the range just above 2 µm through 11 µm. A calculation using our measured index values predicts a value of within 2% of this, the specific value depending on assumptions on what happens to the multiple reflections in this highly refracting prism window. This spectrum is similar qualitatively and quantitatively to transmittance spectra of high-quality Ge in the literature.9 The sharp drop in transmittance to zero near 2 µm is due to the onset of electronic transitions at the semiconductor conduction band edge, and is the lower limit wavelength for use of Ge as thick transmitting windows and optics. To avoid the distorting effects of high absorption on the index measurements below 2 µm, 2.25 µm was used as our lowest measurement wavelength. The absorption structure at 12 µm and longer wavelengths is due to phonon absorption. The upper limit of the measurements was 14 µm. 5

1.0

0.9

0.8

0.7

0.6

0.5

0.4 Transmittance 0.3

0.2

0.1

0.0 2 4 6 8 10 12 14 16 18 20 Wavelength (µm)

Figure 2. Transmittance versus wavelength of the 15 degree Ge prism measured using a Fourier-transform spectrometer and an integrating sphere. The illumination was apertured to 6 mm diameter at the center of the prism, with the first face at 23 degrees to the illumination and the back face 8 degrees to the illumination.

2.3 Measurement Procedures

Deviation angle measurements were made at minimum deviation, transmitting through the prism from both directions by rotating the prism around and measuring the deviations on both sides of the undeviated beam. This required establishing minimum deviation settings on each side by multiple index scans as a function of prism position. On each side, deviation angle measurements were made at 17 wavelengths shown in Table 1, within the range 2.25 µm to 14 µm. The deviation angle at each wavelength was determined by the average of the deviation angle from both sides. This was repeated 5 times. During the measurements, the temperature of the prism and the N2 gas in the sample purge chamber were continuously monitored, but because they were controlled to 22.000 C, there was no need to make temperature corrections. The pressure of the N2 gas in the chamber was nearly the ambient pressure and was not controlled. It drifted with the ambient pressure through the measurement process. This pressure was monitored throughout the measurements, and it was corrected for in the analysis

The calibrated wavelengths from the monochromator were the wavelengths in N2 gas. The wavelengths shown in the first column of Table 1 are wavelengths corrected to vacuum wavelengths. The wavelengths in the second column of Table 1 are wavelengths corrected to air wavelengths at 22.000 °C and a pressure of 101.325 kPa (1 standard atmosphere). To 16,17 make the conversions, the well-established indices of air and N2 gas as functions of temperature and pressure were used.

3. RESULTS

3.1 Index of Refraction

The indices of refraction determined by the procedures discussed above are presented in Table 1. The indices are presented as relative indices in air (column 4) and absolute (vacuum) indices (column 3). The wavelengths are given as wavelengths in air (column 2) and as vacuum wavelengths (column 1). The last column gives, for each wavelength, our estimate of standard (1-sigma) uncertainty of the index. This was determined by summing in quadrature our estimates of the standard uncertainties of the various uncertainty components listed in Table 2. The total uncertainties range from near 2 × 10-5 at the shortest wavelengths, increasing to 8 × 10-5 at 14 µm. The magnitudes of these uncertainties are about what we would have expected on the basis of diffraction-dominated measurements, as discussed in the Introduction.

6

Table 1. Measured relative and absolute indices of refraction, and total standard uncertainties. λvac(µm)a λair(µm)b Indexvac Indexair σ (x10-5) 2.2500 2.2494 4.084061 4.082973 2.5 2.5000 2.4993 4.066938 4.065855 2.2 3.0000 2.9992 4.045788 4.044711 2.1 3.3922 3.3913 4.035707 4.034633 2.0 4.0000 3.9989 4.025809 4.024738 2.2 4.5000 4.4988 4.020564 4.019494 2.3 5.0000 4.9987 4.016859 4.015790 2.4 5.5000 5.4985 4.014126 4.013058 2.6 6.0000 5.9984 4.012056 4.010988 2.7 7.0000 6.9981 4.009179 4.008112 3.0 8.0000 7.9979 4.007305 4.006238 3.9 9.0000 8.9976 4.006008 4.004942 4.3 10.0000 9.9973 4.005056 4.003990 4.8 11.0000 10.9971 4.004362 4.003296 6.4 12.0000 11.9968 4.003801 4.002736 6.9 13.0000 12.9965 4.003375 4.002310 7.5 14.0000 13.9963 4.002958 4.001893 8.0

