Capillary Viscometer

Capillary Viscometer

J Clin Pathol: first published as 10.1136/jcp.41.11.1213 on 1 November 1988. Downloaded from J Clin Pathol 1988;41:1213-1216 Automated measurement of plasma viscosity by capillary viscometer B M COOKE, J STUART Department ofHaematology, Medical School, University ofBirmingham SUMMARY Plasma viscosity has several advantages over the erythrocyte sedimentation rate as a measurement ofan acute phase response of more than 24 hours' duration. A new capillary viscometer (Coulter Viscometer II), which gives an automated measurement ofplasma viscosity, was compared with the selected manual method (Harkness viscometer) of the International Committee for Standardization in Haematology. Automated measurement ofplasma viscosity at 25°C showed close correlation (r = 0-979, p < 0-002) with the selected method for 160 specimens of plasma. Satisfactory precision both within batch and between batch (coefficients ofvariation of 1-7% or less) was obtained at viscosity values up to 5 7 mPa-s. There was no detectable carry over between samples and viscosity values were corrected adequately for ambient temperature for the range 15-32°C. Careful daily cleaning was required to prevent accumulation of protein within the automatic sampling valve of the instrument. Automated measurement of plasma viscosity is an attractive alternative to measurement of the erythrocyte sedimentation rate. Tissue damage induces a local inflammatory response performance with that of the selected method of the that in turn stimulates, through interleukin 1 and other International Committee. mediators, an increase in hepatic synthesis of several by copyright. acute phase proteins.' Quantitative assay of these Patients and methods acute phase proteins in serum is useful for diagnosing and monitoring an inflammatory response.2 In the Venous blood, anticoagulated with the dipotassium early stages ofinflammation (< 24 hours), assay ofthe salt of ethylenediamine tetra-acetic acid (1 -5-2 0 mg/ fast responding C reactive protein is the test ofchoice, ml) was collected from healthy controls, hospital whereas the more complex hyperproteinaemia of the patients with an acute phase or chronic phase later stages of inflammation (> 24 hours) should be inflammatory response, and patients with measured by tests such as the erythrocyte sedi- paraproteinaemia. Blood samples were centrifuged at mentation rate (ESR) and plasma viscosity, which are 1500 g for 10 minutes within five hours of vene- sensitive to the combined effect ofmultiple proteins.3 puncture, and the supernatant plasma was stored at http://jcp.bmj.com/ Measurement of plasma viscosity has several room temperature in a capped plastic tube until advantages over measurement of ESR.2' These assayed (within two hours). include independence from the effects of anaemia, Manual viscosity measurements were made accord- polycythaemia, and erythrocyte deformability; a ing to the recommendations of the International reference range that is independent of sex and less Committee for Standardization in Haematology' dependent on age; a shorter assay time; the facility to using a Harkness capillary viscometer (Coulter store plasma for subsequent assay; and the potential Electronics Ltd, Luton) at 25°C. Automated on September 28, 2021 by guest. Protected for calibration, quality control, and automation. The measurements were made by the new Viscometer II selected method of the International Committee for (Coulter Electronics Ltd). The latter comprises a glass Standardization in Haematology for measuring capillary tube 0X38 mm in internal diameter and 200 plasma viscosity was that of Harkness, and was based mm long through which a plasma sample of 0-5 ml is on a manually operated, temperature controlled, forced at a constant positive pressure of 17-2 kPa. capillary viscometer.8 An automated capillary vis- Plasma viscosity is proportional to its flow time cometer incorporating the Harkness principle is now through the capillary. Readings are made at ambient commercially available, and we have compared its temperature and corrected to a viscosity value at 25 or 37°C, whichever is preferred. The instrument may be used as a manual viscometer, the results being shown on a vacuum Accepted for publication 19 May 1988 fluorescent display. Alternatively, the optional 20 1213 J Clin Pathol: first published as 10.1136/jcp.41.11.1213 on 1 November 1988. Downloaded from 1214 Cooke, Stuart sample carousel and printer give an automated Plasma samples from 160 hospital patients were measure of viscosity. assayed in parallel using a Viscometer II and a Harkness viscometer at 25°C and showed a close Results linear association (r = 0-979, gradient = 0-969, intercept = 0-058). The correlation for 50 of these The Viscometer II required a 30 minute warm up patients, selected to cover the viscosity range 1 38 to period during which it automatically primed itselfwith 5-02 mPs-s, is shown in fig 1. Viscometer II rinse fluid (Coulter Electronics Ltd). Within batch