Table 2 gives our assessment of the principal sources of uncertainty in the index results in Table 1. Column 2 gives the magnitudes of the sources, and column 3 gives the resulting contribution to the index uncertainties. When the uncertainties vary with wavelength, the uncertainties are shown as a range corresponding to wavelengths from the shortest to the longest. The largest contribution to the uncertainty for all wavelengths, except the lowest, comes from the determination of the peak angular position. This increases with wavelength, as the peak widths increase due to diffraction. The other strongly wavelength-varying uncertainty is from the radiation wavelength uncertainty of 0.2 nm, independent of wavelength. This increases rapidly as the wavelength gets shorter and is the largest uncertainty at 2.25 µm.

Table 2. Sources of uncertainty and their impacts on the index uncertainty.

Sources Magnitude of uncertainty (1-σ) Index uncertainty (1-σ) Peak position (0.8 – 5) arcseconds [1/10th of 1.1 - 7.9 x 10-5 diffracted peak FWHM] Prism apex angle 0.2 arcseconds [including λ/120 1.2 x 10-5 (633 nm) RMS figure error] Index 0.91 x 10-5 RMS for 10 mm 0.91 x 10-5 inhomogeneity thickness at 3.39 µm Encoder angle 0.2 arcseconds 0.33 - 0.34 x 10-5 Prism temperature 0.005 °C 0.20 x 10-5 Wavelength 0.0002 µm 1.5 - 0.01 x 10-5 Alignment 0.05 arcseconds 0.08 x 10-5 -5 N2 gas temperature 0.02 °C 0.08 x 10 -5 N2 gas pressure 0.02 kPa 0.02 x 10

7

3.2 Sellmeier Fit and Residuals

The air index values (at 22.000 °C) at the air wavelengths listed in Table 1 were fit by commercial least squares non-linear fitting software to several types of dispersion formulas. It was found that the best fit resulted from a standard 3-term, 6- parameter Sellmeier formula shown in Equation (2). Adding more terms or more adjustable parameters does not significantly improve the fit.

λ λ λ 1 = + + (2) λ 2 λ 2 λ 2 𝐾𝐾1 𝐾𝐾2 𝐾𝐾3 2 2 2 2 1 2 3 The Sellmeier parameters are listed in𝑛𝑛 Table− 3, valid− 𝐿𝐿for 2.25− µ𝐿𝐿m ≤ λ ≤− 𝐿𝐿14 µm. The best fit places two divergent terms (poles) below the transmission region, in the electronic absorption band wavelengths, as expected. One pole at 1.18 µm is close to the band edge, and the other at 0.40 µm is rather lower. The pole at a wavelength longer than the measured wavelengths is at 27.4 µm, in the phonon absorption band region, also not unexpected. For the formula to be consistent with the measurements within the stated uncertainties, all digits of each of the parameters in Table 3 should be used.