precision was determined at four Calibration was also automatic using the manu- levels ofviscosity using plasma from four patients. The facturer's Viscometer II Calibrator. The viscosity of coefficients of variation for 10 aliquots from each high and low control reagents (Viscometer II Hi/Lo specimen of plasma were 0 5% at 1 07 Control, Coulter Electronics Ltd) was measured with mPa-s (plasma diluted with 3 1% w/v trisodium each batch oftest plasmas as a check ofcalibration and citrate), 0-6% at 1-64 mPa-s, 1-3% at 3-78 mPa-s, and to detect drift. Recalibration was usually required 1-7% at 5-66 mPa-s. Between batch precision was once or twice a week. The manufacturer's stated values determined at three measurements of viscosity using for the viscosity of the Viscometer II Calibrator and the manufacturer's low control reagent, calibrator, Hi/Lo Control reagents were checked using the and high control reagent. The coefficients of variation Harkness viscometer. The results (table 1) show close for 10 aliquots of each reagent were 0 53% at 1-20 agreement with the stated values. mPa-s, 0 55% at 1 60 mPa's, and 0 58% at 2-00 mPa-s. All viscosity measurements with the Viscometer II A specimen of normal plasma was diluted, using were corrected by the instrument to a selected 3*1% w/v trisodium citrate to give a low viscosity temperature of 25°C to permit comparison with the Harkness viscometer, the water bath of which was maintained at 25°C. To check the instrument's ability 5-0 to correct to 25°C, the Viscometer II was placed in a 'A temperature controlled room at 1 5°C, 25°C, and then 0 32°C. Viscosity measurements were then made on 10 0 / - aliquots of the manufacturer's low control reagent, 4 by copyright. calibrator, and high control reagent. The mean results E-- (table 2) show adequate correction of the viscosity 1- readings at 15°C and 32°C to within 1-7% of the measured viscosity at 25°C. ,E 30- u . 0 W Table 1 Viscosity values (mPa-s) at 25°Cfor control in vi 120- reagents as stated by manufacturer and measured by 0~ Harkness viscometer 1>20- http://jcp.bmj.com/ Low control High control reagent Calibrator reagent I I I I. I I I I Manufacturer's 10 2 0 30 40 50 stated range 1-17-1-23 1-581-62 1 96-2-04 Plasma viscosity (Harkness viscometer) (mPao s) Mean viscosity (Harkness; Fig I Scattergram with regression line showing correlation = n 10) 120 160 2-02 = 0 p < 0 between values at on September 28, 2021 by guest. Protected 95% Confidence (r 994; 002) plasma viscosity interval 1-196 to 1 204 1596 to 1 604 2-015 to 2 025 25°C measured with a Harkness viscometer and a Viscometer II (n = 50). Table 2 Mean viscosity values measured by Viscometer IIat three temperatures, and expressed as mPa-s at 25°C,for low control reagent, calibrator, and high control reagent Low control reagent Calibrator High control reagent 150C 25°C 320C 150C 25°C 32°C 150C 25°C 32°C Mean(n = 10) 1-20 1-21 1-23 1 61 1 61 1-60 2-04 201 2-00 95% Confidence interval 1-19 to 1-21 1-20 to 1-22 1-22 to 1-24 1-60 to 1-62 1-60 to 1-61 1-59 to 1-61 2-02 to 2-06 2-00 to 2-02 1 99 to 2-01 Coefficient ofvariation (%) 0-26 0 44 0-57 0-83 0-76 0 61 1 49 0 58 0-76 J Clin Pathol: first published as 10.1136/jcp.41.11.1213 on 1 November 1988. Downloaded from Automated measurement ofplasma viscosity 1215 ;VI .0 0-E " 3.0 - 3.0- 0 u - 2.0- 2.0- aE W~ 1.- 1.0- Harkness viscometer Viscometer 11 Fig 2 Ten alternate measurements at 25°C ofplasma samples ofhigh (407 mPa-s; 0) and low (I 15 mPa-s; 0) viscosity, measured with a Harkness viscometer and a Viscometer II. (mean 1[15 mPa-s, 95% confidence interval 1-14 to samples of distilled water at the end of the working 1-16, for 10 replicate measurements). A plasma ofhigh day, was sufficient to overcome this problem. The viscosity (mean 4 07 mPa-s, 95% confidence interval operator's manual has been modified accordingly. 4 05 to 4 09, for 10 measurements) was obtained from The Viscometer II is designed to measure plasma a patient with an IgM paraprotein. When the two viscosity in the range 1-5 mPs-s; viscosity results plasmas were assayed alternately, there was no between 5-7 mPa-s are displayed but with a cautionary evidence of carry over (fig 2) for either instrument. symbol.

View Full Text

Details

  • File Type
    pdf
  • Upload Time
    -
  • Content Languages
    English
  • Upload User
    Anonymous/Not logged-in
  • File Pages
    4 Page
  • File Size
    -

Download

Channel Download Status
Express Download Enable

Copyright

We respect the copyrights and intellectual property rights of all users. All uploaded documents are either original works of the uploader or authorized works of the rightful owners.

  • Not to be reproduced or distributed without explicit permission.
  • Not used for commercial purposes outside of approved use cases.
  • Not used to infringe on the rights of the original creators.
  • If you believe any content infringes your copyright, please contact us immediately.

Support

For help with questions, suggestions, or problems, please contact us