Table 3. Sellmeier constants of Equation (2). Range of validity: 2.25 µm < λ < 14 µm. Sellmeier constants

K1 0.4886331

K2 14.5142535

K3 0.0091224 2 L1 1.393959 µm 2 L2 0.1626427 µm 2 L3 752.190 µm

The residuals of the fit (the difference between the measured values and formula values) are shown in Figure 3. For wavelengths below 10 µm, all residuals are below 1 × 10-5. This is consistent with lowest estimated uncertainty in this range, σ = 2 × 10-5. The largest residual in the total wavelength range is 2.3 × 10-5 at 13 µm. This is also consistent with estimated uncertainty of σ = 7.5 × 10-5 at this wavelength. Though good residual fits have little to say about systematic errors (which could be fit very well if they are smoothly varying or constant), the residuals with magnitudes at least a factor of 2 below all the estimate uncertainties, indicates that the random errors are reasonably assessed, and maybe overestimated. 8

3

2 )

5

− 1 x10 (

0 Residual Residual -1

-2 0 2 4 6 8 10 12 14 Wavelength (µm)

Figure 3. Plot of residuals of the fit to the data of the Sellmeier formula (2) using the parameters in Table 3.

4. COMPARISON WITH OTHER MEASUREMENTS

Comparison of the Sellmeier equation values of this work with the most extensively used data, of Salzberg et al.3,4 and of Icenogle et al.,5, is shown in Figure 4. The plot gives the difference between these data and our Sellmeier formula values at each of the previous author’s measured wavelengths. The error bars correspond to the reported standard uncertainties for each of the measurements. The error bars for the results of this work are barely visible at the scale of the plot. The older results of Salzberg, et al., are all below those of this work, by (7 to 30) × 10-4, well outside their estimated uncertainty of ±3 × 10-4. The Ge material they measured may well have had a substantial difference in quality, and this could account for the difference.

40 n(Salzberg) - n(Burnett) 30 n(Icenogle) - n(Burnett)

20 ) 10 4 −

0 ( x10 n

∆ -10

-20

-30

0 2 4 6 8 10 12 14 Wavlength (µm)

Figure 4. Plot of difference between the measurements of Salzberg et al.4 and of Icenogle et al.,5 and values of the Sellmeier formula of this work. The error bars represent standard uncertainties. 9

The newer data of Icenogle et al., is closer to values of this work, with a trend of lower-than-our-values to higher- than-our-values crossing at about 2.5 µm. Their error bars include, or nearly include, our values for all wavelengths except for the value at 12.36 µm.

Figure 5 gives a similar comparison of the limited set of measurements, 8 µm to 14 µm, of Edwin et al.6 to the results of this work. In this case our values are well within the error bars of Edwin, et al. In fact, the Edwin, et al. values are nearly within our error bars, which are 3 to 8 times smaller.

5

4

3

2 )

4 1 − 0

( x10

n -1 ∆ -2

-3

-4 n(Edwin) - n(Burnett)

-5 7 8 9 10 11 12 13 14 15 Wavelength (µm)

Figure 5. Plot of difference between the measurements of Edwin, et al.6, and values of the Sellmeier formula of this work. The error bars represent standard uncertainties. The heavier-set error bars correspond to the results of this work.

4. CONCLUSIONS

The IR optics community has generally relied on index data for Ge measured 35 or more years ago, for optical design with Ge. It seems that these values should be revisited, especially since improvements in material fabrication may have produced higher quality materials, with potentially different index properties. We have demonstrated that we have developed a refractometry facility and procedures that are at least at the state-of-the art for index measurements in the near- to mid-IR. We have used this facility to measure the index of recently produced high-quality Ge. Our analysis of our measurement uncertainties gives values smaller than those of most previous reported measurements of Ge by about an order of magnitude or more, and our measurements approach diffraction-limited accuracy. This lowered uncertainty results from improvements of a number of the most important sources of uncertainty, including temperature, wavelength, and angle measurements. Our measurements cover the wavelength range 2.25 to 14 µm at 22.000 °C. We provide a Sellmeier formula for this range, with a standard uncertainty ranging from 2.0 × 10-5 near the short wavelength end and 8 × 10-5 at the long wavelength end.

Comparing our new Ge index data to the widely used data starting from the late 1950s, there seems to be convergence of the data over time and more consistency of the uncertainties with the differences. The remaining differences may be genuine, resulting from actual material differences, such as doping and impurity levels. A verified improvement in accuracy should be of value to assess the material differences for the new generation of Ge and other materials. We suspect that the short wavelength band edges would be particularly sensitive to defect and impurity differences, and that this would be reflected in the index behavior in this region. With the increased accuracy of our new facility, we are now measuring a number of Ge samples to explore these potential index differences. 10

ACKNOWLEDGEMENTS

We are pleased to acknowledge the help of Gary Wiese, of Lockheed Corporation, in helping set up our interaction with the IR optics community, helping develop the sample specifications, and getting suppliers involved in the program and getting them to provide samples.

Commercial equipment, instruments, materials, or software are identified to specify adequately the experimental procedure. Such identification does not imply recommendation or endorsement by the National Institute of Standards and Technology or does it imply that the items identified are necessarily the best available for the purpose.

REFERENCES

[1] Briggs, H. B., “Optical Effects in Bulk Silicon and Germanium,” Phys. Rev. 77, 287 (1950). [2] Rank, D. H., Bennett, H. E., and Cronemeyer, D. C., “Index of Refraction of Germanium Measured by an Interference Method,” J. Opt. Soc. Am. 44(1), 13-16 (1954). [3] Salzberg, C. D. and Villa, J. J., “Infrared Refractive Indexes of Silicon Germanium and Modified Selenium Glass,” J. Opt. Soc. Am. 47(3), 244-246 (1957). [4] Salzberg, C. D. and Villa, J. J., “Index of Refraction of Germanium,” J. Opt. Soc. Am. 48, 579 (1958). [5] Icenogle, H. W., Platt, B. C., and Wolfe, W. L., “Refractive indexes and temperature coefficients of germanium and silicon,” Appl. Opt. 15(10), 2348-2351 (1976). [6] Edwin, R. P., Dudermel, M. T., and Lamare, M., “Refractive index measurements of a germanium sample,” Appl. Opt. 17(7), 1066-1068 (1978). [7] Li, H. H., “Refractive Index of Silicon and Germanium and Its Wavelength and Temperature Derivatives,” J. Phys. Chem. Ref. Data. 9(3), 561-658 (1980). [8] Potter, R., “Germanium (Ge)”, in Handbook of Optical Constants of Solids, Vol I., Ed. by E. Palik, Academic Press, New York, 465-478 (1985). [9] Browder, J. S, Ballard, S. S., and Klocek, P., “Germanium, Ge”, in Handbook of Infrared Optical Materials, Ed. by P. Klocek, Marcel Dekker, New York, 262-265 (1991). [10] Daimon, M. and Masumura, A., “High-accuracy measurements of the refractive index and its temperature coefficient of calcium fluoride in a wide wavelength range from 138 to 2326 nm,” Appl. Opt. 41(25), 5275-5281 (2001). [11] Born, M. and Wolf, E, [Principles of Optics, 7th edition], Cambridge University Press, Cambridge, UK & New York, 190-193 (1999). [12] Tentori, D. and Lerma, J. R., “Refractometry by minimum deviation: accuracy analysis,” Opt. Eng. 29, 160-168 (1990). [13] Mangum, B. W. and Furukawa, G. T., “Guidelines for Realizing the International Temperature Scale of 1990 (ITS- 90),” NIST Technical Note 1265, U.S. Government Printing Office, 28-70 (1990). [14] Schwider, J, Burow, R., Elssner, K. E., Spolaczyk, R, and Grzanna, J., “Homogeneity testing by phase shifting interferometry,” Appl. Opt. 24(18), 3059-3061 (1985). [15] Hanssen, L. M., “Integrating-sphere system and method for absolute measurement of transmittance, reflectance, and absorptance of specular samples,” Appl. Opt. 40, 3196-3204 (2001). [16] Birch, K. P. and Downs, M. J., “An Updated Edlen Equation for the Refractive Index of Air,” Metrologia 30 155- 162 (1993). [17] Peck, E. R. and Khanna, B. N., “Dispersion of Nitrogen,” J. Opt. Soc. Am. 56(8), 1059-1063 (1966).